PSV estimation variations following doppler ultrasound setting changes

Size: px
Start display at page:

Download "PSV estimation variations following doppler ultrasound setting changes"

Transcription

1 PSV estimation variations following doppler ultrasound setting changes Poster No.: C-3213 Congress: ECR 2010 Type: Scientific Exhibit Topic: Vascular Authors: S. Mirsadraee, A. Evans, D. O. Kessel; Leeds/UK Keywords: Doppler Phantom, Pulsatile phantom, Test objects DOI: /ecr2010/C-3213 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 8

2 Purpose Doppler ultrasound is regularly used to investigate vascular pathologies and plan treatment. Surrogate measures of stenosis and other haemodynamically important pathologies include peak systolic velocity and pulsatility parameters. There is scant evidence data on the reproducibility of these measures between operators or between machines. Aim of Study Aims of this study was to: 1.develop a portable phantom that simulates near physiological arterial flow in medium size vessels such as carotid arteries. 2.assess the effect of ultrasound equipment settings on velocity measurement reproducibility. Activities have been carried out in three phases: preliminary investigations and design of an example test object construction and validation of the example test object assessment of flow measurement by equipment setting changes Methods and Materials Fluid circuit & Blood mimicking solution A pulsatile blood pump was connected to a 500mls reservoir and the flow phantom. Natural latex rubber tubes were used to connect different components. The reservoir was used to avoid air bubbles within the circuit (Figure 2). Blood mimicking solution (BMS) Page 2 of 8

3 was made from nylon scattering particles (Orgasol) suspended in a fluid base of water, glycerol, dextran and surfactant (Ramnarine et al. 1998). Tissue mimicking material (TMM) The criteria for the selection of TMM included speed of sound, viscosity, safety, and availability. Materials tested were liquid and solid poly vinyl alcohol (PVA), natural latex rubber, and liquid glycerine. Final TMM material consisted of aluminium particles suspended in solid agar gel. Phantom (Figure 1) The blood vessel model was made from moulded latex tube (1-2mm thickness, 1.5mm diameter, 5 cm in length). The tube was connected to connecting tubes and then suspended in the tissue mimicking material (liquid PVA; agar based TMM) within a sealed container. To prevent dryness of TMM PVA and also protecting US probe, the TMM (2.5cm in depth above the tube) was topped with thin layers of latex (2-3mm in thickness) and air dried. Silicon sealant was used to seal potential areas of leak. Validation Ultrasound and computerised tomography were used to evaluate the physical properties of the phantom. Speed of sound, reflectivity, and ability to maintain vessel shape were evaluated in 3cm deep block of TMM and also moulded vessel tube. The BMS and flow were assessed with Duplex ultrasound. An experienced vascular sonographer made multiple peak velocity measurements on four different modern ultrasound machines. Assessment of velocity variations Phantom was perfused at constant flow volume (302+6ml/min) with BMS. A single experienced Doppler ultrasound operator performed multiple peak systolic (PSV) and end diastolic velocity (EDV) measurements (10 measurements per experiment). Measurements were performed with the operator blind to the results. The following parameters: PRF ( ), Doppler gain (35, 45, 55, and 65dB), and scanning angle (55 vs 65 degrees) were assessed. Statistical analysis of variance (ANOVA) was used to compare the velocity values. Images for this section: Page 3 of 8

4 Page 4 of 8

5 Fig. 1: PVA based phantom. Second image shows sagital CT reformat. 1=surface latex cover; 2=liquid PVA; 3=vessel wall Fig. 2: A: Images of the fluid circuit. P=pump; C=connector; Ph= phantom; R=reservoir. B: Colour Doppler signal and pulsatile flow measurements using pulsed Doppler. Page 5 of 8

6 Results Speed of sound in liquid glycerine, solid latex and PVA was significantly higher than in water (2-3 times higher in average). In contrary, the speed of sound in liquid PVA and agar based TMM was similar to that in water. The liquid PVA was relatively echo free unless contaminated with air bubbles. Pulsatile (multiphasic) Doppler flow signal was detected and the velocity was measured (Figure 1,2). Reproducible measurements of velocity were achieved. There was no significant statistical difference in the peak velocity measurements between ultrasound machines (n=4; ANOVA, p-value>0.3). Increasing Doppler gain resulted in significantly higher PSV and EDV (mean PSV values at 65dB 35% higher than at 35dB; 39 vs 56 cm/s; p <0.001). Changing PRF value resulted in significantly different velocity values but there was no linear correlation between these variables. Higher scanning angles (60 vs 55) resulted in significantly higher velocity estimation (38 vs 50 cm/s at 35dB; p<0.004)(figure 2). Images for this section: Fig. 1: A: Images of the fluid circuit. P=pump; C=connector; Ph= phantom; R=reservoir. B: Colour Doppler signal and pulsatile flow measurements using pulsed Doppler. Page 6 of 8

7 Fig. 2: Doppler signal from the flowing BMS. A: Measurement of PSV using different gains and PRF. Increasing gain resulted in significantly higher PSV measurement (ANOVA; P0.3). Page 7 of 8

