Equipment Specifications for FLAT PANEL DIGITAL (Digital Cardiac Cath-Lab Single/Mono Plane with Intra Vascular Ultra Sound)

Size: px
Start display at page:

Download "Equipment Specifications for FLAT PANEL DIGITAL (Digital Cardiac Cath-Lab Single/Mono Plane with Intra Vascular Ultra Sound)"

Transcription

1 Equipment Specifications for FLAT PANEL DIGITAL (Digital Cardiac Cath-Lab Single/Mono Plane with Intra Vascular Ultra Sound) UNSPSC Code: ECRI Code: 1. Description of Function 1.1 Cardiac Catheterization is required to do angiography of blocked vessels in the heart. Flat panel is the latest in digital X-Ray imaging technology. 2. Operational Requirements 2.1 Flat Panel Single Plane Cardiovascular Catheterization Laboratory system with Digital Subtraction Angiography With a Floor/Ceiling mounted arm stand and digital imaging X- Ray System for Cardiovascular Diagnostic and Interventional Procedures. 2.2 System should be complete with pressure injector, Homodynamic & Electrophysiological Study Recorder With Programmed Extra stimulus Stimulator and Radio-Frequency Ablation Generator and advanced integrated IVUS 3. Technical Specifications 3.1 GANTRY Floor/Ceiling Mounted Gantry rotation/angulations: LAO/RAO At least 100/100. Cranial/Caudal ± 45 degrees. Speed of Rotation: Cranial/Caudal: 10deg/sec or higher. LAO/RAO: 10 deg/sec or higher. Non-contact sensing mechanism. Real Time Patient sensing or Mechanical and Electromechanical covers with patient zone. Whole body head to toe coverage without repositioning patient. 3.2 Facility for rotational angiography 3.3 PATIENT SUPPORT TABLE Floating/Floor mounted with carbon fiber tabletop with easy patient transport capability 3.4 X-RAY GENERATOR should be of latest technology with high frequency type with at least 100 kw X-RAY TUBE Focus: Dual Anode Heat: 1.5 MHU or higher Capacity. Maximum heat dissipation capacity of at least 1500 Oiled or water-cooled to ensure continuous operation Latest generation tube, fluoroscopic power shall be at least 3000 watts for noiseless operation to support long-term interventional procedures & for better visibility in obese and deep angulations. 3.5 Image Acquisition System

2 1. Flat Detector of current generation, Good detector high spatial resolution acquisition and display in Matrix of at least 1024x Flat Panel detector system with excellent spatial and contrast resolution. 3. Frame rates selectable from 7-30 frames per seconds. 3.6 Monitors: TFT Monitors 18 inch or higher in exam room Additional monitor for patient database. CD workstation monitor Cath-Recorder Slave Monitor With high resolution matrix 1024 x 1024 Total at least 5 TFT Monitors 3.7 DIGITAL FLUOROSCOPY 8 Bit or higher. Flouro grab and replay facility 3.8 DIGITAL ACQUISITION & DIGITAL PROCESSING Acquisitions 1024 X 1024, True On Line Digital Subtraction Angiography: 1024x1024 with all post processing facility 3.9 Image processing system 1. Series exposures with frame rates of 30/15/10/7.5 fps 2. Pulsed fluoroscopy standard. 3. Applications for maximized radiation protection 4. DICOM 3 ready for patient data acquisition, documentation and archiving 3.10 COMMUNICATION AND ARCHIVING Review /Archiving & Networking. DICOM based CD recording for dynamic recording on CD. CDs to have DICOM software embedded for instant review in any PC. Ability to record DSA runs in CD. Standard line rate Video Output: Image storage capacity of at least 50,000 images at 1024x1024 matrix and Cath Lab should be supplied with state of the art QCA & LVA. Ability to generate single DVD incorporating multiple patient studies capable of review in any PC 3.11 Automated Motorized High Speed and Pressure Injection System HEMODYNAMIC AND ELECTROPHYSIOLOGIC STUDY RECORDER WITH PROGRAMMED EXTRA STIMULUS DELIVERING SYSTEM AND RADIO-FREQUENCY ABLATION GENERATOR Cardiac catheterization Electrophysiological & hemodynamic recorder with 4 invasive pressure recording. Simultaneously, 12 Lead ECG with two monitors in examination room and one monitor in Control room full remote operation. Full hemodynamic analysis and laser printer with full EP Recording and Analyzing System with Programmable Extra Stimulus Delivering Apparatus and Radio-Frequency Ablation System. Minimum 24 Intracardiac inputs 3.13 HEMOXIMETER Hemoximeter for measuring Hb and oxygen saturation during cardiac catheterization complete with all accessories like rinse solution, calibration solution etc for at least 1 year.

