1 High Speed Acquisition of Spectra Triggering and Synchronization of the SamBa Camera to a Scanning Laser Overview This note describes how to use the Sensovation SE-34 for high speed acquisition of Spectra. In essence one can perform hyper-spectral 2-D or 3-D imaging. The use of a CHARGE MULTIPLYING CCD (singleelectron detection) allows very high sensitivity. At the high rates needed for scanning spectroscopy, cooling of the CCD is not needed. Typical applications: Confocal Scanning Microscopy Multi-Photon Microscopy Scanning Laser Opthalmoscopy Live Cell Imaging Slide Based Cytometry Biochip Scanners, etc. One CCD Camera (SamBa SE-34) coupled with a spectrograph can replace a complex and expensive multi-photomultiplier or multiple-apd system, while giving full digitisation of the spectrum. Wavelength ranges to be measured can be software selected. Fig 1. The SamBa SE-34 Charge Multiplied CCD SINGLE ELECTRON DETECTION Fig. 2. Coupled to a Spectrograph (ISA CP-140) to measure spectra 1= info@sensovation.com

2 Typical Laser Scanning Confocal Optical System REPLACE MULTIPLE PMTs WITH THE SamBa SE-34 + Spectrograph! Fig. 3: Typical Laser Scanning Confocal Optical System: Scanning laser systems typically use Photomultipliers for detection, since very fast acquisition times and high sensitivity is required. The dwell time of the laser spot together with its scan speed determine the resolution of the imaging system. For fluorescence imaging, other factors come into play such as the tradeoff between laser power and photobleaching of dyes. As shown in the above example, multiple photomultipliers (PMT) or avalanche photodiodes (APD) are used in order to detect many wavelengths simultaneously. The disadvantages of such systems are: Only few wavelength bands can be measured. As the number of colors go up, optical efficiency & sensitivity is lost. Optical filters fix the wavelengths, and mechanical filter wheels are big, unreliable. Optical filters throw away light that is not in the band to be measured. Expensive and complex. Now there is a better solution 2= info@sensovation.com

3 Sensovation s SamBa SE-34, Charge Multiplying CCD The SamBa SE-34 was designed for high-speed spectral scanning applications to replace multiple PMTs. In order to do this, the following are absolutely mandatory features: Readout Sequence of the CCD is user programmable: You decide where you want to position the spectra on the CCD, how it is read, which wavelength bands. Triggering is precise for synchronization with a fast scanning laser. You can trigger anywhere in the readout sequence! It is necessary here to trigger on SPECTRA, which may relate to a particular sub-area of the CCD! Since this is not an imaging situation, user control of the interface to the host (PC) and the timing is necessary. For example, how do frame grabbers acquire spectra? The SamBa SE-34 allows you to control the data transmission timing. The Charge Multiplying CCD with the capability to detect single electrons is the key to achieving the sensitivity at speeds high enough for these applications. Line binning, unlimited and user programmable This increases sensitivity, and allows ultra-high speed capture of a burst of spectra on the CCD (described later). With the following advantages: Higher performance for multi-wavelength detection Confocal and multi-photon strategies can be preserved (in comparison to normal imaging), Background scatter and crosstalk minimized. Software programmable wavelength bands. Digitization of the wavelength axis better discrimination of dyes Solid State reliability, calibrated and repeatable measurements Compact and less complex Less expensive 3= info@sensovation.com

4 Architecture of the Charge Multiplying CCD ACTIVE AREA Spectrum Rows are summed here! STORAGE AREA On-chip Charge Multiplier Detection of SINGLE ELECTRONS! Fig. 4 Architecture of the Charge Multiplying CCD The architecture of the Texas Instruments TC253SPD Charge Multiplying CCD is shown above. First of all we notice that it is a frame transfer type of CCD, which has an active area (where the pixels are exposed), and a shaded storage area. This frame transfer feature is useful in capturing spectra fast, and guaranteeing that no further light will affect the captured spectra. Secondly, the serial or readout register at the bottom of the storage area is extended such that it has a region 400 multiplication pixels. It is in this region that the electrons are multiplied by a gain of up to x100 as a line is read. The idea is that since the read noise of the output of the CCD is about 20 electrons rms, we need to multiply a single electron by a gain of at least 60 (3 times the read noise) to get a minimum SNR of 3. Then we can say that we can measure single electrons. The high clocking speeds of the TI CCD (12MHz readout, 3MHz line shift) allow us to use it in laser scanning spectroscopy. 4= info@sensovation.com

5 Schemes for the Acquisition of Spectra Case 1: Projection of a single spectrum onto the CCD, with continuous or burstmode acquisition in time. The most likely scheme for acquiring a single spectrum is to position the image of the spectrum in the active area, just above the storage area as shown in the figure above. This situation is shown in Fig. 4 above. The SamBa AS-34 is programmed such that upon receiving an external hardware trigger, one spectra is captured i.e., shifted fast down into the storage area. In this process, the lines containing the spectra are summed ( binned ) into one line. This process can continue continuously as the laser scans, thereby building up a hyper-spectral image of the focal plane. Continuous reading of spectra: Burst rate of capturing up to 500 spectra: 17,000 spectra/s. 150,000 spectra/s. Background Subtraction, Fixed-pattern Correction: Unwanted offset signal comprising of dark response (fixed-pattern variances, dark current) is eliminated by acquiring spectra in the dark, and then subtracting this fixed component from each subsequently acquired spectrum. To do this, the exposure of the CCD is pulsed off, and a dark spectrum stored. Such calibration spectra can be updated at any time, and at the limit each spectra can be acquired paired with a dark spectrum. Such in-line correction of data (data processing) can be implemented in the SamBa on a custom basis. Relative Measurement: In some applications, a change in intensity and/or spectral properties with time is measured. An examples is the detection of kinetic molecular binding via fluorescence intensity or wavelength shift. In these cases, the capture of spectra is timed and synchronized to match the rate of the kinetic event. For example, pairs of spectra with a specific time interval can be captured. Or a burst of spectra can be captured, synchronized to the start of an event. Rejection of background scatter and fluorescence: In other applications, the unwanted varying background caused by the illumination or excitation source must be eliminated. This cannot be done using a fixed calibration. A common technique uses pulsed sources, together with the gating of spectral measurement during the lifetime decay of fluorescence dyes. This can be implemented using the SamBa SE-34 by synchronizing the capture of the spectrum such that the integration occurs during a precise time period after the excitation pulse. The decay of the dye itself can be measured via capture of multiple spectra after the light pulse. In effect, relative measurements can be done on-the-fly, at high speeds. 5= info@sensovation.com

