Determination of the optimal conditions for dental subtraction radiography using a storage phosphor system

Size: px
Start display at page:

Download "Determination of the optimal conditions for dental subtraction radiography using a storage phosphor system"

Transcription

1 Dentomaxillofacial Radiology (1999) 28, 1±5 ã 1999 Stockton Press All rights reserved 0250 ± 832X/99 $ Determination of the optimal conditions for dental subtraction radiography using a storage phosphor system DS Brettle 1 R Ellwood 2 and R Davies 2 1 NW Medical Physics, and 2 Dental Health Unit, Manchester, UK Objectives: To determine the optimal kvp and exposure conditions for digital subtraction radiography system using a storage phosphor system. Method: Signal-to-noise (SNR) measurements were acquired using a Digora system (Soredex, Helsinki, Finland) of large area, low contrast, clinically realistic details against varying degrees of background attenuation for both single exposure unsubtracted and subtracted images. These results were combined with a measure of estimated thyroid dose to derive a gure of merit (FOM) for the unsubtracted and subtracted images. Results: For both unsubtracted and subtraction radiography, an exposure at 50 kvp and 250 mgy produce the best overall FOM. However, using the system at the 60 kvp maximum a FOM at 1000 mgy for unsubtracted radiography and 500 mgy for subtraction radiography gave the best SNR performance. Conclusion: Operating parameters have been derived which allow the user to choose between optimising SNR and dose (50 kvp, 250 mgy for unsubtracted and subtracted radiography) or SNR alone (60 kvp, 1000 mgy for unsubtracted and 60 kvp, 500 mgy for subtracted), for the visualisation of clinically representative details using the Digora system. Keywords: radiography, dental; digital radiography, dental; subtraction technique; doseresponse relationship Introduction During the last two decades greater emphasis has been given to the early detection and prevention of oral disease. Preventive interventions in the early stages of a disease are often both clinically and nancially more e ective than those performed when it is more advanced. 1 The early detection of destructive periodontal disease and caries progression is, therefore, becoming increasingly important. One technique which has been widely investigated in dentistry, since it facilitates earlier detection of hard tissue pathology, is subtraction radiography. Anatomical structural noise is the predominant limiting factor in the diagnosis of small changes in dental hard tissues and its removal, using subtraction radiography, signi cantly improves the diagnostic yield compared with a single radiograph. 2 has Correspondence to: Mr DS Brettle, North West Medical Physics, Christie Hospital, Manchester M20 4BX, UK Received 2 June 1998; accepted 13 September 1998 been applied in a wide range of dental diagnostic tasks. For example, in periodontology it has been used to evaluate periodontal disease progression 3, periodontal surgery 4 and therapeutic agents. 5 Although subtraction radiography has also been used to detect changes in the status of caries 6 it has not found such widespread acceptance in cariology as in periodontology. One important factor in the implementation of subtraction radiography is the easy access to robust digital data. Direct digital systems that facilitate this are becoming increasing available. We have previously evaluated the clinical e ectiveness of a photostimulable phosphor (PSP) direct digital system compared with lm. 7 This type of system behaves in a di erent way to lm, most notably in respect to its wide exposure latitude. Therefore before it can be used for subtraction radiography an investigation into the optimal operating parameters is needed. Radiographic imaging is a noise-limited process where the quality of the radiographic image is determined not only by how well the signal is reproduced but also on the level of noise. Image quality improves with increased signal strength and

