Distinguish two types of uncertainty evaluation Do a design stage uncertainty analysis Calculate error of a derived variable Handle uncertainty/error

Size: px
Start display at page:

Download "Distinguish two types of uncertainty evaluation Do a design stage uncertainty analysis Calculate error of a derived variable Handle uncertainty/error"

Transcription

1 ME343 Chapter 5

2 Distinguish two types of uncertainty evaluation Do a design stage uncertainty analysis Calculate error of a derived variable Handle uncertainty/error propagation from multiple sources

3 Use available information to estimate the uncertainty range, ux, in the interval estimate of the true value: For current analysis, assume that the true value, x is constant (time independent.

4 Statistical Uncertainty analysis (Type A in NIST terminology based on series of repeated measurements, and calibration tests Judgment based Uncertainty Analysis (Type B in NIST terminology based on previous measurement data, experience with relevant materials and instruments, manufacturer's specifications, data provided in calibration and other reports, and uncertainties assigned to reference data taken from handbooks

5

6 Design stage Calibration stage Data acquisition stage Data reduction stage

7 We apply design-stage uncertainty analysis when we don t have the system yet, or have not tested it yet. We either rely on the manufacturer s data, or on our own estimates Major facilities may need to be built and equipment ordered with considerable lead time.

8 It is useful for selecting instruments, selecting measurement techniques, and obtaining an appropriate estimate of the uncertainty likely to exist in the measured data. At this point, the measurement system and associated procedures are but a concept. Usually little is known about the instruments, and in many cases they are still pictures in a catalog.

9 Design-Stage Uncertainty is made up of two principle components. 1. Instrument resolution. Instrument uncertainty, uc, (an estimate of the systematic error for the instrument

10

11

12

13

14

15

16

17

18

19 Treat an error as: Precision error if it can be statistically estimated in some manner (data scatter Bias error otherwise (cannot be directly discerned by statistical methods, e.g. By repeating measurements.

20 Calibration Alternate (concomitant Method Interlaboratory comparisons Experience

21 System repeatability and resolution (measurement system Temporal and spatial variations (measured variable properties Operating and environmental conditions (process Measurement repeatability (measurement procedure and Technique

22 A temperature measurement system is composed of a sensor and a readout devices. The readout device has a claimed accuracy of 0.6 C with a resolution of 0.1 C. the sensor has an off-the-shelf accuracy of 0.5 C. Estimate a design-stage uncertainty in temperature indicated by this combination. The design-stage uncertainty is for the combined system is

23 where (ud R is the design-stage uncertainty of the readout device and (uds is that of the sensor. In either case, the individual design-stage uncertainty is found from Sensor

24

25 The linear displacement of a vehicle due to an applied impact force is measured with a transducer. Transducer specifications are: Input range : 0-5 m Output range: 0-5 V Linearity: ± 0.5% reading Drift: ± 0.05%/ C FSO Sensitivity: ± 0.10% FSO Hysteresis: ± 0.5% reading The transducer output is indicated on a voltmeter (accuracy: within ± 0.1% reading; resolution: 10μV. The expected nominal displacement of the vehicle is to be 4m for an impact force of 000± 100N (95%.

26 The force-displacement relation can be assumed linear. Four replications consisting of 10 measurements are made over the course of one day. The results are given with the ambient temperature for each run. Test run N Mean value [m]

27 At the design-stage, only information known prior to the test are included: For the transducer, we estimate the instrument error from the elemental errors: e L = ± (0.005(4m = 0.01 m e H = ± (0.005(4m = 0.01 m e s = ± (0.001(5m = m e D = ± (0.005/ C(3 C(5m = m so that,

28 The transducer has a sensitivity Kt = 1V/m so that the voltmeter output can be restated in terms of displacement [m]. For the voltmeter: For the transducer-voltmeter system, the designstage uncertainty is:

29 Now we look at the case of a derived quantity that is estimated from the measurement of several primary quantities. The question that needs to be answered is the following: A derived quantity Q is estimated using a formula that involves the primary quantities. a1,a,...an Each one of these is available in terms of the respective best values b1, b,...bn and the respective standard deviations σ1,σ...σn. What is the best estimate for Q and what is the corresponding standard deviation σq?

30 We have, by definition Q =Q(a1,a,...an It is obvious that the best value of Q should correspond to that obtained by using the best values for the a s. Thus, the best estimate for Q given by Q Q( b, b, b3,... b 1 n by definition, we should have:

31 The subscript i indicates the experiment number and the ith estimate of Q is given by Qi = Q(a1i,ai,...ani using a Taylor expansion around the best value as where the partial derivatives are all evaluated at the best values for the a s. If the a s are all independent of one another then the errors in these are unrelated to one another and hence the cross terms.

