1 Rx Recon - Electronic Reconstitution Unit Users Guide HCL #10406 Health Care Logistics, Inc HCL-INTL is a registered trademark of Health Care Logistics, Inc.
2 Table of Contents... Important Caution Components Installation Pre-setting Clean Air Venting Filter Installing Clean Air Vent Filter Purging Calibrating Manual Mode Scan Mode Entering a New Product Modifying an Existing Product Removing a Product Maintenance To Clean Warranty and Specifications Maintenance Schedule Notes Property of Health Care Logistics, Inc. This material cannot be copied without the prior written permission of Health Care Logistics, Inc. Call Free: or HCL INTL ( ) Fax Free:
3 Important The Rx Recon is a fully automated distilled water dispenser that serves to reconstitute antibiotics. To ensure that you will enjoy many years of trouble free operation, we have developed this Users Guide. It contains information on how to operate and maintain your dispenser properly and safely. Please read the entire manual carefully before attempting to use the Rx Recon for the first time. CAUTION Only use distilled water. The Rx Recon unit has been designed for distilled water exclusively. Always keep the cap on the reservoir and a clean air vent filter in the air vent opening to avoid water contamination and a malfunction of the delivery mechanism. The reservoir must be free of dust particles and fibers. Always empty and unplug the unit before attempting to move the Rx Recon. Always replenish the water reservoir with special care to avoid spilling distilled water within the unit. Failure to do so could result in damaging the unit components. Clean the unit reservoir and the flow-thru cylinder every month. HELP If you ever have a question concerning your Rx Recon or if you need service, feel free to call: P.O. Box 25, Circleville, OH Call Free: or HCL INTL ( ) Fax Free:
4 Components 4 Call Free: or HCL INTL ( ) Fax Free:
5 Installation Your Rx Recon unit was shipped with an installation kit that contains the following: wall bracket and screws for mounting the Rx Recon flow-thru cylinder small piece of cotton bar code scanner AC cord particulate filter 6-clean air venting filters graduated cylinder for calibration step cleaning hose for emptying the reservoir Choose the appropriate location to install the Rx Recon unit. Make sure that you keep enough room to easily fill and refill the reservoir and to comfortably prepare the prescription. Manipulate the unit carefully when proceeding. Installing the Rx Recon unit: The holder has to be affixed strongly to support the weight unit at its maximum capacity ( 4.0 liters of distilled water). 1. Once the mounting location is chosen, hold the wall bracket to the surface and mark the mounting holes with a pen or pencil. The unit needs mounted close to an electrical outlet. 2. Drill the marked holes with a 1 8 (3mm) drill bit. 3. Insert the plastic anchors into the drilled holes. Call Free: or HCL INTL ( ) Fax Free:
6 Installation 4. Anchors may need to be gently tapped with a hammer. 5. At this point all six of the anchors should be flush with the wall. 6. Align the holes in the wall bracket with the anchors and attach the bracket using the screws provided. Use a drill gun with a phillips head or a phillips screwdriver to tighten.. st 7. Attach the barcode scanner holder to either side of the unit with the screws provided. The unit has pre-drilled holes for easy location and attachment. r Lip 8. Slide the lip on the back of the unit onto the wall bracket. The lip will slide over the top of the wall bracket. Once the unit is on the bracket slide it down until it cannot go any further. 9. Place a small level on the top of the machine to ensure it is level. Adjust accordingly. 6 Call Free: or HCL INTL ( ) Fax Free:
7 Installation jack outlet 10. Plug the AC cord into the socket on the bottom of the unit and into an electrical outlet to power the unit. 11. Plug the scanner jack into the jack connector outlet on the bottom of the unit. Snap the scanner into the scanner holder. 12. Remove the lid to the water reservoir. 13. Fill the reservoir with distilled water. 14. Loosely insert the small piece of cotton into the flow-thru cylinder air vent to prevent dust and foreign material from entering. 15. Remove the cap on the white outlet tube located on the bottom of the unit. 16. Attach the flow thru cylinder to the white outlet tube. 17. Turn the Rx Recon on by flipping the switch under the unit and turn the barcode scanner on by pressing the button on the scanner. Call Free: or HCL INTL ( ) Fax Free:
8 Installation 18. (A) Follow the initial Rinse and Clean Instructions on pages of this User s Guide. (B) After performing the initial Rinse and Clean follow pre-setting steps on page 8-11 of this User s Guide. Initial Rinse and Clean After unit installation, and as with any new electrical appliance utilizing fluids, before using the Rx Recon for the first time, it should be put through a cleaning cycle as explained on page of this User s Guide. Follow the directions under To Clean. Also note that this procedure should be completed on a monthly basis once the unit is put into operation. Pre-setting Once the unit has been rinsed and cleaned it is ready to be put into functional operation. You must insert a clean air venting filter into the unit, fill the reservoir with distilled water, purge the supply line, and calibrate the unit. Clean Air Venting Filters Clean air venting filters are provided with the unit. They filter the air being drawn into the device, allowing only clean air to enter the system during use. A clean air venting filter should be inserted into the air vent opening prior to using the device. See page 9 for instructions. 8 Call Free: or HCL INTL ( ) Fax Free:
9 Installing a #10419 Clean Air Vent Filter: Pre-setting 1. With a gloved hand, remove the Air Vent Cap by turning it counterclockwise. 2. Insert a Clean Air Vent Filter into the Air Vent Opening. The tapered part of the filter should be at the bottom. 3. Replace the Air Vent Cap. 4. Turn the Air Vent Cap clockwise to tighten. The filter should be changed during the normal recommended maintenance schedule of once a month. (See Maintenance - To Clean on page 16 of this User s Guide. You are provided with six clean air venting filters with the machine. Additional filters can be purchased in quantities of 6 from Health Care Logistics - HCL Catalog # Call Free: or HCL INTL ( ) Fax Free:
