LINE SCALE BILATERAL COMPARISON

Save this PDF as:
 WORD  PNG  TXT  JPG

Size: px
Start display at page:

Download "LINE SCALE BILATERAL COMPARISON"

Transcription

1 LINE SCALE BILATERAL COMPARISON Draft A 01 June 2016

2 1. DOCUMENT CONTROL Version Draft A: 01 June INTRODUCTION 2.1 INRIM (Italy) and INTI (Argentina) aim to perform a bilateral comparison of line scale. The subject is to give metrological equivalence to the measurement of line scale at INTI. 2.2 INTI will be the pilot laboratory. INRIM will be the linking laboratory. (Uruguay) will be the laboratory that protects the data comparison. 2.3 This document describes the technical protocol for the bilateral comparison of line scale between INRIM and INTI. It is based on the protocol for the comparison of line scale (EURAMET.L-K7). 2.4 Both NMIs (INRIM and INTI) have prepared the technical protocol and INTI (pilot laboratory) will be responsible for data evaluation and final report. 3. MEASUREMENT ARTEFACT The measurement artefact is a glass scale from MITUTOYO, belonging to MITUTOYO-Argentine, Figure 1. Figure 1 Glass scale Next are indicated the characteristics of the scale. Figure 2 Model: S/N: Range: 300 mm Resolution: 0,5 mm Thermal expansion coefficient: (8 ± 1) x 10-6 [1 / ºC] Graduation line thickness: 0,09 mm Overall dimension: 320 mm x 20 mm x 5 mm 2 de 6

3 The line scale is contained in a wooden box. Figure 2 Line scale characteristics An image of the lines with an objective of 64 X is showed in the Figure 3 Figure 3 Image of the lines with 64X 4. HANDLING THE ARTEFACT 4.1 General Handling Open the wooden box carefully and only in a clean environment. Use clean room gloves in order to handle the scale and never touch the surface of the scale. It is not allowed to use any type of glue or wax for fixing the scale. When not in use, place the scale back into its container to avoid dust or dirt deposits. 4.2 Cleaning Cleaning the scale with soft tissues or similar should be avoided. If it is necessary to clean the scale before the measurement clean air or other clean gas (preferable to very low pressure) can be used. 4.3 Temperature measurement of the artifact For the measurement of the artefact, it is not allowed to fix the temperature sensors to the artefact using any type of glue, wax or clamping fixture. It is recommended to measure the temperature in the mounting fixtures. 4.4 Storage Use original transportation container to avoid dust deposits. Always try to keep the artefact under good measuring room conditions, i.e. within the room, where it gets calibrated. 5. MEASURAND The measurand is the distance between the center line position of the reference line (position 0 ) and the center line position of the measured line. 3 de 6

4 To increase comparability of the results, all measurements are performed over the center of the shorter lines (e.g. line of 0,5 mm), with a measuring window of 0,1 mm. That is, it had to be tried to apply an effective slit height or CCD image window height of 0,1 mm for the analysis of measurements, Figure 4. The distance between the reference line and the measured line is after calculated over the reference axis (center of measuring section). It is recommended to use the line of 0,5 mm (L=2,02 mm) for the determination of the measuring section. Figure 4 Section of measurement 6. MEASUREMENTS INSTRUCTIONS Before calibration, the line scale must be inspected for damage. It is recommended to support the scale at the Airy points, held only by their gravity forces. If additional clamping of the scale is required during the measurement, e.g. because of a fast moving carriage, it is recommended to lightly pinch the scale on the sides at one of the Airy support points. For alignment purposes of the graduation lines, the lower edge of the graduation lines should be used. It is recommended that the principal references will be the lines of 0 mm and 280 mm, Figure 5. Figure 5 Alignment of the scale 4 de 6

5 The measurement result to be reported is the distance from the zero line to the measured line at 20ºC, according table 1. Distance range Pitch 0 to 10 mm 1 mm 0 to 280 mm 28 mm Table 1 Distances to measure The air and material temperature during measurement should be reported also. 7. MEASUREMENT UNCERTAINTY The uncertainty of measurement shall be estimated according to the ISO Guide to the Expression of Uncertainty in Measurement (GUM) and the complete uncertainty budget must be reported. 8. REPORTING OF RESULTS The report should include: - Description of the measurement method - The reference standard - The traceability to the SI - The results of the quantities to be measured - The associated standard uncertainties, the effective degrees of freedom and the expanded uncertainties. - The uncertainty budget 9. TRANSPORTATION The travelling standard must be transported in the original case and protected from mechanical loads, vibrations, etc. for the transport by plane. 10. CIRCULATION SCHEME The agenda of the calibration is given in the table 2. Laboratory Schedule time INTI 1 st June 2016 August data backup INRIM October 2016 November data backup INTI 2 nd December 2016 December data backup 11. Table 2 Timetable of the comparison (INRIM laboratory and INTI laboratory agree to send a copy of the data to the laboratory, in order to safeguard the data comparison) 5 de 6

