Determination of Dissolved Inorganic Nitrate plus Nitrite (NO3+NO2) in Fresh/Estuarine/Coastal Waters Using Cadmium Reduction

Size: px
Start display at page:

Download "Determination of Dissolved Inorganic Nitrate plus Nitrite (NO3+NO2) in Fresh/Estuarine/Coastal Waters Using Cadmium Reduction"

Transcription

1 Determination of Dissolved Inorganic Nitrate plus Nitrite (NO3+NO2) in Fresh/Estuarine/Coastal Waters Using Cadmium Reduction 1. SCOPE and APPLICATION 1.1 Cadmium reduction is used to quantitatively reduce dissolved nitrate to nitrite which is then measured by colorimetric quantitative analysis of a highly colored azo dye. The method is used to analyze all ranges of salinity. 1.2 A Method Detection Limit (MDL) of mg NO 3 +NO 2 -N/L was determined using the Student s t value (3.14, n=7) times the standard deviation of a minimum of 7 replicates. 1.3 The Quantitation Limit for NO 3 +NO 2 was set at mg NO 3 +NO 2 - N/L. 1.4 The method is suitable NO 3 +NO 2 concentrations to.056 mg NO 3 +NO 2 -N/L. 1.5 This procedure should be used by analysts experienced in the theory and application of aqueous inorganic analysis. A three month training period with an analyst experienced in the analysis of nitrate plus nitrite in aqueous samples by cadmium reduction is required. 1.6 This method can be used for all programs that require analysis of dissolved inorganic nitrate plus nitrite. 1.7 This procedure conforms to EPA Method (1979). 2. SUMMARY 2.1 Filtered samples are passed through a granulated copper-cadmium column to reduce nitrate to nitrite. The nitrite, both that which was reduced from nitrate and nitrite that was originally present, is then determined by diazotizing with sulfanilamide and coupling with N-1- napthylethylenediamine dihydrochloride to form a colored azo dye. 3. DEFINITIONS 3.1 Acceptance Criteria Specified limits placed on characteristics of an item, process, or service defined in a requirement document. (ASQC) 3.2 Accuracy The degree of agreement between an observed value and an accepted reference value. Accuracy includes a combination of random error (precision) and systematic error (bias) components which are due to sampling and analytical operations; a data quality indicator. (QAMS) 3.3 Aliquot A discrete, measured, representative portion of a sample taken for analysis. (EPA QAD Glossary) 3.4 Analytical Range to mg NO 3 +NO 2 -N/L, using black/black sample pump tube and yellow/yellow ammonium chloride diluent pump tube at a Standard Calibration setting of April 14,

2 3.5 Batch Environmental samples, which are prepared and /or analyzed together with the same process and personnel, using the same lot(s) of reagents. A preparation batch is composed of one to 300 environmental samples of the same matrix, meeting the above mentioned criteria and with a maximum time between the start of processing of the first and last sample in the batch to be 10 hours. An analytical batch is composed of prepared environmental samples (extracts, digestates, or concentrates) and/or those samples not requiring preparation, which are analyzed together as a group using the same calibration curve or factor. An analytical batch can include samples originating from various environmental matrices and can exceed 20 samples. (NELAC/EPA) 3.6 Blank- A sample that has not been exposed to the analyzed sample stream in order to monitor contamination during sampling, transport, storage or analysis. The blank is subjected to the usual analytical and measurement process to establish a zero baseline or background value and is sometimes used to adjust or correct routine analytical results. (ASQC) 3.7 Calibrate- To determine, by measurement or comparison with a standard, the correct value of each scale reading on a meter or other device, or the correct value for each setting of a control knob. The levels of the applied calibration standard should bracket the range of planned or expected sample measurements. (NELAC) 3.8 Calibration The set of operations which establish, under specified conditions, the relationship between values indicated by a measuring device. The levels of the applied calibration standard should bracket the range of planned or expected sample measurements. (NELAC) 3.9 Calibration Blank A volume of reagent water fortified with the same matrix as the calibration standards, without analyte added Calibration Curve The graphical relationship between known values, such as concentrations, or a series of calibration standards and their analytical response. (NELAC) 3.11 Calibration Method A defined technical procedure for performing a calibration. (NELAC) 3.12 Calibration Standard A substance or reference material used to calibrate an instrument. (QAMS) Initial Calibration Standard (STD) A series of standard solutions used to initially establish instrument calibration responses and develop calibration curves for individual target analytes Initial Calibration Verification (ICV) An individual standard, distinct from the Initial Calibration Standards (STD), analyzed initially, prior to any sample analysis, which verifies acceptability of the calibration curve or previously established calibration curve. April 14,

3 Continuing Calibration Verification (CCV) An individual standard which is analyzed after every field sample analysis Certified Reference Material A reference material one or more of whose property values are certified by a technically valid procedure, accompanied by or traceable to a certificate or other documentation which is issued by a certifying body. (ISO 17025) 3.14 Colorimeter Detector found in Bran & Luebbe Single-Channel Industrial Colorimeter. Color is quantitatively detected with 199- B phototubes using 550 nm monochromatic filters and 50 mm long flow cell with 1.5 mm internal diameter. Comparisons are made between signals from the colored solution in the flow cell to the signal of air in the reference cell. Signals from the Colorimeter are transmitted to a Recorder Corrective Action Action taken to eliminate the causes of an existing nonconformity, defect or other undesirable situation in order to prevent recurrence. (ISO 8402) 3.16 Deficiency An unauthorized deviation from acceptable procedures or practices. (ASQC) 3.17 Demonstration of Capability A procedure to establish the ability of the analyst to generate acceptable accuracy. (NELAC) 3.18 Detection Limit The lowest concentration or amount of the target analyte that can be determined to be different from zero by a single measurement at a stated degree of confidence Duplicate Analysis The analyses of measurements of the variable of interest performed identically on two sub samples (aliquots) of the same sample. The results from duplicate analyses are used to evaluate analytical or measurement precision but not the precision of sampling, preservation or storage internal to the laboratory. (EPA-QAD) 3.20 External Standard (ES) A pure analyte (potassium nitrate (KN O 3 )) that is measured in an experiment separate from the experiment used to measure the analyte(s) in the sample. The signal observed for a known quantity of the pure external standard is used to calibrate the instrument response for the corresponding analyte(s). The instrument response is used to calculate the concentrations of the analyte(s) in the unknown sample Field Duplicates (FD1 and FD2) Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures. Analyses of FD1 and FD2 provide a measure of the precision associated with sample collection, preservation and storage, as well as with laboratory procedures Field Reagent Blank (FRB) A aliquot of reagent water or other blank matrix that is placed in a sample container in the laboratory and treated as a sample in all respects, including shipment to the sampling site, exposure to the sampling site conditions, storage, preservation, April 14,

