Comments Concerning the Implementation of 21 CFR Part 11
|
|
- Margery Holt
- 7 years ago
- Views:
Transcription
1 Introduction There currently is considerable controversy surrounding 21 CFR Part 11 with respect to the informal interpretations that have been presented by FDA personnel at industry conferences. The lack of a clear and consistent position by the FDA with respect to the regulation has placed many manufacturers in a difficult position with respect to compliance. Of particular concern on the part of these manufacturers is that compliance with the strict interpretation presented by individuals within the Agency may incur a considerable cost without commensurate benefit in terms of increased safety or enhanced product quality. These prescriptive interpretations are in conflict with the Part 11 Compliance Policy Guide (section ) that defines a results based policy that stresses the nature and extent of the deviation and effect on product quality and data integrity. We have addressed some of these controversial questions in this correspondence. Electronic Records and Part 11 Applicability The definition of an electronic record includes any electronic data, however, when that data is subject to compliance with electronic record requirements can be interpreted in several ways. This interpretation is significant with respect to defining when audit trails and additional Part 11 controls are required. To be consistent with predicate regulations, the definition of an electronic record must be based on the intended use of the application and not based on definitions such as the method, media, or duration of storage or transmission techniques. Interpretations that attempt to individually prescribe solutions for differing scenarios of use will make the regulation impractical to implement. Failure to interpret Part 11 based on intended use by introducing new definitions such as storage to durable medium, hybrid records, typewriter excuse, transient views, etc., serves only to complicate the requirements of the regulation and subject them to rapid obsolescence. The following references from Part 11 provide the manufacturer with the responsibility to define the applicability of Part 11 to their quality system. a. Part 11 section 11.1 states: 11.1 (a) The regulations in this part set forth the criteria under which the agency considers electronic records, electronic signatures, and handwritten signatures executed to electronic records to be trustworthy, reliable, and generally equivalent to paper records and handwritten signatures executed on paper. The regulation states how the manufacturer may elect to use electronic records and does not mandate compliance with all elements of Part 11 in every instance where computerized systems are used. Part 11 does not provide authority to the agency to redefine document control requirements of the predicate regulations (21 CFR 820, 21 Page 1 of 5
2 CFR 211, etc.) that have required document review and approval and change control but not audit trails of all changes made during document revisions. b. In addition, in the preamble of the Federal Register Volume 62, Number 54, of Thursday March 20, 1997, section III Comments on the Proposed rule, subelement I Effective Date/Grandfathering, fourth paragraph states the following: The agency emphasizes that these regulations do not require, but rather permit, the use of electronic records and electronic signatures. Firms not confident that their electronic systems meet the minimal requirements of these regulations are free to continue to use traditional signatures and paper documents to meet record keeping requirements. In accordance with this statement, the firm has the ability to control document drafts without audit trails for all changes and then to allow access to the same documents once approved in a controlled electronic environment that ensures the integrity of the document in an adequately validated system. c) The Compliance Policy Guide section (Enforcement Policy: 21 CFR Part 11) states: Part 11 applies to all FDA program areas, but does not mandate electronic record keeping. Part 11 describes the technical and procedural requirements that must be met if a person chooses to maintain records electronically and use electronic signatures. Once again the emphasis is on the use of electronic records in place of paper records and not on the redefinition of record keeping requirements based on Part 11 interpretation. The ability to support changes to documents without maintaining the audit trail of all changes is certainly allowed under the predicate regulations as long as the change history is established. Electronic Record Definition The following are suggested clarifications to the definition of electronic records and the scope of when automated audit trails are required. 1. Manufacturers can define how electronic equipment is used to support the predicate regulations and therefore whether the data must be retained as an electronic record in support of the quality system or paper based record. All electronic equipment used in manufacturing need not be controlled in accordance with Part Raw data from an instrument that cannot be modified based on security and procedural controls need not have secure, computer-generated time-stamped audit trail (11.10(e)) functions implemented. Audit trails for these instruments can be established based on demonstrating the effectiveness of the security and procedural controls that prevent access to modification of run data. Audit trails for these systems can be established by demonstrating that each run generates a unique record that is archived as read-only data. Page 2 of 5
