Determination of Impurities in Semiconductor-Grade Nitric Acid with the NexION 300S/350S ICP-MS

Size: px
Start display at page:

Download "Determination of Impurities in Semiconductor-Grade Nitric Acid with the NexION 300S/350S ICP-MS"

Transcription

1 APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Semiconductor-Grade Nitric Acid with the NexION 300S/350S ICP-MS Introduction Semiconductor devices are currently being designed with smaller line widths and are more susceptible to low-level impurities. Nitric acid (HNO 3) is widely used as a mixture with hydrofluoric acid (HF) to alter between diffusion-limited or rate-limited etching in the semiconductor industry. The mixture is commonly used to etch and expose the critical layer in the front-end processing. In this stage, the actual devices, including transistors and resistors, are created. A typical front-end process includes the following: preparation of the wafer surface, growth of silicon dioxide (SiO 2), patterning and subsequent implantation or diffusion of dopants to obtain the desired electrical properties, growth or deposition of a gate dielectric and via etching. Any metal impurities present would have detrimental effects on the reliability of an IC device. Nitric acid is also commonly used in semiconductor laboratories to carry out analysis of other semiconductor materials, and thus needs to be of high purity and quality. SEMI Standard C specifies the maximum concentration of metal contaminants by element and tier for nitric acid.

2 Inductively coupled plasma mass spectrometry (ICP-MS) is an indispensable analytical tool for quality control because of its superior capability to detect at ultratrace (ng/l or parts-per-trillion) levels and ability to rapidly determine analytes in various process chemicals. Nevertheless, under conventional plasma conditions, argon ions combine with matrix components to generate polyatomic interferences. Some of the common interferences are 38 Ar 1 H + on 39 K +, 40 Ar + on 40 Ca +, and 40 Ar 16 O + on 56 Fe +. While cool plasma has been shown to be effective in reducing argon-based interferences, it is more prone to matrix suppression than hot plasma, especially for refractory elements or elements with high ionization potentials. In addition, other polyatomic interferences may be preferentially formed under the low plasma energy, which are not seen under hot plasma conditions. Collision cells using multipoles and un-reactive gases have proven useful in reducing polyatomic interferences. This approach necessitates the use of kinetic energy discrimination (KED) to remove the unwanted by-products which could result in loss of sensitivity. The Dynamic Reaction Cell (DRC ) is another correction technique which uses a quadrupole mass filter to create Dynamic Bandpass Tuning (DBT), where only ions of a specific mass range pass through the cell, thus allowing only controlled reactions to take place. As a result of this capability, undesirable by-product ions do not form within the cell, even when very reactive gases are used, such as NH 3 or O 2. impurities in HNO 3 can be measured under a single set of hot plasma conditions for all analytes in one analysis. This was best accomplished using both Standard and Reaction modes in a single method. Experimental conditions Normally, the concentration of HNO 3 is around 70%. In this experiment, a five-fold dilution is carried out on 55% ultra pure HNO 3 (Tamapure-AA 10, Tama Chemicals, Tokyo, Japan). Standard solutions were made from a 10 mg/l multi-element standard (PerkinElmer Pure, PerkinElmer, Shelton, CT USA). The instrument used for this experiment was a NexION 300S ICP-MS (PerkinElmer, Shelton, CT USA). Instrumental parameters and sample introduction components are shown in Table 1. Results HNO 3 samples were quantitatively analyzed using additions calibrations; the calibration curves for K, Ca, Fe and Ni are shown in Figures 1 4, indicating good linearity. This is possible with all the polyatomic interferences removed by the reactive NH 3 gas in combination with the bandpass. The PerkinElmer NexION 300 ICP-MS incorporates Universal Cell Technology, which allows the use of Collision mode (with KED), Reaction mode (incorporating DBT), and Standard mode (where a cell gas is not used), and the user has the ability to select whichever mode(s) are most appropriate for the application, and switch between modes within the same analytical method. This application note demonstrates the ability of the NexION 300 ICP-MS to remove interferences so that trace levels of Figure 1. K calibration, with NH 3 cell gas flow of 0.6 ml/min. Table 1. Instrumental parameters and sample introduction components for the NexION 300S ICP-MS. Spray Chamber: Quartz Cyclonic Plasma Gas: 18 L/min Torch: Standard Quartz Auxiliary Gas: 1.1 L/min Torch Injector: 2-mm Quartz Nebulizer Flow: 1.01 L/min Sampler Cone: Platinum RF Power: 1500 W Skimmer Cone: Platinum Integration Time: 1 sec/mass Nebulizer: Meinhard Type A Concentric Quartz Replicates: 3 2

