Physics of Colour and Doppler Lynette Hassall DMU AMS MLI Clinical Applications Specialist Sonosite Australasia Pty Ltd

Size: px
Start display at page:

Download "Physics of Colour and Doppler Lynette Hassall DMU AMS MLI Clinical Applications Specialist Sonosite Australasia Pty Ltd"

Transcription

1 These notes are intended to be an introduction to the subject only. I have tried to make the Doppler physics as simplistic as possible and convey a limited, basic overview of a very complex subject. I am attempting to provide you with a general understanding of how and why Doppler Ultrasound produces an image, and how to optimize those images to achieve the best from your machine, for the benefit of your patients. Please refer to Physics text books and articles for more complete explanations. Reading these notes does not imply any qualification or competence in Ultrasound physics. Attendance at a Practical Ultrasound Training Course is highly recommended Table of contents Concepts... 2 Colour... 6 Colour Power Doppeler... 7 Spectral Doppler Artifacts Getting started - Techniques References and Bibliography 22 1

2 Concepts All Colour and Doppler Applications in Ultrasound work from the Doppler Principle. The Doppler equation calculates the difference in received frequency from the transmitted frequency sent from the probe, measured in Hertz. From this, the Ultrasound machine calculates the velocity of the blood flow (using a complex mathematical formula called the Fast Fourier Transform [FFT]) and displays the result as a colour overlay on the B-mode display, an audible sound, or as a graph (Spectral Display). Red Blood Cells (RBC) act as Rayleigh scatterers this is a special type of ultrasound reflection pattern, the scatter is equal in all directions ie because the reflectors are so small the reflection is not angle dependent the main point is that, for Doppler purposes - the red blood cells act as point sources of ultrasound. We use this reflection pattern when we are obtaining a colour image or Doppler trace. Colour: The colour box is displayed as an overlay onto the B-Mode image. The crystals first send out a pulse of sound which produces the B-mode image, then a selected group of crystals send out a second pulse to form the colour image you see on the screen. Each colour line of sight has numerous sampling sites along it so you need to keep the colour box as narrow as possible to keep the frame rate high, which in turn provides a much faster update process and a more real-time display of the haemodynamics. The colour box may be angled at +15, 0, or -15 depending on the angle of the vessel to the transducer B-Mode plus or or When set up correctly, the colour display provides you with a qualitative impression of the flow characteristics of the vessel you are observing. No numerical value can be 2

3 assigned to the image, but the colour image can assist you in quickly and easily placing the spectral sample volume in the area of fastest flow. When positioning the colour box you are trying to achieve an angle with the direction of the blood flow in the vessel (this is the reason for the ability to change the angle of the colour box in some systems). Imagine the acute angle of the colour box is an arrow and point the arrow towards the most superficial portion of the vessel. This is the first step in optimizing the colour display. If the vessel itself is at an angle to the transducer, or is very deep, a straight box may be used. Pulse Repetition Frequency (PRF) may also be called the Velocity range or Scale. The PRF is the rate at which the lines of sight are sent out and received back to the transducer. The higher the PRF is, the greater the velocity of flow we can accurately display. We need to correlate this with the speed of flow in the vessel we are interrogating too high a PRF and we will not detect slow venous flow, too low a setting for the PRF and we will see aliasing in all vessels. 3

4 Filter also called Wall Filter, High Pass Filter, or Wall Thump filter blood moves within the vessel, the walls of the vessel also move slightly, and so does the surrounding tissue. This movement from the vessel walls and tissue are high amplitude (very strong) but low frequency (they are not moving as fast as the blood within the vessels) in order to avoid displaying these confusing signals, the filter can be set to remove echoes below a certain frequency. If the filter is set too low the wall signal will not be removed, and if it is set too high the signals from slow moving blood in a vein may also be removed. Slower flow towards the wall of the artery may also be removed and we will not see the colour fill to the vessel walls. The wall filter needs to be set appropriately for the vessel you are interrogating a good start is - High for arteries, Low for veins If the angle is incorrect there will be very poor representation of flow 4

5 It is important that you have a basic understanding of the Doppler equation used for this calculation as it helps to understand how to manipulate the probe and machine factors to obtain the best result possible. The Colour image, the Power image, and the Spectral Trace all work on the same principle, only the mode of display has changed. It will also help you to understand why sometimes, even though there is flow in the vessel, you may not be scanning at the correct angle to demonstrate it. This is a form of the Doppler equation calculating Doppler shift. f D ( khz ) = f t ( khz ) 2 u c ( m ( cm / s ) / s ) And may be written as Transmit. Frequency 2 velocity. of. moving. reflector cosine. Doppler. Angle Doppler. Shift = speed. of. sound. in. soft. tissue ƒ (d) Doppler shift the difference between the transmitted and received frequencies - this is measured by the system and is calculated in khz. u - is the velocity of the moving reflector (RBC) relative to the transducer this is what we are attempting to quantify in our spectral Doppler trace. 2 is used due to the pulse-echo effect from the Rayleigh scatterer or Red Blood Cells (RBC) the pulse from the transducer to a moving reflector (RBC) produces a Doppler shift at time of reception of the pulse, then a further Doppler shift occurs when the pulse from the RBC returns to the transducer. c is a constant the assumed speed of sound in soft tissue 1540 m/s ƒ (t) Transmitted frequency - this is the frequency emitted by the transducer Cos θ- the angle of approach of the flow direction of the red blood cells to the Doppler beam must be allowed for, this is done using the cosine of the Doppler angle we are using this to calculate the vector at which the red blood cell is moving relative to the transducer. By changing the formula around to solve the velocity of the moving reflector (written as u ) the system is able to calculate the speed of the reflector in cm/second (using the Fast Fourier Transform [FFT]) U = f ( t ) f ( d ) c 2 cos θ cos The effect of this is to produce a signal which is AUDIBLE. θ Velocity. of. moving. reflector = Transmit Doppler. frequency. shift 1540 m / s 2 Co sin e. Doppler. Angle Make sure the Volume control for your system is ON so that you are able to hear the signal. By listening to the pitch of the sound you will be able to hear when you are sampling and artery ( higher pitch, rhythmic variation), or a vein (lower pitch, whooshing sound). 5

