1 Assessing the Long-Term Performance of the UK Emissions Monitoring Industry from PT Scheme Data Dr Marc Coleman AQE 2013, 13 th 14 th March, Telford UK

2 Contents Description of NPL Proficiency Testing schemes Stack Simulator facility How proficiency is expressed in terms of a z-score for each participant Evaluation of industry performance since inception of each scheme Z-score distributions Subsequent performance of outlier participants Conclusions

3 NPL Operated Schemes There are 3 ANNONYMISED Proficiency Testing (PT) schemes operated at NPL Calibration Gas Gas bottle certified reference materials Tests calibration proficiency Gas Measurement Utilises NPL s Stack Simulator facility Sampling apparatus required Tests not only quantification proficiency but importantly also sampling proficiency Particulate Shims to simulate filters NaCl solutions to simulate probe washings

4 NPL Stack Simulator Facility 2m tall recirculating stack Four 127mm (BSP) ports Temperatures of up to 180 C and flows of up to 12m.s -1 Up to 25% H 2 O can be introduced along with a broad range of analytes / interferents (e.g. HCl, SO 2, NOx, CO, CO 2, O 2, etc.)

5 NPL PT Schemes Coverage Calibration Gas longest running scheme ~70 organizations have participated Representing vast majority of UK stack testing industry

6 Expression of Results Results are expressed in terms of z-scores Consistent with ISO z-scores are segregated into 3 categories 2 & -2 = satisfactory >2 & <3 or <-2 & >-3 = warning 3 & -3 = unsatisfactory On the balance of probabilities it s unlikely that a participant would achieve these levels without there being a real issue to address Consequence of not achieving a satisfactory score is that at the annual surveillance visit UKAS are likely to expect to see evidence of an internal investigation as to the cause

7 Analysis Approach (1) Motivation for work was to determine if any trends had started to emerge across the schemes Is the proficiency of the industry being maintained / improving / deteriorating? Comparing data between rounds Issue evaluating across rounds is that test concentrations and allowable deviations change Selected a fixed concentration present in all rounds Recalculated z-scores using allowable deviations from 2011 Have focused on NO, SO 2 & particulates only as more difficult to measure compared to other species in the schemes (e.g. CO)

8 Analysis Approach (2) Allowable deviations and assigned values for recalculated participant z-scores Scheme Measurand Assigned value Calibration Gas Gas Measurement Allowable deviation SO 2 100ppm ±5ppm 5% NO 100ppm ±5ppm 5% SO 2 40ppm ±5ppm 12% NO 30ppm ±3.5ppm 12% Particulate Shims 140mg ±14mg 10% NaCl 5mg ±1mg 20% Allowable deviation / % For SO 2 & NO the IED requires measurement uncertainty of ±10%, k=1 of the ELV For particulates IED requires ± 15%, k=1 of ELV

9 CALIBRATION GAS (Gas Bottles)

10 Z-Score Distributions for SO 2 and NO Calibration Gas Scheme Outlier at z=12 SO 2 NO Box encompasses 25-75% quartiles Whiskers extend out to farthest score within 25% of box limit * = outlier z-score Greater number of unsatisfactory (>3 & <-3) participant scores for SO 2 than NO Frequency of unsatisfactory scores & outliers decreases with time

11 Tracking Performance SO 2 Plots show subsequent score for any participant achieving a z-score >3 & <- 3 All participants unsatisfactory by 2011 requirements are seen to improve Scheme appears to be having the desired effect NO

12 Interquartile (25-75%) Z-Score Ranges for Calibration Gas SO Round NO Decreased interquartile distribution seen for NO over SO 2 Combined with boxplots implies a greater proficiency of the industry for NO over SO 2 No obvious trend with time in interquartile range Round

13 GAS MEASUREMENT (Stack Simulator)

14 Z-Score Distributions for Gas Measurement Scheme SO NO Narrower NO distributions However, some evidence of a negative bias in SO 2 scores Consistent with sample loss in sampling apparatus but not sufficient to lead to unsatisfactory scores Highlights importance of schemes where sampling is included in the proficiency test To accurately assess proficiency the entire measurement must be tested

15 Interquartile (25-75%) Z-Score Ranges for Gas Measurement SO 2 As for Calibration Gas scheme decreased interquartile distribution seen for NO over SO 2 No trend with time NO

