1 ZEISS AXIOSKOP Microscope s User Manual Managed by For information about this instrument, please contact Dr. Alloysius Budi ext or budiutama@rice.edu
2 Use of the Axioskop (Zeiss) The Axioskop is equipped for viewing samples with transmitted light (from a halogen lamp source) and has optics for: Bright field, Phase Contrast and Differential Interference Contrast (DIC) microscopy. Samples can also be viewed for fluorescence microscopy using the incidentlight system (front a mercury lamp source). 1) For transmitted light microscopy work, press the green on/off light switch (1a) at the back-right side of the microscope while light intensity / level knob is located next to switch (1b). The halogen lamp is in the vented box at the bottom rear of the scope (1c). 1c 1a 1b 2) For fluorescence microscopy work, press green on/off light switch (2a) on the front of the arc lamp power supply box behind the scope (right side- 2b). The mercury lamp is in the vented housing on the rear of the scope and is above the halogen lamp box (2c). *Note: the halogen lamp does not need to be on when only viewing fluorescence 2b 2c 2a 2b Transmitted light microscopy (Brightfield/DIC) Begin with glass slide/specimen securely in place on stage, light at medium brightness/temperature (~6). Ocular's (eyepieces) diopter rings set at "white dot-to-0-position". In most cases start with 10 x objective lens in
3 position. Obtain a "basic" focus using the coarse (large inner black knob), and then if desired, the fine (smaller outer knob) coaxial focusing control knobs (on the back left and right side of the microscope base). *Note: one revolution of the course knob = ~2 mm of stage travel toward any Z-direction Kohler Alignment To achieve maximum even illumination of the sample, the condenser needs to be properly focused and adjusted (Kohler Alignment) 1) Put the specimen in focus in the light path, then pull out the Analyzer [thin black plastic slider with dark filter that is below and to right of the ocular and the large black filter slider]. Turn/rotate counter-clockwise the lower light path. Pull out Analyzer Polarizer using its black lever (at the bottom of the condenser and above the field diaphragm) so they are not in the *Note: for Brightfield/DIC adjustment, the black horizontal metal filter slider should be in the "no-filter/ open position" (last position on left or 5th slot to left; pulled to the extreme right). **Note: After adjustment, for Brightfield use, you can leave analyzer & bottom polarizer in OUT position. For DIC use, analyzer and polarizer can be returned (slide IN) to the light path. 2) Transmitted light can be controlled by adjusting the black circular field diaphragm lens (3a) at the base of the microscope by turning its outer ring clockwise (opens) or counterclockwise (closes). To adjust the condenser and focus the light, put the sample on the stage look into the ocular and focus it. 3a 3b 3b Then close the field diaphragm, center the "small" light spot by turning the left and right silver color centering screws (3b - Two silver screws at 7 & 5 o'clock on the front of the condenser). 3c 3d
4 Raise /lower the condenser apparatus by turning the small black knob (3c - on the left back side of the condenser below the stage) until the circumference of the light image is "sharp and well defined as a hexagon shape." Sharp edge Before After * Note: During this process, the particular condenser ring optics can be selected to match the type of imaging desired: turn the condenser (turret) control ring (3d) to: HDIC for brightfield work; DIC for differential interference contrast work. For Phase Contrast work, use no 1 for 10X, no 2 for 20X and 40X and no 3 for 63X objective. DIC Brightfield Phase 1 Phase 3 **Note: it might not be necessary to adjust/focus the condenser for different objective lenses (magnifications). For example, if condenser is focused for 20x, that could be "acceptable" for 10x and 40x (air objectives); focus for 63x could be "acceptable" for 100x (oil objectives), but NOT for the 10, 20, and 40x air. Minor re-centering with the condenser centering screws can be performed: the light is best centered when the spot is opened up with the field diaphragm ring to about 80-90% full field illumination, and the light circumference is "concentric" with the edge of the field boundary. ***Note: the silver ring on the condenser opens (clockwise) and closes (counter-clockwise) a diaphragm for more or less light, respectively lesser light might give best image (totally counter-clockwise), however, adjust as necessary for the conditions.