3 The detection limits (DLs) and background equivalent concentrations (BECs) were both determined in 10% HNO3, while accounting for the sensitivities in 10% HNO3. DLs were calculated by multiplying the standard deviation by three, and BECs were determined by measuring the signal intensities. Recoveries were determined from 10 ng/l spikes. The results are summarized in Table 2. Stability was determined by continuous introduction into the NexION 300S of 10 ng/l spikes (without rinse) for 10 hours. Figures 5 and 6 show excellent stability, with RSDs of < 3% over 10 hours (Table 2, last column). The stability results, combined with the spike recovery data, highlight the ability of the NexION 300S ICP-MS for the determination of all SEMI-required elements in the HNO3 matrix. Figure 2. Ca calibration, with NH3 cell gas flow of 1 ml/min. Figure 5. Ten-hour long-term stability (normalized intensity) for a 10 ng/l spike for first group of analytes. Figure 3. Fe calibration, with NH3 cell gas flow of 0.6 ml/min. Figure 6. Ten-hour long-term stability (normalized intensity) for a 10 ng/l spike for second group of analytes. Figure 4. Ni calibration, with NH3 cell gas flow of 0.3 ml/min. 3

4 Table 2. Detection limits (DLs) and background equivalent concentrations (BECs) for all analytes in 10% HNO 3, and 10 ng/l spike recoveries. Cell Gas Flow* DL BEC 10 ppt Analyte Mass (ml/min) RPq (ppt) (ppt) Recovery % RSD Li % 2.7 Be % 3.3 B % 3.2 Na % 2.1 Mg % 1.3 Al % 1.5 K % 1.7 Ca % 1.1 Ti % 1.8 V < DL 103% 1.3 Cr % 1.8 Mn % 1.3 Fe % 1.5 Co % 1.4 Ni % 1.8 Cu % 1.9 Zn % 2.9 Ga % 1.1 Ge % 1.8 As % 2.1 Sr % 1.4 Zr < DL 102% 1.4 Nb % 1.2 Mo < DL 98% 1.8 Ru < DL 96% 1.5 Rh % 1.2 Pd % 1.8 Ag % 1.7 Cd < DL 101% 1.5 In < DL 102% 1.6 Sn % 1.7 Sb % 1.6 Ba % 1.3 Ta < DL 103% 2.0 W < DL 100% 1.8 Pt % 1.9 Au % Tl < DL 104% 1.2 Pb % 1.5 Bi < DL 103% 1.8 U < DL 102% 1.9 *Cell gas used is NH 3. 4

5 Conclusion The NexION 300S ICP-MS has shown to be robust and suitable for the routine quantification of ultratrace impurities at the ng/l level in HNO 3. By means of computer-controlled switching between Standard mode and Reaction mode in the Universal Cell, interference-free analysis using hot plasma conditions for all analytes is possible during a single sample run. References 1. SEMI Standard C , SEMI Standards, semi.org/en/index.htm PerkinElmer, Inc. 940 Winter Street Waltham, MA USA P: (800) or (+1) For a complete listing of our global offices, visit Copyright , PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners A_01

How To Use The Nexion 300

How To Use The Nexion 300 APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of

More information

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu

More information

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth

More information

Universal Data Acquisition (UDA)

Universal Data Acquisition (UDA) Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte

More information

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development

More information

ELECTRON CONFIGURATION (SHORT FORM) # of electrons in the subshell. valence electrons Valence electrons have the largest value for "n"!

ELECTRON CONFIGURATION (SHORT FORM) # of electrons in the subshell. valence electrons Valence electrons have the largest value for n! 179 ELECTRON CONFIGURATION (SHORT FORM) - We can represent the electron configuration without drawing a diagram or writing down pages of quantum numbers every time. We write the "electron configuration".

More information

Sample Analysis Design Isotope Dilution

Sample Analysis Design Isotope Dilution Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with

More information

B I N G O B I N G O. Hf Cd Na Nb Lr. I Fl Fr Mo Si. Ho Bi Ce Eu Ac. Md Co P Pa Tc. Uut Rh K N. Sb At Md H. Bh Cm H Bi Es. Mo Uus Lu P F.

