1 BSI: An In Vitro Diagnostics Notified Body A guide to the In Vitro Diagnostic Directive...making excellence a habit.

2 A BSI guide to the In Vitro Diagnostic Directive Introduction In Vitro Diagnostics (IVD) is an essential and fast growing part of the global healthcare system, as they add value to patients, medical professionals and the industry along with enhancing the well-being of the population as a whole. The purpose of the BSI IVD Guide is to provide useful information to In Vitro Device Manufacturers and other interested parties seeking to place products on to the European Market. This requires a clear understanding of the In Vitro Diagnostic Directive (IVDD) 98/79/EC which sets the regulatory requirements for obtaining CE marking. Information is also presented on how to determine if you require a Notified Body to assess conformity prior to affixing a CE mark. If you think you may need a Notified Body, whether your device ranges from simple to complex, BSI s in-house team of experts are here to assist you. We promise to deliver professional, fast and responsive services with the highest quality and most efficient reviews possible. The Aim of the In Vitro Diagnostic Directive (98/79/EC) The In Vitro Diagnostic Directive (IVDD) 98/79/EC was introduced in the later part of 1998 and compliance became mandatory on December 7, The Directive provides regulatory requirements that facilitate the free trade within the European Economic Area (EEA), which comprises the 27 European Union (EU) member states and Iceland, Liechtenstein and Norway as members of the European Free Trade Association (EFTA). The IVDD specifically addresses the safety, quality and performance of In Vitro Diagnostic medical devices (IVDs). The aim of the Directive is to ensure that IVDs do not compromise the health and safety of patients, users and third parties and attain the performance levels specified by the manufacturer. The manufacturer is responsible for ensuring their products comply with the Essential Requirements of the Directive before affixing the CE marking and legally gaining access and free movement within the EEA. However, for some devices the use of a Notified Body may be required for conformity assessments before placing the device on the European market. What is an In Vitro Medical Device? IVDs are medical devices and accessories used to perform tests on samples, such as blood, urine, tissue, taken away from the human body to help detect infection, diagnose a medical condition, prevent disease or monitor drug therapies. The first step is to determine if the product is a medical device as defined by the Directive. The In Vitro Device Directive Article 1, point 2b defines an IVD as any medical device which is a Reagent, Reagent Product Calibrator, Control Material Kit Instrument, Apparatus Equipment, System Whether used alone or in combination, intended by the manufacturer to be used in vitro for the examination of specimen, including blood and tissue donations, derived from the human body, solely or principally for the purpose of providing information: concerning a physiological of pathological state of health or disease, or concerning a congenital abnormality, or to determine the safety and compatibility with potential recipients, or to monitor therapeutic measures. Examples include: HIV test kits Immunoassay analysers Blood gas analysers The calibrators and control materials used to verify the performance of the analysers Specimen receptacles and blood collection tubes Blood glucose meters and strips 2 A Guide to the In Vitro Diagnostic Directive

3 Who Does the Directive Apply To? The Directive applies to all IVDs sold in the EEA regardless of where they are designed and manufactured. If the manufacturer is based outside the EEA you may still be responsible for ensuring your device meets the requirements of the Directive. If you are a distributor and the device is not already CE marked or you change the intended purpose of a device, for example from point of care to self test or you add your own instructions for use or labelling you could become the manufacturer and therefore responsible for meeting the requirements of the Directive. If you are not based in the European Union, you will be required to have an Authorised Representative based in the EU. The Directive lists Essential Requirements to which all IVDs must comply before being placed on the market. These requirements address the design, production, labelling and instructions for use. Not all the Essential Requirements will apply to all devices; the manufacturer determines which are appropriate for their device according to the manufacturer s intended purpose. Manufacturers commonly demonstrate compliance using an Essential Requirement checklist, which considers each Essential Requirement and determines whether it is applicable. One way in which manufacturers can demonstrate that they have met the Essential Requirements is to comply with the relevant standards. Harmonised standards, such as ISO for Quality Management Systems or ISO for Risk Management, have been written especially to help demonstrate compliance with the Directive. If you meet the requirements of these standards in full there is a presumption that you conform to the requirements of the Directive. The Essential Requirements address the risk of the device including: Analytical & diagnostic sensitivity, Analytical & diagnostic specificity, Accuracy, Repeatability, Reproducibility. Classification and the Conformity Assessment Routes There are a number of ways you can demonstrate conformity with the Directive, these involve a choice of testing and quality assurance modules. The choices available depend on the classification of the device. Appendix I describes the different routes available. The Directive groups IVDs into four categories according to the perceived risk associated with the relative danger to public health and/or patient treatment by an IVD failing to perform as intended. The diagram below shows the classification and indicates where a Notified Body is required. Annex II of the Directive contains an inclusive list of products which require certification by a third party called a Notified Body. Only the devices specifically listed in Annex II require a Notified Body, for example PSA is the only cancer marker in List B. Is a Notified Body Required? Annex II List A Notified Body Required Design Dossier Review (Including Compliance to the CTS) Audit of Quality Management System Batch Released by the Notified Body HIV, Hepatitis ABO Blood Grouping Annex II List B Notified Body Required Audit of Technical Documentation & Quality Management System Rubella, PSA, Self Test for Blood Glucose Self Test Notified Body Required Review of Design & Labeling for Lay User Suitability Pregnancy, Cholesterol Home Tests General IVDD Classification No Notified Body Required Manufacturer Self Declares Tests for Hormones, Cardiac Markers, Hematology and Clinical Chemistry Tests Examples A Guide to the In Vitro Diagnostic Directive 3

