1 HDR104 RADIOGRAPHIC EQUIPMENT AND IMAGE RECORDING 1 CHAPTER 1 MAJOR COMPONENTS OF X-RAY GENERATING UNIT PREPARED BY: MR KAMARUL AMIN BIN ABDULLAH SCHOOL OF MEDICAL IMAGING FACULTY OF HEALTH SCIENCES

2 LEARNING OUTCOMES At the end of the lesson, the student should be able to:- Describe what x-ray tube is. Identify the parameter selectors on control panel and their functions. Discuss the characteristics of x-ray couch. Describe the types of tube stand. Describe briefly the principle of Bucky. Slide 2 of 52

3 OUTLINES INTRODUCTION 1.1 X-ray Tube 1.4 Tube Stand 1.2 Control Panel 1.5 Bucky Tray 1.3 X-ray Couch (Table) 1.6 References Slide 3 of 52

4 INTRODUCTION Slide 4 of 52

5 INTRODUCTION Slide 5 of 52

6 1.1 X-ray Tube Specifications It is attached to a tube support. It produces X-rays & Heat. It can operate at high voltage which ranges from kvp. It has an insulation which is lead lined to avoid leakage of radiation. Oil is used to cool X-ray tube during exposures. It has an indicator such as height, kvp, mas and motion. Figure 1: The x-ray tube. Slide 6 of 52

7 1.1 X-ray Tube Specifications (Continued) It is being incorporated with safety interlocks to avoid injury to staff and patients. It has Light Beam Diaphragm to include the area of interest. It also has an Integrated Collimation with various sizes according to the projection and cassettes/films. This equipment is Earthed (Shockproof). It is connected with high tension cables that supplies the high voltage current. Figure 1: The x-ray tube. Slide 7 of 52

8 1.1 X-ray Tube X-ray tube encased within LEAD lined case. Figure 2: The x-ray tube with diagram. Slide 8 of 52

9 1.2 Control Panel There are THREE primary control 1. kvp quality 2. ma 3. Time quantity Auxiliary control (anatomical AEC control) Slide 9 of 52

10 1.2 Control Panel Main Power AEC Bucky Selector kvp Selector ma selector Timer Selector Exposure Switch Focal Spot Selector Anatomical Part Selector Figure 3: The picture shows an example of control panel. Slide 10 of 52

11 1.2 Control Panel Figure 4: Control Panel (Somasundaram, 2009) Figure 5: Control Panel (Medilux, 2010) Slide 11 of 52

12 1.2 Control Panel kvp(kilovoltagepeak) The energy (the penetrating power) of the x-ray beam is controlled by the voltage adjustment. It is labelled in kev (thousand electron volts) and sometimes the level is referred to as kvp (kilovoltage potential). The higher the voltage setting, the more energetic will be the beam of x-ray. It plays major part in affecting contrast of image. Slide 12 of 52

13 1.2 Control Panel ma (MiliAmperage) The second control of the output of the x-ray tube is called the ma (milliamperage) control. This control determines how much current is allowed to flow through the filament which is the cathode side of the tube. The effect of the ma circuit is quite linear. It is more concerned in density aspect of the image. Slide 13 of 52

14 1.2 Control Panel Timer (s) Control length of time machine produces x rays This is usually denoted as an "S" (exposure time in seconds) and is combined with the ma control. e.g using 10 milliampere seconds you could use a 10 ma current with a 1.0 second exposure or a 20 ma current for a 0.5 second exposure or any combination of the two which would result in the number 10. Both of these factors and their combination affect the film in a linear way. Slide 14 of 52

15 1.2 Control Panel AEC (Automatic Exposure Chamber) Variable densities exist within subjects. AEC s monitor the beam of X-rays transmitted from the patient. Production of X-ray radiation can be measured. Located under patient but above cassette. Minimizes inaccurate exposures. Figure 6: Automatic Exposure Chamber Ionization Chamber. Slide 15 of 52

16 1.2 Control Panel Exposure terminates once a sufficient quantity of radiation has been received to produce a radiographic image of required density. Permit accurate exposures Usually three AED s incorporated in table tops and vertical chest stands. Markings usually on surface of table/chest stand. Individual chambers may be selected depending on the exam being performed. Slide 16 of 52

