1 PHS-25CW/3CW/3BW Benchtop ph Meter Instruction Manual BANTE INSTRUMENTS LIMITED Technology to create the future...
2 PHS-W Series Benchtop ph Meter 1 Introduction Thank you for selecting the PHS-W series benchtop ph meter. This manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. The following list describes the standard accessories of the meter. After the unpacking, please check all accessories are complete. If any are damaged or missing, please contact nearest distributor. ACCESSORIES: ph Electrode Temperature Probe (Only for PHS-3BW meter) ph Buffer Pouches (ph4.01/7.00/10.01) Power Adapter Display PHS-W series benchtop ph meter is equipped with a clear and bright LCD display that used to show measured values and mode indicators Electrode slope indicator (After calibration, if the slope less then 90%, the indicator will disappear) 2. Measurement mode indicator 3. Calibration indicator 4. Temperature setting indicator 5. Error alarm indicator 6. Automatic Temperature Compensation indicator Keypad The meter includes 8 function keys, names and symbols describe each function key controls. KEY ON CAL ph mv o C ENTER FUNCTION Turn on the meter, press and hold the key for 5 seconds to turn off meter. Press the key to enter calibration mode. Press the key to return to ph measurement mode. Press the key to enter mv measurement mode. Press the Key to set the temperature value. In the temperature setting mode, press the key to increase the setting value. Confirms the calibration or setting value. In the temperature setting mode, press the key to decrease the setting value.
3 PHS-W Series Benchtop ph Meter 2 Connectors PHS-W series benchtop ph meter provides 3 connectors are located on the rear panel. Listed in the below table are the details of these connectors. ATC ELECTRODE POWER INDEX: CONNECTOR ELECTRODE POWER ATC DESCRIPTION For connecting the ph or ORP electrode For connecting the DC9V power adapter For connecting the temperature probe (Only for PHS-3BW meter) Connecting the Electrode Holder The meter comes with an easy-to-use holder for mounting the ph electrodes and temperature probe. If necessary, follow the steps below to install the electrode holder. 1. The base of the electrode holder with an irregular round hole. 2. The electrode arm has a plastic rod. Plastic Rod 3. Insert the plastic rod into the irregular round hole and swivel the electrode arm 90 o. Electrode holder is now ready to swing into desired position.
4 PHS-W Series Benchtop ph Meter 3 ADJUSTMENT OF ELECTRODE ARM: After installation, if the electrode arm automatically rises or falls, you need to adjust the screws until arm locate at any position. 1. Remove the plastic cover from the electrode arm. Screws 2. Use the screwdriver to tighten the screw moderately. 3. Insert the plastic cover to previous position. Installation is completed. Connecting the ph Electrode 1. Take out the ph electrode from the packaging. Follow the steps below to place the electrode into left or right sides of the electrode arm. A. Insert the electrode B. Hook up the cord C. Hook up the cord 2. Insert the BNC connector into corresponding connector socket. Rotate and push the connector clockwise until it locks. After connection is completed, DO NOT pull on the sensor cord. Always make sure that the connector is clean and dry. Connecting the Power Adapter 1. Before plugging in the power adapter, ensure the its voltage matches your local main voltage. 2. Insert the power connector into corresponding connector socket (Marked Power ), the meter is now ready for use. Power On/Off Press ON key to turn on the meter, the display shows measured values. Press and hold the ON key for 5 seconds, the meter will turn off.
5 PHS-W Series Benchtop ph Meter 4 Prior to Use Remove the protective cap from the bottom of the ph electrode. If the glass sensitive membrane on the electrode dries out, soak the electrode in 3M KCL solution or tap water for at least 15 minutes. DO NOT use distilled or deionized water, it will shorten the life of sensor.... Glass Sensitive Membrane Preparing the ph Buffer Solutions PHS-W series benchtop ph meter is packaged with three ph buffer packets required for calibration. Open the ph7.00 buffer packet, place the powder into a 250ml volumetric flask. Pour distilled water 250ml to scale line, mix the solution until reagent is completely dissolved. Preparation of ph4.01 and standard buffer solutions are the same as above. Prepared standard buffer solutions should be stored in hermetically sealed glass containers. Temperature Compensation In order to get accurate results, you need to enable the manual or automatic temperature compensation function before ph measurement or calibration. MANUAL TEMPERATURE COMPENSATION: 1. Use a high-accuracy thermometer to measure the temperature of sample. 2. Press o C key to enter temperature setting mode. 3. Press or key to set the temperature value of the sample. 4. Press ENTER key to confirm, the meter returns to measurement mode. Setting is completed.
