Lessons in Radiography Using the X-Ray Simulation Program Student Booklet
|
|
- Ethan Lang
- 7 years ago
- Views:
Transcription
1 Lessons in Radiography Using the X-Ray Simulation Program Student Booklet Prepared by the North Central Collaboration for Education in NDT. Partial support for this work was provided by the NSF-ATE (Advanced Technological Education) program through grant #DUE Opinions expressed are those of the authors and not necessarily those of the National Science Foundation.
2 Table of Contents Introduction... 1 Contents of this Training Packet... 1 Lesson One: Film Properties... 2 Introduction... 2 Objectives... 2 Log Relative Exposure Calculations... 4 Sensitometric Curve Calculations for AGFA Film... 5 Simulator Exercise... 6 Laboratory Exercise... 9 Lesson Two: Kilovoltage Effect on Contrast Introduction Objectives Calculations Simulator Exercise Writing Assignment Laboratory Exercise Lesson Three: Milliamperage and Time Relationships Introduction Objective Reference Terms Formulas for Time and milliamprage Calculations Simulator Exercise Laboratory Exercise Lesson Four: Creating an Exposure Chart Introduction Objective Reference Terms Calculations Simulator Exercise Laboratory Exercise Lesson Five: Source-to-Film Relationships Introduction Objective Reference Terms Simulator Exercise Calculations for source to film distance change Laboratory Exercise Lesson Six: Radiographic Equivalency Factors Introduction Objective Reference Terms Calculations Simulator Exercise Laboratory Exercise Lesson Seven: Defect Content Introduction Objective... 38
3 Reference Terms Simulator Exercise Laboratory Exercise Lesson Eight: Defect Shape, and Relationship to X-ray Source Introduction Objective Reference Terms Simulator Exercise Laboratory Exercise Lesson Nine: Image Magnification, Beam Divergence and Distortion Introduction Objective Reference Terms Geometric Enlargement (Magnification) Calculations Geometric Unsharpness Calculations Beam Coverage Calculations Simulator Exercise Laboratory Exercise Lesson Ten: Developing a Radiographic Technique Card Introduction Objective Reference Terms Simulator Exercise Laboratory Exercise... 56
4 Introduction Successful radiography depends on numerous variables that affect the outcome and quality of a radiograph. Many of these variables have a substantial effect on the results while others have only a minor effect. One of the difficulties encountered when learning the basics of radiography is that there is a relatively long lag time between when an exposure is made and when the resulting radiograph is ready for viewing. This makes it a time consuming exercise when the lesson call for multiple exposures to be made in order see the effects that certain exposure variables have on image quality. These lessons use an X- ray inspection simulation program that allows beginning radiographers to experiment with set-up and exposure variables. Simulated radiographs are produced in a matter of seconds providing immediate feed on the impact that these variables have on the image quality. The X-ray simulator (XRSIM) program used for these exercises was developed by researchers at the Center for NDE at Iowa State University. The XRSIM program allows physically accurate radiographic inspections to be simulated using a computer aided design (CAD) model of a part. These lesson plans that make use of the XRSIM program were developed with funding from the National Science Foundation (NSF). Contents of this Training Packet Each of the ten lessons in this packet includes: an introduction to the main subject of the lesson statement of the objectives definition of new terms exercises that make use of the X-ray simulation program exercises for hand-on lab activities 1
5 Lesson One: Film Properties Introduction In this lesson, the relationship between film speed and exposure will be evaluated. Calculations will be made using sensitometric curves to determine the best choice of film for a given exposure or visa-versa. The manufacturer of the film makes no difference, as all films will work in the same manner. Review the film manufacturer s literature available at your school before beginning this exercise. Objectives To develop an understanding of industrial radiographic film properties including the characteristic curves for various films. To learn to calculate changes in film speeds while maintaining a constant density of the radiograph. Reference Terms Radiographic sensitivity is the general term used to describe the smallest detail that can be seen using a radiograph. The term also refers to the ease with which images can be seen or details can be detected. Radiographic sensitivity, in other words, is the amount of information that can be found on a radiograph. There are two independent sets of factors that determine the level of radiographic sensitivity, definition and contrast. Definition refers to the sharpness of the outline in the image. This sharpness depends on the type of screens and film used. The sharpness of the outline also depends on the radiation energy used, the geometry of the radiographic specimen and the spot size of the radiation source. Radiographic contrast is the difference between the areas of radiographic density. This difference in the density depends on two factors, subject contrast and film contrast. Subject contrast is the ratio of X-ray intensities transmitted through two or more portions of different thickness in a specimen. The energy of the radiation used and its distribution, and the intensity of the radiation used have to be taken into account when determining the subject contrast. Film contrast refers to the slope of the characteristic curve of the film. This contrast depends on three factors, the type of film being used, the density of the film, and the way the film is processed. The process used to expose the film is also important. Of particularly importance is whether the film has been exposed using lead screens or fluorescent screens. 2
6 Film grains are the materials that capture the image, which after development becomes a radiograph. Film grains are made-up of extremely fine silver bromide crystals. There can be billions of grains in one square centimeter. The larger the grain size the faster the film. Industry uses a number of film speeds to produce the desired radiographic image depending on the quality of the radiograph sought. Logarithm is the number of the exponent of the power of 10 that a number is raised. Consider a six-inch scale as a linear form of measurement. If we wanted to represent a great distance in the physical distance of six inches, we would use a logarithmic measurement or scale. The logarithmic scale moves by factors of ten, and becomes more compressed the closer to ten the numbers become. Once ten is reached we began counting by one hundred, and then by one thousand. In radiography, and ultrasound the logarithmic scale will be used often. Sensitometric charts are comprised of a group of curves representing the relative sensitivity of various X-ray films to exposure. The curves are a plot of the film density versus the logarithm of the relative exposure they must receive to reach a particular density. Relative exposure means that one of the curves is used as the standard and all other curves are related to this curve by a factor. The charts are used to make comparisons between various films and to make exposure and density calculations when changing from one film to another. 3
7 Log Relative Exposure Calculations Use the sensitometric curves in Figure 1 on page 9 to perform the following calculations: Example. A density of 2.0 was produced using Type I film with an exposure time of 3.0 milliampminutes. A density of 3.5 on the same film is desired. Find the 2.0 density on the left of the chart; move horizontally to film Type I and then move down to the log relative exposure number at the base of the chart (1.30). Repeat these steps to find the log relative exposure number for a density of 3.5 (1.65). Subtract 1.30 from 1.65, which leaves Take the inverse log of this number, which is Now multiply the original time (3.0) by 2.24 to get the new exposure setting of 6.72 ma-minutes (4 ma for 1 minute 41 seconds) to achieve a 3.5 density on the same film. If the density were to be decreased, or if a second faster film were used (further to the left on the chart.), the original exposure time would be divided by the exposure factor. 1. If a density of 2.0 is produced using Type I film and an exposure of 25-mA-minutes, what exposure setting is needed to produce a density of 1.5 using the same film? 2. If an exposure of 8-mAm produces a density of 2.5 when a Type I film is used, what exposure is needed to produce the same density using a type 2 film? 3. Using a Type II film, a density of 2.0 is produced with an exposure of 11-mAm. What exposure is required to increase the density to 3.5 using the same film? 4. If a density of 1.5 is produced using Type II film and an exposure of 8-mAm, what exposure is necessary to produce a density of 2.5 on type 3 film? 5. If a density of 2.25 is produced using Type III film and an exposure of 21-mAm, what exposure in necessary to produce the same density on type II film? 6. If an exposure of 5-mAm produces a density of 2.0 using a Type I film is, what exposure is necessary to produce the same density using a type III film? 7. Using a Type II film, an exposure of 9-mAm produces a density of 2.5. What exposure is necessary to increase the density to 3.0 on the same film? 8. If an exposure of: 7-mAm density of 3.0 on a Type II film, what exposure is required to produce a 2.0 density on a type 3 film? 9. Using a Type I film, if an exposure of 14-mAm produces a density of 3.0, what exposure is needed to reduce the density to 1.5 on the same film? 10. If a density of 2.4 is produced on a type III film with an exposure of 26-mAm, what exposure is required to produce the same density on a type II film? 4
