Specimen Labeling Errors in Surgical Pathology An 18-Month Experience

Size: px
Start display at page:

Download "Specimen Labeling Errors in Surgical Pathology An 18-Month Experience"

Transcription

1 Anatomic Pathology / Specimen Labeling Errors Specimen Labeling Errors in Surgical Pathology An 18-Month Experience Lester J. Layfield, MD, 1 and Gina M. Anderson, HT(ASCP) 2 Key Words: Errors; Labeling; Surgical pathology; Laboratories DOI: /AJCPHLQHJ0S3DFJK Abstract Elimination of medical errors is important for pathologists. Errors occurring in surgical pathology involve specimen defects, specimen labeling, processing, diagnosis, and reporting defects. Errors occur during prelaboratory, laboratory, and postlaboratory phases. We reviewed our experience with mislabeled specimens in the laboratory for an 18-month period. The percentage of error was calculated on a per case, block, and slide basis. Errors were characterized by site and as incorrect patient or site. The study involved 75 labeling errors (0.25% of cases) that were detected. Of the 75 errors, 55 (73%) involved patient name, and 18 (24%) involved site. The majority of mislabelings (52 [69%]) occurred in the gross room. Although infrequent, labeling errors involved misidentification of patient or specimen source. Of the errors, 73% (55/75) of errors resulted in slides assigned to an incorrect patient. Most errors occurred in the gross room. Newer technologies such as bar coding and radio frequency chip methods may reduce the frequency of specimen labeling errors. Medical error reduction has become a major focus for organized medicine since the publication of the Institute of Medicine s report on medical errors in However, in a recent review of errors in anatomic pathology, errors involving specimen labeling were only briefly mentioned. 2 Labeling errors can result in inappropriate therapy or the withholding of therapy in patients with unrecognized malignancies. Identification errors involving laboratory specimens may involve misidentification of a patient or the patient s specimen 3 or the site from which the specimen was obtained. Such errors may result in significant patient inconvenience or harm. Identification errors are frequent by anecdotal evidence, 3 but few studies have documented the frequency of such errors in the anatomic pathology laboratory. Identification errors fall into 2 categories represented by patient identification errors and specimen identification errors. In a patient identification error, a specimen is labeled with the incorrect patient name or identification number. In a specimen identification error, a specimen is misidentified as to site of origin or time of collection but the specimen is correctly associated with patient name and/or unique identification number. Identification errors can be classified as preanalytic, analytic, and postanalytic. All 3 types are of concern to pathologists. Preanalytic, or clinical, errors can be addressed by standardization of the specimen collection process and feedback to clinical staff. Only the analytic errors are directly addressable by laboratory professionals and surgical pathologists. A number of authors have addressed labeling errors in laboratory medicine, including Q-Probes analyses undertaken by the College of American Pathologists (CAP). 4-6 These authors have described methods for error reduction. 4-6 Of the Q-Probes accessioning errors, 10% were due to misidentification. The 466 Am J Clin Pathol 2010;134: DOI: /AJCPHLQHJ0S3DFJK

2 Anatomic Pathology / Original Article majority of routine practices for the identification of labeling and specimen identification errors probably underestimates the frequency of occurrence of such errors, and many errors may go undetected. 3 Labeling errors may be detected by the laboratory staff, pathologists, or clinicians. Labeling errors detected before patient harm or inconvenience has occurred are frequently designated near misses. 3 Few studies have addressed the frequency of labeling errors for anatomic pathology specimens. In a CAP Q-Probes program, specimen identification and accessioning deficiencies were found in 60,042 of 1,004,115 cases accessioned (6.0%). 3,7 Errors associated with specimen misidentification accounted for 9.6% of the total errors. 7 Labeling errors have been categorized as class 1, typographical errors with no clinical consequences; class 2, minor errors unlikely to have clinical consequences; and class 3, errors that are significant and have the potential to detrimentally impact patient care. 8 Francis et al 8 documented the frequency of preanalytic errors in a high-volume endoscopy center. In a study of 8,231 specimen containers, they documented a 0.09% rate for class 3 errors in which specimen bottles were mislabeled for patient name or unique identification number. They described a method for significantly reducing this error rate for mislabeling with significant clinical implications. 8 A few studies have documented error rates for specimen identification and information transmission in surgical pathology and cytopathology specimens. 9,10 In observational studies by Raab et al 9 and Smith and Raab, 10 no specimen was found to be totally free of defect, although the frequency and type of defect varied considerably. Approximately 1.5% of specimen containers lacked an accurate patient name or second identifier. 9 These near-miss errors required accessioners to spend considerable time (15-45 minutes) correcting these defects. 9 In the study by Smith and Raab, 10 the rate of operator-dependent near-miss events in the gross room was 0.6 events per specimen. The operator-dependent errors were subcategorized into accessioning (wrong accession number written on requisition form), set up (requisition paired with incorrect specimen container), and grossing (different requisition forms in grossing area at the same time). These near misses or potential errors could have had significant impact on patient care in that an incorrect diagnosis might have been rendered for a patient. Makary et al 11 reviewed a total of 21,351 surgical specimens and detected 91 errors in specimen labeling. Of these errors, 11 (12%) involved specimens associated with an incorrect patient assignment. In their study, specimens most commonly mislabeled were breast, skin, and colon. 11 The majority of labeling errors occurred in biopsy specimens. Specimen labeling errors within the laboratory can occur at several points of specimen processing. Within the gross room, specimen containers can be paired with cassettes labeled with an incorrect case number (wrong patient) or an incorrect part number (wrong site). In the histology laboratory, cassettes at the cutting station can be paired with incorrectly pencil-labeled slides (wrong patient or site), or a correctly pencil-labeled slide can have the wrong paper label applied (wrong patient or site). A surgical pathologist may pick up an incorrect slide and dictate a report with an incorrect diagnosis for the patient s laboratory data being reviewed. Errors may also occur during transcription when dictations are transcribed to the wrong report number and patient. Examples of most of these have been described in the literature. Meyer et al 12 reported a 0.59% error rate in placing paper labels on correctly pencil-labeled glass slides. We investigated the frequency of labeling errors in an anatomic pathology practice processing more than 29,000 specimens in an 18-month period. We analyzed the types of specimens involved, laboratory location where the error occurred, the training level of the person making the error, the potential impact on the patient, and the person detecting the error. Herein, we report the results of that study. Materials and Methods The University of Utah Department of Pathology and ARUP Laboratories (Salt Lake City) maintain extensive quality assurance (QA) and quality control (QC) records documenting all labeling errors in the preanalytic, analytic, and postanalytic phases of anatomic pathology specimen analysis and reporting. In addition, the QA records include data on all federal and CAP mandates concerning data required on requisition forms and anatomic pathology reports. We reviewed the QA/QC records for an 18-month period. Only labeling errors occurring in the laboratory phase were analyzed. During this period, 29,479 cases were accessioned, and the QA/QC data concerning these cases were reviewed. Errors were characterized as mislabeled cassettes/blocks (gross room errors), mislabeled slides (histology laboratory), and mislabeled reports. Errors were characterized as incorrect patient or correct patient but incorrect specimen site or location designation. Finally, the type of specimen (biopsy vs resection) and tissue site were recorded along with the training level of the staff making and detecting the errors. Results The 29,479 cases accessioned were associated with 109,354 blocks and 248,013 slides. Characteristics of the errors are shown in Table 1. The percentage of error was calculated on a per case, per block, and per slide basis. The study included 75 labeling errors that were detected (0.25% of Am J Clin Pathol 2010;134: DOI: /AJCPHLQHJ0S3DFJK 467

