1 QUALITY THAT COUNTS

2 Daily Maintenance (1) Morning Start up Check reagent containers, fill if needed and be sure to rinse bottles prior to refilling When filling Trypan Blue reagent container filter stock Trypan Blue prior to use Empty waste container In Liquid Management perform Prime Visually check syringe and all lines NOTE Please keep all reagent containers and sample cups covered and free of dust!

3 Daily Maintenance (2) Runs throughout the day Before each run Empty waste container if greater than 50% full Check that system has been primed and there are no air pockets or bubbles in capillary tubes After each run (10-20 samples) Perform Standard Clean (1.5 min) Perform Chamber Clean 2 (0.5 min) Empty waste container

4 Daily Maintenance (3) End of Day Shutdown Perform Standard Clean (1.5 min) Perform Chamber Clean 2 (0.5 min) Perform LM Shutdown Empty waste container

5 Weekly Maintenance (1) At the Start of the week (Monday) Clean the reagent containers with distilled or de-ionized water and refill with new reagents. Make sure to replace the Trypan Blue solution each week. Clean external surface of capillary tubes, which reach into the sample vial & reagent containers, with a lint-free cloth. Remove and clean syringe (using 70% ethanol and distilled or de-ionized water).

6 Weekly Maintenance (2) At the End of the week (Friday) Perform Standard Clean (1.5 min) Perform Chamber Clean 2 (0.5 min) Perform Intensive Clean (20 min) Perform LM Shutdown Empty waste container

7 (Bi-)Monthly Maintenance Perform two Intensive Cleans with 0.1N to 1N HCl instead of Cedex Detergent Procedure Replace Detergent in reservoir with 0.1N to 1N HCl Run Prime Start Intensive Clean Replace HCl in reservoir with detergent Run Prime Run Standard Clean Use Syringe Kit to Backflush flow cell in reverse direction

8 Syringe Backflush Routine Detach the tubing and flangeless fittings from position 8 on the multi-way valve (far bottom right hand side of multi-way valve while looking at front of Cedex). Attach tubing and flangeless fittings from position 8 to receiving port of tubing (supplied in kit). Place open end of tubing into empty beaker. Unscrew flangeless nut and ferrule from syringe, supplied in kit. Fill the syringe with DI, particle-free water. Slide the flangeless nut over frosted tubing leading into waste container (In addition to the frosted tubing, there is also clear tubing leading into the waste container). Do not use this tubing with the nut threads facing the open end of the tubing. Slip the flangeless ferrule over the tubing, with the tapered position of the ferrule facing toward the nut.

9 Syringe Backflush Routine Insert the frosted tubing with the flangeless nut and ferrule in place into the receiving port of the syringe filled with DI, particle-free water and tighten the nut finger tight. Alternate pushing liquid and air from the syringe through the flowcell by rotating the syringe. Pull the plunger in the reverse direction to allow more air into the syringe. Continue pushing liquid and air through the syringe. When syringe is empty, detach syringe from tubing and reattach tubing and flangeless fittings finger tight back into position 8. Perform a Standard Clean and check to make sure that there are no leaks coming from position number 8.

10 Version 1.0, January, 2006 This Cleaning and Maintenance Standard Operation Procedure is valid for Cedex 1.3, 1.5, 1.6, 1.7 (Standard) and 1.10 (HiRes). Your individual innovatis contact persons: Service & Support: Product Management: Sales: Phone: Phone: Phone:

11 innovatis AG Meisenstr Bielefeld Germany Phone: +49 (0) Fax: +49 (0) innovatis Inc. 301 Lindenwood Drive, Suite 115 Malvern, PA USA Phone: Toll free Fax: Toll free Phone:

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588).

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588). User Bulletin ABI PRISM 310 Genetic Analyzer 01/2007 SUBJECT: Syringe Care and Maintenance Overview Safety Information Warranty Information In this User Bulletin This user bulletin describes the care,

More information

Replacement Instructions ABI 431A/433A Waste Port

Replacement Instructions ABI 431A/433A Waste Port Replacement Instructions ABI 431A/433A Waste Port February 1996 (updated 09/2002) IMPORTANT Do not use the installation instructions contained within the waste port assembly packaging because they are

More information

AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU680 Weekly (Analyzer and ISE) Maintenance

AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU680 Weekly (Analyzer and ISE) Maintenance AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance Step 1 Action Perform the Manually Clean the ISE Mix Bar, Liquid Level Sensors, Sample Pot and Sample Pot Tubing procedure AU680 Weekly (Analyzer