B I N G O B I N G O. Hf Cd Na Nb Lr. I Fl Fr Mo Si. Ho Bi Ce Eu Ac. Md Co P Pa Tc. Uut Rh K N. Sb At Md H. Bh Cm H Bi Es. Mo Uus Lu P F. Hf Cd Na Nb Lr Ho Bi Ce u Ac I Fl Fr Mo i Md Co P Pa Tc Uut Rh K N Dy Cl N Am b At Md H Y Bh Cm H Bi s Mo Uus Lu P F Cu Ar Ag Mg K Thomas Jefferson National Accelerator Facility - Office of cience ducation

More information

Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist

Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist Application of ICP-MS as a Tool for Environmental Monitoring Bill Spence Environmental Market Specialist Introduction Knowledge of the concentration of metallic species in the environment has been important

More information

VALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol

VALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol Page 1 van 36 pages VALIDATION REPORT Investigation of the major interferences for proposed Horizontal ICP AES method Identification code Validation in accordance with Reference method Scope Reference

More information

The 30-Minute Guide to ICP-MS

The 30-Minute Guide to ICP-MS Technical Note The 30-Minute Guide to ICP-MS ICP-Mass Spectrometry A Worthy Member of the Inorganic Analysis Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been gaining

More information

MILESTONE CLEAN CHEMISTRY LINE DUOPUR TRACECLEAN ULTRA-TRACE INSERTS. A Breakthrough in Reduction and Control of the Analytical Blank MILESTONE

MILESTONE CLEAN CHEMISTRY LINE DUOPUR TRACECLEAN ULTRA-TRACE INSERTS. A Breakthrough in Reduction and Control of the Analytical Blank MILESTONE MILESTONE CLEAN CHEMISTRY LINE A Breakthrough in Reduction and Control of the Analytical Blank DUOPUR TRACECLEAN ULTRA-TRACE INSERTS MILESTONE MILESTONE CLEAN CHEMISTRY LINE An innovative and complete

More information

Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT

Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.

More information

ICP - Mass Spectrometry APPLICATION NOTE. Authors. Introduction. Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA

ICP - Mass Spectrometry APPLICATION NOTE. Authors. Introduction. Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA APPLICATION NOTE ICP - Mass Spectrometry Authors Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA Benefits of the NexION 300X/350X ICP-MS Coupled with the prepfast In-line Auto-dilution/calibrationSystem

More information

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published

More information

aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS

aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS aurora M90 A new era in ICP-MS Innovation with Integrity ICP-MS Your Partner in ICP-MS Solutions Bruker continues to find new and novel ways to meet your changing needs. As a leader in elemental analysis

More information

NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight.

NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight. NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight. When your instrument is more efficient and productive, so too are your scientists and your lab. And nothing puts the flow in your

More information

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Why? Chemists are concerned with mass relationships in chemical reactions, usually run on a macroscopic scale (grams, kilograms, etc.). To deal with

More information

Electronegativity and Polarity

Electronegativity and Polarity and Polarity N Goalby Chemrevise.org Definition: is the relative tendency of an atom in a molecule to attract electrons in a covalent bond to itself. is measured on the Pauling scale (ranges from 0 to

More information

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers Key Questions & Exercises Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers 1. The atomic weight of carbon is 12.0107 u, so a mole of carbon has a mass of 12.0107 g. Why doesn t a mole of

More information

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,

More information

SUS. Company Profile. Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0)208/658535 Fax 0049(0)208/658536

SUS. Company Profile. Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0)208/658535 Fax 0049(0)208/658536 SUS Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0208/658535 Fax 0049(0208/658536 Company Profile SUS was founded in 1986 in Oberhausen in the Ruhr area (close to Düsseldorf, in order to meet

More information

OLIVÉR BÁNHIDI 1. Introduction

OLIVÉR BÁNHIDI 1. Introduction Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY

More information

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially

More information

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,

More information

CLASS TEST GRADE 11. PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change

CLASS TEST GRADE 11. PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change CLASS TEST GRADE PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change MARKS: 45 TIME: hour INSTRUCTIONS AND INFORMATION. Answer ALL the questions. 2. You may use non-programmable calculators. 3. You may

More information

Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy

Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy In support of the toy industry, many laboratories are being called upon to analyze a wide variety of