8 Sensitometric Curve Calculations for AGFA Film Using the film chart Figure 2 on page 10, calculate the exposure times needed to accommodate the density or film type changes. Example: An exposure of D-4 film for 1.5 minute produced a radiograph with a density of 1.5. The specification requires a density of 2.5. Use the sensitometric curves to select a new exposure time. First find the 1.5 density point then move horizontally to the D-4 film. Then move down the chart to locate the log number (2.51) for that exposure. Next move to the new density 2.5 and move horizontally to the D-4 film, and then down the chart to locate the log number (2.75). Now subtract the two log values and take the inverse log of this value ( = Take the inverse log of 0.24 to get ). Round this number to the nearest tenth and multiply it by the original exposure time (1.5 X 1.7 = 2.55 minutes) to get the exposure time needed for a 2.5 density on your radiograph. 1. If a density of 1.5 is produced on D8 film with an exposure of 9-mAm, what exposure is required to produce a density of 3.0 on D4 film? 2. What exposure is required to change the density on D5 film to 3.5 when an exposure of 12-mAm produces a density of 2.3? 3. If a density of 3.0 is produced with an exposure of 22-mAm when D3 film is used, what exposure will produce a density of 3.0 when D7 film is used? 4. When D2 film is used, a density of 2.25 is produced with an exposure of 17-mAm. What exposure is required to produce a density of 3.5 on D4 film? 5. Using D4 film, what exposure is required to increase the density to 3.0 when an exposure of 7-mAm produces a density of 1.5? 6. If an exposure of 3-mAm produces a density of 3.5 on D7 film, what exposure is necessary to produce a density of 3.0 on D5 film? 7. If a 24-mAm exposure produces a density of 1.5 on a D4 film, what exposure will produce a density of 2.5 on D7 film? 8. If a density of 2.0 is produced on D8 film with an exposure of 6-mAm, what exposure is needed to produce a density of 2.5 on D5 film? 9. Using D4 film, what exposure is required to raise the density to 3.5 when an exposure of 6-mAm produces a density of 2.5? 10. If a 7-mAm exposure produces a density of 3.5 on D5 film, what exposure is needed to produce a density of 2.0 on D3 film? 5
9 Simulator Exercise Use the XRSIM program to evaluate the relationship between film speed and exposure time by completing the following exercise. 1. Review the AGFA sensitometric curves in Figure 2. This chart shows the speeds and characteristics of various films. Review Table 1, which shows ratios of film speeds. Notice in Table 1 that all films are referenced to D-7 film. 2. In the XRSIM program, open the stepwedge11cm CAD file and select 2024Al (aluminum alloy 2024) as the material. Below your material select the display tab, and check translucent to allow the defect to be viewed within the part. Using the defect tab select the sphere and choose air as the material. Locate the defect below the surface on step five. Select the detector tab and change the film to D-7. Use the top view and click and drag the film to expand it so the film extends beyond the part (The stepwedge may need to be moved slightly to reveal the film.) Use the HOMX160 generator with an initial setting of 60 kv, 1 ma, and 7 seconds to produce an image with a density reasonably close to 2.5 on step five of the stepwedge. Use the view/slice option to evaluate the density of the stepwedge. NOTE: Remember to change the name of the XRSIM density image (xyz file) each time you produce a simulated radiograph or the system will save one file over the other. To open multiple images for viewing, go to open, change file type to density image and select the desired file. Open the slice option by left clicking on the Imageslice icon. When the slice window opens, select horizontal and move your cursor over step six on the stepwedge to measure the density. 3. Using the film characteristic curves, calculate a change of density from 2.5 to 3.5 on D-7 film. Produce an image using XRSIM to confirm the calculation. 4. Change the detector to D-5 (which has a smaller grain and is a slower film) and use the appropriate calculations to determine the setting needed to generate a density of 3.5 on step five. Produce an image to confirm the calculation. 5. Manufacturer s film speed ratio charts give an overview of the characteristic curve relationships. On Chart 3, note the speed ratio differences for the films produced by this company. Notice that the manufacturer has chosen D-7 as the standard or 0 film. D-7 will be the standard against witch other films will be compared. For example, if you exposed a D-7 film for 33-milliamp seconds, the chart indicates that a D-5 film would need to be exposed 1.8 times as long. Note the differences between the other films. Also note the change of exposure ratios as kilovoltage is increased. Why is this? 6
10 6. Using the film characteristic curves for AGFA films (Figure 1), compare two films, D- 4 and D-7. If a density of 2.0 were achieved when D-4 film is exposed for 43 mas at 100 kv, what would be the correct exposure if D-7 film were used? 7. After calculating the time difference look on the "relative exposure factor" chart and find the ratio between the two films. 8. Was the time change 3 to 1 as predicted by the exposure factor chart? Density Figure 1. Film sensitometric (characteristic) curves for three typical film types. 7
11 Figure 2. Film sensitometric curves for Agfa Structurix industrial X-ray films. Film Type Radiation 100 KV 200 KV ir 192 D D D D D D Figure 3. Approximate Relative Exposure Factors for Agfa Structurix Industrial X-ray films. 8
12 Laboratory Exercise This exercise requires the use of an X-ray generator. Prior to the operation of this equipment, the required safety training must have been completed and instructions on the safe operation of the X-ray equipment must have been received. 1. Produce a film radiograph of a 0.1 to 0.5-inch steel stepwedge using the X-ray system of your choosing in the laboratory. Select an exposure setting (kv, ma and time) for any one of the steps from an existing exposure chart for the system. Place a lead letter on the edge of the step and make the exposure. Did your exposure chart correctly predict the density you achieved? If not, what was the difference in density? 2. Using the manufacture exposure chart for the film used above, and the density produced on your first exposure, calculate the correct exposure time for a change of film (faster or slower). Show your calculations, and new time below. 3. Produce a second radiograph using the calculated change in time to compensate for the change of film. Allowing for some deviation (+/- 5 percent), did your calculations produce the density predicted? If not, identify some of the possible reasons that for the discrepancy. 9
13 Introduction LESSON TWO: KILOVOLTAGE EFFECT ON CONTRAST In this lesson, the relationship between energy of the X-rays and the image contrast will be evaluated. The energy of the X-rays is controlled by the electrical potential of the X-ray tube. This potential is generally in the range of thousands of volts. X-ray generators will have a control that allows the voltage to be adjusted and this is generally labeled in kilovolts. Therefore, it is common to speak of the energy of X-rays in terms of the voltage used to produce the X-rays and this value is often referred to as the kilovoltage. The kilovoltage has an effect on the radiographic image because the energy used to produce the X-rays is related to the penetrating power of the X-rays. The first requirement in developing a radiographic procedure is to determine what level of radiation is necessary to penetrate the object being inspected and produce an image in a practical exposure time. If too high of kilovoltage is used, image quality deteriorates. A number of methods have been developed for determining the kilovoltage needed to radiograph an object of certain thickness and material. The two methods that will be covered here are the Half-Value Layer Method (HVL) and the Fixed Exposure Method. Objectives To evaluate how X-ray energy (kilovoltage) effects radiograph quality. To investigate how different radiographic films affect image contrast. To lean to apply several multiple methods of determining the kilovoltage required for an exposure. Reference Terms Kilovoltage is a measure of the electrical potential between the anode and the cathode in the X-ray tube. The higher the kilovoltage, the shorter the wavelength and the greater the penetrating power of the X-rays produced. The energy level used for an exposure will be greatly influenced by the material being radiographed because the X- rays must have enough energy to penetrate through the material being inspected. The kilovoltage should be chosen carefully because: 1. scatter increases as the energy increases scatter increases 2. contrast sensitivity decreases as energy increases Latitude refers to the amount of thickness variation that a particular film will be able to image in one exposure. As latitude increases contrast will diminish, and vice versa. Half Value Layer is the thickness of a given material that will reduce the radiation passing through it by one half. The energy of the ray, and the atomic make-up of the material determine half value layers. As kilovoltage increases the half value layer will decrease. 10
