The Impact of Hydrocarbon Dew Point on Sample Conditioning for BTU Analysis
|
|
- Lorin McBride
- 7 years ago
- Views:
Transcription
1 The Impact of Hydrocarbon Dew Point on Sample Conditioning for BTU Analysis Shane Hale, Emerson Process Management Introduction No matter how good the analyzer is, the results can only be as good as the sample it analyzes. If the composition of the gas changes between the sample point and the analyzer, the analysis and the resulting calculations (such as energy flow) will not be representative of what is flowing through the line. The most common cause of the composition changing in the sample handling system is related to the hydrocarbon dew point (HCDP) of the sample. If at any stage the natural gas is allowed to go below the HCDP, the heavy components will drop out of the gas phase (and into the liquid phase) and the composition of the gas will change. Understanding the impact of HCDP and sample composition is critical for designing sample handling systems. Understanding the different methods for determining the HCDP and using it in practical situations can provide operational tools to improve measurement and protect against sample handling issues. Determining the HCDP The first step to designing a sample handling system is determining what the HCDP of the sample will be. Typically, an equation of state is used to determine what the theoretical HCDP of a composition is. When using an equation of state to determine the HCDP, the composition used is very important. Often the typical or nominal composition is used to design the sample system. However, the system should be designed for the worst case scenario. An analogy is the seat-belt. The seat-belt is designed to perform correctly when things are going wrong, not when everything is going well. The sample system should also be designed to perform when an upset or abnormal condition results in dirty or heavy sample and protect the analyzer while maintaining the composition so the abnormal mixture is correctly reported. The API 14.1 standard recommends that the natural gas sample should be maintained at least 30 F above the expected HCDP (API 14.1, 2006, Section 6.6). Users should use the gas with the highest HCDP as a reference point and then design the system to maintain the sample 30 F above that. Many people assume the HCDP tracks the BTU content. Figure 1 shows that for a large sample of more than 1000 natural gas analysis there is a loose relationship of higher cricondentherm with BTU; however, the relationship is not strong enough to use as a basis of design of the sample system. Figure 1 - BTU and Cricondentherm chart showing a loose relationship of BTU to the cricondentherm. Pipeline Quality Gas Transmission For pipeline quality gas transmission measurement, the large number of interconnects in the transmission system means the original source of the gas is often unknown and impossible to determine. A commonly used
2 tariff specification for HCDP in the United States is 15 F (-9.4 C), so one could assume this as the HCDP for the worst case scenario. However, this tariff is not used in all locations, is often not continuously monitored and thus not enforced, and producers are usually given relief to keep injecting off-spec gas while they bring their process under control. These factors result in the probability that gas with a HCDP greater than the tariff specification will be sampled at some stage. By definition, the HCDP of the gas in the pipeline must be at or below the flowing gas temperature and sample probes should be designed and located to only sample the gas phase. Using the flowing gas temperature is therefore a better basis to design the sample system to account for the car-crash occasions when the sample goes outside of the normally operating conditions. Most pipelines will have flowing temperatures lower than 90 F (32 C), so this can be used as the expected highest HCDP gas. Adding 30 F to this temperature will result in an API 14.1 compliant sample system temperature of 120 F (~50 C). If pipeline temperatures greater than 90 F (32 C) are expected, the sample system design temperature should be altered to suit. Raw Production Gas The raw gas directly from the well can be significantly richer and contain more contaminates than pipeline quality gas. Very often, the stream is two-phase or very close to the HCDP. Special attention must be paid to removing the sample with as little liquid hydrocarbon as possible and removing the liquids at the same pressure and temperature as the flowing stream. If the temperature or pressure of the sample is changed prior to removing the liquids, then the composition of the gas phase will change and make any resultant analysis unrepresentative of the stream gas. The easiest way to achieve this is to reject the liquids at the inside the stream using sample probes with a liquid-membrane tip. Measured versus Theoretical HCDP The determination of the HCDP falls into two distinct methods. The measured HCDP method relies on cooling the gas sample until the formation of liquids can be detected in a chilled-mirror analyzer. The theoretical method uses the composition of the gas and an equation of state (EOS) to calculate the theoretical HCDP. With the chilled-mirror method, the formation of hydrocarbon liquids is detected on a mirror either by an operator (a manual HC dew-scope) or light-based detector technology. The accuracy of this method is determined by the rate of cooling of the mirror, and the repeatability of the hydrocarbon mist detection. The results of manual dewscopes are heavily dependent on the experience (and patience) of the operator, but the latest generation of automated methods are highly repeatable. Figure 2 - Cricondentherm pressure for samples with cricondentherm between 0 F and 40 F using the PR EOS. The chilled-mirror is a physical measurement so will provide only the HCDP at the pressure at which the measurement is conducted. Typically the pressure on an automatic chilled-mirror system will be set at 400 PSI (2,757 kpa) that approximates the usual cricondentherm pressure (Figure 3) for pipeline quality natural gas so the results are close to the calculated cricondentherm. However, there is no method to extrapolate this result to other pressures (such as the pipeline pressure or delivery pressure). Additionally, a visible mist must be formed on the mirror for the detection circuit to register the dew point, which means that liquid hydrocarbons have already
