Shielding for Diagnostic X-rays: UK Guidance. Jerry Williams

Size: px
Start display at page:

Download "Shielding for Diagnostic X-rays: UK Guidance. Jerry Williams"

Transcription

1 Shielding for Diagnostic X-rays: UK Guidance Jerry Williams Royal Infirmary of Edinburgh

2 BIR/ IPEM Working Party Report British Institute of Radiology Institute of Physics and Engineering in Medicine Working Party: David Sutton, Jerry Williams, Colin Martin, Don McIntosh, Tony Cotterill, Graham Hart, David Gallacher Publication date: 2000

3 BIR/ IPEM Working Party Report Content: Design criteria and dose constraints Primary & Secondary radiations Building materials X-ray transmission ss factors Assessment of shielding Worked examples Working Party: David Sutton, Jerry Williams, Colin Martin, Don McIntosh, Tony Cotterill, Graham Hart, David Gallacher Publication date: 2000

4 Radiation Sources Radiography (film/ screen) Radiography + Fluoroscopy Angiography CT (single slice) Mammography Dental DEXA

5 This talk Design criteria Transmission factors Shielding materials Pi Primary and Secondary Radiations

6 Design Criteria: Dosimetric Quantities Dose limits Effective dose, E msv Dose monitoring Operational quantity Personal dose equivalent, H p (d) msv Shielding calculations l Air kerma, K mgy

7 Design Criteria Dose constraint (people) Based on public dose limit (1 msv/ year) Applied to everyone (staff + public) 0.3 mgy/ year (6 µgy/ week) Occupancy factor 5% Dose to be < 6 mgy/ year (Controlled area) Dose constraint (film or CR) 0.4 µgy/ day (2 µgy/ week)

8 Occupancy factors Based on individual occupancy Examples: Office 100% Reception area 100% X-ray control room 100% Patient examination room 50% Corridor 20% Toilets/ bathrooms 10% Outdoor area with seating 10% Store rooms 5% Unattended waiting rooms 5%

9 X-ray Transmission factors Adopted Archer s empirical model B + β α = α γ t 1 e β α 1 γ Used Simpkin s α, β, γ values Added data for brick and barium plaster Archer et al (1983) Health Physics, 44, 507. Simpkin (1995) Health Physics, 68, 704

10 Shielding material UK Standard lead thicknesses Code t mm

11 Primary and Secondary Radiations Leakage radiation Primary/ transmitted beams Scatter

12 X-ray Tube Leakage Assumptions: Tube leakage factor: 150 kv/ 3.3 ma Field size: 1000 cm 2 Distance: 1 m from tube/ patient WP Recommendation: Leakage contribution may be ignored

13 Primary & Transmitted Radiation X-ray beam Patient Fluoroscopy Beam restricted to image intensifier/ flat panel detector Detector has > 2mm lead equivalent Mammography Imaging device

14 Radiography: Film dose method Cassette Bucky system Dose to film/ imaging plate 400 speed system Dose 10 µgy Lead equivalence (100 kv) Cassette 0.19 mm Cassette + grid 0.26 mm Cassette/grid/table 0.8 mm Cassette/grid/chest Bucky mm Lead equivalence data derived from Dixon (1994) Med Phys, 21, 1785

15 Radiography: Film dose method Dose to film/ imaging plate 400 speed system Dose 10 µgy Lead equivalence (100 kv) Cassette 0.19 mm Cassette + grid 0.26 mm Cassette/grid/table 0.8 mm Cassette/grid/chest Bucky mm

16 Attenuated primary: limiting HVL 1 Limiting HVL: = ln(2) / α 0.1 α t D = D o e Transmission VL mm Limiting H Lead thickness mm kv

17 Radiography: ESD method Unattenuated primary beam Beam outside patient Entrance Surface Dose (ESD) Situations include Beam not collimated to cassette Beam not directed to Bucky Examples Chest radiography Cross table radiography Out of Bucky radiography (e.g. extremities) Area of wall irradiated Not easy to predict Variable