3 3.14 Latest generation integrated intra vascular ultrasound which is useable and interfacable with all IVUS catheters available in the market for coronary, peripheral and carotid IVUS studies with recording & DICOM facility and with at least 10 IVUS catheters provided free of cost and all controlling consoles of IVUS machine should be mounted on the Cath Table and also in the monitoring room. 4. System Configuration Accessories, spares and consumables 4.1 Main Unit C-Arm with Gantry, X-RAY generator and Tube and integrated Flat panel detector Automated Motorized High Speed Injection System Examination Light LEAD glass size- 200 X 100 cm 4.5 Digital Storage and Achieving system with DICOM CD RECORDER WITH 100 STARTER CDS Hemoximeter ACT machine and cartridges/tubes Flat Panel integrated. 4.9 Haemodynamic and Electrophysiologic Study Recorder with programmed extra stimulus delivering system and RF Ablation generator -01 set 4.10 Patient Table TFT Monitor Protective apron of at least 0.50 mm lead equivalent Protective gonad shield at least 0.35 mm lead equivalent Thyroid collars Lead Eye glasses Radiation protection device for the operator at the table and for legs below the table 4.17 Advanced two way communication system in cath lab and monitoring room

4 5 Environmental factors 5.1 The unit shall be capable of operating continuously in ambient temperature of 10 0 C C and relative humidity of 15-90% 5.2 The unit shall be capable of being stored continuously in ambient temperature of 0 50 deg C and relative humidity of 15-90% 5.3 Shall meet IEC : 2001(Or Equivalent BIS) General Requirements of Safety for Electromagnetic Compatibility The chosen supplier would be asked to undertake a turnkey Project wherein necessary civil work modifications like False Ceiling, Wall Tiling, Anti Static Flooring and finishing works would be provided by them under the supervision of the support staff e.g. CPWD (Civil). 2. The supplier would also provide the Scrub area and the Catheter wash area. 3. The supplier also would provide the necessary furniture like tables, computer chairs, cupboards, catheter hang wall mounts etc Appropriate Air-conditioning would be provided by the supplier and maintained throughout the Warranty period of the cath-lab. 2. The entire Cath-Lab including the Air Conditioning should be connected to the Generator of the hospital. 5.6 Proper shielding should have to be done by the supplier to minimize radiation leakage as per AERB and BARC regulations. 6 Power Supply 6.1 Power input to be VAC(Single Phase), / V (3 Phase)/ 50Hz as appropriate fitted with Indian plug 6.2 Reset table over current breaker shall be fitted for protection 6.3 Suitable Servo controlled Stabilizer/CVT for the entire cath lab. 6.4 Online UPS of suitable rating conforming to shall be supplied for the entire cath lab system including X-ray generation with a minimum power back up of 30 minutes. 6.5 The Power requirements involve laying a 125 KVA Cable from the substation to the Cath-Lab and making a Bus-Bar and a Power Distribution Board and this would be done by the supplier as a turn key project under the supervision of the support staffs e.g. PWD (Elect) 7 Standards, Safety and Training 7.1 Should be FDA or CE approved product 7.2 Electrical safety conforms to standards for electrical safety IEC General Requirements

5 7.3 Manufacturer should have ISO certification for quality standards. 7.4 Shall comply with AERB and BARC guidelines. 8 Documentation 8.1 User manual in English 8.2 Service manual in English 8.3 List of important spare parts and accessories with their part number and costing 8.4 Certificate of Calibration and inspection from the factory 8.5 Log book with instructions for daily, weekly, monthly and quarterly maintenance checklist. The job description of the hospital technician and company service engineer should be clearly spelt out. 8.6 List of Equipments available for providing calibration and routine maintenance support as per manufacturer documentation in service / technical manual. 8.7 User List and performance certificate of at least 5 cath labs installation in the past five years from government institutions should be submitted along with the techno commercial bid.

Technical Specifications for Flat Panel Single Plane Cardiac Cath-Lab along with Accessories

Technical Specifications for Flat Panel Single Plane Cardiac Cath-Lab along with Accessories Technical Specifications for Flat Panel Single Plane Cardiac Cath-Lab along with Accessories Sr. No Specifications as per tender Latest state of the art, single plane ceiling mounted C-arm Latest state

More information

KS-70. Catheterization Table GENERAL

KS-70. Catheterization Table GENERAL PD505-009A Catheterization Table KS-70 GENERAL This Catheterization Table KS-70, which is combined with X-ray high voltage generator, X-ray tube assembly, X-ray image system (Image Intensifier, Flat Panel

More information

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950 HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple

More information

MEDICAL IMAGING. 90/30 Remote Controlled RF System

MEDICAL IMAGING. 90/30 Remote Controlled RF System MEDICAL IMAGING 90/30 Remote Controlled RF System The Easy solution for your Remote Controlled room Apollo EZ represents the Easy approach for the radiographic/fluoroscopic room and it comes from the experience