5 The best DIC images are determined empirically, and important factors are: positions of the prism (adjusted by turning the little silver screw above each objective lens red arrows in left image) and the condenser aperture diaphragm. Turn the prism screw for the desired image, and turn the diaphragm ring 3a [more open diaphragm gives more contrast and depth of focus]. On the right side of the microscope base, there are 4 buttons for the filter magazine in the illuminating beam path, and from front-to-back are: neutral density filters 0.015, 0.06, 0.25, and a conversion filter which converts artificial halogen light (color temperature of 3200 K) into natural outdoor light (color temperature of 5500 K). When push buttons are IN or OUT, the corresponding filters are IN or OUT as well. Fluorescence microscopy After turning ON the mercury arch lamp, record in the log hook the number on the time counter and actual time of day. * Note: once the arc lamp is ON, it must stay on for at least 30 minutes before being turned off when the lamp is turned OFF, it must remain off (to cool) for at least 30 minutes before being turned ON again. filter. Pull out Analyzer ** Note: Pull out the analyzer (thin black plastic slider with dark filter which is below and to the right of the ocular and the large black metal filter slider) from the light path because it not only blocks light, but continued exposure to UV light will damage the
6 When the arc lamp is turned on, and the slide/specimen is placed in position on the stage, illumination of the sample from the objective lens will be readily apparent. The slider, black and marked L on left side and R on right side, can be used to block / interrupt the illumination beam path (move to far right position) when not viewing the sample to prevent photobleaching. The green filter/middle position slightly filters the light, but the open position (far right) is probably best when examining the sample. *Note: in fluorescence microscopy, the less exposure of the sample / fluorophore to excitation light, the less photobleaching and subsequent loss of fluorescence and intensity of signal. The fluorescence reflector or tiller slider [large horizontal black metal box between ocular and objective lenses] has 4 positions for fluorescence that contain various excitation filter / dichroic mirror / emission filter sets, from right to left: // ~ um ex. (green em.) // ~ nm ex. (red em.) // ~ nm ex. (green/yellow em) // ~ nm ex. (blue em.) Decide which filter set is appropriate/best for exciting and detecting the fluorophore(s) being used, and move the filter slider to that position. [See fluorescence reflector slider sheet for filter set details.] *Note: It should not be necessary to change/adjust the illumination beam, however, the following controls affect the diameter and centering of the light. Levers for adjusting the luminous field diaphragm (upper, left and right side) move by pushing or pulling levers to the right to reduce and to the left to increase the illuminated area [ normally," the area of light should just fill the field of view, as for the halogen lamp beam with light microscopy]. Use the two silver centering screws (lower, left and right side) to position the luminous field diaphragm - to move and align the light beam, alternate twisting both screws.
7 Zeiss Axioskop Fluorescence Reflector Slider The first (right-most) position: (for FITC/ AF488/ GFP - type fluorophores) Excitation filter: nm Dichroic mirror: 495 nm Emission filter: nm Second position: (for Rhodamine/ AF594/ Texas red - type fluorophores) Excitation filter: nm Dichroic mirror: 580 nm Emission filter: 590 nm LP Third position: (for YFP - type fluorophores) Excitation filter: nm Dichroic mirror: 510 nm Emission filter: 520 nm LP Fourth position: (for DAPI / Hoeschst - type fluorophores) Excitation filter: 300(365)390 nm Dichroic mirror: 395 nm Emission filter: 420 nm LP Fifth position: (for DIC / Brightfield) OPEN (no filters)
8 Objectives for Zeiss Axioskop Plan - NEOFLUAR 10X, NA 03 Ph 1 (air; n=1) Phase contrast: condenser position 1 Plan - NEOFLUAR 20X, NA 0.5 Ph 2 (air; n=1) Phase contrast: condenser position 2 Plan - NEOFLUAR 40X, NA 0.75 Ph 2 (air; n=1) Phase contrast: condenser position 2 Plan - NEOFLUAR 63X, NA 1.25 Ph 2 (oil; n=1.518) Phase contrast: condenser position 2 Plan - NEOFLUAR 100X, NA 1.30 (oil; n=1.518) *NOTE: NA = numerical aperture; n = refractive index)
Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012
Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012 Figure 1.1 illustrates the parts of an upright compound microscope and indicates the terminology that I use in these notes. Figure 1.1. Parts of a Compound
More informationMicroscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES
BIODIVERSITY I BIOL1051 Microscopy Professor Marc C. Lavoie marc.lavoie@cavehill.uwi.edu MAJOR FUNCTIONS OF MICROSCOPES MAGNIFY RESOLVE: => INCREASE CONTRAST Microscopy 1. Eyepieces 2. Diopter adjustment
More informationMICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.