14 Calculations Half-Value Layer Method (HVL) The HVL Method is a simplification of a very complex calculation and only takes a few variables into consideration. For example, it is assumed that the energy is a single wavelength and scatter is not a concern. The HVL Method uses values that have been previously calculated and can be found in Attenuation Coefficient Tables contained in various references such as the Nondestructive Testing Handbook, Volume Three on Radiography and Radiation Testing. Using this method: ideally the energy level is selected so that the part thickness to be penetrated is 5 Half-Value Layers. satisfactory results can be obtained if the deviation from this rule is no greater than a factor of 2 half-value layers. Therefore, the target range is 3 to 7 half-value layers, but at a value of three image contrast is questionable and at a value of 7 exposure time will be excessive. the attenuation coefficients for the material must be located from reference materials and converted to Half-Value Layers using the following formula. HVL AttenuationCoefficient Below are linear attenuation factors to be used with aluminum and steel at a number of kilovoltage settings. Note from the table below that kilovoltage selection has a great affect on the HVL. Kilovoltage Material Attenuation coefficient 50 Al Al Al Al St St St St St St 1.15 Example: A steel part that is 1.27 cm thick must be inspected. Choose a kilovoltage that should produce a radiograph in a reasonable amount of time. From experience, 150 kv is know to provide good penetrating power and should produce a radiograph in a relatively short period. From the table, 1.54 is the value that corresponds to 150 kev for steel. Using the formula, divided by 1.54 equals 0.45 cm cm is the half-value layer for 150 kv. Divided the part thickness by the half-value layer to get the number of half value layers for the chosen kilovoltage setting divided by.45 equals 2.8 or rounded up 3.0 halfvalue layers. This value is at the lower limit of the range and could result in a radiograph with questionable contrast sensitivity. The calculation should be repeated using a lower kilovoltage. 11
15 Work the calculation at 100kv and determine the number of half value layers. Half value layers at 100 kv. Which setting is closer to five half value layers 100 or 150 kv? Fixed Exposure Method In this method, the kilovoltage is adjusted to maintain a constant film density for radiographs of objects with different material thicknesses and properties. Time, milliamperage, and source-to-film distance are held constant. This method is not as widely used as the HVL Method because it requires considerable work experience to develop the knowledge to produce acceptable radiographs. Without these skills this method requires much trial and error. As you can imagine, this method would require numerous exposures and added time without the aid of the simulator. 12
16 Simulator Exercise Simulator Stepwedge Thickness. Step Step Step Step Step Step Step Step Step Step cm 1.0 cm 1.5 cm 2.0 cm 2.5 cm 3.0 cm 3.5 cm 4.0 cm 4.5 cm 5.0 cm 1. Using the simulator, open the stepwedge11cm CAD file and select 2024Al (aluminum alloy 2024) as the material. Select step nine on the stepwedge and place a void (air) defect 0.55 cm by 0.45 cm by 0.65 cm. in the center of the step. 2. Using the HVL Method, calculate five half-value layers to get the optimum kilovoltage for step nine on the stepwedge. Create an image with a density of 2.5 on the step using D-7 film. Save the image in a file (xy1) for comparison with images you will produce later. Show your calculations below. Optimum Kilovoltage for step nine. 3. Create two more images using D-7 film and maintaining a 2.5 density. Create the first image with the kilovoltage reduced by 50 percent and the second image with an increase in kilovoltage 3 times the optimum value calculated above. Which exposure produced the most easily recognizable defect? 4. Using the slice option to evaluate density, record the contrast ratio between the defect and the stepwedge for each exposure. Exposure 1. Exposure 2. Exposure 3. Was the exposure with the greatest contrast sensitivity the one where the defect was easiest to see? 5. Using the slice option to evaluate density of the steps. Which exposure produced the greatest number of steps within a density range of 1.0 to 4.0? Exposure 1. Exposure 2. Exposure 3. Did this exposure have the highest or lowest contrast sensitivity? 13
17 9. Note 6. Recalculate and repeat your three exposures using D-2 film. Show your calculations below. Which exposure produced the most easily recognizable defect? 7. Using the slice option to evaluate density, record the contrast ratio between the defect and the stepwedge for each exposure. Exposure 1. Exposure 2. Exposure 3. Was the exposure with the greatest contrast sensitivity the one where the defect was easiest to see? 8. Using the slice option to evaluate density of the steps. Which exposure produced the greatest number of steps within a density range of 1.0 to 4.0? Exposure 1. Exposure 2. Exposure 3. Did this exposure have the highest or lowest contrast sensitivity? any differences seen when changing films. Describe the contrast and latitude differences between the two films. 14
18 Writing Assignment Describe your observations of this laboratory activity in a two-paragraph summary, and be prepared to discuss those observations during the next class. In your summary, discuss which method of determining kilovoltage feel is best, and discuss why. Summarize your observations on why latitude increases and contrast decreases as kilovoltage increases. 15
19 Laboratory Exercise Using the Half-Value Layer method, calculate 3, 5 and 7 half-value layers and note the corresponding kilovoltage for a steel or aluminum stepwedge. Using an exposure chart developed for the system, compare the kilovoltage range for your materials. Are the kilovoltages reasonably close? Write a paragraph that discusses the correlation of the kilovoltage values produced by the half-value method and those on the exposure chart. Note any similarities and differences. 16
20 LESSON THREE: MILLIAMPERAGE AND TIME RELATIONSHIPS Introduction In this lesson, the relationship between milliamps and the time of an exposure will be examined. An understanding will be developed of the interaction between the amperage and time controls, and their relationship to the density of a radiograph. Calculations to adjust for time or milliamperage to obtain a certain density on the radiograph will be introduced. Objective To develop an understanding of the effects of milliamperage and time changes on the density of a radiograph. To learn to calculate the approximate density change on a radiograph when the milliamperage or time setting of the exposure are changed. Reference Terms Milliamperage is a term used to describe the amount of radiation produced by the X-ray system. The milliamperage is actually the units of the electrical current that is used to produce the radiation. Milliamp-minutes (mam) are the product of the amount of milliamps and the amount of time used to make the exposure. Since the exposure time is directly tied to the amount of current used (i.e. the more current used, the less time needed to make an exposure), it is convenient to note the product of these two value rather that the two individual values. Formulas for Time and milliamprage Calculations There is roughly a one-to-one direct relationship between a milliamperage or time change and the resulting change in image density. If time is held constant and only the milliamperage is varied, the following equation can be used. M M 1 2 D D Where: M 1 = Milliamp original M 2 = Milliamp new D 1 = Film density original D 2 = Film density new
21 Example calculation: With a ma setting of 3.0, a density of 1.0 is produced. What ma setting would be needed to increase the density to 2.0 with time and Kv remaining constant? 3.0 / x = 1.0 / 2.0. Crossmultiply and divide. 3.0 times 2.0 = 6. Divide 6 by 1.0 = 6.0. Your new ma setting will be 6.0 ma. Use the previous equation to calculate the missing variable. 1. M 1 = 5.0 M 2 =? D 1 = 3.0 D 2 = 2.4" 2. M 1 =? M 2 = 3.0 D 1 = 2.1 D 2 = M 1 = 6.0 M 2 = 4.0 D 1 = 3.8 D 2 =? 4. M 1 = 5.0 M 2 = 3.0 D 1 =? D 2 = M 1 = 2.0 M 2 = 4.0 D 1 =? D 2 = 3.3 Time will now be the variable, and milliamperage will remain constant. The calculation will be the same as above except that time replaces milliamperage as one of the variables. T1 T 2 D D 1 2 Where: T 1 = Time (in minutes or seconds) original T 2 = Time (in minutes or seconds) new D 1 = Film density original D 2 = Film density new 18
