1 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 General Purpose ph/orp Sensor Specifications Figure Insertion/Submersion Sensor Sensor Type: General purpose 3900 Measured Range: ph AccuGlass 0-14 ORP to mv Percent Linearity Over ph Range: Range Linearity % % % % % % % Operating Temperature: -10 to 100 C (14 to 212 F) Automatic temperature compensation -10 to 100 C (14 to 212 F) Maximum Pressure: 790 kpa [abs] (100 psig) at 100 C (212 F) CRN rating: 60 psig up to 100 C (212 F). Materials of Construction: stainless steel, glass, Teflon, polyphenylene sulfide (PPS), and EPDM Materials of ORP: Platinum Process Connections: Front facing: 3 4 and 1 MNPT Rear facing: 1 MNPT Integral Cable: 32 ft (10m) cable with integral SMART preamp; 15 ft (4.7m) cable without preamp sensor VP8 Cable: use XX, 2.5 ft (.8m) to 100 ft (31m) Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb) OPERATIONAL NOTE The 3900 Sensor responds to changes in ph at a minimum conductivity of 0.1mS/cm in deionized water. Sample flow rate must be controlled to 2 gph (7.6L/hr). The offset is approximately -0.2 ph after 3 months at ambient temperature. CAUTION: SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user. WARNING Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!

2 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 CAUTION The solution used during calibration is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water. ATEX DIRECTIVE: Special Conditions for safe use 1. All ph/orp sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a build up of an electrostatic charge. 2. All ph/orp sensor models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test to earth. This must be taken into consideration at installation. Storage 1. It is recommended that electrodes be stored in their original shipping containers until needed. 2. Do not store at temperatures below -10 C (14 F). 3. Electrodes should be stored with a protective cap containing KCl solution (PN ). 4. For overnight storage, immerse the sensor in tap water or 4 ph buffer solution. 5. ph glass electrodes slowly deteriorate in storage. There is no specific expiration date. However, the calibration procedure described below should be followed to determine that the sensor calibrates properly. Electrode preparation 1. Remove electrode from shipping container. 2. Remove the protective boot covering the electrode bulb. 3. Rinse away salt film with clean water, then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there. Installation For sensor orientation, see Figure 2. For sensor dimensions, see Figure 3. For mounting options, see Figures 4-7. For wiring, see Figures For additional wiring information, please visit our website at Wrap the sensor threads with six or seven (6 7) turns of Teflon tape to prevent leakage. Do not over tighten the sensor into it s receptacle. Hand tighten the sensor, and then tighten one or two (1 2) turns with a wrench. Two point buffer calibration Select two stable buffer solutions, preferably ph and 10.0 (ph buffers other than ph 4.0 and ph 10.0 can be used as long as the ph values are at least two ph units apart). NOTE: A ph 7 buffer solution reads a mv value of approx. zero, and ph buffers read approximately ± 59.1 mv for each ph unit above or below ph 7. Check the ph buffer manufacturer specifications for millivolt values at various temperatures since it may affect the actual value of the buffer solution mv/ph value. 1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/ transmitter. 2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water. 3. Repeat steps 1 and 2 using the second buffer solution. 4. The theoretical slope value, according to the Nernst equation for calculating ph, is approximately 59.1 mv/ph. Over time the sensor will age, both in the process and in storage, which will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mv/ph. Recommended ph Sensor Standardization For maximum accuracy, the sensor can be standardized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor s slope but will simply adjust the analyzer s reading to match that of the known process ph.

3 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Maintenance for ph electrode Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement. 1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent. 2. To remove scale deposits, soak electrodes for 1 to 5 minutes in a 5% hydrochloric acid solution. 3. Temperature effect on life expectancy: If glass electrode life expectancy is 100% at 25 C (77 F), then it will be approximately 25% at 80 C (176 F), and approximately 10% at 100 C (212 F). CAUTION Hydrochloric acid is toxic and highly corrosive. Avoid skin contact, wear protective gloves. Use only in a well ventilated area. Do not inhale fumes. In case of an accident, consult a doctor immediately. ORP Calibration 1. After making an electrical connection between the sensor and the instrument, obtain a standard solution of saturated quinhydrone PN R508-8OZ (475mV). This can also be made quite simply by adding a few crystals of quinhydrone to either ph 4 or ph 7 buffer. Quinhydrone is only slightly soluble, therefore only a few crystals will be required. 2. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize. 3. Standardize the instrument to the solution value shown in the table, right. The resulting potentials, measured with a clean platinum electrode and Figure 2. Sensor Orientation saturated KCl/AgCl reference electrode, should be within +/- 20 millivolts of the value shown in the table below. Solution temperature must be noted to ensure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used. 4. Remove the sensor from the buffer, rinse, and install in the process. Maintenance for ORP Electrode Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement. 1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent. 2. To remove scale deposits, soak electrodes for 1 to 5 minutes in a 5% hydrochloric acid solution. 3. ORP (metallic) electrodes should be polished with moistened baking soda. Figure 3. Sensor Dimensions 3/4 IN MNPT 1 IN MNPT WRENCH FLATS 1.30 IN ACROSS 1 IN MNPT TEMP COMP SOLN GND REFERENCE JUNCTION ph ELECTRODE

4 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 3. Sensor Dimensions 3/4 IN MNPT 1 IN MNPT WRENCH FLATS 1.30 IN ACROSS 1 IN MNPT TEMP COMP SOLN GND REFERENCE JUNCTION ph ELECTRODE Figure 4. Typical Flow Through insertion installation using PN Pipe Tee (sensor must be installed at least 10 above the horizon) Horizontal Pipe Tee (PN ) Pressure/Temperature Ratings psig (kpa [abs]) F ( C) 150 (1136) 150 (65) 128 (984) 160 (71) 102 (805) 170 (77) 80 (653) 180 (82) 57 (494) 200 (93) 48 (432) 210 (99) Figure 5. Low Flow Cell PN INCH MILLIMETER Inlet and outlet connections are stainless steel and take 1/4-inch OD tubing. Flow cell is polycarbonate with 1/4-inch FNPT fittings. Wetted materials: Body & Nut Polyester/Polycarbonate Fittings 316 SST Seals Silicone Flow cell ratings: Temperature: F (0-70 C) Max. Pressure: 90 PSIG (721 kpa [abs]) Flow rate: 2-5 GPH ( LPH) 4