3 dropped out before the detection circuit is tripped. For this reason, the results of the physical measurement of HCDP will be lower than those calculated from the theoretical models. Another practical factor for consideration is that a chilled-mirror device is a dedicated analyzer for a single measurement and the user must determine if the additional cost of the analyzer and associated sample handling system can be justified. An alternative is to use the theoretical EOS method to calculate the HCDP using the composition of the gas. The composition is almost always reported using a gas chromatograph (online or offline) for the determination of gas quality, energy, compressibility, and other physical properties, so extending the use of the composition for the determination of HCDP can be an attractive option. Several equations of state (EOS) can be used for calculating the theoretical HCDP for natural gas mixtures with the most common being the Peng-Robinson (PR) and the Soave-Redlich-Kwong (SRK). The SRK equation will tend to calculate values about 5 F (3 C) higher than the PR equation (refer Figure 3), but the PR is generally accepted to calculate values more representative of typical natural gas mixtures. Figure 3 - Difference between the cricondentherm from the SRK and PR Equations of state for samples between 0 F and 40 F. The advantage of using the theoretical method is it allows the calculation of the cricondentherm temperature and pressure as well as the HCDP at any pressure. Gas chromatographs are available that will calculate the HCDP at four user set pressures and the cricondentherm temperature and pressure. However, the level of analysis is very important to determining the HCDP with reasonable accuracy. The typical gas chromatograph used in natural gas custody transfer applications performs as C6+ in which the individual hydrocarbon species up to normal-pentane, nitrogen, and carbon dioxide is measured, and any hydrocarbon larger than n-pentane is lumped together as C6+. For the determination of energy content and other physical properties, the C6+ value is then mathematically split into fractions of hexanes, heptanes, and octanes. This is acceptable for energy content, relative density, compressibility, and the other typical calculations as the C6+ fraction is very small (less than 0.1%) and using fixed assumptions for these components results in very small errors. However, the concentration of the heavier hydrocarbons dramatically affects the hydrocarbon dew point value as it is the heavy components that will drop out into the liquid phase. Using fixed ratios for the determination of the C6, C7, and C8 concentrations will result in large errors the HCDP calculation, well beyond practical use (Error! Reference source not found.). Extended analysis gas chromatographs capable of measuring the heavier hydrocarbon species permit a much more accurate calculation of the HCDP. In many laboratories, measurement of the hydrocarbons up to C12 or even C14 is conducted using a Thermal Conductivity Detector (TCD) for the C1 to C5 measurement, and a Flame Ionization Detector (FID) for the heavier hydrocarbons. The analysis cycle time can be from 10 minutes to over 40 minutes that is usually not practical for online use in custody transfer where the typical C6+ analysis is between three and five minutes.
4 Figure 4 - The HCDP calculation errors from using a C6+ fixed ratio split to determine the C6, C7 and C8 values. Online gas chromatographs that use parallel analysis paths, dual TCDs, and analysis times of five minutes to measure up to C9+ have been available for more than ten years. This level of extended analysis provides an alternative to the poor HCDP determination using the C6+ analysis, and the excessively long analysis times for C12 or C14 analysis found in the laboratory. Using the C9+ analysis to determine the HCDP results in a more accurate and consistent values that can be used practically in the field. Further accuracy can be achieved using fixed fractions of C9, C10, and C11 for the C9+ component that can be based on experience or routine extended laboratory analysis. Using HCDP in Operations Whereas the main purpose for HCDP determination is for monitoring the gas for tariff specifications, the online calculation of the HCDP can also provide useful diagnostics for measurement operations. Flow meters do not measure the flow accurately if there is two-phase flow. The drop-out of heavy hydrocarbons causes loss-andunaccounted-for errors across the pipeline as the energy value of the gas exiting the pipeline is less than the energy value of the gas entering the network. Liquid hydrocarbons can result in hydrate formation and other operational issues. Determining the HCDP online provides an effective tool to mitigate these issues. Using the stream pressure as one of the pressures for the calculation of the HCDP (through an analog input or Modbus serial link) provides a useful indicator of potential problems. For example, if the calculated HCDP is the equal to or higher than the pipeline temperature, it is a sign that there are liquid hydrocarbons in the pipeline that should be addressed by operations. Figure 5 shows a real-world example of where the stream was two-phase when the sample was extracted. In this situation, the sample probe did not exclude liquids, and the sample was heated prior to the pressure reduction. The reported composition will be incorrect as some liquids have been