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

FUEL CELL EXPERIMENTATION

FUEL CELL EXPERIMENTATION Grade Level: 6-8 Group/Team Size: 2 Time Required: 120 minutes Summary: This activity will guide students through four experiments with a hydrogen fuel cell to learn how to fill the unit, decompose water

More information

The Drink-Aide The NEW look of independence

The Drink-Aide The NEW look of independence The Drink-Aide The NEW look of independence Introducing Drink-Aide What is Drink-Aide? The Drink-Aide is a water bottle that attaches easily to a wheelchair for use by persons with little or no upper

More information

Bydureon 2 mg powder and solvent for prolonged-release suspension for injection

Bydureon 2 mg powder and solvent for prolonged-release suspension for injection INSTRUCTIONS FOR THE USER Your step by step guide Bydureon 2 mg powder and solvent for prolonged-release suspension for injection If you have questions about taking Bydureon Refer to the Common questions

More information

Operator Quick Guide EC SENSOR

Operator Quick Guide EC SENSOR Operator Quick Guide EC SENSOR Revision G - 24/04/2008 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra

More information

REPAIR AND MAINTENANCE MANUAL

REPAIR AND MAINTENANCE MANUAL REPAIR AND MAINTENANCE MANUAL Professional Pest Control Application Equipment Prime Line Sprayers B&G Equipment Company 135 Region South Drive, Jackson, GA 30233 800.544.8811 Phone 678.688.5601 Fax 678.688.5633

More information

Installation Instructions 6028.801

Installation Instructions 6028.801 DAZZLE Installation Instructions 08.80 Spread Lavatory Faucet with Speed Connect Drain* Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American

More information

Atrium Pneumostat Chest Drain Valve. Discharge Instructions

Atrium Pneumostat Chest Drain Valve. Discharge Instructions Atrium Pneumostat Chest Drain Valve Discharge Instructions Your Physician Name: Contact Number for Emergencies: Introduction Your physician has decided to change your chest drainage system to a smaller

More information

Sample Collection Procedures for Insulating Fluid Samples

Sample Collection Procedures for Insulating Fluid Samples Sample Collection Procedures for Insulating Fluid Samples GENERAL Sampling electrical insulating oil for laboratory analysis requires care to ensure that the sample is representative of the product being

More information

Before repairing your toilet, you must determine

Before repairing your toilet, you must determine Before repairing your toilet, you must determine which type of toilet you have. Pressurized Toilets If you have a pressurized toilet, it is recommended that you call a licensed plumbing contractor to repair

More information

Cleaning Instructions, Pilot Replacement and Valve Change. Model No.: 233010 Natural Gas

Cleaning Instructions, Pilot Replacement and Valve Change. Model No.: 233010 Natural Gas Cleaning Instructions, Pilot Replacement and Valve Change Model No.: 233010 Natural Gas 8mm open-end wrench 9mm open-end wrench 10mm open-end wrench 12mm open-end wrench 13mm open-end wrench Phillips screw

More information

Agilent High Sensitivity DNA Kit Guide

Agilent High Sensitivity DNA Kit Guide Agilent High Sensitivity DNA Kit Guide Agilent High Sensitivity DNA Agilent Technologies Notices Agilent Technologies, Inc. 2009, 2013 No part of this manual may be reproduced in any form or by any means

More information

Installation Instructions 4508 4508S

Installation Instructions 4508 4508S SYMPHONY Spread Lavatory Faucet with Speed Connect Drain Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American Standard faucets. Speed Connect

More information

INSULIN INJECTION KNOW-HOW

INSULIN INJECTION KNOW-HOW 0-0- INSULIN INJECTION KNOW-HOW Learning how to Congratulations for making the move to insulin therapy. It won t be long before you start enjoying better blood sugar control, more energy, and a host of

More information

Installing a Capillary Column into an ECD and a Micro ECD

Installing a Capillary Column into an ECD and a Micro ECD Installing a Capillary Column into an ECD and a Micro ECD Procedure: Preparing capillary columns You must prepare your capillary column before installation. Proper preparation assures that the column end

More information

ON TRACK Reprocessing Inservice / Competency for GI Endoscopes (EVIS, EXERA, EXERA II)

ON TRACK Reprocessing Inservice / Competency for GI Endoscopes (EVIS, EXERA, EXERA II) THIS CHECKLIST IS DESIGNED FOR USE SOLELY AS A CUSTOMER EDUCATIONAL TOOL, AND IS NOT INTENDED TO REPLACE OR ANY WAY MODIFY THE OLYMPUS INSTRUCTION MANUAL/REPROCESSING MANUAL. BE SURE TO FOLLOW THE DETAILED