3 Layfield and Anderson / Specimen Labeling Errors Table 1 Characteristics of 75 Laboratory Labeling Errors Characteristic Type of error Wrong patient 55 Wrong site 18 Incorrect label number but same patient and site 2 Where error occurred Gross room 52 Histology laboratory 19 Pathology office 4 Level of training or position of person making error Certified histology technologist 12 Laboratory assistant, gross room 46 Certified pathology assistant 2 Resident pathologist 4 Transcriptionist 4 Laboratory assistant, histology 7 Level of training or position of person recognizing error Pathologist 33 Certified histotechnologist 39 Transcriptionist 1 Clinician 1 Not recorded 1 Potential impact on patient if error not identified None 62 Wrong therapy 13 cases, 0.068% of blocks, and 0.030% of slides). The majority of mislabeling (52 [69%]) occurred within the gross room owing to incorrect labeling of the cassette submitted to the histology laboratory. These errors were detected in the histology laboratory, by a surgical pathologist or the referring clinician on reading the patient s surgical pathology report. Errors occurring within the histology laboratory were most commonly detected by the sign-out surgical pathologist. The majority of labeling errors occurred in biopsies of the skin, esophagus, kidney, and colon. The distribution of labeling errors by tissue and specimen type reflected the distribution of case types received in surgical pathology. Hence, most errors occurred with small biopsy specimens from the endoscopy and dermatology clinics, which supplied the majority of specimens examined in the surgical pathology laboratory. Discussion No. of Errors Since the Institute of Medicine s report on medical errors in 1999, 1 increased attention has been given to potential sources of error in medical practice. The CAP has performed Q-Probes studies investigating labeling errors in anatomic and clinical pathology. 6,7 Labeling errors may occur in the preanalytic or analytic phases of specimen processing in surgical pathology, as well as during the reporting process. Labeling errors involving the accurate linkage of the patient with the surgical specimen may result in delay of diagnosis or institution of inappropriate therapy. 2,3 Labeling errors resulting in inappropriate identification of a specimen as to tissue type, site, or time of specimen acquisition may result in less harm but still seriously poor outcomes for a patient. Errors in correct lateralization (right vs left) can result in serious complications such as treatment of the wrong organ or limb. We investigated the percentage of mislabeled specimens in a surgical practice during a period of 18 months. We found that approximately 0.25% of cases were associated with a labeling error in the laboratory phase of analysis. In a study by Makary et al 11 evaluating 21,351 surgical specimens, 11 (0.05% of specimens) mislabelings were detected in which labeling errors resulted in assignment of the specimen to an incorrect patient. In their study, the errors were not specifically assigned to the preanalytic or analytic phase of specimen analysis. In our study, we investigated only labeling errors occurring during the analytic or laboratory phase of analysis. The preanalytic phase of specimen labeling was specifically studied in a large endoscopy center. 8 Francis et al 8 studied more than 8,000 specimen containers and demonstrated a 0.09% rate for incorrect specimen labeling. A few other studies had addressed labeling issues in the anatomic pathology laboratory, 7 and the findings are comparable. It seems that the incidence of specimen mislabeling in the preanalytic phase is relatively similar to our experience of errors in the laboratory phase of specimen analysis. The major or class 3 labeling error rate of 0.09% described by Francis et al 8 is lower than our 0.25% incidence of major mislabeling in the analytic phase. Differences in study design may explain our higher rate of error. In the study by Francis et al, 8 the error rate was based on a per specimen bottle rate, whereas our rate was calculated on a per case basis. Because several specimen bottles may be present for a given case, the 2 rates are not directly comparable. For gastrointestinal (GI) biopsy specimens, the number of specimen bottles usually closely approximates the block number. Hence, a better comparison between our study and that by Francis et al 8 is equating our rate of labeling errors per block with their rate of major labeling errors per specimen bottle. In this case, our mislabeling rate per block of 0.068% is similar to the 0.09% reported by Francis et al. 8 This rate is also similar to the 0.05% rate of error for incorrect patient labeling reported by Makary et al. 11 Anatomic pathology labeling errors can occur at a number of points in specimen preparation and analysis. During the grossing in procedure, tissues can be placed in an incorrectly labeled cassette, resulting in assignment to the wrong patient or the wrong site in the correct patient. In the former instance, the patient can receive an incorrect diagnosis resulting in an inappropriate therapy or follow-up. In the latter case, therapy may be directed at the wrong site or stage may be incorrectly assigned. 468 Am J Clin Pathol 2010;134: DOI: /AJCPHLQHJ0S3DFJK