22 Use the equation above to determine the missing variable. 1. T 1 = 2 T 2 = 8.0 D 1 = 1.1 D 2 =? 2. T 1 = 15.0 T 2 =? D 1 = 2.1 D 2 = T 1 =? T 2 = 9.0 D 1 = 2.4 D 2 = T 1 =? T 2 = 12.0 D 1 = 2.6 D 2 = 2.1 Now put the two calculations together to address all three variables at the same time. In this calculation, remember that the milliamps and time are a product and need to be multiplied before working the calculation. M 1 T1 D1 M T D 1. M 1 = 3.0, T 1 = 12.0 M 2 = 4.0, T 2 = 5.0 D 1 = 3.6 D 2 =? 2. M 1 = 1.0, T 1 = 63.0 M 2 = 4.0, T 2 = 20.0 D 1 =? D 2 = M 1 = 4.0, T 1 = 40.0 M 2 = 4.0, T 2 = 27.0 D 1 = 3.9 D 2 =? 4. M 1 = 3.0, T 1 = 15.0 M 2 = 5.0, T 2 =? D 1 = 2.9 D 2 =
23 Simulator Exercise After participating in class discussion and viewing demonstrations with the simulator, complete the calculations for change of density, milliamperage, and time. Next, complete the following activity using the X-ray simulation program. 1. Use the following settings as a starting point to produce three images: Part CAD File Part Material Source to film distance Detector Energy Current Time Stepwedge11cm Titanium (Ti) 100 cm. D-5 film 150 kv 1 ma 7 seconds These settings should generate an image with a density close to Change the milliamperage by increasing the setting to 2.0. What is the density on step eight? 3. Reduce your milliamps to 1.0 and increase the time to 14 seconds. What is the density? 4. Was the resulting density the same for an increase in milliamps as it was for and increase in time? Explain the relationship between time and milliamperage. 5. An industry rule-of-thumb is that if the exposure time is doubled, the density will approximately double. Use the slice option to verify the density results. 6. Continue with the settings used above (1ma and 7 seconds). a. Calculate the change in time needed to produce a density of 2.5 b. Calculate the change in time needed to produce a density of Write a statement noting the results and conclusions drawn from this lesson. 20
24 Laboratory Exercise Name: 1. Produce two radiographs on any X-ray system in the laboratory using the steel or aluminum stepwedge. First use an available exposure chart to expose a step near the center of the stepwedge. 2. Choose any three steps on the processed radiograph and calculate a density change to raise or lower the density to 2.0 for the steps chosen. Show your work and results below. 3. Once your calculations are complete expose and process the stepwedge and film. Note the density on the appropriate step. Did you produce a density change as calculated? Attach your film and explain your findings below. 21
25 Lesson Four: Creating an Exposure Chart Introduction This lesson will explain how to create a radiographic exposure chart. As material thickness changes exposure times must be adjusted to produce an acceptable density. Exposure charts provide the proper milliamperage and time settings for a constant kilovoltage over a range of material thickness. Exposure charts are developed with the following held constant: kilovoltage, density, film type, film processing, source to film distance, use of lead screens, and material type. The materials thickness is the variable and the chart provides a fast and reliable way of determining exposure time that produces a certain density. Objective To learn to interpret radiographic exposure charts to select exposure settings. To learn to make density conversion calculations. To create radiographic exposure charts that will be used as guides for establishing exposure settings for a particular X-ray system, film and film processing parameters. Reference Terms Densitometer is an instrument for measuring photographic densities. A number of different types are available commercially. An important property is reliability; that is, the densitometer should reproduce readings from day to day. X-Ray Exposure Chart is a chart that shows the relationship between material thickness, kilovoltage, and exposure time. Film Development is the process used to turn a latent image into a viewable radiograph. Processing variables such as manual or automatic processing, time and temperature will affect the final radiograph and the outcome of the exposure chart. Film Speed is a system used to classify the relative exposure times needed to expose various films. The film speed is dependant on the grain size of the film and larger the film grain size, the less exposure time required to produce a certain density of the processed film. 22
26 Calculations Density Conversion Calculations Use the following equation to solve for the missing information. M 1 T1 D1 M T D Example: If 20.0 mam produces a density of 2.0, what exposure (amperage times time) would be needed to produce a density of 4.0? 20.0mAm? mam 2 min 4 min (20.0 mam times 4.0 min divided by 2 min = X) The new amperage would be 40.0 mam. 1. If 20.0 mam produces a density of 2.0, what exposure would be necessary to produce a density of 4.0? 2. If 15.0 mam produces a density of 2.3, what exposure would be necessary to produced a density of 3.6? 3. If 24.0 mam produces a density of 3.0, what exposure would be necessary to produced a density of 2.0? 4. If 10.0 mam produces a density of 1.5, what exposure would be necessary to produced a density of 3.1? 5. If 15.0 mam produces a density of 1.3, what exposure would be necessary to produced a density of 2.6? 6. If 22.0 mam produces a density of 1.8, what exposure would be necessary to produced a density of 2.6? 7. If 8.0 mam produces a density of 2.1, what exposure would be necessary to produced a density of 2.0? 8. If 18.0 mam produces a density of 1.3, what exposure would be necessary to produced a density of 2.6? 9. If 21.0 mam produces a density of 1.7, what exposure would be necessary to produced a density of If 14.0 mam produces a density of 2.6, what exposure would be necessary to produced a density of 3.5? 23
27 Simulator Exercise 1. Create two exposure charts using the simulator. Review the exposure chart on the following page before making your exposure chart. Use step eight on the simulator stepwedge. The step thicknesses are: Step cm Step cm Step cm Step cm Step cm Step cm Step cm Step cm Step cm Step cm 1. Place a 0.8 by 0.8 by 0.8 cm defect on the center of step eight for ease of location. To locate the defect, left click on the defect tab and then open. Choose the sphere, and it will appear on your stepwedge. Under the Flaw Operation menu, choose scale enter the dimensions for the flaw. Left click and drag the defect to the position you choose. To aid in location of the defect use the translucent option. Left click on the sample tab. At the lower right of the window there is a box marked translucent, left click. This will let you see the defect as you move it around in the stepwedge. 2. Using the knowledge gained in Lessons 1 and 2, calculate and choose the appropriate kilovoltage for aluminum and steel material. Choose an appropriate film. Film speed and type must always be considered when making an exposure chart in order to keep the exposure time reasonable. Denser materials, such as steel usually require a faster film than do less dense materials such as aluminum. Industry production demands and sensitivity requirements will dictate which film should be used. It often is not practical to x-ray steel with a very fine grain film that takes an extended time to expose, when a faster film will locate the needed defects. 3. Calculate the kilovoltage for steps one, eight and ten to determine the kilovoltage range for the span of thicknesses of the stepwedge. Record the kilovoltage. Steel Thickness step one Kilovoltage Thickness step eight Kilovoltage Thickness step ten Kilovoltage Aluminum Thickness step one Kilovoltage Thickness step eight Kilovoltage Thickness step ten Kilovoltage 5. Select the calculated kilovoltage for step number eight of the steel stepwedge and set the milliamperage and time to produce a density as close to 2.0 as practical on the step (not on the defect). 24
28 6. Create the image and use the slice option to record the densities from the steps beginning with the lightest. 7. Use the density conversion calculations to determine the exposure needed to produce a density of 2.0 on each of the steps. 8. Plot the exposure versus thickness on log graph paper. On the vertical edge of the log graph paper, plot the exposure in milliamp-minutes and on the horizontal line of the graph plot the material thickness. Place a mark at the intersection of these values and continue until all points for that kilovoltage setting are marked. Now connect the points to produce a kilovoltage reference line for a range of material thickness and time settings. The exposure chart on the next page can serve as an example. 9. Next, select the optimum kilovoltage calculated for step one and create another line on the chart. Repeat the process for step ten of the stepwedge. When complete the chart should show three kilovoltage lines. Remember to record all information including film, development method used, screens used, and source-to-film distance (SFD) in an information box. 10. Repeat the process for the aluminum material and produce a second chart. 11. Once the charts are complete select one thickness (step) for each kilovoltage listed on the exposure charts and produce an image using the simulator. Using the slice option record the density of the steps. Theoretically, the densities should all be 2.0 but some variation is to be expected. Exposure chart 1 Material Step thickness: Density Step thickness: Density Step thickness: Density Exposure chart 2 Material Step thickness: Density Step thickness: Density Step thickness: Density 25