MICROSCOPY I. OBJECTIVES To demonstrate skill in the proper utilization of a light microscope. To demonstrate skill in the use of ocular and stage micrometers for measurements of cell size. To recognize
More informationChapter 12 Filters for FISH Imaging
Chapter 12 Filters for FISH Imaging Dan Osborn The application of in situ hybridization (ISH) has advanced from short lived, non-specific isotopic methods, to very specific, long lived, multiple color
More informationNational Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104
National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR MODELS 106, 106-L 107, 107-L 108, 108-L 109-L ELEMENTARY
More informationChapter 4. Microscopy, Staining, and Classification. Lecture prepared by Mindy Miller-Kittrell North Carolina State University
Chapter 4 Microscopy, Staining, and Classification 2012 Pearson Education Inc. Lecture prepared by Mindy Miller-Kittrell North Carolina State University Microscopy and Staining 2012 Pearson Education Inc.
More informationCare and Use of the Compound Microscope
Revised Fall 2011 Care and Use of the Compound Microscope Objectives After completing this lab students should be able to 1. properly clean and carry a compound and dissecting microscope. 2. focus a specimen
More informationCompound microscope (Hund)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 Compound microscope (Hund) 15 16 17 18 19 20 1) Eyepieces (magnifies 10x), one with diopter adjustment, 2) Interp[upillary adjustment, 3) Head, 4) Revolving nosepiece,
More informationOPERATING INSTRUCTIONS XJL 101/101A REFLECTED LIGHT METALLURGICAL MICROSCOPE
OPERATING INSTRUCTIONS XJL 0/0A REFLECTED LIGHT METALLURGICAL MICROSCOPE M.R.C.LTD. OFFICES: HAHYSTADRUT 84, HOLON 58394 P.O.B. 684, TEL-AVIV 606, ISRAEL TEL: 97-3-559305,97-3-55955 FAX: 97-3-559459 www.mrclab.com
More informationThe Applied Imaging Cytogenetic Workstation
The Applied Imaging Cytogenetic Workstation Introduction to the Genus F.I.S.H. Imaging System. The Genus Imaging System is a powerful and easy way to capture Fluorescence in-situ Hybridisation pictures
More informationMT-30 & MT-90 Series. Advanced Academic Microscopes/ Advanced Academic Polarizing Microscope INSTRUCTION MANUAL
Introduction With your purchase of an MT-30/MT-90 series type microscope you have chosen for a quality product. The MT-30/MT-90 series type microscopes are developed for use at schools and laboratories.
More informationNational Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104
National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR MODELS 156, 156-S, 157 COMPOUND BIOLOGICAL MICROSCOPES
More informationHow To Use An Asbestos Microscope
Asbestos Microscopes and Accessories Pyser-SGI has been supplying microscopes and accessories into Asbestos Laboratories for over 40 years PS12 Stage Micrometer with UKAS Certificate of Calibration - For
More informationTHE COMPOUND MICROSCOPE
THE COMPOUND MICROSCOPE In microbiology, the microscope plays an important role in allowing us to see tiny objects that are normally invisible to the naked eye. It is essential for students to learn how
More informationSP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE)
SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE) We greatly appreciate your purchase of this Tamron lens. The Tamron SP AF 300mm F/2.8 LD [IF] is a fast telephoto lens developed for Canon AF single-reflex
More informationChapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012
Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012 You are sitting at your microscope working at high magnification trying to sort out the three-dimensional compartmentalization
More informationA new advance in routine inspections INVERTED MICROSCOPE CKX41/CKX31
A new advance in routine inspections INVERTED MICROSCOPE CKX41/CKX31 Phase contrast Relief contrast Incorporation of advanced UIS2 optics ensures the highest level of clarity for cell checking applications.