5 sampled and then vaporized by the sample system, resulting in a sample that is analyzed by the gas chromatograph richer than the gas phase in the pipeline resulting in an over-reporting of energy content. Figure 5 - Example of a sample taken when the stream was two phase. The incorrect composition will result in errors in the energy and other physical properties calculations leading to incorrect fiscal reporting. Additionally, the flow-meter may also read incorrectly because of the interference of the liquid hydrocarbons in the measurement (depending on the technology used). Modern ultrasonic meters can determine when there is two-phase flow from the flow profile and other diagnostic indicators and provide an alarm. However, trending the HCDP from an online GC allows operations to determine when the HCDP is being approached and take measures to avoid the issue escalating to a two-phase flow condition (such as shutting in a rich gas supply). Conclusion When designing a sample system, knowing what the HCDP of the sample can be expected to be will help determine what components are required to ensure a good sample reaches the analyzers. The online analysis with an online HCDP analyzer provides an accurate and repeatable determination for HCDP for use in reporting gas quality tariff values. However, the determination of HCDP at the pipeline pressure and the cricondentherm can only be done using the theoretical EOS models. The typical C6+ compositional analysis used in natural gas custody transfer is not sufficient to calculate the HCDP with enough accuracy, as the heavy components higher than normal-pentane affect the HCDP dramatically and assuming them from a fixed ratio of the C6+ value introduces excessive errors. The extended C12 or C14 analysis available in the lab provide accurate results, but the extended times are impractical for use in online custody transfer. Using an online C9+ analysis provides a compromise that permits an accurate determination of the cricondentherm and the HCDP at flowing conditions. Using the HCDP value at flowing conditions and comparing to the stream temperature can determine if the sample handling system is changing the composition of the gas, can be used to avoid two-phase flow, and thus avoid measurement and operational issues.
Fundamentals of Gas Chromatography
Application Note Oil & Gas Fundamentals of Gas Chromatography Figure 1 - The Function Components of a Gas Chromatograph Process Line Sample System Chromatograph Oven GC Controller Probe Sample Return Slip
More informationChromatograph Maintenance and Troubleshooting Class # 5050
Chromatograph Maintenance and Troubleshooting Class # 5050 Shane Hale Business Development Manager Emerson Process Management Gas Chromatographs Division 5650 Brittmoore Drive Houston, TX 77041 USA Introduction
More informationFiscal Measurement Natural Gas
White Paper FSG-WP-0012 February 2014 Fiscal Measurement Natural Gas Fiscal Measurement How much? Fiscal Measurement must not be confused with Custody Transfer; in fact, fiscal measurement is a more general
More informationNatural Gas Processing Solutions Using the 700XA Gas Chromatograph
Application Note Oil & Gas Natural Gas Processing Solutions Using the 700XA Gas Chromatograph Optimizing the efficiency and profitability of a natural gas processing plant requires responding to the changing
More informationBTU ANALYSIS USING A GAS CHROMATOGRAPH
BTU ANALYSIS USING A GAS CHROMATOGRAPH By Joseph E. Landes Southern Petroleum Laboratories, Inc. Technical Director Hydrocarbon and Field Services 8820 Interchange Drive Houston, Texas 77054 Introduction
More informationPRACTICAL HYDROCARBON DEW POINT SPECIFICATION FOR NATURAL GAS TRANSMISSION LINES
PRACTICAL HYDROCARBON DEW POINT SPECIFICATION FOR NATURAL GAS TRANSMISSION LINES Jerry A. Bullin and Carl Fitz Bryan Research & Engineering, Inc. Bryan, Texas, U.S.A. Todd Dustman Questar Pipeline Company
More informationLESSONS LEARNED FROM THE API 14.1 GAS SAMPLING RESEARCH PROJECT. Darin L. George and Eric Kelner
LESSONS LEARNED FROM THE API 14.1 GAS SAMPLING RESEARCH PROJECT Darin L. George and Eric Kelner Southwest Research Institute 6220 Culebra Road San Antonio, Texas, USA Introduction Since 1999, the Gas Technology
More informationOptimization of Natural Gas Processing Plants Including Business Aspects
Page 1 of 12 Optimization of Natural Gas Processing Plants Including Business Aspects KEITH A. BULLIN, Bryan Research & Engineering, Inc., Bryan, Texas KENNETH R. HALL, Texas A&M University, College Station,
More informationTECHNIQUES FOR NATURAL GAS SAMPLING A DISCUSSION OF FIELD METHODS FOR OBTAINING SPOT SAMPLES. Randall Messman SW Regional Sales Manager
TECHNIQUES FOR NATURAL GAS SAMPLING A DISCUSSION OF FIELD METHODS FOR OBTAINING SPOT SAMPLES Randall Messman SW Regional Sales Manager PGI International 16101 Vallen Drive Houston, TX 77041 Purpose Natural
More informationEmerson Gas Chromatograph Solutions For Natural Gas Transmission Applications. Superior Performance, Repeatability, and Reliability
Emerson Gas Chromatograph Solutions For Natural Gas Transmission Applications Superior Performance, Repeatability, and Reliability Struggling to Reduce Costs While Meeting Tighter Energy Measuremen What
More informationTHE HUMIDITY/MOISTURE HANDBOOK
THE HUMIDITY/MOISTURE HANDBOOK Table of Contents Introduction... 3 Relative Humidity... 3 Partial Pressure... 4 Saturation Pressure (Ps)... 5 Other Absolute Moisture Scales... 8 % Moisture by Volume (%M
More informationA TRACEABLE CALIBRATION PROCEDURE FOR HYDROCARBON DEW POINT METERS
23rd World Gas Conference, Amsterdam 2006 A TRACEABLE CALIBRATION PROCEDURE FOR HYDROCARBON DEW POINT METERS Main author H.J. Panneman N.V. Nederlandse Gasunie Gasunie Engineering & Technology Groningen,
More information1. Standard conditions are 29.92 in. Hg (760 mm Hg, 14.696 psia) and 68 F (20 C).