2 2 reduced noise. Therefore image quality should be quanti ed through some measure of the relationship between the two factors. Such a measure is the signalto-noise ratio (SNR) which in its simplest form is the ratio of signal amplitude to the standard deviation in the noise uctuations. 8 Signal-to-noise-ratio is preferable to subjective techniques for this type of study as it allows quantitative measurements which are intrinsically linked to the physical performance of the system. Additionally, details can be chosen which simulate the radiographic response of clinical features and as such re ect actual clinical performance. However, the physical performance of a system is only part of the overall picture and any system that is to be used clinically also needs to consider the risk to the patient. An acceptable measure of risk for an intra-oral radiographic examination would be the dose to the thyroid. The aim of this study was therefore to identify the kvp and radiation exposure for a digital subtraction radiography system using the Digora system which provides the optimal combination of image quality and patient safety. Methods Test object An aluminium step-wedge comprising six steps ranging from 0 ± 2 cm in 4 mm steps was constructed (Figure 1). A 2 mm thick lead stop, equal in area to a step, was included in the base of the step-wedge along one of the long sides. This, combined with the 0 cm step, ensured the detector received exposures ranging from 0 ± 100% of the entrance exposure. Four 5 mm diameter aluminium details on a 0.1 mm Al base plate were placed on each step. The thickness of the details was 0.1, 0.3, 0.5 and 0.7 mm. The test object was designed to cover the range and subtleties of densities that may occur in a clinical image whilst allowing measurement of SNR. Images were acquired for each combination of 50, 60 and 70 kvp and 60, 125, 250, 500, 1000 and 2000 mgy entrance exposures. The X-ray units and the exposure settings are shown in Table 1. The exposure was measured with a 4000 M+ dosemeter (Victoreen Inc., Cleveland, Ohio, USA). The operating range of the Digora (Soredex, Helsinki, Finland) system was de ned before use by reading an image plate uniformly exposed to the highest exposure and kvp to be used clinically. For this study the operating range was set at 70 kvp and 7 mas as determined by the Radiology Department, Manchester Dental Hospital. The same image plate was used for all exposures. Three identical images, containing both the stepwedge and the test details, were acquired for each setting. For the subtraction images a single base image i.e. an image of the stepwedge with the test details removed, was acquired at the same exposure settings as the original unsubtracted images and subtracted from the corresponding original image. All the subtraction images were aligned using the software described elsewhere. 9 Figure 1 Diagram of the test object and a corresponding radiograph at 60 kvp. For details of the test object see Methods Signal to noise ratio The signal-to-noise-ratio (SNR) measurements were calculated using a sampling aperture covering a pixel region. This region was large enough to cover the majority of the detail without including the edges. An identical aperture was also used for measurements on the background. The SNR was calculated using equation 1 where S Sig =mean grey level for the detail, Table 1 Systems and parameters used to achieve the range of radiographic factors Exposure times required to achieve indicated entrance dose(s) kvp Unit Filtration (mm Al) FOD (cm) ma 60 mgy 125 mgy 250 mgy 500 mgy 1000 mgy 2000 mgy Oralix 50 a Heliodent MD b Heliodent MD b a Gendex, Monza, Italy. b Sirona, Bensheim, Germany ±

3 S BG =mean grey level for the background and s BG =standard deviation for the background SNR ˆ SSig S BG BG 1 The signal-noise ratio was measured for the 0.5 mm detail thickness at three di erent step thicknesses 4, 8 and 12 mm Al. This detail was chosen after all the images had been acquired because it was visible on all images and representative of subtle clinical changes. The step thicknesses were similarly chosen to represent a range of bone presentations which covered the typical clinical range. Eight separate SNR measurements were made for each detail choosing di erent background regions each time to include background variability. This was done for each of the three repeat images, so that 24 measurements were made in total for each detail. The mean SNR was recorded along with the standard deviation of the 24 SNR measurements as a measure of spread. FOM ˆ SNR 4SNR 12 2 Thyroid dose Where: SNR 4 =the SNR of the 0.5 mm detail on the 4 mm step SNR 12 =the SNR of the 0.5 mm detail on the 12 mm step This gure thus re ects the performance of each set of operating parameters across the range of potential clinical appearances relative to risk to the patient. 3 Thyroid doses To estimate the relative risk to the thyroid gland for each of the exposures the NRPB 10 organ data was used in conjunction with the XDOSE software (v2.1) (National Radiation Laboratory, Christchurch, NZ). The examination chosen was the lateral head radiograph because the thyroid gland was not directly irradiated in this view. Calculations were made for each of the imaging conditions used. Figure 3 50 kvp SNRs against entrance exposure for the 0.5 mm Figure of merit calculation We now have two measures, SNR and thyroid dose, to describe the performance of the subtraction system under simulated clinical conditions. The use of a gure of merit (FOM) allows a range of data to be presented in a single gure which is designed to increase with increasing SNR and decrease with increasing thyroid dose. The highest FOM is therefore the optimum balance between bene t and risk. The FOM used is developed from the FOM de nition given by Chakraborty and Barnes 11 and is de ned as: Figure 4 60 kvp SNRs against entrance exposure for the 0.5 mm Figure 2 Relative thyroid dose against entrance exposure for a lateral head radiograph without the thyroid gland in the primary beam Figure 5 70 kvp SNRs against entrance exposure for the 0.5 mm