4 and all analytical procedures. The purpose of the FRB is to determine if method analytes or other interferences are present in the field environment Holding time The maximum time that samples may be held prior to analysis and still be considered valid. (40 CFR Part 136) The time elapsed from the time of sampling to the time of extraction or analysis, as appropriate Instrument Detection Limit (IDL) The minimum quantity of analyte of the concentration equivalent which gives an analyte signal equal to 3.14 times 7 replicates that make up the standard deviation of the background signal at the selected wavelength, mass, retention time absorbance line, etc Laboratory Duplicates (LD1 and LD2) Two aliquots of the same sample taken in the laboratory and analyzed separately with identical procedures. Analyses of LD1 and LD2 indicate precision associated with laboratory procedures, but not with sample collection, preservation, or storage procedures Laboratory Reagent Blank (LRB) A blank matrix (i.e., DI water) that is treated exactly as a sample including exposure to all glassware, equipment, solvents, and reagents that are used with other samples. The LRB is used to determine if method analytes or other interferences are present in the laboratory environment, the reagents, or the instrument Laboratory Control Sample (LCS) A sample matrix, free from the analytes of interest, spiked with verified known amounts of analytes from a source independent of the calibration standard or a material containing known and verified amounts of analytes. The LCS is generally used to establish intra-laboratory or analyst-specific precision and bias or to assess the performance of all or a portion of the measurement system. (NELAC) 3.28 Limit of Detection (LOD) The lowest concentration level that can be determined by a single analysis and with a defined level of confidence to be statistically different from a blank. (ACS) 3.29 Limit of Quantitation (LOQ) The minimum levels, concentrations, or quantities of a target variable (target analyte) that can be reported with a specified degree of confidence. The LOQ is set at 3 to 10 times the LOD, depending on the degree of confidence desired Linear Dynamic Range (LDR) The absolute quantity over which the instrument response to an analyte is linear. This specification is also referred to as the Linear Calibration Range (LCR) Manifold The module whose configuration of glass connectors, fittings, mixing coils, tubing and Cadmium-Copper reduction column precisely reduces the nitrate in the sample to nitrite, followed by color production. April 14,

5 3.32 Material Safety Data Sheets (MSDS) Written information provided by vendors concerning a chemical s toxicity, health hazards, physical properties, fire, and reactivity data including storage, spill, and handling precautions May Denotes permitted action, but not required action. (NELAC) 3.34 Method Detection Limit (MDL) The minimum concentration of an analyte that can be identified, measured, and reported with 99% confidence that the analyte concentration is greater than zero. (Standard Methods) 3.35 Must Denotes a requirement that must be met. (Random House College Dictionary) 3.36 Precision The degree to which a set of observations or measurements of the same property, obtained under similar conditions, conform to themselves; a data quality indicator. Precision is usually expressed as standard deviation, variance or range, in either absolute or relative terms. (NELAC) 3.37 Preservation Refrigeration, freezing, and/or reagents added at the time of sample collection (or later) to maintain the chemical and or biological integrity of the sample Proportioning Pump A peristaltic pump that mixes and advances samples and reagents through proscribed precision pump tubes proportionately for the reactions to take place and for the concentration to be measured Quality Control Sample (QCS) A sample of analyte of known and certified concentration. The QCS is obtained from a source external to the laboratory and different from the source of calibration standards. It is used to check laboratory performance with externally prepared test materials Recorder A graphic recorder used to record electronic output from the colorimeter Run Cycle Typically a day of operation the entire analytical sequence from sampling the first standard to the last sample of the day Sampler An automated rotational device that moves sample cups sequentially to aspirate an aliquot into the proscribed analytical stream. As the loaded sample tray rotates, a metal probe dips into the sample cup and aspirates sample for a preset time, rises from the sample cup and aspirates air for approximately one second and goes into a deionized water-filled wash receptacle, where deionized water is aspirated. After another preset interval, the probe rises from the wash receptacle, aspirates air and moves into the next sample cup. The sampler moves at a rate of 40 samples per hour with a sample to wash solution ratio of 9: Sensitivity The capability of a test method or instrument to discriminate between measurement responses representing different levels (concentrations) of a variable of interest. April 14,

6 4 INTERFERENCES 3.44 Shall Denotes a requirement that is mandatory whenever the criterion for conformance with the specification requires that there be no deviation. (ANSI) 3.45 Should Denotes a guideline or recommendation whenever noncompliance with the specification is permissible. (ANSI) 3.46 Standard Reference Material (SRM) Material which has been certified for specific analytes by a variety of analytical techniques and/or by numerous laboratories using similar analytical techniques. These may consist of pure chemicals, buffers, or compositional standards. The materials are used as an indication of the accuracy of a specific analytical technique. 4.1 Suspended matter in the sample will restrict flow through the apparatus. All samples must be filtered See Section Concentrations of sulfide, iron, copper or other metals above several milligrams per liter lower reduction efficiency, yielding inaccurate concentrations for those samples and, also, subsequent analyses. Frequent checks of column efficiency and re-analyses of affected samples are necessary. 5 SAFETY 5.1 Safety precautions must be taken when handling reagents, samples and equipment in the laboratory. Protective clothing including lab coats, safety glasses and enclosed shoes should be worn. In certain situations, it will be necessary to also use gloves and/or a face shield. If solutions come in contact with eyes, flush with water continuously for 15 minutes. If solutions come in contact with skin, wash thoroughly with soap and water. Contact Solomons Rescue Squad (911) if emergency treatment is needed and also inform the CBL Business Manager of the incident. Contact the CBL Business Manager if additional treatment is required. 5.2 The toxicity or carcinogenicity of each reagent used in this procedure may not have been fully established. Each chemical should be regarded as a potential health hazard and exposure should be as low as reasonably achievable. Cautions are included for known hazardous materials and procedures. 5.3 Do not wear jewelry when troubleshooting electrical components. Even low voltage points are dangerous and can injure if allowed to short circuit. 5.4 The following hazard classifications are listed for the chemicals used in this procedure. Detailed information is provided on Material Safety Data Sheets (MSDS). April 14,

