1 Belgian dosimetry Audits in RadioTherapy B. Schaeken, R. Cuypers, S. Lelie, W. Schroeyers, F. Sergent, S. Vynckier, A. Rijnders, D. Verellen, H. Janssens
2 Organisation: national audit - 31 locations - 79 linac s photon beams electron beams Visited audits / 3 y time span; all Belgian clinical beams Basic mechanical checks: isocentrum; lasers; telemeter; light field correspondence Basic dosimetry checks: dose measurements in water on beam axis at pre-defined depths - photon beams: 11 dose measurements all types - electron beams: 2 dose measurements in reference conditions
3 first large-scale audit based on alanine/ EPR dosimetry requested dose: 4 Gy, 4 pellets: u c (D w ) = 1.1 % (k =1) MU should be calculated with TPS as would be done for patients sad ssd dref alanine 1 cm x 1 cm dref sad set-up ssd set-up δ (D meas -D center )/ D center within optimal level: δ 3% dose determined with individual base functions Anton: Appl.Rad.Isot 62, 25; Phys.Med.Biol 51, 26 measured dose expressed as dose normalized amplitudes
4 sleeve b fading fading b T T Q Q M j j M j j b b m D k k k k k k k with k D m m A = = Α = =, 1 1 PTB-traceable
5 Dosimetry audits: photon beams Exp. Experiment Depth N r (cm) Irradiation distance Field size Accessory N r det Requested dose ref 1 Ref. field dref ssd or sad 1cm x 1cm no 1 4 Gy Tray factor dref ssd or sad 1cm x 1cm tray 2 4 Gy Energy open beam 4 Energy wedged beam 1&2 ssd or sad 1cm x 1cm no 3,4 4 Gy 3.3 1&2 ssd or sad 1cm x 1cm wedge 5,6 4 Gy Output factor1 8 ssd or sad 6cm x 6cm no 7 4 Gy Output factor2 8 ssd or sad 8cm x 2cm no 8 4 Gy Output factor3 8 ssd or sad 2cm x 8cm no 9 4 Gy Output factor4 8 sad or ssd 2cm x 2cm no 1 4 Gy MLC 1 symm/irreg 1 MLC 2 asymm/irreg 11 MLC 3 asymm/irreg.+ wedge 8 ssd or sad circular no 11 4 Gy ssd or sad 15cm x 12cm no 12 4 Gy ssd or sad 12cm x 8 cm wedge 13 4 Gy 3.11
6 Dosimetry audits: electron beams BELdART Exp Nr Experiment Depth (cm) Irradiation distance Field size accesso ry Nr det Requested dose ref 12 Ref. field_ MeV1 dref ssd or clinical practice 1cm x 1cm no 14 4 Gy Ref. field_ MeV2 (R 5 > 7 cm) dref ssd or clinical practice 1cm x 1cm no 15 4 Gy 4.14 involving 68 alanine pellets
7 1. Actual status: 36 linacs: Varian: 11; Siemens: 7; Elekta: 16; Novalis: 1; BrainLabAB/MHI Vero : 1 Dosimetry was checked in 64 photon beams: 4x 4MV; 1x 5MV; 32x 6MV; 1x 1MV; 17x 15MV; 9x 18MV 54 electron beams: 2x 4MeV; 1x 5MeV; 18x 6MeV; 1x 7MeV; 3x 8MeV; 2x 9MeV; 3x 12MeV; 1x 14MeV; 9x 15MeV; 12x 18MeV; 1x 2MeV; 1x 25MeV For 1 st and 2 nd run measurements in photon beams: D meas / D stated =.999, σ =.13 (#83) D alanine / D ionometry = 1.1, σ =.8 (#98)
8 2. Traceability: Alanine vs. ionisation cham ber: re fe re nce and non re fe re nce beams N = 98 mean = 1,1 σ =,6 range =,31 Alanine vs. ionisation chamber: ref. beams N = 17 mean =,998 σ =,5 range =,16 5,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 D w, alanine /D w, ion chamber
9 3.Results MV: Photon ref beam output N = 56 mean = 1, σ =,1 range =, Quality Index N = 62 mean =,997 σ =,1 range =, ,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 D w, meas /D w, center,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 QI meas /QI center (TPR2/1)
10 3.Results MV: Photon wedged beam output N = 12 mean =,998 σ =,17 range =, Photon beam output with collimator opening N = 372 mean = 1, σ =,15 range =,85 5 2,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 D w, meas /D w, center,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 D w, meas /D w, center
11 3.Results MV: MLC1 MLC2 MLC3 Photon irr. Field output MLC1 MLC2 MLC3 N = mean = 1,,998,993 σ =,16,15,17 range =,71,77,72,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 D w, meas /D w, center