17 1.2 Control Panel Figure 7: The position of the chambers. Slide 17 of 52

18 1.3 X-ray Couch (Table) Figure 8: The picture shows the x-ray table/couch. Slide 18 of 52

19 1.3 X-ray Couch (Table) Specifications It must be strong enough to support large patients or obese. It should have low attenuation materials (such as Carbon Fibre) as to improve quality of x-ray beam. It has Rise and Fall facilities. Figure 9: X-ray Couch Slide 19 of 52

20 1.3 X-ray Couch (Table) Specifications (Continued) It has Kick buttons. Incorporated with an Integrated Bucky tray. Insulated to avoid electrical shocked. It is easy to clean as to reduce cross infection. Figure 9: X-ray Couch It is associated with pressure switches. Slide 20 of 52

21 1.4 Tube Stand Figure 10: Ceiling Mounted X-ray Tube Figure 11: Floor Mounted X-ray Tube Slide 21 of 52

22 1.4 Tube Stand Advantages Flexibility Ease of use and Durable Safety features/interlocks Specifications of modern X-ray tube Integral features Wide range of movements Disadvantages Expensive Interlock too sensitive Figure 10: Ceiling Mounted X- ray Tube Slide 22 of 52

23 1.4 Tube Stand Advantages Inexpensive. Can be used with upright Bucky. Require less space. Disadvantages Less flexible. Not user friendly for complicated cases. Figure 11: Floor Mounted X-ray Tube Slide 23 of 52

24 1.5 Bucky Tray Figure 12: Table Bucky Figure 13: Erect Bucky It actually acts as cassette holder. Slide 24 of 52

25 1.5 Bucky Tray Integrated within table top. Lock mechanism ensures cassette is secure. Grid mechanism incorporated within cassette tray. Grid reduces scatter radiation from certain examinations:- Abdomen Pelvis Hips Figure 14: Bucky Tray with Grid. Slide 25 of 52

26 1.5 Bucky Tray Important that X-ray tube and cassette tray are aligned Set distance between X-ray tube and Cassette tray Oscillating movement ensures scatter is absorbed before reaching film Figure 15: The x-ray tube and bucky. Slide 26 of 52

27 1.5 References No. REFERENCES 1 Richard R. Carlton, Arlene McKenna Adler (2005) Principles of Radiographic Imaging, Delmar 2 Bushong, S. C. (2008). Radiologic science for technologists. Canada: Elsevier. Slide 27 of 52

28 SUMMARY There are FIVE major components of x-ray generating unit:- 1. X-ray Tube 2. Control Panel 3. X-ray Couch 4. Tube Stand 5. Bucky Tray Slide 28 of 52

29 NEXT SESSION PREVIEW CHAPTER 2: ELECTROSTATIC Slide 29 of 52

30 APPENDIX FIGURE Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 SOURCE B7-2b.JPG pg Slide 30 of 52

31 APPENDIX FIGURE Figure 10 Figure 11 Figure 12 Figure 13 Figure 14 Figure 15 Figure 16 Figure 17 Figure 18 Figure 19 Figure 20 Figure 21 SOURCE ion_2inf416.gif Slide 31 of 52

32 APPENDIX FIGURE Figure 22 Figure 23 Figure 24 Figure 25 Figure 26 Figure 27 SOURCE Slide 32 of 52

33 Activity Test Your Knowledge Answer the question. What is the push or pull of an object that can cause it to accelerate called? Mass Speed Force Slide 33 of 52

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the discussion on the major components of the x-ray machine. Today

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Production of X-rays and Interactions of X-rays with Matter

Production of X-rays and Interactions of X-rays with Matter Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage

More information

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the

More information

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION

More information

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses.

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. 1 3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. (A) Any person may apply to the director for approval to conduct an educational program for

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum

More information

Physics 42 Lab 4 Fall 2012 Cathode Ray Tube (CRT)

Physics 42 Lab 4 Fall 2012 Cathode Ray Tube (CRT) Physics 42 Lab 4 Fall 202 Cathode Ray Tube (CRT) PRE-LAB Read the background information in the lab below and then derive this formula for the deflection. D = LPV defl 2 SV accel () Redraw the diagram

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding

More information

X-ray Imaging System. X-Ray Circuit. Principles of Imaging Science II (RAD 120) X-ray Imaging System Circuitry