18 Examples

19 Chest Radiography (film dose) Parameters 100 films/ week 90 kv ESD = 0.15 mgy Film dose = 10 µgy FFD = 1.8 m FSD = 1.4 m Film-wall ll = 0.7 m Best Radiographic Practice

20 Calculation (Film-dose method) Attenuated kerma at wall per week: Film dose x Workload x ISL correction = 0.52 mgy No of HVLs required (6 µgy/ week) = 6.5 Limiting HVL of 90 kv = 0.23 mm Total equivalent lead thickness = 1.5 mm Additional lead shielding = mm

21 Chest Radiography (ESD) Parameters 100 films/ week 90 kv ESD = 0.15 mgy Film dose = 10 µgy FFD = 1.8 m FSD = 1.4 m Film-wall ll = 0.7 m

22 Calculation (ESD method) Worst case assumption: Beam size greater than patient and Bucky system Unattenuated kerma at wall per week: ESD x Workload x ISL correction = 4.7 mgy Maximum transmission permitted = 6x10-3 / 4.7 = Lead 90 kv 1.4 mm

23 Summary Best practice Beam collimated to patient/ image plate/ Bucky 0.8 mm lead Poor practice Beam outside patient and Bucky 1.4 mm lead

24 Scatter

25 NCRP 49 method Scatter Model K s = a K u F 400 K s scatter kerma; K u primary kerma F Field size (cm 2 ); a scatter factor Trout and Kelly (1972), Radiology, 104, 161 K u. F Area-air kerma product (AKP) or Dose area product (DAP) K s = S DAP

26 Why DAP? No assumptions on field size Availability of DAP data Requirement in UK legislation to record patient dose DAP preferred patient dose metric (fluoroscopy and radiography) National surveys of patient dose

27 Scatter factor normalised to DAP (S) Scatter factor, S µgy.(gy.cm 2 ) kvp 100 kvp 85 kvp 70 kvp 50 kvp Angle of scatter S = ( a θ + b θ + c θ + d θ + e ) ( kv 85 ) [ f + 1 ] JR Williams (1996) Br J Radiol, 69, 1032

28 Scatter factor normalised to DAP (S) -1 Scatter factor, S µg Gy.(Gy.cm 2 ) kvp 100 kvp 85 kvp 70 kvp 50 kvp Angle of scatter a -1.04E-07 b 327E E-05 c -2.75E-03 d 8.37E-02 e 158E E+00 f 5.99E-03 S = ( a θ + b θ + c θ + d θ + e ) ( kv 85 ) [ f + 1 ] JR Williams (1996) Br J Radiol, 69, 1032

29 Calculation vs measurement Clinical i l application No of DAP/ week Ave Calc positions Gy cm 2 Dose mgy Measured/ Calc dose Mean Min Max Interventional Radiology (liver disease) Abdominal + lower limb angiograpy Cardiac angiography Cerebral angiography Ba contrast studies (o/c tube) Ba contrast t studies (u/c tube) General Radiography

30 Validation of scatter data by Monte-Carlo S µgy/ Gy cm Scatter fraction 85 kv Measured MC Angle

31 Comparison with NCRP 147 S µgy (G Gy cm 2 ) UK Scatter fraction 85 kv NCRP 147 MonteCarlo Angle

32 Differences in scatter factors? Phantom X-ray spectrum Measurement methods

33 Method in practice Scatter factor, S µgy.(gy.cm 2 ) kvp 100 kvp 85 kvp 70 kvp 50 kvp Angle of scatter S ( aθ + bθ + cθ + d + e) ( kv 85) = θ [ f + 1]

34

35

36 Method in practice (1) 1.0 θ r d dose Relative Scattering angle S max = [( ) ] ( 2 ) kv µgy Gy cm 1 d = 1 m; θ = 117

37

38 θ d = 1 m S ave = S( θ) dθ d θ

39 Method in practice (2) S ave over C-arm rotation (30 to 150 ) S max for parallel beam kv S max S ave

40 To be continued.. Application examples Fluoroscopy Radiography Comparisons with NCRP 147

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

Staff Doses & Practical Radiation Protection in DEXA

Staff Doses & Practical Radiation Protection in DEXA Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /

More information

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma): Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 47 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL Refresher Course

More information

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

Recommendations for a technical quality control program for diagnostic X-ray equipment

Recommendations for a technical quality control program for diagnostic X-ray equipment Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer

More information

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY B. Wall, (Co-ordinator) 1, D. Hart 1, H. Mol 2, A. Lecluyse 2, A. Aroua 3, P. Trueb 3, J. Griebel 4, E. Nekolla 4, P. Gron 5, H. Waltenburg 5, H. Beauvais-March

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

The Design of Diagnostic Medical Facilities where Ionising Radiation is used

The Design of Diagnostic Medical Facilities where Ionising Radiation is used RPII 09/01 The Design of Diagnostic Medical Facilities where Ionising Radiation is used A Code of Practice issued by the Radiological Protection Institute of Ireland June 2009 The Design of Diagnostic

More information

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act

More information

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department

More information

Diagnostic Exposure Tracking in the Medical Record

Diagnostic Exposure Tracking in the Medical Record Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand

More information

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information

More information

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation Safety Characteristics of the NOMAD Portable X-ray System Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality

More information

Required RS Training Info

Required RS Training Info C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of

More information

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida Patient Dose Tracking for Imaging Studies David E. Hintenlang, Ph.D., DABR University of Florida Conflict of Interest Statement No affiliation or financial interests in any of the commercial products or

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)

More information

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581 X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers

More information

SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS

SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS AERB SAFETY CODE NO. AERB/SC/MED-2 (Rev. 1) SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS Approved by the Board on October 5, 2001 This document is subject to review, after a period

More information

National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK

National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK The British Journal of Radiology, 82 (2009), 1 12 COMMENTARY National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK D HART, PhD, M C HILLIER and B F WALL,

More information

Dose Tracking System:

Dose Tracking System: Dose Tracking System: A Paradigm Shift in Patient Dose Monitoring Andrew Kuhls-Gilcrist, Ph.D. XRVL Clinical Science Manager Toshiba America Medical Systems, Inc. 2 Dose Tracking System: A Paradigm Shift

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT

More information

5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21 $1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/ (;32685(6 (XURSHDQ&RPPLVVLRQ

5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21 $1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/ (;32685(6 (XURSHDQ&RPPLVVLRQ 5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21 $1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/ (;32685(6 (XURSHDQ&RPPLVVLRQ European Commission 5DGLDWLRQ3URWHFWLRQ *8,'(/,1(621('8&$7,21$1'75$,1,1*,15$',$7,21 3527(&7,21)250(',&$/(;32685(6

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Alzen,, G., Beiderwellen,, K., Volz,, M., Berthold,, L.D.,

More information

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES Ausra Urboniene, Birute Griciene Lithuania Introduction Medical staff during interventional radiology procedures

More information

Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland

Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland HPA-CRCE-43 Trends in Dental Radiography Equipment and Patient Dose in the UK and Republic of Ireland A review of data collected by the HPA Dental X-Ray Protection Services between 28 and 211 for intra-oral

More information

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

Page: 1 of 6 Page: 1 of 6

Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique

More information

Radiation Dose Measurements in Routine X ray Examinations

Radiation Dose Measurements in Routine X ray Examinations Tenth Radiation Physics & Protection Conference, 27-3 November 21, Nasr City - Cairo, Egypt Radiation Dose Measurements in Routine X ray Examinations H. Osman *1, A. Sulieman 1, I.I. Suliman 2 and A.K.

