IAEA code of practice for dosimetry in diagnostic radiology: recommendations for hospitals and specific issues

Size: px
Start display at page:

Download "IAEA code of practice for dosimetry in diagnostic radiology: recommendations for hospitals and specific issues"

Transcription

1 IAEA code of practice for dosimetry in diagnostic radiology: recommendations for hospitals and specific issues Antti Kosunen Radiation Metrology Laboratory-STUK Radiation safety and dosimetry in medical applications May 2012 TTÜ, Ehitajate tee 5, Tallinn, Estonia RADIATION AND NUCLEAR SAFETY AUTHORITY Presentation Review /introduction to contents of IAEA TRS 457 for clinical dosimetry Experiences of use the IAEA TRS 457 (KAP meters, others) Remarks based on work at the SSDL of STUK Methods by TLD are not reviewed RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 2 1

2 RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 3 IAEA TRS 457, Clinical dosimetry Dosimetry formalism Dosimetry intrumentation Quantities and uncertainties General radiography Fluoroscopy Mammography Computed tomography Dental radiography RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 4 2

3 IAEA TRS 457, Clinical dosimetry Dosimetry phantom patient Guidance for selection of patients for DRL Appendix VI. Application of patient dose measurements Background Guidance for DRLs Guidance for organ dose conversion factors for examinations Summary of using effective dose RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 5 IAEA TRS 457, Clinical dosimetry Assumed that the x- ray equipment is properly maintained according to QA program Assumed that the x ray tube output is linear relative to tube loading, if not complementary measurements are required HVL For interpolations of calibration coefficients For conversions to organ doses (mammography) Not necessrerily for every dose measurement, as information from Quality Control results exist RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 6 3

4 IAEA TRS 457, Clinical dosimetry Examples and worksheets for measurements Dosimeters should be calibrated at the laboratory Calibration interval: Based on national requirements, Usually should not exceed two years RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 7 IAEA TRS 457, Accuracy requirements for dosimetry At reference conditions of dosimetry Risk assessment : Absolute risk Dose surveys Relative risk Interventional procedures Quality assurance measurements Paediatric procedures ( 2sd ) ~20 % 7 % RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 8 4

5 IAEA TRS 457, Dosimetry formalism Air kerma based formalism Beam quality correction factor RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 9 IAEA TRS 457, Selection of instruments, clinical Requirements of International Standards are followed: Requirements for dosemeters at laboratory can be used also as a background information for clinical dosemeters RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 10 5

6 IAEA TRS 457, Measurement uncertainty (accuracy) RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 11 IAEA TRS 457, General radiology, dosimetry quantities RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 12 6

7 IAEA TRS 457, General radiology Incident air kerma Air kerma detector with isotropic response RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 13 IAEA TRS 457, General radiology, phantoms RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 14 7

8 IAEA TRS 457, Review on phantoms RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 15 IAEA TRS 457, General radiology, uncertainty RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 16 8

9 IAEA TRS 457, Fluoroscopy, dosimetry quantities RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 17 IAEA TRS 457, Fluoroscopy Detector with back scatter measurement capability Anti- scatter grid in position if used Focus to table top distance as used clinically Phantom 20 cm, 30 cm for large patients Collimation large, to phantom size Exposure under ABC RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 18 9

10 IAEA TRS 457, Fluoroscopy RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 19 IAEA TRS 457, Fluoroscopy, uncertainties RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 20 10

11 IAEA TRS 457, Fluoroscopy, kerma -area product (P KA ) Calibration of KAP meters in situ Beam area method Tandem method with reference KAP meter Toroi et.al. A tandem calibration method for kerma-area product meters, PMB 53, 2008 RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 21 IAEA TRS 457, Kerma -area product Incident beam KAP meter Transmitted beam 5 mmal total filtration Toroi et.al. Effects of radiation quality on the calibration of kerma-area product meters in x-ray beams, PMB 53, 2008 RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 22 11