More information

Tender Notice No: NEIGR/S&P/03/04/2014-2015 Dated: 13/04/2015 TENDER CORRIGENDUM /ADDENDUM

Tender Notice No: NEIGR/S&P/03/04/2014-2015 Dated: 13/04/2015 TENDER CORRIGENDUM /ADDENDUM North Eastern Indira Gandhi Regional Institute of Health and Medical Sciences (An Autonomous Institute, Ministry of Health and Family Welfare, Government of India) Director s Block, Mawdiangdiang, Shillong

More information

Melting Point. Electrothermal PRODUCT SPECIFICATIONS OPERATION

Melting Point. Electrothermal PRODUCT SPECIFICATIONS OPERATION M A N U A L M E L - T E M P THE LOWEST COST MELTING POINT APPARATUS FOUND ANYWHERE! Low cost unit Temperature range to 500 C Aluminum casting conducts and radiates heat uniformly to capillaries and thermometer.

More information

AXIOM Artis U The little big one for universal angiography and cardiology applications

AXIOM Artis U The little big one for universal angiography and cardiology applications AXIOM Artis U The little big one for universal angiography and cardiology applications AXIOM The versatile system for universal angiography and cardiology applications Unchain your workflow! Ease of Use

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

Ch. 138 CARDIAC CATHETERIZATION SERVICES 28 138.1 CHAPTER 138. CARDIAC CATHETERIZATION SERVICES GENERAL PROVISIONS

Ch. 138 CARDIAC CATHETERIZATION SERVICES 28 138.1 CHAPTER 138. CARDIAC CATHETERIZATION SERVICES GENERAL PROVISIONS Ch. 138 CARDIAC CATHETERIZATION SERVICES 28 138.1 CHAPTER 138. CARDIAC CATHETERIZATION SERVICES Sec. 138.1 Principle. 138.2. Definitions. GENERAL PROVISIONS PROGRAM, SERVICE, PERSONNEL AND AGREEMENT REQUIREMENTS

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

AIS 3000 APC AIS 3000 15kVA 208V w/2 Batt. Module Exp. To 4, Start-Up 5X8, Internal Maintenance Bypass

AIS 3000 APC AIS 3000 15kVA 208V w/2 Batt. Module Exp. To 4, Start-Up 5X8, Internal Maintenance Bypass AIS 3000 APC AIS 3000 15kVA 208V w/2 Batt. Module Exp. To 4, Start-Up 5X8, Internal Maintenance Bypass APC AIS 3000, 12 kw / 15 kva,input 208V 3PH / Output 120V, 208V, 208V 3PH, Interface Port DB-9 RS-232,

More information

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl Cirrus 0.2T MRI for Everyone North America, Asia, Europe contact: kturek@mri-tech.pl MRI-TECH inc. Cirrus MRI system for all your needs: Low costs Low maintenance High quality Open geometry Imaging of

More information

TrueAlarm Fire Alarm Systems

TrueAlarm Fire Alarm Systems TrueAlarm Fire Alarm Systems UL, ULC, CSFM Listed; FM Approved; MEA (NYC) Acceptance* Fire Alarm System Accessories, 4190 Series PC Annunciator with Multi-Client Capability Features Fire alarm control

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION

More information

Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service.

Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service. Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service. Royal College of Radiologists British Society of Interventional Radiology The Vascular Society of Great

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)

More information

SECTION 13850 DETECTION AND ALARM

SECTION 13850 DETECTION AND ALARM SECTION 13850 DETECTION AND ALARM PART 1 GENERAL 1.01 SUMMARY A. Section Includes 1. Control Panel 2 Associated Equipment B. Products Installed But Not Supplied Under This Section 1. Section 16140 - Wiring

More information

Checklist for submission of application form for Registration

Checklist for submission of application form for Registration Checklist for submission of application form for Registration *Incomplete submission may cause delay in processing of the application. Name of the Hospital/Institution: City State Sr.No. Checkpoints Status

More information

Tube Control Measurement, Sorting Modular System for Glass Tube

Tube Control Measurement, Sorting Modular System for Glass Tube Tube Control Measurement, Sorting Modular System for Glass Tube Tube Control is a modular designed system of settled instruments and modules. It comprises measuring instruments for the tube dimensions,

More information

Datex-Ohmeda Critical Care Monitor

Datex-Ohmeda Critical Care Monitor GE Healthcare Datex-Ohmeda Critical Care Monitor Comprehensive monitoring for the challenging Critical Care environment Critical Care monitor offers a flexible range of module frame options, displays and

More information

Product Information. QUADRA-CHEK 3000 Evaluation Electronics For Metrological Applications