INTRODUCTION Emission Monitoring Inc. DETERMINING F-FACTORS FROM GAS CHROMATOGRAPHIC ANALYSES Roger T. Shigehara Emission Monitoring Incorporated Howard F. Schiff TRC Environmental Corporation EPA Method
More information23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
More informationPreparation of Natural Gas Blends Used as Calibration Standards: Sources of Uncertainty and Best Preparation Practices
Preparation of Natural Gas Blends Used as Calibration Standards: Sources of Uncertainty and Best Preparation Practices FINAL REPORT Prepared by: D. L. George SOUTHWEST RESEARCH INSTITUTE Mechanical and
More informationANNEX I GAS MEASUREMENTS AND TESTING AT THE DELIVERY POINT. Table of Contents
ANNEX I GAS MEASUREMENTS AND TESTING AT THE DELIVERY POINT Table of Contents 1. Purpose... 1 2. Scope... 1 3. Definitions, Symbols and Abbreviations... 2 4 General... 6 5. Natural Gas Volumetric Flow Measurement...
More informationProcess Instrumentation
Process Instrumentation Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Process Gas Chromatography Features of Process Chromatographs Agilent 7890A
More informationTemperature. Number of moles. Constant Terms. Pressure. Answers Additional Questions 12.1
Answers Additional Questions 12.1 1. A gas collected over water has a total pressure equal to the pressure of the dry gas plus the pressure of the water vapor. If the partial pressure of water at 25.0
More informationThermal Conductivity Detector
Thermal Conductivity Detector Troubleshooting Tips The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components)
More informationCRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT
CRYOCOOLER CONTAMINATION STUDY S.W.K. Yuan, D.T.Kuo, and A.S. Loc EI Technologies, Cryocooler Group Sylmar, CA 91342 ASTRACT Eliminating or reduction of contamination is essential for the success of cryocoolers
More informationTHE BASICS Q: What is VOC? Q: What are flashing losses/voc emissions from hydrocarbon storage tanks? - 1 -
Calculation of Flashing Losses/VOC Emissions from Hydrocarbon Storage Tanks THE BASICS Q: What is VOC? A: VOC is an acronym that stands for Volatile Organic Compounds. VOC are components of hydrocarbon
More informationThe photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
More informationOIL & GAS PRODUCTION AUDITING
OIL & GAS PRODUCTION AUDITING Wayne C. Dussel Offshore Measurement Supervisor El Paso Production GOM Basic Measurement Audit Steps I. Select location (s) to be reviewed. II. Review contracts, commingling
More informationMeasurements and Modelling of Hydrocarbon Dew Points for Natural Gases
Measurements and Modelling of Hydrocarbon Dew Points for Natural Gases Eirini Skylogianni Natural Gas Technology Submission date: May 2013 Supervisor: Even Solbraa, EPT Co-supervisor: Efstathios Skouras
More informationProcess gas chromatographs. Measurement made easy. Compliance Solutions for 40CFR60 Subpart Ja
Process gas chromatographs. Measurement made easy Compliance Solutions for 40CFR60 Subpart Ja Introduction ABB offers a proven analytical solution to completely meet your Refinery Flare Monitoring requirements
More informationHSE information sheet. Fire and explosion hazards in offshore gas turbines. Offshore Information Sheet No. 10/2008
HSE information sheet Fire and explosion hazards in offshore gas turbines Offshore Information Sheet No. 10/2008 Contents Introduction.. 2 Background of gas turbine incidents in the UK offshore sector...2
More informationChemical Analysis. Natural Gas Analyzer. A Family of optimized GC Solutions. Gas Chromatography. think forward
Chemical Analysis Natural Gas Analyzer A Family of optimized GC Solutions think forward Gas Chromatography Bruker Natural Gas Analyzers Natural gas is bought and sold as a bulk commodity with price based
More informationAccurate Air Flow Measurement in Electronics Cooling
Accurate Air Flow Measurement in Electronics Cooling Joachim Preiss, Raouf Ismail Cambridge AccuSense, Inc. E-mail: info@degreec.com Air is the most commonly used medium to remove heat from electronics
More informationOn-line Dissolved Gas Analysis
On-line Dissolved Gas Analysis accord. to IEC 567/ASTM 3612 Online-Monitoring of transformers Automatic on-line analysis of 11 gases The analysis is done in 2 steps with high selectivity and accuracy fully
More informationIAPWS Certified Research Need - ICRN
IAPWS Certified Research Need - ICRN ICRN 23 Dew Point for Flue Gas of Power-Plant Exhaust The IAPWS Working Group Industrial Requirements and Solutions has examined the published work in the area of dew-point
More informationCalculate Available Heat for Natural Gas Fuel For Industrial Heating Equipment and Boilers
For Industrial Heating Equipment and Boilers Prepared for California Energy Commission (CEC) Prepared By: Southern California Gas Company (A Sempra Energy Utility) E3M Inc. May 2012 i Disclaimer The CEC
