LAB I. INTRODUCTION TO LAB EQUIPMENT

Size: px
Start display at page:

Download "LAB I. INTRODUCTION TO LAB EQUIPMENT"

Transcription

1 LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE In this lab you will learn how to properly operate the basic bench equipment used for characterizing active devices: 1. Oscilloscope (Agilent MSO6032A), 2. Source Measure Unit (SMU) (Keithley 2430), 3. Function generator Agilent 33220A, and a 4. Bread board. You will use these tools to characterize three simple resistive circuits, perform theoretical circuit analyses on them, analyze the results, and present your findings in a concise, organized lab report. 2. OVERVIEW The Background Information section in this lab manual describes the basic operations of each lab equipment. You are expected to learn these basic operations during lab, ideally before moving on to the Lab Procedure section. The lab procedure will test your comprehension of the background materials by asking you to build simple resistive circuits and use the bench equipment to characterize them. Information essential to your understanding of this lab: 1. Background Material Materials necessary for this experiment: 1. Standard bench equipment. 2. Two resistors: 3.3 kω and 5.1 kω. 3. Two 10:1 Oscilloscope Probes. 4. One RG58C/U Coaxial Cable. 5. Two Red & Black Test Lead Pair (Banana-Plug to Alligator-Clip.) 3. BACKGROUND INFORMATION 3.1 BREADBOARD BASICS Breadboards (aka. Solderless board, Prototype board) are simply a set of pre-wired interconnected strips that are accessible through periodically spaced hole in the board. Looking at Figure 1., you can identify which holes form an interconnected strip by the black lines connecting them. By plugging the lead of a component into a hole you will be connected to all the other components in that strip without permanently connecting them. This allows you to build, alter, and test your prototype circuits quickly. Lab I: Introduction to Lab Equipment Page 1

2 There are two basic types of strips. The first type is called connection strip, they typically take up most of the board and are connected horizontally. Each hole can uniquely identified using the labels a-j column labels and 1-63 row labels. NOTE: a-e connection strips are not connected to the f-i connection strips. The second type is called bus strip. ALL the holes in a bus strip are connected vertically. Bus strips are typically labeled A or B and are marked by a red or blue line along their length. Figure 1. A schematic diagram of the breadboard showing buses and strips. 3.2 KEITHLEY SOURCE MEASURE UNIT 2400 The Keithley SMU can be used as a voltage source, a current source, a voltmeter, or an ammeter. Examine Figures 2. & 3. below before moving on to studying the main functions of the Keithley SMU. Figure 2. Keithley SMU button descriptions. Lab I: Introduction to Lab Equipment Page 2

3 Figure 3. Front panel of Keithley 2430 SMU SET VOLTAGE/CURRENT SOURCE CONFIGURATION In order to use the Keithley SMU as a voltage source or a current source, you need to follow the steps given below. 1. Press the V or I button in the Source group. 2. Press the EDIT button (top left): The display value V src or I src should start blinking. If it is not blinking press the EDIT button again. 3. To set your source value, you need to use the following buttons: Select Range: These buttons are used to change the range of the source value by an order of magnitude (i.e. by a factor of 10). Select Digit: The Left and Right arrows in the EDIT group are used to select the digit you wish to alter. Select Number: The Up and Down arrows in the source group are used to change the digit value. Lab I: Introduction to Lab Equipment Page 3

4 Or you can enter the numbers directly using these buttons: 4. Once you set the value, press Enter COMPLIANCE (IMPORTANT!!!) Once you have set your source value, you need to set your compliance value. How do I set the compliance value? Press the Edit button twice. You will see a set of digits blink. Use the same buttons you used to set your source value above to set your compliance value. How do I determine compliance value? Use data sheets to determine the voltage and current limits of your component. Next, use your magical powers of electrical engineering (also known as the mystical art of circuit analysis ) to figure what voltage and current your component will experience. For example the average resistor is rated at a ¼ watt. If you put 1V across that resistor, you need to make sure as a good and employable electrical engineer that you don t put more that 0.25A through it. Therefore, if you set up the SMU as a voltage source delivering 1V to your resistor, your compliance value will be 250mA. What is compliance? Compliance is a safety feature incorporated in the Keithley SMU to protect your circuit components from unexpected high power of operation i.e. it prevents you from unexpectedly frying your circuit. It is a limiting factor input by the user. If you set up an SMU as a voltage source, you must also set the highest current value the SMU is allowed to provide to your circuit; this is called current clamping. If you set up an SMU as a current source, you must set the highest voltage value the SMU is allowed to provide to your circuit; this is called voltage clamping. Lab I: Introduction to Lab Equipment Page 4

5 On the screen, the compliance value is typically displayed to the right of the source value, and in this format: Cmpl: ma (assuming you set up a voltage source.) Once you have set up source and compliance for an SMU, you can push the ON/OFF button at the bottom right corner of the front panel to power your circuit. Check the compliance value in the display. If something blinks, there is a problem. If you turn on your SMU and your circuit attempts to draw more current than is allowed by your compliance value, the Cmpl: text will blink (ex. Cmpl: ma ; here bold text indicates blinking text). This is called breaking real compliance. To overcome this, you need to increase the compliance value or recheck your circuit setup. If the units portion of your compliance value blinks ( Cmpl: ma ), you broke range compliance. It means the compliance value you entered is well above the range of current values being drawn by your circuit. The actual current drawn is below the range of measurement of the SMU. You need to press the AUTO button to allow the Keithley to set the compliance value to some lower value VOLTMETER / AMMETER CONFIGURATION To configure the Keithley SMU as an Ammeter or a Voltmeter, do the following: Voltmeter Setup: 1. Set the SMU up as a current source with zero output current. 2. Then from the control panel area, press the V button in the MEAS group under the display. Ammeter instructions 1. Set the SMU up as a voltage source with zero output voltage. 2. Then from the control panel area, press the I button in the MEAS group under the display. 3.3 AGILENT MSO6000 SERIES OSCILLOSCOPE This section will instruct you on how to operate the Agilent 6000 Series Oscilloscope. Lab I: Introduction to Lab Equipment Page 5