12 IAEA TRS 457, Fluoroscopy, Kerma area- product Over and under couch set-ups Beam area or tandem method RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 23 IAEA TRS 457, KAP -meters Toroi et.al. Effects of radiation quality on the calibration of kerma-area product meters in x-ray beams, PMB 53, 2008 RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 24 12

13 IAEA TRS 457, Kerma -area product (KAP) RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 25 IAEA TRS 457, KAP-meters Large fields for reference KAP meter 30 cm * 30 cm Patient Dose Calibrator PDC-1 Radcal kv Toroi & Kosunen. The energy dependence of the response of a patient dose calibrator, PMB 54, 2009 RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 26 13

14 KAP-meters, Set-up 80 kv, 4.8 mmal, 10*10cm at 30 FDD Toroi et.al. Calibration of kerma area product meters with a patient dose calibrator, IDOS proceedings, Vol2, IAEA, Standards-Applications-and-Quality-Assurance-in-Medical-Radiation-Dosimetry-IDOS- Proceedings-of-an-International-Symposium-Vienna-9-12-November-2010 RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 27 IAEA TRS 457, KAP uncertainties IEC 60580, 25 % (k=2) In examinations with varying voltage, 25 % For specific tube voltage and energy dependence considered: 6 % (k=2) KAP indicators based on calculations need to be calibrated also RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 28 14

15 IAEA TRS 457, Mammography, dosimetry quantities RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 29 IAEA TRS 457, Mammography Plane parallel IC with entrance window (low energy dependence) 45 mm PMMA phantom RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 30 15

16 IAEA TRS 457, Mammography, HVL HVL measurement RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 31 IAEA TRS 457, Mammography, glandular dose RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 32 16

17 IAEA TRS 457, Mammography, uncertainties RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 33 IAEA TRS 457, Computed tomography, dosimetry quantities RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 34 17

18 IAEA TRS 457, Computed tomography CT air kerma indices In air or in a phantom T, Nominal slice thickness N, number of slices/detector rows per scan for MSCT/MDCT RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 35 IAEA TRS 457, Computed tomography, phantoms RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 36 18

19 IAEA TRS 457, Dosimetry formalism RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 37 IAEA TRS 457, Computed tomography RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 38 19

20 IAEA TRS 457, Computed tomography Air kerma lenght product RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 39 IAEA TRS 457, Computed tomography, kerma lenght product RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 40 20

21 AAPM, IAEA, Recent approach for wide CT beam dosimetry Work by ICRU? RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 41 IAEA HHS No5, Computed tomography From :IAEA Human Health Report No:5, Status of computed tomography dosimetry for wide cone beam scanners, 2011 RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 42 21

22 Computed tomography, kerma lenght product, STUK approach Measurement of kerma lenght product for whole examination RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 43 IAEA TRS 457, Dental radiography, dosimetry quantities RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 44 22

23 IAEA TRS 457, Dental radiography TRS include quidance for bitewiew and panoramic examinations Uncertainties Kerma lenght product 13 % - 6 % Kerma area product 13,2 % 6,4 % RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 45 IAEA TRS 457, Implementation and testing Advanced techniques, The development of advanced dosimetry techniques for diagnostic and interventional radiology Coordinated Research Project on Implementation of the International Code of Practice on Dosimetry in Diagnostic Radiology TRS 457 RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 46 23

24 IAEA TRS 457, Implementation and testing Advanced techniques, Paediatrics - clinical Development of guidance General Radiography fluoroscopy CT analysis & uncertainty Interpretation of clinical data Development of guidance development of questionnaire organ dose + eff. Dose & risk SSDL Preparation for intercomparison intercomparison -SSDL - KLP intercomparison -SSDL - KAP new qualities [KAP & mammo] Farmer chambers with RQT I/II RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 47 IAEA TRS 457, Implementation and testing Advanced techniques, Advanced dosimetry Development of guidance MDCT cone beam CT vs CT mammographic tomosynthesis dental dosimetry Skin dose Development of guidance Film characterisation SSDL high dose calibration high dose patients low dose calibration low dose patients uncertainty DICOM Development of guidance survey of DICOM capability Uncertainty formulation Assessment of clinical scenarios assessment of uncertainty budget from scenarios Development of guidance I/II RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 48 24