5 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 6. Jet Spray Cleaner PN (sensor must be installed at least 10 above the horizon) The Jet Spray Cleaner eliminates routine, manual sensor maintenance by cleaning the sensor with water or compressed air. Flow through the cleaner can be controlled by a solenoid valve. NOTE: The Jet Spray Cleaner can be used with Handrail Mounting Assembly (PN , not shown) or can be mounted through conduit as shown below. Jet Spray Cleaner shown with 3900VP ph sensor Figure 7. Low Flow Panel PN SQP10077-LQD Low Flow Panel Specifications Inlet flow Inlet pressure Temperature 3-80 gph ( L/hr) 3 65 psig ( kpa abs)* F (0-50 C) *The minimum inlet pressure is required to open a check valve, which prevents the flow cell from draining if sample flow is lost. Removing the check valve lowers the inlet pressure requirement to a few feet of water head. 5

6 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 NOTE For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product. 1056, 1057, 56, 5081, 6081, 54e, and XMT: and sensors with SMART preamps: Wiring_Diagram.aspx 1055: Figure and 3900VP-01 to 54epH/ORP Figure and 3900VP-02 to 54epH/ORP 6

7 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 10. Sensor Wiring 3900VP-01 gray cable to 54epH/ORP Figure 11. Sensor Wiring 3900VP-02 gray cable to 54epH/ORP 7

8 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure and 3900VP-01 to Xmt Figure and 3900VP-02 to Xmt 8

9 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 14. Sensor Wiring 3900VP-01 with gray cable to XMT Figure 15. Sensor Wiring 3900VP-02 with gray cable to Xmt 9

10 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure and 3900VP-01 to 1055 Figure 17. Two-Input and 3900VP-01 to

11 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure and 3900VP-02 to 1055 Figure 19. Two-Input and 3900VP-02 to

12 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 20. Wiring with Remote Junction Box and Preamp (PN ) to and 3900VP-02 Figure pH/pH Wiring through Remote Junction Boxes to and 3900VP-02 12

13 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure Sensors, Preamplifier in Junction Box ID (inner drain) (No Connection Green White White/Red Red White/Gray Gray Blue Clear Orange 3900 Sensors, Preamplifier in Junction Box For and 3900VP-02 sensors Non-preamp sensors Blue Cable ID (inner drain) Connect to instrument ground or as shown 3 White (RTD Return) 3 RTD RTN /RTN 4 White/Red (RTD Sense) 4 RTD SENSE /SNS 5 Red (RTD in) 5 RTD IN 6 White/Gray (Reference Shield) 6 Ref SHLD / SHLD /GUARD 7 8 Gray (Reference in) 7 REF / REF IN Blue (Solution Ground) 8 GND/SOL GND 9 9 White/Black ( mv/ph Shield) ph SHLD / SHLD / GUARD 10 Black ( mv/ph in) 10 ph/ph in TB 1 BNC No Connection TB 2 Brown (-5VDC) Green (+5 VDC) Extension Cable PN or shown Extension Cable PN can also be used; extend color scheme from sensor or VP8 cable instead of using shown wiring scheme. -5VDC / -V SNSR +5VDC / +V SNSR Rosemount Analytical intruments: 56 / 1056 / 1057 / 1066 / 6081 Remote Junction Box PN or Sensor Head Junction Box PN (Both J-boxes include Preamplifier Board PN ) epH/ORP XMT 13

14 General Purpose ph/orp Sensor October 2014 Figure Sensors, Preamplifier in Sensor Instruction Sheet LIQ_MAN_ABR_3900 Green White White/Red Red White/Gray Gray Blue Clear Orange 3900 Sensors, Preamplifier in Sensor For and 3900VP-01 sensors Preamplifier in sensor Blue Cable Green Connect to instrument ground or as shown White (RTD Return) RTD RTN /RTN White/Red (RTD Sense) Red (RTD in) RTD SENSE /SNS RTD IN Gray (Reference in) Blue (Solution Ground) Clear ( mv/ph Shield) Orange ( mv/ph in) White/Gray (-5V) ID (inner drain) (+5V) Ref SHLD / SHLD /GUARD REF / REF IN GND/SOL GND ph SHLD / SHLD / GUARD ph/ph in -5VDC / -V SNSR +5VDC / +V SNSR Rosemount Analytical intruments: 56 / 1056 / 1057 / 1066 / epH/ORP , 3400HTVP-70, 10 XMT

15 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure Sensors, Preamplifier in Sensor 3900 Sensors, Preamplifier in Sensor, Junction Box used for extending cable For and 3900VP-02 sensors Preamplifier in sensor Blue Cable Green White White/Red Red Gray Blue Clear (Orange shield) Orange White/Gray ID (inner drain) TB TB 2 Remote Junction Box PN (with Extension board for point-to-point wiring) Inner Drain White (RTD Return) White/Red (RTD Sense) Red (RTD in) White/Gray (No connection) Gray (Reference in) Blue (Solution Ground) White/Black (mv/ph Sheild) Black ( mv/ph in) Brown (-5VDC) Green (+5 VDC) Extension Cable PN or shown. Connect to instrument ground or as shown Extension Cable PN can also be used; White/gray is replacement wire color for Brown wire. RTD RTN /RTN RTD SENSE /SNS RTD IN Ref SHLD / SHLD /GUARD REF / REF IN GND/SOL GND ph SHLD / SHLD / GUARD ph/ph in -5VDC / -V SNSR +5VDC / +V SNSR Rosemount Analytical intruments: 56 / 1056 / 1057 / 1066 / epH/ORP XMT 15

16 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure and 3900VP-01 to Models 5081, 4081 or 3081 Figure and 3900VP-02 to Models 5081, 4081 or

17 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 10. Wiring for and 3900VP-01 to 56, 1056 & 1057 ph Figure 11. Wiring for and 3900VP-02 to 56, 1056 & 1057 ph 17