Product Information. QUADRA-CHEK 3000 Evaluation Electronics For Metrological Applications Product Information QUADRA-CHEK 3000 Evaluation Electronics For Metrological Applications April 2016 QUADRA-CHEK 3000 The evaluation electronics for intuitive 2-D measurement The QUADRA-CHEK 3000 evaluation

More information

Digital Dental X-ray. Midmark Animal Health Products

Digital Dental X-ray. Midmark Animal Health Products Digital Dental X-ray Midmark Animal Health Products Disease you can't see... 60% of periodontal disease lies below the gum line, and many issues aren't detectable via probing or visual inspection.* Don't

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

C-Arm 101. Everything you ever need to know! Guide to C-Arms

C-Arm 101. Everything you ever need to know! Guide to C-Arms REVISED 2015 C-Arm 101 Everything you ever need to know! Guide to C-Arms C-Arms 101 Welcome to the Atlantis Worldwide Buyers Guide. This guide is designed to help you learn all you need to know about C-Arms.

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

TECHNICAL SPECIFICATION. StirLIN-1 Economy Liquid Nitrogen System

TECHNICAL SPECIFICATION. StirLIN-1 Economy Liquid Nitrogen System TECHNICAL SPECIFICATION StirLIN-1 Economy Liquid Nitrogen System Reference 80 8120_01 Issue Date August 1, 2015 1. INTRODUCTION Since more than sixty years Stirling Cryogenics has designed and manufactured

More information

Compaq Rack Options and Accessories. Monitors and Keyboards

Compaq Rack Options and Accessories. Monitors and Keyboards 1 Compaq Rack Options and Accessories Compaq offers a wide variety of rack options and accessories that help you to complete your 9000 and 10000 series racks. Monitors and Keyboards TFT5600RKM United States

More information

DIVISION 28 ELECTRONIC SAFETY AND SECURITY SECTION 28 13 00 SECURITY AND ACCESS CONTROL SYSTEM

DIVISION 28 ELECTRONIC SAFETY AND SECURITY SECTION 28 13 00 SECURITY AND ACCESS CONTROL SYSTEM DIVISION 28 ELECTRONIC SAFETY AND SECURITY PART 1 GENERAL 1.01 DESCRIPTION A. Division 28 Contractor shall extend a complete Matrix Systems Access Control System as shown on the Drawings and as specified

More information

A1.5 Classroom Standards

A1.5 Classroom Standards A1.5 Classroom Standards Note: The following standards have been developed and adopted by KU's Instructional and Mediated Spaces Committee (IMSAC), which includes representatives from DCM, Instructional

More information

DR Retrofit Kit 1417 WL Configuration Integrated PACS

DR Retrofit Kit 1417 WL Configuration Integrated PACS DR Retrofit Kit 1417 WL Configuration Integrated PACS Description: The DR Retrofit Kit 1417WL - Configuration Integrated PACS makes the use of CR Cassettes/Imaging Plates or the use of analogue x-ray films,

More information

Bringing together all of the cardiovascular information you need into a single platform

Bringing together all of the cardiovascular information you need into a single platform Care Providers Hospitals Enterprise Image and Information Bringing together all of the cardiovascular information you need into a single platform Organizations want to improve quality of care, boost efficiencies

More information

Understanding Digital Modalities: System Integration and Use

Understanding Digital Modalities: System Integration and Use Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information

More information

HP EliteDisplay LED Backlit Monitors. User Guide

HP EliteDisplay LED Backlit Monitors. User Guide HP EliteDisplay LED Backlit Monitors User Guide 2012, Hewlett-Packard Development Company, L.P. Microsoft and Windows are U.S. registered trademarks of Microsoft Corporation. The only warranties for HP

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

Connect to a remote PC via Ethernet with a monitor, keyboard, and mouse. This network monitor offers both long distance and wireless transmission.

Connect to a remote PC via Ethernet with a monitor, keyboard, and mouse. This network monitor offers both long distance and wireless transmission. 17" (43 cm) LCD Monitor Connect to a remote PC via Ethernet with a monitor, keyboard, and mouse. This network monitor offers both long distance and wireless transmission. Long Distance Transmission Easily

More information

Digital input modules

Digital input modules 8 172 TX-I/O Digital input modules TXM1.8D TXM1.16D Two fully compatible versions: TXM1.8D: 8 inputs, each with a three-color LED (green, yellow or red) TXM1.16D: As TXM1.8X, but 16 inputs, each with a

More information

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen

More information

GE Healthcare. OEC 9900 Elite. Premium Digital Mobile Imaging System.