16 PARTICULATES (Shims and NaCl Washings)

17 Z-Score Distributions for Particulates Scheme Shims Round Outlier at z=-40 Washings Shims performance has been maintained Washings appears to have improved EN requires a weighing accuracy of ±5% Are only testing weighing in scheme so could argue this should be the allowable deviation Washing distributions would increase 4-fold, performance would be poorer Round

18 Tracking Performance Post Unsatisfactory Score As with gases, participants who would be unsatisfactory by 2011 requirements all improve

19 Conclusions The number of unsatisfactory scores for NO & SO 2 has decreased since the inception of the Calibration Gas scheme showing that the proficiency testing is having the desired effect The Stack Simulator (Gas Measurement scheme) data evidenced a negative bias in SO 2 scores (i.e. participant underreading) not seen with the gas bottle (Calibration Gas) scheme Bias not sufficient to lead to unsatisfactory scores Highlighted importance of testing proficiency of entire measurement For particulates number of unsatisfactory scores for washings has decreased The Future PT testing for HCl using the Stack Simulator commenced in Feb 2013 Particulate Stack Simulator (enabling particulate sampling to be directly tested, i.e. as opposed to NaCl solutions to simulate washings) Stack Simulator PT on only wet chemistry SRMs (SO 2, HCl) Speciated VOCs

Policy on Monitoring of Stack Emissions to Air at EPA Licensed Sites

Policy on Monitoring of Stack Emissions to Air at EPA Licensed Sites Policy on Monitoring of Stack Emissions to Air at EPA Licensed Sites 1. OVERVIEW In Ireland at present there are no mandatory minimum criteria specified in relation to the completion of stationary source

More information

Examination syllabuses for Manual Stack emissions monitoring. Environment Agency Version 9 January 2015

Examination syllabuses for Manual Stack emissions monitoring. Environment Agency Version 9 January 2015 Examination syllabuses for Manual Stack emissions monitoring Environment Agency Version 9 January 2015 Record of amendments Version Date Section Amendment number 6 August 09 Level 1 Updated Level 1 syllabus

More information

USE OF REFERENCE MATERIALS IN THE LABORATORY

USE OF REFERENCE MATERIALS IN THE LABORATORY USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently

More information

M20. Technical Guidance Note (Monitoring) Quality assurance of continuous emission monitoring systems - application of EN 14181 and BS EN 13284-2

M20. Technical Guidance Note (Monitoring) Quality assurance of continuous emission monitoring systems - application of EN 14181 and BS EN 13284-2 linear Technical Guidance Note (Monitoring) M20 Quality assurance of continuous emission monitoring systems - application of EN 14181 and BS EN 13284-2 Environment Agency Version 3 June 2015 Foreword This

More information

Manual stack emission monitoring Performance standard for organisations. Environment Agency November 2011 Version 7.2

Manual stack emission monitoring Performance standard for organisations. Environment Agency November 2011 Version 7.2 Manual stack emission monitoring Performance standard for organisations Environment Agency November 2011 Version 7.2 Record of amendments Version Date Section Amendment number 7.1 Jan 11 5.6.8 Added requirement

More information

STACK EMISSIONS MONITORING REPORT

STACK EMISSIONS MONITORING REPORT STACK EMISSIONS MONITORING REPORT Unit 5 Crown Industrial Estate Kenwood Road Stockport SK5 6PH Tel: 0161 443 0980 Fax: 0161 443 0989 Your contact at ESG Andrew Palliser Business Manager Tel: 0161 443

More information

LAB 37 Edition 3 June 2013

LAB 37 Edition 3 June 2013 LAB 37 Edition 3 June 2013 Accreditation Requirements for Sampling and Testing in Accordance with the Drinking Water Testing Specification (DWTS) CONTENTS SECTION PAGE 1 Introduction 2 2 Regulatory requirements

More information

LAQM Helpdesk March 2015

LAQM Helpdesk March 2015 Summary of Laboratory Performance in AIR/WASP NO 2 Proficiency Testing Scheme (April February 2015). Reports are prepared by HSL for BV/NPL on behalf of Defra and the Devolved Administrations. Background

More information

Performance Standards and Test Procedures for Automated Dust Arrestment-Plant Monitors

Performance Standards and Test Procedures for Automated Dust Arrestment-Plant Monitors Performance Standards and Test Procedures for Automated Dust Arrestment-Plant Monitors Certification under MCERTS according to EN 15859 Environment Agency Version 1 August 2010 Record of amendments Version

