Determination of Impurities in Semiconductor-Grade Nitric Acid with the NexION 300S/350S ICP-MS
|
|
- Derrick Casey
- 7 years ago
- Views:
Transcription
1 APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Semiconductor-Grade Nitric Acid with the NexION 300S/350S ICP-MS Introduction Semiconductor devices are currently being designed with smaller line widths and are more susceptible to low-level impurities. Nitric acid (HNO 3) is widely used as a mixture with hydrofluoric acid (HF) to alter between diffusion-limited or rate-limited etching in the semiconductor industry. The mixture is commonly used to etch and expose the critical layer in the front-end processing. In this stage, the actual devices, including transistors and resistors, are created. A typical front-end process includes the following: preparation of the wafer surface, growth of silicon dioxide (SiO 2), patterning and subsequent implantation or diffusion of dopants to obtain the desired electrical properties, growth or deposition of a gate dielectric and via etching. Any metal impurities present would have detrimental effects on the reliability of an IC device. Nitric acid is also commonly used in semiconductor laboratories to carry out analysis of other semiconductor materials, and thus needs to be of high purity and quality. SEMI Standard C specifies the maximum concentration of metal contaminants by element and tier for nitric acid.
2 Inductively coupled plasma mass spectrometry (ICP-MS) is an indispensable analytical tool for quality control because of its superior capability to detect at ultratrace (ng/l or parts-per-trillion) levels and ability to rapidly determine analytes in various process chemicals. Nevertheless, under conventional plasma conditions, argon ions combine with matrix components to generate polyatomic interferences. Some of the common interferences are 38 Ar 1 H + on 39 K +, 40 Ar + on 40 Ca +, and 40 Ar 16 O + on 56 Fe +. While cool plasma has been shown to be effective in reducing argon-based interferences, it is more prone to matrix suppression than hot plasma, especially for refractory elements or elements with high ionization potentials. In addition, other polyatomic interferences may be preferentially formed under the low plasma energy, which are not seen under hot plasma conditions. Collision cells using multipoles and un-reactive gases have proven useful in reducing polyatomic interferences. This approach necessitates the use of kinetic energy discrimination (KED) to remove the unwanted by-products which could result in loss of sensitivity. The Dynamic Reaction Cell (DRC ) is another correction technique which uses a quadrupole mass filter to create Dynamic Bandpass Tuning (DBT), where only ions of a specific mass range pass through the cell, thus allowing only controlled reactions to take place. As a result of this capability, undesirable by-product ions do not form within the cell, even when very reactive gases are used, such as NH 3 or O 2. impurities in HNO 3 can be measured under a single set of hot plasma conditions for all analytes in one analysis. This was best accomplished using both Standard and Reaction modes in a single method. Experimental conditions Normally, the concentration of HNO 3 is around 70%. In this experiment, a five-fold dilution is carried out on 55% ultra pure HNO 3 (Tamapure-AA 10, Tama Chemicals, Tokyo, Japan). Standard solutions were made from a 10 mg/l multi-element standard (PerkinElmer Pure, PerkinElmer, Shelton, CT USA). The instrument used for this experiment was a NexION 300S ICP-MS (PerkinElmer, Shelton, CT USA). Instrumental parameters and sample introduction components are shown in Table 1. Results HNO 3 samples were quantitatively analyzed using additions calibrations; the calibration curves for K, Ca, Fe and Ni are shown in Figures 1 4, indicating good linearity. This is possible with all the polyatomic interferences removed by the reactive NH 3 gas in combination with the bandpass. The PerkinElmer NexION 300 ICP-MS incorporates Universal Cell Technology, which allows the use of Collision mode (with KED), Reaction mode (incorporating DBT), and Standard mode (where a cell gas is not used), and the user has the ability to select whichever mode(s) are most appropriate for the