6 Colour gives directional information for blood flow. This directional information is relative to the face of the transducer ie towards the transducer or away from the transducer. A transducer crystal is able to resonate to produce the sound wave, then stop and listen for the return echo it cannot both transmit and receive simultaneously. The colour box is superimposed over the B-mode Image;- first the B-mode image is generated then a second pulse of sound is transmitted (from a selected number of crystals) to produce the colour display. You can expect the frame rate to be slower due to this double transmit However, the faster the frame rate the more real-time the colour representation of the haemodynamics will be. In order to keep the frame rate as high as possible you should keep the colour box narrowed to as small an area as practical. Any movement is best seen at 0 - movement which is either directly towards or directly away from the transducer. As this is virtually impossible to achieve in real patients an angle equal to or less than 60 is used. If the angle of flow is 90 (perpendicular) to the beam there is no movement relative to the transducer so no colour (or very poor colour representation) will be seen. This is one reason why colour and Doppler should be avoided in Transverse if we are at 90 to the direction of blood flow, we do not see colour even if flow is present. We use the colour bar at the top of the image to indicate direction of flow the colour represented at the top of the bar indicates movement towards the probe, on the bottom is away from the probe. Be aware that the machine does not know whether you are examining an artery or a vein so do not be fooled into thinking that everything that has been encoded red on your image is an artery, or blue is a vein. You control the colour encoding by pressing the Invert button. Check the display for reference. To achieve good colour resolution keep the colour box as small as possible in both height and width. *When in colour mode the gain control will increase the brightness of the colour pixel representation on the screen. 6

7 Colour Power Doppler (CPD) or Amplitude Mapping is also known as Power Doppler, Power Doppler Imaging, or Power for short this is used when we simply want to see whether there is flow in a vessel or structure, or not. CPD is displayed as one colour only so does not give any directional information to the flow. Power is useful because it is not as angle dependent as colour, is more sensitive to low flow states, and has better edge resolution, This means we will detect smaller vessels (such as those in thyroid or testes), more easily, it will also display the trickle flow in stenotic arteries. Power may be used in the transverse view to show flow in the vessel. *To be absolutely sure, you will still need to check in longitudinal to confirm whether you are interrogating a vein or artery In these image we see the radial artery in transverse and longitudinal 7

8 Spectral Doppler also known as Pulsed Wave Doppler - is a graphical representation of flow velocity over time. The brightness of the pixels represents the amplitude (strength) of the returned echo, the gain control will alter this brightness to make the graph easier to see when in Doppler mode. Doppler angle Arteries and veins have characteristic patterns of flow and these are displayed on the graph. Sample Volume The Doppler line of sight is a representation of the beam from Doppler Line of Sight the transducer crystal. The sample volume is the box which represents the area of movement we are sampling, the sample volume is three dimensional, so it is really sampling a cube of tissue. This sample volume needs to be kept as small as possible so extraneous movement is not included in the signal. Wall motion or movement from other vessels in close proximity to the vessel being sampled can be represented on the display and cause confusion. The Doppler angle is the angle between the direction of flow and the line of sight, and must be at 60 or below. Any angle higher than this and the calculations of velocity of flow become increasingly inaccurate. The movement of the blood cells has two components, relative to transducer and relative to the body [towards or away from the heart]. This combination of movement gives us a vector and the angle measurement must be calculated using the cosine of the angle. Due to the cosine part of the Doppler equation the error margin changes with changing angle of insonation, so to maintain comparable measurements the angles must be kept the same eg for ICA/CCA ratio measurements. Any series of follow-up measurements must be also taken using the same Doppler angle for the results to be comparable (so the error margin will remain the same). The 60 angle is recommended, but this angle must also be lined up along the direction of flow within the vessel. This is usually achieved by aligning the Doppler angle with the vessel walls. *However, if there is a stenotic jet, the flow may be eccentrically placed within the vessel and the walls will not be a guide to the correct placement of the angle. This image shows a typical low resistance arterial waveform. The positioning of the trace above the base line indicates continuous positive direction of the flow (relative to the transducer) throughout the cycle. 8

9 By convention, write arterial flow above the baseline and venous flow below the baseline to provide a quick reference point for observers. Differentiation between arteries and veins using ultrasound: To begin with we observe the vessel in B-Mode *Remembering that we always assess the vessel in both transverse and longitudinal planes. Ultrasound is a dynamic examination; so follow the course of the vessel for a minimum of several cm before making any decision. An artery will display Pulsatility in response to systole and diastole Limited response to extrinsic compression, walls will not co-apt easily No change in caliber will be noted with breathing The three layers of the artery walls will be seen the intima and adventitia are seen as hyperechoic lines, separated by a hypoechoic layer the muscular medial layer A vein will display Phasicity in response to breathing, or as response to the Valsalva manoeuvre the vein will increase and decrease in caliber as the patient breathes, if it is close to the trunk. Small peripheral veins will not display phasicity but will show increased caliber if a tourniquet is applied to the limb. The vein walls will co-apt if compression is applied to the vessel. In a normal vein the vessel walls will collapse completely and touch each other (co-apt) if they do not then you are pressing on an artery OR the vein has either acute or chronic thrombus within it. *Compression should only be performed in the transverse plane, if you apply compression with the transducer in the longitudinal plane, it is easy to slip off the plane of the vessel and mistakenly think it is co-apting. No pulsatility will be demonstrated, however the closer you get to the heart, the more likely you will see transmitted pulsations, superimposed over the normal phasicity. Only a single layer to the vessel wall will be seen, there is no muscular layer visible 9

