HPLC in MCAL. Solvents. Fluorescence Detector. Autosampler Manual injector. Pumps. UV Vis diode array Detector
|
|
- Felix Clark
- 7 years ago
- Views:
Transcription
1 HPLC in MCAL Solvents Fluorescence Detector Autosampler Manual injector UV Vis diode array Detector Pumps
2 HPLC Image System control- Data storage HPLC column (packing) Injector Autosampler Sample Solvent(s) Pump(s) Detector(s) (Diode Array & Fluorescence) Waste
3 Separation in HPLC Seperation depends on : Polarity of molecule RP verses NP, Hilic, Hydrophilic-Interaction Chromatography Hic, Hydrophobic-interaction Chromatography Electrical charge (ion exchange). Anion exchanger attracts charged analyte Cation exchanger attracts + charged analyte Molecular size ze exclusion or gel permeation
4 Columns - Heart of System Solid Support (column packing) - Backbone for bonded phases. Usually 10µ, 5µ, 3µ, 2.5µ siloxane or polymeric particles. Porous or solid core Packing has different degrees of polarity depending on functional groups attached Bonded Phases - Functional groups chemically bound to the solid support.
5 HPLC - Modes Normal Phase - Polar stationary phase and non-polar solvent. Reverse Phase (used for caffeine and quinine separation) - Non-polar stationary phase and a polar solvent.
6 C4 C8 C18 Normal Phase verses Reverse Phase HPLC Packing Polar packing Non-polar packing
7 HPLC Solid Support C18RP CH 3 CH 2 CH 2 CH 2 CH 2 (CH 2 ) 14 CH 3 H CH 3 (CH 2 ) 18 Gel Surface CH 3 Bonded phase
8 HPLC Solid Support C18RP End Capping CH 3 CH 2 CH 2 CH 2 CH 2 (CH 2 ) 14 CH 3 H + (CH 2 ) 2 CH 3 (CH 2 ) 18 Gel Surface CH 3 Bonded phase
9 HPLC Solid Support C18RP Polymeric CH 3 CH 2 CH 2 CH 2 CH 2 (CH 2 ) 14 CH 3 (CH 2 ) 18 CH 3 (CH 2 ) 18 Gel Surface CH 3 Bonded phase
10 HPLC Solid Support C18RP Polymeric H CH 2 CH 2 CH 2 CH 2 (CH 2 ) 14 (CH 2 ) 18 CH 3 (CH 2 ) 18 Gel Surface CH 3 Bonded phase
11 Solid Core Particle Solid Core Porous uter Layer
12 Core-Shell Particle
13 Core-Shell Particle
14 HPLC Columns (MCAL) C18 Microsorb (Varian)5 u 100 A 250 x 4.5 mm C18 Polar Synergi (Phenomenex) 4 u 80 A 250 x 4.5 mm C18 RP Particil column (Varian) 3 u 300 A 100 x 2.0 mm C18 Hilic column (Phenomenex) 5 u 200 A 150 x 4.6 mm XB-C18 RP Kinetex (Phenomenex) 2.6 u 100 A 100 x 4.6 mm C18 Microsorb (Varian) 5 u 100 A 250 x 4.5 C8 Polaris (Varian) 3 u A 50 x 20 mm
15 Manual Injector HPLC columns
16 Solvents for RP H 2 most polar with decreasing polarity as solvent B added to mix. Solvent B usually MeH, acetonitrile, iso-propanol Can be isocratic mixture - 1 or 2 pump system or Gradient need at least two pumps and mixing apparatus. Caffeine-quinine exp. uses buffer-acetonitrile or buffer- methanol gradient mixing.
17 The Theoretical Plate Model of Chromatography The plate model supposes that the chromatographic column is contains a large number of separate layers, called theoretical plates (N). Separate equilibrations of the sample between the stationary and mobile phase occur in these "plates". The analyte moves down the column by transfer of equilibrated mobile phase from one plate to the next.