18 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 26. Sensor Wiring 3900VP-01 with gray cable to 1056/56 Figure 27. Sensor Wiring 3900VP-02 with gray cable to 1056/56 18

19 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 28. Sensor Wiring 3900VP-01 with gray cable to 1057 Figure 29. Sensor Wiring 3900VP-01 with gray cable to

20 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 30. ph/orp Sensor Wiring 1066/3900VP-01 (blue cable refer to manual for recommended cable order) Figure 31. ph/orp Sensor Wiring 1066/3900VP-02 (gray cable refer to manual for recommended cable order) 20

21 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 32. ph/orp Sensor Wiring 1066/3900VP-01 (refer to manual for recommended cable order) Figure 33. ph/orp Sensor Wiring 1066/3900VP-01 (blue cable refer to manual for recommended cable order) 21

22 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 34. ph/orp Sensor Wiring 1066/3900VP-02 (refer to manual for recommended cable order) Figure 35. ph/orp Sensor Wiring 1066/3900VP-02 (blue cable refer to manual for recommended cable order) 22

23 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 36. Sensor Wiring 3900VP-01 with gray cable to 5081 Figure 37. Sensor Wiring 3900VP-01 with gray cable to

24 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure and 3900VP-01 to 6081-P Figure and 3900VP-02 to 6081-P 24

25 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 40. Sensor Wiring 3900VP-01 with gray cable to 6081 Figure 41. Sensor Wiring 3900VP-02 with gray cable to

26 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Ordering Information The 3900/3900VP ph/orp Sensor is housed in a PSS plastic body with built-in solution ground for advanced diagnostics. A Pt-100 RTD is offered for temperature compensator. The sensor is available with an integral cable or VP8 connector. Junction box kits with preamplifiers can be ordered separately if an analyzer/ transmitter with an integral preamplifier is not located within 15 ft. of the sensor. SMART preamplifier (-01) is the standard preamplifier option, and is compatible with models XMT, 3081, 5081, 6081, 54e, 1055 and 1056 analyzers/transmitters ph/orp Sensor - New Generation general purpose sensor Code Preamplifier Option 01 Preamplifier 02 No preamplifier Code Measuring Electrode 10 General Purpose Low resistivity GPLR ph Glass (0-14pH) 12 Platinum ORP EXAMPLE VP ph/orp Sensor - New Generation general purpose sensor with VP Code Preamplifier Option 01 Preamplifier 02 No preamplifier Code Measuring Electrode 10 General Purpose Low resistivity GPLR ph Glass (0-14pH) 12 Platinum ORP EXAMPLE 3900VP

27 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Accessories Part Number Description ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector Jet cleaner Insertion adapter Low flow cell Junction Box with Preamplifier for Models 54e, 3081, 4081, 5081, XMT, 1055, CPVC flow through Tee, 1-1/2 in. NPT process connection Sensor handrail assembly Buffer solution, 4.01 ph, 16oz Buffer solution, 6.86 ph, 16oz Buffer solution, 9.18 ph, 16oz Extension cable, gray Extension cable, blue SQP10077-LQD Low Flow Control Panel 27

28 LIQ_MAN_ABR_3900 Rev. E October 2014 facebook.com/emersonrosemountanalytical AnalyticExpert.com twitter.com/raihome youtube.com/user/rosemountanalytical 8 Credit Cards for U.S. Purchases Only. Emerson Process Management 2400 Barranca Parkway Irvine, CA USA Tel: (949) Fax: (949) RosemountAnalytical.com 2014 Rosemount Analytical, Inc. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.

General Purpose ph/orp Sensors

General Purpose ph/orp Sensors General Purpose ph/orp Sensors Product Data Sheet LIQ_PDS_3900 / Rev. F December 2014 General Purpose ph/orp Sensors SMART preamplifier-stores calibration and other data. Extended sensor life provided

More information

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev. Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE

More information

ENDURANCE Conductivity Sensors

ENDURANCE Conductivity Sensors Instruction Manual 400 and 400VP LIQ_MAN_ABR_400_400VP December 2013 ENDURANCE Conductivity Sensors For additional information, please refer visit our website at www.rosemountanalytical.com. CAUTION Sensor/Process

More information

Submersion/Insertion Toroidal Sensor

Submersion/Insertion Toroidal Sensor Instruction Sheet PN 51A-228/rev.P December 2010 Model 228 Submersion/Insertion Toroidal Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS

More information

Submersion/Insertion Toroidal Sensor

Submersion/Insertion Toroidal Sensor Instruction Manual 228 Sensor LIQ_MAN_ABR_228 May 2014 PN 51A-228 Submersion/Insertion Toroidal Sensor For additional information, please visit our website at www.rosemountanalytical.com CAUTION SENSOR/PROCESS

More information

Trace Dissolved Oxygen Sensor

Trace Dissolved Oxygen Sensor Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION

More information

400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors

400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors 400 and 400VP Series Product Data Sheet LIQ-PDS-400-400VP June 2016 400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors No initial calibration required. Factory-measured cell constant ensures

More information

Toroidal Conductivity Sensor

Toroidal Conductivity Sensor Instruction Sheet PN 51A-/rev.C June 2012 Toroidal Conductivity Sensor For additional information, please visit our website at www.emersonprocess.com/rosemountanalytical.com SPECIFICATIONS Wetted Materials:

More information

Ozone Sensor. Model 499A OZ CAUTION CAUTION. Instruction Sheet PN 51A-499AOZ/rev.F January 2007

Ozone Sensor. Model 499A OZ CAUTION CAUTION. Instruction Sheet PN 51A-499AOZ/rev.F January 2007 Instruction Sheet PN 51A-499AOZ/rev.F January 2007 Model 499A OZ Ozone Sensor For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website at www.emersonprocess.com/raihome/liquid/.

More information

Free Chlorine Sensor. Quick Start Guide PN 51A-499ACL-01, Rev. L March 2016

Free Chlorine Sensor. Quick Start Guide PN 51A-499ACL-01, Rev. L March 2016 Free Chlorine Sensor Quick Start Guide PN 51A-499ACL-01, Rev. L March 2016 Quick Start Guide March 2016 For additional information, please visit our website at EmersonProcess.com/LiquidAnalysis. Sensor/process

More information

For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/.