More information

CHROME FRONT BRAKE MASTER CYLINDER KIT

CHROME FRONT BRAKE MASTER CYLINDER KIT -J075 REV. 009-0-0 CHROME FRONT BRAKE MASTER CYLINDER KIT GENERAL Kit Number 5-99D, 5-99D Models These Chrome Master Cylinder Kits are designed to replace the original equipment front brake master cylinder

More information

Volkswagen Jetta, Golf, GTI 1999, 2000 Brake System 47 Brakes - Hydraulic Components (Page GR-47)

Volkswagen Jetta, Golf, GTI 1999, 2000 Brake System 47 Brakes - Hydraulic Components (Page GR-47) 47 Brakes - Hydraulic Components (Page GR-47) FS III front brake calipers, servicing Front brake caliper piston, removing and installing FN 3 front brake calipers, servicing Front caliper piston, removing

More information

Installation and Maintenance Instructions

Installation and Maintenance Instructions Installation and Maintenance Instructions Limited One Year Warranty T&S warrants to the original purchaser (other than for purposes of resale) that such product is free from defects in material and workmanship

More information

Coating and Extraction of Honeycomb Denuders

Coating and Extraction of Honeycomb Denuders QSM Approval: Coating and Extraction of Honeycomb Denuders 1 INTRODUCTION 1.1 The following procedures are used for the coating of honeycomb denuders with citric acid and sodium carbonate solutions, and

More information

CYLINDER REFILL AND INSPECTION

CYLINDER REFILL AND INSPECTION Refill/Inspection Procedure 1 CYLINDER REFILL AND INSPECTION REVISION 2.01 DATE 11.1.2013 1.0 General Information This procedure outlines the process in which Authorized BCA refill centers should complete

More information

Procenter Delivery System* Troubleshooting Charts

Procenter Delivery System* Troubleshooting Charts 1 2 3 4 Procenter Delivery System* Troubleshooting Charts * includes International (Whip) Units Component Page Procenter / Int l. (Whip) Delivery Units*... 2 Assistant s Units...6 Midmark Cuspidor...7

More information

SIBIR MANUAL V110 KE. English page 5 RKE - 1D 822 70 66-02

SIBIR MANUAL V110 KE. English page 5 RKE - 1D 822 70 66-02 SIBIR MANUAL RKE - 1D V110 KE English page 5 822 70 66-02 2 1 A. Flue tube E. Adapter I. Thermostat B. Flue baffle F. Burner base J. Lever arm C. Control knob G. Lamp glass D. Fuel gauge H. Lamp glass

More information

After completing this module, you will be able to: Describe the BD FACScan system and how it works.

After completing this module, you will be able to: Describe the BD FACScan system and how it works. BD FACScan System After completing this module, you will be able to: Describe the BD FACScan system and how it works. Perform instrument startup and shutdown procedures. Describe instrument maintenance

More information

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown) Service Manual Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com

More information

Joplin Rebuild Instructions

Joplin Rebuild Instructions Joplin Rebuild Instructions This is a step by step list of instruction on how to rebuild the Crankbrothers Joplin post. It may not always be required to completely tear-down the post to do the work needed

More information

NITROUS TRANSFER PUMP INSTRUCTIONS

NITROUS TRANSFER PUMP INSTRUCTIONS NITROUS TRANSFER PUMP INSTRUCTIONS SAFETY TIPS Never directly inhale nitrous oxide. When inhaled in large quantities, nitrous oxide can cause respiratory ailments or in extreme cases, death by suffocation.

More information

Instructions for Use. Components of the GENOTROPIN PEN 12

Instructions for Use. Components of the GENOTROPIN PEN 12 Instructions for Use Important Note Please read these instructions completely before using the GENOTROPIN PEN 12. If there is anything you do not understand or cannot do, call the toll-free number listed

More information

TROUBLESHOOTING FLOW CHART

TROUBLESHOOTING FLOW CHART TROUBLESHOOTING FLOW CHART Revised and updated on January 16, 2003, by Tony Leger and Automation Lab Technology. FACScan(FC)\FACSort(FST) FACScalibur (FCB) Fluidics/Pneumatics troubleshooting guide DCM

More information

ArtisanLink Staining System is an automated special stains slide

ArtisanLink Staining System is an automated special stains slide Technical Tips Tips on using the ArtisanLink Special Staining System Jamie Nowacek, BS, HT(ASCP) CM, QIHC, PMP Dako North America, Inc. Carpinteria, CA, USA ArtisanLink Staining System is an automated