application, and switch between modes within the same analytical method. This application note demonstrates the ability of the NexION 300 ICP-MS to remove interferences so that trace levels of Figure 1. K calibration, with NH 3 cell gas flow of 0.6 ml/min. Table 1. Instrumental parameters and sample introduction components for the NexION 300S ICP-MS. Spray Chamber: Quartz Cyclonic Plasma Gas: 18 L/min Torch: Standard Quartz Auxiliary Gas: 1.1 L/min Torch Injector: 2-mm Quartz Nebulizer Flow: 1.01 L/min Sampler Cone: Platinum RF Power: 1500 W Skimmer Cone: Platinum Integration Time: 1 sec/mass Nebulizer: Meinhard Type A Concentric Quartz Replicates: 3 2
3 The detection limits (DLs) and background equivalent concentrations (BECs) were both determined in 10% HNO3, while accounting for the sensitivities in 10% HNO3. DLs were calculated by multiplying the standard deviation by three, and BECs were determined by measuring the signal intensities. Recoveries were determined from 10 ng/l spikes. The results are summarized in Table 2. Stability was determined by continuous introduction into the NexION 300S of 10 ng/l spikes (without rinse) for 10 hours. Figures 5 and 6 show excellent stability, with RSDs of < 3% over 10 hours (Table 2, last column). The stability results, combined with the spike recovery data, highlight the ability of the NexION 300S ICP-MS for the determination of all SEMI-required elements in the HNO3 matrix. Figure 2. Ca calibration, with NH3 cell gas flow of 1 ml/min. Figure 5. Ten-hour long-term stability (normalized intensity) for a 10 ng/l spike for first group of analytes. Figure 3. Fe calibration, with NH3 cell gas flow of 0.6 ml/min. Figure 6. Ten-hour long-term stability (normalized intensity) for a 10 ng/l spike for second group of analytes. Figure 4. Ni calibration, with NH3 cell gas flow of 0.3 ml/min. 3
4 Table 2. Detection limits (DLs) and background equivalent concentrations (BECs) for all analytes in 10% HNO 3, and 10 ng/l spike recoveries. Cell Gas Flow* DL BEC 10 ppt Analyte Mass (ml/min) RPq (ppt) (ppt) Recovery % RSD Li % 2.7 Be % 3.3 B % 3.2 Na % 2.1 Mg % 1.3 Al % 1.5 K % 1.7 Ca % 1.1 Ti % 1.8 V < DL 103% 1.3 Cr % 1.8 Mn % 1.3 Fe % 1.5 Co % 1.4 Ni % 1.8 Cu % 1.9 Zn % 2.9 Ga % 1.1 Ge % 1.8 As % 2.1 Sr % 1.4 Zr < DL 102% 1.4 Nb % 1.2 Mo < DL 98% 1.8 Ru < DL 96% 1.5 Rh % 1.2 Pd % 1.8 Ag % 1.7 Cd < DL 101% 1.5 In < DL 102% 1.6 Sn % 1.7 Sb % 1.6 Ba % 1.3 Ta < DL 103% 2.0 W < DL 100% 1.8 Pt % 1.9 Au % Tl < DL 104% 1.2 Pb % 1.5 Bi < DL 103% 1.8 U < DL 102% 1.9 *Cell gas used is NH 3. 4
5 Conclusion The NexION 300S ICP-MS has shown to be robust and suitable for the routine quantification of ultratrace impurities at the ng/l level in HNO 3. By means of computer-controlled switching between Standard mode and Reaction mode in the Universal Cell, interference-free analysis using hot plasma conditions for all analytes is possible during a single sample run. References 1. SEMI Standard C , SEMI Standards, semi.org/en/index.htm PerkinElmer, Inc. 940 Winter Street Waltham, MA USA P: (800) or (+1) For a complete listing of our global offices, visit Copyright , PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners A_01
How To Use The Nexion 300
APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of
More informationStandard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry
Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
More informationMulti-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
More informationUniversal Data Acquisition (UDA)
Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte
More informationDetermination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
More informationELECTRON CONFIGURATION (SHORT FORM) # of electrons in the subshell. valence electrons Valence electrons have the largest value for "n"!
179 ELECTRON CONFIGURATION (SHORT FORM) - We can represent the electron configuration without drawing a diagram or writing down pages of quantum numbers every time. We write the "electron configuration".
More informationSample Analysis Design Isotope Dilution
Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with
More informationB I N G O B I N G O. Hf Cd Na Nb Lr. I Fl Fr Mo Si. Ho Bi Ce Eu Ac. Md Co P Pa Tc. Uut Rh K N. Sb At Md H. Bh Cm H Bi Es. Mo Uus Lu P F.