12 kq,q in high-energy electron beams: SL25: 4, 6, 8, 1, 12, 15, 18, 2, 22 MeV ROOS chamber cross calibrated vs. Farmer at 18 MeV / NCS-18 protocol Roos/Farmer = 1.18 Alanine (8 Gy): ur(dw) =.3%; Roos/Farmer: ur(dw) = 1.8% (NCS-18/pp69) Two types of holder MeV kq,q
13 Results MeV: All electron beams N = 57 mean =,994 σ =,25 range =, Low energy electron beams N = 29 mean =,996 σ =,21 range =,94,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 4 2,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 D w, meas /D w, center D w, meas /D w, center High energy electron beams N = 28 mean =,992 σ =,28 range =,133 2,9,92,94,96,98 1, 1,2 1,4 1,6 1,8 1,1 D w, meas /D w, center
14 3.Results summary (latest): output measurements Dmeas/ Dcenter 1sd R(min max) # photon ref beam QI (open beam) wedged photon beam regular field sizes MLC electron beams Results are incl. 2 nd and 3 th run measurements If deviations are observed, it s likely that they are systematic, even within optimal level Finding explanations for a deviation needs a detailed look-up Out of tolerance situations disappear in a 2 nd run, although a clear explanation is difficult to find
15 4. Conclusion: alanine/epr was successfully used as transfer dosimetry system allowing sufficient accurate dose measurements results into accordance with (reference) ionometry beam data are well modeled in TPS for the visited centers suits well for mailed dosimetry (practicalities) but the true benefit of alanine/epr lays in complex dose deliveries (helical, IMRT, SRS, brachytherapy...)
16 Acknowledgments: Hilde Beutjens (lab ass) and Mathias Anton (PTB); fully supported by FANC-AFCN Visit us at Info
Worldwide Quality Assurance networks for radiotherapy dosimetry
Worldwide Quality Assurance networks for radiotherapy dosimetry Joanna Izewska Dosimetry and Medical Radiation Physics Section dosimetry@iaea.org Audit in radiotherapy dosimetry Objectives to enhance confidence
Challenges in small field MV photon dosimetry
Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland maria.aspradakis@ksl.ch Why are
Clinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline
Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50
What is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?
Congreso Internacional Conjunto Cancún 2004 LAS/ANS-SNM-SMSR/International Joint Meeting Cancun 2004 LAS/ANS-SNM-SMSR XV Congreso Anual de la SNM y XXII Reunión Anual de la SMSR/XV SNM Annual Meeting and
ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS. Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic
ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic Disclosure I have received research grants from Siemens
MOC Exam Preparation Therapy
MOC Exam Preparation Therapy R. Alfredo C. Siochi, PhD., DABR Overview MOC Exam Requirement Available Guidance What I did What I wished I did 1 MOC Exam Requirement Rolling Window of 10 years Must pass
ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY
TECHNICAL REPORTS SERIES No. 398 Absorbed Dose Determination in External Beam Radiotherapy An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water Sponsored by the
Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care
Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require
3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
ACCELERATORS AND MEDICAL PHYSICS 2
ACCELERATORS AND MEDICAL PHYSICS 2 Ugo Amaldi University of Milano Bicocca and TERA Foundation EPFL 2-28.10.10 - U. Amaldi 1 The icone of radiation therapy Radiation beam in matter EPFL 2-28.10.10 - U.
Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.
