1 Chapter 5 X-ray Production, X-ray Tubes, and Generators
2 X-ray Generation
3 Production: Bremsstrahlung Brake Radiation Interaction between accelerated electron and nucleus of target material Radiation energy loss E k ZZ % Collisional energy loss 820, ,000 Bremsstralung radiation from energetic electron interaction Bremsstralung radiation energy distribution
4 Characteristic Radiation Interaction between accelerated electron and bounded electrons of target material Shell Tungsten (74) Molybdenum Rhodium (42) (45) K L 12.1/11.5/ /2.6/ /3.1/3.0 M 2.8~ ~ ~0.2 E K -E L = =59.3KeV Generation of Characteristic X-ray Filtered spectrum of X-ray
5 X-ray Tube Voltage: quality 20~150KeV Current: quantity Fluoroscopy 1~5mA Radiography 100~1000mA Time: quantity 1/100~ 10sec
6 X-ray Tube Cathode Generate thermal electron Generate thermal electron Filament Current: 10V, ~7A Tube Current: ~1500mA Determine by filament current & KVP Space charge limited/emission limited Determine Focal spot size Small/ Large Filament Biased Cup Current: 10V, ~7A
7 X-ray Tube Cathode Emission limited
8 Anode Metal target with positive potential Tungsten + copper X-ray generation Mostly by heat: more than 90% Tungsten: Z=74 high h melting point, high h atomic number Molybdenum(42), Rhodium(45):Mammography
9 Rotating Anode To spread out heat Rotor/stator 3000~10,000 rpm Increase spot area 300 times Heat by IR radiation To tube insert and surrounding oil bath
10 Anode angle & Focal Spot Size Tilting angle θ: 7~20 Effective focal width=actual focal width Effective focal length=actual focal length sin θ
11 Angle & Field Coverage
12 Focal Spot Size Variation
13 Resolution Measurement Pinhole Image Slit Image Star pattern Bar Pattern
14 Heel Effect Reduction of X-ray intensity toward anode side of the field. Great escaping thickness Position thicker body part to cathode side
15 Filtration Housing: Block X-ray radiation by lead Except X-ray out port Intrinsic filtration: less than 15KeV By housing metal, Glass, Oil Added filtration: To reduce patient dose Blocking out lower X-ray not contributing imaging Compensation filtration Trough filter: thinner for mediastinum More uniform exposure to detector
16 Collimators Restrict the X-ray field into minimum area
17 X-ray Generator Components Auto Exposure Control Overload protection: by temperature sensor, by calculating KVP ma time
18 High Voltage Supply Generate High Voltage: 100V 100KV Generate High Voltage: 100V 100KV Reduce X-ray Ripple
19 Three-Phase X-ray Generator Using 3 Phase AC line Using 3-Phase AC line Reduced ripple: 13~25%
20 High Freq. Inverter Generator More efficient, More compact, Less costly to manufacture, minimized ripple Easy Control for KVP & ma Widely using
21 Automatic Exposure Control Exposure time selection Exposure time selection Using ionization sensor
22 Factors affecting X-ray Characteristics i of generated X-ray Quality & Quantity 1. Target Material 2. Voltage 3. Current 4. Exposure time 5. Beam filtration 6. Generator waveform
23 Target Material Efficiency of bremsstralung Higher interaction with higher Z- material Quantity of bremsstralung radiation Determines characteristic ti radiation energy Quality of characteristic radiation
24 KVp Determine maximum energy in bremsstralung Quality of X-ray Increase in quantity of X-ray yphotons Exposure KVp 2
25 ma & exposure time Tube current: ma Current from cathode to anode Exposure ma: quantity control Exposure time Quantity control like tube current Tube current exposure time mas
26 Filtration & Ripple Beam filtration Reduce photon number (quantity) Shift average energy to higher value (quality) Voltage ripple Lower ripple shift average energy to higher value Affect quality & quantity of X-ray beam
27 Power Rating & Heat Loading Power rating Power = kvp A max Power = 100kVp 600mA=60kW Power limit on cathode filaments Heat Loading Energy(HU) = kvp A max exposure time Energy(HU) = 80kVp 250mA 0.1sec=2000HU For large HU tube Large focal spot, high anode rotation speed, massive anode, large heat sink, large anode diameter
28 X-ray Exposure Rating Chart To protect tube damage: Single/multiple exposure Maximum allowable tube current 3000rpm 10000rpm 0.3mm 03 focal spot 1.2mm focal spot
29 Cooling Chart & Heat Input For continuous usage Estimate cooling time for tube safety Anode X-ray tube housing
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
More informationX-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
More informationX-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
More informationIn the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the
In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the discussion on the major components of the x-ray machine. Today
More informationProduction of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
More informationProduction of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
More informationVoltage nominal maximum for test. Inverse voltage. nominal maximum for test. Focal spots (IEC 60336:2005) small large
TECHNICAL DATA GENERAL INFORMATIONS Voltage nominal maximum for test 1 135 kv When mounting tube inserts adopt proper caution, in order to avoid glass bulb breaking and fragments projection. Please use
More informationX-Mind. Instinct for perfection
X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in
More informationStudy the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
More informationVeraview IC5 HD High definition, digital imaging excellence. Thinking ahead. Focused on life.
