EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy



Similar documents
EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Determination of the Mass Percentage of Copper in a Penny. Introduction

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

Solubility Product Constants

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

EXPERIMENT 7. Identifying a Substance by Acid-Base Titration

Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

COMMON LABORATORY APPARATUS

HS 1003 Part 2 HS 1003 Heavy Metals Test

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

GA/7 Potentiometric Titration

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Measurement and Calibration

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

The Determination of an Equilibrium Constant

Chem 131A: Absorbance of Riboflavin

Solutions and Dilutions

Calibration of Glassware

Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston

Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston

Determination of Aspirin using Back Titration

SYNTHESIS AND ANALYSIS OF A COORDINATION COMPOUND OF COPPER

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

Analysis of Vitamin C Using Iodine. Introduction

CHEM 161: Beer s Law and Analysis of a Sports Drink

EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

FAJANS DETERMINATION OF CHLORIDE

Experiment 7: Titration of an Antacid

Acid Base Titrations

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

FerroVer Method 1 Method to 3.0, 1.0 to 30.0 and 10.0 to mg/l Fe Powder Pillows

Calcium Analysis by EDTA Titration

NITRIC OXIDE and NITROGEN DIOXIDE 6014

Volumetric Analysis. Lecture 5 Experiment 9 in Beran page 109 Prelab = Page 115

Phosphorus Analyzer Procedure Last updated: A. Introduction

Standard Test Method for Analysis of Barium, Calcium, Magnesium, and Zinc in Unused Lubricating Oils by Atomic Absorption Spectrometry 1

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

POLYDIMETHYLSILOXANE

Experiment 6 Coffee-cup Calorimetry

Coordination Compounds with Copper (II) Prelab (Week 2)

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Flame Tests & Electron Configuration

PREPARATION AND PROPERTIES OF A SOAP

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

Experiment 1: Measurement and Density

Determination of calcium by Standardized EDTA Solution

USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

The introduction of your report should be written on the on the topic of the role of indicators on acid base titrations.

SODIUM CARBOXYMETHYL CELLULOSE

Determination of a Chemical Formula

PART I: PREPARATION OF SOLUTIONS AND STANDARDIZATION OF A BASE

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: )

Colorimetric Determination of Iron in Vitamin Tablets

Physical Properties of a Pure Substance, Water

XI. Methods of Analysis DETERMINATION OF POTASSIUM CARBONATE CALCULATIONS REAGENTS PROCEDURE

EXPERIMENT 7 Reaction Stoichiometry and Percent Yield

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Measuring Manganese Concentration Using Spectrophotometry

The Empirical Formula of a Compound

Paper Chromatography: Separation and Identification of Five Metal Cations

EDTA Titrations 1: Standardization of EDTA and Analysis of Zinc in a Supplement Tablet. by Professor David Cash. September, 2008

experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.

AN EXPERIMENT IN ALCHEMY: COPPER TO SILVER TO GOLD 2005, 2000, 1996 by David A. Katz. All rights reserved

Experiment 5 Preparation of Cyclohexene

OLIVÉR BÁNHIDI 1. Introduction

DR/4000 PROCEDURE. CHLORINE, Free

Quaternary Ammonium Compounds DOC

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Esterification Method 1 Method to 2800 mg/l (as acetic acid) Reagent Solution

HYDRATES 2009 by David A. Katz. All Rights reserved. Reproduction permitted for education use provided original copyright is included.

Color, True and Apparent

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

SEPARATION OF A MIXTURE OF SUBSTANCES LAB

Determination of Citric Acid in Powdered Drink Mixes

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq)

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

HOW TO MAKE STANDARD SOLUTIONS FOR CHEMISTRY

Calibration of Volumetric Glassware

Factors Affecting Enzyme Activity

Mixtures and Pure Substances

Solubility Curve of Sugar in Water

Color, True and Apparent

Phosphorus, Reactive (Orthophosphate)

Hydrogen Peroxide Cell-Based Assay Kit

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

Transcription:

EXPERIMENT 4 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium in it. Clean all glassware and rinse thoroughly with deionized water both before and after use. There is sufficient sodium in tap water and even in distilled water to invalidate your results. Remember also to rinse out your plastic wash bottle several times and then fill it with deionized water. SAFETY WARNING CAUTION Although the natural-gas/air flame is rather small, it has such a high temperature that contact of the flesh with even the outer edge of the flame will instantly produce a third-degree burn. Hands should be kept completely out of the chimney or burner housing whenever the flame is burning. NEVER put your hand above the burner housing. UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown sodium solution can be issued. Your name, section number, and your locker number should be written legibly on this flask. The flask does not need to be dry on the inside, but needs to have been rinsed with deionized water thoroughly after it was washed. The flask must be turned in at least 1 lab period before you plan to do the experiment so that the Teaching Assistants will have time to prepare the unknown. Each student will have his or her own unknown to analyze even if working in pairs. BACKGROUND Flame photometry, now more properly called flame atomic emission spectrometry or flame photometry is a relatively old instrumental analysis method. Its origins date back to Bunsen s flame-color tests for the qualitative identification of select metallic elements. Probably the most common example of the atomic emission effect is fireworks for 4 th of July celebrations and other events. As an analytical method, atomic emission is a fast, simple, and sensitive method for the determination of trace metal ions in solution. Because of the very narrow (ca. 0.01 nm) and characteristic emission lines from the gas-phase atoms in the flame plasma, the method is relatively free of interferences from other elements. Typical precision and accuracy for analysis CHE 226 Analytical Chemistry Laboratory 24 Na Atomic Emission