GE Healthcare. OEC 9900 Elite. Premium Digital Mobile Imaging System. GE Healthcare OEC 9900 Elite Premium Digital Mobile Imaging System. Discover the clear difference. For over three decades, GE OEC has been a market leader in mobile C-arm surgical imaging. Now, the OEC

More information

ST. DAVID S MEDICAL CENTER CARDIOLOGY - Special, Invasive, Diagnostic, or High-Risk Procedure Requirements

ST. DAVID S MEDICAL CENTER CARDIOLOGY - Special, Invasive, Diagnostic, or High-Risk Procedure Requirements ST. DAVID S MEDICAL CENTER CARDIOLOGY - Special, Invasive, Diagnostic, or High-Risk Procedure Requirements Cardiac Catheterization & Peripheral Angiography Completion of a fellowship in Cardiovascular

More information

Specification guide no. 9 Active harmonic conditioner from 20 to 480 A

Specification guide no. 9 Active harmonic conditioner from 20 to 480 A Active harmonic conditioner from 20 to 480 A Contents 1 - Installations concerned... 2 1.1 - New installations... 2 1.2 - Existing installations... 2 2 - General... 2 2.1 - Scope... 2 2.2 - Applicable

More information

Multi-Range Programmable DC Power Supplies 9115 Series

Multi-Range Programmable DC Power Supplies 9115 Series Data Sheet Multi-Range Programmable DC Power Supplies 1200 W / 3000 W Multi-Range DC Power Supplies Features & Benefits Any model can replace several supplies on your bench or in your rack. Unlike conventional

More information

RDJ10RF/SET. Wireless room temperature controller with 24-hour time switch and LCD. Programmable, for heating systems

RDJ10RF/SET. Wireless room temperature controller with 24-hour time switch and LCD. Programmable, for heating systems 3 072 RDJ10RF RCR10/433 Wireless room temperature controller with 24-hour time switch and LCD Programmable, for heating systems RDJ10RF/SET Operating modes: Automatic, Comfort, Energy Saving, and Frost

More information

BSE Geluid en Licht Pagina 1 van 5. Description

BSE Geluid en Licht Pagina 1 van 5. Description Description The MiniMAC Profile is an automated moving-head spotlight that provides 12 dichroic color filters, 7 interchangeable rotating gobos, high-speed mechanical shutter, 540 of pan by 270 of tilt,

More information

DVI DA2 and DVI DA 4 User Guide

DVI DA2 and DVI DA 4 User Guide DVI DA2 and DVI DA 4 User Guide This guide describes the installation and operation of the Extron DVI DA2 and DVI DA4 Distribution Amplifiers. Unless stated otherwise, distribution amplifier or the unit

More information

Staff Doses & Practical Radiation Protection in DEXA

Staff Doses & Practical Radiation Protection in DEXA Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /

More information

T7560A,B,C Digital Wall Module

T7560A,B,C Digital Wall Module T7560A,B,C Digital Wall Module HONEYWELL EXCEL 5000 OPEN SYSTEM BEFORE INSTALLATION All wiring must comply with local electrical codes and ordinances or as specified on installation wiring diagrams. Digital

More information

BASIC STANDARDS FOR RESIDENCY TRAINING IN CARDIOLOGY

BASIC STANDARDS FOR RESIDENCY TRAINING IN CARDIOLOGY BASIC STANDARDS FOR RESIDENCY TRAINING IN CARDIOLOGY American Osteopathic Association and the American College of Osteopathic Internists Specific Requirements For Osteopathic Subspecialty Training In Cardiology

More information

Embedded Systems in Healthcare. Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008

Embedded Systems in Healthcare. Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008 Embedded Systems in Healthcare Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008 About the Speaker Working for Philips Research since 1982 Projects

More information

Siemens AG 2011 SINAMICS V60. The perfect solution for basic servo applications. Brochure May 2011 SINAMICS. Answers for industry.

Siemens AG 2011 SINAMICS V60. The perfect solution for basic servo applications. Brochure May 2011 SINAMICS. Answers for industry. The perfect solution for basic servo applications Brochure May 2011 SINAMICS Answers for industry. with 1FL5 servomotors The solution for basic servo applications There is a requirement to automate motion

More information

A PATIENT S GUIDE TO CARDIAC CATHETERIZATION

A PATIENT S GUIDE TO CARDIAC CATHETERIZATION A PATIENT S GUIDE TO CARDIAC CATHETERIZATION The science of medicine. The compassion to heal. This teaching booklet is designed to introduce you to cardiac catheterization. In the following pages, we will

More information

Environmental Product Declaration www.siemens.com/healthcare

Environmental Product Declaration www.siemens.com/healthcare AXIOM Sensis XP Environmental Product Declaration www.siemens.com/healthcare s Progress that is Impressive Ecological Advantages of AXIOM Sensis XP Sensis allows for paperless docu mentation and reporting

More information

ACL 395 Resistivity Meter

ACL 395 Resistivity Meter ACL 395 Resistivity Meter OPERATION MANUAL Meter is warranted for one year from the date of purchase on parts and labor. Calibration is recommended every 12 months. 840 W. 49 th Place Page 1 of 13 info@aclstaticide.com