More information

CVD SILICON CARBIDE. CVD SILICON CARBIDE s attributes include:

CVD SILICON CARBIDE. CVD SILICON CARBIDE s attributes include: CVD SILICON CARBIDE CVD SILICON CARBIDE is the ideal performance material for design engineers. It outperforms conventional forms of silicon carbide, as well as other ceramics, quartz, and metals in chemical

More information

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES ICP OPTICAL ATOMIC EMISSION SPECTROSCOPY Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES HORIBA Scientific Longjumeau, France Keywords: metallurgy, non-ferrous, halogens 1 Introduction

More information

Wipe Analysis to Determine Metal Contamination on Critical Surfaces

Wipe Analysis to Determine Metal Contamination on Critical Surfaces By Albert Dato, Ph.D., Warren York, Jennifer Jew, Laarni Huerta, Brice Norton, and Michael Coste On-wafer metallic contamination is detrimental to the fabrication and performance of semiconductor devices.

More information

Sustainable energy products Simulation based design for recycling

Sustainable energy products Simulation based design for recycling Sustainable energy products Simulation based design for recycling Markus A. Reuter (Prof. Dr. Dr. hc) Director: Technology Management, Outotec Oyj Aalto University (Finland), Central South University (China),

More information

100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals.

100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals. 2.21 Ionic Bonding 100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals. Forming ions Metal atoms lose electrons to form +ve ions. Non-metal

More information

All answers must use the correct number of significant figures, and must show units!

All answers must use the correct number of significant figures, and must show units! CHEM 10113, Quiz 2 September 7, 2011 Name (please print) All answers must use the correct number of significant figures, and must show units! IA Periodic Table of the Elements VIIIA (1) (18) 1 2 1 H IIA

More information

EXPERIMENT 4 The Periodic Table - Atoms and Elements

EXPERIMENT 4 The Periodic Table - Atoms and Elements EXPERIMENT 4 The Periodic Table - Atoms and Elements INTRODUCTION Primary substances, called elements, build all the materials around you. There are more than 109 different elements known today. The elements

More information

Chromium Speciation in Drinking Water by LC-ICP-MS

Chromium Speciation in Drinking Water by LC-ICP-MS APPLICATION NOTE LC-ICP-MS Authors: Helmut Ernstberger Ken Neubauer PerkinElmer, Inc. Shelton, CT Chromium Speciation in Drinking Water by LC-ICP-MS Introduction With the growing concern of pollutants

More information

Standard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry

Standard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry Standard Operation Procedure Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu

More information

Elemental Analyses by ICP-AES

Elemental Analyses by ICP-AES Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to

More information

Using Direct Solid Sampling ICP-MS to Complement SEM- EDX and SIMS in Characterizing Semiconductor Materials

Using Direct Solid Sampling ICP-MS to Complement SEM- EDX and SIMS in Characterizing Semiconductor Materials Using Direct Solid Sampling ICP-MS to Complement SEM- EDX and SIMS in Characterizing Semiconductor Materials Fuhe Li and Scott Anderson Air Liquide America-Balazs Analytical Services, 4649 Landing Parkway,

More information

Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?

Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common? Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction

More information

Analyses on copper samples from Micans

Analyses on copper samples from Micans PDF rendering: DokumentID 1473479, Version 1., Status Godkänt, Sekretessklass Öppen Analyses on copper samples from Micans P. Berastegui, M. Hahlin, M. Ottosson, M. Korvela, Y. Andersson, R. Berger and

More information

B) atomic number C) both the solid and the liquid phase D) Au C) Sn, Si, C A) metal C) O, S, Se C) In D) tin D) methane D) bismuth B) Group 2 metal

B) atomic number C) both the solid and the liquid phase D) Au C) Sn, Si, C A) metal C) O, S, Se C) In D) tin D) methane D) bismuth B) Group 2 metal 1. The elements on the Periodic Table are arranged in order of increasing A) atomic mass B) atomic number C) molar mass D) oxidation number 2. Which list of elements consists of a metal, a metalloid, and

More information

Analytical Method. Determination of metals in workplace air by inductively coupled plasma mass spectrometry (ICP-MS) ANALYTICAL METHOD 362

