Masimo Adaptive Threshold Alarm : Intelligent and Personalised Notification to Reduce Nuisance Alarms



Similar documents
Optimizing Masimo SET SpO2 Alarm Settings on Select GE Monitors

Advanced Alarm Performance. Reduce false alarms and nuisance alarms without delaying actionable alarms

Alarm Management on the Medical Surgical Floor

Patient SafetyNet System

Signal Extraction Technology

Sue Carol Verrillo, RN, MSN, CRRN The Johns Hopkins Hospital November 14, 2014

The Growing Concern Surrounding Medical Alarm Fatigue

The Growing Concern Surrounding Medical Alarm Fatigue

Sound the Alarm: Alarm System Management for Patient Safety

Bringing Order Out of Chaos

Policy Name: Patient Monitoring via the Patient SafetyNet (PSN) Monitoring System

Nellcor Pulse Oximeters with OxiMax Technology from Covidien

Medical Device Alerting: Improving Response & Reducing the Noise

Capnostream 20p Bedside Capnography Monitor. The only integrated, complete picture of oxygenation and ventilation 1

Pulse Oximetry Alarm Reduction Strategy at Boston Medical Center

CARESCAPE VC150 Vital Signs Monitor. Connecting intelligence and care.

Masimo Patient Safetynet HL7 Interface Specifications

GE Healthcare. B40 Patient Monitor Connecting intelligence and care.

High-intensity performance. IntelliVue MP90 patient monitor

University of Michigan Hospital

Alarm Management. Scope and Impact of the Problem. Expected Practice and Nursing Actions*

Evolution of an Integrated System for Alarm and Call Management March 19, 2014

Chapter 4. Regulation of Alarm Systems

5162 E. Working in oxygen-reduced atmospheres BGI/GUV-I 5162 E. Information

Networked versatility. IntelliVue MP60 and MP70 patient monitors

COMPACT GUIDE. Camera-Integrated Motion Analysis

CHAPTER SECURITY ALARM SYSTEMS

Critical Care Therapy and Respiratory Care Section

The Periodic Moving Average Filter for Removing Motion Artifacts from PPG Signals

CA 1050 Owner s Guide

Integrating Defensive Monitoring in the General Care Unit to Improve Failure to Rescue Trends

SHARE THIS WHITEPAPER. Top Selection Criteria for an Anti-DDoS Solution Whitepaper

Solution Title: Predicting Care Using Informatics/MEWS (Modified Early Warning System)

Diagnostics and Troubleshooting Using Event Policies and Actions

Improving. Summary. gathered from. research, and. Burnout of. Whitepaper

Safety Innovations FOUNDATIONHTSI

Connecting Remote Cardiac Monitoring Issues with Care Areas

Boise Municipal Code. Chapter 6-15 FALSE SECURITY ALARMS

Process Solutions. DynAMo Alarm & Operations Management. Solution Note

Michael R. Pinsky, M.D., C.M., Dr.h.c., FCCP, MCCM Professor of Critical Care Medicine, Bioengineering, Anesthesiology, Cardiovascular Diseases, and

UPSentry Smart User s Manual. Shutdown Management Software. for Mac OS X 10.2

ITIL A guide to event management

Delivering operations integrity through better plant safety, availability and compliance across your entire enterprise

Milestone Edge Storage with flexible retrieval

X Series Extensive Capabilities

How to Use Motion Detection in ACTi Cameras

FLORIDA DEPARTMENT OF HEALTH BOARD OF DENTISTRY

Chapter 9 N.C. C. N.O. Single-Pole Double-Throw

Destiny Destiny Owners Manual

ETCO2 Monitoring: Riding the Wave! Disclosure 4/11/2013

Sus tain.ability Honeywell Users Group Asia Pacific. Annemarie Diepenbroek, Greg Belcher Experion Overview and New Solutions

ITIL A guide to Event Management

Analysis of an EDI Transaction

Alarm Management What, Why, Who and How?

PANIC, The ALBA Alarm System

Software-based medical devices from defibrillators

How To Manage Risk

MONITOR ISM / AFx Multi-Tenant Security System User Guide V1.3

Alarm Management. July 2012 / White paper. Make the most of your energy SM

March 2015 MEGATRENDS IN THE OIL AND GAS INDUSTRY

ARTICLE 735. Emergency Alarm System

The purpose of Capacity and Availability Management (CAM) is to plan and monitor the effective provision of resources to support service requirements.