25 Summary I/II IAEA TRS 457 contains a comprehensive description of dosimetry methods used for diagnostic radiology Two options for reference KAP are presented; beam area or use of a reference KAP meter Generally % (k=2) level of uncertainty (KAP meters 25 %) can be achieved for measurements in clinical dosimetry. Accuracy below 10% requires use of reference class dosemeters and/or calibration of dosemeter for user specific radiation qualities. RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 49 Summary II/II For dosimetry of wide beam CT the IAEA TRS457 methods need to be complemented by more recent dosimetry guidance (AAPM, IAEA, others) IAEA TRS 457 has limitations for wide beam CT dosimetry. More recent recommendations exist; AAPM (2010), IAEA (2011) Advanced dosimetry techniques for diagnostic and interventional radiology are under investigation in the coordinated research project by IAEA. RADIATION AND NUCLEAR SAFETY AUTHORITY / AK-STUK 50 25

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric

More information

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department

More information

Diagnostic Exposure Tracking in the Medical Record

Diagnostic Exposure Tracking in the Medical Record Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand

More information

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization

More information

The Exposure Index and Its Standardization

The Exposure Index and Its Standardization The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany ulrich.neitzel@philips.com Content The Exposure Index (EI) What is it? What is it good

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)

More information

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida Patient Dose Tracking for Imaging Studies David E. Hintenlang, Ph.D., DABR University of Florida Conflict of Interest Statement No affiliation or financial interests in any of the commercial products or

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)

More information

Assessing Radiation Dose: How to Do It Right

Assessing Radiation Dose: How to Do It Right Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY TECHNICAL REPORTS SERIES No. 398 Absorbed Dose Determination in External Beam Radiotherapy An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water Sponsored by the

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Recommendations for a technical quality control program for diagnostic X-ray equipment

Recommendations for a technical quality control program for diagnostic X-ray equipment Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads

More information

Digital Breast Tomosynthesis QC Requirements

Digital Breast Tomosynthesis QC Requirements Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

Personal Dosimetry in Dental Radiology

Personal Dosimetry in Dental Radiology Personal Dosimetry in Dental Radiology Position Statement Personal Dosimetry in Dental Radiology Radiological Protection Institute of Ireland May 2011 3 Clonskeagh Square, Dublin 14 Background The RPII

More information

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION

More information

DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE

DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE A web module produced by Committee 3 of the International Commission on Radiological Protection (ICRP Key Points Diagnostic

More information

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY B. Wall, (Co-ordinator) 1, D. Hart 1, H. Mol 2, A. Lecluyse 2, A. Aroua 3, P. Trueb 3, J. Griebel 4, E. Nekolla 4, P. Gron 5, H. Waltenburg 5, H. Beauvais-March

More information

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT

More information

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation Safety Characteristics of the NOMAD Portable X-ray System Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

Determination of the Effective Energy in X-rays Standard Beams, Mammography Level

Determination of the Effective Energy in X-rays Standard Beams, Mammography Level Determination of the Effective Energy in X-rays Standard Beams, Mammography Level Eduardo de Lima Corrêa 1, Vitor Vivolo 1, Maria da Penha A. Potiens 1 1 Instituto de Pesquisas Energéticas e Nucleares

More information

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT

More information

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Atomic Energy Regulatory Board Mumbai 400 094 1 DEFINITIONS 1. Quality Assurance (QA): Planned and systematic actions

More information

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment

More information

centre for radiation protection

centre for radiation protection centre for radiation protection To excel in radiation science so as to: enforce and promote the radiation safety of workers, the public and the environment; ensure that irradiating apparatus and nuclear

More information

RESPONSE BY THE CEO OF ARPANSA TO THE ADVICE ON MEDICAL RADIATION ISSUES from the RADIATION HEALTH AND SAFETY ADVISORY COUNCIL