Hf Cd Na Nb Lr Ho Bi Ce u Ac I Fl Fr Mo i Md Co P Pa Tc Uut Rh K N Dy Cl N Am b At Md H Y Bh Cm H Bi s Mo Uus Lu P F Cu Ar Ag Mg K Thomas Jefferson National Accelerator Facility - Office of cience ducation
More informationApplication of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist
Application of ICP-MS as a Tool for Environmental Monitoring Bill Spence Environmental Market Specialist Introduction Knowledge of the concentration of metallic species in the environment has been important
More informationVALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol
Page 1 van 36 pages VALIDATION REPORT Investigation of the major interferences for proposed Horizontal ICP AES method Identification code Validation in accordance with Reference method Scope Reference
More informationThe 30-Minute Guide to ICP-MS
Technical Note The 30-Minute Guide to ICP-MS ICP-Mass Spectrometry A Worthy Member of the Inorganic Analysis Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been gaining
More informationMILESTONE CLEAN CHEMISTRY LINE DUOPUR TRACECLEAN ULTRA-TRACE INSERTS. A Breakthrough in Reduction and Control of the Analytical Blank MILESTONE
MILESTONE CLEAN CHEMISTRY LINE A Breakthrough in Reduction and Control of the Analytical Blank DUOPUR TRACECLEAN ULTRA-TRACE INSERTS MILESTONE MILESTONE CLEAN CHEMISTRY LINE An innovative and complete
More informationRegional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT
Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.
More informationICP - Mass Spectrometry APPLICATION NOTE. Authors. Introduction. Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA
APPLICATION NOTE ICP - Mass Spectrometry Authors Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA Benefits of the NexION 300X/350X ICP-MS Coupled with the prepfast In-line Auto-dilution/calibrationSystem
More informationNANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
More informationaurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS
aurora M90 A new era in ICP-MS Innovation with Integrity ICP-MS Your Partner in ICP-MS Solutions Bruker continues to find new and novel ways to meet your changing needs. As a leader in elemental analysis
More informationNexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight.
NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight. When your instrument is more efficient and productive, so too are your scientists and your lab. And nothing puts the flow in your
More informationChem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Why? Chemists are concerned with mass relationships in chemical reactions, usually run on a macroscopic scale (grams, kilograms, etc.). To deal with
More informationElectronegativity and Polarity
and Polarity N Goalby Chemrevise.org Definition: is the relative tendency of an atom in a molecule to attract electrons in a covalent bond to itself. is measured on the Pauling scale (ranges from 0 to
More informationChem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers
Key Questions & Exercises Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers 1. The atomic weight of carbon is 12.0107 u, so a mole of carbon has a mass of 12.0107 g. Why doesn t a mole of
More informationAnalysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
More informationSUS. Company Profile. Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0)208/658535 Fax 0049(0)208/658536
SUS Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0208/658535 Fax 0049(0208/658536 Company Profile SUS was founded in 1986 in Oberhausen in the Ruhr area (close to Düsseldorf, in order to meet
More informationOLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
More informationHow To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
More informationA fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,
More informationCLASS TEST GRADE 11. PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change
CLASS TEST GRADE PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change MARKS: 45 TIME: hour INSTRUCTIONS AND INFORMATION. Answer ALL the questions. 2. You may use non-programmable calculators. 3. You may
More informationHeavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy
Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy In support of the toy industry, many laboratories are being called upon to analyze a wide variety of
More informationCVD SILICON CARBIDE. CVD SILICON CARBIDE s attributes include:
CVD SILICON CARBIDE CVD SILICON CARBIDE is the ideal performance material for design engineers. It outperforms conventional forms of silicon carbide, as well as other ceramics, quartz, and metals in chemical
More informationAnalysis of Chlorine, Bromine and Iodine in Water using ICP-AES
ICP OPTICAL ATOMIC EMISSION SPECTROSCOPY Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES HORIBA Scientific Longjumeau, France Keywords: metallurgy, non-ferrous, halogens 1 Introduction
More informationWipe Analysis to Determine Metal Contamination on Critical Surfaces
By Albert Dato, Ph.D., Warren York, Jennifer Jew, Laarni Huerta, Brice Norton, and Michael Coste On-wafer metallic contamination is detrimental to the fabrication and performance of semiconductor devices.