Efficacy of IMRT with dot decimal compensators for radiotherapy of malignant pleural mesothelioma post extrapleural pneumonectomy Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan
High-Energy Electron Beam Therapy Dosimetry with Ionisation Chambers
Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology
The feasibility of a QA program for ISIORT Trials
Baveno, Italy June 22-24, 2012 The feasibility of a QA program for ISIORT Trials Frank W. Hensley 1, Don A. Goer 2, Sebastian Adamczyk 3, Falk Roeder 1, Felix Sedlmayer 4, Peter Kopp 4 1 University Clinics
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams Peter R. Almond Brown Cancer Center, Louisville, Kentucky 40202 Peter J. Biggs Department of Radiation Oncology,
Quality Assurance of accelerators; the technologists responsibility
Quality Assurance of accelerators; the technologists responsibility Christa Timmermans Radiation Technologist, Erasmus MC- Daniel den Hoed, Rotterdam, The Netherlands EORTC-ROG RT technologists Section
Quality Assurance in Stereotactic. Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
External dosimetry Dosimetry in new radiotherapeutic techniques
External dosimetry Dosimetry in new radiotherapeutic techniques Albert Lisbona Medical Physics Department CLCC René Gauducheau 44805 Saint-Herblain France a-lisbona@nantes.fnclcc.fr Objective : To describe
Variance reduction techniques used in BEAMnrc
Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,
Chapter XIII. PDD TAR TMR Dose Free Space
Ver 2.11 Chapter XIII PDD TAR TMR Dose Free Space Joseph F. Buono RTT Allied Health Science Nassau Community College 1 Education Drive Garden City, NY 11530-6793 phone: 516-572 - 7536 oice - 9460 Secretary
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
QA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth
Executive Office of Health and Human s Office of Medicaid www.mass.gov/masshealth May 2015 TO: FROM: RE: Radiation Oncology Centers Participating in Daniel Tsai, Assistant Secretary and Director of (2015
Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y
Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How
Horizon Blue Cross Blue Shield of New Jersey 2012 Radiation Therapy Payment Rules
55875 55876 55876 Description Transperineal placement of needles or catheters into prostate for interstitial radioelement application, with or wout cytosocopy 55920 Placement of interstitial device(s)
Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY
Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY D.I. THWAITES Department of Oncology Physics, Edinburgh Cancer Centre, University of Edinburgh, Western General Hospital, Edinburgh, United Kingdom
Performance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
RADIATION SHIELDING DESIGN 2010
RADIATION SHIELDING DESIGN 2010 ACMP 2010 SAN ANTONIO, TEXAS MAY 23, 2010 DANIEL G. HARRELL SHIELDING CONSTRUCTION SOLUTIONS, INC. 2010 Shielding Construction Solutions 1 2010 Shielding Construction Solutions
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS JOHN L. HORTON Department of Radiation Physics University of Texas
1. Provide clinical training in radiation oncology physics within a structured clinical environment.
Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
SEAAPM Symposium The Practice of Quality Assurance in an Era of Change
SEAAPM Symposium The Practice of Quality Assurance in an Era of Change Directors: Dan Bourland and Ingrid Marshall Vendors, breaks, and food next door Thank you to SEAAPM EXCOM, Commercial Partners, Faculty,
Specifications Varian Treatment Streamlined Treatment Delivery Management Application. Specifications
Specifications Varian Treatment Streamlined Treatment Delivery Management Application i Varian Treatment Streamlined Treatment Delivery Management Application Specifications Specifications Varian Treatment
Principles of dosimetry The ionization chamber
Principles of dosimetry The ionization chamber FYS-KJM 4710 Audun Sanderud Department of Physics Ionometry 1) Ionometry: the measurement of the number of ionizations in substance The number of ionizations
Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation
Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation Deborah Schofield, Laurence Court, Aaron Allen, Fred Hacker, Maria Czerminska Department of Radiation Oncology Dana
Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy
Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy This set of 117 slides is based on Chapter 11 authored by M.D.C. Evans of the IAEA publication (ISBN 92-0-107304-6):