of dilute aqueous solutions with no major interferences present are about ±1-5% relative. Detection limits can be quite low. Good elements typically have detection limits between about 1 ng/ml and 1 µg/ml. The method is suitable for many metallic elements, especially for those metals that are easily excited to higher energy levels at the relatively cool temperatures of some flames Li, Na, K, Rb, Cs, Ca, Cu, Sr, and Ba. Metalloids and nonmetals generally do not produce isolated neutral atoms in a flame, but mostly as polyatomic radicals and ions. Therefore, nonmetallic elements are not suitable for determination by flame emission spectroscopy, except for a very few and under very specialized conditions. Flame photometry is a highly empirical, rather than an absolute, method of analysis such as gravimetry. That is, you must calibrate the method carefully and frequently. Many different experimental variables affect the intensity of light emitted from the flame and that finding its way to the detector. Therefore, careful and frequent calibration is required for good results. INSTRUMENTATION Buck Scientific Flame Photometer, Model PFP7 The PFP7 Flame Photometer is a low-temperature (air/natural gas) flame atomic emission photometer designed for the routine determination of sodium and potassium in aqueous solutions, two very important physiological elements. The normal adult ranges for Na + and K + in plasma are 136-145 mm and 3.5-5.0 mm, respectively. These levels correspond to about 3200 and 170 µg/ml. Plasma is typically diluted 100- to 200-fold prior to analysis. Additional filters are available for this instrument for lithium, calcium, and barium. The low-temperature flame (about 1700 ο C as compared to oxygen/acetylene at 3100 ο ) generates strong emission only from the most easily excited elements. Wavelength isolation is by use of a simple narrow-bandpass interference filter that is designed to transmit only the intense, characteristic sodium-doublet lines at about 589.0 and 589.6 nm. [Separate filters must be used to transmit the calcium line at 442.7 nm or the two potassium lines at 766.5 and 769.9 nm.] The detector is a relatively inexpensive, sturdy p-i-n photodiode. This solid-state device has an intrinsic (non-doped) layer sandwiched between the usual p- and n-doped layers that are in any diode thus the origin of the appellation p-i-n. This arrangement gives the detector greater sensitivity and faster operating speed than standard photodiodes. [http://www.rp-photonics.com/p_i_n_photodiodes.html] The instrument is called a single-channel photometer because it can determine only one element at a time and has a single direct-reading output. The filter must be changed and the instrument recalibrated for a different element. The instrument uses a capillary aspirator to inject the sample into a mixing chamber containing a PTFE spray-impact bead and several PTFE baffles that serve to mix the fuel, oxidant, and sample droplets. This combination generates a sample mist of only the smallest droplets to enter the burner; most of the sample aspirated goes down the drain. Sample solution consumption is 2-6 ml/min. CHE 226 Analytical Chemistry Laboratory 25 Na Atomic Emission

The manufacturer claims the limits of detection for the instrument are 0.2, 0.2, 0.25, 15, and 30 µg/ml for Na, K, Li, Ca, and Ba, respectively. The reproducibility is said to be better than 1% relative standard deviation for 20 consecutive samples using 10 ppm Na set to read 50.0 on the meter. EQUIPMENT NEEDED Wash bottle(s) rinsed several times and then filled with deionized water One 500-mL volumetric flask, from your locker Assorted volumetric and/or graduated transfer pipets (provided for you in the locker designated for this experiment. Five 100-mL volumetric flasks for the standards (experiment locker) Eight to ten small plastic containers for aspirating solutions (experiment locker) PREPARATION OF SOLUTIONS Standard Sodium Stock Solution, 100.0 ppm 1. Accurately (to 0.1 mg) weigh out by difference 0.1271 g of reagent grade NaCl into a small plastic weighing boat. It is very difficult and time consuming to weigh out exactly this amount. Get it as close as you reasonably can, record the exact mass, and correct your concentrations accordingly. Remember: NEVER transfer chemicals inside an analytical balance. 2. Carefully transfer the salt quantitatively into a 500-mL volumetric flask. Use a few squirts of deionized water from your wash bottle on the weighing boat and the sides of the flask to wash all of it down into the flask. [0.100 g Na/L = 100 mg/l = 100 µg/ml = 100 ppm Na). 3. Add about 100 ml of deionized water to the flask, swirl several times, and dissolve all of the salt before diluting to volume with deionized water. This is critical. Sodium Standard Calibration Solutions 1. Use deionized water for the blank. 2. Pipet 1.00, 2.00, 3.00, 4.00, and 5.00 ml of the standard 100-ppm sodium solution into the first, second, third, fourth, and fifth 100-mL volumetric flasks, respectively. 3. Dilute carefully to the mark with deionized water and mix thoroughly. Unknown Solution Obtain the unknown from the instructor and carefully dilute to the 100-mL mark with deionized water. Mix thoroughly. CHE 226 Analytical Chemistry Laboratory 26 Na Atomic Emission

