Spectral Distribution of UV Light During Aging of Medium Pressure UV Lamps

Size: px
Start display at page:

Download "Spectral Distribution of UV Light During Aging of Medium Pressure UV Lamps"

Transcription

1 Spectral Distribution of UV Light During Aging of Medium Pressure UV Lamps Charles M. Sharpless, Karl G. Linden Duke University, Department of Civil and Environmental Engineering 120 Hudson Hall Durham, NC Paris Neofotistos ONDEO Degremont, Inc Emerywood Parkway Richmond, VA ABSTRACT As part of a project to validate a UV reactor for disinfection of drinking water, the germicidal output of four medium pressure lamps housed in a continuously operating reactor was followed as a function of time. Spectra were measured monthly and were used in conjunction with radiometer readings to calculate germicidal fluence rates (using DNA absorbance weighting) at the radiometer sensor position with the lamps operating at full power (4 kw) during each measurement. Based on the measured fluence rates, the power was adjusted so that the UV lamps operated continuously to deliver a germicidal output of approximately 70% of that obtained at full power at the end of an initial 100 h burn-in period when the lamps were new. The experimental protocol was designed to ensure reproducibility of spectrometer and radiometer positioning and response. Under the conditions of this test, lamp aging has been relatively slight. The average depreciation in germicidal fluence rates for all four lamps was 17% after 3980 h of operation. Slightly more depreciation was observed in the UVC than in the UVB region (18% versus 15% for the average of all lamps). Consistent with this, the germicidal fluence rate depreciated slightly more rapidly than the total (200 to 350 nm) fluence rate. These changes indicate a more rapid depreciation of the shorter wavelengths, but the extent to which this is due to inherent changes in the lamp spectra compared to solarization of the quartz sleeves cannot be determined from the present data. Efforts will be made to separate these two effects by examining the quartz sleeves at the end of the tests. INTRODUCTION The way in which UV lamp irradiance depreciates over time has been a topic of interest ever since UV equipment was applied to pathogen inactivation (disinfection). Unfortunately, there is very little information available on the topic and even less has actually been published. In wastewater disinfection, many plants will operate the UV lamps past their expected life, because even with lower germicidal output than originally assumed, most plants will still meet their discharge permits. However, over the past several years more emphasis has been put on accurately assessing dose delivery, especially with the wide acceptance of UV disinfection for drinking water and wastewater reuse applications. This requires more careful monitoring of the UV lamps in the system and precise information regarding the depreciation of the UV lamps over time under worst-case conditions.

2 Currently, there is no industry standard protocol for aging and measuring the depreciation of UV lamps. Several approaches are possible: the UV lamps could be aged and measured in the air; they could be aged in water and measured in air; or they could be aged and measured in water. Lamp operation may also differ during aging; the lamps may or may not be cycled on and off, and the power levels may be set to provide a range of irradiance levels during the aging test. Measurement equipment has not been standardized either. For example, some equipment allows only a narrow band of wavelengths around 254 nm to be measured while others allow measurement of a broad spectrum of wavelengths in the UV-C and UV-B range. In December 2000, a protocol for lamp aging was introduced in the NWRI/AWWARF UV Disinfection guidelines [AWWARF, 2000]. The guidelines specified a narrowly defined method for determining lamp output depreciation over time. The guidelines specified aging in water, addressed dimming and cycling and minimum measuring standards. The NWRI/AWWARF guidelines were the basis for the aging and measurement of the UV lamps outlined in this paper. MATERIALS AND METHODS Overview: The tests were designed to determine the germicidal output of the lamps over time relative to their output after a 100 h burn-in period during which the lamps were operated at full power (4 kw). In the first test after this period, the germicidal output was measured at full power and at 75% power (3 kw), and the data were used to determine the operating power that would provide approximately 70% of the full power germicidal output. Between tests, the lamps were operated at sufficient power to maintain 70% germicidal output (i.e., after every test the operating power was increased slightly to account for lamp depreciation as determined during the test). Tests were conducted at two places. For approximately the first 4000 h of testing, the reactor was installed at the Appomatox River Water Authority drinking water treatment plant (ARWA, Chesterfield, VA) where water was taken post-filtration. A typical absorbance spectrum of the water delivered to the reactor is shown in Figure 1. Problems with maintaining adequate flow led to relocating the reactor to Degremont North American Research & Development (DENARD) where water was taken from the tap and re-circulated through the reactor and a chiller with the addition of approximately 10% fresh tap water to ensure adequate cooling of the lamps. All the data presented here were obtained after operation at ARWA, and tests are presently continuing at DENARD. During tests and daily operation of the reactor, the water flow rate was approximately 100 gal/min, which was sufficient to keep the temperature of the reactor at that of the water (at or below 78 o F). During daily operation, the lamps were subjected to one on/off cycle per day, and one automatic wiper cycle per day. UV Reactor: The UV reactor used for aging and monitoring the UV lamps was a modified Aquaray H 2 O 20 reactor with four medium pressure UV lamps. The UV lamps can be operated up to 4KW power and they have a 50 cm arc length. The UV lamp is designed so both electrical connections are on one side so they are accessed and maintained from one side of the reactor. The Aquaray H 2 O uses two state of the art electronic ballasts to power the UV lamps. The ballasts operate from 480 V, 3 phase, 50/60 Hz power and are designed to operate medium pressure discharge lamps. Each ballast has outputs for two UV lamps. In contrast to conventional ballasts (inductive lamp ballast or transformer with transductor), the lamp with an electronic ballast is operated at high frequency (approx. 250 khz). The lamp does not flicker and dimming is infinitely adjustable to a range between 20% and 100% of the electric power. Each UV lamp in the reactor is monitored by a UV sensor. Due to the modifications