More information

18.6.1 Terms concerned with internal quality control procedures

18.6.1 Terms concerned with internal quality control procedures 18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable

More information

Technical Guidance Note (Monitoring)

Technical Guidance Note (Monitoring) Technical Guidance Note () M2 of stack emissions to air Environment Agency Version 11 November 2015 Foreword This Technical Guidance Note (TGN) is one of a series providing guidance to our staff, monitoring

More information

The Question of Homogeneity inside a Chimney: Application of ISO 13528 to Stack Emission Proficiency Tests

The Question of Homogeneity inside a Chimney: Application of ISO 13528 to Stack Emission Proficiency Tests Hessian Agency for the Environment and Geology Department I3 Air Pollution, Emission Jens Cordes Benno Stoffels Dominik Wildanger The Question of Homogeneity inside a Chimney: Application of ISO 13528

More information

Performance Standards and Test Procedures for Continuous Emission Monitoring Systems

Performance Standards and Test Procedures for Continuous Emission Monitoring Systems Performance Standards and Test Procedures for Continuous Emission Monitoring Systems For gaseous, particulate and flow-rate monitoring systems Environment Agency July 2012 Version 3.4 Record of amendments

More information

CT556 Boiler Efficiency Modern Controls for Package Boiler Operations

CT556 Boiler Efficiency Modern Controls for Package Boiler Operations CT556 Boiler Efficiency Modern Controls for Package Boiler Operations Marc L Hunter Vice President 25 & 26 March 2015 PUBLIC www.rockwellautomation.com www.us.endress.com 2 Agenda Boiler Operations Overview

More information

Hessian Agency for the Environment and Geology Ludwig-Mond-Straße 33 D-34121 Kassel, Germany. Department I3 (Air Pollution, Emission)

Hessian Agency for the Environment and Geology Ludwig-Mond-Straße 33 D-34121 Kassel, Germany. Department I3 (Air Pollution, Emission) page 1 of 11 Hessian Agency for the Environment and Geology Ludwig-Mond-Straße 33 D-34121 Kassel, Germany Editor German version: I3-Wi Status German version: 30.10.2014 Editor English translation: I3-Co

More information

Descriptive statistics Statistical inference statistical inference, statistical induction and inferential statistics

Descriptive statistics Statistical inference statistical inference, statistical induction and inferential statistics Descriptive statistics is the discipline of quantitatively describing the main features of a collection of data. Descriptive statistics are distinguished from inferential statistics (or inductive statistics),

More information

Where the exp subscripts refer to the experimental temperature and pressure acquired in the laboratory.

Where the exp subscripts refer to the experimental temperature and pressure acquired in the laboratory. Molar Volume of Carbon Dioxide Reading assignment: Julia Burdge, Chemistry 3rd edition, Chapter 10. Goals To determine the molar volume of carbon dioxide gas and the amount of sodium carbonate in a sample.

More information

Estimates of Uncertainty of the Calibration of Balances

Estimates of Uncertainty of the Calibration of Balances Estimates of Uncertainty of the Calibration of Balances Chart shows the uncertainty (estimates) of the calibration of representative electronic balances. Chart. Uncertainty of calibration (estimates) Minimum

More information

Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team

Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for

More information

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

ETV Joint Verification Statement

ETV Joint Verification Statement THE ENVIRONMENTAL TECHNOLOGY VERIFICATION PROGRAM U.S. Environmental Protection Agency ETV Joint Verification Statement TECHNOLOGY TYPE: Continuous Emission Monitor APPLICATION: TECHNOLOGY NAME: COMPANY:

More information

The prediction of flue gas emissions from the combustion in the industrial tubular heaters

The prediction of flue gas emissions from the combustion in the industrial tubular heaters Ovidius University Annals of Chemistry Volume, Number 1, pp.137-11, 9 The prediction of flue gas emissions from the combustion in the industrial tubular heaters Mirela VELICU a, Claudia-Irina KONCSAG b

More information

Procedure for Equipment Calibration and Maintenance

Procedure for Equipment Calibration and Maintenance Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory

More information

CMMs and PROFICIENCY TESTING

CMMs and PROFICIENCY TESTING CMMs and PROFICIENCY TESTING Speaker/Author: Dr. Henrik S. Nielsen HN Metrology Consulting, Inc. HN Proficiency Testing, Inc. Indianapolis, Indiana, USA hsnielsen@hn-metrology.com Phone: (317) 849 9577;

More information

PM Fine Database and Associated Visualization Tool

PM Fine Database and Associated Visualization Tool PM Fine Database and Associated Visualization Tool Steven Boone Director of Information Technology E. H. Pechan and Associates, Inc. 3622 Lyckan Parkway, Suite 2002; Durham, North Carolina 27707; 919-493-3144

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

Quality Assurance/Quality Control in Acid Deposition Monitoring

Quality Assurance/Quality Control in Acid Deposition Monitoring Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial

More information

Any potential user should ensure in consultation with the manufacturer that this AMS is suitable for the installation on which it will be installed.