10 Pre-setting Purging: To remove the air from the unit and assure delivery accuracy, connect the flow thru cylinder and activate the unit as follow: Command from the main screen display LCD display shows 1 press <MENU> {1 = New 2 = Change} 2 press <+> {Any Key to Empty} 3 press any key, let the unit run for 20 seconds. {Any Key to Stop} While the water is running, squeeze the hose between the outlet and the flow-thru cylinder to remove the bubbles. 4 press any key to stop 5 press <CLR> to return to the main screen!!!important!!! This step must be done At step 3, press any key to start the water running. While the water is flowing into the flow-thru cylinder, squeeze the hose between the outlet and flow-thru cylinder to remove the bubbles. 10 Call Free: or HCL INTL ( ) Fax Free:
11 Calibrating: Pre-setting IMPORTANT** Fill the reservoir full of distilled water before calibrating and make sure the flow-thru cylinder is empty. The Rx Recon unit must be calibrated after installation, servicing, cleaning or when its reservoir becomes completely empty. Command from the main screen display LCD display shows 1 press <TEST> {TAKE SAMPLE 1 = OK} 2 press <1> {SAMPLING...} 3 when delivery is completed, pour the {Entr TST ML...} water into the graduated cylinder 4 enter the precise quantity as measured by {. decimal point} the graduated cylinder (example: 79.8 ml Enter the fractional amount even if it is 0). Unit will beep if the decimal point is not added. Perform the following test using the manual mode to ensure the calibration is precise. Request 100 ml in one portion and verify volume with your graduated cylinder. Note: Follow Manual Mode Instructions 1 to 4 on page 12 - press <1> in instruction 4. If the reading is not accurate, start the calibration over and make sure that you enter the precise quantity shown by your graduated cylinder. Note: Imprecise delivery by the Rx Recon occurs by imprecise data entry during the calibration step. Follow steps 1 and 2. At step 2 press the test key to start water sampling. The water sample will accumulate in the flow-thru cylinder. Place the graduated cylinder under the outlet tube on the flow thru cylinder and drain the water into the graduated cylinder. Measure the quantity of water in the graduated cylinder and input the quantity into the Rx Recon keypad. Note: The stand in these photos is for photography use only. It is not sold with the Rx Recon. Call Free: or HCL INTL ( ) Fax Free:
12 Manual Mode In the manual mode, you must manually enter the required information to deliver the distilled water quantity needed for your prescription. Command from the main screen display LCD display shows 1 press <1> {Enter ML} 2 enter the ML required by the product label {XXX ML? ENT/CLR} (if less than 100 ml press <ENTER> when done) 3 press <ENTER> if your ML entry is correct {Portion? 1 or 2} (or <CLEAR> if incorrect and enter the correct amount) 4 enter into how many portions you would like {DELIVERING} the delivery, 1 or 2 press <1> to supply 100% of the portion press <2> to supply the first portion at 80% and the second portion at 20%. shows what you have entered 5 if 2 portions were asked for and the first portion {For 2nd - > ENTER} has been delivered press <ENTER> to deliver the remaining 20% {XXX ML X PORTION} 6 press <CLR> to return to main screen O an alarm can be activated to remind that a P second portion needs to be delivered T - press <MENU> twice {Prtn Beep 1 = Y, 2 = N} I - press <1> to unable the alarm O - press <2> to disable N 12 Call Free: or HCL INTL ( ) Fax Free:
13 Scan Mode RX Recon is shipped pre-programmed with a product database loaded in it. In the Scan Mode, the Rx Recon recognizes the product through use of the product specific NDC barcode and delivers the corresponding volume of diluent required in either one or two portions as indicated on the product label. In the scan mode the user has the ability to: - enter new products - modify an existing product - remove a product Command from the main screen display LCD display shows 1 press <2> {SCAN PRODUCT...} 2 scan the product NDC until you hear a beep {DELIVERING} *if the scan is not activated in 8 seconds the LCD display will return to the main menu 3 if the data in the database suggests 2 portions {For 2nd ENTER } for this product, 80% of the required ML will be delivered 4 press <ENTER> to deliver the remaining 20% {XXX ML X PORTION} O an alarm can be activated to remind that a P second portion needs to be delivered T - press <MENU> twice I - press <1> to unable the alarm {Prtn Beep 1 = 2, 2 = N O - press <2> to disable N 5 press <CLR> to return to the main screen 6 if the scanned product is not in the database {NOT IN DATABASE} 7 to enter the product press <MENU> {SCAN PRODUCT} 8 scan the product NDC until you hear a beep {Enter new ML} 9 enter the ML required by the product label just {Portion? 1 or 2} scanned (if less than 100 ml press <ENTER> when done) 10 enter how many portions you would like for {DONE} delivery, 1 or 2 press <1> to supply 100% of the portion and then press <2> to supply the first portion at 80% and the second portion at 20% Call Free: or HCL INTL ( ) Fax Free:
14 Entering a New Product Command from the main screen display LCD display shows 1 press <MENU> {1=NEW 2=CHANGE} 2 press <1> {SCAN PRODUCT...} 3 scan the product NDC barcode until you hear {Enter new ML} a beep 4 enter the ML required by the product label {Portion? 1 or 2} just scanned (if less than 100 ml press <ENTER> when done 5 enter how many portions you would like for delivery, 1 or 2 press <1> to supply 100% of the portion press <2> to supply the first portion at 80% and the second portion at 20%. {...DONE...} and then 6 if the product is already in the database {PRODUCT EXISTS} 7 press <CLR> to return to the main screen NOTE: The maximum number of different products that can be entered into the database is (256). Entering more than the amount will produce a {Database is full} message on the LCD display. Press <CLR> to return to the main screen display. 14 Call Free: or HCL INTL ( ) Fax Free:
15 Modifying and Removing an Existing Product Modifying an Existing Product Command from the main screen display LCD display shows 1 press <MENU> {1=NEW 2=CHANGE} 2 press <2> {SCAN PRODUCT...} 3 scan the product NDC barcode until you hear {XXX ML X PORTION} a beep This is the ML and Portion found in the database 4 press <MENU> {Enter new ML} 5 enter the new ML required by the product {Portion? 1 or 2} label just scanned (if less than 100ml press <ENTER> when done 6 enter how many portions you would like for {...DONE...} delivery, 1 or 2 press <1> to supply 100% of the portion and then press <2> to supply the first portion at 80% and the second portion at 20% Removing a Product Command LCD display shows from the main screen display 1 press <MENU> {1=NEW 2=CHANGE} 2 press <2> {SCAN PRODUCT...} 3 scan the product NDC barcode until you hear {XXX ML X PORTION} a beep This is the ML and Portion found in the database 4 press <MENU> {Enter new ML} 5 enter <0> for NEW ML and press <ENTER> {Portion? 1 or 2} 6 enter <0> for NEW PORTION {...DONE...} and then ERROR - If an error has been made, you will be notified by a two BEEPS and/or the LCD will show {input error xxx} for 2 seconds. The display will then go back to where the error occurred to allow you to re-enter data. Call Free: or HCL INTL ( ) Fax Free:
16 Maintenance - To Clean UNIT REFILL INDICATOR - If the reservoir becomes low on liquid the LCD display will FLASH {!!Water low!!} until the reservoir is filled. No other functions can be performed until the reservoir is replenished. Your new Rx Recon is designed to give you many years of dependable service. However, there are a few things you are expected to do to maintain the units efficiency: Use only distilled water in unit Always keep the cap on the reservoir. To avoid water contamination and a malfunction of the delivery mechanism, the reservoir must be free of dust particles and fibers. Always replenish the reservoir with special care to avoid spilling distilled water within the unit. Failure to do so could result in damaging the unit components. Clean the unit reservoir and the flow-thru cylinder every month as follows: TO CLEAN: The reservoir should be cleaned monthly with a dilute bleach solution mixed in the following ratio: 1 teaspoonful (5mL) per 1500mL of distilled water or 3 teaspoonfuls (15mL) per gallon of distilled water. Note: The stand in these photos is for photography use only. It is not sold with the Rx Recon. 1. Turn the power to the unit off. 2. Fill the reservoir with a one gallon mixture of the dilute bleach solution. To Empty the Reservoir Command 3. Let the solution set in the reservoir for 15 minutes. 4. Turn the unit back on, connect the cleaning hose at the outlet of the flow thru cylinder and open the clip. LCD display shows from the main screen display 5. Empty the reservoir by performing the steps on the right. See page 17 for steps press <MENU> {1=NEW 2=CHANGE} 2 press <+> {Any key to EMPTY} 3 press ANY KEY {Any key to STOP} let it run until it is empty 4 press <CLR> to return to main screen!!!important!!! this step must be done Call Free: or HCL INTL ( ) Fax Free:
17 To Clean 6. Rinse the unit with an additional two gallons of distilled water by filling and emptying the reservoir as shown on the bottom of page 16. 7A. DURING INITIAL INSTALLATION: After rinsing, the unit is now ready to be put into use. Fill the reservoir with distilled water and follow the Pre-setting Instructions on pages 4 & 5 for air filtration, purging, and calibration. 7B. DURING MONTHLY MAINTENANCE: After rinsing, the unit is now ready to be put back into use. Refill the reservoir with distilled water. (Note: the unit needs to be re-purged and recalibrated). 8. Purge the system following the steps on page Calibrate the system by following steps on page The unit is now to be used again for reconstitution. Call Free: or HCL INTL ( ) Fax Free:
18 Warranty and Specifications Warranty Length Health Care Logistics Health Care Logistics Warranty Will Cover Will Not Cover One Year Full Replacement parts Any damage resulting from Warranty on and repair labor costs Rx Recon unit: to correct defects in A. Improper installation From date of materials B. Inadequate use of unit purchase C. using other liquid than distilled water in the reservoir D. dust or fibers within the reservoir E. accident, alteration, abuse One Year Full Replacement parts F. repairs to parts or systems caused Warranty on and repair labor costs by unauthorized modifications Barcode scanner: to correct defects in made to the unit From date of materials purchase Specifications Dimensions Height Length Width Without the graduated cylinder (387mm) (260mm) (143mm) With graduated cylinder (781mm) (260mm) (143mm) Reservoir capacity 4.0 liters (1 gallon) Operating voltage 115V AC 60 HZ 500 M.A. Delivery capacity 8ml/ sec 80ml/10 sec Weight of unit empty 6 pounds (2.7kg) full 16 pounds (7.2kg) 18 Call Free: or HCL INTL ( ) Fax Free:
19 Maintenance Schedule Call Free: or HCL INTL ( ) Fax Free:
20 Notes 20 Call Free: or HCL INTL ( ) Fax Free:
Troubleshooting Guide for Dispenser Model #900172
Troubleshooting Guide for Dispenser Model #900172 Q. The cold water is room temperature. A. There is a cold water thermostat control mounted to the back of the majority of our appliances. Simply unplug
More informationNewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS
NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS ELECTRICAL SAFETY This appliance is for indoor use only. Always turn off the unit and unplug
More informationNewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS
NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS BEFORE USE GENERAL SAFETY INSTRUCTIONS: ALWAYS OPERATE THE UNIT IN AN UPRIGHT POSITION AND PLACE
More informationGlobal Water Instrumentation, Inc.
Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 6 Tipping Bucket: RG200 02/05/10 01-343 - 1 - Congratulations on your
More informationPortable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W
Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CCS Customer Support : 1-800-474-2147 Voltage rating: 115V~60Hz Power rating : 1400W For product inquiries or support
More informationPortable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW
Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MN12CES / MN10CESWW Voltage rating: 120V~60Hz Power rating : 1100W (MN12CES) Power rating : 900W (MN10CESWW) Customer
More informationCordlessPleated FabricShade including Top-Down/Bottom-Up. owner's handbook
CordlessPleated FabricShade including Top-Down/Bottom-Up owner's handbook July, 2013 español página 9 2 INSTALLATION INSTRUCTIONS Step 1. Check Package Contents Missing part? Call 1-800-264-1190 Mounting
More informationUSER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model # 900172
Model # 900172: Page 1 USER MANUAL Bottom Loading Bottled Water Dispenser Model # 900172 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING
More informationPortable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W
MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries
More informationDTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual
DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR Installation and Operation Manual 1 ENGLISH Safety Instructions 2 Features 2-3 Specifications 3 Installation 4-5 Wiring Diagrams 6-7 Warranty 8
More informationBubble King. User Manual
Bubble King User Manual TABLE OF CONTENTS 1. Before You Begin... 3 What Is Included... 3 Unpacking Instructions... 3 Claims... 3 Text Conventions... 3 Icons... 3 Document Information... 3 Product at a
More informationWeather Radio Alarm Clock
1200093 User s Guide Weather Radio Alarm Clock Thank you for purchasing your Weather Radio Alarm Clock from RadioShack. Please read this user s guide before installing, setting up, and using your new weather
More informationTipping Bucket Rain Gauge Models 6011-A 6011-B
Tipping Bucket Rain Gauge Models 6011-A 6011-B User s Manual 1165 National Drive Sacramento, CA 95834 800.824.5873 Table of Contents General Information... 1 Introduction... 1 Installation General... 2
More informationSunMaxx Solar Filling Station Operating Instructions
SunMaxx Solar Filling Operating Instructions Content 1. Declaration of conformity... 2 2. Introduction... 2 3. Transportation and unpacking... 4 4. Mounting and commissioning... 5 5. End of operation...
More informationMODEL G300 BRAKE BLEEDER
MODEL G300 BRAKE BLEEDER Installation, Operation & Repair Parts Information Branick Industries, Inc. 4245 Main Avenue P.O. Box 1937 Fargo, North Dakota 58103 REV060616 P/N: 81-0035G 1 THIS PAGE INTENTIONALLY
More informationUSE &CARE GUIDE. Remote Faucet Pump System. See Important Safeguards on page 2
Remote Faucet Pump System USE &CARE GUIDE See Important Safeguards on page 2 An exclamation point within an equilateral triangle is intended to alert user to the presence of important operating and maintenance
More informationWireless Indoor/ Outdoor Thermometer
Wireless Indoor/ Outdoor Thermometer Owner s Manual Please read before using this equipment. ˆ Contents FCC Information... 3 FCC Declaration of Conformity... 5 Preparation... 5 Installing Batteries...
More information123 Industrial Loop Road Paynesville, MN 56362 Phone: 1-800-864-1649 www.master-mfg.com MASTER MANUFACTURING MASTER GARDNER
123 Industrial Loop Road Paynesville, MN 56362 Phone: 1-800-864-1649 www.master-mfg.com MASTER MANUFACTURING MASTER GARDNER Part Number PCD E3 009B MM Rev 1 Nov. 2010 INTRODUCTION The purpose of this manual
More informationNewAir AC-14100E / AC-14100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS
NewAir AC-14100E / AC-14100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS PARTS LIST Front Back FIG.5 FIG.6 1. Control panel 2. Carrying handle 3. Water tank 4. Casters
More informationHollow or Cinder Block Wall Mount Job Aid
Summary This Job Aid covers: Hollow or Cinder Block Wall Mount Job Aid Hollow or Cinder Block wall Mount Bracket Installing the Mount Bracket into Cinder Block This Job Aid supports the Technician audience.
More informationInstruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function
Important Safeguards READ ALL INSTRUCTIONS BEFORE USE. Instruction Manual Automatic Dispensing Hot Water Pot with Reboil Function Image of SP-3015 & SP-3815 SP-3015: 3.0L SP-3815: 3.8L SP-3017: 3.0L (Stainless
More informationFT-100 Fire Training Fog Machine User Manual English
FT-100 Fire Training Fog Machine User Manual English 2015 Antari Lighting and Effects Ltd. User Manual - English Safety Information Please read the following safety information carefully before operating
More informationRI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
More informationInstallation and Operation Manual. Digital Remote Meter for Monitoring System Performance. Version: RM-1
REMOTE METER TM Installation and Operation Manual. Digital Remote Meter for Monitoring System Performance. Version: RM-1 1098 Washington Crossing Road Washington Crossing, PA 18977 USA www.morningstarcorp.com
More informationLine to Refrigerator Ice/Water Dispenser
Standard 18 Line to Refrigerator Ice/Water Dispenser Pump Module Bottled Water How The System Works The FLOJET Bottled Water Dispensing System was designed to pump water from a commercially available 5-gallon
More informationPortable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CHCSCS
Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CHCSCS Voltage rating: 120V~60Hz Power rating : 1400W(Cooling) Power rating : 1350W(Heating) Customer Support : 1-800-474-21477
More informationAutomatic Phone-Out Home Monitoring Systems
Automatic Phone-Out Home Monitoring Systems Power Outage and Freeze Alarm Model Number: THP202 Power Outage, Freeze and Flood Alarm Product Description Model Number: THP201 These monitoring systems are
More informationOPL BASIC. Dosing System for Professional Laundry machines. Contents
OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page
More informationModel 349KLX Remote Display Digital Scale User Instructions
Model 349KLX Remote Display Digital Scale User Instructions P/N UM349KLX Rev 3-080713 1 Model 349KLX Thank you for your purchase of this Health o meter Professional product. Please read this manual carefully,
More informationINSTRUCTIONS: LocknCharge Laptop Carts
INSTRUCTIONS: LocknCharge Laptop Carts www.lockncharge.com Extra Tools required: Hammer, Philips head screwdriver, medium adjustable spanner. (Allen key supplied) (Panel colours for illustration purposes
More informationHigh Efficiency Vaporizing System and Highly Durable Heaters Ensure 25% Increased Output.
High Efficiency Vaporizing System and Highly Durable Heaters Ensure 25% Increased Output. Modular Design for Easy, Cost Effective Maintenance. Standardized Interface Allows Interchangeable Control Modules
More informationElectric Kettle. Model: SK-1717 INSTRUCTIONS MANUAL
Electric Kettle Model: SK-1717 INSTRUCTIONS MANUAL Thank you for your purchase. Please read this Instruction Manual thoroughly before use and keep in a safe place for future reference. READ ALL INSTRUCTIONS
More informationHow To Service A Paper Shredder
Paper shredder 20-2HS 80C POWERSHRED OPERATING INSTRUCTIONS NOTICE D'UTILISATION MANUAL DEL PROPIETARIO BETRIEBSANLEITUNG 20 80 20C 80C 20-2HS Edition 0/0.70.999.0 Edition 0/0 20-2HS 80C Paper shredder
More informationTS93 EMR T/PT/TDE. Surface applied door closer
TS EMR T/PT/TDE Surface applied door closer Installation instructions: Pull side track mount door closer with smoke detector (EMR T) Push side track mount door closer with smoke detector (EMR PT) Double
More informationUB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS
UB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS: Carefully read these instructions before installing your new air-conditioner. AUSTRALIAN AUTOMOTIVE AIR AL00500054E 1 Table
More informationFCC COMPLIANCE STATEMENT FOR AMERICAN USERS
FCC COMPLIANCE STATEMENT FOR AMERICAN USERS This equipment has been tested and found to comply with the limits for a CLASS A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
More informationOperating Manual Color Changing Fountains
Operating Manual Color Changing Fountains 1-877-80-PONDS www.atlanticwatergardens.com Introduction Thank you for purchasing Atlantic s Color Changing Fountains overflowing features that glow from within.
More informationHP ProDisplay P19A LED Backlit Monitor. User Guide
HP ProDisplay P19A LED Backlit Monitor User Guide 2013 Hewlett-Packard Development Company, L.P. The only warranties for HP products and services are set forth in the express warranty statements accompanying
More informationPlease read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan
Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan TABLE OF CONTENTS INTRODUCTION 2 IMPORTANT SAFEGUARDS...2 PACKAGE CONTAINS..2 NAMES OF PARTS.3
More informationAnthro Tablet Charging Cabinets
Owner's Manual for Anthro Tablet Charging Cabinets Covers part numbers beginning TAB16 and TAB20. Components at a Glance The open views of the front and side of the cabinet below are for the TAB20 cabinet.