More information

INSTRUCTIONS 56-1190-98. Edition AC

INSTRUCTIONS 56-1190-98. Edition AC Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography

More information

CARING FOR YOUR WATER HEATER

CARING FOR YOUR WATER HEATER http://waterheatertimer.org/troubleshoot-rheem-tankless-water-heater.html Water Heater Inspections CARING FOR YOUR WATER HEATER Venting System (Direct Vent Only) The venting system should be inspected

More information

Peripheral Venous Catheter Placement Simulator

Peripheral Venous Catheter Placement Simulator Attention Do not let ink from pens, newspapers, this manual or other sources come in contact with the manikin, as they cannot be cleaned the manikin skin. MW9 Peripheral Venous Catheter Placement Simulator

More information

PLUMBING CARE AND REPAIR

PLUMBING CARE AND REPAIR PLUMBING CARE AND REPAIR No special skills are required for any of the repairs described in this section. Plumbing repairs are seldom dangerous unless leaking water is soaking areas where there are electrical

More information

Operator s Manual. Pressure Injection Cell Model PC8500. Congratulations! Contents

Operator s Manual. Pressure Injection Cell Model PC8500. Congratulations! Contents Operator s Manual Pressure Injection Cell Model PC8500 Congratulations! Congratulations on your purchase of a Next Advance Pressure Injection Cell. Please read this operator s manual which explains proper

More information

User Manual Important: First read the Medication Guide that comes inside your FORTEO carton.

User Manual Important: First read the Medication Guide that comes inside your FORTEO carton. 1 RA082FSAM02 User Manual Important: First read the Medication Guide that comes inside your FORTEO carton. Before you use your new FORTEO delivery device, please read the entire front and back of this

More information

Routine Maintenance for Atomic Absorption Spectrophotometers

Routine Maintenance for Atomic Absorption Spectrophotometers Routine Maintenance for Atomic Absorption Spectrophotometers Application Note Atomic Absorption Author Margaret A. Cunliffe Introduction Instruments in good operating condition are a necessity in any analytical

More information

Line to Refrigerator Ice/Water Dispenser

Line to Refrigerator Ice/Water Dispenser Standard 18 Line to Refrigerator Ice/Water Dispenser Pump Module Bottled Water How The System Works The FLOJET Bottled Water Dispensing System was designed to pump water from a commercially available 5-gallon

More information

How To Replace A Reverse Osmosis Water Tank

How To Replace A Reverse Osmosis Water Tank ALL MODELS REVERSE OSMOSIS WITH NONAIRGAP FAUCET 1. Read all instructions carefully before starting installation. 2. Find the cold water line beneath your sink. The cold water is typically on the right.

More information

USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model # 900172

USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model # 900172 Model # 900172: Page 1 USER MANUAL Bottom Loading Bottled Water Dispenser Model # 900172 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING

More information

National- Spencer Inc.

National- Spencer Inc. 9-27-2010 National- Spencer Inc. 19.2V HEAVY DUTY GREASE GUN PRODUCT SPECIFICATION Charger Input Power 110 VAC Battery Output Power 19.2V Battery Capacity 1500 MAH Battery Pack Charge Time 1 Hour Maximum

More information

CADD-legacy PCA pump Model 6300-100ml Cassette

CADD-legacy PCA pump Model 6300-100ml Cassette HOW TO START THE CADD PUMP Press and hold the stop/start button CADD-legacy PCA pump Model 6300-100ml Cassette Three dashes will count down under the STARTING message STOPPED HOW TO STOP THE CADD PUMP

More information

February 2010. Xerox 8265/8290 Color Wide Format Printer User Maintenance and Cleaning Guide

February 2010. Xerox 8265/8290 Color Wide Format Printer User Maintenance and Cleaning Guide February 2010 Xerox 8265/8290 Color Wide Format Printer User Maintenance and Cleaning Guide 2010 Xerox Corporation. All rights reserved. Xerox and the sphere of connectivity design and all product names

More information

Repairing the Seam of the WaterRower Tank

Repairing the Seam of the WaterRower Tank Repairing the Seam of the WaterRower Tank The following instructions are for repairing a WaterRower that is leaking water around the seam, the 1 overlap between the top and bottom halves of the tank. While

More information

2 (. 3 4 5 . -,.. -,. 3. : 1 2 3 > 4, :!! 1,, 17 Lt MAX 2 3 4,, > 5.,,, 6 :, :1)! 2)! 7,, 1, :,. 2,, Bottom of steam generator 7 8 9 10 11 12 automatically and this phenomenon will continue for

More information

Mississauga Halton/Central West Regional Cancer Program Contact Us Put contact sticker here Chemotherapy Spil or Leak Handling Instructions for Home

Mississauga Halton/Central West Regional Cancer Program Contact Us Put contact sticker here Chemotherapy Spil or Leak Handling Instructions for Home Mississauga Halton/Central West Regional Cancer Program Contact Us Put contact sticker here Chemotherapy Spill or Leak Handling Instructions for Home Introduction: Notes You have been given a Chemo Spill

More information

ClikSTAR - Important facts about your new insulin delivery device.

