1 EUROPEAN COMMISSION DG Internal Market, Industry, Entrepreneurship and SMEs Directorate Consumer, Environmental and Health Technologies Unit Health technology and Cosmetics MEDICAL DEVICES: Guidance document - Qualification and Classification of stand alone software MEDDEV 2.1/6 July 2016 GUIDELINES ON THE QUALIFICATION AND CLASSIFICATION OF STAND ALONE SOFTWARE USED IN HEALTHCARE WITHIN THE REGULATORY FRAMEWORK OF MEDICAL DEVICES Foreword The present guidelines are part of a set of guidelines relating to questions of application of the EU legislation on medical devices. They are legally not binding. The guidelines have been carefully drafted through a process of consultation of the various interested parties (Competent Authorities, Commission services, industry and Notified Bodies in the medical device sector) during which intermediate drafts were circulated and comments were taken up in the document where appropriate. Therefore this document reflects positions taken in particular by the aforementioned interested parties. Due to the participation of the aforementioned interested parties, it is anticipated that these guidelines will be followed within the Member States and, therefore, ensure uniform application of relevant Directive provisions.

2 Content Introduction Definitions and abbreviations Qualification Introduction to criteria for qualification Qualification criteria as medical device Qualification criteria as IVD medical device Classification of stand alone software Software as active therapeutic medical devices Software intended for diagnosis or therapy IVDD and software in conjunction with in vitro diagnostic devices Modules Annex 1: Illustrative examples of qualification and classification for software used in the healthcare environment a) Hospital Information Systems b) Decision Support Software c) Information Systems c.1) Electronic Patient Records Systems c.1.1) Clinical Information Systems CIS/Patient Data Management Systems PDMS c.1.2) Pre-hospital Electrocardiograph (ECG) System c.1.3) Radiological Information System (RIS) c1.4) Picture Archive Communication System (PACS) d) Communication Systems

3 d.1) Telemedicine Systems d.1.1) Telesurgery d.1.2) Video appointment software d.1.3) Home care monitoring, wired or mobile e) Web systems for monitoring of data f) In vitro diagnostic (IVD) software: LIS & WAM f.1) Laboratory Information Systems (LIS) and Work Area Managers (WAM)...24 f.2) Expert system 25 f.3) Interpretation of raw data..26 f.4) Home care monitoring, wired or mobile 26 2

4 Introduction The purpose of this document is to define the criteria for the qualification of stand alone software, when used in healthcare setting, as a medical device and the application of the classification criteria to such software. This document only deals with stand alone software and provides some illustrative examples. Software incorporated in medical devices is outside the scope of this guideline. The criteria specified in this document apply also to mobile applications. Directive 2007/47/EC 1 amended the definition of the term "medical device" used in Directives 90/385/EEC 2 and 93/42/EEC 3. Recital 6 of Directive 2007/47/EC states that "it is necessary to clarify that software in its own right, when specifically intended by the manufacturer to be used for one or more of the medical purposes set out in the definition of a medical device, is a medical device. Stand alone software for general purposes when used in a healthcare setting is not a medical device." Stand alone software shall be qualified as an in vitro diagnostic (IVD) medical device or as an accessory to an IVD provided that it satisfies the definition of an IVD or of an accessory to an IVD as set out in Directive 98/79/EC 4. 1 OJ L 247, , p OJ L 189, , p OJ L 169, , p. 1 3

5 1. Definitions and abbreviations o Intended purpose: intended purpose means the use for which the device is intended according to the data supplied by the manufacturer on the labelling, in the instructions and/or in promotional materials. 5 o Accessory: accessory means an article which whilst not being a device is intended specifically by its manufacturer to be used together with a device to enable it to be used in accordance with the use of the device intended by the manufacturer of the device 6. o Placing on the market: placing on the market means the first making available in return for payment or free of charge of a device other than a device intended for clinical investigation, with a view to distribution and/or use on the Community market, regardless of whether it is new or fully refurbished 7. o Putting into service: putting into service means the stage at which a device has been made available to the final user as being ready for use on the Community market for the first time for its intended purpose 8. o Medical Device: 4 OJ L 331, , p. 1 5 Article 1(2)g of Directive 93/42/EEC 6 Article 1(2)b of Directive 93/42/EEC 7 Article 1(2)h of Directive 93/42/EEC 8 Article 1(2)i of Directive 93/42/EEC 4

6 medical device means any instrument, apparatus, appliance, software, material or other article, whether used alone or in combination, including the software intended by its manufacturer to be used specifically for diagnostic and/or therapeutic purposes and necessary for its proper application, intended by the manufacturer to be used for human beings for the purpose of: diagnosis, prevention, monitoring, treatment or alleviation of disease, diagnosis, monitoring, treatment, alleviation of or compensation for an injury or handicap, investigation, replacement or modification of the anatomy or of a physiological process, control of conception, and which does not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic means, but which may be assisted in its function by such means 9 ; o In Vitro diagnostic medical device: in vitro diagnostic medical device means any medical device which is a reagent, reagent product, calibrator, control material, kit, instrument, apparatus, equipment, or system, whether used alone or in combination, intended by the manufacturer to be used in vitro for the examination of specimens, including blood and tissue donations, derived from the human body, solely or principally for the purpose of providing information: concerning a physiological or pathological state, or concerning a congenital abnormality, or to determine the safety and compatibility with potential recipients, or to monitor therapeutic measures 10. o Active medical device: any medical device operation of which depends on a source of electrical energy or any source of power other than that directly generated by the human body or gravity and which acts by converting this energy. Medical devices intended to transmit energy, substances or other elements between an active medical device and the patient without any significant change, are not considered to be active medical devices. Stand alone software is considered to be an active medical device 11. o Active implantable medical device: 9 Article 1(2)a of Directive 93/42/EEC 10 Article 1(2)b of Directive 98/79/EC 5

7 active implantable medical device means any active medical device which is intended to be totally or partially introduced, surgically or medically, into the human body or by medical intervention into a natural orifice, and which is intended to remain after the procedure 12. o Active therapeutical device: Any active medical device, whether used alone or in combination with other medical devices, to support, modify, replace or restore biological functions or structures with a view to treatment or alleviation of an illness, injury or handicap 13. o Active device for diagnosis: Any active medical device, whether used alone or in combination with other medical devices, to supply information for detecting, diagnosing, monitoring or treating physiological conditions, states of health, illnesses or congenital deformities 14. o Input data: Any data provided to software in order to obtain output data after computation of this data can be considered as an input data. Input data examples (non-exhaustive): Data given through the use of a human data-input device such as a keyboard, mouse, stylus, or touch screen; Data given through speech recognition; Digital document: formatted for general purpose such as Word file or pdf file or jpeg image, formatted for medical purpose such as DICOM file or ECG records or Electronic Health Record, unformatted document. Note that digital documents have to be differentiated from software able to read such documents; Data received from / transmitted by devices. 11 Annex IX, section 1.4, of Directive 93/42/EEC 12 Article 1(2)c of Directive 90/385/EEC 13 Annex IX, section 1.5, of Directive 93/42/EEC 14 Annex IX, section 1.6, of Directive 93/42/EEC 6

8 o Output data: Any data produced by a software can be considered as an output data. Output data examples (non-exhaustive): Screen display data (such as layout with number, characters, picture, graphics etc.); Print data (such as layout with number, characters, picture, graphics etc.); Audio data; Digital document (formatted for a general purpose such as Word file or pdf file or jpeg image, or formatted for medical purpose such as DICOM file or ECG records or Electronic Health Record, unformatted document). Haptic buzzing as an alternative to audio sound o Software: For the purpose of this guideline, software is defined as a set of instructions that processes input data and creates output data. o Stand alone software: For the purpose of this guideline stand alone software means software which is not incorporated in a medical device at the time of its placing on the market or its making available. o Expert function software: For the purpose of this document, the expert function software means software which is able to analyse existing information to generate new specific information according to the intended use of the software. o Software as a medical device: The term Software as a Medical Device (SaMD) is defined as software intended to be used for one or more medical purposes that perform these purposes without being part of a hardware medical device. Definition of the IMDRF/SaMD WG/N10FINAL:2013 : Software as a Medical Device (SaMD): Key Definitions. 7