4 General IVDs General devices include all IVDs other than those covered by Annex II and IVDs for self-testing. For a general IVD the manufacturer self-declares conformity with the relevant Essential Requirements of the Directive, and ensures the device fulfils the applicable obligations described in Annex III. These devices do not require the involvement of a Notified Body. IVDs for Self-Testing The particular requirements for self-testing set out in the Directive exclude self-test devices covered in Annex II. The manufacturer prepares a declaration of conformity in a similar way to the general devices but a Notified Body review is also required. The Notified Body will ensure that the device design (section 6 of Annex III) and the information provided for its use is suitable for non-professional users. This is essential as the lay user may make important decisions based on the results obtained, without consulting a healthcare professional. Alternatively the manufacturer may follow the conformity assessment routes for higher risk products as follows. IVDs in Annex II The annex is sub-divided into two lists, List A and List B. All Annex II IVDs require the involvement of a Notified Body before the product can be placed on the market. IVDs in Annex II List B Moderate Risk For devices under Annex II List B, the manufacturer must follow the applicable obligations imposed by either Annex IV, or by Annex V and VI, or alternatively by Annexes V and VII and must declare and ensure that the device meets the requirements described in these annexes. List B includes - Reagents and reagent products, including related calibration and control material for: Blood groups Anti-Duffy and Anti-Kidd; irregular and anti-erythrocytic antibodies; Rubella and toxoplasmosis; Phenylketonuria; CMV, Chlamydia, HLA tissue groups DR, A and B; PSA; Self-test blood glucose measuring; Devices & software designed specifically for evaluating the risk of trisomy 21. IVDs in Annex II List A High Risk For devices under Annex II List A, manufacturers must follow either Annex IV, or alternatively Annexes V and VII (they cannot follow Annexes V and VI). In addition, the Notified Body must verify the product meets the Common Technical Specification (CTS) and must release each batch of product before it can be placed on the European market. The batch release often requires testing. Manufacturers are expected to comply with the CTS, which are drawn up by an expert group convened by the Commission. The CTS defines performance evaluation criteria and batch release criteria. List A Reagent and reagent products, including calibrators and control materials for: Determining blood groups ABO Rhesus (C, c, D, E, e) Anti Kell The detection, confirmation and quantification of HIV 1 and 2 HTLV I and II Hepatitis B, C and D Additional Requirements of the Directive The Directive includes ongoing obligations for the manufacturer with regards to experience gained in the post-production phase, including implementation of any necessary corrective actions. The manufacturer must maintain a Vigilance System and notify the regulatory authorities of any serious incident which could or has put a patient at risk, or requires a product to be systematically recalled. An incident should be reported to the Competent Authority in the country where the incident has occurred. Manufacturers who do not have a registered place of business in the EU must designate an Authorised Representative to perform certain obligations. The Authorised Representative will be the first point of contact for Competent Authorities. As an example in a vigilance or compliance case, they may be asked to provide documents to the Competent Authority on behalf of the manufacturer. 4 A BSI Guide to the In Vitro Diagnostic Directive

5 What is the BSI ing Process? The BSI CE marking process starts with a Customer Needs Evaluation which includes a discussion to learn more about your organization, its regulatory requirements, any challenges and timeline demands. Plus we can answer questions you may have regarding BSI and the process. The next step is the completion of a Customer Information Form which gathers more detailed information about your company and products; this will enable us to prepare a quotation. Once you have accepted the quotation, an application review takes place and the details of the initial assessment and identification of appropriate assessors occurs. If you have an Annex II List A product the criteria for the verification of manufactured product will also be established using three batches of product. Manufacturer Completes Customer Information Form Quotation, Acceptance & Contract Review Audit QMS Customer Needs Evaluation CONFORMITY ASSESSMENT Technical File Review / Design Examination Certificate Decision Stage 1 Verification of Manufactured Product A Scheme Manager, who is part of the in-house IVD Technology Team, will be assigned to handle each customer s account and is responsible for managing all its activities. This enables customers to have one person to contact at BSI that understands their business, as all the Team members have in-depth IVD regulatory and industry experience. The advantage of this structure is that you have a responsive Team who can answer your questions quickly and process your applications, reviews and visits. BSI conducts post assessment reviews to ensure the quality of our work; the first is a peer review and the second a compliance and quality check to ensure we deliver a consistent service to our clients. All approvals have to be maintained and a program of continuing assessment visits will be established. BSI supports you every step of the way. Whether you are a current customer or transferring to BSI, we have a dedicated team of Healthcare Professionals committed to making the process robust, efficient and predictable. In addition BSI is working with two of the leading test laboratories, providing Annex II List A verification of manufactured product testing. Companies using these laboratories that choose to transfer to BSI can keep their same testing facilities which will enable a smooth and more seamless transition. If you require any further information regarding the IVD Directive, please contact the BSI IVD Team who will be pleased to answer your questions. Certificate Decision Stage 2 Issue of Certificate Continuing Assessment Visits A BSI Guide to the In Vitro Diagnostic Directive 5

6 Appendix A General IVD Device i.e. all devices other than self-testing devices or devices appearing in Annex II Device Annex III EC Declaration of Conformity (self declaration) Self Testing IVD excluding those which appear in Annex II Device Annex IV Full Quality Assurance (ISO 13485) Audit by NB Annex V EC Type Examination by NB Annex III EC Declaration of Conformity Annex VI EC Verification (Product Examination) by NB Annex VII Production Quality Assurance (ISO 13485) Audit by NB Product Design Examination (Annex III Section 6) by NB 6 A BSI Guide to the In Vitro Diagnostic Directive

7 IVDs Appearing in Annex II List B Device Annex IV Quality Assurance (ISO 13485) Audit by NB Annex V EC Type Examination by NB Annex VI EC Verification (Product Examination) by NB Annex VII Production Quality Assurance (ISO 13485) Audit by NB IVDs Appearing in Annex II List A Device Annex IV Full Quality Assurance (ISO 13485) Audit by NB Annex V EC Type Examination by NB Product Design Dossier Examination (Annex IV Section 4) by NB Annex VII Production Quality Assurance (ISO 13485) Audit by NB Batch or Product Verification (Annex IV Section 6) by NB Batch or Product Verification (Annex VII Section 5) by NB If you require any further information regarding the IVD Directive, please contact the BSI IVD team who will be pleased to answer your questions. A BSI Guide to the In Vitro Diagnostic Directive 7