More informationSustainable energy products Simulation based design for recycling
Sustainable energy products Simulation based design for recycling Markus A. Reuter (Prof. Dr. Dr. hc) Director: Technology Management, Outotec Oyj Aalto University (Finland), Central South University (China),
More information100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals.
2.21 Ionic Bonding 100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals. Forming ions Metal atoms lose electrons to form +ve ions. Non-metal
More informationAll answers must use the correct number of significant figures, and must show units!
CHEM 10113, Quiz 2 September 7, 2011 Name (please print) All answers must use the correct number of significant figures, and must show units! IA Periodic Table of the Elements VIIIA (1) (18) 1 2 1 H IIA
More informationEXPERIMENT 4 The Periodic Table - Atoms and Elements
EXPERIMENT 4 The Periodic Table - Atoms and Elements INTRODUCTION Primary substances, called elements, build all the materials around you. There are more than 109 different elements known today. The elements
More informationChromium Speciation in Drinking Water by LC-ICP-MS
APPLICATION NOTE LC-ICP-MS Authors: Helmut Ernstberger Ken Neubauer PerkinElmer, Inc. Shelton, CT Chromium Speciation in Drinking Water by LC-ICP-MS Introduction With the growing concern of pollutants
More informationStandard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry
Standard Operation Procedure Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
More informationElemental Analyses by ICP-AES
Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to
More informationUsing Direct Solid Sampling ICP-MS to Complement SEM- EDX and SIMS in Characterizing Semiconductor Materials
Using Direct Solid Sampling ICP-MS to Complement SEM- EDX and SIMS in Characterizing Semiconductor Materials Fuhe Li and Scott Anderson Air Liquide America-Balazs Analytical Services, 4649 Landing Parkway,
More informationMass spectrometry. What are the principles behind MS? What do all MS instruments have in common?
Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction
More informationAnalyses on copper samples from Micans
PDF rendering: DokumentID 1473479, Version 1., Status Godkänt, Sekretessklass Öppen Analyses on copper samples from Micans P. Berastegui, M. Hahlin, M. Ottosson, M. Korvela, Y. Andersson, R. Berger and
More informationB) atomic number C) both the solid and the liquid phase D) Au C) Sn, Si, C A) metal C) O, S, Se C) In D) tin D) methane D) bismuth B) Group 2 metal
1. The elements on the Periodic Table are arranged in order of increasing A) atomic mass B) atomic number C) molar mass D) oxidation number 2. Which list of elements consists of a metal, a metalloid, and
More informationAnalytical Method. Determination of metals in workplace air by inductively coupled plasma mass spectrometry (ICP-MS) ANALYTICAL METHOD 362
Analytical Method Determination of metals in workplace air by inductively coupled plasma mass spectrometry (ICP-MS) ANALYTICAL METHOD 362 Applicability This method applies to the determination of several
More informationFinnigan ELEMENT2. High Performance High Resolution ICP-MS
m a s s s p e c t r o m e t r y Finnigan ELEMENT2 High Performance High Resolution ICP-MS Analyze Detect Measure Control Finnigan ELEMENT2 High Performance High Resolution ICP-MS Finnigan ELEMENT2 High
More informationNeutralization of Acid Mine Drainage Using Stabilized Flue Gas Desulfurization Material
Neutralization of Acid Mine Drainage Using Stabilized Flue Gas Desulfurization Material W. Wolfe 1, C.-M. Cheng 1, R. Baker 1, T. Butalia 1, J. Massey-Norton 2 1 The Ohio State University, 2 American Electric
More informationNanoparticle Deposition on Packaging Materials by the Liquid Flame Spray
Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray Hannu Teisala a, Mikko Tuominen a, Mikko Aromaa b, Jyrki M. Mäkelä b, Milena Stepien c, Jarkko J. Saarinen c, Martti Toivakka c
More informationChapter 8 - Chemical Equations and Reactions
Chapter 8 - Chemical Equations and Reactions 8-1 Describing Chemical Reactions I. Introduction A. Reactants 1. Original substances entering into a chemical rxn B. Products 1. The resulting substances from
More informationAnalytical Techniques for Trace Elemental Analyses on Wafer Surfaces for Monitoring and Controlling Contamination
Analytical Techniques for Trace Elemental Analyses on Wafer Surfaces for Monitoring and Controlling Contamination J. Wang, M. K. Balazs Balazs Analytical Laboratory, Sunnyvale, CA 94089 P. Pianetta, K.