6 Figure 4. Front Panel of Agilent 6000 Series Oscilloscope OSCILLOSCOPE FRONT PANEL CONTROL Intensity Control (2 in Fig. 4) Rotate clockwise to increase the display intensity; counterclockwise to decrease. You can vary the intensity control to bring out signal detail, much like an analog oscilloscope. Digital channel waveform intensity is not adjustable. Autoscale Key (24 in Fig. 4) When you press the Autoscale key the oscilloscope will quickly determine which channels have activity, and it will turn these channels on and scale them to display the input signals. Vertical Position Control (6 in Fig. 4) Use this knob to change the channel s vertical position on the display. There is one Vertical Position control for each channel. Channel On/Off Key (7 in Fig. 4) Use this key to switch the channel on or off, or to access the channel s menu in the soft-keys. There is one Channel On/Off key for each channel. Vertical Sensitivity (9 in Fig. 4) Use this knob to change the vertical sensitivity (gain) of the channel. Lab I: Introduction to Lab Equipment Page 6

7 AutoProbe Interface (10 in Fig. 4) When you connect a probe to the oscilloscope, the AutoProbe Interface attempts to determine the type of probe and set its parameters in the Probe menu accordingly. Horizontal Delay Control (18 in Fig. 4) When the oscilloscope is RUNning, this control lets you set the acquisition window relative to the trigger point. When the oscilloscope is STOPped, you can turn this knob to pan through the data horizontally. This lets you see the captured waveform before the trigger (turn the knob clockwise) or after the trigger (turn the knob counterclockwise). Horizontal Sweep Speed Control (20 in Fig. 4) Turn this knob to adjust the sweep speed. This will change the time per horizontal division on the display. When adjusted after the waveform has been acquired and the oscilloscope is stopped, this has the effect of stretching out or squeezing the waveform horizontally. Measure Keys (21 in Fig. 4) Press the Cursors key to switch on cursors that you can use for making measurements. Press the Quick Meas key to access a set of predefined measurements Entry Knob (23 in Fig. 4) The entry knob is used to select items from menus and to change values. Its function changes based upon which menu is displayed. Note that the curved arrow symbol above the entry knob illuminates whenever the entry knob can be used to select a value. Use the entry knob to select among the choices that are shown on the softkeys. Softkeys (25 in Fig. 4) The functions of these keys change based upon the menus shown on the display directly above the keys. Probe Attenuation Factor: Some Oscilloscope probe attenuates the incoming signal by a certain factor. In this lab, we use 10:1 probe which attenuates the incoming signal by a factor of 10. By matching the attenuation factor of the oscilloscope to the attenuation of the probe, your measurements will reflect the actual voltage levels at the probe tip. Basic 10:1 probe calibration procedure: 1. Press the Save/Recall key on the front panel and then press the Default Setup Softkey (located directly below the display on the front panel). The oscilloscope is now configured to its default settings. 2. Select a probe and set the physical switch on the lead to 10x. 3. Connect the probe to channel 1, press the (1) button. 4. Ensure the resistance is set at 1Mohm on screen. 5. Press the softkey marked above Probe. Set the probe attenuation factor to 10:1. 6. Connect the oscilloscope probe to the Probe Comp signal terminal on the front panel. Lab I: Introduction to Lab Equipment Page 7

8 7. Connect the probe s ground lead to the ground terminal next to the Probe Comp terminal. 8. Press Autoscale. 9. You should see a waveform on the oscilloscope s display similar to this: 6. If the square wave form is not shaped correctly, it implies that the probes are not compensated to match to the oscilloscope s channels. In order to compensate the probe, you should adjust the trimmer capacitor of the probe (located on the probe BNC connector) for the flattest pulse possible using the plastic flathead driver provided with the probe MEASURING VOLTAGES AND TIME-RELATED PARAMETERS When measuring voltages with the oscilloscope, place the probes in parallel across the component where the voltage signal is being measured. Once you have the signal displayed on the screen, you can use buttons and keys to do the measurements. To measure RMS, DC, or peak to peak voltages with the oscilloscope, use the following method: Press the Quick Meas button on the Measure keys section (21 in Fig. 4). The Select menu appears on the bottom of the screen. Press the button beneath that, or use the Entry Knob (23 in Fig. 3) to select the desired value like RMS, Amplitude, Average, Peak to peak etc. The selected value would be displayed on the bottom of the display. To measure Frequency, period and other time-related parameters with the oscilloscope, use the following method: Press the Quick Meas button on the Measure keys section (21 in Fig. 4). The Select menu appears on the bottom of the screen. Press the button beneath that, or use the Entry Knob (23 in Fig. 3) to select the desired value Frequency, delay, period, Duty cycle etc. The selected value would be displayed on the bottom of the display. For other measurements related to the voltage and time-related parameters, we use cursors. To measure using the cursors do the following: Cursors are horizontal and vertical markers that indicate X-axis Lab I: Introduction to Lab Equipment Page 8

9 values (usually time) and Y-axis values (usually voltage) on a selected waveform source. The position of the cursors can be moved turning the Entry knob. When you press the Cursors key, it will illuminate and the cursors will turn on. To turn cursors off, press this key again until it is not illuminated, or press the Quick Meas key. Cursors are not always limited to the visible display. If you set a cursor, then pan and zoom the waveform until the cursor is off screen, its value will not be changed, and if you pan the waveform back again it will have the cursor in the original place. The following steps guide you through the front-panel Cursors key. You can use the cursors to make custom voltage or time measurements on the signal. 1. Connect a signal to the oscilloscope and obtain a stable display. 2. Press the Cursors key. View the cursor functions in the softkey menu: Mode Sets the cursors to measure voltage and time (Normal), or displays the binary or hexadecimal logic value of the displayed waveforms. Source Selects a channel or math function for the cursor measurements. X Y Selects either the X cursors or the Y cursors for adjustment with the Entry knob. X1 and X2 Adjust horizontally and normally measure time. Y1 and Y2 Adjust vertically and normally measure voltage. X1 X2 and Y1 Y2 Move the cursors together when turning the Entry knob. Lab I: Introduction to Lab Equipment Page 9