6 Rejection of Ambient Light: This can be done by a simple pulsing (strobing or chopping) the light source such that a higher intensity is concentrated in the pulses, and synchronizing the capture of spectra to the pulse frequency the same concept as a lock-in amplifier. The captured Spectra correspond only to the intervals when the source in on. The longer dead (dark) periods in between are discarded. Since ambient light is at DC or have other specific frequencies such as 60Hz, it can be strongly attenuated. Case 2: Projection of a many spectra onto the CCD simultaneously In some applications, a number of spectra must be measured simultaneously often in order to increase throughput. An example of this is the spectral measurement of samples in an array, where one row at a time is projected onto the CCD. In the case of a 96-well micro-plate, either 8 spectra or 12 spectra (a row of wells) are measured simultaneously. An effective use of the 2-dimensional area of the CCD is shown below in this typical projection of the spectra on the CCD: Regions of Interest to be binned and measured Spectral axis Discrete spectra Wavelength Bands are binned here STORAGE AREA On-chip Charge Multiplier Detection of SINGLE ELECTRONS! Fig. 5 Typical projection of multiple spectra on the CCD In this acquisition scheme, all of the discrete objects (samples) in a row are excited/illuminated simultaneously. The resulting spectra are projected onto the active area of the CCD. 6= info@sensovation.com

7 The exposure and readout of the CCD is synchronized to the indexing of the rows of objects perhaps mechanical motion in one direction. The advantage of the frame transfer CCD is that the image in the active area is stored fast, and can be read simultaneously while the mechanical motion and exposure of the following row is occurring. This means that speed and light collection efficiency is increased. Furthermore, the user can programmatically select the wavelength bands to be measured, as well as the precise location of the samples. These settings define Regions of Interests to be measured. Through onchip binning of these areas, sensitivity can be increased while greatly reducing the amount of data and increasing readout speed. Also by varying binning levels, the dynamic range of measurement can be adjusted to the signal intensity. The SamBa allows this flexibility. Case 3: Projection of the spectrum of a continuous line onto the CCD (hyper-spectral line scanning) If one wanted to hyper-spectrally scan an area with a high resolution in the direction of the scan width (in the spatial axis), then a continuous line on the object would be excited/illuminated. The spectra of the emission from this line would be projected onto the CCD as shown below: Regions of interest Spectral Axis Wavelength bands of interest are binned here STORAGE AREA On-chip Charge Multiplier Detection of SINGLE ELECTRONS! Fig. 6: Typical projection of the spectrum of a line onto the CCD The acquisition scheme is similar to case 2 above, where the surface to be measured is scanned in one direction. The lines are read at a high pixel resolution. Regions of interest can be defined, together with on-chip binning of the wavelength bands to be measured. 7= info@sensovation.com

8 Programming the SamBa SE-34 to acquire spectra The key is the Readout Sequence, which can be programmed by the user. This Readout Sequence is a text list comprised of steps, where each step performs shifting of pixels, shifting of lines, etc. Therefore the user can define how the pixels are read, down to individual pixels. There is no limitation to binning, clearing (skipping) pixels or lines, etc., up to the limits of what the CCD architecture will allow. Other steps in the sequence may perform I/O the most useful here are the generation of framegrabber timing signals such as Frame- and Line Enable Signals. The other important step is the wait for trigger. After you decide which Readout Sequence suits your detection scheme, you then send the sequence to the SamBa. With a START command, the sequence is executed in blazing hardware speed and precision, and the spectral data is yours. The SamBa SE-34 can send data to your host computer either via a standard digital framegrabber, or over Ethernet. Scheme 1. Capture a "burst" of spectra and then transfer them to a framegrabber board: In this case, the image of a single spectrum is positioned on the top N rows of the CCD chip, to maximize the number of spectra that can be captured. It is assumed that only the top N rows are exposed (perhaps the rest of the CCD is mechanically shaded, or optically not exposed). For every new sample location (laser scanner position), we "acquire" a spectrum on the CCD by vertically shifting the image say N rows in the direction of the storage area. Upon each trigger, this is repeated until the CCD is "full of spectra". Since the frame transfer CCD has a total vertical resolution of 1000 lines (counting both the storage plus the active areas), you can collect a burst of 1000/N spectra before having to read them. Naturally the better you can focus the spectrum into less lines, the more of them you can collect in one burst. The burst of spectra are then read form the CCD chip, and transferred to the host as one "image". 8= info@sensovation.com