3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage
3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter
QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS
QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS Ben Mijnheer Agnieszka
Our Department: structure and organization
EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled
NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer
Evaluation of complexity and deliverability of IMRT- treatment plans Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer June 11 Abstract Background: Intensity modulated beams are used
RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
Advanced variance reduction techniques applied to Monte Carlo simulation of linacs
MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave
Secondary Neutrons in Proton and Ion Therapy
Secondary Neutrons in Proton and Ion Therapy L. Stolarczyk Institute of Nuclear Physics PAN, Poland on behalf of WG9 EURADOS Acknowledgments EURADOS Workig Group 9 Roger Harrison Jean Marc Bordy Carles
Characterisation of two new ionisation chamber types for use in reference electron dosimetry in the UK
NPL REPORT DQL-RD001 Characterisation of two new ionisation chamber types for use in reference electron dosimetry in the UK Julia A D Pearce September 2004 September 2004 CHARACTERISATION OF TWO NEW IONISATION
Process Optimization in LINAC QA
Process Optimization in LINAC QA STARCHECK The high-resolution ionization chamber matrix for precise profile measurement STARCHECK Excellent, Unpreceden The profiles are measured along the main axes and
Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015
INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
7/21/2014. Quality Metrics and Risk Management with High Risk Radiation Oncology Procedures. Disclosure of Conflict of Interest
Quality Metrics and Risk Management with High Risk Radiation Oncology Procedures V. Sathiaseelan, PhD Chief Physicist Department of Radiation Oncology, Northwestern Memorial Hospital Associate Professor
Independent corroboration of monitor unit calculations performed by a 3D computerized planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and
1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range
1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range The radiation from orthovoltage units is referred to as x-rays, generated by bombarding
Scripps Proton Therapy Center: Configuration and Implementation
Scripps Proton Therapy Center: Configuration and Implementation Anthony Mascia Proton Symposium 2013 Annual AAPM Meeting Indianapolis, Indiana USA Facility Configuration Scripps Proton Therapy Center ***All
UNIVERSITY OF PÉCS, FACULTY OF HEALTH SCIENCES PH.D. SCHOOL OF HEALTH SCIENCES
UNIVERSITY OF PÉCS, FACULTY OF HEALTH SCIENCES PH.D. SCHOOL OF HEALTH SCIENCES HEAD OF THE PH.D. SCHOOL: JÓZSEF BÓDIS, PROF., M.D., D.SC., TEACHER OF THE UNIVERSITY, RECTOR DEVELOPING ADVANCED 3D CONFORMAL
Albany Medical Center Chief Medical Physicist Radiation Oncology Imaging and Related Services
Albany Medical Center Chief Medical Physicist Radiation Oncology Imaging and Related Services POSITION SUMMARY: The Medical Physicist is responsible for assuring the high quality radiation therapy through
TEACHING RADIOTHERAPY PHYSICS CONCEPTS USING SIMULATION: EXPERIENCE WITH STUDENT RADIOGRAPHERS IN LIVERPOOL, UK
TEACHING RADIOTHERAPY PHYSICS CONCEPTS USING SIMULATION: EXPERIENCE WITH STUDENT RADIOGRAPHERS IN LIVERPOOL, UK M C Kirby, PhD Directorate of Medical Imaging and Radiotherapy, School of Health Sciences,
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
Dosimetric impact of the 160 MLC on head and neck IMRT treatments
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Dosimetric impact of the 160 MLC on head and neck IMRT treatments Prema Rassiah-Szegedi, a Martin Szegedi, Vikren Sarkar, Seth Streitmatter,
Clinical Study Monte Carlo-Based Dose Calculation in Postprostatectomy Image-Guided Intensity Modulated Radiotherapy: A Pilot Study
Radiotherapy Volume 215, Article ID 682463, 6 pages http://dx.doi.org/1.1155/215/682463 Clinical Study Monte Carlo-Based Dose Calculation in Postprostatectomy Image-Guided Intensity Modulated Radiotherapy:
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
Qualitätsmanagement-Handbuch
U r Kalibrier- und Messmöglichkeiten Qualitätsmanagement-Handbuch Relative erweiterte Messunsicherheit, die sich aus der Standardmessunsicherheit durch Multiplikation mit dem Erweiterungsfaktor k = 2 ergibt.