More informationBeware of Exaggerated Claims and Uncertified Manufacturers Ratings Regarding Commercial Boiler Efficiency
Beware of Exaggerated Claims and Uncertified Manufacturers Ratings Regarding Commercial Boiler Efficiency For many years competent professional HVAC engineers have been aware of exaggerated or deceptive
More informationCustody Transfer Measurement. with the V-Cone Flowmeter
Custody Transfer Measurement with the V-Cone Flowmeter Stephen A. Ifft McCrometer Inc. Hemet, California, USA Abstract This paper will discuss the approval of the McCrometer V-Cone flowmeter for custody
More information[Elliot and Lira, introduction to Chemical EngineeringThermodynamics, Prentice Hall, 1999].
ASPEN HYSYS- fluid package selection Δένδπο Αποθάζεων για επιλογή θεπμοδςναμικού μονηέλος BIP: mean binary interaction parameters [Elliot and Lira, introduction to Chemical EngineeringThermodynamics, Prentice
More informationECONOMICAL OPTIONS FOR RECOVERING NGL / LPG AT LNG RECEIVING TERMINALS
ECONOMICAL OPTIONS FOR RECOVERING NGL / LPG AT RECEIVING TERMINALS Presented at the 86 th Annual Convention of the Gas Processors Association March 13, 2007 San Antonio, Texas Kyle T. Cuellar Ortloff Engineers,
More informationChapter 4. Chemical Energy
hapter 4 hemical Energy Perhaps the most convenient form in which to store energy is chemical energy. The foods we eat, combined with the oxygen we breathe, store energy that our bodies extract and convert
More informationFractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
More informationControl Device Requirements Charts For Oil and Gas Handling and Production Facilities
Device Charts For Oil and Gas Handling and Production Facilities Purpose/Scope: The purpose of this document is to provide standardized guidance for use by the regulated community and air permit reviewers,
More informationEthanol-Water Phase Separation White Paper
Ethanol-Water Phase Separation White Paper Samir Jain Background The use of ethanol as a fuel in the United States has significantly increased over the past decade. Today, greater than 80% of all retail
More informationApplication of the Orifice Meter for Accurate Gas Flow Measurement page 1. Application of the Orifice Meter for Accurate Gas Flow Measurement.
Application of the Orifice Meter for Accurate Gas Flow Measurement page 1 DANIEL MEASUREMENT AND CONTROL WHITE PAPER Application of the Orifice Meter for Accurate Gas Flow Measurement www.daniel.com Summary
More informationThe Use of Hydrogen Carrier Gas for GC/MS
Technical Note The Use of Hydrogen Carrier Gas for GC/MS Gas Chromatography/ Mass Spectrometry Highlights Guidelines that can mitigate dangers and leverage benefits of using hydrogen Key safety factors
More informationDissolved Gas Analysis Guide for Transformers Filled with Beta Fluid
DSI Ventures, Inc. PHONE: (903) 526-7577 FAX: (903) 526-0021 www.dsiventures.com Dissolved Gas Analysis Guide for Transformers Filled with Beta Fluid Introduction Analysis of dissolved gases in transformer
More informationDaniel Danalyzer Model 700 Gas Chromatograph
Product Data Sheet Model 700 Daniel Danalyzer Model 700 Gas Chromatograph Danalyzer Model 700 Gas Chromatographs offer the broadest range of analysis options available today in a fieldmount gas chromatograph
More informationInternal Combustion Optical Sensor (ICOS)
Internal Combustion Optical Sensor (ICOS) Optical Engine Indication The ICOS System In-Cylinder Optical Indication 4air/fuel ratio 4exhaust gas concentration and EGR 4gas temperature 4analysis of highly
More informationOptimize Pipeline Hydraulics with Multiphase Flow Modeling
Optimize Pipeline Hydraulics with Multiphase Flow Modeling Below are questions asked by attendees during the webinar on February 22, 2012, followed by answers provided by our presenters. Will you also
More informationMETHOD 511.4 EXPLOSIVE ATMOSPHERE
EXPLOSIVE ATMOSPHERE NOTE: Tailoring is essential. Select methods, procedures, and parameter levels based on the tailoring process described in Part One, paragraph 4.2.2, and Appendix C. Apply the general
More informationMeasurement & Analytics. VIS Multiphase Flow Meter High gas content applications solved with no radioactive source - Measurement made easy
Measurement & Analytics VIS Multiphase Flow Meter High gas content applications solved with no radioactive source - Measurement made easy ABB s VIS Multiphase Flow Meter Easy to operate, accurate and fieldproven
More informationGC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
More informationPesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationRate Equations and Detailed Balance
Rate Equations and Detailed Balance Initial question: Last time we mentioned astrophysical masers. Why can they exist spontaneously? Could there be astrophysical lasers, i.e., ones that emit in the optical?
