Effect of various digital processing algorithms on the measurement accuracy of endodontic file length

Size: px
Start display at page:

Download "Effect of various digital processing algorithms on the measurement accuracy of endodontic file length"

Transcription

1 Effect of various digital processing algorithms on the measurement accuracy of endodontic file length Betül İlhan Kal, DDS, a B. Güniz Baksı, PhD, b Nesrin Dündar, PhD, c and Bilge Hakan Şen, PhD, d Izmir, Turkey EGE UNIVERSITY Objective. The aim of this study was to compare the accuracy of endodontic file lengths after application of various image enhancement modalities. Study Design. Endodontic files of three different ISO sizes were inserted in 20 single-rooted extracted permanent mandibular premolar teeth and standardized images were obtained. Original digital images were then enhanced using five processing algorithms. Six evaluators measured the length of each file on each image. The measurements from each processing algorithm and each file size were compared using repeated measures ANOVA and Bonferroni tests (P 0.05). Paired t test was performed to compare the measurements with the true lengths of the files (P 0.05). Results. All of the processing algorithms provided significantly shorter measurements than the true length of each file size (P 0.05). The threshold enhancement modality produced significantly higher mean error values (P 0.05), while there was no significant difference among the other enhancement modalities (P 0.05). Decrease in mean error value was observed with increasing file size (P 0.05). Conclusions. Invert, contrast/brightness and edge enhancement algorithms may be recommended for accurate file length measurements when utilizing storage phosphor plates. (Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;103:280-4) Digital intraoral radiography has been accepted as an alternative to film-based radiography in endodontics. An advantage of digital imaging over conventional radiography is that images are not static, but may be manipulated by image processing to improve image quality. 1 Several image-processing algorithms, mainly adjustments for contrast and brightness have been evaluated for various diagnoses and treatment options. 1-3 It has been suggested that proper application of image processing and filtering may improve both the diagnostic quality and validity of digital images by delineating subtle differences in radiodensity and sharpening the distinction between neighboring tissues. 4-6 Many studies have investigated the effect of image enhancement on the diagnostic quality of digital images obtained either for the detection of periapical lesions and/orprognosisofendodontictreatment. 7,8 Ithasbeen a Assistant professor, Department of Oral Diagnosis and Radiology, b Professor, Department of Oral Diagnosis and Radiology, School of Dentistry, Ege University. c Associate professor, Department of Oral Diagnosis and Radiology, d Professor, Department of Restorative Dentistry and Endodontology, Received for publication Apr 20, 2006; returned for revision May 29, 2006; accepted for publication Jun 1, /$ - see front matter 2007 Mosby, Inc. All rights reserved. doi: /j.tripleo proved that enhanced digital images were diagnostically comparable to film images for the interpretation of root and canal morphology as well as final verification of the outcome of root canal treatment. In studies comparing the image quality of different imaging systems, 1,9,10 ithas been established that digital image enhancement and filtering may increase diagnostic visibility of endodontic file tips. On the contrary, it has been reported by some authors that even the enhancement and filtering cannot aid for the delineation and visibility of thin files. 2,11 Most of these studies have compared the subjective image quality of digital images based on observer performances; and therefore, their major limitation is reliance on subjective assessment of the image. 2,9,12,13 However, it has been advocated that it may be more important to use a computable objective measure to predict diagnostic accuracy rather than subjective assessments. 14 This is particularly important in endodontics because it will influence the final verification of the outcome of root canal treatment as it determines an accurate working length and consequently a possible successful treatment. It is, however, interesting to note that in studies comparing the validity of file length measurements using different imaging systems, the investigators compared the mean error values of the systems only in respect to conventional film measurements. 11,15-17 Itwas clearly stated that when evaluating the accuracy of a diagnostic method, a solid validation method is required to express the true state of disease 280

