1 Running head: DIGITAL BREAST TOMOSYNTHESIS AND MAMMOGRAPHY 1 Digital Breast Tomosynthesis and Mammography November 29 th, 2011
2 BREAST TOMOSYNTHESIS 2 Abstract Breast cancer is an inevitable disease that has been successfully screened, diagnosed and treated through technology advancements. The demand for accurate detection has increased, forcing the demand for more advances in technology. Digital breast tomosynthesis (DBT) has provided the necessary advancements in breast imaging. Breast cancer can now be seen more readily than ever before because of the 3D tomosynthesis images. The collaboration of multiple modalities now provides increasing detection and survival rates.
3 BREAST TOMOSYNTHESIS 3 Digital Breast Tomosynthesis and Mammography Breast cancer is the most prevalent cancer in women in the United States and "remains one of the leading causes of death in women over the age of 40 years" (Hakim et al., 2010, p. W172). Helvie (2010) discusses the importance and value that digital mammography (DM) possesses. With controlled, random trials, DM has proven over and over again the survival benefits for patients. DM is the only imaging modality that has shown these benefits consistently. Digital Breast Tomosynthesis (DBT) is an advancement in mammography that can continue to prove survival benefits. Because DBT is consistent with DM, the superiority of this modality in relation to the detection of breast cancer will continue to surpass Magnetic Resonance Imaging (MRI), Computed Tomography (CT) and Ultrasound. As mammography continues to be a great way for breast cancer detection, the amount of patients experiencing requested recall imaging is high. The amount of missed breast cancers is also a concern for many patients because of the delay in treatment and more aggressive approach necessary due to the delay. Each woman's breast configuration varies greatly causing the diagnosis of breast cancer to be more difficult than one may think. By using large sensor size and high resolution detectors Hakim et al. (2010) discuss that they "offer not only better contrast sensitivity but also great opportunities to develop advanced digital imaging techniques that could improve conspicuity of breast lesions by enabling better tissue visualization through the provision of 3D nonoverlapped tissue information" (p. W172). DBT is designed to create 3D images in order to have the best image quality. "Its main advantages are the improvement of the conspicuity of structures, the possibility of depth localization, and the smaller dynamic range required for each reconstructed slice" (Vecchio et
4 BREAST TOMOSYNTHESIS 4 Fig. 1 Illustration of DBT schematic view NOTE. Helvie, M. 2010, Digital mammography imaging: breast tomosynthesis and advanced applications. Retrieved October 10, 2011, from libpublic3.library.isu.edu/pmc/articles/pmc /?tool= al., 2011, p. 1207). DBT equipment retrieves multiple projection exposures obtained over a limited arc angle. (See Fig. 1) These images will then be reconstructed using unique algorithms. Chan et al. (2010) discuss two different ways to obtain the 3D images. The first is to combine the information obtained from the image analysis of multiple 2D projection view (PV) images. This approach can be beneficial because the projection view images are independent of the method of reconstruction. The reduction of noise and false positives when using the PV approach is due in part to the correlation between the images after the union of information obtained from the multiple views. The second way to obtain the 3D images is to perform the image analysis of the obtained slices after the tomosynthesis reconstruction of the projection views. "This approach takes advantage of the image reconstruction technique in combining the spatial information accurately before image analysis" (Chan et al., 2010, p. 3577). Williams, Judy, Gunn, and Majewski (2010) discuss the discomfort that a patient is in when receiving a mammogram because of the compressive force. With DBT the compressive force is reduced, providing more comfort to the patient during the exam. Less force is needed to be used because of the multiple sections that are formed into 3D images. With DBT, breast masses will be seen more readily as compared to digital mammography alone. "It is considered a possible solution to overcome the main shortcoming in two-
5 BREAST TOMOSYNTHESIS 5 dimensional (2D) conventional mammography: the overlap of anatomical structures on pathologies of interest" (Shaheen et al., 2010, p. 108). The images are presented to the radiologists much like a CT or MRI study. The reconstructed slices can vary in thickness, being as thin as.5 mm. Thin slices will project masses that are often times superimposed with breast tissue when retrieving 2D breast images. As radiologists become familiar with the creation of the DBT 3D images, the detection rate and treatment of DBT patients will greatly increase. According to Gennaro et al. (2010) designing the equipment for tomosythesis has been in the works since the 1930s but has been transformed and used more readily in the past ten years because of the advancements in technology. The design of the DBT machine has progressed rapidly in the past 5 years. The Food and Drug Administration (FDA) approved the Holigic Selenia Dimensions 3D System on February 11, (See Fig. 2) The Affirm upright breast biopsy guidance system and the ImageChecker 3D Calc computer-aided detection are two new products that are specifically created for the new Fig. 2 Hologic Selenia 3D NOTE. Retrieved November 1, 2011 from ologic-gets-fda-nod-for-seleniadimensions/ Selenia machine. Currently Affirm is approved for sale in the United States while Image Checker is still in the works. There are opportunities for radiologists to addend conferences and experience firsthand what reading 3D breast images are like. If imaging departments receive proper education on the advantages of DBT, patients will soon see the benefits of this advancing modality.