6 12. CONTACT INRIM Strada delle Cacce 91 I-10135, Torino Italy Contact: Mr. Massimo Zucco Tel: (968) INTI Av. Vélez Sársfield 1561 X5000JKC, Córdoba Córdoba, Argentina Contact: Mr. Bruno Gastaldi Tel: Av. Italia , Montevideo Uruguay Contact: Mr. Alejandro Acquarone Tel: de 6

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS XVII IMEKO World Congress Metrology in the 3rd Millennium June 22 27, 2003, Dubrovnik, Croatia LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS Tanfer Yandayan TUBITAK-Ulusal

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 7 Contact: Mr S C Sparks Solent Industrial Estate Tel: +44 (0)1489 790296 Hedge End Fax: +44 (0)1489 790294 Southampton E-Mail: info@southcal.co.uk

More information

OBSERVATIONS Serial. No. Reading of meniscus Reading of pointer tip Height h 1 - M.S.R V.S.R T.R(h 1 ) M.S.R V.S.R T.R(h 2 ) h 2 = h (cm)

OBSERVATIONS Serial. No. Reading of meniscus Reading of pointer tip Height h 1 - M.S.R V.S.R T.R(h 1 ) M.S.R V.S.R T.R(h 2 ) h 2 = h (cm) EXPERIMENT 11 AIM To determine the surface tension of water by capillary rise method. Capillary tube and a tipped pointer clamped in a stand, travelling microscope, clean water in a beaker. Surface tension,

More information

Solar Radiation Measurement. Bruce W Forgan, WMO RAV Metrology Workshop, Melbourne, Novemberr 2011

Solar Radiation Measurement. Bruce W Forgan, WMO RAV Metrology Workshop, Melbourne, Novemberr 2011 Solar Radiation Measurement Bruce W Forgan, WMO RAV Metrology Workshop, Melbourne, Novemberr 2011 Why Do We Need Data on Solar Energy? Global Climate System Climate Energy Balance Solar Exposure and Irradiance

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information

CCM.M-K3.1 comparison / 50 kg Mass

CCM.M-K3.1 comparison / 50 kg Mass CCM.M-K3.1 comparison / 50 kg Mass The 7 th of April 2010 G. MANN - T. MADEC - P.A. MEURY Laboratoire Commun de Métrologie LNE-CNAM Abstract This comparison is a subsequent to the CCM.M-K3 comparison performed

More information

To measure an object length, note the number of divisions spanned by the object then multiply by the conversion factor for the magnification used.

To measure an object length, note the number of divisions spanned by the object then multiply by the conversion factor for the magnification used. STAGE MICROMETERS Introduction Whenever there is a need to make measurements with an eyepiece graticule, there is also a need to ensure that the microscope is calibrated. The use of a stage micrometer

More information

Dome Camera with IR Night Vision

Dome Camera with IR Night Vision Dome Camera with IR Night Vision Instruction Manual VER.:1.0, NO.: R040687/2 Thank you for purchasing our product. Before installing this camera, please read this instruction manual carefully to ensure

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK code Location code Customers Sites locations: 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Lim Tiong Thai Tel: +673-2-386475

More information

(Uncertainty) 2. How uncertain is your uncertainty budget?

(Uncertainty) 2. How uncertain is your uncertainty budget? (Uncertainty) 2 How uncertain is your uncertainty budget? Paper Author and Presenter: Dr. Henrik S. Nielsen HN Metrology Consulting, Inc 10219 Coral Reef Way, Indianapolis, IN 46256 Phone: (317) 849 9577,

More information

Torque Key Comparison CCM.T-K1.3

Torque Key Comparison CCM.T-K1.3 Torque Key Comparison CCM.T-K1.3 Measurand Torque: 0 N m, 500 N m, 1000 N m Final Report March 9, 2015 Pilot Institute: Contact Person: E-mail: Co-author: Korea Research Institute of Standards and Science,

More information

technical guide Simple Linear Measurement Instruments

technical guide Simple Linear Measurement Instruments technical guide Simple Linear Measurement Instruments Second edition February 2008 technical guide Simple Linear Measurement Instruments AS TG 1 Edition Statement Edition Amendment Date of Issue ISBN No.

More information

Temperature Rise Measurements

Temperature Rise Measurements Chapter 22 Temperature Rise Measurements 22.1 Background The SAR in tissue simulating liquid can be assessed by means of temperature rise or by measurements of induced electric field. SAR = c T dt (22.1)

More information

Equipment Checks and Calibrations

Equipment Checks and Calibrations Equipment Checks and Calibrations Paul McMullen Sector Manager, Calibration This Session 1. Calibration: What is it? 2. A process of comparison 3. What about adjustment? 4. What is a check? 5. In-house

More information

Easy Machining Center Setup

Easy Machining Center Setup White Paper Document No. MWA-072-EN_01_1404 April 2014 Easy Machining Center Setup Using FANUC s Direct Input of Workpiece Origin Setting Measured and Tool Length Measurement features to easily establish