2 Volume 103, Number 2 Kal et al. 281 and/or measurement. 18,19 Accordingly, they should have compared also the measurements in respect to the true length in order to demonstrate the absolute accuracy of the systems. When those studies about the validity of file length measurements using digital enhancements were evaluated, it was surprising to observe that only one study had evaluated the measurements in respect to the true length of the files. 10 Consequently, it is possible to conclude that the amount and the type of processing that can improve the accuracy of endodontic file length measurements are still debatable. Therefore, the aim of this study was to compare the validity of various processing algorithms with regard to the measurement of endodontic file lengths using Digora storage phosphor plate (SPP) images. MATERIALS AND METHODS Twenty extracted permanent lower premolar teeth with single roots and root canals and endodontic files with ISO sizes 08, 10, and 15 were used for the study. One investigator (B.I.K.) performed the technical procedures for all teeth. Standard access cavities were prepared using a water-cooled diamond fissure bur in a high-speed hand piece. Gates-Glidden drills #2 and #3 were used to enlarge the coronal part of the root canals. Size 08 Hedstrom files (Dentsply Maillefer, Ballaigues, Switzerland) were inserted into each canal until the tip of the file was seen just at the apical foramen. Following the fixation of the files using resin composite material (Filtek Z250, 3M ESPE, St Paul, MN) and a file stopper, the experimental teeth were mounted in acrylic blocks with original premolar and molar teeth at either end creating natural contact points. A 10-mm rectangular orthodontic wire was also mounted in each acrylic block in the same plane as the endodontic file and placed parallel to the root axis. Because there is a radiographic magnification to some extent on any digital image, the wire served as a reference for the measurement of file lengths and as a calibration control for each original and enhanced digital image. The acrylic blocks were then placed on a supporting post (Pentamix, 3M ESPE) with the x-ray cone to allow exact parallel alignment. In order to keep the receptor perpendicular to the beam at all exposures and to provide consistent and reproducible exposition geometry, a Rinn-Endo-ray film holder (Dentsply/Rinn Corporation, Elgin, IL) was used. The standard geometric configuration was fixed at 30 cm source-to-object distance, and zero degrees vertical and horizontal angulations of the x-ray beam. A plexiglass with thickness of 15 mm was inserted between the x-ray tube and the acrylic block to simulate the effect of soft tissue during all exposures. Radiographic images of each experimental Fig. 1. Original and enhanced digital images of file size 08 (O Original; C Contrast/brightness; S Sharpen; E Find edges; I Invert; T Threshold). tooth were obtained with the Digora SPPs (Soredex, Orion Corporation, Helsinki, Finland) with the x-ray unit operating at 65 kvp and 10 ma for 0.16 seconds (Trophy Radiologie, Vincennes, France). Following the exposures, the Digora plates were scanned immediately in the Digora fmx-scanner previously calibrated for a highest exposure of 1 second and the resulting images were transferred as 8-bit TIFF files to a personal computer (LiteOn, Dong Guang, China) with operating system of Microsoft Windows XP (Microsoft Corp, Redmond, WA). Thereafter, the file in each canal was removed and the distance from the file stop to the tip was measured 3 times using a digital caliper to the nearest mm by the same researcher and the mean was recorded as the true length. The same procedure was repeated for sizes 10 and 15 files. Sixty original images were obtained from 20 teeth exposed with 3 different size files. After taking each original radiograph, one researcher manipulated images using the Image J 1.34 software package (National Institutes of Health, Bethesda, MD). Image J was chosen as the processing software because different softwares having similar functions with common names for various processing techniques may have different effects on the image 20 and because original Digora software did not contain sharpen, edge enhancement, and threshold algorithms. Five different processing algorithms, namely contrast and brightness adjustment, find edges, invert, threshold and sharpen were applied as provided by the

3 282 Kal et al. February 2007 Table I. Absolute mean error standard deviation and minimum-maximum values (mm) of endodontic file lengths for various processing algorithms* File size 08 File size 10 File size 15 Mean SD Min-Max Mean SD Min-Max Mean SD Min-Max Original Threshold Sharpness Invert Find edges Con./bright *The numbers in bold show the lowest mean error values for each file size. software (Fig. 1). Thus, there were 360 original and enhanced images to be assessed. Three radiologists and 3 endodontists with a mean age of 35.5 (range 28-56) and mean clinical experience of 11.8 (range 6-28) acted as evaluators and were asked to measure the distance (length) of each file from the apical-most tip of the file to the stopper on each original and enhanced image. The viewing sessions were performed in a darkened room to minimize glare. Digital images were displayed on a 17-inch SVGA color monitor ( pixels; LiteOn) and the measurements were performed using the distance measurement tool provided by the above-mentioned software. All measurements done by observers were first calibrated to the known reference length (10 mm) of the rectangular wire using the above-mentioned software s distance calibration tool. Image J software provided measurements to the nearest mm as well. A mean (from all 6 observers) was calculated for measurements obtained from each image (original and enhanced using 5 different algorithms) and file size. To simplify statistical analysis, the mean error in each reading was calculated as the absolute value of the difference between an observer s measurement and the true length in case negative or positive values would cancel each other out. The mean absolute errors were then compared for all observers for each image processed using different algorithms and each file size using repeated measures analysis of variance (ANOVA) and Bonferroni tests (P.05). Paired t test was performed to compare the observers measurements with the true length for each image processed using different algorithms and each file size with the statistical significance set at.05 for all statistical analyses (SPSS 3.0 for Windows, SPSS Inc, Chiago, IL). RESULTS Repeated measures ANOVA showed significant differences among processing algorithms and 3 file sizes (P.05), but not among the observers (P.05). The results presented in Table I demonstrate the mean absolute error values of 3 endodontic files measured using various enhancement modalities. The inverted images presented the lowest mean error values for size 08 files, while contrast/brightness adjusted images for size 10 files and edge enhanced images for size 15 files (Table I). This means that above-mentioned enhancements provided file lengths that were quite similar to the true length of each file size. However, no significant differences were observed among original, sharpened, inverted, edge enhanced, and contrast/ brightness adjusted images with regard to the mean error value for each of the 3 file sizes (P.05). The threshold enhancement modality produced significantly higher mean error values than other image enhancement modalities for all file sizes, indicating the least accurate measurements (P.05). When the accuracy of the measurements within each image-processing algorithm was compared with respect to the file sizes, a decrease in mean error value was observed with increasing file size for all processing algorithms used in this study (P.05). In other words, the accuracy of measurements increased with the increasing file size. The paired t test revealed that the observers measurements for file sizes 08, 10, and 15 on all images were significantly shorter than the true length (P.05), except for the inverted, contrast/brightness, and edge enhanced images for size 15 file (P.05). DISCUSSION The main aim of image processing is the process of producing images with sufficient detail to reveal the information that is already there but not visible to the naked eye. It is undisputed that diagnostic impact depends on the task, the quality of source data, and the kind of image processing applied. In studies testing the effect of image processing for the evaluation of endodontic length measurements, either contrast and brightness adjustments and/or magnification (zoom) algorithms were evaluated and images were mostly ac-

