C RITICAL C ARE. The GEM Premier TM 3000 with Intelligent Quality Management (iqm TM ): Features, Technical Description & Statistical Validation

Size: px
Start display at page:

Download "C RITICAL C ARE. The GEM Premier TM 3000 with Intelligent Quality Management (iqm TM ): Features, Technical Description & Statistical Validation"

Transcription

1 C RITICAL C ARE 9 Kevin Fallon, Ph.D. Sohrab Mansouri, Ph.D. The GEM Premier TM 3000 with Intelligent Quality Management (iqm TM ): Features, Technical Description & Statistical Validation

2 Kevin Fallon, Ph.D. Sohrab Mansouri, Ph.D. The GEM Premier TM 3000 with Intelligent Quality Management (iqm TM ): Features, Technical Description & Statistical Validation

3 Table of Contents 3 I. Features Page 5 II. Technical Description Page 10 III. Statistical Validation Page 17 IV. Glossary of Acronyms Page 21 Appendix Page 22

4 I. Features 5 What It Is What It Does How It Works The new Intelligent Quality Management system (iqm TM ) from Instrumentation Laboratory (IL), is a fully automated Quality Assurance system for use on IL s GEM Premier 3000 analyzer. iqm replaces the need for conventional external Quality Control (QC) with its combination of software, Process Control (PC) Solutions, and Calibration Validaton Product (CVP) components. Designed to ensure at least the same level of quality as traditional QC methods, iqm significantly reduces time and costs of performing QC, while helping to assure regulatory compliance and improving the quality of test results. iqm monitors performance of all critical components of the GEM Premier The "intelligent" quality management system: Validates cartridge integrity Identifies changes that affect analytical performance and potential failure patterns Automatically performs corrective actions Automatically documents the failure and corrective actions taken Unlike external forms of QC, iqm monitors performance real-time, conducting: System checks Sensor checks Stability checks Pattern checks System checks include fluidics checks to test sample integrity, reagent solutions, and the peristaltic pump. Additionally, system checks test mechanical components of the cartridge, such as the valve position and sample arm position, as well as the instrument heater block and electronics. All system checks are performed on an ongoing basis.

5 6 If any system failure is found, iqm confirms the nature of the failure, then either rejects the cartridge or halts instrument operation. Sensor checks are performed using three different solutions, each at different intervals. PC Solution B is measured after every patient sample, and in the event the instrument is unused, every half hour. Sensor outputs are monitored in PC Solution B every thirty seconds. During these checks, the drift is analyzed to assess the difference between a measured value and the expected value. To further test for drift and slope, iqm uses fresh PC Solution A every four hours, and fresh PC Solution C every twenty-four hours. iqm confirms any sensor failure and either suppresses the result of a failed parameter, or disables any sensor that shows persistent failure. Process stability checks are also performed in order to verify the process control solution stability during the life of a cartridge. In the event of process stability failure, the cartridge is rejected. (Solutions tend to be very stable; the feature simply further ensures accuracy and precision in test results.) Last, iqm performs Failure Pattern Recognition (FPR) Checks. Common sources of failure, such as micro-clots on sensors, any sensor malfunction, or sensor interference, leave "fingerprints" on the sensors. These fingerprints are read as patterns that iqm can detect and attempt to correct. For example, in the event of a micro-clot, iqm will automatically perform a special rinse. If it cannot correct the problem, the sensor will be disabled. Additionally, iqm can use FPR to warn of the presence of interference in a sample. See "Technical Explanation" and "Statistical Validation" for detail and supporting data. Why It Works: Compatibility with the GEM Premier 3000 Designed specifically to assure accurate and precise test results, iqm exemplifies the simplicity and ease-of-use features associated with the GEM Premier 3000 system. The following unique elements make iqm possible on the GEM Premier 3000: All analytical components (e.g., sensors, solutions, tubing, sampler) of the GEM Premier 3000 system are included in a self-contained, multi-use, disposable cartridge (PAK).

6 7 The GEM PAK is a completely closed system. The user cannot introduce changes to the analytical system during the three-week use-life of the PAK. The GEM PAK constitutes a "run," per the National Committee for Clinical Laboratory Standards (NCCLS) definition (the period of time during which an analytical system is expected to be stable). The GEM PAK solution values are traceable to NIST standards and stated in each PAK barcode. Each sensor and each lot of solutions manufactured by IL are functionally tested prior to PAK assembly. In addition, a representative sample of each PAK batch is functionally tested. Therefore, prior to release of each batch of PAKs, IL is able to test the complete analytical system, including the actual combination of analytical elements (e.g., sensors, solutions, tubing, sampler) that will eventually be used by the operator. The GEM Premier 3000 analyzer records comprehensive information about all used PAKs, including all calibrations and millivolt (mv) sensor readings. This has allowed IL to study and understand the patterns generated by any PAK failure. Immediately after insertion of the iqm PAK into the GEM Premier 3000, and before the system will report any results, an external CVP is run to validate the integrity of the PC Solutions and overall performance of the analytical system. Why It's Better Than Conventional QC Methods In order to assure quality results from a blood gas analyzer, hospital staff routinely perform daily liquid QC. This process, as well as monitoring, interpreting and documenting the results, places a significant burden on hospital staff. Traditional QC methods pose many challenges, which may even compromise the quality and accuracy of test results. For example, with traditional QC: Manual process requires skilled labor Aqueous QC ampoules must be run immediately due to the volatility of O 2 and CO 2 Corrective action involves potentially complex processes and skills Manual documentation is needed for regulatory compliance Discrete process may leave errors undetected until a QC or a series of QCs are run (after hours or days) The alternative QC methods advocated by various manufacturers of critical care analyzers, namely on-board automated quality control or electronic QC (EQC), are still inadequate in addressing all of the issues previously mentioned. On-board automated quality control systems are not capable of continually monitoring the system functionality and do not address the manual corrective action and documentation issues. EQC only confirms the operation of the instrument s measurement circuitry and its computational components. However, EQC fails to provide information about the chemical measurement process, which includes the sensors and reagents, and it is a manual, discrete process. iqm has been developed to address all of these issues. iqm is an active quality process control program designed to help ensure that the GEM Premier 3000 analyzer provides reliable results at all times. The quality control process is an integral part of system operation, and reduces time to error detection from hours to minutes.

7 8 Calibrators and Controls iqm continuously monitors operation of the entire testing process including, sensors, fluidics and electronics, and automatically performs and documents corrective actions upon detection of an error. Running traditional liquid quality controls are no longer needed. iqm Traditional QC Electronic QC Auto QC Continuous control Intermittent control Intermittent control Intermittent control Automatic corrective Manual corrective Manual corrective Manual corrective action action action action Automatic Manual Manual Manual documentation documentation documentation documentation Automated Labor-intensive Manual operation Manual maintenance operation operation of external modules Integrated in GEM Additional QC Additional QC Additional QC and/or Premier 3000 PAKs materials required simulators required hardware required Total Analytical May leave errors Electronic component May not check the System QC undetected check only full fluidic pathway iqm cartridges utilize PC Solutions A, B and C to perform multiple functions: To indicate if the system has drifted from the reference points established after cartridge insertion and verification with CVP As a reference, to eliminate natural sensor drift ( calibrate ) To identify abnormal sensor behavior drift and trigger FPR software to determine the cause of the failure, and initiate corrective action To determine the status of the analytical system, rather than a single analysis of an external quality control sample. This is achieved by analyzing the collective values of the three PC Solutions, tested hundreds of times during the life of a PAK, along with FPR. The perception that the same solution cannot be used for two different functions most often originates from experience with other types of diagnostic equipment, such as clinical chemistry systems. In the past, calibrators and controls required manual reconstitution. Reagents were also reconstituted and changed, sometimes within the course of a day. Tubing changes and other maintenance functions that may affect system performance, were also performed routinely. A separate control material was a check that the calibrator was reconstituted correctly, or had not expired, and that other changes to the system did not alter calibration. The GEM Premier 3000, as previously discussed, is a closed system. The operator cannot change any of the internal cartridge components, and the cartridge will not be accepted by the system if the expiration date has past. Therefore, prior concerns that resulted in separate solutions being used for calibrators and controls do not exist with the GEM Premier 3000 iqm cartridge. PC Solutions play important roles within the iqm cartridge. Therefore, a Process Stability check is performed on the solutions, in the very unlikely event that all PC Solutions deteriorated at the same rate. The Process Stability check is a method to verify PC Solution stability throughout the cartridge use-life. The measured oxygen in PC Solution A during the use-life is compared to the very first measured PC Solution A oxygen value during warm-up. The delta must be within allowable limits. The PO 2 in PC Solution A is used for the Process Stability check because:

8 9 Oxygen is considered the most sensitive parameter for detecting deterioration in PC Solutions, since they have no oxygen-buffering capacity. The process of measuring oxygen in PC Solution A utilizes all three PC Solutions. Therefore, deterioration in any of the PC Solutions will affect the measured oxygen in the PC Solution A. In the event of Process Stability failure, the cartridge will be rejected. There are PC Solutions used to perform additional sensor checks that do not perform a secondary function for a particular sensor within the cartridge. However, only ph, PCO 2 and PO 2 sensors require the additional pattern check for identifying malfunctions that are very rare and very slow in progression. Therefore, a PC Solution not used in the initial cartridge verification process is used to further evaluate the performance of these sensors. PC Solution C serves this purpose for ph and PCO 2, and PC Solution A provides an additional measurement for PO 2. The GEM Premier 3000 with iqm initiates a new paradigm in quality management. It is the first clinical instrument to apply the Process Control approach in the laboratory setting: determine the quality requirements, identify the potential sources of error, and design a method of control for each potential failure. Statistical analysis of iqm compared to traditional QC has shown that iqm is better than traditional controls at detecting failures that could result in clinically significant errors. iqm can detect more errors, and detect them sooner. iqm Benefits iqm provides laboratories and point-of-care testing locations with the following key benefits: Quality Assurance Active, continuous, real-time quality process Detects and attempts to correct errors immediately Detects errors sooner than external QC, reducing time to error detection from hours to minutes Ensures that the optimal quality control protocol is followed at all times, regardless of the time of day or operator training Patient Safety Helps assure the quality of each patient result Helps assure that the right medical decision is made for the patient Prevents reporting sample results when there is a risk of reporting an incorrect result Cost Reduction Eliminates all manual processes associated with traditional quality control Performs automatically, with no human intervention Documents failures and corrective actions automatically, with no human intervention Eliminates the labor required to maintain a quality control protocol, train operators, run routine external quality control material (or the maintenance and refilling of mechanical ampoule beakers), document quality control results, troubleshoot when a failure is identified, and document the failure and corrective action taken Produces reports required by regulatory agencies Reduces inventory management and procurement procedures

9 10 II. Technical Description Understanding GEM Premier 3000 Operation This section provides a detailed description of the GEM Premier 3000 operation, a prerequisite in understanding how iqm functions. Sampling Position System description Figure 1. Internal components of the GEM Premier 3000 cartridge The GEM Premier 3000 system includes two components: the instrument and a disposable cartridge. The cartridge measures ph, PCO 2, PO 2, Na +, K +, Ca ++, Glucose, Lactate and Hematocrit. The following provides an overview of the cartridge: All required components for sample analysis are included in the cartridge, including sensors, sampler, pump tubing, distribution valve, waste bag and PC Solutions. The cartridge components and fluidic path are schematically illustrated in Figure 1. The sensor card contains all of the sensors in a gas-tight chamber. The sensors are monitored with three PC Solutions: A, B and C. These solutions are pre-tonometered and contain known quantities of the analytes tested using known reference standards. The solutions are sealed in gas impermeable bags with no headspace, allowing their use over a wide range of ambient temperatures and atmospheric pressures. PC Solution B is also used for rinsing. There is a fourth bag called "Reference Electrode Solution" that contains silver ion. This solution is pumped into the reference channel in the sensor card to form the Ag/Ag+ reference electrode. The sensor card resides in a thermal block, which maintains the temperature at 37 C and provides an electrical interface to the sensors. The peristaltic pump moves various fluids (sample, PC Solutions and reference solutions) into the sensor card and eventually to the waste bag.

10 11 Cartridge Operation When the cartridge is inserted in the instrument, all information associated with the cartridge is read from the cartridge bar code and recorded. This information includes cartridge type (specifying menu, sample capacity and use-life), expiration date and PC Solution values. Once the cartridge bar code is validated (new cartridge with valid expiration date), the sensors are hydrated and calibrated within 30 minutes before the analyzer becomes ready for use. Cartridge Quality Assurance Testing Cartridge Components Recording Cartridge Data Understanding iqm Operation Cartridge quality assurance is achieved primarily by two methods: 1. Testing critical components of the cartridge before assembly 2. Downloading all cartridge data for investigation The critical components of the cartridge, sensor card and PC Solutions, are designed for testing before cartridge assembly. Unlike sensors in other cartridge-based systems, the GEM Premier 3000 sensor card can tolerate many cycles of hydration and drying without any deterioration in sensor performance. Furthermore, the fluid opening valve in the reagent bag is configured for sampling without affecting its solution composition. All manufactured sensor cards are tested in an elaborate and automated computer-controlled testing system. The test utilizes PC Solutions similar to those in the cartridge, plus an additional test solution for performance evaluation. Only the sensor cards that pass stringent criteria based on sensitivity (slope), accuracy, and drift are assembled in cartridges. A statistically significant number of reagent bags from a production lot are tested for quality assurance and for assigning analyte values. The value assignment process is traceable to National Institute of Standards and Technology (NIST) standards. The assigned concentration values are included in the cartridge bar-code. All system and cartridge operations are recorded on the instrument hard drive. The recorded data is very comprehensive. For example, every sensor mv, every sample or PC Solution measurement, and every check or alarm is recorded. This information can be downloaded onto a floppy disk or directly ed from the instrument through the GEMweb feature. In case of cartridge malfunction, the user can send the cartridge data to the complaint investigation department at IL for analysis and failure determination. iqm operation is achieved by adding additional layers of checks to the GEM Premier 3000 operation. iqm uses FPR to identify cartridge malfunctions that are currently undetected by internal system and sensor checks. iqm operation can be summarized as follows: Monitors performance of the system in real-time Identifies potential failure patterns Automatically performs corrective actions Automatically documents the failure and action taken