32 The equation can be written as: Noting that the above equation can be in the form

33 the error propagation formula

34 Two resistors are to be combined to form an equivalent resistance of Readily available are two common resistors rated at and two common resistors rated at 0005%. What combination of resistors (series or parallel would provide the smaller uncertainty in an equivalent 1000 resistance? Case 1 (series resistors R T = R 1 + R If we use the two 500 resistors, then u d 1 The (u 50, d propagation of series = [( = u d R R [(1.u 50 T 1 d1 uncertainty through to u d1 R ( R (1.u d T ] ] 1 71 R T is estimated (95% by

35 Case (paralell resistors If we use the two 000 resistors, then u d 1 The (u 100, d propagation of paralell RT = [( R = [({ u R 35 d uncertainty through to 1 R u R d1 (95% R ( R R1R ( R R 1 T ] 1 }.u d 1 R T is ({ estimated R 1 R 1 R by R1R ( R R 1 }u d ] 1

36 Heat transfer from a rod of diameter D immersed in a fluid can be described by the Nusselt number Nu=hD/k, where h is the heat transfer coefficient and k is the thermal conductivity of the fluid. If h can be measured within 7% (95%, estimate the uncertainty in Nu for the nominal value of h=150 W/m-K. Let D=00.5 mm and k=0.6% W/m-K. u h = (0.07(150 = 10.5 W/m -K u D = m u k = (0.0(0.6 = 0.01 W/m-K We have Nu = f(h,d,k

37 Then, we estimate the Nusselt number here to be Nu = hd/k u Nu = (95% So, Nusselt number can be determined within about 8% in this range of values x0.005 ( ( ( u u k hd ( u k h ( u k D ( u u k N ( u D N ( u h N ( u 1 Nu 1 k D h Nu 1 k u d u h u Nu

Basic RTD Measurements. Basics of Resistance Temperature Detectors

Basic RTD Measurements. Basics of Resistance Temperature Detectors Basic RTD Measurements Basics of Resistance Temperature Detectors Platinum RTD resistances range from about 10 O for a birdcage configuration to 10k O for a film type, but the most common is 100 O at 0

More information

Calibrating DC Shunts: Techniques and Uncertainties

Calibrating DC Shunts: Techniques and Uncertainties Calibrating DC Shunts: Techniques and Uncertainties Jay Klevens Ohm-Labs, Inc. 611 E. Carson St., Pittsburgh, PA 15203 (412) 431-0640 jklevens@ohm-labs.com Abstract Accurate electrical current measurement

More information

Module 1 : Conduction. Lecture 5 : 1D conduction example problems. 2D conduction

Module 1 : Conduction. Lecture 5 : 1D conduction example problems. 2D conduction Module 1 : Conduction Lecture 5 : 1D conduction example problems. 2D conduction Objectives In this class: An example of optimization for insulation thickness is solved. The 1D conduction is considered

More information

Calibration of Dallas sensors

Calibration of Dallas sensors Calibration of Dallas sensors Mariusz Sapinski INFN Sezione di Roma1 Roma, Italy (April 2006) 1. Objectives The goal of this work is to perform a calibration of Dallas sensors. Nine Dallas sensors are

More information

7. DYNAMIC LIGHT SCATTERING 7.1 First order temporal autocorrelation function.

7. DYNAMIC LIGHT SCATTERING 7.1 First order temporal autocorrelation function. 7. DYNAMIC LIGHT SCATTERING 7. First order temporal autocorrelation function. Dynamic light scattering (DLS) studies the properties of inhomogeneous and dynamic media. A generic situation is illustrated

More information

Environmental Monitoring with Sensors: Hands-on Exercise

Environmental Monitoring with Sensors: Hands-on Exercise Environmental Monitoring with Sensors: Hands-on Exercise Now that you ve seen a few types of sensors, along with some circuits that can be developed to condition their responses, let s spend a bit of time

More information

Series-Parallel Circuits. Objectives

Series-Parallel Circuits. Objectives Series-Parallel Circuits Objectives Identify series-parallel configuration Analyze series-parallel circuits Apply KVL and KCL to the series-parallel circuits Analyze loaded voltage dividers Determine the