9 The SamBa SE-34 readout sequence would look like this: READOUT SEQUENCE, BURST OF 50 SPECTRA, NO BINNING SEQ:BEG SEQ:STEP CLRPULSE Clear entire CCD. SEQ:LOOP BEG 1000/N Loop to capture 1000/N spectra, one per trigger. SEQ:STEP TRIG Wait for rising edge trigger. SEQ:STEP INTTIMER Integrate. SEQ:STEP VSTORE,20 Capture a spectrum by shifting the image N rows toward the storage area. SEQ:LOOP BEG 3 Loop to send 3 dummy lines to framegrabber SEQ:STEP SETLEN,1 This is simply a framegrabber timing requirement. SEQ:STEP HREAD,SIZE_X SEQ:STEP SETLEN,0 SEQ:STEP SETFEN Readout the CCD. SEQ:LOOP BEG 1000 Loop to read 1000 lines SEQ:STEP VSERIAL,1 SEQ:STEP HCLR,710 + START_X Charge Multiply & skip to start of line SEQ:STEP SETLEN,1 Read a line SEQ:STEP HREAD,SIZE_X A programmable line length is read. SEQ:STEP SETLEN,0 SEQ:STEP SETFEN,0 SEQ:STEP REPEAT,1 SEQ:END Assuming that N=20 lines, since the vertical shift speed is 3MHz, with this sequence you can capture 150,000 spectra/s max. burst rate (not including additional desired integration time). Each frame contains 1000 rows, where there are 20 rows/spectrum. An improvement of this is if you perform vertical binning, in which the spectrum is summed into one line on the CCD chip. This allows you to store more spectra on the chip, as well as increase sensitivity at the same time. This line-binning occurs in the first row of the storage area, located between the active and storage areas. Note however that if the intensity is too bright, this summed line will be the first to bloom. it is better to position the image of the spectrum at the bottom N rows of the CCD chip, near the storage area. 9= info@sensovation.com

10 The SamBa SE-34 Readout Sequence would look like this: READOUT SEQUENCE, BURST OF 500 SPECTRA, BINNING SEQ:BEG SEQ:STEP CLRPULSE Clear entire CCD. SEQ:LOOP BEG 500 Loop to capture 500 spectra, one per trigger. SEQ:STEP TRIG Wait for rising edge trigger. SEQ:STEP INTTIMER Integrate. SEQ:STEP VSUM,N Capture a spectrum by binning N rows into one row of the storage area. SEQ:STEP VSERIAL,1 Shift the binned row once toward the serial register. SEQ:LOOP BEG 3 Loop to send 3 dummy lines to framegrabber SEQ:STEP SETLEN,1 This is simply a framegrabber timing requirement. SEQ:STEP HREAD,SIZE_X SEQ:STEP SETLEN,0 SEQ:STEP SETFEN Readout the CCD, all 500 rows of the storage area. SEQ:LOOP BEG 500 SEQ:STEP VSERIAL,1 SEQ:STEP HCLR,710 + START_X Charge Multiply & skip to start of line SEQ:STEP SETLEN,1 Read line SEQ:STEP HREAD,SIZE_X Desired line length is read SEQ:STEP SETLEN,0 SEQ:STEP SETFEN,0 SEQ:STEP REPEAT,1 SEQ:END Scheme 2: Continuous acquisition of spectra. In this case, the image of the spectrum is positioned at the bottom N rows of the CCD chip i.e., closest to the storage area. Only these N rows are exposed. For every new sample location (laser scanner position), we "acquire" a spectrum on the CCD by vertically shifting the image N rows in the direction of the storage area. Here again we will bin the N rows into one such that the time to read one spectrum is minimized, and sensitivity is maximized. Also, if we only read a portion of the line (the first n pixels), speed will be increased. Upon each trigger, one spectrum is read. 10= info@sensovation.com

11 Here the focusing of the spectrum into a smaller number of lines is not as critical to speed, since this is limited by the time to read a line. Since you can control the Frame Enable and Line Enable in the Readout Sequence, you can either transmit the spectra to the IMAQ board as a continuous stream of lines (like a line-scan camera) or formatted as "images", where an image comprises of many spectra. Remembering that in imaging mode many framegrabbers (including the IMAQ-1422) require a number of dummy lines before the frame, in order to synchronize. This will of course cause periodic loss of spectra, which may not be acceptable. It is therefore better if the framegrabber can operate in a streaming line scanning mode, in which there is no concept of frames. Such a sequence is shown below, where the Frame Enable signal is not used at all. It is to be verified that the framegrabber can receive this. The SamBa SE-34 readout sequence would look like this: READOUT SEQUENCE, CONTINUOUS SPECTRA, BINNING SEQ:BEG SEQ:STEP TRIG Wait for rising edge trigger. SEQ:STEP INTTIMER Integrate. SEQ:STEP VSUM,N Capture a spectrum by binning N rows into one first row of the storage area. SEQ:STEP VSERIAL,1 Shift the binned row once toward the serial register. SEQ:STEP HCLR,24 + START_X Charge Multiply & skip to start of line SEQ:STEP SETLEN,1 SEQ:STEP HREAD,SIZE_X Read the binned spectrum (one line) SEQ:STEP SETLEN,0 SEQ:STEP REPEAT,1 SEQ:END With this Sequence you can sample spectra continuously (656 pixels/spectrum) at 17,000 spectra/s max. (with no additional desired integration time). Note that since this is a continuous mode, the CCD is not initially cleared in the readout sequence. You may want to use a "clearing sequence" prior to loading this sequence. Nevertheless, the first few spectra will likely not be valid, until all of the CCD lines have been read once. 11= info@sensovation.com