For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. Instruction Sheet PN 51A-499AOZ/rev.H September 2010 Model 499A OZ Ozone Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS APPLICATION

More information

Sanitary and CIP Toroidal Sensor

Sanitary and CIP Toroidal Sensor Instruction Sheet PN 51A-225/rev.I December 2010 Model 225 Sanitary and CIP Toroidal Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS

More information

Total Chlorine Sensor

Total Chlorine Sensor Instruction Sheet PN 51A-499ACL02/rev.G May 2007 Model 499A CL-02 Total Chlorine Sensor For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website

More information

Dissolved Oxygen Sensors

Dissolved Oxygen Sensors Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit

More information

PART 1 - INTRODUCTION...

PART 1 - INTRODUCTION... Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART

More information

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes Signet 2818-2823 Conductivity/Resistivity Electrodes 2818 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - Polypro - SS on 10 and 20 cell - Tri-clamp 1-1½ in., 2 in. - Opt. ½ in.

More information

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes Signet 2818-2823 Conductivity/Resistivity Electrodes 2818 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - Polypro - SS on 10 and 20 cell - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT

More information

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes Signet 2818-2823 Conductivity/Resistivity Electrodes 2818 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - Polypro - SS on 10 and 20 cell - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT

More information

Telephone: 760 930-6501 Fax: 760 930-0615 IQ Scientific Instruments, Inc. 2075-E Corte del Nogal Carlsbad, CA 92011

Telephone: 760 930-6501 Fax: 760 930-0615  IQ Scientific Instruments, Inc. 2075-E Corte del Nogal Carlsbad, CA 92011 IQ Scientific Instruments, Inc. 2075-E Corte del Nogal Carlsbad, CA 92011 Telephone: 760 930-6501 Fax: 760 930-0615 www.phmeters.com Revision 4.5 2005 IQ Scientific Instruments, Inc. MODEL IQ140 ph Meter

More information

GLOBE FLOWCELL. Operating Manual. ROLA-CHEM Flowcell PN /17/2015 page: 1 of 8

GLOBE FLOWCELL. Operating Manual. ROLA-CHEM Flowcell PN /17/2015 page: 1 of 8 GLOBE FLOWCELL Operating Manual ROLA-CHEM Flowcell PN 550182 09/17/2015 page: 1 of 8 Table of Contents Section title Page 1. Safety Instructions 2 2. Description 2 3. Installation 3 4. Use of Flowcell

More information

Model Q46P/R. Sealed Reference System... for Longer Sensor Life

Model Q46P/R. Sealed Reference System... for Longer Sensor Life Call 800.959.0299 to speak with a sales representative or visit us on the web at www.analyticaltechnology.com ph/orpmonitor Model Q46P/R INTRODUCTION. Measurement and control of ph is important in a wide

More information

TB2CS Conductivity sensor 2-Electrode. Superior accuracy in low level conductivity measurement ranges

TB2CS Conductivity sensor 2-Electrode. Superior accuracy in low level conductivity measurement ranges Data sheet TB2CS Conductivity sensor 2-Electrode Superior accuracy in low level conductivity measurement ranges Measurement range from 0 to 199.9 µs/cm and 0 to 19.99 ms/cm Resolution to 0.001 µs/cm in

More information

Denver Instrument Company. BASIC ph Meter. Operation Manual 300408.1 Rev. G

Denver Instrument Company. BASIC ph Meter. Operation Manual 300408.1 Rev. G Denver Instrument Company BASIC Meter Operation Manual 300408.1 Rev. G BASIC Meter Quick Reference 1. Connect power cable to meter power jack and to AC power source. 2. Connect a glass electrode to the

More information

PRESSURE AND TEMPERATURE SENSORS FOR THE E340 BOILER ROOM CONTROL

PRESSURE AND TEMPERATURE SENSORS FOR THE E340 BOILER ROOM CONTROL ES-3481 JUNE 27, 2005 PRESSURE AND TEMPERATURE SENSORS FOR THE E340 BOILER ROOM CONTROL DESCRIPTION The PS348 and TS348 Series of pressure and temperature sensors are designed for use with the E340 Boiler

More information

PRESSURE AND TEMPERATURE SENSORS for use with the BURNERLOGIX Burner Management System and PPC4000 Series Fuel/Air Ratio Controller

PRESSURE AND TEMPERATURE SENSORS for use with the BURNERLOGIX Burner Management System and PPC4000 Series Fuel/Air Ratio Controller BLZPTS-1 NOVEMBER 19, 2013 TS350 TS752 BLPS PRESSURE AND TEMPERATURE SENSORS for use with the BURNERLOGIX Burner Management System and PPC4000 Series Fuel/Air Ratio Controller DESCRIPTION The BLPS, TS350

More information

CALIBRATING TOROIDAL CONDUCTIVITY SENSORS

CALIBRATING TOROIDAL CONDUCTIVITY SENSORS Application Data Sheet ADS 43-025 October 2010 Theory CALIBRATING TOROIDAL CONDUCTIVITY SENSORS BACKGROUND A toroidal conductivity sensor consists of two wire wound toroids encased in a plastic body. When

More information

Install NEMA 4X enclosure for the Sievers * InnovOx Online TOC Analyzer

Install NEMA 4X enclosure for the Sievers * InnovOx Online TOC Analyzer 6060 Spine Road Boulder, CO 80301 USA USA 1 800 255 6964 Europe 0161 864 6800 Worldwide 303 444 2009 Install NEMA 4X enclosure for the Sievers * InnovOx Online TOC Analyzer Introduction Purpose: To install

More information

CONTACTING CONDUCTIVITY/RESISTIVITY SENSORS

CONTACTING CONDUCTIVITY/RESISTIVITY SENSORS CONTACTING CONDUCTIVITY/RESISTIVITY SENSORS Applications Drinking Water Power Industrial Water Ultimate accuracy from ultra-pure to high conductivity applications. High Performance Design These sensors