3 required for the lamp measurement, special sensor ports were used in order to bring the sensor port closer to the UV lamp reducing the water gap to 3.8 cm (Figure 2). Figure 1: Absorbance spectrum of post-filter AWRA water (October, 2002) Absorbance (cm -1 ) Figure 2: Schematic of the sensor port arrangement. Radiometer/sensor in sensor port Sensor port Reactor wall Quartz sleeve UV lamp Test Equipment: Transmittance spectra of optical components used in the tests (filters and screen diffusers) were measured in the laboratory at Duke University on a Cary Bio-100 double-beam scanning spectrometer (Varian, Inc.). The same instrument was used to record the absorbance spectra of water from the reactor. An IL1700 radiometer system was used (International Light, Inc., Newburyport, MA). The detector was an SED240 equipped with a W-diffuser and a QNDS1 filter. The detector response function as determined by the manufacturer during its most recent calibration (July, 2002) is shown in Figure 3 along with the transmittance spectrum of the QNDS1 filter.

4 Figure 3: Transmittance spectrum of the QNDS1 filter and SED240 detector response function including W-diffuser %T SED 240 Response (ma cm 2 mw -1 ) A hand-held fiber-optic spectrometer (S2000, Ocean Optics, Inc., Dunedin, FL) was used without the fiber-optic cable to obtain lamp emission spectra. The spectral response of the spectrometer was calibrated in the laboratory at Duke against a NIST traceable D 2 lamp (Thermo Oriel, Stratford, CT). The spectrometer was used in scope mode, which directly records counts from the diode array detector. Counts were converted into spectral irradiance (mw cm -2 nm -1 ) using the conversion factors determined with the calibrated D 2 lamp. The radiation intensity was too high to connect the spectrometer directly to the sensor port, so an adapter was made that housed two diffuser screens with meshes crossed at 45 o. While the individual screens have an essentially flat spectral response, the crossed screens show a slight wavelength dependence of the transmittance. The transmittance spectrum of the crossed screens is shown in Figure 4. Figure 4: Transmittance spectrum of the crossed screens in the spectrometer adapter %T

5 Testing Procedure: Each test followed a standard procedure beginning with turning off the reactor and draining the water. The reactor was then opened, and the quartz sleeves were cleaned with a citric acid solution followed by a water rinse. Sleeves and lamps were visually inspected for wear while the reactor was open. In the case of lamp 2, the mounts were positioned such that the return wire for the electrodes passed in front of the sensor window during normal operation so the lamp was rotated 180 o during tests to prevent this from interfering with the measurements. The regular sensor ports were then replaced with sensor ports identical to the ones used in daily operation but that were used only during tests to prevent them from solarizing or otherwise aging. At this point, the reactor was sealed and filled; tests were conducted at a flow rate of approximately 100 gal/min, which was also the regular operating flow rate. Each lamp was tested individually at 4 kw (full power) while the other lamps remained off. Several measurements were made for each lamp. The radiation intensity was quantified three different ways: by radiometer, by standard sensor included with the reactor, and by DVGW-type sensor. Other data collected include the emission spectra of the lamps and their operating current. The germicidal and 200 to 350 nm fluence rates were determined by combining radiometric and spectrometric measurements, and the method is described in detail below. Briefly, the radiometer detector was placed in the reactor sensor port(s) to measure the apparent fluence rate; the term apparent is used because the value measured by the radiometer is a function of the water transmittance, the optics and sensitivity of the detector, and the spectral distribution of the radiation. Corrections for these parameters were made mathematically as described below. The calibration plane of the detector defined the plane in which the fluence rate was measured. To ensure reproducibility of the detector placement from test-to-test, a PVC mount was constructed that could be screwed into the sensor ports of the reactor (Figure 5). A similar arrangement was adopted for mounting the spectrometer used to obtain emission spectra of the lamps (uncorrected for water transmittance). Figure 5: Arrangement of the radiometer detector during tests. Reactor Sensor Port Calibration Plane W-Diffuser Detector Mount SED240 Detector QNDS1 Filter Irradiance Calculation: The basic method for calculating the fluence rate using a radiometer when the lamp emission spectrum is known has been described previously (Linden and Darby, 1997). The method involves calculating the radiometer sensitivity factor from equation 1.