18 Chromatographic Resolution Depends n: 1.Column efficiency (N) peak width Particle diameter Flow rate Sol. Viscosity Temp. Col. Length Col. diameter 2. Solvent efficiency (α) retention time Stationary phase Mobile phase Solvent composition Temperature
19 Column Efficiency -Theoretical plate numbers (N) Theoretical plate numbers are indirect measure of peak width for a peak at a specific retention time. Columns with high plate numbers are considered to be more efficient, that is, have higher column efficiency, than columns with a lower plate count. A column with a high number of theoretical plates will have a narrower peak at a given retention time than a column with a lower N number. N=5.545(t r /w h ) 2
20 Column Efficiency -Height equivalent to theoretical plates Measure of the resolving power of the column HETP = L/N L = Length of column N = Theoretical plates At each theoretical plate one equilibrium distribution of solute between mobile phase and stationary phase occurs.
21 Peak Seperation (α) n stationary phases where the alphas (α) are small, more efficient columns (higher N) are beneficial. Column efficiency is a function of: a) the column dimensions (diameter, length and film thickness), b) the type of carrier solvent and its flow rate or average linear velocity, and c) the compound and its retention.
22 Band broadening theory (Van Deemter equation) column band broadening originates from three main sources: multiple path of an analyte through the column packing; molecular diffusion; effect of mass transfer between phases.
23 Multiple path (Eddy diffusion) The velocity of mobile phase in the column may vary significantly across the column diameter, depending on the particle shape, porosity, and the whole bed structure.
24 Measure of the Resolving Power of the Column HETP = A + B / u + C u A = eddy diffusion take different paths (different lengths). B = longitudinal diffusion analyte diffuses from centre to outer edge. C = resistance to mass transfer: analyte strong affinity for stationary and veloscity of mobile phase is high. U = avg veloscity of mobile phase
25 Resolving Power α α N N
26 Column efficiency is reported as plates per meter (N/Meter). Smaller average packing particle size = Larger N Broader particle size distribution = Smaller N Better packing procedures = Larger N
27 Polarity - Elotropic Series Water is at the polar end of mobile-phase-solvent scale, Hexane, an aliphatic hydrocarbon, is at the non-polar end. In between, single solvents, as well as miscible-solvent mixtures
28 Stationary Phase Polarity Spectrum lica has an active, hydrophilic [water-loving] surface containing acidic silanol [silicon-containing analog of alcohol] The polarity of the silica surface may be modified selectively by chemically bonding to it less polar functional group like n-octylsilyl- [C8], and n-octadecylsilyl- [C18, DS] moieties. The latter is a hydrophobic [water-hating], very non-polar packing.
29 HPLC Experiment Seperation and identification of caffeine, quinine and added theophylline (internal standard) in commercial drinks Comparison of different quantitation methods (external standard, internal standard) Uv verses fluorescent detection Diode array detection
30 HPLC Experiment The separation is done on a C18 RP column. The solvent is a 20% methanol in water (low ph with TEA added) increased to 30% methanol. The caffeine comes off column before the quinine.
31 HPLC Exp: Caffeine and Quinine Quinine Rt = 11.6 min Caffeine wavelength 206 Rt = 9.2 min Caffeine wavelength 273 Rt = 9.2 min
32 Diode Array Detector Caffeine
33 Detection with HPLC Ultraviolet light source (deuterium lamp 190 to 950nm) Visible light source (quartz iodide 350 nm and beyond) Flurescent light (Xenon lamp) Excitation and emmission wavelength selection
34 Concentration f Analytes Concentration is determined by measuring the area of known concentrations of caffeine and quinine, and making a standard curve. Concentration of caffeine is also determined by adding an internal standard (IS) and calculating caffeine based on the response factor of the IS.
Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
More informationThe Theory of HPLC. Gradient HPLC
The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims
More informationGuide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
More informationPurification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
More informationMCAL Spectrophotometry. Spectrophotometry
MCAL Spectrophotometry Instruments include: Cary 50 UV-vis Spectrophotometer Eclipse Spectrofluorometer HPLC Diode Array and Fluorescence ICP-OES with CCD detection Spectrophotometry The instruments all
More informationSPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322. i.chianella.1998@cranfield.ac.
