1 Performance Verification of Esco Pharmacon Downflow Booth via Surrogate Air Monitoring Exposure to Lactose During Material Transfers Involving 25 kg Of Active Pharmaceutical Ingredient By XQ Lin, Yayi Nestiti April 2009 Introduction Downflow booths are used in the pharmaceutical, fine chemical and food industries, for operations such as sampling, grinding, dispensing and filling, which generate airborne particles; when processes involve hazardous, toxic or sensitizing materials and when operators, adjoining areas require protection from exposure to aerosols of the process materials. Surrogate Powder Containment testing on an Esco Pharmacon Downflow Booth (model DFB-1.5S1-10-C, serial number ) was conducted to assess the performance characteristics of these booths in containing and controlling airborne concentrations of particulate powders generated during typical powder handling and transfer operations. Three test iterations were conducted, each involving the transfer of 25 kg of lactose from a bulk product drum into a receiving drum. The tests results prove that the booth delivers excellent containment significantly below industry-accepted criteria. Performance Testing The downflow booth was first tested to ensure it meets filter integrity, airflow and operator comfort performance criteria. HEPA Filter Test This test ensures that the HEPA filters meet filter integrity requirements for allowable penetration of PAO (Poly Alpha Olefin) test aerosol. The upstream face of each filter was challenged with an aerosol generated from Shell Ondina EL oil, producing particles with a mass diameter of less than 0.8 µm. The volume of aerosol produced was sufficient to provide a challenge of between 20 µg/l and 50 µg/l. A forward light scattering photometer with a sampling probe was then used to scan each filter downstream for leaks. Aerosol penetration was % for the all downflow HEPA filters (rear left, rear right, front left, front right) and bleed HEPA filter 1. These measurements were within industry accepted acceptance criteria, that no reading of the photometer shall exceed % of the challenge value.
2 Downflow Test This test ensures that the downflow velocity across the filter faces is of a sufficient velocity and is uniform. Downflow velocity was measured at a distance of 100mm (4 ) below the filter face, with a total of 20 points measured across 4 filters, using the manufacturer s grid. The average velocity for all 20 points was 0.5m/s (98fpm), with a maximum deviation of 16%. This met industry accepted criteria, which specifies the downflow booth s average downflow velocity should be 0.5m/s (98fpm) ± 10%, and that individual readings for each filter should have a maximum deviation of 20% from the filter s average downflow velocity. Bleed Airflow Test This test ensures the bleed airflow volume is within the manufacturer s specifications. Bleed airflow velocity was measured at a distance of 100mm (4 ) from the filter face, using a calibrated anemometer positioned according to the manufacturer s grid. The measured bleed airflow volume was 436 m 3 /h (257 cfm). The bleed airflow volume as a percentage of booth total airflow (4676 m 3 /h, or 2752 cfm) was 9.32%. This met the manufacturer s acceptance criteria, which specified that bleed airflow volume should be 10% ± 2% of the total airflow volume. Light Intensity Test This test ensures that light intensity, at a working height of 1200mm (4 ) from the floor, is sufficient for operator comfort. The average measured light intensity, according to the manufacturer s grid, was 891 lux (83 ft. candles). This met the manufacturer s acceptance criteria, which specified that the average light intensity should be more than 500 lux (46 ft. candles), and that no light level should be less than 40% of the average. Noise Level Test This test ensures that the noise level, measured at 1200mm (4 ) above the floor, and 1000mm (3 3 ) from the front centre, using a calibrated sound meter, is not excessive for operator comfort. The measured noise level was 65.3dBA. This met the manufacturer s acceptance criteria that the noise level should not exceed 68dBA.
3 Surrogate Testing The surrogate testing method is in accordance with the ISPE Good Practice Guideline Assessing the Particulate Containment Performance of Pharmaceutical Equipment. Surrogate Material According to the ISPE Good Practice Guideline Assessing the Particulate Containment Performance of Pharmaceutical Equipment, lactose monohydrate (lactose) is preferred as a surrogate test agent to identify the "range" of potential exposures and containment during a powder handling operation. Although not a crystalline material, lactose is recommended because its particle size and bulk density are comparable to many Active Pharmaceutical Ingredients (APIs). Lactose is also non-toxic, safe to handle and is readily detectable at very low concentrations, thus enabling the assessment of very short duration tasks. The surrogate material used met the requirements specified in the ISPE Good Practice Guide (Lactose-313, NF Monohydrate; Foremost USA cat # W406313). The lactose used had a particle size distribution of 75% (by weight) less than 37 micrometers; 24% between 75 and 37 micrometers; and one percent larger than 75 micrometers. Air Sampling Air sampling locations were set up as illustrated in Figure 1, and consisted of: 1. Operating Breathing Zone (OBZ): the sampler was attached to the operator s shirt s collar and was placed less than 300mm (12 ) from the mouth. 2. Inside booth 200mm (8 ) off left side wall & 200mm (8 ) outside safe work zone. 3. Inside booth, center, 200mm (8 ) outside safe work zone. 4. Inside booth 200mm (8 ) off right side wall & 200mm (8 ) outside safe work zone. 5. Outside booth 200mm (8 ) outside booth in front of left side wall. 6. Outside booth 1500mm (5 ) outside and in the center of the booth. 7. Outside booth 200mm (8 ) in front of right side wall. 8. Background samples inside and outside the booth, midway between the booth side walls, 200mm (8 ) from floor and 2000mm (6 7 ) from open face of booth.