6 SF 1 = RRF(λ) P(λ) (1) λ Here, SF is the sensor factor, λ represents the wavelength, RRF is the relative response factor of the radiometer (relative to the value used on the IL1700 during measurement to convert current to irradiance), and P is the lamp probability function (i.e., the normalized emission spectrum over the range of wavelengths to which the radiometer detector responds). The fluence rate at the detector is obtained by multiplying this factor by the irradiance read off the radiometer. For these tests, additional calculations were necessary, because of the use of the QNDS1 filter, the screens in the spectrometer adapter, and alteration of the observed lamp spectra due to absorption of radiation in the water. To correct for the use of the QNDS1, the SF is altered according equation SF = (2) RRF(λ) TQNDS (λ) P(λ) λ Here, T QNDS (λ) is the percent transmittance of the QNDS1 filter at each wavelength. To correct for the screens in the spectrometer adapter, P(λ) is altered according to equation 3. P(λ) TAS(λ) P' (λ) = (3) P(λ) λ TAS(λ) Here, T AS (λ) is the percent transmittance of the adapter screens. Note that this equation defines normalized values of P (λ). The spectral distribution of the fluence rate uncorrected for the water absorbance was then calculated from equation 4. E (λ) = P' (λ) MR SF (4) Here, MR is the radiometer reading during the test. Finally, correction for the water absorbance was made using equations 5 and 6 (Morowitz, 1950). E(λ) E' (λ) = (5) S 3.8a(λ) (1 10 ) S = ln(10) 3.8 a(λ) Here, E (λ) is the fluence rate at the radiometer detector calibration plane in the absence of absorption by the water, and a(λ) is the absorbance coefficient of the water (cm -1 ). The fluence rate from 200 to 350 nm was calculated as the sum of the E (λ) values. Germicidal fluence rate, E G, was calculated by weighting the E (λ) values by the relative absorbance spectrum of DNA (with 254 nm given a weighting factor of unity) and summing according to equation 7. E G λ (6) = E'(λ)A' (λ) (5) DNA

7 In this equation, A DNA (λ) are the absorption coefficients of DNA relative to 254 nm. The DNA spectrum used in these calculations is shown in Figure 6; the spectrum from 200 to 300 nm was taken from Jagger, 1967 and above 300 nm it was fit with a gaussian curve. Figure 6: Germicidal weighting curve [from Jagger 1967] used to calculate E G. This is the DNA absorption spectrum relative to 254 nm Germicidal Weighting Factor RESULTS AND DISCUSSION Figure 7 shows how the observed spectrum of lamp 1 changed during the first 3980 h of operation. The spectra shown in the figure are corrected for all the effects described in the preceeding section and represent the spectral distribution of the fluence rate at the detector calibration plane (i.e, they are E (λ) values, the irradiance expected if the reactor were filled with a non-uv absorbing medium). There is a distinct drop in the fluence rate at all wavelengths that is most pronounced below 240 nm. Similar changes were also observed for the other three lamps. It is not possible from the present data to determine whether these changes are due solely to lamp aging or to a combination of aging and solarization of the quartz sleeves. For the purposes of these tests, such a distinction is not necessary. However, when the tests are completed the sleeve transmittance will be measured and compared to that of new sleeves to assess what fraction of the observed depreciation can be attributed to sleeve aging and what to lamp aging. Of particular interest in these tests are the rates at which germicidal, total (200 to 350 nm), UVB (280 to 315 nm), and UVC (200 to 280 nm for these lamps) fluence rates change in the reactor during full-time operation of the system. The change in UVA fluence rate cannot be adequately assessed from the data due to the limited response range of the SED240 detector (Figure 3). For lamp 1, the change in germicidal and total fluence rates is shown in Figure 8. From the figure, it is clear that the germicidal fluence rate depreciates more rapidly than the 200 to 350 nm fluence rate, consistent with the observed spectral changes (Figure 7).

8 Figure 7: Spectral changes for lamp 1 over the first 3980 hours of operation hours 3980 hours 1.4 Irradiance (mw cm -2 ) Figure 8: Change in germicidal and total fluence rate with time for lamp Total Germicidal 0.95 Relative Irradiance Hours The relative change in germicidal versus total irradiance is reflected in a more rapid decrease of the UVC fluence rate compared to the UVB fluence rate, which is shown in Figure 9. These changes are similar to previously reported trends (Schenck, 1987), but the depreciation rates observed here are much slower than expected based on the results discussed by Schenck. The reason for this discrepancy is not presently known, but the trends observed here are consistent with the manufacturer s life expectancy of 6000 h (or more) for these lamps to operate effectively for UV disinfection.

9 Figure 9: Change in UVB and UVC fluence rate with time for lamp Relative Irradiance UVC UVB Hours CONCLUSION Because few data are available in the published literature, the aging profile of UV lamps is not well understood. This ongoing study provides information on the manner in which MP UV lamps age with regard to their total output as well as changes in their spectral distribution. Over 4000 hours of investigation, the MP lamps tested decreased in irradiance by approximately 17%. Based on one previous report (Schenck, 1987), these lamps appear to be aging fairly slowly. Slightly more depreciation has been observed in the UVC than in the UVB region (18% versus 15% for the average of all lamps). Consistent with this, the germicidal fluence rate depreciated slightly more rapidly than the total (200 to 350 nm) fluence rate. In general, shorter wavelengths tended to age more rapidly, which may be due to a combination of lamp and sleeve aging. LITERATURE CITED AWWARF (American Water Works Research Foundation) (2000) Ultraviolet Disinfection Guidelines for Drinking Water and Water Reuse, Project Report #2674; American Water Works Association, Denver, CO. Linden, K.G.; Darby, J.L. (1997) J. Environ. Eng.-ASCE, 123(11): Morowitz, H. J. (1950) Science, 111: Jagger, J. H. (1967) Introduction to Research in UV Photobiology; Prentice-Hall, Inc., Englewood Cliffs, N.J. Schenck, G.O. (1987) "Ultraviolet Sterilization" in Handbook of Water Purification, 2nd ed.; Lorch, W. (Ed.); Wiley and Sons, NY.