8 Conclusion A pulsatile flow Doppler phantom has been developed and reproducible signals were detected. Different equipment demonstrated reproducible velocity measurements. Scanning technique can significantly alter the results of the ultrasound Doppler velocity measurements. Laboratories using velocity measurements to make management decisions should establish a measurement error range. Future directions: Carotid test objects that reproduce vascular stenoses are made to study flow dynamics. The test object is X-ray and MR compatible and will be used for multimodality assessment of luminal stenosis and flow restrictions. References Lubbers, K Ramnarine and PR Hoskins, Blood mimicking fluid for flow Doppler test object. Eur J Ultrasound 7 (1998) Volume 7, Supplement 1, February 1998, pp (1) Personal Information Mirsadraee S, Kessel D, Evans A Leeds Teaching Hospitals and University of Leeds, Leeds, United Kingdom This study has been supported by the 2007 BSIR Research and Educational Bursary. Page 8 of 8

CT findings in Differential Diagnosis between Tuberculous Pleurisy and Malignant Effusion

CT findings in Differential Diagnosis between Tuberculous Pleurisy and Malignant Effusion CT findings in Differential Diagnosis between Tuberculous Pleurisy and Malignant Effusion Poster No.: E-0084 Congress: ESTI 2012 Type: Scientific Exhibit Authors: S. S. Shim, Y. Kim; Seoul/KR Keywords:

More information

Utilization management for successful process optimization in radiology

Utilization management for successful process optimization in radiology Utilization management for successful process optimization in radiology Poster No.: C-0737 Congress: ECR 2011 Type: Scientific Paper Authors: H.-P. Busch 1, W. Frewer 1, A. van Est 2 ; 1 Trier/DE, 2 Eindhoven/NL

More information

Patterns of nodal spread in thoracic malignancies

Patterns of nodal spread in thoracic malignancies Patterns of nodal spread in thoracic malignancies Poster No.: C-0977 Congress: ECR 2010 Type: Educational Exhibit Topic: Chest Authors: R. dos Santos, M. Duarte, J. Alpendre, J. Castaño, Z. Seabra, Â.

More information

Normal vascular variants of the upper extremity

Normal vascular variants of the upper extremity Normal vascular variants of the upper extremity Poster No.: C-1039 Congress: ECR 2014 Type: Educational Exhibit Authors: P. Paixao, A. P. Gomes, M. S. C. Sousa, I. Santiago, A. S. C. C. Germano; Amadora/PT

More information

OA related pain medication intake in subjects from the OAI incidence cohort - association with focal knee lesions and cartilage T2 measurements

OA related pain medication intake in subjects from the OAI incidence cohort - association with focal knee lesions and cartilage T2 measurements OA related pain medication intake in subjects from the OAI incidence cohort - association with focal knee lesions and cartilage T2 measurements Poster No.: C-0007 Congress: ECR 2012 Type: Scientific Exhibit

More information

An Image Based Semiautomatic Online Cancer Staging Program for Common Female Pelvic Malignancies

An Image Based Semiautomatic Online Cancer Staging Program for Common Female Pelvic Malignancies An Image Based Semiautomatic Online Cancer Staging Program for Common Female Pelvic Malignancies Poster No.: C-1050 Congress: ECR 2012 Type: Educational Exhibit Authors: R. Talanow; Lexington, KY/US Keywords:

More information

MRI scanning of the claustrophobic patients

MRI scanning of the claustrophobic patients MRI scanning of the claustrophobic patients Poster No.: C-0549 Congress: ECR 2014 Type: Educational Exhibit Authors: M. minov, A. Doreski, M. Popovska, G. Markoski, S. 1 2 3 2 2 1 2 2 3 Jovanoska, I. miladinovski

More information

Needle crystal detector technology in mammography further dose reduction and clinical image quality with

Needle crystal detector technology in mammography further dose reduction and clinical image quality with Needle crystal detector technology in mammography further dose reduction and clinical image quality with different beam qualities (W/Rh vs. Mo/Rh) Poster No.: C-1461 Congress: ECR 2013 Type: Scientific

More information

Eye lens dose measurements in Interventional Cardiology

Eye lens dose measurements in Interventional Cardiology Eye lens dose measurements in Interventional Cardiology Poster No.: 479 Congress: ESCR 2014 Type: Scientific Poster Authors: N. Kollaros 1, E. Carinou 2, C. Plemmenos 1, I. Stathopoulos 1, V. Keywords:

More information

Pancreatic masses: What is there besides cancer

Pancreatic masses: What is there besides cancer Pancreatic masses: What is there besides cancer Poster No.: C-0201 Congress: ECR 2010 Type: Educational Exhibit Topic: Abdominal Viscera (Solid Organs) Authors: M. A. Portilha, C. Ruivo, I. Santiago, M.

More information

Preoperative evaluation of future remnant liver function by the contrast enhance ratio in hepatocellular image.