4 Anatomic Pathology / Original Article In our study, 73% of errors resulted in incorrect linkage of a specimen with a patient. Of these wrong patient errors, 69% occurred in the gross room (Table 1). In the majority of cases, the labeled cassettes were switched between 2 patients whose specimens were sequentially grossed in. In all cases, these switched cases were small biopsy specimens in which the prelabeled cassettes had been placed with the incorrect paperwork and specimen containers. The staff members working in the gross room include pathology residents, certified pathology assistants, and laboratory assistants. The laboratory assistants had at least 2 years of college and received on-the-job training but were not certified. Of the specimen labeling errors, 46 (88%) of 52 were made by laboratory assistants. While laboratory assistants prepare the majority of small cases, they seem to be overrepresented in cases involving mismatching of specimen with labeled cassette. Wrong site mislabelings were less frequent and were due to incorrect assignment of a subset number or letter to a cassette. The majority of these errors were made by laboratory assistants (11/18 [61%]), with most of the other wrong site errors being made by pathology residents. Raab et al 9 in a direct observation of specimen accessioning and grossing documented that 1.7% of specimen containers lacked an accurate patient name or second identifier, 3.4% lacked a date of collection, and 7.3% lacked a physician name. In that study, the frequency of specimen defect types varied considerably by preanalytic collection site, and patient name defects occurred only in some clinics. They considered the root cause of the majority of defects to be the result of lack of standardized preanalytic processes and lack of redundant checks. 9 Smith and Raab 10 in a similar observational study demonstrated that near-miss events were 0.6 events per specimen. These errors were subcategorized into accessioning, set-up, and grossing errors. Errors in these areas occurred at a rate of 5.5 events per specimen. The most important of these errors were assignment of a wrong accession number to a requisition form, the requisition paired with an incorrect specimen container, and 2 or more requisition forms in the grossing area at the same time. They concluded that traditional batch flow design in surgical pathology grossing facilities led to unacceptably high rates of process and operator-dependent near misses. Our experience with mislabeled specimens is similar in that the majority of errors (69%) occurred in the gross room and involved the mislabeling of cassettes with subsequent block and slide being associated with an incorrect patient or incorrect site. These near-miss errors were usually recognized in the histopathology laboratory by histotechnologists or subsequently by surgical pathologists signing out the cases. Of the mislabelings, 19 (25%) occurred in the histology laboratory. All mislabelings occurred during 1 of 2 procedures. The first procedure was performed by certified histotechnologists and involved blocks being matched with slides prelabeled by pencil on the frosted glass end of the slide. These pencil-labeled slides were matched with the blocks at the time of tissue sectioning and the sections transferred to the glass slides. In 12 cases (63% of histology errors), a certified histotechnologist mislabeled a glass slide before cutting the section or transferred the cut section onto the incorrectly pencil-labeled slide. In 7 cases (37% of histology labeling errors), a laboratory assistant placed an incorrect preprinted label on the pencil-labeled glass slide. This error represented 9% of all labeling errors and occurred in 0.003% of all slides prepared. Meyer et al, 12 in a retrospective review of slide labeling in Papanicolaou testing, found incorrect attachment of a paper label to a previously pencil-labeled slide to occur in 0.59% of slides. They developed an alternative technique in which the paper label was placed at the opposite end from the pencil labeling of the slide. This method seemed to substantially reduce labeling errors to nearly zero. Of the labeling errors, 4 (5% of total errors) occurred in the pathology office where office staff placed an incorrect case number (wrong patient) on outside slides submitted for internal review. All 4 cases were identified by the sign-out pathologist because the tissue on the slide did not match the tissue described in the outside pathology report and the outside slide number did not match the accession number on the report. GI tract biopsy specimens were the most frequent type of specimen associated with mislabeling. Mislabeled GI biopsy specimens represented 16 (21%) of 75 mislabeled specimens. This parallels the frequency of GI biopsies in the case workload of approximately 20%. All errors in the labeling of GI biopsy specimens involved assigning the material to the wrong patient. This clustering of labeling errors in GI biopsy specimens may have resulted from GI biopsy specimens arriving in batches from the endoscopy clinic and processed in the pathology gross room as a group. Because all were grossly similar, the chances for error in matching cassettes with paperwork and specimens were relatively high. Other specimens with a relatively high incidence of labeling errors were renal and skin biopsy specimens. Similar factors of small specimen, similar appearance, and batch processing may have increased the likelihood of specimen mislabeling (cassette mislabeling). Smith and Raab, 10 in an observational study of gross room practices, concluded that batch flow processing led to unacceptably high rates of nearmiss errors. Training level appeared to influence the likelihood that a person would make a labeling error. While the majority of labeling errors in the gross room were made by laboratory assistants (88%), the laboratory assistants also grossed in the majority of specimen cassettes (68%). Thus, their relatively high rate of labeling error is in part explained by grossing in a relatively high percentage of total cassettes/tissue Am J Clin Pathol 2010;134: DOI: /AJCPHLQHJ0S3DFJK 469

5 Layfield and Anderson / Specimen Labeling Errors specimens. Similarly, differences in the number of errors between certified histotechnologists and histology laboratory assistants could not be explained by training level. Of 19 errors occurring in the histology laboratory, 12 (63%) were made by certified histotechnologists and only 7 (37%) by laboratory assistants. The potential for adverse impact due to a labeling error varied considerably in this series. Of 75 labeling errors, 62 (83%) would not have had a significant impact on patient care if they had not been identified in the surgical pathology laboratory or sign-out process. Reasons for a lack of impact were discordance between patient sex and reported specimen source (prostate gland in a female patient), diagnosis without subsequent impact on patient care (cholelithiasis in a patient known not to have had a cholecystectomy), and incorrect diagnosis identical to the actual patient diagnosis (eg, normal vs normal or adenomatous polyp vs adenomatous polyp). Also, diagnoses in which the site was incorrect but did not affect stage, margin status, or need for additional surgery did not have an impact on patient outcome. Of the 75 errors, 13 (17%) could have had a significantly negative impact on a patient. The most common of these errors occurred when 2 specimens from different patients but from the same site were switched and mislabeled. Such errors occurred most commonly in batched specimens and included gastrointestinal and renal biopsy specimens. An example of such a clinically significant labeling error with patient impact was the switching of 2 colonic biopsy specimens from different patients. One biopsy showed an adenocarcinoma, and the other contained only a hyperplastic polyp. Clearly, patient follow-up and therapy would have been significantly different for these 2 patients. The use of the Lean method has been shown to reduce errors in anatomic pathology. 13,14 The Lean method, when coupled with automated systems, can substantially reduce errors in histology laboratories 13 and in other components of anatomic pathology specimen tracking, processing, and reporting. 14 Mislabeled specimens in surgical pathology represent an infrequent but important source of medical error. In the analytic component of specimen processing, most errors occur in the gross room. This clustering of labeling errors may be due to batch processing in which similar specimens may be confused, or it may be a function of staff training level. Avoidance of batch processing and the use of newer technologies such as bar codes on specimen containers, requisition forms, cassettes, and slides or the use of radio frequency chip technology may significantly reduce the incidence of specimen mislabeling. From the 1 Department of Pathology, University of Utah School of Medicine, and 2 ARUP Laboratories, Salt Lake City. Address reprint requests to Dr Layfield: Dept of Pathology, University of Utah School of Medicine, Huntsman Cancer Hospital, 1950 Circle of Hope, Rm 6755, Salt Lake City, UT References 1. Kohn LT, Corrigan JM, Donaldson MS, eds. To Err Is Human: Building a Safer Health System. Washington, DC: National Academy Press; Sirota RL. A framework for error in anatomic pathology. Pathol Case Rev. 2009;14: Valenstein PN, Sirota RL. Identification errors in pathology and laboratory medicine. Clin Lab Med. 2004;24: Specimen labeling QI project cuts error rate. Healthcare Benchmarks Qual Improv. 2006;13: Quillen K, Murphy K. Quality improvement to decrease specimen mislabeling in transfusion medicine. Arch Pathol Lab Med. 2006;130: Wagar EA, Stankovic AK, Raab S, et al. Specimen labeling errors: a Q-Probes analysis of 147 clinical laboratories. Arch Pathol Lab Med. 2008;132: Nakhleh RE, Zarbo RJ. Surgical pathology specimen identification and accessioning: a College of American Pathologists Q-Probes study of 1,004,115 cases from 417 institutions. Arch Pathol Lab Med. 1996;120: Francis DL, Prabhakar S, Sanderson SO. A quality initiative to decrease pathology specimen-labeling errors using radiofrequency identification in a high-volume endoscopy center. Am J Gastroenterol. 2009;104: Raab SS, King AM, Grzybicki DM. Root cause analysis of surgical pathology identification and information defects [abstract]. Mod Pathol. 2009;22(suppl 1):366A. Abstract Smith ML, Raab SS. Near-miss event rates in a traditional surgical pathology accessioning and gross examination laboratory [abstract]. Mod Pathol. 2009;22(suppl 1):366A. Abstract Makary MA, Epstein J, Pronovost PJ, et al. Surgical specimen identification errors: a new measure of quality in surgical care. Surgery. 2007;141: Meyer E, Underwood RS, Padmanabhan V. Patient misidentification in Papanicolaou tests: a systems-based approach to reducing errors. Arch Pathol Lab Med. 2009;133: Hassell LA, Glass CF, Yip C, et al. The combined positive impact of Lean methodology and Ventana Symphony autostainer on histology lab workflow. BMC Clin Pathol. 2010;10:2. doi: / Condel JL, Sharbaugh DT, Raab SS. Error-free pathology: applying Lean production methods to anatomic pathology. Clin Lab Med. 2004;24: Am J Clin Pathol 2010;134: DOI: /AJCPHLQHJ0S3DFJK

Factors Affecting Patient Safety and Error Reduction in Surgical Pathology

Factors Affecting Patient Safety and Error Reduction in Surgical Pathology Patient Safety and Error Reduction in Surgical Pathology Raouf E. Nakhleh, MD Context. National patient safety goals and error reduction efforts should be addressed by each surgical pathology laboratory.