Analytical Method. Determination of metals in workplace air by inductively coupled plasma mass spectrometry (ICP-MS) ANALYTICAL METHOD 362 Analytical Method Determination of metals in workplace air by inductively coupled plasma mass spectrometry (ICP-MS) ANALYTICAL METHOD 362 Applicability This method applies to the determination of several

More information

Finnigan ELEMENT2. High Performance High Resolution ICP-MS

Finnigan ELEMENT2. High Performance High Resolution ICP-MS m a s s s p e c t r o m e t r y Finnigan ELEMENT2 High Performance High Resolution ICP-MS Analyze Detect Measure Control Finnigan ELEMENT2 High Performance High Resolution ICP-MS Finnigan ELEMENT2 High

More information

Neutralization of Acid Mine Drainage Using Stabilized Flue Gas Desulfurization Material

Neutralization of Acid Mine Drainage Using Stabilized Flue Gas Desulfurization Material Neutralization of Acid Mine Drainage Using Stabilized Flue Gas Desulfurization Material W. Wolfe 1, C.-M. Cheng 1, R. Baker 1, T. Butalia 1, J. Massey-Norton 2 1 The Ohio State University, 2 American Electric

More information

Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray

Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray Hannu Teisala a, Mikko Tuominen a, Mikko Aromaa b, Jyrki M. Mäkelä b, Milena Stepien c, Jarkko J. Saarinen c, Martti Toivakka c

More information

Chapter 8 - Chemical Equations and Reactions

Chapter 8 - Chemical Equations and Reactions Chapter 8 - Chemical Equations and Reactions 8-1 Describing Chemical Reactions I. Introduction A. Reactants 1. Original substances entering into a chemical rxn B. Products 1. The resulting substances from

More information

Analytical Techniques for Trace Elemental Analyses on Wafer Surfaces for Monitoring and Controlling Contamination

Analytical Techniques for Trace Elemental Analyses on Wafer Surfaces for Monitoring and Controlling Contamination Analytical Techniques for Trace Elemental Analyses on Wafer Surfaces for Monitoring and Controlling Contamination J. Wang, M. K. Balazs Balazs Analytical Laboratory, Sunnyvale, CA 94089 P. Pianetta, K.

More information

Quartz Glass. Tubes and Rods

Quartz Glass. Tubes and Rods Quartz Glass Tubes and Rods PH 300, GE 214, QI PN GVB Solutions in Glass Schlackstrasse 3 52080 Aachen Germany +49-241/9108588 +49-241/9108589 E- info@g-v-b.de www.g-v-b.de Table of contents General Information

More information

Module 7 Wet and Dry Etching. Class Notes

Module 7 Wet and Dry Etching. Class Notes Module 7 Wet and Dry Etching Class Notes 1. Introduction Etching techniques are commonly used in the fabrication processes of semiconductor devices to remove selected layers for the purposes of pattern

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

From Quantum to Matter 2006

From Quantum to Matter 2006 From Quantum to Matter 006 Why such a course? Ronald Griessen Vrije Universiteit, Amsterdam AMOLF, May 4, 004 vrije Universiteit amsterdam Why study quantum mechanics? From Quantum to Matter: The main

More information

Nomenclature and Formulas of Ionic Compounds. Section I: Writing the Name from the Formula

Nomenclature and Formulas of Ionic Compounds. Section I: Writing the Name from the Formula Purpose: Theory: Nomenclature and Formulas of Ionic Compounds 1. To become familiar with the rules of chemical nomenclature, based on the classification of compounds. 2. To write the proper name of the

More information

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications

More information

Determination of elements by ICP-AES and ICP-MS

Determination of elements by ICP-AES and ICP-MS September 2003 HORIZONTAL - 19 Determination of elements by ICP-AES and ICP-MS Henk J van de Wiel National Institute of Public Health and the Environment (RIVM) Bilthoven, The Netherlands 1 HORIZONTAL

More information

NET IONIC EQUATIONS. A balanced chemical equation can describe all chemical reactions, an example of such an equation is:

NET IONIC EQUATIONS. A balanced chemical equation can describe all chemical reactions, an example of such an equation is: NET IONIC EQUATIONS A balanced chemical equation can describe all chemical reactions, an example of such an equation is: NaCl + AgNO 3 AgCl + NaNO 3 In this case, the simple formulas of the various reactants