Any potential user should ensure in consultation with the manufacturer that this AMS is suitable for the installation on which it will be installed. Test report: 936/21210692/A of 30 March 2011 First certification: 29 July 2011 Run of validity until: 28 July 2016 Publication BAnz. 29 July 2011, No 113, page 2725, chapter I, No 4.1 Authorised application

More information

Calibration Certificate

Calibration Certificate P.O. Box 87-0087 Phone:856-686-1600 Fax:856-686-1601 www.troemner.com e-mail: troemner@troemner.com SECTION 1: Page 1 of 7 Pages SECTION 2: APPROVED SIGNATORY Joseph Moran SECTION 3: PERSON PERFORMING

More information

Method 202 Dry Impinger Method for Determining Condensable Particulate Emissions from Stationary Sources. By: Paul Ottenstein

Method 202 Dry Impinger Method for Determining Condensable Particulate Emissions from Stationary Sources. By: Paul Ottenstein Air Quality Issues in the Rocky Mountain Region PM2.5 Source Testing Method 201A Determination PM 10 and PM 2.5 Emissions from Stationary Sources (Constant Sampling Rate Procedure) Method 202 Dry Impinger

More information

EN 15267 - A new unified testing and approval scheme for automated measuring systems

EN 15267 - A new unified testing and approval scheme for automated measuring systems EN 15267 - A new unified testing and approval scheme for automated measuring systems Richard Gould, Technical Advisor, Environment Agency, Lutra House, Dodd Way, Walton Summit Industrial Estate, Preston,

More information

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon

More information

10 20 30 40 50 60 Mark. Use this information and the cumulative frequency graph to draw a box plot showing information about the students marks.

10 20 30 40 50 60 Mark. Use this information and the cumulative frequency graph to draw a box plot showing information about the students marks. GCSE Exam Questions on Frequency (Grade B) 1. 200 students took a test. The cumulative graph gives information about their marks. 200 160 120 80 0 10 20 30 50 60 Mark The lowest mark scored in the test

More information

EXPERIMENT 9 Evaluation of the Universal Gas Constant, R

EXPERIMENT 9 Evaluation of the Universal Gas Constant, R Outcomes EXPERIMENT 9 Evaluation of the Universal Gas Constant, R After completing this experiment, the student should be able to: 1. Determine universal gas constant using reaction of an acid with a metal.

More information

Continuous Emissions Monitoring - Program 77

Continuous Emissions Monitoring - Program 77 Program Description Program Overview Coal-fired power plants are in increased need of robust, accurate, and certifiable continuous emissions monitors (CEMs) for mercury, particulate matter (PM), acid gases,

More information

Scientific and analytical consultancy services for industry

Scientific and analytical consultancy services for industry Scientific and analytical consultancy services for industry EXPERTISE ON SITE AND IN THE LABORATORY Anchem Laboratories is a multi-disciplined accredited Laboratory offering a range of specialised environmental

More information

Analytical Chemistry Lab Reports

Analytical Chemistry Lab Reports Analytical Chemistry Lab Reports Format and Calculations John Collins collinj2@gmail.com Measurement Analytical chemistry is entirely about measurement, what these measurements signify, and the understanding

More information

Environment Protection Agency Office of Environmental Enforcement (OEE) Air Emissions Monitoring Guidance Note (AG2)

Environment Protection Agency Office of Environmental Enforcement (OEE) Air Emissions Monitoring Guidance Note (AG2) Environment Protection Agency Office of Environmental Enforcement (OEE) Air Emissions Monitoring Guidance Note (AG2) Version 2 (DRAFT FOR CONSULTATION) Environmental Protection Agency Johnstown Castle