More informationBasic Hydraulics and Pneumatics
Basic Hydraulics and Pneumatics Module 1: Introduction to Pneumatics PREPARED BY IAT Curriculum Unit March 2011 Institute of Applied Technology, 2011 ATM 1122 Basic Hydraulics and Pneumatics Module 1:
More informationApplication Requirement
Hazardous Gas Monitors Sensor Selection Overview Safety Monitoring (LEL) Toxic Limit Detection (PEL) Leak Detection Personal Safety Application Requirement Exposure Assessment (TWA) Ambient Air Quality
More information!"#$ Reservoir Fluid Properties. State of the Art and Outlook for Future Development. Dr. Muhammad Al-Marhoun
Society of Petroleum Engineers SPE 2001 2002 Distinguished Lecturer Program 4 July 2002 Reservoir Fluid Properties State of the Art and Outlook for Future Development Dr. Muhammad Al-Marhoun King Fahd
More informationEffects of Ultra-High Turndown in Hydronic Boilers By Sean Lobdell and Brian Huibregtse January 2014
Effects of Ultra-High Turndown in Hydronic Boilers By Sean Lobdell and Brian Huibregtse January 2014 Today, hydronic boilers are producing greater efficiencies than previously thought possible, all due
More informationThermodynamics - Example Problems Problems and Solutions
Thermodynamics - Example Problems Problems and Solutions 1 Examining a Power Plant Consider a power plant. At point 1 the working gas has a temperature of T = 25 C. The pressure is 1bar and the mass flow
More informationGas Detection for Refining. HA University
Gas Detection for Refining HA University Refinery Process and Detection Needs Refining i Crude Oil Final Products Coke Asphalt Waxes, Lubricating Oils and Greases Kerosene, Jet Fuel, Diesel Fuel, Home
More informationThe DL8000 Preset Controller. The Ideal Device for Controlling and Measuring the Loading of Liquid Hydrocarbons into Transporting Tankers
The DL8000 Preset Controller The Ideal Device for Controlling and Measuring the Loading of Liquid Hydrocarbons into Transporting Tankers Reliability, ruggedness, flexibility, and accuracy all in one package
More informationCOMPRESSED GASES. 1.2 The contents of each cylinder and container must be clearly identified (by tag or stamp) on the cylinder.