6 BREAST TOMOSYNTHESIS 6 Algorithm Reconstruction Now that DBT is emerging at a fast pace, many details are being explored in order to have the most efficient imaging equipment and techniques. The algorithm reconstruction that takes place after the multiple images have been captured is a crucial area for the creation of the 3D images. According to Sidky et. al (2009) iterative reconstruction is most advantageous for DBT. The incomplete data that is originally obtained from a tomographic system lacks the projection data that is much needed. As a result, regularization of the image takes on two roles: (1) Selection of a unique image among those that agree with the projection data and (2) the traditional role where the image is regularized while relaxing consistency with the available data. In the first role, the image regularization is lowered while the image is constrained to a given data agreement. In the second role the data constraint on the image is relaxed, allowing for further minimization of the image regularization (Sidkey et. al, 2009, p. 4921). In order to obtain data consistency as well as physical consistency, the reconstructed algorithm uses projection onto convex sets (POCS). Because the algorithm has steep descent components and the POCS step sizes needed to be more easily controlled, B A another element was added to the POCS to ensure quality images. The adaptive steepest decent-pocs Fig. 3 Microcalcifications shown in both A. DM and B. DBT NOTE. Helvie, M. 2010, Digital mammography imaging: breast tomosynthesis and advanced applications. Retrieved October 10, 2011, from libpublic3.library.isu.edu/pmc/articles/pmc /?tool= (ASC-POCS) algorithm "allows for the separation of the two roles for the regularizer in tomographic image reconstruction from incomplete projection
7 BREAST TOMOSYNTHESIS 7 data" (Sidky et al., 2009, p. 4921). Now that ASC-POCS has been adopted as a new algorithm, the imaging of microcalcifications will be greatly affected. (See Fig. 3) The contrast of images will increase, showing better quality images as well as more diagnostic images. The most significant part of this advancement is the concept of decreasing patient dose as well as reducing the x-ray intensity. There is hope that an even more efficient algorithm can be produced based off of the ASC-POCS algorithm. Imaging Dense Breast Tissue Women with dense breast tissue are the most difficult patients to image. There are many opportunities for dense breast tissue to overlap on itself, hiding tumors and lesions. With DBT, not only are cancers shown more readily, the margins of the cancers are shown giving the radiologist more information without additional mammography views. Because the number of women with dense A C B D breast tissue is rapidly increasing; finding a way to accurately image these patients is a priority. According to Helvie (2010) "The primary benefit of DBT would be expected to be for Fig. 4 (A) Mediolateral oblique (MLO) DM (B) MLO DBT (C) Cranial caudal (CC) DM (D) CC DBT. This patient has invasive ductal cancer. NOTE. Helvie, M. 2010, Digital mammography imaging: breast tomosynthesis and advanced applications. Retrieved October 10, 2011, from libpublic3. library.isu.edu/pmc/articles/pmc /?tool=
8 BREAST TOMOSYNTHESIS 8 noncalcified mammographic findings such as masses, asymmetries, and distortion" (p. 919). By improving sensitivity, more cancers will be detected in dense breast tissue and the number of callbacks will decrease. (See Fig. 4) Imaging Non-Dense Breast Tissue Women with non-dense breast tissue, or fatty tissue, are ideal patients for mammography. Although fatty tissue can be imaged with digital mammography with higher detection rates compared to dense tissue, the detection of smaller lesions is greater with DBT. "This is a variant of improved sensitivity as a decrease in size at time of detection may be associated with improvement in clinical outcome" (Helvie, 2010, p. 919). Another advantage of DBT compared Fig. 5 DM vs DBT on a patient who has a mass NOTE. Helvie, M. 2010, Digital mammography imaging: breast tomosynthesis and advanced applications. Retrieved October 10, 2011, from libpublic3.library. isu.edu/pmc/articles/pmc /?tool= to DM when imaging fatty tissue is the obvious border of the lesion. (See Fig. 5) The diagnostic differences between DM and DBT are clear. Radiation Dose When new radiology equipment is introduced into the medical field, many patients, technologists and physicians question the dose the patient will receive. The balance between image quality and patient dose is always a major concern. The FDA has set a limit of 300 mrad per exposure for breast imaging; 150 to 250 mrad being the usual range for 2D mammography. Although DBT does not exceed the FDA limit, it is a slightly higher initial dose to patients compared to conventional mammography.