PROCEDURE Carefully follow the instructions provided you for use of the instrument and measure the emission intensity for the blank (deionized water), each standard, and the unknown(s). 1. When you are approaching the time to begin taking emission readings, call the Teaching Assistant to light the flame, stabilize the flame photometer, and instruct you in its proper and safe use. The instrument should have been turned on and the flame lit for 15 minutes [aspirating deionized water] to ensure stability 2. Thoroughly rinse all the equipment you will use in this experiment, first with lots of distilled water, secondly with deionized water from a rinse bottle. 3. Fill the tall, 25-mL, capped polyethylene vials with the Blank (deionized water), the five standards (1, 2, 3, 4, and 5 ppm Na) and the unknown solution(s) in that order and place them in the plastic holder designed for them. Because water droplets cling to the vials, their insides will need to be pre-rinsed with small amounts of their solutions first. Put a ml or two into a vial, cap it, shake it, then shake the contents into the sink. Do this at least 3 times for each vial. 4. Aspirate deionized water until the meter reading stabilizes, this may take 30-90 sec. Use the blank knob to set the meter reading to 0.00. Then aspirate the highest standard (5 ppm) until the meter reading has stabilized. Use the fine sensitivity knob to set the meter reading to 5.00. [The coarse sensitivity switch should be in the correct setting and not have to be switched.] 5. Repeat the two-step calibration procedure with deionized water and the 5 ppm standard as many times as it takes to get them both stabilized at 0.00 and 5.00, respectively. 6. Aspirate the blank, the 5 standards, and the unknown(s) in that order. Take three replicate readings of each solution once the meter reading has stabilized. There will be some bounce (noise) in the readings, especially at the higher concentrations. 7. For the second calibration run, place the unknown solution(s) between the two standards whose readings bracket that of unknown(s), so that the concentrations of the solutions aspirated now all increase monotonically. Atomic emission instruments work best when going from low to high concentrations. 8. Repeat the whole process of calibration and taking triplicate readings as before at least 1 or 2 more times. The more data you have to review, the better you will be able to detect and eliminate determinate error inaccuracy in the final reported value. 9. INSTRUMENT ALERT. The aspirator compartment does not drain properly sometimes. The symptom to watch for: You are aspirating some solution and the reading starts drifting around. You then aspirate deionized water, but there is still a drifting and significant signal for Na. If you open the little window on the flame chimney, you can see some of the yellowish emission of Na. This means that the aspirator compartment needs to be flushed, that is, get it draining again. Usually this can be accomplished by gently wiggling the drain tygon tubing that is attached to the bottom of the aspirator-burner, just a little below where it is attached. If you have continuing trouble, ask the TA for assistance. CHE 226 Analytical Chemistry Laboratory 27 Na Atomic Emission

10. When completely done with the experiment, aspirate deionized water to clean out the aspirator/burner, clean the work areas up thoroughly, and notify a TA that the instrument is ready to be shut down. 11. Thoroughly rinse all the glass and plastic ware provided you for the experiment with deionized water. Drain the water out and put the equipment back in the drawer. HAZARDOUS WASTE DISPOSAL NO hazardous materials are used or generated in this experiment. All we have are dilute solutions of ordinary table salt. Empty ALL the solutions down a sink drain with cold water running. Thoroughly rinse all glass or plastic ware you used with deionized water. DATA TREATMENT Prepare a calibration curve by plotting the emission intensities as a function of Na concentration. Determine the concentration of sodium in the unknown sample by reading the concentration of the sample which corresponds to its emission intensity from the calibration curve. Depending on the drift in the instrument and other factors, it may be better to average all three values for each solution and obtain one final value for the unknown, or to get three separate values for the unknown, each using its own calibration curve, and average the three values. Try both to see if one approach seems to be better than the other. If the plot appears to be reasonably linear, or at least that portion of it that includes your unknown, use the Excel LINEST function to do a linear-least-squares fit to the data, which will also provide you some quality parameters for the fit. Report the best estimate for the average concentration of sodium in ppm (µg/ml) and the associated standard deviation of the value. TEXT REFERENCE D. A. Skoog, D. M. West, F. J. Holler, and S. R. Crouch, Analytical Chemistry: An Introduction, 7th ed., Chapter 23, pp. 594-631. Revised March 8, 2007 Copyright by The Department of Chemistry,, 2007 CHE 226 Analytical Chemistry Laboratory 28 Na Atomic Emission