4 4 Figure 6 The same measurements as Figure 3 after subtraction of Figure 9 Relationship of gure of merit (FOM) to thyroid entrance dose for unsubtracted images Figure 7 The same measurements as Figure 4 after subtraction of Figure 10 Relationship of gure of merit (FOM) to thyroid entrance dose for subtracted images Discussion Figure 8 The same measurements as Figure 5 after subtraction of Results Thyroid doses Figure 2 shows the relative thyroid doses for each examination calculated using the NRPB data. SNR measurements The results of the SNR study are shown in Figure 3 ± 8. The combination of the SNR and the thyroid dose data is shown as the FOM in Figures 9 and 10. The aim of this study was to determine the optimal radiographic parameters for subtraction dental radiography using a photostimulable phosphor direct digital system. To constrain the study, it was decided to restrict the parameters that could be varied to kvp and exposure. SNR was chosen as the parameter to optimise as it is a quantitative measure which has a strong basis in imaging theory. 8 The requirement to measure SNR a ected the design of the test object. More speci cally, the details had to be large enough to get relevant measurements, yet not so large as to be clinically unrealistic or prevent the acquisition of background measurements. Using a step-wedge ensured a range of radiographic contrasts consistent with those achievable in dental radiography. Similarly, the thickness of the details were chosen to be clinically relevant and span the threshold of detection. Aluminium was used for the details and the steps because it has a similar attenuation coe cient to tissues containing calcium. The operation of the Digora system is not as straightforward as a conventional radiographic system. It requires the operating range to be set before use and manipulates the data range accordingly. This process, combined with the complex mechanisms of photostimulable phosphor luminescence, makes the