10 Cursor Examples 1. Cursors measure pulse widths other than middle threshold points 2. Cursors measure frequency of pulse ringing MEASURING CURRENTS The Oscilloscope can only measure current indirectly, by reading the voltage across a resistor while it is in a circuit and then applying Ohm s Law to find the current. If you have two signals and want to find the phase between similar points select the source of measurement for cursor 1 as channel 1 and the source for cursor 2 as channel 2. The difference readout is the delay between the two signals. If you divide that delay by the period then you have the phase value as a fraction of 360, or 2π radians. If you would like to represent that in degrees all you have to do is convert it from radians to degrees. Lab I: Introduction to Lab Equipment Page 10

11 3.4. FUNCTION GENERATOR AGILENT 33220A The function generator is used to generate signals for your circuits. You will need to know how to set the function generator to get sine, square, triangle or ramp signals. In addition, you will have to set up the frequency, the amplitude, offset voltage and the duty cycle. The default settings for this instrument are a sinewave of 1 khz, with an amplitude of 100 mv and a DC offset of 0.0 V. Figure 5. Front panel of the Agilent 33220A function generator. The function generator is very easy to use since each function has a specific button. If you want to select a waveform, just look for the button with the desired waveform such as a sine wave, a square wave, triangle wave, or ramp wave. Then, just press its button. All that you have to do now is set the parameters for the waveform. To set the frequency, amplitude, offset or the duty cycle you need to do the following: 1. Press the appropriate gray buttons beneath the display screen (Freq/Period, Ampl/Hi Level, Offset/Lo Level, or Duty Cycle). 2. You may enter the value one of two ways. a.) Turn the knob and the highlighted digit will change. You may select a different digit by using the < or the > buttons. b.) You can also key in the digit by using number buttons. 3. Press Output button on the bottom right of the front panel (right next to Sync cable) and make sure the light is on. Lab I: Introduction to Lab Equipment Page 11

12 IMPEDANCE MATCHING (IMPORTANT!) In order to make sure you read the exact value of the amplitude output by the function generator, You should make sure the output impedance of the function generator is matched to the impedance of the connected circuits. This function generator has 50 Ω output impedance. It has been configured by the manufacturer to deliver the voltage signal when a load of 50 Ω is attached to it. In the case of large impedance circuits the function generator may deliver up to twice the voltage that you have set it up to deliver. In our case, we use a series connected 5.1 kω resistor and 3.3 k Ω resistor, which is much higher than 50 Ω. Hence, when you set 1 V pp on the function generator, you will observe twice the amplitude (2 V pp ) on the oscilloscope. In order to overcome this, you need to set the function generator to have High Z output impedance. To do this, press the Utility button and press the output setup and you can change the output impedance to the High Z output mode. 4. PREPARATION There is no preparation for this lab except for reading and learning the background material. 5. PROCEDURE Before proceeding with the lab, please familiarize yourself with setting up the bench equipment. Refer to Section 3 for details. 5.1 FUNCTION GENERATOR AND OSCILLOSCOPE Use the function generator and the oscilloscope to perform the following tasks. 1. Build circuit A shown below in Figure Set the function generator to generate a sinusoidal signal with a frequency of 100 Hz and peak-to-peak voltage of 5V. 3. Set up one probe across the whole circuit, and another across R2. 4. Subtract Channel 2 signals from Channel 1 signals using the Oscilloscope. 5. Measure the voltages and time related parameters asked for on the Instructor Verification Sheet. Obtain TA Signature. Lab I: Introduction to Lab Equipment Page 12

13 Figure 6. Circuit A 5.2 KEITHLEY SMU Use the two Keithley SMUs to perform the following tasks: Using the circuit B of Figure 7 set up a Keithley SMU as a voltage source of 10 V DC. Figure out the compliance by evaluating the circuit. Use the second Keithley SMU to measure the voltages in R1 and R2. Measure the current in the circuit directly from the Keithley SMU used as the voltage source. Record values on IV sheet Using the circuit C of Figure 8, set up a Keithley SMU as a current source of 5 ma DC. Set up the other Keithley to measure the current in R1 and in R2. Record Values on IV Sheet Measure the impedance of your two resistors using the Ohmmeter setting of a Keithley SMU. Record the values in the Instructor Verification Sheet. Get TA Signature. Figure 7. Circuit 'B' Figure 8. Circuit 'C' Lab I: Introduction to Lab Equipment Page 13

14 Notice that Circuits B and C are source transforms of each other. You should be able to compare and contrast the voltage and current measurements. 6. LAB REPORT Type a lab report with a cover sheet containing your name, class (including section number), date of the lab, and the report due date. Use the following outline to draft sections of your lab report: Abstract: Briefly describe the purpose of the lab, the analysis you performed, and your findings. Introduction: Briefly mention the bench equipment you used in the lab and their basic functions in your own words. Procedure: You do not need to provide a procedure section for this lab. Data Presentation: Report all the measured data collected. Make sure it is well presented, has units and labels - and is easily discernable which values are from a particular section of the procedure. Please use Excel, Matlab or another software to help generate well-formed tables. Analysis: Perform theoretical circuit analysis on each circuit you characterized i.e. use the measured values of your resistors (5.2.3) to find the theoretical voltage and current values for circuit A, B, and C. Do show work typed equations, units etc. Include the circuit diagrams in your descriptions, if needed. Compare your calculated values to the measured values using percent error calculations. Be sure to organize your analyses appropriately according to procedure section number. Conclusions: What conclusions can you draw about using bench equipment from your direct experience of setting it up and using it to characterize circuits? What do the results of your circuit analyses tell you about your bench equipment? Lab I: Introduction to Lab Equipment Page 14

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

Oscilloscope, Function Generator, and Voltage Division

Oscilloscope, Function Generator, and Voltage Division 1. Introduction Oscilloscope, Function Generator, and Voltage Division In this lab the student will learn to use the oscilloscope and function generator. The student will also verify the concept of voltage