More information

T0118 T2118 T3118. Instruction Manual

T0118 T2118 T3118. Instruction Manual Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature, relative humidity and other derived humidity

More information

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Excelart Vantage Magnet Configuration: Ultra-short-bore Strength (or W x H): 1.5 T Homogeneity, ppm V-RMS: Dimensions of maximum

More information

Duct Humidity Transmitter

Duct Humidity Transmitter SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers

More information

ACL 800 MEGOHMMETER OPERATION MANUAL

ACL 800 MEGOHMMETER OPERATION MANUAL ACL 800 MEGOHMMETER OPERATION MANUAL Meter is warranted for one year from the date of purchase on parts and labor. Calibration is recommended every twelve months. NOTE: THIS MANUAL IS NOT TO BE SEPARATED

More information

Pocket Tach 99 (PT99) Non-Contact Tachometer

Pocket Tach 99 (PT99) Non-Contact Tachometer CE DECLARATION OF CONFORMITY As Manufacturer: Monarch Instrument Division of Monarch International Inc. 15 Columbia Drive, Amherst NH 03031 USA declares under Monarch s sole responsibility that the product:

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

4x1 USB HDMI KVM Switch Quick Installation Guide

4x1 USB HDMI KVM Switch Quick Installation Guide Introduction 4x1 USB HDMI KVM Switch Quick Installation Guide The 4x1 USB HDMI KVM Switch shares one HDMI display between four HDMI sources. Key Features and Benefits Share an HDTV or display, USB keyboard

More information

SECTION 16853 ZONED (DC LOOP) FIREALARM SYSTEM

SECTION 16853 ZONED (DC LOOP) FIREALARM SYSTEM SECTION 16853 ZONED (DC LOOP) FIRE-ALARM SYSTEM PART 1 - GENERAL 1.1 RELATED DOCUMENTS A. Drawings and general provisions of the Contract, including General and Supplementary Conditions and Division 1

More information

TRACTION NETWORK MONITORING AND PROTECTION SYSTEM SMTN-3 CITY ELECTRIC TRANSPORT RAILWAYS METRO INDUSTRY

TRACTION NETWORK MONITORING AND PROTECTION SYSTEM SMTN-3 CITY ELECTRIC TRANSPORT RAILWAYS METRO INDUSTRY TRACTION NETWORK MONITORING AND PROTECTION SYSTEM SMTN-3 CITY ELECTRIC TRANSPORT RAILWAYS METRO INDUSTRY 2 TRACTION NETWORK MONITORING AND PROTECTION SYSTEM Traction network monitoring and protection system,

More information

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581 X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers

More information

Drayton Digistat +2RF/+3RF

Drayton Digistat +2RF/+3RF /+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer

More information

SUTTER MEDICAL CENTER, SACRAMENTO Department of Cardiovascular Disease Cardiology - Delineation of Privileges

SUTTER MEDICAL CENTER, SACRAMENTO Department of Cardiovascular Disease Cardiology - Delineation of Privileges INITIAL: [ ] RENEWED: [ ] DATE: ADDITIONAL: [ ] Privileges are granted for Sutter General Hospital, Sutter Memorial Hospital, Sutter Center for Psychiatry, Sutter Oaks Midtown and the Capitol Pavilion

More information

8-Port PoE Network Switch

8-Port PoE Network Switch Page 1 of 12 8-Port PoE Network Switch User s Guide Document Number 91I00-1 Rev A August, 2009 FCC Warning: Page 2 of 12 This devices has been tested and found to comply with the regulations for Class

More information

Cardius. XPO Series. Cardius 3 XPO

Cardius. XPO Series. Cardius 3 XPO Cardius XPO Series Cardius 3 XPO Three imaging systems, all of which provide outstanding image quality, superior efficiency, and increased patient comfort. Cardius 2 XPO Includes the standard features

More information

Videoconferencing Room Guide

Videoconferencing Room Guide Videoconferencing Room Guide IVN RoomGuide.doc Page 1 Introduction This guide will serve as a tool to plan the design and implementation of a videoconference room. The information in this guide should

More information

Now bring the advantages of digital workflow to cardiology

Now bring the advantages of digital workflow to cardiology CARESTREAM Cardiology PACS Now bring the advantages of digital workflow to cardiology The single system advantage Now there s ONE Solution. 2 A large and growing number of health care facilities are benefiting

More information

NC State University Design and Construction Guidelines Division 26 Fire Alarm Systems

NC State University Design and Construction Guidelines Division 26 Fire Alarm Systems NC State University Design and Construction Guidelines Division 26 Fire Alarm Systems 1.0 Purpose A. The following guideline provides the minimum standards and requirements for fire alarm systems. 2.0