Veraview IC5 HD High definition, digital imaging excellence Thinking ahead. Focused on life. Thinking ahead. Focused on life. 2 3 IC5 HD Super High-Speed with High Definition Clarity For dental radiology
More informationX-ray Imaging System. X-Ray Circuit. Principles of Imaging Science II (RAD 120) X-ray Imaging System Circuitry
Principles of Imaging Science II (RAD 120) X-ray Imaging System Circuitry X-ray Imaging System Operating console Set x-ray tube current (quantity) and voltage (quality) Controls line compensation, kvp,
More informationPLANMED SOPHIE MAMMOGRAPHY X-RAY UNIT TECHNICAL MANUAL 788405/11 LAT LAT SIN DEX. kv kv. mas. mas CTL
LAT mas 8 3 2 1 LAT mas PLANMED SOPHIE MAMMOGRAPHY X-RAY UNIT OBL CC OBL SIN DEX kv kv TECHNICAL MANUAL En 788405/11 TABLE OF CONTENTS Chapter A Chapter B Chapter C GENERAL & TECHNICAL DATA 1 WARNINGS...............................................
More information5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)
More informationLectures about XRF (X-Ray Fluorescence)
1 / 38 Lectures about XRF (X-Ray Fluorescence) Advanced Physics Laboratory Laurea Magistrale in Fisica year 2013 - Camerino 2 / 38 X-ray Fluorescence XRF is an acronym for X-Ray Fluorescence. The XRF technique
More informationRadiation Strip Thickness Measurement Systems
Radiation Strip Thickness Measurement Systems During the past years we have increased our sales of radiometric Vollmer strip thickness measurement systems, i.e. X-ray or isotope gauges, dramatically. Now,
More informationHIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950
HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple
More informationScan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
More informationIon Beam Sputtering: Practical Applications to Electron Microscopy
Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION
More informationX-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION
X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION David Bernard Dage Precision Industries Inc. Fremont, California d.bernard@dage-group.com ABSTRACT The x-ray inspection of PCB assembly processes
More informationVacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
More informationDiscontinued. LUXEON V Portable. power light source. Introduction
Preliminary Technical Datasheet DS40 power light source LUXEON V Portable Introduction LUXEON is a revolutionary, energy efficient and ultra compact new light source, combining the lifetime and reliability
More informationImage Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
More informationAdvanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF)
Advanced Physics Laboratory XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF) Bahia Arezki Contents 1. INTRODUCTION... 2 2. FUNDAMENTALS... 2 2.1 X-RAY PRODUCTION... 2 2. 1. 1 Continuous radiation...
More informationCHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
More informationSIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES
Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 34 SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES S. Cornaby 1, A. Reyes-Mena 1, P. W. Moody 1,
More informationEDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system
EDS system Most common X-Ray measurement system in the SEM lab. Major elements (10 wt% or greater) identified in ~10 secs. Minor elements identifiable in ~100 secs. Rapid qualitative and accurate quantitative
More informationSpectra of pyroelectric X-ray generator
Abstract Spectra of pyroelectric X-ray generator Nagaychenko V.I., Sanin V.M., Yegorov A.M., Shchagin A.V. Kharkov Institute of Physics and Technology, Kharkov 6118, Ukraine The construction of X-ray generator
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:
More informationAAPM REPORT NO. 14 PERFORMANCE SPECIFICATIONS AND ACCEPTANCE TESTING FOR X-RAY GENERATORS AND AUTOMATIC EXPOSURE CONTROL DEVICES
AAPM REPORT NO. 14 PERFORMANCE SPECIFICATIONS AND ACCEPTANCE TESTING FOR X-RAY GENERATORS AND AUTOMATIC EXPOSURE CONTROL DEVICES Published for the American Association of Physics in Medicine by the American
More informationContents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays
J. E. Wilhjelm Ørsted TU Technical University of enmark, Bldg. 348, K-2800 Kongens Lyngby, enmark. X-ray and Computed Tomography Contents History and characterization of X-rays Conventional (projection)
More informationA VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY
A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY I. BIBICU 1, G. NICOLESCU 2, L. CIOLACU 2, L. SERBINA 2 1 National Institute for Materials Physics, Bucharest 77125, Romania, bibicu@infim.ro 2
More informationChapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
More informationIR Edixeon Emitter. 1W Edixeon
High Power LED 1W Edixeon IR Edixeon Emitter Date:2006/06/01 Version:2.0 Device No. : 3-RD-01-E0009 Believe SRL Via Lago di Trasimeno, 21 - Schio (VI) - Italy TEL: +39/0445/579035 FAX: +39/0445/575708
More informationPERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
More information3 Atomic Structure 15
3 Atomic Structure 15 3.1 Atoms You need to be familiar with the terms in italics The diameter of the nucleus is approximately 10-15 m and an atom 10-10 m. All matter consists of atoms. An atom can be
More information50W EdiStar. Approved By Customer. Designer Checker Approval. Ultra High Power LED
Ultra High Power LED 50W EdiStar EdiStar Emitter Approved By Customer Designer Checker Approval Date:2006/12/04 Version:Preliminary V0.1 4F, No. 800, Chung-Cheng Rd, Chung-Ho, Taipei 235, Taiwan Tel: 886-2-8227-6996
More informationPHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.
PHYS 222 Spring 2012 Final Exam Closed books, notes, etc. No electronic device except a calculator. NAME: (all questions with equal weight) 1. If the distance between two point charges is tripled, the
More informationIntroduction to Powder X-Ray Diffraction History Basic Principles
Introduction to Powder X-Ray Diffraction History Basic Principles Folie.1 History: Wilhelm Conrad Röntgen Wilhelm Conrad Röntgen discovered 1895 the X-rays. 1901 he was honoured by the Noble prize for
More informationDPX Series Service Training. Theory and System Overview
DPX Series Service Training Theory and System Overview Rev F - 2/2006 DPX consists of 4 Primary Modules X-Ray Generation Tube Power Supplies X-Ray Detection Detector Counting (AGS) Communication Communications
More informationIAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)
IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization
More informationExperimental study of beam hardening artefacts in photon counting breast computed tomography
Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del Guerra a,n. Lanconelli b, A. Lauria c, G. Mettivier c, M.C. Montesi c, D. Panetta a,
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
More informationName: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography
Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,