Assessing Scheduled Support of Medical Equipment

Spirodoc. 3D Laboratory for respiratory analysis two functional modes: doctor and patient. Four devices in one. Spirometer with Touch Screen display

FireScope + ServiceNow: CMDB Integration Use Cases

TECHNICAL SERVICE BULLETIN

Kendall DL Cable and Lead Wire System. A Disposable Cardiac Monitoring Solution Designed for Superior Clinical Performance. Safe, Simple, Secure

Improved Outcomes and Reduced Costs with Contact-free Continuous Patient Monitoring on a Medical-Surgical Hospital Unit

Monitoring Windows Event Logs

Guaranteeing your equipment is future-proof

GUIDELINES ON MEDICAL DEVICES

CH CONSCIOUS SEDATION

Intelligent Monitoring Software. IMZ-RS400 Series IMZ-RS401 IMZ-RS404 IMZ-RS409 IMZ-RS416 IMZ-RS432

Clinical Alarms Improvement Initiative

Features. Emerson Solutions for Abnormal Situations

Find what matters. Information Alchemy Turning Your Building Data Into Money

Everybody has the right to

Trend Micro Hosted Security Stop Spam. Save Time.

Focus on Safety 08: Incident Notification and Prevention Aug Creating a Safe and Sustainable Environment for Science

ICS for LAUSD EOC and DOC Operation

Welch Allyn Connex Vital Signs Monitor 6000 Series and Accessory Cable Management Stand. Advanced accessory management that stands on its own.

Connect care for early intervention

GE Healthcare. Dash 3000, 4000 & High-acuity mobile patient monitoring

Remote Guarding. The traditional guarding functions that you depend on can now be performed remotely.

Guide to manual tasks high impact function (HIF) audit 2010 March 2010

MONITOR ALARM FATIGUE: STANDARDIZING USE OF PHYSIOLOGICAL MONITORS AND DECREASING NUISANCE ALARMS. Critical Care Management. 1.

Transcription:

Masimo Adaptive Threshold Alarm : Intelligent and Personalised Notification to Reduce Nuisance Alarms Summary Nuisance alarms and false alarms can desensitise clinicians, causing actionable alarms to be ignored Masimo SET Measure-through Motion and Low Perfusion pulse oximetry is proven to reduce false alarms during challenging conditions Many true alarms represent brief desaturations that do not require clinical intervention, which can desensitise clinicians to actionable events Actionable alarms are characterised by significant and sustained SpO2 drops from the patient baseline or SpO2 below a severe threshold value Traditional alarm configurations rely on fixed thresholds and notification delays and are based on a one-size-fits-all approach Reducing fixed thresholds and increasing notification delays decreases alarm frequency but may cause clinically significant events to be missed Masimo Adaptive Threshold Alarm automatically determines each patient s baseline saturation and audibly alarms when saturation significantly changes or when fixed alarm thresholds have been crossed With Masimo Adaptive Threshold Alarm, nuisance alarms can be reduced, freeing clinicians to focus on patients who need care Background Pulse oximetry alarms can be placed into two primary categories: 1) True alarms alarms that occur due to correct SpO2 or pulse rate values 2) False alarms alarms that occur due to incorrect SpO2 or pulse rate values True alarms can be placed into two sub-categories: Actionable alarms True alarms which require clinician notification and potential intervention; Nuisance alarms True alarms that the clinician deems not actionable 1 ECRI Institute has identified nuisance alarms as one of the top technology hazards in hospitals today. 2 While responding to actionable alarms is critical to prevent patient injury or death, the frequency of false and nuisance alarms can increase workload and desensitise clinicians to all alarms, putting patients at significant risk. Multiple studies have established the performance of Masimo SET Measure-through Motion and Low Perfusion pulse oximetry with true alarm sensitivity of 95% and false alarm reduction of 97% 3 during challenging conditions. Masimo SET provides a solid foundation upon which to separate true alarms from false alarms, and therefore allows alarm management strategies to focus on notification of potentially actionable alarms. While Masimo believes that all true alarms are potentially actionable, we also believe in empowering clinicians to create alarm settings that reduce the frequency of nuisance alarms, depending on the care area, patient population, and individual patient.