8 Global expertise Certification services ISO QMS auditing CE marking Health Canada CMDCAS Japan PAL FDA 510k Third-Party Review Programme FDA Accredited Persons Inspections Australia EU CAB Hong Kong CAB Russian Registration Certification Taiwan TCP Training courses IVD/Technical File CE marking and the Medical Device Directives Auditing Clinical Data Requirements, PMS and Vigilance Device Drug Combinations Environmental Risk Management Software German Medical Device Regulation BSI/USA/307/MS/0112/E Your partner in worldwide compliance: Call BSI today on or visit to start your partnership BSI Group America Inc Sunset Hills Road, Suite 200 Reston, VA USA Tel: Fax: healthcare@bsigroup.com BSI Group Canada Inc Airport Road, Suite 102 Mississauga, ON L4V 1E1 Canada Tel: Fax: inquiry.canada@bsigroup.com BSI Group México, S. de R.L. de C.V. Torre Mayor Av. Paseo de la Reforma No. 505 Piso 41 -Suite CCol. Cuauhtémoc, C.P México, D.F. Tel: Fax: informacion.msmexico@bsigroup.com BSI Group Canada Inc. Rua Gomes de Carvalho, andar Vila Olímpia São Paulo - SP CEP: Tel: Fax: bsibrasil@bsibrasil.com.br BSI is a trusted partner to industry and government with a focus to support their business objectives through the transfer of knowledge of best practices, assurance services to identify and measure performance indicators, training services to aid the building of organizational competency and enable continuous improvement, and the tools to monitor, enhance and report on compliance against the organization s management system objectives. Copyright 2012 The British Standards Institution. All Rights Reserved. The BSI certification mark may be used on your stationery, literature and vehicles when you have successfully achieved certification and conform with applicable guidelines. The mark shall never be applied directly on the product or service.

Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC

Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC August 2013 Contents 1 Introduction...3 2 Scope of the directive...3 2.1 What is an in vitro diagnostic medical device?... 3 2.2 Specimen

More information

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES A WHITE PAPER ON THE REQUIREMENTS, REGULATIONS AND OPPORTUNITIES CONTAINED IN EC DIRECTIVE 98/79/EC AND A BRIEF COMPARISON

More information

Want to know more about the Notified Body?

Want to know more about the Notified Body? NBG/V 2/07/14 BSI Notified Body (0086) Kitemark Court Davy Avenue Knowlhill Milton Keynes MK5 8PP United Kingdom T: +44 845 080 9000 E: eu.medicaldevices@bsigroup.com W: medicaldevices.bsigroup.com Want

More information

Future Regulatory Challenges for In VitroDiagnostics. Trilateral Workshop, Berlin 16&17 June 2010

Future Regulatory Challenges for In VitroDiagnostics. Trilateral Workshop, Berlin 16&17 June 2010 Future Regulatory Challenges for In VitroDiagnostics Trilateral Workshop, Berlin 16&17 June 2010 Doris-Ann Williams Director General The British In Vitro Diagnostics Association 2 European Regulation of

More information

PROPOSED DOCUMENT. Global Harmonization Task Force. Title: Principles of In Vitro Diagnostic (IVD) Medical Devices Classification

PROPOSED DOCUMENT. Global Harmonization Task Force. Title: Principles of In Vitro Diagnostic (IVD) Medical Devices Classification SG1(PD)/N045R12 PROPOSED DOCUMENT Global Harmonization Task Force Title: Principles of In Vitro Diagnostic (IVD) Medical Devices Classification Authoring Group: Study Group 1 of the Global Harmonization

More information

Medical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day

Medical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day Medical Devices Notified Bodies and the CE certification Process for Medical Devices European Surgical Robotics Demonstration Day 27 th March 2014, Leuven, Belgium 1 Bart Mersseman Product Manager Medical

More information

2.0 Rationale, Purpose and Scope... 5. 4.0 Definitions... 6. 5.0 General Principles... 7

2.0 Rationale, Purpose and Scope... 5. 4.0 Definitions... 6. 5.0 General Principles... 7 Table of Contents Principles of IVD Medical Devices Classification 1.0 Introduction... 4 2.0 Rationale, Purpose and Scope... 5 2.1 Rationale... 5 2.2 Purpose... 5 2.3 Scope... 5 3.0 References... 5 4.0

More information

SMF Awareness Seminar 2014

SMF Awareness Seminar 2014 SMF Awareness Seminar 2014 Clinical Evaluation for In Vitro Diagnostic Medical Devices Dr Jiang Naxin Health Sciences Authority Medical Device Branch 1 In vitro diagnostic product means Definition of IVD

More information

How To Know If A Mobile App Is A Medical Device

How To Know If A Mobile App Is A Medical Device The Regulation of Medical Device Apps Prepared for West of England Academic Health Science Network and University of Bristol June 2014 June 2014 1 Table of Contents 1 Purpose...3 2 Scope...3 3 The Regulation

More information

2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002

2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002 STATUTORY INSTRUMENTS 2002 No. 618 CONSUMER PROTECTION The Medical Devices Regulations 2002 Made - - - - - 20th May 2002 Laid before Parliament 21st May 2002 Coming into force - - 13th June 2002 ARRANGEMENT

More information

Medical Devices: CE Marking Step-by Step

Medical Devices: CE Marking Step-by Step Medical Devices: CE Marking Step-by Step Sylvia Mohr Standards Specialist U.S. Mission to the European Union 1 Introduction CE marking in 8 Steps Overview intended purpose - technical documentation classification

More information

GUIDELINES ON MEDICAL DEVICES. IVD GUIDANCES : Borderline issues A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES

GUIDELINES ON MEDICAL DEVICES. IVD GUIDANCES : Borderline issues A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single Market : regulatory environment, standardisation and New Approach Pressure equipment, medical devices, metrology MEDDEV. 2.14/1 rev. 1 January

More information

(OJ L 169, 12.7.1993, p. 1)

(OJ L 169, 12.7.1993, p. 1) 1993L0042 EN 11.10.2007 005.001 1 This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents B COUNCIL DIRECTIVE 93/42/EEC of 14 June 1993 concerning

More information

MEDICAL DEVICES INTERIM REGULATION

MEDICAL DEVICES INTERIM REGULATION Issued by the Saudi Food and Drug Authority Board of Directors decree number 1-8-1429 and dated 27 December 2008 The Interim was published in Umm Al-QURA Journal year 86 Issue No:4249 dated in 17 April

More information

Safeguarding public health The Regulation of Software as a Medical Device

Safeguarding public health The Regulation of Software as a Medical Device Safeguarding public health The Regulation of Software as a Medical Device Dr Head of New and Emerging Technologies Medicines and Healthcare Products Regulatory Agency (MHRA) Crown Copyright 2013 About

More information

CE Marking: Your Key to Entering the European Market

CE Marking: Your Key to Entering the European Market CE Marking: Your Key to Entering the European Market N. Gerard Zapiain U.S. Department of Commerce International Trade Administration Office of Microelectronics, Medical Equipment and Instrumentation I.

More information

GLOBAL FUND QUALITY ASSURANCE POLICY FOR DIAGNOSTICS PRODUCTS. (Issued on 14 December 2010, amended on 5 February 2014)

GLOBAL FUND QUALITY ASSURANCE POLICY FOR DIAGNOSTICS PRODUCTS. (Issued on 14 December 2010, amended on 5 February 2014) GLOBAL FUND QUALITY ASSURANCE POLICY FOR DIAGNOSTICS PRODUCTS (Issued on 14 December 2010, amended on 5 February 2014) BASIC PRINCIPLES 1. Grant funds provided by the Global Fund may only be used to procure

More information

Registration of Class B Medical Devices

Registration of Class B Medical Devices Registration of Class B Medical Devices 1 INTRODUCTION Medical devices are classified based on a rule based risk classification system into four risk classes Class A to D with Class A being the lowest

More information

EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below.

EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below. EMKI COMPETENCE in MD EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below. Annex II.: EC DECLARATION OF CONFORMITY

More information

Quality Management System Certification. Understanding Quality Management System (QMS) certification

Quality Management System Certification. Understanding Quality Management System (QMS) certification Quality Management System Certification Understanding Quality Management System (QMS) certification The medical device manufacturing sector is one of the most regulated sectors in which significant quality

More information

Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca

Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca Software Regulated as a Medical Device Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca Therapeutic Products Directorate Director General S. Sharma

More information

Preparing for Unannounced Inspections from Notified Bodies

Preparing for Unannounced Inspections from Notified Bodies Preparing for Unannounced Inspections from Notified Bodies Europe has introduced further measures for unannounced audits of manufacturers by notified bodies. With this in mind, James Pink, VP Europe-Health

More information

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES Foreword GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES These guidance notes do not aim to be a definite interpretation of National Laws and/or regulations and are for guidance purpose only.

More information

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE White paper produced by Maetrics For more information, please contact global sales +1 610 458 9312 +1 877 623 8742 globalsales@maetrics.com

More information

BSI Road Show: September 8 th to 15 th, 2014

BSI Road Show: September 8 th to 15 th, 2014 BSI Road Show: September 8 th to 15 th, 2014 Post Market Surveillance (including PMCF): common non compliances Ibim Tariah Ph.D Technical Director, Healthcare Solutions Itoro Udofia Ph.D Global Head, Orthopaedics

More information

EPF Position Statement on the European Commission s proposal for a Regulation on In Vitro Medical Devices

EPF Position Statement on the European Commission s proposal for a Regulation on In Vitro Medical Devices EPF Position Statement on the European Commission s proposal for a Regulation on In Vitro Medical 03/09/2013 In Vitro Diagnostic medical devices include all tests performed to provide a diagnosis by assessing

More information

Med-Info Regular Information Bulletins for the Medical Device Industry

Med-Info Regular Information Bulletins for the Medical Device Industry Med-Info Regular Information Bulletins for the Medical Device Industry Product Service Globalization of Medical Device Approval Mutual Recognition Agreements (MRA) between the EU and the USA and Australia

More information

Med-Info. Council Directive 93/42/EEC on Medical Devices. TÜV SÜD Product Service GmbH

Med-Info. Council Directive 93/42/EEC on Medical Devices. TÜV SÜD Product Service GmbH Med-Info International expert information for the Medical Device industry Council Directive 93/42/E on Medical Devices Practice-oriented summary of the most important aspects and requirements contained

More information

THE POST-MARKET PRIORITY Understanding and Meeting Demand for Effective Post-Market Clinical Follow-Up

THE POST-MARKET PRIORITY Understanding and Meeting Demand for Effective Post-Market Clinical Follow-Up Healthcare THE POST-MARKET PRIORITY Understanding and Meeting Demand for Effective Post-Market Clinical Follow-Up Introduction In his recent paper, Basic Anatomy of a Design Dossier, Dr. Hamish Forster,

More information

Reporting Changes to your Notified Body

Reporting Changes to your Notified Body Reporting Changes to your Notified Body Vicky Medley - Head of QMS, Medical Devices September 2015 1) Requirements & Guidance 2) Reporting Changes - What, When, Why & How 3) Some examples 4) Summary &

More information

DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL

DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL 29.3.2014 Official Journal of the European Union L 96/149 DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 26 February 2014 on the harmonisation of the laws of the Member States relating

More information

The European Union s EMC Directive 2014/30/EU. Ross Carlton, Sr. EMC Engineer National Instruments