12 Triggering When using asynchronous external triggering, there are two ways that the integration time can be defined: a) Programmed integration time. As shown in the above sequences, a desired integration time is set via the SCPI command: ACQ:TINT <time in ms, with range of to 65535> Then within the Readout Sequence, this step is used immediately after the trigger: SEQ:STEP TRIG SEQ:STEP This sequence would typically continue after the integration step to store the spectrum (capture it on the CCD) is, and eventually read it. b) One can use two trigger pulses to define an integration time. Note that while waiting for trigger, the CCD is integrating: SEQ:STEP TRIG SEQ:STEP TRIG Or: SEQ:STEP TRIG SEQ:STEP CLRPULSE (clears the ENTIRE CCD in 5) SEQ:STEP TRIG Trigger Timing: Note that the external trigger input is rising edge triggered, and requires a 1 us high pulse width minimum. This input is the LED of an opto-coupler, and the external driver must be able to drive 10mA. There is a 10us max delay from rising edge of the trigger signal to clocking of the CCD. This is mainly due to the opto-coupler and debouncing. 12= info@sensovation.com

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn

More information

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Some limitations (disadvantages) of NSOM A trade-off

More information

ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution

ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution Product Information Version 1.0 ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution ZEISS Axiocam 506 color Sensor Model Sensor Pixel

More information

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology product overview family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology scmos knowledge base scmos General Information PCO scmos cameras are a breakthrough

More information

How To Use An Edge 3.1 Scientific Cmmos Camera

How To Use An Edge 3.1 Scientific Cmmos Camera edge 3.1 scientific CMOS camera high resolution 2048 x 1536 pixel low noise 1.1 electrons global shutter USB 3.0 small form factor high dynamic range 27 000:1 high speed 50 fps high quantum efficiency

More information

AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz

AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz AN1200.04 Application Note: FCC Regulations for ISM Band Devices: Copyright Semtech 2006 1 of 15 www.semtech.com 1 Table of Contents 1 Table of Contents...2 1.1 Index of Figures...2 1.2 Index of Tables...2

More information

Computer Vision. Image acquisition. 25 August 2014. Copyright 2001 2014 by NHL Hogeschool and Van de Loosdrecht Machine Vision BV All rights reserved

Computer Vision. Image acquisition. 25 August 2014. Copyright 2001 2014 by NHL Hogeschool and Van de Loosdrecht Machine Vision BV All rights reserved Computer Vision Image acquisition 25 August 2014 Copyright 2001 2014 by NHL Hogeschool and Van de Loosdrecht Machine Vision BV All rights reserved j.van.de.loosdrecht@nhl.nl, jaap@vdlmv.nl Image acquisition

More information

SSO Transmission Grating Spectrograph (TGS) User s Guide

SSO Transmission Grating Spectrograph (TGS) User s Guide SSO Transmission Grating Spectrograph (TGS) User s Guide The Rigel TGS User s Guide available online explains how a transmission grating spectrograph (TGS) works and how efficient they are. Please refer

More information

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps 37 500:1 > 70 % pco. low noise high resolution high speed high dynamic range

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps 37 500:1 > 70 % pco. low noise high resolution high speed high dynamic range edge 4.2 LT scientific CMOS camera high resolution 2048 x 2048 pixel low noise 0.8 electrons USB 3.0 small form factor high dynamic range 37 500:1 high speed 40 fps high quantum efficiency > 70 % edge

More information

Understanding Line Scan Camera Applications

Understanding Line Scan Camera Applications Understanding Line Scan Camera Applications Discover the benefits of line scan cameras, including perfect, high resolution images, and the ability to image large objects. A line scan camera has a single

More information

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy Microscopy from Carl Zeiss AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy New Dimensions in Performance AxioCam MR from Carl Zeiss Both

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

High Definition Imaging

High Definition Imaging High Definition Imaging Scientific CMOS Camera Photon Technology International www.pti-nj.com Scientific CMOS Camera The new HDI camera is a breakthrough in scientific imaging cameras, due to its distinctive

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

S2000 Spectrometer Data Sheet

S2000 Spectrometer Data Sheet Description The Ocean Optics OEM S2000 Spectrometer includes the linear CCD-array optical bench, plus the circuits necessary for spectrometer operation. The result is a compact, flexible system with no

More information

Webex-based remote instrument control guide

Webex-based remote instrument control guide Webex-based remote instrument control guide 1) Once a microscopic imaging experiment is scheduled, the host (i.e., Hunter Bioimaging Facility) will send an email containing a Webex meeting link to the

More information

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL Robert Bagnell 2006 This tutorial covers the following CLSM topics: 1) What is the optical principal behind CLSM? 2) What is the spatial resolution in X, Y,

More information

COOKBOOK. for. Aristarchos Transient Spectrometer (ATS)

COOKBOOK. for. Aristarchos Transient Spectrometer (ATS) NATIONAL OBSERVATORY OF ATHENS Institute for Astronomy, Astrophysics, Space Applications and Remote Sensing HELMOS OBSERVATORY COOKBOOK for Aristarchos Transient Spectrometer (ATS) P. Boumis, J. Meaburn,

More information

Raman Spectroscopy Basics

Raman Spectroscopy Basics Raman Spectroscopy Basics Introduction Raman spectroscopy is a spectroscopic technique based on inelastic scattering of monochromatic light, usually from a laser source. Inelastic scattering means that

More information

High speed infrared camera for contactless temperature measurement on rotating tires

High speed infrared camera for contactless temperature measurement on rotating tires DIAS Infrared GmbH Publications No. 9 1 High speed infrared camera for contactless temperature measurement on rotating tires Helmut Budzier, Dresden University of Technology, Institute for Solid-State

More information

APPLICATION NOTE. Basler racer Migration Guide. Mechanics. www.baslerweb.com. Flexible Mount Concept. Housing

APPLICATION NOTE. Basler racer Migration Guide. Mechanics. www.baslerweb.com. Flexible Mount Concept. Housing 62 62 APPLICATION NOTE www.baslerweb.com Basler racer Migration Guide This paper describes what to consider when replacing the Basler L100 Camera Link or the Basler runner Gigabit Ethernet (GigE) line