More information

CAP Laboratory Improvement Programs. Mislabeling of Cases, Specimens, Blocks, and Slides. A College of American Pathologists Study of 136 Institutions

CAP Laboratory Improvement Programs. Mislabeling of Cases, Specimens, Blocks, and Slides. A College of American Pathologists Study of 136 Institutions CAP Laboratory Improvement Programs Mislabeling of Cases, Specimens, Blocks, and Slides A College of American Pathologists Study of 136 Institutions Raouf E. Nakhleh, MD; Michael O. Idowu, MD; Rhona J.

More information

Automated Patient Barcode Tracking and Sample Preparation Using the Gilson GX 281 Barcode Matching System

Automated Patient Barcode Tracking and Sample Preparation Using the Gilson GX 281 Barcode Matching System Automated Patient Barcode Tracking and Sample Preparation Using the Gilson GX 281 Barcode Matching System Keywords Introduction Application Note CL0413 GX 281 Liquid Handler, sample tracking, barcode matching,

More information

Surgical specimen identification errors: A new measure of quality in surgical care

Surgical specimen identification errors: A new measure of quality in surgical care Surgical specimen identification errors: A new measure of quality in surgical care Martin A. Makary, MD, MPH, a,d,e Jonathan Epstein, MD, b Peter J. Pronovost, MD, PhD, a,c,d,e E. Anne Millman, MS, a,d,e

More information

Error Reduction and Prevention in Surgical Pathology. Raouf E. Nakhleh, MD Professor of Pathology Mayo Clinic Florida

Error Reduction and Prevention in Surgical Pathology. Raouf E. Nakhleh, MD Professor of Pathology Mayo Clinic Florida Error Reduction and Prevention in Surgical Pathology Raouf E. Nakhleh, MD Professor of Pathology Mayo Clinic Florida Disclosure None 2 Course Objectives At the end of the presentation participants should

More information

Patient Misidentification in Laboratory Medicine. A Qualitative Analysis of 227 Root Cause Analysis Reports in the Veterans Health Administration

Patient Misidentification in Laboratory Medicine. A Qualitative Analysis of 227 Root Cause Analysis Reports in the Veterans Health Administration Patient Misidentification in Laboratory Medicine A Qualitative Analysis of 227 Root Cause Analysis Reports in the Veterans Health Administration Edward J. Dunn, MD, ScD; Paul J. Moga, DO, PhD N Context.

More information

PATIENT SAFETY DIAGNOSTIC ERROR SURGICAL PATHOLOGY

PATIENT SAFETY DIAGNOSTIC ERROR SURGICAL PATHOLOGY PATIENT SAFETY & DIAGNOSTIC ERROR in SURGICAL PATHOLOGY Richard J. Zarbo, M.D., D.M.D. Rita D Angelo, M.S., ASQE, SSBB Pathology and Laboratory Medicine Henry Ford Health System Detroit, Michigan Zarbo-D

More information

The Henry Ford Production System Measures of Process Defects and Waste in Surgical Pathology as a Basis for Quality Improvement Initiatives

The Henry Ford Production System Measures of Process Defects and Waste in Surgical Pathology as a Basis for Quality Improvement Initiatives Anatomic Pathology / THE HENRY FORD PRODUCTION SYSTEM The Henry Ford Production System Measures of Process Defects and Waste in Surgical Pathology as a Basis for Quality Improvement Initiatives Rita D

More information

Delphic AP. Designer software for your pathology workflow. We Believe the Possibilities.

Delphic AP. Designer software for your pathology workflow. We Believe the Possibilities. We Believe the Possibilities. Delphic AP Designer software for your pathology workflow. Sysmex Software Solutions. Shaping the advancement of healthcare. Delphic AP 1 Delphic AP Delphic AP has been specifically

More information

Patient Safety and the Laboratory

Patient Safety and the Laboratory College of American Pathologists Laboratory Accreditation Program Patient Safety and the Laboratory May 21, 2008 Copyright 2008 College of American Pathologists (CAP). All rights are reserved. Participants

More information

Customer Satisfaction in Anatomic Pathology. A College of American Pathologists Q-Probes Study of 3065 Physician Surveys From 94 Laboratories

Customer Satisfaction in Anatomic Pathology. A College of American Pathologists Q-Probes Study of 3065 Physician Surveys From 94 Laboratories Customer Satisfaction in Anatomic Pathology A College of American Pathologists Q-Probes Study of 065 Physician Surveys From 9 Laboratories Richard J. Zarbo, MD, DMD; Raouf E. Nakhleh, MD; Molly Walsh,

More information

We Believe the Possibilities. Delphic AP

We Believe the Possibilities. Delphic AP We Believe the Possibilities. Delphic AP Designer Software for your pathology workflow. Delphic AP Tracks the specimen process cycle Labels cassettes and slides Enables sophisticated report creation Send

More information

LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009

LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009 LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009 Paul Bachner, MD, FCAP Professor & Chair Department of Pathology & Laboratory Medicine University of Kentucky

More information

Center for Clinical Standards and Quality /Survey & Certification Group

Center for Clinical Standards and Quality /Survey & Certification Group DEPARTMENT OF HEALTH & HUMAN SERVICES Centers for Medicare & Medicaid Services 7500 Security Boulevard, Mail Stop C2-21-16 Baltimore, Maryland 21244-1850 Center for Clinical Standards and Quality /Survey

More information

2009 LAP Audioconference Series. How to Prepare and Comply with Your Quality Management Plan

2009 LAP Audioconference Series. How to Prepare and Comply with Your Quality Management Plan 2009 LAP Audioconference Series How to Prepare and Comply with Your Quality Management Objectives: As a result of participating in this session, you will be able to: Explain the reasons why the QM is important

More information

Relational Coupling by 2-D Barcodes A Novel Approach to Positive Patient and Specimen Identification in Tissue and Cytology Specimens

Relational Coupling by 2-D Barcodes A Novel Approach to Positive Patient and Specimen Identification in Tissue and Cytology Specimens Relational Coupling by 2-D Barcodes A Novel Approach to Positive Patient and Specimen Identification in Tissue and Cytology Specimens Lee KC, Wong SN, Leung YW, Lam WL Department of Pathology, Princess

More information

Leica CEREBRO Sample Tracking. Patient Safety. Enhanced Productivity. Complete Flexibility.