9 2. Qualification 2.1 Introduction to criteria for qualification The purpose of this part of the guidance is to clarify the distinction between different types of medical software, namely medical software that is: - part of MD or IVD - accessories - standalone Software - not a Medical Device Stand alone software must have a medical purpose to be qualified as medical device. It should be noted that only the intended purpose as described by the manufacturer of the product is relevant for the qualification and classification of any device and not by virtue of the way it may be called. Stand alone software that does not meet the definition of a medical device or of an IVD medical device but is intended by the manufacturer to be an accessory to a medical device, or an IVD medical device, falls respectively under the scope of Directive 93/42/EEC or Directive 98/79/EC. It is to be noted that to be qualified as an IVD medical device, stand alone software must first fulfil the definition of a medical device. Where a given product does not fall under the definition of medical device, or is excluded by the scope of the Directives, other Community and/or national legislation may be applicable Qualification criteria as medical device Software can be used for a large variety of medical purposes 15. In that respect the arguments do not differ from those used for other medical devices. Stand alone software can directly control an apparatus (e.g. radiotherapy treatment), can provide immediate decision triggering information (e.g. blood glucose meters), or can provide support for healthcare professionals (e.g. ECG interpretation). Not all stand alone software used within healthcare can be qualified as a medical device. 15 MEDDEV 2.1/1: Definitions of "medical devices", "accessory" and "manufacturer" 8

10 Stand alone software may run on different operating systems or in virtual environments. These operating systems or virtual environments do not impact the qualification criteria. Stand alone software might also be an accessory of a medical device. The risk related to a malfunction of the stand alone software used within healthcare is in itself not a criterion for its qualification or not as a medical device. It is, therefore, necessary to clarify some criteria for the qualification of stand alone software as medical devices. The decision diagram (Figure 1) gives some guidance regarding the necessary steps to qualify stand alone software as medical device. 9

11 Figure 1: A decision diagram to assist qualification of software as medical device. 9

12 Legend of the flowchart: MD: Medical Device * Medical Device Directives means the 3 following directives: o DIRECTIVE 93/42/EEC of 14 June 1993 concerning medical devices o DIRECTIVE 98/79/EC of 27 October 1998 on in vitro diagnostic medical devices o DIRECTIVE 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member States relating to active implantable medical devices 10

13 Decision step 1: If the product is a software according to the definition of this document, then it may be a medical device; if the product is not a software according to the definition of this document, then it is not covered by the medical device directives.. Decision step 2: if the software is incorporated into a medical device rather than stand alone software, it must be considered as part of that medical device in the regulatory process of that device. If it is stand alone software 16, proceed to decision step 3. Decision step 3: if the software does not perform an action on data, or performs an action limited to storage, archival, communication 17, simple search or lossless compression (i.e. using a compression procedure that allows the exact reconstruction of the original data) it is not a medical device. Altering the representation of data for embellishment purposes does not make the software a medical device. In other cases, including where the software alters the representation of data for a medical purpose, it could be a medical device. Simple search refers to the retrieval of records by matching record metadata against record search criteria, e.g. library functions. Simple search does not include software which provides interpretative search results, e.g. to identify medical findings in health records or on medical images. Software which is intended to create or modify medical information might be qualified as a medical device. If such alterations are made to facilitate the perceptual and/or interpretative tasks performed by the healthcare professionals when reviewing medical information, (e.g. when searching the image for findings that support a clinical hypothesis as to the diagnosis or evolution of therapy) the software could be a medical device. Note: the display of images usually involves alterations to the representation because techniques are used such as contrast stretching, edge enhancement, gray scale manipulation, smoothing, sharpening, zooming and re-sizing. Alterations may include reconstruction, lossy compression, filtering, pattern recognition, modelling, interpolation, transformation, classification (e.g. scoring of tumors against specific criteria), segmentation, registration (e.g. mapping a data set to a model or atlas or to another data 16 See chapter 2 - Definitions and Abbreviations for a definition of stand alone software. 17 Communication: The flow of information from one point, known as the source, to another, the receiver; Source: IEEE

14 set, e.g. registering an MRI image on a CT image), calculations, quantification, qualification (e.g. comparison of data against references), rendering, visualisation, interpretation, etc.. Decision step 4: an example of software for the benefit of individual patients is software intended to be used for the evaluation of patient data to support or influence the medical care provided to that patient. Examples of software which are not considered as being for the benefit of individual patients are those which aggregate population data, provide generic diagnostic or treatment pathways, scientific literature, medical atlases, models and templates as well as software for epidemiologic studies or registers. Decision step 5: if the manufacturer specifically intends the software to be used for any of the purposes listed in Article 1(2)a of Directive 93/42/EEC, then the software shall be qualified as a medical device. However, if only a non-medical purpose is intended by the manufacturer, such as invoicing or staff planning, it is not a medical device. Note: A task such as ing, web or voice messaging, data parsing, word processing, and back-up is by itself not considered as being a medical purpose, according to Directive 93/42/EEC. Decision step 6: if the software is an accessory to a medical device, it is not a medical device, but it falls under Directive 93/42/EEC. The legal definition of 'putting into service' requires that a device is made available to the final user/operator as being ready for use on the Community market. Software made available to the user over the internet (directly or via download) or via in vitro diagnostic commercial services, which is qualified as a medical device, is subject to the medical devices directives Qualification criteria as IVD medical device Stand alone software fulfilling the definition of medical device and intended to be used for the purpose of providing information derived from in vitro examination of a specimen derived from the human body falls under Directive 98/79/EC. Provided that stand alone software is intended specifically by its manufacturer to be used together with an IVD medical device to enable that device to be used in accordance with its intended purpose, this stand alone falls under the scope of the IVD Directive and shall be treated as an IVD device in its own right. 12

15 Example: analysis and interpretation of the optical density delivered by an ELISA reader, line or spot pattern of a blot. 13

16 Figure 2: Decision diagram to assist qualification of stand alone software as IVD device. 13

17 Decision step 1: please, start by making reference to figure 1. Decision step 2: Stand alone software with expert function: This stand alone software might provide information on e.g. differential diagnosis, prediction of the risks of developing a disease, prediction of the percentage of efficiency or failure (e.g. of a treatment), identifying species of bacteria. It may qualify differently based on the fact that the existing information is obtained from either IVD medical devices, other medical devices or both. Decision step 3: if the information provided by the software is based on data obtained from IVD medical devices only, the software is an IVD medical device or an accessory of an IVD medical device. If data are obtained from both IVD medical devices and from medical devices are analysed together for the purpose of providing information according to the definition of an IVD medical device, this software is an IVD medical device (e.g. evaluation of the risk of Trisomy 21). As described in Annex I B 3.1, for stand alone software that is intended for use in combination with other devices or equipment, the whole combination must be safe and must not impair the specified performances of the devices. As a concretization of these general requirements of the IVDD; the clinical evidence of this combined intended use of medical devices shall be performed for each of the medical devices from which the data is obtained. Decision step 4: if the information provided by the software is based on data obtained from medical devices only, the software would be a medical device. Note 1: Stand alone software that only collects results obtained from one or several IVD devices (directly and/or manually), and transmits without modification this information to a centralised database (e.g. Laboratory Information Management System, LIMS) or to healthcare providers is not an IVD medical device. 14