3 3. Even when a manufacturer is using electronic records, the applicability of audit trails requires clarification. Audit trails should be defined in two different categories: a) The first category is raw data that is captured by equipment or manually entered based on observed information and can be changed on-line requires all changes to the data to be tracked as secure, computer-generated time-stamped audit trails (11.10(e)). b) The second category is documents that are developed in word processing systems are subject to review and approval in accordance with document controls established in quality system regulations (21 CFR part 820, 21 CFR part 210/211, etc.). A revision history is required for these documents but not an audit trail for all changes made by any personnel throughout the life of the document such as for drafts and preliminary document versions. These documents have been formally reviewed prior to release. As such, these documents, when used as an electronic record, should include read only access to the documents on-line and manual controls to ensure that the versions provided on-line are accurate. 4. Data collected for assessment of machine performance and other quality system data not explicitly required by regulation, need not be controlled as electronic records if the data is only used to define potential corrective and/or preventive actions that are later subject to validation. 5. Monitoring systems that are used as the basis for real-time process adjustments must be validated. 6. Where automated systems have been validated and the results are controlled, incoming machine data and results of intermediate processing from the automated system need not be controlled as an electronic record. 7. Paper based output of a computer system is acceptable given validation has been conducted. If the process definition requires only summary data from the process, then only this data need be transferred (electronically or hard copy) to the target quality record. Issues Regarding Specific Articles of the Part 11 Regulation 11.10(b) The ability to generate accurate and complete copies of records in both human readable and electronic form suitable for inspection, review, and copying by the agency. ASCII files that can be exported must be acceptable for copying electronically or else potential copyright problems may arise. Unless the FDA standardizes on select applications that they use and therefore the format of data they need, it will be impossible to provide copies of all applications to the FDA (c) Use of secure, computer-generated, time-stamped audit trails to independently record the date and time of operator entries and actions that create, modify, or delete electronic records. Record changes shall not obscure previously recorded information. Page 3 of 5
4 This portion of the regulation is the one that has generated the most controversy. Where raw data is required by predicate regulations (21 CFR Part 58 Good Laboratory Practices) an audit trail is clearly necessary. Where documents are reviewed and approved separately (such as for on-line procedures), the electronic version need not include an audit trail for all changes made. The current regulatory requirement to provide a description of changes made to documents has been recognized as an acceptable practice and is compliant with the predicate regulation. (See previous discussion items regarding audit trails.) Controls for open systems - Procedures and controls to ensure authenticity, integrity, and as appropriate, confidentiality. Include additional measures beyond requirements such as document encryption and use of digital signature standards The prescribed controls for open systems are certainly reasonable. What is unreasonable is the interpretation that any system that provides access via the Internet is considered to be an open system. This interpretation is not consistent with the definition as provided in section 11.3 of the regulation. Access to an electronic record system via does not constitute an open system and is acceptable provided security controls are in place to restrict external access by unauthorized personnel Signature manifestations - Signed electronic records contain the following information associated with the signing: (1) Name (2) Date and time (3) Meaning (author, review, approval) This is certainly reasonable requirement; however, more prescriptive requirements such as that the time is local and that the time is to the nearest second are not appropriate or beneficial General requirements - Electronic signatures shall be unique and not reused or reassigned Not reused or reassigned should include a time period such as defined for record retention. Two years would seem to be adequate time to ensure that all operations associated with an ID can be conclusively identified to a unique owner Controls for identification codes/ passwords - Use of transaction safeguards to prevent unauthorized use of passwords and/or identification codes, and to detect and report in an immediate and urgent manner any attempts at their unauthorized use to the system security unit, and, as appropriate, to organizational management. Amplifying guidance is needed to clarify these requirements. Is a function that log-offs a user after three unsuccessful log-on attempts acceptable? When are reports to be sent to Page 4 of 5