More information

ph or O.R.P. Transmitter

ph or O.R.P. Transmitter ph or O.R.P. Transmitter Type 8202 can be combined with... Programmable outputs: two transistor and single or dual analog 4-20 ma (Process + Temp) Removable backlit display Universal fluidic process connection

More information

Digital inductive conductivity transmitter

Digital inductive conductivity transmitter Digital inductive conductivity transmitter Optimal solution for conductivity measurements in difficult fluids (polluted, dirty,...) PEEK/PPA version for CIP applications Large range of process connections

More information

Multi-Parameter Electrochemical / Optical Water Quality System

Multi-Parameter Electrochemical / Optical Water Quality System Product Data Sheet PDS 71-WQS/rev.H September 2011 Model WQS Multi-Parameter Electrochemical / Optical Water Quality System COMPLETE SYSTEM INCLUDES sensors, cable, and instruments. MEASURES ph, ORP, Conductivity,

More information

Instruction Manual PN 51-381/rev.E March 2010 Model 381. ph/orp Sensor

Instruction Manual PN 51-381/rev.E March 2010 Model 381. ph/orp Sensor Instruction Manual PN 51-381/rev.E March 2010 Model 381 ph/orp Sensor ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products to meet

More information

ph Measurement and Control

ph Measurement and Control ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components

More information

Combination ph/redox (ORP) Sensors

Combination ph/redox (ORP) Sensors Data Sheet SS/AP00 Issue 4 Combination ph/redox (ORP) Sensors AP00 Series Long life large PTFE junction provides resistance to fouling double junction with gelled electrolyte provides better-defined junction

More information

Sartorius Basic ph Meter PB-20

Sartorius Basic ph Meter PB-20 98648-009-18 Sartorius Basic ph Meter PB-20 Operation Manual Contents Page General View 4 Warning and Safety Information 5 Basic ph Meter Quick Reference 6 Installing and Maintaining Electrodes 8 Standardizing

More information

Introduction to Oxidation Reduction Potential Measurement

Introduction to Oxidation Reduction Potential Measurement Introduction to Oxidation Reduction Potential Measurement Dr. Axel W. Bier, Hach-Lange, Application Specialist Theory Oxidation Reduction Potential (ORP or Redox Potential) measures an aqueous system s

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Model IQ120 and IQ125 Professional ph Meter with Silicon Chip Sensor (IQ125 additional instructions printed in RED) IQ Scientific Instruments, Inc. 2075-E Corte del Nogal Carlsbad,

More information

SOIL PH METER Model : PH-212

SOIL PH METER Model : PH-212 SOIL PH METER Model : PH-212 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 2-1 Meter...1 2-2 Soil ph Electrode... 3 3. FRONT PANEL DESCRIPTIONS... 4 3-1 Display...4 3-2 Power On/OFF Button...

More information

ACTIVE ELECTRODELESS SENSORS Instruction Manual

ACTIVE ELECTRODELESS SENSORS Instruction Manual W A L C H E M IWAKI America Inc. Active Electrodeless Sensors ACTIVE ELECTRODELESS SENSORS Instruction Manual Five Boynton Road Hopping Brook Park Holliston, MA 01746 USA TEL: 508-429-1110 FAX: 508-429-7433

More information

Milwaukee USER MANUAL

Milwaukee USER MANUAL Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the

More information

Process Instrumentation

Process Instrumentation Table of Contents Page T1.2 Introduction T1.4 HI 504 T1.8 Panel Mounted Process Controllers T1.12 Wall Mounted Process Controllers T1.33 Dosing Pumps T1.50 Accessories for Dosing Pumps T1.58 Transmitters

More information

Sanitary Type Electronic Flow Switches

Sanitary Type Electronic Flow Switches Industries Pharmaceutical Sanitary Type Electronic Flow Switches Food and Beverage Biotechnology Semiconductor Water Treatment S Series Ameritrol, Inc. Instruments and Controls Features No Moving Parts

More information

INSTRUCTION MANUAL. ph/mv METER

INSTRUCTION MANUAL. ph/mv METER INSTRUCTION MANUAL ph/mv METER QUICK OPERATION 1. Open shipping box - take out instrument. 2. Any damage? - Report to freight company. - Call Distributor/Dealer 3. Turn instrument on. If meter pegs, -

More information

DOE-45PA Dissolved Oxygen Sensor

DOE-45PA Dissolved Oxygen Sensor DOE-45PA Dissolved Oxygen Sensor - 2 - TABLE OF CONTENTS Part 1 - Introduction... 3 Figure 1-1 D.O. SYSTEM DIAGRAM WITH SUBMERSIBLE SENSOR... 3 Part 2 Electrical Connection... 4 Figure 2-1 SUBMERSIBLE

More information

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter Thermal Mass meter and Temperature Transmitter Description The MTI10 insertion thermal mass flowmeter and the MTL10 in-line thermal mass flowmeter provide accurate mass flow measurement of clean, dry gases

More information

OXIGENO DISUELTO. Dissolved Oxygen. Install & forget. Sistema autolimpiable. Autoclean system

OXIGENO DISUELTO. Dissolved Oxygen. Install & forget. Sistema autolimpiable. Autoclean system Sistema autolimpiable Dissolved Oxygen Autoclean system Applications: - water treatment - activated sludge - de-nitrification - fish pond Install & forget For special applications, the D.O monitor can

More information

Powers Controls TH 192 HC Heating/Cooling Room Thermostat

Powers Controls TH 192 HC Heating/Cooling Room Thermostat Powers Controls TH 192 HC Heating/Cooling Room Thermostat Technical Instructions Document No. 155-066P25 TH 192-2 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 192 HC thermostats are proportional

More information

Gems Sensors offers a complete line of products for these industries:

Gems Sensors offers a complete line of products for these industries: Model 830 Instruction Bulletin No. 197070 Rev. A Wet/Wet Differential Pressure Transducer Gems Sensors offers a complete line of products for these industries: Industrial HVAC Test & Measurement Barometric