5 interpretation of the Digora results more complex to analyse than a conventional lm-based system. The subtracted results are not as easy to interpret as the unsubtracted. Normally, when two temporally separated images containing small signal changes are subtracted, the signal-to-noise ratio will reduce due to the additive nature of uncorrelated noise sources. This is clearly not happening in the subtracted images. A possible explanation of this could be that the system is only weakly quantum limited, with stronger nonstochastic noise sources. This is indicated in the unsubtracted image by the large deviations in measured SNR. Sources of this noise could be scatter from adjacent steps or dirt on the image plates and/or the test object. Clearly, when the removal of anatomical noise is also considered, the subtracted system may well produce greater clinical SNRs than the unsubtracted images. The SNR results on their own do not give a total picture of the ideal exposure conditions. Obviously, increasing the exposure will increase the detail SNR but it will also increase patient dose. To determine the optimum exposure conditions for clinical use, this factor needs to be considered. This was achieved by estimating the dose to the thyroid for each of the exposure levels used. These doses were not absolute measurements but relative indicators. The results achieved were predictable, with higher exposures giving higher doses and greater penetration, i.e. higher kv, increasing the number of photons reaching the thyroid, and consequently the dose. Combining the results of the SNR measurements with the estimates of thyroid dose allowed a gure of merit to be derived which was sensitive to maximum SNR across the greatest range of steps (clinical representations) and risk to the patient. The results for FOM show that the best overall results for single unsubtracted exposure were obtained at 50 kvp and 250 mgy. However at 60 kvp and 1000 MGy the SNRs are better across the range, albeit for seven times the dose (Figure 9). The results at 70 kv were considerably worse than those at 50 and 60 kvp, suggesting that this may not be a suitable technique to chose. A similar case exists for the subtracted results with an optimum maximum at 50 kvp and 250 mgy (Figure 10). However signi cantly better SNRs were obtained at 60 kvp at the cost of increases in dose of up to four times. For both unsubtracted and subtraction radiography it appears that if optimum SNR/dose performance is desirable then 50 kvp should be used. If optimum SNR is required, then 60 kvp is preferable. In addition to the e ects of varying the exposure parameters, other factors also need to be considered. These include geometric alignment and its e ect on quanti cation of detail volume. Future investigations could also consider the e ects of dual energy subtraction which not only utilises the advantages of subtraction but increases relative radiographic contrast by exploiting the energy-dependent absorption properties of pathological features. Acknowledgements The authors would like to thank the sta of the Radiology Department, Manchester Dental Hospital, for their help 5 References 1. The Department of Health for England. An oral health strategy for England. London. Department of Health GroÈ ndahl K, GroÈ ndahl H-G, Webber RL. Digital subtraction radiography for diagnosis of periodontal bone lesions with simulated high-speed systems. Oral Surg Oral Med Oral Pathol 1983; 55: 313 ± Je coat MK, Reddy MS. A comparison of probing and radiographic methods for detection of periodontal disease progression. Current Opinion in Dentistry 1991; 1: 45 ± BraÈ gger U, Pasquali L, Kornman KS. Remodelling of interdental alveolar bone after periodontal ap procedures assessed by means of computer-assisted densitometric image analysis (CADIA). J Clin Periodontol 1988; 15: 558 ± Je coat MK, Reddy MS, Haigh S et al. A comparison of topical ketorolac systemic urbiprofen, and placebo for the inhibition of bone loss in adult periodontitis. J Periodontol 1995; 66: 329 ± Wenzel A., Halse A. Digital subtraction radiography after stannous uoride treatment for occlusal caries diagnosis. Oral Surg Oral Med Oral Pathol. 1992; 74: 824 ± Brettle DS, Workman A, Ellwood RP, Launders JH, Horner K, Davies RM. The imaging performance of a storage phosphor system for dental radiography. Br. J. Radiol. 1996; 69: 256 ± Workman A, Brettle DS. Review Article: Physical performance measures of radiographic imaging systems. Dentomaxillofac Radiol. 1997; 26: 139 ± Ellwood RP, Davies RM, Worthington HV. Evaluation of a dental subtraction radiography system. J Periodont Res. 1997; 32: 241 ± Hart D, Jones DG, Wall BF. Normalised organ doses for medical X-ray examinations calculated using Monte Carlo techniques. London HMSO NRPB - SR Chakraborty DP, Barnes GT. An energy sensitive cassette for dual-energy mammography Medical Physics. 1989; 16: 7 ± 13.

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Page: 1 of 6 Page: 1 of 6

Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique

More information

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

There is no question: X Rays help dentists care for children. Select x rays for individual s needs, not merely as a routine

There is no question: X Rays help dentists care for children. Select x rays for individual s needs, not merely as a routine 1 Medical Physicists Should Join the Movement Image Gently in Dentistry Safe, Effective Dental X Rays for Children There is no question: X Rays help dentists care for children. As imaging professionals

More information

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department

More information

EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE SYSTEM

EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE SYSTEM www.fob.usp.br/jaos or www.scielo.br/jaos J Appl Oral Sci. 2008;16(5):350-4 EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

Dental Radiography Core Subject. Digital Radiography

Dental Radiography Core Subject. Digital Radiography Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of

More information

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Radiography: 2D and 3D Metrics of Performance Towards Quality Index AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination

More information

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

European Academy of DentoMaxilloFacial Radiology

European Academy of DentoMaxilloFacial Radiology European Academy of DentoMaxilloFacial Radiology Framework for Specialist Training in Dental and Maxillofacial Radiology Background The scope of DentoMaxilloFacial Radiology DMFR (Dental and Maxillofacial

More information

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational

More information

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department

More information

Possibilities and Limits of Digital Industrial Radiology: - The new high contrast sensitivity technique - Examples and system theoretical analysis

Possibilities and Limits of Digital Industrial Radiology: - The new high contrast sensitivity technique - Examples and system theoretical analysis DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France Possibilities and Limits of Digital Industrial Radiology: - The new high contrast

More information

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation Safety Characteristics of the NOMAD Portable X-ray System Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric

More information

Systematic X-ray diagnostics as flexible and individual as your dental practice. XIOS PLUS the intraoral sensor system

Systematic X-ray diagnostics as flexible and individual as your dental practice. XIOS PLUS the intraoral sensor system C A D / C AM SYSTEMS INSTRUMENT S HYGIENE SYSTEMS TRE ATMENT CENTERS IM AGING SYSTEMS XIOS PLUS the intraoral sensor system Systematic X-ray diagnostics as flexible and individual as your dental practice.

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum

More information

Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland

Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland HPA-CRCE-43 Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland A review of data collected by the HPA Dental X-Ray Protection Services between 28 and 211 for intra-oral

More information

Patient Exposure Doses During Diagnostic Radiography

Patient Exposure Doses During Diagnostic Radiography Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University

More information

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS A. Slavtchev 1, I. Manolov 2 1 National Centre of Radiobiology and Radiation Protection 2 Medrom Ltd. Abstract In recent years the number

More information

Systematic X-ray diagnostics as flexible and individual as your dental practice.

Systematic X-ray diagnostics as flexible and individual as your dental practice. CAD/CAM SYSTEMS INSTRUMENTS HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS XIOS PLUS THE INTRAORAL SENSOR SYSTEM Systematic X-ray diagnostics as flexible and individual as your dental practice. T h

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality

More information

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment

More information

A comparison of 18 different x-ray detectors currently used in dentistry

A comparison of 18 different x-ray detectors currently used in dentistry A comparison of 18 different x-ray detectors currently used in dentistry Allan G. Farman, BDS, PhD, DSc, MBA, a and Taeko T. Farman, DMD, PhD, MS, b Louiville, Ky THE UNIVERSITY OF LOUISVILLE Purpose.

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

Oftentimes, as implant surgeons, we are

Oftentimes, as implant surgeons, we are CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the

More information

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Kodak s dental radiograph series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at

More information

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information

More information

Digital Dental X-ray. Midmark Animal Health Products

Digital Dental X-ray. Midmark Animal Health Products Digital Dental X-ray Midmark Animal Health Products Disease you can't see... 60% of periodontal disease lies below the gum line, and many issues aren't detectable via probing or visual inspection.* Don't

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

Effective dose and risk assessment from film tomography used for dental implant diagnostics

Effective dose and risk assessment from film tomography used for dental implant diagnostics Effective dose and risk assessment from film tomography used for dental implant diagnostics N.L. Frederiksen, B.W. Benson and T.W. Sokolowski* The Baylor College of Dentistry and "The Sammons Cancer Center

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

INTRODUCTION. Radiology. Radiology

INTRODUCTION. Radiology. Radiology Radiology Braz Oral Res 2006;20(3):207-13 Radiology Reproducibility of pixel values for two photostimulable phosphor plates in consecutive standardized scannings Reprodutibilidade dos valores de pixels

More information

Integration of multiple direct digital imaging sources in a picture archiving and communication system (PACS)

Integration of multiple direct digital imaging sources in a picture archiving and communication system (PACS) (2003) 32, 337 342 q 2003 The British Institute of Radiology http://dmfr.birjournals.org TECHNICAL REPORT Integration of multiple direct digital imaging sources in a picture archiving and communication