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

Lab 1: The Digital Oscilloscope

Lab 1: The Digital Oscilloscope PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix

More information

RC Circuits and The Oscilloscope Physics Lab X

RC Circuits and The Oscilloscope Physics Lab X Objective RC Circuits and The Oscilloscope Physics Lab X In this series of experiments, the time constant of an RC circuit will be measured experimentally and compared with the theoretical expression for

More information

Experiment 2 Diode Applications: Rectifiers

Experiment 2 Diode Applications: Rectifiers ECE 3550 - Practicum Fall 2007 Experiment 2 Diode Applications: Rectifiers Objectives 1. To investigate the characteristics of half-wave and full-wave rectifier circuits. 2. To recognize the usefulness

More information

Beginners Guide to the TDS 210 and TDS 220 Oscilloscopes

Beginners Guide to the TDS 210 and TDS 220 Oscilloscopes Beginners Guide to the TDS 210 and TDS 220 Oscilloscopes By David S. Lay P. Eng Foreword This guide contains information to help you become familiar with using digital oscilloscopes. You should work through

More information

Lab E1: Introduction to Circuits

Lab E1: Introduction to Circuits E1.1 Lab E1: Introduction to Circuits The purpose of the this lab is to introduce you to some basic instrumentation used in electrical circuits. You will learn to use a DC power supply, a digital multimeter

More information

Fundamentals of Signature Analysis

Fundamentals of Signature Analysis Fundamentals of Signature Analysis An In-depth Overview of Power-off Testing Using Analog Signature Analysis www.huntron.com 1 www.huntron.com 2 Table of Contents SECTION 1. INTRODUCTION... 7 PURPOSE...

More information

ε: Voltage output of Signal Generator (also called the Source voltage or Applied

ε: Voltage output of Signal Generator (also called the Source voltage or Applied Experiment #10: LR & RC Circuits Frequency Response EQUIPMENT NEEDED Science Workshop Interface Power Amplifier (2) Voltage Sensor graph paper (optional) (3) Patch Cords Decade resistor, capacitor, and

More information

Lab 3 Rectifier Circuits

Lab 3 Rectifier Circuits ECET 242 Electronic Circuits Lab 3 Rectifier Circuits Page 1 of 5 Name: Objective: Students successfully completing this lab exercise will accomplish the following objectives: 1. Learn how to construct

More information

Annex: VISIR Remote Laboratory

Annex: VISIR Remote Laboratory Open Learning Approach with Remote Experiments 518987-LLP-1-2011-1-ES-KA3-KA3MP Multilateral Projects UNIVERSITY OF DEUSTO Annex: VISIR Remote Laboratory OLAREX project report Olga Dziabenko, Unai Hernandez

More information

Inductors in AC Circuits

Inductors in AC Circuits Inductors in AC Circuits Name Section Resistors, inductors, and capacitors all have the effect of modifying the size of the current in an AC circuit and the time at which the current reaches its maximum

More information

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER 2014 Amplifier - 1 FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER The objectives of this experiment are: To understand the concept of HI-FI audio equipment To generate a frequency response curve for an audio

More information

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz 1 Sine wave with a DC offset f = frequency in Hz A = DC offset voltage (average voltage) B = Sine amplitude Vpp = 2B Vmax = A +

More information

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b DIODE CIRCUITS LABORATORY A solid state diode consists of a junction of either dissimilar semiconductors (pn junction diode) or a metal and a semiconductor (Schottky barrier diode). Regardless of the type,

More information

Lab 3 - DC Circuits and Ohm s Law

Lab 3 - DC Circuits and Ohm s Law Lab 3 DC Circuits and Ohm s Law L3-1 Name Date Partners Lab 3 - DC Circuits and Ohm s Law OBJECTIES To learn to apply the concept of potential difference (voltage) to explain the action of a battery in

More information

EXPERIMENT NUMBER 8 CAPACITOR CURRENT-VOLTAGE RELATIONSHIP

EXPERIMENT NUMBER 8 CAPACITOR CURRENT-VOLTAGE RELATIONSHIP 1 EXPERIMENT NUMBER 8 CAPACITOR CURRENT-VOLTAGE RELATIONSHIP Purpose: To demonstrate the relationship between the voltage and current of a capacitor. Theory: A capacitor is a linear circuit element whose

More information

RLC Series Resonance

RLC Series Resonance RLC Series Resonance 11EM Object: The purpose of this laboratory activity is to study resonance in a resistor-inductor-capacitor (RLC) circuit by examining the current through the circuit as a function

More information

Figure 1: Multiple unsynchronized snapshots of the same sinusoidal signal.

Figure 1: Multiple unsynchronized snapshots of the same sinusoidal signal. 1 Oscilloscope Guide Introduction An oscilloscope is a device used to observe and measure time-dependent electronic signals. It is essentially an enhanced voltmeter which displays a graph of potential

More information

Experiment #11: LRC Circuit (Power Amplifier, Voltage Sensor)

Experiment #11: LRC Circuit (Power Amplifier, Voltage Sensor) Experiment #11: LRC Circuit (Power Amplifier, Voltage Sensor) Concept: circuits Time: 30 m SW Interface: 750 Windows file: RLC.SWS EQUIPMENT NEEDED Science Workshop Interface Power Amplifier (2) Voltage

More information

CHAPTER 11: Flip Flops

CHAPTER 11: Flip Flops CHAPTER 11: Flip Flops In this chapter, you will be building the part of the circuit that controls the command sequencing. The required circuit must operate the counter and the memory chip. When the teach

More information

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change

More information

Electronic WorkBench tutorial

Electronic WorkBench tutorial Electronic WorkBench tutorial Introduction Electronic WorkBench (EWB) is a simulation package for electronic circuits. It allows you to design and analyze circuits without using breadboards, real components

More information

DCM555 - Data Communications Lab 8 Time Division Multiplexing (TDM) Part 1 - T1/DS1 Signals