More information

Solutions for Storage Tanks

Solutions for Storage Tanks www.tankgaugingsys.com Solutions for Storage Tanks 5020 Top Mounted Sour Service Mechanical Level Gauge TGS 020 1.) POINT SWITCH CSA Class 1, Div.1 Magnetically Activated Proximity Switches for high/low

More information

Powers Controls TH 192 S Single Temperature Room Thermostat

Powers Controls TH 192 S Single Temperature Room Thermostat Technical Instructions Document No. 155-065P25 TH 192-1 Powers Controls TH 192 S Single Temperature Room Thermostat 50 60 70 80 TH0355R1 70 80 POWERS Description The TH 192 S thermostats are proportional

More information

Powers Controls Free Energy Band TH 193 HC Heating/Cooling Room Thermostat

Powers Controls Free Energy Band TH 193 HC Heating/Cooling Room Thermostat Document No. 155-068P25 TH 193-4 Powers Controls Free Energy Band TH 193 HC Heating/Cooling Room Thermostat 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 193 HC thermostats are proportional dual

More information

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION INSTALLATION INSTRUCTIONS CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION The CVS Controls 67CFR Filter regulator is a pressure reducing supply regulator typically used for pneumatic

More information

High Purity ph Sensor

High Purity ph Sensor Instruction Manual PN 51-320HP/rev.J May 2006 Model 320HP High Purity ph Sensor ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products

More information

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01 Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery

More information

JACKET/FRAME ASSEMBLY See page 4

JACKET/FRAME ASSEMBLY See page 4 PSD-3-148 WIRING/HARNESS & PROBES See page 8 PROPANE GAS TRAIN ASSEMBLY See page 6 COMMERCIAL POWER BURNER BOILER REPLACEMENT PARTS LIST COPPER BOILERS FOR HYDRONIC HEATING & HOT WATER SUPPLY LEGEND MODELS

More information

Welcome. Basics about ph-measurement. Dipl.-Ing. Manfred Schleicher

Welcome. Basics about ph-measurement. Dipl.-Ing. Manfred Schleicher Welcome Basics about ph-measurement Dipl.-Ing. Manfred Schleicher This presentation helps for starting with ph-measurement Index Definition ph-value ph- electrode and temperature depency of ph measurement

More information

TWO-WIRE TRANSMITTERS FOR ph, ORP, CONDUCTIVITY, OXYGEN, OZONE, AND CHLORINE

TWO-WIRE TRANSMITTERS FOR ph, ORP, CONDUCTIVITY, OXYGEN, OZONE, AND CHLORINE Product Data Sheet 71-5081/rev.K April 2006 Model 5081 TWO-WIRE TRANSMITTERS FOR ph, ORP, CONDUCTIVITY, OXYGEN, OZONE, AND CHLORINE Model 5081 Family of Two-wire Transmitters CHOICE OF COMMUNICATION PROTOCOL:

More information

SE 655(X) / SE 656(X) Toroidal Conductivity Sensors

SE 655(X) / SE 656(X) Toroidal Conductivity Sensors User Manual English SE 655(X) / SE 656(X) Toroidal Conductivity Sensors Latest Product Information: www.knick.de Description 3 The SE 655(X) and SE 656(X) toroidal conductivity sensors are particularly

More information

Electronic Flow Switches

Electronic Flow Switches Industries Electronic Flow Switches Petrochemical Refining Oil Production Water Treatment Pharmaceutical Food and Beverage Pulp and Paper Power Production Gas Processing Mining Biotechnology Semiconductor

More information

Electronic Level Switches

Electronic Level Switches Electronic Level Switches LX Series Applications Universal High, Low Level Oil/Water Interface Emulsion from Oil/Water Air/Foam Interface Foam/Liquid Interface Solids detection in a Liquid Top Mounted

More information

Humidity Sensor/Transmitter Assembly

Humidity Sensor/Transmitter Assembly Humidity Sensor/Transmitter Assembly Wall/Duct Space Outside Air (OSA) Overview Minco humidity and humidity/temperature transmitters are designed using an advanced microprocessor. Digital signal processing

More information

Self Purging Liquid Trap (SPLT ) User s Manual

Self Purging Liquid Trap (SPLT ) User s Manual Self Purging Liquid Trap (SPLT ) User s Manual High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be released unexpectedly and with extreme force. High

More information

NT PRESSURE TRANSDUCERS FOR DEIONIZED WATER SYSTEMS

NT PRESSURE TRANSDUCERS FOR DEIONIZED WATER SYSTEMS NT PRESSURE TRANSDUCERS FOR DEIONIZED WATER SYSTEMS NT PRESSURE TRANSDUCERS FOR DEIONIZED WATER SYSTEMS reliable pressure measurement Reliable Pressure Measurement Concerns such as uptime, contamination

More information

HG600/HG800. Hydro-Power Unit. Installation and Operation Guide

HG600/HG800. Hydro-Power Unit. Installation and Operation Guide HG600/HG800 Hydro-Power Unit Installation and Operation Guide Copyright 04 HydroSpin, All rights reserved. The data contained in this document is proprietary to HydroSpin. It is disclosed to the receiving

More information

UltraBasic Benchtop ph Meter Operation Manual

UltraBasic Benchtop ph Meter Operation Manual UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.

More information

Product Data. A/RH Series Relative Humidity Transmitters

Product Data. A/RH Series Relative Humidity Transmitters Relative Humidity DISPLAYS A CI Product Data A/RH Series Relative Humidity Transmitters First in Sensor Quality The A/RH Series Relative Humidity transmitters, convert a resistance change to a linear to

More information

Baumann 87000 Flexsleev Sanitary Control Valve

Baumann 87000 Flexsleev Sanitary Control Valve Product Bulletin D103345X012 Baumann 87000 Flexsleev Sanitary Control Valve The Baumann 87000 control valve (figures 1 and 2) is excellent for throttling high purity liquid or gaseous media commonly found

More information

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference

More information

VOYAGER 570G. 744A Sprayer Control

VOYAGER 570G. 744A Sprayer Control VOYAGER 570G 744A Sprayer Control U S E R M A N U A L U S E R M A N U A L Table of Contents CHAPTER 1 - INTRODUCTION...1 SYSTEM CONFIGURATIONS...1 KIT CONTENTS...3 CONTROL HOUSING ASSEMBLY...5 CHAPTER