The European Union s EMC Directive 2014/30/EU. Ross Carlton, Sr. EMC Engineer National Instruments The European Union s EMC Directive 2014/30/EU Ross Carlton, Sr. EMC Engineer National Instruments EU Directives Overview Historical perspective The Old Approach o Detailed guidance with all necessary technical

More information

CE mark now! Mandatory CE marking of construction products 1 July 2013

CE mark now! Mandatory CE marking of construction products 1 July 2013 CE mark now! Mandatory CE marking of construction products 1 July 2013 Mandatory CE marking of construction products 1 July 2013 From 1 July 2013, construction products covered by a harmonised standard

More information

Manual for ITC's clients, 2013. Conformity assessment of ACTIVE IMPLANTABLE MEDICAL DEVICES pursuant to Council Directive 90/385/EEC

Manual for ITC's clients, 2013. Conformity assessment of ACTIVE IMPLANTABLE MEDICAL DEVICES pursuant to Council Directive 90/385/EEC Manual for ITC's clients, 2013 Conformity assessment of ACTIVE IMPLANTABLE MEDICAL DEVICES pursuant to Council Directive 90/385/EEC Institute for Testing and Certification, inc., Czech Republic, EU 1.

More information

How to apply for Medical Device License in Taiwan

How to apply for Medical Device License in Taiwan How to apply for Medical Device License in Taiwan According to Pharmaceutical Affairs Act (PAA), all medical devices regardless their classification shall apply for Medical Device License before they are

More information

Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994

Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994 Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical SUR-G0014-1 12 NOVEMBER 2013 This guide does not purport to be an interpretation of law and/or regulations

More information

Medical Device Regulatory Requirements for Taiwan

Medical Device Regulatory Requirements for Taiwan Medical Device Regulatory Requirements for Taiwan Disclaimer: The information contained on this website is derived from public sources and is current to the best of the knowledge of the Department of Commerce

More information

RECALLS in EUROPE. Past, present, near & further future. Gert Bos BSI Medcon May 2012

RECALLS in EUROPE. Past, present, near & further future. Gert Bos BSI Medcon May 2012 RECALLS in EUROPE Past, present, near & further future Gert Bos BSI Medcon May 2012 A C P D Vigilance, vigilant: Continually watchful or on guard; always prepared for possible danger Content 2 Current

More information

Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System

Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System NBOG s Best Practice Guide applicable for AIMDD, MDD, and IVDD 2014-3 Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System 1 Introduction The

More information

INTEROPERABILITY UNIT

INTEROPERABILITY UNIT INTEROPERABILITY UNIT MODULES FOR THE PROCEDURES FOR ASSESSMENT OF CONFORMITY, SUITABILITY FOR USE AND EC VERIFICATION TO BE USED IN THE TECHNICAL SPECIFICATIONS FOR INTEROPERABILITY Reference: Version

More information

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? CEMO Congress, Paris, February 3rd, 2011 Dr. Martine Dehlinger-Kremer Vice President Global Regulatory Affairs Agenda

More information

GUIDELINES ON MEDICAL DEVICES

GUIDELINES ON MEDICAL DEVICES EUROPEAN COMMISSION DIRECTORATE GENERAL for HEALTH and CONSUMERS Consumer Affairs Health technology and Cosmetics MEDDEV 2.12/2 rev2 January 2012 GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP

More information

Vigilance Reporting. Vicky Medley - Head of QMS, Medical Devices. September 2015. Copyright 2015 BSI. All rights reserved.

Vigilance Reporting. Vicky Medley - Head of QMS, Medical Devices. September 2015. Copyright 2015 BSI. All rights reserved. Vigilance Reporting Vicky Medley - Head of QMS, Medical Devices September 2015 2 Why? 3 protecting and improving public health https://www.gov.uk/government/organisations/medicines-and-healthcareproducts-regulatory-agency/about

More information

DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL

DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL L 218/82 EN Official Journal of the European Union 13.8.2008 DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL DECISION No 768/2008/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of

More information

CE Marking Changing & Evolving Requirements September 2013 2013 BSI Healthcare Roadshow

CE Marking Changing & Evolving Requirements September 2013 2013 BSI Healthcare Roadshow CE Marking Changing & Evolving Requirements September 2013 2013 BSI Healthcare Roadshow Existing Directives 2 Directive 93/42/EEC 1. Check Scope of Medical Device Directive (Article 1) 2. Determine Device

More information

Quality Management System Certification. Understanding Quality Management System (QMS) certification

Quality Management System Certification. Understanding Quality Management System (QMS) certification Quality Management System Certification Understanding Quality Management System (QMS) certification The medical device manufacturing sector is one of the most regulated sectors in which significant quality

More information

MEDICAL DEVICE GUIDANCE

MEDICAL DEVICE GUIDANCE May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained

More information

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES Foreword This guidance document is informative and advisory and has no legal authority. Individual national enforcement authorities are bound

More information

THE CZECH MARKET WITH IN VITRO DIAGNOSTIC MEDICAL DEVICES

THE CZECH MARKET WITH IN VITRO DIAGNOSTIC MEDICAL DEVICES THE CZECH MARKET WITH IN VITRO DIAGNOSTIC MEDICAL DEVICES 1 Motto: Finances used in relation to IVD are not costs but true investments in health. Czech Association of In Vitro Diagnostic Medical Suppliers

More information

BSI Unannounced Audits

BSI Unannounced Audits BSI Unannounced Audits Frequently Asked Questions (FAQs) BSI Group Kitemark Court, Davy Avenue Knowlhill, Milton Keynes MK5 8PP, United Kingdom T: +44 845 080 900 Certification.sales@bsigroup.com bsigroup.com

More information

Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR)

Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR) Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR) Paul Brooks Healthcare Solutions Disclaimer The new regulations are not finalized and subject to change http://medicaldevices.bsigroup.com/en-us/resources/whitepapers-and-articles/

More information

Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device

Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device List of Contents Page 1. Introduction 1.1 The Medical Device Administrative Control System and the 3 proposed legislation