World-first Proton Pencil Beam Scanning System with FDA Clearance
Hitachi Review Vol. 58 (29), No.5 225 World-first Proton Pencil Beam Scanning System with FDA Clearance Completion of Proton Therapy System for MDACC Koji Matsuda Hiroyuki Itami Daishun Chiba Kazuyoshi
Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015)
MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015) CODE CODE TYPE Description 0330 REVCODE RADIOLOGY/THERAPEUTIC - GENERAL CLASSIFICATION 0333 REVCODE RADIOLOGY/THERAPEUTIC
IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect Vincent.bossier@iba-group.com. Protect, Enhance and Save Lives
Vincent Bossier System Architect Vincent.bossier@iba-group.com IBA Proton Therapy Biomed days 2015 Protect, Enhance and Save Lives 1 Agenda AN INTRODUCTION TO IBA WHY PROTON THERAPY CLINICAL WORKFLOW TREATMENT
Therapy with protons and ion beams
Therapy with protons and ion beams Biomedical Physics Lecture WiSe 2012/13 many slides from J. Willkens (TUM) Key quesions: How does ion cancer therapy work? What is the physics behind it? How can physics
Status of Radiation Safety System at
Status of Radiation Safety System at Taiwan Photon Source Joseph C. Liu Radiation and Operation Safety Division National Synchrotron Radiation Research Center, Taiwan NSRRC layout 1.5 GeV, 120m, 400 ma
Analysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA
Cancer Research and Oncology: Open Access Received: Nov 16, 2015, Accepted: Jan 12, 2016, Published: Jan 15, 2016 CancerResOncol, Volume 1, Issue 1 http://crescopublications.org/pdf/crooa/crooa-1-001.pdf
Medical Physicist Courses in India
Goverment of India Atomic Energy Regulatory Board Radiological Safety Division Date: 27/06/2014 Contents Medical Physicist Courses in India 1 Minimum qualification for Medical Physicist/RSO... 2 1.1 Medical
Development of on line monitor detectors used for clinical routine in proton and ion therapy
Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of
A Thesis. entitled. Retrofitted Micro-MLC SRS System. George H. Hancock
A Thesis entitled Development of a Comprehensive Linac-based Quality Assurance Program for a Retrofitted Micro-MLC SRS System. by George H. Hancock Submitted to the Graduate Faculty as partial fulfillment
Veterinary Oncology: The Lumps We Hate To Treat
Veterinary Oncology: The Lumps We Hate To Treat Michelle Turek, DVM, DACVIM (Oncology), DACVR (Radiation Oncology) College of Veterinary Medicine University of Georgia Athens, GA Veterinary Oncology Veterinary
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
METHODS FOR IN VIVO DOSIMETRY IN EXTERNAL RADIOTHERAPY
METHODS FOR IN VIVO DOSIMETRY IN EXTERNAL RADIOTHERAPY METHODS FOR IN VIVO DOSIMETRY IN EXTERNAL RADIOTHERAPY Jan VAN DAM Emeritus, Division of Radiation Physics - Department of Radiotherapy University
$ " %#&&%"'( &%")$#*((+,-.##/ $01&0112 344 5*/(&%"2&0116 '(! (## " *# )#! &%") 78 ' ** #* ( 9! '(' :! 9* ## '$ %# & &$%%&%!