More information

UNCERTAINTY AND CONFIDENCE IN MEASUREMENT

UNCERTAINTY AND CONFIDENCE IN MEASUREMENT Ελληνικό Στατιστικό Ινστιτούτο Πρακτικά 8 ου Πανελληνίου Συνεδρίου Στατιστικής (2005) σελ.44-449 UNCERTAINTY AND CONFIDENCE IN MEASUREMENT Ioannis Kechagioglou Dept. of Applied Informatics, University

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

Using NTC Temperature Sensors Integrated into Power Modules

Using NTC Temperature Sensors Integrated into Power Modules Using NTC Temperature Sensors Integrated into Power Modules Pierre-Laurent Doumergue R&D Engineer Advanced Power Technology Europe Chemin de Magret 33700 Mérignac, France Introduction Most APTE (Advanced

More information

THE STRAIN GAGE PRESSURE TRANSDUCER

THE STRAIN GAGE PRESSURE TRANSDUCER THE STRAIN GAGE PRESSURE TRANSDUCER Pressure transducers use a variety of sensing devices to provide an electrical output proportional to applied pressure. The sensing device employed in the transducers

More information

Experimental test preparation, performance and data processing

Experimental test preparation, performance and data processing Chapter 4 Experimental test preparation, performance and data processing 4.1 Introduction This chapter is dedicated to the experimental test preparation, performance, and data processing and analysis.

More information

Experimental Uncertainties (Errors)

Experimental Uncertainties (Errors) Experimental Uncertainties (Errors) Sources of Experimental Uncertainties (Experimental Errors): All measurements are subject to some uncertainty as a wide range of errors and inaccuracies can and do happen.

More information

Determining Measurement Uncertainty for Dimensional Measurements

Determining Measurement Uncertainty for Dimensional Measurements Determining Measurement Uncertainty for Dimensional Measurements The purpose of any measurement activity is to determine or quantify the size, location or amount of an object, substance or physical parameter

More information

Typical Aluminum GFM Mass Flow Meter

Typical Aluminum GFM Mass Flow Meter MASS FLOW METERS Design Features n Rigid metallic construction. n Maximum pressure of 1000 psig (70 bars). n Leak integrity 1 x 10-9 of helium. n NIST traceable certification. n Built-in tiltable LCD readout.

More information

BJT AC Analysis. by Kenneth A. Kuhn Oct. 20, 2001, rev Aug. 31, 2008

BJT AC Analysis. by Kenneth A. Kuhn Oct. 20, 2001, rev Aug. 31, 2008 by Kenneth A. Kuhn Oct. 20, 2001, rev Aug. 31, 2008 Introduction This note will discuss AC analysis using the beta, re transistor model shown in Figure 1 for the three types of amplifiers: common-emitter,

More information

Uncertainty Estimation and Calculation. Gerald Recktenwald Portland State University Department of Mechanical Engineering gerry@me.pdx.

Uncertainty Estimation and Calculation. Gerald Recktenwald Portland State University Department of Mechanical Engineering gerry@me.pdx. Uncertainty Estimation and Calculation Gerald Recktenwald Portland State University Department of Mechanical Engineering gerry@me.pdx.edu These slides are a supplement to the lectures in ME 449/549 Thermal

More information

(Refer Slide Time 1:13 min)

(Refer Slide Time 1:13 min) Mechanical Measurements and Metrology Prof. S. P. Venkateshan Department of Mechanical Engineering Indian Institute of Technology, Madras Module - 1 Lecture - 1 Introduction to the Study of Mechanical

More information

APPLICATION NOTE. Measuring Current Output Transducers with Campbell Scientific Dataloggers. App. Note Code: 2MI-B Revision: 1

APPLICATION NOTE. Measuring Current Output Transducers with Campbell Scientific Dataloggers. App. Note Code: 2MI-B Revision: 1 App. Note Code: 2MI-B Revision: 1 APPLICATION NOTE Measuring Current Output s with Campbell Scientific Dataloggers 815 W. 1800 N. Logan, Utah 84321-1784 (435) 753-2342 FAX (435) 750-9540 Copyright (C)

More information

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS XVII IMEKO World Congress Metrology in the 3rd Millennium June 22 27, 2003, Dubrovnik, Croatia LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS Tanfer Yandayan TUBITAK-Ulusal

More information

FIELD TESTS OF FUEL EFFICIENCY MAGNETS J A CRABB JULY 1997 SWEEG REPORT 80

FIELD TESTS OF FUEL EFFICIENCY MAGNETS J A CRABB JULY 1997 SWEEG REPORT 80 FIELD TESTS OF FUEL EFFICIENCY MAGNETS J A CRABB JULY 1997 SWEEG REPORT 80 List of symbols T f average temperature at the flow header, C T r average temperature at the return header, C f average water