6 PHS-W Series Benchtop ph Meter 5 AUTOMATIC TEMPERATURE COMPENSATION: Place the temperature probe into the circular hole of electrode arm. Insert the connector of temperature probe into corresponding connector socket (Marked ATC ). The ATC indicator will show on the display, the meter is now switched to automatic temperature compensation mode. ATC ATC ph Calibration PHS-W series benchtop ph meter can perform 2 points calibration. First calibration point must use ph7.00 (or ph6.86) buffer solution. If your sample is belong to acidic substances, second point use the ph4.01 buffer solution, alkaline substances use the ph10.01 (or ph9.18) buffer solution. The meter will automatically recognize and calibrate to following standard buffer values. USA Standard Buffer Options: ph4.01, 7.00, NIST Standard Buffer Options: ph4.01, 6.86, 9.18 The meter must be calibrated prior to first use or new electrode replaced. To ensure accuracy, regular calibration is recommended. Do not reuse calibration solution after calibration, contaminants in solution will affect the calibration and eventually the accuracy of the measurement. 1. Rinse the ph electrode with distilled water. 2. Press CAL key, the meter enters calibration mode, the display shows required calibration solution (e.g., ph7.00).
7 PHS-W Series Benchtop ph Meter 6 3. Dip the ph electrode into the ph7.00 standard buffer solution, the end of the electrode must be completely immersed into the calibration solution. Stir the sensor gently. 4. Press ENTER key, the meter begins calibration. Wait for the reading to stabilize, the displayed value will flash 3 times indicating that first calibration point is completed. The display shows --- indicating that the meter waits for second point calibration. 5. Rinse the ph electrode with distilled water, dip the electrode into the ph4.01 or standard buffer solution. Stir the sensor gently. Press ENTER key to confirm. 6. Wait for the reading to stabilize, the displayed value will flash 3 times, the meter automatically shows electrode slope and returns to measurement mode. Calibration is completed. Electrode Slope EXIT THE CALIBRATION: During the calibration process, if you want to exit the calibration, press ph key, the meter will return to measurement mode immediately. Electrode indicator will show average slope of the ph electrode after calibration. When the electrode or calibration result does not meet measurement requirements, the indicator will disappear on the display.
8 PHS-W Series Benchtop ph Meter 7 ph Measurement 1. Rinse the ph electrode with distilled water to remove any impurities adhering to the probe body. 2. Dip the electrode into the sample solution, stir the sensor gently to create a homogeneous sample. 3. Wait for the measured value to stabilize, record the reading on the display. mv Measurement 1. In the ph measurement mode, press mv key, the meter will switch to mv measurement mode. 2. Rinse the electrode thoroughly with distilled water. 3. Dip the electrode into the sample solution, stir the sensor gently. 4. Wait for the measured value to stabilize, record the reading on the display. Electrode Care and Maintenance Since ph electrode is susceptible to dirt and contamination, clean as necessary depending on the extent and condition of use. AFTER MEASURING: Rinse the ph electrode in distilled water, store the electrode into the electrode storage solution. CLEANING THE ELECTRODE: Salt deposits: Soak the electrode in warm tap water to dissolve deposits, then thoroughly rinse with distilled water. Oil or Grease film: Wash the electrode bulb gently in some detergents and water. If necessary, using the alcohol to clean the electrode bulb, then rinse with distilled water. Place the sensor in the electrode storage solution for 30 minutes. Clogged reference junction: Heat a diluted KCl solution to 60 o C to 80 o C. Place the sensing part of the electrode into the heated solution for about 10 minutes. Allow the electrode to cool in some unheated KCl solution. Protein deposits: Prepare a 1% pepsin solution in 0.1M of HCL. Place the electrode in the solution for 10 minutes. Rinse the sensor with distilled water. REACTIVATING THE ELECTRODE: If stored and cleaned properly, the electrode should be ready for immediate use. However, a dehydrated bulb may cause sluggish response. To rehydrate the bulb, immerse the electrode in a ph4.01 buffer solution for 10 to 30 minutes. If this fails, the electrode requires activation. Soak the ph electrode in 0.1M HCl for 5 minutes. Remove and rinse with deionized water, then place in 0.1M NaOH for 5 minutes. Remove and rinse again, and soak in electrode storage solution for 30 minutes. HOW TO PREPARE THE ELECTRODE STORAGE SOLUTION: 1. Dissolve grams of potassium chloride reagent (KCL) in the 100mL distilled water. 2. Pour the 50mL of KCL solution into the beaker. 3. Pour the 50mL of ph4.01 standard buffer solution into the same beaker. 4. Stir the solution until the solution has thoroughly mixed. Preparation is completed.
