A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials



Similar documents
Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

ORTEC. High Resolution Detection Systems for Interdiction of Nuclear Material Trafficking. Abstract. Introduction

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

ORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software

Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software

Overview of Nuclear Detection Needs for Homeland Security. Abstract


Micro-Detective. Ultra-Light, High-Fidelity Hand-Held Radioisotope Identifier

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA

INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION

ORTEC Radioactivity and its Measurements in Foodstuffs Where Does Radioactivity in Food Come From? Natural Radioactivity: Industrial Radioactivity:

Antoine Henri Becquerel was born in Paris on December 15, 1852

Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)

Tutorial 4.6 Gamma Spectrum Analysis

Radiological mapping and characterization at the Barsebäck nuclear power plant

Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N

RADIATION DETECTION & ANALYSIS SYSTEMS FOR WATER

identifinder R-Series Handheld Radiation Detectors

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Radiological Protection Principles concerning the Natural Radioactivity of Building Materials

Micro-Detective-HX. Enhanced Capability, Ultra-Light, High-Fidelity Hand-Held Radioisotope Identifier

Performance of a portable, Digital-Signal-Processing MCA with Safeguards Germanium Detectors

GammaVision 8. Gamma Spectroscopy Software. Compatible, Efficient, and Defendable Results for Gamma Spectroscopy Applications.

EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES. L. Csedreki 1. Abstract. I. Introduction

arxiv: v1 [physics.ins-det] 14 Jan 2014

Networks and Connections in the Counting Room

ORTEC. Micro-trans-SPEC. Ultra-Light, Battery-Powered, Portable, LN 2 -Free, HPGe Gamma Spectrometer

An Innovative, portable MCA based on Digital Signal Processing. R.D. Bingham, R.M. Keyser, T.R. Twomey

Waste Management 04 Conference, February 29 - March 4, 2004, Tucson, AZ Copyright WM Symposia, Inc. All Rights Reserved. Reprinted with permission.

For convenience, we may consider an atom in two parts: the nucleus and the electrons.

Experiment 10. Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado

RN-222 RELEASE TO THE ENVIRONMENT: COMPARISON BETWEEN DIFFERENT GRANITE SOURCES

Laboratory Technologies, Inc. Multi-Wiper. Time Saving Multiple Detector Wipe Counter

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy

How To Monitor For Radioactive Material In International Mail

Environmental Radiation Monitoring in Taiwan

CONCENTRATION OF U-238, U-235, RA-226, TH-232 AND K- 40 FOR SOME GRANITE SAMPLES IN EASTERN DESERT OF EGYPT.

Activities at temporary job site utilizing U.S. NRC Radioactive Material License # Amendment (,

FORMERLY UTILIZED SITES REMEDIAL ACTION PROGRAM

Application Note. ipix A Gamma imager to support various applications

PROSPECT: Precision Reactor Oscillation and Spectrum experiment

Nuclear data for science & technology: Centres for development

Eckert & Ziegler Reference & Calibration Sources

Application Note. Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System

Measurement of Germanium Detector Efficiency

Radioactivity III: Measurement of Half Life.

Radiation Protection s Benefits in the Production of Iodine-131

[2] At the time of purchase of a Strontium-90 source, the activity is Bq.

Measurement of RF Emissions from a Final Coat Electronics Corrosion Module

SAFE MANAGEMENT OF SMOKE DETECTORS CONTAINING RADIOACTIVE SOURCES

Global Value 7. Productivity Suite for GammaVision. Optimizing Gamma Spectrometry Processes through Secure Data Management and Measurement Automation.

P O L Y T E C R.M. S.R.L.