More information

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES TSI and TSI logo are registered trademarks of TSI Incorporated. SmartTune is a trademark of TSI Incorporated. THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES IFA 300 Constant Temperature Anemometry

More information

A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW. 1998 ASME Fluids Engineering Division Summer Meeting

A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW. 1998 ASME Fluids Engineering Division Summer Meeting TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW Proceedings of FEDSM 98: June -5, 998, Washington, DC FEDSM98 49 ABSTRACT The pressure

More information

AP Physics 1 and 2 Lab Investigations

AP Physics 1 and 2 Lab Investigations AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks

More information

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration

More information

EXPERIMENTAL ANALYSIS OF HEAT TRANSFER ENHANCEMENT IN A CIRCULAR TUBE WITH DIFFERENT TWIST RATIO OF TWISTED TAPE INSERTS

EXPERIMENTAL ANALYSIS OF HEAT TRANSFER ENHANCEMENT IN A CIRCULAR TUBE WITH DIFFERENT TWIST RATIO OF TWISTED TAPE INSERTS INTERNATIONAL JOURNAL OF HEAT AND TECHNOLOGY Vol.33 (2015), No.3, pp.158-162 http://dx.doi.org/10.18280/ijht.330324 EXPERIMENTAL ANALYSIS OF HEAT TRANSFER ENHANCEMENT IN A CIRCULAR TUBE WITH DIFFERENT

More information

Case study Realistic and best uncertainties with humidity instruments

Case study Realistic and best uncertainties with humidity instruments Case study Realistic and best uncertainties with humidity instruments Domen Hudoklin September 2014 Brdo, Slovenia Humidity calibrations Dewpoint primary - against dewpoint generator secondary - against

More information

M&V of Hot Water Boiler Plant. Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd

M&V of Hot Water Boiler Plant. Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd M&V of Hot Water Boiler Plant Boban Ratkovich, P. Eng, CEM, BESA, LEED AP President CES Engineering Ltd AIA Quality Assurance Learning Objectives 1. Case study examples of procedures and obstacles in measuring

More information

Section 3. Sensor to ADC Design Example

Section 3. Sensor to ADC Design Example Section 3 Sensor to ADC Design Example 3-1 This section describes the design of a sensor to ADC system. The sensor measures temperature, and the measurement is interfaced into an ADC selected by the systems

More information

Estimates of Uncertainty of the Calibration of Balances

Estimates of Uncertainty of the Calibration of Balances Estimates of Uncertainty of the Calibration of Balances Chart shows the uncertainty (estimates) of the calibration of representative electronic balances. Chart. Uncertainty of calibration (estimates) Minimum

More information

Natural Convection. Buoyancy force

Natural Convection. Buoyancy force Natural Convection In natural convection, the fluid motion occurs by natural means such as buoyancy. Since the fluid velocity associated with natural convection is relatively low, the heat transfer coefficient

More information

R&D Engineer. equipment. the power

R&D Engineer. equipment. the power Application Note APT0406 Using NTC Temperature sensor integrated into power module Pierre-Laurent Doumergue R&D Engineer Microsemi Power Module Products 26 rue de Campilleau 33 520 Bruges, France Introduction:

More information

arxiv:astro-ph/0509450 v1 15 Sep 2005

arxiv:astro-ph/0509450 v1 15 Sep 2005 arxiv:astro-ph/0509450 v1 15 Sep 2005 TESTING THERMO-ACOUSTIC SOUND GENERATION IN WATER WITH PROTON AND LASER BEAMS K. GRAF, G. ANTON, J. HÖSSL, A. KAPPES, T. KARG, U. KATZ, R. LAHMANN, C. NAUMANN, K.

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

Balance Accuracy. FlexiWeigh

Balance Accuracy. FlexiWeigh FlexiWeigh The balance is an integral part of the FlexiWeigh workstation. The role and limitations of the balance are important to understand in this environment. This technical note describes the different

More information

Fibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques

Fibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques Fibre Bragg Grating Sensors An ntroduction to Bragg gratings and interrogation techniques Dr Crispin Doyle Senior Applications Engineer, Smart Fibres Ltd. 2003 1) The Fibre Bragg Grating (FBG) There are

More information

Performance ratio. Contents. Quality factor for the PV plant

Performance ratio. Contents. Quality factor for the PV plant Performance ratio Quality factor for the PV plant Contents The performance ratio is one of the most important variables for evaluating the efficiency of a PV plant. Specifically, the performance ratio