5 organizational management? Are there any other safeguards that would be expected from the agency? Summary The FDA s publishing of new regulations in the recent past has been accompanied by significant amplifying guidance documents. These guidance documents have very effectively served to specify the implementation requirements. Part 11 requirements have been promulgated without accompanying guidance documents and industry is left to interpret according to inconsistent statements of a few FDA individuals. Guidance on Part 11 is essential to facilitate industry compliance. ASQ Biomedical Division as a professional society would like to assist in providing better information to industry and establishing consistent and reasonable guidance for implementation of this regulation. Please let us know how this service may be accomplished in conjunction with agency programs. We look forward to your feedback. Page 5 of 5
Full Compliance Contents
Full Compliance for and EU Annex 11 With the regulation support of Contents 1. Introduction 2 2. The regulations 2 3. FDA 3 Subpart B Electronic records 3 Subpart C Electronic Signatures 9 4. EU GMP Annex
More informationAgilent MicroLab Software with Spectroscopy Configuration Manager and Spectroscopy Database Administrator (SCM/SDA)
Agilent MicroLab Software with Spectroscopy Configuration Manager and Spectroscopy Database Administrator (SCM/SDA) Compliance with 21 CFR Part 11 Introduction Part 11 in Title 21 of the Code of Federal
More information21 CFR PART 11 ELECTRONIC RECORDS, ELECTRONIC SIGNATURES 21.11.2013. 21 CFR Part 11 Compliance PLA 2.1
21 CFR PART 11 ELECTRONIC RECORDS, ELECTRONIC SIGNATURES Compliance of PLA 2.1 21.11.2013 21 CFR Part 11 Compliance PLA 2.1 SEC. 11.2 IMPLEMENTATION. (a) For records required to be maintained but not submitted
More informationSelf-Assessment of eresearch Compliance with 21 CFR Part 11, Electronic Record; Electronic Signatures
Self-Assessment of eresearch Compliance with 21 CFR Part 11, Electronic Record; Electronic Signatures Subpart A General Provisions Sec. 11.1 Scope. (a) The regulations in this part set forth the criteria
More informationFDA Title 21 CFR Part 11:Electronic Records; Electronic Signatures; Final Rule (1997)
www.qadata.co.za Introduction FDA Title 21 CFR Part 11:Electronic Records; Electronic Signatures; Final Rule (1997) INITIAL REGULATION RELEASED High profile audit findings Industry complaints to wasting
More informationElectronic records and electronic signatures in the regulated environment of the pharmaceutical and medical device industries
White Paper No 01 I December 2010 Implementation of 21 CFR Part 11 in the epmotion Software Electronic records and electronic signatures in the regulated environment of the pharmaceutical and medical device
More informationFILEHOLD DOCUMENT MANAGEMENT SYSTEM 21 CFR PART 11 COMPLIANCE WHITE PAPER
FILEHOLD DOCUMENT MANAGEMENT SYSTEM 21 CFR PART 11 COMPLIANCE WHITE PAPER Copyright 2012 FileHold Systems Inc. All rights reserved. For further information about this manual or other FileHold Systems products,
More informationInfinityQS SPC Quality System & FDA s 21 CFR Part 11 Requirements
InfinityQS SPC Quality System & FDA s 21 CFR Part 11 Requirements www.infinityqs.com Copyright InfinityQS International Table of Contents Overview... FDA s 21 CFR Part 11 Requirements... PART 11 ELECTRONIC
More informationImplementation of 21CFR11 Features in Micromeritics Software Software ID
Implementation of 21CFR11 Features in Micromeritics Software Software ID PART 11 ELECTRONIC RECORDS; ELECTRONIC SIGNATURES Subpart A General Provisions Sec. 11.1 Scope. 11.2 Implementation. 11.3 Definitions.
More informationTools to Aid in 21 CFR Part 11 Compliance with EZChrom Elite Chromatography Data System. White Paper. By Frank Tontala
Tools to Aid in 21 CFR Part 11 Compliance with EZChrom Elite Chromatography Data System White Paper By Frank Tontala Agilent Technologies Software & Informatics Life Sciences & Chemical Analysis Group
More information21 CFR Part 11 Compliance Using STATISTICA
21 CFR Part 11 Compliance Using STATISTICA Last Updated: April 2003 This document was updated to reflect the FDA s latest guidance (released February, 2003) and the withdrawal of previous guidance.! STATSOFT
More informationPOLICY ISSUES IN E-COMMERCE APPLICATIONS: ELECTRONIC RECORD AND SIGNATURE COMPLIANCE FDA 21 CFR 11 ALPHATRUST PRONTO ENTERPRISE PLATFORM
W H I T E P A P E R POLICY ISSUES IN E-COMMERCE APPLICATIONS: ELECTRONIC RECORD AND SIGNATURE COMPLIANCE FDA 21 CFR 11 ALPHATRUST PRONTO ENTERPRISE PLATFORM This white paper is written for senior executives
More informationThe Impact of 21 CFR Part 11 on Product Development
The Impact of 21 CFR Part 11 on Product Development Product development has become an increasingly critical factor in highly-regulated life sciences industries. Biotechnology, medical device, and pharmaceutical
More informationEmpower TM 2 Software
Empower TM 2 Software 21 CFR PART 11 COMPLIANCE ASSESSMENT Revision A, December, 2005 1 of 14 Waters Corporation Note: Information presented in this document assumes that the appropriate Empower 2 System
More information21 CFR Part 11 Electronic Records & Signatures
Gap Analysis - Checklist 21 CFR Part 11 Electronic Records & Signatures his document is a proposal and starting point only. he type and extent of documentation depends on the process environment. he proposed
More informationTIBCO Spotfire and S+ Product Family
TIBCO Spotfire and S+ Product Family Compliance with 21 CFR Part 11, GxP and Related Software Validation Issues The Code of Federal Regulations Title 21 Part 11 is a significant regulatory requirement
More informationImplement best practices by using FileMaker Pro 7 as the backbone of your 21 CFR 11 compliant system.