Preoperative evaluation of future remnant liver function by the contrast enhance ratio in hepatocellular image. Preoperative evaluation of future remnant liver function by the contrast enhance ratio in hepatocellular image. Poster No.: C-0564 Congress: ECR 2011 Type: Scientific Exhibit Authors: S. Matsushima, Y.

More information

Determination of bone age using MRI of hand/wrist: a pilot study

Determination of bone age using MRI of hand/wrist: a pilot study Determination of bone age using MRI of hand/wrist: a pilot study Poster No.: C-0963 Congress: ECR 2011 Type: Scientific Exhibit Authors: E. Tomei, M. Marini, A. Stagnitti, A. sartori, L. bertana, N. Al

More information

Comparison of radiation dose from X-ray, CT, and PET/ CT in paediatric patients with neuroblastoma using a dose monitoring program

Comparison of radiation dose from X-ray, CT, and PET/ CT in paediatric patients with neuroblastoma using a dose monitoring program Comparison of radiation dose from X-ray, CT, and PET/ CT in paediatric patients with neuroblastoma using a dose monitoring program Poster No.: C-0591 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific

More information

Diffuse infiltration in multiple myeloma treatment response assessment with "total-spine" contrast enhanced MR imaging

Diffuse infiltration in multiple myeloma treatment response assessment with total-spine contrast enhanced MR imaging Diffuse infiltration in multiple myeloma treatment response assessment with "total-spine" contrast enhanced MR imaging Poster No.: C-0276 Congress: ECR 2011 Type: Scientific Paper Authors: P. P. Arcuri,

More information

Acute abdominal pain in the elderly patient: Impact of early MDCT examination on diagnosis and management

Acute abdominal pain in the elderly patient: Impact of early MDCT examination on diagnosis and management Acute abdominal pain in the elderly patient: Impact of early MDCT examination on diagnosis and management Poster No.: C-1464 Congress: ECR 2010 Type: Topic: Scientific Exhibit GI Tract Authors: A. Pinto,

More information

Development and implementation of a help desk system in a radiology department in a high complexity hospital: a South American experience

Development and implementation of a help desk system in a radiology department in a high complexity hospital: a South American experience Development and implementation of a help desk system in a radiology department in a high complexity hospital: a South American experience Poster No.: C-2409 Congress: ECR 2013 Type: Authors: Keywords:

More information

Role of 3D volumetry CT in the correlation between postoperative gastric volume and weight loss in obese patients undergoing gastric sleeve surgery.

Role of 3D volumetry CT in the correlation between postoperative gastric volume and weight loss in obese patients undergoing gastric sleeve surgery. Role of 3D volumetry CT in the correlation between postoperative gastric volume and weight loss in obese patients undergoing gastric sleeve surgery. Poster No.: C-2145 Congress: ECR 2014 Type: Scientific

More information

DICOM metadata-mining in PACS for computed radiography X-Ray exposure analysis: a mammography multisite study

DICOM metadata-mining in PACS for computed radiography X-Ray exposure analysis: a mammography multisite study DICOM metadata-mining in PACS for computed radiography X-Ray exposure analysis: a mammography multisite study Poster No.: B-0276 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Paper M. R.

More information

Angio-CT preoperative planning of inferior limb free flaps in plastic surgery

Angio-CT preoperative planning of inferior limb free flaps in plastic surgery Angio-CT preoperative planning of inferior limb free flaps in plastic surgery Poster No.: C-3046 Congress: ECR 2010 Type: Educational Exhibit Topic: Vascular Authors: L. Saba, M. Atzeni, D. Ribuffo, G.

More information

Comparison between photoacoustic mammography images of breasts cancers and histological staining patterns of CD31 and carbonic anhydrase IX

Comparison between photoacoustic mammography images of breasts cancers and histological staining patterns of CD31 and carbonic anhydrase IX Comparison between photoacoustic mammography images of breasts cancers and histological staining patterns of CD31 and carbonic anhydrase IX Poster No.: C-1893 Congress: ECR 2013 Type: Scientific Exhibit

More information

Metastatic malignant melanoma revisited

Metastatic malignant melanoma revisited Metastatic malignant melanoma revisited Poster No.: C-2446 Congress: ECR 2010 Type: Educational Exhibit Topic: Musculoskeletal Authors: C. Paulino, B. Gonçalves, P. Marques, M. Gonçalo, F. CaseiroAlves;

More information

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

Measure #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care

Measure #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care Measure #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care 2016 PQRS OPTIONS FOR INDIVIDUAL MEASURES: CLAIMS, REGISTRY

More information

Abdominal CT Perfusion: Effects of Breath Control Technique

Abdominal CT Perfusion: Effects of Breath Control Technique Abdominal CT Perfusion: Effects of Breath Control Technique Poster No.: C-0068 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Yoshikawa, T. Kanda, Y. Ohno, Y. Fujisawa, N. Negi, M. 1 1 2 1 1 1

More information

Ultrasound Phantoms: B-Mode, Doppler and Others. Heather M. Pierce, B.S. Norfolk, VA

Ultrasound Phantoms: B-Mode, Doppler and Others. Heather M. Pierce, B.S. Norfolk, VA Ultrasound Phantoms: B-Mode, Doppler and Others Heather M. Pierce, B.S. Norfolk, VA Overview Design of ultrasound phantoms Types of phantoms available Use of phantoms in a QA/QC program Advantages and

More information

VFI Technology to Change the Way You Work

VFI Technology to Change the Way You Work Technology to Change the Way You Work Vascular Ultrasound Made Easier Vector Flow Imaging VFI VFI is a ground-breaking technology that can revolutionize the workflow for many Doppler ultrasound applications.