4 Figure 1. Plan View of Sampling Locations
5 Air sampler cassettes to capture lactose (25 mm, 1.0 µm PTFE filter) were connected to the pumps. The pumps were operated at a flow rate of approximately 2.0 liters per minute (0.07cfm). Powder Handling Operation Fiber drums were used in this test and were fitted with two liners. The inner liner holds the lactose and the outer liner provides redundant protection. Two drums (bulk product drum and receiving drum) were moved into the downflow booth. The receiving drum was positioned as close as possible to the exhaust located at the back of the booth. The bulk drum was positioned beside the receiving drum, to minimize the distance required for lactose transfer. The operator de-lidded the drums and staged the liners. Lactose was scooped by hand using a 1000 cc scoop until most of the material was transferred. The liner containing the remaining material was then lifted out of the bulk drum, and the remaining material was directly discharged to the receiving drum. The liner was then placed back into the empty bulk drum. Finally, the receiving drum was tied off and the drums were re-lidded. The powder handling operation was repeated three times to yield a greater number of test samples. At the end of each iteration, the operator remained in the booth for 15 minutes to ensure all dust emitted from the task was collected. The operator wore several layers of impermeable gloves. To prevent cross contamination, a layer of gloves was removed after each iteration and/or after performing a task that results in a high level of dust on the gloves. Used gloves were disposed into a lined waste collection bin, sited at the corner of the booth. In total, 25 air samples were collected for analysis 9 outside booth samples, 9 inside booth samples, 3 operator breathing zone samples, 2 background samples and 2 field blanks (as controls). Samples were analyzed for lactose at an independent laboratory, ESA Laboratories, Inc, USA, using High Performance Liquid Chromatography (HPLC). Analytical data from the laboratory is appended to this report.
6 Summary of Results Date Air Sample # Type of Sample Description Time (min) Iteration 1 (Average Environmental Conditions: Temperature: 29 0 C, Humidity: 73%) Concentration of Lactose (µg/m3) 3/20/ Background Background air sample (inside) /20/ Background Background air sample (outside) 114 < /20/ OBZ Drum to drum transfer of 25 Kg lactose by hand scooping 3/20/ Inside Static sample inside booth: left 40 < /20/ Inside Static sample inside booth: center 40 < /20/ Inside Static sample inside booth: right 40 < /20/ Outside Static sample outside booth: left 40 < /20/ Outside Static sample outside booth: center 40 < /20/ Outside Static sample outside booth: right 40 <0.024 Iteration 2 (Average Environmental Conditions: Temperature: 30 0 C, Humidity: 72%) 3/20/ OBZ Drum to drum transfer of 25 kg lactose by hand scooping 3/20/ Inside Static sample inside booth: left 36 < /20/ Inside Static sample inside booth: center 36 < /20/ Inside Static sample inside booth: right 36 < /20/ Outside Static sample outside booth: left 36 < /20/ Outside Static sample outside booth: center 36 < /20/ Outside Static sample outside booth: right 36 <0.027 Iteration 3 (Average Environmental Conditions: Temperature: 31 0 C, Humidity: 70%) 3/20/ OBZ Drum to drum transfer of 25 kg lactose by hand scooping 3/20/ Inside Static sample inside booth: left 38 < /20/ Inside Static sample inside booth: center 38 < /20/ Inside Static sample inside booth: right 38 < /20/ Outside Static sample outside booth: left 38 < /20/ Outside Static sample outside booth: center 38 < /20/ Outside Static sample outside booth: right 38 < /20/2009 Field blank #1 Control sample NA < /20/2009 Field blank #2 Control sample NA <0.002 This test assesses the performance characteristics of the Esco Pharmacon Downflow Booth in containing and controlling airborne concentrations of particulate powder generated during a typical powder handling and transfer operation. The powder handling operation was transferring 25kg of lactose from a bulk product drum to a receiving drum in three iterations. Drum to drum transfers by hand scooping or direct discharge from a drum liner are common tasks performed at
7 a majority of pharmaceutical facilities. Being an open and manual process, reliant on operator technique (instead of manufacturer-specific equipment), it challenges the downflow booth to a reasonable level and provides a suitable parallel for real world tasks. Summary of Exposure Outcomes Iteration Background Duration 40 minutes 36 minutes 38 minutes 114 minutes Operator Breathing Zone 6.35 µg/m 3 36 µg/m µg/m 3 NA Inside Range µg/m µg/m µg/m µg/m 3 Inside Mean µg/m µg/m µg/m 3 Outside Range µg/m µg/m µg/m µg/m 3 Outside Mean µg/m µg/m µg/m 3 Exposure values inside the booth and in the operator breathing zone were within industry accepted criteria (<100 µg/m 3 ). These values prove that the booth provides good containment. In particular, the operator breathing zone values were significantly below industry accepted criteria. Exposure values outside the booth were relatively low indicating the booth provided good containment, and that samples inside the booth did not falsely register external contamination.