More informationPortable Evaporative Air Cooler. OWNER S MANUAL Read and save these instructions before use. Model: CL30XC
Portable Evaporative Air Cooler OWNER S MANUAL Read and save these instructions before use Model: CL30XC Power rating: 250 Watts Voltage rating: 230 Volt, 50Hz Made in P.R.C. QUICK START GUIDE Fill with
More informationDeluxe Bottle Warmer
Deluxe Bottle Warmer IMPORTANT SAFEGUARDS This product is for household use only. When using electrical appliances, basic safety precautions should always be followed including the following: Read all
More informationPart Name/Description Part Number Quantity. Power Cable 4000950-5 1
Note: Indented items indicate parts included in an assembly listed above Part Name/Description Part Number Quantity Power Cable 4000950-5 1 Raven Harness Adapter Kit 4100525 1 Installation Instructions
More informationB150 Series Commercial Brewer. Direct Plumb Kit
B150 Series Commercial Brewer Welcome Keurig Authorized Dealer! Direct Plumb Kit This manual provides: Installation of the Direct Plumb Kit (DPK) that allows you to connect your customer s B150 brewer
More informationService Guide. Gateway M275
Service Guide Gateway M275 Contents Replacing Gateway M275 Components.................................... 1 Identifying the convertible tablet PC model...................................... 2 Identifying
More informationCAD-05 Kit GSM Auto Dialer. Owner s Manual
CAD-05 Kit GSM Auto Dialer Owner s Manual CAD-05 Kit Manual.indd 1 Warnings: This device complies with Part 15 of the FCC rules, Operation of this device is subject to the following conditions: 1. This
More informationFG MOISTURE MONITOR Installation & Operation Manual
FG MOISTURE MONITOR Installation & Operation Manual Issue 3.0 7/20/10 1 Contents SERVICE AND TECHNICAL SUPPORT... 2 INSTALLATION:... 3 MOISTURE SENSOR INSTALLATION:... 3 SENSOR CONNECTOR:... 5 MONITOR
More informationSBC90. Abrasive Blast Cabinet Assembly & Operating Instructions
SBC90 Abrasive Blast Cabinet Assembly & Operating Instructions READ ALL INSTRUCTIONS AND WARNINGS BEFORE USING THIS PRODUCT. SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE. This manual provides important
More informationTassimo The Art of the Perfect Cup
USER INSTRUCTIONS Tassimo The Art of the Perfect Cup Thank you for purchasing the Tassimo hot beverage machine. With Tassimo, you can now enjoy your favourite hot beverage any time. Rich filter coffee,
More informationEVAPORATIVE AIR COOLER INSTRUCTION MANUAL
EVAPORATIVE AIR COOLER INSTRUCTION MANUAL Model: SF-614P Please read this manual thoroughly before using this product. Keep in a safe place for future reference. CONTENTS A. PRODUCT INTRODUCTION 2 B. TECHNICAL
More informationMilwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
More informationBasic Spring Motor Roller Shades
Comprehensive Roller Shade Installation Guide Basic Spring Motor Roller Shades ATTENTION!!! READ CAREFULLY! This shade has a reliable long-lasting Spring Motor. The Spring Motor must have proper tension
More information767 Diagnostic System
767 Diagnostic System 3.5v 767 Wall Transformer Patent Pending Listed to: UL 60601-1 CSA C22.2 No. 601 Thank you for purchasing the Welch Allyn 3.5v 767 Wall Transformer. This manual is meant to provide
More informationRevolutionary Tabletop Automation. Prescription Validation, Counting and Filling System
Revolutionary Tabletop Automation Prescription Validation, Counting and Filling System Eyecon is a revolutionary pharmacy automation system Eyecon is fast becoming the preferred choice of pharmacies that
More informationName of Equipment Silver King Model SKMCD1P/C1. This equipment chapter is to be inserted in the appropriate section of the Equipment Manual.
Name of Equipment Silver King Model SKMCD1P/C1 This equipment chapter is to be inserted in the appropriate section of the Equipment Manual. Manufactured exclusively for McDonald s By Silver King Refrigeration,
More informationHydraulic Air Riveter. Operation Manual RC197 RC198. Recommend for mass production line industrial area. Page 1
Hydraulic Air Riveter Recommend for mass production line industrial area Operation Manual RC197 RC198 Page 1 CONTENTS 1 Important Safety Rules 2 Tool Specification 2.1 Nomenclature 2.2 Technical Data 3
More informationUSER MANUAL WARNING! CONTENTS MODEL 1 SPECIFICATIONS READ ALL INSTRUCTIONS BEFORE PROCEEDING. Non-Programmable Single Stage Heat/Cool Thermostat
Builder MODEL 1010 Series Non-Programmable Single Stage Heat/Cool Thermostat USER MANUAL Compatible with low voltage single stage gas, oil or electric heating or cooling systems, including single stage
More informationInstallation Operation Maintenance Troubleshooting
Installation Operation Maintenance Troubleshooting Version 3, 2.1, June/09 Apr/09 1416 1402 Table of Contents Important information 2 Weights & Dimensions 3 Operator s Instructions 4 Flow Diagram 5 Purging
More informationGSM Autodialer Professional GJD700 Speech & Text Autodialer
Text Edit message GSM Autodialer Professional GJD700 Speech & Text Autodialer Introduction The GSM Autodialer Professional works in conjunction with standard alarm systems and makes use of your preferred
More informationAD-01 Slave Auto Dialer. Owner s Manual
AD-01 Slave Auto Dialer Owner s Manual AD-01 Slave Manual.indd 1 10/15/2009 10:20:44 AM 2 AD-01 Slave Manual.indd 2 10/15/2009 10:20:44 AM Features: Programmable entry/exit delay time; select up to 9 (32
More informationDehumidifier Users manual. For Models: DH45S DH65S
Dehumidifier Users manual For Models: DH45S DH65S 950-0062-revD Jan. 9 2007 FORWARD The appearance of the units that you purchase might be slightly different from the ones described in the Manual, but
More informationWireless Alarm System. Wireless Keypad. User s Manual. Choice ALERT. Control all sensors, accessories and remote transmitters from one location
Wireless Alarm System Wireless Keypad User s Manual 1 2 3 4 Choice ALERT Control all sensors, accessories and remote transmitters from one location Table of Contents Important Safeguards 4 Introduction
More informationPlease keep these operating instructions in a safe place.