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

Interlaboratory studies

Interlaboratory studies Interlaboratory studies Department of Food Chemistry and Analysis, ICT Prague Vladimír Kocourek Prague, 2012 Interlaboratory studies Various titles: Interlaboratory proficiency test or studies Interlaboratory

More information

RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES

RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES Reviewed by: Natalija Jovičić Zarić MSc, Quality Manager Approved by: Milica Lukešević, Acting Director Any

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK code Location code Customers Sites locations: 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Lim Tiong Thai Tel: +673-2-386475

More information

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS. PURPOSE The purpose of this report is to verify the performance characteristics of the Sievers M-Series TOC Analyzers. Performance was quantified

More information

An overview of Euro VI for trucks over 3.5t. Brought to you by Mercedes-Benz

An overview of Euro VI for trucks over 3.5t. Brought to you by Mercedes-Benz An overview of Euro VI for trucks over 3.5t Brought to you by Mercedes-Benz Contents What is Euro VI? What is Euro VI? 01 What s different about Euro VI? 02 When do you need to think about it? 03 How is

More information

ATMOSPHERIC EMISSIONS FROM GAS FIRED HOME HEATING APPLIANCES

ATMOSPHERIC EMISSIONS FROM GAS FIRED HOME HEATING APPLIANCES ATMOSPHERIC EMISSIONS FROM GAS FIRED HOME HEATING APPLIANCES Stefano Cernuschi, Stefano Consonni, Giovanni Lonati, Michele Giugliano, Senem Ozgen DIIAR Environmental Sect., Politecnico di Milano, P.za

More information

STATE OF COLORADO PORTABLE ANALYZER MONITORING PROTOCOL

STATE OF COLORADO PORTABLE ANALYZER MONITORING PROTOCOL STATE OF COLORADO PORTABLE ANALYZER MONITORING PROTOCOL Determination of Nitrogen Oxides, Carbon Monoxide and Oxygen Emissions from Natural Gas-Fired Reciprocating Engines, Combustion Turbines, Boilers,

More information

Any potential user should ensure in consultation with the manufacturer that this AMS is suitable for the installation on which it will be installed.

Any potential user should ensure in consultation with the manufacturer that this AMS is suitable for the installation on which it will be installed. Test report: 936/21206693/B of 28 March 2011 First certification: 29 July 2011 Run of validity until: 28 July 2016 Publication BAnz. 29 July 2011, No 113, page 2725, chapter I, No 4.2 Authorised application

More information

Check-list for auditors. Audit visit at candidate registered laboratories

Check-list for auditors. Audit visit at candidate registered laboratories Check-list for auditors Audit visit at candidate registered laboratories 1 Contents Introduction Laboratory documentation required by the standard(s), available for inspection by the auditor Items to be

More information

Mercury Speciation in Flue Gas from Coal-fired Power Stations

Mercury Speciation in Flue Gas from Coal-fired Power Stations Mercury Speciation in Flue Gas from Coal-fired Power Stations Matthew A Dexter, Warren T Corns and Peter B Stockwell P S Analytical Arthur House Crayfields Industrial Estate Main Road Orpington BR5 3HP

More information

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure

More information

(a) Method 1 Sample and Velocity Traverses for Stationary Sources.

(a) Method 1 Sample and Velocity Traverses for Stationary Sources. While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the

More information

HS 1003 Part 2 HS 1003 Heavy Metals Test

HS 1003 Part 2 HS 1003 Heavy Metals Test HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:

More information

Selection, Use and Interpretation of Proficiency Testing (PT) Schemes

Selection, Use and Interpretation of Proficiency Testing (PT) Schemes Selection, Use and Interpretation of Proficiency Testing (PT)Schemes es Second Ed dition 2011 Selection, Use and Interpretation of Proficiency Testing (PT) Schemes Second Edition 2011 Editors Ian Mann

More information

IMPLEMENTING ISO 14001:2004. www.aecos.co.uk

IMPLEMENTING ISO 14001:2004. www.aecos.co.uk IMPLEMENTING ISO 14001:2004 www.aecos.co.uk What is an Environmental Management System? A systematic framework to manage the immediate and long term environmental impacts of an organisation s products,

More information

Quality Assurance Program Quality Assurance Plan Periodic Audits Standard Operating Procedures Preventive Maintenance Outline Quality Assurance Plan (QAP) What is a QAP Why is a QAP required? What does

More information

ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi

ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi The Innovation Hub, Pretoria Presentation layout 1. Introduction ISO/IEC 17025 2. Requirements

More information

ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES

ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES AUTHORITY: VICE PRESIDENT EFFECTIVE DATE: 2015/01/01 DOCUMENT NUMBER: MA 2002 TABLE OF CONTENTS Foreword... 3 1. Application... 4 2.