9 PHS-W Series Benchtop ph Meter 8 Optional ph Electrodes PHS-W series benchtop ph meter is equipped with a general purpose ph electrode, if it can not fully meet your measurement requirements, please refer to the table below to select an applicable sensor. SAMPLE TYPE P11 P12 P13 P14 P16 P17 P18 P19 P20 P21 Agar Beer Blood Products Bread, Dough Cement Cosmetics Dairy Products Education Fats/Cream Field Use Fish Products Lab Flasks Low Ionic Meat, Cheese Micro Samples Paint Photographic Soil Surface Test Tubes Tris Buffer Viscose Samples Troubleshooting LCD DISPLAY CAUSE CORRECTIVE ACTION Electrode dried out Measured value is out of range Incorrect ph buffer solutions Electrode is broken Soak the ph electrode in 3M KCL solution for 10 minutes Check the electrode whether clogged, dirty or broken Using the fresh ph buffer solutions for calibration Replace the ph electrode
10 PHS-W Series Benchtop ph Meter 9 Specifications Model PHS-25CW PHS-3CW PHS-3BW Range 0.00~14.00pH Accuracy ± 0.05pH ± 0.01pH ± 0.01pH ph Resolution Calibration Points Calibration Solutions 0.01pH 2 points USA (ph4.01/7.00/10.01) or NIST (ph4.01/6.86/9.18) Temperature Compensation 0~100 C, Manual 0~100 C, Manual 0~100 C, Automatic Range -1999~1999mV mv Accuracy ± 10mV ± 1mV ± 1mV Resolution 1mV Range ~100 C Temperature Accuracy ± 1 C Resolution C Others Connector Power Requirements Dimensions Weight BNC DC9V, using AC adapters, 220VAC/50Hz 210(L) 205(W) 65(H)mm 1.5kg
11 PHS-W Series Benchtop ph Meter 10 Addendum: How to select the USA or NIST buffer standard PHS-W series benchtop ph meter contain two ph buffer standards that can be selected. The factory default parameter is USA standard. If you expect convert into another standard, please follow the steps below to set. 1. Press ON key to turn on the meter. 2. Press and hold the CAL key, then pull out the power adapter. 3. Insert the power adapter again. Undo the CAL key after 3 seconds. 4. Press ON key to turn on the meter. Setting is completed.
12 Hazardous Substance Statement Bante Instruments is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as cadmium, lead, mercury and the like. New technologies and design parameters are also promoting these efforts and we expect to have little or no such materials in our product in the coming years. We welcome our customer suggestions on how to speed up these efforts. Warranty The warranty period for meter is one year from the date of shipment. Above warranty does not cover sensor and calibration solutions. For more information, please contact nearest authorized distributor. BANTE INSTRUMENTS LIMITED Phone: Website: The information in this document is subject to change without notice. Copyright Bante Instruments Limited, All rights reserved.
