1 Certified Sep-Pak CartridgeS Contents I. INTRODUCTION II. USING CERTIFIED SEP-PAK CARTIDGES III. STRATEGIES FOR SOLID-PHASE EXTRACTION IV. FACTORS AFFECTING SEPARATION EFFICIENCY V. STORAGE AND DISPOSAL OF USED CARTRIDGES VI. Ordering information VII. ADDITIONAL INFORMATION VII. LITERATURE REFERENCES I. INTRODUCTION Waters Certified Sep-Pak cartridges are convenient, reproducible, disposable solid-phase extraction (SPE) devices for sample preparation. Certified Sep-Pak cartridges are manufactured in a Waters ISO and ISO 9002-registered facility in compliance with cgmp guidelines for the U.S. Food and Drug Administration for Class 1 Medical Devices. Certified Sep-Pak cartridges are available in many styles to accommodate manual and automated solid-phase extraction and are available in 3 cc and 6 cc formats in the following chemistries: Reversed-Phase: C 18 Normal-Phase: Silica, Alumina A, Alumina B, Alumina N, Florisil Certified Sep-Pak Cartridges 1
2 II. USING CERTIFIED SEP-PAK CARTRIDGES Below is a five step guideline for using Certified Sep-Pak cartridges. Each separation is different and not all steps may be required for your application. 1. Prepare sample: The sample can be applied to the Certified Sep-Pak cartridge in either liquid or a gas phase. If the sample is a solid, it must be dissolved or extracted prior to loading. 2. Condition/equilibrate cartridge: A conditioning step is required for reversed-phase sorbents (e.g., C 18 ). Use a strong solvent to wet the stationary phase. Conditioning is followed by an equilibration step using a weak solvent. 3. Load sample: The sample is loaded onto the cartridge. 4. Wash: The wash step removes interferences while retaining the analyte. If weakly retained interferences are present in the sample matrix, wash with an appropriate weak solvent. 5. Elute: Use an appropriate stronger solvent to elute the analyte while retaining more strongly bound interferences. III. STRATEGIES FOR SOLID-PHASE EXTRACTION Purification: Use as a chemical filter to retain the component of interest while interferences elute, or to retain the interferences while the analyte passes through unretained. a. Retention-Cleanup-Elution Strategy As the sample is loaded onto the cartridge, the analytes of interest are retained by the sorbent. If needed, an optimized series of washes are used to remove matrix interference from the cartridge. A strong solvent is used to elute the analytes from the cartridge. Sample enrichment results when the final elution volume is smaller than the load volume. Stationary Phase Particles Load Sample (Black) Step Elute 1 Step Elute 2 Step Elute 3 One cartridge can separate all three dyes b. Pass-Through Cleanup Strategy NOTE: Different strength solvents can be used to separate the dyes. Pass-through cleanup methods optimize matrix retention while the analytes of interest pass-through the cartridge unretained. No sample enrichment occurs during the solid-phase extraction (SPE) step. 1. Sample is passed through sorbent and collected No sample enrichment 2. Matrix interferences are retained on sorbent Trace enrichment or concentration: Use to concentrate an analyte from a dilute solution. This is done when the analyte concentration is below the detection limit of the analytical method. Fractionation: Use with a step gradient of increasing solvent strength to selectivity elute and isolate analytes based on differences in polarity. Solvent exchange: Use to adsorb the analyte and elute with a desirable solvent. This is helpful of the sample is dissolved in a solvent that is incompatible with the analytical method. Pass-Through Certified Sep-Pak Cartridges 2
3 c. SPE Procedure Steps The following section describes the steps involved in a complete solid-phase extraction procedure: 1. Pretreatment Solid samples (soil, tissue, etc.) Shake, sonicate, or use soxhlet extraction. - Extract sample with polar organic solvent (methanol, acetonitrile) for polar analytes. - Extract sample with organic solvent and drying agent (dichloromethane, acetone) for non-polar analytes and multi-residue extraction. Non-Aqueous liquid If the sample is soluble in water, dilute it with water for reversed-phase SPE. If the sample is soluble in hexane, dilute it with (or exchange to) hexane for SPE. Wastewater Filter or centrifuge as necessary. 