4 Volume 103, Number 2 Kal et al. 283 quired using charge-coupled device (CCD) sensors. 8,10,21 The present study is the first for evaluating the influence of multiple image-processing algorithms on the measurement accuracy of endodontic file lengths and first for comparing the effect of various processing algorithms on images acquired using only Digora SPPs. In the present study, the enhancement modalities were applied as originally provided by the software to avoid adverse effect of the variety of observer experience and individual image preferences in digital radiography. The use of automatic settings for each enhancement gave an equal chance of comparable performance to all observers. Since every other parameter including the radiation source, display, and measurement conditions and software were kept constant, the results reflect the direct effect of processing algorithms on the measurement accuracy of endodontic files. Although different algorithms provided varying levels of accuracy for each different file thicknesses, all processing algorithms used in this study presented significantly shorter measurements than the true length for each of the file sizes. The only exception occurred with the use of inverted, edge enhanced, and contrast-brightness enhanced images for file size 15, which presented measurements closest to the true length. On the other hand, it was found that the threshold algorithm gave rise to significantly shorter measurements than the other 4 algorithms for each of the file sizes. It was shown that fine details may be lost in threshold enhanced images, particularly file tips easily fade from the image. 8 Accordingly, the threshold modality by rounding off the single image points (pixels) according to a threshold may have made the file tip appear less clear and thus the file length somewhat shorter on digital images. There is evidence that certain enhancements of digital images may improve accuracy of detection and quantification of carious lesions, for the detection of periapical bone lesions and the visualization of root canal anatomy. 7,22,23 However, none of the image-processing algorithms significantly improved the measurement accuracy of size 08 and 10 files in this study and only few of them increased the measurement accuracy of size 15 files. It has been stated that size 08 and 10 files are not adequate for working length radiographs because small file tips fade out and are usually not visible. 24 However, it is sometimes inevitable to use these files in small canals such as in mesiobuccal roots of maxillary and mesial roots of mandibular molars. There are 2 reasons why no difference was found with the processing algorithms for thin files. First, pixel size of Digora images as given by the manufacturer is mm 5%. 25 Onthe other hand, the thicknesses of the size 08, 10, and 15 files at the tip are given as 0.08, 0.10, and 0.15 mm respectively. According to the Nyquist frequency, the spatial resolution of the system must equal twice the spatial frequency of the smallest detail to be detected. 26 Therefore, the minimum size of the detectable detail with Digora images would be 0.14 mm, which is approximate to the tip of file size 15. As can be expected, detectability of the file tip and consequently the accuracy of length measurement increased as the file got thicker and covered more pixels of the Digora image. Another reason for the underestimation of file length measurements for thin files may be that the Digora storage phosphor plates when exposed with exposure times as recommended by the manufacturer, resultant original images exhibit less contrast between different image areas and increased amount of quantum noise. 27,28 This increased amount of noise appearing as a fog on the image cannot be eliminated; in other words, faint details of the thin file tips could not be delineated in images with originally impaired density even with the aid of processing algorithms. At this point, it should be emphasized that with the use of particular processing algorithms (invert and contrastbrightness adjustment) the mean error values were reduced to mm for size 08 and mm for size 10 files, respectively. Although statistically significant, mm divergence from the true length could be regarded as clinically insignificant for size 10 files. Since the performance of processing algorithms improved with the increasing file size, with the use of above-mentioned processing algorithms sizes 10 and above may be recommended when using SPPs for endodontic file length determinations. Exposing the Digora SPPs above the exposure times recommended by the manufacturer might have optimized the response of the resultant images to the digital processing but such exposures exceeded those needed clinically. Another possibility of using size 08 and thinner files for endodontic length determination using storage phosphor plates may be provided by using files of a highdensity metal alloys. In the present study, the measurement of file lengths using 5 image-processing filters were undertaken with the goal of finding the best enhancement algorithm that would reliably increase the measurement accuracy of endodontic file length. In the clinical situation, discrimination of small file tips may become more difficult because of problems in selection of the optimal exposure time and the effects of scatter radiation, as well as differences in bone density. Clinical studies to compare the effects of various processing filters with SPP images will be required for further evaluation of the clinical usefulness of various processing algorithms.