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide
More informationph Measurement and Control
ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components
More informationUser Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
More informationExStik TM ph Waterproof Meters
User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL
More informationDirect ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
More informationNitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
More informationMilwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
More informationMilwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
More informationMeasuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3
Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration
More informationUltraBasic Benchtop ph Meter Operation Manual
UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.
More informationCare and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features 2 Quick guide 2 To operate 3 Cleaning 4 Battery replacement 4 Calibration 5 Error messages 5 Troubleshooting guide 6 Technical specifications 6 Information about
More informationSenTix 950 SenTix 980
SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the
More informationCare and use guide. Contents. page
Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement
More informationInstruction Manual Temp 4/5/6 Temperature Meter Series
Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.
More informationInstruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...
Instruction Manual ph 5+ ph/ C ph 6+ ph/ C/mV Ion 6+ ph/ C/mV/Ion Technology Made Easy... 68X243633 Rev 1-1 3/2011 Part of Thermo Fisher Scientific Preface This instruction manual serves to explain the
More informationSINGLE RANGE DS METERS Operation Instructions
SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax
More informationBluelab Soil ph Meter Instruction Manual
Instruction Manual www.getbluelab.com Table of Contents 1.0 Information about measuring the ph of soils/media. 2 2.0 Introduction to Bluelab Soil ph Meter 4 3.0 Preparing the Bluelab Soil ph Meter for
More informationUSEPA Electrode Method Method 8156
ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method
More informationOPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S
Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and
More informationPortable Waterproof ph/dissolved Oxygen Meter
Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35632-Series Portable Waterproof /Dissolved Oxygen Meter WTERPROOF REDY MES % T /Dissolved Oxygen/ /
More informationUser Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
More informationPROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:
PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA
More informationMODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4
More informationResolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or
More informationph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.
Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE
More informationElectrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986
Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986 Fann Instrument Company Electrical Stability Tester Instruction Manual 2012 Fann Instrument
More informationDirect Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor
Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process
More informationCare and use guide. soil ph meter. Contents. page
soil ph meter Care and use guide Contents page Features Bluelab Soil ph Meter Introduction 3 Preparing for use 3 Calibration 4 Information about measuring the ph of soils/media 6 Measuring the ph value
More informationModel 4510 Conductivity/TDS Meter Operating Manual
Model 4510 Conductivity/TDS Meter Operating Manual SAFETY Please read this information carefully prior to installing or using this equipment. 1. The unit described in this manual is designed to be operated
More informationCadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationph 200 ph-/redox-/temperature Meter Operating instructions
ph 200 ph-/redox-/temperature Meter Operating instructions 2 Table of Contents Page 1. Scope of delivery...4 2. System description...4 3. Connections...5 4. Display...7 5. Keypad...7 6. Start up...8 7.
More informationColor, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method
More informationUser's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600
User's Guide ExStik DO600 Dissolved Oxygen Meter II RECALL Dissolved Oxygen DO600 Introduction Congratulations on your purchase of the ExStik DO600 dissolved oxygen / temperature meter which simultaneously
More informationph Measurements of Common Substances
Chem 100 Section Experiment 10 Name Partner s Name Introduction ph Measurements of Common Substances The concentration of an acid or base is frequently expressed as ph. Historically, ph stands for the
More informationColor, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater
More informationCompact ph Measuring Instruments
testo 205, 206, 230 Compact ph Measuring Instruments With innovative probe engineering ph C The new ph measuring instruments with innovative probe engineering Measurement of the ph value plays an important
More informationEnzyme Action: Testing Catalase Activity
Enzyme Action: Testing Catalase Activity Experiment 6A Many organisms can decompose hydrogen peroxide (H 2 O 2 ) enzymatically. Enzymes are globular proteins, responsible for most of the chemical activities
More informationMAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725. Read this manual thoroughly before using and save it for future reference
MAP COFFEE MIA CAPSULE COFFEE MACHINE INSTRUCTION MANUAL MODEL: CM4725 Read this manual thoroughly before using and save it for future reference Purchase MAP capsules at Harvey Norman stores nationwide
More informationELECTROCHEMISTRY. Theory and Practice
ELECTROCHEMISTRY Theory and Practice Dr. Axel Bier Lab Product Application Manager - Electrochemistry For more information, please visit: www.hach.com/smartprobes P a g e 2 CONTENT INTRODUCTION 3 HOW TO
More informationUser's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
More informationHandbook. Thermo Scientific. ph Electrode Selection Handbook
Handbook Thermo Scientific ph Electrode Selection Handbook Table of Contents ph Theory... 4 ph Electrode Theory... 4 ph versus Temperature Theory... 5 ph Electrode Selection Guide... 6 Thermo Scientific
More informationCONDUCTIVITY SENSOR. Description D0382. Figure 1. The Conductivity Sensor
CONDUCTIVITY SENSOR Description D0382 Figure 1. The Conductivity Sensor Short Description The Conductivity Sensor can be used to measure either solution conductivity or total ion concentration of aqueous
More informationThermistor Thermometer MODEL NO. 93210-00
Thermistor Thermometer MODEL NO. 93210-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail:
More informationAutoRanging Digital MultiMeter
Owner's Manual AutoRanging Digital MultiMeter Model No. 82139 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance
More informationRelated concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.