10 Veins have valves, and a valve sinus (widening) at the level of the valve, arteries do not. Colour variation between arteries and veins The flow towards the transducer is demonstrated as either red or blue in colour mode. Unfortunately the system does not automatically detect arterial flow and render the arteries red or venous flow and render the veins blue you need to determine whether the vessel is a vein or an artery and then choose which colour to display the vessel appropriately. By convention and for convenience - we tend to display the arteries red and the veins blue so as not to confuse observers. (If you look at any anatomy text book, arteries are always coloured red, and veins blue). On the colour display we observe the hue of the colour slow flow is displayed in darker hues, while faster flow is displayed as lighter brighter hues. In a typical artery we may demonstrate a darker colour close to the walls, shading to a brighter colour in the centre indicating parabolic flow fastest in the middle of the artery. Arteries will display a pulsatile flow in response to the cardiac cycle. In systole there will be an increased flow with a change in colour from a darker to a lighter red. In diastole you may see a flash of the opposite colour as we get a reverse component to the flow in diastole. This rhythmic change in colour in conjunction with the cardiac cycle is typical of the appearance of the arteries in colour mode. Veins will display a monophasic colour pattern, there will be no rhythmic change in flow pattern with cardiac cycle, instead the flow may cease completely if the limb is dependent, if a tourniquet is applied, or if the patient performs the valsalva manoeuvre. Another way to check whether you are observing a vein or an artery is to see if the flow augments with distal compression. If you are observing an arm vein and squeeze the hand then the flow in the vein will increase and a lighter hue colour will result. This augmentation does not occur with arteries. Flow in the vein and artery will be in different directions you should have one blue and one red vessel In this image we have the vertebral artery and vein running side by side, demonstrating different flow direction as different colours. 10

11 Spectral variation between arteries and veins When you are interrogating the arteries and veins using spectral Doppler you will see characteristic flow patterns in arteries and veins (more on this subject later) If you have the sound turned on you will hear a difference in the sound generated by the vessels. Veins have a characteristic, low pitched whooshing sound, arteries have a higher pitched, rhythmic variation which increases in pitch in systole, decreases in diastole and corresponds with the cardiac cycle. Turbulent or stenotic areas of flow will demonstrate loud sounds, almost like the sound the static on your TV screen makes if you are off channel. (this is difficult to describe on paper, but unmistakable when you have heard it) A point to note is that you do not NEED to view the vessel in Colour mode, prior to positioning the Doppler sample volume. You can simply position the sample volume within the vessel without activating the colour control. However, the Colour image is used to give us a quick look at the vessel to provide us with a quantitative impression of the flow characteristics. The colour image also allows us to position our sample volume quickly and accurately, within the fastest flow. The Doppler scale setting should be adjusted so the waveforms fill the window, without aliasing this allows us to measure the velocities more accurately, and obtain a better appreciation of the haemodynamics. 11

12 Spectral Doppler The sample volume needs to be kept as small as possible to accurately and unambiguously represent the maximum flow velocity in the vessel you are interrogating. If the sample volume is too wide then you are obtaining signals from all structures moving within the volume, including the vessel walls, and possibly any other vessel or moving structure adjacent to it. In this example the sample volume is much too large for the vessel. The wall movement, the slower blood flow at the periphery, and the very much faster flow in the centre of the vessel are going to be displayed. This will produce a very noisy spectral trace with a very wide range of velocities displayed. This wide range of velocities has been termed spectral broadening. True spectral broadening occurs at an area of stenosis, due to the turbulence of the flow, but if your sample volume is too large this can be produced artifactually. In this second example the sample volume has been decreased in size and now only the fastest flow in the centre of the vessel will be displayed, producing a much cleaner signal, with no artifactual spectral broadening. When you have the colour set up correctly you should take the time to systematically set up the Doppler line of sight, sample volume and Doppler angle to achieve the best Doppler signal possible for that vessel Align the Doppler line of sight so that it is parallel to the sides of the colour box When this is done set the Doppler sample volume so that it is as small as possible, and completely within the centre of the vessel we do not want to inadvertently sample the walls or surrounding tissue 12

13 Finally align the Doppler angle to 60 to the Doppler line of sight, BUT ALSO ALONG the direction of the fastest flow within the vessel. A normal vessel will usually have parabolic flow, which means that if you line up the Doppler angle with the vessel walls, the angle to the direction of flow should be approximately correct Remember this flow may not necessarily be aligned with the vessel walls if there is a stenosis and an eccentric jet, or if there is a bifurcatiorn or tortuosity of the vessel walls. Do not align the Doppler angle across the line of flow as you will only display a noisy spectral trace, with spectral broadening. Doppler angle across direction of flow Doppler angle aligned with direction of flow What do the signals mean? The graphical display gives us a picture of velocity along the y -axis and time across the x -axis of the graph. The brightness of the pixels on the graph displays the strength or amplitude of the returned signal this may be changed using the Gain control, when in Pulsed Wave mode. The Baseline represents zero flow direction with reference to the transducer; it can be referred to as the base line or the zero crossing line. Velocity Y axis Time Baseline X axis 13