11 12 iqm operation begins following cartridge verification after warm-up. External CVP is used to independently verify cartridge calibration. Four ampoules of CVP (two levels for ph, blood gases, electrolytes and metabolites, and two levels for hematocrit) must pass before the cartridge can be used for sample reporting. If one or more of the analytes fails CVP testing, those channels will not be available for sample analysis. Once cartridge calibration is verified, iqm monitors the status of the calibration during cartridge use-life. Upon detecting a problem, the analyzer automatically performs corrective actions. The following are examples of corrective actions: Automatically performing a special rinse cycle, then subsequently verifing cartridge function if micro-clots are detected Permanently disabling a failed sensor if its functionality cannot be recovered Rejecting a cartridge for process stability failure Alerting the operator if an interfering substance in the sample is detected During cartridge operation, the instrument automatically and continuously performs various checks that can be categorized into four groups: 1. System checks 2. Sensor checks 3. FPR checks 4. Process Stability checks System Checks Sensor Checks System checks include basic functionality of the instrument and the cartridge Examples of these checks include: Cartridge fluidic checks to assess sample integrity, detect the presence of PC Solutions, and verify peristaltic pump functionality Cartridge mechanical checks, such as proper operation of the distribution valve and sampler arm Instrument heater-block checks, such as heater-block temperature and mv-output threshold from sensors (e.g., outputs at rail) Instrument electronic checks, such as analog/digital electronic calibration verification and communication between processors Any failure in the system checks will lead to a corrective action. The corrective action will include verification of the failure followed by one of the following steps: Cartridge rejected, in case of cartridge-related failure Instrument operation halted, in case of instrument-related system failure Sensor checks address functionality of the sensors. PC Solutions A, B and C are automatically brought into the sensor card at various intervals to verify sensor operation. The solution that is residing in the sensor card is measured and the drift is determined. (Drift is the delta between the measured and expected value.) The expected values are obtained from either the cartridge bar-code or from prior PC Solution measurements. Sensor checks are performed with the following frequencies: PC Solution B is measured most frequently. Fresh B is measured, at minimum, every 30 minutes or after every sample. Furthermore, PC solution B is measured every 30 seconds while residing in the sensor card in between fresh B measurements.

12 13 PC Solution A is measured, at minimum, every 4 hours. All sensor slope values are also measured and checked. Slope, which is an indicator of sensor sensitivity, must be within allowable limits. PC Solution C is measured, at minimum, every 24 hours. PC Solution C is primarily used for measuring low-level oxygen and for verifying performance of the ph and PCO 2 sensors. It is also used for conditioning the glucose and lactate inner membranes, which provide rejection of interference compounds present in blood. It should be noted that the PC Solution B and PC Solution A measurements are transparent to the operator. The measurement can be interrupted at any time to run a sample (except the A measurement within the first 4 hours of cartridge life, which is uninterruptible). The PC Solution C measurement is uninterruptible, but the operator has the option of selecting the exact time of day for performing PC Solution C. The process of verifying sensor operation by measuring PC Solutions is very similar to the process of sample measurement. As indicated in Figure 2, the sample path and the PC Solution path into the sensor card is identical. In addition, the algorithms for calculating the PC Solution values are similar to those for sample calculations using prior calibration values. After performing sensor checks, the analyzer conducts a number of actions. If all measured values are within allowable limits, the sensors will be calibrated and, as a result, the drifts will be set to zero. If any measurement or slope value is outside the allowable limits, the following corrective actions will take place: Parameter result in subsequent sample report will be suppressed If the failure fits with a known pattern, specific corrective actions will follow (see FPR checks). If not, fresh solution will be brought into the sensor card for up to two more times and measured If the failure persists in two consecutive C measurements, in three consecutive A measurements, or 15 consecutive B measurements, the failed parameter will be permanently disabled Home Position Figure 2. Similarities between sample path and PC Solution path

13 14 FPR Checks Micro-Clot Patterns FPR checks were developed through years of investigating field complaints. As indicted in the previous section, GEM analyzers (GEM Premier 3000 as well as the GEM Premier) provide comprehensive information about used cartridges. This information, plus in-house investigations about the cause of failures, provided the background for the FPR checks. Two distinct failure patterns were identified: micro-clot-related failures and certain sensor malfunctions that are not well detected by other internal checks or by external QC. Furthermore, certain interference patterns were also specified. Micro-clots are small blood clots or fibrin strands that adhere to a sensor and induce a change in sensor characteristics, such as sluggish response or sensitivity change. Micro-clot patterns are distinct for various sensors. Sensorcheck failures (drift errors) are used to identify the presence of micro-clots. In the event of micro-clot pattern detection, iqm automatically initiates a special rinse cycle, using PC Solution C (PC Solution B is used for normal rinse). PC Solution C has added surfactants for better clot removal. Upon completion of the rinse cycle, iqm checks for a clot pattern on the affected sensor. If a clot pattern still remains, the affected sensor is disabled. If a clot pattern is not detected, the sensor status becomes green (ready for measurement). Figure 3 illustrates an example of micro-clot formation on the oxygen sensor and the pursuing corrective actions. In this example, the detected pattern Rinse Ready B B B B B B B B B B B B B PO 2 Output (mv) Sample # 17 A Sample # 21 A Sample # 19 Sample # 20 Sample # 18 Clot Removal Verified Clot Detected 10 Clot Verified Cartridge Use Life (hr) Measured B PO 2 (mmhg) B drift Micro-Clot Pattern Detected - Negative B drift error after sample - Positive A drift error Corrected 120 Measured A A drift A & B Sequence Figure 3. Example of Micro-Clot Pattern for PO 2 Sensor

14 15 Sensor Malfunction Patterns Interference Patterns was a negative drift error in the PC Solution B oxygen check after sample #19, followed by a positive drift error in the oxygen PC Solution A check which was performed automatically and immediately after the PC Solution B check failure. After confirming the presence of a micro-clot (by the PC Solution A check), the special rinse cycle was initiated. Following the rinse cycle, the removal of the clot was verified by re-running PC Solution B and A checks. In this example, the entire process of detection, clot removal and verification took about 10 minutes. Only ph, PCO 2 and PO 2 sensors need an additional pattern check for identifying malfunctions. Existing sensor checks are found adequate for detecting any malfunction in other sensors. It should be noted that normal sensor malfunctions for ph and PCO 2 sensors are identified with the existing sensor checks. The specific malfunctions for which iqm is checking in these sensors are very rare and very slow in progression. Therefore, the daily PC Solution C check is adequate in detecting these malfunctions. In case of a sensor malfunction pattern, the affected sensor is permanently disabled by iqm. Two interference patterns are checked, one for positively charged lipophilic compounds, and one for negatively charged lipophilic compounds. These compounds can cause false readings for a few of the ion-selective electrodes. Benzalkonium Chloride is a good example of a positively charged lipophilic compound that can cause false-elevated readings for sodium and ionized calcium. Sodium Thiopental is an example of a negatively charged lipophilic compound. For an interference pattern to be positively identified, all conditions must be met. In the event of interference pattern detection in a sample, iqm notifies the user. Figure 4 illustrates an example of Benzalkonium interference on the ionized calcium sensor. The detected pattern was a positive drift error in the ionized calcium PC Solution B check, along with certain drift movements in the sodium and potassium sensors. Following the interference detection, PC Solution B is performed at a higher than normal frequency (every 3 minutes versus every half-hour). 50 Sample with Benzalkonium 45 Sensor Output (mv) Sample B drift error Cartridge Life (hours) Figure 4. Example of Interference Caused by a Positively Charged Lipophilic Compound

15 16 Process Stability Check Process Stability check is a method of verifying the PC Solution stability throughout the cartridge use-life. The measured oxygen in PC Solution A during the use-life is compared to the very first measured PC Solution A during warm-up. The delta has to be within allowable limits. The PO 2 in PC Solution A is used for the process stability check because: Oxygen is considered the most sensitive parameter for detecting deterioration in PC Solutions, since they have no oxygen-buffering capacity The process of measuring oxygen in PC Solution A utilizes all three PC Solutions, so that deterioration in any of the PC Solutions will affect the measured oxygen in PC Solution A. In the event of process stability failure, the cartridge will be rejected.