9 BREAST TOMOSYNTHESIS 9 Hendrick (2010) discusses the radiation doses received from screen-film mammography (SFM), digital mammography (DM) and DBT. A two view DM exam administers a lower dose to the breast compared to that of a SFM exam while a single view DBT exam administers a comparable dose as a two view DM exam. Depending on the DBT image acquisition strategy, the number of images requested can vary between a single view DBT, two view DBT or DBT and DM combined. Each patient needs to evaluate their individual risks and benefits when evaluating breast imaging. According to Hendrick (2010) "on average, approximately 10% of women in the United States are recalled on the basis of screening mammography findings" (p. 250). Recall imaging usually involves two or three additional views with a higher dose. Things to consider before receiving breast imaging include previous recall image rate, previous radiation exposure and breast tissue density. "If DBT leads to reduction in recall rate or improvement in sensitivity and specificity, a minimally higher dose may be acceptable" (Helvie, 2010, p. 919). Case Study Bertolini et al. (2010) conducted a study comparing DBT and Full-field digital mammography (FFDM) using images from both breast modalities having a contrast detail phantom as a guiding tool. A pixel pitch size of 10µm was used with a 24 cm x 29 cm dimension. The same pixel pitch size was used for the FFDM images, while the DBT had a pixel pitch size of 88 µm after reconstruction. According to Bertolini et al. "The DBT images were reconstructed at 1-mm in-plane spacing (between slices)" (p. 59). As the machine was in motion, it obtained 15 slices across the arc. The contrast detail (CD) phantom was designed to detect masses and calcifications. To decipher between the two, multiple holes of different depths and diameters were included on the
10 BREAST TOMOSYNTHESIS 10 phantom. (See Fig. 6) If the object diameter was greater than 0.5mm, the classification would be a mass, if the object diameter was less than 0.5mm, the classification would be a calcification. "The CD phantom was positioned between two thick layers containing a breast-simulation Fig. 6 FFDM image of phanton without confounding material NOTE. Bertolini et al., Contrast detail phantom comparison on a commercially available unit. Digital breast tomosynthesis (DBT) versus full-field digital mammography (FFDM). Retrieved October 17, 2011 from sstoken=6lll9ll89ol9ov9xzooh3fvlzjz28hlvkk&show =Object&ErrorURL=http%3A%2F%2Flinksource.ebsco.com%2Ferr or.aspx. Reprinted with permission. material. This confounding layer was obtained by properly treating surgical breast specimens. The phantom contained a set of radiopaque markers for image registration" (Bertolini, 2010, p.59). Both FFDM and DBT used automatic exposure control (AEC) when producing images. The acquisition pattern for both imaging modalities was equal, having acquired images from 0, 90, 180, 270 of CD phantom rotation. A technique of 29 kilovoltage peak (kvp) with an anti-scatter grid was used for FFDM using AEC and a technique of 32 kvp with no grid was selected for DBT. (See Fig. 7) "Two different Fig. 7 Phantom with material in place (C) FFDM (D) DBT NOTE. Bertolini et al., Contrast detail phantom comparison on a commercially available unit. Digital breast tomosynthesis (DBT) versus full-field digital mammography (FFDM). Retrieved October 17, 2011 from isu.edu/direct.asp?accesstoken=6lll9ll89ol9ov9xzooh3fv LZJZ28HLVKK&Show=Object&ErrorURL=http%3A%2F%2Flink source.ebsco.com%2ferror.aspx. Reprinted with permission. representations of the same object were obtained with the two techniques" (Bertolini, 2010,
11 BREAST TOMOSYNTHESIS 11 p.60). The comparison of the FFDM image and the DBT image is outstanding. The differences in noise, contrast and spacial system resolution seen on the FFDM and DBT images is exceptional. Conclusion With the advances taking place in medical imaging, particularly breast imaging, breast cancer victims will have a much greater survival rate. Now that DBT has been approved by the FDA, better breast image quality is available. Patients with different breast size, shape and density are now able to be at ease knowing that better breast imaging options are obtainable. DBT is the future of breast imaging and is going to surpass current mammography expectations.