DCM555 - Data Communications Lab 8 Time Division Multiplexing (TDM) Part 1 - T1/DS1 Signals DCM555 - Data Communications Lab 8 Time Division Multiplexing (TDM) Part 1 - T1/DS1 Signals Name: St. #: Section: (Note: Show all of your calculations, express your answer to the appropriate number of

More information

Electronics. Discrete assembly of an operational amplifier as a transistor circuit. LD Physics Leaflets P4.2.1.1

Electronics. Discrete assembly of an operational amplifier as a transistor circuit. LD Physics Leaflets P4.2.1.1 Electronics Operational Amplifier Internal design of an operational amplifier LD Physics Leaflets Discrete assembly of an operational amplifier as a transistor circuit P4.2.1.1 Objects of the experiment

More information

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS

LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS LAB 7 MOSFET CHARACTERISTICS AND APPLICATIONS Objective In this experiment you will study the i-v characteristics of an MOS transistor. You will use the MOSFET as a variable resistor and as a switch. BACKGROUND

More information

Physics 3330 Experiment #2 Fall 1999. DC techniques, dividers, and bridges R 2 =(1-S)R P R 1 =SR P. R P =10kΩ 10-turn pot.

Physics 3330 Experiment #2 Fall 1999. DC techniques, dividers, and bridges R 2 =(1-S)R P R 1 =SR P. R P =10kΩ 10-turn pot. Physics 3330 Experiment #2 Fall 1999 DC techniques, dividers, and bridges Purpose You will gain a familiarity with the circuit board and work with a variety of DC techniques, including voltage dividers,

More information

Lab Exercise 1: Acoustic Waves

Lab Exercise 1: Acoustic Waves Lab Exercise 1: Acoustic Waves Contents 1-1 PRE-LAB ASSIGNMENT................. 2 1-3.1 Spreading Factor: Spherical Waves........ 2 1-3.2 Interference In 3-D................. 3 1-4 EQUIPMENT........................

More information

MATERIALS. Multisim screen shots sent to TA.

MATERIALS. Multisim screen shots sent to TA. Page 1/8 Revision 0 9-Jun-10 OBJECTIVES Learn new Multisim components and instruments. Conduct a Multisim transient analysis. Gain proficiency in the function generator and oscilloscope. MATERIALS Multisim

More information

22.302 Experiment 5. Strain Gage Measurements

22.302 Experiment 5. Strain Gage Measurements 22.302 Experiment 5 Strain Gage Measurements Introduction The design of components for many engineering systems is based on the application of theoretical models. The accuracy of these models can be verified

More information

User Manual. TDS 200-Series Digital Real-Time Oscilloscope 071-0398-03

User Manual. TDS 200-Series Digital Real-Time Oscilloscope 071-0398-03 User Manual TDS 200-Series Digital Real-Time Oscilloscope 071-0398-03 This document supports firmware version FV:v1.00 and above. www.tektronix.com Copyright Tektronix, Inc. All rights reserved. Tektronix

More information

Agilent U1610/20A Handheld Digital Oscilloscope Quick Start Guide

Agilent U1610/20A Handheld Digital Oscilloscope Quick Start Guide Agilent U1610/20A Handheld Digital Oscilloscope Quick Start Guide Verify that you received the following items in the shipment of your handheld scope: 1 Power cable 1 Li-Ion battery pack, 10.8 V (included

More information

Step Response of RC Circuits

Step Response of RC Circuits Step Response of RC Circuits 1. OBJECTIVES...2 2. REFERENCE...2 3. CIRCUITS...2 4. COMPONENTS AND SPECIFICATIONS...3 QUANTITY...3 DESCRIPTION...3 COMMENTS...3 5. DISCUSSION...3 5.1 SOURCE RESISTANCE...3

More information

Lab 4 - Data Acquisition

Lab 4 - Data Acquisition Spring 11 Lab 4 - Data Acquisition Lab 4-1 Lab 4 - Data Acquisition Format This lab will be conducted during your regularly scheduled lab time in a group format. Each student is responsible for learning

More information

Physics 120 Lab 6: Field Effect Transistors - Ohmic region

Physics 120 Lab 6: Field Effect Transistors - Ohmic region Physics 120 Lab 6: Field Effect Transistors - Ohmic region The FET can be used in two extreme ways. One is as a voltage controlled resistance, in the so called "Ohmic" region, for which V DS < V GS - V

More information

Reading assignment: All students should read the Appendix about using oscilloscopes.

Reading assignment: All students should read the Appendix about using oscilloscopes. 10. A ircuits* Objective: To learn how to analyze current and voltage relationships in alternating current (a.c.) circuits. You will use the method of phasors, or the vector addition of rotating vectors

More information

Lab 3: Introduction to Data Acquisition Cards

Lab 3: Introduction to Data Acquisition Cards Lab 3: Introduction to Data Acquisition Cards INTRODUCTION: In this lab, you will be building a VI to display the input measured on a channel. However, within your own VI you will use LabVIEW supplied

More information

Laboratory 4: Feedback and Compensation

Laboratory 4: Feedback and Compensation Laboratory 4: Feedback and Compensation To be performed during Week 9 (Oct. 20-24) and Week 10 (Oct. 27-31) Due Week 11 (Nov. 3-7) 1 Pre-Lab This Pre-Lab should be completed before attending your regular

More information

Dash 18X / Dash 18 Data Acquisition Recorder

Dash 18X / Dash 18 Data Acquisition Recorder 75 Dash 18X / Dash 18 Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 3.1 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH

More information

AC CIRCUITS - CAPACITORS AND INDUCTORS

AC CIRCUITS - CAPACITORS AND INDUCTORS EXPRIMENT#8 AC CIRCUITS - CAPACITORS AND INDUCTORS NOTE: Two weeks are allocated for this experiment. Before performing this experiment, review the Proper Oscilloscope Use section of Experiment #7. Objective

More information

Objectives: Part 1: Build a simple power supply. CS99S Laboratory 1

Objectives: Part 1: Build a simple power supply. CS99S Laboratory 1 CS99S Laboratory 1 Objectives: 1. Become familiar with the breadboard 2. Build a logic power supply 3. Use switches to make 1s and 0s 4. Use LEDs to observe 1s and 0s 5. Make a simple oscillator 6. Use