16 III. Statistical Validation 17 Statistical Evaluation of Drift Limits 2 Statistical Method 2 As indicted in the previous sections, drift limits are used as a trigger for sensor and FPR checks, and for subsequent iqm corrective actions. Therefore, drift limits should be optimized for high probability of error detection and low probability of false rejection. This section explains how statistical control methods are used for evaluation of drift limits. Drift limits on PC Solutions A and B can be characterized as a single measurement of a control material. Statistical control methods are then used to develop probabilities for error detection and false rejection. This approach allows a comparison of performance expected for iqm with performance of traditional QC procedures. The method is as follows: Define the quality requirement in terms of total allowable error (TEa) - ph, PCO 2, Na +, and K + = CLIA 88 limits - PO 2 = 10% - Ca ++ = 9% - Glucose = 12% or 12 mg/dl - Lactate = 15% or 0.4 mmol/l - Hematocrit = 2% Define method performance -Mean and SD values are obtained from the data collected from 24 cartridges representing a wide variety of uses from customers in the field and in-house tests Predict QC performance - Calculate Method Sigma = TEa/SD -Calculate Control Limit = Drift Limit/SD - Determine probability of false rejection (Pfr) from normal probability distribution (from tables of areas under normal curve, or z charts) Pfr = Prob(z Control Limit) -Determine probability of error detection (Ped) with 95% confidence from normal probability distribution Ped = 1 Prob(z (Method Sigma Control Limit 1.65)) - Calculate Average Run Length for rejectable quality ARLr = 1/Ped - Determine average detection time (unit of time for detecting error that can be compared to traditional QC) Average detection time = ARLr x sampling time As specified previously, sampling time for Solution A is between 1 to 4 hours and for Solution B is between 3 and 30 minutes (3 minutes when there is a sample in between B measurements). For a given TEa, the drift limits must provide a high probability of error detection (Ped 1) and a low probability of false rejection (Pfr 0). Tables 1 and 2 provide a summary of the analysis of the drift limits for PC Solutions A and B. Results of the drift limit analysis indicate that the probability of false rejection is close to zero for all parameters in PC Solutions A and B. PC Solution B is the primary means for error detection because of high measurement frequency. Probability of error detection is high in the PC Solution B. Even for sodium and glucose with low Ped values, the average error detection time is within 10 minutes.

17 18 Table 1. Statistical analysis of drift limits for PC Solution A ph PCO 2 PO 2 Na + K + Ca ++ Glu Lac Hct Mean SD TEa Method Sigma Drift Limit Control Limit Pfr z for Ped Ped ARLr ATDr (4 hr) *Compare with a typical QC program where QC is performed only every 8 hours Table 2. Statistical analysis of drift limits for PC Solution B ph PCO 2 PO 2 Na + K + Ca ++ Glu Lac Hct unit mmhg mmhg mmol/l mmol/l mmol/l mg/dl mmol/l % Mean SD TEa Method Sigma Drift Limit Control Limit Pfr Ped ARLr Average Detection Time (min)* Evaluation of Cartridge Malfunction Detection The effectiveness and efficacy of iqm error detection was investigated using the existing field complaint data files for the GEM Premier 3000 with Phase II cartridge configuration. (Phase II cartridge configuration has the capability to include glucose and lactate). It should be noted that the FPR checks were originally developed from field and in-house data using the GEM Premier 3000 and Phase I cartridge configurations. The investigation focused on all reported field QC failures from US accounts beginning with the release date of Phase II cartridges in June 2001, to the end of March (Only QC failures were selected because iqm is expected to detect cartridge malfunctions currently identified by running external quality controls.) Approximately 7000 cartridges were shipped to U.S. GEM Premier 3000 accounts during this period. Summary data is outlined below. A total of 164 QC-related cartridge failures were reported to IL From 164 reported QC failures, IL obtained 144 data disks From 144 obtained data disks, IL confirmed 117 real QC failures (others either had no QC failure or had failures other than QC) From 117 confirmed QC failures, 79 passed initial QC out of warm-up, but failed QC later

18 19 Data from the 79 GEM Premier 3000 cartridges that had confirmed QC failures during cartridge use-life was extensively analyzed. The FPR checks were applied to the data to determine if iqm could detect any malfunction. The results are summarized below. 68 of the data files exhibited micro-clot failure patterns. - PO 2 : 46 - Hematocrit: 16 - Ca ++ : 2 - ph: 1 - PO 2 and Hematocrit: 2 - PO 2, Na + and Ca ++ : 1 6 showed a sensor failure pattern. - PCO 2 : 4 - ph: 2 3 demonstrated one QC level at the limit of the acceptable range with a subsequent QC failure. - PO 2 : 2 - Glucose: 1 2 had no identifiable failure pattern; both QC failures were marginal and occurred at only one level. - Na + : 1 - Ca ++ : 1 This analysis indicates that FPR was able to detect malfunction in 77 out of 79 reported QC failures. It is interesting to note that in most cases, FPR flagged the failure immediately after the sample that caused the malfunction; while the operator typically became aware of the problem only later when the QC was run and failed. In some cases, QC had been run several hours after occurrence of the malfunction. Furthermore, no false-positive flag was generated by FPR in all 79 investigated cases. No obvious malfunction was detected in the data from the two cartridges that FPR did not flag but had reported QC failure. All system parameters were found within specifications. Blood results for the parameter with reported QC failure were examined and found reasonable. In both cases, the reported QC failures were marginal and at one level only. It was concluded that the reported QC failures in those two cartridges could not represent a serious cartridge malfunction. The failures could be considered as a false-positive QC failure. iqm Error Detection During Limited Distribution During a limited distribution phase, iqm was implemented in 20 hospital sites. These selected sites represented high volume locations, many with a history of pre-analytical issues tha may affect patient results. The issues included samples containing micro-clots and samples contaminated with Benzalkonium. Data from 42 cartridges from the various clinical sites, using various cartridge types, were analyzed at IL. Cartridge files with reported iqm failures were of particular interest in the investigation. Two-thirds of the cartridge data collected showed no cartridge issue, while one-third had at least one reported problem during cartridge use-life. The iqm Corrective Action Report was used to identify any cartridge issue. Additionally, the sensor output data files, the sensor calibration data files, and the system logs were examined to confirm the validity of iqm reported errors.

19 20 iqm correctly flagged the presence of micro-clots in 11 cartridges. Five of the micro-clots were successfully rinsed out, while 6 could not be removed, and the analyzer automatically disabled the affected sensors. There were 5 flags for Benzalkonium interference on the sodium and ionized calcium measurements. The sodium and ionized calcium values in those flagged samples were abnormally high. In all cases, iqm flagged the problem within minutes of the occurrence. iqm was very effective in identifying cartridge issues promptly. The automatic corrective action function of iqm was effective in removing about one-half of the detected micro-clots. Summary iqm is an active, real-time quality process control program that allows for immediate error detection and correction, thus further enhancing quality assurance in the GEM Premier 3000 analyzer.