18 Note 2: Stand alone software (e.g. LIMS) managing the feed-back to an IVD (e.g. retesting of samples) based on collected IVD results is not an IVD medical device as per decision step 5 of figure 1. Note 3: Software intended to modify the representation of available IVD results is not considered as an IVD medical device as per decision step 5 of Figure 1, e.g. basic operations of arithmetic (e.g. mean, conversion of units) and/or plotting of results in function of time, and/or a comparison of the result to the limits of acceptance set by the user. Note 4: Stand alone software intended for archiving patient results or for transferring results from the home environment to the healthcare provider is not an IVD device. 3. Classification of stand alone software Stand alone software that meets the definition of a medical device shall be considered as an active medical device. 18 This means that rules 9, 10, 11 and 12 of Annex IX to Directive 93/42/EEC may apply. Clause 2.3 of the implementing rules in Annex IX states that software 'which drives a medical device or influences the use of a device, falls automatically into the same class 19, as the device it drives. 3.1 Software as active therapeutic medical devices According to rule 9 of Annex IX to Directive 93/42/EEC, all active therapeutic devices intended to administer or exchange energy are in Class IIa unless their characteristics are such that they may administer or exchange energy to or from the human body in a potentially hazardous way, taking account of the nature, the density and site of application of the energy, in which case they are in Class IIb. 18 Annex IX, section 1.4, of Directive 93/42/EEC 19 Annex IX, section 2.3, of Directive 93/42/EEC 15

19 All active devices intended to control or monitor the performance of active therapeutic devices in Class IIb, or intended directly to influence the performance of such devices are in Class IIb according to implementing rule 2.3. Example: radiotherapy planning system used to calculate the dose of ionizing radiation to be administered to the patient, insulin dosage planning stand alone software Software intended for diagnosis or therapy According to rule 10 of Annex IX to Directive 93/42/EEC, active devices intended for diagnosis are in Class IIa: - if they are intended to image in vivo distribution of radiopharmaceuticals. Example: Clinical application of registration of PET datasets on CT datasets for follow-up tumour treatment. - if they are intended to allow direct diagnosis or monitoring of vital physiological processes, unless they are specifically intended for monitoring of vital physiological parameters, where the nature of variations is such that it could result in immediate danger to the patient, for instance variations in cardiac performance, respiration, activity of CNS in which case they are in Class IIb. Examples: - Software for the presentation of the heart rate or other physiological parameters during routine checkups (Class IIa); - Software for the presentation of the heart rate or other physiological parameters for intensive care monitoring (Class IIb). Active devices intended to emit ionizing radiation and intended for diagnostic and therapeutic interventional radiology including devices which control or monitor such devices, or which directly influence their performance, are in Class IIb. Rule 11 of Annex IX to Directive 93/42/EEC states that all active devices intended to administer and/or remove medicines, body liquids or other substances to or from the body are in Class IIa, unless this is done in a manner: 16

20 - that is potentially hazardous, taking account of the nature of the substances involved, of the part of the body concerned and of the mode of application in which case they are in Class IIb. Stand alone software which drives such medical device or influences the use of such device falls automatically into the same class as the device it drives (implementing rule 2.3). Rule 12 of Annex IX to Directive 93/42/EEC states that all other active devices are in Class I. Class I stand alone software can also come with a measuring function. Example: orthopaedic planning software to measure interpedicular distance or sagittal diameter of the spinal canal. 3.2 IVDD and software in conjunction with in-vitro diagnostic devices Stand alone software qualified as in vitro diagnostic medical devices should be regulated according to Directive 98/79/EC. When intended for evaluating the risk of trisomy 21, such software are specifically mentioned in Annex II List B of Directive 98/79/EC and shall therefore follow the conformity assessment procedure described in Article 9 of this Directive. 4. Modules Some stand alone software may break down into a significant number of applications for the user where each of these applications is correlated with a module. Some of these modules have a medical purpose, some not. Such software may be intended to cover many needs, e.g.: - Collect and maintain administrative patient details; - Keep on file the medical history of the patient; 17

21 - Invoicing and other accounting functions; - Provide a link to the social security system for reimbursement; - Provide a link to drug prescription systems (with possible link to drug dispensing outlets); - Provide expert system assistance for medical decision making (e.g. radiotherapy dose planner). This raises the issue as to whether the whole product can be CE marked when not all applications have a medical purpose. Computer programmes used in healthcare mostly have applications which consist of both medical device and non-medical device modules. The modules which are subject to the medical device Directives (Figures 1 and 2) must comply with the requirements of the medical device Directives and must carry the CE marking. The non-medical device modules are not subject to the requirements for medical devices. It is the obligation of the manufacturer to identify the boundaries and the interfaces of the different modules. The boundaries of the modules which are subject to the medical device Directives should be clearly identified by the manufacturer and based on the intended use. If the modules which are subject to the medical device Directives are intended for use in combination with other modules of the whole software structure, other devices or equipment, the whole combination, including the connection system, must be safe and must not impair the specified performances of the modules which are subject to the medical device Directives i.e. essential requirements 3.1 of Directive 98/79/EC and essential requirements 9.1 of Directive 93/42/EEC 18

22 Annex 1: Illustrative examples of qualification for software used in the healthcare environment The sector of software pursuing a medical purpose is rapidly evolving. The list of examples provided below is not exhaustive. The examples have been drafted in the light of today's state of the art, in order to give the reader a better understanding of the application of the principles set out in the guideline. In the light of the technological progress, further examples will be regularly added into the Manual on borderline and classification in the Community regulatory framework for medical devices 21. a) Hospital Information Systems Hospital Information Systems mean, in this context, systems that support the process of patient management. Typically they are intended for patient admission, for scheduling patient appointments, for insurance and billing purposes. These Hospital Information Systems are not qualified as medical devices. However they may be used with additional modules, as described hereafter. These modules might be qualified in their own right as medical devices. b) Decision Support Software

23 In general, they are computer based tools which combine medical knowledge databases and algorithms with patient specific data. They are intended to provide healthcare professionals and/or users with recommendations for diagnosis, prognosis, monitoring and treatment of individual patients. Based on steps 3, 4, and 5 of Figure 1, they are qualified as medical devices. - Radiotherapy treatment planning systems 22 are intended to calculate the dosage of ionizing irradiation to be applied to a specific patient. They are considered to control, monitor or directly influence the source of ionizing radiation and are qualified as medical devices. - Drug (e.g.: Chemotherapy) planning systems are intended to calculate the drug dosage to be administered to a specific patient and therefore are qualified as medical devices. - Computer Aided Detection systems are intended to provide information that may suggest or exclude medical conditions and, therefore, qualified as medical devices. For example, such systems would be able to automatically read x-ray images or interpret ECGs. c) Information Systems Information systems that are intended only to store, archive and transfer data are not qualified as medical devices in themselves. However they may be used with additional modules. These modules might be qualified in their own right as medical devices. c.1) Electronic Patient Record Systems Electronic patient record systems are intended to store and transfer electronic patient records. They archive all kinds of documents and data related to a specific patient. The electronic patient records themselves are not computer programs, therefore, they should not be 22 See EN «Requirements for the safety of radiotherapy treatment planning systems» 20

24 qualified as a medical device i.e. an electronic patient record that simply replaces a patient s paper file does not meet the definition of a medical device. The modules used with electronic patient record system modules that might be qualified in their own right as medical devices are for example: - an image viewer with functionality for diagnosis based on digital images; - a medication module. c.1.1) Clinical Information Systems CIS / Patient Data Management Systems - PDMS A CIS/PDMS is a software based system primarily intended for e.g. intensive care units to store and transfer patient information generated in association with the patient s intensive care treatment. Usually the system contains information such as patient identification, vital intensive care parameters and other documented clinical observations. These CIS/PDMS are not qualified as medical devices. Modules that are intended to provide additional information that contributes to diagnosis, therapy and follow-up (e.g. generate alarms) are qualified as medical devices. c.1.2) Pre-hospital Electrocardiograph (ECG) System A system for managing pre-hospital ECG is a software based system intended for ambulance services to store and transfer information from patients, which are connected to an ECG monitor, to a doctor at remote location. Usually the system contains information about patient identification, vital parameters and other documented clinical observations. These Pre-hospital Electrocardiograph (ECG) Systems are not qualified as medical devices. Modules that provide patient's treatment information to the paramedics in the ambulance to start the patient s treatment while the patient is being transported are qualified as Medical Devices. c.1.3) Radiological Information System (RIS) 21