2. Condition For reversed-phase sorbents, preconditioning of the sorbent with an organic solvent, such as methanol, acetonitrile, isopropanol, or tetrahydrofuran, is usually necessary to obtain reproducible results. Without this step, a highly aqueous solvent cannot penetrate the hydrophobic surface and wet the sorbent. Thus, only a small fraction of the sorbent surface area would be available for interaction with the analyte. For the same reason, it is important not to let silica-based SPE cartridges dry out between the solvation step and the addition of the sample. A complete preconditioning of a reversed-phase cartridge includes the solvation step and an equilibration with a low-strength solvent such as water or buffer. 3. Load When the analytes of interest are not retained by the sorbent, this is called analyte breakthrough. For some methods, such as pass-through cleanup, analyte breakthrough is desirable and is maximized for those specific methods. However, in all other cases, analyte breakthrough is unwanted and contributes to poor recovery and method reproducibility. Breakthrough occurs when: There is too high an organic concentration in the load solution for very polar analytes. Dilute sample at least 1:1 with water or buffer prior to loading. The analytes are bound to proteins, they may pass through the sorbent. Ensure that analytes are not bound to proteins by acidifying or basifying the sample. Sorbent is overloaded by the matrix component. Therefore, it is important to choose the correct sorbent mass (see table 1). The flow rate of the load step is too fast. There is not enough contact time between the analytes and the sorbent. Look at the drops and adjust the vacuum so that you see discrete droplets, not a stream of liquid. Table 1. Choice of Cartridges Based on Sample Size Sample Size Cartridge ml 3 cc/200 mg or 6 cc/500 mg ml 3 cc/200 mg or 6 cc/500 mg ml 6 cc/500 mg (LP) or 6 cc/1 g 4. Wash The wash steps are designed to remove unwanted matrix components that remain from the loading step. The ideal wash solvent removes only the matrix while keeping the analytes bound to the sorbent. For complex samples this is impossible, so the wash steps are optimized using ph, solvent strength, and solvent polarity to remove as much matrix as possible while maintaining acceptable analyte recovery. Certified Sep-Pak Cartridges 3
4 5. Elute Once the interferences are washed off the cartridge, a strong solvent is introduced to elute the analytes of interest. The volume and flow rate of the eluting solvents should be precisely controlled as in the load step to ensure reproducible results. Table 2. Guidelines on the Various Types of Separation Mechanisms Analyte Reversed Phase Normal Phase Ion Exchange Low to moderate polarity/ hydrophobic Moderate to highly polar/uncharged Charged or ionized Matrix Aqueous Non-polar organic solvent Aqueous/low ionic strength Condition/Equilibrate 1. Solvate polar organic 2. Water Non-polar organic Low ionic strength buffer Wash Aqueous/buffer Non-polar Low ionic strength buffer Elute Increase polar organic content in steps Increase moderate to high polarity organic content in steps Stronger buffers ionic strength or ph to neutralize the charge IV. FACTORS AFFECTING SEPARATION EFFICIENCY Separation efficiency varies with flow rate and sample load. Here are some guidelines to consider when optimizing your separation. a. Flow Rate If the sample flow rate is too high, components may not interact sufficiently with he sorbent. The result is loss of resolution, analyte breakthrough, and poor recovery. To condition the cartridge, use a flow rate of 5-10 ml/min. To load and elute Certified Sep-Pak Vac 3 cc 200 mg, and 500 mg cartridges, use a flow rate or ml/min. To load and elute all other cartridges, use a flow rate or 2-10 ml/min. b. Sample Load V. STORAGE AND DISPOSAL OF USED CARTRIDGES Cartridges stored in their original sealed pouch remain stable for long periods. Once opened, special care must be taken to ensure that the SPE cartridges are not exposed to the laboratory environment for long periods. The highly active sorbents will readily accumulate background contaminants from the environment and will affect the quality of the device. Waters recommends that you open and use only those cartridges that will be needed. If this is not possible, store the exposed cartridges in clean contaminant free container such as a desiccator. Dispose of used cartridges safely in accordance with applicable government or local regulations. Cartridge capacity depends upon the mass of all components in the sample, volume of solvent, solvent strength, sorbent type, sorbent quality, and retention factor (k) of components. Overloading results in sample breakthrough which leads to variability in sample recovery. Certified Sep-Pak Cartridges 4