More informationQuartz Glass. Tubes and Rods
Quartz Glass Tubes and Rods PH 300, GE 214, QI PN GVB Solutions in Glass Schlackstrasse 3 52080 Aachen Germany +49-241/9108588 +49-241/9108589 E- info@g-v-b.de www.g-v-b.de Table of contents General Information
More informationModule 7 Wet and Dry Etching. Class Notes
Module 7 Wet and Dry Etching Class Notes 1. Introduction Etching techniques are commonly used in the fabrication processes of semiconductor devices to remove selected layers for the purposes of pattern
More informationSample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
More informationFrom Quantum to Matter 2006
From Quantum to Matter 006 Why such a course? Ronald Griessen Vrije Universiteit, Amsterdam AMOLF, May 4, 004 vrije Universiteit amsterdam Why study quantum mechanics? From Quantum to Matter: The main
More informationNomenclature and Formulas of Ionic Compounds. Section I: Writing the Name from the Formula
Purpose: Theory: Nomenclature and Formulas of Ionic Compounds 1. To become familiar with the rules of chemical nomenclature, based on the classification of compounds. 2. To write the proper name of the
More informationPortable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE
Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications
More informationDetermination of elements by ICP-AES and ICP-MS
September 2003 HORIZONTAL - 19 Determination of elements by ICP-AES and ICP-MS Henk J van de Wiel National Institute of Public Health and the Environment (RIVM) Bilthoven, The Netherlands 1 HORIZONTAL
More informationNET IONIC EQUATIONS. A balanced chemical equation can describe all chemical reactions, an example of such an equation is:
NET IONIC EQUATIONS A balanced chemical equation can describe all chemical reactions, an example of such an equation is: NaCl + AgNO 3 AgCl + NaNO 3 In this case, the simple formulas of the various reactants
More informationDimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
More information3. What would you predict for the intensity and binding energy for the 3p orbital for that of sulfur?
PSI AP Chemistry Periodic Trends MC Review Name Periodic Law and the Quantum Model Use the PES spectrum of Phosphorus below to answer questions 1-3. 1. Which peak corresponds to the 1s orbital? (A) 1.06
More informationCoating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
More informationNuclear ZPE Tapping. Horace Heffner May 2007
ENERGY FROM UNCERTAINTY The uncertainty of momentum for a particle constrained by distance Δx is given, according to Heisenberg, by: Δmv = h/(2 π Δx) but since KE = (1/2) m v 2 = (1/(2 m) ) (Δmv) 2 ΔKE
More informationElemental Scientific. Stability. Elemental Scientific
Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dilution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast inline
More informationAgilent 720/730 Series ICP-OES
Agilent 720/730 Series ICP-OES Flexible. Productive. Robust. Specifications Introduction Agilent 720/730 Series ICP-OES spectrometers are manufactured according to a quality management system certified
More informationChapter 8 Atomic Electronic Configurations and Periodicity
Chapter 8 Electron Configurations Page 1 Chapter 8 Atomic Electronic Configurations and Periodicity 8-1. Substances that are weakly attracted to a magnetic field but lose their magnetism when removed from
More informationNomenclature of Ionic Compounds
Nomenclature of Ionic Compounds Ionic compounds are composed of ions. An ion is an atom or molecule with an electrical charge. Monatomic ions are formed from single atoms that have gained or lost electrons.