21 CRF 11 Electronic Records and Signatures Implement best practices by using FileMaker Pro 7 as the backbone of your 21 CFR 11 compliant system. By Todd Duell What does Title 21 of the Code of Federal
More informationAssessment of Vaisala Veriteq vlog Validation System Compliance to 21 CFR Part 11 Requirements
/ WHITE PAPER Assessment of Vaisala Veriteq vlog Validation System Compliance to 21 CFR Part 11 Requirements The 21 CFR Part 11 rule states that the FDA view is that the risks of falsification, misinterpretation,
More information21 CFR Part 11 Implementation Spectrum ES
21 CFR Part 11 Implementation Spectrum ES INFRARED SPECTROSCOPY T E C H N I C A L N O T E Introduction Compliance with 21 CFR Part 11 is mandatory for pharmaceutical companies and their suppliers to sell
More informationOracle WebCenter Content
Oracle WebCenter Content 21 CFR Part 11 Certification Kim Hutchings US Data Management Phone: 888-231-0816 Email: khutchings@usdatamanagement.com Introduction In May 2011, US Data Management (USDM) was
More informationHow To Control A Record System
Thermo Scientific Qtegra Intelligent Scientific Data Solution (ISDS) Software for 21 CFR Part 11 Compliant Laboratories Technical Note 43106 Key Words Compliance, Electronic Records, 21 CFR Part 11 Goal
More informationDeclaration of Conformity 21 CFR Part 11 SIMATIC WinCC flexible 2007
Declaration of Conformity 21 CFR Part 11 SIMATIC WinCC flexible 2007 SIEMENS AG Industry Sector Industry Automation D-76181 Karlsruhe, Federal Republic of Germany E-mail: pharma.aud@siemens.com Fax: +49
More informationrsdm and 21 CFR Part 11
rsdm and 21 CFR Part 11 Meeting the 21 CFR Part 11 Burden without Overburdening The right solutions for smaller biopharma. Nothing more. Nothing less. Prepared by: Ken VanLuvanee www.virtualregulatorysolutions.com
More informationDeltaV Capabilities for Electronic Records Management
January 2013 Page 1 DeltaV Capabilities for Electronic Records Management This paper describes DeltaV s integrated solution for meeting FDA 21CFR Part 11 requirements in process automation applications
More informationCompliance Matrix for 21 CFR Part 11: Electronic Records
Compliance Matrix for 21 CFR Part 11: Electronic Records Philip E. Plantz, PhD, Applications Manager David Kremer, Senior Software Engineer Application Note SL-AN-27 Revision A Provided By: Microtrac,
More informationDeltaV Capabilities for Electronic Records Management
September 2004 Page 1 An integrated solution for meeting FDA 21CFR Part 11 requirements in process automation applications using a configurable off-the-shelf (COTS) solution Emerson Process Management.
More information21 CFR Part 11 White Paper
21 CFR Part 11 White Paper Version V8.00 SR1 ProLeiT AG Einsteinstrasse 8, D-91074 Herzogenaurach, Germany Phone: +49 (0) 9132 777-0 Fax: +49 (0) 9132 777-150 E-Mail: info@proleit.com Internet: http://www.proleit.com
More informationA ChemoMetec A/S White Paper September 2013
NucleoCounter NC-3000, NucleoView NC-3000 Software and Code of Federal Regulation 21 Part 11; Electronic Records, Electronic Signatures (21 CFR Part 11) A ChemoMetec A/S White Paper September 2013 ChemoMetec
More informationAutoSave. Achieving Part 11 Compliance. A White Paper
AutoSave Achieving Part 11 Compliance A White Paper Synopsis This whitepaper provides information related to FDA regulation 21 CFR Part 11 (Part 11) for organizations considering MDT software solutions.
More informationElectronic Document and Record Compliance for the Life Sciences
Electronic Document and Record Compliance for the Life Sciences Kiran Thakrar, SoluSoft Inc. SoluSoft, Inc. 300 Willow Street South North Andover, MA 01845 Website: www.solu-soft.com Email: solusoftsales@solu-soft.com
More informationIntland s Medical Template
Intland s Medical Template Traceability Browser Risk Management & FMEA Medical Wiki Supports compliance with IEC 62304, FDA Title 21 CFR Part 11, ISO 14971, IEC 60601 and more INTLAND codebeamer ALM is
More informationInfoCenter Suite and the FDA s 21 CFR part 11 Electronic Records; Electronic Signatures
InfoCenter Suite and the FDA s 21 CFR part 11 Electronic Records; Electronic Signatures Overview One of the most popular applications of InfoCenter Suite is to help FDA regulated companies comply with
More informationGuidance for Industry. 21 CFR Part 11; Electronic. Records; Electronic Signatures. Time Stamps
Guidance for Industry 21 CFR Part 11; Electronic Records; Electronic Signatures Time Stamps Draft Guidance This guidance document is being distributed for comment purposes only. Comments and suggestions
More informationThis interpretation of the revised Annex
Reprinted from PHARMACEUTICAL ENGINEERING The Official Magazine of ISPE July/August 2011, Vol. 31 No. 4 www.ispe.org Copyright ISPE 2011 The ISPE GAMP Community of Practice (COP) provides its interpretation
More informationRisk-Based Approach to 21 CFR Part 11
3109 W. Dr. Martin Luther King, Jr. Blvd., Suite 250 Tampa, FL 33607 USA Tel: 813/960-2105 Fax: 813/264-2816 www.ispe.org Risk-Based Approach to 21 CFR Part 11 The 21 CFR Part 11 regulation is a comprehensive
More information21 CFR Part 11 Checklist
21 CFR Part 11 Checklist GUIDE NOTOCORD - 113 Chemin de Ronde - 78290 Croissy-sur-Seine www.notocord.com my.notocord.com support@notocord.com +33 (0)1 34 80 00 00 1. Preliminary information 1.1. Purpose
More informationCompliance Response Edition 07/2009. SIMATIC WinCC V7.0 Compliance Response Electronic Records / Electronic Signatures. simatic wincc DOKUMENTATION
Compliance Response Edition 07/2009 SIMATIC WinCC V7.0 Compliance Response Electronic Records / Electronic Signatures simatic wincc DOKUMENTATION Compliance Response Electronic Records / Electronic Signatures
More informationFDA 21 CFR Part 11 Electronic records and signatures solutions for the Life Sciences Industry
FDA 21 CFR Part 11 Electronic records and signatures solutions for the Life Sciences Industry The Rule 21 CFR Part 11 Handwritten signature means the scripted name or legal mark of an individual handwritten
More informationImplementing Title 21 CFR Part 11 (Electronic Records ; Electronic Signatures) in Manufacturing Presented by: Steve Malyszko, P.E.