8 Conclusion A comprehensive battery of tests including filter integrity, downflow, bleed airflow, light intensity and noise level was carried out to verify the performance of the Esco Pharmacon Downflow booth. Booth performance was established to be within industry accepted and manufacturer s criteria. A surrogate powder test involving the hand transfer of 25kg of lactose, in accordance with the ISPE Good Practice Guidelines, was then carried out. Multiple air sampling positions were strategically located at the operator breathing zone, inside, and outside the booth. Laboratory analysis of the sampling discs recovered indicated that booth performance was excellent, with all sampling locations, including the critical operator breathing zone, reading less than 100 µg/m 3. The result is specified for lactose. The downflow booth s containment with other compounds can only be known by testing with those compounds. Real world results may differ depending on the material handled and process. References Floura, Hari & Kremer, John. Performance Verification of a Downflow Booth via Surrogate Testing. The Official Magazine of ISPE, November/December Vol 28 No. 6 Floura, Hari Performance Verification of a Downflow Booth via Surrogate Air Monitoring Guideline Document. Floura LLC Containment Consultants. ISPE Good Practice Guide: Assessing the Particulate Containment Performance of Pharmaceutical Equipment, International Society for Pharmaceutical Engineering (ISPE), First Edition, January 2005,
9 Date Results of lactose surrogate air sampling downflow booth Type of Sample Activity Time Pre- Post- Sample Air flow Flow Flow # (min) rate rate Rate Sample amount of lactose airbone concentration volume (L) (ng/cassette) of lactose (µg/m3) Iteration #1 3/20/ Background background area air sample (inside) /20/ Background background area air sample (outside) < /20/ OBZ drum to drum transfer of 25 Kg lactose by hand scooping (iteration #1) 3/20/ Area (in) Static area air sample inside booth: LEFT side (iteration #1) < /20/ Area (in) Static area air sample inside booth: CENTER (iteration #1) < /20/ Area (in) Static area air sample inside booth: RIGHT side (iteration #1) < /20/ Area (out) Static area air sample outside booth: LEFT side (iteration #1) < /20/ Area (out) Static area air sample outside booth: CENTER (iteration #1) < /20/ Area (out) Static area air sample outside booth: RIGHT side (iteration #1) < /20/2009 Field Blank #1 N/A N/A < Iteration #2 3/20/ OBZ Drum to drum transfer of 25 kg lactose by hand scooping (iteration #2) 3/20/ Area (in) Static area air sample inside booth: LEFT side (iteration #2) < /20/ Area (in) Static area air sample inside booth: CENTER (iteration #2) < /20/ Area (in) Static area air sample inside booth: RIGHT side (iteration #2) < /20/ Area (out) Static area air sample outside booth: LEFT side (iteration #2) < /20/ Area (out) Static area air sample outside booth: CENTER (iteration #2) < /20/ Area (out) Static area air sample outside booth: RIGHT side (iteration #2) < /20/2009 Field Blank #1 N/A N/A < Iteration #3 3/20/ OBZ Drum to drum transfer of 25 kg lactose by hand scooping (iteration #3) 3/20/ Area (in) Static area air sample inside booth: LEFT side (iteration #3) < /20/ Area (in) Static area air sample inside booth: CENTER (iteration #3) < /20/ Area (in) Static area air sample inside booth: RIGHT side (iteration #3) < /20/ Area (out) Static area air sample outside booth: LEFT side (iteration #3) < /20/ Area (out) Static area air sample outside booth: CENTER (iteration #3) < /20/ Area (out) Static area air sample outside booth: RIGHT side (iteration #3) < /20/2009 Field Blank #1 N/A N/A <
10 Representative Photos Transferring of Lactose ESCO Pharmacon Downflow Booth Setup Operator de-lidded the drum and staged the liners of both drums Operator was scooping the lactose from bulk product drum (left picture) to receiving drum (right picture)
11 The inner liner was taken out from the bulk product drum (left picture) to be discharged directly to receiving drum The liners was tied off for the receiving drum (left picture) and both of drums were re-lidded
12 The sampler cassette on operator breathing zone (left picture) and position of overall samplers (right picture)
13 ESA LABORATORIES, INC. Page 1 22 ALPHA ROAD CHELMSFORD, MA (978) FAX: (978) PO#/RELEASE: YAYI NESTITI/NA ESAL BATCH#: CLIENT JOB#: NA DATE RECEIVED : DATE ANALYZED : YAYI NESTITI DATE REPORTED : ESCO MICRO PTE LTD 21 CHANGI SOUTH STREET 1 SINGAPORE, SINGAPORE Revised Test Report Paul A. Ullucci Laboratory Director TEST CODE TEST DESCRIPTION REFERENCE ANALYTICAL METHOD REPORTING LIMIT LACTOSE LACTOSE ESA L40102REVC 2.0NG LACTOSE-M LACTOSE IN MASS ESA SOP L40102REVC 2.0NG SAMP DATE LACTOSE LACTOSE-M NO. CLLCT SAMPLE ID/OTHER UG/M3 NG MESSAGES < < <0.024 < <0.025 < <0.022 < <0.026 < <0.024 < <0.024 < <0.027 <2.0
14 ESA LABORATORIES, INC. Page 2 22 ALPHA ROAD CHELMSFORD, MA (978) FAX: (978) PO#/RELEASE: YAYI NESTITI/NA ESAL BATCH#: CLIENT JOB#: NA DATE RECEIVED : DATE ANALYZED : YAYI NESTITI DATE REPORTED : ESCO MICRO PTE LTD 21 CHANGI SOUTH STREET 1 SINGAPORE, SINGAPORE Revised Test Report TEST CODE TEST DESCRIPTION REFERENCE ANALYTICAL METHOD REPORTING LIMIT LACTOSE LACTOSE ESA L40102REVC 2.0NG LACTOSE-M LACTOSE IN MASS ESA SOP L40102REVC 2.0NG SAMP DATE LACTOSE LACTOSE-M NO. CLLCT SAMPLE ID/OTHER UG/M3 NG MESSAGES <0.028 < <0.025 < <0.029 TR< <0.027 < <0.027 < <0.025 < <0.026 < <0.024 < <0.027 <2.0