29 Example X-ray Exposure Chart 100 milliamp seconds YX - 5 Film / No screens Auto Process 85 degrees Density: 2.0 FFD: 39 inches Source: X-ray / Model 160 Serial No. RA Material: Aluminum 1 0 ½ 1 1 ½ 2 2 ½ 3 3 ½ Thickness (inch) 26
30 Laboratory Exercise Name: Produce an exposure charts for steel or aluminum on one of the laboratory X-ray systems using a process similar to that used in the simulator exercise. Use a densitometer to measure the film density that corresponds to each step of the stepwedge on the radiograph. Generate lines for at least four kilovoltage settings on the chart. Kilovoltage settings should be in increments of 10 kilovolts or more. Remember to list any variable in the information box on the chart. When finished, turn in the following documents: Calculations used to develop the exposure chart. One exposure chart, for aluminum or for steel. Charts will show four kilovoltage settings over a given thickness range. A two paragraph description of your experience developing an exposure chart 27
31 LESSON FIVE: SOURCE-TO-FILM RELATIONSHIPS Introduction In this lesson, Newton s inverse square law will be used to calculate the film density when the source-to-film distance (SFD) has changed. Newton s inverse square law states that the intensity of the electromagnetic energy is inversely proportional to the square of the distance from the source. Therefore, reducing the SFD by one-half will increase the intensity of the radiation by a factor of four. Since the exposure time is directly related the radiation intensity, reducing the SFD by one-half would also reduce the exposure time by a factor of four. Often it is not practical to make exposures at the distance listed on an exposure chart. Reducing the SFD distance is sometimes necessary to speed the radiography of extremely thick parts, and increasing the SFD may be necessary when the physical shape of the part does not permit using distances listed on the exposure chart. Objective To develop an understanding of the effects of distance change on the density of a radiograph. To calculate and maintain a density on a radiograph when the source-to-film distance has been altered. Reference Terms Milliamperage-Distance Relationship is the dependence between the milliamperage required for a given exposure and the source-to-film distance. This relationship has been standardized to meet with manufactures ratings on the various x-ray tubes. Time-Distance Relationship is the dependence between the time required for a specific exposure and the source-to-film distance. The exposure time is indirectly proportional to the square of source-to-film distance. Newton s Inverse Square Law describes the radiation intensity as a function of distance from the source. The intensity reaching the specimen is inversely proportional to the square of the distance. Anytime the distance doubles the intensity of the radiation reaching the film is changed by a factor of four. Milliamprage, Time, Distance is the relationship between the milliamprage, and time required for a given exposure and source- to- film distance. This relationship can be easily calculated for any source- to- film distance change. Source-To-Film Distance is the measured distance the x-ray tube is from the film to be exposed. This distance will change for various reasons. Often there is a need to speed up production, hence the distance will be reduced thereby shorting the exposure time at a given density. 28
32 Simulator Exercise 1. After reading and classroom discussion, the following calculations should be completed and evaluated in class. 2. The following problem should be calculated and the answers confirmed on the number eight step of the aluminum stepwedge and D-3 film. Place a "void" flaw in the step for ease of location. Choose a kilovoltage setting appropriate for the thickness you are exposing (See lesson on kilovoltage selection). Establish a density of 2.5 at the default setting of 100 cm. This image will be saved in the density file as xyz.xbm. Verify the density using the slice option. Now change the source to film distance to (35cm), recalculate the new time setting for the distance change. Using the full and top view note the change in the radiation cone coverage of your film. Produce an image on the simulator using your new calculations. Name your image file xy1.xbm. Remember you must name the file before you produce the simulated image or the image will write over the xyz.xbm file. Check your density. Did the density remain at 2.5? if not what was the density? was it within 10 percent of your calculated change? 3. Next select the stepwedge choose Fe as your material. Establish a 2.5 density using AA 400 film on step seven. Place a void flaw in the step for ease of location. Calculate a change of distance to 125 cm while maintaining a 2.5 density. Save this image in a density file xy2.xbm. Using the full and top views note the change in the radiation cone. Retrieve the three radiographic image files from the density file and write a short description of the results. 4. Using the attached chart for change of distance produce another image with AA 400 film and a source to film distance of 25 inches (remember to convert cm to inches). The density on step seven should be near 2.5. Using the chart attached recalculate a change of distance to 40 inches. Did you maintain the 2.5 density? Write a short description of the results comparing the calculation method to use of the chart to determine Milliamp and time settings for a distance change. 29
33 Calculations for source to film distance change According to Newton s inverse square law, the intensity of the radiation intensity is inversely proportional to the square of the distance from the source. This can be represented by the following equation. Intensity 1 D 2 Since the exposure of a radiograph is inversely related to the radiation intensity (i.e. as intensity increases, the exposure must decreases to produce a constant density), it can be written that the exposure is directly proportional to the square of the source-to-film distance as represented by this equation. 2 Exposure SFD This equation can then be used to make calculations for changes in the source-to-film distance by setting two equations equal to each other. This equation is Exposure1 2 SFD1 Exposure SFD Example: an exposure for 16mAm with a source to film distance of 20 inches produces a density of 2.5. It is necessary to move the part closer to the source to reduce the exposure time but the density must be kept at 2.5. What would the new time be if the distance were moved to 10 inches from the source and all other factors remained the same? (As before cross multiply and divide. Four milliamp-minutes would be the new time for the distance change.) Use the above equation to solve for the missing variable in order to keep the density constant. 1. If 5.5 mam at a source-to-film distance of 12 inches produces a density of 3.0, what exposure is needed to produce the same density if the source is moved to 24 inches? 2. If 12.0 mam at a source-to-film distance of 21 inches produces a density of 2.50, what exposure is needed to produce the same density if the source is moved to 14 inches? 3. If 16.5 mam at a source-to-film distance of 18 inches produces a density of 3.3, what exposure is needed to produce the same density if the source is moved to 12 inches? 4. If 8.1 mam at a source-to-film distance of 16 inches produces a density of 3.6, what exposure is needed to produce the same density if the source is moved to 9 inches? 5. If 2.0 mam at a source-to-film distance of 6 inches produces a density of 2.0, what exposure is needed to produce the same density if the source is moved to 11 inches? 6. If 15.5 mam at a source-to-film distance of 21 inches produces a density of 3.0, what exposure is needed to produce the same density if the source is moved to 12 inches? 30
34 MILLIAMPERAGE-TIME AND DISTANCE RELATIONS When distances are changed from a given value to another given value on the chart the original time can be multiplied or divided by the factor given in the chart. New Dist. Old Dist. 25" 30" 35" 40" 45" 50" 25" " " " " "
35 Laboratory Exercise Name: 1. Produce a radiograph of a stepwedge using the X-ray system, film, and source-to-film distance of your choice. 2. Change the source-to-film distance and calculate a new exposure setting to produce a radiograph with densities similar to the first radiograph. 3. Compare the two exposures and write a two paragraph on effects of the source-to-film distance on exposure density. 32