Implementing Title 21 CFR Part 11 (Electronic Records ; Electronic Signatures) in Manufacturing Presented by: Steve Malyszko, P.E. President & CEO Agenda Introduction Who is Malisko Engineering? Title
More informationNova Southeastern University Standard Operating Procedure for GCP. Title: Electronic Source Documents for Clinical Research Study Version # 1
Nova Southeastern University Standard Operating Procedure for GCP Title: Electronic Source Documents for Clinical Research Study Version # 1 SOP Number: OCR-RDM-006 Effective Date: August 2013 Page 1 of
More informationEnabling SharePoint for 21 CFR Part 11 Compliance - Electronic Signature Use Case
Enabling SharePoint for 21 CFR Part 11 Compliance - Electronic Signature Use Case Sudeep Nambiar Technical Strategist www.linkedin.com/in/sudeepnambiar/ Thanks to our Sponsors! Platinum: Gold: Silver:
More informationSupplement to the Guidance for Electronic Data Capture in Clinical Trials
Supplement to the Guidance for Electronic Data Capture in Clinical Trials January 10, 2012 Drug Evaluation Committee, Japan Pharmaceutical Manufacturers Association Note: The original language of this
More informationEclipsys Sunrise Clinical Manager Enterprise Electronic Medical Record (SCM) and Title 21 Code of Federal Regulations Part 11 (21CFR11)
Eclipsys Sunrise Clinical Manager Enterprise Electronic Medical Record (SCM) and Title 21 Code of Federal Regulations Part 11 (21CFR11) The title 21 code of federal regulations part 11 deals with an institutions
More informationGuidance for Industry. 21 CFR Part 11; Electronic Records; Electronic Signatures. Electronic Copies of Electronic Records
Guidance for Industry 21 CFR Part 11; Electronic Records; Electronic Signatures Electronic Copies of Electronic Records Draft Guidance This guidance document is being distributed for comment purposes only.
More informationSolidWorks Enterprise PDM and FDA 21CFR Part 11
T E C H N I C A L P A P E R SolidWorks Enterprise PDM and FDA 21CFR Part 11 This Technical Paper discusses the technical solutions provided by SolidWorks Enterprise PDM to address the FDA 21 CFR Part 11
More informationREGULATIONS COMPLIANCE ASSESSMENT
ALIX is free software: you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation. REGULATIONS COMPLIANCE ASSESSMENT BUSINESS
More informationManual 074 Electronic Records and Electronic Signatures 1. Purpose
1. Purpose The purpose of this document is to provide an interpretation of FDA 21 CFR Part 11, Electronic Records; Electronic Signatures (ER/ES) and to provide guidance for acceptable practices in the
More informationScreenMaster RVG200 Paperless recorder FDA-approved record keeping. Measurement made easy
Information INF13/147 EN ScreenMaster RVG200 Paperless recorder FDA-approved record keeping Measurement made easy Guidance on the use of the RVG200 paperless recorder for electronic record keeping in FDA-approved
More informationGuidance for Industry Computerized Systems Used in Clinical Investigations
Guidance for Industry Computerized Systems Used in Clinical Investigations U.S. Department of Health and Human Services Food and Drug Administration (FDA) Office of the Commissioner (OC) May 2007 Guidance
More informationSoftware Manual Part IV: FDA 21 CFR part 11. Version 2.20
Software Manual Part IV: FDA 21 CFR part 11 Version 2.20 OPTIMA Software Manual Part IV: FDA 21 CFR part 11 BMG LABTECH This manual was designed to guide OPTIMA users through the software features related
More informationGuidance for Industry Part 11, Electronic Records; Electronic Signatures Scope and Application
Guidance for Industry Part 11, Electronic Records; Electronic Signatures Scope and Application U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research
More informationCHAPTER 11 COMPUTER SYSTEMS INFORMATION TECHNOLOGY SERVICES CONTROLS
11-1 CHAPTER 11 COMPUTER SYSTEMS INFORMATION TECHNOLOGY SERVICES CONTROLS INTRODUCTION The State Board of Accounts, in accordance with State statutes and the Statements on Auditing Standards Numbers 78
More informationINFORMATION TECHNOLOGY CONTROLS
CHAPTER 14 INFORMATION TECHNOLOGY CONTROLS SCOPE This chapter addresses requirements common to all financial accounting systems and is not limited to the statewide financial accounting system, ENCOMPASS,
More informationSoftware. For the 21 CFR Part 11 Environment. The Science and Technology of Small Particles
Software For the 21 CFR Part 11 Environment The Science and Technology of Small Particles 21 CFR Part 11 Solution confirm Software The Code of Federal Regulations Title 21, Part 11, was implemented by
More informationCompliance Response SIMATIC SIMATIC PCS 7 V8.1. Electronic Records / Electronic Signatures (ERES) Edition 03/2015. Answers for industry.