More information

Medical Devices. IRELAND Arthur Cox

Medical Devices. IRELAND Arthur Cox Medical Devices IRELAND Arthur Cox CONTACT INFORMATION Colin Kavanagh Arthur Cox Earlsfort Centre Earlsfort Terrace Dublin 2, Ireland +353.1.618.0548 Colin.Kavanagh@arthurcox.com www.arthurcox.com 1. Definition

More information

Presented by Rosemarie Bell 24 April 2014

Presented by Rosemarie Bell 24 April 2014 Global Good Manufacturing Practice A Comparability Study to Link Good Manufacturing Practice Standards for World Wide Compliance Within the Cellular Therapy Industry Presented by Rosemarie Bell 24 April

More information

A GUIDE FOR THE QUALITY ASSURANCE

A GUIDE FOR THE QUALITY ASSURANCE WHO/BCT/03.02 DISTR.: GENERAL A GUIDE FOR THE QUALITY ASSURANCE OF SINGLE USE INJECTION EQUIPMENT Final - 17 February 2003 World Health Organization, Department of Blood Safety and Clinical Technology,

More information

In vitro diagnostic reagent, calibrator and control material stability

In vitro diagnostic reagent, calibrator and control material stability In vitro diagnostic reagent, calibrator and control material stability Pauline Armstrong Randox Laboratories, Ltd. Crumlin The stability of in vitro medical diagnostic devices whether they are reagents,

More information

Medical Certification: Bringing genomic microcores to clinical use OI- VF- WP- 011

Medical Certification: Bringing genomic microcores to clinical use OI- VF- WP- 011 Medical Certification: Bringing genomic microcores to clinical use VoltedFlow GmbH Industriestrasse 23, 6055, Alpnach Dorf, Switzerland 1 Summary VoltedFlow has developed solutions to vastly speed up DNA

More information

New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01)

New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01) Safeguarding public health New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01) Tony Orme, Senior GDP Inspector Inspection, Enforcement and Standards Division

More information

CANADIAN RULE BASED CLASSIFICATION SYSTEM (IVDD) Life Sciences British Columbia NRC-Industry Research Assistance Program

CANADIAN RULE BASED CLASSIFICATION SYSTEM (IVDD) Life Sciences British Columbia NRC-Industry Research Assistance Program CANADIAN RULE BASED CLASSIFICATION SYSTEM (IVDD) Life Sciences British Columbia NRC-Industry Research Assistance Program Health Canada Regulations on Medical Devices Vancouver, B.C. October 29, 2007 Nancy

More information

What is ATEX? The European Regulatory Framework for Manufacture, Installation and Use of Equipment in Explosive Atmospheres

What is ATEX? The European Regulatory Framework for Manufacture, Installation and Use of Equipment in Explosive Atmospheres What is ATEX? The European Regulatory Framework for Manufacture, Installation and Use of Equipment in Explosive Atmospheres Ron Sinclair MBE Chair IECEx ExTAG ATEX? ATEX = Atmosphères Explosibles ATEX

More information

Final Document. Software as a Medical Device (SaMD): Key Definitions. Date: 9 December 2013. Despina Spanou, IMDRF Chair. IMDRF/SaMD WG/N10FINAL:2013

Final Document. Software as a Medical Device (SaMD): Key Definitions. Date: 9 December 2013. Despina Spanou, IMDRF Chair. IMDRF/SaMD WG/N10FINAL:2013 Final Document Title: Authoring Group: Software as a Medical Device (SaMD): Key Definitions IMDRF SaMD Working Group Date: 9 December 2013 Despina Spanou, IMDRF Chair This document was produced by the

More information

Your essential guide to CE Marking

Your essential guide to CE Marking Your essential guide to CE Marking CE marking A guide for window and door fabricators This guide provides essential information that will help you with your preparations for CE marking. A flowchart is

More information

Med-Info. New Act on Medical Devices in Japan. TÜV SÜD Product Service GmbH. International expert information for the Medical Device industry

Med-Info. New Act on Medical Devices in Japan. TÜV SÜD Product Service GmbH. International expert information for the Medical Device industry Med-Info International expert information for the Medical Device industry New Act on Medical Devices in Japan New Act on Medical Devices (PMD Act) to be launched on November 25, 2014 Practice-oriented

More information

EU DIRECTIVE ON GOOD CLINICAL PRACTICE IN CLINICAL TRIALS DH & MHRA BRIEFING NOTE

EU DIRECTIVE ON GOOD CLINICAL PRACTICE IN CLINICAL TRIALS DH & MHRA BRIEFING NOTE EU DIRECTIVE ON GOOD CLINICAL PRACTICE IN CLINICAL TRIALS DH & MHRA BRIEFING NOTE Purpose 1. The Clinical Trials Directive 2001/20/EC heralds certain additional responsibilities for the Medicines and Healthcare

More information

GLOBAL DATABASE ON BLOOD SAFETY

GLOBAL DATABASE ON BLOOD SAFETY GLOBAL DATABASE ON BLOOD SAFETY Summary Report 1998 1999 World Health Organization Blood Transfusion Safety 1211 Geneva 27, Switzerland Tel: +41 22 791 4385 Fax: +41 22 791 4836 http://www.who.int/bct/bts

More information

AMERICAN INTERNATIONAL COMPANIES Page 1 ORGAN PROCUREMENT ORGANIZATION & TISSUE BANK SUPPLEMENTAL APPLICATION

AMERICAN INTERNATIONAL COMPANIES Page 1 ORGAN PROCUREMENT ORGANIZATION & TISSUE BANK SUPPLEMENTAL APPLICATION AMERICAN INTERNATIONAL COMPANIES Page 1 This is a Supplemental Application which becomes part of the Lexington Miscellaneous Facilities Application. The Applicant represents that the statements and facts

More information

PUBLICATIONS. Introduction

PUBLICATIONS. Introduction Page: 1 of 10 Introduction 1. The requirements of this document supplement the Factory Production Control requirements given in Table 1 of PN111 by giving the specific requirements for Factory Production