X-ray Imaging System. X-Ray Circuit. Principles of Imaging Science II (RAD 120) X-ray Imaging System Circuitry Principles of Imaging Science II (RAD 120) X-ray Imaging System Circuitry X-ray Imaging System Operating console Set x-ray tube current (quantity) and voltage (quality) Controls line compensation, kvp,

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams EPEX Direct-to-Digital System for General Radiography Featuring... Versatility for a complete range of general radiographic exams The Hologic EPEX General Radiography System is a direct-to-digital, fully

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

1 Safety instructions. 2 Device components. 3 Function. Leader in Led solutions. Compact presence detector. Operating instructions

1 Safety instructions. 2 Device components. 3 Function. Leader in Led solutions. Compact presence detector. Operating instructions Compact presence detector Operating instructions 1 Safety instructions Electrical equipment may only be installed and fitted by electrically skilled persons. Failure to observe the instructions may cause

More information

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine

More information

Position Classification Standard for Diagnostic Radiologic Technologist Series, GS-0647

Position Classification Standard for Diagnostic Radiologic Technologist Series, GS-0647 Position Classification Standard for Diagnostic Radiologic Technologist Series, GS-0647 Table of Contents SERIES DEFINITION... 2 EXCLUSIONS... 2 TITLES... 3 OCCUPATIONAL INFORMATION... 3 GRADING OF POSITIONS...

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)

More information

Running head: OBESITY AND MEDICAL IMAGING 1. Obesity and Medical Imaging. November 13, 2012

Running head: OBESITY AND MEDICAL IMAGING 1. Obesity and Medical Imaging. November 13, 2012 Running head: OBESITY AND MEDICAL IMAGING 1 Obesity and Medical Imaging November 13, 2012 OBESITY AND MEDICAL IMAGING 2 Abstract Obesity is a growing problem in the United States, and it affects the health

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Kodak s dental radiograph series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at

More information

33-10-06-03. General requirements. 1. Administrative Controls.

33-10-06-03. General requirements. 1. Administrative Controls. 33-10-06-03. General requirements. 1. Administrative Controls. a. Registrant. The registrant shall be responsible for directing the operation of the x-ray systems which have been registered with the department.

More information

HERZ-Thermal Actuators

HERZ-Thermal Actuators HERZ-Thermal Actuators Data Sheet 7708-7990, Issue 1011 Dimensions in mm 1 7710 00 1 7710 01 1 7711 18 1 7710 80 1 7710 81 1 7711 80 1 7711 81 1 7990 00 1 7980 00 1 7708 11 1 7708 10 1 7708 23 1 7709 01

More information

Lab 4: Magnetic Force on Electrons

Lab 4: Magnetic Force on Electrons Lab 4: Magnetic Force on Electrons Introduction: Forces on particles are not limited to gravity and electricity. Magnetic forces also exist. This magnetic force is known as the Lorentz force and it is

More information

Diagnostic X-ray facilities as per quality control performances in Tanzania

Diagnostic X-ray facilities as per quality control performances in Tanzania JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 4, FALL 2006 Diagnostic X-ray facilities as per quality control performances in Tanzania Yesaya Y. Sungita, 1 Simon S.L. Mdoe, 1 and Peter

More information

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25 Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25 CHAPTER 227. RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT, X-RAY GAUGING EQUIPMENT, ELECTRON MICROSCOPES AND X-RAY CALIBRATION SYSTEMS Sec. 227.1.

More information

portable x-ray survey report

portable x-ray survey report department of HealtH and Human services centers for medicare & medicaid services form approved omb no. 0938-0027 portable x-ray survey report provider number H1 date surveyed (1) initial (2) survey W H2

More information

RADIATION SAFETY STANDARD

RADIATION SAFETY STANDARD Radiation Safety Act 1999 RADIATION SAFETY STANDARD NM004:2010 Standard for radiation apparatus used to carry out industrial radiography Preface Under section 17 of the Radiation Safety Act 1999, a possession

More information

VCR Ion Beam Sputter Coater

VCR Ion Beam Sputter Coater VCR Ion Beam Sputter Coater Sputtering Process and Rates 2 Vacuum System 3 Loading the Sputter Chamber 4 Sputter Coating 5 Removing Samples from Chamber 6 Appendix A: VCR High Vacuum Gauge Conditioning

More information

How To Become An Xray Technician

How To Become An Xray Technician MICHIGAN CIVIL SERVICE COMMISSION JOB SPECIFICATION X-RAY TECHNICIAN JOB DESCRIPTION Employees in this job function as radiological technologists with responsibility for operating X-ray and fluoroscopic

More information

3PH2 or 3PHE 3 Phase Control Board

3PH2 or 3PHE 3 Phase Control Board 3PH2 or 3PHE 3 Phase Control Board The 3-Phase Control Board is used to control automatic doors or gates that operate with 3-Phase or high current single-phase motors in high usage residential and commercial

More information

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.