More information

ECEN 1400, Introduction to Analog and Digital Electronics

ECEN 1400, Introduction to Analog and Digital Electronics ECEN 1400, Introduction to Analog and Digital Electronics Lab 4: Power supply 1 INTRODUCTION This lab will span two lab periods. In this lab, you will create the power supply that transforms the AC wall

More information

[F/T] [5] [KHz] [AMP] [3] [V] 4 ) To set DC offset to -2.5V press the following keys [OFS] [+/-] [2] [.] [5] [V]

[F/T] [5] [KHz] [AMP] [3] [V] 4 ) To set DC offset to -2.5V press the following keys [OFS] [+/-] [2] [.] [5] [V] FG085 minidds Function Generator Manual of Operation Applicable Models: 08501, 08501K, 08502K, 08503, 08503K Applicable Firmware Version: 1 ) 113-08501-100 or later (for U5) 2 ) 113-08502-030 or later

More information

GENERAL SCIENCE LABORATORY 1110L Lab Experiment 6: Ohm s Law

GENERAL SCIENCE LABORATORY 1110L Lab Experiment 6: Ohm s Law GENERAL SCIENCE LABORATORY 1110L Lab Experiment 6: Ohm s Law OBJECTIVES: To verify Ohm s law, the mathematical relationship among current, voltage or potential difference, and resistance, in a simple circuit.

More information

Lab 1: Introduction to PSpice

Lab 1: Introduction to PSpice Lab 1: Introduction to PSpice Objectives A primary purpose of this lab is for you to become familiar with the use of PSpice and to learn to use it to assist you in the analysis of circuits. The software

More information

Using an Oscilloscope

Using an Oscilloscope Using an Oscilloscope The oscilloscope is used to measure a voltage that changes in time. It has two probes, like a voltmeter. You put these probes on either side of the thing that you want to measure

More information

RF Measurements Using a Modular Digitizer

RF Measurements Using a Modular Digitizer RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.

More information

Measuring Electric Phenomena: the Ammeter and Voltmeter

Measuring Electric Phenomena: the Ammeter and Voltmeter Measuring Electric Phenomena: the Ammeter and Voltmeter 1 Objectives 1. To understand the use and operation of the Ammeter and Voltmeter in a simple direct current circuit, and 2. To verify Ohm s Law for

More information

TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin

TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin (Updated 7/19/08 to delete sine wave output) I constructed the 1 MHz square wave generator shown in the Appendix. This

More information

Lab 5 Operational Amplifiers

Lab 5 Operational Amplifiers Lab 5 Operational Amplifiers By: Gary A. Ybarra Christopher E. Cramer Duke University Department of Electrical and Computer Engineering Durham, NC. Purpose The purpose of this lab is to examine the properties

More information

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON. WHY DIFFERENTIAL? Voltage, The Difference Whether aware of it or not, a person using an oscilloscope to make any voltage measurement is actually making a differential voltage measurement. By definition,

More information

EE 242 EXPERIMENT 5: COMPUTER SIMULATION OF THREE-PHASE CIRCUITS USING PSPICE SCHEMATICS 1

EE 242 EXPERIMENT 5: COMPUTER SIMULATION OF THREE-PHASE CIRCUITS USING PSPICE SCHEMATICS 1 EE 242 EXPERIMENT 5: COMPUTER SIMULATION OF THREE-PHASE CIRCUITS USING PSPICE SCHEMATICS 1 Objective: To build, simulate, and analyze three-phase circuits using OrCAD Capture Pspice Schematics under balanced

More information

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT 4 Understand single-phase alternating current (ac) theory Single phase AC

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2006 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

Frequency Response of Filters

Frequency Response of Filters School of Engineering Department of Electrical and Computer Engineering 332:224 Principles of Electrical Engineering II Laboratory Experiment 2 Frequency Response of Filters 1 Introduction Objectives To

More information

Agilent InfiniiVision 3000 X-Series Oscilloscopes. User's Guide

Agilent InfiniiVision 3000 X-Series Oscilloscopes. User's Guide Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide s1 Notices Agilent Technologies, Inc. 2005-2013 No part of this manual may be reproduced in any form or by any means (including electronic

More information

PropScope USB Oscilloscope

PropScope USB Oscilloscope USB Oscilloscope v1.0 December 2009 Manual by Hanno Sander 3 Table of Contents ForeWord... Part I Welcome... 4 Part II Getting... Started 6 2.1 Installation... 7 2.2 8 Connect...

More information

RC & RL Transient Response

RC & RL Transient Response EE 2006 University of Minnesota Duluth ab 8 1. Introduction R & R Transient Response The student will analyze series R and R circuits. A step input will excite these respective circuits, producing a transient

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2009 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

Dash 8Xe / Dash 8X Data Acquisition Recorder

Dash 8Xe / Dash 8X Data Acquisition Recorder 75 Dash 8Xe / Dash 8X Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 2.0 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH

More information

Lab 1: DC Circuits. Student 1, student1@ufl.edu Partner : Student 2, student2@ufl.edu

Lab 1: DC Circuits. Student 1, student1@ufl.edu Partner : Student 2, student2@ufl.edu Lab Date Lab 1: DC Circuits Student 1, student1@ufl.edu Partner : Student 2, student2@ufl.edu I. Introduction The purpose of this lab is to allow the students to become comfortable with the use of lab

More information

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE 46 2007. Test Method for AC to DC Power Supplies

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE 46 2007. Test Method for AC to DC Power Supplies ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE 46 2007 Test Method for AC to DC Power Supplies NOTICE The Society of Cable Telecommunications Engineers (SCTE)

More information

Essential Electrical Concepts

Essential Electrical Concepts Essential Electrical Concepts Introduction Modern vehicles incorporate many electrical and electronic components and systems: Audio Lights Navigation Engine control Transmission control Braking and traction

More information

Lab #9: AC Steady State Analysis

Lab #9: AC Steady State Analysis Theory & Introduction Lab #9: AC Steady State Analysis Goals for Lab #9 The main goal for lab 9 is to make the students familar with AC steady state analysis, db scale and the NI ELVIS frequency analyzer.