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

Vascular Laboratory Fundamentals. Robert Mitchell MD, RPVI Duke University Medical Center 5/2/08

Vascular Laboratory Fundamentals. Robert Mitchell MD, RPVI Duke University Medical Center 5/2/08 Vascular Laboratory Fundamentals Robert Mitchell MD, RPVI Duke University Medical Center 5/2/08 Goals of Lecture Understand basic ultrasound principles Understand normal and abnormal arterial hemodynamics

More information

MR imaging of primary sclerosing cholangitis (PSC) using the hepatobiliary specific contrast agent Gd-EOB-DTPA

MR imaging of primary sclerosing cholangitis (PSC) using the hepatobiliary specific contrast agent Gd-EOB-DTPA MR imaging of primary sclerosing cholangitis (PSC) using the hepatobiliary specific contrast agent Gd-EOB-DTPA Poster No.: C-0019 Congress: ECR 2010 Type: Educational Exhibit Topic: Abdominal Viscera (Solid

More information

Horton General Hospital Deep Vein Thrombosis (DVT) and Pulmonary Embolism (PE) Information for patients

Horton General Hospital Deep Vein Thrombosis (DVT) and Pulmonary Embolism (PE) Information for patients Horton General Hospital Deep Vein Thrombosis (DVT) and Pulmonary Embolism (PE) Information for patients What is a Deep Vein Thrombosis (DVT)? A DVT is a blood clot which forms in a deep vein, usually in

More information

Differential diagnosis of vertebral compression fracture using in-phase/opposed-phase and Short TI inversion recovery imaging

Differential diagnosis of vertebral compression fracture using in-phase/opposed-phase and Short TI inversion recovery imaging Differential diagnosis of vertebral compression fracture using in-phase/opposed-phase and Short TI inversion recovery imaging Poster No.: C-0795 Congress: ECR 2013 Type: Scientific Exhibit Authors: A.

More information

Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010

Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 1 Ny teknologi: Ultralyd måler m blodstrøm Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 2 Conventional imaging methods of blood flow using ultrasound

More information

Second look ultrasound examination for breast lesions: MRI and pathologic correlation

Second look ultrasound examination for breast lesions: MRI and pathologic correlation Second look ultrasound examination for breast lesions: MRI and pathologic correlation Poster No.: C-0559 Congress: ECR 2015 Type: Scientific Exhibit Authors: E. Serrano Tamayo, E. López Soriano, M. Muñoz

More information

Diagnostic performance of MRI in differentiating metastatic from acute osteoporotic compression fractures of the spine

Diagnostic performance of MRI in differentiating metastatic from acute osteoporotic compression fractures of the spine Diagnostic performance of MRI in differentiating metastatic from acute osteoporotic compression fractures of the spine Poster No.: C-1399 Congress: ECR 2013 Type: Scientific Exhibit Authors: J. Martel,

More information

Ultrasound Fantomen. www.pi-medical.nl

Ultrasound Fantomen. www.pi-medical.nl Ultrasound Fantomen bootsen de echografische eigenschappen van menselijk weefsel na. In het weefselequivalente materiaal zijn structuren aangebracht die verschillende kwaliteitscontroles en trainingen

More information

Flow Assurance & Operability

Flow Assurance & Operability Flow Assurance & Operability Erosion due to sand production Date Business Name Overview 1. What causes erosion? 2. Sand production and transport 3. Sand management 4. Instrumentation / monitoring of sand

More information

Ultrasound Technologies. PD1v. vascular pocket doppler. Operating Instructions

Ultrasound Technologies. PD1v. vascular pocket doppler. Operating Instructions Ultrasound Technologies PD1v vascular pocket doppler Operating Instructions Before using your Pocket Doppler for the first time, please read these operating instructions carefully. 1 INTRODUCTION The UltraTec

More information

Edwards FloTrac Sensor & Edwards Vigileo Monitor. Measuring Continuous Cardiac Output with the FloTrac Sensor and Vigileo Monitor

Edwards FloTrac Sensor & Edwards Vigileo Monitor. Measuring Continuous Cardiac Output with the FloTrac Sensor and Vigileo Monitor Edwards FloTrac Sensor & Edwards Vigileo Monitor Measuring Continuous Cardiac Output with the FloTrac Sensor and Vigileo Monitor 1 Topics System Configuration Physiological Principles Pulse pressure relationship

More information

Extracranial Cerebrovascular Duplex Ultrasound Evaluation

Extracranial Cerebrovascular Duplex Ultrasound Evaluation VASCULAR TECHNOLOGY PROFESSIONAL PERFORMANCE GUIDELINES Extracranial Cerebrovascular Duplex Ultrasound Evaluation This Guideline was prepared by the Professional Guidelines Subcommittee of the Society