14 The arterial wave forms are traditionally written above the base line, and venous waveforms are written below the baseline. High resistance arterial Waveform Low resistance arterial Waveform Venous Waveform Arterial waveforms usually show a rhythmic pattern of flow which corresponds with the cardiac cycle, a high resistance waveform displays a sharp upstroke in systole, with reverse flow (or no flow) in the diastolic phase. The third component of elastic recoil of the vessel walls, displays as a second smaller peak, before the cycle starts again. The low resistance waveform displays as a continuous forward flow throughout the whole cycle, with the velocity never dropping below the baseline. Venous flow displays as a monophasic continuous flow towards the heart. There is variation in velocity with respiration, or augmentation (flow increase) on distal compression. The closer you are to the heart, the more likely there will be transmitted pulsations superimposed on the phasic pattern. A display demonstrating spectral broadening shows infill of the spectral window (the clear area under the spectral trace) there are no clearly defined velocities so this appearance demonstrates either Turbulence Disturbed pattern of blood flow at a bifurcation Post stenotic area Sample volume is too large Sample volume too close to wall of vessel Doppler angle is across the lines of blood flow Spectral Broadening A pattern such as this one with spectral broadening and a high amplitude spike would indicate that the sample volume has been placed too close to the pulsing wall of an artery Wall Thump / Spectral Broadening 14

15 Artifacts. As with B-mode, colour mode and Doppler mode also have intrinsic artifacts associated with them. Attenuation As the ultrasound beam travels through tissue it is attenuated because the signals from red blood cells are much weaker than those from surrounding tissues, you need to use a lower frequency than you would expect for a vessel of given depth. You may also need to increase the gain to produce a colour image from a deep vessel. In a deep vessel you may not be able to demonstrate colour flow due to this beam attenuation. Place your Doppler sample volume into the vessel and try to obtain a spectral trace anyway, as the Spectral Doppler has a higher intensity and you may get the information needed. Acoustic shadowing Has a different effect sometimes the structure causing the acoustic shadowing will not be visible in the image, due to the angles of the lines of sight for the colour box and Doppler differing from the B-mode this becomes especially confusing if you can see the vessel clearly in the B-Mode image. For example a Subclavian artery may have a shadow over it from the clavicle, you would expect, because you can see the artery in B- Mode, that you would receive a Doppler signal from it however If the angle is incorrect you are trying to place your Doppler signal through the bone and so will produce acoustic shadowing for the Doppler signal different to that from the B-mode. In this first illustration the B-mode shadowing is clearly defined you can see the Subclavian vessel and the shadow from the clavicle 15

16 In this second illustration, when we activate the colour box, we observe a colour acoustic shadow and will not see colour in the vessel, even though the vessel is still clearly visible in the B-mode image. Colour signal does not travel through bone no colour fill is seen in vessel This image shows an abrupt lack of colour within the colour box, even though the vessel walls and lumen can still be seen in B-mode 16

17 Aliasing Is an indication that you are sampling too slowly for the velocities you are trying to document. Aliasing is demonstrated on your colour image by a change in colour from paler red hues, through white, to paler blue hues. This may not necessarily be an indication that you have disease present, with increased flow velocities, it may be that you have the settings incorrect. Aliasing is not necessarily a bad thing if you are trying to find flow in a vein and have your settings for low velocity flow, if you slide onto an artery the flow will be too fast, and you will see the aliasing and realize that you are not in the vein. If the vessel is stenotic then aliasing may provide a clue which tells you that there is an area of increased flow velocity and you need to examine that portion of the vessel carefully. In Doppler trace the signal will wrap around the baseline. Look at the soft keys, you have PRF (which stands for Pulse Repetition Frequency) and to the left of this control there is a separate control with labels for Low, Medium and High (these are the Sensitivity settings). The Scale controls the big jumps large changes in velocity The PRF control is the fine tuning for the colour flow. In simple terms if you have very fast flow change to HIGH, if you are looking for very slow flow or small vessels change to LOW. To overcome aliasing Lower the baseline Increase the PRF change the scale from Med to High Reduce the transmitted frequency of the beam change the Res or Gen setting to Pen Reduce the depth of the sample volume (although this should already be at the most superficial depth possible, for good B-mode optimization) Change of direction The colour display always provides directional information in relation to its motion - either towards or away from the transducer If you observe the colour bar at the corner of your image you can decide what colour has been allocated to flow towards the transducer. This colour bar displays red on the top, so red in this illustration represents flow towards the transducer 17

18 Tortuosity - A tortuous vessel can be confusing as it seems to have two directions of flow within it. If you think logically, you understand that the blood flow is in the same direction with respect to the body, so any confusing signal in the vessel needs to be interpreted - In this illustration of a tortuous vessel, the overall flow is towards the marker side of the image - however the flow begins towards the transducer (displayed as red), briefly it has no motion at all relative to the transducer (so no colour is displayed -shown as a black area), then the flow is away from the transducer (displayed as blue). To alleviate this problem the colour box can made smaller laterally, and the colour box angled appropriately, so only one side of the image has colour encoding at any one time. The flow hasn t changed, only our perceptions. be Alternatively the CPD may be activated to avoid confusion. 18

19 Electrical Interference Colour and Doppler signals are subject to interference from other sources of electrical signals. These signals are usually composed of straight lines superimposed over the colour or Doppler display. Very little can be done to minimize these artifacts other than recognize them for what they are. One possible solution is to unplug the system from the mains electrical supply and run on battery power. Incorrect Filter setting Incorrect adjustment of the Filter may cause errors in registration of the flow. In vessels with low flow, a high filter setting may filter out all of the flow signal, so no flow will be seen in the vessel. In a vessel with higher flow, the high filter may limit the colour fill and artifactually cause a loss of colour close to the vessel walls, mimicking fresh thrombus formation. In doppler With a spectral trace we chop off the lower velocities of flow and do not represent these in the image we see a black area close to the baseline 19

