1 PAGE: 1 of 11 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS 7.0 PROCEDURE 8.0 CALCULATIONS 9.0 QUALITY ASSURANCE/QUALITY CONTROL 10.0 DATA VALIDATION 11.0 HEALTH AND SAFETY 12.0 REFERENCES 13.0 APPENDIX A - Digi-Sense Digital ph/mv/orp Operator s Manual REPLACES: SOP #2077, Revision 0.0; 12/19/96. U.S. EPA Contract EP-W

2 PAGE: 2 of SCOPE AND APPLICATION This standard operating procedure (SOP) describes the measurement of ph in aqueous media using a COLE-Palmer Digi-Sense digital ph/mv/orp meter. These are standard (i.e. typically applicable) operating procedures which may be varied or changed as required, dependent upon site conditions, equipment limitations, or limitations imposed by the procedure. In all instances, the ultimate procedures employed should be documented and associated with the final report. Mention of trade names or commercial products does not constitute U.S. Environmental Protection Agency (U.S. EPA) endorsement or recommendation for use. 2.0 METHOD SUMMARY The electromotive force (EMF) produced in the glass electrode system varies linearly with ph. This linear relationship is described by plotting the measured EMF against the ph of different buffers. Sample ph is determined by extrapolation (1). 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING and STORAGE No preservation is necessary. Samples should be stored in glass or polyethylene bottles, maintained at 4 o C and analyzed immediately. Analyses are typically performed at room temperature. All samples and calibration buffers should be allowed to equilibrate to ambient temperature prior to analysis. The amount of the sample to be collected and the proper sample container type (i.e., glass, plastic), chemical preservation, and storage requirements are dependent on the matrix being sampled and the parameter(s) of interest, and are discussed in ERT/SERAS SOP #2003, Sample Storage, Preservation and Handling for water matrices. 4.0 INTERFERENCES AND POTENTIAL PROBLEMS The combination electrode is relatively free from interference from color, turbidity, colloidal matter, oxidants, reductants, and high salinity, except for a sodium error at ph >10 (1). Measurements of ph are affected by temperature in two ways: mechanical effects that are caused by changes in the properties of the electrodes, and chemical effects caused by equilibrium changes. Standard ph buffers have a specific ph at indicated temperatures. 5.0 EQUIPMENT/APPARATUS The following are standard materials and equipment required for testing: ph meter, Model No

3 PAGE: 3 of REAGENTS Combination ph electrode Magnetic stir plate Magnetic stir bars Thermometer 9V battery At least two standard known ph buffers are needed to calibrate the meter. The three most common are ph 4.00, 7.00, and buffers. The buffers selected should bracket the ph of the samples. A saturated aqueous potassium chloride (KCl) solution is used for filling the combination electrode. If separate glass and reference electrodes are used, the reference electrode is filled with saturated aqueous KCl. Deionized or distilled water should be used for rinsing the probe between samples. 7.0 PROCEDURE 7.1 ph Calibration Procedure 1. Press ON/OFF key to turn meter on. 2. Connect the ph probe to the meter. 3. Immerse the probe in buffer Press the ph/mv key to select ph. 5. Adjust temp EC control to the temperature of buffer Note: Both buffers should be at the same temperature. 6. Adjust the standardize control to read 7.00 ph on the display. 7. Rinse the ph probe with deionized or distilled water and blot dry. 8. Immerse the probe in the second buffer (4.00 or 10.00) 9. Adjust the temp EC control to the temperature of the second buffer. 10. Allow the reading to stabilize, and then adjust the SLOPE control to the value of the second buffer. 11. Rinse probe with deionized or distilled water. Calibration is now complete. The ph meter should be calibrated at the beginning of every eight-hour operating period, and the calibration must be checked at least once during every subsequent eight-hour period.