More information

Subclavian Steal Syndrome By Marta Thorup

Subclavian Steal Syndrome By Marta Thorup Subclavian Steal Syndrome By Marta Thorup Definition Subclavian steal syndrome (SSS), is a constellation of signs and symptoms that arise from retrograde flow of blood in the vertebral artery, due to proximal

More information

Ultrasound - Vascular

Ultrasound - Vascular Scan for mobile link. Ultrasound - Vascular Vascular ultrasound uses sound waves to evaluate the body s circulatory system and help identify blockages and detect blood clots. A Doppler ultrasound study

More information

3 rd Russian-Bavarian Conference on Bio-Medical Engineering

3 rd Russian-Bavarian Conference on Bio-Medical Engineering 3 rd Russian-Bavarian Conference on Bio-Medical Engineering Blood Pressure Estimation based on Pulse Transit Time and Compensation of Vertical Position Dipl.-Inform. Med. Christian Douniama Dipl.-Ing.

More information

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen.

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen. Scan for mobile link. Ultrasound - Abdomen Ultrasound imaging of the abdomen uses sound waves to produce pictures of the structures within the upper abdomen. It is used to help diagnose pain or distention

More information

Comparison to other anterior segment imaging devices

Comparison to other anterior segment imaging devices found helpful in a number of areas of ophthalmology. However, in the field of anterior segment tumors its usefulness is confined to a subset of tumors that are confined to the iris, hypo pigmented, and

More information

Applications of Doppler Ultrasound in Fetal Growth Assessment. David Cole

Applications of Doppler Ultrasound in Fetal Growth Assessment. David Cole Applications of Doppler Ultrasound in Fetal Growth Assessment David Cole Aims The aim of this presentation is to consider the use of Doppler ultrasound to investigate and monitor those pregnancies at risk

More information

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Various Technics of Liquids and Solids Level Measurements. (Part 3) (Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,

More information

Factors Affecting Blood Pressure. Vessel Elasticity Blood Volume Cardiac Output

Factors Affecting Blood Pressure. Vessel Elasticity Blood Volume Cardiac Output Factors that Affect Pressure Graphics are used with permission of: Pearson Education Inc., publishing as Benjamin Cummings (http://www.aw-bc.com) Page 1. Introduction pressure is affected by several factors:

More information

Respiratory dynamic MRI for determining aortic invasion of thoracic neoplasms

Respiratory dynamic MRI for determining aortic invasion of thoracic neoplasms Respiratory dynamic MRI for determining aortic invasion of thoracic neoplasms Poster No.: C-0760 Congress: ECR 2013 Type: Educational Exhibit Authors: Y. J. Hong, J. Hur, H.-J. Lee, Y. J. Kim, B. W. Choi;

More information

Vascular Technology (VT) Content Outline Anatomy & physiology 20% Cerebrovascular Cerebrovascular normal anatomy Evaluate the cerebrovascular vessels

Vascular Technology (VT) Content Outline Anatomy & physiology 20% Cerebrovascular Cerebrovascular normal anatomy Evaluate the cerebrovascular vessels Vascular Technology (VT) Content Outline Anatomy & physiology 20% normal anatomy Evaluate the cerebrovascular vessels hemodynamics Evaluate the cerebrovascular vessels for normal perfusion normal anatomy

More information

Fellow TEE Review Workshop Hemodynamic Calculations 2013. Director, Intraoperative TEE Program. Johns Hopkins School of Medicine

Fellow TEE Review Workshop Hemodynamic Calculations 2013. Director, Intraoperative TEE Program. Johns Hopkins School of Medicine Fellow TEE Review Workshop Hemodynamic Calculations 2013 Mary Beth Brady, MD, FASE Director, Intraoperative TEE Program Johns Hopkins School of Medicine At the conclusion of the workshop, the participants

More information

Customer: Cardiff Hospital NHS (United Kingdom) Machine:

Customer: Cardiff Hospital NHS (United Kingdom) Machine: Customer: Cardiff Hospital NHS (United Kingdom) Machine: Bench for leak test Model: Easy TS Page 1 of 7 Date: 6 th March 2012 1. Introduction. Easy TS is a compact and reliable table top bench designed

More information

Scanning Acoustic Microscopy Training

Scanning Acoustic Microscopy Training Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed

More information

Percutaneous Abscess Drainage

Percutaneous Abscess Drainage Scan for mobile link. Percutaneous Abscess Drainage An abscess is an infected fluid collection within the body. Percutaneous abscess drainage uses imaging guidance to place a thin needle through the skin

More information

Vascular System The heart can be thought of 2 separate pumps from the right ventricle, blood is pumped at a low pressure to the lungs and then back

Vascular System The heart can be thought of 2 separate pumps from the right ventricle, blood is pumped at a low pressure to the lungs and then back Vascular System The heart can be thought of 2 separate pumps from the right ventricle, blood is pumped at a low pressure to the lungs and then back to the left atria from the left ventricle, blood is pumped