More information

The Electrical Properties of Materials: Resistivity

The Electrical Properties of Materials: Resistivity The Electrical Properties of Materials: Resistivity 1 Objectives 1. To understand the properties of resistance and resistivity in conductors, 2. To measure the resistivity and temperature coefficient of

More information

Pressure Transducer to ADC Application

Pressure Transducer to ADC Application Application Report SLOA05 October 2000 Pressure Transducer to ADC Application John Bishop ABSTRACT Advanced Analog Products/OpAmp Applications A range of bridgetype transducers can measure numerous process

More information

Check-list for auditors. Audit visit at candidate registered laboratories

Check-list for auditors. Audit visit at candidate registered laboratories Check-list for auditors Audit visit at candidate registered laboratories 1 Contents Introduction Laboratory documentation required by the standard(s), available for inspection by the auditor Items to be

More information

Temperature Calibration; Depths of Immersion

Temperature Calibration; Depths of Immersion Temperature Calibration; epths of Immersion Author: John P. Tavener Company Isothermal Technology Limited, Pine Grove, Southport, England Abstract Of all the sources of errors and uncertainties in thermal

More information

Experiment 6 ~ Joule Heating of a Resistor

Experiment 6 ~ Joule Heating of a Resistor Experiment 6 ~ Joule Heating of a Resistor Introduction: The power P absorbed in an electrical resistor of resistance R, current I, and voltage V is given by P = I 2 R = V 2 /R = VI. Despite the fact that

More information

IC Temperature Sensor Provides Thermocouple Cold-Junction Compensation

IC Temperature Sensor Provides Thermocouple Cold-Junction Compensation IC Temperature Sensor Provides Thermocouple Cold-Junction Compensation INTRODUCTION Due to their low cost and ease of use thermocouples are still a popular means for making temperature measurements up

More information

Carbon Cable. Sergio Rubio Carles Paul Albert Monte

Carbon Cable. Sergio Rubio Carles Paul Albert Monte Carbon Cable Sergio Rubio Carles Paul Albert Monte Carbon, Copper and Manganine PhYsical PropERTieS CARBON PROPERTIES Carbon physical Properties Temperature Coefficient α -0,0005 ºC-1 Density D 2260 kg/m3

More information

Temperature Rise Measurements

Temperature Rise Measurements Chapter 22 Temperature Rise Measurements 22.1 Background The SAR in tissue simulating liquid can be assessed by means of temperature rise or by measurements of induced electric field. SAR = c T dt (22.1)

More information

Department of Chemical Engineering, National Institute of Technology, Tiruchirappalli 620 015, Tamil Nadu, India

Department of Chemical Engineering, National Institute of Technology, Tiruchirappalli 620 015, Tamil Nadu, India Experimental Thermal and Fluid Science 32 (2007) 92 97 www.elsevier.com/locate/etfs Studies on heat transfer and friction factor characteristics of laminar flow through a circular tube fitted with right

More information

The accelerometer designed and realized so far is intended for an. aerospace application. Detailed testing and analysis needs to be

The accelerometer designed and realized so far is intended for an. aerospace application. Detailed testing and analysis needs to be 86 Chapter 4 Accelerometer Testing 4.1 Introduction The accelerometer designed and realized so far is intended for an aerospace application. Detailed testing and analysis needs to be conducted to qualify

More information

INTRODUCTION TO ERRORS AND ERROR ANALYSIS

INTRODUCTION TO ERRORS AND ERROR ANALYSIS INTRODUCTION TO ERRORS AND ERROR ANALYSIS To many students and to the public in general, an error is something they have done wrong. However, in science, the word error means the uncertainty which accompanies

More information

Experiment 4: Sensor Bridge Circuits (tbc 1/11/2007, revised 2/20/2007, 2/28/2007) I. Introduction. From Voltage Dividers to Wheatstone Bridges

Experiment 4: Sensor Bridge Circuits (tbc 1/11/2007, revised 2/20/2007, 2/28/2007) I. Introduction. From Voltage Dividers to Wheatstone Bridges Experiment 4: Sensor Bridge Circuits (tbc //2007, revised 2/20/2007, 2/28/2007) Objective: To implement Wheatstone bridge circuits for temperature measurements using thermistors. I. Introduction. From

More information

Measuring Temperature withthermistors a Tutorial David Potter

Measuring Temperature withthermistors a Tutorial David Potter NATIONAL INSTRUMENTS The Software is the Instrument Application Note 065 Measuring Temperature withthermistors a Tutorial David Potter Introduction Thermistors are thermally sensitive resistors used in

More information

The Determination of Uncertainties in Charpy Impact Testing

The Determination of Uncertainties in Charpy Impact Testing Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials Code of Practice No. 06 The Determination of Uncertainties in Charpy Impact Testing M.A. Lont

More information

COMMUNICATING using MEASUREMENTS In Engineering we use a great many measuring instruments.