20 IV. Glossary of Acronyms 21 ARLr: CAR: CVP: EQC: FPR: iqm: IL: Average Run Length for Rejectable Quality Corrective Action Report Calibration Validation Product Electronic Quality Control Failure Pattern Recognition Intelligent Quality Management Instrumentation Laboratory NCCLS: National Committee for Clinical Laboratory Standards NIST: PC: Ped: Pfr: QC: SD: TEa: National Institute of Standards and Technology Process Control Probability of Error Detection Probability of False Rejection Quality Control Standard Deviation Total Allowable Error

21 22 Appendix AI. Interferences Note: the following information is based on data obtained during the development of iqm cartridges; please refer to the complete GEM Premier 3000 Operator s Manual (Revison 3) for complete specifications on all available GEM Premier 3000 cartridges. The following substances can potentially interfere with sample analysis: Benzalkonium Chloride and Benzalkonium Heparin: Arterial lines and sampling devices coated with these substances may interfere with Na + and Ca ++ determinations, causing falsely elevated Na + and Ca ++ readings. Following sample analysis, and analysis of PC Solution B, if Benzalkonium Chloride or Benzalkonium Heparin patterns are detected, the following message will be displayed on the analyzer, and will persist until acknowledged by the operator: Sensor Interference Detected for Na and ica on last sample likely due to Benzalkonium The GEM Premier 3000 offers the operator the ability to enable flagging of patient results if an interference pattern is detected. In addition, this option, when enabled, delays the reporting of results until PC Solution B is evaluated following sample analysis. If flagging of patient results is enabled, the following message (plus progress bar) will be presented while the post-analysis PC Solution B check is underway: Checking for presence of interference and micro clots This message will remain displayed until the analysis of PC Solution B, is complete. If an interfering substance pattern is detected, the affected blood result(s) will be flagged. In addition, the analyzer will beep three times to alert the operator. The following message disappears only after operator acknowledgment: Sensor Interference Detected for Na and ica on last sample likely due to Benzalkonium Thiopental sodium may interfere with the Na +, K +, PCO 2 and Ca ++ readings. Thiopental Sodium is also known by other names, including: Thiomebumal Sodium, Penthiobarbital Sodium, Thiopentone Sodium, Thionembutatal, Pentothal Sodium, Nesdonal Sodium, Intraval Sodium, Traoanal, and Thiothal Sodium. Following sample analysis and analysis of PC Solution B, if the associated pattern is detected in PC Solution B, the following message will be displayed on the analyzer, and will persist until operator acknowledgment: Sensor Interference Detected for xxxxx on last sample (where xxxx is the analyte or analytes affected) When flagging of patient results for an interference is enabled, the following message (plus progress bar) will be presented while the post analysis PC Solution B check is underway: Checking for presence of interference and micro clots This message will remain displayed until the PC Solution B is complete. At that point, processing will continue until results are displayed. If Thiopental Sodium is detected, the affected blood result(s) will be flagged. In addition, the analyzer will beep three times to alert the operator. The following message disappears only after operator acknowledgment. Sensor Interference Detected for xxxxx on last sample (where xxxx is the analyte or analytes affected)

22 23 AII. Performance Characteristics Summary GEM CVP Precision Precision data were generated at IL using GEM CVP: 2 levels for ph, blood gases, electrolytes and metabolites and 2 levels for Hematocrit. Based on NCCLS guidelines, the verification material levels were run in singlet once a day for 14 days (twice on day 1) for a total of 15 replicates on each of 9 different GEM Premier 3000 instruments (N=135). The table below lists the combined results of the 9 instruments. Note: SD is used for ph since differences are so small that %CV would be misleading. GEM CVP Level 1: Parameter Mean Day-to-Day %CV (or SD) Total %CV (or SD) ph (SD) (SD) PCO 2 (mmhg) PO 2 (mmhg) Na + (mmol/l) K + (mmol/l) Ca ++ (mmol/l) Glucose (mg/dl) Lactate (mmol/l) GEM CVP Level 2: Parameter Mean Day-to-Day %CV (or SD) Total %CV (or SD) ph (SD) (SD) PCO 2 (mmhg) PO 2 (mmhg) Na + (mmol/l) K + (mmol/l) Ca ++ (mmol/l) Glucose (mg/dl) Lactate (mmol/l) GEM CVP Level 3: Parameter Mean Day-to-Day %CV Total %CV Hematocrit (%) GEM CVP Level 4: Parameter Mean Day-to-Day %CV Total %CV Hematocrit (%)

23 24 AIII. Method Comparison Arterial, venous, heart bypass and liver transplant blood samples were obtained from patients using heparinized syringes. The table and graphs below demonstrate that the GEM Premier 3000 using an iqm cartridge is statistically similar in performance to a reference analyzer. Analyte N Slope Intercept r Sample Range ph PCO 2 (mmhg) PO 2 (mmhg) Na + (mmol/l) K + (mmol/l) Ca ++ (mmol/l) Glucose (mg/dl) Lactate (mmol/l) Hct (%) ph Method Comparison ph Units iqm GEM ph Units Reference y = x r = n = mmhg iqm GEM PCO Method Comparison 2 mmhg Reference y = 1.674x r = n =

24 PO Method Comparison mmhg iqm GEM mmhg Reference y = x r = n = Na Method Comparison mmol/l iqm GEM mmol/l Reference y = x r = n = K Method Comparison 7.5 mmol/l iqm GEM mmol/l Reference y = x r = n =

25 Ca ++ Method Comparison 1.5 mmol/l iqm GEM y = 0.962x r = n = mmol/l Reference 450 Glucose Method Comparison mg/dl iqm GEM mg/dl Reference y = x r = n = mmol/l iqm GEM Lactate Method Comparison mmol/l Reference y = x r = n =

26 27 % iqm GEM Hematocrit Method Comparison % Reference y = x r = n =

27 28 AIV. iqm Reports iqm Delta Charts An iqm Delta Chart is generated for each Process Control Solution and analyte combination. A Delta Chart, depicted in Figure 5, includes: 1. Analyzer (GEM Premier 3000) 2. Analyte name 3. Analyzer serial number and name, if the analyzer is given a unique name by the facility 4. Nominal target value for the analyte 5. Report month and year 6. PC Solution 7. The number of times the PC Solution was run each day (If the number of points exceeds 99, 99 will be displayed) 8. The maximum delta between the Process Control Solution target value and the actual measured value for the selected analyte each day, represented by a short horizontal line 9. The mean delta between the Process Control Solution target value and the actual measured values for the selected analyte each day, depicted by a round, bolded dot 10. The maximum delta between the Process Control Solution target value and the actual measured value for the selected analyte each day, represented by a short horizontal line 11. A vertical line, which connects the maximum, mean and minimum delta points 12. The maximum allowable delta for the analyte 13. The minimum allowable delta for the analyte 14. The date of cartridge insertion (indicated by an arrow), along with the cartridge lot number } Figure 5. iqm Delta Chart: Components.

28 29 Delta points outside the designated limits will not be shown on the Delta Chart, but will be included in the iqm Corrective Action Report. This is because the GEM Premier 3000 requires the use of static graphs (graphs with ranges that do not change). Please refer to Figure 6 for graphical representations of the following situations: 1. It is possible to have the minimum and maximum delta values coincide with one another and the mean delta result. This happens when: The same delta value is obtained for an analyte each time a PC Solution is run. Only one delta value is obtained for an analyte through the course of a day, which is expected for PC Solution C. 2. A minimum or maximum delta result (represented by a horizontal line) may coincide with a daily mean delta result. An example of this occurring is: K + is measured in PC Solution B 90 times in one day 89 x the delta value = 0 1 x the delta value = In this case, the mean and minimum delta values will be the same (0). The horizontal line representing the minimum delta value will intersect the round, bolded dot representing the mean value. Furthermore, since the minimum delta value is 0, the horizontal line representing the minimum delta result will coincide with the 0 horizontal line on the graph. 3. If the maximum and/or minimum delta obtained during a day coincides with the maximum or minimum allowable delta limits, then the horizontal line will blend with the upper/lower graph line limits. Figure 6. iqm Delta Chart: Specific Delta Value Representations.