Conductivity of strong and weak electrolytes TEC Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Principle It is possible to
More informationRefer to page 6 to page 19 for detailed instructions.
CODE:GZ0000297065 Refer to page 6 to page 19 for detailed instructions. Preface This manual describes the operation of the COMPACT CONDUCTIVITY METER, B-771. Be sure to read this manual before using the
More informationPRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION
INDUSTRIAL COATINGS RA260 WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION is water emulsifiable rust preventative oil, which provides outstanding corrosion protection for all phosphate metals.
More informationChem 100 Lab Experiment #9 - ACID/BASE INDICATORS
Lab #9 Chem 100 Lab Experiment #9 - ACID/BASE INDICATORS Name: Purpose: In this laboratory we will investigate how indicators can be used to test for the presence of acids or bases in a number of common
More informationOperating Instructions
Operating Instructions FiveEasyPlus ph meter FEP20 Contents Contents 1. Introduction 3 2. Safety measures 3 3. Installation 4 3.1 Installing the electrode arm (optional for FEP20) 4 4. Operating the FiveEasyPlus
More informationMeasurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
More informationMODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation
More informationPART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
More informationph 300 and 310 Portable Waterproof ph/mv/ C Meter
Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35618-Series 300 and 310 Portable Waterproof /mv/ Meter REDY MES WTERPROOF T /mv/ Data Meter 300 Series
More informationMixtures and Pure Substances
Unit 2 Mixtures and Pure Substances Matter can be classified into two groups: mixtures and pure substances. Mixtures are the most common form of matter and consist of mixtures of pure substances. They
More informationMeasuring the Conductivity of Pure Water Using USP<645>
Measuring the Conductivity of Pure Water Using USP Water Analysis Instruments, Thermo Fisher Scientific Application Note 002 Key Words Water conductivity, ultrapure water (UPW), USP, water quality,
More informationHygro-Thermometer + InfraRed Thermometer Model RH101
User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative
More informationAIRMoN, ph and Conductivity Field Measurement
AIRMoN, ph and Conductivity Field Measurement Items needed: Sample to be measured (sample volume 50 ml) Field Observer Form (FOF) for the sample to be measured conductance cell ph vials, 4 conductivity
More informationph and Acidity in Tomato Sauce
ph and Acidity in Tomato Sauce DOC316.52.93100 Endpoint potentiometric titration Scope and application: Tomato sauce 1. Introduction This working procedure refers to the determination of titratable acidity
More informationIn this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..
Identification of a Substance by Physical Properties 2009 by David A. Katz. All rights reserved. Permission for academic use provided the original copyright is included Every substance has a unique set
More informationTroubleshooting Salt Generators
Troubleshooting Salt Generators NOTE Turn off power to unit prior to attempting service or repair. Problems and Corrective Action Problem Possible Cause Corrective Action Low or no chlorine. Low stabilizer
More informationCare, Maintenance, and Storage of YSI 6-Series Probes and Sondes
Tech Note Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes Probe Care and Maintenance Periodic cleaning and membrane changes are required to keep your sensors operating properly. Follow
More informationPREPARATION AND PROPERTIES OF A SOAP
(adapted from Blackburn et al., Laboratory Manual to Accompany World of Chemistry, 2 nd ed., (1996) Saunders College Publishing: Fort Worth) Purpose: To prepare a sample of soap and to examine its properties.
