Determination of mercury in aqua regia and nitric acid digests with cold-vapour atomic spectrometry or cold-vapour atomic fluorescence spectrometry
|
|
- Noreen Wells
- 1 years ago
- Views:
Transcription
1 Determination of mercury in aqua regia and nitric acid digests with cold-vapour atomic spectrometry or cold-vapour atomic fluorescence spectrometry 1 Scope This horizontal standard specifies a method for the determination of mercury in nitric acid digest or aqua regia extract, with cold-vapour atomic absorption spectrometry method or cold-vapour atomic fluorescence spectrometry. 2 Principle Mercury is reduced to the elemental state by tin(ii)-chloride solution, and liberated from solution in a closed system. The mercury vapour passes through a cell positioned in the light path of an atomic absorption spectrometer. Its absorbance at a wavelength of 253,7 nm is measured. The absorbance signal is a function of mercury concentration. Alternatively, after the reduction step, the mercury vapour is injected into the cell of an atomic fluorescence spectrometer where the mercury atoms are excited by radiation of a specific wavelength. The intensity of the fluorescence radiation is a function of mercury concentration. NOTE Tin(II)-chloride as a reduction substance is preferred in this standard because sodium borohydride reduces many elements common in soil extract solutions to the elemental state, which cause matrix problems under particular circumstances. 3 Reagents and gases All reagents shall be of recognised analytical grade. Use deionized water or water distilled from an all-glass apparatus, complying with grade 2 as defined in ISO The water used for blank determinations, and for preparing reagents and standard solutions shall have a mercury concentration that is negligible compared with the lowest concentration to be determined in the sample solutions. Purity of argon or nitrogen should be 99,99 %. 3.1 Hydrochloric acid, 37 %; ~ 1,18 g/ml The same batch of hydrochloric acid shall be used throughout the procedre. 3.2 Nitric acid, 65 %; ~ 1,40 g/ml The same batch of nitric acid shall be used throughout the procedre.
2 3.2.1 Nitric acid, diluted solution Pour 500 ml of water into a 1000 ml volumetric flask. Add 250 ml of nitric acid (3.2) with caution, mix and fill to the mark with water. 3.3 Aqua regia, diluted solution Add 21 ml hydrochloric acid (3.1) and 7 ml nitric acid (3.2) to 500 ml of water in a 1000 ml volumetric flask, mix and fill to the mark with water. 3.4 Tin(II)-chloride solution, 10 % (m/v) Dissolve 10 g of SnCl 2.2H 2 O in 30 ml of hydrochloric acid (3.1), transfer to a 100 ml volumetric flask and fill to the mark with water. The blank concentration of mercury can be reduced by bubbling a stream of nitrogen through the solution for 30 minutes, if necessary. Prepare this solution on the day of use. 3.5 Mercury, stock solution corresponding to 1000 mg/l of mercury Two sources of stock solutions can be used: - commercially available stock solutions. - stock solutions prepared in the laboratory from pure elements or stoichiometrically defined and dried salts or oxides. Dissolve 100 mg ± 0,4 mg of mercury metal (minimum purity 99,99 %) in a 100 ml volumetric flask with 17 ml nitric acid (3.2) and fill to the mark with water. NOTE Commercially available stock solutions have the advantage that they remove the need to handle toxic mercury. However, special care needs to be taken that these solutions are supplied with a certified composition from a reputable source and are checked on a regular basis. 3.6 Mercury, standard solution corresponding to 20 mg/l of mercury Pipette 2 ml of the stock mercury solution (3.5) into a 100 ml volumetric flask, add 10 ml nitric acid (3.2.1), mix and fill to the mark with water. 3.7 Mercury, standard solution corresponding to 0,2 mg/l of mercury Pipette 1 ml of the standard mercury solution (3.6) into a 100 ml volumetric flask, add 10 ml nitric acid (3.2.1), mix and fill to the mark with water. 4 Apparatus 4.1 Usual laboratory apparatus