More information

Aids needed for demonstrations: viscous fluid (water), tubes (pipes), injections, paper, stopwatches, vessels,, weights

Aids needed for demonstrations: viscous fluid (water), tubes (pipes), injections, paper, stopwatches, vessels,, weights 1 Viscous and turbulent flow Level: high school (16-17 years) hours (2 hours class teaching, 2 hours practical excercises) Content: 1. Viscous flow 2. Poiseuille s law 3. Passing from laminar to turbulent

More information

CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION

CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION CINDY WEILAND AND SANDRA L. KATANICK Continued innovations in noninvasive testing equipment provide skilled sonographers and physicians with the technology

More information

Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator

Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator Kenji MAWATARI, Koich SAMESHIMA, Mitsuyoshi MIYAI, Shinya MATSUDA Abstract We developed a new inkjet head by applying MEMS

More information

As published in PIM International

As published in PIM International As published in PIM International www.pim-international.com 64 Powder Injection Moulding International September 2012 Rapid prototyping of highperformance ceramics opens new opportunities for the CIM industry

More information

Duplication Images in Vascular Sonography

Duplication Images in Vascular Sonography Article Duplication Images in Vascular Sonography Jonathan M. Rubin, MD, PhD, Jing Gao, MD, Keith Hetel, MD, Robert Min, MD Objective. The purpose of this study was to determine the characteristics and

More information

Ultrasound in Vascular Surgery. Torbjørn Dahl

Ultrasound in Vascular Surgery. Torbjørn Dahl Ultrasound in Vascular Surgery Torbjørn Dahl 1 The field of vascular surgery Veins dilatation and obstruction (varicose veins and valve dysfunction) Arteries dilatation and narrowing (aneurysms and atherosclerosis)

More information

Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK.

Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK. The Physics of Flow Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK. Introduction Flow is defined as the quantity of fluid

More information

Convincing Deans that Ultrasound should be in the Medical Student Curriculum. Richard Hoppmann Dean University of South Carolina School of Medicine

Convincing Deans that Ultrasound should be in the Medical Student Curriculum. Richard Hoppmann Dean University of South Carolina School of Medicine Convincing Deans that Ultrasound should be in the Medical Student Curriculum Richard Hoppmann Dean University of South Carolina School of Medicine Convincing Deans that Ultrasound should be in the Medical

More information

Radial-axial Radial mixing is based on the premise that the fluids to be mixed enter the mixer in the correct proportions simultaneously

Radial-axial Radial mixing is based on the premise that the fluids to be mixed enter the mixer in the correct proportions simultaneously Brochure E-0501 1 Introduction. Static mixers are used for a wide range of applications including mixing, heat exchange and dispersion, due to numerous unique innovations our products are especially suitable

More information

PIPELINE LEAKAGE DETECTION USING FIBER-OPTIC DISTRIBUTED STRAIN AND TEMPERATURE SENSORS WHITE PAPER

PIPELINE LEAKAGE DETECTION USING FIBER-OPTIC DISTRIBUTED STRAIN AND TEMPERATURE SENSORS WHITE PAPER PIPELINE LEAKAGE DETECTION USING FIBER-OPTIC DISTRIBUTED STRAIN AND TEMPERATURE SENSORS WHITE PAPER Lufan Zou and Taha Landolsi OZ Optics Limited, 219 Westbrook Road, Ottawa, ON, Canada, K0A 1L0 E-mail:

More information

Hemodynamic Monitoring: Principles to Practice M. L. Cheatham, MD, FACS, FCCM

Hemodynamic Monitoring: Principles to Practice M. L. Cheatham, MD, FACS, FCCM SUMMARY HEMODYNAMIC MONITORING: FROM PRINCIPLES TO PRACTICE Michael L. Cheatham, MD, FACS, FCCM Director, Surgical Intensive Care Units Orlando Regional Medical Center Orlando, Florida Fluid-filled catheters

More information

Vascular ultrasound tests

Vascular ultrasound tests Critical Care, Theatre & Diagnostics Vascular ultrasound tests Information for patients Your doctor has referred you for a vascular ultrasound test. This leaflet explains what happens during the test,

More information

Micro-Optical Sensor Use in Boundary Layer Flows with Polymers and Bubbles

Micro-Optical Sensor Use in Boundary Layer Flows with Polymers and Bubbles 2 nd International Symposium on Seawater Drag Reduction Busan, Korea, 23-26 MAY 2005 Micro-Optical Sensor Use in Boundary Layer Flows with Polymers and Bubbles D. Modarress, P. Svitek (Measurement Science

More information

Mixing in the process industry: Chemicals Food Pharmaceuticals Paper Polymers Minerals Environmental. Chemical Industry:

Mixing in the process industry: Chemicals Food Pharmaceuticals Paper Polymers Minerals Environmental. Chemical Industry: Mixing Notes: Chapter 19 Robert P. Hesketh Mixing in the process industry: Chemicals Food Pharmaceuticals Paper Polymers Minerals Environmental Chemical Industry: Paints and Coatings Synthetic Rubbers