5 284 Kal et al. February 2007 REFERENCES 1. Li G, Sanderink GCH, Welander U, McDavid WD, Näsström K. Evaluation of endodontic files in digital radiographs before and after employing three image processing algorithms. Dentomaxillofac Radiol 2004;33: Friedlander LT, Love RM, Chandler NP. A comparison of phosphor-plate digital images with conventional radiographs for the perceived clarity of fine endodontic files and periapical lesions. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2002;93: Hintze H, Wenzel A, Frydenberg M. Accuracy of caries detection with four storage phosphor systems and E-speed radiographs. Dentomaxillofac Radiol 2002;31: Wenzel A. Computer-aided image manipulation of intraoral radiographs to enhance diagnosis in dental practice: a review. Int Dent J 1993;43: Shrout MK, Russell CM, Potter BJ, Powell BJ, HildeboLt CF. Digital enhancement of radiographs: can it improve caries diagnosis? J Am Dent Assoc 1996;127: Eickholz P, Riess T, Lenhard M, Hassfeld S, Staehle HJ. Digital radiography of interproximal bone loss: validity of different filters. J Clin Periodontol 1999;26: Kullendorff B, Nilsson M. Diagnostic accuracy of direct digital dental radiography for the detection of periapical bone lesions. II. Effects on diagnostic accuracy after application of image processing. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996;82: Shearer AC, Horner K, Wilson NH. Radiovisiography for length estimation in root canal treatment: an in-vitro comparison with conventional radiography. Int Endod J 1991;24: Ellingsen MA, Harrington GW, Hollender LG. Radiovisiography versus conventional radiography for detection of small instruments in endodontic length determination. Part 1. In vitro evaluation. J Endod 1995;21: Woolhiser GA, Brand JW, Hoen MM, Geist JR, Pikula AA, Pink FE. Accuracy of film-based, digital and enhanced digital images for endodontic length determination. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2005;99: Lozano A, Forner L, Liena C. In vitro comparison of root-canal measurements with conventional and digital radiology. Int Endod J 2002;35: Ellingsen MA, Hollender LG, Harrington GW. Radiovisiography versus conventional radiography for detection of small instruments in endodontic length determination. Part 2. In vivo evaluation. J Endod 1995;21: Leddy BJ, Miles DA, Newton CW, Brown CE. Interpretation of endodontic file lengths using Radiovisiography. J Endod 1994;20: Cosman PC, Gray RM, Olshen RA. Evaluating quality of compressed medical images: SNR, subjective rating and diagnostic accuracy. Proceedings of the IEEE 1994;82: Versteeg KH, Sanderink GCH, van Ginkel FC, van der Stelt PF. Estimating distances on direct digital images and conventional radiographs. J Am Dent Assoc 1997;128: Vandre RH, Pajak JC, Abdel-Nabi H, Farman TT, Farman AG. Comparison of observer performance in determining the position of endodontic files with physical measures in the evaluation of dental x-ray imaging systems. Dentimaxillofac Radiol 2000;29: Eikenberg S, Vandre R. Comparison of digital dental x-ray systems with self-developing film and manual processing for endodontic file length determination. J Endod 2000;26: Wenzel A, Hintze H. The choice of gold standard for evaluating tests for caries diagnosis. Dentomaxillofac Radiol 1999;28: Wenzel A, Verdonschot EH. Some considerations in the evaluation of diagnostic tests in dentistry. Dentomaxillofac Radiol 1994;23: Lehmann TM, Troeltsch E, Spitzer K. Image processing and enhancement provided by commercial dental software programs. Dentomaxillofac Radiol 2002;31: Hedrick RT, Dove SB, Peters DD, McDavid WD. Radiographic determination of canal length direct digital radiography versus conventional radiography. J Endod 1994;20: Naoum HJ, Chandler NP, Love RM. Conventional versus storage phosphor-plate digital images to visualize the root canal system contrasted with a radiopaque medium. J Endod 2003;29: Møystad A, Svanaes DB, Risnes S, Larheim TA, Gröndahl HG. Detection of aproximal caries with a storage phosphor system. A comparison of enhanced digital images with dental x-ray film. Dentimaxillofac Radiol 1996;25: Walton R, Torabinejad M. Principles and practice of endodontics. Philadelphia: WB Saunders; p Digora FMX, technical data. Available at: Accessed April 19, Versteeg CH, Sanderink GCH, van der Stelt PF. Efficacy of digital intra-oral radiography in clinical dentistry. J Dent 1997;25: Borg E, Gröndahl HG. Endodontic measurements in digital radiographs acquired by a photostimulable, storage phosphor system. Endod Dent Traumatol 1996;12: Chotas HG, Floyd CE Jr, Dobbins JT 3rd, Ravin CE. Digital chest radiography with photostimulable storage phosphors: signal-to-noise ratio as a function of kilovoltage with matched exposure risk. Radiology 1993;186: Reprint requests: Bilge Hakan Şen, PhD Ege Universitesi Dishekimligi Fakultesi Endodonti Bilim Dali Bornova, Izmir, Turkey bilge.hakan.sen@ege.edu.tr

Dental Radiography Core Subject. Digital Radiography

Dental Radiography Core Subject. Digital Radiography Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of

More information

Oftentimes, as implant surgeons, we are

Oftentimes, as implant surgeons, we are CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the

More information

ENDODONTOLOGY INTRODUCTION. Original Research ABSTRACT

ENDODONTOLOGY INTRODUCTION. Original Research ABSTRACT Original Research Radiovisiography and ultrasound real time imaging in the differential diagnosis of periapical lesions of endodontic origin A comparative in vivo study SHILPA H. BHANDI * SUBHASH T.S.

More information

There is no question: X Rays help dentists care for children. Select x rays for individual s needs, not merely as a routine

There is no question: X Rays help dentists care for children. Select x rays for individual s needs, not merely as a routine 1 Medical Physicists Should Join the Movement Image Gently in Dentistry Safe, Effective Dental X Rays for Children There is no question: X Rays help dentists care for children. As imaging professionals

More information

EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE SYSTEM

EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE SYSTEM www.fob.usp.br/jaos or www.scielo.br/jaos J Appl Oral Sci. 2008;16(5):350-4 EVALUATION OF THE EFFECTS OF PROCESSING DELAYS AND PROTECTIVE PLASTIC CASES ON IMAGE QUALITY OF A PHOTOSTIMULABLE PHOSPHOR PLATE

More information

Determination of the optimal conditions for dental subtraction radiography using a storage phosphor system

Determination of the optimal conditions for dental subtraction radiography using a storage phosphor system Dentomaxillofacial Radiology (1999) 28, 1±5 ã 1999 Stockton Press All rights reserved 0250 ± 832X/99 $12.00 http://www.stockton-press.co.uk/dmfr Determination of the optimal conditions for dental subtraction

More information

Integration of multiple direct digital imaging sources in a picture archiving and communication system (PACS)

Integration of multiple direct digital imaging sources in a picture archiving and communication system (PACS) (2003) 32, 337 342 q 2003 The British Institute of Radiology http://dmfr.birjournals.org TECHNICAL REPORT Integration of multiple direct digital imaging sources in a picture archiving and communication

More information

Periapical radiography

Periapical radiography 8 Periapical radiography Periapical radiography describes intraoral techniques designed to show individual teeth and the tissues around the apices. Each film usually shows two to four teeth and provides

More information

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S.