More information

Experiment1: Introduction to laboratory equipment and basic components.

Experiment1: Introduction to laboratory equipment and basic components. Experiment1: Introduction to laboratory equipment and basic components. 1 OBJECTIVES. This experiment will provide exposure to the various test equipment to be used in subsequent experiments. A primary

More information

Op-Amp Simulation EE/CS 5720/6720. Read Chapter 5 in Johns & Martin before you begin this assignment.

Op-Amp Simulation EE/CS 5720/6720. Read Chapter 5 in Johns & Martin before you begin this assignment. Op-Amp Simulation EE/CS 5720/6720 Read Chapter 5 in Johns & Martin before you begin this assignment. This assignment will take you through the simulation and basic characterization of a simple operational

More information

RGB for ZX Spectrum 128, +2, +2A, +3

RGB for ZX Spectrum 128, +2, +2A, +3 RGB for ZX Spectrum 128, +2, +2A, +3 Introduction... 2 Video Circuitry... 3 Audio Circuitry... 8 Lead Wiring... 9 Testing The Lead... 11 Spectrum +2A/+3 RGB Differences... 12 Circuitry Calculations...

More information

Numerical Parameters Analysis of Boonton 4540 Peak Power Meter

Numerical Parameters Analysis of Boonton 4540 Peak Power Meter Application Note Numerical Parameters Analysis of Boonton 4540 Peak Power Meter Mazumder Alam Product Marketing Manager, Boonton Electronics Introduction The Boonton 4540 series RF peak power meters consisting

More information

Fig. 1 Analogue Multimeter Fig.2 Digital Multimeter

Fig. 1 Analogue Multimeter Fig.2 Digital Multimeter ELECTRICAL INSTRUMENT AND MEASUREMENT Electrical measuring instruments are devices used to measure electrical quantities such as electric current, voltage, resistance, electrical power and energy. MULTIMETERS

More information

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law.

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law. 260 17-1 I. THEORY EXPERIMENT 17 QUALITATIVE STUDY OF INDUCED EMF Along the extended central axis of a bar magnet, the magnetic field vector B r, on the side nearer the North pole, points away from this

More information

School of Electrical and Information Engineering. The Remote Laboratory System. Electrical and Information Engineering Project 2006.

School of Electrical and Information Engineering. The Remote Laboratory System. Electrical and Information Engineering Project 2006. School of Electrical and Information Engineering The Remote Laboratory System Electrical and Information Engineering Project 2006 By Ben Loud Project Number: Supervisors: 2006A Jan Machotka Zorica Nedic

More information

LABORATORY 2 THE DIFFERENTIAL AMPLIFIER

LABORATORY 2 THE DIFFERENTIAL AMPLIFIER LABORATORY 2 THE DIFFERENTIAL AMPLIFIER OBJECTIVES 1. To understand how to amplify weak (small) signals in the presence of noise. 1. To understand how a differential amplifier rejects noise and common

More information

Electronics. Basic Concepts. Yrd. Doç. Dr. Aytaç GÖREN Yrd. Doç. Dr. Levent ÇETİN

Electronics. Basic Concepts. Yrd. Doç. Dr. Aytaç GÖREN Yrd. Doç. Dr. Levent ÇETİN Electronics Basic Concepts Electric charge Ordinary matter is made up of atoms which have positively charged nuclei and negatively charged electrons surrounding them. Charge is quantized as the subtraction

More information

Measuring Impedance and Frequency Response of Guitar Pickups

Measuring Impedance and Frequency Response of Guitar Pickups Measuring Impedance and Frequency Response of Guitar Pickups Peter D. Hiscocks Syscomp Electronic Design Limited phiscock@ee.ryerson.ca www.syscompdesign.com April 30, 2011 Introduction The CircuitGear

More information

Experiment 8 : Pulse Width Modulation

Experiment 8 : Pulse Width Modulation Name/NetID: Teammate/NetID: Experiment 8 : Pulse Width Modulation Laboratory Outline In experiment 5 we learned how to control the speed of a DC motor using a variable resistor. This week, we will learn

More information

Dayton Audio is proud to introduce DATS V2, the best tool ever for accurately measuring loudspeaker driver parameters in seconds.

Dayton Audio is proud to introduce DATS V2, the best tool ever for accurately measuring loudspeaker driver parameters in seconds. Dayton Audio is proud to introduce DATS V2, the best tool ever for accurately measuring loudspeaker driver parameters in seconds. DATS V2 is the latest edition of the Dayton Audio Test System. The original

More information

SERIES-PARALLEL DC CIRCUITS

SERIES-PARALLEL DC CIRCUITS Name: Date: Course and Section: Instructor: EXPERIMENT 1 SERIES-PARALLEL DC CIRCUITS OBJECTIVES 1. Test the theoretical analysis of series-parallel networks through direct measurements. 2. Improve skills

More information

The RIDZ 8x2 Audio Switcher

The RIDZ 8x2 Audio Switcher The RIDZ 8x2 Audio Switcher Engineering Manual Support Number 800-765-2930 International 712-852-2813 Table of Contents General Information for the RIDZ (8 x 2) Switcher..... 3 Input 9 on the RIDZ Switcher....6

More information

Automotive Sensor Simulator. Automotive sensor simulator. Operating manual. AutoSim

Automotive Sensor Simulator. Automotive sensor simulator. Operating manual. AutoSim Automotive sensor simulator Operating manual AutoSim Contents Introduction.. page 3 Technical specifications.... page 4 Typical application of AutoSim simulator..... page 4 Device appearance... page 5

More information

EE101 Labs and ECEbot Assembly/Testing Instructions

EE101 Labs and ECEbot Assembly/Testing Instructions EE101 Labs and ECEbot Assembly/Testing Instructions by Montana State University Department of Electrical and Computer Engineering A Montana Space Grant Consortium Project December 3, 2008 Beginning in

More information

EE 1202 Experiment #4 Capacitors, Inductors, and Transient Circuits

EE 1202 Experiment #4 Capacitors, Inductors, and Transient Circuits EE 1202 Experiment #4 Capacitors, Inductors, and Transient Circuits 1. Introduction and Goal: Exploring transient behavior due to inductors and capacitors in DC circuits; gaining experience with lab instruments.