WM2012 Conference, February 26 -March 1, 2012, Phoenix, Arizona, USA

IDM-200 -V. Interchangeable HPGe Detector Module (IDM) for Systems Integrators

Perspectives of a method for measuring soil-gas radon by an opened counting vial

SAFE MANAGEMENT OF SMOKE DETECTORS CONTAINING RADIOACTIVE SOURCES

Natural radioactivity in waste from former non-uranium mining areas in Norway

CHAPTER 7 SCINTILLATION COUNTING

SECURITY ARRANGEMENTS OF RADIATION SOURCES

Comparison of natural radioactivity removal methods for drinking water supplies: A review

International Symposium on Nuclear Security

Alpha/Beta Counting Manual Systems

MATERIALS LICENSE. 1. Honeywell International, Inc. 3. License Number: SUB-526, Amendment 6-1

Introduction to Geiger Counters

PTP. Professional Training Programs. The source for professional radiological sciences training Courses

Atomic and Nuclear Physics Laboratory (Physics 4780)

Lecture 2 Macroscopic Interactions Neutron Interactions and Applications Spring 2010

COMPENDIUM OF EPA-APPROVED ANALYTICAL METHODS FOR MEASURING RADIONUCLIDES IN DRINKING WATER

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:

Quality Assurance Testing of AtomLab Dose Calibrators

NUCLEAR BOREHOLE LOGGING TECHNIQUES DEVELOPED BY CSIRO-EXPLORATION AND MINING FOR IN SITU EVALUATION OF COAL AND MINERAL DEPOSITS

13C NMR Spectroscopy

ENRICHMENT AND URANIUM MASS FROM NMIS FOR HEU METAL

State of the art & future developments in the detection of nuclear/radioactive materials

A Review of Emerging Gamma Detector Technologies for Airborne. Radiation Monitoring

IAEA Support for the Establishment of Nuclear Security Education

Selection, use and maintenance of portable monitoring instruments

THYROID UPTAKE SYSTEMS LIST OF CONTENTS:

FUNDAMENTALS OF A STATE S NUCLEAR SECURITY REGIME: OBJECTIVE AND ESSENTIAL ELEMENTS

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

Basic Nuclear Concepts

Medical Physics and Radioactivity

Objectives 404 CHAPTER 9 RADIATION

Detecting nuclear and radiological materials

Radiation Detection and Measurement

States Environmental Monitoring at Nuclear Power Plants

NRC s Program for Remediating Polluted Sites J.T. Greeves, D.A. Orlando, J.T. Buckley, G.N. Gnugnoli, R.L. Johnson US Nuclear Regulatory Commission

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL

CRIIRAD report N Analyses of atmospheric radon 222 / canisters exposed by Greenpeace in Niger (Arlit/Akokan sector)

Transcription:

Illicit Trafficking, Sub/cross-national threats, Poster presentation A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials Ronald M. Keyser, Timothy R. Twomey, Daniel L. Upp ORTEC 801 South Illinois Avenue Oak Ridge, TN, 37831 USA email: Ron.keyser@ortec-online.com, tim.twomey@ortec-online.com, dan.upp@ortec-online.com Abstract: The three steps in the interdiction process for illicit trafficking of radioactive and nuclear materials are detection, location and identification. Many handheld radioisotope identifiers (RIDs) have been introduced with the claim to the latter two tasks for gamma-emitting sources. Recent emphasis has been on identification in order to reduce the false positive rate. Previous work compared the search mode of an HPGe-based and a NaI-based RID for neutron and gamma-ray sources. In the present work, the performance of the same two instruments in the identification of radionuclides is compared. The requirements for the identification are given in ANSI N42.34 and related standards. The tests in the standard are for single nuclides and certain mixtures of nuclides for both shielded and unshielded sources. The two systems have been compared under the standard test conditions for a variety of the sources and configurations specified. Results comparing the performance of both systems, with the standard and with each other will be given. The results show the HPGe system provides the correct identification more rapidly and on less spectral data than the NaI system and that it easily exceeds the ANSI N42.34 requirement. Keywords: radioisotope; identifiers; germanium detectors; HPGe; illicit trafficking 1. Introduction The Detective EX is a handheld radioisotope identifier (RID) based on a high purity germanium detector (HPGe) and moderated 3 He tubes for the gamma ray and neutron detection respectively. The size of the HPGe and neutron detectors was based on the efficiency requirements of ANSI N42.34 for the detection of differing amounts of material and on the ability to correctly identify the various nuclides in mixtures. The mixtures specified in the standard are those which could be used to hide prohibited material by masking it with other, innocent, radioactive materials. HPGe is arguably the only material available today which has both the necessary high resolution and adequate detection efficiency in a single detector. The neutron detectors were not used for these identifications. Because of the relatively small size of the HPGe detector (about 12% relative efficiency), there is a widely held misconception that accurate nuclide identification will require a long counting time in order to obtain the necessary spectrum counts. However, these results show that the improved resolution of the HPGe more than compensates for the fewer overall counts than in the scintillator spectrum. In the expected mode of operation of the RIDs, the location of the source is determined in the search mode and then the identification of the nuclides is determined in the identify mode. Prior work [1,2,3,4]