Power consumption (kw)

Power consumption (kw) OS Series Series description LIT OS systems are applied to wastewater disinfection. The models of this series were developed on base of high-efficiency low-pressure amalgam lamps and have all components

More information

Application Note (A5)

Application Note (A5) Application Note (A5) Discrepancies Associated with Using UVA and UVB Meters to Measure Output of Sunlight, Solar Simulators, etc. Revision: A November 1991 Gooch & Housego 4632 36 th Street, Orlando,

More information

UTAH DIVISION OF DRINKING WATER UV REACTOR MONITORING AND REPORTING PROTOCOL (Last Update February 17, 2011)

UTAH DIVISION OF DRINKING WATER UV REACTOR MONITORING AND REPORTING PROTOCOL (Last Update February 17, 2011) UTAH DIVISION OF DRINKING WATER UV REACTOR MONITORING AND REPORTING PROTOCOL (Last Update February 17, 2011) This document provides clarifications and specifications on the UV monitoring and reporting

More information

UV Disinfection Systems A Series. Main Applications: Drinking Water, Process Water, Warm Water / Legionellae. IT T Industries Engineered for life

UV Disinfection Systems A Series. Main Applications: Drinking Water, Process Water, Warm Water / Legionellae. IT T Industries Engineered for life DIN EN ISO 9001 DIN EN ISO 9001 DIN EN ISO 14001 UV Disinfection Systems A Series Main Applications: Drinking Water, Process Water, Warm Water / Legionellae DIN EN ISO 14001 UV Disinfection Systems A Series

More information

COLUMBIA UNIVERSITY HEALTH SCIENCES DIVISION

COLUMBIA UNIVERSITY HEALTH SCIENCES DIVISION COLUMBIA UNIVERSITY HEALTH SCIENCES DIVISION ENVIRONMENTAL HEALTH & SAFETY Working Safely with Ultraviolet Radiation Policy and Procedures Purpose: To provide information and guidelines for the safe use

More information

Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July 2001. Reprinted from Photonics Spectra, Laurin Publishing

Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July 2001. Reprinted from Photonics Spectra, Laurin Publishing Reprint (R22) Avoiding Errors in UV Radiation Measurements By Thomas C. Larason July 2001 Reprinted from Photonics Spectra, Laurin Publishing Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1

More information

Basics of UV Disinfection

Basics of UV Disinfection Local Section Seminar Basics of UV Disinfection April 17, 2014 Joseph Moore, P.E. DuBois-Cooper Associates, Inc. 1 Presentation Outline Basics of Ultraviolet (UV) Disinfection Components of UV Disinfection

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

Sap Steady UV Unit for Maple Sap

Sap Steady UV Unit for Maple Sap Cornell Maple Bulletin 203 (2007) Sap Steady UV Unit for Maple Sap by BRIAN CHABOT Overview Components The Sap-Steady unit from FPE Inc. (Macedon NY) has been designed to kill bacteria yeast, and some

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry

Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry Models UVS-1, MoniSpec-UV & MoniSpec-UVd Model UVS-1 general information

More information

Introduction to Use of UV Light For the Control of Air Handler Contamination. What is UVGI or Ultraviolet Light?

Introduction to Use of UV Light For the Control of Air Handler Contamination. What is UVGI or Ultraviolet Light? Introduction to Use of UV Light For the Control of Air Handler Contamination By Chris Willette Since the energy crisis of the 70 s, we have been forced to build our homes tighter and tighter to save on

More information

UV, An Effective Approved Method of Disinfection

UV, An Effective Approved Method of Disinfection UV, An Effective Approved Method of Disinfection INTRODUCTION Ultraviolet (UV) irradiation is a powerful technology that has been successfully deployed in several diverse industries such as pharmaceutical,

More information

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION G. Thuillier1, D. Bolsee2 1 LATMOS-CNRS, France 2 Institut d Aéronomie Spatiale

More information

0.00 18 TO18 0.00 _ nothing special M 0.20 49.00 A UVA 24.00 18ISO90 TO18 6.00 Lens concentr. Lens D 0.50 49.00 B UVB 24.

0.00 18 TO18 0.00 _ nothing special M 0.20 49.00 A UVA 24.00 18ISO90 TO18 6.00 Lens concentr. Lens D 0.50 49.00 B UVB 24. Basic Information That guide assists you selecting the right ilicon Carbide (ic) based UV photodiode for your application. Basically this selection is between active area, spectral behaviour, packaging

More information

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007 PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow

More information

T5 Watt-Miser Linear Fluorescent lamps

T5 Watt-Miser Linear Fluorescent lamps GE Lighting T5 Watt-Miser Linear Fluorescent lamps DATA SHEET T5 Watt-Miser High Efficiency 13W, 20W, 26W, 33W T5 Watt-Miser High Output 21W, 36W, 46W, 51W, 76W Product information T5 Watt-Miser lamps

More information

AQUADA UV Eliminate Bacteria in Your Drinking Water!