More information

AxioCam HR The Camera that Challenges your Microscope

AxioCam HR The Camera that Challenges your Microscope Microscopy from Carl Zeiss AxioCam HR The Camera that Challenges your Microscope Documentation at the edge of the visible The Camera for Maximum Success: AxioCam HR Low light fluorescence, live cell imaging,

More information

Note monitors controlled by analog signals CRT monitors are controlled by analog voltage. i. e. the level of analog signal delivered through the

Note monitors controlled by analog signals CRT monitors are controlled by analog voltage. i. e. the level of analog signal delivered through the DVI Interface The outline: The reasons for digital interface of a monitor the transfer from VGA to DVI. DVI v. analog interface. The principles of LCD control through DVI interface. The link between DVI

More information

Applications: X-ray Microtomography, Streak Tube and CRT Readout, Industrial & Medical Imaging X-RAY GROUP

Applications: X-ray Microtomography, Streak Tube and CRT Readout, Industrial & Medical Imaging X-RAY GROUP Now Powered by LightField FEATURES BENEFITS Back Illuminated CCD (248B) For highest sensitivity Front illuminated CCD (248F) Affordable technology for moderate light level applications Ultra low noise

More information

Trigger-to-Image Reliability (T2IR)

Trigger-to-Image Reliability (T2IR) sensors cameras frame grabbers processors software vision solutions TECHNOLOGY PRIMER Trigger-to-Image Reliability (T2IR) Learn more about: What is Trigger-to-Image Reliability (T2IR)? Understanding T2IR

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

A Laser Scanner Chip Set for Accurate Perception Systems

A Laser Scanner Chip Set for Accurate Perception Systems A Laser Scanner Chip Set for Accurate Perception Systems 313 A Laser Scanner Chip Set for Accurate Perception Systems S. Kurtti, J.-P. Jansson, J. Kostamovaara, University of Oulu Abstract This paper presents

More information

Fluorescent Array Imaging Reader

Fluorescent Array Imaging Reader Fluorescent Array Imaging Reader Multiplex Enabled For 2-Color Fluorescent Microarrays Extended Dynamic Range Automated Spot Analysis Compact and Affordable Microarray Analysis With FLAIR FLAIR Sensovation

More information

MH - Gesellschaft für Hardware/Software mbh

MH - Gesellschaft für Hardware/Software mbh E.d.a.s.VX Data acquisition on board road and track vehicles The E.d.a.s.VX System is designed for portable applications running on 12 Volts DC, and is capable of measuring at selectable rates up to 30,000,000

More information

RF Measurements Using a Modular Digitizer

RF Measurements Using a Modular Digitizer RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.

More information

Spectral Measurement Solutions for Industry and Research

Spectral Measurement Solutions for Industry and Research Spectral Measurement Solutions for Industry and Research Hamamatsu Photonics offers a comprehensive range of products for spectroscopic applications, covering the, Visible and Infrared regions for Industrial,

More information

White paper. CCD and CMOS sensor technology Technical white paper

White paper. CCD and CMOS sensor technology Technical white paper White paper CCD and CMOS sensor technology Technical white paper Table of contents 1. Introduction to image sensors 3 2. CCD technology 4 3. CMOS technology 5 4. HDTV and megapixel sensors 6 5. Main differences

More information

Introduction to CCDs and CCD Data Calibration

Introduction to CCDs and CCD Data Calibration Introduction to CCDs and CCD Data Calibration Dr. William Welsh San Diego State University CCD: charge coupled devices integrated circuit silicon chips that can record optical (and X-ray) light pixel =

More information

Characterizing Digital Cameras with the Photon Transfer Curve

Characterizing Digital Cameras with the Photon Transfer Curve Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently

More information

Optimal Vision Using Cameras for Intelligent Transportation Systems

Optimal Vision Using Cameras for Intelligent Transportation Systems WHITE PAPER www.baslerweb.com Optimal Vision Using Cameras for Intelligent Transportation Systems Intelligent Transportation Systems (ITS) require intelligent vision solutions. Modern camera technologies

More information

Recording the Instrument Response Function of a Multiphoton FLIM System

Recording the Instrument Response Function of a Multiphoton FLIM System Recording the Instrument Response Function of a Multiphoton FLIM System Abstract. FLIM data analysis in presence of SHG signals or extremely fast decay components requires the correct instrument response

More information

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ. Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ., Raleigh, NC One vital step is to choose a transfer lens matched to your

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012

Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012 Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012 You are sitting at your microscope working at high magnification trying to sort out the three-dimensional compartmentalization

More information

Leica TCS SP5 Confocal Laser Scanning Microscope User Guide 1. BASIC IMAGE ACQUISITION

Leica TCS SP5 Confocal Laser Scanning Microscope User Guide 1. BASIC IMAGE ACQUISITION Leica TCS SP5 Confocal Laser Scanning Microscope User Guide 1. BASIC IMAGE ACQUISITION This manual is the FIRST section of a THREE part Leica TCS - SP5 User Guide edited by Donald Pottle Leica True Confocal

More information

Spike-Based Sensing and Processing: What are spikes good for? John G. Harris Electrical and Computer Engineering Dept

Spike-Based Sensing and Processing: What are spikes good for? John G. Harris Electrical and Computer Engineering Dept Spike-Based Sensing and Processing: What are spikes good for? John G. Harris Electrical and Computer Engineering Dept ONR NEURO-SILICON WORKSHOP, AUG 1-2, 2006 Take Home Messages Introduce integrate-and-fire

More information

RFSPACE CLOUD-IQ #CONNECTED SOFTWARE DEFINED RADIO

RFSPACE CLOUD-IQ #CONNECTED SOFTWARE DEFINED RADIO CLOUD-IQ #CONNECTED SOFTWARE DEFINED RADIO 1 - SPECIFICATIONS Cloud-IQ INTRODUCTION The Cloud-IQ is a high performance, direct sampling software radio with an ethernet interface. It offers outstanding