More information

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

3. What would you predict for the intensity and binding energy for the 3p orbital for that of sulfur?

3. What would you predict for the intensity and binding energy for the 3p orbital for that of sulfur? PSI AP Chemistry Periodic Trends MC Review Name Periodic Law and the Quantum Model Use the PES spectrum of Phosphorus below to answer questions 1-3. 1. Which peak corresponds to the 1s orbital? (A) 1.06

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

Nuclear ZPE Tapping. Horace Heffner May 2007

Nuclear ZPE Tapping. Horace Heffner May 2007 ENERGY FROM UNCERTAINTY The uncertainty of momentum for a particle constrained by distance Δx is given, according to Heisenberg, by: Δmv = h/(2 π Δx) but since KE = (1/2) m v 2 = (1/(2 m) ) (Δmv) 2 ΔKE

More information

Elemental Scientific. Stability. Elemental Scientific

Elemental Scientific. Stability. Elemental Scientific Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dilution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast inline

More information

Agilent 720/730 Series ICP-OES

Agilent 720/730 Series ICP-OES Agilent 720/730 Series ICP-OES Flexible. Productive. Robust. Specifications Introduction Agilent 720/730 Series ICP-OES spectrometers are manufactured according to a quality management system certified

More information

Chapter 8 Atomic Electronic Configurations and Periodicity

Chapter 8 Atomic Electronic Configurations and Periodicity Chapter 8 Electron Configurations Page 1 Chapter 8 Atomic Electronic Configurations and Periodicity 8-1. Substances that are weakly attracted to a magnetic field but lose their magnetism when removed from

More information

Nomenclature of Ionic Compounds

Nomenclature of Ionic Compounds Nomenclature of Ionic Compounds Ionic compounds are composed of ions. An ion is an atom or molecule with an electrical charge. Monatomic ions are formed from single atoms that have gained or lost electrons.

More information

Unit 2 Matter and Chemical Change. Unit Test

Unit 2 Matter and Chemical Change. Unit Test Unit Test Student Name Class Section 1 Properties of Matter 1. Aluminum foam is used to create lighter, safer cars. The reason that a lighter car is a safer car is because aluminum foam is A. less rigid

More information

EXPERIMENT 7 Reaction Stoichiometry and Percent Yield

EXPERIMENT 7 Reaction Stoichiometry and Percent Yield EXPERIMENT 7 Reaction Stoichiometry and Percent Yield INTRODUCTION Stoichiometry calculations are about calculating the amounts of substances that react and form in a chemical reaction. The word stoichiometry

More information

Work Assignment 5-03. Prepared for: Office of Air Quality Planning and Standards U.S. Environmental Protection Agency Research Triangle Park, NC

Work Assignment 5-03. Prepared for: Office of Air Quality Planning and Standards U.S. Environmental Protection Agency Research Triangle Park, NC ERG No.: 0143.04.005 EPA Contract No.: 68-D-00-264 Standard Operating Procedure for the Determination of Metals In Ambient Particulate Matter Analyzed By Inductively Coupled Plasma/Mass Spectrometry (ICP/MS)

More information

Role of Hydrogen Bonding on Protein Secondary Structure Introduction

Role of Hydrogen Bonding on Protein Secondary Structure Introduction Role of Hydrogen Bonding on Protein Secondary Structure Introduction The function and chemical properties of proteins are determined by its three-dimensional structure. The final architecture of the protein

More information

METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY

METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY METHOD 6010B INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY 1.0 SCOPE AND APPLICATION 1.1 Inductively coupled plasma-atomic emission spectrometry (ICP-AES) determines trace elements, including

More information

Multicrystalline solar silicon production for development of photovoltaic industry

Multicrystalline solar silicon production for development of photovoltaic industry Multicrystalline solar silicon production for development of photovoltaic industry A.I. Nepomnyaschikh Institute of Geochemistry, Siberian Branch, Russian Academy of Sciences, E-mail: ainep@igc.irk.ru

More information

Find a pair of elements in the periodic table with atomic numbers less than 20 that are an exception to the original periodic law.

Find a pair of elements in the periodic table with atomic numbers less than 20 that are an exception to the original periodic law. Example Exercise 6.1 Periodic Law Find the two elements in the fifth row of the periodic table that violate the original periodic law proposed by Mendeleev. Mendeleev proposed that elements be arranged

More information

Assessing the Fate of Silver Nanoparticles in Surface Water using Single Particle ICP-MS

Assessing the Fate of Silver Nanoparticles in Surface Water using Single Particle ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Authors: Madjid Hadioui, Ph.D. Université de Montréal Kevin Wilkinson, Ph.D. Université de Montréal Chady Stephan, Ph.D. PerkinElmer, Inc. Assessing the Fate of

More information

It takes four quantum numbers to describe an electron. Additionally, every electron has a unique set of quantum numbers.