4 PAGE: 4 of Operation 1. Push the ON/OFF key to turn the meter on. 2. Push the ph/mv key until the annunciators indicate the desired mode. For millivolt or ORP (oxidation-reduction potential) measurement: Press the ph/mv key until the display shows millivolts. Verify the probe connection, then rinse the probe with distilled water and blot dry. Immerse the probe in the sample to be measured. Allow the displayed reading to stabilize, then take the reading. 7.3 Temperature Compensation Temperature compensation can be set manually by the temperature o C adjustment over a range of 0 o to 100 o C. 7.4 Sample Measurement 1. Measure the temperature of the sample to be measured 2. Rinse the probes with distilled or deionized water. Blot dry. With the meter on, place the electrode in the sample solution to be measured, which is being magnetically stirred. 3. If the meter is calibrated using ph 4.00 and ph 7.00 buffers and the sample reading is >7.00, the meter will be recalibrated using ph 7.00 and buffers. Adjust the temp EC control the match the sample solution temperature. The sample ph will be displayed. Record the reading once the meter has stabilized. For millivolt or ORP measurement: Press the ph/mv key until the display show millivolts. Verify the probe connection, then rinse the probe with distilled water and blot dry. Immerse the probe in the sample to be measured. Allow the display reading to stabilize, then take the reading. 7.5 Battery Replacement The ph meter uses a 9 volt battery with a life of 2000 hours. If the low battery indicator is on, stop operation and replace the internal battery with a new 9 volt battery. 7.6 Cleaning the Probe The glass bulb is the sensitive part of the probe, it should always be kept clean. Rinse the probe with deionized or distilled water after use. Before storage, rinse the probe with tap or distilled water, shake dry and place the probe in the protective cap, which should be filled with a KCl solution or equivalent probe storage solution.

5 PAGE: 5 of CALCULATIONS If KCl or equivalent storage solution is not available, use a 4.00 ph buffer, 7.00 ph buffer, or tap water. Distilled water should never be used. The value displayed is read directly as ph. The temperature of the samples and calibration buffers should be identical to assure accuracy. Record sample temperature with ph value. 9.0 QUALITY ASSURANCE/QUALITY CONTROL All data must be documented on field data sheets or within site or laboratory notebooks. All instrumentation must be operated in accordance with the manufacturer s instructions. Equipment check-out procedures and calibration activities must be performed. (See section 7.1). Duplicate samples should be processed with the frequency of one in twenty samples. Duplicate samples will be used to determine precision DATA VALIDATION By careful use of a laboratory ph meter with good electrodes, a precision of ±0.02 ph unit and accuracy of ± 0.05 ph unit can be achieved. However ±0.1 ph unit represents the limit of accuracy under normal conditions, especially for measurement of water and poorly buffered solutions (1) HEALTH AND SAFETY General laboratory safety practices should be followed. Waste samples should be handled with care due to the uncertainty of the properties and contents involved. Refer to the specific material safety data sheet (MSDS) for the hazardous properties of any chemical or reagent utilized in this analysis. All excess samples, used samples, and waste material generated during analysis must be disposed in accordance with ERT/SERAS SOP #1501, Hazardous Waste Management. When working with potentially hazardous materials, follow U.S. EPA, OSHA, and Lockheed Martin health and safety procedures. More specifically, refer to ERT/SERAS SOP #3013, Laboratory Safety Program REFERENCES (1) American Public Health Association (APHA) Standard Methods for the Examination of Water and Wastewater, 18th Edition. Baltimore, Maryland: Port City Press. (2) Cole-Parmer Instrument Co. Digi-Sense Digital ph/mv/orp Meter Operator s Manual, A , Edition 2692.

6 PAGE: 6 of APPENDIX A - Digi-Sense Digital ph/mv/orp Operator s Manual

7 PAGE: 7 of 11 APPENDIX A Digi-Sense Digital ph/mv/orp Meter Operator s Manual SOP #2077 November 2000

8 PAGE: 8 of 11

9 PAGE: 9 of 11

10 PAGE: 10 of 11

11 PAGE: 11 of 11

USEPA Electrode Method Method 8156

USEPA Electrode Method Method 8156 ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method

More information

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination

More information

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process

More information

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure

More information

Measurement by Ion Selective Electrodes

Measurement by Ion Selective Electrodes Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable

More information

AIRMoN, ph and Conductivity Field Measurement

AIRMoN, ph and Conductivity Field Measurement AIRMoN, ph and Conductivity Field Measurement Items needed: Sample to be measured (sample volume 50 ml) Field Observer Form (FOF) for the sample to be measured conductance cell ph vials, 4 conductivity

More information

EQUIPMENT CALIBRATION/MAINTENANCE

EQUIPMENT CALIBRATION/MAINTENANCE EQUIPMENT CALIBRATION/MAINTENANCE The following chart provides a guide for calibration of standard equipment used in the Fish Health Laboratory. All Intervals suggested are subject to individual manufacturer