15 ESA LABORATORIES, INC. Page 3 22 ALPHA ROAD CHELMSFORD, MA (978) FAX: (978) PO#/RELEASE: YAYI NESTITI/NA ESAL BATCH#: CLIENT JOB#: NA DATE RECEIVED : DATE ANALYZED : YAYI NESTITI DATE REPORTED : ESCO MICRO PTE LTD 21 CHANGI SOUTH STREET 1 SINGAPORE, SINGAPORE Revised Test Report TEST CODE TEST DESCRIPTION REFERENCE ANALYTICAL METHOD REPORTING LIMIT LACTOSE LACTOSE ESA L40102REVC 2.0NG LACTOSE-M LACTOSE IN MASS ESA SOP L40102REVC 2.0NG SAMP DATE LACTOSE LACTOSE-M NO. CLLCT SAMPLE ID/OTHER UG/M3 NG MESSAGES <0.026 < <0.025 < BLANK <2.0NG < BLANK <2.0NG <2.0 RESULTS ARE BLANK CORRECTED ON ESAL'S MATRIX BLANK END OF REPORT
Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested
DOP Solutions Technical Document: Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested Issue 1: Prepared by Neil Stephenson and Tim Triggs and edited by
More informationAir and Water Balancing - Sound and Vibration Testing - Cleanroom Services - Building Commissioning Building Envelope Testing Thermal Imaging
NEUDORFER ENGINEERS, INC. provides our customer with unbiased, professional, performance testing of Cleanrooms. We are committed to providing you, our customer, with value-added expertise in an effort
More informationEnvironmental Monitoring
Chapter 10 Environmental Monitoring By the end of this chapter, you will be able to: l List the types of environmental testing l Name three types of checks that are carried out to ensure the isolator is
More informationEACO DOP / PAO TESTING GUIDELINE 2013
2013 Foreword The guideline promotes industry standards for training, equipment, and methodologies for HEPA filter integrity testing of negative air units, vacuums and any other HEPA filtered equipment
More informationCETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010
CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010 1. Background Class II Type B Biosafety Cabinets (BSCs) are unique laboratory
More informationENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS
ENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS PRESENTATION TO RACI 20 TH APRIL 2016 PAUL MORGAN DIRECTOR CRITICAL CERTIFICATION PTY LTD CHANGES TO ISO14644 PART 1 CHANGES
More informationA Report of Air Sampling for Mercury Vapor and Particulates During Fluorescent Tubes Disposal Operation Maywood, Illinois Survey date: 9/28/2000
A Report of Air Sampling for Mercury Vapor and Particulates During Fluorescent Tubes Disposal Operation Maywood, Illinois Survey date: 9/28/2000 Prepared for: Air Cycle Corporation 2000 South 25 th Avenue,
More informationWIBObarrier OCS - Open Containment Systems WIBObarrier CCS - Closed Containment Systems
WIBOjekt and WIBObarrier Air Curtain Systems. The perfect protection for person, product and environment. Filling Weighing Sampling Research Safe handling with high potent products / API WIBObarrier OCS
More informationEnvironmental Monitoring of Clean Rooms
Environmental Monitoring of Raul Duarte President DDK Scientific, Corp. Copyright DDK Scientific, Corp. 2008, 2009, DDK Scientific, Corp. Proprietary A manufacturing facility for pharmaceutical products
More informationAsbestos in the Workplace: A Guide to Removal of Friable Asbestos Containing Material
Asbestos in the Workplace: A Guide to Removal of Friable Asbestos Containing Material Revised November 21, 2013 Application Code of Practice Where asbestos is present or believed to be present in a workplace
More informationHEPA FILTER DEFECTS ARE COMMON
HEPA FILTER DEFECTS ARE COMMON Kenneth F. Cherry 1 and Kenneth A. Smigielski 2 1 NCEC Inc. 2 First Solar LLC. ABSTRACT HEPA filters (High Efficiency Particulate Air) filters are claimed by vendors to have
More informationCLEAN ROOM MONITORING REGULATORY STANDARDS
CLEAN ROOM MONITORING REGULATORY STANDARDS Air Classification as per Schedule M Maximum permitted number of particles / m3 equal or above at rest in operation 0.5µm 5.0µm 0.5µm 5.0µm A 3,520 29 3,500 29
More informationTHE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS
THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS INDOOR AIR QUALITY INSTRUMENTS UNDERSTANDING, ACCELERATED BREATHE A LITTLE EASIER WITH TSI Indoor air quality is a growing concern. With the increasing
More informationWhitepaper Cross-Contamination
Whitepaper Cross-Contamination Protection in HVAC I N D U S T R Y I N S I G H T S Norman A. Goldschmidt September 30, 2011 Principal, Engineering www.geieng.com I N D U S T R Y I N S I G H T S Genesis
More informationThe Question of Homogeneity inside a Chimney: Application of ISO 13528 to Stack Emission Proficiency Tests
Hessian Agency for the Environment and Geology Department I3 Air Pollution, Emission Jens Cordes Benno Stoffels Dominik Wildanger The Question of Homogeneity inside a Chimney: Application of ISO 13528
More informationClass II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health
Background Class II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health This document provides guidelines for the design and installation of Class
More informationSENTRY AIR SYSTEMS, INC. PHARMACEUTICAL COMPOUNDING CONTAINMENT SYSTEMS SIMPLE SOLUTIONS FOR CLEANER AIR. www.sentryair.com 1.800.799.