More information

Output Ripple and Noise Measurement Methods for Ericsson Power Modules

Output Ripple and Noise Measurement Methods for Ericsson Power Modules Output Ripple and Noise Measurement Methods for Ericsson Power Modules Design Note 022 Ericsson Power Modules Ripple and Noise Abstract There is no industry-wide standard for measuring output ripple and

More information

U1602A Handheld Oscilloscopes, 20 MHz

U1602A Handheld Oscilloscopes, 20 MHz Products & Services Technical Support Buy Industries About Agilent United States Home >... > Oscilloscopes > U1600A Series handheld oscilloscopes (2 models) > U1602A Handheld Oscilloscopes, 20 MHz Product

More information

Ph 3504 Nuclear Magnetic Resonance and Electron Spin Resonance

Ph 3504 Nuclear Magnetic Resonance and Electron Spin Resonance Ph 3504 Nuclear Magnetic Resonance and Electron Spin Resonance Required background reading Tipler, Llewellyn, section 12-3 (you only need to read the part labeled Nuclear Magnetic Resonance on pages 596-597

More information

User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR

User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR Content General safety summary...1 Introduction...2 Chapter 1 Getting started...3 System Requirements...4 Installing Hardware...5 Installing Software...8

More information

DSO-1062D/DSO-1102D/DSO-1202D different bandwidths. Digital Oscilloscope User Manual

DSO-1062D/DSO-1102D/DSO-1202D different bandwidths. Digital Oscilloscope User Manual DSO-1062D/DSO-1102D/DSO-1202D different bandwidths Digital Oscilloscope User Manual Contents Contents Contents... i Copyright Declaration... iv Chapter 1 Safety Tips... 1 1.1 General Safety Summary...

More information

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12 Table of Contents The Basics of Electricity 2 Using a Digital Multimeter 4 IDEAL Digital Multimeters An Introduction The Basics of Digital Multimeters is designed to give you a fundamental knowledge of

More information

LABORATORY 10 TIME AVERAGES, RMS VALUES AND THE BRIDGE RECTIFIER. Bridge Rectifier

LABORATORY 10 TIME AVERAGES, RMS VALUES AND THE BRIDGE RECTIFIER. Bridge Rectifier LABORATORY 10 TIME AVERAGES, RMS VALUES AND THE BRIDGE RECTIFIER Full-wave Rectification: Bridge Rectifier For many electronic circuits, DC supply voltages are required but only AC voltages are available.

More information

TIMING SIGNALS, IRIG-B AND PULSES

TIMING SIGNALS, IRIG-B AND PULSES TIMING SIGNALS, IRIG-B AND PULSES Document No. PD0043200B July 2013 Arbiter Systems, Inc. 1324 Vendels Circle, Suite 121 Paso Robles, CA 93446 U.S.A. (805) 237-3831, (800) 321-3831 http://www.arbiter.com

More information

EXPERIMENT 7 OHM S LAW, RESISTORS IN SERIES AND PARALLEL

EXPERIMENT 7 OHM S LAW, RESISTORS IN SERIES AND PARALLEL 260 7- I. THEOY EXPEIMENT 7 OHM S LAW, ESISTOS IN SEIES AND PAALLEL The purposes of this experiment are to test Ohm's Law, to study resistors in series and parallel, and to learn the correct use of ammeters

More information

Your Multimeter. The Arduino Uno 10/1/2012. Using Your Arduino, Breadboard and Multimeter. EAS 199A Fall 2012. Work in teams of two!

Your Multimeter. The Arduino Uno 10/1/2012. Using Your Arduino, Breadboard and Multimeter. EAS 199A Fall 2012. Work in teams of two! Using Your Arduino, Breadboard and Multimeter Work in teams of two! EAS 199A Fall 2012 pincer clips good for working with breadboard wiring (push these onto probes) Your Multimeter probes leads Turn knob

More information

Agilent 54621A/22A/24A/41A/42A Oscilloscopes and Agilent 54621D/22D/41D/42D Mixed-Signal Oscilloscopes. User s Guide

Agilent 54621A/22A/24A/41A/42A Oscilloscopes and Agilent 54621D/22D/41D/42D Mixed-Signal Oscilloscopes. User s Guide User s Guide Publication Number 54622-97036 September 2002 For Safety Information and Regulatory information, see the pages behind the Index. Copyright Agilent Technologies 2000-2002 All Rights Reserved

More information

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 2-1 General Specifications...1 2-2 Electrical Specifications... 2 3. FRONT PANEL DESCRIPTION... 4 3-1

More information

= V peak 2 = 0.707V peak

= V peak 2 = 0.707V peak BASIC ELECTRONICS - RECTIFICATION AND FILTERING PURPOSE Suppose that you wanted to build a simple DC electronic power supply, which operated off of an AC input (e.g., something you might plug into a standard

More information

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION 1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.

More information

Capacitive Touch Sensor Project:

Capacitive Touch Sensor Project: NOTE: This project does not include a complete parts list. In particular, the IC described here does not come in a dual-inline-package (DIP), and so a gull-wing package has to be soldered to an adaptor

More information

Embest DSO2300 USB Oscilloscope

Embest DSO2300 USB Oscilloscope Embest DSO2300 USB Oscilloscope - 8-bit, 100Ms/s, 50MHz, 2-channel USB1.1/2.0 Compatible Digital Storage Oscilloscope - Multi-functions Including Logic Analyzer, Spectrum Analyzer (FFT), Record & Playback

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 2011 Interference - 1 INTERFERENCE OF SOUND WAVES The objectives of this experiment are: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves. To observe interference phenomena

More information