More information

STRAIN GAGE INSTALLATION HOLE DRILLING

STRAIN GAGE INSTALLATION HOLE DRILLING Via Delle Calandre 63 50041 Calenzano (Florence) Italy Tel: +39.055.8826-302 Fax: -303 www.sintechnology.com info@sintechnology.com STRAIN GAGE INSTALLATION HOLE DRILLING How to install strain gages for

More information

3R UAV01 Ultrasonic Anemometer for Davis VP2 User Manual. 14/05/2012 Rev. A

3R UAV01 Ultrasonic Anemometer for Davis VP2 User Manual. 14/05/2012 Rev. A User Manual 14/05/2012 Rev. A Darrera, S.A. C/ de l Església, 5-7-9 08950 Esplugues de Llobregat Barcelona (Spain) Tel: +34 934 734 532 Fax: +34 934 735 708 info@darrera.com www.darrera.com Copyright Darrera,

More information

Keywords: Torque, calibration, uncertainty, torque tools, torque testers. Contents

Keywords: Torque, calibration, uncertainty, torque tools, torque testers. Contents Measurement uncertainty in torque calibration Luca G. Bochese, ASQ CQE, CRE (Boch - Milano), info@boch.net, www.boch.net Dr. A. Bochese ( Feanor - Tallinn), info@feanor.com, www.feanor.com Keywords: Torque,

More information

EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa

EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa Final Report of EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa Yasin DURGUT 1 Nenad PETROVSKI 2 Vanco KACARSKI 2 (1) National Metrology Institute (UME),

More information

PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES

PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES British Made Portable Microscopes with Image Erection PCB Through - Hole Scope Depthscope Brinell Hardness Impression Microscope Cathode Ray Tube Inspector Portable Video Scope Metallurgical Microscope

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

National Performance Evaluation Facility for LADARs

National Performance Evaluation Facility for LADARs National Performance Evaluation Facility for LADARs Kamel S. Saidi (presenter) Geraldine S. Cheok William C. Stone The National Institute of Standards and Technology Construction Metrology and Automation

More information

Mounting Instructions for SP4 Power Modules

Mounting Instructions for SP4 Power Modules Mounting Instructions for SP4 Power Modules Pierre-Laurent Doumergue R&D Engineer Microsemi Power Module Products 26 rue de Campilleau 33 520 Bruges, France Introduction: This application note gives the

More information

DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements

DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements DMS 680 Universal length measuring system gauge inspection for»iso 9000«requirements Universal length measuring system DMS 680 Wide range of applications For periodic inspection of gauges, reference gauges,

More information

Mercury TM 1500P PCB-Mount Digital Encoders

Mercury TM 1500P PCB-Mount Digital Encoders Mercury TM 1500P PCB-Mount Digital Encoders Factory Set Resolution to 0.50μm Reflective Linear and Rotary Encoders Sensor the size of a Dime Resolution Factory Set: x4, x8, x20, or x40 Linear: 5μm, 2.5μm,

More information

Good Practice Guide No. 41. CMM Measurement Strategies. David Flack. Issue 2

Good Practice Guide No. 41. CMM Measurement Strategies. David Flack. Issue 2 Good Practice Guide No. 41 CMM Measurement Strategies David Flack Issue 2 Measurement Good Practice Guide No. 41 CMM measurement strategies David Flack Engineering Measurement Division National Physical

More information

ASSEMBLY INSTRUCTIONS

ASSEMBLY INSTRUCTIONS 00MK-MA0140-EN00 ASSEMBLY INSTRUCTIONS Universal system with fixing screws, for sloped roof Rev.02 February 2014 Table of contents General instructions 2 General safety instructions 3 Universal system

More information

Check-list for auditors. Audit visit at candidate registered laboratories

Check-list for auditors. Audit visit at candidate registered laboratories Check-list for auditors Audit visit at candidate registered laboratories 1 Contents Introduction Laboratory documentation required by the standard(s), available for inspection by the auditor Items to be

More information

SP AF 17~50 mm F/2.8 XR Di-II LD Aspherical [IF] (Model A16)

SP AF 17~50 mm F/2.8 XR Di-II LD Aspherical [IF] (Model A16) SP AF 17~50 mm F/2.8 XR Di-II LD Aspherical [IF] (Model A16) Thank you for purchasing the Tamron lens as the latest addition to your photographic equipment. Di II lens (Digitally Intigrated) series are

More information

The Interpretation of Optical Data Sheets. Carl Zeiss AG, Udo Schellenbach, PH-V

The Interpretation of Optical Data Sheets. Carl Zeiss AG, Udo Schellenbach, PH-V The Interpretation of Optical Data Sheets Some Facts about Zeiss Since 1896 famous for Camera Lenses Mr. Carl Zeiss founded Zeiss 1846 and started collaboration with Prof. Ernst Abbe 1866 Worldwide 48

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

REPORT OF CALIBRATION N 1ST Test No. 821/

REPORT OF CALIBRATION N 1ST Test No. 821/ UNITED STATES DEPARTMENT OF COMMERCE National Institute of Standards and Technology Gaithersburg, Maryland 20899- REPORT OF CALIBRATION N 1ST Test No. 821/275185-07 For: (I) NIST Committee Meter Bar Page

More information

KALIBRERINGSBEVIS. 25 temperature loggers Ecoguard model TS-020000, identification se table of results.