More informationEnzyme Action: Testing Catalase Activity
Enzyme Action: Testing Catalase Activity Experiment 6A Many organisms can decompose hydrogen peroxide (H 2 O 2 ) enzymatically. Enzymes are globular proteins, responsible for most of the chemical activities
More informationDissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
More informationFully Automatic Washing Machine User manual
Fully Automatic Washing Machine User manual This manual is for HWMP55-918 Please read this manual carefully before using. Retain it for future reference. CONTENTS CONTENTS Inside cover Parts 1 Safety precautions
More informationAcid Base Titrations
Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually
More informationOperating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso
Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso SevenEasy Conductivity Italiano Español Français Deutsch English Contents Contents Page 1. Introduction...
More informationELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:
BOECO PORTABLE PH/ORP/TEMP METER MODEL PT-380 A Simultaneous displays ph or mv and temperature. A Convenient calibration with automatic buffer recognition of European and US buffer sets. Calibration data
More informationUSER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE
More informationDimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
More informationInstruction Manual HI 9146. Portable Waterproof Microprocessor Dissolved Oxygen Meter
Instruction Manual HI 9146 Portable Waterproof Microprocessor Dissolved Oxygen Meter w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction
More informationChloride and Salinity
INTRODUCTION Chloride Chloride and Chloride, in the form of the Cl ion, is one of the major inorganic anions, or negative ions, in saltwater and freshwater. It originates from the dissociation of salts,
More informationBLOOD COLLECTION MIXER
USER S MANUAL BLOOD COLLECTION MIXER Model CM735 No. CAT.CM73522Ce Centron Technologies Corporation 319-25 Sadang-4-dong, Dongjak-ku Seoul, Korea 156-823 Tel. +82-2.522.7807 Fax +82-2.522.7806 Table of
More informationph and Acidity in Wine and Fruit Juice
ph and Acidity in Wine and Fruit Juice DOC316.52.93089 Endpoint potentiometric titration Scope and application: Wine and fruit juice 1. Introduction This working procedure refers to the titration of total
More informationDrop Counter Sensor Product Number: ENDRP-AD100
imagine explore learn Drop Counter Sensor Product Number: ENDRP-AD100 Overview The Drop Counter sensor is an optical sensor that accurately records the number of drops of titrant added during a titration.
More informationExperiment 6 Coffee-cup Calorimetry
6-1 Experiment 6 Coffee-cup Calorimetry Introduction: Chemical reactions involve the release or consumption of energy, usually in the form of heat. Heat is measured in the energy units, Joules (J), defined
More informationDISSOLVED OXYGEN PROBE
51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe
More informationInstruction Manual DO 110. Hand-held Dissolved Oxygen Meter with RS 232C. 68X361305 Rev. 0-11/05. Technology Made Easy...