36 Introduction LESSON SIX: RADIOGRAPHIC EQUIVALENCY FACTORS This lesson addresses the physical properties of materials and how these properties affect the ability of X-rays to pass through a given amount of material. The material properties of primary interest in radiography are the density and the atomic number. The higher the density of a material, the more difficult it is for radiation to penetrate the material. This concept was studied in Radiation Safety when calculating shielding. Technique charts and other materials used in industry to control radiographic procedures usually address only one material. When presented with a new material of a different density, new exposure times must be determined. This lesson will explain how radiographic equivalency factors can be used to compensate for changes in materials properties when exposure parameters for other materials are known. Objective To investigate the effect that material properties have on radiographic density. To learn to compensate for material changes when making exposures by using the exposure settings at a given kilovoltage for one material to estimate the settings for a second material. Reference Terms X-ray Absorption is a term used to describe the ability of a material to stop or absorb x- rays. X-ray absorption depends on three factors: the thickness of the material, the density of the material and the atomic number of the material. The atomic nature of the material influences the X-ray absorption more than the other two factors combined. Exposure Factor is a term used to describe the combined effect of milliamprage, time and source-to-film distance. A change in any of these factors will affect the density of the radiograph. Image quality and resolution is affected by the contrast, sensitivity, and definition of the image. Equivalence Factor is the value by which the thickness of a material is multiplied to give the thickness of a "standard" material. Equivalence factors provide a means of relating material of different densities so that exposure estimates can be made. 33
37 APPROXIMATE RADIOGRAPHIC EQUIVALENCE FACTORS Material 50 kv 100 kv 150 kv 220 kv 400 kv Magnesium Aluminum (pure) Aluminum Titanium Steel Steel Copper Zinc Brass 1.4* 1.3* 1.3* Inconel X Zirconium 2.3 Lead 14.0 Uranium Aluminum is taken as the standard metal at 50kV and 100kV. Steel becomes the standard at the higher voltages and at gamma ray energy levels. The thickness of the material of interest is multiplied by the corresponding factor to obtain the approximate equivalent thickness of the standard metal. The exposure settings that would normally be used to produce a radiograph of the standard metal at this thickness is used to produce the radiograph of the material of interest. *Tin or lead alloyed in the brass will increase these factors. Example: To radiograph 0.5 inch of copper at 220 kv, multiply 0.5 inch by 1.4 to obtain an equivalency thickness of 0.7 inch for steel. The exposure settings that would be used to produce a radiograph of 0.7 inch of steel are; therefore, use to produce the radiograph of 0.5 inch of copper. 34
CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
More informationX-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
More informationProduction of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
More informationStandard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
More informationSession 7 Bivariate Data and Analysis
Session 7 Bivariate Data and Analysis Key Terms for This Session Previously Introduced mean standard deviation New in This Session association bivariate analysis contingency table co-variation least squares
More informationAcceleration of Gravity Lab Basic Version
Acceleration of Gravity Lab Basic Version In this lab you will explore the motion of falling objects. As an object begins to fall, it moves faster and faster (its velocity increases) due to the acceleration
More informationPlots, Curve-Fitting, and Data Modeling in Microsoft Excel
Plots, Curve-Fitting, and Data Modeling in Microsoft Excel This handout offers some tips on making nice plots of data collected in your lab experiments, as well as instruction on how to use the built-in
More informationPhysics 42 Lab 4 Fall 2012 Cathode Ray Tube (CRT)
Physics 42 Lab 4 Fall 202 Cathode Ray Tube (CRT) PRE-LAB Read the background information in the lab below and then derive this formula for the deflection. D = LPV defl 2 SV accel () Redraw the diagram
More informationSpectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
More informationGeometric Optics Converging Lenses and Mirrors Physics Lab IV
Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The
More information1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002
05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical
More informationStudying Topography, Orographic Rainfall, and Ecosystems (STORE)
Studying Topography, Orographic Rainfall, and Ecosystems (STORE) Basic Lesson 3: Using Microsoft Excel to Analyze Weather Data: Topography and Temperature Introduction This lesson uses NCDC data to compare
More informationATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
More informationThis activity will show you how to draw graphs of algebraic functions in Excel.
This activity will show you how to draw graphs of algebraic functions in Excel. Open a new Excel workbook. This is Excel in Office 2007. You may not have used this version before but it is very much the
More informationMD5-26 Stacking Blocks Pages 115 116
MD5-26 Stacking Blocks Pages 115 116 STANDARDS 5.MD.C.4 Goals Students will find the number of cubes in a rectangular stack and develop the formula length width height for the number of cubes in a stack.
More informationUsing the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
More informationMODEL 9020 & 9020-OEM CORRATER TRANSMITTER USER MANUAL
Serial Number MODEL 9020 & 9020-OEM CORRATER TRANSMITTER USER MANUAL ROHRBACK COSASCO SYSTEMS, INC. 11841 East Smith Avenue Santa Fe Springs, CA 90670 Tel: (562) 949-0123 (800) 635-6898 Fax: (562) 949-3065
More informationNumber Sense and Operations
Number Sense and Operations representing as they: 6.N.1 6.N.2 6.N.3 6.N.4 6.N.5 6.N.6 6.N.7 6.N.8 6.N.9 6.N.10 6.N.11 6.N.12 6.N.13. 6.N.14 6.N.15 Demonstrate an understanding of positive integer exponents
More informationCoating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
More informationAtomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
More informationProduction of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
More informationX-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION
X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION David Bernard Dage Precision Industries Inc. Fremont, California d.bernard@dage-group.com ABSTRACT The x-ray inspection of PCB assembly processes
More informationSummary of important mathematical operations and formulas (from first tutorial):
EXCEL Intermediate Tutorial Summary of important mathematical operations and formulas (from first tutorial): Operation Key Addition + Subtraction - Multiplication * Division / Exponential ^ To enter a
More informationThea Omni Light. Thea Spot Light. Light setup & Optimization
Light setup In this tutorial we will learn how to setup lights inside Thea Studio and how to create mesh lights and optimize them for faster rendering with less noise. Let us have a look at the different
More informationRadiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
More informationGAMMA-RAY SPECTRA REFERENCES
GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.
More informationFREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5
Physics 161 FREE FALL Introduction This experiment is designed to study the motion of an object that is accelerated by the force of gravity. It also serves as an introduction to the data analysis capabilities
More informationObjectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials.
Lab 3 Tension Test Objectives Concepts Background Experimental Procedure Report Requirements Discussion Objectives Experimentally determine the yield strength, tensile strength, and modules of elasticity
More informationValor Christian High School Mrs. Bogar Biology Graphing Fun with a Paper Towel Lab
1 Valor Christian High School Mrs. Bogar Biology Graphing Fun with a Paper Towel Lab I m sure you ve wondered about the absorbency of paper towel brands as you ve quickly tried to mop up spilled soda from
More informationUltrasonic Wave Propagation Review
Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic
More informationELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.
More informationScope and Sequence KA KB 1A 1B 2A 2B 3A 3B 4A 4B 5A 5B 6A 6B
Scope and Sequence Earlybird Kindergarten, Standards Edition Primary Mathematics, Standards Edition Copyright 2008 [SingaporeMath.com Inc.] The check mark indicates where the topic is first introduced
More informationInvestigating Relationships of Area and Perimeter in Similar Polygons
Investigating Relationships of Area and Perimeter in Similar Polygons Lesson Summary: This lesson investigates the relationships between the area and perimeter of similar polygons using geometry software.