SIMATIC SIMATIC PCS 7 V8.1 Electronic Records / Electronic Signatures (ERES) Compliance Response Edition 03/2015 Answers for industry. Compliance Response Electronic Records / Electronic Signatures (ERES)
More informationPART 10 COMPUTER SYSTEMS
PART 10 COMPUTER SYSTEMS 10-1 PART 10 COMPUTER SYSTEMS The following is a general outline of steps to follow when contemplating the purchase of data processing hardware and/or software. The State Board
More informationFor technical assistance, please contact: Thermo Nicolet Corporation 5225 Verona Road Madison WI 53711-4495
The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. Thermo Nicolet Corporation shall not be liable
More informationGuidance for Industry COMPUTERIZED SYSTEMS USED IN CLINICAL TRIALS
Guidance for Industry COMPUTERIZED SYSTEMS USED IN CLINICAL TRIALS U.S. Department of Health and Human Services Food and Drug Administration Center for Biologic Evaluation and Research (CBER) Center for
More informationShiny Server Pro: Regulatory Compliance and Validation Issues
Shiny Server Pro: Regulatory Compliance and Validation Issues A Guidance Document for the Use of Shiny Server Pro in Regulated Clinical Trial Environments June 19, 2014 RStudio, Inc. 250 Northern Ave.
More informationWaters Empower 2 Software Seamlessly Manages Regulated Data to Aid in 21 CFR Part 11 Compliance
THE ROLE OF WATERS EMPOWER 2 SOFTWARE IN ASSISTING IN 21 CFR PART 11 COMPLIANCE Waters Empower 2 Software Seamlessly Manages Regulated Data to Aid in 21 CFR Part 11 Compliance Summary The objective of
More informationWaters Empower Software Seamlessly Manages Regulated Data to Aid in 21 CFR Part 11 Compliance
THE ROLE OF WATERS EMPOWER SOFTWARE IN ASSISTING IN 21 CFR PART 11 COMPLIANCE Waters Empower Software Seamlessly Manages Regulated Data to Aid in 21 CFR Part 11 Compliance Summary The objective of this
More information21 CFR Part 11 Deployment Guide for Wonderware System Platform 3.1, InTouch 10.1 and Historian 9.0
Deployment Guide 21 CFR Part 11 Deployment Guide for Authors: Invensys Operations Management and TSD (Total System Design, an Optimation Company) Table of Contents by Section 1 Before you Begin...............................................................................................
More informationCoSign for 21CFR Part 11 Compliance
CoSign for 21CFR Part 11 Compliance 2 Electronic Signatures at Company XYZ Company XYZ operates in a regulated environment and is subject to compliance with numerous US government regulations governed
More informationImplementing CitectSCADA to meet the requirements of FDA 21 CFR Part 11
Implementing CitectSCADA to meet the requirements of FDA 21 CFR Part 11 Version 2.0 July 2008 FDA 21 CFR Part 11 READY Reversed out on solid background colour Background In 1991, members of the pharmaceutical
More informationooo- I542 (314 Brlngrng tnnovation to patient care worldwide April 23,2003
Dockets Manalgement Branch (HFA-305) Food and Drug Administration 5630 Fishers Lane, Rm. 1061 Rockville, MD 20852. RE: Docked+ Nos. 03D-0060,99D-1458,OOD-1538,OOD-1543,OOD-1542, and 000-l 539; Draft Guidance
More information1/30/2013. Agenda. Electronic Signatures/ Informed Consent
2013 UC Compliance & Audit Symposium Recordkeeping Matters: Part 11 Compliance Rachel Nosowsky ~ Acting Deputy General Counsel Rachel.Nosowsky@ucop.edu ~ (510) 987-9407 Agenda Electronic Signatures/Informed
More informationLabChip GX/GXII with LabChip GxP Software
Regulatory Compliance LabChip GX/GXII with LabChip GxP Software Supporting Regulatory Compliance Caliper LabChip GX/GXII suite of instruments provides automated electrophoresis to analyze quality, size,
More informationSympatec GmbH System-Partikel-Technik WINDOX 4. Electronic Records/ Electronic Signatures Compliance Assessment Worksheet for 21 CFR Part 11
Sympatec GmbH System-Partikel-Technik WINDOX 4 Electronic Records/ Electronic Signatures Compliance Worksheet for 21 CFR Part 11 Note Complete or partial duplication of the present documentation and
More informationMeeting the FDA s Requirements for Electronic Records and Electronic Signatures (21 CFR Part 11)
Meeting the FDA s Requirements for Electronic Records and Electronic Signatures (21 CFR Part 11) Executive Summary...3 Background...4 Internet Growth in the Pharmaceutical Industries...4 The Need for Security...4
More informationComputerized Systems Used in Medical Device Clinical Investigations
Computerized Systems Used in Medical Device Clinical Investigations Presented by Jonathan Helfgott Consumer Safety Officer Division of Bioresearch Monitoring Office of Compliance Center for Devices and
More informationThermal Analysis. http://www.fda.gov. Subpart A General Provisions 11.1 Scope. 11.2 Implementation. 11.3 Definitions.