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S. Bitewing Radiography B.E. DIXON B.D.S., M.Sc., D.P.D.S. Main Indications Detection of Dental Caries Monitoring progression of caries Assessment of existing restorations Assessment of Periodontal status

More information

Removal of a fractured instrument with a new extractor: clinical cases

Removal of a fractured instrument with a new extractor: clinical cases Clinical Removal of a fractured instrument with a new extractor: clinical cases Dominique Martin 1 Introduction The fracture of a root canal instrument during endodontic treatment is a surgical accident

More information

Drastimal Surface Dental Caries in the Mixed Dentition

Drastimal Surface Dental Caries in the Mixed Dentition Scientific Article Comparison of direct digital and conventional radiography for the detection of proximal surface caries in the mixed dentition Kristen K. Uprichard, DMD Brad J. Potter, DDS, MS Carl M.

More information

Accuracy of space analysis with emodels and plaster models

Accuracy of space analysis with emodels and plaster models ORIGINAL ARTICLE Accuracy of space analysis with emodels and plaster models S. Russell Mullen, a Chris A. Martin, b Peter Ngan, c and Marcia Gladwin d Leesburg, Va, and Morgantown, WVa Introduction: The

More information

Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose!

Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose! Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose! Dale A. Miles BA, DDS, MS, FRCD(C) Diplomate, American Board of Oral and Maxillofacial Radiology It used to be that panoramic images

More information

Digital Radiography in Dentistry: Moving from Film-based to Digital Imaging

Digital Radiography in Dentistry: Moving from Film-based to Digital Imaging Digital Radiography in Dentistry: Moving from Film-based to Digital Imaging Gail F. Williamson, RDH, MS Continuing Education Units: 4 hours Online Course: www.dentalcare.com/en-us/dental-education/continuing-education/ce350/ce350.aspx

More information

Clinical Accuracy of Ipex Apex Locator for Measurement of Root Canal Length of Primary Molars

Clinical Accuracy of Ipex Apex Locator for Measurement of Root Canal Length of Primary Molars Original Article Clinical Accuracy of Ipex Apex Locator for Measurement of Root Canal Length of Primary Molars M. Karami Nogorani 1, M. Zare Jahromi 2, Z. Dehghan 3, R. Talaei 4. 1 Associate Professor,

More information

Key words maxillary sinus, posterior teeth roots. Introduction. Tikrit Journal for Dental Sciences 1(2012)81-88

Key words maxillary sinus, posterior teeth roots. Introduction. Tikrit Journal for Dental Sciences 1(2012)81-88 Tikrit Journal for Dental Sciences 1(2012)81-88 The Relation of Maxillary Posterior Teeth Roots to the Maxillary Sinus Floor Using Panoramic and Computed Tomography Imaging in a Sample of Kurdish People

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

The Benefits of Using Arap Locator in Root Canal Treatment

The Benefits of Using Arap Locator in Root Canal Treatment UNSURPASSED ACCURACY UNRIVALED RELIABILITY 1 1. Data on file. Unsurpassed accuracy 1 with convenient, easy-to-read display Fully automatic multifrequency measurements Compact design with small footprint

More information

Condylar position in children with functional posterior crossbites: before and after crossbite correction*

Condylar position in children with functional posterior crossbites: before and after crossbite correction* PEDIATRIC DENTISTRY/Copyright 1980 by The American Academy of Pedodontics/Vol. 2, No. 3 Condylar position in children with functional posterior crossbites: before and after crossbite correction* David

More information

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Standard Test Method for Classification of Film Systems for Industrial Radiography 1 Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately

More information

Endodontics. Colleagues for Excellence. Access Opening and Canal Location

Endodontics. Colleagues for Excellence. Access Opening and Canal Location Endodontics Colleagues for Excellence Spring 2010 Access Opening and Canal Location Published for the Dental Professional Community by the American Association of Endodontists Cover artwork: Rusty Jones,

More information

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come

More information

DENTAL Cone beam 3D X-RAY SYSTEM with

DENTAL Cone beam 3D X-RAY SYSTEM with VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic

More information

In Practice: How Going Digital Will Affect the Dental Office -- Farman et al. 139 (3): 14S -- The Journal of the American Dental Association

In Practice: How Going Digital Will Affect the Dental Office -- Farman et al. 139 (3): 14S -- The Journal of the American Dental Association QUICK SEARCH: Author: Keyword(s): [advanced] HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS Year: Vol: Page: J Am Dent Assoc, Vol 139, No suppl_3, 14S-19S. 2008 American Dental Association

More information

Dental Radiology Related With Pedodontics

Dental Radiology Related With Pedodontics Dental Radiology Related With Pedodontics Presented by: Dr. Rajeev Kumar Singh Radiology The branch of medicine that deals with the use of X- rays and radioactive substances in the diagnosis and treatment

More information

What Parents Should Know about the Safety of Dental Radiology.

What Parents Should Know about the Safety of Dental Radiology. What Parents Should Know about the Safety of Dental Radiology. There are many different types of x-ray images (pictures) that can be taken of children in the dental office to assist in diagnosis. These

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y. 11. The Effect of Abutment Tooth Connection with Extracoronal Attachment using the Three Dimensional Finite Element Method - Part 2. The Construction of Finite Element Model from CT Data - Ando A., Nakamura

More information

Intraoral Radiography: Positioning and Radiation Protection. A Peer-Reviewed Publication Written by Gail F. Williamson, RDH, MS

Intraoral Radiography: Positioning and Radiation Protection. A Peer-Reviewed Publication Written by Gail F. Williamson, RDH, MS Earn 4 CE credits This course was written for dentists, dental hygienists, and assistants. Intraoral Radiography: Positioning and Radiation Protection A Peer-Reviewed Publication Written by Gail F. Williamson,

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

A comparison of 18 different x-ray detectors currently used in dentistry

A comparison of 18 different x-ray detectors currently used in dentistry A comparison of 18 different x-ray detectors currently used in dentistry Allan G. Farman, BDS, PhD, DSc, MBA, a and Taeko T. Farman, DMD, PhD, MS, b Louiville, Ky THE UNIVERSITY OF LOUISVILLE Purpose.