RESPONSE BY THE CEO OF ARPANSA TO THE ADVICE ON MEDICAL RADIATION ISSUES from the RADIATION HEALTH AND SAFETY ADVISORY COUNCIL RESPONSE BY THE CEO OF ARPANSA TO THE ADVICE ON MEDICAL RADIATION ISSUES from the RADIATION HEALTH AND SAFETY ADVISORY COUNCIL INTRODUCTION The practice of radiology, nuclear medicine and radiotherapy

More information

Radiation survey meters used for environmental monitoring

Radiation survey meters used for environmental monitoring NKS-257 ISBN 978-87-7893-329-4 Radiation survey meters used for environmental monitoring Hans Bjerke (editor) (1) Thorgeir Sigurdsson, (2) Kurt Meier Pedersen (3) Jan-Erik Grindborg, Linda Persson, (4)

More information

Practical exercise: Effective dose estimate in CT

Practical exercise: Effective dose estimate in CT Practical exercise: Effective dose estimate in CT TRAINING COURCE PROGRAM 19 20 May 2011, Sofia, Bulgaria Virginia Tsapaki Medical Physics Dpt Konstantopoulio General Hospital email: virginia@otenet.gr

More information

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Fernanda B. C. Nonato, Claudinei T. Cescon and Linda V. E. Caldas, Instituto de Pesquisas Energéticas e Nucleares,

More information

Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose

Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose Monitoring Program for Computed Tomography, Digital Radiography, and Nuclear Medicine by Olav Christianson Medical Physics

More information

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES Ausra Urboniene, Birute Griciene Lithuania Introduction Medical staff during interventional radiology procedures

More information

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act

More information

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

Cone Beam Reconstruction Jiang Hsieh, Ph.D. Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition

More information

Eye lens dose measurements in Interventional Cardiology

Eye lens dose measurements in Interventional Cardiology Eye lens dose measurements in Interventional Cardiology Poster No.: 479 Congress: ESCR 2014 Type: Scientific Poster Authors: N. Kollaros 1, E. Carinou 2, C. Plemmenos 1, I. Stathopoulos 1, V. Keywords:

More information

The Calibration Chain: Role of BIPM,

The Calibration Chain: Role of BIPM, The Calibration Chain: Role of BIPM, PSDLs and ADCLs J. Seuntjens and M. McEwen McGill University & National Research Council CANADA Clinical Dosimetry for Radiotherapy, 2009 AAPM Summer School PSDL ADCL

More information

Radiation Safety Issues for Radiologic Technologists

Radiation Safety Issues for Radiologic Technologists Radiation Safety Issues for Radiologic Technologists Greg Sackett, M.S., CHP Medical Physicist Radiation Worker Risks? 1 Patient Risks? Acute Effects? Delayed Effects? Patient Questions? Radiation Dose

More information

Patient Exposure Doses During Diagnostic Radiography

Patient Exposure Doses During Diagnostic Radiography Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University

More information

Qualitätsmanagement-Handbuch

Qualitätsmanagement-Handbuch U r Kalibrier- und Messmöglichkeiten Qualitätsmanagement-Handbuch Relative erweiterte Messunsicherheit, die sich aus der Standardmessunsicherheit durch Multiplikation mit dem Erweiterungsfaktor k = 2 ergibt.

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

Staff Doses & Practical Radiation Protection in DEXA

Staff Doses & Practical Radiation Protection in DEXA Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /

More information

AAPM protocol for 40 300 kv x-ray beam dosimetry in radiotherapy and radiobiology

AAPM protocol for 40 300 kv x-ray beam dosimetry in radiotherapy and radiobiology AAPM protocol for 40 300 kv x-ray beam dosimetry in radiotherapy and radiobiology C.-M. Ma, Chair a) Radiation Oncology Dept., Stanford University School of Medicine, Stanford, California 94305-5304 and

More information

WHERE IN THE WORLD JILL LIPOTI?