More information

BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS

BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS PLEASE READ COMPLETE INSTRUCTIONS BEFORE YOU BEGIN LEAVE INSTALLATION INSTRUCTIONS AND USER'S GUIDE WITH OWNER ALL ELECTRIC WIRING AND PLUMBING MUST BE DONE

More information

Stainless Steel Single and Dual Circulation Kits

Stainless Steel Single and Dual Circulation Kits Instruction Sheet P/N 160780 01 Stainless Steel Single and Dual Circulation Kits Introduction The single and dual high-pressure circulation kits allow you to vary and control the circulation rate of coating

More information

Replacement parts. WM97+ gas-fired water boiler Boiler Manual. OBTAIN PARTS ONLY THROUGH WEIL-McLAIN THE BOILER CONTAINS CERAMIC FIBER MATERIALS

Replacement parts. WM97+ gas-fired water boiler Boiler Manual. OBTAIN PARTS ONLY THROUGH WEIL-McLAIN THE BOILER CONTAINS CERAMIC FIBER MATERIALS Replacement parts DO NOT SERVICE THE BOILER WITHOUT A WM97+ MAINTENANCE KIT AVAILABLE Failure to adhere to these guidelines can result in severe personal injury, death or substantial property damage. The

More information

Mobrey Ultrasonic. MSP900SH Level and MSP900FH Flow Transmitters. Product Data Sheet February 2015 IP2032, Rev BA

Mobrey Ultrasonic. MSP900SH Level and MSP900FH Flow Transmitters. Product Data Sheet February 2015 IP2032, Rev BA Mobrey Ultrasonic MSP900SH Level and MSP900FH Flow Transmitters Product Data Sheet February 2015 IP2032, Rev BA Non-contacting measurement with no moving parts Fast and simple to install and configure

More information

GENERAL USE AND LABORATORY ph ELECTRODES PRODUCT SPECIFICATION SHEET

GENERAL USE AND LABORATORY ph ELECTRODES PRODUCT SPECIFICATION SHEET GENERAL USE AND LABORATORY ph ELECTRODES The right sensor makes the difference. Different applications require different types of ph sensors to ensure measurement accuracy and extend instrument lifetime.

More information

Fisher 2100 Pneumatic and 2100E Electric Liquid Level Switches

Fisher 2100 Pneumatic and 2100E Electric Liquid Level Switches 2100 and 2100E Level Switches Product Bulletin Fisher 2100 Pneumatic and 2100E Electric Liquid Level Switches Fisher 2100 on-off pneumatic switch and 2100E electric switch sense (shown in figure 1) high

More information

WEL Series ph/orp ELECTRODES

WEL Series ph/orp ELECTRODES Iwaki America Inc. WEL Series ph/orp ELECTRODES Instruction Manual Five Boynton Road Hopping Brook Park Holliston, MA 01746 TEL: 508-429-1110 Web: www.walchem.com Notice 2016 WALCHEM, An Iwaki America

More information

430 ph Meter. Instruction Manual

430 ph Meter. Instruction Manual 40 ph Meter Instruction Manual This manual contains complete instructions for setting up and using the 40 ph meter. Applications information is also available. The information contained in this manual

More information

Services The PT, ph-temperature driver measures a ph sensor and a temperature using a platinum RTD.

Services The PT, ph-temperature driver measures a ph sensor and a temperature using a platinum RTD. PT: erature Safety +/-1VDC maximum on field wiring terminals. 24 VDC maximum on internal card surfaces. 1.1 Installation Services The PT, -erature driver measures a sensor and a temperature using a platinum.

More information

In-Line Electronic Flow Switches

In-Line Electronic Flow Switches In-Line Electronic Flow Switches Industries Petrochemical Refining Oil Production Water Treatment Pharmaceutical Food and Beverage Pulp and Paper Power Production Gas Processing Mining Biotechnology Semiconductor

More information

Introduction. Safety Guidelines. Part Numbers. Operation and Maintenance Manual. Operation and Maintenance Manual

Introduction. Safety Guidelines. Part Numbers. Operation and Maintenance Manual. Operation and Maintenance Manual Operation and Maintenance Manual Introduction Read all instructions thoroughly. Installation of the OilTector must comply with all Federal, State and Local Codes, Regulations and Practices. The OilTector

More information

OxyGuard. Dissolved Oxygen Probe

OxyGuard. Dissolved Oxygen Probe OxyGuard Dissolved Oxygen Probe Standard type USER MANUAL D02 Standard Probe manual gb 0903 CONTENTS 1. DESCRIPTION 1.1 The OxyGuard Probe... 1 1.3 Which kind of transmitter can be used with the OxyGuard

More information

Installation Instructions / Warranty

Installation Instructions / Warranty Installation Instructions / Warranty Vertical Spray Bidet Fitting Rough 1322X181 Citterio 39224xx0 Citterio 39225xx0 Starck 10220xx0 Steel 35220800 Uno 38220xx0 Terrano 37224xx0 Terrano 37225xx0 Carlton

More information

Level Plus. Magnetostrictive Liquid-Level Sensors with Temposonics Technology. M-Series Model MR Transmitter with Analog Output.