More information

Making Multiple Code Reading Easy. Webinar

Making Multiple Code Reading Easy. Webinar Making Multiple Code Reading Easy Webinar Today s Agenda Introduction How DataMan Makes Multiple Code Reading Easy Multiple Code Reading Applications Product Demonstration Videos Q&A 2 Introduction Introduction

More information

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules Abstract J.L. Crozier, E.E. van Dyk, F.J. Vorster Nelson Mandela Metropolitan University Electroluminescence (EL) is a useful

More information

Advances in scmos Camera Technology Benefit Bio Research

Advances in scmos Camera Technology Benefit Bio Research Advances in scmos Camera Technology Benefit Bio Research scmos camera technology is gaining in popularity - Why? In recent years, cell biology has emphasized live cell dynamics, mechanisms and electrochemical

More information

Single Photon Counting Module COUNT -Series

Single Photon Counting Module COUNT -Series Description Laser Components COUNT series of s has been developed to offer a unique combination of high photon detection efficiency, wide dynamic range and ease of use for photon counting applications.

More information

Make Life Easier by Using Modern Features of the SPCM Software

Make Life Easier by Using Modern Features of the SPCM Software Make Life Easier by Using Modern Features of the SPCM Software Abstract. Over a long period of time the SPCM operating software of the Becker & Hickl TCSPC systems has continuously been upgraded with new

More information

Microsecond Decay FLIM: Combined Fluorescence and Phosphorescence Lifetime Imaging

Microsecond Decay FLIM: Combined Fluorescence and Phosphorescence Lifetime Imaging Microsecond Decay FLIM: Combined Fluorescence and Phosphorescence Lifetime Imaging Abstract. We present a lifetime imaging technique that simultaneously records fluorescence and phosphorescence lifetime

More information

WHITE PAPER. Are More Pixels Better? www.basler-ipcam.com. Resolution Does it Really Matter?

WHITE PAPER. Are More Pixels Better? www.basler-ipcam.com. Resolution Does it Really Matter? WHITE PAPER www.basler-ipcam.com Are More Pixels Better? The most frequently asked question when buying a new digital security camera is, What resolution does the camera provide? The resolution is indeed

More information

Realtime FFT processing in Rohde & Schwarz receivers

Realtime FFT processing in Rohde & Schwarz receivers Realtime FFT in Rohde & Schwarz receivers Radiomonitoring & Radiolocation Application Brochure 01.00 Realtime FFT in Rohde & Schwarz receivers Introduction This application brochure describes the sophisticated

More information

CNC-STEP. "LaserProbe4500" 3D laser scanning system Instruction manual

CNC-STEP. LaserProbe4500 3D laser scanning system Instruction manual LaserProbe4500 CNC-STEP "LaserProbe4500" 3D laser scanning system Instruction manual 2 Hylewicz CNC-Technik Siemensstrasse 13-15 D-47608 Geldern Fon.: +49 (0) 2831 133236 E-Mail: info@cnc-step.com Website:

More information

Wide range of applications from Real time imaging of low light fluorescence to Ultra low light detection

Wide range of applications from Real time imaging of low light fluorescence to Ultra low light detection Electron Multiplying CCD Camera C100-13, -14 Wide range of applications from Real time imaging of low light fluorescence to Ultra low light detection Great Stability High Sensitivity Low Noise High Resolution

More information

scmos discover new ways of seeing andor.com Features and Benefits Scientific CMOS - Fast, sensitive, compact and light. Specifications Summary

scmos discover new ways of seeing andor.com Features and Benefits Scientific CMOS - Fast, sensitive, compact and light. Specifications Summary Ultra Sensitive Imaging Low Light Imaging Features and Benefits Compact and light Ideal for integration into space restrictive set-ups 1.2 e - read noise Lower detection limit than any CCD 5.5 megapixel

More information

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,

More information

Dynamic IR Scene Projector Based Upon the Digital Micromirror Device

Dynamic IR Scene Projector Based Upon the Digital Micromirror Device Dynamic IR Scene Projector Based Upon the Digital Micromirror Device D. Brett Beasley, Matt Bender, Jay Crosby, Tim Messer, and Daniel A. Saylor Optical Sciences Corporation www.opticalsciences.com P.O.

More information

Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:

Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance: Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T

More information

Basler. Line Scan Cameras

Basler. Line Scan Cameras Basler Line Scan Cameras High-quality line scan technology meets a cost-effective GigE interface Real color support in a compact housing size Shading correction compensates for difficult lighting conditions

More information

Software Manual RS232 Laser Merge Module. Document # SU-256521-09 Rev A

Software Manual RS232 Laser Merge Module. Document # SU-256521-09 Rev A Laser Merge Module Document # SU-256521-09 Rev A The information presented in this document is proprietary to Spectral Applied Research Inc. and cannot be used for any purpose other than that for which

More information

TOSHIBA CCD LINEAR IMAGE SENSOR CCD(Charge Coupled Device) TCD1304AP

TOSHIBA CCD LINEAR IMAGE SENSOR CCD(Charge Coupled Device) TCD1304AP TOSHIBA CCD LINEAR IMAGE SENSOR CCD(Charge Coupled Device) TCD1304AP TCD1304AP The TCD1304AP is a high sensitive and low dark current 3648 elements linear image sensor. The sensor can be used for POS scanner.