12 BREAST TOMOSYNTHESIS 12 References Bertolini, M., Nitrosi, A., Giovanni, B., Botti, A., Tassoni, D., Sghedoni, R., & Zuccoli, G. (2011). Contrast detail phantom comparison on a commercially available unit. Digital breast tomosynthesis (DBT) versus full-field digital mammography (FFDM). Journal of Digital Imaging, 24(1), Retrieved from Direct.asp?AccessToken=6LLL9LL89OL9OV9XZOOH3FVLZJZ28HLVKK&Show=O bject&errorurl=http%3a%2f%2flinksource.ebsco.com%2ferror.aspx Chan, H.P., Wu, Y.T., Sahiner, B., Wei, J., Hlevie, M., Zhang, Y.,...Way, T. (2010). Characterization of masses in digital breast tomosynthesis: comparison of machine learning in projection views and reconstructed slices. Medical Physics, 37(7), Retrieved from Gennaro, G., Toledano, A., Maggio, C.D., Baldan, E., Bezzon, E., La Grassa, M.,...Muzzio, P.C. (2010). Digital breast tomosynthesis versus digital mammography: a clinical performance study. European Society of Radiology, 20, doi: /s Hakim, C.M., Chough, D.M., Ganott, M.A., Sumkin, J.H., Zulie, M.L., & Gur, D. (2010). Digital breast tomosynthesis in the diagnostic environment: a subjective side-by-side review. American Journal of Roentgenology, 195, doi: /AJR Helvie, M.A. (2010). Digital mammography imaging: breast tomosynthesis and advanced applications. National Institutes of Health, 48(5), doi:10/1016/j.rcl Hendrick, E.R. (2010). Radiation doses and cancer risks from breast imaging studies. Radiology, 257, Retrieved from Hologic The Women's Health Company Web site, (n.d.). Hologic showcases its 3d mammography (breast tomosynthesis) at national consortium of breast centers (NCBC)
13 BREAST TOMOSYNTHESIS 13 meeting. Retrieved from year.2011_173-id html Shaheen, E., Zanca, F., Sisini, F., Zhang, G., Jacobs, J., & Bosmans, H. (2010). Simulation of 3D objects into breast tomosynthesis images. Radiation Protection Dosimetry, 139, doi: /rpd/ncq084 Sidky, E.Y., Pan, X., Reiser, I.S., Nishikawa, R.M., Moore, R.H., & Kopans, D.B. (2009). Enhanced imaging of microcalcifications in digital breast tomosynthesis through improved image-reconstruction algorithms. Medical Physics, 36(11), doi: / Vecchio, S., Albanese, A., Vignoli, P., & Taibi, A. (2010). A novel approach to digital breast tomosynthesis for simultaneous acquisition of 2D and 3D images. European Society of Radiology, 21, doi: /s y Williams, M.B., Judy, P.G., Gunn, S., & Majewski, S. (2010). Dual-modality breast tonosynthesis. Radiology, 255, Retrieved from pmc/articles/pmc /?tool=pubmed
Mammography. What is Mammography?
Scan for mobile link. Mammography Mammography is a specific type of breast imaging that uses low-dose x-rays to detect cancer early before women experience symptoms when it is most treatable. Tell your
More informationRole of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
More informationHow To Compare Mammography To A 3D System
Radiology Advisory Panel Meeting Hologic Selenia Dimensions 3D System with C-View Software Module FDA Review Robert Ochs, PhD Branch Chief Mammography, Ultrasound, and Imaging Software Branch Division
More informationHologic Selenia Dimensions C-View Software Module. October 24, 2012
Hologic Selenia Dimensions C-View Software Module October 24, 2012 Introduction and Agenda Peter Soltani, Ph.D. Senior VP & GM, Breast Health Hologic, Inc. Agenda Technology Overview Clinical Overview
More informationPersonalized Breast Screening Service
Frequently Asked Questions WHAT IS BREAST DENSITY? Breasts are made up of a mixture of fibrous, glandular and fatty tissue. Your breasts are considered if you have predominantly fibrous or glandular tissue
More informationA Guide to Breast Imaging: The Latest Technology for Screening and Detecting Breast Cancer
A Guide to Breast Imaging: The Latest Technology for Screening and Detecting Breast Cancer Sally Herschorn, MD Associate Professor of Radiology University of Vermont College of Medicine Medical Director
More informationBreast Imaging Made Brief and Simple. Jane Clayton MD Associate Professor Department of Radiology LSUHSC New Orleans, LA
Breast Imaging Made Brief and Simple Jane Clayton MD Associate Professor Department of Radiology LSUHSC New Orleans, LA What women are referred for breast imaging? Two groups of women are referred for
More informationQ.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
More informationPatient Prep Information
Stereotactic Breast Biopsy Patient Prep Information Imaging Services Cannon Memorial Hospital Watauga Medical Center Table Weight Limits for each facility Cannon Memorial Hospital Watauga Medical Center
More informationBREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY
BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY N. GEERAERT a,b,c, R. KLAUSZ a, S. MULLER a, I. BLOCH c, H. BOSMANS b a GE Healthcare, Buc, France b Department
More informationBreast Density Legislation: Implications for primary care providers
Breast Density Legislation: Implications for primary care providers Deborah J. Rhodes MD Associate Professor of Medicine 2012 MFMER slide-1 Disclosure Relevant financial relationship(s) None Off-label
More informationBreast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography 1
Francesco Sardanelli, MD Franco Zandrino, MD Andrea Imperiale, MD Emma Bonaldo, MD Maria G. Quartini, RT Nadia Cogorno, RT Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography
More informationMAMMOGRAPHY GOALS AND OBJECTIVES
MAMMOGRAPHY GOALS AND OBJECTIVES GOALS: After completion of the mammography rotations, the resident will be able to: 1. Demonstrate learning of the knowledge-based objectives-(practice Base Learning) 2.