ClikSTAR - Important facts about your new insulin delivery device. ClikSTAR - Important facts about your new insulin delivery device. Instruction for Use ClikSTAR Insulin delivery device Before you start: Read these instructions and follow them completely each time you

More information

Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen Series: 220000-450000

Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen Series: 220000-450000 Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen Series: 220000-450000 10 mm open end wrench 12 mm open end wrench 9/16 open end wrench 5/8 open end wrench 11/16 open end wrench 9/16

More information

Installing a new 928 Motorsports All-Aluminum Radiator

Installing a new 928 Motorsports All-Aluminum Radiator 928 Motorsports Supercharger Installation Copyright 2008, 928 Motorsports, LLC All Rights Reserved Installing a new 928 Motorsports All-Aluminum Radiator Copyright 2008 928 Motorsports, LLC. Toll-Free

More information

Scotchgard Multi-Layer Protective Film 1004

Scotchgard Multi-Layer Protective Film 1004 Scotchgard Multi-Layer Protective Film 1004 3 Technical Bulletin September, 2008 Scotchgard Multi-Layer Protective Film 1004 Application Kit (3M Part No. 70-0710-4666-1) includes: (clockwise) Heavy Duty

More information

- Do not immerse cords, pump or any electrical wire or device in water or other liquid.

- Do not immerse cords, pump or any electrical wire or device in water or other liquid. TRANSFER PUMP INSTRUCTIONS FOR THE SUPER TRANSFER PUMP 1. Pump (water may be present in the pump and lines due to testing at factory) 2. Instructions. 3. Inlet thicker wall hose (1), outlet hose (1) 4.

More information

Coal fired steam locomotive CHARYBDIS

Coal fired steam locomotive CHARYBDIS Coal fired steam locomotive CHARYBDIS Owners Manual Contents Locomotive running tools Controls Running in Preparing to run your locomotive Running your locomotive Once you have finished your running session

More information

Cell Sorting. S3e Cell Sorter Quick Guide

Cell Sorting. S3e Cell Sorter Quick Guide Cell Sorting S3e Cell Sorter Quick Guide STARTUP Checking Fluidics STARTUP There are three bulk fluid containers (Figure 1) that may require attention before starting the system. If fluidics need to be

More information

Installation and Operation Manual

Installation and Operation Manual Model WHED20 How to install, operate and maintain your Undersink Drinking Water Filter System Do not return unit to store If you have any questions or concerns when installing, operating or maintaining

More information

PARTS NEEDED FOR CONVERTING FROM LP TO NG

PARTS NEEDED FOR CONVERTING FROM LP TO NG 1 6 7 8 3 5 4 10 9 2 PARTS NEEDED FOR CONVERTING FROM LP TO NG Number Description 1 30" NG Manifold 2 NG side burner valve 3 NG Regulator 4 Rear I/R NG orifice 4 Rear I/R NG orifice 5 Side burner NG orifice

More information

HUMULIN 70/30 KwikPen

HUMULIN 70/30 KwikPen 1 Instructions for Use HUMULIN 70/30 KwikPen (70% human insulin isophane suspension 30% human insulin injection [rdna origin]) Read the Instructions for Use before you start taking HUMULIN 70/30 and each

More information

USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102

USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE

More information

1-800-862-2731 Administration of Meropenem For Child

1-800-862-2731 Administration of Meropenem For Child 1 Drug Name: _Meropenem 1-800-862-2731 Administration of Meropenem For Child Dosage: milligrams every hours over 5 minutes Key Points: FLUSHING: Meropenem Heparin flush 1. Always wash your hands with an

More information

Instruction Manual. Cable Splicing. Models 4500-9, 4500-9-HDF1, 4500-9-HDF2 4500-9-SS1

Instruction Manual. Cable Splicing. Models 4500-9, 4500-9-HDF1, 4500-9-HDF2 4500-9-SS1 Instruction Manual Cable Splicing Models 4500-9, 4500-9-HDF1, 4500-9-HDF2 4500-9-SS1 No part of this instruction manual may be reproduced, by any means, without the written consent of Geokon, Inc. The

More information

How does your heart pump blood in one direction?