7 Chemical Health Flammability Reactivity Contact Storage Sodium Hydroxide White Stripe Copper Sulfate Green Ammonium Chloride Green Sulfanilamide Green N Green napthylethylenediamine dihydrochloride Brij Green Phosphoric Acid White Hydrochloric Acid White Acetone Red Cadmium Red Potassium nitrate Yellow Sodium nitrite Yellow Chloroform Blue On a scale of 0 to 4, the substance is rated on four hazard categories: health, flammability, reactivity, and contact. (0 is non-hazardous and 4 is extremely hazardous) STORAGE Red Flammability Hazard: Store in a flammable liquid storage area. Blue Health Hazard: Store in a secure poison area. Yellow Reactivity Hazard: Keep separate from flammable and combustible materials. White Contact Hazard: Store in a corrosion-proof area. Green Use general chemical storage (On older labels, this category was orange). Striped Incompatible materials of the same color class have striped labels. These products should not be stored adjacent to substances with the same color label. Proper storage must be individually determined. 6 EQUIPMENT AND SUPPLIES 6.1 Technicon Bran & Luebbe AutoAnalyzer II sampler (now owned by Seal Analytical), proportioning pump, manifold and colorimeter capable of analyzing for nitrate plus nitrite are used in this laboratory. A PMC Industries Flat Bed Linear recorder is used to record electronic output from the colorimeter. 6.2 Freezer, capable of maintaining C. 6.3 Lab ware All reusable lab ware (glass, Teflon, plastic, etc) should be sufficiently clean for the task objectives. This laboratory cleans all lab ware related to this method with a 10% HCl (v/v) acid rinse. 7 REAGENTS AND STANDARDS 7.1 Purity of Water Unless otherwise indicated, references to water shall be understood to mean reagent water conforming to Specification D 1193, Type I. Freshly prepared water should be used for making the standards April 14,

8 intended for calibration. The detection limits of this method will be limited by the purity of the water and reagents used to make the standards. 7.2 Purity of Reagents Reagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that all reagents shall conform to specifications of the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available. Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high purity to permit its use without compromising the accuracy of the determination. 7.3 Alkaline Water Sodium hydroxide (NaOH, pellets) 0.20±0.02 g Deionized water up to 1000 ml Add 0.20 g of sodium hydroxide pellets to 1000 ml of deionized water. Write name of preparer, preparation date, reagent manufacturer, manufacturer lot number in the Analytical Reagent log book. The reagent is stable for six months. 7.4 Copper Sulfate Reagent, 2% Copper sulfate (CuSO 4 5H 2 O) 2 g Deionized water up to 100 ml In a 100 ml volumetric flask, dissolve 2 g of copper sulfate in ~80 ml of deionized water. Dilute to 100 ml with deionized water. Write name of preparer, preparation date, reagent manufacturer, manufacturer lot number in the Analytical Reagent log book. The reagent is stable for six months. 7.5 Ammonium Chloride Reagent Ammonium Chloride (NH 4 Cl) 10 g Deionized water up to 1000 ml Copper Sulfate Reagent, 2% 6 drops Sodium Hydroxide 2 pellets In a 1000 ml volumetric flask, dissolve 10 g of concentrated ammonium chloride to ~800 ml of Deionized Water. Dilute to 1000 ml with Deionized Water. Attain a ph balance of 8.5. Add 6 drops of Copper Sulfate Reagent, 2% and 2 pellets NaOH. Write name of preparer, preparation date, reagent manufacturer, manufacturer lot number in the Analytical Reagent log book. The reagent is stable for six months. 7.6 Color Reagent Sulfanilamide (C 6 H 8 N 2 O 2 S) 20 g Phosphoric Acid (H 3 PO 4 ), concentrated (80%) 200 ml N-1-napthylethylenediamine dihydrochloride (C 12 H 14 N 2 2HCl) 1 g Deionized water up to 2000 ml Brij-35, 30% 1 ml In a 2000 ml volumetric flask, add 200 ml concentrated phosphoric acid and 20 g of sulfanilamide to ~1500 ml deionized water. Dissolve completely. Add 1 g of N-1-napthylethylenediamine dihydrochloride and dissolve. Dilute to 2000 ml with deionized water and add 1 ml of 30% Brij-35. Write name of preparer, preparation date, reagent manufacturers, April 14,

9 manufacturers lot numbers in the Analytical Reagent log book. Make fresh every 3 months. Store at 4ºC. 7.7 Nitrate Stock Standard, 5000 µm Potassium nitrate (KNO 3 ), primary standard grade, dried at 45ºC g Deionized water up to 1000 ml In a 1000 ml volumetric flask, dissolve g of potassium nitrate in ~800 ml of deionized water. Dilute to 1000 ml with deionized water (1 ml contains 5 µmoles N). Write name of preparer, preparation date, standard manufacturer, manufacturer lot number in the Analytical Standard log book. Make fresh every 4 months or when < 20% remains in bottle. 7.8 Secondary Nitrate Standard Stock Nitrate Standard 0.80 ml Deionized water up to 100 ml In a volumetric flask, dilute 0.80 ml of Stock Nitrate Standard to 100 ml with deionized water to yield a concentration of 40 µm NO 3 N/L (0.56 mg N/L). Write name of preparer, preparation date, standard manufacturer, manufacturer lot number in the Analytical Standard log book. Make fresh every 4 weeks. 7.9 Working Nitrate Standard Dilute 1, 2.5, 5, 7.5 and 10 ml of Secondary Standard to 100 ml with deionized water to yield concentrations of 0.4 µm N ( mg N/L), 1.0 µm N (0.014 mg N/L), 2.0 µm N (0.028 mg N/L), 3.0 µm N (0.042 mg N/L) and 4.0 µm N (.056 mg N/L). Write name of preparer, preparation date, standard manufacturer, manufacturer lot number in the Analytical Standard log book. Make fresh every 4 weeks Stock Nitrite Standard Sodium nitrite (NaNO 2 ), primary standard grade, dried at 45ºC g Deionized water up to 1000 ml In a 1000 ml volumetric flask, dissolve g of sodium nitrite in ~800 ml of deionized water. Dilute to 1000 ml with deionized water (1 ml contains 5 µmoles N). Add 1 ml of chloroform as a preservative. Write name of preparer, preparation date, standard manufacturer, manufacturer lot number in the Analytical Standard log book. Make fresh every 4 months or when < 20% remains in bottle. 7.11Secondary Nitrite Standard Stock Nitrite Standard 0.80 ml Deionized water up to 100 ml In a volumetric flask, dilute 0.70 ml of Stock Nitrite Standard to 100 ml with deionized water to yield a concentration of 35 µm NO 2 N/L (0.49 mg N/L). Write name of preparer, preparation date, standard April 14,