20 Mirror image artifact Occurs when there is a large impedance mismatch between structures The result is a structure being displayed twice one a mirror image of the other Soft tissue / bone interface Soft tissue / Lung interface Diaphragm / Lung Interface Bladder / Rectum or Bowel Interface The most interesting thing about the mirror image artifact is that it is present in all modes B-Mode, Power, Colour AND Doppler and we can achieve a spectral trace from the artifact, as well as the from the real vessel. Note in the last image, the sample volume is placed within the artifactual vessel Colour Flash Artifact Colour is written within the colour box over EVERYTHING that is moving, it does not simply pick out the blood flow and display it for us. Because the colour is written over everything within the colour box that moves, we get a flash artifact if we move the transducer, or if the patient twitches or moves their muscles, or if peristalsis occurs, or if breathing occurs, or if the heart beats. To avoid this artifact keep your transducer face very still, ask the patient to keep still, but realize that sometimes this artifact is unavoidable, as we can t always control breathing, peristalsis or heart motion 20

21 Getting started - Techniques Physics of Colour and Doppler 1 - Optimise the B-Mode Image You will not be able to obtain a good colour image or Doppler trace if you have not understood the principles of optimization in B-mode. 2 Apply generous quantities of gel to the skin surface. 3 Adjust the depth so that the structure you are interested in is as large as possible on the screen (the object fills the screen) 4 Optimize the Gain and Time Gain Compensation (TGC Near and Far Gain) controls to achieve a balanced image 5 Survey the vessel in both longitudinal and transverse 6 Observe the B-mode characteristics of the vessel and try to determine whether it is a vein or an artery 7 Turn into longitudinal and display the full length of the vessel across the screen. Appearance on screen 8 Tip or tilt the transducer produce an angle with the vessel you may need to make a puddle of gel and rest one side of the transducer in it to obtain an angle, this provides contact with the gel, and does not compress small superficial vessels. Appearance on screen 9 Activate the colour Box 10 Point the arrow side of the colour box towards the most superficial part of the vessel, to optimize the angle between Colour Doppler lines of sight and flow 21

22 If there is no flow displayed, Lift the probe pressure are you pushing too hard and compressing the vessel? Decrease the PRF or Scale to Low Turn the wall filter to Low Increase the colour Gain Augment the vessel distally If there is still no flow displayed Change to CPD and reassess 11 When you have colour flow within the colour box, optimize the colour parameters to achieve the optimum colour image. Decide whether you are looking at a vein or an artery and make sure the colour is set up so that if the vessel is an artery it is displayed as red, and if it is a vein display it as blue. 12 Narrow the colour box, also make it as short vertically as practical, to keep the frame rate as high as possible. You will not be able to display a good Spectral trace if you start with a poor Colour Image. Colour and Spectral Doppler work on the same principle so Optimise the colour first then the Spectral Doppler. 13 Activate the pulsed wave Doppler 14 Align the Doppler line of sight so that it is parallel to the sides of the colour box 15 Then set the Doppler sample volume so that it is as small as possible, and completely within the vessel you are interrogating we do not want to inadvertently sample the walls or surrounding tissue 16 Finally align the Doppler angle to 60 to the Doppler line of sight BUT ALSO ALONG the direction of the fastest flow within the vessel. *Remember this may not necessarily be aligned with the vessel walls if there is a stenosis with an eccentric jet, or bifurcation or tortuosity within the vessel itself. There is no use lining up the Doppler angle across the line of flow as you will only display a noisy spectral trace. 17 Look at the trace on the screen o Is there spectral broadening? o Is it arterial or venous? 18 Listen to the audible signal 19 Optimise the signal you are achieving. o Decrease the size of the sample volume o Align the angle correctly to the flow o Increase the Gain 20 Freeze the image 21 Measure velocities (if necessary) 22 Document, by Saving the image. 22

23 References & Bibliography Allen, P; Dubbins, P; Pozniak, M; McDicken, W N; Clinical Doppler Ultrasound, 2003, Churchill Livingstone (Pub) Gent, R Applied Physics and Technology of Diagnostic Ultrasound (1997) Openbook (Pub) Kremkau F Diagnostic Ultrasound Principles and Instruments (2006) 7 th Ed, Saunders Elsevier (Pub) O Regan, S Haemodynamics of peripheral arterial waveforms Article soundeffects issue 4, 2008 pp Rose, M Did you put colour on that? Article soundeffects Issue 2, 2006, pp Stockdale, A Doppler at 60 - Magical or mythical Article soundeffects issue 3, 2006, pp Szabo T, Diagnostic Ultrasound Imaging Inside Out (2004) Elsevier (Pub) Zweibel, W Introduction to Vascular Ultrasonography 4 th Ed (2000) Saunders (Pub) 23

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

Vascular Laboratory Fundamentals. Robert Mitchell MD, RPVI Duke University Medical Center 5/2/08

Vascular Laboratory Fundamentals. Robert Mitchell MD, RPVI Duke University Medical Center 5/2/08 Vascular Laboratory Fundamentals Robert Mitchell MD, RPVI Duke University Medical Center 5/2/08 Goals of Lecture Understand basic ultrasound principles Understand normal and abnormal arterial hemodynamics

More information

Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect

Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect Objectives: PS-7.1 Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect Illustrate ways that the energy of waves is transferred by interaction with

More information

VFI Technology to Change the Way You Work

VFI Technology to Change the Way You Work Technology to Change the Way You Work Vascular Ultrasound Made Easier Vector Flow Imaging VFI VFI is a ground-breaking technology that can revolutionize the workflow for many Doppler ultrasound applications.

More information

AP1 Waves. (A) frequency (B) wavelength (C) speed (D) intensity. Answer: (A) and (D) frequency and intensity.