More informationRadiation Safety Characteristics of the NOMAD Portable X-ray System
Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality
More informationCathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
More informationMeasurement of Charge-to-Mass (e/m) Ratio for the Electron
Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic
More informationELECTRICAL FUNDAMENTALS
General Electricity is a form of energy called electrical energy. It is sometimes called an "unseen" force because the energy itself cannot be seen, heard, touched, or smelled. However, the effects of
More informationX-Ray Tubes for Medical Imaging AAPM 2013. Rolf Behling Philips Healthcare, Hamburg, Germany MO-F-141-1
X-Ray Tubes for Medical Imaging AAPM 2013 MO-F-141-1 Rolf Behling Philips Healthcare, Hamburg, Germany August, 2013 1 Abstract Why do we find 500+ types of X-ray tubes on the market? Why still vacuum technology
More informationInterested in conducting your own webinar? Email webinars@bnpmedia.com
Interested in conducting your own webinar? Email webinars@bnpmedia.com IR Applications in the Real World Archive www.process-heating.com Add to your favorites IR Applications in the Real World What it
More informationPrinciples of Ion Implant
Principles of Ion Implant Generation of ions dopant gas containing desired species BF 3, B 2 H 6, PH 3, AsH 3, AsF 5 plasma provides positive ions (B 11 ) +, BF 2+, (P 31 ) +, (P 31 ) ++ Ion Extraction
More informationSolid Oxide Fuel Cell Gas Turbine Hybrid Power Plant. M. Henke, C. Willich, M. Steilen, J. Kallo, K. A. Friedrich
www.dlr.de Chart 1 > SOFC XIII > Moritz Henke > October 7, 2013 Solid Oxide Fuel Cell Gas Turbine Hybrid Power Plant M. Henke, C. Willich, M. Steilen, J. Kallo, K. A. Friedrich www.dlr.de Chart 2 > SOFC
More informationProceedings of the Sixth Workshop on RF Superconductivity, CEBAF, Newport News, Virginia, USA
FIELD EMISSION IN RF CAVITIES : OBSERVATION OF LIGHT SPOTS AT HIGH ELECTRIC FIELDS T. Junquera, A. Le Goff Institut de Physique Nucleaire (CNRS-IN2P3) 91406 Orsay France B. Bonin, H. Safa, J. Tan DSM/DAPNIA/SEA
More informationThis document is the author s post-print version of this article, i.e. the final draft version after review. The final document will be printed in
1 2 3 4 5 6 7 This document is the author s post-print version of this article, i.e. the final draft version after review. The final document will be printed in Nuclear Instruments and Methods in Physics
More informationCONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination
More information7. What is the current in a circuit if 15 coulombs of electric charge move past a given point in 3 seconds? (1) 5 A (3) 18 A (2) 12 A (4) 45 A
1. Compared to the number of free electrons in a conductor, the number of free electrons in an insulator of the same volume is less the same greater 2. Most metals are good electrical conductors because
More informationPerformance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
More informationRigaku XRD-System Instruction Manual v4/19/03. The Krishnan Group/Wilcox 132, University of Washington
Rigaku XRD-System Instruction Manual v4/19/03 The Krishnan Group/Wilcox 132, University of Washington Contents: - General information - Safety - How to turn on the X-rays - How to turn off the X-rays -
More informationMDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
More informationDU 694. Technical Data Specifications (888-59) (889-03) PX 1494: (874-53) (874-57) (874-54) (874-58) (874-55) (874-59) (874-56) (874-60)
Technical Data Specifications DU 694 (888-59) (889-03) Technical Data 0 KHU Insert / Focal Spots Données Techniques Tubes 4 KJ 0,4/0,8 Foyers Technische Daten 4 KJ Röhre 0,4/0,8 Brennpunkte Datos Técnicos
More informationHIGH CURRENT OPERATION OF THE ACSI TR30 CYCLOTRON
HIGH CURRENT OPERATION OF THE ACSI TR30 CYCLOTRON 18 th International Conference on Cyclotrons and their Applications, Sicily, 2007 Vasile Sabaiduc, Project Manager, RF Senior Engineer Advanced Cyclotron
More informationThe quadrature signals and the index pulse are accessed through five 0.025 inch square pins located on 0.1 inch centers.