More information

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Standard Test Method for Classification of Film Systems for Industrial Radiography 1 Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

COLLEGE ON MEDICAL PHYSICS

COLLEGE ON MEDICAL PHYSICS COLLEGE ON MEDICAL PHYSICS (30 August - 22 September 2004) Miramare - Trieste, Italy Directors: A. Benini, P. Sprawls and S. Tabakov Local Organizer: L. Bertocchi FINAL PROGRAMME (all lectures were held

More information

Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography 1

Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography 1 Francesco Sardanelli, MD Franco Zandrino, MD Andrea Imperiale, MD Emma Bonaldo, MD Maria G. Quartini, RT Nadia Cogorno, RT Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography

More information

Fundamentals of Cone-Beam CT Imaging

Fundamentals of Cone-Beam CT Imaging Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

Integrated Management & Educational Consultancy

Integrated Management & Educational Consultancy Integrated Management & Educational Consultancy Services MANDEC Radiology Update Seminar for The Dental Team Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email:admin@imecs.freeserve.co.uk

More information

Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose!

Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose! Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose! Dale A. Miles BA, DDS, MS, FRCD(C) Diplomate, American Board of Oral and Maxillofacial Radiology It used to be that panoramic images

More information

Digital Mammography Update: Design and Characteristics of Current Systems

Digital Mammography Update: Design and Characteristics of Current Systems : Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,

More information

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Development of on line monitor detectors used for clinical routine in proton and ion therapy Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary

More information

ENDODONTOLOGY INTRODUCTION. Original Research ABSTRACT

ENDODONTOLOGY INTRODUCTION. Original Research ABSTRACT Original Research Radiovisiography and ultrasound real time imaging in the differential diagnosis of periapical lesions of endodontic origin A comparative in vivo study SHILPA H. BHANDI * SUBHASH T.S.

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

Digital panoramic X-ray for practical diagnostics

Digital panoramic X-ray for practical diagnostics ORTHOPHOS XG 3 instruments treatment centers Imaging systems cad/cam systems ORTHOPHOS XG 3 Digital panoramic X-ray for practical diagnostics System concept Standard panoramic X-ray with proven technology.

More information

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

Contract: FIGM-CT-2000-00061. Measures for Optimising Radiological Information and Dose in Digital Imaging and Interventional Radiology (DIMOND III)

Contract: FIGM-CT-2000-00061. Measures for Optimising Radiological Information and Dose in Digital Imaging and Interventional Radiology (DIMOND III) Contract: FIGM-CT-2000-00061 Measures for Optimising Radiological Information and Dose in Digital Imaging and Interventional Radiology (DIMOND III) Introduction The DIMOND III project has supported and

More information

VistaScan Perio Plus Image plate scanner more time for patient care

VistaScan Perio Plus Image plate scanner more time for patient care VistaScan Perio Plus Image plate scanner more time for patient care Compressed air Suction Imaging Dental care Hygiene Efficiency for work in the dental practice Up to 8 images at once Digital X-ray with

More information

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S.

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S. Bitewing Radiography B.E. DIXON B.D.S., M.Sc., D.P.D.S. Main Indications Detection of Dental Caries Monitoring progression of caries Assessment of existing restorations Assessment of Periodontal status

More information

QUALITY ASSURANCE IN DENTAL RADIOGRAPHY. Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000

QUALITY ASSURANCE IN DENTAL RADIOGRAPHY. Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000 QUALITY ASSURANCE IN DENTAL RADIOGRAPHY Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000 Quality Assurance (QA) Ensure consistently adequate diagnostic information whilst keeping doses

More information

Best Practices in Digital Radiography

Best Practices in Digital Radiography WHITE PAPER Best Practices in Digital Radiography Tracy L. Herrmann, M.Ed., R.T.(R); Terri L. Fauber, Ed.D., R.T.(R)(M); Julie Gill, Ph.D., R.T.(R)(QM); Colleen Hoffman, R.T.(R)(M)(CT); Denise K. Orth,