More informationHarlem Hospital Center Integrated Radiology Residency Program Mammography Educational goals and objectives
Harlem Hospital Center Integrated Radiology Residency Program Mammography Educational goals and objectives Rotation 1 (Radiology year 1/2) Knowledge Based Objectives: At the end of the rotation, the resident
More informationMammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection
Digital Mammography Upright Stereo Mammography For 35+ years, screenfilm has been the gold standard for breast cancer detection IN THE BEGINNING Mammography technology has come a long way since the first
More informationLow-dose CT Imaging. Edgar Fearnow, M.D. Section Chief, Computed Tomography, Lancaster General Hospital
Lung Cancer Screening with Low-dose CT Imaging Edgar Fearnow, M.D. Section Chief, Computed Tomography, Lancaster General Hospital Despite recent declines in the incidence of lung cancer related to the
More informationDose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
More informationRunning head: OBESITY AND MEDICAL IMAGING 1. Obesity and Medical Imaging. November 13, 2012
Running head: OBESITY AND MEDICAL IMAGING 1 Obesity and Medical Imaging November 13, 2012 OBESITY AND MEDICAL IMAGING 2 Abstract Obesity is a growing problem in the United States, and it affects the health
More informationHow To Write A Draft Report On Breast Cancer Screening
Health Technology Assessment Appropriate Imaging for Breast Cancer Screening in Special Populations Draft Evidence Report: & December 10, 2014 Health Technology Assessment Program (HTA) Washington State
More informationSustaining a High-Quality Breast MRI Practice
Sustaining a High-Quality Breast MRI Practice Christoph Lee, MD, MSHS Associate Professor of Radiology Adjunct Associate Professor, Health Services University of Washington September 11, 2015 Overview
More informationConcepts for High-Resolution Low-Dose CT of the Breast
RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi
More informationSurveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,
More informationStereotactic Breast Biopsy
Scan for mobile link. Stereotactic Breast Biopsy Stereotactic breast biopsy uses mammography a specific type of breast imaging that uses low-dose x-rays to help locate a breast lump or abnormality and
More informationDigital Mammogram National Database
Digital Mammogram National Database Professor Michael Brady FRS FREng Medical Vision Laboratory Oxford University Chairman: Mirada Solutions Ltd PharmaGrid 2/7/03 ediamond aims construct a federated database
More informationACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
More informationQ: What differentiates a diagnostic from a screening mammography procedure?
The following Q&As address Medicare guidelines on the reporting of breast imaging procedures. Private payer guidelines may vary from Medicare guidelines and from payer to payer; therefore, please be sure
More informationDigital Breast Tomosynthesis
Digital Breast Tomosynthesis White Paper A clinical assessment based on literature by Dr. Finn Lindhardt, senior consultant, Breast Center Regionshospitalet Viborg June 2010 This article summarizes the
More informationScan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
More informationDigital Mammography Update: Design and Characteristics of Current Systems
: Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,
More informationOBJECTIVES By the end of this segment, the community participant will be able to:
Cancer 101: Cancer Diagnosis and Staging Linda U. Krebs, RN, PhD, AOCN, FAAN OCEAN Native Navigators and the Cancer Continuum (NNACC) (NCMHD R24MD002811) Cancer 101: Diagnosis & Staging (Watanabe-Galloway
More informationR/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
More informationfor breast cancer detection
Microwave imaging for breast cancer detection Sara Salvador Summary Epidemiology of breast cancer The early diagnosis: State of the art Microwave imaging: The signal processing algorithm 2D/3D models and
More informationSupports screening, diagnostic, and multimodality workflows
GE Healthcare Universal Viewer Breast Imaging Product Data Sheet Supports screening, diagnostic, and multimodality workflows Introduction Centricity PACS with Universal Viewer 1 puts clinical insight within
More informationHow To Decide If You Should Get A Mammogram
American Medical Women s Association Position Paper on Principals of Breast Cancer Screening Breast cancer affects one woman in eight in the United States and is the most common cancer diagnosed in women
More informationIAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.
IAC Ch 41, p.1 641 41.7 (136C) X-ray machines used for stereotactically guided breast biopsy. 41.7(1) Definitions. In addition to the definitions provided in rules 641 38.2(136C), 641 40.2(136C), and 641
More information2014 Meeting of Mid-Atlantic States Radiation Control Programs March 25, 2014
2014 Meeting of Mid-Atlantic States Radiation Control Programs March 25, 2014 Interpreting Physicians Final Requirements for Interpreting Physicians Meet Initial Qualifications License Certification OR
More informationVI. FREQUENTLY ASKED QUESTIONS CONCERNING BREAST IMAGING AUDITS
ACR BI-RADS ATLAS VI. FREQUENTLY ASKED QUESTIONS CONCERNING BREAST IMAGING AUDITS American College of Radiology 55 ACR BI-RADS ATLAS A. All Breast Imaging Modalities 1. According to the BI-RADS Atlas,
More informationBreast Sonography general goal. Optimizing Breast Sonography. BUS indications -- all. Breast Sonography specific goals.