How does your heart pump blood in one direction? Have a Heart How does your heart pump blood in one direction? Description How does your heart move blood in one direction, around in a loop? In this activity, you will make a model of one of the heart

More information

Overview PARTS LIST. B. Lever mounting base C. Flush handle assembly D. Grey/Blue float stop E. Grey float (Full Flush) F. Flush valve washer

Overview PARTS LIST. B. Lever mounting base C. Flush handle assembly D. Grey/Blue float stop E. Grey float (Full Flush) F. Flush valve washer Overview READ ENTIRE INSTRUCTIONS BEFORE STARTING INSTALLATION PARTS LIST A. Flush valve B. Lever mounting base C. Flush handle assembly D. Grey/Blue float stop E. Grey float (Full Flush) F. Flush valve

More information

Instructions for Use

Instructions for Use 1 Instructions for Use HUMULIN R U-500 KwikPen insulin human injection U-500 (500 units/ml, 3 ml pen) Important: Know your dose of HUMULIN R U-500 insulin. The Pen delivers your dose in insulin units.

More information

Aseptic Technique. A GMP/GTP Training Module

Aseptic Technique. A GMP/GTP Training Module Aseptic Technique A GMP/GTP Training Module Aseptic Technique The GMP Facility manufactures products for clinical use These products must meet a number of requirements, one of which is that they are sterile

More information

ÄKTA system Training Guide

ÄKTA system Training Guide ÄKTA system Training Guide Explorer 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bioscience ÄKTAexplorer ÄKTAexplorer Buffers Injector Box amersham bioscience Monitor Conductivity

More information

INGROUND, ROPED, AND HOLELESS JACK SEAL REPLACEMENT

INGROUND, ROPED, AND HOLELESS JACK SEAL REPLACEMENT D. L. Martin Company INGROUND, ROPED, AND HOLELESS JACK SEAL REPLACEMENT TYPICAL HYDRAULIC JACK HEAD ASSEMBLY 1 TOOLS REQUIRED: 1) Container to hold hydraulic fluid. 2) Clean rags and protective covers

More information

SDS-PAGE Protocol Mutated from the SDS-PAGE protocol written by the Lord of the Flies

SDS-PAGE Protocol Mutated from the SDS-PAGE protocol written by the Lord of the Flies SDS-PAGE Protocol Mutated from the SDS-PAGE protocol written by the Lord of the Flies Pouring the resolving gel 1. Clean glass plates with soap and water, then with ethanol. Assemble the glass plates and

More information

ESS Method 130.1: General Auto Analyzer Procedures

ESS Method 130.1: General Auto Analyzer Procedures ESS Method 130.1: General Auto Analyzer Procedures Environmental Sciences Section Inorganic Chemistry Unit Wisconsin State Lab of Hygiene 465 Henry Mall Madison, WI 53706 Revised October 1992 ESS Method

More information

Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen. Series: 150000 200605 LP Gas

Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen. Series: 150000 200605 LP Gas Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen Series: 150000 200605 LP Gas Tools needed Bottle brush Non-abrasive scouring pad Small utility brush Heavy-duty pipe cleaners Air hose

More information

Ozone Generator Manual

Ozone Generator Manual ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 Ozone Generator Manual 80003-118 MCN-282, 05/14/04 Table of Contents 1.0 INTRODUCTION... 1 1.1 Unpack...1 1.2 Check Order...1 1.3 Serial Numbers...1

More information

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance Bench Autoclave Standard Operating Procedure For Installation, Use and Maintenance 1. Introduction This SOP is intended for use with the following model, in a laboratory context: Type: Nuve Bench Top Steam

More information

3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE

3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE 4 3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE A. Changing the Control Board B. Replacing a Heating Element C. Thermistor Replacement D. High Limit Switch Replacement E. Level Detector Replacement

More information

NAIAD INFLATABLES (NZ) LTD

NAIAD INFLATABLES (NZ) LTD Inner Tube Removal and Fitting Original System The buoyancy tubes add significantly to the stability of the craft. It is the owners responsibility to ensure that the various components that make up the

More information

Troubleshooting Guide for Dispenser Model #900172

Troubleshooting Guide for Dispenser Model #900172 Troubleshooting Guide for Dispenser Model #900172 Q. The cold water is room temperature. A. There is a cold water thermostat control mounted to the back of the majority of our appliances. Simply unplug

More information

PAPER CHROMATOGRAPHY

PAPER CHROMATOGRAPHY PAPER CHROMATOGRAPHY INTRODUCTION Chromatography is a technique that is used to separate and to identify components of a mixture. This analytical technique has a wide range of applications in the real

