Assessment of left ventricular function by different speckle-tracking software



Similar documents
Disclosure. All the authors have no conflict of interest to disclose in this study.

Normal ranges of left ventricular global longitudinal strain: A meta-analysis of 2484 subjects

How To Measure Global Strain Of The Heart Through Speckle Tracking

Assessment of Right Ventricular Systolic Function by Tissue Doppler Echocardiography

Quantification of regional myocardial function by strain rate and strain for evaluation of. artery disease. Automated versus manual analysis during

Low-gradient severe aortic stenosis with normal LVEF: A disturbing clinical entity

Fall 3 Ekokardiografi

Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging

Magnetic Resonance Quantitative Analysis. MRV MR Flow. Reliable analysis of heart and peripheral arteries in the clinical workflow

Vivid S6. Cardiovascular ultrasound system. imagination at work. GE Healthcare

FUNÇÃO VENTRICULAR ESQUERDA POR ECO-3D/4D - UM NOVO PARADIGMA DE AVALIAÇÃO? LEFT VENTRICULAR FUNCTION BY 3D/4D-ECHO - A NEW PARADIGM OF ASSESSMENT?

Cardiology Fellowship Manual. Goals & Objectives -Cardiac Imaging- 1 Page

Vs. K. Farsalinos, D. Tsiapras, S. Kyrzopoulos, M. Savvopoulou, E. Avramidou, D. Vassilopoulou, V. Voudris

Diagnostic and Therapeutic Procedures

Pocket-size ultrasound, a new diagnostic tool in clinical practice

HeAT: A Software Assistant for the Analysis of LV Remodeling after Myocardial Infarction in 4D MR Follow-Up Studies

Dysfunction of aortic valve prostheses

Company profile. Quote from the analyst of. Company Milestone. Feel the Difference. Tel: Fax:

GE Healthcare. Vivid 7 Dimension Cardiovascular Ultrasound System

Clinical Training for Visage 7 Cardiac. Visage 7

Charlotte Bjork Ingul Gothenburg, Sweden Klingenbergsvei 9, 7043 Trondheim, Norway

Three Dimensional Ultrasound Imaging

Blood Flow Hemodynamics, Cardiac Mechanics, and Doppler Echocardiography

MYOCARDIAL PERFUSION COMPUTED TOMOGRAPHY PhD course in Medical Imaging. Anne Günther Department of Radiology OUS Rikshospitalet

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

1p36 and the Heart. John Lynn Jefferies, MD, MPH, FACC, FAHA

RACE I Rapid Assessment by Cardiac Echo. Intensive Care Training Program Radboud University Medical Centre NIjmegen

BIOMEDICAL ULTRASOUND

Medical management of CHF: A New Class of Medication. Al Timothy, M.D. Cardiovascular Institute of the South

Recommendations of the European Association of Echocardiography

Fetal Left Ventricular Mass Determination on 2-Dimensional Echocardiography Using Area-Length Calculation Methods

Atrial Septal Aneurysm: A Study in Five Hundred Adult Patients

Principles of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University

ECHOCARDIOGRAPHY PROPERTY OF ELSEVIER SAMPLE CONTENT - NOT FINAL CHAPTER 6. Hisham Dokainish, MD, FACC, FASE

Real-Time 3-Dimensional Transesophageal Echocardiography in the Evaluation of Post-Operative Mitral Annuloplasty Ring and Prosthetic Valve Dehiscence

Australian and New Zealand College of Anaesthetists (ANZCA) Guidelines on Training and Practice of Perioperative Cardiac Ultrasound in Adults

INHERIT. The Lancet Diabetes & Endocrinology In press

Signal-averaged electrocardiography late potentials

Focused Transthoracic Echocardiography by Sports Medicine Physicians

Applications of Doppler Ultrasound in Fetal Growth Assessment. David Cole

DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA.

Note: The left and right sides of the heart must pump exactly the same volume of blood when averaged over a period of time

Duplication Images in Vascular Sonography

Press. Siemens solutions support diagnosis and treatment of cardiovascular diseases

Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice

Recommendations for Evaluation of the Severity of Native Valvular Regurgitation with Two-dimensional and Doppler Echocardiography

Deriving the 12-lead Electrocardiogram From Four Standard Leads Based on the Frank Torso Model

GUIDELINES AND STANDARDS

ACLS Defibrillation Protocols With the ZOLL Rectilinear Biphasic Waveform AHA/ERC Guidelines 2005

CHAPTER THREE COMMON DESCRIPTIVE STATISTICS COMMON DESCRIPTIVE STATISTICS / 13

Cardiac Tool User s Guide. Cardiac Tool. Version 1.0

X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1)

Exchange solutes and water with cells of the body

Provider Checklist-Outpatient Imaging. Checklist: Nuclear Stress Test, Thallium/Technetium/Sestamibi (CPT Code )

Is Left Ventricular Diastolic Dysfunction a clinical predictor of Syncope?