WHERE IN THE WORLD JILL LIPOTI? WHERE IN THE WORLD IS JILL LIPOTI? HELLO FROM NEW JERSEY CRCPD - National Symposium on Fusion Imaging and Multimodalities February 18-20, 2004 Kansas City, Missouri New Jersey s Requirements As They Pertain

More information

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information

More information

Radiation Dose Measurements in Routine X ray Examinations

Radiation Dose Measurements in Routine X ray Examinations Tenth Radiation Physics & Protection Conference, 27-3 November 21, Nasr City - Cairo, Egypt Radiation Dose Measurements in Routine X ray Examinations H. Osman *1, A. Sulieman 1, I.I. Suliman 2 and A.K.

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration

SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration Implementation Page #: Page 1 of 10 Last Reviewed/Update 1. Purpose and Scope The purpose of this document is to describe the Medical Physics and Radiation Safety program at Boston University (BU) and

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

Challenges in small field MV photon dosimetry

Challenges in small field MV photon dosimetry Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland maria.aspradakis@ksl.ch Why are

More information

Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards

Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards Presented: August 3, 2011 Daniel Kassiday United States Food and Drug Administration Center for Devices and Radiological

More information

SOURCES AND EFFECTS OF IONIZING RADIATION

SOURCES AND EFFECTS OF IONIZING RADIATION SOURCES AND EFFECTS OF IONIZING RADIATION United Nations Scientific Committee on the Effects of Atomic Radiation UNSCEAR 2008 Report to the General Assembly with Scientific Annexes VOLUME I UNITED NATIONS

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems

CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems GE s Industrial X-ray and CT Forum, Cologne July 7 th 2015 Oliver Brunke GE Sensing & Inspection Technologies

More information

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS

APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS SAFETY REPORTS SERIES No. 39 APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL

More information

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma): Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 47 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL Refresher Course

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

GE Healthcare. DoseWatch. Gathering Radiation Dose Data GE Healthcare DoseWatch Gathering Radiation Dose Data Gathering Radiation Dose Data After your organization crosses the bridge of Why? for patient radiation dose management either due to regulation, risk

More information

Fundamentals of Cone-Beam CT Imaging

Fundamentals of Cone-Beam CT Imaging Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum

More information

USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY

USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY GUIDE ST 5.3 / 4 MAY 2007 USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY 1 Ge n e r a l 3 2 Safety licences and exemption from licensing 3 3 Safety licence-exempt use of radiation 3

More information

5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21 $1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/ (;32685(6 (XURSHDQ&RPPLVVLRQ

5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21 $1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/ (;32685(6 (XURSHDQ&RPPLVVLRQ 5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21 $1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/ (;32685(6 (XURSHDQ&RPPLVVLRQ European Commission 5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21$1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/(;32685(6

More information

IAEA HUMAN HEALTH SERIES

IAEA HUMAN HEALTH SERIES IAEA HUMAN HEALTH SERIES No. 19 Quality Assurance Programme for Computed Tomography: Diagnostic and Therapy Applications IAEA HUMAN HEALTH SERIES PUBLICATIONS The mandate of the IAEA human health programme

More information

Gestión global de la dosis en TC. Sistema de registro y gestión

Gestión global de la dosis en TC. Sistema de registro y gestión Gestión global de la dosis en TC. Sistema de registro y gestión IV Jornada de Protección Radiológica Hospitalaria SARH FEA Radiofísica Hospital Virgen de las Nieves. Granada jalmansa.lopez@gmail.com 22

More information

Monitoring Patient Radiation Dose in VA. Charles M. Anderson MD, PhD Chief Consultant for Diagnostic Services Veterans Health Administration

Monitoring Patient Radiation Dose in VA. Charles M. Anderson MD, PhD Chief Consultant for Diagnostic Services Veterans Health Administration Monitoring Patient Radiation Dose in VA Charles M. Anderson MD, PhD Chief Consultant for Diagnostic Services Veterans Health Administration TWENTY ONE/NOVEMBER 2011 Computed Tomography (CT), Magnetic Resonance

More information

Quality Assurance of Radiotherapy Equipment

Quality Assurance of Radiotherapy Equipment Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria

More information

SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS

SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS AERB SAFETY CODE NO. AERB/SC/MED-2 (Rev. 1) SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS Approved by the Board on October 5, 2001 This document is subject to review, after a period

More information

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Quality control of CT systems by automated monitoring of key performance indicators: a two-year study Patrik Nowik, a Robert Bujila,

More information

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement.