More information

4. 3. Quality control in the laboratory. Unit 4: Quality assurance and quality control

4. 3. Quality control in the laboratory. Unit 4: Quality assurance and quality control 4. 3 Quality control in the laboratory Quality control is used in an analytical laboratory to check that the results being produced are fit for purpose. A laboratory will typically have its own inhouse

More information

R&S ZNC Vector Network Analyzer Specifications

R&S ZNC Vector Network Analyzer Specifications ZNC3_dat-sw_en_5214-5610-22_v0300_cover.indd 1 Data Sheet 03.00 Test & Measurement R&S ZNC Vector Network Analyzer Specifications 04.09.2012 13:39:47 CONTENTS Definitions... 3 Measurement range... 4 Measurement

More information

EN 15267 Type Approval & Certification of AMS (QAL1)

EN 15267 Type Approval & Certification of AMS (QAL1) EN 15267 Type Approval & Certification of AMS (QAL1) Mr. Karsten Pletscher TÜV Rheinland Energie und Umwelt GmbH Am Grauen Stein, 51105 Cologne Germany Phone: ++49 221 806 2592 karsten.pletscher@de.tuv.com

More information

Shape of Data Distributions

Shape of Data Distributions Lesson 13 Main Idea Describe a data distribution by its center, spread, and overall shape. Relate the choice of center and spread to the shape of the distribution. New Vocabulary distribution symmetric

More information

EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy

EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy EXPERIMENT 4 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium

More information

Advantage of Using Water-Emulsified Fuel on Combustion and Emission Characteristics

Advantage of Using Water-Emulsified Fuel on Combustion and Emission Characteristics Advantage of Using Water-Emulsified Fuel on Combustion and Emission Characteristics T. Yatsufusa *1, T. Kumura, Y. Nakagawa, Y. Kidoguchi The University of Tokushima Abstract In the present study, the

More information

SO 3 -Monitoring in Flue Gas of a Power Plant Application & Results

SO 3 -Monitoring in Flue Gas of a Power Plant Application & Results of a Power Plant Application & Results : Introduction - SO 3 / H 2 SO4 - The SO 3 Challenge : Spectral evaluation - Components to be monitored - Calibration : Analyzer Setup : Field Test Results : Benefits

More information

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting. Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description

More information

The Effect of Temperature on Conductivity Measurement

The Effect of Temperature on Conductivity Measurement The Effect of on Measurement Authors: John J. Barron & Colin Ashton Technical Services Department, Reagecon Diagnostics Ltd, Shannon Free Zone, County Clare, Ireland Abstract The conductivity of all solutions

More information

Importance and Method of using Reference Materials in measuring reliability of Analytical results and Why Custnmers should use them.

Importance and Method of using Reference Materials in measuring reliability of Analytical results and Why Custnmers should use them. Analytical Challenges ill Metallurgy Sluart Mars/and Importance and Method of using Reference Materials in measuring reliability of Analytical results and Why Custnmers should use them. Mintek Analytical

More information

PTA proficiency testing for metal testing laboratories

PTA proficiency testing for metal testing laboratories PTA proficiency testing for metal testing laboratories BRIGGS Philip (Proficiency Testing Australia PO Box 7507, Silverwater, Rhodes NSW 2128 Australia ) Abstract: Proficiency testing is used internationally

More information

Operating Performance and Latest Technology of DeNOx Plants for Coal-Fired Boilers

Operating Performance and Latest Technology of DeNOx Plants for Coal-Fired Boilers 192 Operating Performance and Latest Technology of DeNOx Plants for Coal-Fired Boilers Operating Performance and Latest Technology of DeNOx Plants for Coal-Fired Boilers Masayuki Hirano Yasuyoshi Kato

More information

AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL

AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL FUEL CONSUMPTION MEASUREMENT AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL Description The combination of AVL Fuel Mass Flow Meter and AVL Fuel Temperature Control is a high precise and continuous

More information

Determination of Insoluble Solids in Pretreated Biomass March 2008 Material

Determination of Insoluble Solids in Pretreated Biomass March 2008 Material National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Determination of Insoluble Solids