AP1 Waves. (A) frequency (B) wavelength (C) speed (D) intensity. Answer: (A) and (D) frequency and intensity. 1. A fire truck is moving at a fairly high speed, with its siren emitting sound at a specific pitch. As the fire truck recedes from you which of the following characteristics of the sound wave from the

More information

BASIC PHYSICAL PRINCIPLES OF ULTRASOUND

BASIC PHYSICAL PRINCIPLES OF ULTRASOUND BASIC PHYSICAL PRINCIPLES OF ULTRASOUND BASIC PHYSICAL PRINCIPLES OF ULTRASOUND INTRODUCTION C ompetent performance of an ultrasound examination of any type is reliant on an understanding of the basic

More information

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL

More information

Duplication Images in Vascular Sonography

Duplication Images in Vascular Sonography Article Duplication Images in Vascular Sonography Jonathan M. Rubin, MD, PhD, Jing Gao, MD, Keith Hetel, MD, Robert Min, MD Objective. The purpose of this study was to determine the characteristics and

More information

A Fishy Tale. Observing the Circulatory System of a Goldfish with a Compound Light Microscope

A Fishy Tale. Observing the Circulatory System of a Goldfish with a Compound Light Microscope A Fishy Tale Observing the Circulatory System of a Goldfish with a Compound Light Microscope A Fishy Tale About this Lesson In this lesson, students will explore a computer animation of the human body

More information

Waves Sound and Light

Waves Sound and Light Waves Sound and Light r2 c:\files\courses\1710\spr12\wavetrans.doc Ron Robertson The Nature of Waves Waves are a type of energy transmission that results from a periodic disturbance (vibration). They are

More information

Ultrasound Technologies. PD1v. vascular pocket doppler. Operating Instructions

Ultrasound Technologies. PD1v. vascular pocket doppler. Operating Instructions Ultrasound Technologies PD1v vascular pocket doppler Operating Instructions Before using your Pocket Doppler for the first time, please read these operating instructions carefully. 1 INTRODUCTION The UltraTec

More information

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

FXA 2008. UNIT G485 Module 4 5.4.3 Ultrasound. Candidates should be able to :

FXA 2008. UNIT G485 Module 4 5.4.3 Ultrasound. Candidates should be able to : 1 Candidates should be able to : ULTRASOUND Describe the properties of ultrasound. ULTRASOUND is any sound wave having a frequency greater than the upper frequency limit of human hearing (20 khz). Describe

More information

Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide)

Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide) Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide) OVERVIEW Students will measure a sound wave by placing the Ward s DataHub microphone near one tuning fork A440 (f=440hz). Then

More information

Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010

Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 1 Ny teknologi: Ultralyd måler m blodstrøm Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 2 Conventional imaging methods of blood flow using ultrasound

More information

Rouleaux flow, Sagittal. Rouleaux flow, transverse. Doppler Overview. Doppler Overview. Why we hate Doppler physics!!

Rouleaux flow, Sagittal. Rouleaux flow, transverse. Doppler Overview. Doppler Overview. Why we hate Doppler physics!! Doppler Technology & Optimization Doppler From A Z Donald T. Milburn, RDCS, RVT, FSDMS Prescott, Az. San Diego local ultrasound society May 12, 2009 Donald T. Milburn, RDCS, RVT, NACT, FSDMS Director,

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

Scanning Acoustic Microscopy Training

Scanning Acoustic Microscopy Training Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed

More information

v = fλ PROGRESSIVE WAVES 1 Candidates should be able to :

v = fλ PROGRESSIVE WAVES 1 Candidates should be able to : PROGRESSIVE WAVES 1 Candidates should be able to : Describe and distinguish between progressive longitudinal and transverse waves. With the exception of electromagnetic waves, which do not need a material

More information

Ultrasound. Sound waves

Ultrasound. Sound waves Ultrasound Basic Idea Send waves into body which are reflected at the interfaces between tissue Return time of the waves tells us of the depth of the reflecting surface History First practical application,

More information

Ultrasound. for Vascular Access and Regional Anesthesia. Brian A. Pollard, md, med

Ultrasound. for Vascular Access and Regional Anesthesia. Brian A. Pollard, md, med Ultrasound Guidance for Vascular Access and Regional Anesthesia Brian A. Pollard, md, med ULTRASOUND GUIDANCE for Vascular Access and Regional Anesthesia Brian A. Pollard BSc, MD, MEd, FRCPC www.usrabook.com

More information

Encoded Phased Array Bridge Pin Inspection

Encoded Phased Array Bridge Pin Inspection Encoded Phased Array Bridge Pin Inspection James S. Doyle Baker Testing Services, Inc. 22 Reservoir Park Dr. Rockland, MA 02370 (781) 871-4458; fax (781) 871-0123; e-mail jdoyle@bakertesting.com Product

More information

Copyright 2008 Pearson Education, Inc., publishing as Pearson Addison-Wesley.

Copyright 2008 Pearson Education, Inc., publishing as Pearson Addison-Wesley. Chapter 20. Traveling Waves You may not realize it, but you are surrounded by waves. The waviness of a water wave is readily apparent, from the ripples on a pond to ocean waves large enough to surf. It

More information

Introduction to acoustic imaging

Introduction to acoustic imaging Introduction to acoustic imaging Contents 1 Propagation of acoustic waves 3 1.1 Wave types.......................................... 3 1.2 Mathematical formulation.................................. 4 1.3

More information

Monitoring of Cerebral Blood Flow. Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex)

Monitoring of Cerebral Blood Flow. Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex) Monitoring of Cerebral Blood Flow Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex) Ultrasound in Tissue Some Facts: blood cell tissue probe ultrasound travels at a constant