29 30 iqm Corrective Action Report (CAR) The Corrective Action Report contains information for significant events that occur during the cartridge on-board use-life. The CAR, shown in Figure 7 includes: 1. Analyzer, analyzer serial number and name 2. Month and year 3. Date and time event occurred 4. Cartridge Lot Number 5. Detected failure description 6. Operator ID, if entered (if the function included operator interaction) 7. Corrective action description 8. Corrective action result 9. A Cartridge Removal entry that lists the total number of Process Control solution adjustments made during the cartridge use-life. The adjustments represent the total number of minor drift errors that are corrected by reanalyzing Process Control solution B. Events, such as interfering substance detection, micro-clot detection, and fatal errors, will be listed and described as individual events on the log { Figure 7. iqm Corrective Action Report-Components. Calibration Validation Product Reports Monthly CVP reports can be printed. A CVP report contains results for all CVP ampoules run within the specified month.

30 31 References 1. GEM Premier 3000 Operating Software, Volume 2, Cartridge Internal Operations, Software Version 5.2, Revision 2.7. Biographies of Authors 2. Westgard JO, Fallon KD, Mansouri S. Validation of iqm Active Process Control Technology Point of Care, 2003; Vol 2 (1): Thiopental Interference in GEM Blood Gas Systems by H. Hanford, Technical Note # Tietz Fundamental of Clinical Chemistry by C.A. Burtis and E.R. Ashwood, Forth Edition (1996), page 828. Kevin Fallon, Ph.D. Dr. Fallon was a member of Instrumentation Laboratory s scientific staff for 25 years, most recently serving as the Director of Scientific Affairs. Additionally, he served on the faculty of the University of Texas Medical School and was Director of the STAT Labs at Hermann Hospital in Houston. Dr. Fallon continues to offer his expertise to IL on a consulting basis. Sohrab Mansouri, Ph.D. Dr. Mansouri has over 20 years of experience in the design, development and implementation of chemical and biochemical sensors for medical applications. He has spent the past 13 years advancing the GEM technology and is responsible for the development of the Premier and Premier 3000 cartridges.

31 2003 Instrumentation Laboratory - Printed in Italy - Speed /03

32 Part. No

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

Portex Arterial Blood Sampling Kits

Portex Arterial Blood Sampling Kits Portex Arterial Blood Sampling Kits Safe Application. Simple Activation. Consistent Results. Portex Arterial Blood Sampling Kits > Compatible with Multi-parameter Analyzers > Balanced Heparin Provides

More information

GEM Premier 3000, now with iqm The New Standard for Critical Care Testing

GEM Premier 3000, now with iqm The New Standard for Critical Care Testing GEM Premier 3000, now with iqm The New Standard for Critical Care Testing ph, Blood Gas Electrolytes Metabolites Hematocrit CO-Oximetry Coagulation GEM Premier 3000, now with iqm Simple. Smart. Standar

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

Ecology Quality Assurance Glossary

Ecology Quality Assurance Glossary Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability

More information

Blood Glucose Monitoring: The Facts about Accuracy

Blood Glucose Monitoring: The Facts about Accuracy Blood Glucose Monitoring: The Facts about Accuracy Summary The purpose of this document is to provide the facts about the accuracy of self-monitoring of blood glucose (SMBG) systems. Accuracy can be defined

More information

ph and Acidity in Wine and Fruit Juice

ph and Acidity in Wine and Fruit Juice ph and Acidity in Wine and Fruit Juice DOC316.52.93089 Endpoint potentiometric titration Scope and application: Wine and fruit juice 1. Introduction This working procedure refers to the titration of total

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

Biochemistry Validation Form

Biochemistry Validation Form Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August

More information

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination

More information

How to Verify Performance Specifications

How to Verify Performance Specifications How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated

More information

YSI 2900 Series Biochemistry Analyzers FERMENTATION CONTROL FOOD & BEVERAGE CLINICAL RESEARCH

YSI 2900 Series Biochemistry Analyzers FERMENTATION CONTROL FOOD & BEVERAGE CLINICAL RESEARCH YSI 2900 Series Biochemistry Analyzers FERMENTATION CONTROL FOOD & BEVERAGE CLINICAL RESEARCH Fast and Accurate: The YSI Gold Standard YSI has earned a reputation as the Gold Standard in bio-analytical

More information

How To Test For Creatinine

How To Test For Creatinine Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 CentraCare Health (CCH) adopts the following policy/procedure for: St. Cloud Hospital Original: 5/02 Minor Revision: Full Review: 1/15 Replaces:

More information

ASN Dialysis Advisory Group ASN DIALYSIS CURRICULUM

ASN Dialysis Advisory Group ASN DIALYSIS CURRICULUM ASN Dialysis Advisory Group ASN DIALYSIS CURRICULUM 0 ASN Dialysis Curriculum Dialysis Circuit Review Bessie A. Young, MD, MPH, FACP Associate Professor, University of Washington Director Home Dialysis,

More information

Neutralizing an Acid and a Base

Neutralizing an Acid and a Base Balancing Act Teacher Information Objectives In this activity, students neutralize a base with an acid. Students determine the point of neutralization of an acid mixed with a base while they: Recognize

More information

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC Common Problems with Online Water Quality Analyzers Greg Macy Director AquatiPro LLC Objectives Diagnosing Colorimetric Analyzers Identifying Common Problems with Sodium Analyzers Troubleshooting ph Analyzers

More information

Enzyme Action: Testing Catalase Activity

Enzyme Action: Testing Catalase Activity Enzyme Action: Testing Catalase Activity Experiment 6A Many organisms can decompose hydrogen peroxide (H 2 O 2 ) enzymatically. Enzymes are globular proteins, responsible for most of the chemical activities

More information

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K092353 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY B. Purpose for Submission: This is a new 510k application for a new indication for the MONOLISA Anti-HAV IgM EIA

More information

Transmitter Interface Program

Transmitter Interface Program Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following

More information

An Introduction to. Metrics. used during. Software Development

An Introduction to. Metrics. used during. Software Development An Introduction to Metrics used during Software Development Life Cycle www.softwaretestinggenius.com Page 1 of 10 Define the Metric Objectives You can t control what you can t measure. This is a quote

More information

ph Alkalinity of Water

ph Alkalinity of Water ph Alkalinity of Water DOC316.52.93085 Based on ISO standard 9963-1:1994 ph-metric Titration 0.4 to 20 mmol/l of Total Alkalinity 1. Introduction Alkalinity of water is its acid-neutralizing capacity.

More information

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history. Revision History COPY The top row of this table shows the most recent changes to this controlled document. For previous revision history information, archived versions of this document are maintained by

More information

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

Relative Humidity Calibration Kit

Relative Humidity Calibration Kit Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration

More information

SOFTWARE-IMPLEMENTED SAFETY LOGIC Angela E. Summers, Ph.D., P.E., President, SIS-TECH Solutions, LP

SOFTWARE-IMPLEMENTED SAFETY LOGIC Angela E. Summers, Ph.D., P.E., President, SIS-TECH Solutions, LP SOFTWARE-IMPLEMENTED SAFETY LOGIC Angela E. Summers, Ph.D., P.E., President, SIS-TECH Solutions, LP Software-Implemented Safety Logic, Loss Prevention Symposium, American Institute of Chemical Engineers,

More information

ph and Acidity in Tomato Sauce

ph and Acidity in Tomato Sauce ph and Acidity in Tomato Sauce DOC316.52.93100 Endpoint potentiometric titration Scope and application: Tomato sauce 1. Introduction This working procedure refers to the determination of titratable acidity

More information

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.