More informationUnit 2 Matter and Chemical Change. Unit Test
Unit Test Student Name Class Section 1 Properties of Matter 1. Aluminum foam is used to create lighter, safer cars. The reason that a lighter car is a safer car is because aluminum foam is A. less rigid
More informationEXPERIMENT 7 Reaction Stoichiometry and Percent Yield
EXPERIMENT 7 Reaction Stoichiometry and Percent Yield INTRODUCTION Stoichiometry calculations are about calculating the amounts of substances that react and form in a chemical reaction. The word stoichiometry
More informationWork Assignment 5-03. Prepared for: Office of Air Quality Planning and Standards U.S. Environmental Protection Agency Research Triangle Park, NC
ERG No.: 0143.04.005 EPA Contract No.: 68-D-00-264 Standard Operating Procedure for the Determination of Metals In Ambient Particulate Matter Analyzed By Inductively Coupled Plasma/Mass Spectrometry (ICP/MS)
More informationRole of Hydrogen Bonding on Protein Secondary Structure Introduction
Role of Hydrogen Bonding on Protein Secondary Structure Introduction The function and chemical properties of proteins are determined by its three-dimensional structure. The final architecture of the protein
More informationMETHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY
METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY 1.0 SCOPE AND APPLICATION 1.1 Inductively coupled plasma-atomic emission spectrometry (ICP-AES) determines trace elements, including
More informationMulticrystalline solar silicon production for development of photovoltaic industry
Multicrystalline solar silicon production for development of photovoltaic industry A.I. Nepomnyaschikh Institute of Geochemistry, Siberian Branch, Russian Academy of Sciences, E-mail: ainep@igc.irk.ru
More informationFind a pair of elements in the periodic table with atomic numbers less than 20 that are an exception to the original periodic law.
Example Exercise 6.1 Periodic Law Find the two elements in the fifth row of the periodic table that violate the original periodic law proposed by Mendeleev. Mendeleev proposed that elements be arranged
More informationAssessing the Fate of Silver Nanoparticles in Surface Water using Single Particle ICP-MS
APPLICATION NOTE ICP - Mass Spectrometry Authors: Madjid Hadioui, Ph.D. Université de Montréal Kevin Wilkinson, Ph.D. Université de Montréal Chady Stephan, Ph.D. PerkinElmer, Inc. Assessing the Fate of
More informationIt takes four quantum numbers to describe an electron. Additionally, every electron has a unique set of quantum numbers.
So, quantum mechanics does not define the path that the electron follows; rather, quantum mechanics works by determining the energy of the electron. Once the energy of an electron is known, the probability
More informationIntroduction to VLSI Fabrication Technologies. Emanuele Baravelli
Introduction to VLSI Fabrication Technologies Emanuele Baravelli 27/09/2005 Organization Materials Used in VLSI Fabrication VLSI Fabrication Technologies Overview of Fabrication Methods Device simulation
More informationNUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility
Use of Absorption Mechanisms to Decrease Heavy Metal Mobility by Jianmin Wang Honglan Shi Joe G. Burken NUTC R304 A National University Transportation Center at Missouri University of Science and Technology
More informationEXPERIMENT 8: Activity Series (Single Displacement Reactions)
EPERIMENT 8: Activity Series (Single Displacement Reactions) PURPOSE a) Reactions of metals with acids and salt solutions b) Determine the activity of metals c) Write a balanced molecular equation, complete
More informationAccurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic Absorption Spectrometry
application Note Atomic Absorption Authors Jijun Yao Renkang Yang Jianmin Chen PerkinElmer, Inc. Shelton, CT 06484 USA Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic
More informationThe R&D Notebook. Ultrapure Water for Boron and Silica Sensitive Laboratory Applications
The R&D Notebook A publication of the Laboratory Division of Millipore RD3 Ultrapure for Boron and Silica Sensitive Laboratory Applications Ichiro Kano () and Daniel Darbouret (2) () Research & Development,
More information8. Relax and do well.