More information

Using the NI 17xx Smart Camera Direct Drive Lighting Controller

Using the NI 17xx Smart Camera Direct Drive Lighting Controller Using the NI 17xx Smart Camera Direct Drive Lighting Controller Overview The use of proper lighting is often overlooked when designing a machine vision application. More robust and accurate inspections

More information

The CCD-S3600-D(-UV) is a

The CCD-S3600-D(-UV) is a Advanced Digital High-Speed CCD Line Camera CCD-S3600-D(-UV) USB 2.0 Plug & Play The CCD-S3600-D(-UV) is a complete and ready to use high-speed digital CCD line scan camera. The device includes many advanced

More information

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (no PiN and pinned Diodes) Peter Fischer P. Fischer, ziti, Uni Heidelberg, Seite 1 Overview Reminder: Classical Photomultiplier

More information

IMPROVED CCD DETECTORS FOR HIGH SPEED, CHARGE EXCHANGE SPECTROSCOPY STUDIES ON THE DIII D TOKAMAK

IMPROVED CCD DETECTORS FOR HIGH SPEED, CHARGE EXCHANGE SPECTROSCOPY STUDIES ON THE DIII D TOKAMAK IMPROVED CCD DETECTORS FOR HIGH SPEED, CHARGE EXCHANGE SPECTROSCOPY STUDIES ON THE TOKAMAK by K.H. Burrell P. Gohil, R.J. Groebner, D.H. Kaplan, D.G. Nilson,* J.I. Robinson, and D.M. Thomas General Atomics,

More information

Confocal Microscopy and Atomic Force Microscopy (AFM) A very brief primer...

Confocal Microscopy and Atomic Force Microscopy (AFM) A very brief primer... Confocal Microscopy and Atomic Force Microscopy (AFM) of biofilms A very brief primer... Fundamentals of Confocal Microscopy Based on a conventional fluorescence microscope Fluorescent Microscope Confocal

More information

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?

More information

Ultrasonic Wave Propagation Review

Ultrasonic Wave Propagation Review Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic

More information

hsfc pro 12 bit ultra speed intensified imaging

hsfc pro 12 bit ultra speed intensified imaging hsfc pro 12 bit ultra speed intensified imaging four MCP-image intensifier camera modules ultra fast shutter down to 3 ns in single mode excellent sensitivity of the system allows single photon detection

More information

Acoustic GHz-Microscopy: Potential, Challenges and Applications

Acoustic GHz-Microscopy: Potential, Challenges and Applications Acoustic GHz-Microscopy: Potential, Challenges and Applications A Joint Development of PVA TePLa Analytical Systems GmbH and Fraunhofer IWM-Halle Dr. Sebastian Brand (Ph.D.) Fraunhofer CAM Fraunhofer Institute

More information

Silicon Seminar. Optolinks and Off Detector Electronics in ATLAS Pixel Detector

Silicon Seminar. Optolinks and Off Detector Electronics in ATLAS Pixel Detector Silicon Seminar Optolinks and Off Detector Electronics in ATLAS Pixel Detector Overview Requirements The architecture of the optical links for the ATLAS pixel detector ROD BOC Optoboard Requirements of

More information

Avalanche Photodiodes: A User's Guide

Avalanche Photodiodes: A User's Guide !"#$%& Abstract Avalanche Photodiodes: A User's Guide Avalanche photodiode detectors have and will continue to be used in many diverse applications such as laser range finders and photon correlation studies.

More information

Flow cytometry basics fluidics, optics, electronics...

Flow cytometry basics fluidics, optics, electronics... Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence

More information

MONOCHROME RGB YCbCr VIDEO DIGITIZER

MONOCHROME RGB YCbCr VIDEO DIGITIZER Active Silicon SNAPPER-PMC-8/24 MONOCHROME RGB YCbCr VIDEO DIGITIZER High quality analogue video acquisition board with square pixel sampling for CCIR, EIA (RS-170) standards, and nonstandard video formats.

More information

Counters and Decoders

Counters and Decoders Physics 3330 Experiment #10 Fall 1999 Purpose Counters and Decoders In this experiment, you will design and construct a 4-bit ripple-through decade counter with a decimal read-out display. Such a counter

More information

Embedded Systems Design Course Applying the mbed microcontroller

Embedded Systems Design Course Applying the mbed microcontroller Embedded Systems Design Course Applying the mbed microcontroller Serial communications with SPI These course notes are written by R.Toulson (Anglia Ruskin University) and T.Wilmshurst (University of Derby).

More information

A VERYbrief history of the confocal microscope 1950s

A VERYbrief history of the confocal microscope 1950s Confocal Microscopy Confocal Microscopy Why do we use confocal microscopy? A brief history of the confocal microscope Advantages/disadvantages of a confocal microscope Types of confocal microscopes The

More information

Cameras for Industries

Cameras for Industries Progressive Scan Industrial Camera Series Featuring GigE Vision Interface GigE Vision Cameras for Industries High speed (1Gbps) & long distance (100m) transfer Available PoE (Power over Ethernet) Power

More information

PoS(PhotoDet 2012)068

PoS(PhotoDet 2012)068 Characterization of the Hamamatsu R11265 multi-anode photomultiplier tube with single photon signals Luca CADAMURO, Marta CALVI, Andrea GIACHERO,, Matteo MAINO, Clara MATTEUZZI, Gianluigi PESSINA Dipartimento

More information

LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes

LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes LBS-300 Beam Sampler for C-mount Cameras P/N SP90183, SP90184, SP90185 and SP90186 YAG Focal Spot Analysis Adapter P/N SP90187, SP90188, SP90189, SP90190, SP90191 User Notes Ophir-Spiricon Inc. 60 West

More information

Alpy guiding User Guide. Olivier Thizy (olivier.thizy@shelyak.com) François Cochard (francois.cochard@shelyak.com)

Alpy guiding User Guide. Olivier Thizy (olivier.thizy@shelyak.com) François Cochard (francois.cochard@shelyak.com) Alpy guiding User Guide Olivier Thizy (olivier.thizy@shelyak.com) François Cochard (francois.cochard@shelyak.com) DC0017A : april 2013 Alpy guiding module User Guide Olivier Thizy (olivier.thizy@shelyak.com)

More information

COMPENSATION MIT Flow Cytometry Core Facility

COMPENSATION MIT Flow Cytometry Core Facility COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,

More information

How an electronic shutter works in a CMOS camera. First, let s review how shutters work in film cameras.