78/0
The ESTRO-EQUAL Quality Assurance Network for Photon and Electron Radiotherapy Beams in Germany
Strahlentherapie und Onkologie Original Article The ESTRO-EQUAL Quality Assurance Network for Photon and Electron Radiotherapy Beams in Germany Ivaldo H. Ferreira 1, Jürgen Richter 2, Andrée Dutreix 1,
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
THE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
The Calibration Chain: Role of BIPM,
The Calibration Chain: Role of BIPM, PSDLs and ADCLs J. Seuntjens and M. McEwen McGill University & National Research Council CANADA Clinical Dosimetry for Radiotherapy, 2009 AAPM Summer School PSDL ADCL
Thomas Rockwell Mackie. Department of Medical Physics University of Wisconsin Madison WI trmackie@wisc.edu
Dosimetry for IMRT Thomas Rockwell Mackie Department of Medical Physics University of Wisconsin Madison WI trmackie@wisc.edu Financial Disclosure I am a founder and Chairman of TomoTherapy Inc. (Madison,
Radiosurgery 8/16/2011. Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago
Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago Radiosurgery The use of radiation as a surgical tool Small volumes of tissues
Gamma Knife and Axesse Radiosurgery
Gamma Knife and Axesse Radiosurgery John C Flickinger MD Departments of Radiation Oncology & Neurological Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Origin of Radiosurgery
Quality Assurance for Treatment Planning Systems
Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline
Radiation Therapy. 1. Introduction. 2. Documentation of Compliance. 3. Didactic Competency Requirements. 4. Clinical Competency Requirements
PRIMARY CERTIFICATION DIDACTIC AND CLINICAL COMPETENCY REQUIREMENTS EFFECTIVE JANUARY 2014 Radiation Therapy 1. Introduction Candidates for certification and registration are required to meet the Professional
OPTICAL IMAGING IN BREAST CANCER: FLUORESCENCE & ČERENKOV. Outline. Example 1: Detecting lymph node involvement. with. Exogenous molecular imaging
OPTICAL IMAGING IN BREAST CANCER: FLUORESCENCE & ČERENKOV Lymph node imaging Whole breast RT Brian W. Pogue PhD Outline 1. Dual-probe fluorescence imaging for lymph node cancer detection 2. Cerenkov Imaging
Perturbation correction factors for the NACP-02 plane-parallel ionization chamber in water in high-energy electron beams
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 51 (2006) 1221 1235 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/51/5/012 Perturbation correction factors for the NACP-02 plane-parallel ionization
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review
Digital vs. Analogue Control Systems
Digital vs. Analogue Control Systems Presented at the 2011 Annual Meeting of the American College of Medical Physics, Chattanooga, TN, May 1, 2011 Ivan A. Brezovich, PhD, Dept. of Rad Onc, Univ of Alabama
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS W. STRYDOM Department of Medical Physics, Medical University of Southern Africa, Pretoria, South Africa W. PARKER, M. OLIVARES Department of Medical
Evolution of Head and Neck Treatment Using Protons. Mayank Amin, M.Sc,CMD
Evolution of Head and Neck Treatment Using Protons Mayank Amin, M.Sc,CMD Facility Layout Gantry Room 3 Fixed Beams Room 4 HEBT Gantry Room 2 Gantry Room 1 Synchrotron Linac Treatment Planning Imaging Area
HADRON THERAPY FOR CANCER TREATMENT
HADRON THERAPY FOR CANCER TREATMENT Seminar presented by Arlene Lennox at Fermilab on Nov 21, 2003 CANCER STAGES LOCAL TUMOR REGIONAL METASTASIS SYSTEMIC DISEASE CANCER TREATMENT SURGERY RADIATION THERAPY
Physics and Dosimetry of SBRT: Simulation and Localization. Moyed Miften, PhD University of Colorado
Physics and Dosimetry of SBRT: Simulation and Localization Moyed Miften, PhD University of Colorado Team Kelly Stuhr Bodo Reitz Quentin Diot Tracey Schefter Laurie Gaspar Brian Kavanagh Rad Onc Staff at
IAEA-TECDOC-1540. Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems April 2007 IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
OBJECTIVE OUTLINE PHOTONEUTRON PRODUCTION
NEUTRON SHIELDING DESIGN AND EVALUATIONS Nisy E. Ipe, Ph.D. Consultant, Shielding Design, Dosimetry & Radiation Protection San Carlos, CA, U.S.A. Email: nisy@comcast.net http://www.shieldingconsultant.com/
arxiv:astro-ph/0509450 v1 15 Sep 2005
arxiv:astro-ph/0509450 v1 15 Sep 2005 TESTING THERMO-ACOUSTIC SOUND GENERATION IN WATER WITH PROTON AND LASER BEAMS K. GRAF, G. ANTON, J. HÖSSL, A. KAPPES, T. KARG, U. KATZ, R. LAHMANN, C. NAUMANN, K.
Assessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
Transition from 2D to 3D Brachytherapy in Cervical Cancers: The Vienna Experience. Richard Pötter MD. BrachyNext, Miami, 2014.
Transition from 2D to 3D Brachytherapy in Cervical Cancers: The Vienna Experience Richard Pötter MD BrachyNext, Miami, 2014 Disclosures Richard Pötter, MD, does not have any financial relationships or