SENTRY AIR SYSTEMS, INC. PHARMACEUTICAL COMPOUNDING CONTAINMENT SYSTEMS SIMPLE SOLUTIONS FOR CLEANER AIR 1.800.799.4609 www.sentryair.com Pharmaceutical Compounding Safety Solutions Sentry Air Systems
More informationNDSU TABLE OF CONTENTS
NDSU TABLE OF CONTENTS 1.0 INTRODUCTION 2.0 DEFINITIONS AND DESCRIPTIONS 3.0 OPERATIONS AND MAINTENANCE PROCEDURES FOR ACM 4.0 FIBER RELEASE EPISODES 5.0 DEMOLITION INFORMATION 6.0 REMARKS 1.0 INTRODUCTION
More informationCORESTA RECOMMENDED METHOD Nº 81
CORESTA RECOMMENDED METHOD Nº 81 ROUTINE ANALYTICAL MACHINE FOR E-CIGARETTE AEROSOL GENERATION AND COLLECTION DEFINITIONS AND STANDARD CONDITIONS (June 2015) 0. INTRODUCTION This Method includes the requirements
More informationContainment Solutions for Potent Materials
Containment Solutions for Potent Materials By Gary Partington Walker Barrier Systems 608-562-7761 gpartington@walkerbarrier.com www.walkerbarrier.com Why do I need containment? Legal NIOSH (National Institute
More informationThermoSorb Desiccant Dryers
ThermoSorb Desiccant Dryers Sierra Why dry compressed air? Contamination reduces efficiency The air we breathe contains contamination in the form of water vapour and airborne particles. During the compression
More informationAir Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection
Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection A wide range of products to meet the needs of transferring animals, operators,
More informationCONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS
CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS UNDERSTANDING, ACCELERATED TSI SETS THE STANDARD FOR PARTICLE COUNTING Performance Semiconductor, hard disk drive, and flat panel manufacturers
More informationFume Hood Questions and Answers
Fume Hood Questions and Answers 1. When should I use a fume hood? Use a chemical fume hood anytime your work involves: Chemicals with a National Fire Protection Association (NFPA) Health rating of 3 or
More informationContamination Control and Environmental Monitoring Systems
Life Science Contamination Control and Environmental Monitoring Systems Hospital Electronics Brookhaven Instruments deliver, install and validate particle counters, Microbial Air Samplers and Monitoring
More informationMETHOD OF TEST FOR UNIT WEIGHT OF FRESH CONCRETE
STATE OF CALIFORNIA BUSINESS, TRANSPORTATION AND HOUSING AGENCY DEPARTMENT OF TRANSPORTATION DIVISION OF ENGINEERING SERVICES Transportation Laboratory 5900 Folsom Blvd. Sacramento, California 95819-4612
More informationCelCulture. CelCulture CO2 Incubator Model CCL-170_-_. CO2 Incubator with Integrated Cooling System Solution for Highly Specialized Applications
1 CO2 Incubator Model CCL-170_-_ CO2 Incubator with Integrated Cooling System Solution for Highly Specialized Applications Welcome to Esco Esco s Vision is to provide enabling technologies for scientific
More informationStainless Steel Worksurface 3 Feet Hood A10-WS-PR3 A10-WS-SR3 4 Feet Hood A10-WS-PR4 A10-WS-SR4 6 Feet Hood A10-WS-PR6 A10-WS-SR6.
Product Sampler #1 Laminar Flow Hoods / Laminar Flow Cabinets: 1- Benchtop Vertical Laminar Flow Hoods- Air Recirculation Available s: Clean Bench Class 100 Bench-top Vertical Laminar Flow Hoods with Air-recirculation
More informationFume Extraction: Guide to Safely Managing Solder Fumes in the Workplace
Fume Extraction: Guide to Safely Managing Solder Fumes in the Workplace Our thanks to OK International for allowing us to reprint the following. Why do we need fume extraction? Today s Electronics workplace
More informationABSTRACT INTRODUCTION
Double Door Transfer Ports (DPTE ) Use to Maximize ALARA in Contained Transfer Operations P.A. Westover, SRNL -Presenting; B.M Allen, SRNL; S.D. Chunglo, la Calhène Inc., Minnesota ABSTRACT Nuclear material
More informationLaminar Flow Cabinets
Laminar Flow Cabinets The World s Most Practical Selection of Benchtop Laminar Flow Cabinets. 24 36 48 Simple, Effective Protection for Samples and Work Processes, Uncompromised Performance, Economical
More informationHigh Efficiency Particulate Air (HEPA) Filter Velocity Reduction Study
High Efficiency Particulate Air (HEPA) Filter Velocity Reduction Study International SEMATECH Manufacturing Initiative Advanced Materials Research Center, AMRC, International SEMATECH Manufacturing Initiative,
More informationInstallation and Operating Manual Model F1800C Model F1800PG Model F1802C Model F1802PG Model F1806C Model F1806PG
Installation and Operating Manual Model F1800C Model F1800PG Model F180C Model F180PG Model F1806C Model F1806PG www.ipsystemsusa.com Contents Safety Warnings... Items supplied... Description... Installation...4
More informationPrinciples of Ventilation. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike License. Your use of this material constitutes acceptance of that license and the conditions of use of materials on this
More informationAPC APPLICATION NOTE #92
#92 Best Practices for Designing Data Centers with the InfraStruXure InRow RC By John Niemann Abstract The InfraStruXure InRow RC is designed to provide cooling at the row and rack level of a data center
More informationChapter 12 Industrial Ventilation
Chapter 1 Industrial Ventilation Introduction Industrial Ventilation is an important method for reducing employee exposures to airborne contaminants. Ventilation is used to reduce or remove contaminates
More informationDRAFT 8/13/2015. Proposed Addition of Non-Mandatory Appendix FC-B has been withdrawn from publication in the next edition.