More information

Describing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power

Describing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power Parameters used to completely characterize a sound wave Describing Sound Waves Chapter 3 Period Frequency Amplitude Power Intensity Speed Wave Length Period Defined as the time it take one wave vibrate

More information

(Refer Slide Time: 2:03)

(Refer Slide Time: 2:03) Control Engineering Prof. Madan Gopal Department of Electrical Engineering Indian Institute of Technology, Delhi Lecture - 11 Models of Industrial Control Devices and Systems (Contd.) Last time we were

More information

LAB 6: GRAVITATIONAL AND PASSIVE FORCES

LAB 6: GRAVITATIONAL AND PASSIVE FORCES 55 Name Date Partners LAB 6: GRAVITATIONAL AND PASSIVE FORCES And thus Nature will be very conformable to herself and very simple, performing all the great Motions of the heavenly Bodies by the attraction

More information

v = λ f this is the Golden Rule for waves transverse & longitudinal waves Harmonic waves The golden rule for waves Example: wave on a string Review

v = λ f this is the Golden Rule for waves transverse & longitudinal waves Harmonic waves The golden rule for waves Example: wave on a string Review L 23 Vibrations and Waves [3] resonance clocks pendulum springs harmonic motion mechanical waves sound waves golden rule for waves musical instruments The Doppler effect Doppler radar radar guns Review

More information

Subclavian Steal Syndrome By Marta Thorup

Subclavian Steal Syndrome By Marta Thorup Subclavian Steal Syndrome By Marta Thorup Definition Subclavian steal syndrome (SSS), is a constellation of signs and symptoms that arise from retrograde flow of blood in the vertebral artery, due to proximal

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

Basics of Ultrasound Imaging

Basics of Ultrasound Imaging 2 Basics of Ultrasound Imaging Vincent Chan and Anahi Perlas Introduction... Basic Principles of B-Mode US... Generation of Ultrasound Pulses... Ultrasound Wavelength and Frequency... Ultrasound Tissue

More information

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law.

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law. 260 17-1 I. THEORY EXPERIMENT 17 QUALITATIVE STUDY OF INDUCED EMF Along the extended central axis of a bar magnet, the magnetic field vector B r, on the side nearer the North pole, points away from this

More information

Waves-Wave Characteristics

Waves-Wave Characteristics 1. What is the wavelength of a 256-hertz sound wave in air at STP? 1. 1.17 10 6 m 2. 1.29 m 3. 0.773 m 4. 8.53 10-7 m 2. The graph below represents the relationship between wavelength and frequency of

More information

LAB 6 - GRAVITATIONAL AND PASSIVE FORCES

LAB 6 - GRAVITATIONAL AND PASSIVE FORCES L06-1 Name Date Partners LAB 6 - GRAVITATIONAL AND PASSIVE FORCES OBJECTIVES And thus Nature will be very conformable to herself and very simple, performing all the great Motions of the heavenly Bodies

More information

Transmission Line and Back Loaded Horn Physics

Transmission Line and Back Loaded Horn Physics Introduction By Martin J. King, 3/29/3 Copyright 23 by Martin J. King. All Rights Reserved. In order to differentiate between a transmission line and a back loaded horn, it is really important to understand

More information

physics 1/12/2016 Chapter 20 Lecture Chapter 20 Traveling Waves

physics 1/12/2016 Chapter 20 Lecture Chapter 20 Traveling Waves Chapter 20 Lecture physics FOR SCIENTISTS AND ENGINEERS a strategic approach THIRD EDITION randall d. knight Chapter 20 Traveling Waves Chapter Goal: To learn the basic properties of traveling waves. Slide

More information

BIOMEDICAL ULTRASOUND

BIOMEDICAL ULTRASOUND BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic

More information

Page 1. Introduction The blood vessels of the body form a closed delivery system that begins and ends at the heart.

Page 1. Introduction The blood vessels of the body form a closed delivery system that begins and ends at the heart. Anatomy Review: Blood Vessel Structure & Function Graphics are used with permission of: Pearson Education Inc., publishing as Benjamin Cummings (http://www.aw-bc.com) Page 1. Introduction The blood vessels

More information

Robot Perception Continued

Robot Perception Continued Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart

More information

PRINCIPLES OF DOPPLER ECHOCARDIOGRAPHY AND THE DOPPLER EXAMINATION #1

PRINCIPLES OF DOPPLER ECHOCARDIOGRAPHY AND THE DOPPLER EXAMINATION #1 PRINCIPLES OF DOPPLER ECHOCARDIOGRAPHY AND THE DOPPLER EXAMINATION #1 Joseph A. Kisslo, MD David B. Adams, RDCS INTRODUCTION Doppler echocardiography is a method for detecting the direction and velocity

More information

1) The time for one cycle of a periodic process is called the A) wavelength. B) period. C) frequency. D) amplitude.

1) The time for one cycle of a periodic process is called the A) wavelength. B) period. C) frequency. D) amplitude. practice wave test.. Name Use the text to make use of any equations you might need (e.g., to determine the velocity of waves in a given material) MULTIPLE CHOICE. Choose the one alternative that best completes

More information

Answer the following questions during or after your study of Wave Properties. 4. How are refraction and the speed of wave in different media related?