AQUADA UV Eliminate Bacteria in Your Drinking Water! AQUADA UV Eliminate Bacteria in Your Drinking Water! Protect your Family from Micro-Organisms Legionella pneumophila Legionella pneumophila even extremely resistant bacteria are safely destroyed in your

More information

Ultra-Violet Products Ltd 2066 W 11th Street, Upland, CA 91786 Unit 1, Trinity Hall Farm Estate

Ultra-Violet Products Ltd 2066 W 11th Street, Upland, CA 91786 Unit 1, Trinity Hall Farm Estate UV/Air Circulator For the UV PCR Workstation User Instructions UVP, LLC Ultra-Violet Products Ltd 2066 W 11th Street, Upland, CA 91786 Unit 1, Trinity Hall Farm Estate Tel: (800) 452-6788 / (909) 946-3197

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

Measuring Laser Power and Energy Output

Measuring Laser Power and Energy Output Measuring Laser Power and Energy Output Introduction The most fundamental method of checking the performance of a laser is to measure its power or energy output. Laser output directly affects a laser s

More information

Understanding UV Monitoring for Air and Water UV Treatments José Goin,* Anatoly Skirda and Eugene Tohktuev INTRODUCTION

Understanding UV Monitoring for Air and Water UV Treatments José Goin,* Anatoly Skirda and Eugene Tohktuev INTRODUCTION Understanding UV Monitoring for Air and Water UV Treatments José Goin,* Anatoly Skirda and Eugene Tohktuev Apprise Technologies, 4802 Oneota Street, Duluth, MN, 55807 * Corresponding author: jgoin@apprisetech.com

More information

UVC LEDs for. Environmental Monitoring

UVC LEDs for. Environmental Monitoring UVC LEDs for Environmental Monitoring Environmental monitoring relies extensively on molecular spectroscopy for tracking air quality, water and wastewater quality, and detecting hazardous substances. Advances

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

Project 2B Building a Solar Cell (2): Solar Cell Performance

Project 2B Building a Solar Cell (2): Solar Cell Performance April. 15, 2010 Due April. 29, 2010 Project 2B Building a Solar Cell (2): Solar Cell Performance Objective: In this project we are going to experimentally measure the I-V characteristics, energy conversion

More information

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control

More information

Duron UV System YoU adjust to changing conditions it S time YoUr UV SYStem DoeS too

Duron UV System YoU adjust to changing conditions it S time YoUr UV SYStem DoeS too Duron UV System You adjust to changing conditions It s time your UV system does too Your operating conditions change. Duron UV System adjusts. Combining intelligent control with the most advanced UV lamp

More information

2 Absorbing Solar Energy

2 Absorbing Solar Energy 2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could

More information

Steps in the design process. The draft Ultraviolet Disinfection Guidance Manual (USEPA, 2003) recommends the following steps in the design:

Steps in the design process. The draft Ultraviolet Disinfection Guidance Manual (USEPA, 2003) recommends the following steps in the design: DESIGN OF UV DISINFECTION SYSTEMS FOR DRINKING WATER In EPA s economic analyses for the proposed LT2ESWTR, EPA estimated that 500 to 1000 filtration plants will choose UV disinfection as part of their

More information

Electromagnetic Radiation (EMR) and Remote Sensing

Electromagnetic Radiation (EMR) and Remote Sensing Electromagnetic Radiation (EMR) and Remote Sensing 1 Atmosphere Anything missing in between? Electromagnetic Radiation (EMR) is radiated by atomic particles at the source (the Sun), propagates through

More information

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this

More information

T5 LongLast. GE Lighting DATA SHEET

T5 LongLast. GE Lighting DATA SHEET GE Lighting T5 LongLast DATA SHEET Linear Fluorescent Lamps T5 LongLast High Efficiency 14W, 21W, 28W, 35W T5 LongLast High Output 24W, 39W, 49W, 54W, 8W Product information T5 LongLast lamps are triphosphor

More information

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment.

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment. Helium-Neon Laser Experiment objectives: assemble and align a 3-mW HeNe laser from readily available optical components, record photographically the transverse mode structure of the laser output beam,

More information

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements. T e c h n i c a l N o t e Reflectance Measurements of Materials Used in the Solar Industry UV/Vis/NIR Author: Dr. Jeffrey L. Taylor PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA Selecting

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

The Calculation of G rms

The Calculation of G rms The Calculation of G rms QualMark Corp. Neill Doertenbach The metric of G rms is typically used to specify and compare the energy in repetitive shock vibration systems. However, the method of arriving

More information

Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V.

Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V. Location: Contact: 1240 Hach Hall Steve Veysey, 1234 Hach Hall Safety All researchers working

More information

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters

More information

Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument

Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument Technical Note INTRODUCTION Direct measurements of nucleic acid samples at OD 260 or protein samples at OD

More information

Introduction to Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectrometry Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,

More information

A Look at Accelerated Photostability Testing for Packaged Food and Drinks

A Look at Accelerated Photostability Testing for Packaged Food and Drinks A Look at Accelerated Photostability Testing for Packaged Food and Drinks By Dr. Oliver Rahäuser and Dr. Artur Schönlein Atlas Material Testing Technology GmbH Vogelsbergstr. 22, 63589 Linsengericht-Altenhaßlau,

More information

Most of this article will be about the lowpressure mercury vapor UV lamps and how

Most of this article will be about the lowpressure mercury vapor UV lamps and how 1-26-06 UV Lights & Lamps - A Manufacturer s Perspective By Don Newsome, FMS #14 President, UV SYSTEMS, Inc. Definition of terms. For clarity, this article uses the engineering terms lamp and light or