More information

Emerging Device Topics for Regulatory Consideration.. Janine Jamieson May 2015

Emerging Device Topics for Regulatory Consideration.. Janine Jamieson May 2015 Emerging Device Topics for Regulatory Consideration. Janine Jamieson May 2015 Disclaimer These are my personal views and not necessarily those of MHRA as an organisation. 2 European regulation of combination

More information

FINAL DOCUMENT. Global Harmonization Task Force. Title: Label and Instructions for Use for Medical Devices

FINAL DOCUMENT. Global Harmonization Task Force. Title: Label and Instructions for Use for Medical Devices GHTF/SG1/N70:2011 FINAL DOCUMENT Global Harmonization Task Force Title: Label and Instructions for Use for Medical Devices Authoring Group: Study Group 1 of the Global Harmonization Task Force Endorsed

More information

MEDICAL DEVICES SECTOR

MEDICAL DEVICES SECTOR MEDICAL DEVICES SECTOR MDS - G1 GUIDANCE FOR MEDICAL DEVICE IMPORTERS Version 2 Our mission is to ensure the safety of food; the safety, quality and efficacy of drugs; and the safety and effectiveness

More information

Med-Info. Malaysia Medical Device Regulations. TÜV SÜD Product Service GmbH. International expert information for the medical device industry

Med-Info. Malaysia Medical Device Regulations. TÜV SÜD Product Service GmbH. International expert information for the medical device industry Med-Info International expert information for the medical device industry Malaysia Medical Device Regulations Passed in 2012, the Medical Device Act (Act 737) and the Medical Device Authority Act 2012

More information

A New Framework for the EU EMC Directive

A New Framework for the EU EMC Directive A New Framework for the EU EMC Directive The European Union New Legislative Framework and New EMC Directive Abstract Changes in the European Union legislative process over the past decade have resulted

More information

Medical Device Software Do You Understand How Software is Regulated?

Medical Device Software Do You Understand How Software is Regulated? Medical Device Software Do You Understand How Software is Regulated? By Gregory Martin Agenda Relevant directives, standards, and guidance documents recommended to develop, maintain, and validate medical

More information

Clinical evaluation Latest development in expectations EU and USA

Clinical evaluation Latest development in expectations EU and USA Clinical evaluation Latest development in expectations EU and USA Medical Devices: staying ahead of regulatory developments Gert Bos BSI Israel 22 April - Herzliya Copyright 2012 BSI. All rights reserved.

More information

An introduction to the regulation of apps and wearables as medical devices

An introduction to the regulation of apps and wearables as medical devices An introduction to the regulation of apps and wearables as medical devices Introduction Phones are increasingly equipped with a range of devices capable of being used as sensors, including gyroscopes,

More information

Adoption by GCP Inspectors Working Group for consultation 14 June 2011. End of consultation (deadline for comments) 15 February 2012

Adoption by GCP Inspectors Working Group for consultation 14 June 2011. End of consultation (deadline for comments) 15 February 2012 10 December 2013 EMA/INS/GCP/600788/2011 Compliance and Inspection Reflection paper on the use of interactive response technologies (interactive voice/web response systems) in clinical trials, with particular

More information

GUIDELINES ON MEDICAL DEVICES EVALUATION OF CLINICAL DATA : A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES

GUIDELINES ON MEDICAL DEVICES EVALUATION OF CLINICAL DATA : A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single Market : regulatory environment, standardisation and New Approach Pressure equipment, medical devices, metrology MEDDEV. 2.7.1 April 2003 GUIDELINES

More information

Reflection paper on the Use of Interactive Response Technologies (Interactive Voice/Web Response Systems) in Clinical Trials

Reflection paper on the Use of Interactive Response Technologies (Interactive Voice/Web Response Systems) in Clinical Trials 1 2 3 5 August 2011 EMA/INS/GCP/600788/2011 Compliance and Inspection 4 5 6 7 Reflection paper on the Use of Interactive Response Technologies (Interactive Voice/Web Response Systems) in Clinical Trials

More information

Regulatory Considerations for Medical Device Software. Medical Device Software

Regulatory Considerations for Medical Device Software. Medical Device Software Medtec Ireland 2015 Wireless Medical Devices Regulatory Considerations for Medical Device Software Kenneth L. Block, RAC October 7, 2015 Galway, Ireland Offices: Dallas, Texas (12 employees) Tokyo, Japan

More information

Qualified Persons in the Pharmaceutical Industry Code of Practice 2009, updated August 2015

Qualified Persons in the Pharmaceutical Industry Code of Practice 2009, updated August 2015 Qualified Persons in the Pharmaceutical Industry Code of Practice 2009, updated August 2015 *QP Code of Practice 2008 updated Aug15 Page 1 of 13 Code of Practice for Qualified Persons 1. INTRODUCTION 2.

More information

How To Inspect A Blood Bank

How To Inspect A Blood Bank Site visit inspection report on compliance with HTA minimum standards Belfast Cord Blood Bank HTA licensing number 11077 Licensed for the procurement, processing, testing, storage, distribution and import/export

More information

TG 47-01. TRANSITIONAL GUIDELINES FOR ISO/IEC 17021-1:2015, ISO 9001:2015 and ISO 14001:2015 CERTIFICATION BODIES

TG 47-01. TRANSITIONAL GUIDELINES FOR ISO/IEC 17021-1:2015, ISO 9001:2015 and ISO 14001:2015 CERTIFICATION BODIES TRANSITIONAL GUIDELINES FOR ISO/IEC 17021-1:2015, ISO 9001:2015 and ISO 14001:2015 CERTIFICATION BODIES Approved By: Senior Manager: Mpho Phaloane Created By: Field Manager: John Ndalamo Date of Approval:

More information

PRCP CE. Version 5.0 of 16 June 2013. Date of entry into force: 1 July 2013

PRCP CE. Version 5.0 of 16 June 2013. Date of entry into force: 1 July 2013 npo Founded by SECO and BBRI PARTICULAR RULES FOR THE CERTIFICATION OF PRODUCTS AND PRODUCTION CONTROL SYSTEMS WITHIN DE FRAMEWORK OF THE EUROPEAN CONSTRUCTION PRODUCTS REGULATION N 305/2011 PRCP CE Version

More information

ASSESSMENT OF QUALITY RISK MANAGEMENT IMPLEMENTATION

ASSESSMENT OF QUALITY RISK MANAGEMENT IMPLEMENTATION PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 038-1 26 March 2012 AIDE-MEMOIRE ASSESSMENT OF QUALITY RISK MANAGEMENT IMPLEMENTATION PIC/S March 2012 Reproduction

More information

Basic Information about the European Directive 93/42/EEC on Medical Devices

Basic Information about the European Directive 93/42/EEC on Medical Devices Basic Information about the European Directive 93/42/EEC on Medical Devices mdc medical device certification GmbH Kriegerstraße 6 D-70191 Stuttgart, Germany Phone: +49-(0)711-253597-0 Fax: +49-(0)711-253597-10

More information

Diagnostic Tests. Brad Spring Director, Regulatory Affairs

Diagnostic Tests. Brad Spring Director, Regulatory Affairs Regulatory Challenges in the Development of Diagnostic Tests Industry Perspective Brad Spring Director, Regulatory Affairs September 28, 2011 Balancing Evidence of Safety & Effectiveness and Time to Market

More information

Gas Detector System, Portable/Fixed Oxygen Analysis and Gas Detection Equipment

Gas Detector System, Portable/Fixed Oxygen Analysis and Gas Detection Equipment STANDARD FOR CERTIFICATION No. 2.9 TYPE APPROVAL PROGRAMME NO. 7-473.30/31 / MED A.1/3.30/54 Gas Detector System, Portable/Fixed Oxygen Analysis and Gas Detection Equipment SEPTEMBER 2011 The content of

More information

The Regulation of Medical Devices in the UK

The Regulation of Medical Devices in the UK The Regulation of Medical Devices in the UK 1. What are the main requirements for a medical device to enter into the market and how are they regulated? Are there any licensing, marketing authorisation,

More information

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians Prepared by the Advisory Committee on Laboratory Medicine College of Physicians & Surgeons of Alberta Serving the Public by guiding

More information

EXTENSION OF PERIODIC INSPECTION INTERVAL FOR LPG CYLINDERS

EXTENSION OF PERIODIC INSPECTION INTERVAL FOR LPG CYLINDERS EXTENSION OF PERIODIC INSPECTION INTERVAL FOR LPG CYLINDERS VCA guidance for gaining authorisation in Great Britain Page 1 of 7 EXTENSION OF PERIODIC INSPECTION INTERVAL FOR LPG CYLINDERS VCA guidance

More information

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i

More information

(Legislative acts) DECISIONS

(Legislative acts) DECISIONS 4.11.2011 Official Journal of the European Union L 287/1 I (Legislative acts) DECISIONS DECISION No 1104/2011/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 25 October 2011 on the rules for access

More information

Comparative analysis between the possible regulatory approaches to GMP compliance TITOLO PRESENTAZIONE

Comparative analysis between the possible regulatory approaches to GMP compliance TITOLO PRESENTAZIONE Comparative analysis between the possible regulatory approaches to GMP compliance TITOLO PRESENTAZIONE Dr. Fulvio CARLOTTI, GNOSIS SpA, Corporate QA Director September 26, 2014 Scope of GMP GMP compliance

More information

Medical Device Approvals in Brazil: A Review and Update

Medical Device Approvals in Brazil: A Review and Update Medical Device Approvals in Brazil: A Review and Update Medical Device Approvals in Brazil: A Review and Update With a population of more than 190 million people and a mature healthcare system, Brazil

More information

A Revised EMC Directive from Europe

A Revised EMC Directive from Europe TÜV America Inc. Phone: (978) 739-7000 5 Cherry Hill Drive Fax: (978) 777-8441 Danvers, MA 01923 E-mail: info@tuvam.com www.tuvamerica.com A Revised EMC Directive from Europe INTRODUCTION The Commission

More information

APPENDIX A GUIDANCE DOCUMENT

APPENDIX A GUIDANCE DOCUMENT APPENDIX A GUIDANCE DOCUMENT Infectious Substances International Civil Aviation Organization Technical Instructions for the Safe Transport of Dangerous Goods by Air, 2005-2006 Note: This guidance document

More information

5th Edition NetCord-FACT International Standards for Cord Blood Collection, Banking, and Release for Administration. Summary of Changes

5th Edition NetCord-FACT International Standards for Cord Blood Collection, Banking, and Release for Administration. Summary of Changes 5th Edition NetCord-FACT International Standards for Cord Blood Collection, Banking, and Release for Administration Summary of Changes This document summarizes the changes made to the 5th edition of the

More information

The Act on Medical Devices

The Act on Medical Devices The Act on Medical Devices The Act on Medical Devices of 2 nd August 1994 (Federal Law Gazette I, p. 1963), in the version of 7 th August 2002 (Federal Law Gazette I, p. 3146), last amended by Article

More information

Shipping of Infectious Substances and Patient Specimens

Shipping of Infectious Substances and Patient Specimens Shipping of Infectious Substances and Patient Specimens If you ship infectious substances or patient specimens you should be aware of the International Air Transport Association (IATA) Dangerous Goods

More information

ANNEX F. Organic agriculture (Art. 11) ARTICLE 1. Objectives

ANNEX F. Organic agriculture (Art. 11) ARTICLE 1. Objectives ANNEX F Organic agriculture (Art. 11) ARTICLE 1 Objectives Without prejudice to their obligations with regard to products not originating in the Member States or other legislative provisions in force,

More information