FETAL HEART ASSESSMENT USING THREE-DIMENSIONAL ULTRASOUND

Distance Learning Program Anatomy of the Human Heart/Pig Heart Dissection Middle School/ High School

CARDIOLOGY PROCEDURES REQUIRING PRECERTIFICATION

AI CPT Codes. x x MRI Magnetic resonance (eg, proton) imaging, temporomandibular joint(s)

Heart Failure EXERCISES. Ⅰ. True or false questions (mark for true question, mark for false question. If it is false, correct it.

12-Lead EKG Interpretation. Judith M. Haluka BS, RCIS, EMT-P

STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT

Specific Basic Standards for Osteopathic Fellowship Training in Cardiology

Atrial Fibrillation 2014 How to Treat How to Anticoagulate. Allan Anderson, MD, FACC, FAHA Division of Cardiology

CPT Radiology Codes Requiring Review by AIM Effective 01/01/2016

Mónica Cristina Cruz Lemini

Computed Tomography, Head Or Brain; Without Contrast Material, Followed By Contrast Material(S) And Further Sections

Cardiovascular Guidelines for DOT Physical Exams By Maureen Collins MSN, APRN, BC

Philips CX50 CompactXtreme System. A New Class of Compact Echocardiography

Mobile Echocardiography

How To Write An Echocardiogram

Epicardial Echocardiography and Epiaortic Ultrasonography

Diagnostic Imaging Prior Review Code List 3 rd Quarter 2016

ECE 533 Project Report Ashish Dhawan Aditi R. Ganesan

In clinical medicine, detailed information on the morphological

PHARMACOLOGICAL Stroke Prevention in Atrial Fibrillation STROKE RISK ASSESSMENT SCORES Vs. BLEEDING RISK ASSESSMENT SCORES.

CARDIAC SONOGRAPHER SERIES

How To Analyse Data From A Heart Study

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Anatomi & Fysiologi The cardiovascular system (chapter 20) The circulation system transports; What the heart can do;

Transcription:

European Journal of Echocardiography (2010) 11, 417 421 doi:10.1093/ejechocard/jep226 Assessment of left ventricular function by different speckle-tracking software Ana Manovel*, David Dawson, Benjamin Smith, and Petros Nihoyannopoulos* Department of Cardiology, Hammersmith Hospital, NHLI, Imperial College London, Du Cane Road, London W12 0NN, UK Received 9 November 2009; accepted after revision 5 December 2009; online publish-ahead-of-print 27 February 2010 Aims Two-dimensional (2D) speckle echocardiography enables objective assessment of left ventricular function through the analysis of myocardial strain, which can be measured by different speckle-tracking software. The aim of this study was to compare two different commercially available cardiac ultrasound systems and their manufacturerspecific speckle-tacking software for the quantification of global myocardial strain in a healthy population.... Methods Twenty-eight healthy subjects (age: 38 + 12, 64% males) underwent two 2D echocardiograms within the same day and results using different cardiac ultrasound systems: Vivid 7 (GE Ultrasound, Horten, Norway) and Artida 4D (Toshiba Medical Systems). Standard apical and short-axis views of the left ventricle were obtained in each subject with a frame-rate range of 60 + 20 frames/s. Global longitudinal, radial, and circumferential strain values were analysed using their respective speckle-tracking software for Vivid (2D-strain EchoPac PC v.7.0.1, GE Healthcare, Horten, Norway) and Toshiba systems (2D Wall Motion Tracking, Toshiba Medical Systems). Global strain values were estimated from the average of regional left ventricular strain values. Agreement between the two systems and software was assessed by Bland Altman method. Mean left ventricular ejection fraction was 59 + 7%. Global longitudinal, radial, and circumferential strain values were, respectively, 221.95 + 1.8, 46.97 + 5.5, and 223.18 + 3.3% when using 2D-strain EchoPac and 222.28 + 2.1, 40.74 + 4.3, and 227.17 + 4.7% when 2D Wall Motion Tracking was used (P ¼ NS). Limits of agreement between both speckle-tracking software were narrower for global longitudinal strain (22.25 to 3.65) than for radial and circumferential strain (22.23 to 12.44 and 21.36 to 10.54, respectively).... Conclusion Two commercially available speckle-tracking software appear to be comparable when quantifying left ventricular function in a healthy population. Global longitudinal strain is a more robust parameter than radial and circumferential strain for the assessment of myocardial function when different cardiac ultrasound systems are used for analysis. ----------------------------------------------------------------------------------------------------------------------------------------------------------- Keywords Left ventricular function Speckle tracking Myocardial strain Introduction The advent of two-dimensional speckle echocardiography (2DSE) has allowed a comprehensive and quantitative assessment of left ventricular function. 1 3 On the basis of a frame-to-frame tracking of myocardial acoustic markers and the subsequent spatial and temporal image processing, current 2DSE software enable the analysis of myocardial strain in its longitudinal, circumferential, and radial dimensions. 4 Over the past 5 years, numerous studies have demonstrated that impaired myocardial strain is an indicator of global and regional dysfunctions in a variety of cardiac conditions, although different speckle-tracking software have been commonly used. 5,6 The accuracy and rapid analysis of 2DSE convert this technique into a suitable method for objective assessment of myocardial function in clinical practice. Nonetheless, consistency of reported data and reference values are needed before its clinical application can be widely accepted. Vendor diversity of currently available speckle-tracking software leads, therefore, to investigate whether quantitative values obtained from every system are similar. Scarce studies have been performed in the past to assess the comparability of systems for tissue Doppler imaging, 7 and no data have been reported using 2DSE. We, therefore, sought to examine the consistency of 2D speckle-derived myocardial strain values in healthy subjects when * Corresponding authors. Juan Ramon Jimenez Hospital, Ronda Norte s/n, Huelva, Spain. Tel: þ34 677886796; fax: þ34 959016785. E-mail address: ajmanosan@yahoo.es (A.M.); Tel: þ44 2083832115; fax: þ44 2083834392. E-mail address: petros@imperial.ac.uk Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2010. For permissions please email: journals.permissions@oxfordjournals.org