More information

Case Report(s): Uncomplicated Crown Fractures

Case Report(s): Uncomplicated Crown Fractures Case Report(s): Uncomplicated Crown Fractures Tooth fractures can be classified as follows: Uncomplicated crown fracture = fracture limited to the crown of the tooth with dentin exposure but no pulp exposure.

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland

Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland HPA-CRCE-43 Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland A review of data collected by the HPA Dental X-Ray Protection Services between 28 and 211 for intra-oral

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

Tooth anatomy risk factors influencing root canal working length accessibility

Tooth anatomy risk factors influencing root canal working length accessibility Int J Oral Sci (2011) 3: 135-140. www.ijos.org.cn ORIGINAL ARTICLE Tooth anatomy risk factors influencing root canal working length accessibility Lu Tang 1,2, Tuo-qi Sun 1,2, Xiao-jie Gao 1, Xue-dong Zhou

More information

Eliminating effect of Er, Cr: YSGG laser irradiation on the smear layer of dentin

Eliminating effect of Er, Cr: YSGG laser irradiation on the smear layer of dentin Eliminating effect of Er, Cr: YSGG laser irradiation on the smear layer of dentin Jingtao Yu, MD, DDS, and Lu Yan, MD, DDS College of Stomatology, China Medical University, Shenyang, P. R. China Purpose:

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Kodak s dental radiograph series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at

More information

Volume 3, Issue 4 US $6.00

Volume 3, Issue 4 US $6.00 Volume 3, Issue 4 US $6.00 Editor: Allan G. Farman, BDS, PhD (odont.), DSc (odont.), Diplomate of the American Board of Oral and Maxillofacial Radiology, Professor of Radiology and Imaging Sciences, Department

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Digital Dental X-ray. Midmark Animal Health Products

Digital Dental X-ray. Midmark Animal Health Products Digital Dental X-ray Midmark Animal Health Products Disease you can't see... 60% of periodontal disease lies below the gum line, and many issues aren't detectable via probing or visual inspection.* Don't

More information

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Dale A. Miles BA, DDS, MS, FRCD (C) Diplomate, American Board of Oral and Maxillofacial Radiology Diplomate, American Board

More information

Simplified Positioning for Dental Radiology

Simplified Positioning for Dental Radiology Simplified Positioning for Dental Radiology Prepared by: Animal Dental Care Tony M. Woodward DVM, Dipl. AVDC 5520 N. Nevada Ave. Suite 150 Colorado Springs, CO 80918 (719) 536-9949 tw@wellpets.com www.wellpets.com

More information

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit I-Max Touch Range PAN / CEPH / 3D digital panoramic unit Evolutive 3 in 1 panoramic unit 3D A new dimension for a complete diagnosis I-Max Touch 3D Evolutive, simple, fast The panoramic unit realizes complete

More information

Analysis of forces developed during root canal filling by different operators

Analysis of forces developed during root canal filling by different operators original article Analysis of forces developed during root canal filling by different operators Maria Rosa Felix de Sousa G. GUIMARÃES, DDS, MSc 1 Henner Alberto GOMIDE, DDS, MSc, PhD 2 Maria Antonieta

More information

DENTAL RADIOGRAPHIC EXAMINATIONS: RECOMMENDATIONS FOR PATIENT SELECTION AND LIMITING RADIATION EXPOSURE

DENTAL RADIOGRAPHIC EXAMINATIONS: RECOMMENDATIONS FOR PATIENT SELECTION AND LIMITING RADIATION EXPOSURE DENTAL RADIOGRAPHIC EXAMINATIONS: RECOMMENDATIONS FOR PATIENT SELECTION AND LIMITING RADIATION EXPOSURE REVISED: 2012 AMERICAN DENTAL ASSOCIATION Council on Scientific Affairs U.S. DEPARTMENT OF HEALTH

More information

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,

More information

The efficacy of gutta-percha removal using ProFiles

The efficacy of gutta-percha removal using ProFiles Blackwell Science, Ltd The efficacy of gutta-percha removal using ProFiles J. J. Ferreira, J. S. Rhodes & T. R. Pitt Ford Department of Conservative Dentistry, Guy s, King s and St. Thomas Dental Institute,

More information

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational

More information

X-ray (Radiography) - Chest

X-ray (Radiography) - Chest Scan for mobile link. X-ray (Radiography) - Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray produces images of

More information

Clinical Efficacy of Three Different Electronic Apex Locators in Comparison with Radiographic Working Length Determination

Clinical Efficacy of Three Different Electronic Apex Locators in Comparison with Radiographic Working Length Determination Research Article Clinical Efficacy of Three Different Electronic Apex Locators in Comparison with Radiographic Working Length Determination Ayca Yilmaz, (contact author) aycasimsek@gmail.com Selçuk Gokyay

More information

Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography 1

Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography 1 Francesco Sardanelli, MD Franco Zandrino, MD Andrea Imperiale, MD Emma Bonaldo, MD Maria G. Quartini, RT Nadia Cogorno, RT Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography

More information

CHAPTER 10 RESTS AND PREPARATIONS. 4. Serve as a reference point for evaluating the fit of the framework to the teeth.