Leica CEREBRO Sample Tracking. Patient Safety. Enhanced Productivity. Complete Flexibility. P P P Leica CEREBRO Sample Tracking Patient Safety. Enhanced Productivity. Complete Flexibility. Patient Safety Protect their health, and your reputation. Don't let identification errors lead to patient

More information

Automating Anatomic Pathology

Automating Anatomic Pathology Automating Anatomic Pathology J. Mark Tuthill, MD Division of Pathology Informatics, Henry Ford Hospital Detroit, MI 48202 Pathology Informatics Summit May, 2015 Disclosures In accordance with ACCME guidelines,

More information

Voice recognition technology (VRT) geared for medical

Voice recognition technology (VRT) geared for medical Voice Recognition Technology Implementation in Surgical Pathology Advantages and Limitations Meenakshi Singh, MD; Timothy R. Pal, MD N Context. Voice recognition technology (VRT) has been in use for medical

More information

Wrong Blood In Tube The Tip of the Iceberg

Wrong Blood In Tube The Tip of the Iceberg Wrong Blood In Tube The Tip of the Iceberg Dr Paula Bolton-Maggs SHOT Medical Director 14 th IHS Montréal SHOT Mission Statement (Serious Hazards of Transfusion) To improve patient safety in blood transfusion

More information

Monitoring customer satisfaction is an important and

Monitoring customer satisfaction is an important and CAP Laboratory Improvement Programs Hospital Nursing Satisfaction With Clinical Laboratory Services A College of American Pathologists Q-Probes Study of 162 Institutions Bruce A. Jones, MD; Molly K. Walsh,

More information

Histology Asset Tracking March 3, 2013

Histology Asset Tracking March 3, 2013 Block and Slide Tracking in Anatomic Pathology Professor of Pathology Associate Director, Anatomic Pathology Director, Pathology Informatics Yale University School of Medicine 2012 College of American

More information

The reason for the success of several well-known. Tracking in Anatomic Pathology

The reason for the success of several well-known. Tracking in Anatomic Pathology Tracking in Anatomic Pathology Liron Pantanowitz, MD; Alexander C. Mackinnon Jr, MD, PhD; John H. Sinard, MD, PhD Bar code based tracking solutions, long present in clinical pathology laboratories, have

More information

Identifying and Avoiding Errors in Surgical Pathology

Identifying and Avoiding Errors in Surgical Pathology Identifying and Avoiding Errors in Surgical Pathology Jeffrey L. Myers, M.D. A. James French Professor & Director, Division of Anatomic Pathology The University of Michigan ERRORS IN SURGICAL PATHOLOGY

More information

Laboratory Quality Metrics

Laboratory Quality Metrics Laboratory Quality Metrics Laila O. Abdel-Wareth. MBBCh, FCAP, FRCPC, FRCPath, EMHCA Chief Medical Officer- Pathology & Laboratory Medicine Services SEHA Chair of Pathology & Laboratory Medicine Institute

More information

September 5, 2014. Linda Stewart Vice President, National Lab Program UnitedHealthcare Group P.O. Box 1459 Minneapolis, MN 55440 1459

September 5, 2014. Linda Stewart Vice President, National Lab Program UnitedHealthcare Group P.O. Box 1459 Minneapolis, MN 55440 1459 September 5, 2014 Linda Stewart Vice President, National Lab Program UnitedHealthcare Group P.O. Box 1459 Minneapolis, MN 55440 1459 Re: UnitedHealthcare Laboratory Benefit Management Program Dear Ms.

More information

Lisa Yerian, MD May 12, 2011

Lisa Yerian, MD May 12, 2011 What are the current best practice models and opportunities for driving efficiency (revenue/cost, work/product/time) our care? Lisa Yerian, MD May 12, 2011 Efficiency Extent to which time or effort is

More information

ProPath selects LKConnect

ProPath selects LKConnect ProPath selects LKConnect Case Study LKConnect for Dermatology helps a pathology group retain clients and acquire new business by facilitating physician practice workflows. Overview For more than forty

More information

Frozen Section Diagnosis

Frozen Section Diagnosis Frozen Section Diagnosis Dr Catherine M Corbishley Honorary Consultant Histopathologist St George s Healthcare NHS Trust and lead examiner final FRCPath Practical 2008-2011 Frozen Section Diagnosis The

More information

EARLY ONLINE RELEASE

EARLY ONLINE RELEASE EARLY ONLINE RELEASE Note: This article was posted on the Archives Web site as an Early Online Release. Early Online Release articles have been peer reviewed, copyedited, and reviewed by the authors. Additional

More information

laboratorymedicine> august 2001> number 8> volume 32

laboratorymedicine> august 2001> number 8> volume 32 interpretation [management/administration and training generalist] A Four-Part Approach to Competency Assessment Jennifer Schiffgens, MBA, MT(ASCP), 1 and Valerie Bush, PhD, MT(ASCP) 2 From 1 Clinical

More information

Streamlining the PATH Line: Onboarding Barcoding in Anatomic Pathology. By: Ary Franklin, HT(ASCP) & Carol Beth Taylor, MHS, PA/HTL(ASCP), GBLSS

Streamlining the PATH Line: Onboarding Barcoding in Anatomic Pathology. By: Ary Franklin, HT(ASCP) & Carol Beth Taylor, MHS, PA/HTL(ASCP), GBLSS Streamlining the PATH Line: Onboarding Barcoding in Anatomic Pathology By: Ary Franklin, HT(ASCP) & Carol Beth Taylor, MHS, PA/HTL(ASCP), GBLSS In the beginning Barcoding Who decides? Not us! What barcode

More information

Captivator EMR Device

Captivator EMR Device Device Clinical Article and Abstract Summary Endoscopic Mucosal Bergman et al: EMR Training Tips Bergman et al: EMR Learning Curve ASGE: EMR & ESD Guidelines Bergman et al: Captivator EMR vs Cook Duette

More information

SAMPLE. Accuracy in Patient and Sample Identification; Approved Guideline. This guideline describes the essential elements of systems and processes

SAMPLE. Accuracy in Patient and Sample Identification; Approved Guideline. This guideline describes the essential elements of systems and processes March 2010 Accuracy in Patient and Sample Identification; Approved Guideline This guideline describes the essential elements of systems and processes required to ensure accurate patient identification.

More information

Incident Management: Developing a Plan

Incident Management: Developing a Plan LABFACTS 71 Incident Management: Developing a Plan INTRODUCTION In light of the Institute of Medicine studies focusing on reducing medical errors, COLA criterion #285 has been changed effective May 1,

More information

Case Study Building a Patient Safety-Centered AP Lab

Case Study Building a Patient Safety-Centered AP Lab Case Study Building a Patient Safety-Centered AP Lab Lean processes brought swift and impressive gains in productivity, patient safety and error reduction: Zero mislabeling errors in microtomy post go-live

More information

LabGuide 71. Incident Management: Developing a Plan

LabGuide 71. Incident Management: Developing a Plan LabGuide 71 Incident Management: Developing a Plan INTRODUCTION In light of the Institute of Medicine studies focusing on reducing medical errors, COLA developed Accreditation criterion QA 20 to focus

More information

THE rising concern and attention to patient

THE rising concern and attention to patient J Nurs Care Qual Vol. 26, No. 1, pp. 13 21 Copyright c 2011 Wolters Kluwer Health Lippincott Williams & Wilkins Decreasing Mislabeled Laboratory Specimens Using Barcode Technology and Bedside Printers

More information

Interoperability between Anatomic Pathology Laboratory Information Systems and Digital Pathology Systems

Interoperability between Anatomic Pathology Laboratory Information Systems and Digital Pathology Systems Interoperability between Anatomic Pathology Laboratory Information Systems and Digital Pathology Systems Authors: Jesus Ellin (Yuma Regional Medical Center), Anna Haskvitz (Sunquest Information Systems,

More information

Put all your results in one basket. Orchard Pathology simplifies the complexities of clinical, molecular, and pathology testing and reporting.