More information

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration

More information

ph Measurement and Control

ph Measurement and Control ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components

More information

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Milwaukee USER MANUAL

Milwaukee USER MANUAL Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the

More information

DR/4000 PROCEDURE. CHLORINE, Free

DR/4000 PROCEDURE. CHLORINE, Free DR/4000 PROCEDURE CHLORINE, Free Method 10069 DPD method* UHR (0.1 10.0 mg/l as Cl 2 ) Scope and Application: For testing higher levels of free chlorine (hypochlorous acid and hypochlorite ion) in drinking

More information

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference

More information

GA/7 Potentiometric Titration

GA/7 Potentiometric Titration Rev. 7/99 7-1 INTRODUCTION GA/7 Potentiometric Titration The potentiometric titration is a useful means of characterizing an acid. The ph of a solution is measured as a function of the amount of titrant

More information

ph and Acidity in Tomato Sauce

ph and Acidity in Tomato Sauce ph and Acidity in Tomato Sauce DOC316.52.93100 Endpoint potentiometric titration Scope and application: Tomato sauce 1. Introduction This working procedure refers to the determination of titratable acidity

More information

UltraBasic Benchtop ph Meter Operation Manual

UltraBasic Benchtop ph Meter Operation Manual UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.

More information

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date: PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA

More information

ph Measurements of Common Substances

ph Measurements of Common Substances Chem 100 Section Experiment 10 Name Partner s Name Introduction ph Measurements of Common Substances The concentration of an acid or base is frequently expressed as ph. Historically, ph stands for the

More information

Acid Dissociation Constants and the Titration of a Weak Acid

Acid Dissociation Constants and the Titration of a Weak Acid Acid Dissociation Constants and the Titration of a Weak Acid One of the most important applications of equilibria is the chemistry of acids and bases. The Brønsted-Lowry acid-base theory defines an acid

More information

ph and Acidity in Wine and Fruit Juice

ph and Acidity in Wine and Fruit Juice ph and Acidity in Wine and Fruit Juice DOC316.52.93089 Endpoint potentiometric titration Scope and application: Wine and fruit juice 1. Introduction This working procedure refers to the titration of total

More information

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration. S HIFT INTO NEUTRAL 15. Acid-Base Titration Shift into Neutral Student Instruction Sheet Challenge Discover the concentration of an unknown acid solution using acid-base titration. Equipment and Materials

More information

9. Analysis of an Acid-Base Titration Curve: The Gran Plot

9. Analysis of an Acid-Base Titration Curve: The Gran Plot 9. Analysis of an Acid-Base Titration Curve: The Gran Plot In this experiment, you will titrate a sample of pure potassium hydrogen phthalate (Table 10-4) with standard NaOH. A Gran plot will be used to

More information

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506 Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking

More information

18 Conductometric Titration

18 Conductometric Titration Lab Activity 18 CONDUCTOMETRIC TITRATION LAB ACTIVITY 18 Conductometric Titration Background Titration is the a method of determining the concentration of an unknown solution (the analyte) by reacting

More information

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required

More information

ExStik TM ph Waterproof Meters

ExStik TM ph Waterproof Meters User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL

More information

Features, Limitations & Potential Errors

Features, Limitations & Potential Errors MINIMIZING USER ERRORS IN ph MEASUREMENTS Features, Limitations & Potential Errors of the Tools of ph Measurements Advanced Sensor Technologies, Inc. 603 North Poplar Street Orange CA 92868-1011 USA General

More information

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL 1 of 8 The quality assurance plan as presented in the Quality Manual should assure that constant and consistent test conditions are met and verified and should be tailored to the laboratory s activities.

More information

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation Introduction The importance of keeping cooling tower water in proper chemical

More information

Measuring the Conductivity of Pure Water Using USP<645>

Measuring the Conductivity of Pure Water Using USP<645> Measuring the Conductivity of Pure Water Using USP Water Analysis Instruments, Thermo Fisher Scientific Application Note 002 Key Words Water conductivity, ultrapure water (UPW), USP, water quality,

More information

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. Titration Curves PURPOSE To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. GOALS 1 To gain experience performing acid-base titrations with a ph meter. 2

More information

SenTix 950 SenTix 980

SenTix 950 SenTix 980 SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the

More information

Acid Base Titrations

Acid Base Titrations Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually

More information

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev. Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE

More information

6 H2O + 6 CO 2 (g) + energy

6 H2O + 6 CO 2 (g) + energy AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic

More information

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen 1. Check the log-book to see if somebody else has already checked the meter for accuracy within the timeframes below. If the meter hasn

More information

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01 Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method

More information

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater

More information

Commonwealth of Kentucky. Wastewater Laboratory Certification Manual

Commonwealth of Kentucky. Wastewater Laboratory Certification Manual Commonwealth of Kentucky Wastewater Laboratory Certification Manual March 2013 This page intentionally left blank ii TABLE OF CONTENTS I. INTRODUCTION I-1 II. GENERAL LABORATORY REQUIREMENTS II-1 1.0 Personnel

More information

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for

More information

Portable Waterproof ph/dissolved Oxygen Meter

Portable Waterproof ph/dissolved Oxygen Meter Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35632-Series Portable Waterproof /Dissolved Oxygen Meter WTERPROOF REDY MES % T /Dissolved Oxygen/ /

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,

More information

User Guide. 9708 Dissolved Oxygen Probes

User Guide. 9708 Dissolved Oxygen Probes User Guide 9708 Dissolved Oxygen Probes Introduction The Thermo Scientific Orion 9708 dissolved oxygen probe simplifies measurements of dissolved oxygen, particularly Biochemical Oxygen Demand (BOD).

More information

Standard buffer solutions for ph-meter calibration. Reliable Traceable to NIST Tested according to DIN 19266 and DIN 19268 Supplied with certificate

Standard buffer solutions for ph-meter calibration. Reliable Traceable to NIST Tested according to DIN 19266 and DIN 19268 Supplied with certificate Reliable Traceable to NIST Tested according to DIN 19266 and DIN 19268 Supplied with certificate ph value is probably the most common of all routinely performed measurements in laboratories. Since ph-value

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from

More information

Experiment 6 Titration II Acid Dissociation Constant

Experiment 6 Titration II Acid Dissociation Constant 6-1 Experiment 6 Titration II Acid Dissociation Constant Introduction: An acid/base titration can be monitored with an indicator or with a ph meter. In either case, the goal is to determine the equivalence

More information

Experiment 4 (Future - Lab needs an unknown)

Experiment 4 (Future - Lab needs an unknown) Experiment 4 (Future - Lab needs an unknown) USING A ph TITRATION TO DETERMINE THE ACID CONTENT OF SOFT DRINKS 2 lab periods Reading: Chapter 9, 185-197; Chapter 10, pg 212-218; Chapter 14 pg 317-323,

More information

Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.

Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Conductivity of strong and weak electrolytes TEC Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Principle It is possible to

More information

Global Water Instrumentation, Inc.

Global Water Instrumentation, Inc. Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 Temperature Sensor: WQ101 ph Sensor: WQ201 Conductivity Sensor: WQ301

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1 MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry EXPERIMENT #5 THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1 I. PURPOSE OF THE EXPERIMENT In this experiment

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide

More information

1.4 This method is capable of measuring DO in the range of 0.20 to 20 mg/l.

1.4 This method is capable of measuring DO in the range of 0.20 to 20 mg/l. Luminescence Measurement of Dissolved Oxygen in Water and Wastewater and for Use in the Determination of BOD 5 and cbod 5 Revision 1.2 October 2011 1. Scope and Application 1.1 This method is for the measurement

More information

DISSOLVED OXYGEN PROBE

DISSOLVED OXYGEN PROBE 51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe

More information

EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1 Introduction This lab is designed for you to discover the properties of electrochemical cells. It requires little previous knowledge of electrochemical

More information

PART 1 - INTRODUCTION...

PART 1 - INTRODUCTION... Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART

More information

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE Department of Bioresources Engineering Rm 111 Worrilow Hall 531 S. College Ave. Newark, DE 19716-2140

More information

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately: BOECO PORTABLE PH/ORP/TEMP METER MODEL PT-380 A Simultaneous displays ph or mv and temperature. A Convenient calibration with automatic buffer recognition of European and US buffer sets. Calibration data

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy... Instruction Manual ph 5+ ph/ C ph 6+ ph/ C/mV Ion 6+ ph/ C/mV/Ion Technology Made Easy... 68X243633 Rev 1-1 3/2011 Part of Thermo Fisher Scientific Preface This instruction manual serves to explain the