3 All glassware shall be carefully cleaned for low trace element determinations, e.g. by immersion in 5 % (V/V) aqueous nitric acid solution for a minimum of six hours, followed by rinsing with water before use. The nitric acid shall be replaced each week. Grade A volumetricglassware shall be used throughout. 4.2 Atomic absorption spectrometer (AAS) Equipped with an element hollow cathode or, preferably, an electrodeless discharge lamp (which give a greater and more stable light intensity) operated at a current recommended by the lamp and instrument manufacturer, an automatic background correction device and a computerised read out. 4.3 Atomic fluorescence spectrometer (AFS) It is equipped with a specific Hg lamp, a fixed 254 filter, a photomultiplier tube for the detection of fluorescence radiation and a computerised read out. Operate at a current recommended by the lamp and instrument manufacturer. 4.4 Cold-vapour generator, batch system or an automated flow injection analysis system (FIAS) Which should be adaptable to the atomic absorption spectrometer (4.2) or to the atomic fluorescence spectrometer (4.3), according to the detection technique used for the determination of mercury. A flow-controlled nitrogen (3) or argon stream (3) is used as an inert carrier to transport mercury vapour into the cell. Time-controlled addition of tin(ii)-chloride reducing solution (3.4) in combination with automatic start of the read signal of the spectrometer is required. Systems with a 10 cm quartz cell with quartz windows adapted for atomic absorption spectrometry (4.2) are commercially available. In this case, the quartz cell is electrically heated to a temperature of 100 C to avoid condensation of water. For the atomic fluorescence spectrometer (4.3), condensation in the cell is avoided by the inclusion of a drying tube in the detection system. NOTE If an automated system is used, where the reaction occurred continuously (FIA-System), the concentration of stannous chloride solution, reaction time and the gas-liquid separator configuration must be optimised due to the slow reaction kinetics of the reducing system. Warning - It is essential that the manufacturer's safety recommendations are strictly observed. 5 Procedure 5.1 Test solution Transfer 10 ml nitric acid or aqua regia soil extracts according to ISO or ISO 11466, respectively, into a 100 ml volumetric flask and fill to the mark with water.
4 5.2 Blank test solution Carry out a blank test at the same time as the extraction with nitric acid or aqua regia using cleaned quartz sand instead of the soil sample and following the sample procedure, using the same quantities of all reagents for the determination, but omitting the test portion. Transfer 10 ml of the blank test solution to a 100 ml volumetric flask and fill to the mark with water. 5.3 Preparation of the calibration solutions Before each batch of determinations, prepare a blank calibration solution and from the 0,2 mg/l mercury standard solution (3.7) at least five calibration solutions covering the range of concentrations to be determined. Pipette 1 ml, 2 ml, 3 ml, 5 ml and 8 ml of mercury standard solution (3.7) into a series of 100 ml volumetric flasks. Fill to the mark with diluted nitric acid solution (3.2.1) or diluted aqua regia solution (3.3) and mix well. These solutions correspond to mercury concentrations of 2 µg/l, 4 µg/l, 6 µg/l, 10 µg/l and 16 µg/l, respectively. Using 10 ml of each solution for the cold-vapour generator these solutions correspond to 20 ng, 40 ng, 60 ng, 100 ng and 150 ng of mercury, respectively. 5.4 Calibration Cold vapour atomic absorption spectrometer (CV-AAS) Set up the atomic absorption spectrometer (4.2) and the cold-vapour generator (4.4) according to the manufacturer's instructions, select a wavelength of 253,7 nm and adjust the quartz cell, the gas flow and the reaction time and flow rate of tin(ii)-chloride solution (3.4) to get acceptable sensitivity. Transfer 10 ml of blank calibration solution (5.3) into the cold-vapour generation bottle, connect the bottle to the system and start the time-controlled programme. The signal of the blank calibration solution should be negligible compared to with the lowest standard solution. Adjust the response of the instrument to zero signal. Repeat the measurement with each standard calibration solution (5.3) at least twice and record the signal and, if the two values fall within an accepted range, average the values Cold vapour atomic fluorescence spectrometer (CV-AFS) Set up the atomic fluorescence spectrometer (4.3) and the cold-vapour generator (4.4) according to the manufacturer's instructions. Adjust the quartz cell, the gas flow and the flow rate of stannous chloride. Wait until the system is in equilibrium. Aspirate a blank and start the measurement procedure. The signal should be negligible. Set the instrument to zero. Read the signal for each standard solution at least twice. If the two values fall within an accepted range, average the values.