25 A RIS is a software based database used in radiology departments to store and transfer radiological images and patient information. The system normally includes functions for patient identification, scheduling, examination results and imaging identification details. These Radiological Information Systems are not qualified as medical devices. However, if such a system includes additional modules they might qualify as medical devices according to step 3 of of Figure 1. c.1.4) Picture Archive Communication System (PACS) The Manual on Borderline and Classification in the Community Regulatory Framework for Medical Devices 23, already addresses the issue. d) Communication Systems The healthcare sector uses communication systems (e.g. systems, mobile telecommunication systems, video communication systems, paging etc.) to transfer electronic information. Different types of messages are being sent such as prescriptions, referrals, images, patient records, etc. Most of the communication systems handle other types of messages other than medical information. This communication system is intended for general purposes, and is used for transferring both medical and non-medical information. Communication systems are normally based on software for general purposes, and do not fall within the definition of a medical device. Communication system modules might be used with other modules that might be qualified in their own right as medical devices. Example: software generating alarms based on the monitoring and analysis of patient specific physiological parameters. d.1) Telemedicine Systems

26 Telemedicine Systems are intended to allow monitoring and/or delivery of healthcare to patients at locations remote from where the healthcare professional is located. d.1.1) Telesurgery Telesurgery is intended to conduct a surgical procedure from a remote location. Virtual reality technology may be used to support a remote surgeon to control a surgical robot performing the surgical procedure. Telesurgery systems should be qualified as medical devices according to steps 3, 4, and 5 of Figure 1. Remote control software used in combination with telesurgery robots is qualified as a medical device. Communication modules themselves are not medical devices. Other modules that are intended to influence the surgery procedure are qualified as medical devices. d.1.2) Video appointment software Video appointment software is intended to perform remote consultations between clinics and patients. It does not fall under the Medical Devices Directives. d.1.3) Home care monitoring, wired or mobile The telecommunication system (mobile, wireless, wire, etc ) is not as such a medical device. e) Web systems for monitoring of data A web system for the monitoring of clinical data typically interacts with a medical device (e.g. implanted devices or homecare devices), and uses a transmitter to send the information over the internet, a landline telephone or a mobile network. The information is collected and stored on a web server usually run by an external party who is generally the manufacturer of the system. The information can be reached by authorized health professionals or the patient through an internet connection. 23

27 - Monitoring of performance of medical devices: Modules that are intended to monitor the medical performance of medical devices fall under the Medical Devices Directives. This includes the clinical performance and failures that could affect medical performance of the device. One example of such product is a web system for monitoring of active implants such as pacemakers or Intra cardiac defibrillators (ICDs). - Monitoring of non medical performance of medical devices: Modules that are intended to monitor non medical performance of medical devices do not fall under the scope of Medical Devices Directives. Example: software for the monitoring of medical devices in hospital systems for the purpose of maintenance and repair. Software modules on server(s) might be qualified in their own right as medical devices depending on their intended purpose. f) In vitro diagnostic (IVD) software: LIS & WAM f.1) Laboratory Information Systems (LIS) and Work Area Managers (WAM) Laboratory Information Systems (LIS) and Work Area Managers (WAM) mean, in this context, systems that support the process from patient sample to patient result. Typically they have pre-analytical functions for ordering, sorting and distribution of test samples. The main task is the management and validation of incoming information obtained from IVD analysers connected to the system, such as calibration, quality control, product expiry and feedback (e.g. retesting of samples needed) through interconnections with various analytical instruments (technical and clinical validation). Finally the post analytical process takes care of communication of laboratory results, statistics and optional reporting to external databases. The software normally supports the following functions: Ordering of laboratory tests, samples with labels and sorting; Technical and clinical validation, connection to analytic instruments; 24

28 Laboratory results and reports on paper, fax or electronic records that can be directly returned to e.g. the ordering clinic's patient record; Analytical instruments can be interfaced with Hospital Information Systems (HIS), Electronic Patient Record Systems, Infectious control databases etc. Note: Software intended to modify the representation of available IVD results is not considered an IVD medical device, e.g. basic operations of arithmetic (e.g. mean, conversion of units) and/or plotting of results in function of time, and/or a comparison of the result to the limits of acceptance set by the user. The results are available, readable and understandable without the intervention of the software. Laboratory Information Systems (LIS) and Work Area Managers (WAM) are not qualified as medical devices in themselves. However they may be used with additional modules. These modules might be qualified in their own right as medical devices. f.2) Expert system Where software is intended to capture and analyze together several results obtained for one patient by one or more IVD devices by means of in vitro examination of body samples (possibly combined with information from medical devices) to provide information falling within the definition of an IVD medical device, e.g. differential diagnosis, this software is considered as an IVD in itself. Examples: - software that integrates genotype of multiple genes to predict risk of developing a disease or medical condition; - software that uses an algorithm to characterize viral resistances to various drugs, based on a nucleotide sequence generated by genotyping assays. This software serves to generate new information (virus resistance profile) from available information on the genotype of the virus; - software intended to be used in microbiology for the identification of clinical isolates and/or the detection antimicrobial resistances. The information provided by the software is based on data obtained with IVD medical devices only or possibly combined with 25

29 information from medical devices. The software is an IVD medical device or an accessory of an IVD medical device. f.3) Interpretation of raw data In the case where software is necessary to render raw data, obtained from an IVD by means of in vitro examination of body samples, readable for the user, this software is to be considered as an accessory to an IVD when it is specifically intended to be used together with this IVD to enable it to be used in accordance with its intended purpose. Example: software intended for the analysis and interpretation of ELISA reader optical density results, line patterns or spot patterns of a blot. f.4) Home care monitoring, wired or mobile Stand alone software intended for archiving patient results or for transferring results from the home environment to the healthcare provider is not an IVD device. The results are available, readable and understandable by the user without the intervention of the software. 26

How To Know If A Mobile App Is A Medical Device

How To Know If A Mobile App Is A Medical Device The Regulation of Medical Device Apps Prepared for West of England Academic Health Science Network and University of Bristol June 2014 June 2014 1 Table of Contents 1 Purpose...3 2 Scope...3 3 The Regulation

More information

Software within the medical device regulatory framework in the EU

Software within the medical device regulatory framework in the EU Software within the medical device regulatory framework in the EU E-HEALTH WEEK (AMSTERDAM, 10 JUNE 2016) Erik Hansson, Deputy Head of Unit, DG GROWTH - Health Technology and Cosmetics, European Commission

More information

Safeguarding public health The Regulation of Software as a Medical Device

Safeguarding public health The Regulation of Software as a Medical Device Safeguarding public health The Regulation of Software as a Medical Device Dr Head of New and Emerging Technologies Medicines and Healthcare Products Regulatory Agency (MHRA) Crown Copyright 2013 About

More information

Medical Device Software Do You Understand How Software is Regulated?