CHEM 1314 3:30 pm Section Exam II ohn II. Gelder October 16, 2002 Name TA's Name Lab Section INSTRUCTIONS: 1. This examination consists of a total of 8 different pages. The last three pages include a periodic
More informationAPPENDIX B: EXERCISES
BUILDING CHEMISTRY LABORATORY SESSIONS APPENDIX B: EXERCISES Molecular mass, the mole, and mass percent Relative atomic and molecular mass Relative atomic mass (A r ) is a constant that expresses the ratio
More informationHigh Sensitivity ICP-MS: Overcome the problem of complex samples
High Sensitivity ICP-MS: Overcome the problem of complex samples Meike Hamester, Andrew Toms and René Chemnitzer Bruker Daltonics Berlin, Germany - Milton, Canada Historical Product Roadmap 2003 2005 2010
More informationORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software
ORTEC DET-SW-UPG Latest Software Features Three Search Modes: Gamma/Neutron total count rate. SNM search mode. Sliding average "monitor" mode. (NEW) User choice of identification schemes: Classify mode
More information1332 CHAPTER 18 Sample Questions
1332 CHAPTER 18 Sample Questions Couple E 0 Couple E 0 Br 2 (l) + 2e 2Br (aq) +1.06 V AuCl 4 + 3e Au + 4Cl +1.00 V Ag + + e Ag +0.80 V Hg 2+ 2 + 2e 2 Hg +0.79 V Fe 3+ (aq) + e Fe 2+ (aq) +0.77 V Cu 2+
More informationFlow Injection Analysis
Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since
More informationAPI 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
More informationUNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
More informationLead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
More informationImplementation Of High-k/Metal Gates In High-Volume Manufacturing
White Paper Implementation Of High-k/Metal Gates In High-Volume Manufacturing INTRODUCTION There have been significant breakthroughs in IC technology in the past decade. The upper interconnect layers of
More informationSample Exercise 2.1 Illustrating the Size of an Atom
Sample Exercise 2.1 Illustrating the Size of an Atom The diameter of a US penny is 19 mm. The diameter of a silver atom, by comparison, is only 2.88 Å. How many silver atoms could be arranged side by side
More informationTable 1. Toxicology of Arsenic Species. Arsenic Species
www.perkinelmer.com Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Speciation of Five Arsenic Compounds in Urine by HPLC/ICP-MS Introduction High-performance
More informationIII. Wet and Dry Etching
III. Wet and Dry Etching Method Environment and Equipment Advantage Disadvantage Directionality Wet Chemical Solutions Atmosphere, Bath 1) Low cost, easy to implement 2) High etching rate 3) Good selectivity
More informationMODERN ATOMIC THEORY AND THE PERIODIC TABLE
CHAPTER 10 MODERN ATOMIC THEORY AND THE PERIODIC TABLE SOLUTIONS TO REVIEW QUESTIONS 1. Wavelength is defined as the distance between consecutive peaks in a wave. It is generally symbolized by the Greek
More informationElectronegativity and Polarity MAIN Idea A chemical bond s character is related to each atom s
Section 8.5 Objectives Describe how electronegativity is used to determine bond type. Compare and contrast polar and nonpolar covalent bonds and polar and nonpolar molecules. Generalize about the characteristics
More informationCATALOGUE REFERENCE MATERIALS
IPT Institute for Technological Research CATALOGUE REFERENCE MATERIALS 2014 Laboratory of Metrological References IPT s Reference Materials www.ipt.br/nmr.htm Av. Prof. Almeida Prado n0 532 Predio 31 Cidade
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationElemental Analysis Manual for Food and Related Products
Elemental Analysis Manual for Food and Related Products 4.7 Inductively Coupled Plasma-Mass Spectrometric Determination of Arsenic, Cadmium, Chromium, Lead, Mercury, and Other Elements in Food Using Microwave
More informationSecondary Ion Mass Spectrometry
Secondary Ion Mass Spectrometry A PRACTICAL HANDBOOK FOR DEPTH PROFILING AND BULK IMPURITY ANALYSIS R. G. Wilson Hughes Research Laboratories Malibu, California F. A. Stevie AT&T Bell Laboratories Allentown,
More informationElectrochemistry - ANSWERS
Electrochemistry - ANSWERS 1. Using a table of standard electrode potentials, predict if the following reactions will occur spontaneously as written. a) Al 3+ + Ni Ni 2+ + Al Al 3+ + 3e - Al E = -1.68
More informationSampling and preparation of heterogeneous waste fuels?
Sampling and preparation of heterogeneous waste fuels? Is it possible to accomplish a representative and relevant composition data? Evalena Wikström, Lennart Gustavsson and Jolanta Franke Aim of the project
More informationEtching Etch Definitions Isotropic Etching: same in all direction Anisotropic Etching: direction sensitive Selectivity: etch rate difference between
Etching Etch Definitions Isotropic Etching: same in all direction Anisotropic Etching: direction sensitive Selectivity: etch rate difference between 2 materials Other layers below one being etch Masking
More informationContamination. Cleanroom. Cleanroom for micro and nano fabrication. Particle Contamination and Yield in Semiconductors.
Fe Particles Metallic contaminants Organic contaminants Surface roughness Au Particles SiO 2 or other thin films Contamination Na Cu Photoresist Interconnect Metal N, P Damages: Oxide breakdown, metal
More information