COMMUNICATING using MEASUREMENTS In Engineering we use a great many measuring instruments. COMMUNICATING using MEASUREMENTS In Engineering we use a great many measuring instruments. Scales Verniers Micrometers Gauges Comparators Thermometers Thermistors + Indicator Thermocouples + Indicator

More information

Iterative calculation of the heat transfer coefficient

Iterative calculation of the heat transfer coefficient Iterative calculation of the heat transfer coefficient D.Roncati Progettazione Ottica Roncati, via Panfilio, 17 44121 Ferrara Aim The plate temperature of a cooling heat sink is an important parameter

More information

Soil Suction. Total Suction

Soil Suction. Total Suction Soil Suction Total Suction Total soil suction is defined in terms of the free energy or the relative vapor pressure (relative humidity) of the soil moisture. Ψ = v RT ln v w 0ω v u v 0 ( u ) u = partial

More information

Temperature Accuracy of Thermistors and RTDs Application Notes

Temperature Accuracy of Thermistors and RTDs Application Notes Thermistors and RTDs are devices used to measure temperature in modern Heating, Ventilating, Air Conditioning and Refrigeration (HVAC/R) systems. The electrical resistance of both devices is determined

More information

G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014

G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014 Page 1 of 31 G104 - Guide for Estimation of Measurement Uncertainty In Testing December 2014 2014 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without

More information

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors Tom Wise on behalf of Daya Bay collaboration Day Bay is a reactor based experiment dedicated to measurement of the last

More information

V out = V in x R 2 (R 1 + R 2 ) V o = V i R 1 FXA 2008 POTENTIAL DIVIDER CIRCUIT 1. Candidates should be able to : SUPPLYING A FIXED PD

V out = V in x R 2 (R 1 + R 2 ) V o = V i R 1 FXA 2008 POTENTIAL DIVIDER CIRCUIT 1. Candidates should be able to : SUPPLYING A FIXED PD POTENTIAL DIIDER CIRCUIT 1 Candidates should be able to : SUPPLYING A FIXED PD Draw a simple potential divider circuit. Explain how a potential divider circuit can be used to produce a variable pd. The

More information

Robot Perception Continued

Robot Perception Continued Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart

More information

Application Note 142 August 2013. New Linear Regulators Solve Old Problems AN142-1

Application Note 142 August 2013. New Linear Regulators Solve Old Problems AN142-1 August 2013 New Linear Regulators Solve Old Problems Bob Dobkin, Vice President, Engineering and CTO, Linear Technology Corp. Regulators regulate but are capable of doing much more. The architecture of

More information

METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS

METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS Contents Foreword --------------------------------------------------------------------------------------------------------------------

More information

How To Calculate Thermal Resistance On A Pb (Plastipo)

How To Calculate Thermal Resistance On A Pb (Plastipo) VISHAY BEYSCHLAG Resistive Products 1. INTRODUCTION Thermal management is becoming more important as the density of electronic components in modern printed circuit boards (PCBs), as well as the applied

More information

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM Force measurement Forces VECTORIAL ISSUES In classical mechanics, a force is defined as "an action capable of modifying the quantity of movement of a material point". Therefore, a force has the attributes

More information

How to calibrate an RTD or Platinum Resistance Thermometer (PRT)

How to calibrate an RTD or Platinum Resistance Thermometer (PRT) How to calibrate an RTD or Platinum Resistance Thermometer (PRT) Application Note Introduction There are two types of calibrations applicable to PRTs characterization and tolerance testing. The type of

More information

CMMs and PROFICIENCY TESTING

CMMs and PROFICIENCY TESTING CMMs and PROFICIENCY TESTING Speaker/Author: Dr. Henrik S. Nielsen HN Metrology Consulting, Inc. HN Proficiency Testing, Inc. Indianapolis, Indiana, USA hsnielsen@hn-metrology.com Phone: (317) 849 9577;

More information

Current Sensor: ACS755xCB-050

Current Sensor: ACS755xCB-050 Package CB-PFF 5 4 The Allegro ACS75x family of current sensors provides economical and precise solutions for current sensing in industrial, automotive, commercial, and communications systems. The device