Medical Device Software Do You Understand How Software is Regulated? Medical Device Software Do You Understand How Software is Regulated? By Gregory Martin Agenda Relevant directives, standards, and guidance documents recommended to develop, maintain, and validate medical

More information

2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002

2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002 STATUTORY INSTRUMENTS 2002 No. 618 CONSUMER PROTECTION The Medical Devices Regulations 2002 Made - - - - - 20th May 2002 Laid before Parliament 21st May 2002 Coming into force - - 13th June 2002 ARRANGEMENT

More information

REFERENCE DOCUMENT. (DRAFT for Public Comments) Title: White Paper on Medical Device Software Regulation Software Qualification and Classification

REFERENCE DOCUMENT. (DRAFT for Public Comments) Title: White Paper on Medical Device Software Regulation Software Qualification and Classification AHWP/WG1/R001:2014 REFERENCE DOCUMENT (DRAFT for Public Comments) Title: White Paper on Medical Device Software Regulation Software Qualification and Classification Author: Work Group 1, Pre-Market Submission

More information

Medical Information Systems. guidance for qualification and classification of standalone software with a medical purpose

Medical Information Systems. guidance for qualification and classification of standalone software with a medical purpose approvals authorisation clinical trials communication competence cosmetics dialogue directives efficacy environment evaluation guidelines harmonisation health economics herbals homeopathics information

More information

GUIDELINES ON MEDICAL DEVICES. IVD GUIDANCES : Borderline issues A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES

GUIDELINES ON MEDICAL DEVICES. IVD GUIDANCES : Borderline issues A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single Market : regulatory environment, standardisation and New Approach Pressure equipment, medical devices, metrology MEDDEV. 2.14/1 rev. 1 January

More information

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES Foreword This guidance document is informative and advisory and has no legal authority. Individual national enforcement authorities are bound

More information

Med-Info. Council Directive 93/42/EEC on Medical Devices. TÜV SÜD Product Service GmbH

Med-Info. Council Directive 93/42/EEC on Medical Devices. TÜV SÜD Product Service GmbH Med-Info International expert information for the Medical Device industry Council Directive 93/42/E on Medical Devices Practice-oriented summary of the most important aspects and requirements contained

More information

Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC

Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC August 2013 Contents 1 Introduction...3 2 Scope of the directive...3 2.1 What is an in vitro diagnostic medical device?... 3 2.2 Specimen

More information

(OJ L 169, 12.7.1993, p. 1)

(OJ L 169, 12.7.1993, p. 1) 1993L0042 EN 11.10.2007 005.001 1 This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents B COUNCIL DIRECTIVE 93/42/EEC of 14 June 1993 concerning

More information

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES Foreword GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES These guidance notes do not aim to be a definite interpretation of National Laws and/or regulations and are for guidance purpose only.

More information

Medical Devices: CE Marking Step-by Step

Medical Devices: CE Marking Step-by Step Medical Devices: CE Marking Step-by Step Sylvia Mohr Standards Specialist U.S. Mission to the European Union 1 Introduction CE marking in 8 Steps Overview intended purpose - technical documentation classification

More information

Final Document. Software as a Medical Device (SaMD): Key Definitions. Date: 9 December 2013. Despina Spanou, IMDRF Chair. IMDRF/SaMD WG/N10FINAL:2013

Final Document. Software as a Medical Device (SaMD): Key Definitions. Date: 9 December 2013. Despina Spanou, IMDRF Chair. IMDRF/SaMD WG/N10FINAL:2013 Final Document Title: Authoring Group: Software as a Medical Device (SaMD): Key Definitions IMDRF SaMD Working Group Date: 9 December 2013 Despina Spanou, IMDRF Chair This document was produced by the

More information

An introduction to the regulation of apps and wearables as medical devices

An introduction to the regulation of apps and wearables as medical devices An introduction to the regulation of apps and wearables as medical devices Introduction Phones are increasingly equipped with a range of devices capable of being used as sensors, including gyroscopes,

More information

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? CEMO Congress, Paris, February 3rd, 2011 Dr. Martine Dehlinger-Kremer Vice President Global Regulatory Affairs Agenda

More information

Regulatory Strategy for Drug-Device Combinations By John Worroll

Regulatory Strategy for Drug-Device Combinations By John Worroll THE MEDICAL DEVICE CONSULTING, AUDITING & TRAINING EXPERTS Regulatory Strategy for Drug-Device Combinations By John Worroll Introduction Drug-device combinations have always been a challenge to the regulatory

More information

Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca

Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca Software Regulated as a Medical Device Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau sarah.chandler@hc-sc.gc.ca Therapeutic Products Directorate Director General S. Sharma

More information

Introduction Processes and KPIs Health Care Value Chain. Rafael Provencio

Introduction Processes and KPIs Health Care Value Chain. Rafael Provencio Introduction Processes and KPIs Health Care Value Chain Rafael Provencio Table of contents 1. Regulatory Framework EU 2. Hospital Value Chain 3. Processes 4. KPIs 2 What is the goal of the regulatory framework?

More information

COCIR GLOSSARY OF TERMS

COCIR GLOSSARY OF TERMS COCIR GLOSSARY OF TERMS 2 European Coordination Committee of the Radiological, 13 COCIR TELEMEDICINE TOOLKIT 2011 Part 2 GLOSSARY OF TERMS Introduction.............................. 14 Part I: Telemedicine

More information

Regulatory Considerations for Medical Device Software. Medical Device Software

Regulatory Considerations for Medical Device Software. Medical Device Software Medtec Ireland 2015 Wireless Medical Devices Regulatory Considerations for Medical Device Software Kenneth L. Block, RAC October 7, 2015 Galway, Ireland Offices: Dallas, Texas (12 employees) Tokyo, Japan

More information

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE White paper produced by Maetrics For more information, please contact global sales +1 610 458 9312 +1 877 623 8742 globalsales@maetrics.com

More information

MEDICAL DEVICES INTERIM REGULATION

MEDICAL DEVICES INTERIM REGULATION Issued by the Saudi Food and Drug Authority Board of Directors decree number 1-8-1429 and dated 27 December 2008 The Interim was published in Umm Al-QURA Journal year 86 Issue No:4249 dated in 17 April

More information

Fra udvikling til kommercialisering af kliniske apps Apps som medicinsk udstyr? 29/1-2015. Brian Hedegaard, DELTA

Fra udvikling til kommercialisering af kliniske apps Apps som medicinsk udstyr? 29/1-2015. Brian Hedegaard, DELTA Fra udvikling til kommercialisering af kliniske apps Apps som medicinsk udstyr? 29/1-2015 Brian Hedegaard, DELTA Awareness please.. Brian Hedegaard Senior Consultant - Medical devices Approval Management

More information

US & CANADA: REGULATION AND GUIDELINES ON MEDICAL SOFTWARE AND APPS OR

US & CANADA: REGULATION AND GUIDELINES ON MEDICAL SOFTWARE AND APPS OR US & CANADA: REGULATION AND GUIDELINES ON MEDICAL SOFTWARE AND APPS OR A MEDICAL DEVICE IS A MEDICAL DEVICE IS A MEDICAL DEVICE AHWP Medical SW Workshop Taipei, Taiwan November 3, 2012 John G. Abbott,

More information

Navigating the New EU Rules for Medical Device Software

Navigating the New EU Rules for Medical Device Software Navigating the New EU Rules for Medical Device Software Mathias Klümper and Erik Vollebregt examine how European Directive 2007/47/EC will affect requirements for medical device software. Directive 2007/47/EC

More information

BSI: An In Vitro Diagnostics Notified Body. A guide to the In Vitro Diagnostic Directive....making excellence a habit.