More information

Federal Wage System Job Grading Standard For Laboratory Working, 3511. Table of Contents

Federal Wage System Job Grading Standard For Laboratory Working, 3511. Table of Contents Federal Wage System Job Grading Standard For Laboratory Working, 3511 Table of Contents WORK COVERED... 2 WORK NOT COVERED...2 TITLES... 2 GRADE LEVELS... 2 NOTE TO USERS... 3 LABORATORY WORKER, GRADE

More information

Partner: Jack 17 November 2011. Determination of the Molar Mass of Volatile Liquids

Partner: Jack 17 November 2011. Determination of the Molar Mass of Volatile Liquids Partner: Jack 17 November 2011 Determination of the Molar Mass of Volatile Liquids Purpose: The purpose of this experiment is to determine the molar mass of three volatile liquids. The liquid is vaporized

More information

Cleaning & Sanitisation

Cleaning & Sanitisation Cleaning & Sanitisation Notice: The information and/or procedures presented in the following demonstration(s) should be performed by a trained Water Cooler Service Technician only. Never attempt to service

More information

www.servicechamp.com

www.servicechamp.com 1-800-221-0216 Fax: 1-800-472-2281 www.servicechamp.com Service Champ Part 52081 Service Interval every 30,000 miles / Chevrolet 1992-2002 6.5 liter Description and Operation The fuel filter element separates

More information

Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin)

Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin) 1 Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin) Read the Instructions for Use before you start taking HUMALOG and each time you get a refill. There may be new information.

More information

Fuel Injection Pump, Rotary (005-014)

Fuel Injection Pump, Rotary (005-014) Fuel Injection Pump, Rotary View Related Topic Page 1 of 30 Fuel Injection Pump, Rotary (005-014) Table of Contents Summary General Information Preparatory Steps Remove Front Gear Train Rear Gear Train

More information

After the reactor has been running and the air in the chamber has been purged the co2 can be introduced and the dial in process can begin.

After the reactor has been running and the air in the chamber has been purged the co2 can be introduced and the dial in process can begin. About Calcium Reactors: A calcium reactor is a device that dissolves calcium carbonate by using co2 gas to create an acidic condition within the reaction chamber. A circulation pump is used to move the

More information

60" TulleTM. Instruction Manual. A Kichler Select ceiling fan

60 TulleTM. Instruction Manual. A Kichler Select ceiling fan 60" TulleTM A Kichler Select ceiling fan Kichler Lighting 7711 East Pleasant Valley Road P.O. Box 318010 Cleveland, Ohio 44131-8010 Customer Service 866.558.5706 8:30 AM to 5:00 PM EST, Monday - Friday

More information

FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE. Battery operation instructions:

FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE. Battery operation instructions: FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE Battery operation instructions: 1. The removing and inserting of batteries is to be conducted by the adults only. Unscrew the screw holding the battery pack s cover

More information

Chill 4 Hole Bath Shower Mixer

Chill 4 Hole Bath Shower Mixer Chill 4 Hole Bath Shower Mixer Fitting Instructions & Contents List Please keep these instructions for future reference and request of replacement parts Contents Page 1. Introduction 02 2. Safety Note

More information

SC, FC and ST * Hot Melt Fiber Optic Connectors

SC, FC and ST * Hot Melt Fiber Optic Connectors SC, FC and ST * Hot Melt Fiber Optic Connectors Instructions September 1998 78-8073-7873-8 2 Contents 1.0 Safety and Helpful Hints...4 2.0 Field Termination Kit Contents...4 3.0 STÑInstallation on 900

More information

rainbowsystem.com Rexair LLC Troy, Michigan 2011 Rexair LLC R14388A-0 Printed in U.S.A.

rainbowsystem.com Rexair LLC Troy, Michigan 2011 Rexair LLC R14388A-0 Printed in U.S.A. rainbowsystem.com Rexair LLC Troy, Michigan 2011 Rexair LLC R14388A-0 Printed in U.S.A. customer care number As part of Rexair s Customer Care program, each AquaMate receives a Customer Care serial number

More information

X-Plain Foley Catheter Male Reference Summary

X-Plain Foley Catheter Male Reference Summary X-Plain Foley Catheter Male Reference Summary Introduction A Foley catheter is a tube that is put through the urinary opening and into your bladder to drain urine. Your doctor may have placed or may ask

More information

Watts Pure Water Systems

Watts Pure Water Systems Watts Pure Water Systems WM-120-PT Installation and Operation Manual Premier Water Systems Do not use with water that is microbiologically unsafe or of unknown quality without adequate disinfection before

More information

Instructions for Use PROCRIT (PRO KRIT) (epoetin alfa)

Instructions for Use PROCRIT (PRO KRIT) (epoetin alfa) Instructions for Use PROCRIT (PROKRIT) (epoetin alfa) Use these Instructions for Use if you or your caregiver has been trained to give PROCRIT injections at home. Do not give yourself the injection unless

More information

Instructions and User Guide. Select Series: Vascular Access Ultrasound Training Models BPO100, BPBV110, BPP120, BPNB150. June 8, 2011 Rev 2.