10 manufacturer, manufacturer lot number in the Analytical Standard log book. Make fresh every 4 weeks. 8 SAMPLE COLLECTION, PRESERVATION, AND STORAGE 8.1 Water collected for NO 3 +NO 2 should be filtered through a Whatman GF/F glass fiber filter (nominal pore size 0.7 m), or equivalent. 8.2 Water collected for NO 3 +NO 2 should be frozen at -20 C. The AutoAnalyzer vial container should be clean and sample rinsed. 8.3 Frozen NO 3 +NO 2 samples may be stored up to 28 days. It has been shown that frozen QCS samples up to a year old still fall well within the control limits. 8.4 NO 3 +NO 2 samples may be refrigerated at 4 C for no longer than one day. 9 QUALITY CONTROL 9.1 The laboratory is required to operate a formal quality control (QC) program. The minimum requirements of this program consist of an initial demonstration of laboratory capability and the continued analysis of laboratory instrument blanks and calibration standard material, analyzed as samples, as a continuing check on performance. The laboratory is required to maintain performance records that define the quality of data generated. 9.2 Initial Demonstration of Capability The initial demonstration of capability (NO 3 +NO 2 ) is used to characterize instrument performance (MDLs) and laboratory performance (analysis of QC samples) prior to the analyses conducted by this procedure Linear Dynamic Range LDR (Linear Calibration Range) should be established for NO 3 +NO 2 using appropriate calibration curve of a blank and five standards Quality Control Sample (QCS/SRM) When using this procedure, a quality control sample is required to be analyzed at the beginning and end of the run, to verify data quality and acceptable instrument performance. If the determined concentrations are not within 10% of the certified values, performance of the determinative step of the method is unacceptable. The source of the problem must be identified and corrected before either proceeding with the initial determination of MDLs or continuing with analyses Method Detection Limits (MDLs) MDLs should be established for NO 3 +NO 2 using a low level ambient water sample, typically three to five times higher than the estimated MDL. To determine the MDL values, analyze seven replicate aliquots of water. Perform all calculations defined in the procedure (Section 10) and report the concentration values in the appropriate units. Calculate the MDL as follows: April 14,

11 MDL = St (n-1,1-α=0.99) Where, S = Standard deviation of the replicate analyses. n=number of replicates t (n-1,1-α=0.99) = Student s t value for the 99% confidence level with n-1 degrees of freedom (t=3.14 for 7 replicates.) MDLs should be determined annually, whenever there is a significant change in instrumental response, change of operator, or a new matrix is encountered. 9.3 Assessing Laboratory Performance Laboratory Reagent Blank (LRB) The laboratory must analyze at least one LRB with each batch of samples. The LRB consists of Nanopure water treated the same as the samples. An amount of analyte above the MDL found in LRB indicates possible reagent or laboratory environment contamination. LRB data are used to assess and correct contamination from the laboratory environment. LRB above the MDL requires that the source of the problem must be identified and corrected before proceeding with analyses Quality Control Sample (QCS)/ Standard Reference Material (SRM) When using this procedure, a quality control sample is required to be analyzed at the beginning of the run and end of the run, to verify data quality and acceptable instrument performance. If the determined concentrations are not within 3 of the certified values, performance of the determinative step of the method is unacceptable. The source of the problem must be identified and corrected before either proceeding with the initial determination of MDLs or continuing with the analyses. The results of these QCS/SRM samples shall be used to determine batch acceptance The QCS are obtained from a source external to the laboratory and different from the source of calibration standards Control Charts The Accuracy Control Chart for QCS/SRM samples is constructed from the average and standard deviation of the 20 most recent QCS/SRM measurements. The accuracy chart includes upper and lower warning levels (WL=±2s) and upper and lower control levels (CL=±3s). These values are derived from stated values of the QCS/SRM. The standard deviation (s) is specified relative to statistical confidence levels of 95% for WLs and 99% for CLs. Set up an accuracy chart by using percent recovery since the concentration of the QCS/SRM varies. Enter QCS/SRM results on the chart each time the sample is analyzed Continuing Calibration Verification (CCV) Following every samples, one CCV of 4.0 µm NO 3 (.056 mg N/L) is analyzed to assess instrument performance. The CCVs are made from the same April 14,

12 material as calibration standards (KNO 3 ), and are to be within TV 3. Failure to meet the criteria requires correcting the problem, including reanalysis of any affected samples. If not enough sample exists, the data must be qualified if reported Reduction Efficiency Verification (REV) The REVs are made from NaNO 2, 35 µm NO 2 (0.49 mg N/L) and are to be within TV 3s of the equivalent CCV, 35 µm NO 3 (0.49 mg N/L). Failure to meet the criteria requires correcting the problem Reagent Blank The Reagent Blank Control Chart for Reagent Blank samples is constructed from the average and standard deviation of the 20 most recent Reagent Blank measurements. The accuracy chart includes upper and lower warning levels (WL=±2s) and upper and lower control levels (CL=±3s). The standard deviation (s) is specified relative to statistical confidence levels of 95% for WLs and 99% for CLs. Enter Reagent Blank results on the chart each time the Reagent Blank is analyzed. 9.4 Assessing Analyte Recovery - % Recovery Analyte recovery is assessed through percent recoveries of laboratory spikes of samples Percent Recovery for each spiked sample should fall within %. Where: %SR = (Actual/Expected) x Assessing Analyte Precision Relative Percent Difference (RPD) Analyte replication is assessed through duplicate analyses of samples Relative Percent Difference RPD = (Laboratory Duplicate Result 1 Laboratory Duplicate Result 2)/[(Laboratory Duplicate Result 1 + Laboratory Duplicate Result 2)/2] X Corrective Actions for Out of Control Data Control limit If one measurement exceeds Accuracy Control Chart CL, repeat the analysis immediately. If the repeat measurement is within the CL, continue analyses; if it exceeds the CL, discontinue analyses and correct the problem Warning limit If two out of three successive points exceed Accuracy Control Chart WL, analyze another sample. If the next point is within WL, continue analyses; if the next point exceeds the WL, evaluate potential bias and correct the problem Trending If seven successive Accuracy Control Chart measurements are on the same side of the central line, discontinue analyses and correct the problem. April 14,