BSI: An In Vitro Diagnostics Notified Body. A guide to the In Vitro Diagnostic Directive....making excellence a habit. BSI: An In Vitro Diagnostics Notified Body A guide to the In Vitro Diagnostic Directive...making excellence a habit. A BSI guide to the In Vitro Diagnostic Directive Introduction In Vitro Diagnostics (IVD)

More information

Health Science Career Field Allied Health and Nursing Pathway (JM)

Health Science Career Field Allied Health and Nursing Pathway (JM) Health Science Career Field Allied Health and Nursing Pathway (JM) ODE Courses Possible Sinclair Courses CTAG Courses for approved programs Health Science and Technology 1 st course in the Career Field

More information

Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System

Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System NBOG s Best Practice Guide applicable for AIMDD, MDD, and IVDD 2014-3 Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System 1 Introduction The

More information

Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994

Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994 Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical SUR-G0014-1 12 NOVEMBER 2013 This guide does not purport to be an interpretation of law and/or regulations

More information

GUIDELINES ON MEDICAL DEVICES

GUIDELINES ON MEDICAL DEVICES EUROPEAN COMMISSION DG Internal Market, Industry, Entrepreneurship and SMEs Consumer, Environmental and Health Technologies Health technology and Cosmetics MEDDEV 2.7/3 revision 3 May 2015 GUIDELINES ON

More information

CE Marking Changing & Evolving Requirements September 2013 2013 BSI Healthcare Roadshow

CE Marking Changing & Evolving Requirements September 2013 2013 BSI Healthcare Roadshow CE Marking Changing & Evolving Requirements September 2013 2013 BSI Healthcare Roadshow Existing Directives 2 Directive 93/42/EEC 1. Check Scope of Medical Device Directive (Article 1) 2. Determine Device

More information

Legal compliance for developers. Training materials (prepared by Tilburg University)

Legal compliance for developers. Training materials (prepared by Tilburg University) Legal compliance for developers using FI-STAR ehealth platform Training materials (prepared by Tilburg University) Target audience: Target audience and objectives developers & other potential users of

More information

Medical Certification: Bringing genomic microcores to clinical use OI- VF- WP- 011

Medical Certification: Bringing genomic microcores to clinical use OI- VF- WP- 011 Medical Certification: Bringing genomic microcores to clinical use VoltedFlow GmbH Industriestrasse 23, 6055, Alpnach Dorf, Switzerland 1 Summary VoltedFlow has developed solutions to vastly speed up DNA

More information

Medical Software Development. International standards requirements and practice

Medical Software Development. International standards requirements and practice Medical Software Development International standards requirements and practice Food and Drug Administration What? A public health agency Why? Protect American consumers How? By enforcing the Federal Food,

More information

The Act on Medical Devices

The Act on Medical Devices The Act on Medical Devices The Act on Medical Devices of 2 nd August 1994 (Federal Law Gazette I, p. 1963), in the version of 7 th August 2002 (Federal Law Gazette I, p. 3146), last amended by Article

More information

CE Marking: Your Key to Entering the European Market

CE Marking: Your Key to Entering the European Market CE Marking: Your Key to Entering the European Market N. Gerard Zapiain U.S. Department of Commerce International Trade Administration Office of Microelectronics, Medical Equipment and Instrumentation I.

More information

Medical Devices. IRELAND Arthur Cox

Medical Devices. IRELAND Arthur Cox Medical Devices IRELAND Arthur Cox CONTACT INFORMATION Colin Kavanagh Arthur Cox Earlsfort Centre Earlsfort Terrace Dublin 2, Ireland +353.1.618.0548 Colin.Kavanagh@arthurcox.com www.arthurcox.com 1. Definition

More information

The Shifting Sands of Medical Software Regulation

The Shifting Sands of Medical Software Regulation The Shifting Sands of Medical Software Regulation Suzanne O Shea Ralph Hall September 10, 2014 What Software is Regulated by FDA? FDA regulates medical devices. FDA regulates software that meets the definition

More information

Fields of Education. Last updated August 2011

Fields of Education. Last updated August 2011 Fields of Education Last updated August 2011 Monash University is required to report to the Department of Education, Employment and Workplace Relations (DEEWR) the number of higher degree by research (HDR)

More information

FACULTY OF ALLIED HEALTH SCIENCES

FACULTY OF ALLIED HEALTH SCIENCES FACULTY OF ALLIED HEALTH SCIENCES 102 Naresuan University FACULTY OF ALLIED HEALTH SCIENCES has focused on providing strong professional programs, including Medical established as one of the leading institutes

More information

The Medical Products Agency s Working Group on Medical Information Systems. Project summary

The Medical Products Agency s Working Group on Medical Information Systems. Project summary The Medical Products Agency s Working Group on Medical Information Systems Project summary Proposal for guidelines regarding classification of software based information systems used in health care This

More information

Medical Technologies Evaluation Programme Methods guide

Medical Technologies Evaluation Programme Methods guide Issue date: April 2011 Medical Technologies Evaluation Programme Methods guide National Institute for Health and Clinical Excellence MidCity Place 71 High Holborn London WC1V 6NA www.nice.org.uk National

More information

EU Law and Life Sciences Mobile Health Apps. Webinar - Brussels, 9 July 2014

EU Law and Life Sciences Mobile Health Apps. Webinar - Brussels, 9 July 2014 EU Law and Life Sciences Mobile Health Apps Webinar - Brussels, 9 July 2014 Dr. Dr. Adem Koyuncu, M.D. Partner Prof. Damien Geradin Partner Agenda Covington at a glance Mobile health apps - opportunities

More information

Future Regulatory Challenges for In VitroDiagnostics. Trilateral Workshop, Berlin 16&17 June 2010

Future Regulatory Challenges for In VitroDiagnostics. Trilateral Workshop, Berlin 16&17 June 2010 Future Regulatory Challenges for In VitroDiagnostics Trilateral Workshop, Berlin 16&17 June 2010 Doris-Ann Williams Director General The British In Vitro Diagnostics Association 2 European Regulation of

More information

Unit/Standard Number. Proficiency Level Achieved: (X) Indicates Competency Achieved to Industry Proficiency Level

Unit/Standard Number. Proficiency Level Achieved: (X) Indicates Competency Achieved to Industry Proficiency Level 1 Secondary Task List 100 INTRODUCTION TO HEALTHCARE 101 Identify the roles and educational/credentialing requirements of various health care practitioners. 102 Describe the 's job description, personal

More information

SMF Awareness Seminar 2014

SMF Awareness Seminar 2014 SMF Awareness Seminar 2014 Clinical Evaluation for In Vitro Diagnostic Medical Devices Dr Jiang Naxin Health Sciences Authority Medical Device Branch 1 In vitro diagnostic product means Definition of IVD

More information

Regulatory Issues in Genetic Testing and Targeted Drug Development

Regulatory Issues in Genetic Testing and Targeted Drug Development Regulatory Issues in Genetic Testing and Targeted Drug Development Janet Woodcock, M.D. Deputy Commissioner for Operations Food and Drug Administration October 12, 2006 Genetic and Genomic Tests are Types

More information

CE-Marked Product List (MDD)

CE-Marked Product List (MDD) CE-Marked Product List (MDD) On March 21st the revised Directive 93/42/EEC concerning medical devices which includes the amendment M5 Directive 2007/47/EC of the European Parliament and of the Council

More information

GUIDELINES ON MEDICAL DEVICES EVALUATION OF CLINICAL DATA : A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES

GUIDELINES ON MEDICAL DEVICES EVALUATION OF CLINICAL DATA : A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single Market : regulatory environment, standardisation and New Approach Pressure equipment, medical devices, metrology MEDDEV. 2.7.1 April 2003 GUIDELINES

More information

Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR)

Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR) Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR) Paul Brooks Healthcare Solutions Disclaimer The new regulations are not finalized and subject to change http://medicaldevices.bsigroup.com/en-us/resources/whitepapers-and-articles/

More information

CENTER FOR CONNECTED HEALTH POLICY

CENTER FOR CONNECTED HEALTH POLICY CENTER FOR CONNECTED HEALTH POLICY The Center for Connected Health Policy (CCHP) is a public interest nonprofit organization that develops and advances telehealth policy solutions to promote improvements

More information

TOTAL PROCESS MANAGEMENT FOR YOUR NEWBORN SCREENING LABORATORY. Specimen Gate software

TOTAL PROCESS MANAGEMENT FOR YOUR NEWBORN SCREENING LABORATORY. Specimen Gate software TOTAL PROCESS MANAGEMENT FOR YOUR NEWBORN SCREENING LABORATORY Specimen Gate software MANAGE YOUR ENTIRE SCREENING PROCESS Newborn screening generates vast amounts of data that require continuous tracking.