5 5.5 Plotting the calibration graph Plot a graph having the mercury concentrations of the calibration solutions (5.3), from which the blank calibration solution reading has been subtracted, as abscissa, and the corresponding signal values as ordinate. 5.6 Test portion measurement Transfer 10 ml of blank test solution (5.2) into the cold-vapour generator bottle, connect the bottle to the system and start the time-controlled programme and record the signal. Repeat the measurement with each test solution (5.1) at least twice, record the signal and, if the two values fall within an accepted range, average the values. If the mercury concentration of the test solution exceeds the range of calibration, the test solution shall be diluted with diluted nitric acid solution (3.2.1) or diluted aqua regia solution (3.3) accordingly. NOTE The standard addition method ca be used to check the absence of matrix effects. If the analytical results according to the standard addition method and the calibration curve method are equal, the calibration curve method can be applied. 5.7 Calculation By reference to the calibration graph obtained, determine the concentration of the element corresponding to the signal of the test solution (5.1) and of the blank test solution (5.2). Calculate the mercury content (w) of the sample, expressed as mg/kg per dried matter using the equation: where w(hg) = (p 1 - p 0 ).f.v.c / m.1000 w is the mass fraction of mercury in the sample, in mg/kg dry matter; p 1 is the mercury concentration, in µg/l, corresponding to the signal of the test portion (5.1); p 0 is the mercury concentration, in µg/l, corresponding to the signal of the same blank (5.2); f V m C f = 10 is the dilution factor of the test solution (5.1) used in this standard; is the volume, in l, of the test portion taken for the analysis (0,1 l); is the mass of the sample in kg, treated according to ISO or...; is the correction factor for the dry soil sample, C = 100/wdm; wdm is the dry matter content of the soil expressed as a percentage according to ISO
6 5.8 Expression of results Values shall be rounded to 0,01 mg/kg, but only significant figures should be expressed. Example: w(hg) = 0,45 mg/kg w(hg) = 12,5 mg/kg 6 Precision For aqua regia the informations given in ISO can be used, for nitric acid an interlaboratory comparison should be carried through for the current sample types. 7 Test report The test report shall contain the following information: a) a reference to this horizontal standard; b) a reference to the method used (nitric acid or aqua regia); c) complete identification of the sample; d) the results of the determination; e) any details not specified in this standard or which are optional, as well as any factor which may have affected the results.
HS 1003 Part 2 HS 1003 Heavy Metals Test
HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:
Colorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
Determination of the Mass Percentage of Copper in a Penny. Introduction
Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Water Quality 2 - Determination of Ca, Mg, Fe, and Na by Flame Atomic Spectrophotometry
Water Quality 2 - Determination of Ca, Mg, Fe, and Na by Flame Atomic Spectrophotometry Introduction Atomic spectroscopy is one of the most widely used methods for quantitative elemental analysis. There
Water Quality 1 - Spectrophotometric Determination of Iron in Drinking Water
Water Quality 1 - Spectrophotometric Determination of Iron in Drinking Water Introduction The safety of drinking water is a very important public health issue. The United States and World Health Organization
POLYDIMETHYLSILOXANE
POLYDIMETHYLSILOXANE Prepared at the 37th JECFA (1990), published in FNP 52 (1992) superseding specifications prepared at the 29th JECFA (1985), published in FNP 34 (1986). Metals and arsenic specifications
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)
Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:
Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy
Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy Objectives In this lab, you will use fluorescence spectroscopy to determine the mass of riboflavin in a vitamin pill. Riboflavin fluorescence
3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
POLYDIMETHYLSILOXANE
POLYDIMETHYLSILOXANE Prepared at the 69 th JECFA (2008), published in FAO JECFA Monographs 5 (2008), superseding specifications prepared at the 37th JECFA (1990), published in the Combined Compendium of
USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls
Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
Phenolphthalein-NaOH Kinetics
Phenolphthalein-NaOH Kinetics Phenolphthalein is one of the most common acid-base indicators used to determine the end point in acid-base titrations. It is also the active ingredient in some laxatives.
EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline
EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution
DMA-80 Direct Mercury Analyzer
DMA-80 Direct Mercury Analyzer Milestone Srl Milestone Srl 1 Introduction Mercury in the atmosphere can be from manmade sources (coal-fired power plants, municipal incinerators, industrial boilers) or
Determination of Aflatoxins in NeemAzal and NeemAzal-T/S
Determination of Aflatoxins in NeemAzal and NeemAzal-T/S PLC-Method compiled by: Trifolio-M Gmb Sonnenstraße 22 35633 Lahnau, Germany Page 1 of 9 General Information about Aflatoxins Neem seed kernels
In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..
Identification of a Substance by Physical Properties 2009 by David A. Katz. All rights reserved. Permission for academic use provided the original copyright is included Every substance has a unique set
Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
Rev 2016-09-23. Experiment 3
Experiment 3 PREPARATION OF A STANDARD SODIUM HYDROXIDE SOLUTION, DETERMINATION OF PURITY OF IMPURE KHP AND STANDARDIZATION OF HYDROCHLORIC ACID 2 lab periods Reading: Chapter 1 (pg 22-24), Chapter 8,
Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy**
Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy** Objectives In this lab, you will use fluorescence spectroscopy to determine the mass and percentage of riboflavin in a vitamin pill.
FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows
Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for
FMA-80 Fluorescence Mercury Analyzer
FMA-8 Fluorescence Mercury Analyzer www.milestonesrl.com MILESTONE H E L P I N G C H E M I S T S The new Milestone FMA-8 The mercury problem A wide range of mercury analyzers Milestone offers a wide range
Solutions and Dilutions
Learning Objectives Students should be able to: Content Design a procedure for making a particular solution and assess the advantages of different approaches. Choose the appropriate glassware to ensure
Carolina s Solution Preparation Manual
84-1201 Carolina s Solution Preparation Manual Instructions Carolina Biological Supply Company has created this reference manual to enable you to prepare solutions. Although many types of solutions may
Pentosans in wood and pulp. 1. Scope
T 223 cm-01 TENTATIVE STANDARD 1943 OFFICIAL STANDARD 1948 REVISED 1978 CLASSICAL METHOD 1984 REAFFIRMED - 2001 2001 TAPPI The information and data contained in this document were prepared by a technical
NITRIC OXIDE and NITROGEN DIOXIDE 6014
NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm
ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND
#3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric
Expt. 4: ANALYSIS FOR SODIUM CARBONATE
Expt. 4: ANALYSIS FOR SODIUM CARBONATE Introduction In this experiment, a solution of hydrochloric acid is prepared, standardized against pure sodium carbonate, and used to determine the percentage of
Expt B: Determining the Equilibrium Constant of Bromothymol Blue. Keywords: Equilibrium Constant, ph, indicator, spectroscopy
Objectives: Keywords: Equilibrium Constant, ph, indicator, spectroscopy Prepare all solutions needed for measurement of the equilibrium constant for bromothymol Make the required spectroscopic measurements
Na + O O - Cl SODIUM CHLORITE SYNONYMS DEFINITION DESCRIPTION FUNCTIONAL USES CHARACTERISTICS
SODIUM CHLORITE New specifications prepared at the 68 th JECFA (2007) and published in FAO JECFA Monographs 4 (2007). An ADI of 0.03 mg/kg bw for chlorite was established at the 68 th JECFA (2007). SYNONYMS
NATIONAL STANDARD FOR FOOD SAFETY OF THE PEOPLE S REPUBLIC OF CHINA
NATIONAL STANDARD FOR FOOD SAFETY OF THE PEOPLE S REPUBLIC OF CHINA GB 5413.9 2010 National food safety standard Determination of vitamin A, D, E in foods for infants and young children, milk and milk
Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems
Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES FIMS 100 and 400 Flow Injection Mercury Systems Delivering the Performance You Need The determination and monitoring of sub-ppb concentrations