Quick Assembly Two and Three Channel Optical Encoders Technical Data HEDM-550x/560x HEDS-550x/554x HEDS-560x/564x Features Two Channel Quadrature Output with Optional Index Pulse Quick and Easy Assembly
More informationX-Ray Diffraction HOW IT WORKS WHAT IT CAN AND WHAT IT CANNOT TELL US. Hanno zur Loye
X-Ray Diffraction HOW IT WORKS WHAT IT CAN AND WHAT IT CANNOT TELL US Hanno zur Loye X-rays are electromagnetic radiation of wavelength about 1 Å (10-10 m), which is about the same size as an atom. The
More informationRadiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
More informationEquipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764
Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine
More informationCALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY
243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution
More informationInstruction Sheet 554 828
08/13-W07-Hund Instruction Sheet 554 828 X-ray image sensor (554 828) Precision slide X-ray image sensor (554 829) Software Computed Tomography Pro (554 820) Sensor head on precision slide with USB module
More informationGeneration of X-Rays (prepared by James R. Connolly, for EPS400-002, Introduction to X-Ray Powder Diffraction, Spring 2005)
A Bit of History A good discussion of the early x-ray discoveries may be found in Chapter 1 of Moore and Reynolds (1997). I have borrowed freely from a variety of sources for this section. An online sketch
More informationElectron Microscopy SEM and TEM
Electron Microscopy SEM and TEM Content 1. Introduction: Motivation for electron microscopy 2. Interaction with matter 3. SEM: Scanning Electron Microscopy 3.1 Functional Principle 3.2 Examples 3.3 EDX
More informationRadiological Worker Training Radiological Safety Training for Radiation Producing (X-Ray) Devices
NOT MEASUREMENT SENSITIVE Appendix C December 2008 DOE HANDBOOK Radiological Worker Training Radiological Safety Training for Radiation Producing (X-Ray) Devices U.S. Department of Energy Washington, D.C.
More informationExploring the Properties of the TV Monitor and Remote Control
Name: LUMINESCENCE It s Cool Light! Class: Visual Quantum Mechanics ACTIVITY 11 Exploring the Properties of the TV Monitor and Remote Control Goal We now explore the properties of a small device that allows
More informationHSeries. High Power High Quality. Ultra-High Speed, Sensing Ionizer SJ-H Series
NEW Ultra-High Speed, Sensing Ionizer SJ-H Series High Power High Quality Suitable for high-speed static elimination in wide areas, including clean room environments HSeries The highest static elimination
More information3. Diodes and Diode Circuits. 3. Diodes and Diode Circuits TLT-8016 Basic Analog Circuits 2005/2006 1
3. Diodes and Diode Circuits 3. Diodes and Diode Circuits TLT-8016 Basic Analog Circuits 2005/2006 1 3.1 Diode Characteristics Small-Signal Diodes Diode: a semiconductor device, which conduct the current
More informationContents. 12. Lot Number 10. 13. Reel Packing Structure 11. 14. Precaution for Use 13. 15. Hazard Substance Analysis 14. 16. Revision History 18
Rev : 00 ISSUE NO : DATE OF ISSUE : 2009. 04. 10 S P E C I F I CATION MODEL : SLHNNWW629T1S0U0S0 [Rank : (S0), (U0), (S0)] HIGH POWER LED - SUNNIX6 CUSTOMER : CUSTOMER CHECKED CHECKED APPROVED SAMSUNG
More informationRESOURCE MANUAL FOR COMPLIANCE TEST PARAMETERS DIAGNOSTIC X-RAY SYSTEMS. Reprinted. July 15, 1999. Diagnostic Devices Branch
RESOURCE MANUAL FOR COMPLIANCE TEST PARAMETERS OF DIAGNOSTIC X-RAY SYSTEMS Reprinted July 15, 1999 Diagnostic Devices Branch Division of Enforcement I Office of Compliance 2098 Gaither Road Rockville,
More informationHistory of the Atom & Atomic Theory
Chapter 5 History of the Atom & Atomic Theory You re invited to a Thinking Inside the Box Conference Each group should nominate a: o Leader o Writer o Presenter You have 5 minutes to come up with observations
More informationElectricity. Investigating spontaneous gas discharge in air as a function of pressure. LD Physics Leaflets P3.9.2.1. 0210-Sel
Electricity Electrical conduction in gases Gas discharge at reduced pressure LD Physics Leaflets P3.9.2.1 Investigating spontaneous gas discharge in air as a function of pressure Objects of the experiments
More informationFig.1. The DAWN spacecraft
Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics
More informationUse of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
More informationTOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] info@rmjordan.com [e-mail] This
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationHygro-Thermometer + InfraRed Thermometer Model RH101
User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative
More informationRadiography: 2D and 3D Metrics of Performance Towards Quality Index
AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..