More information

APMI CODE OF PRACTICE ON ADVERTISING OF MEDICINAL PRODUCTS ISSUE 1

APMI CODE OF PRACTICE ON ADVERTISING OF MEDICINAL PRODUCTS ISSUE 1 APMI CODE OF PRACTICE ON ADVERTISING OF MEDICINAL PRODUCTS ISSUE 1 August 2010 1. Introduction... 2 2. Scope... 3 3. Definitions... 4 4. General rules of advertising of medicinal products... 6 5. Advertising

More information

How to apply for Medical Device License in Taiwan

How to apply for Medical Device License in Taiwan How to apply for Medical Device License in Taiwan According to Pharmaceutical Affairs Act (PAA), all medical devices regardless their classification shall apply for Medical Device License before they are

More information

Medical Information Systems

Medical Information Systems Medical Information Systems Introduction The introduction of information systems in hospitals and other medical facilities is not only driven by the wish to improve management of patient-related data for

More information

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES A WHITE PAPER ON THE REQUIREMENTS, REGULATIONS AND OPPORTUNITIES CONTAINED IN EC DIRECTIVE 98/79/EC AND A BRIEF COMPARISON

More information

Guidelines for conformity assessment of In Vitro Fertilisation (IVF) and Assisted Reproduction Technologies (ART) products

Guidelines for conformity assessment of In Vitro Fertilisation (IVF) and Assisted Reproduction Technologies (ART) products EUROPEAN COMMISSION DG HEALTH & CONSUMERS Directorate B - Consumer Affairs Unit B2 Health Technology and Cosmetics MEDICAL DEVICES: Guidance document MEDDEV 2.2/4 January 2012 Guidelines for conformity

More information

Introduction to Information and Computer Science: Information Systems

Introduction to Information and Computer Science: Information Systems Introduction to Information and Computer Science: Information Systems Lecture 1 Audio Transcript Slide 1 Welcome to Introduction to Information and Computer Science: Information Systems. The component,

More information

MEDICAL DEVICE GUIDANCE

MEDICAL DEVICE GUIDANCE May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained

More information

Want to know more about the Notified Body?

Want to know more about the Notified Body? NBG/V 2/07/14 BSI Notified Body (0086) Kitemark Court Davy Avenue Knowlhill Milton Keynes MK5 8PP United Kingdom T: +44 845 080 9000 E: eu.medicaldevices@bsigroup.com W: medicaldevices.bsigroup.com Want

More information

Nurse Practitioner Mentor Guideline NPAC-NZ

Nurse Practitioner Mentor Guideline NPAC-NZ Nurse Practitioner Mentor Guideline NPAC-NZ Purpose To provide a framework for the mentorship of registered nurses to prepare for Nurse Practitioner (NP) registration from the Nursing Council of New Zealand.

More information

Medical/Clinical Assistant CIP 51.0801 Task Grid

Medical/Clinical Assistant CIP 51.0801 Task Grid 1 Secondary Task List 100 PHARMACOLOGY 101 Measure drug dosages using droppers, medicine cups, syringes, and other specialized devices. 102 Utilize correct technique to administer medications using the

More information

GUIDELINES ON A MEDICAL DEVICES VIGILANCE SYSTEM

GUIDELINES ON A MEDICAL DEVICES VIGILANCE SYSTEM EUROPEAN COMMISSION DG ENTERPRISE AND INDUSTRY Directorate F-Consumer Good Unit F3- Cosmetic and Medical Devices MEDICAL DEVICES: Guidance document MEDDEV 2.12-1 rev 5 April 2007 GUIDELINES ON A MEDICAL

More information

Medical Device Software

Medical Device Software Medical Device Software Bakul Patel Senior Policy Advisor 1 Overview Medical devices and software Oversight principles and Current approach Trends, Challenges and opportunities Addressing challenges 2

More information

On Behalf of: InTouch Health

On Behalf of: InTouch Health On Behalf of: InTouch Health White Paper FDA Regulation of Mobile Health Technologies The Current Regulatory Framework as Applied to InTouch Health s Telehealth Solutions June 15, 2012; Updated June 15,

More information

MHRA GMP Data Integrity Definitions and Guidance for Industry January 2015

MHRA GMP Data Integrity Definitions and Guidance for Industry January 2015 MHRA GMP Data Integrity Definitions and Guidance for Industry Introduction: Data integrity is fundamental in a pharmaceutical quality system which ensures that medicines are of the required quality. This

More information

CDRH Regulated Software

CDRH Regulated Software CDRH Regulated Software An Introduction John F. Murray Jr. CDRH Software Compliance Expert CDRH Regulates Software in the following areas Medical Devices Automation of Production Systems Automation of

More information

OCCUPATIONAL GROUP: Health Services. CLASS FAMILY: Allied Health CLASS FAMILY DESCRIPTION:

OCCUPATIONAL GROUP: Health Services. CLASS FAMILY: Allied Health CLASS FAMILY DESCRIPTION: OCCUPATIONAL GROUP: Health Services CLASS FAMILY: Allied Health CLASS FAMILY DESCRIPTION: This family of positions is comprised of health care practitioners with formal education and clinical training

More information

TRANSATLANTIC TRADE AND INVESTMENT PARTNERSHIP

TRANSATLANTIC TRADE AND INVESTMENT PARTNERSHIP DISCLAIMER: The EU reserves the right to make subsequent modifications to this text and to complement its proposals at a later stage, by modifying, supplementing or withdrawing all, or any part, at any

More information

American Psychological Association D esignation Criteria for Education and

American Psychological Association D esignation Criteria for Education and American Psychological Association D esignation Criteria for Education and Training Programs in Preparation for Prescriptive Authority Approved by APA Council of Representatives, 2009 Criterion P: Program

More information

Vigilance Reporting. Vicky Medley - Head of QMS, Medical Devices. September 2015. Copyright 2015 BSI. All rights reserved.

Vigilance Reporting. Vicky Medley - Head of QMS, Medical Devices. September 2015. Copyright 2015 BSI. All rights reserved. Vigilance Reporting Vicky Medley - Head of QMS, Medical Devices September 2015 2 Why? 3 protecting and improving public health https://www.gov.uk/government/organisations/medicines-and-healthcareproducts-regulatory-agency/about

More information

Guidelines for the Use of Controlled Substances in the Treatment of Pain Adopted by the New Hampshire Medical Society, July 1998

Guidelines for the Use of Controlled Substances in the Treatment of Pain Adopted by the New Hampshire Medical Society, July 1998 Guidelines for the Use of Controlled Substances in the Treatment of Pain Adopted by the New Hampshire Medical Society, July 1998 Section I: Preamble The New Hampshire Medical Society believes that principles

More information

Improving quality, protecting patients

Improving quality, protecting patients Improving quality, protecting patients Standards of proficiency for Healthcare Science Practitioners 31 July 2014 Version 1.0 Review date: 31 July 2015 Foreword I am pleased to present the Academy for

More information

12040/1/15 REV 1 LES/ns 1 DG B 3B

12040/1/15 REV 1 LES/ns 1 DG B 3B Council of the European Union Brussels, 21 September 2015 (OR. en) Interinstitutional File: 2012/0266 (COD) 12040/1/15 REV 1 NOTE From: To: General Secretariat of the Council PHARM 36 SAN 281 MI 567 COMPET

More information

Check List. Telehealth Credentialing and Privileging Sec. 482.12. Conditions of Participation Governing Body

Check List. Telehealth Credentialing and Privileging Sec. 482.12. Conditions of Participation Governing Body Check List Telehealth Credentialing and Privileging Sec. 482.12. Conditions of Participation Governing Body The Centers for Medicare and Medicaid Services (CMS) final rule on credentialing and privileging

More information

GUIDELINES ON A MEDICAL DEVICES VIGILANCE SYSTEM

GUIDELINES ON A MEDICAL DEVICES VIGILANCE SYSTEM EUROPEAN COMMISSION DG Health and Consumers (SANCO) Directorate B-Consumer Affairs Unit B2- Health Technology and Cosmetics MEDICAL DEVICES: Guidance document MEDDEV 2.12-1 rev 8 January 2013 GUIDELINES