KALIBRERINGSBEVIS. 25 temperature loggers Ecoguard model TS-020000, identification se table of results. utfärdat av riksmätplats 01 issued by a Swedish National Laboratory Contact person Lars-Erik Josefson 2014-03-27 1 (3) MTv4F006969-K01 Measurement Technology +46 10 516 54 56 larserik.josefson@sp.se EcoGuard

More information

2 Axis Solar Tracker / 12 Kw

2 Axis Solar Tracker / 12 Kw 2 Axis Solar Tracker / 12 Kw 4 Rows 14 metres 4 rows x 14 metres Quantity / Panel Type: 68 units Photovoltaic panel 1.600 x 790 mm Panel power / Total power: 165 W 11,2 Kw Total Surface: 94 m² Diameter

More information

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Verification of Impact Testing Machine

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Verification of Impact Testing Machine NABL 122-14 NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Verification of Impact Testing Machine Reviewed

More information

Machine Standard Configuration Standard Configuration Machine Alignment Capability Process Alignment Capability Core Cycle Time Maximum Print Area Printer Construction ISCAN TM Machine Control Operating

More information

AF 70~300 mm F/4-5.6 Di LD Macro 1:2 (Model A17)

AF 70~300 mm F/4-5.6 Di LD Macro 1:2 (Model A17) AF 70~300 mm F/4-5.6 Di LD Macro 1:2 (Model A17) Thank you for purchasing the Tamron lens as the latest addition to your photographic equipment. Di II lens (Digitally Integrated) series are designed for

More information

S Davidson, M Perkin, M Buckley

S Davidson, M Perkin, M Buckley No. 71 Measurement Good Practice Guide The Measurement of Mass and Weight S Davidson, M Perkin, M Buckley The National Physical Laboratory is operated on behalf of the DTI by NPL Management Limited, a

More information

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and

More information

Stage Metrology Concepts: Application Specific Testing. Hans Hansen Sr. Systems Design Engineer BSME, MSE 2/28/2008

Stage Metrology Concepts: Application Specific Testing. Hans Hansen Sr. Systems Design Engineer BSME, MSE 2/28/2008 Stage Metrology Concepts: Application Specific Testing Hans Hansen Sr. Systems Design Engineer BSME, MSE 2/28/2008 Topics Review of Stage Error Definitions Setting Stage Specifications for Applications

More information

Halogen Lamps. Lamp Construction

Halogen Lamps. Lamp Construction Halogen Lamps Lamp Construction Halogen lamps are the most advanced and multi-purpose incandescent lamps. Although they belong to the incandescent family of lamps, they are designed to provide a superior

More information

SKF Shaft Alignment Tools

SKF Shaft Alignment Tools SKF Shaft Alignment Tools Accurate shaft alignment really matters Reduce machinery breakdowns and increase your uptime It s a fact. Shaft misalignment is responsible for up to % of all costs related to

More information

Compliant Terminal Technology Summary Test Report

Compliant Terminal Technology Summary Test Report Engineering Report ER04100 October 5th, 2004 Revision A Copyright Autosplice Inc., September 2004 Table of Contents Summary Overview 3 Compliant Terminal Specifications 3 Test Plan 4 Test Conditions 4

More information

Rear Impact Guard TEST METHOD 223. Standards and Regulations Division. Issued: December 2003

Rear Impact Guard TEST METHOD 223. Standards and Regulations Division. Issued: December 2003 Transport Canada Safety and Security Road Safety Transports Canada Sécurité et sûreté Sécurité routière Standards and Regulations Division TEST METHOD 223 Rear Impact Guard Issued: December 2003 Standards

More information

MT-30 & MT-90 Series. Advanced Academic Microscopes/ Advanced Academic Polarizing Microscope INSTRUCTION MANUAL

MT-30 & MT-90 Series. Advanced Academic Microscopes/ Advanced Academic Polarizing Microscope INSTRUCTION MANUAL Introduction With your purchase of an MT-30/MT-90 series type microscope you have chosen for a quality product. The MT-30/MT-90 series type microscopes are developed for use at schools and laboratories.

More information

RI27-J CNG REDUCER INSTALLATION PROVISIONS & WARNINGS

RI27-J CNG REDUCER INSTALLATION PROVISIONS & WARNINGS RI27-J CNG REDUCER INSTALLATION PROVISIONS & WARNINGS Rev.01.02 BIGAS INTERNATIONAL AUTOGAS SYSTEMS S.r.l. Via di Le Prata, 62/66-50041 Calenzano Firenze ITALY Tel. 0554211275 - Fax 0554215977 http: www.bigas.it

More information

Products for Effective Asset Management

Products for Effective Asset Management HellermannTyton RFID SOLUTIONS Products for Effective Asset Management Asset Management There is an ever increasing need by business to ensure the traceability and record-keeping of calibration, location

More information

18-270mm F/3.5-6.3 Di II VC PZD for Canon, Nikon (Model B008) 18-270mm F/3.5-6.3 Di II PZD for Sony (Model B008)

18-270mm F/3.5-6.3 Di II VC PZD for Canon, Nikon (Model B008) 18-270mm F/3.5-6.3 Di II PZD for Sony (Model B008) R 18-270mm F/3.5-6.3 Di II VC PZD for Canon, Nikon (Model B008) 18-270mm F/3.5-6.3 Di II PZD for Sony (Model B008) Thank you for purchasing the Tamron lens as the latest addition to your photographic equipment.