More information

Effect of various digital processing algorithms on the measurement accuracy of endodontic file length

Effect of various digital processing algorithms on the measurement accuracy of endodontic file length Effect of various digital processing algorithms on the measurement accuracy of endodontic file length Betül İlhan Kal, DDS, a B. Güniz Baksı, PhD, b Nesrin Dündar, PhD, c and Bilge Hakan Şen, PhD, d Izmir,

More information

Measuring Size and Distance in Digital Mammography

Measuring Size and Distance in Digital Mammography On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval

More information

DENTAL RADIOGRAPHIC EXAMINATIONS: RECOMMENDATIONS FOR PATIENT SELECTION AND LIMITING RADIATION EXPOSURE

DENTAL RADIOGRAPHIC EXAMINATIONS: RECOMMENDATIONS FOR PATIENT SELECTION AND LIMITING RADIATION EXPOSURE DENTAL RADIOGRAPHIC EXAMINATIONS: RECOMMENDATIONS FOR PATIENT SELECTION AND LIMITING RADIATION EXPOSURE REVISED: 2012 AMERICAN DENTAL ASSOCIATION Council on Scientific Affairs U.S. DEPARTMENT OF HEALTH

More information

Radiation Safety In Dental Practice

Radiation Safety In Dental Practice Radiation Safety In Dental Practice A study guide and excerpts from The California Radiaton Control Regulations pertaining to dental practice Radiation Safety Protection Program template December 2014

More information

DENTAL Cone beam 3D X-RAY SYSTEM with

DENTAL Cone beam 3D X-RAY SYSTEM with VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic

More information

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Guidelines on education and training for digital radiology Author:

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

What Parents Should Know about the Safety of Dental Radiology.

What Parents Should Know about the Safety of Dental Radiology. What Parents Should Know about the Safety of Dental Radiology. There are many different types of x-ray images (pictures) that can be taken of children in the dental office to assist in diagnosis. These

More information

Radiation protection principles observance in Iranian dental schools

Radiation protection principles observance in Iranian dental schools Iran. J. Radiat. Res., 2010; 8 (1): 51-57 Radiation protection principles observance in Iranian dental schools A. Eskandarlou 1, K. Ghazi-Khanlou Sani 2*, A.R. Mehdizadeh 3 1School of Dentistry, Department

More information

Improving digital image quality for larger patient sizes without compromises

Improving digital image quality for larger patient sizes without compromises Improving digital image quality for larger patient sizes without compromises Christiaan Kok, Technology Physicist Grids Smit Röntgen, Best, The Netherlands 7.5 7 6.5 fiber grids: 25 lp/cm with variable

More information

The convenient and compact CR system for veterinary practices

The convenient and compact CR system for veterinary practices Divario CR-Fvet with dicom PACS DX-R Software CR Systems for the Future The convenient and compact CR system for veterinary practices Divario CR-F20/ F40/ F60/ F80 vet Divario CR-F20/ F40/ F60/ F80 vet

More information

Digital radiography conquers the veterinary world

Digital radiography conquers the veterinary world Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing

More information

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects. RADIATION SAFETY: HOW TO EDUCATE AND PROTECT YOURSELF AND YOUR STAFF John Farrelly, DVM, MS, ACVIM (Oncology), ACVR (Radiation Oncology) Cornell University Veterinary Specialists The Veterinary Cancer

More information

Production of X-rays and Interactions of X-rays with Matter

Production of X-rays and Interactions of X-rays with Matter Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage

More information

Short-cut design of small hydroelectric plants

Short-cut design of small hydroelectric plants Renewable Energy 19 (2000) 545±563 www.elsevier.com/locate/renene Short-cut design of small hydroelectric plants N.G. Voros*, C.T. Kiranoudis, Z.B. Maroulis Department of Chemical Engineering, National