Please keep these operating instructions in a safe place. Dear Customer: Check this product for visible damage immediately upon receipt. Inform the shipper if there is any shipping damage. Note that damage
More informationSCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT
REV. 05-04- SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT GENERAL Kit Numbers 95-08B, 9006-09B, 876-0A Models For model fitment information, see the P&A Retail Catalog or the Parts and Accessories
More informationExpress5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001
Express5800/120Ed Rack Mount Kit Installation Procedures PN: 455-01607-001 Proprietary Notice and Liability Disclaimer The information disclosed in this document, including all designs and related materials,
More informationUpgrading a GainMaker Line Extender from a Line-Powered Source to a 120 V AC-Powered Source Installation Instructions
Upgrading a GainMaker Line Extender from a Line-Powered Source to a 120 V AC-Powered Source Installation Instructions Overview Audience These installation instructions are intended for all cable system
More informationInstallation Operation Maintenance Troubleshooting Version 2.1, Apr/09 DAYTONA
Installation Operation Maintenance Troubleshooting Version 2.1, Apr/09 1607 DAYTONA Congratulations! All of us at Summit Dental Systems want you to know that your 1607 Daytona Cuspidor has been built with
More informationBottom Loading Water Dispenser
Bottom Loading Water Dispenser Model # 601000 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING DISPENSER. SAVE THIS MANUAL FOR FUTURE
More informationUSER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR
USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting
More informationuser s manual Battery Case model #: SPB3200 Battery Case Charger for Samsung Galaxy S 4
user s manual model #: SPB3200 Charger for Samsung Galaxy S 4 What s Included Unpack the battery case and make sure all accessories are put aside so they will not be lost. hello. USB to Micro USB Cable
More informationUSER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE
More informationSETTING UP YOUR 6000 SERIES TIME RECORDER
INTRODUCTION The Pyramid 6000 Series Time Recorder is an electronic time recorder designed to make payroll processing easier and more efficient. The 6000 series will meet your business needs whether your
More informationel-720 Carbon Dioxide Gas Leak Detector User s Guide
el-720 Carbon Dioxide Gas Leak Detector User s Guide Congratulations on the purchase of the AccuTools el-720 Carbon Dioxide Gas Leak Detector, the most technologically advanced instrument of its kind.
More informationFRONT BUMPER INSTALLATION INSTRUCTIONS 2007-2011 DODGE / MERCEDES SPRINTER
Aluminess Products Inc 9402 Wheatlands Ct. #A Santee, CA 92071 619-449-9930 FRONT BUMPER INSTALLATION INSTRUCTIONS 2007-2011 DODGE / MERCEDES SPRINTER Please read before beginning Stainless steel hardware
More informationWireless Alarm System. Panic Button. User s Manual. Choice ALERT. Control all sensors, accessories and remote transmitters from one location
Wireless Alarm System Panic Button User s Manual Choice ALERT Control all sensors, accessories and remote transmitters from one location Table of Contents Important Safeguards 4 Introduction 5 Activating
More informationHYLA NST Cleaning System
Owner s Manual HYLA NST Cleaning System The HYLA NST Cleaning System aspirates and cleans the air through a waterbased filtration process. The system is intended for household use only. Applications: Usual
More informationOperating Instructions Drill rig DRU160
Operating Instructions Drill rig DRU160 Index 000 / 001 Original operating instructions 10988825 en / 20.10.2009 Congratulations! With a Hydrostress unit from TYROLIT you have chosen a tried and tested
More informationYour Simple Guide to Battery. www.firstalertpro.com. Replacement. Customer Care: 1-800-852-0086. www.firstalertpro.
Previous Menu Your Simple Guide to Battery www.firstalertpro.com Replacement p e t s ts ep -by Customer Care: 1-800-852-0086 FA/1565 9/00 www.firstalertpro.com Table of Contents: page Introduction...............................