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement. DISCLAIMER: This publication is based on sources and information believed to be reliable, but the AAPM, the editors, and the publisher disclaim any warranty or liability based on or relating to the contents

More information

The Challenge of CT Dose Records

The Challenge of CT Dose Records The Challenge of CT Dose Records Kimberly Applegate, MD, MS, FACR Professor of Radiology and Pediatrics Emory University *financial disclosures: -Springer Textbook contracts -AIM advisory board for patient

More information

Production of X-rays and Interactions of X-rays with Matter

Production of X-rays and Interactions of X-rays with Matter Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage

More information

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Alzen,, G., Beiderwellen,, K., Volz,, M., Berthold,, L.D.,

More information

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses.

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. 1 3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. (A) Any person may apply to the director for approval to conduct an educational program for

More information

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

Safety and Radiation Protection Office Working with X-Ray Equipment

Safety and Radiation Protection Office Working with X-Ray Equipment Safety and Radiation Protection Office Working with X-Ray Equipment Radiation Safety Rules May 2016 Radiation Safety Rules X-Ray Equipment General Rules for the protection of persons exposed to ionising

More information

Computed Tomography Radiation Safety Issues in Ontario

Computed Tomography Radiation Safety Issues in Ontario Computed Tomography Radiation Safety Issues in Ontario Healthcare Human Factors Group Centre for Global ehealth Innovation University Health Network Toronto, ON, Canada June 16, 2006 Table of Contents

More information

C A M P E P. Standards for Accreditation of Graduate Educational Programs in Medical Physics

C A M P E P. Standards for Accreditation of Graduate Educational Programs in Medical Physics Standards for Accreditation of Graduate Educational Programs in Medical Physics Preamble Revised July 2015 Medical Physics is a branch of physics that applies the concepts and principles of physics to

More information

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes. Radiologic Technologist Overview The Field - Specialty Areas - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Radiologic technologists

More information

Measuring Patient Exposure in Interventional Radiology

Measuring Patient Exposure in Interventional Radiology Measuring Patient Exposure in Interventional Radiology GAFCHROMIC XR-R, a Wide-Area Dosimetry Film to Assess Peak Skin Dose (PSD) Xiang Yu, Ph.D. Director, Advanced Materials International Specialty Products

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools 3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground

More information

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Performance evaluation and quality assurance of Varian enhanced dynamic wedges JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department

More information

3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?

3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC? Disclosures Understanding IAC CT Accreditation Serve as one of two AAPM representatives to IAC CT Board of Directors, Serve as IAC representative to AAPM for focus group on accreditations Provide 3-hour

More information

Update on ACR Digital Mammography QC Manual

Update on ACR Digital Mammography QC Manual Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual

More information

Diagnostic X-ray facilities as per quality control performances in Tanzania

Diagnostic X-ray facilities as per quality control performances in Tanzania JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 4, FALL 2006 Diagnostic X-ray facilities as per quality control performances in Tanzania Yesaya Y. Sungita, 1 Simon S.L. Mdoe, 1 and Peter

More information

CURRICULUM VITAE. Kalpana M. Kanal, Ph.D., DABR

CURRICULUM VITAE. Kalpana M. Kanal, Ph.D., DABR May 31, 2005 ADDRESS: PERSONAL: CURRICULUM VITAE Kalpana M. Kanal, Ph.D., DABR Department of Radiology University of Washington Medical Center 1959 NE Pacific Avenue Seattle, WA 98195-7115 (206) 616-1966

More information