Thermal Analysis 21 CFR 11 Compliance 21 CFR Part 11 Electronic Records; Electronic Signatures General concept The U.S. Federal Food and Drug Administration (FDA) has issued regulations that provide criteria
More informationGuidance for Industry
Guidance for Industry Electronic Source Data in Clinical Investigations U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for
More informationFDA Regulation of Electronic Source Data in Clinical Investigations
FDA Regulation of Electronic Source Data in Clinical Investigations Q1 Productions Second Annual Innovations in Clinical Data Management Conference, Arlington, VA Mahnu Davar Presented on Oct. 28, 2014
More informationThe biggest challenges of Life Sciences companies today. Comply or Perish: Maintaining 21 CFR Part 11 Compliance
S E P T E M B E R 2 0 1 3 Comply or Perish: The biggest challenges of Life Sciences companies today are maintaining a robust product pipeline and reducing time to market while complying with an increasing
More informationREQUEST FOR BOARD ACTION
REQUEST FOR BOARD ACTION HENDERSON COUNTY BOARD OF COMMISSIONERS MEETING DATE: 23 March 2005 SUBJECT: ATTACHMENT(S): HIPAA 1. Proposed Resolution adopting policies 2. Proposed policies SUMMARY OF REQUEST:
More informationAccess Control BUSINESS REQUIREMENTS FOR ACCESS CONTROL
AU7087_C013.fm Page 173 Friday, April 28, 2006 9:45 AM 13 Access Control The Access Control clause is the second largest clause, containing 25 controls and 7 control objectives. This clause contains critical
More informationOLEY VALLEY SCHOOL DISTRICT
No. 829 SECTION: OPERATIONS OLEY VALLEY SCHOOL DISTRICT TITLE: ELECTRONIC SIGNATURES ADOPTED: February 19, 2015 REVISED: 829. ELECTRONIC TRANSACTIONS, RECORDS AND SIGNATURES 1. Purpose 2. Authority 73
More informationSpectroscopy Configuration Manager (SCM) Software. 21 CFR Part 11 Compliance Booklet
Spectroscopy Configuration Manager (SCM) Software 21 CFR Part 11 Compliance Booklet Notices Agilent Technologies, Inc. 2015 No part of this manual may be reproduced in any form or by any means (including
More informationResearch and the HIPAA Security Rule Prepared for the Association of American Medical Colleges by Daniel Masys, M.D. Professor and Chairman,
Research and the HIPAA Security Rule Prepared for the Association of American Medical Colleges by Daniel Masys, M.D. Professor and Chairman, Department of Biomedical Informatics Vanderbilt University School
More informationClinical database/ecrf validation: effective processes and procedures
TITOLO SLIDE Testo Slide Testo Slide Testo Slide Clinical database/ecrf validation: effective processes and procedures IV BIAS ANNUAL CONGRESS Padova September, 26 th 2012 PQE WORKSHOP: What's new in Computerized
More informationData Management PACT Workshop: Design & Operation of GMP Cell Therapy Facilities April 10 th -11 th, 2007
Data Management PACT Workshop: Design & Operation of GMP Cell Therapy Facilities April 10 th -11 th, 2007 Data Management Discuss Database Development Design Process Tips Data Normalization Reporting Ideas
More informationSCADA Compliance Tools For NERC-CIP. The Right Tools for Bringing Your Organization in Line with the Latest Standards
SCADA Compliance Tools For NERC-CIP The Right Tools for Bringing Your Organization in Line with the Latest Standards OVERVIEW Electrical utilities are responsible for defining critical cyber assets which
More informationAuthorized. User Agreement
Authorized User Agreement CareAccord Health Information Exchange (HIE) Table of Contents Authorized User Agreement... 3 CareAccord Health Information Exchange (HIE) Polices and Procedures... 5 SECTION
More informationDepartment of Health and Human Services
Thursday, February 20, 2003 Part II Department of Health and Human Services Office of the Secretary 45 CFR Parts 160, 162, and 164 Health Insurance Reform: Security Standards; Final Rule VerDate Jan2003
More informationGLP Records Storage and Retrieval
GLP Records Storage and Retrieval Cindy Green BBS UNITED, ALEX HAYDEN/GETTY IMAGES GLP Topics addresses topics associated with good laboratory practice requirements. We intend this column to be a useful
More informationIssues to Address: The Privacy Concerns of Individuals
July 21, 2009 The Honorable Michael J. Astrue Commissioner Social Security Administration 6401 Security Boulevard Baltimore, MD 21235-7703 Dear Mike: As you requested, the ABA explored the issues related