AG-1-201X Revision of ASME AG-1-2012 Code on Nuclear Air and Gas Treatment DRAFT 8/13/2015 Appendix FC-B has been withdrawn from publication in the next edition. Proposed revisions to Section FC-B have
More informationHEPA Filters and Filter Testing A Comparison of Factory Tests and In-Service Tests Bulletin: PB2007-0504
HEPA Filters and Filter Testing A Comparison of Factory Tests and In-Service Tests Bulletin: PB2007-0504 Introduction HEPA filters, once known as absolute filters, were originally developed as the particulate
More informationExperimental Evaluation Of The Frost Formation
Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 2014 Experimental Evaluation Of The Frost Formation Yusuke Tashiro Mitsubishi
More informationCLASSIFICATION OF CLEANROOMS
CLASSIFICATION OF CLEANROOMS Cleanrooms are classified by the cleanliness of their air. The method most easily understood and most universally applied is the one suggested in the earlier versions (A to
More informationUnderstanding USP 797
DDK Scientific, Corp. Raul Duarte June 2008 Copyright DDK Scientific, Corp. 2008, 2009 DDK Scientific, Corp. Proprietary The purpose of USP 797 is to prevent harm and fatality to patients that could result
More information191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES
191: Calibration and Standards in Measurement Richard Paton National Engineering Laboratory, Scotland, UK 1 General Principles 1 2 Gravimetric Calibration of Liquid meters 2 3 Volumetric Calibration of
More informationAPPENDIX A-1 SAMPLE CLEANROOM SPECIFICATIONS MICROELECTRONICS & SEMICONDUCTORS
APPENDIX A-1 SAMPLE CLEANROOM SPECIFICATIONS SECTION xxxxx CLEANROOM PERFORMANCE TESTING PART 1 -- GENERAL 1.1 WORK INCLUDED A. This Section specifies the requirements of the NEBB Certified Cleanroom Performance
More informationVALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING. F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group
VALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group tel + 3226431600 E-Mail: francis.smal@be.snclavalin.com VALIDATION OF CLEAN
More informationSPILLS & SPILL KITS. Spills -General Guidelines:
Spills -General Guidelines: If handled properly, a spill may be nothing more than a nuisance. If handled improperly, a spill can seriously disrupt your activities and the work of your colleagues. At worst,
More informationHazardous Drug. Cleanroom. Standards
Hazardous Drug Cleanroom Standards Module 3 Table of Contents Section A... 1 A.1 International Standards Organization (ISO) Classifications... 1 A.2 Hazardous Drug Cleanroom and Anteroom... 1 A.2.1 Location
More informationWaste Oil Handling, Spill Containment & Lighting
58 Waste Oil Handling, Spill Containment & Lighting 372000 SAMOA Mobile Waste Oil Drainer & Extractor Drains and/or extracts waste oil from cars and light commercial vehicles Drain bowl capacity of 20
More informationSPECIAL APPLICATIONS
Page 1 SPECIAL APPLICATIONS The protection of microelectronics clean rooms using aspirating smoke detection systems requires special consideration. The physical configuration of the room(s) and the direction
More informationBenjamin Fontes, Geoffrey Lyon and Maryjo Lanzillotta Yale University
Benjamin Fontes, Geoffrey Lyon and Maryjo Lanzillotta Yale University Yale Cell Sorting Biosafety Cross Training Learning phase Program Collaboration Setting up a program Partnership Maintaining the program
More informationCleaning. By the end of this chapter, you will be able to: Introduction. Definitions. Chapter 9
Chapter 9 By the end of this chapter, you will be able to: l Define the terms cleaning, decontamination, disinfection l List the factors that affect disinfection l Describe the aims of a cleaning schedule
More informationBackground on QRM. Pharma EXPO 28 30 September 2015 Las Vegas, NV. Your QRM Program. Paul H. Ahlijanian
Using Quality Risk Management to Evaluate Compliance for Multiproduct and Inventory Facilities in the Pharmaceutical Industry Paul H. Ahlijanian Senior Manager, Pharmaceutical Sciences Quality Assurance
More informationAccurate Air Flow Measurement in Electronics Cooling
Accurate Air Flow Measurement in Electronics Cooling Joachim Preiss, Raouf Ismail Cambridge AccuSense, Inc. E-mail: info@degreec.com Air is the most commonly used medium to remove heat from electronics
More informationOPTICAL PARTICLE SIZER MASS CALIBRATION METHOD
OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO 12501-1/4) at TSI. This method
More informationDuring Various Aluminum Fabricating Operations
Guidelines for Handling Aluminum Fines Generated During Various Aluminum Fabricating Operations F-1 Guidelines for Handling Aluminum Fines Generated During Various Aluminum Fabricating Operations SCOPE
More informationCHAPTER 13 LAND DISPOSAL
CHAPTER 13 LAND DISPOSAL Supplemental Questions: Which of Shakespeare's plays is the source of the opening quote? The Tempest [1611-1612],Act: I, Scene: i, Line: 70. 13-1. Cite four reasons landfills remain
More informationNatural Air Drying Corn Paul E. Sumner University of Georgia Biological and Agricultural Engineering Department
Natural Air Drying Corn Paul E. Sumner University of Georgia Biological and Agricultural Engineering Department Many small farmers prefer to dry crops using unheated or natural air drying in bins in layers
More informationHEPA Filters for Pharma Applications Cleanrooms & Safety Applications. Gordon Farquharson June 2012
HEPA Filters for Pharma Applications Cleanrooms & Safety Applications Gordon Farquharson June 2012 1 Air Filters at work Images courtesy of CamfilFarr 2 How filters work Sieving Effect Straining 3 Filter
More informationNewson Gale. Is it ever Safe to use Plastic Drums and Containers in Hazardous areas? White Paper. Author Details:
White Paper Is it ever Safe to use Plastic Drums Author Details: Mike O Brien, Head of Product Marketing for If you have any questions relating to the topics discussed in this article, you can contact
More informationoil liquid water water liquid Answer, Key Homework 2 David McIntyre 1
Answer, Key Homework 2 David McIntyre 1 This print-out should have 14 questions, check that it is complete. Multiple-choice questions may continue on the next column or page: find all choices before making
More informationDRYER VENTING SPECIFICATIONS DRYER SAFETY
DRYER VENTING SPECIFICATIONS Table of Contents DRYER SAFETY...1 INSTALLATION REQUIREMENTS...4 Venting Requirements...5 DRYER INSPECTION AND CLEANING...7 Frequency of Exhaust System Cleaning...7 Inspecting
More informationTesting HEPA/ULPA filters at the manufacturer s facility and in situ on the user s premises