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

Digital Radiography Image Quality: Image Acquisition

Digital Radiography Image Quality: Image Acquisition Digital Radiography Image Quality: Image Acquisition Mark B. Williams, PhD a, Elizabeth A. Krupinski, PhD b, Keith J. Strauss, MS c, William K. Breeden, III, MS d, Mark S. Rzeszotarski, PhD e, Kimberly

More information

The Convenient and Compact CR System for Veterinary Practices

The Convenient and Compact CR System for Veterinary Practices The Convenient and Compact CR System for Veterinary Practices Fire CR+ 20 / 40 / 60 / 80 vet with a throughput of 20 to 70 cassettes per hour and the option of a space - saving wall mount The small and

More information

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors Application Note Case Study Technology Primer White Paper X-RAY IMAGING Emerging Digital Technology - CMOS Detectors Image Sensors X-Ray DETECTORS Scanners Image Processing Custom Solutions In all domains

More information

Required RS Training Info

Required RS Training Info C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness

More information

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.

More information

Radiation Safety In Dental Practice

Radiation Safety In Dental Practice Radiation Safety In Dental Practice A study guide and excerpts from The California Radiaton Control Regulations pertaining to dental practice Radiation Safety Protection Program template December 2013

More information

Professional Education in CBCT

Professional Education in CBCT Professional Education in CBCT Hugh Devlin Where do you get information? 1. GOOGLE 2. WIKIPEDIA 3. JOURNALS eg DMFR but we are all busy people Recent New York Times article Refers to a JADA issue devoted

More information

CONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME

CONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME CONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME NHSBSP Publication No 60 (Version 2) April 2005 Published by: NHS Cancer Screening Programmes The Manor House 260 Ecclesall Road

More information

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT

More information

Residency Competency and Proficiency Statements

Residency Competency and Proficiency Statements Residency Competency and Proficiency Statements 1. REQUEST AND RESPOND TO REQUESTS FOR CONSULTATIONS Identify needs and make referrals to appropriate health care providers for the treatment of physiologic,

More information

IONISING RADIATION. X-rays: benefit and risk

IONISING RADIATION. X-rays: benefit and risk IONISING RADIATION X-rays: benefit and risk Impress Federal Office for Radiation Protection Postfach 10 01 49 D - 38201 Salzgitter Telephone: + 49 (0) 30 18333-0 Fax: + 49 (0) 30 18333-1885 Website: www.bfs.de

More information

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Radiographic Image Production X-ray photons emitted from the x-ray tube interact with the body, exit the patient (exit beam) and interact

More information

Recommendations for a technical quality control program for diagnostic X-ray equipment

Recommendations for a technical quality control program for diagnostic X-ray equipment Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer

More information

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Dale A. Miles BA, DDS, MS, FRCD (C) Diplomate, American Board of Oral and Maxillofacial Radiology Diplomate, American Board

More information

Parkway College of Nursing and Allied Health School of Nursing and Allied Health. BSc (Hons) Diagnostic Radiography and Imaging Module Synopses

Parkway College of Nursing and Allied Health School of Nursing and Allied Health. BSc (Hons) Diagnostic Radiography and Imaging Module Synopses Parkway College of Nursing and Allied Health School of Nursing and Allied Health 4FHH1157 Anatomy, Physiology and Pathology for Imaging 1 Develop a broad understanding of anatomy, physiology, pathology

More information

Use of lead shielding for adult chest CT. patient & radiographer experiences

Use of lead shielding for adult chest CT. patient & radiographer experiences Use of lead shielding for adult chest CT Dose measurements and patient & radiographer experiences Previous published work Doshi S, Negus I, Oduko J, Foetal radiation dose from CT pulmonary angiography

More information

The Measurement of Sensitivity in Fluorescence Spectroscopy

The Measurement of Sensitivity in Fluorescence Spectroscopy The Measurement of Sensitivity in Fluorescence Spectroscopy Among instrumental techniques, fluorescence spectroscopy is recognized as one of the more sensitive. In fluorescence, the intensity of the emission

More information