More information

Quaternary Ammonium Compounds DOC316.53.01128

Quaternary Ammonium Compounds DOC316.53.01128 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower

More information

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS 5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS Sl. No. Contents Preamble 5.1 Aim 5.2 Introduction 5.2.1 Environmental Significance 5.3 Principle 5.4 Materials Required 5.4.1 Apparatus Required 5.4.2

More information

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND #3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric

More information

Care and use guide. Contents. page

Care and use guide. Contents. page Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement

More information

Odor (Threshold Odor, Consistent Series)

Odor (Threshold Odor, Consistent Series) METHOD #: 140.1 (Issued 1971) TITLE: ANALYTE: INSTRUMENTATION: Odor (Threshold Odor, Consistent Series) Odor N/A STORET No. 60 C: 00086 Room Temp: 00085 1.0 Scope and Application 1.1 This method is applicable

More information

Professional Plus. Calibration Tips

Professional Plus. Calibration Tips Professional Plus Calibration Tips Table of Contents Introduction...1 Calibration Worksheet...1 Temperature...3 Calibration Tips...3 Troubleshooting Tips...3 Conductivity...4 Calibration Tips...4 Troubleshooting

More information

A Potentiometric Analysis of Fluoride Ion in Toothpaste

A Potentiometric Analysis of Fluoride Ion in Toothpaste CHEM 311L Quantitative Analysis Laboratory Revision 2.0 A Potentiometric Analysis of Fluoride Ion in Toothpaste In this laboratory exercise, we will analyze a toothpaste sample for its Fluoride Ion (F

More information

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater Salicylate Method * 1. Enter the stored program number for ammonia nitrogen (NH 3 -N). 2. Press: 64 ENTER The display

More information

Experiment 9 Electrochemistry I Galvanic Cell

Experiment 9 Electrochemistry I Galvanic Cell 9-1 Experiment 9 Electrochemistry I Galvanic Cell Introduction: Chemical reactions involving the transfer of electrons from one reactant to another are called oxidation-reduction reactions or redox reactions.

More information

Chem 1B Saddleback College Dr. White 1. Experiment 8 Titration Curve for a Monoprotic Acid

Chem 1B Saddleback College Dr. White 1. Experiment 8 Titration Curve for a Monoprotic Acid Chem 1B Saddleback College Dr. White 1 Experiment 8 Titration Curve for a Monoprotic Acid Objectives To learn the difference between titration curves involving a strong acid with a strong base and a weak

More information

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso SevenEasy Conductivity Italiano Español Français Deutsch English Contents Contents Page 1. Introduction...

More information

6.3 SPECIFIC ELECTRICAL CONDUCTANCE

6.3 SPECIFIC ELECTRICAL CONDUCTANCE SC 1 6.3 SPECIFIC ELECTRICAL CONDUCTANCE By D.B. Radtke, J.V. Davis, and F.D. Wilde Page Specific electrical conductance...sc 3 6.3.1 Equipment and supplies...4 6.3.1.A Maintenance, cleaning, and storage...6

More information

ph: Measurement and Uses

ph: Measurement and Uses ph: Measurement and Uses One of the most important properties of aqueous solutions is the concentration of hydrogen ion. The concentration of H + (or H 3 O + ) affects the solubility of inorganic and organic

More information

Solubility Product Constants

Solubility Product Constants Solubility Product Constants PURPOSE To measure the solubility product constant (K sp ) of copper (II) iodate, Cu(IO 3 ) 2. GOALS 1 To measure the molar solubility of a sparingly soluble salt in water.

More information

III. Chemical Kinetics

III. Chemical Kinetics WARNING NOTICE: The experiments described in these materials are potentially hazardous and require a high level of safety training, special facilities and equipment, and supervision by appropriate individuals.

More information

GUIDELINES FOR USE BY CLASS A LICENSEES IN MEETING SNP REQUIREMENTS JULY 1996

GUIDELINES FOR USE BY CLASS A LICENSEES IN MEETING SNP REQUIREMENTS JULY 1996 QUALITY ASSURANCE (QA) AND QUALITY CONTROL (QC) GUIDELINES FOR USE BY CLASS A LICENSEES IN MEETING SNP REQUIREMENTS AND FOR SUBMISSION OF A QA/QC PLAN JULY 1996 DEPARTMENT OF INDIAN AND NORTHERN AFFAIRS