Level Plus. Magnetostrictive Liquid-Level Sensors with Temposonics Technology. M-Series Model MR Transmitter with Analog Output. Level Plus Magnetostrictive Liquid-Level Sensors with Temposonics Technology M-Series Model MR Transmitter with Analog Output Data Sheet SENSORS Document Part Number 550677 Revision J Features 4 to 20

More information

Instruction Manual LIQ-MAN-5081P/rev.J January 2015. Model 5081-P. Two-Wire ph/orp Transmitter

Instruction Manual LIQ-MAN-5081P/rev.J January 2015. Model 5081-P. Two-Wire ph/orp Transmitter Instruction Manual LIQ-MAN-5081P/rev.J January 2015 Model 5081-P Two-Wire ph/orp Transmitter ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests

More information

Series Regulators - Pressure Reducing

Series Regulators - Pressure Reducing 26-1000 Series Regulators - Pressure Reducing D26100538X012 Specifications For other materials or modifications, please consult TESCOM. OPERATING PARAMETERS Pressure rating per criteria of ANSI/ASME B31.3

More information

31000 and Proximity Probe Housings Bently Nevada* Asset Condition Monitoring

31000 and Proximity Probe Housings Bently Nevada* Asset Condition Monitoring 31000 and 32000 Proximity Probe Housings Bently Nevada* Asset Condition Monitoring Description The 31000/32000 Proximity Probe Housing Assemblies are recommended when mounting proximity probes through

More information

Table of Contents accumet

Table of Contents accumet AR50 USER Manual Table of Contents accumet Introduction 4 Unpacking the Meter 5 Specifications 6 Getting Started Connectors Attaching ph and ISE electrodes Attaching Conductivity Probes 8 10 12 Using the

More information

Product Data. Model 575 Series Submersible Level Transmitters ELECTRONIC PRESSURE MEASUREMENT PRODUCTS DESCRIPTION FEATURES APPLICATIONS

Product Data. Model 575 Series Submersible Level Transmitters ELECTRONIC PRESSURE MEASUREMENT PRODUCTS DESCRIPTION FEATURES APPLICATIONS EPP - 1 DECRIPTION The odel 575 is used in general water/wastewater applications. The transmitter is offered in a standard non-clogging snub nose end, flush diaphragm or protected flush diaphragm with

More information

ph/orp Electrode System 7600 Series

ph/orp Electrode System 7600 Series Instruction Manual IM/7650 Issue 9 ph/orp Electrode System 7600 Series ABB The Company We are an established world force in the design and manufacture of instrumentation for industrial process control,

More information

4909 316SS Conductivity Cell Insertion/Removal Assembly Operations Manual. 70-82-25-68 Rev. 1 7/99

4909 316SS Conductivity Cell Insertion/Removal Assembly Operations Manual. 70-82-25-68 Rev. 1 7/99 4909 316SS Conductivity Cell Insertion/Removal Assembly Operations Manual 70-82-25-68 Rev. 1 7/99 Contents Copyright, Notices, and Trademarks Printed in U.S.A. Copyright 1999 by Honeywell Inc. Revision

More information

Installation Method 1: Lays Into the Ceiling Grid. Package Contents

Installation Method 1: Lays Into the Ceiling Grid. Package Contents 5731 Benjamin Center Drive Tampa, Florida 33634 Phone:813.349.1900 Fax:813.349.1907 Version: A Release Date: 2013-07-24 READ THIS MANUAL CAREFULLY AND KEEP IT ACCESSIBLE FOR FURTHER REFERENCE! USE AS DIRECTED:

More information

BLUE RIBBON CORP. BC001 Birdcage Installation Manual

BLUE RIBBON CORP. BC001 Birdcage Installation Manual Product Overview: This manual is applicable for the BC001 Birdcage Submersible Level Transmitters. If the product you have has a different prefix part number than listed above please contact the factory

More information

Z SERIES PRESSURE TANKS INSTALLATION INSTRUCTIONS

Z SERIES PRESSURE TANKS INSTALLATION INSTRUCTIONS www.burcam.com 2190 Dagenais Blvd. West TEL: 514.337.4415 LAVAL (Quebec) FAX: 514.337.4029 CANADA H7L 5X9 infoburcam.com This product has been carefully packaged at the factory to prevent damage during

More information

INLINE flowmeter for continuous flow measurement

INLINE flowmeter for continuous flow measurement INLINE flowmeter for continuous flow measurement Economic integration in pipe systems without any additional piping 3-wire frequency pulse version to directly interface with PLC s (both PNP and NPN) Connection

More information

Sanitary Type Electronic Level Switches

Sanitary Type Electronic Level Switches Sanitary Type Electronic Level Switches SL Series Applications Universal High, Low Level Organics/Water Interface Emulsion from Organics/Water Air/Foam Interface Foam/Liquid Interface Solids detection

More information

Electrochemistry. Standard ph Electrodes. Speciality ph & ORP Electrodes. electrodes

Electrochemistry. Standard ph Electrodes. Speciality ph & ORP Electrodes. electrodes Standard ph Electrodes All are supplied with 1 m of cable and BNC plug (+ 2 banana plugs built-in ATC) SP10B, SP10T SP10B: ph SP10T: ph + ATC (Pt) SP20B, SP20T SP20B: ph SP20T: ph + ATC (Pt) Speciality

More information

2 WAY (2/2) AND 3 WAY (3/2) BRASS AND STAINLESS STEEL SOLENOID VALVES

2 WAY (2/2) AND 3 WAY (3/2) BRASS AND STAINLESS STEEL SOLENOID VALVES 2 WAY (2/2) AND 3 WAY (3/2) BRASS AND STAINLESS STEEL SOLENOID VALVES DIRECT ACTING Brochure N-289 SPECIFICATIONS Port Size: 1/8 and1/4 Solenoid Valve for air, oil, water and other fluids Flow Direction:

More information

before tightening the bolts. This will help eliminate the potential for any interference between the flywheel housing and the oil pan.

before tightening the bolts. This will help eliminate the potential for any interference between the flywheel housing and the oil pan. SERIES 60 SERVICE MANUAL before tightening the bolts. This will help eliminate the potential for any interference between the flywheel housing and the oil pan. 1. Drain Plug 6. Bolt 2. Nut 7. Plug 3. Isolator

More information

Models SFP33 & SFP32 Pressure Transmitters

Models SFP33 & SFP32 Pressure Transmitters SFP33 Flow-Through Pressure Transmitter SFP32 Single-Ended Pressure Transmitter DS-PR-SFP-eng April, 2016 Data Sheet Models SFP33 & SFP32 Pressure Transmitters Pressure Brooks Seal Free Pressure Measurement

More information

MICROCHEM 2 SERIES 4000

MICROCHEM 2 SERIES 4000 MICROCHEM 2 SERIES 4000 The MicroChem 2 is a versitile multiparameter instrument capable of being configured as a transmitter, PID controller, or swimming pool controller. Specifically designed for drinking

More information