Adpative threshold alarm Limitations of Fixed Alarm Thresholds Since the inception of pulse oximetry, default pulse oximetry alarm thresholds have been based on two key assumptions: 1) A patient's baseline oxygen saturation value is 98 to 100% 2) When the oxygen saturation crosses 90%, it is clinically significant These assumptions are applied through the concepts of a fixed low SpO2 alarm threshold and a fixed delay in alarm notification. However, this approach creates challenges in the care of diverse patient populations. For example, one patient dropping from SpO2 of 93% to 89% triggers the same alarm as another patient dropping from 99% to 89%. This one-size-fits-all approach leads to some true alarms that a clinician deems non-actionable. This can cause clinicians to desensitised to true alarms that are in fact actionable. Nuisance alarms can be reduced by lowering alarm thresholds and increasing alarm delays. 4 This can significantly reduce alarm frequency but does not fully address the limitations of the core assumptions with fixed alarm thresholds and alarm delays. Neither of these strategies adjusts alarm activation based on the individual patient s condition. In addition, lowering fixed thresholds too far may cause clinicians to miss relevant events available with highly accurate SpO2 measurements with Masimo SET. A recent study in Anesthesiology conducted at a large academic medical center demonstrated a 48% reduction in ICU transfers from a post-surgical general care floor using Masimo SET pulse oximetry. 5 Nuisance alarms were substantially reduced by setting surveillance alarm conditions lower than traditional threshold alarms (80%) and setting a delay (15 sec) to the onset of the audible alarm at the bedside. Although significant improvements in patient outcomes were realised, this strategy reduced nuisance alarms at the expense of sensitivity to prolonged desaturation events above the surveillance settings and potentially actionable alarms. Peer editorial comments on this study suggested that broad adoption of this surveillance strategy may be limited by the unconventional low SpO2 alarm setting. 6 Masimo Adaptive Threshold Alarm Objectives and Assumptions An optimal alarm strategy would reduce nuisance alarms while still identifying clinically relevant events. Masimo Adaptive Threshold Alarm was designed with this intent and the ability to allow the clinician to determine who defines a "nuisance" alarm. This strategy is based on different core assumptions than traditional fixed alarm threshold configurations: 1) Each patient has their own baseline normal saturation 2) Changes from baseline are the primary indicator of deterioration 3) Severely low or fast desaturations still require immediate notification Description Masimo s Adaptive Threshold Alarm is an intelligent and automatically customised alarm solution that significantly reduces the incidence of nuisance alarms while preserving actionable alarms. The Adaptive Threshold Alarm is designed to learn the patient s baseline value during the first 1.5-2 minutes of continuous monitoring and thereafter updates the baseline value based on the recent history of the parameter values. An Adaptive Threshold Limit is continuously determined for that patient based on the low SpO2 threshold alarm setting, the Rapid Desat setting, and the current baseline value. SpO2 measurement values that fall below the Adaptive Threshold Limit will trigger an audible alarm following the user-configured alarm delay.

The Adaptive Threshold Alarm allows shallow desaturations to persist for a longer period of time than deep desaturations but will immediately alarm under either of two conditions: 1) A rapid desaturation (user configured to be 5 or 10%) that exceeds configured rapid desaturation criteria (rate of decline) and/or; 2) A prolonged shallow desaturation event. Adaptive Threshold Alarm will also monitor desaturation recovery and adjust the audible alarm accordingly. This feature further reduces nuisance alarms by allowing the patient s saturation to self-recover to a safe setting without activating an audible alarm. Use of the Adaptive Threshold Alarm is optional and can be switched OFF, in which case traditional alarms limits and delays are used. 100 98 96 94 92 SpO2 % 90 88 86 84 82 80 100 150 200 250 300 Baseline SpO2 Time in Minutes Fixed Threshold Alarm Event Adaptive Threshold Alarm Event Adaptive Threshold Fixed Threshold Figure 2. The behavior of the Adaptive Threshold Alarm with the alarm delay set to 10 seconds is compared to conventional alarm settings for alarm frequency.