5 VI. ORDERING INFORMATION Description Part Number Alumina A Certified Sep-Pak Vac, 3 cc, 500 mg, 50/box Certified Sep-Pak Vac, 6 cc, 500 mg, 30/box Certified Sep-Pak Vac, 6 cc, 1 g, 30/box Alumina B Certified Sep-Pak Vac, 3 cc, 500 mg, 50/box Certified Sep-Pak Vac, 6 cc, 500 mg, 30/box Certified Sep-Pak Vac, 6 cc, 1 g, 30/box Alumina N Certified Sep-Pak Vac, 3 cc, 500 mg, 50/box Certified Sep-Pak Vac, 6 cc, 500 mg, 30/box Certified Sep-Pak Vac, 6 cc, 1 g, 30/box Florisil Certified Sep-Pak Vac, 3 cc, 500 mg, 50/box Certified Sep-Pak Vac, 6 cc, 500 mg, 30/box Certified Sep-Pak Vac, 6 cc, 1 g, 30/box Silica Certified Sep-Pak Vac, 3 cc, 200 mg, 50/box Certified Sep-Pak Vac, 3 cc, 500 mg, 50/box Certified Sep-Pak Vac, 6 cc, 500 mg, 30/box Certified Sep-Pak Vac, 6 cc, 1 g, 30/box C18 Certified Sep-Pak Vac, 3 cc, 200 mg, 50/box Certified Sep-Pak Vac, 3 cc, 500 mg, 50/box Certified Sep-Pak Vac, 6 cc, 500 mg, 30/box Certified Sep-Pak Vac, 6 cc, 1 g, 30/box VII. ADDITIONAL INFORMATION Sep-Pak accessories are available to help with your solid-phase extraction. Ordering information for these products is below. Sep-Pak Cartridge Accessories Description Qty. Part No. 30 cc Reservoir 48/pkg WAT cc Reservoir 12/pkg WAT Reservoir Adapters for 3, and 6 cc VAC 12/pkg WAT Male-Male Adapter 100/pkg WAT Male Luer Plugs WAT Female Luer Plugs WAT Extraction Manifolds 20 position extraction manifold without rack WAT position extraction manifold with 13 x 75 mm test tube rack WAT position extraction manifold with 13 x 100 mm test tube rack WAT Description Qty. Part No. 20 position extraction manifold with 16 x 75 mm test tube rack WAT position extraction manifold with 16 x 100 mm test tube rack WAT Accessories and Spare Parts for the Waters Extraction Manifolds Needle valves, 20/pkg (required when using silicabased SPE cartridges) WAT Needle tips, 20/pkg WAT Cover, 20 position without gauge assembly WAT Gauge assembly, vacuum WAT Reservoir, glass with outlet valve WAT Outlet valve kit WAT Gasket for cover WAT Ejector tool WAT Luer Plugs 25/pkg WAT Rubber ball ring (for vacuum gauge assembly) WAT Reversible Vial Rack for 1 ml or 4 ml autosampler vials WAT x 75 mm test tube rack WAT x 100 mm test tube rack WAT x 75 mm test tube rack WAT x 100 mm test tube rack WAT Adapter (to attach reservoir to 1, 3, and 6 cc Sep-Pak Vac cartridges) 100/pkg WAT Adapter (to attach reservoir to 12, 20, and 35 cc Sep-Pak Vac cartridges) 12/pkg WAT Vacuum pump (110 V, 60 Hz) 10/pkg WAT Vacuum pump (220 V, 50 Hz) WAT Vacuum pump (220 V, 50 Hz) WAT For more detailed information, consult the Chromatography Columns and Supplies Catalog by visiting (literature code EN), or contact your Waters representative. If you require technical assistance, or have any questions regarding sample preparation, visit or call us at VIII. LITERATURE REFERENCES Description Certified Sep-Pak Brochure Sorbent Selection Guide Wall Chart A Sample Preparation Primer and Guide to Solid-Phase Extraction Methods Development Sample Preparation Solutions Brochure Sep-Pak Sample Extraction Products Brochure Food Safety Application Notebook Environmental Chromatography Methods Guide Literature Code EN EN WA EN EN EN EN Certified Sep-Pak Cartridges 5
6 Sales Offices Austria and European Export (Central South Eastern Europe, CIS and Middle East) Australia Belgium Brazil Canada x2205 China CIS/Russia / Czech Republic Denmark Finland France Germany Hong Kong The Netherlands Norway Poland Puerto Rico Singapore Spain Sweden Switzerland Taiwan United Kingdom All other countries: Waters Corporation U.S.A Hungary India and India Subcontinent Ireland Italy Japan Korea Mexico Waters Corporation. Waters, The Science of What s Possible, and Sep-Pak are trademarks of Waters Corporation. Florisil is a trademark of U.S. Silica. Sep-Pak Cartridges are covered by U.S. Patents # 4,211,658 and/or # 4,250,035 and several foreign patents. March EN Rev A VW-PDF Waters Corporation 34 Maple Street Milford, MA U.S.A. T: F: Certified Sep-Pak Cartridges 6
Oasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
More informationThermo Scientific HyperSep Solid Phase Extraction Method Development Guide
chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo
More information[ Care and Use Manual ]
PREP Calibration Mix DIOS Low i. Introduction Pre-packaged PREP Calibration Mixtures eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample
More informationPurification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
More informationProteinChip Energy Absorbing Molecules (EAM)
ProteinChip Energy Absorbing Molecules (EAM) Instruction Manual Catalog #C30-00001, #C30-00002, #C30-00003, #C30-00004 For technical support, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD
More informationGuide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
More information[ Care and Use Manual ]
The Waters GPC Cleanup System uses two Envirogel GPC Cleanup Columns to perform routine cleanup of environmental samples as specified by EPA Method 3640A, GPC Cleanup. The columns, 19 x 150 mm and 19 x
More informationPrepTip. Reverse Phase PrepTip User Guide
PrepTip Reverse Phase PrepTip User Guide All text, photographs and illustrations are copyrighted by Harvard Apparatus, Inc. 2004. PrepTip is a trademark of Harvard Apparatus, Inc. Harvard Apparatus 84
More information[NUGENESIS SAMPLE MANAGEMENT ] AMPLE IMPROVING LAB EFFICIENCY, ANAGEMENT ACCELERATING BUSINESS DECISIONS. bigstock.com $69
[NUGENESIS SAMPLE MANAGEMENT ] AMPLE ACCELERATING BUSINESS DECISIONS IMPROVING LAB EFFICIENCY, ANAGEMENT bigstock.com $69 [ N U GENESIS SAMPLE MANAGEMENT ] MANAGING SAMPLES SHOULD NEVER BE A BARRIER TO
More informationApplicationNOTE Introduction
Introduction Honey, like other foods, is subject to strict quality control measures before it can be sold commercially. It is a natural product, produced solely by bees from flower and tree pollen. Human
More informationThermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationORGANIC SAMPLE PREPARATION
ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,
More information[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS. Simplifed
[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS Simplifed The QuEChERS method (an acronym for Quick, Easy, Cheap, Effective, Rugged, and Safe) followed by dispersive solid-phase extraction
More information[ Care and Use Manual ]
Analytical columns and cartridges i. Introduction Waters offers a large selection of analytical columns and cartridges packed with silica-based packing materials. Follow this instruction sheet to obtain
More informationTransforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide
Transforming a ChromLab Software -D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide Instructional Guide Chromatography Bulletin 675 ChromLab
More informationENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual
ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual Catalog numbers 780-1070 780-1650 Please read these instructions before you use ENrich SEC high-resolution size exclusion
More informationMEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process
- Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples
More informationDetermination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)
Application Note 340 Determination of Fat in Dried Milk Products INTRODUCTION Many extraction techniques for the determination of fat in food are labor-intensive or require long extraction times. The Roese-Gottlieb
More informationFast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
More informationWaters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
More informationHow To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
More informationApplication Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u
Application Note USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u Summary A regeneration study of MEP HyperCel mixed-mode sorbent was conducted
More informationA Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
More informationForeign Taxes Paid and Foreign Source Income INTECH Global Income Managed Volatility Fund