Optimizing general goal Cindy Rapp BS, RDMS, FAIUM, FSDMS University of Colorado Hospital Denver, Colorado to make a more specific diagnosis than can be made with clinical and mammographic findings alone
More informationDigital Breast Tomosynthesis QC Requirements
Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist
More informationEvaluation of Stylus for Radiographic Image Annotation
Evaluation of Stylus for Radiographic Image Annotation Gautam S. Muralidhar, 1 Gary J. Whitman, 2 Tamara Miner Haygood, 2 Tanya W. Stephens, 2 Alan C. Bovik, 3 and Mia K. Markey 1 We evaluated the use
More informationBreast cancer close to the nipple: Does this carry a higher risk ofaxillary node metastasesupon diagnosis?
Breast cancer close to the nipple: Does this carry a higher risk ofaxillary node metastasesupon diagnosis? Erin I. Lewis, BUSM 2010 Cheri Nguyen, BUSM 2008 Priscilla Slanetz, M.D., MPH Al Ozonoff, Ph.d.
More informationMammography Applications for MAMMOMAT 1000/3000 Nova
Mammography Applications for MAMMOMAT 1000/3000 Nova Introduction This booklet is intended as an application handbook for use with Siemens MAMMOMAT 1000 and 3000 Nova. The booklet contains advice on positioning,
More informationAn Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
More informationNational Medical Policy
National Medical Policy Subject: Breast Tomosynthesis (Digital), 3-D Mammography Policy Number: NMP 526 Effective Date: December 2013 Update: April 2016 This National Medical Policy is subject to the terms
More informationRegistration of X ray mammograms and 3 dimensional speed of sound images of the female breast
Diploma Thesis 2008 Registration of X ray mammograms and 3 dimensional speed of sound images of the female breast by Marie Holzapfel - Student number: 164122 - - Course: TIT05ANS - Company: Forschungszentrum
More informationSegmentation of Breast Tumor from Mammographic Images Using Histogram Peak Slicing Threshold
International Journal of Computer Sciences and Engineering Open Access Research Paper Volume-4, Special Issue-1 E-ISSN: 2347-2693 Segmentation of Breast Tumor from Mammographic Images Using Histogram Peak
More informationInfrared Thermography Not a Useful Breast Cancer Screening Tool
Contact: Jeanne-Marie Phillips Sharon Grutman HealthFlash Marketing The American Society of Breast Surgeons 203-977-3333 877-992-5470 Infrared Thermography Not a Useful Breast Cancer Screening Tool Mammography
More informationBREAST IMAGING. Developed by the Ad Hoc Committee on Resident and Fellow Education of the Society of Breast Imaging
BREAST IMAGING Developed by the Ad Hoc Committee on Resident and Fellow Education of the Society of Breast Imaging Stephen A. Feig, M.D., Chair Ferris Hall, M.D. Debra Ikeda, M.D. Ellen Mendelson, M.D.
More informationThe Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.
Radiologic Technologist Overview The Field - Specialty Areas - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Radiologic technologists
More informationX-ray (Radiography) - Bone
Scan for mobile link. X-ray (Radiography) - Bone Bone x-ray uses a very small dose of ionizing radiation to produce pictures of any bone in the body. It is commonly used to diagnose fractured bones or
More informationPRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,
More informationCT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationREGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
More informationCONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME
CONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME NHSBSP Publication No 60 (Version 2) April 2005 Published by: NHS Cancer Screening Programmes The Manor House 260 Ecclesall Road
More informationUpdate on ACR Digital Mammography QC Manual
Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual
More informationMDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
More informationPreparing for your Breast Tomosynthesis
Preparing for your Breast Tomosynthesis For patients at the Rapid Diagnostic Centre UHN Read this resource to learn: How to prepare What to expect during breast tomosynthesis What questions to ask your
More informationThe Principles of Contrast Mammography
The Principles of Contrast Mammography Andrew Smith, Ph.D. Vice President Imaging Science, Hologic Introduction Multiple clinical studies have demonstrated that the use of screening mammography reduces
More informationTest Sensitivity in the Computer-Aided Detection of Breast Cancer from Clinical Mammographic Screening: a Meta-analysis
Test Sensitivity in the Computer-Aided Detection of Breast Cancer from Clinical Mammographic Screening: a Meta-analysis Corresponding Author: Jacob Levman 1,2, PhD 1. Institute of Biomedical Engineering
More informationBREAST IMAGING SOLUTIONS. Integrated Breast Imaging Workflow Solutions