More information

Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D104 706114 / 00 01 / 2012

Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D104 706114 / 00 01 / 2012 Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D104 706114 / 00 01 / 2012 Contents 1 Preliminary note 3 1.1 Symbols used 3 1.2 Warning signs used 3 2 Safety instructions

More information

Cathode Ray Tube. Introduction. Functional principle

Cathode Ray Tube. Introduction. Functional principle Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the

More information

USER S MANUAL HSC-24A

USER S MANUAL HSC-24A AIRREX AIR CONDITIONER USER S MANUAL HSC-24A Thank you for purchasing an AIRREX AIR CONDITIONER. BEFORE operation please read this user s manual carefully. Keep this manual readily available. It is ESSENTIAL

More information

JB-70 Dental X-Ray System

JB-70 Dental X-Ray System Dental X-Ray System User Manual 00-02-1568 Rev. U 0120 ECN: P3250 Attention: The equipment must only be installed and operated in accordance with the safety procedures and operating instructions given

More information

JEFFERSON COLLEGE. Radiographic Physics

JEFFERSON COLLEGE. Radiographic Physics JEFFERSON COLLEGE COURSE SYLLABUS RAD160 Radiographic Physics 3 Credit Hours Revised by: Janet E. Akers BS RT (R)(M) Date: October 3, 2013 Kenny Wilson, Director, Health Occupation Programs Dena McCaffrey,

More information

Laserlyte-Flex Alignment System

Laserlyte-Flex Alignment System Laserlyte-Flex Alignment System LaserLyte-Flex The LaserLyte-Flex Alignment System is a unique, interchangeable, low cost plug and play laser system. Designed specifically for aligning and positioning

More information

MEDICAL IMAGING. 90/30 Remote Controlled RF System

MEDICAL IMAGING. 90/30 Remote Controlled RF System MEDICAL IMAGING 90/30 Remote Controlled RF System The Easy solution for your Remote Controlled room Apollo EZ represents the Easy approach for the radiographic/fluoroscopic room and it comes from the experience

More information

E/M Experiment: Electrons in a Magnetic Field.

E/M Experiment: Electrons in a Magnetic Field. E/M Experiment: Electrons in a Magnetic Field. PRE-LAB You will be doing this experiment before we cover the relevant material in class. But there are only two fundamental concepts that you need to understand.

More information

FXA 2008. UNIT G484 Module 2 4.2.3 Simple Harmonic Oscillations 11. frequency of the applied = natural frequency of the

FXA 2008. UNIT G484 Module 2 4.2.3 Simple Harmonic Oscillations 11. frequency of the applied = natural frequency of the 11 FORCED OSCILLATIONS AND RESONANCE POINTER INSTRUMENTS Analogue ammeter and voltmeters, have CRITICAL DAMPING so as to allow the needle pointer to reach its correct position on the scale after a single

More information

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Two-Body System: Two Hanging Masses

Two-Body System: Two Hanging Masses Specific Outcome: i. I can apply Newton s laws of motion to solve, algebraically, linear motion problems in horizontal, vertical and inclined planes near the surface of Earth, ignoring air resistance.