13 9.6.4 When external QCS samples are out of control, correct the problem. Reanalyze the samples analyzed between the last in-control measurement and the out-of-control one When external CCV samples are out of control, correct the problem. Reanalyze the samples analyzed between the last in-control measurement and the out-of-control one. 9.7 General Operation - To assure optimal operation and analytical results, the Reagent Blank and CCV are tracked daily in the raw data file, copied to Reagent Blank and CCV Control Charts. 10 CALIBRATION AND STANDARDIZATION 10.1 Calibration Daily calibration must be performed before sample analysis may begin. Six point calibrations are used with the Technicon Bran & Luebbe AutoAnalyzer II. Type I water is used as the zero point in the calibration Working Nitrate Standards Dilute Dilute 1, 2.5, 5, 7.5 and 10 ml of Secondary Standard to 100 ml with deionized water to yield concentrations of 0.4 µm N ( mg N/L), 1.0 µm N (0.014 mg N/L), 2.0 µm N (0.028 mg N/L), 3.0 µm N (0.042 mg N/L) and 4.0 µm N (0.056 mg N/L) Prepare standard curve by plotting response on recorder of each and every standard processed through the manifold against NO 3 N/L concentration in standards Compute sample NO 3 +NO 2 N/L concentration by comparing sample response on recorder with standard curve. If NO 3 N/L concentration is required, subtract NO 2 N/L concentration from NO 3 +NO 2 N/L concentration. 11 PROCEDURE NEW REDUCTION COLUMN PREPARATION 11.1 Prepare Copper-Cadmium Column Use good quality cadmium filings of mesh size Clean 10 g of cadmium with 20 ml of acetone. Rinse twice with 20 ml of deionized water. Next, clean cadmium with 50 ml of 1 N Hydrochloric Acid for 1 minute. Cadmium turns silver in color. Decant Hydrochloric Acid and wash the cadmium with another 50 ml of 1 N Hydrochloric Acid for l minute Decant 1 N Hydrochloric Acid and wash the cadmium several times with deionized water Decant deionized water and add 20 ml of 2% (w/v) Copper Sulfate (CuSO 4 5H 2 O). Wash the cadmium until no blue color remains in the solution Decant Copper Sulfate solution and add another 20 ml of 2% (w/v) Copper Sulfate (CuSO 4 5H 2 O). Wash the cadmium until no blue color remains in the solution. The cadmium will be dark brown in color Decant Copper Sulfate solution and wash thoroughly (~10 times) with deionized water. April 14,

14 11.7 Set up Manifold, following general procedure of manufacturer in the following prescribed order Insert a glass wool plug at the outlet end of the column. Fill the reductor column tubing (22 cm length of inch ID Tygon tubing) with Ammonium Chloride Reagent and transfer the prepared cadmium granules to the column using a Pasteur pipette or some other method that prevents contact of cadmium granules with air. Do not allow any air bubbles to be trapped in column. Pack entire column uniformly with filings such that, visually, the packed filings have separation gaps ~1mm Ammonium Chloride Reagent initiates analytical sample stream from 1.20 ml/min Yellow/Yellow pump tube Air is injected from 0.32 ml/min Black/Black pump tube Sample is added from 0.32 ml/min Black/Black pump tube Mixing occurs in five turn coil Air bubbles are de-bubbled from analytical sample stream using 0.60 ml/min Red/Red pump tube De-bubbled analytical sample stream passes through 22 cm reductor column Air is injected from 0.32 ml/min Black/Black pump tube Color Reagent is added from 0.32 ml/min Black/Black pump tube Mixing occurs in twenty-two turn coil Analytical sample stream enters 1.5 mm ID, 50 mm long Flow Cell pulled by 0.80 ml/min waste line. Bubbles and remainder of sample stream exit by gravity Color of analytical sample stream is quantitatively read at 550 nm by Colorimeter with 199-B Phototube, electronic output recorded on strip chart of Recorder Attach pump tubes to end rails of Proportioning Pump. Put platen on Proportioning Pump. With deionized water running through the sample line and Ammonium Chloride Reagent running through its designated line, attach the column. Make sure there are no air bubbles in the valve and attach the column to the intake side of the valve first. Open the valve to allow Ammonium Chloride Reagent stream to flow through the column. Allow deionized water to run through the Color Reagent line Turn on Colorimeter and Recorder Check for good flow characteristics (good bubble pattern) after insertion of air bubbles beyond the column. If the column is packed too tightly, an inconsistent flow pattern will result. Allow Ammonium Chloride Reagent to flow through Column, manifold and Colorimeter for one hour At conclusion of that hour, condition the column with approximately 100 mg N/L (KNO 3 ) for 5 minutes, followed by approximately 100 mg N/L (NaNO 2 ) for 5 minutes. Turn Baseline Knob on Colorimeter to obtain 0 deflection on Recorder Attach Color Reagent line to Color Reagent. At Colorimeter Standard Calibration setting of 1.00, note deflection on Recorder. Reject Color Reagent April 14,

15 if deflection is more than 8 out of total 100 chart units. Turn Baseline Knob on Colorimeter to obtain 0 deflection on Recorder At Colorimeter Standard Calibration setting of 1.00, analyze Secondary Nitrate Standard (40 µm NO 3 N/L (0.56 mg N/L)) and Secondary Nitrite Standard (40 µm NO 2 N/L (0.56 mg N/L)). If peak height of Secondary Nitrate Standard is <90% of peak height of Secondary Nitrite Standard, prepare new cadmium reduction column Set Colorimeter Standard Calibration setting at Analyze Working Nitrate Standards. Prepare standard curve by plotting response on recorder of standards processed through the manifold against NO 3 N/L concentration in standards Analyze samples. Compute sample NO 3 N/L concentration by comparing sample response on Recorder with standard curve At the end of the run, at Colorimeter Standard Calibration setting 1.00, analyze Secondary Nitrate Standard (40 µm NO 3 N/L (0.56 mg N/L)) and Secondary Nitrite Standard (40 µm NO 2 N/L (0.56 mg N/L)). If peak height of Secondary Nitrate Standard is <90% of peak height of Secondary Nitrite Standard, reject all sample concentrations and prepare a new cadmium reduction column Allow deionized water to flow through the sample line for 10 minutes. Close the valve to the column, diverting flow. Allow deionized water to flow through sample, Ammonium Chloride and Color Reagent lines for one minute. Turn Proportioning Pump switch to fast pump for its allotted time Turn off Sampler, Colorimeter and Recorder. Release and remove Proportioning Pump platen. Release pump tube holders from end rails. 12 PROCEDURE DAILY OPERATION 12.1 Attach pump tubes to end rails of Proportioning Pump. Put platen on Proportioning Pump. Allow deionized water to run through the sample line, Ammonium Chloride Reagent to run through its line and deionized water to run through the Color Reagent line. Check for good flow characteristics (good bubble pattern). Open the valve to allow Ammonium Chloride Reagent stream to flow through the column Turn on Colorimeter and Recorder. Set Colorimeter Standard Calibration setting to Let liquid pump through the column, Manifold and Colorimeter for one hour At the conclusion of that hour, turn Baseline Knob on Colorimeter to obtain 0 deflection on Recorder Attach Color Reagent line to the Color Reagent. At a Colorimeter Standard Calibration setting of 1.00, note deflection on the Recorder. Reject Color Reagent if deflection is more than 8 out of total 100 chart units. Turn Baseline Knob on the Colorimeter to obtain 0 deflection on Recorder At Colorimeter Standard Calibration setting 1.00, analyze Secondary Nitrate Standard (35 µm NO 3 N/L (0.49 mg N/L)) and Secondary Nitrite Standard (35 µm NO 2 N/L (0.49 mg N/L)). If the peak height of Secondary April 14,