SPE and HPLC Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322 i.chianella.1998@cranfield.ac.uk Solid-Phase Extraction- SPE Simple, fast and efficient sample preparation
More informationThe Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC
The Essential CHROMacademy Guide Mobile Phase Optimization Strategies for Reversed Phase HPLC Webcast Notes Type your questions in the Submit Question box, located below the slide window You can enlarge
More informationComparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns
omparison of the Reversed-Phase Selectivity of Solid ore HPL olumns M. Dolci, L. Pereira, Thermo Fisher Scientific, Runcorn, heshire, UK Technical Note 20788 Key Words Accucore, solid core, column characterization,
More informationLUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)
October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for
More informationHPLC Basics, Equipment, and Troubleshooting
HPLC Basics, Equipment, and Troubleshooting HPLC Basics, Equipment, and Troubleshooting explains chromatography in practical terms from the ground up HPLC Basics, Equipment, and Troubleshooting HPLC Basics,
More informationDiffusion and Fluid Flow
Diffusion and Fluid Flow What determines the diffusion coefficient? What determines fluid flow? 1. Diffusion: Diffusion refers to the transport of substance against a concentration gradient. ΔS>0 Mass
More informationSolid Core Technical Compendium
150-C4 150-C18 Phenyl-Hexyl XL-C18 PFP 150-Amide-HILIC C0 Urea-HILIC Phenyl-X HILIC Polar Premium C8 XL-C8 aq C18 RP-MS Solid Core Technical Compendium Want fast, high-resolution separations, plus a range
More informationFractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
More informationThermo Scientific Syncronis HPLC Columns. Technical Manual
Thermo Scientific Syncronis HPLC Columns Technical Manual Thermo Scientific Syncronis HPLC Columns When developing a new method, one of the most important goals for the chromatographer is to achieve a
More informationIntroduction to HPLC. Basic Module in Bioorganic Chemistry January 16, 2007 Johannes Ranke. Introduction to HPLC p.1/17
Introduction to HPLC Basic Module in Bioorganic Chemistry January 16, 2007 Johannes Ranke Introduction to HPLC p.1/17 Sources R. Kellner, J.-M. Mermet, M. Otto, H. M. Widmer (Eds.), Analytical Chemistry,
More informationCapillary HPLC Introduction
Capillary HPLC Introduction Capillary HPLC Liquid chromatography/mass spectrometry, LC/MS, is a revolutionary tool in the chemical and life sciences. LC/MS is accelerating chemical research by providing
More informationSolid Phase Extraction Products PAGE: 1. Introduction of Solid Phase Extraction (SPE) Why Choose Nano-Micro Tech SPE
Solid Phase Extraction Products PAGE: 1 Introduction of Solid Phase Extraction (SPE) SPE has been used increasingly in chemical analysis and purification broadly, and become the most popular technology
More informationChapter 21 Principles of Chromatography and Mass Spectrometry Problems 1,2,3,4,8,9,11,13
Chapter 21 Principles of Chromatography and Mass Spectrometry Problems 1,2,3,4,8,9,11,13 21-1 What is Chromatography Process to separate and purify compounds in a mixture by passing them trough a material
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationStandard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS
Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLCDAD and HPLCMS/MS Jongheon Shin College of Pharmacy Seoul National University Summary The standard analytical
More informationReversed Phase Chromatography
Reversed Phase Chromatography Principles and Methods 18-1134-16 Edition AA Reversed Phase Chromatography Contents 1. Introduction... 5 Theory of reversed phase chromatography... 6 The matrix... 9 The ligands...