More information

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration

More information

Quality Assurance/Quality Control in Acid Deposition Monitoring

Quality Assurance/Quality Control in Acid Deposition Monitoring Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial

More information

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation Introduction The importance of keeping cooling tower water in proper chemical

More information

Enzyme Action: Testing Catalase Activity

Enzyme Action: Testing Catalase Activity Enzyme Action: Testing Catalase Activity Experiment 6A Many organisms can decompose hydrogen peroxide (H 2 O 2 ) enzymatically. Enzymes are globular proteins, responsible for most of the chemical activities

More information

Individualized Quality Control Plan

Individualized Quality Control Plan NOVEMBER 2014 CLIA Individualized Quality Control Plan WHAT IS AN IQCP? OVERVIEW Discovering, Deciphering and Designing an Individualized Quality Control Plan (IQCP) IQCP provides a framework for customizing

More information

USE OF REFERENCE MATERIALS IN THE LABORATORY

USE OF REFERENCE MATERIALS IN THE LABORATORY USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002 AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002 First Edition MONITORING AND LABORATORY DIVISION August

More information

Fundamentals of Mass Flow Control

Fundamentals of Mass Flow Control Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of

More information

ArtisanLink Staining System is an automated special stains slide

ArtisanLink Staining System is an automated special stains slide Technical Tips Tips on using the ArtisanLink Special Staining System Jamie Nowacek, BS, HT(ASCP) CM, QIHC, PMP Dako North America, Inc. Carpinteria, CA, USA ArtisanLink Staining System is an automated

More information

APPENDIX N. Data Validation Using Data Descriptors

APPENDIX N. Data Validation Using Data Descriptors APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection

More information

GA/7 Potentiometric Titration

GA/7 Potentiometric Titration Rev. 7/99 7-1 INTRODUCTION GA/7 Potentiometric Titration The potentiometric titration is a useful means of characterizing an acid. The ph of a solution is measured as a function of the amount of titrant

More information

Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10

Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10 Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan SECTION 10 DATA QUALITY ASSURANCE AND QUALITY CONTROL (QA/QC) Hawai`i HEER TGM 10-1 December 31, 2008 CONTENTS Acronyms

More information

ADVANCED CONTROL TECHNIQUE OF CENTRIFUGAL COMPRESSOR FOR COMPLEX GAS COMPRESSION PROCESSES

ADVANCED CONTROL TECHNIQUE OF CENTRIFUGAL COMPRESSOR FOR COMPLEX GAS COMPRESSION PROCESSES ADVANCED CONTROL TECHNIQUE OF CENTRIFUGAL COMPRESSOR FOR COMPLEX GAS COMPRESSION PROCESSES by Kazuhiro Takeda Research Manager, Research and Development Center and Kengo Hirano Instrument and Control Engineer,

More information

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests 2010 Laboratory Accreditation Program Audioconference Accreditation Requirements for Waived Vs. Non-waived Tests Objectives: After participating in this audioconference you will be able to: Define the

More information

Measurement by Ion Selective Electrodes

Measurement by Ion Selective Electrodes Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable

More information

To help you understand the role documents and records play in the quality system and the monitoring of programs.

To help you understand the role documents and records play in the quality system and the monitoring of programs. Module 15 Documents and Records Purpose Pre-requisite Modules Learning Objectives Content Outline Handout Notes on Customization To help you understand the role documents and records play in the quality

More information

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known

More information

How is Dew Point Measured Reliably?

How is Dew Point Measured Reliably? How is Dew Point Measured Reliably? Air compression increases vapor pressure and thus dew point. = water vapor Vapor pressure Frost point Saturation - 0 C + Dew point Monitoring with alarming and real

More information

BEST PRACTICES IN LEAK DETECTION AND REPAIR (LDAR) PROGRAMS. November 06, 2014 By: François Thibodeau, Eng.

BEST PRACTICES IN LEAK DETECTION AND REPAIR (LDAR) PROGRAMS. November 06, 2014 By: François Thibodeau, Eng. BEST PRACTICES IN LEAK DETECTION AND REPAIR (LDAR) PROGRAMS November 06, 2014 By: François Thibodeau, Eng. 2 PRESENTATION OVERVIEW 1. Essential Components of a LDAR Program 2. LDAR Best Practices a. A

More information

USEPA Electrode Method Method 8156

USEPA Electrode Method Method 8156 ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method

More information

Inferred ph in Steam Plant Water Chemistry Monitoring

Inferred ph in Steam Plant Water Chemistry Monitoring Application Data Sheet ADS 4900-87/rev.B January 2009 Power Industry Inferred ph in Steam Plant Water Chemistry Monitoring INTRODUCTION Inferred ph means ph calculated from straight and cation conductivity.

More information

Control Device Requirements Charts For Oil and Gas Handling and Production Facilities

Control Device Requirements Charts For Oil and Gas Handling and Production Facilities Device Charts For Oil and Gas Handling and Production Facilities Purpose/Scope: The purpose of this document is to provide standardized guidance for use by the regulated community and air permit reviewers,

More information

Maximizing return on plant assets

Maximizing return on plant assets Maximizing return on plant assets Manufacturers in nearly every process industry face the need to improve their return on large asset investments. Effectively managing assets, however, requires a wealth

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

TBA-FX8 Automated Clinical Chemistry Analyzer with Flexibility to Respond to Demands of Large-Scale Clinical Laboratories

TBA-FX8 Automated Clinical Chemistry Analyzer with Flexibility to Respond to Demands of Large-Scale Clinical Laboratories TBA-FX8 Automated Clinical Chemistry Analyzer with Flexibility to Respond to Demands of Large-Scale Clinical Laboratories HIGASHI Tetsuya MATSUMOTO Satoshi Automated clinical chemistry analyzers are instruments

More information

Procedure Manual for the i-stat System

Procedure Manual for the i-stat System Procedure Manual for the i-stat System This Procedure Manual is intended to be a template for the Procedure Manual required by CLIA and laboratory accreditation bodies. This Procedure Manual should be

More information

Quaternary Ammonium Compounds DOC316.53.01128

Quaternary Ammonium Compounds DOC316.53.01128 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower

More information

Supplier Quality Agreements

Supplier Quality Agreements Supplier Quality Agreements Dan O Leary CBA, CQA, CQE, CRE, SSBB, CIRM President Ombu Enterprises, LLC Dan@OmbuEnterprises.com www.ombuenterprises.com 603-209-0600 1 Speaker Biography Dan O Leary Dan O

More information

Single-Use Flow-Through Cell ph FTC-SU-HP8

Single-Use Flow-Through Cell ph FTC-SU-HP8 Single-Use Flow-Through Cell ph FTC-SU-HP8 SENSOR PROBES Instruction Manual Single-Use Flow-Through Cell ph FTC-SU-HP8 Specification: ph sensor stick integrated in a single-use flow-through cell Document

More information

Introduction to Vicinity

Introduction to Vicinity White Paper September 9, 2002 Table of Contents Introduction to Vicinity... 1 Bill of Material (BOM) vs. Formula... 2 QC testing... 2 Unit of measure... 3 Lot tracking... 3 Hybrid manufacturing... 3 Vicinity