CHAPTER 10 RESTS AND PREPARATIONS. 4. Serve as a reference point for evaluating the fit of the framework to the teeth. CHAPTER 10 RESTS AND DEFINITIONS A REST is any rigid part of an RPD framework which contacts a properly prepared surface of a tooth. A REST PREPARATION or REST SEAT is any portion of a tooth or restoration

More information

While the prosthetic rehabilitation of

While the prosthetic rehabilitation of Restoring Mandibular Single Teeth with the Inclusive Tooth Replacement Solution Go online for in-depth content by Bradley C. Bockhorst, DMD While the prosthetic rehabilitation of full-arch cases provides

More information

4-1-2005. Dental Clinical Criteria and Documentation Requirements

4-1-2005. Dental Clinical Criteria and Documentation Requirements 4-1-2005 Dental Clinical Criteria and Documentation Requirements Table of Contents Dental Clinical Criteria Cast Restorations and Veneer Procedures... Pages 1-3 Crown Repair... Page 3 Endodontic Procedures...

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:

More information

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,

More information

Digital panoramic X-ray for practical diagnostics

Digital panoramic X-ray for practical diagnostics ORTHOPHOS XG 3 instruments treatment centers Imaging systems cad/cam systems ORTHOPHOS XG 3 Digital panoramic X-ray for practical diagnostics System concept Standard panoramic X-ray with proven technology.

More information

Healing Abutment Selection. Perio Implant Part I. Implant Surface Characteristics. Single Tooth Restorations. Credit and Thanks for Lecture Material

Healing Abutment Selection. Perio Implant Part I. Implant Surface Characteristics. Single Tooth Restorations. Credit and Thanks for Lecture Material Healing Abutment Selection Perio Implant Part I Credit and Thanks for Lecture Material Implant Surface Characteristics!CAPT Robert Taft!CAPT Greg Waskewicz!Periodontal Residents NPDS and UMN!Machined Titanium!Tiunite!Osseotite

More information

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM

More information

ProOne Digital X-ray Imaging Hardware Diagrams and System Requirements

ProOne Digital X-ray Imaging Hardware Diagrams and System Requirements Digital X-ray Imaging Hardware Diagrams and System Requirements Digital X-Ray Note: An Uninterrupted Power Supply (UPS) / Power Conditioner is recommended for the X-Ray. Install a new Planmeca Power Conditioner

More information

CS 9000. 3D Panoramic Cephalometric. Innovation, in reach

CS 9000. 3D Panoramic Cephalometric. Innovation, in reach CS 9000 3D Panoramic Cephalometric Innovation, in reach Innovation made simple We believe in innovation. We always have. In fact, our products have consistently distinguished themselves as groundbreaking

More information

Measuring Size and Distance in Digital Mammography

Measuring Size and Distance in Digital Mammography On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval

More information

We Look Forward to Seeing You at the Hawaii Dental Association Convention. Win a Round Trip for Two to the Neighbor Island!*

We Look Forward to Seeing You at the Hawaii Dental Association Convention. Win a Round Trip for Two to the Neighbor Island!* January 2015 Upcoming HDS Holidays Christmas: Thursday, December 25 Friday, December 26 New Year s: Thursday, January 1 Friday, January 2 HDS Online and DenTel will be available. New Dentists Endodontist

More information

VistaScan Perio Plus Image plate scanner more time for patient care

VistaScan Perio Plus Image plate scanner more time for patient care VistaScan Perio Plus Image plate scanner more time for patient care Compressed air Suction Imaging Dental care Hygiene Efficiency for work in the dental practice Up to 8 images at once Digital X-ray with

More information

Dental caries is an infectious disease caused

Dental caries is an infectious disease caused Emerging Methods of Caries Diagnosis George K. Stookey, Ph.D.; Carlos González-Cabezas, D.D.S., Ph.D. Abstract: Current diagnostic tools used in dental caries detection are not sensitive enough to diagnose

More information

Integrated Management & Educational Consultancy

Integrated Management & Educational Consultancy Integrated Management & Educational Consultancy Services MANDEC Radiology Update Seminar for The Dental Team Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email:admin@imecs.freeserve.co.uk

More information

Intraoral Radiographic Techniques

Intraoral Radiographic Techniques Intraoral Radiographic Techniques Allan G. Farman, BDS, EdS., MBA, PhD; Sandra A. Kolsom, CDA-Emeritus, RDA; ADAA 2014 Council on Education Continuing Education Units: 4 hours Online Course: www.dentalcare.com/en-us/dental-education/continuing-education/ce119/ce119.aspx

More information

A Review of Intraoral Radiology

A Review of Intraoral Radiology Earn 2 CE credits This course was written for dentists, dental hygienists, and assistants. A Review of Intraoral Radiology A Peer-Reviewed Publication Written by Jeffery B. Price, DDS, MS Abstract Dentists

More information

Systematic X-ray diagnostics as flexible and individual as your dental practice. XIOS PLUS the intraoral sensor system

Systematic X-ray diagnostics as flexible and individual as your dental practice. XIOS PLUS the intraoral sensor system C A D / C AM SYSTEMS INSTRUMENT S HYGIENE SYSTEMS TRE ATMENT CENTERS IM AGING SYSTEMS XIOS PLUS the intraoral sensor system Systematic X-ray diagnostics as flexible and individual as your dental practice.