More informationNational Optical & Scientific Instrument Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104
National Optical & Scientific Instrument Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR STEREOSCOPIC MICROSCOPES MODEL NUMBERS 400 400TL
More informationEXPERIMENT #1: MICROSCOPY
EXPERIMENT #1: MICROSCOPY Brightfield Compound Light Microscope The light microscope is an important tool in the study of microorganisms. The compound light microscope uses visible light to directly illuminate
More informationMITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY
MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY Adapted from Foundations of Biology I; Lab 6 Introduction to Microscopy Dr. John Robertson, Westminster College Biology Department,
More informationLight and its effects
Light and its effects Light and the speed of light Shadows Shadow films Pinhole camera (1) Pinhole camera (2) Reflection of light Image in a plane mirror An image in a plane mirror is: (i) the same size
More informationAP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
More informationPrinciples of Microscopy and Confocal and Fluorescence Microscopy
Principles of Microscopy and Confocal and Fluorescence Microscopy Content This course in Light Microscopy follows the series of successful courses in Light Microscopy, Confocal and Fluorescence Microscopy
More informationVISM Evolution Scope Series
1 VISM Evolution Scope Series Congratulations on the purchase of your New VISM Evolution (EVO) Series Scope! The EVO Series of Scopes give you many great high end features and various magnification ranges
More informationVISM CQB Scope Series
1 VISM CQB Scope Series Congratulations on the purchase of your new VISM CQB Scope! The CQB Series of Scopes give you many great options so you can choose the scope that best fits your needs. Backed by
More informationZeiss Laser Scanning Microscope 510. Zuidhorst 178. For more information contact. Tom Groothuis. ZH153, t.a.m.groothuis@utwente.nl
Zeiss Laser Scanning Microscope 510 Zuidhorst 178 For more information contact Tom Groothuis ZH153, t.a.m.groothuis@utwente.nl Start Hardware Remove the blue dust cap from the microscope. If needed, turn
More informationCONFOCAL LASER SCANNING MICROSCOPY TUTORIAL
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL Robert Bagnell 2006 This tutorial covers the following CLSM topics: 1) What is the optical principal behind CLSM? 2) What is the spatial resolution in X, Y,
More informationFeatures: Mounting the Optic
MICRO GREEN DOT with INTEGRATED RED LASER The Patented Micro Green Dot with Integrated Red Laser is a compact optical system packed with many features. The Green Dot is designed as the primary targeting
More informationLecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.
Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Some limitations (disadvantages) of NSOM A trade-off
More informationNational Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104
National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR MODELS 130, 131, 132, 138 & 139 COMPOUND MICROSCOPES
More informationName Class Date Laboratory Investigation 4B Chapter 4: Cell Structure
Name Class Date Laboratory Investigation 4B Chapter 4: Cell Structure The Microscope: A Tool of the Scientist You may refer to pages 66-67, 72-73 in your textbook for a general discussion of microscopes.
More informationInfrared Viewers. Manual
Infrared Viewers Manual Contents Introduction 3 How it works 3 IR viewer in comparison with a CCD camera 4 Visualization of infrared laser beam in mid-air 4 Power Density 5 Spectral sensitivity 6 Operation
More informationReview: Amscope Metallurgical Trinocular Microscope. Michael Reese Much FRMS EMS Bethlehem, Pennsylvania USA
Review: Amscope Metallurgical Trinocular Microscope Michael Reese Much FRMS EMS Bethlehem, Pennsylvania USA Amscope is an importer and distributor of microscopes and microscopy products and accessories
More informationWAVELENGTH OF LIGHT - DIFFRACTION GRATING
PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant
More informationMeasuring the Point Spread Function of a Fluorescence Microscope
Frederick National Laboratory Measuring the Point Spread Function of a Fluorescence Microscope Stephen J Lockett, PhD Principal Scientist, Optical Microscopy and Analysis Laboratory Frederick National
More informationZeiss 780 Training Notes
Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn
More informationTS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 Instructions
TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 ENG Instructions Thank you for purchasing a Canon product. Canon s TS-E lenses are tilt-shift lenses designed for EOS cameras. The tilt-shift mechanism enables
More informationeuromex microscopen bv The Netherlands www.euromex.com page1/16
euromex microscopen bv The Netherlands www.euromex.com page1/16 Introduction The Oxion series has been designed with all kind of Life Sciences applications and great durability in mind. This resulted in
More informationMicroscope Lab Introduction to the Microscope Lab Activity
Microscope Lab Introduction to the Microscope Lab Activity Wendy Kim 3B 24 Sep 2010 http://www.mainsgate.com/spacebio/modules/gs_resource/ CellDivisionMetaphase.jpeg 1 Introduction Microscope is a tool
More informationM A R C O. CP-670 AUTOMATIC CHART PROJECTOR Instruction Manual
M A R C O CP-670 AUTOMATIC CHART PROJECTOR Instruction Manual CONTENTS GENERAL DESCRIPTION............... 1 INSTALLATION...................... 2 Mounting the Projector Positioning the Projector Positioning
More informationProducts - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available!
Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available! Arrayit SpotLight Two-Color Microarray Scanners provide the market s most affordable
More informationScience In Action 8 Unit C - Light and Optical Systems. 1.1 The Challenge of light
1.1 The Challenge of light 1. Pythagoras' thoughts about light were proven wrong because it was impossible to see A. the light beams B. dark objects C. in the dark D. shiny objects 2. Sir Isaac Newton
More informationMACRO PHOTO LENS. MP-E 65mm f/2.8 1-5 ENG. Instruction
MACRO PHOTO LENS MP-E 65mm f/2.8 1-5 ENG Instruction Thank you for purchasing a Canon product. Canon Macro Photo Lens MP-E 65mm f/2.8 1-5 is a high-magnification macro lens that can magnify the subject
More informationThe Krasnogorsk-3. Operating Instructions. Sold by NCS Products, New York
The Krasnogorsk-3 Operating Instructions Sold by NCS Products, New York [6] Viewfinder [7] Light Meter FPS Dial [8] Light Meter Power Switch [9] Light Meter ASA/ISO Knob [10] Pistol Grip [11] Lens Focus
More informationExercise 2. The Compound Light Microscope
6 Exercise 2 The Compound Light Microscope INTRODUCTION: Student Learning Objectives: After completing this exercise students will: a. Demonstrate proficient use of the microscope using low, high dry,
More informationPETRI V6 posted 7-25-'03
PETRI V6 posted 7-25-'03 This camera manual library is for reference and historical purposes, all rights reserved. This page is copyright by mike@butkus.org M. Butkus, NJ. This page may not be sold or
More informationConfocal Microscopy and Atomic Force Microscopy (AFM) A very brief primer...
Confocal Microscopy and Atomic Force Microscopy (AFM) of biofilms A very brief primer... Fundamentals of Confocal Microscopy Based on a conventional fluorescence microscope Fluorescent Microscope Confocal
More informationEF70-200mm F2.8L IS USM. Instruction
EF70-200mm F2.8L IS USM Instruction Thank you for purchasing a Canon USM lens. The Canon EF 70-200mm f/2.8l IS USM is a high-performance, telephoto zoom lens dedicated to Canon EOS cameras. It is equipped
More informationCALIBRATION FOR LAL20X & LAL24X
CALIBRATION AND FAULT FINDING FOR LAL20X & LAL24X DUMPY LEVELS MEASURING EXPERTS SINCE 1869 How The LAL20X & LAL24X Works The Automatic level is called Automatic because it requires only a simple basic
More informationOlympus BX51 Microscope & DP70 Digital Camera System
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL 001.847.913.0777 for Refurbished & Certified Lab Equipment Olympus BX51 Microscope & DP70 Digital Camera System The BX51 was voted
More informationUser Guide LUXXOR VIDEO MICROSCOPE. 2 Luxxor Video Microscope Set Up
2 Luxxor Video Microscope Set Up User Guide LUXXOR VIDEO MICROSCOPE Column When removed from its packaging, the Luxxor Video Microscope will be configured as shown, with the Vertical Slide and Slide Stop
More informationChapter 17: Light and Image Formation
Chapter 17: Light and Image Formation 1. When light enters a medium with a higher index of refraction it is A. absorbed. B. bent away from the normal. C. bent towards from the normal. D. continues in the
More informationRevision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away.
Revision problem Chapter 18 problem 37 page 612 Suppose you point a pinhole camera at a 15m tall tree that is 75m away. 1 Optical Instruments Thin lens equation Refractive power Cameras The human eye Combining
More informationThe How Do Guide to AxioVision. Author: Brian Svedberg Image Analysis Specialist Carl Zeiss MicroImaging, Inc.