It takes four quantum numbers to describe an electron. Additionally, every electron has a unique set of quantum numbers. So, quantum mechanics does not define the path that the electron follows; rather, quantum mechanics works by determining the energy of the electron. Once the energy of an electron is known, the probability

More information

Introduction to VLSI Fabrication Technologies. Emanuele Baravelli

Introduction to VLSI Fabrication Technologies. Emanuele Baravelli Introduction to VLSI Fabrication Technologies Emanuele Baravelli 27/09/2005 Organization Materials Used in VLSI Fabrication VLSI Fabrication Technologies Overview of Fabrication Methods Device simulation

More information

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility Use of Absorption Mechanisms to Decrease Heavy Metal Mobility by Jianmin Wang Honglan Shi Joe G. Burken NUTC R304 A National University Transportation Center at Missouri University of Science and Technology

More information

EXPERIMENT 8: Activity Series (Single Displacement Reactions)

EXPERIMENT 8: Activity Series (Single Displacement Reactions) EPERIMENT 8: Activity Series (Single Displacement Reactions) PURPOSE a) Reactions of metals with acids and salt solutions b) Determine the activity of metals c) Write a balanced molecular equation, complete

More information

Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic Absorption Spectrometry

Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic Absorption Spectrometry application Note Atomic Absorption Authors Jijun Yao Renkang Yang Jianmin Chen PerkinElmer, Inc. Shelton, CT 06484 USA Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic

More information

The R&D Notebook. Ultrapure Water for Boron and Silica Sensitive Laboratory Applications

The R&D Notebook. Ultrapure Water for Boron and Silica Sensitive Laboratory Applications The R&D Notebook A publication of the Laboratory Division of Millipore RD3 Ultrapure for Boron and Silica Sensitive Laboratory Applications Ichiro Kano () and Daniel Darbouret (2) () Research & Development,

More information

8. Relax and do well.

8. Relax and do well. CHEM 1314 3:30 pm Section Exam II ohn II. Gelder October 16, 2002 Name TA's Name Lab Section INSTRUCTIONS: 1. This examination consists of a total of 8 different pages. The last three pages include a periodic

More information

APPENDIX B: EXERCISES

APPENDIX B: EXERCISES BUILDING CHEMISTRY LABORATORY SESSIONS APPENDIX B: EXERCISES Molecular mass, the mole, and mass percent Relative atomic and molecular mass Relative atomic mass (A r ) is a constant that expresses the ratio

More information

High Sensitivity ICP-MS: Overcome the problem of complex samples

High Sensitivity ICP-MS: Overcome the problem of complex samples High Sensitivity ICP-MS: Overcome the problem of complex samples Meike Hamester, Andrew Toms and René Chemnitzer Bruker Daltonics Berlin, Germany - Milton, Canada Historical Product Roadmap 2003 2005 2010

More information

ORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software

ORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software ORTEC DET-SW-UPG Latest Software Features Three Search Modes: Gamma/Neutron total count rate. SNM search mode. Sliding average "monitor" mode. (NEW) User choice of identification schemes: Classify mode

More information

1332 CHAPTER 18 Sample Questions

1332 CHAPTER 18 Sample Questions 1332 CHAPTER 18 Sample Questions Couple E 0 Couple E 0 Br 2 (l) + 2e 2Br (aq) +1.06 V AuCl 4 + 3e Au + 4Cl +1.00 V Ag + + e Ag +0.80 V Hg 2+ 2 + 2e 2 Hg +0.79 V Fe 3+ (aq) + e Fe 2+ (aq) +0.77 V Cu 2+

More information

Flow Injection Analysis

Flow Injection Analysis Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since

More information

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation

More information

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

Implementation Of High-k/Metal Gates In High-Volume Manufacturing

Implementation Of High-k/Metal Gates In High-Volume Manufacturing White Paper Implementation Of High-k/Metal Gates In High-Volume Manufacturing INTRODUCTION There have been significant breakthroughs in IC technology in the past decade. The upper interconnect layers of