More information

Table of Contents. TransformerIQ FAQs 1

Table of Contents. TransformerIQ FAQs 1 [12/12] FAQs Table of Contents FAQ 1. What size transformer will TransformerIQ monitor?... 2 FAQ 2. Are there different TransformerIQ versions?... 2 FAQ 3. How are the units powered?... 2 FAQ 4. What is

More information

K ODAK D IRECTV IEW CR 850 SYSTEM

K ODAK D IRECTV IEW CR 850 SYSTEM CR 850 SYSTEM A fast, premium single-cassette CR system HEALTH IMAGING A BETTER V IEW OF L IFE. CR 850 SYSTEM A FAST, PREMIUM SINGLE-CASSETTE CR SYSTEM FOR PRODUCTIVE X-RAY IMAGING Boost productivity with

More information

Bloomberg 17 Inch Flat Panel Display. 29 April 2005 Version: 1.1

Bloomberg 17 Inch Flat Panel Display. 29 April 2005 Version: 1.1 Bloomberg 17 Inch Flat Panel Display 29 April 2005 Version: 1.1 1 Flat Panel Technical Information Description The BLOOMBERG 17-inch Flat Panel Display, which is comprised of two high-quality independent

More information

FIRE ALARM AND DETECTION SYSTEMS SECTION 16721

FIRE ALARM AND DETECTION SYSTEMS SECTION 16721 PART 1 - GENERAL 1.01 WORK INCLUDED FIRE ALARM AND DETECTION SYSTEMS SECTION 16721 A. Provide a complete fully addressable, power limited, fire detection and evacuation system. The system shall be connected

More information

A. All equipment and materials used shall be standard components that are regularly manufactured and used in the manufacturer s system.

A. All equipment and materials used shall be standard components that are regularly manufactured and used in the manufacturer s system. ENDURA - WS5050 PC WORKSTATION TECHNICAL SPECIFICATIONS SECURITY SYSTEM DIVISION 28 ELECTRONIC SAFETY AND SECURITY LEVEL 1 28 20 00 ELECTRONIC SURVEILLANCE LEVEL 2 28 23 00 VIDEO SURVEILLANCE LEVEL 3 28

More information

The Hybrid Cath Lab: A Revolution in Cardiac Care

The Hybrid Cath Lab: A Revolution in Cardiac Care Surgery Artis zee in the OR In the interest of improved patient outcomes, increased workflow efficiency, and implementation of cutting-edge technology, The Prince Charles Hospital in Brisbane, Australia

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

LNG Monitoring. Fiber-Optic Leakage Detection System. Pipeline leakage detection. Regasification and liquefaction monitoring

LNG Monitoring. Fiber-Optic Leakage Detection System. Pipeline leakage detection. Regasification and liquefaction monitoring LNG Monitoring Fiber-Optic Leakage Detection System Pipeline leakage detection Regasification and liquefaction monitoring Tank annulus and base slab monitoring Spill containment control Intelligent Solutions

More information

GE Healthcare. Vivid 7 Dimension Cardiovascular Ultrasound System

GE Healthcare. Vivid 7 Dimension Cardiovascular Ultrasound System GE Healthcare Vivid 7 Dimension Cardiovascular Ultrasound System New dime Introducing multi-dimensional and 4D imaging. The ability to incorporate multi-dimensional and 4D imaging into your clinical routine

More information

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination

More information

Sun Group Android Wi-Fi Digital Signage Touch Screen Series. SUN-SG8xxxxAS(W) SUN Group (Taiwan) -Android Touch Screen LCD Signage

Sun Group Android Wi-Fi Digital Signage Touch Screen Series. SUN-SG8xxxxAS(W) SUN Group (Taiwan) -Android Touch Screen LCD Signage Sun Group Android Wi-Fi Digital Signage Touch Screen Series Model No. SUN-SG8xxxxAS(W) SUN Group (Taiwan) -Android Touch Screen LCD Signage SUN Group Series: Size Model 15.6" SUN-SG8156ASW Android Wi-Fi

More information

Problem / Solution. The problem:

Problem / Solution. The problem: Problem / Solution Solar Photovoltaic (PV) technology has shown significant development in recent years and this has led to many UK homeowners investing in Solar PV installations. Solar electricity is

More information

FVSU-33201301 REPAIR & UPGRADE FIBER OPTIC 280000-1 SECTION 280000 FIRE ALARM SYSTEM CONNECTIVITY

FVSU-33201301 REPAIR & UPGRADE FIBER OPTIC 280000-1 SECTION 280000 FIRE ALARM SYSTEM CONNECTIVITY FVSU-33201301 REPAIR & UPGRADE FIBER OPTIC 280000-1 SECTION 280000 FIRE ALARM SYSTEM CONNECTIVITY PART 1 - GENERAL 1.1 SUMMARY A. Section Includes: 1. Fire-alarm control unit. 2. Addressable interface

More information

How To Use A U.S. Cell Phone At Home

How To Use A U.S. Cell Phone At Home U.S. Cellular Home Phone 1 Contents Getting to Know Your Device...5 Appearance... 5 LED Indicators... 6 Device Installation...7 Before You Begin... 7 Installing the Battery... 7 Installing the Power Adapter...