Put all your results in one basket. Orchard Pathology simplifies the complexities of clinical, molecular, and pathology testing and reporting. Put all your results in one basket. Orchard Pathology simplifies the complexities of clinical, molecular, and pathology testing and reporting. Designed for the Comprehensive Pathology Laboratory Orchard

More information

RE: Australian Safety and Quality Goals for Health Care: Consultation paper

RE: Australian Safety and Quality Goals for Health Care: Consultation paper 10 February 2012 Mr Bill Lawrence AM Acting CEO Australian Commission on Safety and Quality in Health Care GPO Box 5480 Sydney NSW 2001 Email: goals@safetyandquality.gov.au Dear Mr Lawrence RE: Australian

More information

Liron Pantanowitz, MD

Liron Pantanowitz, MD Whole Slide Imaging: Education, Tumor Boards, & Presentations Liron Pantanowitz, MD Associate Professor of Pathology & Biomedical Informatics, University of Pittsburgh Medical Center, Pittsburgh, PA, USA.

More information

PREDICTIVE ANALYTICS AT THE BEDSIDE: The 5 W s. John W. Cromwell, MD

PREDICTIVE ANALYTICS AT THE BEDSIDE: The 5 W s. John W. Cromwell, MD PREDICTIVE ANALYTICS AT THE BEDSIDE: The 5 W s John W. Cromwell, MD Disclosures Entac Medical, Inc. - Chief Medical Officer Apervita, Inc. - Consultant Dell, Inc. - Consultant Cromwell Lab Analytics 2009

More information

HER2 Status: What is the Difference Between Breast and Gastric Cancer?

HER2 Status: What is the Difference Between Breast and Gastric Cancer? Ask the Experts HER2 Status: What is the Difference Between Breast and Gastric Cancer? Bharat Jasani MBChB, PhD, FRCPath Marco Novelli MBChB, PhD, FRCPath Josef Rüschoff, MD Robert Y. Osamura, MD, FIAC

More information

Our Vision To be the Western Colorado and Eastern Utah laboratory services provider of choice.

Our Vision To be the Western Colorado and Eastern Utah laboratory services provider of choice. Community Hospital provides laboratory services focused on patient-guest care. We are committed to providing accurate, timely, cost-effective, laboratory testing for your patients. Our knowledge and expertise

More information

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical

More information

How to report Upper GI EMR/ESD specimens

How to report Upper GI EMR/ESD specimens Section of Pathology and Tumour Biology How to report Upper GI EMR/ESD specimens Dr.H.Grabsch Warning. Most of the criteria, methodologies, evidence presented in this talk are based on studies in early

More information

Section 7: Program Objectives and Supervision of Residents at USC University Hospital

Section 7: Program Objectives and Supervision of Residents at USC University Hospital Section 7: Program Objectives and Supervision of Residents at USC University Hospital Program Objectives, Goals, and Supervision of Residents at USC University Hospital Page 7-1 SECTION 7: PROGRAM OBJECTIVES

More information

VENTANA Digital Pathology. Reliable. Efficient. Comprehensive.

VENTANA Digital Pathology. Reliable. Efficient. Comprehensive. VENTANA Digital Pathology Reliable. Efficient. Comprehensive. In the quest to improve the lives of all patients afflicted with cancer, every second counts. Fast, informed clinical decisions are essential

More information

Copyright 2014, AORN, Inc. Page 1 of 5

Copyright 2014, AORN, Inc. Page 1 of 5 AORN Position Statement on One Perioperative Registered Nurse Circulator Dedicated to Every Patient Undergoing an Operative or Other Invasive Procedure POSITION STATEMENT The goal of perioperative nursing

More information

Quality Management Tools

Quality Management Tools Quality Management Tools Through our programs, laboratories are able to compare their performance against other laboratories and continuously demonstrate the value they add to their institutions. Raouf

More information

KIDNEY FUNCTION RELATION TO SIZE OF THE TUMOR IN RENAL CELL CANCINOMA

KIDNEY FUNCTION RELATION TO SIZE OF THE TUMOR IN RENAL CELL CANCINOMA KIDNEY FUNCTION RELATION TO SIZE OF THE TUMOR IN RENAL CELL CANCINOMA O.E. Stakhvoskyi, E.O. Stakhovsky, Y.V. Vitruk, O.A. Voylenko, P.S. Vukalovich, V.A. Kotov, O.M. Gavriluk National Canсer Institute,

More information

How To Use Praxis

How To Use Praxis Praxis EMR Case Studies For more client case studies in a variety of medical specialties, go to http://www.praxisemr.com/testimonials.htm Our most recent and extensive case study covers Jeremy Bradley,

More information

May 8, 2013. The Honorable Fred Upton Chairman House Committee on Energy and Commerce United States House of Representatives Washington, DC 20151

May 8, 2013. The Honorable Fred Upton Chairman House Committee on Energy and Commerce United States House of Representatives Washington, DC 20151 May 8, 2013 The Honorable Dave Camp Chairman House Committee on Ways and Means United States House of Representatives Washington D.C. 20515 The Honorable Fred Upton Chairman House Committee on Energy and

More information

ATOMS A Laboratory Specimen Collection and Management System

ATOMS A Laboratory Specimen Collection and Management System ATOMS A Laboratory Specimen Collection and Management System Cicada Cube Pte Ltd 20 Ayer Rajah Crescent, #09-26 Technopreneur Centre, Singapore 139664 Tel: 65-67787833; Fax: 65-67797229 Email: info@cicadacube.com

More information

Reducing Errors in Blood Specimen Labeling: A Multihospital Initiative

Reducing Errors in Blood Specimen Labeling: A Multihospital Initiative Reducing Errors in Blood Specimen Labeling: A Multihospital Initiative ABSTRACT Patient blood specimen identification is critical for quality patient care. Misidentified specimens can result in delayed

More information

COMMISSION ON LABORATORY ACCREDITATION. Laboratory Accreditation Program LABORATORY GENERAL CHECKLIST

COMMISSION ON LABORATORY ACCREDITATION. Laboratory Accreditation Program LABORATORY GENERAL CHECKLIST Revised: 09/27/2007 COMMISSION ON LABORATORY ACCREDITATION Laboratory Accreditation Program LABORATORY GENERAL CHECKLIST Disclaimer and Copyright Notice The College of American Pathologists (CAP) Checklists

More information

chapter 5. Quality control at the population-based cancer registry

chapter 5. Quality control at the population-based cancer registry chapter 5. Quality control at the population-based cancer registry All cancer registries should be able to give some objective indication of the quality of the data that they have collected. The methods

More information

Affiliated Dermatology : A Practice with a Technology Focus and Family at Heart

Affiliated Dermatology : A Practice with a Technology Focus and Family at Heart Affiliated Dermatology : A Practice with a Technology Focus and Family at Heart left to right: Suzanne Murry, Operations Manager, Stacie Schimke, Chief Histotechnologist, Dr. Richard L. Averitte, Jr.,

More information

CME/SAM. Laboratory Medicine Quality Indicators. A Review of the Literature. Shahram Shahangian, PhD, MS, and Susan R. Snyder, PhD, MBA.