How an electronic shutter works in a CMOS camera. First, let s review how shutters work in film cameras. How an electronic shutter works in a CMOS camera I have been asked many times how an electronic shutter works in a CMOS camera and how it affects the camera s performance. Here s a description of the way

More information

Email: tjohn@mail.nplindia.ernet.in

Email: tjohn@mail.nplindia.ernet.in USE OF VIRTUAL INSTRUMENTS IN RADIO AND ATMOSPHERIC EXPERIMENTS P.N. VIJAYAKUMAR, THOMAS JOHN AND S.C. GARG RADIO AND ATMOSPHERIC SCIENCE DIVISION, NATIONAL PHYSICAL LABORATORY, NEW DELHI 110012, INDIA

More information

MANUAL FOR RX700 LR and NR

MANUAL FOR RX700 LR and NR MANUAL FOR RX700 LR and NR 2013, November 11 Revision/ updates Date, updates, and person Revision 1.2 03-12-2013, By Patrick M Affected pages, ETC ALL Content Revision/ updates... 1 Preface... 2 Technical

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

TOSHIBA CCD Image Sensor CCD (charge coupled device) TCD2955D

TOSHIBA CCD Image Sensor CCD (charge coupled device) TCD2955D Preliminary TOSHIBA CCD Image Sensor CCD (charge coupled device) TCD2955D The TCD2955D is a high sensitive and low dark current 4240 elements 6 line CCD color image sensor which includes CCD drive circuit

More information

Little LFO. Little LFO. User Manual. by Little IO Co.

Little LFO. Little LFO. User Manual. by Little IO Co. 1 Little LFO User Manual Little LFO by Little IO Co. 2 Contents Overview Oscillator Status Switch Status Light Oscillator Label Volume and Envelope Volume Envelope Attack (ATT) Decay (DEC) Sustain (SUS)

More information

Fig.1. The DAWN spacecraft

Fig.1. The DAWN spacecraft Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics

More information

Lab 1: The Digital Oscilloscope

Lab 1: The Digital Oscilloscope PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix

More information

Introduction to reflective aberration corrected holographic diffraction gratings

Introduction to reflective aberration corrected holographic diffraction gratings Introduction to reflective aberration corrected holographic diffraction gratings By Steve Slutter, Wu Jiang, and Olivier Nicolle The reflective diffraction grating is the heart of most spectroscopy systems

More information

High Speed, Real-Time Machine Vision. Perry West Automated Vision Systems, Inc.

High Speed, Real-Time Machine Vision. Perry West Automated Vision Systems, Inc. High Speed, Real-Time Machine Vision Perry West Automated Vision Systems, Inc. Table of Contents Introduction...1 Latency...3 Imaging...5 Triggering The Stimulus...5 External Sensor...5 Part Detection

More information

ECONseries Low Cost USB DAQ

ECONseries Low Cost USB DAQ ECONseries Low Cost USB Data Acquisition Modules ECONseries Low Cost USB DAQ The ECONseries is a flexible yet economical series of multifunction data acquisition modules. You choose the number of analog

More information

High-speed Photography with a Still Digital Camera

High-speed Photography with a Still Digital Camera High-speed Photography with a Still Digital Camera Matthew Moore North Carolina School of Science and Math 1219 Broad St. Durham, NC 27705 Sponsored By Dr. Loren Winters 12/2/99 Abstract High-speed photography

More information

OPERATING INSTRUCTIONS Nikon TiE Deconvolution Microscope CCRB 1-220G

OPERATING INSTRUCTIONS Nikon TiE Deconvolution Microscope CCRB 1-220G OPERATING INSTRUCTIONS Nikon TiE Deconvolution Microscope CCRB 1-220G Conventions Through these notes bold font refers to a software button and italics refer to a hardware switch. The software makes extensive

More information

STEPPER MOTOR SPEED AND POSITION CONTROL

STEPPER MOTOR SPEED AND POSITION CONTROL STEPPER MOTOR SPEED AND POSITION CONTROL Group 8: Subash Anigandla Hemanth Rachakonda Bala Subramanyam Yannam Sri Divya Krovvidi Instructor: Dr. Jens - Peter Kaps ECE 511 Microprocessors Fall Semester

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

Lecture forum InnovationPoint

Lecture forum InnovationPoint InnovationPoint Opto-Mechanical Design Software Applications and Features, an Overview... Christoph GERHARD Product Manager Business Unit Catalog LINOS Photonics GmbH & Co. KG Christoph GERHARD, Product

More information

FFT Spectrum Analyzers

FFT Spectrum Analyzers FFT Spectrum Analyzers SR760 and SR770 100 khz single-channel FFT spectrum analyzers SR760 & SR770 FFT Spectrum Analyzers DC to 100 khz bandwidth 90 db dynamic range Low-distortion source (SR770) Harmonic,

More information

MRC High Resolution. MR-compatible digital HD video camera. User manual

MRC High Resolution. MR-compatible digital HD video camera. User manual MRC High Resolution MR-compatible digital HD video camera User manual page 1 of 12 Contents 1. Intended use...2 2. System components...3 3. Video camera and lens...4 4. Interface...4 5. Installation...5

More information