Pretreatment and extraction for the determination of extractable ammonia, nitrate and nitrite
Determination of Nutrients Determination of Ammonia, Nitrate and nitrite Pretreatment and extraction for the determination of extractable ammonia, nitrate and nitrite Introduction This document is developed
Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
EDICT ± OF GOVERNMENT
EDICT ± OF GOVERNMENT Inordertopromotepubliceducationandpublicsafety,equal justiceforal,abeterinformedcitizenry,theruleoflaw,world tradeandworldpeace,thislegaldocumentisherebymade availableonanoncommercialbasis,asitistherightofal
Analysis of Vitamin C Using Iodine. Introduction
Analysis of Vitamin C Using Iodine Introduction Vitamin C (ascorbic acid) is oxidized to dehydroascorbic acid using a mild oxidizing agent such as iodine. The oxidation is a two- electron process, following
Apparatus error for each piece of equipment = 100 x margin of error quantity measured
1) Error Analysis Apparatus Errors (uncertainty) Every time you make a measurement with a piece of apparatus, there is a small margin of error (i.e. uncertainty) in that measurement due to the apparatus
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method
Coordination Compounds with Copper (II) Prelab (Week 2)
Coordination Compounds with Copper (II) Prelab (Week 2) Name Total /10 SHOW ALL WORK NO WORK = NO CREDIT 1. What is the purpose of this experiment? 2. Write the generic chemical formula for the coordination
Determination of calcium by Standardized EDTA Solution
Determination of calcium by Standardized EDTA Solution Introduction The classic method of determining calcium and other suitable cations is titration with a standardized solution of ethylenediaminetetraacetic
A 21 a - Whey Protein Nitrogen Index
A 21 a - Whey Protein Nitrogen Index GEA Niro Method No. A 21 a Revised: November 2009 1. Principle The undenaturated Whey Protein Nitrogen Index (WPN) is a measure of the heat treatment applied to the
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater
Acid Base Titrations
Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually
Energy (J) -8E-19 -1.2E-18 -1.6E-18 -2E-18
Spectrophotometry Reading assignment:. http://en.wikipedia.org/wiki/beer-lambert_law Goals We will study the spectral properties of a transition metal-containing compound. We will also study the relationship
Chem 131A: Absorbance of Riboflavin
Chem 131A: Absorbance of Riboflavin Purpose: The purpose of this experiment is to: 1) Familiarize the student with the use of the HP 8452 diode array spectrophotometer, 2) examine the limitations of the
EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy
EXPERIMENT 4 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium
SODIUM CARBOXYMETHYL CELLULOSE
SODIUM CARBOXYMETHYL CELLULOSE Prepared at the 28th JECFA (1984), published in FNP 31/2 (1984) and in FNP 52 (1992). Metals and arsenic specifications revised at the 55 th JECFA (2000). An ADI not specified
Mercury Determination in House Dust, NIST SRM-2584, USEPA Method 7473, Using the Teledyne Leeman Labs Hydra II C Combustion CVAAS
Application Note - AN1507 Mercury Determination in House Dust, NIST SRM-2584, USEPA Method 7473, Using the Teledyne Leeman Labs Hydra II C Combustion CVAAS Page 1 Jeff Forsberg, Product Manager, John McQuatters,
Recovery of Elemental Copper from Copper (II) Nitrate
Recovery of Elemental Copper from Copper (II) Nitrate Objectives: Challenge: Students should be able to - recognize evidence(s) of a chemical change - convert word equations into formula equations - perform
5. Which temperature is equal to +20 K? 1) 253ºC 2) 293ºC 3) 253 C 4) 293 C
1. The average kinetic energy of water molecules increases when 1) H 2 O(s) changes to H 2 O( ) at 0ºC 3) H 2 O( ) at 10ºC changes to H 2 O( ) at 20ºC 2) H 2 O( ) changes to H 2 O(s) at 0ºC 4) H 2 O( )
Determination of calcium and magnesium found in everyday multivitamins
Determination of calcium and magnesium found in everyday multivitamins Liz Blair and Spencer Knutson Department of Chemistry, Concordia College, 901 8 th St S, Moorhead, MN 56562 Abstract The amounts of