More informationRADIATION SAFETY STANDARD
Radiation Safety Act 1999 RADIATION SAFETY STANDARD NM004:2010 Standard for radiation apparatus used to carry out industrial radiography Preface Under section 17 of the Radiation Safety Act 1999, a possession
More informationIntroduction to vacuum gauges. Vacuum Gauges where the Pressure Readings are Independent of the Type of Gas (Mechanical Vacuum Gauges)
Introduction to vacuum gauges Vacuum Gauges where the Pressure Readings are Independent of the Type of Gas (Mechanical Vacuum Gauges) BOURDON Vacuum Gauge The inside of a tube which is bent into a circular
More informationCATHODIC PROTECTION TRANSFORMER RECTIFIER (CPTR)
CATHODIC PROTECTION TRANSFORMER RECTIFIER (CPTR) Fig1. CPTR SMART CONTROL PRECISION CUSTOMIZED MANUAL OR AUTO SCR OR DIODE DRY TYPE SWITCH MODE DRY TYPE SCR OR DIODE OIL TYPE 1 INDEX PAGE 1. Cover 2. Index
More informationE/M Experiment: Electrons in a Magnetic Field.
E/M Experiment: Electrons in a Magnetic Field. PRE-LAB You will be doing this experiment before we cover the relevant material in class. But there are only two fundamental concepts that you need to understand.
More informationElectron Charge to Mass Ratio Matthew Norton, Chris Bush, Brian Atinaja, Becker Steven. Norton 0
Electron Charge to Mass Ratio Matthew Norton, Chris Bush, Brian Atinaja, Becker Steven Norton 0 Norton 1 Abstract The electron charge to mass ratio was an experiment that was used to calculate the ratio
More informationOPERATION MANUAL INFRARED THERMOMETER
OPERATION MANUAL INFRARED THERMOMETER Features: Precise non-contact measurements Built-in laser pointer Automatic selection range and Resolution to 0.1ºor 1º / switchable button Automatic Data Hold & Auto
More informationTHEORY OF XRF. Getting acquainted with the principles. Peter Brouwer
THEORY OF XRF Getting acquainted with the principles Peter Brouwer THEORY OF XRF Getting acquainted with the principles Peter Brouwer First published in The Netherlands under the title Theory of XRF. Copyright
More informationAmptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
More information3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses.
1 3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. (A) Any person may apply to the director for approval to conduct an educational program for
More informationData Sheet. AEDR-8000 Series Reflective Surface Mount Optical Encoder. Description. Features. Applications
AEDR-8000 Series Reflective Surface Mount Optical Encoder Data Sheet Description By introducing an additional PCB top-up thickness attached to the encoder, the AEDR-8000 Encoder series could easily solve
More informationView of ΣIGMA TM (Ref. 1)
Overview of the FESEM system 1. Electron optical column 2. Specimen chamber 3. EDS detector [Electron Dispersive Spectroscopy] 4. Monitors 5. BSD (Back scatter detector) 6. Personal Computer 7. ON/STANDBY/OFF
More informationX-ray Tomography in Industrial Metrology
verlag moderne industrie X-ray Tomography in Industrial Metrology Precise, Economical and Universal Ralf Christoph and Hans Joachim Neumann This book was produced with technical collaboration of Werth
More informationMajor: Radiography Degree: Bachelor of Science in Radiography
Major: Radiography Degree: Bachelor of Science in Radiography Prerequisites MAC 1105 SPC 2608 BSC 2085C BSC 2086C STA 2014 CHM 2045C MCB 2010C PSY 2012 PHY 2053 PHY 2053L (32 Semester Hours) College Algebra
More informationLenses and Apertures of A TEM
Instructor: Dr. C.Wang EMA 6518 Course Presentation Lenses and Apertures of A TEM Group Member: Anup Kr. Keshri Srikanth Korla Sushma Amruthaluri Venkata Pasumarthi Xudong Chen Outline Electron Optics
More informationActivitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies
Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies (Ci), where 1 Ci = 3.7x10 10 disintegrations per second.
More information