The How Do Guide to AxioVision Author: Brian Svedberg Image Analysis Specialist Carl Zeiss MicroImaging, Inc. How do I grab a single image? How do I grab a single image? Click on the Camera icon to switch
More informationSP AF 90mm F/2.8 Di Macro 1:1 (Model 272E)
SP AF 90mm F/2.8 Di Macro 1:1 (Model 272E) Thank you for purchasing this Tamron lens as the latest addition to your photographic equipment. Before using your new lens, please read the contents of this
More informationBasic Optics System OS-8515C
40 50 30 60 20 70 10 80 0 90 80 10 20 70 T 30 60 40 50 50 40 60 30 C 70 20 80 10 90 90 0 80 10 70 20 60 50 40 30 Instruction Manual with Experiment Guide and Teachers Notes 012-09900B Basic Optics System
More informationP R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)
INSURANCE SCAM OPTICS - LABORATORY INVESTIGATION P R E A M B L E The original form of the problem is an Experimental Group Research Project, undertaken by students organised into small groups working as
More informationLenses and Apertures of A TEM
Instructor: Dr. C.Wang EMA 6518 Course Presentation Lenses and Apertures of A TEM Group Member: Anup Kr. Keshri Srikanth Korla Sushma Amruthaluri Venkata Pasumarthi Xudong Chen Outline Electron Optics
More information18-270mm F/3.5-6.3 Di II VC PZD for Canon, Nikon (Model B008) 18-270mm F/3.5-6.3 Di II PZD for Sony (Model B008)
R 18-270mm F/3.5-6.3 Di II VC PZD for Canon, Nikon (Model B008) 18-270mm F/3.5-6.3 Di II PZD for Sony (Model B008) Thank you for purchasing the Tamron lens as the latest addition to your photographic equipment.
More informationLeica TCS SP5 Confocal Laser Scanning Microscope User Guide 1. BASIC IMAGE ACQUISITION
Leica TCS SP5 Confocal Laser Scanning Microscope User Guide 1. BASIC IMAGE ACQUISITION This manual is the FIRST section of a THREE part Leica TCS - SP5 User Guide edited by Donald Pottle Leica True Confocal
More informationAF 70~300 mm F/4-5.6 Di LD Macro 1:2 (Model A17)
AF 70~300 mm F/4-5.6 Di LD Macro 1:2 (Model A17) Thank you for purchasing the Tamron lens as the latest addition to your photographic equipment. Di II lens (Digitally Integrated) series are designed for
More informationLeica DM LS2 and DM LB2. Life Science System Microscopes with New Ergonomic Concept
Leica DM LS2 and DM LB2 Life Science System Microscopes with New Ergonomic Concept Worldwide Customer Base Two universal specialists The universal system microscopes Leica DM LS2 and DM LB2 can be used
More informationEF70-300mm f/4-5.6l IS USM
EF70-300mm f/4-5.6l IS USM ENG Instruction Thank you for purchasing a Canon product. Dedicated to EOS cameras, the Canon EF70-300mm f/4-5.6l IS USM lens is a highperformance telephoto zoom lens equipped
More informationTEC APO140. This telescope is closer to optical perfection more than any instrument I have ever used before. R.Renzi, Italy.
140 manuals 2013_110 manuals 1/13/2013 10:40 PM Page 1 TEC APO140 This telescope is closer to optical perfection more than any instrument I have ever used before. R.Renzi, Italy. OWNER S MANUAL 2013 140
More informationFilters for Digital Photography
Filters for Digital Photography LICHTFILTER Whether for analog or Digital Photography: The best results are achieved by using correction filters - not by digitally enhancing in a software program as once
More informationEF70-300mm f/4-5.6 IS USM
EF70-300mm f/4-5.6 IS USM ENG Instruction Thank you for purchasing a Canon product. Dedicated to EOS cameras, the Canon EF70-300mm f/4-5.6 IS USM lens is a highperformance telephoto zoom lens equipped
More informationPowerful Upright Microscope Solutions for Life Science and Clinical Applications
Powerful Upright Microscope Solutions for Life Science and Clinical Applications Leica DM4 B Leica DM6 B Short training time, intuitive usage and high quality images are decisive criteria when choosing
More informationBiomedical & X-ray Physics Kjell Carlsson. Light Microscopy. Compendium compiled for course SK2500, Physics of Biomedical Microscopy.