More information

Advanced Solutions for Gas Monitoring

Advanced Solutions for Gas Monitoring Continuous Emissions Monitoring Stack Testing Process Monitoring Quality Control Engine Exhaust Gas Monitoring Workplace Air Quality Monitoring Combustion Research Emergency Rescue Services Advanced Solutions

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 7 Contact: Mr S C Sparks Solent Industrial Estate Tel: +44 (0)1489 790296 Hedge End Fax: +44 (0)1489 790294 Southampton E-Mail: info@southcal.co.uk

More information

Experis Gas Mixtures. When accuracy, stability and reliability are a must

Experis Gas Mixtures. When accuracy, stability and reliability are a must Experis Gas Mixtures When accuracy, stability and reliability are a must Making the Difference For over 70 years, Air Products has enabled customers to become more productive, energy efficient and sustainable.

More information

PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS

PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS Maria Belli Istituto Superiore per la Protezione e la Ricerca Ambientale (former APAT) ISPRA activities production and characterization

More information

Sensory evaluation of building products - Results of a German round robin test

Sensory evaluation of building products - Results of a German round robin test Sensory evaluation of building products - Results of a German round robin test Johannes Kasche 1, Arne Dahms 1, Wolfgang Horn 2, Oliver Jann 2, Dirk Müller 1 1 Technische Universität Berlin, Germany 2

More information

Quantifying Uncertainty in Analytical Measurement

Quantifying Uncertainty in Analytical Measurement EURACHEM / CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition QUAM:000.P1 EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition Editors S L R Ellison

More information

How to Verify Performance Specifications

How to Verify Performance Specifications How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated

More information

R&S SMZ Frequency Multiplier Specifications

R&S SMZ Frequency Multiplier Specifications Test & Measurement Data Sheet 02.00 R&S SMZ Frequency Multiplier Specifications CONTENTS Definitions... 3 Specifications... 4 RF performance...4 Frequency...4 Level...4 Spectral purity...4 Connectors...5

More information

Implications of changing natural gas prices in the United States electricity sector for SO 2. and life cycle GHG emissions: Supplementary Information

Implications of changing natural gas prices in the United States electricity sector for SO 2. and life cycle GHG emissions: Supplementary Information Implications of changing natural gas prices in the United States electricity sector for SO 2, NO X and life cycle GHG emissions: Supplementary Information Aranya Venkatesh, Paulina Jaramillo, W Michael

More information

MATHESON STACK EMISSION CALIBRATION PROGRAM. Experience the MATHESON Commitment to Supply Chain Excellence

MATHESON STACK EMISSION CALIBRATION PROGRAM. Experience the MATHESON Commitment to Supply Chain Excellence MATHESON STACK EMISSION CALIBRATION PROGRAM MATHESON STACK EMISSION CALIBRATION PROGRAM MATHESON is a single source for industrial, medical, specialty, and electronics gases as well as gas handling equipment,

More information

R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 LABORATORIES

R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 LABORATORIES Page 1 of 7 R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 November 2009 2009 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without

More information

This document was prepared in conjunction with work accomplished under Contract No. DE-AC09-96SR18500 with the U. S. Department of Energy.

This document was prepared in conjunction with work accomplished under Contract No. DE-AC09-96SR18500 with the U. S. Department of Energy. This document was prepared in conjunction with work accomplished under Contract No. DE-AC09-96SR18500 with the U. S. Department of Energy. DISCLAIMER This report was prepared as an account of work sponsored

More information

Thermal environments. www.swema.com swema@swema.com Tel. +46 8 94 00 90

Thermal environments. www.swema.com swema@swema.com Tel. +46 8 94 00 90 ISO 7730 Moderate Thermal Environments Thermal environments Swema equipment for the measurement of Moderate Thermal Environments is designed to comply with the standards: ISO7726 - Thermal environments

More information

Guide To Preparation of Stock Standard Solutions

Guide To Preparation of Stock Standard Solutions chemias ft Guide To Preparation of Stock Standard Solutions First Edition May 2011 Na+ 1000 ppm Guide To Preparation Of Stock Standard Solutions By: CHEMIASOFT May 2011 Page 2 of 61 Page 3 of 61 Table

More information

Texas Commission on Environmental Quality Page 1 Chapter 117 - Control of Air Pollution from Nitrogen Compounds

Texas Commission on Environmental Quality Page 1 Chapter 117 - Control of Air Pollution from Nitrogen Compounds Texas Commission on Environmental Quality Page 1 SUBCHAPTER G: GENERAL MONITORING AND TESTING REQUIREMENTS DIVISION 1: COMPLIANCE STACK TESTING AND REPORT REQUIREMENTS 117.8000, 117.8010 Effective June

More information

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas

A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas Paper #46 Heidi C. Hayes Air Toxics, Ltd., 180 Blue Ravine Rd. Ste. B, Folsom, CA 95630 Diane J.