More information

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL

More information

FLUID FLOW STREAMLINE LAMINAR FLOW TURBULENT FLOW REYNOLDS NUMBER

FLUID FLOW STREAMLINE LAMINAR FLOW TURBULENT FLOW REYNOLDS NUMBER VISUAL PHYSICS School of Physics University of Sydney Australia FLUID FLOW STREAMLINE LAMINAR FLOW TURBULENT FLOW REYNOLDS NUMBER? What type of fluid flow is observed? The above pictures show how the effect

More information

Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment

Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment Fluid Structure Interaction VI 3 Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment J. Hengstler & J. Dual Department of Mechanical and Process

More information

Math Matters: Dissecting Hydrometer Calculations

Math Matters: Dissecting Hydrometer Calculations Math Matters: Dissecting Hydrometer Calculations By Jonathan Sirianni, Laboratory Assessor In the 16th century Galileo and the Tower of Pisa made famous the fact that two objects, no matter their mass,

More information

Lecture 5 Hemodynamics. Description of fluid flow. The equation of continuity

Lecture 5 Hemodynamics. Description of fluid flow. The equation of continuity 1 Lecture 5 Hemodynamics Description of fluid flow Hydrodynamics is the part of physics, which studies the motion of fluids. It is based on the laws of mechanics. Hemodynamics studies the motion of blood

More information

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers Flow and Flow rate. Laminar and Turbulent flow Laminar flow: smooth, orderly and regular Mechanical sensors have inertia, which can integrate out small variations due to turbulence Turbulent flow: chaotic

More information

ASSOCIATE OF APPLIED SCIENCE IN DIAGNOSTIC MEDICAL SONOGRAPHY (DMS AAS) 130 quarter credit units / 2,160 clock hours / 84 weeks (20-32 hours per

ASSOCIATE OF APPLIED SCIENCE IN DIAGNOSTIC MEDICAL SONOGRAPHY (DMS AAS) 130 quarter credit units / 2,160 clock hours / 84 weeks (20-32 hours per ASSOCIATE OF APPLIED SCIENCE IN DIAGNOSTIC MEDICAL SONOGRAPHY (DMS AAS) 130 quarter credit units / 2,160 clock hours / 84 weeks (20-32 hours per week) Educational Objective: The Associate of Applied Science

More information

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. ASGMT / Averaging Pitot Tube Flow Measurement Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. Averaging Pitot Tube Meters Introduction

More information

Steven R. McManus Environmental Product Support Teledyne Isco

Steven R. McManus Environmental Product Support Teledyne Isco Steven R. McManus Environmental Product Support Teledyne Isco Flow Measurement Applications and Technologies. Methods of flow measurement Technologies of flow measurement Flow Rate Applications Channel

More information

Evaluation of Initial Setting Time of Fresh Concrete

Evaluation of Initial Setting Time of Fresh Concrete Evaluation of Initial Setting Time of Fresh Concrete R R C Piyasena, P A T S Premerathne, B T D Perera, S M A Nanayakkara Abstract According to ASTM 403C, initial setting time of concrete is measured based

More information

Ultrasound Simulators

Ultrasound Simulators Ultrasound Simulators Tripp Bell, MD University of South Carolina School of Medicine Objectives Give a brief history of simulation in medicine Describe the types of ultrasound simulators Show how ultrasound

More information

Knowledge Domain: Plumbing Unit: Leaking Skill: Epoxy

Knowledge Domain: Plumbing Unit: Leaking Skill: Epoxy Knowledge Domain: Plumbing Unit: Leaking Skill: Epoxy Tools and Parts Required: 1) Jar 2) Oversized lid 3) Epoxy 4) Rubber tube 5) Soapy water 6) Safety goggles 7) Latex gloves 8) Scrap stick to mix epoxy

More information

DEEPSEALSM SERVICE SOLUTIONS

DEEPSEALSM SERVICE SOLUTIONS DEEP WATER CHEMICAL SOLUTIONS DEEPSEALSM SERVICE SOLUTIONS (DEEP SEA ENGINEERING ASSURANCE LEADERSHIP) Solving challenges. Deepwater Solutions from Multi-Chem ABOUT HALLIBURTON Founded in 1919, Halliburton

More information

Ultrasonic Wave Propagation Review

Ultrasonic Wave Propagation Review Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic

More information

Section Two: Arterial Pressure Monitoring

Section Two: Arterial Pressure Monitoring Section Two: Arterial Pressure Monitoring Indications An arterial line is indicated for blood pressure monitoring for the patient with any medical or surgical condition that compromises cardiac output,

More information

Collision of a small bubble with a large falling particle

Collision of a small bubble with a large falling particle EPJ Web of Conferences 67, 212 (214) DOI: 1.11/ epjconf/ 21467212 C Owned by the authors, published by EDP Sciences, 214 Collision of a small bubble with a large falling particle Jiri Vejrazka 1,a, Martin

More information

Surgery Support System as a Surgeon s Advanced Hand and Eye

Surgery Support System as a Surgeon s Advanced Hand and Eye Surgery Support System as a Surgeon s Advanced Hand and Eye 8 Surgery Support System as a Surgeon s Advanced Hand and Eye Kazutoshi Kan Michio Oikawa Takashi Azuma Shio Miyamoto OVERVIEW: A surgery support

More information

Choosing optimal rapid manufacturing process for thin-walled products using expert algorithm

Choosing optimal rapid manufacturing process for thin-walled products using expert algorithm Choosing optimal rapid manufacturing process for thin-walled products using expert algorithm Filip Górski, Wiesław Kuczko, Radosław Wichniarek, Adam Dudziak, Maciej Kowalski, Przemysław Zawadzki Poznan

More information

How does your heart pump blood in one direction?