CHEM 43303. The Oxidation of Formic Acid by Bromine
CHEM 43303 Experiment No. 01 The Oxidation of Formic Acid by Bromine 1. Introduction Bromine oxidises formic (methanoic) acid to carbon dioxide according to the following stoichiometric equation HCOOH
I. ACID-BASE NEUTRALIZATION, TITRATION
LABORATORY 3 I. ACID-BASE NEUTRALIZATION, TITRATION Acid-base neutralization is a process in which acid reacts with base to produce water and salt. The driving force of this reaction is formation of a
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
Lab #11: Determination of a Chemical Equilibrium Constant
Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a
INTERNATIONAL OLIVE COUNCIL
INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 3 February 2015 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org
Standard Test Method for Analysis of Barium, Calcium, Magnesium, and Zinc in Unused Lubricating Oils by Atomic Absorption Spectrometry 1
Designation: D 4628 02 An American National Standard Standard Test Method for Analysis of Barium, Calcium, Magnesium, and Zinc in Unused Lubricating Oils by Atomic Absorption Spectrometry 1 This standard
OXIDATION-REDUCTION TITRATIONS-Permanganometry
Experiment No. Date OXIDATION-REDUCTION TITRATIONS-Permanganometry INTRODUCTION Potassium permanganate, KMnO 4, is probably the most widely used of all volumetric oxidizing agents. It is a powerful oxidant
ANALYSIS OF WATER FOR CHLORIDE AND SULFATE IONS
Test Procedure for ANALYSIS OF WATER FOR CHLORIDE AND SULFATE IONS TxDOT Designation: Tex-619-J Effective Date: August 2005 1. SCOPE 1.1 Use this method to analyze water for chloride and sulfate ions to
Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow
Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow Parameter and code: Phosphorus, total-in-bottom-material, dry weight, I-6600-88 (mg/kg as P): 00668 1. Application This method is used
Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40
Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry - Calibration Curve Procedure. The second document
The Determination of the Percent of Salicylic Acid in Aspirin
Salicylic acid The Determination of the Percent of Salicylic Acid in Aspirin Aspirin or acetylsalicylic acid, may be prepared by the reaction of salicylic acid and either acetic acid or acetic anhydride
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student. LBS 171L Section 9 TA: Dana October 27, 2005
How uch Phosphate is the Body Being Exposed to During a Lifetime by Showering? Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student LBS 171L Section 9 TA: Dana October 7, 005 [Note:
DMA-1 Direct Mercury Analyzer
DMA-1 Direct Mercury Analyzer www.milestonesrl.com MILESTONE H E L P I N G C H E M I S T S 8 good reasons to choose the milestone DmA-1 1. no sample preparation 2. ease of use 3. lowest cost of analysis
Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1
Experiment 13H FV 1/25/2011(2-run) THE REACTION OF RED FOOD COLOR WITH BLEACH 1 PROBLEM: Determine the rate law for the chemical reaction between FD&C Red Dye #3 and sodium hypochlorite. LEARNING OBJECTIVES:
Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston
Chemistry 118 Laboratory University of Massachusetts, Boston STOICHIOMETRY - LIMITING REAGENT --------------------------------------------------------------------------------------------------------------------------------------------
Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)
Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based
AOAC Official Method 934.01. Moisture in Animal Feed
Loss on Drying at 95-100 C -Final Action AOAC Official Method 934.01 Moisture in Animal Feed Dry amount of sample containing 2 g dry material to constant weight at 95-100 C under pressure
TECHNICAL CHLORPYRIFOS
TECHNICAL WHO/SIT/21.R3 TECHNICAL CHLORPYRIFOS Full specification WHO/SIT/21.R3 Revised 10 December 1999 1. Specification 1.1 Description The material shall consist of chlorpyrifos together with related
Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.