More information

Measurement of RF Emissions from a Final Coat Electronics Corrosion Module

Measurement of RF Emissions from a Final Coat Electronics Corrosion Module Engineering Test Report No. 37802-02 Rev. A Measurement of RF Emissions from a Final Coat Electronics Corrosion Module For : Canadian Auto Preservation, Inc. 390 Bradwick Drive Concord, Ontario CANADA

More information

Mission Hill Water Treatment Plant & Kalamalka Lake Pump Station Upgrades

Mission Hill Water Treatment Plant & Kalamalka Lake Pump Station Upgrades Mission Hill Water Treatment Plant Treatment Plant Pilot Testing Regional District of North Okanagan, assisted by Sandwell Engineering Services and four manufacturers conducted pilot scale water treatment

More information

Measuring of optical output and attenuation

Measuring of optical output and attenuation Measuring of optical output and attenuation THEORY Measuring of optical output is the fundamental part of measuring in optoelectronics. The importance of an optical power meter can be compared to an ammeter

More information

Laser diffuse reflection light scanner with background suppression. Dimensioned drawing

Laser diffuse reflection light scanner with background suppression. Dimensioned drawing Specifications and description HRTL 3B Laser diffuse reflection light scanner with background suppression Dimensioned drawing We reserve the right to make changes DS_HRTL3B_en.fm en 01-2010/12 50114049

More information

Hello and Welcome to this presentation on LED Basics. In this presentation we will look at a few topics in semiconductor lighting such as light

Hello and Welcome to this presentation on LED Basics. In this presentation we will look at a few topics in semiconductor lighting such as light Hello and Welcome to this presentation on LED Basics. In this presentation we will look at a few topics in semiconductor lighting such as light generation from a semiconductor material, LED chip technology,

More information

Today, healthcare facilities must be prepared to face

Today, healthcare facilities must be prepared to face Alternate Methods for USING GERMICIDAL UV-C Disinfection Lamps TO SUPPLEMENT HEPA Filtration By Dave Shagott Today, healthcare facilities must be prepared to face potential large-scale threats from chemical,

More information

D-39740-2011. Next-generation LED phototherapy NanoBlu 500

D-39740-2011. Next-generation LED phototherapy NanoBlu 500 D-39740-2011 Next-generation LED phototherapy NanoBlu 500 02 Is there a powerful phototherapy device that s really compact? D-3534-2011 03 BETTER TREaTMENTS SuPPoRT BETTER outcomes The effective treatment

More information

The Physics of Energy sources Renewable sources of energy. Solar Energy

The Physics of Energy sources Renewable sources of energy. Solar Energy The Physics of Energy sources Renewable sources of energy Solar Energy B. Maffei Bruno.maffei@manchester.ac.uk Renewable sources 1 Solar power! There are basically two ways of using directly the radiative

More information

Quantifying Bacterial Concentration using a Calibrated Growth Curve

Quantifying Bacterial Concentration using a Calibrated Growth Curve BTEC 4200 Lab 2. Quantifying Bacterial Concentration using a Calibrated Growth Curve Background and References Bacterial concentration can be measured by several methods, all of which you have studied

More information

Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2

Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2 Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2 D. K. McDermitt, J. M. Welles, and R. D. Eckles - LI-COR, inc. Lincoln, NE 68504 USA Introduction Infrared analysis

More information

Copyright 1999 2010 by Mark Brandt, Ph.D. 12

Copyright 1999 2010 by Mark Brandt, Ph.D. 12 Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits

More information

ConstantColor CMH MR16

ConstantColor CMH MR16 GE Lighting ConstantColor CMH MR16 Reflector Ceramic Metal Halide Lamps W and 35W DATA SHEET Product information ConstantColor CMH lamps combine HPS technology (providing stability, efficiency & uniformity)

More information

5.5. San Diego (8/22/03 10/4/04)

5.5. San Diego (8/22/03 10/4/04) NSF UV SPECTRORADIOMETER NETWORK 23-24 OPERATIONS REPORT 5.5. San Diego (8/22/3 1/4/4) The 23-24 season at San Diego includes the period 8/22/3 1/4/4. In contrast to other network sites, San Diego serves

More information

Department of Engineering Enzo Ferrari University of Modena and Reggio Emilia

Department of Engineering Enzo Ferrari University of Modena and Reggio Emilia Department of Engineering Enzo Ferrari University of Modena and Reggio Emilia Object: Measurement of solar reflectance, thermal emittance and Solar Reflectance Index Report Reference person: Alberto Muscio

More information

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Guaranteed specifications Design overview Double beam, ratio recording, double out-of-plane Littrow monochromator UV-Vis-NIR spectrophotometer (Agilent Cary

More information

The photoionization detector (PID) utilizes ultraviolet

The photoionization detector (PID) utilizes ultraviolet Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).