CME/SAM. Laboratory Medicine Quality Indicators. A Review of the Literature. Shahram Shahangian, PhD, MS, and Susan R. Snyder, PhD, MBA. Clinical Chemistry / Laboratory Medicine Quality Indicators Laboratory Medicine Quality Indicators A Review of the Literature Shahram Shahangian, PhD, MS, and Susan R. Snyder, PhD, MBA Key Words: Laboratory

More information

Framework for action. Right patient right care

Framework for action. Right patient right care Framework for action Contents Summary 1 Background 2 Role of the NPSA 3 The NPSA project 5 Manual checking 6 Technologies 8 The interdependence of technology and human factors 10 The National Programme

More information

Provincial Quality Management Programs for Mammography, Colonoscopy and Pathology in Ontario

Provincial Quality Management Programs for Mammography, Colonoscopy and Pathology in Ontario Provincial Quality Management Programs for Mammography, Colonoscopy and Pathology in Ontario Quality Management Partnership Consultation Materials: Heal th System Ad mi nistrato r s October 20, 2014 Table

More information

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR CliniCAl laboratories 2nd Edition Effective 1 April 2010 International Patient Safety Goals (IPSG) Goals The following is a list of all goals.

More information

General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations

General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations General Information Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations Our Background Foundation Laboratory is an advanced, accredited and certified clinical diagnostic

More information

Online Supplement to Clinical Peer Review Programs Impact on Quality and Safety in U.S. Hospitals, by Marc T. Edwards, MD

Online Supplement to Clinical Peer Review Programs Impact on Quality and Safety in U.S. Hospitals, by Marc T. Edwards, MD Online Supplement to Clinical Peer Review Programs Impact on Quality and Safety in U.S. Hospitals, by Marc T. Edwards, MD Journal of Healthcare Management 58(5), September/October 2013 Tabulated Survey

More information

Quality and peer review The Canadian perspective

Quality and peer review The Canadian perspective Disclosure Quality and peer review The Canadian perspective Co-founder and shareholder in Real Time Medical Inc. MIR Barcelona October 11, 2013 David A Koff MD FRCPC Chief and Chair Radiology at McMaster

More information

If you are signing for a minor child, you refers to your child throughout the consent document.

If you are signing for a minor child, you refers to your child throughout the consent document. CONSENT TO PARTICIPATE IN A CLINICAL RESEARCH STUDY Adult Patient or Parent, for Minor Patient INSTITUTE: National Cancer Institute PRINCIPAL INVESTIGATOR: Raffit Hassan, M.D. STUDY TITLE: Tissue Procurement

More information

Medical Errors for the Laboratorian 2014 Update

Medical Errors for the Laboratorian 2014 Update Medical Errors for the Laboratorian 2014 Update This course meets the 2-hours Medical Errors continuing education requirement for healthcare providers in the state of Florida. Objectives: Explain the document

More information

DATE APPROVED: DATE EFFECTIVE: Date of Approval. REFERENCE NO. MOH/04 PAGE: 1 of 7

DATE APPROVED: DATE EFFECTIVE: Date of Approval. REFERENCE NO. MOH/04 PAGE: 1 of 7 SOURCE: Ministry of Health DATE APPROVED: DATE EFFECTIVE: Date of Approval REPLACESPOLICY DATED: 1 POLICY TITLE: Incident/Accident Reporting REFERENCE NO. MOH/04 PAGE: 1 of 7 REVISION DATE(s): Ministry

More information

Improving Cardiac Surgery Patient Flow through Computer Simulation Modeling

Improving Cardiac Surgery Patient Flow through Computer Simulation Modeling Improving Cardiac Surgery Patient Flow through Computer Simulation Modeling Dana Khayal, Fatma Almadhoun, Lama Al-Sarraj and Farayi Musharavati Abstract In this paper, computer simulation modeling was

More information

Materials and Methods

Materials and Methods Anatomic Pathology / SYNOPTIC TOOL FOR HEAD AND NECK CANCERS Usefulness of a Synoptic Data Tool for Reporting of Head and Neck Neoplasms Based on the College of American Pathologists Cancer Checklists

More information

Epi procolon The Blood Test for Colorectal Cancer Screening

Epi procolon The Blood Test for Colorectal Cancer Screening Epi procolon The Blood Test for Colorectal Cancer Screening Epi procolon is an approved blood test for colorectal cancer screening. The US Preventive Services Task Force, the American Cancer Society and

More information

Biovigilance Component Hemovigilance Module Incident Reporting

Biovigilance Component Hemovigilance Module Incident Reporting Biovigilance Component Hemovigilance Module Incident Reporting National Center for Emerging and Zoonotic Infectious Diseases Division of Healthcare Quality Promotion 1 Objectives Review key terms used

More information

Why Does Certification Matter?

Why Does Certification Matter? Why Does Certification Matter? Jones & Bartlett Learning, CHAPTER LLC 1 OBJECTIVES/TOPICS TO COVER: Describe the differences between certification, registration, and licensure. List the benefits of pharmacy

More information

What is a definition of risk?

What is a definition of risk? What is a definition of risk? Definition of Risk Risk is the probability or threat of any negative occurrence caused by internal or external vulnerabilities interfering with achieving objectives that may

More information

Below are some of the ways pathologists contribute to the six domains (listed in the order they appear on page 2 of the Strategy):

Below are some of the ways pathologists contribute to the six domains (listed in the order they appear on page 2 of the Strategy): January 10, 2014 Center for Medicare and Medicaid Services U.S. Department of Health and Human Services 7500 Security Boulevard Baltimore, Maryland 21244-18559 Re: CMS Quality Strategy Submitted electronically

More information

Strategies for LEADERSHIP. Hospital Executives and Their Role in Patient Safety

Strategies for LEADERSHIP. Hospital Executives and Their Role in Patient Safety Strategies for LEADERSHIP Hospital Executives and Their Role in Patient Safety 1 Effective Leadership for Patient Safety Creating and Leading Significant Change Dear Colleague: In 1995, two tragic medication

More information

FULLY AUTOMATED PARALLEL DUAL STAINING DETECTION SYSTEM. ChromoPlex 1 Dual Detection for BOND

FULLY AUTOMATED PARALLEL DUAL STAINING DETECTION SYSTEM. ChromoPlex 1 Dual Detection for BOND ChromoPlex 1 Dual Detection for BOND FULLY AUTOMATED PARALLEL DUAL STAINING DETECTION SYSTEM View multiple antibodies on a single slide to deliver a comprehensive clinical result. 1 MULTIPLY YOUR CAPABILITIES

More information

BREAST CANCER PATHOLOGY

BREAST CANCER PATHOLOGY BREAST CANCER PATHOLOGY FACT SHEET Version 4, Aug 2013 This fact sheet was produced by Breast Cancer Network Australia with input from The Royal College of Pathologists of Australasia I m a nurse and know

More information

POLICY TITLE PATIENT SAFETY EVENT MANAGEMENT

POLICY TITLE PATIENT SAFETY EVENT MANAGEMENT Page 1 of 7 INTRODUCTION Fraser Health is committed to providing quality care to its patients 1, as described in its Vision, Mission and Values, recognizing that safe health care is its first priority.