Halogen Free: What, Why and How Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. We will discuss What are Halogens and how we interact with them. Why Halogens, within consumer
4026 Synthesis of 2-chloro-2-methylpropane (tert-butyl chloride) from tert-butanol
4026 Synthesis of 2-chloro-2-methylpropane (tert-butyl chloride) from tert-butanol OH + HCl Cl + H 2 O C 4 H 10 O C 4 H 9 Cl (74.1) (36.5) (92.6) Classification Reaction types and substance classes nucleophilic
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Grace Analytical Lab 536 South Clark Street 10th Floor Chicago, IL 60605 April 15, 1994 Revision 2 Standard Operating Procedure
Quaternary Ammonium Compounds DOC316.53.01128
Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower
MILESTONE H E L P I N G C H E M I S T S
MILESTONE H E L P I N G C H E M I S T S 8 good reasons to choose the Milestone DMA-80 1. No sample preparation The DMA-80 does not require any sample preparation or other wet chemistry prior the analysis.
XI. Methods of Analysis DETERMINATION OF POTASSIUM CARBONATE CALCULATIONS REAGENTS PROCEDURE
XI. NOTE: An automatic titrator may be utilized for ease of analysis, especially if this test is performed often. Contact Technical Service for further information concerning automatic titration. Methods
LACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e
1 SIDE BY SIDE COMPARISON TABLE FOR THE DETERMINATION OF CYANIDE BY UV DIGESTION AND AMPEROMETRIC DETECTION FIA METHOD - 10-204-00-5-D (ASTM D7511-09 e ) TOPIC Method ASTM D7511-09e LACHAT METHOD NUMBER
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014 This experiment explores quality assurance practices which are commonly
Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid
Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid Introduction Many metals react with acids to form hydrogen gas. In this experiment, you will use the reactions
Flow Injection Analysis
Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of total nitrogen in food and crude protein calculation (Kjeldahl method) Responsible person: Assoc.Prof. Ing.Kateřina Riddellová,
Measuring Manganese Concentration Using Spectrophotometry
Measuring Manganese Concentration Using Spectrophotometry Objectives To use spectroscopy to determine the amount of Manganese is an unknown sample. Scenario Your have just joined a "Green Team" at SMC
HOW TO MAKE STANDARD SOLUTIONS FOR CHEMISTRY
HOW TO MAKE STANDARD SOLUTIONS FOR CHEMISTRY Phillip Bigelow Chemists make two common types of "standard solutions": Molar solutions Normal solutions Both of these solutions are concentrations (or strengths
Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.
Conductivity of strong and weak electrolytes TEC Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Principle It is possible to
Experiment 3 Introduction to Density INTRODUCTION
Experiment 3 Introduction to Density INTRODUCTION The purpose of this experiment is to understand the meaning and significance of the density of a substance. Density is a basic physical property of a homogeneous
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution
Volatile Acids DOC316.53.01144 Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution Scope and application: For digestor sludges. 1 Adapted from The Analyst, 87, 949 (1962).
GLUCOSE OXIDASE
www.megazyme.com GLUCOSE OXIDASE ASSAY PROCEDURE K-GLOX 04/14 (200 Assays per Kit) or (1960 Auto-Analyser Assays per Kit) or (2000 Microplate Assays per Kit) Megazyme International Ireland 2014 INTRODUCTION:
The Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
Mouse C-Peptide I EIA
KAMIYA BIOMEDICAL COMPANY Mouse C-Peptide I EIA For the quantitative determination of C-Peptide I in mouse plasma, serum and urine. Cat. No. KT-367 For Research Use Only. 1 Rev. 091207 PRODUCT INFORMATION
QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.
QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition
EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM ICS English version Digestion of soil, sludge, and biowaste and waste for the extraction of aqua regia soluble elements This draft European Standard
Laboratory Procedure Manual
Laboratory Procedure Manual Analyte: Matrix: Method: Method No.: Lead Dust Wipes Modification of the Automated AAII-25 Colorimetric method Nitric and Hydrochloric Acid Digestion followed by Atomic Absorption
Gas Laws. E k = ½ (mass)(speed) 2. v101613_10am
Gas Laws v101613_10am Objective: In this lab you will become familiar with the Ideal Gas Law and Dalton s Law of Partial Pressures. You will be able to use the information collected along with stoichiometry