418 A. Manovel et al. analysed by two different cardiac ultrasound systems and their respective software. Methods Study population Thirty-three subjects (age: 38 + 12, 64% males) with the absence of structural cardiac disease were included in the study. All participants underwent two transthoracic echocardiograms within the same day using different cardiac ultrasound systems. Global longitudinal, radial, and circumferential strain values were analysed in all subjects using the specific 2D speckle-tracking software of each system. Echocardiography Transthoracic echocardiograms were performed using two commercially available ultrasound transducer and equipment (M3S probe, Vivid 7, GE Healthcare, Horten, Norway; 3 MHz/PST-30SBT probe, 4D Artida, Toshiba Medical Systems). Standard three apical and shortaxis views of the left ventricle were obtained using a frame-rate of 40 80 frames/s. Three cardiac cycles at each plane were stored in cineloop format in order to subsequently select the images of better quality for off-line speckle-tracking analysis. The same transducer frequency (1.7 3.6 MHz) was used in both systems, and special care was taken to use similar sector width, focus position, and frame-rate range during image acquisition. Two-dimensional speckle-tracking analysis Speckle-tracking analysis was performed by the respective customized 2DSE software for Vivid (2D-strain EchoPac PC v.7.0.1, GE Healthcare, Horten, Norway) and Toshiba system (2D Wall Motion Tracking, Toshiba Medical Systems). The endocardial border was manually traced, and a region of interest was drawn to include the entire myocardium in all cases. The software algorithms automatically segmented the left ventricular planes into equidistant segments and then performed speckle tracking on a frame-to-frame basis. When using 2D-strain Echopac, images were accepted for analysis when global tracking quality index equalled 1, which implied that all segments approved for speckle analysis were tracked reliably. When 2D Wall Motion Tracking was used, exclusion of the studies was done upon visual assessment, when abnormal curves were believed to be artefactual. Longitudinal strain values were obtained from the three apical views, and radial and circumferential strain measurements from the basal, mid, and apical short-axis planes. Global strain values were calculated from the average of regional values and stated in per cent (%). Intraobserver and interobserver reproducibility of longitudinal, radial, and circumferential strain values was analysed with repeated measurements by the same observer at two different time points and by a second independent observer in five randomly selected subjects. Both observers were blinded to the results of the other software package and previous strain results when assessing reproducibility. Statistical analysis All values were expressed as mean + SD. Agreement between the two speckle-tracking methods was assessed by Bland Altman analysis. 8 The mean difference and limits of agreement between the measurements derived from each system were calculated. Intraobserver and interobserver variability values were estimated as the absolute difference between the corresponding two measurements in per cent of their mean in each patient and then average over the five patients. Table 1 Clinical and echocardiographic characteristics of study patients (n 5 28) Age (years) 38 + 12 Male (%) 18 (64) LVEF (%) 59 + 7 FS (%) 33 + 4 S a (cm/s) 10 + 2 E/A 1.5 + 0.4 E a /A a 2.0 + 0.6 LVEF, left ventricular ejection fraction; FS, fractional shortening; S a, systolic tissue velocity at lateral mitral annulus; E/A, ratio of early and late diastolic waves of mitral inflow; E a /A a, ratio of early and late diastolic tissue velocities at lateral mitral annulus. Statistical analysis was performed using SPSS software (version 15.0, SPSS Inc, Chicago, IL, USA). Results Baseline characteristics All study subjects showed normal systolic function with LVEF, fractional shortening, and myocardial systolic velocities by tissue Doppler within normal limits. Diastolic function, as assessed by mitral valve inflow pattern and E/A ratio at the lateral mitral annulus, was also normal in all subjects (Table 1). Average heart rate showed no differences when echocardiogram was performed with both ultrasound systems (64 + 11 vs. 68 + 12 bpm, P ¼ NS). Speckle-derived strain analysis Speckle analysis was feasible in 28 subjects from initially recruited 33, since 5 were excluded for speckle analysis due to inadequate tracking by either of the software. Time required for speckle analysis and interpretation of all strain curves was 7 + 2 and 9 + 3 min for Vivid and Toshiba systems, respectively. Similar peak global longitudinal strain was found when 2D-strain EchoPac and 2D Wall Motion Tracking software were used (221.95 + 1.8 vs. 222.28 + 2.1%, respectively; P ¼ 0.56). Representative examples of longitudinal strain measured by the two systems are shown in Figure 1. For measurements derived from short-axis planes, radial strain appeared to be lower (46.97 + 5.5 vs. 40.74 + 4.3%, P ¼ 0.19) and circumferential strain higher (223.18 + 3.3 vs. 227.17 + 4.7%, P ¼ 0.06) when using Toshiba system than those values obtained from 2D-strain EchoPac. Nevertheless, neither of them reached statistical significance. The comparability of global strain measurements between both speckle-tracking software showed good agreement for global longitudinal strain, with an average difference of 0.33 + 1.48 (mean + 2 SD) and limits of agreement of 22.25 to 3.65. The limits were found to be wider for radial and circumferential strain (22.23 to 12.44 and 21.36 to 10.54, respectively) than for longitudinal strain measurements. Bland Altman plots are depicted in Figure 2.

Assessment of LV function 419 Figure 1 Example of two-dimensional speckle-derived longitudinal strain curves obtained from apical four-chamber view using 2D-strain EchoPac (Vivid) (A) and 2D Wall Motion Tracking (Toshiba) (B) software. Images on the left side represent the region of interest included for speckle analysis, which is automatically divided into six segments and codified in different colours by either system. Graphs on the right side show global and regional longitudinal strain curves. Global values are depicted as white lines (dotted line in A and solid line in B) and regional strain curves in different colours corresponding to each myocardial segment. Intraobserver and interobserver variability in the strain measurements was,6 + 2 and 6 + 4% for Vivid system, and,8 + 9 and 7 + 11% for Toshiba system, respectively. Discussion The main results of this study are that two different commercially available software show good comparability for global longitudinal strain as a parameter of overall left ventricular function, and that lower agreement is found for myocardial deformation parameters derived from short-axis views than for longitudinal strain. These findings are important in view of future clinical applications of 2DSE, particularly for patient diagnosis and follow-up in centres where diversity of cardiac ultrasound systems exist. To our knowledge, this is the first study which compares the 2DSE-derived myocardial strain measurements between different available software. Advantages and pitfalls of speckle-derived strain measurements Owing to the angle independency tracking, 2DSE enables to assess myocardial deformation in 3D and at different cross-sectional views throughout the left ventricle, which provides more