More information

Healthcare Market Overview

Healthcare Market Overview Healthcare Market Overview INTRODUCTION For healthcare organizations that require instantaneous availability of accurate information, mtuitive extends the value of legacy investments with capture and delivery

More information

Explanation of your PAP smear

Explanation of your PAP smear Explanation of your PAP smear Approximately 5-10% of PAP smears in the United States are judged to be abnormal. Too often, the woman who receives this news worries that she already has, or will develop,

More information

Continuing Education Opportunities

Continuing Education Opportunities www.cap.org Continuing Education Opportunities Surveys Education Programs When your lab participates in Surveys, every member of your team can enroll in education activities and earn CME/CE at no additional

More information

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests 2010 Laboratory Accreditation Program Audioconference Accreditation Requirements for Waived Vs. Non-waived Tests Objectives: After participating in this audioconference you will be able to: Define the

More information

Disclosing Medical Errors: 2008 and Beyond. Wendy Levinson, MD Professor of Medicine University of Toronto

Disclosing Medical Errors: 2008 and Beyond. Wendy Levinson, MD Professor of Medicine University of Toronto Disclosing Medical Errors: 2008 and Beyond Wendy Levinson, MD Professor of Medicine University of Toronto Goals 1. To describe physicians attitudes and experience re disclosure. 2. To describe changing

More information

Travel Distance to Healthcare Centers is Associated with Advanced Colon Cancer at Presentation

Travel Distance to Healthcare Centers is Associated with Advanced Colon Cancer at Presentation Travel Distance to Healthcare Centers is Associated with Advanced Colon Cancer at Presentation Yan Xing, MD, PhD, Ryaz B. Chagpar, MD, MS, Y Nancy You MD, MHSc, Yi Ju Chiang, MSPH, Barry W. Feig, MD, George

More information

Best Practice Partnership Program

Best Practice Partnership Program Best Practice Partnership Program Empowering Community- Based Pathologists Keeping Medicine Local Losing outreach to centralized reference laboratories? Wondering how molecular testing & digital pathology

More information

DOCUMENT CONTROL PAGE

DOCUMENT CONTROL PAGE Review Circulation Application Ratification Author Minor Amendment Supersedes Title DOCUMENT CONTROL PAGE Title: Directorate of Laboratory Medicine Sample Acceptance Policy Version: 009 Reference Number:

More information

Subject: Clinical Laboratory Personnel Standards: Draft Proposed Regulations:

Subject: Clinical Laboratory Personnel Standards: Draft Proposed Regulations: Ronald Chapman, MD, MPH Director, California Department of Public Health and State Health Officer California Department of Public Health Post Office Box 997377, MS-0507 Sacramento, CA 97377 Subject: Clinical

More information

Preparation "Clinical Laboratory Technologist and Technician Overview"

Preparation Clinical Laboratory Technologist and Technician Overview Clinical Laboratory Technologist and Technician Overview The Field - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Clinical laboratory

More information

Failure Modes and Effects Analysis (FMEA)

Failure Modes and Effects Analysis (FMEA) Failure Modes and Effects Analysis (FMEA) Sasa Mutic Washington University School of Medicine St. Louis Missouri Failure Modes and Effects Analysis Objectives: To motivate the use of FMEA and to provide

More information

Errors in Pathology and Laboratory Medicine: Consequences and Prevention

Errors in Pathology and Laboratory Medicine: Consequences and Prevention Journal of Surgical Oncology 2004;88:161 181 Errors in Pathology and Laboratory Medicine: Consequences and Prevention SANDRA C. HOLLENSEAD, MD, WILLIAM B. LOCKWOOD, PhD, MD, AND RONALD J. ELIN, MD, PhD,*

More information

EARLY ONLINE RELEASE

EARLY ONLINE RELEASE EARLY ONLINE RELEASE Note: This article was posted on the Archives Web site as an Early Online Release. Early Online Release articles have been peer reviewed, copyedited, and reviewed by the authors. Additional

More information

South Carolina Society of Health-System Pharmacists Position Statement on Pharmacy Technicians

South Carolina Society of Health-System Pharmacists Position Statement on Pharmacy Technicians South Carolina Society of Health-System Pharmacists Position Statement on Pharmacy Technicians The safety and health of the citizens of South Carolina are vital concerns for all pharmacists. Without appropriate

More information

The Ontario Cancer Registry moves to the 21 st Century

The Ontario Cancer Registry moves to the 21 st Century The Ontario Cancer Registry moves to the 21 st Century Rebuilding the OCR Public Health Ontario Grand Rounds Oct. 14, 2014 Diane Nishri, MSc Mary Jane King, MPH, CTR Outline 1. What is the Ontario Cancer

More information

Clinical Decision Support (CDS) to improve colorectal cancer screening

Clinical Decision Support (CDS) to improve colorectal cancer screening Clinical Decision Support (CDS) to improve colorectal cancer screening NIH Collaboratory Grand Rounds Sept 26, 2014 Presented by: Tim Burdick MD MSc OCHIN Chief Medical Informatics Officer Adjunct Associate

More information

E.H.R. s and Improving Patient Safety - What Has Been the Real Impact?

E.H.R. s and Improving Patient Safety - What Has Been the Real Impact? E.H.R. s and Improving Patient Safety - What Has Been the Real Impact? Presented by: Mary Erickson, RN, HTS Accounting Manager HTS, a division of Mountain Pacific Quality Health Foundation 1 Understand

More information

Lean Strategies Used to Optimize Automation. Why Lean Six Sigma? Laboratory Goals. Decreased TAT. Accurate Results. LEAN Goals.

Lean Strategies Used to Optimize Automation. Why Lean Six Sigma? Laboratory Goals. Decreased TAT. Accurate Results. LEAN Goals. Lean Strategies Used to Optimize Automation Linda Stubbs, ASQ CSSBB Senior Workflow Specialist Why LEAN Six Sigma? Why Lean Six Sigma? Laboratory Goals Decreased TAT Accurate Results LEAN Goals Increase

More information

Electronic Document Management J. Mark Tuthill, MD, Henry Ford Health System

Electronic Document Management J. Mark Tuthill, MD, Henry Ford Health System Electronic Document Management J. Mark Tuthill, MD, Henry Ford Health System Electronic document management is important for the clinical laboratory in many ways. In this session we will review the definition

More information

Examples: collecting the right sample on the right patient; labeling the sample accurately;

Examples: collecting the right sample on the right patient; labeling the sample accurately; Quality Assessment: Basics & EMR is called by several names: and are just two of them. Quality Assessment (QA) is an on-going program designed to monitor and evaluate the quality or accuracy of all laboratory

More information