How does your heart pump blood in one direction? Have a Heart How does your heart pump blood in one direction? Description How does your heart move blood in one direction, around in a loop? In this activity, you will make a model of one of the heart

More information

Upper Extremity Arterial Duplex Evaluation

Upper Extremity Arterial Duplex Evaluation VASCULAR TECHNOLOGY PROFESSIONAL PERFORMANCE GUIDELINES Upper Extremity Arterial Duplex Evaluation This Guideline was prepared by the Professional Guidelines Subcommittee of the Society for Vascular Ultrasound

More information

X-ray (Radiography) - Abdomen

X-ray (Radiography) - Abdomen Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,

More information

Monitoring of Cerebral Blood Flow. Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex)

Monitoring of Cerebral Blood Flow. Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex) Monitoring of Cerebral Blood Flow Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex) Ultrasound in Tissue Some Facts: blood cell tissue probe ultrasound travels at a constant

More information

Connectivity in a Wireless World. Cables Connectors 2014. A Special Supplement to

Connectivity in a Wireless World. Cables Connectors 2014. A Special Supplement to Connectivity in a Wireless World Cables Connectors 204 A Special Supplement to Signal Launch Methods for RF/Microwave PCBs John Coonrod Rogers Corp., Chandler, AZ COAX CABLE MICROSTRIP TRANSMISSION LINE

More information

ICH Topic Q4B Annex 5 Disintegration Test General Chapter. Step 3

ICH Topic Q4B Annex 5 Disintegration Test General Chapter. Step 3 European Medicines Agency June 2008 EMEA/CHMP/ICH/308895/2008 ICH Topic Q4B Annex 5 Disintegration Test General Chapter Step 3 ANNEX 5 TO NOTE FOR EVALUATION AND RECOMMENDATION OF PHARMACOPOEIAL TEXTS

More information

Oil Sight Glass BENEFITS

Oil Sight Glass BENEFITS 8 L-4090 Schroeder es provide maintenance and lubrication management professionals a complete and immediate visual oil analysis. Constructed of durable cast acrylic, they withstand most petroleum products

More information

BIOMEDICAL ULTRASOUND

BIOMEDICAL ULTRASOUND BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic

More information

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,

More information

10 is a soft polymer clay that can easily be moulded by hands (S1-Left). The simplest option is to create

10 is a soft polymer clay that can easily be moulded by hands (S1-Left). The simplest option is to create Electronic Supplementary Material (ESI) for Lab on a Chip. This journal is The Royal Society of Chemistry 2014 1 Angry pathogens, how to get rid of them: introducing microfluidics for 2 waterborne pathogen

More information

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Janet E. Semmens Sonoscan, Inc. 2149 E. Pratt Boulevard Elk Grove Village, IL 60007 USA Phone: (847)

More information

ASSESSMENT REPORT. Program: AAS Diagnostic Medical Sonography Cardiac Vascular Tract

ASSESSMENT REPORT. Program: AAS Diagnostic Medical Sonography Cardiac Vascular Tract ASSESSMENT REPORT Date Submitted: February 9, 2015 School: Engelstad School of Health Sciences Program: AAS Diagnostic Medical Sonography Cardiac Vascular Tract Submitted By: Tracy Lopez, Program Director

More information

EFSUMB EUROPEAN FEDERATION OF SOCIETIES FOR ULTRASOUND IN MEDICINE AND BIOLOGY Building a European Ultrasound Community

EFSUMB EUROPEAN FEDERATION OF SOCIETIES FOR ULTRASOUND IN MEDICINE AND BIOLOGY Building a European Ultrasound Community MINIMUM TRAINING REQUIREMENTS FOR THE PRACTICE OF MEDICAL ULTRASOUND IN EUROPE Appendix 8: Vascular Ultrasound Level 1 Training and Practice Practical training should involve at least two half day ultrasound

More information

A Measurement of 3-D Water Velocity Components During ROV Tether Simulations in a Test Tank Using Hydroacoustic Doppler Velocimeter

A Measurement of 3-D Water Velocity Components During ROV Tether Simulations in a Test Tank Using Hydroacoustic Doppler Velocimeter A Measurement of 3-D Water Velocity Components During ROV Tether Simulations in a Test Tank Using Hydroacoustic Doppler Velocimeter Leszek Kwapisz (*) Marek Narewski Lech A.Rowinski Cezary Zrodowski Faculty

More information