420 A. Manovel et al. dimension as assessed by 2DSE. 4,9 In our study, the value of global longitudinal strain obtained in healthy subjects is the same as reported previously. In addition, the present research shows that global longitudinal strain measurement by 2DSE seems to be consistent when two different 2D-strain software are used for analysis. Strain measurements derived from short-axis views have demonstrated, however, less agreement than longitudinal strain values in this subpopulation. Nonetheless, it is important to mention that when global radial and circumferential measurements were compared with those obtained in previous studies using the same software, similar values were found. 10,11 The observed variation between software when radial and circumferential strain are analysed might be somehow expected owing to the limitations inherent to 2D speckle-tracking. Two-dimensional speckle echocardiography is highly influenced by 2D image quality, and tracking in the lateral direction is particularly dependent on lateral resolution. The fact that speckles are tracked almost within the same scan line when longitudinal strain is assessed might explain the high reproducibility of this parameter. Radial and circumferential strain measurements are, on the other hand, more influenced by the line density and, therefore, highly subjected to the appropriateness of parameter setting for image acquisition. It is also important to note that the considerable out-of-plane motion of acoustic markers in short-axis views could also represent a limitation for the consistency of findings when transmural and circumferential motion are analysed. In addition, the presence of a transmural deformation gradient and differently oriented myofibre layers within the left ventricular wall should be kept in mind, 12,13 since the initial position from where myocardial acoustic markers start to be tracked might differ between software. When validated against cardiac magnetic resonance, transmural strain and circumferential strain by 2DSE have also been reported to have only a fair correlation. 14,15 Figure 2 Bland Altman plots represent the agreement between 2D-strain EchoPac (Vivid) and 2D Wall Motion Tracking (Toshiba) measurements for global longitudinal, radial, and circumferential strain. Dotted horizontal lines denote the bias (mean inter-technique difference) and solid horizontal lines depict the 95% limits of agreement (2 SD around the mean difference). GLS, global longitudinal strain; GRS, global radial strain; GCS, global circumferential strain. Sub-index a: Vivid system; subindex b: Toshiba system. comprehensive assessment of myocardial function than Dopplerbased modalities. Previous studies have shown that the normal myocardium exhibits 20% of global deformation in its longitudinal Clinical relevance of the findings and comparability between systems Since the advent of new cardiac imaging technology based on tissue characterization, the comparison of values from different software manufacturers has been scarcely reported. A previous study using tissue Doppler imaging showed that different systems have sufficient accuracy and agreement for the estimation of myocardial velocities, although variations in measurements of strain and strain rate were present. 7 Two-dimensional speckle echocardiography employs different calculation algorithms from Doppler-based modalities, and whether variations on strain measurements by 2DSE exist was unknown. This is the first study comparing 2D speckle analysis by two different systems. The fact that we have obtained a good agreement and reproducibility for longitudinal strain measurements between EchoPac and 2D Wall Motion Tracking software suggests that this parameter could be suitable for the assessment of global left ventricular function independently whether Vivid 7 (GE Healthcare) or Artida 4D (Toshiba Medical Systems) cardiac ultrasound system is used for analysis. This statement is important, as longitudinal myocardial fibres are known to be first affected in many cardiac conditions. 16,17 Moreover, long-axis function constitutes a substantial