More informationAuthentication of Documents/Use of Professional Stamps
Authentication of Documents/Use of Professional Stamps 1 Introduction The intention of this guideline is to amplify and clarify requirements for authentication of documents and use of the stamp, based
More informationHealth Insurance Portability and Accountability Act Enterprise Compliance Auditing & Reporting ECAR for HIPAA Technical Product Overview Whitepaper
Regulatory Compliance Solutions for Microsoft Windows IT Security Controls Supporting DHS HIPAA Final Security Rules Health Insurance Portability and Accountability Act Enterprise Compliance Auditing &
More informationData Management Unit Research Institute for Health Sciences, Chiang Mai University
Data Management Unit Research Institute for Health Sciences, Chiang Mai University Clinical Data Management is the process of handling data from clinical trials. The inherent goal of any clinical data
More informationBUSINESS ASSOCIATE AGREEMENT HIPAA Protected Health Information
BUSINESS ASSOCIATE AGREEMENT HIPAA Protected Health Information I. PREAMBLE ( Covered Entity ) and ( Business Associate ) (jointly the Parties ) wish to enter into an Agreement to comply with the requirements
More informationCompliance in the BioPharma Industry. White Paper v1.0
in the BioPharma Industry White Paper v1.0 July 2005 I. Introduction...3 II. Overview of Regulations...3 III. Overview of Validation...4 a. Validation...4 b. Security s Part Within Validation...6 IV. Introduction
More informationHIPAA BUSINESS ASSOCIATE AGREEMENT
HIPAA BUSINESS ASSOCIATE AGREEMENT This HIPAA Business Associate Agreement ( Agreement ) is by and between ( Covered Entity ) and Xelex Digital, LLC ( Business Associate ), and is effective as of. WHEREAS,
More informationData Management and Good Clinical Practice Patrick Murphy, Research Informatics, Family Health International
Data Management and Good Clinical Practice Patrick Murphy,, Family Health International 1 What is GCP? Good Clinical Practice is an international ethical and scientific quality standard for designing,
More informationRecent Developments Affecting the Disclosure of Test Data and Materials: Comments Regarding the 1996 Statement on the Disclosure of Test Data 1
Recent Developments Affecting the Disclosure of Test Data and Materials: Comments Regarding the 1996 Statement on the Disclosure of Test Data 1, American Psychological Association In t r o d u c t i o
More informationResults Oriented Change Management
Results Oriented Change Management Validating Change Policy through Auditing Abstract Change management can be one of the largest and most difficult tasks for a business to implement, monitor and control
More informationHIPAA Security. 5 Security Standards: Organizational, Policies. Security Topics. and Procedures and Documentation Requirements
HIPAA Security S E R I E S Security Topics 1. Security 101 for Covered Entities 2. Security Standards - Administrative Safeguards 3. Security Standards - Physical Safeguards 4. Security Standards - Technical
More informationHESI: Fetal Imaging Workshop 21 CFR Part 11 Electronic Records & Signatures. Presented by: Jonathan S. Helfgott
HESI: Fetal Imaging Workshop 21 CFR Part 11 Electronic Records & Signatures Presented by: Jonathan S. Helfgott Director of Regulatory Affairs, Stage 2 Innovations Jonathan@stage2innovations.com Disclaimer
More informationClient Advisory October 2009. Data Security Law MGL Chapter 93H and 201 CMR 17.00
Client Advisory October 2009 Data Security Law MGL Chapter 93H and 201 CMR 17.00 For a discussion of these and other issues, please visit the update on our website at /law. To receive mailings via email,
More informationImplementing an Audit Trail within a Clinical Reporting Tool Paul Gilbert, Troy A. Ruth, Gregory T. Weber DataCeutics, Inc.
Paper AD12 Implementing an Audit Trail within a Clinical Reporting Tool Paul Gilbert, Troy A. Ruth, Gregory T. Weber DataCeutics, Inc., Pottstown, PA ABSTRACT This paper is a follow-up to Overview of a
More informationNIST 800-53A: Guide for Assessing the Security Controls in Federal Information Systems. Samuel R. Ashmore Margarita Castillo Barry Gavrich
NIST 800-53A: Guide for Assessing the Security Controls in Federal Information Systems Samuel R. Ashmore Margarita Castillo Barry Gavrich CS589 Information & Risk Management New Mexico Tech Spring 2007
More information