More information

Checklist for the Qualification of Digital Detector Array Systems

Checklist for the Qualification of Digital Detector Array Systems Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality

More information

Powertome Assisted Atraumatic Tooth Extraction

Powertome Assisted Atraumatic Tooth Extraction Powertome Assisted Atraumatic Tooth Extraction White et al Jason White, DDS 1 2 3 Abstract Background: While traditional dental extraction techniques encourage minimal trauma, luxated elevation and forceps

More information

Veraviewepocs 3D R100 & F40

Veraviewepocs 3D R100 & F40 Veraviewepocs 3D R100 & F40 Innovative 3D Reuleaux Full Arch FOV Thinking ahead. Focused on life. Veraviewepocs 3D R100 A New Frontier in X-ray Diagnostics Veraviewepocs 3D R100 has changed the shape of

More information

C-shaped root canal system in mandibular second molars in a Chinese population evaluated by cone-beam computed tomography

C-shaped root canal system in mandibular second molars in a Chinese population evaluated by cone-beam computed tomography doi:10.1111/j.1365-2591.2011.01896.x C-shaped root canal system in mandibular second molars in a Chinese population evaluated by cone-beam computed tomography Q. Zheng 1, L. Zhang 1, X. Zhou 2, Q. Wang

More information

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,

More information

Scanners and How to Use Them

Scanners and How to Use Them Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types

More information

ENGLISH. Planmeca ProMax 3D concept

ENGLISH. Planmeca ProMax 3D concept ENGLISH Planmeca ProMax 3D concept Profound understanding of anatomy Planmeca ProMax 3D concept is an intelligent and multipurpose X-ray unit series designed to obtain complete information on patient anatomy

More information

Taylor Dental Assisting School Course Description

Taylor Dental Assisting School Course Description Taylor Dental Assisting School Course Description Entry Level Dental Assisting The Entry Level Dental Assisting Course is divided into Twenty Six (26) modules of four hours each. Total time is One Hundred

More information

Retrospective analysis of factors influencing the eruption of delayed permanent incisors after supernumerary tooth removal

Retrospective analysis of factors influencing the eruption of delayed permanent incisors after supernumerary tooth removal Retrospective analysis of factors influencing the eruption of delayed permanent incisors after supernumerary tooth removal R.A.E. BRYAN*, B.O.I. COLE**, R.R. WELBURY* ABSTRACT. Aim This was to assess the

More information

Veraview IC5 HD High definition, digital imaging excellence. Thinking ahead. Focused on life.

Veraview IC5 HD High definition, digital imaging excellence. Thinking ahead. Focused on life. Veraview IC5 HD High definition, digital imaging excellence Thinking ahead. Focused on life. Thinking ahead. Focused on life. 2 3 IC5 HD Super High-Speed with High Definition Clarity For dental radiology

More information

THE SELECTION OF PATIENTS FOR DENTAL RADIOGRAPHIC EXAMINATIONS

THE SELECTION OF PATIENTS FOR DENTAL RADIOGRAPHIC EXAMINATIONS THE SELECTION OF PATIENTS FOR DENTAL RADIOGRAPHIC EXAMINATIONS REVISED: 2004 AMERICAN DENTAL ASSOCIATION Council on Dental Benefit Programs Council on Scientific Affairs U.S. DEPARTMENT OF HEALTH AND HUMAN

More information

ENGLISH. Planmeca ProMax 3D Mid

ENGLISH. Planmeca ProMax 3D Mid ENGLISH Planmeca ProMax 3D Mid Genuine all-in-one unit Planmeca ProMax 3D Mid is a genuine all-in-one CBVT (Cone Beam Volumetric Tomography) unit including 3D imaging, digital panoramic, digital cephalometric

More information

X-ray (Radiography), Chest

X-ray (Radiography), Chest X-ray (Radiography), Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray makes images of the heart, lungs, airways,

More information

Residency Competency and Proficiency Statements

Residency Competency and Proficiency Statements Residency Competency and Proficiency Statements 1. REQUEST AND RESPOND TO REQUESTS FOR CONSULTATIONS Identify needs and make referrals to appropriate health care providers for the treatment of physiologic,

More information

Measuring Dental Implant Safety and Performance

Measuring Dental Implant Safety and Performance ISSN 181-513. Russian Journal of Biomechanics. 015. Vol. 19, No. 3: 36-43 MOBILITY OF DENTAL IMPLANTS: NEW CAPABILITY OF A KNOWN DEVICE V.A. Eroshin 1, M.V. Dzhalalova 1, A.V. Boyko 1, S.D. Arutyunov,

More information

Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery. Consequences of tooth loss.

Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery. Consequences of tooth loss. Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery Molars The wide occlusal surface is designed for food grinding. The surface needs to be aligned with the

More information

For citation purposes, the electronic version is the definitive version of this article: www.cda-adc.ca/jcda/vol-73/issue-5/409.

For citation purposes, the electronic version is the definitive version of this article: www.cda-adc.ca/jcda/vol-73/issue-5/409. Professional I s s u e s Some Current Legal Issues that May Affect Oral and Maxillofacial Radiology: Part 1. Basic Principles in Digital Dental Radiology David S. MacDonald-Jankowski, BDS, BSc (Hons),

More information

Exhibit 9 ADEA Competencies for Entry into the Allied Dental Professions (As approved by the 2010 ADEA House of Delegates)

Exhibit 9 ADEA Competencies for Entry into the Allied Dental Professions (As approved by the 2010 ADEA House of Delegates) Exhibit 9 ADEA Competencies for Entry into the Allied Dental Professions (As approved by the 2010 ADEA House of Delegates) Introduction In 1998 99, the Section on Dental Hygiene of the American Association

More information

Clinical usefulness of teleradiology in general dental practice

Clinical usefulness of teleradiology in general dental practice Imaging Science in Dentistry 2013; 43: 99-104 http://dx.doi.org/10.5624/isd.2013.43.2.99 Clinical usefulness of teleradiology in general dental practice Jin-Woo Choi Department of Oral and Maxillofacial

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information