More information

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation

More information

Determining the Identity of an Unknown Weak Acid

Determining the Identity of an Unknown Weak Acid Purpose The purpose of this experiment is to observe and measure a weak acid neutralization and determine the identity of an unknown acid by titration. Introduction The purpose of this exercise is to identify

More information

Appendix C. Vernier Tutorial

Appendix C. Vernier Tutorial C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer

More information

ph units constitute a scale which allows scientists to determine the acid or base content of a substance or solution. The ph 0

ph units constitute a scale which allows scientists to determine the acid or base content of a substance or solution. The ph 0 ACID-BASE TITRATION LAB PH 2.PALM INTRODUCTION Acids and bases represent a major class of chemical substances. We encounter them every day as we eat, clean our homes and ourselves, and perform many other

More information

FT 2000. FIELD MEASUREMENT OF RESIDUAL CHLORINE

FT 2000. FIELD MEASUREMENT OF RESIDUAL CHLORINE FT 2000. FIELD MEASUREMENT OF RESIDUAL CHLORINE Use in conjunction with: FT 1000 General Field Testing and Measurement FQ 1000 Field Quality Control Requirements FS 1000 General Sampling Procedures FD

More information

Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets

Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets Chemistry 119: Experiment 7 Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets Vitamin C is another name for ascorbic acid (C 6 H 8 O 6, see below ), a weak acid that can be determined by titration

More information

Lesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling

Lesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling Lesson 5: Water Conductivity and Total Dissolved Solids Water Quality Sampling Time Frame: Two 45-50 minute class periods Grade Level: 8 th 12 th Grade Overview: There are a wide variety of inorganic substances

More information

SINGLE RANGE DS METERS Operation Instructions

SINGLE RANGE DS METERS Operation Instructions SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax

More information

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters 1. SCOPE and APPLICATION 1.1 Gravimetric analysis is used to determine total suspended

More information

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Grace Analytical Lab 536 South Clark Street 10th Floor Chicago, IL 60605 April 15, 1994 Revision 2 Standard Operating Procedure

More information

Relative Humidity Calibration Kit

Relative Humidity Calibration Kit Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration

More information

Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes

Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes Tech Note Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes Probe Care and Maintenance Periodic cleaning and membrane changes are required to keep your sensors operating properly. Follow

More information

Electrical Conductivity of Aqueous Solutions

Electrical Conductivity of Aqueous Solutions Electrical Conductivity of Aqueous Solutions PRE-LAB ASSIGNMENT: Reading: Chapter 4.-4.3 in Brown, LeMay, Bursten & Murphy.. Using Table in this handout, determine which solution has a higher conductivity,.

More information

PRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION

PRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION INDUSTRIAL COATINGS RA260 WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION is water emulsifiable rust preventative oil, which provides outstanding corrosion protection for all phosphate metals.

More information

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq)

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq) Weak Acid Titration v120413 You are encouraged to carefully read the following sections in Tro (2 nd ed.) to prepare for this experiment: Sec 4.8, pp 158-159 (Acid/Base Titrations), Sec 16.4, pp 729-43

More information

PN 44-6033/rev. D December 2010. Theory and Practice of ph Measurement

PN 44-6033/rev. D December 2010. Theory and Practice of ph Measurement PN 44-6033/rev. D December 2010 Theory and Practice of ph Measurement THEORY AND PRACTICE OF ph MEASUREMENT TABLE OF CONTENTS THEORY AND PRACTICE OF ph MEASUREMENT TABLE OF CONTENTS Section Title Page

More information

Microwave Meter. Instruction Manual

Microwave Meter. Instruction Manual Microwave Meter 840046 Instruction Manual Microwave Meter 840046 Copyright 2009 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted

More information

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC Common Problems with Online Water Quality Analyzers Greg Macy Director AquatiPro LLC Objectives Diagnosing Colorimetric Analyzers Identifying Common Problems with Sodium Analyzers Troubleshooting ph Analyzers

More information

Laboratory Exercise: Calibration of a Thermometer

Laboratory Exercise: Calibration of a Thermometer CHEM 109 Introduction themistry Revision 3.1 Laboratory Exercise: Calibration of a Thermometer In this exercise we will calibrate a stem-type thermometer and then use it to correctly measure the Air temperature

More information

Instruction Manual Temp 4/5/6 Temperature Meter Series

Instruction Manual Temp 4/5/6 Temperature Meter Series Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.

More information