More information

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD TN-100 The Fundamentals of Infrared Spectroscopy The Principles of Infrared Spectroscopy Joe Van Gompel, PhD Spectroscopy is the study of the interaction of electromagnetic radiation with matter. The electromagnetic

More information

Ultraviolet Spectroscopy

Ultraviolet Spectroscopy Ultraviolet Spectroscopy The wavelength of UV and visible light are substantially shorter than the wavelength of infrared radiation. The UV spectrum ranges from 100 to 400 nm. A UV-Vis spectrophotometer

More information

INFRARED PARTS MANUAL

INFRARED PARTS MANUAL INFRARED PARTS MANUAL PIR325 FL65 GLOLAB CORPORATION Thank you for buying our Pyroelectric Infrared components. The goal of Glolab is to produce top quality electronic kits, products and components. All

More information

Tech Bulletin. Understanding Solar Performance

Tech Bulletin. Understanding Solar Performance Tech Bulletin Understanding Solar Performance Bekaert solar control window films use advanced technology to benefit consumers with quality solutions that enhance comfort and decrease energy use. By understanding

More information

Sound Power Measurement

Sound Power Measurement Sound Power Measurement A sound source will radiate different sound powers in different environments, especially at low frequencies when the wavelength is comparable to the size of the room 1. Fortunately

More information

Ces deux analyseurs permettent le contrôle de la qualité de l eau et des effluents.

Ces deux analyseurs permettent le contrôle de la qualité de l eau et des effluents. OptiLIS 201 & 301 Online multi-parameter water analyzers These two standalone analyzers enable you to control continuously both fresh water and effluents quality Ces deux analyseurs permettent le contrôle

More information

GE ConstantColor CMH Precise MR16 20 & 35W

GE ConstantColor CMH Precise MR16 20 & 35W GE Lighting GE ConstantColor CMH Precise MR16 2 & 35W DATA SHEE T TECHNOLOGY ConstantColor CMH lamps combine HPS technology (providing stability, efficiency & uniformity) and Metal Halide Technology (providing

More information

Fig.1. The DAWN spacecraft

Fig.1. The DAWN spacecraft Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics

More information

Application Note AN1

Application Note AN1 TAKING INVENTIVE STEPS IN INFRARED. MINIATURE INFRARED GAS SENSORS GOLD SERIES UK Patent App. No. 799A USA Patent App. No. 9/78,7 World Patents Pending SENSOR OVERVIEW Application Note AN The Dynament

More information

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in

More information

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation? From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation? From lowest energy to highest energy, which of the following correctly

More information

2 Spectrophotometry and the Analysis of Riboflavin

2 Spectrophotometry and the Analysis of Riboflavin 2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound

More information

TYPE 65UV5-1004 CEX Integrated Flame Scanner with Internal Flame Relay and Analog Output

TYPE 65UV5-1004 CEX Integrated Flame Scanner with Internal Flame Relay and Analog Output 133-683 JUNE 28, 2006 TYPE 65UV5-1004 CEX Integrated Flame Scanner with Internal Flame Relay and Analog Output DESCRIPTION APPLICATION The Fireye 65UV5-1004 flame scanner is a microprocessor based flame

More information

CONSERVATION AND LIGHTING

CONSERVATION AND LIGHTING CONSERVATION AND LIGHTING Light is essential for the examination and enjoyment of collection items. But in a museum light also means damage: dyes and pigments fade or change appearance and the materials

More information

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

WAVELENGTH OF LIGHT - DIFFRACTION GRATING PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant

More information

Nucleic Acid Purity Assessment using A 260 /A 280 Ratios

Nucleic Acid Purity Assessment using A 260 /A 280 Ratios Nucleic Acid Purity Assessment using A 260 /A 280 Ratios A common practice in molecular biology is to perform a quick assessment of the purity of nucleic acid samples by determining the ratio of spectrophotometric

More information

Applications. Architectural

Applications. Architectural GE Lighting Arcstream Single Ended Metal Halide lamps 7W and 15W DATA SHEET Product information Arcstream Single Ended lamp consists of a compact high pressure metal halide discharge operating in a quartz

More information

INFRARED SPECTROSCOPY (IR)

INFRARED SPECTROSCOPY (IR) INFRARED SPECTROSCOPY (IR) Theory and Interpretation of IR spectra ASSIGNED READINGS Introduction to technique 25 (p. 833-834 in lab textbook) Uses of the Infrared Spectrum (p. 847-853) Look over pages

More information

Calibration of a High Dynamic Range, Low Light Level Visible Source

Calibration of a High Dynamic Range, Low Light Level Visible Source Calibration of a High Dynamic Range, Low Light Level Visible Source Joe LaVeigne a, Todd Szarlan a, Nate Radtke a a Santa Barbara Infrared, Inc., 30 S. Calle Cesar Chavez, #D, Santa Barbara, CA 93103 ABSTRACT

More information

Single-phase (220...240 V) voltage monitoring: Undervoltage Overvoltage Window mode (overvoltage + undervoltage) Voltage fault memory selectable

Single-phase (220...240 V) voltage monitoring: Undervoltage Overvoltage Window mode (overvoltage + undervoltage) Voltage fault memory selectable Features 70.11 70.31 70.41 Electronic voltage monitoring relays for single and three-phase applications Multifunctional types, providing the flexibility of monitoring Undervoltage, Overvoltage, Window

More information

Alfa Laval PureBallast 3.1

Alfa Laval PureBallast 3.1 Alfa Laval PureBallast 3.1 Ballast water treatment system 1000 m 3 /h system Alfa Laval PureBallast 3.1 is an automated inline treatment system for the biological disinfection of ballast water. Operating

More information

Using the Spectrophotometer

Using the Spectrophotometer Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to

More information

Experiment #5: Qualitative Absorption Spectroscopy

Experiment #5: Qualitative Absorption Spectroscopy Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions

More information

What is Solar Control?