More informationOptical Power Meter. Specification & User Manual
Optical Power Meter Specification & User Manual Page 1 of 9 Copyright 2011 reserves the right to modify specifications without prior notice Table of Contents 1. Description and Features.......3 2. Specification......4
More informationFUTURE CALL PICTURE CARE PHONE MODEL: FC-1007 USER MANUAL
FUTURE CALL PICTURE CARE PHONE MODEL: FC-1007 USER MANUAL Please follow instructions for repairing if any otherwise do not alter or repair any parts of device except specified. IMPORTANT SAFETY INSTRUCTIONS
More information50 TO 60 INSTALLATION INSTRUCTIONS v1.0
50 TO 60 INSTALLATION INSTRUCTIONS v1.0 PLEASE READ Thank you for purchasing The TV Shield. Please read the following, before starting the installation process. UYou will need to unscrew the ten (10) black
More informationNESCAFÉ Solution 8/60 Beverage System. Operations Manual
NESCAFÉ Solution 8/60 Beverage System Operations Manual 1 INDEX & WARNINGS INDEX Warnings SECTION: 1 INDEX & WARNINGS SECTION: 2 BASIC PARTS OF THE MACHINE SECTION: 3 LOADING PRODUCTS SECTION: 4 DISPENSING
More information149mm. Walk-Thru Assembly Gate. MODEL NO: 1161, 1167 Owner s Manual. www.regalo-baby.com
149mm 210mm Walk-Thru Assembly Gate MODEL NO: 1161, 1167 Owner s Manual READ ALL INSTRUCTIONS BEFORE ASSEMBLY AND USE OF GATE. KEEP INSTRUCTIONS FOR FUTURE USE. www.regalo-baby.com... Regalo International,
More informationStart Here. 3200 Series. All-in-One Printer. Print Copy Scan. Includes Information on the: ESP 3250 All-in-One Printer ESP 3260 All-in-One Printer
Start Here All-in-One Printer 3200 Series MS SD/HC MMC Print Copy Scan Includes Information on the: ESP 3250 All-in-One Printer ESP 3260 All-in-One Printer Note: A USB 2.0 cable is required to connect
More informationSafety, Operation and Maintenance Manual with Parts List
Safety, Operation and Maintenance Manual with Parts List 20-Gallon Wet/Dry Vac Important Information and Safety Instructions PLEASE READ BEFORE USE! # 961130020 9/10-Rev 1 20-Gallon Wet/Dray Vac TABLE
More informationMAC FAUCETS. Owner s Manual. Models: FA444
MAC FAUCETS Owner s Manual Models: FA444 Installation procedure: 1. Take a moment to view drawings and read special Installation note on page 3 of this manual before proceeding with installation. Connect
More informationArctic Leash Instruction Manual Alaskan Products Company LLC.
Retractable Cord Reel for Engine Block Heaters Arctic Leash Instruction Manual Alaskan Products Company LLC. INSTALLATION INSTRUCTIONS This instruction manual is a reference guide for installing and operating
More informationSCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT
REV. 2015-04-23 SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT GENERAL Kit Numbers 29299-08, 29098-09 Models For model fitment information, see the P&A Retail Catalog or the Parts and Accessories
More informationAccessory Manual Ventilation Section Model Comfort Circle
Accessory Manual Ventilation Section Model Comfort Circle Manual version:. R a Copyright and trademarks All the information and drawings in this manual are the property of Biddle and may not be used (other
More informationRemote Alarm Monitoring System
Remote Alarm Monitoring System Operation and Service Manual Model RA1 RA4 A A Version HELMER SCIENTIFIC 14400 Bergen Boulevard Noblesville, IN 46060 USA PH +1.317.773.9073 FAX +1.317.773.9082 USA and Canada
More informationOperating Guide Manuel de l utilisateur
Operating Guide Manuel de l utilisateur Model RF-110 Modèle RF-110 English Thank you for purchasing this Ringmaster; please read this operating guide carefully before use and keep it for future reference.
More informationDigital Photo Picture Frame. Operation Manual
Digital Photo Picture Frame Operation Manual 20070309 CONGRATULATIONS on your purchase of a Polaroid 7 LCD Digital Photo Picture Frame. Please read carefully and follow all warnings and instructions in
More informationACW200C / ACW200CH. Innovative AutoDrain Technology. Owner s Manual Portable Air Conditioner with Heat
Owner s Manual Portable Air Conditioner with Heat ACW200C / ACW200CH It is important that you read these instructions before using your new purchase and we strongly recommend that you keep them in a safe
More informationCustomer Service 800.780.8889 or visit us online at smithnoble.com. Wood Shutters. Step by Step Installation. Troubleshooting Tips
reflect your style TM Customer Service 800.780.8889 or visit us online at smithnoble.com Wood Shutters Step by Step Installation Troubleshooting Tips Easy Care Instructions STEP 1 Your New Shutter Thank
More informationReplacement Instructions. Warning: During this procedure, keep small parts away from children.
apple ibook G4 Memory Card Replacement Instructions Follow the instructions in this sheet carefully. Failure to follow these instructions could damage your equipment and void its warranty. Note: Written
More informationFASCINATION 700 HVLP TANNING PRO SYSTEM USER MANUAL
FASCINATION 700 HVLP TANNING PRO SYSTEM USER MANUAL Congratulations on choosing the Fascination 700 HVLP Tanning Pro System! Your system includes the following items: 1 Fascination 700 HVLP Tanning Pro
More informationTechnical Update TAA.TU.11093 Rev. 1
http://www.gambro.com/en/usa_tech/ 800-525-2623 303-222-6500 Technical Update TAA.TU.11093 Rev. 1 Effective: 05 APR 2013 CO# 13084 Product: Subject: From: Phoenix Dialysis System Required Electrical Safety
More informationPost Mount Light Installation*
Post Mount Light Installation* *For the general installation of most Post Mount Spotlights, many vehicles may need slight modifications to these instructions. You will need the following tools: High torque
More informationCORNER FRIDGE REFRIGERATION UNIT INSTALLATION AND OPERATION INSTRUCTIONS
CORNER FRIDGE REFRIGERATION UNIT INSTALLATION AND OPERATION INSTRUCTIONS D E F Y I N G C O N V E N T I O N Congratulations on your new Corner Fridge Your new corner fridge may have different functions
More informationRules for Safe Operation
Rules for Safe Operation Important: Do not attempt to operate the CleanStation until you have read and understand all of the instructions in this manual. Failure to do so may result in injury. The machine