Assessment of LV function 421 component of left ventricular performance, despite the preservation of motion derived from circumferentially oriented myofibres and left ventricular ejection fraction. The findings of this study might be the first step in the searching of conventional strain values for different cardiac conditions, allowing the use of different software available in each institution. This is of notable interest in view of the future clinical application of speckle-tracking echocardiography. With regard to radial and circumferential strain, the clinical significance of the maximal intertechnique difference observed is to be clarified before accepting the use of either speckle-tracking software for the same clinical purpose. 18 Study limitations The results of this study have been obtained from a small group of healthy volunteers. Therefore, extensive validation studies in different clinical settings are necessary in order to be able to extrapolate conclusions to the variety of cardiac conditions. This study could be the guide of future research in this particular setting. Owing to the dependency of 2D speckle analysis on image quality, special care was taken to keep to a minimum the difference in parameters setting between systems when acquiring the images. However, some inter-technique variations were still present. Of note, the consideration of a frame-rate range rather than an exact frame-rate value in the studies and the impossibility to keep the same region of interest thickness between systems are some of the aspects which could have affected our results. Finally, although the algorithms used by the different software packages are supposed to be similar in terms of the speckletracking analysis, which is based on a block-matching approach of the speckle patterns within the myocardium, and the strain calculation as the relative change in length between individual points (1 ¼ L(t) 2 L 0 )/L 0 ), some software-inherent differences in data processing might exist. Conclusion Two different commercially available cardiac ultrasound systems appear to be comparable when quantifying left ventricular function by 2DSE-derived global longitudinal strain. Until more research is available, strain values obtained from short-axis views should be considered more cautiously in the assessment of myocardial function when different speckle-tracking software are used for analysis. Further, similar studies on different cardiac conditions are needed in order to be able to provide reference values for the clinical application of 2DSE. Funding The first author was recipient of a grant from the Spanish Society of Cardiology. Conflict of interest: none declared. References 1. Adamu U, Schmitz F, Becker M, Kelm M, Hoffmann R. Advanced speckle tracking echocardiography allowing a three-myocardial layer-specific analysis of deformation parameters. Eur J Echocardiogr 2009;10:303 8. 2. Perry R, De Pasquale CG, Chew DP, Joseph MX. Assessment of early diastolic left ventricular function by two-dimensional echocardiographic speckle tracking. Eur J Echocardiogr 2008;9:791 5. 3. Bijnens BH, Cikes M, Claus P, Sutherland GR. Velocity and deformation imaging for the assessment of myocardial dysfunction. Eur J Echocardiogr 2009;10:216 26. 4. Hurlburt HM, Aurigemma GP, Hill JC, Narayanan A, Gaasch WH, Vinch CS et al. Direct ultrasound measurement of longitudinal, circumferential, and radial strain using 2-dimensional strain imaging in normal adults. Echocardiography 2007;24: 723 31. 5. Choi JO, Cho SW, Song YB, Cho SJ, Song BG et al. Longitudinal 2D strain at rest predicts the presence of left main and three vessel coronary artery disease in patients without regional wall motion abnormality. Eur J Echocardiogr 2009;10:695 701. 6. Carasso S, Yang H, Woo A, Vannan MA, Jamorski M, Wigle ED et al. Systolic myocardial mechanics in hypertrophic cardiomyopathy: novel concepts and implications for clinical status. J Am Soc Echocardiogr 2008;675 83. 7. Kjaergaard J, Korinek J, Belohlavek M, Oh JK, Sogaard P, Hassager C. Accuracy, reproducibility, and comparability of Doppler tissue imaging by two high-end ultrasound systems. J Am Soc Echocardiogr 2006;19:322 8. 8. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307 10. 9. Reisner SA, Lysyansky P, Agmon Y, Mutlak D, Lessick J, Friedman Z. Global longitudinal strain: a novel index of left ventricular systolic function. J Am Soc Echocardiogr 2004;17:630 3. 10. Becker M, Bilke E, Kühl H, Katoh M, Kramann R et al. Analysis of myocardial deformation based on pixel tracking in two dimensional echocardiographic images enables quantitative assessment of regional left ventricular function. Heart 2006;92:1102 8. 11. Serri K, Reant P, Lafitte M, Berhouet M, Le Bouffos V et al. Global and regional myocardial function quantification by two-dimensional strain: application in hypertrophic cardiomyopathy. J Am Coll Cardiol 2006;47:1175 81. 12. Sengupta PP, Khandheria BK, Korinek J, Wang J, Jahangir A, Seward JB et al. Apex-to-base dispersion in regional timing of left ventricular shortening and lengthening. J Am Coll Cardiol 2006;47:163 72. 13. Delfino JG, Fornwalt BK, Eisner RL, Leon AR, Oshinski JN. Determination of transmural, endocardial, and epicardial radial strain and strain rate from phase contrast MR velocity data. J Magn Reson Imaging 2008;27:522 8. 14. Cho GY, Chan J, Leano R, Strudwick M, Marwick TH. Comparison of twodimensional speckle and tissue velocity based strain and validation with harmonic phase magnetic resonance imaging. Am J Cardiol 2006;97:1661 6. 15. Becker M, Hoffmann R, Kuhl HP et al. Analysis of myocardial deformation based on ultrasonic pixel tracking to determine transmurality on chronic myocardial infarction. Eur Heart J 2006;27:2560 6. 16. Nagueh SF, McFalls J, Meyer D, Hill R, Zoghbi WA, Tam JW et al. Tissue Doppler imaging predicts the development of hypertrophic cardiomyopathy in subjects with subclinical disease. Circulation 2003;108:395e8. 17. Nakai H, Takeuchi M, Nishikage T, Lang RM, Otsuji Y. Subclinical left ventricular dysfunction in asymptomatic diabetic patients assessed by two-dimensional speckle tracking echocardiography: correlation with diabetic duration. Eur J Echocardiogr 2009;10:926 32. 18. Bland JM, Altman DG. A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. Comput Biol Med 1990;20:337 40.