More information

X-ray (Radiography) - Bone

X-ray (Radiography) - Bone Scan for mobile link. X-ray (Radiography) - Bone Bone x-ray uses a very small dose of ionizing radiation to produce pictures of any bone in the body. It is commonly used to diagnose fractured bones or

More information

Integrated Solar Radiant Systems

Integrated Solar Radiant Systems Integrated Solar Radiant Systems William Shady PE President Topics Radiant heating Indoor air quality Radiant Cooling Project Photos Questions and answers The goal for our clients Healthy Comfort Why Radiant

More information

UPDATEUPDATEUPDATE. Converting the tower chassis to a rack-mount chassis. A - M5 13 8 To mount outer member to rack

UPDATEUPDATEUPDATE. Converting the tower chassis to a rack-mount chassis. A - M5 13 8 To mount outer member to rack UPDATEUPDATEUPDATE Converting the tower chassis to a rack-mount chassis Rail overview Converting your Server to Rack-Mount The rail kit includes the following: Two rail units Five packages of screws (labeled

More information

SALES SPECIFICATION. SC7640 Auto/Manual High Resolution Sputter Coater

SALES SPECIFICATION. SC7640 Auto/Manual High Resolution Sputter Coater SALES SPECIFICATION SC7640 Auto/Manual High Resolution Sputter Coater Document Number SS-SC7640 Issue 1 (01/02) Disclaimer The components and packages described in this document are mutually compatible

More information

PowlVac Vacuum Integrity Tester

PowlVac Vacuum Integrity Tester Instructions IB-60025B PowlVac Vacuum Integrity Tester POWELL ELECTRICAL MANUFACTURING COMPANY 8550 MOSLEY DRIVE HOUSTON, TEXAS 77075 USA PHONE (713) 944-6900 FAX (713) 947-4453 www.powellelectric.com

More information

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act

More information

XRS-3 X-RAY SOURCE OPERATOR S MANUAL

XRS-3 X-RAY SOURCE OPERATOR S MANUAL XRS-3 X-RAY SOURCE OPERATOR S MANUAL CONTENTS ITEM PAGE 1.0 INTRODUCTION... 2 2.0 WARNINGS... 2 EXCLUSION ZONE DIAGRAM. 4 3.0 PHYSICAL DESCRIPTION... 5 3.1 HIGH VOLTAGE PULSER/TUBEHEAD... 5 3.2 BASE...

More information

XR200 X-RAY SOURCE OPERATOR S MANUAL

XR200 X-RAY SOURCE OPERATOR S MANUAL XR200 X-RAY SOURCE OPERATOR S MANUAL CONTENTS ITEM PAGE 1.0 INTRODUCTION... 2 2.0 WARNINGS... 2 EXCLUSION ZONE DIAGRAM... 4 3.0 PHYSICAL DESCRIPTION... 5 3.1 HIGH VOLTAGE PULSER/TUBEHEAD... 5 3.2 BASE...

More information

Analytical X-ray Safety Refresh Training. Radiation Safety Officer Department of Environmental Health and Safety

Analytical X-ray Safety Refresh Training. Radiation Safety Officer Department of Environmental Health and Safety Analytical X-ray Safety Refresh Training Radiation Safety Officer Department of Environmental Health and Safety TWO TYPES OF X-RAYS Man-made X-rays Electrons are accelerated from the cathode by high voltage,

More information

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information

More information

Required RS Training Info

Required RS Training Info C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness

More information

SP1790JK 900MHz Wireless Indoor/Outdoor Speakers. User Manual INTRODUCTION FEATURES IMPORTANT SAFETY INFORMATION

SP1790JK 900MHz Wireless Indoor/Outdoor Speakers. User Manual INTRODUCTION FEATURES IMPORTANT SAFETY INFORMATION SP1790JK 900MHz Wireless Indoor/Outdoor Speakers INTRODUCTION This 900 MHz digital hybrid wireless speaker system uses the latest wireless technology that enables you to enjoy music and TV sound anywhere

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation Safety Characteristics of the NOMAD Portable X-ray System Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality

More information

Instruction Manual Differential Pressure Transmitter

Instruction Manual Differential Pressure Transmitter DE28 Instruction Manual Differential Pressure Transmitter Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

W850. Maintenance Guide

W850. Maintenance Guide W850 Maintenance Guide August 2011 www.lexmark.com Contents...3 Cleaning the exterior of the printer...3 Storing supplies...3 Conserving supplies...3 Checking the status of supplies...4 Checking the status

More information

The photoionization detector (PID) utilizes ultraviolet

The photoionization detector (PID) utilizes ultraviolet Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).