More information

R22. K Control. Indoor Unit. Nomenclature. Compatibility PL H 3 G K H B. Unit style Heat Pump Horse Power

R22. K Control. Indoor Unit. Nomenclature. Compatibility PL H 3 G K H B. Unit style Heat Pump Horse Power R22. K Control. Indoor Unit. Nomenclature. PL H 3 G K H B Compatibility Unit style Heat Pump Horse Power Control Boost Heaters R22. K Control. Outdoor Unit. Nomenclature. PU H 3 Y K A Compatibility Outdoor

More information

Industrial Flat Panel

Industrial Flat Panel Product Data Sheet The s are rugged industry-standard LCD monitors in a wide screen design. Installation: - Equally suited for installation in the machine, in control cabinets, in consoles, to gantries,

More information

Laboratory Construction Guidelines Vicon MX Systems

Laboratory Construction Guidelines Vicon MX Systems Laboratory Construction Guidelines Vicon MX Systems These Vicon MX Laboratory Construction Guidelines illustrate how to prepare a life sciences laboratory for installation of your new Vicon MX motion measurement

More information

HEAT MONITORING AND CONTROL INSTRUMENTS FOR TRANSFORMERS MB 103 TEMPERATURE CONTROL UNIT PT 100 TEMPERATURE SENSOR

HEAT MONITORING AND CONTROL INSTRUMENTS FOR TRANSFORMERS MB 103 TEMPERATURE CONTROL UNIT PT 100 TEMPERATURE SENSOR HEAT MONITORING AND CONTROL INSTRUMENTS FOR TRANSFORMERS MB 103 TEMPERATURE CONTROL UNIT PT 100 TEMPERATURE SENSOR 2 MB 103 TRANSFORMER TEMPERATURE CONTROL UNIT WITH CONTROL SIGNAL Recent approval of new

More information

Planning Information. Passenger and Freight Elevators.

Planning Information. Passenger and Freight Elevators. A company of ThyssenKrupp Elevator Thyssen Aufzüge tk This planning information will give you an overview of the most important planning criteria of passenger and freight elevators. However you should

More information

AC3200 INTELLIGENT BROADBAND AMPLIFIER

AC3200 INTELLIGENT BROADBAND AMPLIFIER Kari Mäki 9.4.2013 1(7) AC3200 INTELLIGENT BROADBAND AMPLIFIER AC3200 is the latest leading-edge addition to AC family with extended frequency and gain ranges and integrated electrical controls in both

More information

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003)

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003) (Part # S-LIB-M003) The smart sensor is designed to work with the HOBO Weather Station logger. The smart sensor has a plug-in modular connector that allows it to be added easily to a HOBO Weather Station.

More information

T a b l e o f C o n t e n t s

T a b l e o f C o n t e n t s w w w. g e n t e c h e a l t h c a r e. c o m M E D I C A L M O N I T O R I N G, A L A R M A N D C O N T R O L S Y S T E M S I n t e r n a t i o n a l E d i t i o n Technology for a Better Future Company

More information

GeekDesk Max Instructions for Assembly & Use

GeekDesk Max Instructions for Assembly & Use GeekDesk Max Instructions for Assembly & Use Work Sitting OR Standing! With the touch of a button... Elevate Your Work Instructions release 1.5, 2012-08-09 WELCOME! Thank you for purchasing a GeekDesk!

More information

Generic - Hearing Loop - (AFILS) U.S. System Specification

Generic - Hearing Loop - (AFILS) U.S. System Specification This document is a generic specification for any Hearing Loop (Audio Frequency Induction Loop System). For the remainder of the document, we will refer to using the term Hearing Loop rather than Audio

More information

22" Widescreen LED Monitor VA2261-2

22 Widescreen LED Monitor VA2261-2 22" Widescreen LED Monitor VA2261-2 The ViewSonic VA2261-2 is an eye-care and environmentally friendly 22" (21.5" viewable) widescreen monitor that features Flicker-Free and Blue Light technology, eliminating

More information

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric

More information

Spetec Laminar Flow Systems. Operating Instructions

Spetec Laminar Flow Systems. Operating Instructions Spetec Laminar Flow Systems Operating Instructions 1 Spetec Laminar Flow Systems Operating Instructions Thank you for your trust in the Spetec clean room systems! Your chosen system is ideally suited for

More information