Income INTECH Global Income Managed Volatility Fund Australia 0.0066 0.0375 Austria 0.0045 0.0014 Belgium 0.0461 0.0138 Bermuda 0.0000 0.0059 Canada 0.0919 0.0275 Cayman Islands 0.0000 0.0044 China 0.0000
More informationDetermination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
More informationInnovative vs Traditional
And in the red corner, introducing the challenger NASDAQ April 4 2011 243.47 $ 1.38 $ 0.58 % 0.07 $ 0.02 $ 39.4 % Technology DVD Digital Streamline Instant Play Internet Access Unlimited Selection No Late
More informationSolid Phase Extraction Products PAGE: 1. Introduction of Solid Phase Extraction (SPE) Why Choose Nano-Micro Tech SPE
Solid Phase Extraction Products PAGE: 1 Introduction of Solid Phase Extraction (SPE) SPE has been used increasingly in chemical analysis and purification broadly, and become the most popular technology
More informationWATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
More informationThis study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA
An Automated Method for the Fractionation of Extractable Petroleum Hydrocarbons (EPH) from Water and Soil Extracts Using the Gilson GX-274 ASPEC System Keywords: Aliphatic Hydrocarbons, Aromatic Hydrocarbons,
More informationSPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322. i.chianella.1998@cranfield.ac.
SPE and HPLC Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0) 1234 758322 i.chianella.1998@cranfield.ac.uk Solid-Phase Extraction- SPE Simple, fast and efficient sample preparation
More informationPall Aria Integrated MF/RO Systems for Cost-Effective Treatment of Sea and Brackish Waters
Pall Aria Integrated MF/RO Systems for Cost-Effective Treatment of Sea and Brackish Waters Pall Aria Integrated MF/RO Systems for Cost-Effective Treatment of Sea and Brackish Waters Water scarcity resulting
More informationCONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
More informationINSTRUCTIONS 56-1190-98. Edition AC
Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography
More informationSupported Payment Methods
Supported Payment Methods Global In the global payments market, credit cards are the most popular payment method. However, BlueSnap expands the payment selection by including not only the major credit
More informationGet the benefits of Norgren s unique range of Online services
Get the benefits of Norgren s unique range of Online services Make your job easier and save time - everything you need to select, design and purchase Norgren pneumatics is in one convenient location, available
More informationSupported Payment Methods
Sell Globally in a Snap Supported Payment Methods Global In the global payments market, credit cards are the most popular payment method. However, BlueSnap expands the payment selection by including not
More informationWhat is Eluent Generation?
What is Eluent Generation? Monika Verma Thermo Fisher Scientific, Sunnyvale, CA, USA White Paper 70607 Executive Summary Eluent generation is an electrolytic process that allows the production of high
More informationThe Theory of HPLC. Gradient HPLC
The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims
More informationAffi-Prep Protein A Matrix Instruction Manual
Affi-Prep Protein A Matrix Instruction Manual Catalog Numbers 156-0005 156-0006 Bio-Rad Laboratories, 2000 Alfred Nobel Dr., Hercules, CA 94547 LIT-230 Rev B Table of Contents Section 1 Introduction...1
More informationHow To Install A Gel Doc Xr+ Camera On A Computer Or Mouse Or Tv (For A Powerbook) With A Zoom Lens (For An Ipa) With An Iphone Or Ipa (For Mac) With The Zoom Lens
Molecular Imager Gel Doc XR+ Installation Guide To install your Gel Doc XR+, perform these seven steps in the following order: 1. Unpack the Gel Doc XR+ system. 2. Install Image Lab software. 3. Set up
More informationExtraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
More informationHigh-Throughput 3-D Chromatography Through Ion Exchange SPE
High-Throughput 3-D Chromatography Through Ion Exchange SPE Application Note 205 Luke Roenneburg and Alan Hamstra (Gilson, Inc.) Introduction 2-dimensional (2-D) separation is the separation of a sample
More informationSolid Phase Extraction for Water Analysis. presented by