More information

Scanner Specifications

Scanner Specifications Scanner Parts Scanner cover with built in transparency unit (TPU) Option The Multi Photo Feeder (part number B813372) lets you load up to 24 photos or business cards and scan them automatically. See the

More information

Temperature Scales. temperature scales Celsius Fahrenheit Kelvin

Temperature Scales. temperature scales Celsius Fahrenheit Kelvin Ch. 10-11 Concept Ch. 10 #1, 3, 7, 8, 9, 11 Ch11, # 3, 6, 11 Problems Ch10 # 3, 5, 11, 17, 21, 24, 25, 29, 33, 37, 39, 43, 47, 59 Problems: CH 11 # 1, 2, 3a, 4, 5, 6, 9, 13, 15, 22, 25, 27, 28, 35 Temperature

More information

Using Equotip Hardness Test Blocks Abstract 1. Introduction 2. Verification of Leeb hardness testers according to standards

Using Equotip Hardness Test Blocks Abstract 1. Introduction 2. Verification of Leeb hardness testers according to standards Using Equotip Hardness Test Blocks Heinz-Horst Pollok, consultant for material testing and instrument electronics (Cologne, Germany) Dr. Ralph T. Mennicke, product manager for material testing, Proceq

More information

DEBUNKING THE TWO GREAT MYTHS ABOUT CALIBRATION: TRACEABILITY TO NIST IF YOU CANNOT ADJUST, YOU CANNOT CALIBRATE! Bucherview Metrology Services

DEBUNKING THE TWO GREAT MYTHS ABOUT CALIBRATION: TRACEABILITY TO NIST IF YOU CANNOT ADJUST, YOU CANNOT CALIBRATE! Bucherview Metrology Services DEBUNKING THE TWO GREAT MYTHS ABOUT CALIBRATION: TRACEABILITY TO NIST IF YOU CANNOT ADJUST, YOU CANNOT CALIBRATE! BY DEFINITION First, let s get everyone on the same page with the definitions of two critical

More information

RI27 CNG REDUCER INSTALLATION PROVISIONS & WARNINGS

RI27 CNG REDUCER INSTALLATION PROVISIONS & WARNINGS RI27 CNG REDUCER INSTALLATION PROVISIONS & WARNINGS Rev.01.02 BIGAS INTERNATIONAL AUTOGAS SYSTEMS S.r.l. Via di Le Prata, 62/66-50041 Calenzano Firenze ITALY Tel. 0554211275 - Fax 0554215977 http: www.bigas.it

More information

00MK-MA0190-EN00. Mounting System ASSEMBLY INSTRUCTIONS. Universal Light system with locking brackets, for sloped roof

00MK-MA0190-EN00. Mounting System ASSEMBLY INSTRUCTIONS. Universal Light system with locking brackets, for sloped roof 00MK-MA0190-EN00 Mounting System ASSEMBLY INSTRUCTIONS Universal Light system with locking brackets, for sloped roof Rev.02 November 2014 Table of contents General instructions 2 General safety instructions

More information

NRC Publications Archive Archives des publications du CNRC

NRC Publications Archive Archives des publications du CNRC NRC Publications Archive Archives des publications du CNRC Digital 3D Imaging and Modeling: A Metrological Approach Beraldin, Jean-Angelo Publisher s version / Version de l'éditeur: Time Compression Technologies

More information

Frameless Kitchen Cabinet Installation Guidelines

Frameless Kitchen Cabinet Installation Guidelines Frameless Kitchen Cabinet Installation Guidelines Easy to follow step-by-step kitchen cabinet installation guidelines TOOL AND MATERIAL LIST Hammer or nail gun Screwdriver Standard, Phillips 2 Level &

More information

VITOSOL r 200-T SP2A. VITOSOL 200-T Type SP2A

VITOSOL r 200-T SP2A. VITOSOL 200-T Type SP2A Technical Data Manual Model Nos. and pricing: see Price List Vacuum tube collector based on the heat pipe principle For the utilisation of solar energy VITOSOL r 200-T SP2A Product may not be exactly as

More information

CMMs and PROFICIENCY TESTING

CMMs and PROFICIENCY TESTING CMMs and PROFICIENCY TESTING Speaker/Author: Dr. Henrik S. Nielsen HN Metrology Consulting, Inc. HN Proficiency Testing, Inc. Indianapolis, Indiana, USA hsnielsen@hn-metrology.com Phone: (317) 849 9577;

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Old Bracknell Lane West Bracknell Berkshire RG12 7AH Contact: Mr M Trotter Tel: +44 (0)1344 459314 Fax: +44 (0)1344 465556 E-Mail: mtrotter@bis.fm