showed that high resolution gamma-ray detectors have the same or greater sensitivity than the most common size of low resolution detectors and that the HPGe detector systems can correctly perform the identification, even for mixtures, in less time than the requirements in the standard. This work compares the time for correct analysis for the HPGe, high resolution RID with a scintillator based RID. 2. Experimental Configuration The two RIDs were described earlier [1] and are shown in Fig. 1. The tests were conducted using the two instruments sequentially. The sources were placed in front of the instrument being tested at 10 cm from the front face of the housing (not necessarily the detector) and aligned with the active center of the detector as defined by the manufacturer. The sources were positioned with low mass and low atomic number materials to reduce scatter. The DU samples were positioned so that the maximum surface area was facing the detector. When the point sources were used with the DU, the point sources were positioned beside the DU to eliminate any shielding by the DU. The average dose, as recorded by the instruments, was about 50 :Sv/h above background, consistent with the ANSI N42.34 [5] requirements. The Detective EX continuously collects and analyzes the spectrum. Once a second, the current analysis result is updated on the display. The running live and real time are also displayed. For each source, the identification was started with the source in place and Figure 1. The Detective and scintillator-based RIDs. the screen was continuously observed so that when the correct result was displayed, the real time (representing the data collection time) could be recorded. The identification was repeated and the results shown are the average of the multiple analyses. The scintillator-based RID operates by collecting a spectrum for a fixed time, storing the spectrum, analyzing the spectrum and displaying the result. For this test, the collection time was set to the minimum time and the manual time extension mode was used to extend the time of the spectrum collection until a result was observed. The recorded times are only for the spectrum collection and do not include the extra time for processing, typically about 25 seconds Both instruments were operated at the same confidence level on the peak areas for the peak to be recognized as present in the spectrum. This level was set such that the uncertainty of the net peak area was less than 5 sigma for most peaks. This level ensures a low level of false positives. For either instrument, if the correct analysis was not observed in 180 seconds, the process was stopped and the result recorded as a failure. The maximum counting time is not given in ANSI N42.34, but is listed in the IAEA draft TECDOC [6] as 60 seconds for unshielded sources. Some of the sources were point sources and some were extended sources. The list of sources is shown in Table 1. Source Nominal activity Form Correct report Cs 137 1 uci Point source Cs 137 Co 60 1 uci Point source Co 60 Ba 133 1 uci Point source Ba 133 Depleted uranium 98 g Bar (5 x 5 x 30) Depleted uranium Depleted uranium 18 g Thin plate Depleted uranium

Mixed 1888 0.5 uci each Point source Am241, Cs137, Co60, Cd109 Thorium 200 g natural thorium Bulk powder, about Th 232, Th 228, or Ra 226 20 mm x 15 mm diameter Natural Uranium Unknown Uranium in glass Uranium Natural Uranium 5000 bq Compressed powder, 5 mm x 40 mm diameter Uranium 2.1. Example spectra As soon as the correct analysis was observed in the Detective, the spectrum was manually saved. An example spectrum for identifying 60 Co is shown in Fig. 2. In this spectrum, note the relatively low number of counts needed to make the identification. The DU and 133 Ba combination is shown in Fig. 3. In this spectrum, note the closely spaced gamma-ray peaks in the spectrum, which are easily separated by the HPGe detector. Figure 2. Short time spectrum of 60 Co from HPGe detector. Figure 3. HPGe Spectrum of depleted uranium and 133 Ba. The corresponding spectra for the scintillator system are shown in Figs. 4 and 5. Figure 4. Scintillator spectrum of 60 Co.