More informationTraining Courses 2014 HPLC GC SPE
Training Courses 2014 HPLC GC SPE 2 1 3 5 4 6 7 5 10 15 20 min 8 2 Courses & Presenters HPLC How to Run HPLC Methods Price: 249 + VAT Code: SS0-5943 Page: 4 How to Develop HPLC Methods Price: 249 + VAT
More informationAutomated Method Development Utilizing Software-Based Optimization and Direct Instrument Control
Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control Dr. Frank Steiner, 1 Andreas Brunner, 1 Fraser McLeod, 1 Dr. Sergey Galushko 2 1 Dionex Corporation, Germering,
More informationChapter 28: High-Performance Liquid Chromatography (HPLC)
Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion
More informationApplication Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
More informationMethod Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Paula Hong and Kenneth J. Fountain Waters Corporation, 34 Maple St., Milford, MA, USA APPLICATION
More informationAn Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.
An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols. Jaap de Zeeuw, Ron Stricek and Gary Stidsen Restek Corp Bellefonte, USA To analyze basic compounds at
More informationAspects of industrial purification of peptides using large-scale chromatography. Lars Andersson and Jonas Persson
Aspects of industrial purification of peptides using large-scale chromatography Introduction By Lars Andersson and Jonas Persson PolyPeptide Laboratories (Sweden) AB PO Box 30089 SE-200 61 LIMHAMN SWEDEN
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationOasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
More informationUnderstanding Selectivity in Reversed Phase Chromatography A Simplified Look at Column Selection
Understanding Selectivity in Reversed Phase Chromatography A Simplified Look at Column Selection Dr. Frank Dorman, Pennsylvania State University Ty Kahler, Randy Romesberg, Vernon Bartlett, Mike Wittrig,
More informationSelectivity and Detectability Optimizations in HPLC
Selectivity and Detectability Optimizations in HPLC SATINDER AHUJA Development Department Pharmaceuticals Division CIBA-GEIGY Corporation Suflern, New York WILEY A WILEY-INTERSCIENCE PUBLICATION JOHN WILEY
More informationSimultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC
Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC Mackenzie Ree and Erik Stoa Department of Chemistry, Concordia College, 901 8 th St S,
More informationContents. List of Contributors XIII The structure of The HPLC-Expert" Preface XVII
V List of Contributors XIII The structure of The HPLC-Expert" Preface XVII XV 1 LC/MS Coupling 1 1.1 State of the Art in LC/MS 1 Oliver Schmitz 1.1.1 Introduction 1 1.1.2 Ionization Methods at Atmospheric
More informationDetermination of Caffeine in Beverages HPLC-1
Determination of in s HPLC-1 Introduction This experiment provides an introduction to the application of High Performance Liquid Chromatography (HPLC) to the solution of complex analytical problems. Cola
More informationSPE Reference Manual & Users Guide. Reference Manual & Users Guide
SM SPE Reference Manual & Users Guide SPE Reference Manual & Users Guide Reference Manual & Users Guide SPE Table of Contents Page No. 1. Objective and Purpose of This Guide 1 2. Solid Phase Extraction
More informationBackground Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
More informationSize Exclusion Chromatography
Size Exclusion Chromatography TOYOPEARL Resins for SEC TOYOPEARL TOYOPEARL TOYOPEARL HW-55 TOYOPEARL HW-65 TOYOPEARL TOSOH BIOSCIENCE TOSOH BIOSCIENCE LLC 56 Keystone Drive Montgomeryville, PA 896-967
More informationAppendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
More informationMEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process
- Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples
More informationUHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
More informationEXAMINATION IN COURSE KJ 2053 CHROMATOGRAPHY
NTNU Norwegian University of Science and Technology Faculty of Natural Sciences and Technology Department of chemistry Page 1 (of 13) Inquiries concerning the contents of the test to : Assoc. professor
More informationHPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
More informationBulk Materials for Preparative and Process Chromatography
Bulk Materials for Preparative and Process Chromatography Silica Based HPLC Bulk Media Characteristics Available in both reversed and normal phases High chemical stability for low leaching High loading
More informationThermo Scientific HyperSep Solid Phase Extraction Method Development Guide
chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo
More informationEXPERIMENT 1 - Determination of the purity and identity of organic compounds by melting point and/or analytical thin layer chromatography
EXPERIMENT 1 - Determination of the purity and identity of organic compounds by melting point and/or analytical thin layer chromatography PART A Melting points and mixed melting points. As discussed in
More informationProtocol: HPLC (amino acids)
Page 1 of 6 1. Chemicals Composition M MW gram volume (ml) product code 0-pthaldialdehyde C 8H 6O 2 134.13 Sigma P0657 Perchloric acid HClO 4 3.5 100.46 Dilute 70% stock solution 1:1 with demi to obtain
More informationProblem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:
Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T
More informationAnalyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
More informationThe Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
More information# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
More information1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
More informationPRODUCT INFORMATION Chromalite resins for reverse-phase chromatography, adsorption and SPE
PRODUCT INFORMATION Chromalite resins for reverse-phase chromatography, adsorption and SPE Presents an overview of the advantages of Purolite s Chromalite polymeric resins over silica for the purification
More informationHow To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
More informationReversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
More informationLecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder
Lecture Chromo-3: Gas Chromatography CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder Outline Introduction Instrument overview Carrier gas Sample
More informationElectrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
More informationFerda ORHUN Mineral Research and Exploration Institute of Turkey
GAS CHROMATOGRAPHY, EXPERIMENTAL STUDIES ON THE FLOW RATE EFFECTS OF CARRIER GAS AND APPLICATION OF THE METHOD TO HYDROCARBON ANALYSIS OF SOME NATURAL GASES IN TURKEY Ferda ORHUN Mineral Research and Exploration
More informationSIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
More informationAMINO ACID ANALYSIS By High Performance Capillary Electrophoresis
AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis Analysis of Amino Acid Standards Label free analysis using the HPCE-512 ABSTRACT Capillary electrophoresis using indirect UV detection
More informationLIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93)
INTRODUCTION HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: 1-13-93) Headache, sore muscles, arthritis pain... How do you spell relief?
More informationFundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
More informationUnique Bonding Chemistry
columns Acclaim HPLC Columns Reversed-phase silica columns for pharmaceutical, nutraceutical, environmental, food, and life science applications in HPLC. High hydrophobicity, low polarity phases LC/MS
More informationGENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
More informationDetermination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC
Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC Abstract: High performance liquid chromatography (HPLC) was used to determine the identity and concentrations of food dyes
More informationGC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
More informationRECITATION NOTES FOR EXPERIMENT # 5 A&B THIN LAYER CHROMATOGRAPHY
RECITATION NOTES FOR EXPERIMENT # 5 A&B THIN LAYER CHROMATOGRAPHY Have your lab textbook available for quick reference to specific pages, indicated in red. BASIC PRINCIPLES OF CHROMATOGRAPHY Chromatography
More informationThin Layer Chromatography.
Thin Layer Chromatography. Thin layer chromatography, or TLC, is a method for analyzing mixtures by separating the compounds in the mixture. TLC can be used to help determine the number of components in
More informationTABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ACKNOWLEDGEMENT ABSTRACT LIST OF TABLES LIST OF FIGURES LIST OF SYMBOLS LIST OF APPENDICES
vii TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ii ACKNOWLEDGEMENT iv ABSTRACT v TABLE OF CONTENTS vii LIST OF TABLES xii LIST OF FIGURES xiv LIST OF SYMBOLS xvii LIST OF APPENDICES xviii 1 INTRODUCTION
More informationIAM Chromatography. HPLC Separation Tools for Membrane Protein Purification and Drug Membrane Permeability Prediction
IAM Chromatography Immobilized Artificial Membrane (IAM) technology is an innovative approach to chromatography in which the chromatographic surface emulates the lipid environment of the cell membrane.