More information

Infrared Viewers. Manual

Infrared Viewers. Manual Infrared Viewers Manual Contents Introduction 3 How it works 3 IR viewer in comparison with a CCD camera 4 Visualization of infrared laser beam in mid-air 4 Power Density 5 Spectral sensitivity 6 Operation

More information

Wetroom Panel (Concealed Fix)

Wetroom Panel (Concealed Fix) Wetroom Panel (Concealed Fix) Instruction Manual DXFCPCF Important Information Toughened glass is completely safe for use in our shower enclosures and bath screens; providing our products are installed

More information

Model Year: 2010 Model: Prius Doc ID: RM000001Y3B015X. Title: ALIGNMENT / HANDLING DIAGNOSIS: FRONT WHEEL ALIGNMENT: ADJUSTMENT (2010 Prius)

Model Year: 2010 Model: Prius Doc ID: RM000001Y3B015X. Title: ALIGNMENT / HANDLING DIAGNOSIS: FRONT WHEEL ALIGNMENT: ADJUSTMENT (2010 Prius) Last Modified: 5-3-2010 6.4 N From: 200904 Model Year: 2010 Model: Prius Doc ID: RM000001Y3B015X Title: ALIGNMENT / HANDLING DIAGNOSIS: FRONT WHEEL ALIGNMENT: ADJUSTMENT (2010 Prius) ADJUSTMENT If the

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS Knife Sharpener MMS INS8500GB 03-2010-03 Instructions Knife Sharpener MMS 1 Instructions Knife Sharpener MMS 2 CONTENTS 1. INTENDED USE... 4 2. SYMBOLS... 4 3. SAFETY NOTES... 4

More information

Laddomat 21-60 Charging unit

Laddomat 21-60 Charging unit Laddomat 21-60 Charging unit User and installation instructions NOTE! Diagrams in this brochure only describe connection principles. Each installation must be measured and carried out according to the

More information

MEASURING MACHINES. Pratt & Whitney METROLOGY LABORATORY. Measurement Systems, Inc.

MEASURING MACHINES. Pratt & Whitney METROLOGY LABORATORY. Measurement Systems, Inc. METROLOGY LABORATORY Pratt & Whitney Measurement s, Inc. METROLOGY LABORATORY The Standard of Accuracy Pratt & Whitney Metrology Laboratory Machines are the standard of accuracy to which all other gages

More information

CT Traceability - Prof. Wim Dewulf, Group T - KU Leuven

CT Traceability - Prof. Wim Dewulf, Group T - KU Leuven CT Traceability - Calibration and Accuracy Prof. Wim Dewulf, Group T - KU Leuven Outline Introduction: terminology and procedures Voxel size calibration Edge offset calibration Conclusions Outline Introduction:

More information

pinta ACOUSTIC PANEL [1/7]

pinta ACOUSTIC PANEL [1/7] [1/7] Too much echo in the workplace? Staff unable to concentrate on their work? The answer is: ACOUSTIC PANEL. Service Call our telephone hotline for advice on your individual situation. Animated installation

More information

Realization of a UV fisheye hyperspectral camera

Realization of a UV fisheye hyperspectral camera Realization of a UV fisheye hyperspectral camera Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM Outline Purpose of the instrument Required specs Hyperspectral technique Optical

More information

Good Practice Guide No. 40. Callipers and micrometers. David Flack. Issue 2

Good Practice Guide No. 40. Callipers and micrometers. David Flack. Issue 2 Good Practice Guide No. 40 Callipers and micrometers David Flack Issue 2 Measurement Good Practice Guide No. 40 Callipers and Micrometers David Flack Engineering Measurement Division National Physical

More information

CQ-Series. User Manual. American Weigh Scales, Inc. CQ-350 (350x0.1g) CQ-500 (500x0.1g) Digital Pocket Scale . = + CALC AC CE % TARE OFF ON MC MR CALI

CQ-Series. User Manual. American Weigh Scales, Inc. CQ-350 (350x0.1g) CQ-500 (500x0.1g) Digital Pocket Scale . = + CALC AC CE % TARE OFF ON MC MR CALI American Weigh Scales, Inc. CQ-Series Digital Pocket Scale 7 4 1 0. 8 9 5 6 2 3 x = - CALI + CALC MODE M CAL AC CE % TARE OFF ON MC MR CQ-350 (350x0.1g) CQ-500 (500x0.1g) User Manual CQ-Series Manual Thank

More information

M S. Laser Alignment Standards & Procedures

M S. Laser Alignment Standards & Procedures M M S Machine Monitoring Systems Ltd Tel: +44 (0) 1869 250801 Unit 4 Business Point Fax: +44(0) 1869 250802 P.O. Box 92 Bicester OX26 5EW Laser Alignment Standards & Procedures Contents Introduction Laser

More information

Multi-touch Overlay Kit

Multi-touch Overlay Kit OWNER S MANUAL Multi-touch Overlay Kit Please read this manual carefully before operating the your set and retain it for future reference. KT-T320 KT-T430 KT-T490 KT-T550 www.lg.com 2 TABLE OF CONTENTS