More information

Guidance for monitoring enclosed landfill gas flares, LFTGN05

Guidance for monitoring enclosed landfill gas flares, LFTGN05 Guidance for monitoring enclosed landfill gas flares, LFTGN05 Document Owner: National Services/ Knowledge, Strategy and Planning (KSP) Version History: Document Date Summary of Changes Version Published

More information

HISTOGRAMS, CUMULATIVE FREQUENCY AND BOX PLOTS

HISTOGRAMS, CUMULATIVE FREQUENCY AND BOX PLOTS Mathematics Revision Guides Histograms, Cumulative Frequency and Box Plots Page 1 of 25 M.K. HOME TUITION Mathematics Revision Guides Level: GCSE Higher Tier HISTOGRAMS, CUMULATIVE FREQUENCY AND BOX PLOTS

More information

Guido Baum Phone: +49(0)221-806 2592 Mail: guido.baum@de.tuv.com www.qal1.de

Guido Baum Phone: +49(0)221-806 2592 Mail: guido.baum@de.tuv.com www.qal1.de Determination of CO 2 and N 2 O emissions with CEM Systems within the scope of greenhouse gas trading CEM 2014 Istanbul Thursday, 15 th May 2014 TÜV Rheinland Energie und Umwelt GmbH Guido Baum Phone:

More information

Iowa Fine Particulate Monitoring Network Design Values 2011-2013. Iowa DNR Ambient Air Monitoring Group

Iowa Fine Particulate Monitoring Network Design Values 2011-2013. Iowa DNR Ambient Air Monitoring Group Iowa Fine Particulate Monitoring Network Design Values 2011-2013 Iowa DNR Ambient Air Monitoring Group What is Fine Particulate Matter (PM 2.5 )? The term particulate matter (PM) includes both solid particles

More information

Marine after-treatment from STT Emtec AB

Marine after-treatment from STT Emtec AB Marine after-treatment from STT Emtec AB For Your Vessel and the Environment 6 7 8 1 11 1 10 9 1. Pick up. Flow direction valve. Filters. Cooler. Condensate trap 6. Flow meter 7. EGR-valve 8. Secondary

More information

List of EA Publications. And International. Documents

List of EA Publications. And International. Documents EA/INF-01 List of EA Publications and International Documents Publication Reference EA-INF/01: 2016 List of EA Publications And International Documents PURPOSE This publication gives the list of EA documents

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1 Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification

More information

Handbook for Emission Measurements for Macedonia

Handbook for Emission Measurements for Macedonia EMISSION MEASUREMENT HANDBOOK 1 Macedonia 12.6.2012 Handbook for Emission Measurements for Macedonia Contents 1. Introduction... 4 2. Requirements for measurement site... 5 2.1 General... 5 2.2 Access

More information

LANCOM4 PORTABLE FLUE GAS MONITORING. NO NO2 NOx CO2 H2S SO2 CxHy. CO low CO high O 2 QUALITY CUSTOMER SOLUTIONS WWW.LANCOM4.COM

LANCOM4 PORTABLE FLUE GAS MONITORING. NO NO2 NOx CO2 H2S SO2 CxHy. CO low CO high O 2 QUALITY CUSTOMER SOLUTIONS WWW.LANCOM4.COM PORTABLE FLUE GAS MONITORING WWW.LANCOM4.COM i CO low CO high O 2 NO NO2 NOx CO2 H2S SO2 CxHy QUALITY CUSTOMER SOLUTIONS AMETEK LAND HAS BEEN BUILDING PRECISION MEASURING EQUIPMENT SINCE 1947. WE ARE SPECIALISTS

More information

Factors Affecting Precipitation of Calcium Carbonate

Factors Affecting Precipitation of Calcium Carbonate Factors Affecting Precipitation of Calcium Carbonate John A. Wojtowicz Chemcon Laboratory tests with clear solutions showed that precipitation of calcium carbonate does not occur in the ph range 7.5 to

More information