Page 1 of 5 COMPRESSED GASES A compressed gas is defined as any mixture of gases in a container with a pressure exceeding 40 psi. at 70 o F, or 104 psi. at 130 o F; or any flammable liquid with an absolute
More informationAPPLIED THERMODYNAMICS TUTORIAL 1 REVISION OF ISENTROPIC EFFICIENCY ADVANCED STEAM CYCLES
APPLIED THERMODYNAMICS TUTORIAL 1 REVISION OF ISENTROPIC EFFICIENCY ADVANCED STEAM CYCLES INTRODUCTION This tutorial is designed for students wishing to extend their knowledge of thermodynamics to a more
More informationCarbon Dioxide Membrane Separation for Carbon Capture using Direct FuelCell Systems
Carbon Dioxide Membrane Separation for Carbon Capture using Direct FuelCell Systems DFC Technology Used as Electrochemical Membrane for CO 2 Purification and Capture during Power Generation FCE s Direct
More information1 Introduction The Scientific Method (1 of 20) 1 Introduction Observations and Measurements Qualitative, Quantitative, Inferences (2 of 20)
The Scientific Method (1 of 20) This is an attempt to state how scientists do science. It is necessarily artificial. Here are MY five steps: Make observations the leaves on my plant are turning yellow
More informationDanalyzer Model 700 Gas Chromatograph
Danalyzer Model 700 Gas Chromatograph Page 1 www.daniel.com Danalyzer Model 700 Gas Chromatograph Danalyzer Model 700 Gas Chromatographs offer the broadest range of analysis options available today in
More informationGulf Coast Conference Prepared by: A. Coderre (Clearstone Engineering Ltd)
APPLICATIONS OF FAST TRANSPORTABLE CHROMATOGRAPHY Gulf Coast Conference Prepared by: A. Coderre (Clearstone Engineering Ltd) Introduction: Adam Coderre M.A.Sc., Mechanical Engineering Specializing in emissions
More informationLECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS
LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS 1. Define an accumulator and explain its function A hydraulic accumulator is a device that stores the potential energy of an incompressible fluid
More informationUNDERSTANDING REFRIGERANT TABLES
Refrigeration Service Engineers Society 1666 Rand Road Des Plaines, Illinois 60016 UNDERSTANDING REFRIGERANT TABLES INTRODUCTION A Mollier diagram is a graphical representation of the properties of a refrigerant,
More informationConductivity Sensor Calibrations to Meet Water Industry Requirements
Conductivity Sensor Calibrations to Meet Water Industry Requirements Victor M. Braga Technical Service/Training Manager Mettler-Toledo Thornton Inc. 36 Middlesex Turnpike Bedford, Massachusetts 01730 Phone
More informationSpecifications for Refrigerants
AHRI Standard 700 2015 Standard for Specifications for Refrigerants IMPORTANT SAFETY DISCLAIMER AHRI does not set safety standards and does not certify or guarantee the safety of any products, components
More informationCalculating Volatile Organic Compounds (VOC) Flash Emissions from Crude Oil and Condensate Tanks at Oil and Gas Production Sites
Air Permit Reference Guide APDG 5942 Calculating Volatile Organic Compounds (VOC) Flash Emissions from Crude Oil and Condensate Tanks at Oil and Gas Production Sites Air Permits Division Texas Commission
More informationHow To Make A High Co 2 Gas Blend
ECONOMICAL OPTION FOR CO 2 / METHANE SEPARATION IN PRODUCED GAS CONTAINING A HIGH CO 2 FRACTION F. Patrick Ross, P.E. TPR Consulting 9907 Sagecourt Drive Houston, Texas 77089 (713) 870-9208 pat.ross@att.net
More informationGravimetric determination of pipette errors
Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with
More informationEXPERIMENT 13: THE IDEAL GAS LAW AND THE MOLECULAR WEIGHT OF GASES
Name Section EXPERIMENT 13: THE IDEAL GAS LAW AND THE MOLECULAR WEIGHT OF GASES PRE-LABORATORY QUESTIONS The following preparatory questions should be answered before coming to lab. They are intended to
More informationSiemens AG, 2008. Case Study December 2008
Siemens AG, 2008 Process Analytics throughout the entire natural gas pipeline supply chain Case Study December 2008 Natural Gas (NG) Monitoring NG during pipeline transmission Natural Gas is a vital component
More informationSelect the Right Relief Valve - Part 1 Saeid Rahimi
Select the Right Relief Valve - Part 1 Saeid Rahimi 8-Apr-01 Introduction Selecting a proper type of relief valve is an essential part of an overpressure protection system design. The selection process
More informationROC800L Liquid Application Software
ROC800:SW1 Product Data Sheet D301576X012 January 2015 ROC800L Liquid Application Software The ROC800L Liquid Application Software Suite enables the ROC800 platform to perform API compliant liquid flow
More informationThe Empirical Formula of a Compound
The Empirical Formula of a Compound Lab #5 Introduction A look at the mass relationships in chemistry reveals little order or sense. The ratio of the masses of the elements in a compound, while constant,
More informationDynamic Modeling Of An Ozone Disinfection Facility
Dynamic Modeling Of An Ozone Disinfection Facility Michael Etheridge Black & Veatch 8400 Ward Parkway Kansas City, MO 64114 Presented to the Fourteenth Ozone World Congress August 25 th, 1999 ABSTRACT
More informationHow does solar air conditioning work?
How does solar air conditioning work? In a conventional air conditioning system; The working fluid arrives at the compressor as a cool, low-pressure gas. The compressor is powered by electricity to squeeze
More informationASimple Guide to Oil Refining
ASimple Guide to Oil Refining We all know that motor oil and gasoline come from crude oil. What many people do not realize is that crude oil is also the starting point for many diverse products such as
More informationEquations of State. Equations of State (EoS)
Equations of State (EoS) Equations of State From molecular considerations, identify which intermolecular interactions are significant (including estimating relative strengths of dipole moments, polarizability,
More informationDENSITY MEASUREMENTS OF LIQUID FUELS TO DETERMINE TEMPERATURE CONVERSION FACTORS FOR LEGAL METROLOGY
XX IMEKO World Congress Metrology for Green Growth September 9 14, 2012, Busan, Republic of Korea DENSITY MEASUREMENTS OF LIQUID FUELS TO DETERMINE TEMPERATURE CONVERSION FACTORS FOR LEGAL METROLOGY H.