Biomedical & X-ray Physics Kjell Carlsson Light Microscopy Compendium compiled for course SK2500, Physics of Biomedical Microscopy by Kjell Carlsson Applied Physics Dept., KTH, Stockholm, 2007 No part
More informationLaserlyte-Flex Alignment System
Laserlyte-Flex Alignment System LaserLyte-Flex The LaserLyte-Flex Alignment System is a unique, interchangeable, low cost plug and play laser system. Designed specifically for aligning and positioning
More informationLBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes
LBS-300 Beam Sampler for C-mount Cameras P/N SP90183, SP90184, SP90185 and SP90186 YAG Focal Spot Analysis Adapter P/N SP90187, SP90188, SP90189, SP90190, SP90191 User Notes Ophir-Spiricon Inc. 60 West
More informationBulb Replacement REPLACING BULBS WARNING
Bulb Replacement REPLACING BULBS Check the operation of all exterior lamps before you drive the vehicle. Caution: Before attempting to replace a bulb, ensure that both the affected lamp and the vehicle's
More informationZeiss 710 Confocal Microscope User Guide
Cornell University - Biotechnology Resource Center Imaging Facility Zeiss 710 Confocal Microscope User Guide B46 Weill Hall BRC Imaging Facility 8-6-2014 Table of Contents Startup General Procedure 2 Microscope
More informationFading Red Tones on Color LaserJet 2600s
Procedures Fading Red Tones on Color LaserJet 2600s HP Color LaserJets 2600n, 1600, and 2605 share the same basic Canon engine design. One aspect of the design brings the common problem of scanner optic
More informationUV100A Ultraviolet Air Treatment System
UV100A Ultraviolet Air Treatment System INSTALLATION INSTRUCTIONS APPLICATION When installed in forced air heating and cooling systems, the UV100A Ultraviolet Air Treatment System kills airborne microorganism
More informationForensic Science: The Basics. Microscopy
Forensic Science: The Basics Microscopy Chapter 6 Jay A. Siegel,Ph.D. Power point presentation by Greg Galardi, Peru State College, Peru Nebraska Presentation by Greg Galardi, Peru State College CRC Press,
More informationBasic Manual Control of a DSLR Camera
Basic Manual Control of a DSLR Camera Naixn 2008 Photographers don t just take photographs - they make them! Produced by Yon Ankersmit for curious-eye.com 2009 Digital Single Lens Reflex Camera The basic
More informationEF-S18-55mm f/3.5-5.6 IS II COPY ENG. Instruction
EF-S18-55mm f/3.5-5.6 IS II ENG Instruction Thank you for purchasing a Canon product. The Canon EF-S18-55mm f/3.5-5.6 IS II lens is a high-performance standard zoom lens developed for digital SLRs compatible
More informationOperating Manual for UVEX-p (JAN Scientific, Inc. Aug 2014)
Operating Manual for UVEX-p (JAN Scientific, Inc. Aug 2014) [The beige shaded areas introduce the user interface in detail. For turn on and run operation, you may skip these descriptions. The blue shaded
More informationEXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab
EXPERIMENT O-6 Michelson Interferometer Abstract A Michelson interferometer, constructed by the student, is used to measure the wavelength of He-Ne laser light and the index of refraction of a flat transparent
More informationEF-S18-135mm f/3.5-5.6 IS
EF-S18-135mm f/3.5-5.6 IS ENG Instruction Thank you for purchasing a Canon product. The Canon EF-S18-135mm f/3.5-5.6 IS lens is a high-performance high-magnification zoom lens equipped with an Image Stabilizer,
More informationLABORATORY MICROSCOPEs
LABORATORY MICROSCOPEs 2011 OPTIKA LABORATORY MICROSCOPES B-350 SERIES - Entry-level upright laboratory microscopes page 5 B-500 SERIES - High quality upright laboratory microscopes page 13 B-600 SERIES-
More informationPhysics 10. Lecture 29A. "There are two ways of spreading light: to be the candle or the mirror that reflects it." --Edith Wharton
Physics 10 Lecture 29A "There are two ways of spreading light: to be the candle or the mirror that reflects it." --Edith Wharton Converging Lenses What if we wanted to use refraction to converge parallel
More informationSP AF 17~50 mm F/2.8 XR Di-II LD Aspherical [IF] (Model A16)
SP AF 17~50 mm F/2.8 XR Di-II LD Aspherical [IF] (Model A16) Thank you for purchasing the Tamron lens as the latest addition to your photographic equipment. Di II lens (Digitally Intigrated) series are
More informationQuick Start Guide. Digital Inverted Microscope. The Microscope Evolved.