BREAST IMAGING SOLUTIONS Integrated Breast Imaging Workflow Solutions Seamless Workflow Solutions Customized to Your Needs Hologic understands that a well structured connectivity plan benefits your whole
More informationUpdate on MQSA and ACR Mammography Accreditation
MQSA - Who s Who Update on MQSA and ACR Mammography Accreditation The Law: Mammography Quality Standards Act (MQSA) The Regulator: US Food and Drug Administration (FDA) Pamela L. Platt, BSRT(R)(M)(CV)
More informationThe American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)
The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015) Table of Contents Stereotactic Breast Biopsy Accreditation Program
More informationAmerican College of Radiology Mammography Accreditation Program
American College of Radiology Mammography Accreditation Program Objectives Overview of the ACR Mammography Accreditation Program Interrelationship between mammography accreditation, FDA certification and
More informationINTERDISCIPLINARY CONFERENCE. Florence/Firenze, Italy Nov 27-29, 2012 Centro Congressi al Duomo, Firenze BREAST SEMINAR SERIES
Since breast cancer is not a systemic disease from inception, when the imagers find in situ and 1-14 mm invasive breast cancer, it is the surgeon, specialized in the treatment of breast diseases, who should
More informationA Breakthrough in Breast Specimen Imaging
BreAST HEALTH SOLUTIONS A Breakthrough in Breast Specimen Imaging Trident Specimen Radiography System Superb Image Quality for Streamlined Workflow The Trident system s sharp, high quality images allow
More informationBreast Ultrasound: Benign vs. Malignant Lesions
October 25-November 19, 2004 Breast Ultrasound: Benign vs. Malignant Lesions Jill Steinkeler,, Tufts University School of Medicine IV Breast Anatomy Case Presentation-Patient 1 62 year old woman with a
More informationU.S. Bureau of Labor Statistics. Radiology Tech
From the: U.S. Bureau of Labor Statistics Radiology Tech What They Do Radiologic technologists (RTs) perform diagnostic imaging examinations, such as x rays, on patients. Duties RTs typically do the following:
More informationCertification protocol for breast screening and breast diagnostic services
Certification protocol for breast screening and breast diagnostic services Authors N. Perry R. Holland M. Broeders H. Rijken M. Rosselli del Turco C. de Wolf This is a revised version of the original EUREF
More informationD. FREQUENTLY ASKED QUESTIONS
ACR BI-RADS ATLAS D. FREQUENTLY ASKED QUESTIONS 1. Under MQSA, is it necessary to include a numeric assessment code (i.e., 0, 1, 2, 3, 4, 5, or 6) in addition to the assessment category in all mammography
More informationPreprocessing of Screening Mammograms Based on Local Statistical Models
Preprocessing of Screening Mammograms Based on Local Statistical Models Jiří Grim Institute of Information Theory and Automation Czech Academy of Sciences P.O. BOX 18, 18208 Prague 8 Czech Republic grim@utia.cas.cz
More informationCoronis 5MP Mammo. The standard of care for digital mammography
Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment
More informationThinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
More informationMammography Education, Inc.
Mammography Education, Inc. 2011 LÁSZLÓ TABÁR, M.D.,F.A.C.R (Hon) 3D image of a milk duct MULTIMODALITY DETECTION and DIAGNOSIS of BREAST DISEASES PRAGUE, Czech Republic Crown Plaza, Prague June 29 - July
More informationGreat Strides MARKEY COMPREHENSIVE BREAST CARE CENTER: SETTING A NEW STANDARD OF CARE
Great Strides MARKEY COMPREHENSIVE BREAST CARE CENTER: SETTING A NEW STANDARD OF CARE Markey Cancer Center s Comprehensive Breast Care Center is the first health care provider in Kentucky to offer new
More informationSUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
More informationX-ray (Radiography) - Abdomen
Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,
More informationBreastScreen and You. Information about mammographic screening
BreastScreen and You Information about mammographic screening This information is about screening for breast cancer. It aims to help you choose whether or not you take part in the BreastScreen Australia
More informationFirst floor, Main Hospital North Services provided 24/7 365 days per year
First floor, Main Hospital North Services provided 24/7 365 days per year General Radiology (X-ray) Fluoroscopy Ultrasound (Sonography) Nuclear Medicine P.E.T. imaging Computed Tomography (CT scan) Magnetic
More informationBreast Cancer Screening in Low- and Middle-Income Countries A Framework To Choose Screening Strategies
Breast Cancer Screening in Low- and Middle-Income Countries A Framework To Choose Screening Strategies Richard Wender, MD Session code: www.worldcancercongress.org A Five Step Framework to Guide Screening
More informationRadiologic Technology (RAD, CLE) courses are open only to Radiologic Technology majors.