More information

SCHATZ - TEST Tool testing systems. SCHATZ -certest-w. Dynamic testing of torque tools

SCHATZ - TEST Tool testing systems. SCHATZ -certest-w. Dynamic testing of torque tools SCHATZ - TEST Tool testing systems SCHATZ -certest Dynamic testing of torque tools SCHATZ -calitest Calibrator for indicating and click-type torque tools SCHATZ -combitest Mobile test center for all torque

More information

The new ISO standard 17123-8 for checking GNSS field measuring systems

The new ISO standard 17123-8 for checking GNSS field measuring systems The new ISO standard 1713-8 for checking GNSS field measuring systems Hans Heister Institute for Geodesy - Geodetic Metrology - UniBw München D-85577 Neubiberg h.heister@unibw.de FIG Working Week - Stockholm

More information

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003)

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003) (Part # S-LIB-M003) The smart sensor is designed to work with the HOBO Weather Station logger. The smart sensor has a plug-in modular connector that allows it to be added easily to a HOBO Weather Station.

More information

UNIVERSAL METAL TOOL STAND

UNIVERSAL METAL TOOL STAND UNIVERSAL METAL TOOL STAND 46075 ASSEMBLY INFORMATION Visit our website at: http://www.harborfreight.com Read this material before using this product. Failure to do so can result in serious injury. Save

More information

AF mm F/4-5.6 LD MACRO 1:2 (Model 572D)

AF mm F/4-5.6 LD MACRO 1:2 (Model 572D) AF 70-300mm F/4-5.6 LD MACRO 1:2 (Model 572D) Thank you for purchasing this Tamron lens. Please read this owner's manual carefully before using your new lens. Proper care and maintenance of this lens will

More information

Building an Off-Center Fixture for Turning Pendants

Building an Off-Center Fixture for Turning Pendants Building an Off-Center Fixture for Turning Pendants Turning a pendant off-center with most available metal pendant chucks means that you will have a significant amount of mass off center, which will limit

More information

Installation and Operation Manual. Digital Remote Meter for Monitoring System Performance. Version: RM-1

Installation and Operation Manual. Digital Remote Meter for Monitoring System Performance. Version: RM-1 REMOTE METER TM Installation and Operation Manual. Digital Remote Meter for Monitoring System Performance. Version: RM-1 1098 Washington Crossing Road Washington Crossing, PA 18977 USA www.morningstarcorp.com

More information

Calibration of AFM with virtual standards; robust, versatile and accurate. Richard Koops VSL Dutch Metrology Institute Delft

Calibration of AFM with virtual standards; robust, versatile and accurate. Richard Koops VSL Dutch Metrology Institute Delft Calibration of AFM with virtual standards; robust, versatile and accurate Richard Koops VSL Dutch Metrology Institute Delft 19-11-2015 VSL Dutch Metrology Institute VSL is the national metrology institute

More information

digital quality control fail pass at your fingertips Smart Metrology Solutions.

digital quality control fail pass at your fingertips Smart Metrology Solutions. digital quality control fail pass µphase smartgage The clever innovation in digital metrology by FISBA, which lets you smarten up your quality management and economize your production processes, yet spares

More information

Assembly of LPCC Packages AN-0001

Assembly of LPCC Packages AN-0001 Assembly of LPCC Packages AN-0001 Surface Mount Assembly and Handling of ANADIGICS LPCC Packages 1.0 Overview ANADIGICS power amplifiers are typically packaged in a Leadless Plastic Chip Carrier (LPCC)

More information

ATL Fuel Level Sender Probes

ATL Fuel Level Sender Probes T E C H N I C A L S P E C I F I C A T I O N The ATL EL-AD-151 (Resistance Output) and EL-AD-152 (Voltage Output) Fuel Level Senders are highly advanced sensors for continuously measuring the contents of

More information

Types of 3D Scanners and 3D Scanning Technologies.

Types of 3D Scanners and 3D Scanning Technologies. Types of 3D Scanners and 3D Scanning Technologies. There are many types of 3D scanners and 3D scanning technologies. Some are ideal for short range scanning while others are better suited for mid or long

More information

GIBEN PRESENTS THE INDUSTRIAL PANEL SAW MODEL

GIBEN PRESENTS THE INDUSTRIAL PANEL SAW MODEL GIBEN PRESENTS THE INDUSTRIAL PANEL SAW MODEL This and every photograph is representative only and are not intended as a scope of supply. They are merely an aid memoir of the various innovations available

More information

MS9000SAN Rework Station

MS9000SAN Rework Station MS9000SAN Rework Station MS9000SAN is the all-round rework system which almost all SMD can be reworked. And the original ITTS auto profiler system is operates of the system easily and exactly. Many kinds,

More information

EQUIPMENT AND PROCEDURE FOR ON-SITE INSPECTION OF MAGNETIC SEPARATORS

EQUIPMENT AND PROCEDURE FOR ON-SITE INSPECTION OF MAGNETIC SEPARATORS EQUIPMENT AND PROCEDURE FOR ON-SITE INSPECTION OF MAGNETIC SEPARATORS 1. Scope 1.1. This standard presents a unified approach to magnetic measurement of permanent magnetic and electromagnetic separation