More information

Checklist for submission of application form for Registration

Checklist for submission of application form for Registration Checklist for submission of application form for Registration *Incomplete submission may cause delay in processing of the application. Name of the Hospital/Institution: City State Sr.No. Checkpoints Status

More information

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment

More information

Production of X-rays and Interactions of X-rays with Matter

Production of X-rays and Interactions of X-rays with Matter Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage

More information

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination

More information

X-ray (Radiography) - Abdomen

X-ray (Radiography) - Abdomen Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,

More information

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT

More information

Airport Scanner Safety

Airport Scanner Safety Scan for mobile link. Airport Scanner Safety Airport scanners When it comes to airport scanner safety, there are four questions you should consider, including: What are airport scanners? Are there different

More information

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization

More information

Low-dose CT for Pulmonary Embolism

Low-dose CT for Pulmonary Embolism Low-dose CT for Pulmonary Embolism Gautham Gautham P. P. Reddy, Reddy, MD, MD, MPH MPH University University of of Washington Washington Introduction Introduction CT CT accounts accounts for for > 50%

More information

DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE

DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE A web module produced by Committee 3 of the International Commission on Radiological Protection (ICRP Key Points Diagnostic

More information

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Guidelines on education and training for digital radiology Author:

More information

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology S Peer 1, R Peer 1, P Torbica 1, W Jaschke 1, E Vaño 2 Innsbruck University

More information

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Annual MQSA Inspection Questions NMQAAC April 19, 2004 1 1 1 1 1 1 1 1 0 1 0 1 0 1 0 1 Annual MQSA Inspection Questions NMQAAC April 1, 00 1.0 Inspection Information 1.1 Name and Address 1. Equipment Registration.0 Facility Inspection Download.0 Facility Inspections.0

More information

Radiation Safety Issues for Radiologic Technologists

Radiation Safety Issues for Radiologic Technologists Radiation Safety Issues for Radiologic Technologists Greg Sackett, M.S., CHP Medical Physicist Radiation Worker Risks? 1 Patient Risks? Acute Effects? Delayed Effects? Patient Questions? Radiation Dose

More information

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY rothenbl@mskcc.org

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum

More information

There is no question: X Rays help dentists care for children. Select x rays for individual s needs, not merely as a routine

There is no question: X Rays help dentists care for children. Select x rays for individual s needs, not merely as a routine 1 Medical Physicists Should Join the Movement Image Gently in Dentistry Safe, Effective Dental X Rays for Children There is no question: X Rays help dentists care for children. As imaging professionals

More information

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses.

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. 1 3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. (A) Any person may apply to the director for approval to conduct an educational program for

More information

Computed Tomography Radiation Safety Issues in Ontario

Computed Tomography Radiation Safety Issues in Ontario Computed Tomography Radiation Safety Issues in Ontario Healthcare Human Factors Group Centre for Global ehealth Innovation University Health Network Toronto, ON, Canada June 16, 2006 Table of Contents

More information

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Atomic Energy Regulatory Board Mumbai 400 094 1 DEFINITIONS 1. Quality Assurance (QA): Planned and systematic actions

More information

Digital Breast Tomosynthesis QC Requirements

Digital Breast Tomosynthesis QC Requirements Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Use of lead shielding for adult chest CT. patient & radiographer experiences

Use of lead shielding for adult chest CT. patient & radiographer experiences Use of lead shielding for adult chest CT Dose measurements and patient & radiographer experiences Previous published work Doshi S, Negus I, Oduko J, Foetal radiation dose from CT pulmonary angiography

More information

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects. RADIATION SAFETY: HOW TO EDUCATE AND PROTECT YOURSELF AND YOUR STAFF John Farrelly, DVM, MS, ACVIM (Oncology), ACVR (Radiation Oncology) Cornell University Veterinary Specialists The Veterinary Cancer

More information

Understanding Digital Modalities: System Integration and Use

Understanding Digital Modalities: System Integration and Use Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information

More information

SOURCES AND EFFECTS OF IONIZING RADIATION

SOURCES AND EFFECTS OF IONIZING RADIATION SOURCES AND EFFECTS OF IONIZING RADIATION United Nations Scientific Committee on the Effects of Atomic Radiation UNSCEAR 2008 Report to the General Assembly with Scientific Annexes VOLUME I UNITED NATIONS

More information

X-Rays Benefits and Risks. Techniques that use x-rays

X-Rays Benefits and Risks. Techniques that use x-rays X-Rays Benefits and Risks X-rays are a form of electromagnetic radiation, just like light waves and radiowaves. Because X-rays have higher energy than light waves, they can pass through the body. X-rays