More information

Charges, voltage and current

Charges, voltage and current Charges, voltage and current Lecture 2 1 Atoms and electrons Atoms are built up from Positively charged nucleus Negatively charged electrons orbiting in shells (or more accurately clouds or orbitals) -

More information

The Three Heat Transfer Modes in Reflow Soldering

The Three Heat Transfer Modes in Reflow Soldering Section 5: Reflow Oven Heat Transfer The Three Heat Transfer Modes in Reflow Soldering There are three different heating modes involved with most SMT reflow processes: conduction, convection, and infrared

More information

Wednesday 16 January 2013 Afternoon

Wednesday 16 January 2013 Afternoon Wednesday 16 January 2013 Afternoon A2 GCE PHYSICS B (ADVANCING PHYSICS) G494/01 Rise and Fall of the Clockwork Universe *G411660113* Candidates answer on the Question Paper. OCR supplied materials: Data,

More information

Replacing the vacuum tube

Replacing the vacuum tube wunder cm7 This microphone has been meticulously created to meet the demand for the U47 sound using historically correct components around the heart of the microphone the original, specially selected EF14

More information

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come

More information

STATIC AND KINETIC FRICTION

STATIC AND KINETIC FRICTION STATIC AND KINETIC FRICTION LAB MECH 3.COMP From Physics with Computers, Vernier Software & Technology, 2000. INTRODUCTION If you try to slide a heavy box resting on the floor, you may find it difficult

More information

SECTION 16720 - FIRE AND SMOKE ALARM SYSTEM. City of San Diego, CWP Guidelines

SECTION 16720 - FIRE AND SMOKE ALARM SYSTEM. City of San Diego, CWP Guidelines PART 1 -- GENERAL 1.1 WORK OF THIS SECTION SECTION 16720 - City of San Diego, CWP Guidelines A. The WORK of this Section includes providing manual and automatic fire alarm and smoke detection systems meeting

More information

Ducoterra Radiant Heating Panel Installation Manual

Ducoterra Radiant Heating Panel Installation Manual Ducoterra Radiant Heating Panel Installation Manual 1. Introduction Your new radiant heating panels are designed to heat living and working spaces rapidly and efficiently by radiant heating. Like the sun,

More information

LUCCI AIRFUSION QUEST II CEILING FAN

LUCCI AIRFUSION QUEST II CEILING FAN LUCCI AIRFUSION QUEST II CEILING FAN WITH IR REMOTE INSTALLATION OPERATION MAINTENANCE WARRANTY INFORMATION CAUTION READ INSTRUCTIONS CAREFULLY FOR SAFE INSTALLATION AND FAN OPERATION. V1.0 QUEST II IR

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

The Basics of Scanning Electron Microscopy

The Basics of Scanning Electron Microscopy The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single

More information

Abdomen X-Ray (AXR) Collimation is ideally from diaphragms to lower border of the symphysis pubis and the lateral skin margins.

Abdomen X-Ray (AXR) Collimation is ideally from diaphragms to lower border of the symphysis pubis and the lateral skin margins. Abdomen X-Ray (AXR) Collimation is ideally from diaphragms to lower border of the symphysis pubis and the lateral skin margins. LMP of child-bearing age female patients should be checked. 1. Acute abdomen

More information

AS COMPETITION PAPER 2007 SOLUTIONS

AS COMPETITION PAPER 2007 SOLUTIONS AS COMPETITION PAPER 2007 Total Mark/50 SOLUTIONS Section A: Multiple Choice 1. C 2. D 3. B 4. B 5. B 6. A 7. A 8. C 1 Section B: Written Answer Question 9. A mass M is attached to the end of a horizontal

More information

Radiation Safety Committee College of William and Mary

Radiation Safety Committee College of William and Mary Radiation Safety Committee College of William and Mary Policy for the operation of Analytical X-ray Equipment The following rules govern the use of analytical X-ray equipment at the College of William

More information

XRS-4 X-RAY SOURCE OPERATOR S MANUAL

XRS-4 X-RAY SOURCE OPERATOR S MANUAL XRS-4 X-RAY SOURCE OPERATOR S MANUAL CONTENTS ITEM PAGE 1.0 INTRODUCTION... 2 2.0 WARNINGS... 2 DUTY CYCLE. 2 3.0 PHYSICAL DESCRIPTION... 3 HIGH VOLTAGE PULSER/TUBEHEAD..... 3 BASE... 3 BATTERY PACK...