More information

Sigma Control PC INSIDE. 97 psi 187 F R on load

Sigma Control PC INSIDE. 97 psi 187 F R on load Sigma Control PC INSIDE 97 psi 187 F R on load Innovation Sigma Control with a PC inside At Kaeser, we pride ourselves on being the world s leading innovator in air system technology. Over twenty-five

More information

How to calibrate an RTD or Platinum Resistance Thermometer (PRT)

How to calibrate an RTD or Platinum Resistance Thermometer (PRT) How to calibrate an RTD or Platinum Resistance Thermometer (PRT) Application Note Introduction There are two types of calibrations applicable to PRTs characterization and tolerance testing. The type of

More information

ORGANIC SAMPLE PREPARATION

ORGANIC SAMPLE PREPARATION ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,

More information

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure

More information

A Potentiometric Analysis of Fluoride Ion in Toothpaste

A Potentiometric Analysis of Fluoride Ion in Toothpaste CHEM 311L Quantitative Analysis Laboratory Revision 2.0 A Potentiometric Analysis of Fluoride Ion in Toothpaste In this laboratory exercise, we will analyze a toothpaste sample for its Fluoride Ion (F

More information

CALIBRATION PRINCIPLES

CALIBRATION PRINCIPLES 1 CALIBRATION PRINCIPLES After completing this chapter, you should be able to: Define key terms relating to calibration and interpret the meaning of each. Understand traceability requirements and how they

More information

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Equipment / Procedure: Filename: Assessor s name & date: Reviewer s name & date: Cosmed K4 2012.1201SOP_cosmed_K4 Richard Metcalfe, December 2012 Tom Nightingale, December

More information

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

The Facts on Equine Drug Testing

The Facts on Equine Drug Testing The Facts on Equine Drug Testing CONTACT Travis Mays, MS Interim Section Head Analytical Chemistry, Drug Testing Laboratory Toxicology Laboratory 979.845.3414 tmays@ LIMITATIONS IN DRUG TESTING While advancements

More information

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,

More information

Welcome to the topic on creating key performance indicators in SAP Business One, release 9.1 version for SAP HANA.

Welcome to the topic on creating key performance indicators in SAP Business One, release 9.1 version for SAP HANA. Welcome to the topic on creating key performance indicators in SAP Business One, release 9.1 version for SAP HANA. 1 In this topic, you will learn how to: Use Key Performance Indicators (also known as

More information

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist This checklist is intended to help laboratories improve the quality of their testing programs with regard to meat and poultry testing samples, in order to ensure those laboratories are at least equal to

More information

Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope... 1. 2 Introduction... 1. 3 Standard Precautions... 1. 4 Definitions...

Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope... 1. 2 Introduction... 1. 3 Standard Precautions... 1. 4 Definitions... Vol. 27 No. 26 Replaces H3-A5 Vol. 23 No. 32 Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition This document provides procedures for the collection

More information

Basic Lessons in Laboratory Quality Control

Basic Lessons in Laboratory Quality Control Bio-Rad Laboratories QC education Basic Lessons in Laboratory Quality Control QC Workbook Basic Lessons in Laboratory Quality Control Written by Greg Cooper, CLS, MHA Manager of Clinical Standards and

More information

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND #3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric

More information

Benefits of Cold Aisle Containment During Cooling Failure

Benefits of Cold Aisle Containment During Cooling Failure Benefits of Cold Aisle Containment During Cooling Failure Introduction Data centers are mission-critical facilities that require constant operation because they are at the core of the customer-business

More information

ph Measurement and Control

ph Measurement and Control ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components

More information

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506 Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking

More information

Welcome to the fully integrated autoimmune laboratory. Saves time by reducing hands-on functions NOVA Lite Barcoded IFA Slides

Welcome to the fully integrated autoimmune laboratory. Saves time by reducing hands-on functions NOVA Lite Barcoded IFA Slides The Integrated Lab Connected Welcome to the fully integrated autoimmune laboratory. Saves time by reducing hands-on functions NOVA Lite Barcoded IFA Slides NOVA View Reduces costs with automation efficiencies

More information

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Data Quality Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Why do we care? Program goals are key what do you want to do with the data? Data utility increases directly

More information

Procedure for Equipment Calibration and Maintenance

Procedure for Equipment Calibration and Maintenance Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory

More information

Quality Requirements of a Point of Care Testing Service

Quality Requirements of a Point of Care Testing Service Quality Requirements of a Point of Care Testing Service Adil I. Khan MSc, PhD Assistant Professor of Pathology Director of Point of Care Testing & Clinical Chemistry Laboratory, Temple University Episcopal

More information

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban

More information

Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1

Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1 ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas

More information

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory

More information

2A.1 Features of Chiller Control Components... 104. 2A.3 Chilled-Water Supply Temperature Control... 107. 2A.4 Cooling-Water Supply Control...

2A.1 Features of Chiller Control Components... 104. 2A.3 Chilled-Water Supply Temperature Control... 107. 2A.4 Cooling-Water Supply Control... Appendix 2A: Chiller Control Principles... 104 2A.1 Features of Chiller Control Components... 104 2A.2 Start-up and Shutdown... 105 2A.2.1 Start-up...105 2A.2.2 Shutdown...106 2A.3 Chilled-Water Supply

More information

Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer

Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer Titration of a Diprotic Acid: Identifying an Unknown Computer 25 A diprotic acid is an acid that yields two H + ions per acid molecule. Examples of diprotic acids are sulfuric acid, H 2 SO 4, and carbonic

More information

MEDICAL DEVICE GUIDANCE

MEDICAL DEVICE GUIDANCE May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained

More information

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration. S HIFT INTO NEUTRAL 15. Acid-Base Titration Shift into Neutral Student Instruction Sheet Challenge Discover the concentration of an unknown acid solution using acid-base titration. Equipment and Materials

More information

18 Conductometric Titration

18 Conductometric Titration Lab Activity 18 CONDUCTOMETRIC TITRATION LAB ACTIVITY 18 Conductometric Titration Background Titration is the a method of determining the concentration of an unknown solution (the analyte) by reacting

More information

OPTI CCA-TS Analyzer. Operator s Manual

OPTI CCA-TS Analyzer. Operator s Manual OPTI CCA-TS Analyzer Operator s Manual OPERATOR S MANUAL REVISION LOG (Please record any changes made to this manual) Revision Release Date Approved by Description A Sept 2004 Released per CO 040316 B

More information

Chloride and Salinity

Chloride and Salinity INTRODUCTION Chloride Chloride and Chloride, in the form of the Cl ion, is one of the major inorganic anions, or negative ions, in saltwater and freshwater. It originates from the dissociation of salts,

More information

Licensing for BarTender s Automation Editions. Understanding Printer-Based Licensing WHITE PAPER

Licensing for BarTender s Automation Editions. Understanding Printer-Based Licensing WHITE PAPER Licensing for BarTender s Automation Editions Understanding Printer-Based Licensing and How to Configure Seagull License Server WHITE PAPER Contents Introduction to Printer-Based Licensing 3 Available

More information

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM Inform II BLOOD GLUCOSE MONITORING SYSTEM Quick Reference Guide for Healthcare Professionals 2 Table of Contents Important... 4 General Information... 6 ACCU-CHEK Inform II Meter...10 How to Perform Patient

More information

FMC Technologies Measurement Solutions Inc.

FMC Technologies Measurement Solutions Inc. Table of Contents 1 Revision History... 4 2 Company... 5 2.1 Introduction... 5 2.2 Quality Policy... 6 2.3 Quality Organization... 6 2.4 Company Processes... 7 3 Scope and Exclusions... 8 3.1 Scope...

More information