More information

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

GE Healthcare. DoseWatch. Gathering Radiation Dose Data GE Healthcare DoseWatch Gathering Radiation Dose Data Gathering Radiation Dose Data After your organization crosses the bridge of Why? for patient radiation dose management either due to regulation, risk

More information

Determination of the optimal conditions for dental subtraction radiography using a storage phosphor system

Determination of the optimal conditions for dental subtraction radiography using a storage phosphor system Dentomaxillofacial Radiology (1999) 28, 1±5 ã 1999 Stockton Press All rights reserved 0250 ± 832X/99 $12.00 http://www.stockton-press.co.uk/dmfr Determination of the optimal conditions for dental subtraction

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

RADIATION SAFETY STANDARD

RADIATION SAFETY STANDARD Radiation Safety Act 1999 RADIATION SAFETY STANDARD NM004:2010 Standard for radiation apparatus used to carry out industrial radiography Preface Under section 17 of the Radiation Safety Act 1999, a possession

More information

MALAYSIAN STANDARD CODE OF PRACTICE FOR RADIATION PROTECTION - MEDICAL X-RAY DIAGNOSIS (FIRST REVISION) Copyright 2007

MALAYSIAN STANDARD CODE OF PRACTICE FOR RADIATION PROTECTION - MEDICAL X-RAY DIAGNOSIS (FIRST REVISION) Copyright 2007 MALAYSIAN STANDARD MS 838:2007 CODE OF PRACTICE FOR RADIATION PROTECTION - MEDICAL X-RAY DIAGNOSIS (FIRST REVISION) ICS: 13.280 Descriptors: code of practice, medical diagnosis, radiation protection, x-ray

More information

Patient Prep Information

Patient Prep Information Stereotactic Breast Biopsy Patient Prep Information Imaging Services Cannon Memorial Hospital Watauga Medical Center Table Weight Limits for each facility Cannon Memorial Hospital Watauga Medical Center

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

Radiation Protection Series

Radiation Protection Series Radiation The Radiation is published by the Australian Radiation and Nuclear Safety Agency (ARPANSA) to promote practices which protect human health and the environment from the possible harmful effects

More information

Patient Exposure Doses During Diagnostic Radiography

Patient Exposure Doses During Diagnostic Radiography Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality

More information

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

Cone Beam Reconstruction Jiang Hsieh, Ph.D. Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition

More information

Dose Modulation Technique in CT

Dose Modulation Technique in CT Content Dose Modulation Technique in CT Short Overview - Overview - Dose Saving Features - Take Home Point - Conclusions Senior Application specialist Siemens Limited, Taiwan For internal use only / Copyright

More information

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes. IAC Ch 41, p.1 641 41.7 (136C) X-ray machines used for stereotactically guided breast biopsy. 41.7(1) Definitions. In addition to the definitions provided in rules 641 38.2(136C), 641 40.2(136C), and 641

More information

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine

More information

X-ray (Radiography) - Bone

X-ray (Radiography) - Bone Scan for mobile link. X-ray (Radiography) - Bone Bone x-ray uses a very small dose of ionizing radiation to produce pictures of any bone in the body. It is commonly used to diagnose fractured bones or

More information

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes. Radiologic Technologist Overview The Field - Specialty Areas - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Radiologic technologists

More information

Update on ACR Digital Mammography QC Manual

Update on ACR Digital Mammography QC Manual Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual

More information

Diagnostic x-ray imaging relies on the attenuation of

Diagnostic x-ray imaging relies on the attenuation of X-Ray Imaging Physics for Nuclear Medicine Technologists. Part 2: X-Ray Interactions and Image Formation* J. Anthony Seibert, PhD; and John M. Boone, PhD Department of Radiology, University of California

More information

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950 HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational

More information

The Exposure Index and Its Standardization

The Exposure Index and Its Standardization The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany ulrich.neitzel@philips.com Content The Exposure Index (EI) What is it? What is it good

More information