More information

Sample Exercise 2.1 Illustrating the Size of an Atom

Sample Exercise 2.1 Illustrating the Size of an Atom Sample Exercise 2.1 Illustrating the Size of an Atom The diameter of a US penny is 19 mm. The diameter of a silver atom, by comparison, is only 2.88 Å. How many silver atoms could be arranged side by side

More information

Table 1. Toxicology of Arsenic Species. Arsenic Species

Table 1. Toxicology of Arsenic Species. Arsenic Species www.perkinelmer.com Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Speciation of Five Arsenic Compounds in Urine by HPLC/ICP-MS Introduction High-performance

More information

III. Wet and Dry Etching

III. Wet and Dry Etching III. Wet and Dry Etching Method Environment and Equipment Advantage Disadvantage Directionality Wet Chemical Solutions Atmosphere, Bath 1) Low cost, easy to implement 2) High etching rate 3) Good selectivity

More information

MODERN ATOMIC THEORY AND THE PERIODIC TABLE

MODERN ATOMIC THEORY AND THE PERIODIC TABLE CHAPTER 10 MODERN ATOMIC THEORY AND THE PERIODIC TABLE SOLUTIONS TO REVIEW QUESTIONS 1. Wavelength is defined as the distance between consecutive peaks in a wave. It is generally symbolized by the Greek

More information

Electronegativity and Polarity MAIN Idea A chemical bond s character is related to each atom s

Electronegativity and Polarity MAIN Idea A chemical bond s character is related to each atom s Section 8.5 Objectives Describe how electronegativity is used to determine bond type. Compare and contrast polar and nonpolar covalent bonds and polar and nonpolar molecules. Generalize about the characteristics

More information

CATALOGUE REFERENCE MATERIALS

CATALOGUE REFERENCE MATERIALS IPT Institute for Technological Research CATALOGUE REFERENCE MATERIALS 2014 Laboratory of Metrological References IPT s Reference Materials www.ipt.br/nmr.htm Av. Prof. Almeida Prado n0 532 Predio 31 Cidade

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

Elemental Analysis Manual for Food and Related Products

Elemental Analysis Manual for Food and Related Products Elemental Analysis Manual for Food and Related Products 4.7 Inductively Coupled Plasma-Mass Spectrometric Determination of Arsenic, Cadmium, Chromium, Lead, Mercury, and Other Elements in Food Using Microwave

More information

Secondary Ion Mass Spectrometry

Secondary Ion Mass Spectrometry Secondary Ion Mass Spectrometry A PRACTICAL HANDBOOK FOR DEPTH PROFILING AND BULK IMPURITY ANALYSIS R. G. Wilson Hughes Research Laboratories Malibu, California F. A. Stevie AT&T Bell Laboratories Allentown,

More information

Electrochemistry - ANSWERS

Electrochemistry - ANSWERS Electrochemistry - ANSWERS 1. Using a table of standard electrode potentials, predict if the following reactions will occur spontaneously as written. a) Al 3+ + Ni Ni 2+ + Al Al 3+ + 3e - Al E = -1.68

More information

Sampling and preparation of heterogeneous waste fuels?

Sampling and preparation of heterogeneous waste fuels? Sampling and preparation of heterogeneous waste fuels? Is it possible to accomplish a representative and relevant composition data? Evalena Wikström, Lennart Gustavsson and Jolanta Franke Aim of the project

More information

Etching Etch Definitions Isotropic Etching: same in all direction Anisotropic Etching: direction sensitive Selectivity: etch rate difference between

Etching Etch Definitions Isotropic Etching: same in all direction Anisotropic Etching: direction sensitive Selectivity: etch rate difference between Etching Etch Definitions Isotropic Etching: same in all direction Anisotropic Etching: direction sensitive Selectivity: etch rate difference between 2 materials Other layers below one being etch Masking

More information

Contamination. Cleanroom. Cleanroom for micro and nano fabrication. Particle Contamination and Yield in Semiconductors.

Contamination. Cleanroom. Cleanroom for micro and nano fabrication. Particle Contamination and Yield in Semiconductors. Fe Particles Metallic contaminants Organic contaminants Surface roughness Au Particles SiO 2 or other thin films Contamination Na Cu Photoresist Interconnect Metal N, P Damages: Oxide breakdown, metal

More information