More information

EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa

EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa Final Report of EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa Yasin DURGUT 1 Nenad PETROVSKI 2 Vanco KACARSKI 2 (1) National Metrology Institute (UME),

More information

AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL

AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL FUEL CONSUMPTION MEASUREMENT AVL FUEL MASS FLOW METER & FUEL TEMPERATURE CONTROL Description The combination of AVL Fuel Mass Flow Meter and AVL Fuel Temperature Control is a high precise and continuous

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information

SCS Directory Accreditation number: SCS 0115

SCS Directory Accreditation number: SCS 0115 International standard: ISO/IEC 17025:2005 Swiss standard: SN EN ISO/IEC 17025:2005 METRON SA Calibration Laboratory Via Luserte Sud 7 CH-6572 Quartino Head: Angelo Capone Responsible for MS: Angelo Capone

More information

An Overview of Calibration Methods and Procedures for Process and Inventory Weigh Systems

An Overview of Calibration Methods and Procedures for Process and Inventory Weigh Systems BLH NOBEL Weighing Systems Brands of VPG Process Weighing Solutions for Process Weighing and Force Measurement Handbook TC0010 Procedures for Process and Inventory Weigh Systems Allen-Bradley Automation

More information

High-ohmic/high-voltage resistors

High-ohmic/high-voltage resistors FEATURES High pulse loading capability Small size. APPLICATIONS Where high resistance, high stability and high reliability at high voltage are required High humidity environment White goods Power supplies.

More information

User checks and maintenance of laboratory balances

User checks and maintenance of laboratory balances Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National

More information

Procon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS

Procon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS Procon Engineering Technical Document PELR 1002 TERMS and DEFINITIONS The following terms are widely used in the weighing industry. Informal comment on terms is in italics and is not part of the formal

More information

Carbon Dioxide and an Argon + Nitrogen Mixture. Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10

Carbon Dioxide and an Argon + Nitrogen Mixture. Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10 Carbon Dioxide and an Argon + Nitrogen Mixture Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10 Measurement of C p /C v for Argon, Nitrogen, Carbon Dioxide and an Argon + Nitrogen Mixture

More information

16. Industrial Platinum Resistance Thermometers

16. Industrial Platinum Resistance Thermometers 129 16. Industrial Platinum Resistance Thermometers Industrial platinum resistance thermometers (IPRTs) are constructed from platinum of a lower quality than is used for SPRTs; typically, values of W(H

More information

Estimating Weighing Uncertainty From Balance Data Sheet Specifications

Estimating Weighing Uncertainty From Balance Data Sheet Specifications Estimating Weighing Uncertainty From Balance Data Sheet Specifications Sources Of Measurement Deviations And Uncertainties Determination Of The Combined Measurement Bias Estimation Of The Combined Measurement

More information

International Journal of Latest Research in Science and Technology Volume 4, Issue 2: Page No.161-166, March-April 2015

International Journal of Latest Research in Science and Technology Volume 4, Issue 2: Page No.161-166, March-April 2015 International Journal of Latest Research in Science and Technology Volume 4, Issue 2: Page No.161-166, March-April 2015 http://www.mnkjournals.com/ijlrst.htm ISSN (Online):2278-5299 EXPERIMENTAL STUDY

More information

American Society of Agricultural and Biological Engineers

American Society of Agricultural and Biological Engineers ASAE S580.1 NOV2013 Testing and Reporting Solar Cooker Performance American Society of Agricultural and Biological Engineers ASABE is a professional and technical organization, of members worldwide, who

More information

1 Calculus of Several Variables

1 Calculus of Several Variables 1 Calculus of Several Variables Reading: [Simon], Chapter 14, p. 300-31. 1.1 Partial Derivatives Let f : R n R. Then for each x i at each point x 0 = (x 0 1,..., x 0 n) the ith partial derivative is defined

More information

Guidelines on the Calibration of Static Torque Measuring Devices

Guidelines on the Calibration of Static Torque Measuring Devices European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET cg-14 Version.0 (03/011) Previously EA-10/14 Calibration Guide EURAMET cg-14

More information

A Simple Current-Sense Technique Eliminating a Sense Resistor

A Simple Current-Sense Technique Eliminating a Sense Resistor INFINITY Application Note AN-7 A Simple Current-Sense Technique Eliminating a Sense Resistor Copyright 998 A SIMPE CURRENT-SENSE TECHNIQUE EIMINATING A SENSE RESISTOR INTRODUCTION A sense resistor R S,