Instruction Manual DO 110 Hand-held Dissolved Oxygen Meter with RS 232C Technology Made Easy... 68X361305 Rev. 0-11/05 Preface This manual serves to explain the use of the hand-held series meters. The
More informationINSTRUCTIONS 56-1190-98. Edition AC
Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography
More information9. Analysis of an Acid-Base Titration Curve: The Gran Plot
9. Analysis of an Acid-Base Titration Curve: The Gran Plot In this experiment, you will titrate a sample of pure potassium hydrogen phthalate (Table 10-4) with standard NaOH. A Gran plot will be used to
More informationPro1030 USER MANUAL. English
Pro1030 USER MANUAL English Contents Warranty...i Introduction...1 Item #605182 Rev A, January 2013 For the latest version of this manual, visit ysi.com 2013 YSI Incorporated. The YSI logo is a registered
More informationCORNER FRIDGE REFRIGERATION UNIT INSTALLATION AND OPERATION INSTRUCTIONS
CORNER FRIDGE REFRIGERATION UNIT INSTALLATION AND OPERATION INSTRUCTIONS D E F Y I N G C O N V E N T I O N Congratulations on your new Corner Fridge Your new corner fridge may have different functions
More informationLaboratory Evaluation Checklist - PSP
Laboratory Evaluation Checklist - PSP PUBLIC HEALTH SERVICE U.S. FOOD AND DRUG ADMINISTRATION SHELLFISH PROGRAM IMPLEMENTATION BRANCH SHELLFISH SAFETY TEAM 5100 PAINT BRANCH PARKWAY COLLEGE PARK, MD 20740-3835
More informationIntroduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of
Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of the blood 7 Procedure to test glucose 7 Procedure to
More informationMicrowave Meter. Instruction Manual
Microwave Meter 840046 Instruction Manual Microwave Meter 840046 Copyright 2009 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted
More informationInstruction manual. VWR unstirred water baths
Instruction manual VWR unstirred water baths Introduction VWR laboratory thermostatic baths provide the means to maintain samples at a precise temperature from just above room temperature up to 99 0 C.
More informationHousehold Acids and Bases
Household Acids and Bases GRADE LEVEL INDICATORS Experiment Demonstrate that the ph scale (0-14) is used to measure acidity and classify substances or solutions as acidic, basic, or neutral. 21 Develop
More informationCQ-Series. User Manual. American Weigh Scales, Inc. CQ-350 (350x0.1g) CQ-500 (500x0.1g) Digital Pocket Scale . = + CALC AC CE % TARE OFF ON MC MR CALI
American Weigh Scales, Inc. CQ-Series Digital Pocket Scale 7 4 1 0. 8 9 5 6 2 3 x = - CALI + CALC MODE M CAL AC CE % TARE OFF ON MC MR CQ-350 (350x0.1g) CQ-500 (500x0.1g) User Manual CQ-Series Manual Thank
More informationOperator Quick Guide EC SENSOR
Operator Quick Guide EC SENSOR Revision G - 24/04/2008 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra
More informationProfessional Plus. Calibration Tips
Professional Plus Calibration Tips Table of Contents Introduction...1 Calibration Worksheet...1 Temperature...3 Calibration Tips...3 Troubleshooting Tips...3 Conductivity...4 Calibration Tips...4 Troubleshooting
More informationManual Ranging MultiMeter
Owner s Manual Manual Ranging MultiMeter Model 82345 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product.! Safety! Operation! Maintenance!
More informationDuct Humidity Transmitter
SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers
More informationINVERTER GENERATOR OWNER S MANUAL FOR YOUR SAFETY PLEASE READ THESE INSTRUCTIONS CAREFULLY AND RETAIN THEM FOR FUTURE USE.
INVERTER GENERATOR OWNER S MANUAL FOR YOUR SAFETY PLEASE READ THESE INSTRUCTIONS CAREFULLY AND RETAIN THEM FOR FUTURE USE. GENERATOR SAFETY EXHAUST GAS PRECAUTIONS Only use outdoors! Exhaust fumes can
More informationThermocouple Thermometer Models
Technology Made Easy... Thermocouple Thermometer Models 60010-00 (Type J) 60010-10 (Type K) 60010-20 (Type T) 68X309905 Rev 1 04/04 TABLE OF CONTENTS 1. INTRODUCTION 4 2. SAFETY PRECAUTIONS 6 3. SPECIFICATIONS
More informationUSEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls
Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required
More informationWide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer
User Manual Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer MODEL 42515 Introduction Congratulations on your purchase of the Model 42515 IR Thermometer. The Model 42515 IR thermometer
More informationQuaternary Ammonium Compounds DOC316.53.01128
Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower
More informationExperiment 12- Classification of Matter Experiment
Experiment 12- Classification of Matter Experiment Matter can be classified into two groups: mixtures and pure substances. Mixtures are the most common form of matter and consist of mixtures of pure substances.
More informationInstruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements
Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction
More informationDetermining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
More information