Impact on Alarm Frequency The alarm frequency of Adaptive Threshold Alarms was compared against fixed alarm thresholds of 90%, 88%, and 85% with alarm delays of 0 and 15 seconds, with and without Rapid Desat setting of 10%. A database of Masimo SET SpO2 values collected from post-surgical general care areas in 10 Hospitals and over 32 million data points was used. First, alarm occurrence was evaluated, defined as the occurrence of an audible alarm of any duration. Table 1 shows a comparison of three alarm configurations: low threshold alarm with a 15 second delay, low threshold alarm with 15 second delay and a Rapid Desat setting of 10%, and Adaptive Threshold Alarm with a 15 second delay and a Rapid Desat setting of 10%. The greatest improvement in reduction of alarm frequency is when the Adaptive Threshold Alarm is activated at the low SpO2 threshold setting of 90%. Table 1. Adaptive Threshold Alarm: Reduction in Alarm Occurrence Low SpO2 Alarm Setting (with No Alarm Delay, Rapid Desat off) Rapid Desat off Reduction in Alarm Occurrence Rapid Desat at 10% Adaptive Threshold Alarm ( Rapid Desat at 10%) 90% 70% 68% 86% 88% 85% 83% 92% 85% 94% 94% 96% Alarm duration is defined as the total time in which an audible alarm would sound, if the clinician did not silence the alarm. An additional benefit to Adaptive Threshold Alarms is the duration in audio annunciation at the bedside. Table 2 shows a comparison of the reduction of audio alarm duration after activation for the same alarm configurations as in Table 1. Table 2. Adaptive Threshold Alarm Performance: Reduction in Alarm Duration Low SpO2 Alarm Setting (with No Alarm Delay, Rapid Desat off) Rapid Desat off Reduction in Alarm Duration Rapid Desat at 10% Adaptive Threshold Alarm ( Rapid Desat at 10%) 90% 41% 40% 86% 88% 77% 76% 93% 85% 94% 94% 97% The reductions in both the occurrence of alarms and the duration of audible alarms at bedside provide a potential improvement in the patient s environment by reducing the number of nuisance alarm events.

The reduction in alarm occurrence and duration with Adaptive Threshold Alarm is greatest when compared to traditional SpO2 alarm setting of 90%. The reduced impact of Adaptive Threshold Alarm at lower settings may occur because fewer alarms occur as the fixed threshold is moved lower. Researchers have demonstrated that it is possible to reduce low SpO2 alarm limits to 80% with a 15 second alarm delay to primarily detect extreme hypoxia events. 5 These alarm settings resulted in a 98% reduction in overall alarm occurrences in the Masimo Database, but may result in missed alarms for prolonged desaturations above 80%. In comparison, applying Adaptive Threshold Alarm to the same data set with a low SpO2 threshold setting of 88% achieved a 92% reduction in alarm frequency. In this comparison, the additional alarms also allow increased sensitivity to potential actionable events for SpO2 levels between 80% and 88%. This approach to alarm functionality addresses concerns raised over the 80% low SpO2 settings while still reducing the large majority of nuisance alarms. Conclusion Masimo SET Measure-through Motion and Low Perfusion technology provides an important foundation to any alarm management strategy because it significantly reduces false alarms and increases true alarm detection. With Masimo Adaptive Threshold Alarm, true alarms can be separated by the severity of desaturation event, based on the patient s baseline value. Masimo Adaptive Threshold Alarm is a patient-centric, smart alarm that addresses the reduction of nuisance alarms without sacrificing sensitivity to sustained shallow desaturation events that can be missed by just lowering alarm thresholds. Unlike other alarm filtering attempts that either suspended alarms in the presence of motion artifact or integrate desaturation events with time, Adaptive Threshold Alarm adjusts to patient baseline values. References 1 IEC 60601-1-08 General requirements for basic safety and essential performance Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems. 2006. 2 ECRI Institute. "Top 10 Technology Hazards for 2011." Health Devices, Vol 39, No 11;386-398. 3 Shah N et al. Journal of Clinical Anesthesia. 2012. In press. 4 Whitepaper: "Masimo Advanced Alarm Performance: An Evidence-Based Approach to Reduce False Alarms and Nuisance Alarms." Masimo Corp. 2010. 5 Taenzer A et al. "Impact of Pulse Oximetry Surveillance on Rescue Events and Intensive Care Unit Transfers A Before-and-After Concurrence Study." Anesthesiology, 2010; 112:282 7. 6 Abenstein P et al. "An Ounce of Prevention May Equate to a Pound of Cure Can Early Detection and Intervention Prevent Adverse Events?" Anesthesiology, 2010; 112: 272-3.

2011 Masimo Corporation. All rights reserved. Masimo U.S. Tel: 1 877 4 Masimo info-america@masimo.com Masimo International Tel: +41 32 720 1111 info-international@masimo.com 6230-6255C-1111