Solid Phase Extraction for Water Analysis presented by Dr. Bernd Thierfelder Outline 1. Introduction in SPE 2. Retention Mechanisms / Sorbent Selectivity 3. strata -X for Water Analysis 4. Applications
More informationImage Lab Software for the GS-900 Densitometer
Image Lab Software for the GS-900 Densitometer Quick Start Guide Catalog # 170-9690 Bio-Rad Technical Support For help and technical advice, please contact the Bio-Rad Technical Support department. In
More informationPall in the Brewery (more value per hectolitre) DE Free Clarification FB1795
Pall in the Brewery (more value per hectolitre) DE Free Clarification FB1795 The Pall World Pall Corporation is a world leader in the design, production and supply of filters, membranes and systems for
More informationAutomation of Solid Phase Extraction and Column Chromatographic Cleanup
Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for
More informationDissolved and precipitated oxalate
Accepted 2005 Process liquors from bleach plants Dissolved and precipitated oxalate Using Ion Chromatography 0 Introduction In bleach plants of pulp mills with a high degree of system closure, there is
More informationSPE Reference Manual & Users Guide. Reference Manual & Users Guide
SM SPE Reference Manual & Users Guide SPE Reference Manual & Users Guide Reference Manual & Users Guide SPE Table of Contents Page No. 1. Objective and Purpose of This Guide 1 2. Solid Phase Extraction
More informationDiafiltration: A Fast, Efficient Method for Desalting, or Buffer Exchange of Biological Samples
Scientific & Technical Report PN 33289 Diafiltration: A Fast, Efficient Method for Desalting, or Buffer Exchange of Biological Samples By Larry Schwartz, Senior Technical Manager, Pall Life Sciences Diafiltration
More informationUltraClean PCR Clean-Up Kit
UltraClean PCR Clean-Up Kit Catalog No. Quantity 12500-50 50 Preps 12500-100 100 Preps 12500-250 250 Preps Instruction Manual Please recycle Version: 02212013 1 Table of Contents Introduction... 3 Protocol
More informationAgilent 4338B Milliohm Meter
Agilent 4338B Milliohm Meter 10 µω to 100 kω Technical Overview Introduction Ideal for precise measurements of extremely low resistances using an ac test signal, the Agilent Technologies 4338B suits bench-top
More informationInstallation Qualification/Operational Qualification Protocols and Instructions. Experion Automated Electrophoresis Station and Software
Installation Qualification/Operational Qualification Protocols and Instructions Experion Automated Electrophoresis Station and Software Validation Kit (Catalog #700-7051) Security Edition Software (Catalog
More informationSPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
More informationRESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml
GE Healthcare Life Sciences Instructions 71-7146-00 AI Ion Exchange Columns RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml Introduction RESOURCE Q and S are pre-packed columns for separating biomolecules
More informationBasics of SPE Technology & Mechanisms. Pieter Grobler, Sigma-Aldrich RSA
Basics of SPE Technology & Mechanisms Pieter Grobler, Sigma-Aldrich RSA Agenda The Importance of Sample Prep Overview of SPE Technology SPE Strategies Understanding Retention Mechanisms Analytical Chromatography
More informationGas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
More informationThermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
More informationPLEUR EVAC CHEST DRAINAGE SYSTEMS. Great Innovations to Enhance Chest Drainage
PLEUR EVAC CHEST DRAINAGE SYSTEMS Great Innovations to Enhance Chest Drainage 2 INTRODUCTION Teleflex Great Innovations to Enhance Chest Drainage Trusted brands make Teleflex a reliable and strong partner.
More informationExtraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
More informationAnalysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )
Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 ) Andrew Aubin Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S Fast analysis of
More informationSupplies 1. Toner cartridges. Printing. Using color. Paper handling. Maintenance. Troubleshooting. Administration. Index
Toner cartridges Supplies 1 Warning: Do not touch the cartridge roller located under the cover. Cover Your printer uses four separate toner cartridges, one for each color: cyan, yellow, magenta, and black.