More information

Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz

Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity

More information

OPTIMIZATION OF DIAMETER RATIO FOR ALPHA-TYPE STIRLING ENGINES

OPTIMIZATION OF DIAMETER RATIO FOR ALPHA-TYPE STIRLING ENGINES OPTIMIZATION OF DIAMETER RATIO FOR ALPHA-TYPE STIRLING ENGINES VLAD MARIO HOMUTESCU* DAN-TEODOR BĂLĂNESCU* * Gheorghe Asachi Technical University of Iassy Department of of ermotechnics ermal Engines and

More information

Battery Thermal Management System Design Modeling

Battery Thermal Management System Design Modeling Battery Thermal Management System Design Modeling Gi-Heon Kim, Ph.D Ahmad Pesaran, Ph.D (ahmad_pesaran@nrel.gov) National Renewable Energy Laboratory, Golden, Colorado, U.S.A. EVS October -8, 8, 006 Yokohama,

More information

The Effect of Forced Air Cooling on Heat Sink Thermal Ratings

The Effect of Forced Air Cooling on Heat Sink Thermal Ratings zpero 1 The Effect of Forced Air Cooling on Heat Sink Thermal Ratings By Paul Bachman, Fellow Engineer & Ronnie Haiduk, Applications Engineer, Crydom, Inc. ABSTRACT A heat sink s ability to dissipate thermal

More information

High-ohmic/high-voltage resistors

High-ohmic/high-voltage resistors FEATURES These resistors meet the safety requirements of: UL1676 (range 510 kω to 11 MΩ) EN60065 BS60065 (U.K.) NFC 92-130 (France) VDE 0860 (Germany) High pulse loading capability Small size. APPLICATIONS

More information

HM1500LF Relative Humidity Module

HM1500LF Relative Humidity Module Small size product Product free from Lead, Cr(6+), Cd and Hg Humidity calibrated within +/-2% @55%RH Typical 1 to 3.6 Volt DC output for 0 to 100% RH at 5Vdc supply Ratiometric to voltage supply from 4.75Vdc

More information

R 60 DoMC-01 rev.1 Additional requirements from the United States. Accuracy class III L. Revision number Date of the revision Nature of the revision

R 60 DoMC-01 rev.1 Additional requirements from the United States. Accuracy class III L. Revision number Date of the revision Nature of the revision R 60 DoMC-01 rev.1 Additional requirements from the United States Accuracy class III L Revision number Date of the revision Nature of the revision Rev.0 29/09/2006 Initial document Rev.1 03/03/2014 Update

More information

9. Force Measurement with a Strain gage Bridge

9. Force Measurement with a Strain gage Bridge 9. Force Measurement with a Strain gage Bridge Task 1. Measure the transfer characteristics of strain-gage based force sensor in a range 0 10 kg 1. 2. Determine the mass of an unknown weight. 3. Check

More information

Load Cell Accuracy in Relation to the Conditions of Use

Load Cell Accuracy in Relation to the Conditions of Use VPG TRANSDUCERS Load Cells and Weigh Modules Technical Note VPGT-0 Load Cell Accuracy in Relation to the Conditions of Use. Load Cell Accuracy Load cells are ranked, according to their overall performance

More information

Digital to Analog and Analog to Digital Conversion

Digital to Analog and Analog to Digital Conversion Real world (lab) is Computer (binary) is digital Digital to Analog and Analog to Digital Conversion V t V t D/A or DAC and A/D or ADC D/A Conversion Computer DAC A/D Conversion Computer DAC Digital to

More information

FREESTUDY HEAT TRANSFER TUTORIAL 3 ADVANCED STUDIES

FREESTUDY HEAT TRANSFER TUTORIAL 3 ADVANCED STUDIES FREESTUDY HEAT TRANSFER TUTORIAL ADVANCED STUDIES This is the third tutorial in the series on heat transfer and covers some of the advanced theory of convection. The tutorials are designed to bring the

More information

A MEASUREMENT ASSURANCE PROGRAM FOR CALIBRATION BY THE TRANSFER METHOD CS800

A MEASUREMENT ASSURANCE PROGRAM FOR CALIBRATION BY THE TRANSFER METHOD CS800 A EAUREENT AURANCE PROGRA FOR CALIBRATION BY THE TRANFER ETHOD C800 JEFF PEERY DEC. 6, 011 This document outlines the easurement Assurance Program (AP for eaetrics C800 flo measurement system a liquid

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information