More informationUpdates to Graphing with Excel
Updates to Graphing with Excel NCC has recently upgraded to a new version of the Microsoft Office suite of programs. As such, many of the directions in the Biology Student Handbook for how to graph with
More informationCutting Edge Digital Radiography Technology in Daily NDT Use
Cutting Edge Digital Radiography Technology in Daily NDT Use Ron PINCU, Eli DAYAN, Ofra KLEINBERGER Vidisco Ltd. Or-Yehuda, Israel 60375 Tel: +972 (3) 533-3001 Email: ndt@vidisco.com ABSTRACT Digital Radiography
More informationPennsylvania System of School Assessment
Pennsylvania System of School Assessment The Assessment Anchors, as defined by the Eligible Content, are organized into cohesive blueprints, each structured with a common labeling system that can be read
More informationMicrosoft Excel Tutorial
Microsoft Excel Tutorial by Dr. James E. Parks Department of Physics and Astronomy 401 Nielsen Physics Building The University of Tennessee Knoxville, Tennessee 37996-1200 Copyright August, 2000 by James
More informationExperiment: Crystal Structure Analysis in Engineering Materials
Experiment: Crystal Structure Analysis in Engineering Materials Objective The purpose of this experiment is to introduce students to the use of X-ray diffraction techniques for investigating various types
More informationProper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center
Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster Jesse Garant, JG&A Metrology Center Traditional Metrology and Inspection Tactile Devices (Touch Probe)
More informationLab 4: Magnetic Force on Electrons
Lab 4: Magnetic Force on Electrons Introduction: Forces on particles are not limited to gravity and electricity. Magnetic forces also exist. This magnetic force is known as the Lorentz force and it is
More informationLearning Module 4 - Thermal Fluid Analysis Note: LM4 is still in progress. This version contains only 3 tutorials.
Learning Module 4 - Thermal Fluid Analysis Note: LM4 is still in progress. This version contains only 3 tutorials. Attachment C1. SolidWorks-Specific FEM Tutorial 1... 2 Attachment C2. SolidWorks-Specific
More informationPhysics Lab Report Guidelines
Physics Lab Report Guidelines Summary The following is an outline of the requirements for a physics lab report. A. Experimental Description 1. Provide a statement of the physical theory or principle observed
More informationSpreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different)
Spreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different) Spreadsheets are computer programs that allow the user to enter and manipulate numbers. They are capable
More informationThe Bending Strength of Pasta
The Bending Strength of Pasta 1.105 Lab #1 Louis L. Bucciarelli 9 September, 2003 Lab Partners: [Name1] [Name2] Data File: Tgroup3.txt On the cover page, include your name, the names of your lab partners,
More informationIndicator 2: Use a variety of algebraic concepts and methods to solve equations and inequalities.
3 rd Grade Math Learning Targets Algebra: Indicator 1: Use procedures to transform algebraic expressions. 3.A.1.1. Students are able to explain the relationship between repeated addition and multiplication.
More informationMap Patterns and Finding the Strike and Dip from a Mapped Outcrop of a Planar Surface
Map Patterns and Finding the Strike and Dip from a Mapped Outcrop of a Planar Surface Topographic maps represent the complex curves of earth s surface with contour lines that represent the intersection
More informationX-ray 100KV. Speed as compared to that of type IX100 used as a standard 100 under each exposure conditions.
DATA SHEET INDUSTRIAL RADIOGRAPHIC SYSTEMS RADIOGRAPHIC FILM IX20, IX25, IX50, IX80, IX, IX, IX29 AND IX59 Fujifilm industrial X-ray films feature revolutionary new film technology. The combination of
More informationAbsorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Calibration Curve Procedure
Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Calibration Curve Procedure Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry. Make sure you
More informationThe Basics of Scanning Electron Microscopy
The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single
More informationThe Circumference Function
2 Geometry You have permission to make copies of this document for your classroom use only. You may not distribute, copy or otherwise reproduce any part of this document or the lessons contained herein
More informationA Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion
A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion Objective In the experiment you will determine the cart acceleration, a, and the friction force, f, experimentally for
More informationHow do you compare numbers? On a number line, larger numbers are to the right and smaller numbers are to the left.
The verbal answers to all of the following questions should be memorized before completion of pre-algebra. Answers that are not memorized will hinder your ability to succeed in algebra 1. Number Basics
More informationUsing Excel (Microsoft Office 2007 Version) for Graphical Analysis of Data
Using Excel (Microsoft Office 2007 Version) for Graphical Analysis of Data Introduction In several upcoming labs, a primary goal will be to determine the mathematical relationship between two variable
More informationCorrelation key concepts:
CORRELATION Correlation key concepts: Types of correlation Methods of studying correlation a) Scatter diagram b) Karl pearson s coefficient of correlation c) Spearman s Rank correlation coefficient d)
More informationPHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.
More informationIntroduction to Solid Modeling Using SolidWorks 2012 SolidWorks Simulation Tutorial Page 1
Introduction to Solid Modeling Using SolidWorks 2012 SolidWorks Simulation Tutorial Page 1 In this tutorial, we will use the SolidWorks Simulation finite element analysis (FEA) program to analyze the response
More informationScientific Graphing in Excel 2010
Scientific Graphing in Excel 2010 When you start Excel, you will see the screen below. Various parts of the display are labelled in red, with arrows, to define the terms used in the remainder of this overview.
More informationReview of Fundamental Mathematics
Review of Fundamental Mathematics As explained in the Preface and in Chapter 1 of your textbook, managerial economics applies microeconomic theory to business decision making. The decision-making tools
More informationSHORE A DUROMETER AND ENGINEERING PROPERTIES
SHORE A DUROMETER AND ENGINEERING PROPERTIES Written by D.L. Hertz, Jr. and A.C. Farinella Presented at the Fall Technical Meeting of The New York Rubber Group Thursday, September 4, 1998 by D.L. Hertz,
More informationYears after 2000. US Student to Teacher Ratio 0 16.048 1 15.893 2 15.900 3 15.900 4 15.800 5 15.657 6 15.540
To complete this technology assignment, you should already have created a scatter plot for your data on your calculator and/or in Excel. You could do this with any two columns of data, but for demonstration
More informationIllinois State Standards Alignments Grades Three through Eleven
Illinois State Standards Alignments Grades Three through Eleven Trademark of Renaissance Learning, Inc., and its subsidiaries, registered, common law, or pending registration in the United States and other
More informationDigital Energy ITI. Instrument Transformer Basic Technical Information and Application
g Digital Energy ITI Instrument Transformer Basic Technical Information and Application Table of Contents DEFINITIONS AND FUNCTIONS CONSTRUCTION FEATURES MAGNETIC CIRCUITS RATING AND RATIO CURRENT TRANSFORMER
More informationThe photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
More informationHandbook on the Ultrasonic Examination. Austenitic Welds
Handbook on the Ultrasonic Examination Austenitic Welds The International Institute of Welding Edition Handbook On the Ultrasonic Examination of Austenitic Welds Compiled by COMMISSION V Testing, Measurement,
More informationTensile Testing Laboratory
Tensile Testing Laboratory By Stephan Favilla 0723668 ME 354 AC Date of Lab Report Submission: February 11 th 2010 Date of Lab Exercise: January 28 th 2010 1 Executive Summary Tensile tests are fundamental
More informationNASA Explorer Schools Pre-Algebra Unit Lesson 2 Student Workbook. Solar System Math. Comparing Mass, Gravity, Composition, & Density
National Aeronautics and Space Administration NASA Explorer Schools Pre-Algebra Unit Lesson 2 Student Workbook Solar System Math Comparing Mass, Gravity, Composition, & Density What interval of values
More informationImage Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
More informationMathematics Navigator. Misconceptions and Errors
Mathematics Navigator Misconceptions and Errors Introduction In this Guide Misconceptions and errors are addressed as follows: Place Value... 1 Addition and Subtraction... 4 Multiplication and Division...