More informationPCR and Sequencing Reaction Clean-Up Kit (Magnetic Bead System) 50 preps Product #60200
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com PCR and Sequencing Reaction Clean-Up Kit (Magnetic Bead System)
More informationCell Sorting. S3e Cell Sorter Quick Guide
Cell Sorting S3e Cell Sorter Quick Guide STARTUP Checking Fluidics STARTUP There are three bulk fluid containers (Figure 1) that may require attention before starting the system. If fluidics need to be
More informationImage Lab Software How to Obtain Stain-Free Gel and Blot Images. Instructions
Image Lab Software How to Obtain Stain-Free Gel and Blot Images Instructions Table of Contents Obtaining Stain-Free Gel images...1 5 Obtaining Stain-Free Blot Images...6 9 If the Gel Doc XR+ or ChemiDoc
More informationClinical and Laboratory Practices Workshops. Helping to improve laboratory skills and the quality of care in resource-limited settings
Clinical and Laboratory Practices Workshops Helping to improve laboratory skills and the quality of care in resource-limited settings BD is committed to collaborating with governmental and non-governmental
More informationBT Premium Event Call and Web Rate Card
BT Managed Event and BT Self-Managed Event (also referred to as Express, Plus and Premium) Conference Bridge and Call for Booked Audio Conferencing Services will comprise the following for each phone-conference:
More informationAutomated Method Development Utilizing Software-Based Optimization and Direct Instrument Control
Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control Dr. Frank Steiner, 1 Andreas Brunner, 1 Fraser McLeod, 1 Dr. Sergey Galushko 2 1 Dionex Corporation, Germering,
More informationCNE Progress Chart (CNE Certification Requirements and Test Numbers) (updated 18 October 2000)
Novell Education CNE Progress Chart (CNE Certification Requirements and Test Numbers) (updated 18 October 2000) Prerequisite for all certifications In order to receive a Novell certification the candidate
More informationHiPer Ion Exchange Chromatography Teaching Kit
HiPer Ion Exchange Chromatography Teaching Kit Product Code: HTC001 Number of experiments that can be performed: 5 Duration of Experiment: Protocol: 5-6 hours Storage Instructions: The kit is stable for
More informationQuantum View Manage Administration Guide
2010 United Parcel Service of America, Inc. UPS, the UPS brandmark and the color brown are trademarks of United Parcel Service of America, Inc. All rights reserved. Quantum View Manage Administration Guide
More information50 g 650 L. *Average yields will vary depending upon a number of factors including type of phage, growth conditions used and developmental stage.
3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Phage DNA Isolation Kit Product # 46800, 46850 Product Insert
More informationBrochure More information from http://www.researchandmarkets.com/reports/1339929/
Brochure More information from http://www.researchandmarkets.com/reports/1339929/ The 2011 World Forecasts of Machine Tools That Remove Material by Laser or Light, Photon, Ultrasonic, Electro-Discharge,
More informationBrochure More information from http://www.researchandmarkets.com/reports/3278449/
Brochure More information from http://www.researchandmarkets.com/reports/3278449/ The 2016 World Forecasts of Hand-Operated Date, Sealing, or Numbering Stamps; Devices for Printing or Embossing Labels;
More informationHigh-voltage Differential Probes TMDP0200 - THDP0200 - THDP0100 - P5200A - P5202A - P5205A - P5210A
High-voltage Differential Probes TMDP0200 - THDP0200 - THDP0100 - P5200A - P5202A - P5205A - P5210A BNC interface (P5200A probes) TekVPI interface (TMDP and THDP Series probes) TekProbe interface (P5202A,
More informationReport on Government Information Requests
Report on Government Information July 1 - December 31, 2014 apple Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software
More informationBuilding a Global Internet Company: Driving Traffic to Your Site. Benjamin Edelman Harvard Business School
Building a Global Internet Company: Driving Traffic to Your Site Benjamin Edelman Harvard Business School Mumbai, India January 20, 2012 About me Assistant Professor, Harvard Business School Teaching:
More informationAgilent 4339B/4349B High Resistance Meters
Agilent 4339B/4349B High Resistance Meters Technical Overview Within Budget Without Compromise Introducing the Agilent Technologies 4339B and 4349B High Resistance Meters Used for Making Ultra- High Resistance
More informationReport on Government Information Requests
Report on Government Information January 1 - June 30, 2015 Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software and