More information

WSD130 SOLDERING STATION

WSD130 SOLDERING STATION NOTES S2032ACH 8/98 Rev. 11/01 WSD130 SOLDERING STATION 1. Power Switch 2. Digital Display 3. UP Button 4. DOWN Button 5. Optical Regulator, Right 6. Connection for Soldering Iron, Right 7. Selection Switch

More information

Data Modul. QMatrix Multi touch: touch screen capacitivo sotto vetro per TFT - Opto Forum 2009-25 Marzo 2009

Data Modul. QMatrix Multi touch: touch screen capacitivo sotto vetro per TFT - Opto Forum 2009-25 Marzo 2009 Data Modul QMatrix Multi touch: touch screen capacitivo sotto vetro per TFT - Opto Forum 2009-25 Marzo 2009 Munich Weikersheim 1 Manufacturer, integrator, distributor, worldwide partner - Our Brands -

More information

Agilent Split Post Dielectric Resonators for Dielectric Measurements of Substrates. Application Note

Agilent Split Post Dielectric Resonators for Dielectric Measurements of Substrates. Application Note Agilent Split Post Dielectric Resonators for Dielectric Measurements of Substrates Application Note l Introduction The split post dielectric resonator (SPDR) provides an accurate technique for measuring

More information

Pet hair clipper. Model 96822. Diagrams within this manual may not be drawn proportionally.

Pet hair clipper. Model 96822. Diagrams within this manual may not be drawn proportionally. Pet hair clipper Model 96822 Cleaning And Operation Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly from the

More information

Metrology and Calibration of Mechatronic Systems. Metrology and Calibration of Mechatronic Systems overview (2014-12-08) Adrian M.

Metrology and Calibration of Mechatronic Systems. Metrology and Calibration of Mechatronic Systems overview (2014-12-08) Adrian M. Metrology and Calibration of Mechatronic Systems Mechatronics Training Curriculum Premium Advanced Feedforward Control (2 days) Workshop Mechatronics System Design (5d) Advanced Advanced Motion Control

More information

MAGNA-FLO SAND FILTER SIDE MOUNTED MULTIPORT VALVE

MAGNA-FLO SAND FILTER SIDE MOUNTED MULTIPORT VALVE MAGNA-FLO SAND FILTER SIDE MOUNTED MULTIPORT VALVE Assembly & Operating Instructions F.G Series: 550mm - 21 650mm - 25 710mm - 28 A. Assembly 1. Prepare a firm base for the sand filter, capable of withstanding

More information

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE)

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE) SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE) We greatly appreciate your purchase of this Tamron lens. The Tamron SP AF 300mm F/2.8 LD [IF] is a fast telephoto lens developed for Canon AF single-reflex

More information

Understanding Infrared Camera Thermal Image Quality

Understanding Infrared Camera Thermal Image Quality Electrophysics Resource Center White Paper Noise{ Clean Signal Understanding Infared Camera Electrophysics Resource Center: Abstract You ve no doubt purchased a digital camera sometime over the past few

More information

HOW TO MAKE A MOTOR BRACKET

HOW TO MAKE A MOTOR BRACKET HOW TO MAKE A MOUNTING BRACKET FOR AN ELECTRIC MOTOR Often, brackets supplied with purchased items (e.g. electric motors or servos) are not ideally suited for fitting into our models or they are not supplied

More information

EasyTrack 2 Accessories List

EasyTrack 2 Accessories List EasyTrack 2 Accessories List DATA LOGGER ET6061 EasyTrack2 Data Logger (6 Channel) Measurement Range -150 to 500 C (-238 F to 932 F) Accuracy ±0.5 C (0.9 F). Type K thermocouple. 6000 readings per channel.

More information

Installation Guide. DEVIreg 530. Electronic Thermostat. www.devi.com

Installation Guide. DEVIreg 530. Electronic Thermostat. www.devi.com Installation Guide DEVIreg 530 Electronic Thermostat www.devi.com The English language is used for the original instructions. Other languages are a translation of the original instructions. (Directive

More information

Clinic Washer-Disinfectors. Washer-Disinfector Line

Clinic Washer-Disinfectors. Washer-Disinfector Line Clinic Washer-Disinfectors Washer-Disinfector Line Easy-to-Use Washer-Disinfectors for Clinics The Clinic Range of Washer Disinfectors provides superior cleaning and disinfection combined with innovative

More information

Experiment 446.1 SURFACE TENSION OF LIQUIDS. Experiment 1, page 1 Version of June 17, 2016

Experiment 446.1 SURFACE TENSION OF LIQUIDS. Experiment 1, page 1 Version of June 17, 2016 Experiment 1, page 1 Version of June 17, 2016 Experiment 446.1 SURFACE TENSION OF LIQUIDS Theory To create a surface requires work that changes the Gibbs energy, G, of a thermodynamic system. dg = SdT

More information