More informationVarious Technics of Liquids and Solids Level Measurements. (Part 3)
(Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,
More informationAlgebra 1 2008. Academic Content Standards Grade Eight and Grade Nine Ohio. Grade Eight. Number, Number Sense and Operations Standard
Academic Content Standards Grade Eight and Grade Nine Ohio Algebra 1 2008 Grade Eight STANDARDS Number, Number Sense and Operations Standard Number and Number Systems 1. Use scientific notation to express
More informationMAG Magnetic Fields revised July 24, 2012
MAG Magnetic Fields revised July 24, 2012 (You will do two experiments; this one (in Rock 402) and the Magnetic Induction experiment (in Rock 403). Sections will switch rooms and experiments half-way through
More informationMeasurement with Ratios
Grade 6 Mathematics, Quarter 2, Unit 2.1 Measurement with Ratios Overview Number of instructional days: 15 (1 day = 45 minutes) Content to be learned Use ratio reasoning to solve real-world and mathematical
More informationAP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
More informationPrinciple of Thermal Imaging
Section 8 All materials, which are above 0 degrees Kelvin (-273 degrees C), emit infrared energy. The infrared energy emitted from the measured object is converted into an electrical signal by the imaging
More informationGEAROLOGY 4-1 WORMS AND WORM GEARS WORMS AND WORM GEARS
GEAROLOGY 4-1 4 4-2 GEAROLOGY COMMON APPLICATIONS: Worm and worm gear sets are used in many, everyday products including: electrical mixers, hubometers, right Now that you have an understanding of two
More informationCommon Core Unit Summary Grades 6 to 8
Common Core Unit Summary Grades 6 to 8 Grade 8: Unit 1: Congruence and Similarity- 8G1-8G5 rotations reflections and translations,( RRT=congruence) understand congruence of 2 d figures after RRT Dilations
More informationCreating Charts in Microsoft Excel A supplement to Chapter 5 of Quantitative Approaches in Business Studies
Creating Charts in Microsoft Excel A supplement to Chapter 5 of Quantitative Approaches in Business Studies Components of a Chart 1 Chart types 2 Data tables 4 The Chart Wizard 5 Column Charts 7 Line charts
More informationTutorial for Tracker and Supporting Software By David Chandler
Tutorial for Tracker and Supporting Software By David Chandler I use a number of free, open source programs to do video analysis. 1. Avidemux, to exerpt the video clip, read the video properties, and save
More informationThe atomic packing factor is defined as the ratio of sphere volume to the total unit cell volume, or APF = V S V C. = 2(sphere volume) = 2 = V C = 4R
3.5 Show that the atomic packing factor for BCC is 0.68. The atomic packing factor is defined as the ratio of sphere volume to the total unit cell volume, or APF = V S V C Since there are two spheres associated
More informationwith functions, expressions and equations which follow in units 3 and 4.
Grade 8 Overview View unit yearlong overview here The unit design was created in line with the areas of focus for grade 8 Mathematics as identified by the Common Core State Standards and the PARCC Model
More informationPart 1: Background - Graphing
Department of Physics and Geology Graphing Astronomy 1401 Equipment Needed Qty Computer with Data Studio Software 1 1.1 Graphing Part 1: Background - Graphing In science it is very important to find and
More informationCommon Core State Standards for Mathematics Accelerated 7th Grade
A Correlation of 2013 To the to the Introduction This document demonstrates how Mathematics Accelerated Grade 7, 2013, meets the. Correlation references are to the pages within the Student Edition. Meeting
More informationThe Effect of Forced Air Cooling on Heat Sink Thermal Ratings
zpero 1 The Effect of Forced Air Cooling on Heat Sink Thermal Ratings By Paul Bachman, Fellow Engineer & Ronnie Haiduk, Applications Engineer, Crydom, Inc. ABSTRACT A heat sink s ability to dissipate thermal
More informationStructural Integrity Analysis
Structural Integrity Analysis 1. STRESS CONCENTRATION Igor Kokcharov 1.1 STRESSES AND CONCENTRATORS 1.1.1 Stress An applied external force F causes inner forces in the carrying structure. Inner forces
More informationCAMI Education linked to CAPS: Mathematics
- 1 - TOPIC 1.1 Whole numbers _CAPS curriculum TERM 1 CONTENT Mental calculations Revise: Multiplication of whole numbers to at least 12 12 Ordering and comparing whole numbers Revise prime numbers to
More informationMagnetic Fields and Their Effects
Name Date Time to Complete h m Partner Course/ Section / Grade Magnetic Fields and Their Effects This experiment is intended to give you some hands-on experience with the effects of, and in some cases
More informationReflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
More informationThe promise of ultrasonic phased arrays and the role of modeling in specifying systems
1 modeling in specifying systems ABSTRACT Guillaume Neau and Deborah Hopkins This article illustrates the advantages of phased-array systems and the value of modeling through several examples taken from
More informationAmerican Society of Agricultural and Biological Engineers
ASAE S580.1 NOV2013 Testing and Reporting Solar Cooker Performance American Society of Agricultural and Biological Engineers ASABE is a professional and technical organization, of members worldwide, who
More informationTemperature Scales. The metric system that we are now using includes a unit that is specific for the representation of measured temperatures.
Temperature Scales INTRODUCTION The metric system that we are now using includes a unit that is specific for the representation of measured temperatures. The unit of temperature in the metric system is
More informationChapter 8 Design of Concrete Mixes
Chapter 8 Design of Concrete Mixes 1 The basic procedure for mix design is applicable to concrete for most purposes including pavements. Concrete mixes should meet; Workability (slump/vebe) Compressive
More informationis the power reference: Specifically, power in db is represented by the following equation, where P0 P db = 10 log 10
RF Basics - Part 1 This is the first article in the multi-part series on RF Basics. We start the series by reviewing some basic RF concepts: Decibels (db), Antenna Gain, Free-space RF Propagation, RF Attenuation,
More informationTensile Testing of Steel
C 265 Lab No. 2: Tensile Testing of Steel See web for typical report format including: TITL PAG, ABSTRACT, TABL OF CONTNTS, LIST OF TABL, LIST OF FIGURS 1.0 - INTRODUCTION See General Lab Report Format
More informationAtomic Force Microscope and Magnetic Force Microscope Background Information
Atomic Force Microscope and Magnetic Force Microscope Background Information Lego Building Instructions There are several places to find the building instructions for building the Lego models of atomic
More informationWaves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide)
Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide) OVERVIEW Students will measure a sound wave by placing the Ward s DataHub microphone near one tuning fork A440 (f=440hz). Then
More informationNON DESTRUCTIVE TESTING & ASNT WELD INSPECTION & AWS
NON DESTRUCTIVE TESTING & ASNT WELD INSPECTION & AWS What is ASNT? What is ASNT? The American Society for Nondestructive Testing, Inc. What is ASNT? ASNT is the world's largest technical society for nondestructive
More informationMATH 60 NOTEBOOK CERTIFICATIONS
MATH 60 NOTEBOOK CERTIFICATIONS Chapter #1: Integers and Real Numbers 1.1a 1.1b 1.2 1.3 1.4 1.8 Chapter #2: Algebraic Expressions, Linear Equations, and Applications 2.1a 2.1b 2.1c 2.2 2.3a 2.3b 2.4 2.5
More informationPre-Algebra 2008. Academic Content Standards Grade Eight Ohio. Number, Number Sense and Operations Standard. Number and Number Systems
Academic Content Standards Grade Eight Ohio Pre-Algebra 2008 STANDARDS Number, Number Sense and Operations Standard Number and Number Systems 1. Use scientific notation to express large numbers and small
More information