More informationEmpore. 96-Well Solid Phase Extraction Plates. Technical Information. Instructions for Use. Product Characteristics. Product Description
Technical Information Empore 96-Well Solid Phase Extraction Plates Instructions for Use Product Description Empore 96-Well Solid Phase Extraction Plates are designed for high throughput solid phase extraction
More informationTHE WORLD S LEADING CAR DESIGN MAGAZINE
THE WORLD S LEADING CAR DESIGN MAGAZINE DISTRIBUITED IN MORE THAN 60 COUNTRIES EUROPE: Austria, Belgium, Cyprus, Denmark, Finland, France, Germany, United Kingdom, Greece, Ireland, Iceland, Italy, Latvia,
More informationHiTrap Heparin HP, 1 ml and 5 ml
Instructions 71-7004-00 AU HiTrap affinity columns HiTrap Heparin HP, 1 ml and 5 ml HiTrap Heparin HP is a prepacked ready to use, column for preparative affinity chromatography. The special design of
More informationOCTOBER 2010. Russell-Parametric Cross-Sectional Volatility (CrossVol ) Indexes Construction and Methodology
OCTOBER 2010 Russell-Parametric Cross-Sectional Volatility (CrossVol ) Indexes Construction and Methodology SEPTEMBER 2010 Russell-Parametric Cross-Sectional Volatility (CrossVol) Indexes Construction
More information105 Adopted: 27.07.95
105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline
More informationExperiment 5: Column Chromatography
Experiment 5: Column Chromatography Separation of Ferrocene & Acetylferrocene by Column Chromatography Reading: Mohrig, Hammond & Schatz Ch. 18 pgs 235-253 watch the technique video on the course website!
More informationWizard DNA Clean-Up System INSTRUCTIONS FOR USE OF PRODUCT A7280.
Technical Bulletin Wizard DNA Clean-Up System INSTRUCTIONS FOR USE OF PRODUCT A7280. PRINTED IN USA. Revised 4/06 AF9TB141 0406TB141 Wizard DNA Clean-Up System All technical literature is available on
More informationA Novel Approach to Low Volume Sample Preparation
A Novel Approach to Low Volume Sample Preparation Tony Edge, Jon Bardsley, Ken Meadows, Luisa Pereira International Symposium on Chromatography Salzburg, Austria September 2014 1 The world leader in serving
More informationPore size and BET surface area. analysis at your fingertips. SA 3100. Gas Adsorption Analyzer
Pore size and BET surface area analysis at your fingertips. SA 3100 Gas Adsorption Analyzer The History of Beckman Coulter Particle Characterization Introducing The SA 3100 Surface Area and Pore Size Analyzer
More informationReport on Government Information Requests
Report on Government Information July 1 - December 31, 2015 Apple takes our commitment to protecting your data very seriously and we work incredibly hard to deliver the most secure hardware, software and
More informationAutoSpec Premier. Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements?
AutoSpec Premier Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements? Looking for sensitivity, selectivity and dynamic range? The AutoSpec Premier is the latest
More informationEarning Customer Loyalty
Earning Customer Loyalty Loyalty cannot be purchased. It s something we earn. When it comes to attracting and retaining dedicated customers, we rely on excellent service to build solid relationships. A
More informationamplification tech Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye
amplification tech note 6047 Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye Liz Jordan and Richard Kurtz, Gene Expression Division, io-rad
More informationKeysight Technologies Using Fine Resolution to Improve Thermal Images. Application Note
Keysight Technologies Using Fine Resolution to Improve Thermal Images Application Note Introduction Digital imaging opened up new possibilities and has been very much part of our lives since social media
More informationWho We Are. Denis Thiery Chairman and Chief Executive Officer
Who We Are Denis Thiery Chairman and Chief Executive Officer Founded in 1924, Neopost has grown to become a global leader in mailing solutions and a major player in digital communications and shipping
More informationSelective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
More informationThe Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.
The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen. Mark Crawford Single Presenter Name Abstract With the recent decline in hive bee populations this year the investigation of acaricides
More informationDetection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
More informationMERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM AN ONLINE TOOL DESIGNED TO TAKE THE WORK OUT OF YOUR COMPENSATION REVIEW PROCESS
MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM AN ONLINE TOOL DESIGNED TO TAKE THE WORK OUT OF YOUR COMPENSATION REVIEW PROCESS MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM www.imercer.com/cars Mercer
More information