More information

CE Marking and Technical Standardisation

CE Marking and Technical Standardisation GAMBICA Technical Guide CE Marking and Technical Standardisation Guidelines for application to Electrical Power Drive Systems Edition 4 : 2015 CE Marking and Technical Standardisation Guidelines for application

More information

Cloud Development of Medical Systems By Oleg Kruk, Embedded Research Lab Leader, DataArt

Cloud Development of Medical Systems By Oleg Kruk, Embedded Research Lab Leader, DataArt Cloud Development of Medical Systems By Oleg Kruk, Embedded Research Lab Leader, DataArt Abstract Wireless electronic medical devices have made remote medicine a reality. Disease prevention, monitoring

More information

Telemedicine, EMR & Remote Connectivity

Telemedicine, EMR & Remote Connectivity Telemedicine, EMR & Remote Connectivity Saroj Mishra, MS, FACS Prof & Head, Dept. of Endocrine Surgery & Faculty I/C, SGPGI Telemedicine Program Sanjay Gandhi Postgraduate Institute of Medical Sciences

More information

2.0 Rationale, Purpose and Scope... 5. 4.0 Definitions... 6. 5.0 General Principles... 7

2.0 Rationale, Purpose and Scope... 5. 4.0 Definitions... 6. 5.0 General Principles... 7 Table of Contents Principles of IVD Medical Devices Classification 1.0 Introduction... 4 2.0 Rationale, Purpose and Scope... 5 2.1 Rationale... 5 2.2 Purpose... 5 2.3 Scope... 5 3.0 References... 5 4.0

More information

medexter clinical decision support

medexter clinical decision support medexter Arden Syntax training course Klaus-Peter Adlassnig and Karsten Fehre Medexter Healthcare Borschkegasse 7/5 A-1090 Vienna www.medexter.com Arden Syntax training course, Vienna, 17 June 2015 Computers

More information

PROPOSED DOCUMENT. Global Harmonization Task Force. Title: Principles of In Vitro Diagnostic (IVD) Medical Devices Classification

PROPOSED DOCUMENT. Global Harmonization Task Force. Title: Principles of In Vitro Diagnostic (IVD) Medical Devices Classification SG1(PD)/N045R12 PROPOSED DOCUMENT Global Harmonization Task Force Title: Principles of In Vitro Diagnostic (IVD) Medical Devices Classification Authoring Group: Study Group 1 of the Global Harmonization

More information

Changing The World One Hospital At a Time

Changing The World One Hospital At a Time Changing The World One Hospital At a Time miplatform Integrated Medical Image and Information Management Platform PACS/RIS System Mobile Imaging System Image-based Conferencing System TeleMed/TeleRad System

More information

CANADIAN RULE BASED CLASSIFICATION SYSTEM (IVDD) Life Sciences British Columbia NRC-Industry Research Assistance Program

CANADIAN RULE BASED CLASSIFICATION SYSTEM (IVDD) Life Sciences British Columbia NRC-Industry Research Assistance Program CANADIAN RULE BASED CLASSIFICATION SYSTEM (IVDD) Life Sciences British Columbia NRC-Industry Research Assistance Program Health Canada Regulations on Medical Devices Vancouver, B.C. October 29, 2007 Nancy

More information

Rule 5.2 Definitions. For the purpose of Chapter 5 only, the following terms have the meanings indicated:

Rule 5.2 Definitions. For the purpose of Chapter 5 only, the following terms have the meanings indicated: Part 2635 Chapter 5: Practice of Telemedicine Rule 5.1 Preamble. These regulations are intended to authorize M.D. and D.O. licensees of the Mississippi State Board of Medical Licensure to practice telemedicine

More information

CE MARKING OF MEDICAL DEVICES Guide

CE MARKING OF MEDICAL DEVICES Guide CE MARKING OF MEDICAL DEVICES Guide Edition 2005 CE marking of medical devices Instructions for use LNE / G-MED 1/63 CONTENTS 1. Introduction to the guide 2. Regulation 2.1. European directives regarding

More information

Challenge of FUJIFILM in Medical ICT

Challenge of FUJIFILM in Medical ICT Challenge of FUJIFILM in Medical ICT Sep.10, 2014 FORWARD-LOOKING STATEMENTS Forward-looking statements, such as those relating to earnings forecasts and other projections contained in this material, are

More information

MHRA GMP Data Integrity Definitions and Guidance for Industry March 2015

MHRA GMP Data Integrity Definitions and Guidance for Industry March 2015 MHRA GMP Data Integrity Definitions and Guidance for Industry Introduction: Data integrity is fundamental in a pharmaceutical quality system which ensures that medicines are of the required quality. This

More information

HIPAA Notice of Privacy Practices

HIPAA Notice of Privacy Practices HIPAA Notice of Privacy Practices THIS NOTICE DESCRIBES HOW MEDICAL INFORMATION ABOUT YOU MAY BE USED AND DISCLOSED AND HOW YOU CAN GET ACCESS TO THIS INFORMATION. PLEASE REVIEW IT CAREFULLY. This Notice

More information

EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below.

EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below. EMKI COMPETENCE in MD EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below. Annex II.: EC DECLARATION OF CONFORMITY

More information

MISSISSIPPI LEGISLATURE REGULAR SESSION 2016

MISSISSIPPI LEGISLATURE REGULAR SESSION 2016 MISSISSIPPI LEGISLATURE REGULAR SESSION 2016 By: Representative Mims To: Public Health and Human Services HOUSE BILL NO. 1187 1 AN ACT TO AMEND SECTION 73-25-34, MISSISSIPPI CODE OF 1972, 2 TO REVISE THE

More information

The U.S. FDA s Regulation and Oversight of Mobile Medical Applications

The U.S. FDA s Regulation and Oversight of Mobile Medical Applications The U.S. FDA s Regulation and Oversight of Mobile Medical Applications The U.S. FDA s Regulation and Oversight of Mobile Medical Applications As smart phones and portable tablet computers become the preferred

More information

COMMISSION STAFF WORKING DOCUMENT. on the existing EU legal framework applicable to lifestyle and wellbeing apps. Accompanying the document

COMMISSION STAFF WORKING DOCUMENT. on the existing EU legal framework applicable to lifestyle and wellbeing apps. Accompanying the document EUROPEAN COMMISSION Brussels, 10.4.2014 SWD(2014) 135 final COMMISSION STAFF WORKING DOCUMENT on the existing EU legal framework applicable to lifestyle and wellbeing apps Accompanying the document GREEN

More information

Regulation of Medical Devices involving Software in Australia an Overview

Regulation of Medical Devices involving Software in Australia an Overview Regulation of Medical Devices involving Software in Australia an Overview John Jamieson Conformity Assessment Branch Therapeutic Goods Administration PO Box 100, Woden ACT 2606 John.Jamieson@health.gov.au

More information

Hitachi Healthcare Company - Our Strategy and Opportunities -

Hitachi Healthcare Company - Our Strategy and Opportunities - Innovative collaboration for better healthcare [Session III: Some tools for the future healthcare system] Hitachi Healthcare Company - Our Strategy and Opportunities - Masahiro Kitano Senior Vice President

More information

Medical College of Georgia Augusta, Georgia School of Medicine Competency based Objectives

Medical College of Georgia Augusta, Georgia School of Medicine Competency based Objectives Medical College of Georgia Augusta, Georgia School of Medicine Competency based Objectives Medical Knowledge Goal Statement: Medical students are expected to master a foundation of clinical knowledge with

More information

Continua and IHE Joint Recommendation on Remote Monitoring

Continua and IHE Joint Recommendation on Remote Monitoring Continua and IHE Joint Recommendation on Remote Monitoring Arto Holopainen - Mega Elektroniikka Oy Mega Elektroniikka Oy www.megaemg.com 3.2.2010 www.ehit.fi Version One Healthcare Record Standards Disease

More information