More information

AAPM REPORT NO. 14 PERFORMANCE SPECIFICATIONS AND ACCEPTANCE TESTING FOR X-RAY GENERATORS AND AUTOMATIC EXPOSURE CONTROL DEVICES

AAPM REPORT NO. 14 PERFORMANCE SPECIFICATIONS AND ACCEPTANCE TESTING FOR X-RAY GENERATORS AND AUTOMATIC EXPOSURE CONTROL DEVICES AAPM REPORT NO. 14 PERFORMANCE SPECIFICATIONS AND ACCEPTANCE TESTING FOR X-RAY GENERATORS AND AUTOMATIC EXPOSURE CONTROL DEVICES Published for the American Association of Physics in Medicine by the American

More information

Procurement Specification for a High Security Lift Arm Barrier (HSLAB) EB950 CR Armstrong High Security Lift Arm Barrier

Procurement Specification for a High Security Lift Arm Barrier (HSLAB) EB950 CR Armstrong High Security Lift Arm Barrier Procurement Specification for a High Security Lift Arm Barrier (HSLAB) EB950 CR Armstrong High Security Lift Arm Barrier A. Requirement This document is to be used to specify the physical and operational

More information

1. At which temperature would a source radiate the least amount of electromagnetic energy? 1) 273 K 3) 32 K 2) 212 K 4) 5 K

1. At which temperature would a source radiate the least amount of electromagnetic energy? 1) 273 K 3) 32 K 2) 212 K 4) 5 K 1. At which temperature would a source radiate the least amount of electromagnetic energy? 1) 273 K 3) 32 K 2) 212 K 4) 5 K 2. How does the amount of heat energy reflected by a smooth, dark-colored concrete

More information

MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference

MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725 Read this manual thoroughly before using and save it for future reference Purchase MAP capsules at Harvey Norman stores nationwide

More information

STATE OF NEBRASKA STATUTES RELATING TO MEDICAL RADIOGRAPHY PRACTICE ACT

STATE OF NEBRASKA STATUTES RELATING TO MEDICAL RADIOGRAPHY PRACTICE ACT 2010 STATE OF NEBRASKA STATUTES RELATING TO MEDICAL RADIOGRAPHY PRACTICE ACT Department of Health and Human Services Division of Public Health Licensure Unit 301 Centennial Mall South, Third Floor PO Box

More information

KS-70. Catheterization Table GENERAL

KS-70. Catheterization Table GENERAL PD505-009A Catheterization Table KS-70 GENERAL This Catheterization Table KS-70, which is combined with X-ray high voltage generator, X-ray tube assembly, X-ray image system (Image Intensifier, Flat Panel

More information

Disassembly of Waste Electrical and Electronic Equipment (WEEE) Manual

Disassembly of Waste Electrical and Electronic Equipment (WEEE) Manual Disassembly of Waste Electrical and Electronic Equipment (WEEE) Manual EU Waste Electronic and Electrical Equipment Directive require producers to provide information of the different electronic and electrical

More information

GenStore Archive Cabinet

GenStore Archive Cabinet GenStore Archive Cabinet Setup and Maintenance Manual Version A April 2014 Two GenStore Archives shown mounted on optional table stand. 2 Contents Getting Started... 4 Using this Manual...4 Safety Precautions...4

More information

X-ray (Radiography), Chest

X-ray (Radiography), Chest X-ray (Radiography), Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray makes images of the heart, lungs, airways,

More information

Series 510 Submersible Level Transmitters

Series 510 Submersible Level Transmitters Series 510 Submersible Level Transmitters The Series 510 Submersible Level Transmitters are solid state instruments designed for direct submergence into many types of liquid for quick, accurate and reliable

More information

Modern Physics Laboratory e/m with Teltron Deflection Tube

Modern Physics Laboratory e/m with Teltron Deflection Tube Modern Physics Laboratory e/m with Teltron Deflection Tube Josh Diamond & John Cummings Fall 2010 Abstract The deflection of an electron beam by electric and magnetic fields is observed, and the charge

More information

Overview. Components of an alarm system Technology behind the equipment Communication Platforms Industry Terminology False Alarm Tips DSC products

Overview. Components of an alarm system Technology behind the equipment Communication Platforms Industry Terminology False Alarm Tips DSC products Basic Security 101 Overview Components of an alarm system Technology behind the equipment Communication Platforms Industry Terminology False Alarm Tips DSC products Components of an Alarm System Main Control

More information