Publication Series Viledon Testing HEPA/ULPA filters at the manufacturer s facility and in situ on the user s premises Dipl.-Ing. Thomas Schroth Dr.-Ing. Thomas Caesar Lecture at 4th CLEANROOMS Europe
More informationCOMPOSITION / INFORMATION ON INGREDIENTS
MATERIAL SAFETY DATA SHEET DAMTITE WATERPROOFER- WHITE 01071, 01211, 01451 DAMTITE POWDER FOUNDATION WATERPROOFER GRAY 02451 SECTION I. COMPANY & PRODUCT INFORMATION Product Identification Manufacturer
More informationSIBATA Product lineup ASBESTOS SAMPLING
SIBATA Product lineup ASBESTOS SAMPLING Procedures for measurement of suspended asbestos Purposes and methods of Asbestos measurement for air environment in Japan Item Industrial Hygiene measurement Building
More informationTHE ENVIRONMENTAL TECHNOLOGY VERIFICATION PROGRAM
THE ENVIRONMENTAL TECHNOLOGY VERIFICATION PROGRAM ETV JOINT VERIFICATION STATEMENT TECHNOLOGY TYPE: APPLICATION: TECHNOLOGY NAME: COMPANY: BAGHOUSE FILTRATION PRODUCTS CONTROL OF PM 2.5 EMISSIONS BY BAGHOUSE
More informationLAMINAR FLOW CABINETS
LAMINAR FLOW CABINETS Vertical Laminar Flow Cabinets VLF 550 VLF650 VLF1000 VLF1200 VLF 1500 VLF1800 Monmouth Scientific's brand new range of Vertical laminar Flow (VLF) cabinets is available in six different
More informationETV Joint Verification Statement
THE ENVIRONMENTAL TECHNOLOGY VERIFICATION PROGRAM U.S. Environmental Protection Agency ETV Joint Verification Statement TECHNOLOGY TYPE: APPLICATION: TECHNOLOGY NAME: COMPANY: BAGHOUSE FILTRATION PRODUCTS
More informationGenStore Archive Cabinet
GenStore Archive Cabinet Setup and Maintenance Manual Version A April 2014 Two GenStore Archives shown mounted on optional table stand. 2 Contents Getting Started... 4 Using this Manual...4 Safety Precautions...4
More informationLABORATORY DETERMINATION OF CALIFORNIA BEARING RATIO
LABORATORY DETERMINATION OF CALIFORNIA BEARING RATIO STANDARD IS: 2720 (Part 16) 1979. DEFINITION California bearing ratio is the ratio of force per unit area required to penetrate in to a soil mass with
More informationDemand Based Static Pressure Reset Control for Laboratories
LLC Demand Based Static Pressure Reset Control for Laboratories Laboratories are generally the largest users of energy in any facility. Whether in a pharmaceutical research facility, a university or a
More informationJANUARY 2013 PREPARATION OF A SITE MASTER FILE FOR A MANUFACTURER OF COSMETIC PRODUCTS
JANUARY 2013 PREPARATION OF A SITE MASTER FILE FOR A MANUFACTURER OF COSMETIC PRODUCTS 1 WHAT IS A SITE MASTER FILE? A Site Master File (SMF) is a document prepared by the manufacturer containing specific
More informationApproval Block. Prepared by: Signature Date. Evan Parnell 08 NOV 2013. Reviewed by: Signature Date. Approved by: Signature Date
ATS-ATP-3460 Page: 1 of 5 Approval Block Prepared by: Signature Date Evan Parnell 08 NOV 2013 Reviewed by: Signature Date Brian Flynn 08 NOV 2013 Approved by: Signature Date Kristal Jewell 08 NOV 2013
More informationOutotec HIGmills; A Fine Grinding Technology
23 rd International Mining Congress & Exhibition of Turkey 16-19 April 2013 ANTALYA Outotec HIGmills; A Fine Grinding Technology H. Lehto Outotec Finland Oy A. Paz Outotec Pty Ltd, Perth, Australia I.
More informationCHAPTER 4 VENTILATION
CHAPTER 4 VENTILATION SECTION 401 GENERAL 401.1 Scope. This chapter, in conjunction with the building code, shall govern the ventilation of spaces within a building intended to be occupied. Mechanical
More informationBIOLOGICAL SAFETY CABINET PROGRAM
BIOLOGICAL SAFETY CABINET PROGRAM TABLE OF CONTENTS: 1. INTRODUCTION 2. PURPOSE AND SCOPE 3. SELECTION AND INSTALLATION 4. USE OF BIOLOGICAL SAFETY CABINETS 5. CERTIFICATION OF BIOLOGICAL SAFETY CABINETS
More informationCartridge Change Schedule
3 Cartridge Change Schedule How to establish a Cartridge Change Schedule Exposure Assessment for Respirator Selection and Chemical Cartridge Changeout Schedule An exposure assessment is a process used
More informationPROCEDURAL STANDARDS FOR CERTIFIED TESTING OF CLEANROOMS
PROCEDURAL STANDARDS FOR CERTIFIED TESTING OF CLEANROOMS 2009 THIRD EDITION National Environmental Balancing Bureau 8575 Grovemont Circle Gaithersburg, Maryland 20877 301-977-3698 301-977-9589 FAX www.nebb.org
More informationThermoStar Refrigeration Air Dryers
ThermoStar Refrigeration Air Dryers Why dry compressed air? Contamination Reduces Efficiency The air we breathe contains contamination in the form of water vapour and airborne particles. These problems
More informationcleanroom validation Cleanroom Validation
Cleanroom Validation cleanroom applications laboratories production environments laminar flow boxes cleanroom concepts microelectronics pharmaceutical industry, food industry nuclear industry, aerospace
More informationG U I D E L CERTIFICATION EDUCATION TRAINING. Procurement Guideline. National Air Filtration Association Recommended Practice
G U National Air Filtration Association Recommended Practice I D E L I N E Procurement Guideline TRAINING EDUCATION CERTIFICATION HVAC Air Filter Purchasing Specification Guideline 1.0 Purpose To provide
More informationIngredient % content Hazard data Cross linked Sodium Polyacrylite 70-90 CAS No:9003-04-7 Non chemical fillers 20-30 Zeolite 1-10
Body Spill Granules Composition Ingredient % content Hazard data Cross linked Sodium Polyacrylite 70-90 CAS No:9003-04-7 Non chemical fillers 20-30 Zeolite 1-10 Hazards Identification Eye contact: May
More informationRegulations Concerning
Regulations Concerning Airborne Particle Counting Two types of activities performed with an airborne particle counter: Classification focuses primarily on the environment Particle Counting Applications
More informationPANDA SPRAY BOOTH FILTER INFORMATION
PANDA SPRAY BOOTH FILTER INFORMATION Our Spray Booth has been designed to reduce the risk of potential hazardous situations occurring. You must adhere to the specifications chart listed below when replacing
More informationDuPont Personal Protection TECHNICAL BULLETIN NUCLEAR RISKS
DuPont Personal Protection TECHNICAL BULLETIN NUCLEAR RISKS Specific questions have arisen related to the recent disaster which has occurred in Japan. A question frequently asked relates to protection
More informationMechanical Systems Competency 1.20
Competency 1.20 Mechanical systems personnel shall demonstrate a working level knowledge of the safety and health fundamentals of mechanical systems and/or components. 1. Supporting Knowledge and Skills
More informationDOE-HDBK-1122-99. Module 2.18 Air Sampling Equipment. Instructor s Guide
Course Title: Radiological Control Technician Module Title: Air Sampling Equipment Module Number: 2.18 Objectives: 2.18.01 Identify the factors that affect the operator's selection of a portable air sampler.