More informationAir Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike License. Your use of this material constitutes acceptance of that license and the conditions of use of materials on this
More informationFast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
More informationSUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
More informationAddressing the World Helium Shortage For Gas Chromatography
Addressing the World Helium Shortage For Gas Chromatography Helium Conservation and Converting GC Methods to Nitrogen and Hydrogen Carrier Gas James D. McCurry, Ph.D. Senior Scientist Agilent Technologies
More informationINSTRUCTOR S MANUAL FOR LABORATORY PROJECTS USING CHROMATOGRAPHY
INSTRUCTOR S MANUAL FOR LABORATORY PROJECTS USING CHROMATOGRAPHY Contents Page Learning Objectives 1 Introduction 2 Appendix 1: Laboratory Project Proposal 5 Appendix 2: Peer and Self-Assessment of Laboratory
More informationINSTRUCTIONS 56-1190-98. Edition AC
Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography
More informationChromQuest 5.0 Chromatography Data System
ChromQuest 5.0 Chromatography Data System User Guide for the Surveyor LC CHROM-97257 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. Surveyor and ChromQuest are trademarks
More informationABRF 2004 Poster# P55-S Displacement Chromatography Effects Can Cause Highly Selective Sampling of Peptides During Solid Phase Extraction Cleanup
ABRF 2004 Poster# P55-S Displacement Chromatography Effects Can Cause Highly Selective Sampling of Peptides During Solid Phase Extraction Cleanup by Andrew J. Alpert PolyLC Inc./ 9151 Rumsey Road, ste.
More informationExpectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
More informationAnalysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
More informationSeparation by Chromatography Methods
Separation by Chromatography Methods Analytical Biochemistry 3.1 Principle of Separation techniques 3.2 Methods Based on Polarity (3.2.1-3.2.3) Biochemistry and Molecular Biology 11.5 Partition Chromatography
More informationAdvantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites
Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Carmen T. Santasania and David S. Bell SUPELCO, 595 North Harrison Road, Bellefonte, PA 16823 T404045 GUS Abstract
More informationOverview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
More informationUltra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
More informationUV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this
More informationDevelopment of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation
IJPAR Vol.4 Issue 4 Oct Dec - 2015 Journal Home page: ISSN: 2320-2831 Research Article Open Access Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation V.
More informationHPLC Column Comparison Screening Study for Reversed Phase Columns
HPLC Column Comparison Screening Study for Reversed Phase Columns Carmen T. Santasania, Eric M. Snyder, Matilal Sarker, and Tracy L. Ascah Supelco, 595 North Harrison Road, Bellefonte, PA 16823, USA T403058
More informationA Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
More informationThermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance
Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid
More informationQUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd
QUANTITATIVE AMINO ACID ANALYSIS Aurélie Lolia Applications Manager, Biochrom Ltd QUANTITATIVE AMINO ACID ANALYSIS Principles of amino acid analysis Ion exchange chromatography The Biochrom 30 physiological
More informationThe LC Handbook Guide to LC Columns and Method Development
Agilent CrossLab combines the innovative laboratory services, software, and consumables competencies of Agilent Technologies and provides a direct connection to a global team of scientific and technical
More informationGas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
More informationUsing Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
More informationTransfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs
Agilent Application Solution Transfer of a USP method for prednisolone from normal phase PLC to SFC using the Agilent 126 Infinity ybrid SFC/UPLC System Saving time and costs Application Note Pharmaceutical
More informationSPECTROSCOPY. Light interacting with matter as an analytical tool
SPECTROSCOPY Light interacting with matter as an analytical tool Electronic Excitation by UV/Vis Spectroscopy : X-ray: core electron excitation UV: valance electronic excitation IR: molecular vibrations
More informationHypercarbColumns 100% porous graphitic carbon for extended separation capabilities
Columns 00% porous graphitic carbon for extended separation capabilities Exceptional retention of very polar analytes Separates structurally related substances ph stable from 0 to Ideal for high temperature
More informationIntelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
More informationPrepTip. Reverse Phase PrepTip User Guide
PrepTip Reverse Phase PrepTip User Guide All text, photographs and illustrations are copyrighted by Harvard Apparatus, Inc. 2004. PrepTip is a trademark of Harvard Apparatus, Inc. Harvard Apparatus 84
More information