More informationValidation of process gas systems
Validation of Process Gas Systems By Jeff Hargroves Alza Corporation Validation of process gas systems involve documenting the expected system behavior, and verifying that the system performs as expected.
More information48 Practice Problems for Ch. 17 - Chem 1C - Joseph
48 Practice Problems for Ch. 17 - Chem 1C - Joseph 1. Which of the following concentration measures will change in value as the temperature of a solution changes? A) mass percent B) mole fraction C) molality
More informationNatural Gas Information Contents
Natural Gas Information Contents What is natural gas Natural Gas Components Physical Properties of Natural Gas Different Forms of Natural Gas The Use of Natural Gas Co-generation System Natural Gas and
More informationPetrochemical Industry Ethylene Plant
Petrochemical Industry Ethylene Plant Application note The Ethylene Plant provides the base feedstock for a number of critical products used throughout the petrochemical industry. Compounds ranging from
More informationNational Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW
NPL REPORT AS 3 Comparison of methods for the measurement of hydrocarbon dew point of natural gas Andrew Brown Martin Milton Gergely Vargha Richard Mounce Chris Cowper Andrew Stokes Andy Benton Mike Bannister
More informationFUNDAMENTALS OF ENGINEERING THERMODYNAMICS
FUNDAMENTALS OF ENGINEERING THERMODYNAMICS System: Quantity of matter (constant mass) or region in space (constant volume) chosen for study. Closed system: Can exchange energy but not mass; mass is constant
More informationSEPARATION OF A MIXTURE OF SUBSTANCES LAB
SEPARATION OF A MIXTURE OF SUBSTANCES LAB Purpose: Every chemical has a set of defined physical properties, and when combined they present a unique fingerprint for that chemical. When chemicals are present
More informationOil and Gas Custody Transfer
By Emerie Dupuis Daniel Measurement and Control, Emerson Process Management Oil and Gas Custody Transfer When money changes hands, flow measurement accuracy matters. When flow measurement is performed
More informationRefinery Equipment of Texas. Mini - Refinery Feasibility Overview
Mini - Refinery Feasibility Overview Introduction This paper is intended to provide information, answer questions, and assist the owner or project developer in making informed buying decisions. A mini-refinery
More informationTurbine Inlet Cooling
Section 2 Turbine Inlet Cooling Reference Charts 32 Relative system costs for an F-class combined cycle 33 Change in power output with ambient air temperature 34 Example of MS7001 design heat rate, flow
More informationSpectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
More informationNetwork-Wide Change Management Visibility with Route Analytics
Network-Wide Change Management Visibility with Route Analytics Executive Summary Change management is a hot topic, and rightly so. Studies routinely report that a significant percentage of application
More informationCommercial refrigeration has been in the environmental. Refrigerant. as a. Basics Considerations PART 1:
PART 1: CO 2 Commercial refrigeration has been in the environmental spotlight for more than a decade, especially as leakage studies have revealed the true effects of hydrofluorocarbon (HFC) emissions.
More informationSTOICHIOMETRY OF COMBUSTION
STOICHIOMETRY OF COMBUSTION FUNDAMENTALS: moles and kilomoles Atomic unit mass: 1/12 126 C ~ 1.66 10-27 kg Atoms and molecules mass is defined in atomic unit mass: which is defined in relation to the 1/12
More informationChapter 19 Purging Air from Piping and Vessels in Hydrocarbon Service
BP Lower 48 Onshore Operations Safety Manual Page 4.19 1 Chapter 19 Purging Air from Piping and Vessels in Hydrocarbon Service I. General Requirements A. After motor vehicle accidents and underground excavation
More informationInferred ph in Steam Plant Water Chemistry Monitoring
Application Data Sheet ADS 4900-87/rev.B January 2009 Power Industry Inferred ph in Steam Plant Water Chemistry Monitoring INTRODUCTION Inferred ph means ph calculated from straight and cation conductivity.
More informationTemperature Calibration; Depths of Immersion
Temperature Calibration; epths of Immersion Author: John P. Tavener Company Isothermal Technology Limited, Pine Grove, Southport, England Abstract Of all the sources of errors and uncertainties in thermal
More informationReal-time Video Monitoring Increases the Efficiency of SCADA Process Management
Harry Hsiao, Moxa Product Manager (harry.hsiao@moxa.com) Traditional SCADA (Supervisory Control and Data Acquisition) systems have been an important part of automation control systems for many years, and
More informationEXPERIMENT 15: Ideal Gas Law: Molecular Weight of a Vapor
EXPERIMENT 15: Ideal Gas Law: Molecular Weight of a Vapor Purpose: In this experiment you will use the ideal gas law to calculate the molecular weight of a volatile liquid compound by measuring the mass,
More information