More informationHead Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids
Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction Last lab you investigated flow loss in a pipe due to the roughness
More informationOperations Unit Custodial Services Staff and Managers/Supervisors who administer the policy.
Facilities Management Policy (FMP) Dalhousie Green Cleaning Area: Operations Title: Green Cleaning Policy Issued by: Director of Operations, Facilities Management Date Issued: 2013-03- 11 Last Revision
More informationIN-DEPTH SURVEY REPORT OF A LOCAL EXHAUST VENTILATION DEVICE FOR SUPPRESSING RESPIRABLE AND CRYSTALLINE SILICA DUST FROM POWERED SAWS
This Survey Report and any recommendations made herein are for the specific facility evaluated and may not be universally applicable. Any recommendations made are not to be considered as final statements
More informationMAC 10. Original & IQ FAMILY OF FAN FILTER UNITS
MAC 10 Original & IQ FAMILY OF FAN FILTER UNITS used in: microelectronics medical device semiconductor disk drive laboratory pharmaceutical markets hospitals clean rooms Table of Contents MAC 10 Original
More informationUSER GUIDE. LED Light Meter. Model LT45
USER GUIDE LED Light Meter Model LT45 Introduction Congratulations on your purchase of the Extech LT45 LED Light Meter that measures light from LED lamps in addition to fluorescent, metal halide, high
More information7. Chemical Waste Disposal Procedures
7. Chemical Waste Disposal Procedures Substances that Can and Cannot be Disposed of via Sink Drains or in Regular Solid Refuse Containers 1. It is prohibited to pour down a sink or floor drain, or place
More informationISO-9001:2000 certified. EE300 Series Electromechanical Inventory Level Measuring System. Aplus Finetek Sensor, Inc.
ISO-9001:2000 certified EE300 Series Electromechanical Inventory Level Measuring System Aplus Finetek Sensor, Inc. PRODUCT INTRODUCTION OPERATING PRINCIPLE The EE300 series level measuring system is a
More informationA.17. OXIDIZING PROPERTIES (SOLIDS)
A.17. OXIDIZING PROPERTIES (SOLIDS) 1. METHOD 1.1. INTRODUCTION It is useful to have preliminary information on any potentially explosive properties of the substance before performing this test. This test
More informationPI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3)
PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3) Yes No 1. Is access to the laboratory limited or restricted at the discretion of the laboratory director when experiments are in progress? Yes No
More informationReceptor Height: 1.5 m in Dispersion Modeling
Receptor Height: 1.5 m in Dispersion Modeling J.C. Laul SB-EWM, Los Alamos National Laboratory P.O. Box 1663, Los Alamos, NM 875057 jclaul@lanl.gov, 505-665-9791 LA-UR-12-01402 Abstract EPIcode (chemicals)
More informationExplosion proof enclosures
1 of 7 Explosion proof enclosures DE8 C 2 of 7 The Ex d enclosures are rugged and designed for harsh environment like: Oil and gas industry Chemical industry Pharmaceutical industry Agribusiness Without
More informationCalculations and OELs
Calculations and Dr. Peter Bellin, PhD, CIH EOH 466A The Occupational Environment Definitions Particulates: dust, fume, mist Vapors Gases Particle size considerations Inspirable Particulate Mass (IPM)
More informationFINAL DRAFT. PILOT STUDY TO ESTIMATE ASBESTOS EXPOSURE from VERMICULITE ATTIC INSULATION. Research Conducted in 2001 and 2002.
PILOT STUDY TO ESTIMATE ASBESTOS EXPOSURE from VERMICULITE ATTIC INSULATION Research Conducted in 2001 and 2002 Prepared for: Fibers and Organics Branch National Program Chemicals Division Office of Pollution
More informationExposure and Control of Ten High Priority MSHA Contaminants in Metal Mines. Presented by: Shannon E. Newton, MPH, CIH
Exposure and Control of Ten High Priority MSHA Contaminants in Metal Mines Presented by: Shannon E. Newton, MPH, CIH Topics What are these hazards? How can they harm you? Why do we care? How do we tell
More informationSample preparation for X-ray fluorescence analysis
Technical articles Sample preparation for X-ray fluorescence analysis III. Pressed and loose powder methods Gakuto Takahashi* 1. Introduction There are two main sample preparation techniques for measurement
More information