Instructions and User Guide. Select Series: Vascular Access Ultrasound Training Models BPO100, BPBV110, BPP120, BPNB150. June 8, 2011 Rev 2. Instructions and User Guide Select Series: Vascular Access Ultrasound Training Models BPO100, BPBV110, BPP120, BPNB150 June 8, 2011 Rev 2.1 Contents 3 Chapter 1: Overview 3 Giving you the confidence only

More information

Back Pack Sprayer. Operator's Manual MODELS MS - 40 MS - 50

Back Pack Sprayer. Operator's Manual MODELS MS - 40 MS - 50 Back Pack Sprayer Operator's Manual MODELS MS - 40 MS - 50 WARNING DANGER Read rules for safe operation and all instructions carefully. ECHO provides this Operator's Manual which must be read and understood

More information

CALIBRATION FOR LAL20X & LAL24X

CALIBRATION FOR LAL20X & LAL24X CALIBRATION AND FAULT FINDING FOR LAL20X & LAL24X DUMPY LEVELS MEASURING EXPERTS SINCE 1869 How The LAL20X & LAL24X Works The Automatic level is called Automatic because it requires only a simple basic

More information

STATFlow. Direct-to-Drain System. For use with STATIM 2000 and STATIM 5000 Autoclaves. Installation Guide and Operator s Manual

STATFlow. Direct-to-Drain System. For use with STATIM 2000 and STATIM 5000 Autoclaves. Installation Guide and Operator s Manual STATFlow Direct-to-Drain System For use with STATIM 2000 and STATIM 5000 Autoclaves Installation Guide and Operator s Manual STATFlow Installation Guide and Operator s Manual 95-113193 Rev 2.0. Copyright

More information

Installation Guide. Wyse Z Class 2 and 4-GB RAM Upgrade Option Kit. Products: Z90DW, Z90SW, Z90D7, Z90S7. Issue: 122211 PN: 883884-32L Rev.

Installation Guide. Wyse Z Class 2 and 4-GB RAM Upgrade Option Kit. Products: Z90DW, Z90SW, Z90D7, Z90S7. Issue: 122211 PN: 883884-32L Rev. Installation Guide Wyse Z Class 2 and 4-GB RAM Upgrade Option Kit Products: Z90DW, Z90SW, Z90D7, Z90S7 Issue: 122211 PN: 883884-32L Rev. A Copyright Notices 2011, Wyse Technology Inc. All rights reserved.

More information

149mm. Walk-Thru Assembly Gate. MODEL NO: 1161, 1167 Owner s Manual. www.regalo-baby.com

149mm. Walk-Thru Assembly Gate. MODEL NO: 1161, 1167 Owner s Manual. www.regalo-baby.com 149mm 210mm Walk-Thru Assembly Gate MODEL NO: 1161, 1167 Owner s Manual READ ALL INSTRUCTIONS BEFORE ASSEMBLY AND USE OF GATE. KEEP INSTRUCTIONS FOR FUTURE USE. www.regalo-baby.com... Regalo International,

More information

Hot, Warm and Cold Mineral Water Cooler [Models Cool18, Cool25, Cool36, Cool50, Cool75, Cool100]

Hot, Warm and Cold Mineral Water Cooler [Models Cool18, Cool25, Cool36, Cool50, Cool75, Cool100] Hot, Warm and Cold Mineral Water Cooler [Models Cool18, Cool25, Cool36, Cool50, Cool75, Cool100] PLEASE READ BEFORE INSTALLATION TO PREVENT DAMAGE TO THE COOLER HOT WATER TANK STEAM EXAUST VENT (CAUTION!)

More information

PREVENTATIVE MAINTENANCE

PREVENTATIVE MAINTENANCE PREVENTATIVE MAINTENANCE PREVENTIVE MAINTENANCE Sandwich Unit: Proper Operation Turn on at least 45 minutes before loading Load only pre-chilled food 34-37 F Move quickly from walk-in to sandwich unit

More information