The scintillator spectrum does not show the separation of the gammaray lines, making the identification difficult. 3. Results 3.1 Single nuclides Figure 5. Scintillator spectrum of depleted uranium and 133 Ba. For the single nuclide cases, both RIDs were able to correctly identify the nuclide, however the Detective determined the result in less time than the scintillator-based system. 3.2 DU and DU with other nuclides The HPGe RID was able to identify DU as depleted uranium both alone and with other nuclides in less than 1 minute, and in some cases less than a few seconds. The time to identify increases as the number of gammaray lines in the spectrum increases, but remains at or near the 60 second requirement except for the DU and mixed nuclide standard. The DU plus mixed 1888 standard is identified in about 100 seconds as DU plus the 4 nuclides shown in Table 1. The scintillator RID reported unknown in all cases where DU was the sample. For the mixtures with the 60 Co 133 Ba and natural thorium, the Figure 6. Comparison of time to identify single nuclides alone and with DU by HPGe and Scintillator RIDs. addin nuclide was reported with no DU reported. For the mixture of 137 Cs and DU, the scintillator RID only reported 137 Cs. These results are shown in Fig. 6. 3.3 Other nuclides and combinations Several other combinations of nuclides were tested. The results are shown in Fig. 7. The HPGe RID is able to identify the material in less than 130 seconds in all cases and generally less than 40 seconds. The background test was done in a room with concrete floors, walls and ceiling and the expected result was 40 K. The background dose is about 5 :Sv/h. Figure 7. Comparison of time to identify multiple nuclide mixtures, uranium, and natural material by HPGe and Scintillator RIDs.

The scintillator RID reported unknown for the small DU sample and for the natural uranium sample. For the mixed gamma, only 241 Am and 60 Co were reported with 109 Cd and 137 Cs missing. For the background case, no nuclides were reported. For the thorium ore, any of the possible components (e.g., 232 Th, 228 Th, or 226 Ra) were accepted as the correct result, since the different requirements can dictate which result is displayed. 4. Conclusions The results show that the HPGe-based RID can correctly analyze single nuclide samples and mixtures of materials in less time than a RID based on scintillation detectors. More importantly, the depleted uranium sample was correctly identified by the HPGe and not identified as DU by the scintillator, both in mixtures and even when DU was the only material present. This is possible because of the greatly improved resolution of HPGe compared to scintillators, which greatly improves the signal-to-noise ratio. Further work will concentrate on more complex mixtures. 4. References 1 Keyser, R. M., Twomey, T. R., and Upp, D. L., An Improved Handheld Radioisotope Identifier (RID) for both Locating and Identifying Radioactive Materials, HPS Midyear Meeting, January 2005 2 Keyser, R. M., Twomey, T. R., and Upp, D. L., Performance of Light-Weight, Battery-Operated, High Purity Germanium Detectors for Field Use, INMM Annual Meeting, July 2003 3 Twomey, T. R. and Keyser, R. M., Hand-Held Radio Isotope Identifiers for Detection and Identification of Illicit Nuclear Materials Trafficking: Pushing the Performance Envelope, World Customs Organization Conference, Sept. 2004 4 Keyser, R. M. and Twomey, T. R., Performance of a Portable HPGe-Based Nuclide Identifier on Multiple Nuclide Mixtures, INMM Annual Meeting, July 2004 5 American National Standards Institute, Performance Criteria for Hand-held Instruments for the Detection and Identification of Radionuclides, ANSI N42.34, January, 2004 6 International Atomic Energy Agency, Technical / Functional Specifications for Border Radiation Monitoring Equipment IAEA-TECDOC-xxxx, Rev 20E final draft April 2005