RADIOLOGIC TECHNOLOGY (A.A.S Degree) Director: Prof. Virginia Mishkin, M.S., R.T. (R) (M) (QM) A radiologic technologist is a skilled professional who provides a specialized health care service. This rewarding
More informationim3d S.p.A. La Ricerca fa Impresa
im3d S.p.A. La Ricerca fa 26 gennaio 2011 Alberto Bert Outline Introduction CAD COLON CAD BREAST DTS Introduction La Ricerca fa Who we are and what we do The company im3d S.p.A. is a company developing
More informationSTUDY PLAN FOR THE CERTIFICATE OF THE HIGHER SPECIALIZATION IN ( Diagnostic Radiology)
STUDY PLAN FOR THE CERTIFICATE OF THE HIGHER SPECIALIZATION IN ( Diagnostic Radiology) plan number :15/11/97/NT I-GENERAL RULES AND CONDITIONS: 1- This plan conforms to the regulations of granting the
More informationThe Possibilities Tomosynthesis Brings to Lung Cancer Screening
Special Report The Possibilities Brings to Lung Cancer Screening Low-dose is a Useful Tool in Lung Cancer Screening A Technical Evaluation of Low-dose with the SONIALVISION safire Joint Industrial-Academic
More informationMulti-slice Helical CT Scanning of the Chest
Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.
More informationOfficial reprint from UpToDate www.uptodate.com 2013 UpToDate
Official reprint from UpToDate www.uptodate.com 2013 UpToDate Patient information: Breast cancer screening (The Basics) Written by the doctors and editors at UpToDate What is breast cancer screening? Breast
More informationFRIEND TO FRIEND CPT CODES 2015 2016. Diagnostic digital breast tomosynthesis, unilateral (list separately in addition to code for primary procedure)
FRIEND TO FRIEND CPT CODES 2015 2016 CPT CODE SERVICE DESCRIPTION FEE EFFECTIVE G0101 Screening pelvic examination $36.69 01 Jan 16 G0202 Mammography, screening, digital, bilateral (2 view film study of
More informationEvaluation and Management of the Breast Mass. Gary Dunnington,, M.D. Department of Surgery Internal Medicine Ambulatory Conference December 4, 2003
Evaluation and Management of the Breast Mass Gary Dunnington,, M.D. Department of Surgery Internal Medicine Ambulatory Conference December 4, 2003 Common Presentations of Breast Disease Breast Mass Abnormal
More informationA Three-Dimensional Correlation Method for Registration of Medical Images in Radiology
A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami
More informationMammograms & Breast Health. An Information Guide for Women U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES. Centers for Disease Control and Prevention
U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention Mammograms & Breast Health An Information Guide for Women This booklet was developed by the Centers for Disease Control
More informationRecommendations for cross-sectional imaging in cancer management, Second edition
www.rcr.ac.uk Recommendations for cross-sectional imaging in cancer management, Second edition Breast cancer Faculty of Clinical Radiology www.rcr.ac.uk Contents Breast cancer 2 Clinical background 2 Who
More informationPresentation Objective: Benefits to New Employee. Goal for Participants. Benefits to the Administration
Imaging Department Quick Step How to Create and Implement a Rapid Integration Orientation New Employee Orientation builds a bridge between the old job and the new job for better quality in health care
More informationNicole Kounalakis, MD
Breast Disease: Diagnosis and Management Nicole Kounalakis, MD Assistant Professor of Surgery Goal of Breast Evaluation The goal of breast evaluation is to classify findings as: normal physiologic variations
More informationOffice Visits. Breast
Early Detection Works Reimbursement Fee Schedule Effective for services on or after July 1, 2015 Program guidelines require that be the payor of last resort. Program funds cannot be used to supplement
More informationPrepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
More informationACR BI-RADS ATLAS MAMMOGRAPHY MAMMOGRAPHY II. REPORTING SYSTEM. American College of Radiology 121
ACR BI-RADS ATLAS II. REPORTING SYSTEM American College of Radiology 121 2013 122 American College of Radiology ACR BI-RADS ATLAS A. REPORT ORGANIZATION (Guidance chapter, see page 147) The reporting system
More informationPurchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage
More informationAutomated Breast Volume Scanning 3D Ultrasound of the Breast
Automated Breast Volume Scanning 3D Ultrasound of the Breast Roel Mus, MD Radboud University Nijmegen Medical Centre (RUNMC), Nijmegen, The Netherlands Matthieu Rutten, MD, PhD Jeroen Bosch Ziekenhuis,
More informationCardiac CT for Calcium Scoring
Scan for mobile link. Cardiac CT for Calcium Scoring Cardiac computed tomography (CT) for Calcium Scoring uses special x-ray equipment to produce pictures of the coronary arteries to determine if they
More informationFirst Warning Systems, Inc.
First Warning Systems, Inc. Breast Cancer Early Detection System US EU/UK & Russia Executive Summary September 2012 i 1. EXECUTIVE SUMMARY 1.1 FIRST WARNING SYSTEM INTRODUCTION First Warning System (FWS)
More information