16 Nitrate Standard is <90% of the peak height of Secondary Nitrite Standard, prepare a new cadmium reduction column Set Colorimeter Standard Calibration setting at Analyze Working Nitrate Standards. Prepare standard curve by plotting response on recorder of standards processed through the manifold against NO 3 N/L concentration in standards in Excel Analyze samples. Compute sample NO 3 N/L concentration by comparing sample response on Recorder with standard curve in Excel At the end of the run, at a Colorimeter Standard Calibration setting of 1.00, analyze Secondary Nitrate Standard (35 µm NO 3 N/L (0.49 mg N/L)) and Secondary Nitrite Standard (35 µm NO 2 N/L (0.49 mg N/L)). If the peak height of Secondary Nitrate Standard is <90% of the peak height of Secondary Nitrite Standard, reject all sample concentrations and prepare a new cadmium reduction column Allow deionized water to flow through the sample line for 10 minutes. Close the valve to the column, diverting flow. Allow deionized water to flow through the sample, Ammonium Chloride and Color Reagent lines for one minute. Turn Proportioning Pump switch to fast pump for its allotted time Turn off Sampler, Colorimeter and Recorder. Release and remove Proportioning Pump platen. Release pump tube holders from end rails. 13 DATA ANALYSIS AND CALCULATIONS 13.1 Upon completion of all analysis, results are saved to a Microsoft Excel daily report file. The file is named by the run date. The daily report file for analytical batch of January 1, 2015 would be named AAIINO23. Peak heights for each sample on chart recorder paper are noted and entered into the report file. Compute sample NO 3 N/L concentration by comparing sample response on chart recorder paper with standard curve in Excel. The analyst examines each row of data. Results are eliminated that are outside the limits of the calibration range. 14 METHOD PERFORMANCE 14.1 On 29 separate dates from January through December 2008, 29 replicate analyses of SPEX Corporation QC 6-42 NUT 1 were performed by Cadmium Reduction. This produced a mean value of mg NO 3 -N/L, SD , Relative Percent Difference of 4.1% from the expected value of ± 10%. This is a mean recovery of 96% For some estuarine samples analyzed by Cadmium Reduction in 2008, the mean difference in concentration between 87 duplicates analyzed on 29 separate dates was mg NO 3 -N/L. The standard deviation of the difference between duplicates was NO 3 -N/L. 15 REFERENCES April 14,

17 15.1 Technicon Industrial Method No W/A Tentative Technicon Industrial Systems. Tarrytown, New York, USEPA Method No in Methods for chemical analysis of water and wastes. United States Environmental Protection Agency, Office of Research and Development. Cincinnati, Ohio. Report No. EPA-600/ March pp. April 14,

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters 1. SCOPE and APPLICATION 1.1 Gravimetric analysis is used to determine total suspended

More information

2. SUMMARY 3. DEFINITIONS

2. SUMMARY 3. DEFINITIONS Determination of Total Dissolved Phosphorus (TDP) in Fresh/Estuarine/Coastal Waters Using Alkaline Persulfate Digestion of Phosphorus to Orthophosphate (PO4) 1. SCOPE and APPLICATION 1.1 Potassium persulfate

More information

METHOD 353.2 DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY

METHOD 353.2 DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY METHOD 353.2 DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING

More information

ESS Method 130.1: General Auto Analyzer Procedures

ESS Method 130.1: General Auto Analyzer Procedures ESS Method 130.1: General Auto Analyzer Procedures Environmental Sciences Section Inorganic Chemistry Unit Wisconsin State Lab of Hygiene 465 Henry Mall Madison, WI 53706 Revised October 1992 ESS Method

More information

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Grace Analytical Lab 536 South Clark Street 10th Floor Chicago, IL 60605 April 15, 1994 Revision 2 Standard Operating Procedure

More information

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow Parameter and code: Phosphorus, total-in-bottom-material, dry weight, I-6600-88 (mg/kg as P): 00668 1. Application This method is used

More information

Ecology Quality Assurance Glossary

Ecology Quality Assurance Glossary Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability

More information

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE Department of Bioresources Engineering Rm 111 Worrilow Hall 531 S. College Ave. Newark, DE 19716-2140

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from

More information

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for

More information

LACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e

LACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e 1 SIDE BY SIDE COMPARISON TABLE FOR THE DETERMINATION OF CYANIDE BY UV DIGESTION AND AMPEROMETRIC DETECTION FIA METHOD - 10-204-00-5-D (ASTM D7511-09 e ) TOPIC Method ASTM D7511-09e LACHAT METHOD NUMBER

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

Determination of the Mass Percentage of Copper in a Penny. Introduction

Determination of the Mass Percentage of Copper in a Penny. Introduction Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny

More information

NITRIC OXIDE and NITROGEN DIOXIDE 6014

NITRIC OXIDE and NITROGEN DIOXIDE 6014 NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method

More information

Phosphorus Analyzer Procedure Last updated: 8.7.15. A. Introduction

Phosphorus Analyzer Procedure Last updated: 8.7.15. A. Introduction Phosphorus Analyzer Procedure Last updated: 8.7.15 A. Introduction The phosphorus analyzer can be used to measure soluble reactive phosphorus (SRP) aka orthophosphate (PO 3-4 ) only. Additionally, this

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater

More information

Colorimetric Determination of Iron in Vitamin Tablets

Colorimetric Determination of Iron in Vitamin Tablets Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered

More information

Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material

Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material PM 2.5 Nylon Filter Cleaning Date: August 25,2009 Page I of8 Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material Environmental and Industrial Sciences Division

More information

HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.

HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0. HEXANES Prepared at the 51st JECFA (1998), published in FNP 52 Add 6 (1998) superseding specifications prepared at the 14th JECFA (1970), published in NMRS 48B (1971) and in FNP 52 (1992). ADI "limited

More information

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014 SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014 This experiment explores quality assurance practices which are commonly

More information

4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES

4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES 4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES Sl. No. Contents Preamble 4.1 Aim 4.2 Introduction 4.2.1 Environmental Significance 4.3 Principle 4.4 Materials Required 4.4.1 Apparatus Required 4.4.2 Chemicals

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS Bernie B. Bernard, Heather Bernard, and James M. Brooks TDI-Brooks International/B&B Laboratories Inc. College Station,

More information

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

Flow Injection Analysis

Flow Injection Analysis Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since

More information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,

More information

DYES AND DYEING 2003 by David A. Katz. All rights reserved. Permission for classroom use provided original copyright is included.

DYES AND DYEING 2003 by David A. Katz. All rights reserved. Permission for classroom use provided original copyright is included. DYES AND DYEING 2003 by David A. Katz. All rights reserved. Permission for classroom use provided original copyright is included. Dyeing of textiles has been practiced for thousands of years with the first

More information

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506 Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking

More information

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure

More information

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine

More information

Coating and Extraction of Honeycomb Denuders

Coating and Extraction of Honeycomb Denuders QSM Approval: Coating and Extraction of Honeycomb Denuders 1 INTRODUCTION 1.1 The following procedures are used for the coating of honeycomb denuders with citric acid and sodium carbonate solutions, and

More information

Quaternary Ammonium Compounds DOC316.53.01128

Quaternary Ammonium Compounds DOC316.53.01128 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower

More information

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND #3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric

More information

HS 1003 Part 2 HS 1003 Heavy Metals Test

HS 1003 Part 2 HS 1003 Heavy Metals Test HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:

More information

APPENDIX N. Data Validation Using Data Descriptors

APPENDIX N. Data Validation Using Data Descriptors APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection

More information

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution

More information

Human Free Testosterone(F-TESTO) ELISA Kit

Human Free Testosterone(F-TESTO) ELISA Kit Human Free Testosterone(F-TESTO) ELISA Kit Catalog Number. MBS700040 For the quantitative determination of human free testosterone(f-testo) concentrations in serum, plasma. This package insert must be

More information

Carolina s Solution Preparation Manual

Carolina s Solution Preparation Manual 84-1201 Carolina s Solution Preparation Manual Instructions Carolina Biological Supply Company has created this reference manual to enable you to prepare solutions. Although many types of solutions may

More information

SUCRALOSE. White to off-white, practically odourless crystalline powder

SUCRALOSE. White to off-white, practically odourless crystalline powder SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA

More information

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of total nitrogen in food and crude protein calculation (Kjeldahl method) Responsible person: Assoc.Prof. Ing.Kateřina Riddellová,

More information

INSTRUCTIONS 56-1190-98. Edition AC

INSTRUCTIONS 56-1190-98. Edition AC Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography

More information

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2 Uscn Life Science Inc. Wuhan Website: www.uscnk.com Phone: +86 27 84259552 Fax: +86 27 84259551 E-mail: uscnk@uscnk.com ELISA Kit for Human Prostaglandin E1(PG-E1) Instruction manual Cat. No.: E0904Hu

More information

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,

More information

Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media

Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media Enhanced Organic Precursor Removals Using Aged Media Page 1 Enhanced Organic Precursor Removals Using Aged Media O b j e c t i v e s The main goals of this project was to evaluate the dissolved organic

More information

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev. K-ASSAY Mouse IgM ELISA For the quantitative determination of IgM in mouse biological samples Cat. No. KT-407 For Research Use Only. 1 Rev. 072309 K-ASSAY PRODUCT INFORMATION Mouse IgM ELISA Cat. No. KT-407

More information

Tamsulosin Hydrochloride Capsules

Tamsulosin Hydrochloride Capsules . nal Revision Bulletin Official October 1, 2011 Tamsulosin 1 standard solution, and shake well. Centrifuge at 1500 rpm for 10 min, and use the supernatant, passing it if Tamsulosin Hydrochloride Capsules

More information

COLORIMETER. Description 0358BT. Figure 1. The Colorimeter

COLORIMETER. Description 0358BT. Figure 1. The Colorimeter COLORIMETER Description 0358BT Figure 1. The Colorimeter Introduction The Colorimeter is designed to determine the concentration of a solution by analyzing its color intensity. Monochromatic light from

More information

Where does it come from?

Where does it come from? What is Phosphorus? Phosphorus (P) is a nutrient that is vital to human, animal, and plant growth. It s one of the most common substances found in nature. It s found in our water, our food, and our bodies.

More information

Continuous-Flow Analyzers

Continuous-Flow Analyzers _ Continuous-Flow Analyzers _ Rapid Control of Fertilizer Quality with the AutoAnalyzer Legislation and quality requirements have led to the need for close control of fermentation and finished products

More information

DR/4000 PROCEDURE. CHLORINE, Free

DR/4000 PROCEDURE. CHLORINE, Free DR/4000 PROCEDURE CHLORINE, Free Method 10069 DPD method* UHR (0.1 10.0 mg/l as Cl 2 ) Scope and Application: For testing higher levels of free chlorine (hypochlorous acid and hypochlorite ion) in drinking

More information

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU

More information

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater Salicylate Method * 1. Enter the stored program number for ammonia nitrogen (NH 3 -N). 2. Press: 64 ENTER The display

More information

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination

More information

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8) Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:

More information

To determine the mass of iron in one adult dose of either a ferrous sulfate or. ferrous gluconate iron supplement using a colorimetric technique.

To determine the mass of iron in one adult dose of either a ferrous sulfate or. ferrous gluconate iron supplement using a colorimetric technique. Lab: Colorimetric Analysis of Iron in Iron Supplements Purpose To determine the mass of iron in one adult dose of either a ferrous sulfate or ferrous gluconate iron supplement using a colorimetric technique.