What is Solar Control? A better environment inside and out. Solar, Safety and Security Window Films: Tech Bulletin Understanding Solar Performance Solar Gard solar control window films use advanced technology to benefit consumers

More information

FTIR Instrumentation

FTIR Instrumentation FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation

More information

Solar ultraviolet-b radiation monitoring in Khorram Abad city in Iran

Solar ultraviolet-b radiation monitoring in Khorram Abad city in Iran Solar ultraviolet-b radiation monitoring in Khorram Abad city in Iran M. Gholami * and L. Yoosefi Iran. J. Radiat. Res., 2009; 7 (3): 171-175 Department of Medical Physics, School of Medicine, Lorestan

More information

UNDERSTANDING AND CONTROLLING COMMON-MODE EMISSIONS IN HIGH-POWER ELECTRONICS

UNDERSTANDING AND CONTROLLING COMMON-MODE EMISSIONS IN HIGH-POWER ELECTRONICS Page 1 UNDERSTANDING AND CONTROLLING COMMON-MODE EMISSIONS IN HIGH-POWER ELECTRONICS By Henry Ott Consultants Livingston, NJ 07039 (973) 992-1793 www.hottconsultants.com hott@ieee.org Page 2 THE BASIC

More information

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions Description The LS-1 Series of tungsten halogen light sources are versatile, white-light lamps optimized for use in the

More information

instruments Analytical Instruments for Science

instruments Analytical Instruments for Science instruments Analytical Instruments for Science instruments Contents PAGE NO. Introduction 4 T60 UV-Vis Spectrophotometer 6 T70 UV-Vis Spectrophotometer 10 T80 UV-Vis Spectrophotometer 14 T90+ UV-Vis Spectrophotometer

More information

We bring quality to light. ISP Series Integrating Spheres

We bring quality to light. ISP Series Integrating Spheres We bring quality to light. ISP Series Integrating Spheres Theory ISP-Series ISP 75 / ISP 100 / ISP 150L ISP 250 ISP 500 ISP 1000 ISP 2000 The Integrating Sphere The working principle Using integrating

More information

Experience. The Atlas Difference.

Experience. The Atlas Difference. XENON TEST INSTRUMENTS Lightfastness Weatherfastness Photostability Experience. The Atlas Difference. WORLD S MOST USED FL ATBED WEATHERING INSTRUMENTS Weathering is the adverse response of a material

More information

Airborne Sound Insulation

Airborne Sound Insulation Airborne Sound Insulation with XL2-TA Sound Level Meter This application note describes the verification of the airborne sound insulation in buildings with the XL2-TA Sound Level Meter. All measurements

More information

FIBRE-OPTICS POWER METER INSTRUCTION MANUAL

FIBRE-OPTICS POWER METER INSTRUCTION MANUAL FIBRE-OPTICS POWER METER INSTRUCTION MANUAL FIBRE-OPTICS POWER METER ELLMAX ELECTRONICS LTD. Unit 29, Leyton Business Centre, Etloe Road, Leyton, London E10 7BT, England Tel: 020-8539 0136 Fax: 020-8539

More information

OPERATING PRINCIPLES FOR PHOTOELECTRIC SENSORS

OPERATING PRINCIPLES FOR PHOTOELECTRIC SENSORS OPERATING PRINCIPLES FOR PHOTOELECTRIC SENSORS Photoelectric These sensors use light sensitive elements to detect objects and are made up of an emitter (light source) and a receiver. Four types of photoelectric

More information

Ultraviolet (UV) Radiation Safety

Ultraviolet (UV) Radiation Safety Ultraviolet (UV) Radiation Safety April 2005 Compiled by Myung Chul Jo Environmental Health and Safety University of Nevada Reno Page 1 of 8 Table of Contents 1. UVRadiation 3 2. Common sources of UV radiation

More information

Light penetration into the soybean canopy and its influence on soybean rust infection

Light penetration into the soybean canopy and its influence on soybean rust infection Light penetration into the soybean canopy and its influence on soybean rust infection Jim Marois David Wright, Sheeja George, Pratibha Srivastava, Phil Harmon, Carrie Harmon, Heather Young (UF) Andy Schuerger

More information

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003)

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003) (Part # S-LIB-M003) The smart sensor is designed to work with the HOBO Weather Station logger. The smart sensor has a plug-in modular connector that allows it to be added easily to a HOBO Weather Station.

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

Single Channel Loop Detector

Single Channel Loop Detector Single Channel Loop Detector Model - LD100 Series The LD100 is a single channel inductive loop detector designed for parking and access control applications. The detector is connected to an inductive loop

More information

Client: RENOLIT Italia Srl. Work completion: 01/10/2008 (or after receipt of the weathered samples) Notes: Preliminary Report

Client: RENOLIT Italia Srl. Work completion: 01/10/2008 (or after receipt of the weathered samples) Notes: Preliminary Report Object: laboratory measurement on alkorplan 35276 Client: RENOLIT Italia Srl Reference person: Prof. Paolo Tartarini Work start: 01/08/2008 Work completion: 01/10/2008 (or after receipt of the weathered

More information

Fibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques

Fibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques Fibre Bragg Grating Sensors An ntroduction to Bragg gratings and interrogation techniques Dr Crispin Doyle Senior Applications Engineer, Smart Fibres Ltd. 2003 1) The Fibre Bragg Grating (FBG) There are

More information