Digital Inverted Microscope Quick Start Guide 18421 Bothell-Everett Hwy. Suite 150 Mill Creek, WA 98012-6825 Phone: (866) 614-4022 or (425) 368-0444 Fax: (425) 368-0555 E-mail: info@amgmicro.com Web: www.amgmicro.com
More informationFirst let us consider microscopes. Human eyes are sensitive to radiation having wavelengths between
Optical Differences Between Telescopes and Microscopes Robert R. Pavlis, Girard, Kansas USA icroscopes and telescopes are optical instruments that are designed to permit observation of objects and details
More informationRESEARCH MACRO ZOOM SYSTEM MICROSCOPE MVX10. Macro View. The first true macro fluorescence imaging system
The first true macro fluorescence imaging system RESEARCH MACRO ZOOM SYSTEM MICROSCOPE MVX10 Macro View The ultimate in macro fluorescence im The new MVX 10 MacroView from Oly Researchers are interested
More informationCopyright 2008 OLYMPUS CORPORATION All Rights Reserved. Paper Craft Assembly Manual
Paper Craft Assembly Manual Before assembly: 1/9 Tips on assembly: Print all the parts (Photo-quality paper is recommended for ink-jet printers and paper of medium thickness (0.20 mm) for laser printers.)
More informationISO9001 ISO14001. Environmentally-friendly product All optical components in CX41 microscope use lead-free eco-glass.
Specifications Item Optical system Illumination CX41 UIS2 (Universal Infinity-corrected) optical system Built-in transmitted Koehler illuminator 6V30W halogen bulb 100-120V/220-240Vg 0.85/0.45A 50/60Hz
More informationNightspot High performance floodlights for discharge lamps and LEDs
Nightspot High performance floodlights for discharge lamps and LEDs Design: Roy Fleedwood The Nightspot series of floodlights comes in three sizes and with rotationally symmetrical reflectors: - Nightspot
More informationEF-S55-250mm f/4-5.6 IS STM COPY ENG. Instructions
EF-S55-250mm f/4-5.6 IS STM ENG Instructions Thank you for purchasing a Canon product. The Canon EF-S55-250mm f/4-5.6 IS STM is a high-performance telephoto zoom lens that makes movie shooting comfortable.
More informationST-80 MICROSCOPE With Electronic Eyepiece #6810. User Guide
ST-80 MICROSCOPE With Electronic Eyepiece #6810 User Guide The ST-80 Microscope Thank you for purchasing your student microscope from ioptron. The ST-80 Microscope is both versatile and easy to use with
More informationEF24-105mm f/4l IS USM
EF24-105mm f/4l IS USM ENG Instruction Thank you for purchasing a Canon product. The Canon EF24-105mm f/4l IS USM lens is a high-performance standard zoom lens developed for EOS cameras. It is equipped
More informationMicroscopy and Cellular Morphology
Microscopy and Cellular Morphology As we discussed in class, many organisms on the planet exist as single cells and are referred to as microorganisms bacteria, protozoans, among others. When a single microorganism
More informationsensors Miniature Sensors - S100 The universal miniature photoelectric sensor
Miniature Sensors - S1 The universal miniature photoelectric sensor Two threaded front mounting holes Two slotted rear mounting holes Anti-tampering sensor (no adjustment) Standard optic functions and
More informationMAVO-MONITOR / MAVO-SPOT Instrument Set for Contact or Distant Measurements of Luminances
Operating Instructions MAVO-MONITOR / MAVO-SPOT Instrument Set for Contact or Distant Measurements of Luminances 15043 1/1.00 10 9 8 1 7 6 1 Display 2 Slider switch cd/m² segment test 3 ON/OFF switch 4
More informationThe new standard in research microscopy
SYSTEM MICROSCOPE Olympus is about life. About photographic innovations that capture precious moments of life. About advanced medical technology that saves lives. About information- and industry-related
More informationCONSERVATION AND LIGHTING
CONSERVATION AND LIGHTING Light is essential for the examination and enjoyment of collection items. But in a museum light also means damage: dyes and pigments fade or change appearance and the materials
More informationOperating Instructions Bedienungsanleitung Mode d emploi
Operating Instructions Bedienungsanleitung Mode d emploi DW 400 www.bron-kobold.com Operating instructions DW 400 Before use Please read all the information contained in these operating instructions carefully.
More informationGE Wireless Devices Battery Replacement
60-506-319.5 Crystal Smoke Detector Two 9V Duracell 9V 1. Twist counter-clockwise until detector become loose from base. 2. Replace batteries observing correct polarity. 3. Replace detector by twisting
More informationBio 321 Lightmicroscopy Electronmicrosopy Image Processing
Bio 321 Lightmicroscopy Electronmicrosopy Image Processing Urs Ziegler Center for Microscopy and Image Analysis Light microscopy (Confocal Laser Scanning Microscopy) Light microscopy (Confocal Laser Scanning
More information