More information

Experiment 8 - Double Displacement Reactions

Experiment 8 - Double Displacement Reactions Experiment 8 - Double Displacement Reactions A double displacement reaction involves two ionic compounds that are dissolved in water. In a double displacement reaction, it appears as though the ions are

More information

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. Halogen Free: What, Why and How Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. We will discuss What are Halogens and how we interact with them. Why Halogens, within consumer

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

Analytical Chemistry Lab Reports

Analytical Chemistry Lab Reports Analytical Chemistry Lab Reports Format and Calculations John Collins collinj2@gmail.com Measurement Analytical chemistry is entirely about measurement, what these measurements signify, and the understanding

More information

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date: PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA

More information

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. IgM ELISA For the quantitative determination of IgM in human serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Calibrator

More information

Total Cyanide by Segmented Flow Injection Analysis, On-Line UV Digestion, and Amperometric Detection

Total Cyanide by Segmented Flow Injection Analysis, On-Line UV Digestion, and Amperometric Detection Methodology Total Cyanide by Segmented Flow Injection Analysis, On-Line UV Digestion, and Amperometric Detection (Cartridge Part #A001471, #A002654, #A002683) 1.0 Scope and Application 1.1 This method

More information

Analysis of Vitamin C Using Iodine. Introduction

Analysis of Vitamin C Using Iodine. Introduction Analysis of Vitamin C Using Iodine Introduction Vitamin C (ascorbic acid) is oxidized to dehydroascorbic acid using a mild oxidizing agent such as iodine. The oxidation is a two- electron process, following

More information

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

Measurement by Ion Selective Electrodes

Measurement by Ion Selective Electrodes Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable

More information

PROTEINS (LOWRY) PROTOCOL

PROTEINS (LOWRY) PROTOCOL 1 PROTEINS (LOWRY) PROTOCOL 1. INTRODUCTION The Lowry Assay: Protein by Folin Reaction (Lowry et al., 1951) has been the most widely used method to estimate the amount of proteins (already in solution

More information

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Catalog No: E0479r 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

ENCM 31073-Environmental Impact Assessment and Environmental Monitoring. Determination of nitrate and phosphate levels in well water/surface water

ENCM 31073-Environmental Impact Assessment and Environmental Monitoring. Determination of nitrate and phosphate levels in well water/surface water ENCM 31073-Environmental Impact Assessment and Environmental Monitoring ntalmonitoringenvironmental Monitoring - Practical Number 4 Determination of nitrate and phosphate levels in well water/surface water

More information

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,

More information

Acid Base Titrations

Acid Base Titrations Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually

More information

1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.

1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil. March 2012 8 1 of 17 1.0 SCOPE 1.1 This procedure describes the preparation and analysis of solid matrices and wipe samples. 1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.

More information

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit Catalog No. CSB-E15852c (96T) This immunoassay kit allows for the in vitro quantitative determination of canine CK-MB concentrations in serum and plasma.

More information

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. Titration Curves PURPOSE To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. GOALS 1 To gain experience performing acid-base titrations with a ph meter. 2

More information

Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid

Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid Introduction Many metals react with acids to form hydrogen gas. In this experiment, you will use the reactions

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

PART 1 - INTRODUCTION...

PART 1 - INTRODUCTION... Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART

More information

Properties of Acids and Bases

Properties of Acids and Bases Lab 22 Properties of Acids and Bases TN Standard 4.2: The student will investigate the characteristics of acids and bases. Have you ever brushed your teeth and then drank a glass of orange juice? What

More information

An Introduction to Standards and Quality Control for the Laboratory Barbara Martin

An Introduction to Standards and Quality Control for the Laboratory Barbara Martin An Introduction to Standards and Quality Control for the Laboratory Barbara Martin Hach Company, 2002. All rights reserved. Lit. No. 2426 eac/rb 8/02 1ed In memory of Clifford C. Hach (1919-1990) inventor,

More information

Quality Assurance/Quality Control in Acid Deposition Monitoring

Quality Assurance/Quality Control in Acid Deposition Monitoring Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial

More information

Medical Cannabis Laboratory Approval Program

Medical Cannabis Laboratory Approval Program Medical Cannabis Laboratory Approval Program Application Process and Required Documentation After the publication of the Medical Cannabis Laboratory Application, currently expected by February 16, 2015,

More information

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry - Calibration Curve Procedure. The second document

More information

Phosphates (ortho- and total)

Phosphates (ortho- and total) INTRODUCTION Phosphates (ortho- and total) Phosphorus is an essential nutrient for all aquatic plants and algae. Only a very small amount is needed, however, so an excess of phosphorus can easily occur.

More information

Laboratory Procedure Manual

Laboratory Procedure Manual Laboratory Procedure Manual Analyte: Matrix: Method: Method No.: Lead Dust Wipes Modification of the Automated AAII-25 Colorimetric method Nitric and Hydrochloric Acid Digestion followed by Atomic Absorption

More information

ph Measurement and Control

ph Measurement and Control ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components

More information

A Beer s Law Experiment

A Beer s Law Experiment A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Introduction: The determination of protein concentration is frequently required in biochemical work. Several methods

More information

Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10

Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10 Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan SECTION 10 DATA QUALITY ASSURANCE AND QUALITY CONTROL (QA/QC) Hawai`i HEER TGM 10-1 December 31, 2008 CONTENTS Acronyms

More information

ERDOSTEINE - MONOGRAPH.

ERDOSTEINE - MONOGRAPH. STRUCTURAL FORMULA (±1S-(2-[N-3-(2-oxotetrahydro thienyl)]acetamido)-thioglycolic acid) C 8 H 11 NO 4 S 2 M.W. = 249.307 DESCRIPTION Color : White to ivory white Appearance : Microcrystalline powder SOLUBILITY

More information

4.2 Bias, Standards and Standardization

4.2 Bias, Standards and Standardization 4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard

More information

Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student. LBS 171L Section 9 TA: Dana October 27, 2005

Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student. LBS 171L Section 9 TA: Dana October 27, 2005 How uch Phosphate is the Body Being Exposed to During a Lifetime by Showering? Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student LBS 171L Section 9 TA: Dana October 7, 005 [Note:

More information

Safety Data Sheet Avesta Neutralization Agent 502

Safety Data Sheet Avesta Neutralization Agent 502 Safety Data Sheet Avesta Neutralization Agent 502 This Safety Data Sheet contains information to help users understand the potential hazards relating to this product and provides advice for risk management.

More information