Answer the following questions during or after your study of Wave Properties. 4. How are refraction and the speed of wave in different media related? Wave Properties Student Worksheet Answer the following questions during or after your study of Wave Properties. 1. A person standing 385 m from a cliff claps her hands loudly, only to hear the sound return

More information

Reflection and Refraction

Reflection and Refraction Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,

More information

Project 4.2.1: Heart Rate

Project 4.2.1: Heart Rate Project 4.2.1: Heart Rate Introduction Even before you were born, one of the first things your doctor did when you went for an office visit was listen to your heart. Your heart rate, the number of times

More information

INTRODUCTION TO RENDERING TECHNIQUES

INTRODUCTION TO RENDERING TECHNIQUES INTRODUCTION TO RENDERING TECHNIQUES 22 Mar. 212 Yanir Kleiman What is 3D Graphics? Why 3D? Draw one frame at a time Model only once X 24 frames per second Color / texture only once 15, frames for a feature

More information

Atomic Force Microscope

Atomic Force Microscope Atomic Force Microscope (Veeco Nanoman) User Manual Basic Operation 4 th Edition Aug 2012 NR System Startup If the system is currently ON To start the NanoScope software, double-click the NanoScope startup

More information

Fellow TEE Review Workshop Hemodynamic Calculations 2013. Director, Intraoperative TEE Program. Johns Hopkins School of Medicine

Fellow TEE Review Workshop Hemodynamic Calculations 2013. Director, Intraoperative TEE Program. Johns Hopkins School of Medicine Fellow TEE Review Workshop Hemodynamic Calculations 2013 Mary Beth Brady, MD, FASE Director, Intraoperative TEE Program Johns Hopkins School of Medicine At the conclusion of the workshop, the participants

More information

Resonance in a Closed End Pipe

Resonance in a Closed End Pipe Experiment 12 Resonance in a Closed End Pipe 12.1 Objectives Determine the relationship between frequency and wavelength for sound waves. Verify the relationship between the frequency of the sound, the

More information

Ultrasonic Wave Propagation Review

Ultrasonic Wave Propagation Review Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic

More information

Our Human Body On-site student activities Years 5 6

Our Human Body On-site student activities Years 5 6 Our Human Body On-site student activities Years 5 6 Our Human Body On-site student activities: Years 5-6 Student activity (and record) sheets have been developed with alternative themes for students to

More information

Direct and Reflected: Understanding the Truth with Y-S 3

Direct and Reflected: Understanding the Truth with Y-S 3 Direct and Reflected: Understanding the Truth with Y-S 3 -Speaker System Design Guide- December 2008 2008 Yamaha Corporation 1 Introduction Y-S 3 is a speaker system design software application. It is

More information

Clinical Training for Visage 7 Cardiac. Visage 7

Clinical Training for Visage 7 Cardiac. Visage 7 Clinical Training for Visage 7 Cardiac Visage 7 Overview Example Usage 3 Cardiac Workflow Examples 4 Remove Chest Wall 5 Edit Chest Wall Removal 6 Object Display Popup 7 Selecting Optimal Phase 8 Thick

More information

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT 4 Understand single-phase alternating current (ac) theory Single phase AC

More information

Experiment 1: SOUND. The equation used to describe a simple sinusoidal function that propagates in space is given by Y = A o sin(k(x v t))

Experiment 1: SOUND. The equation used to describe a simple sinusoidal function that propagates in space is given by Y = A o sin(k(x v t)) Experiment 1: SOUND Introduction Sound is classified under the topic of mechanical waves. A mechanical wave is a term which refers to a displacement of elements in a medium from their equilibrium state,

More information

The Sonometer The Resonant String and Timbre Change after plucking

The Sonometer The Resonant String and Timbre Change after plucking The Sonometer The Resonant String and Timbre Change after plucking EQUIPMENT Pasco sonometers (pick up 5 from teaching lab) and 5 kits to go with them BK Precision function generators and Tenma oscilloscopes

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 1/2016 Sound 1/8 INTERFERENCE OF SOUND WAVES PURPOSE: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves and to observe interference phenomena with ultrasonic sound waves.

More information

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen.

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen. Scan for mobile link. Ultrasound - Abdomen Ultrasound imaging of the abdomen uses sound waves to produce pictures of the structures within the upper abdomen. It is used to help diagnose pain or distention

More information

Experiment 5: Magnetic Fields of a Bar Magnet and of the Earth

Experiment 5: Magnetic Fields of a Bar Magnet and of the Earth MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2005 Experiment 5: Magnetic Fields of a Bar Magnet and of the Earth OBJECTIVES 1. To examine the magnetic field associated with a

More information

After a wave passes through a medium, how does the position of that medium compare to its original position?

After a wave passes through a medium, how does the position of that medium compare to its original position? Light Waves Test Question Bank Standard/Advanced Name: Question 1 (1 point) The electromagnetic waves with the highest frequencies are called A. radio waves. B. gamma rays. C. X-rays. D. visible light.

More information

PICC- Peripherally Inserted Central Catheter PROCEDURAL CONSENT FORM. A. Interpreter / cultural needs. B. Procedure. C. Risks of the procedure

PICC- Peripherally Inserted Central Catheter PROCEDURAL CONSENT FORM. A. Interpreter / cultural needs. B. Procedure. C. Risks of the procedure DO NOT WRITE IN THIS BINDING MARGIN v3.00-03/2011 SW9266 Facility: PICC- Peripherally Inserted Central Catheter A. Interpreter / cultural needs An Interpreter Service is required? Yes No If Yes, is a qualified

More information

The EasySense unit can detect that the Smart Q Heart Rate Sensor is connected and the range it is set to.

The EasySense unit can detect that the Smart Q Heart Rate Sensor is connected and the range it is set to. Heart Rate Sensor Heart Rate Sensor (Product No PC-3147) Pulse rate Range: 0 to 200 bpm Resolution: 1 bpm Waveform Range: -2000 to 2000 mv Resolution: 1 mv Introduction The Smart Q Heart Rate Sensor monitors

More information

Crystal Optics of Visible Light

Crystal Optics of Visible Light Crystal Optics of Visible Light This can be a very helpful aspect of minerals in understanding the petrographic history of a rock. The manner by which light is transferred through a mineral is a means

More information

Physics 9e/Cutnell. correlated to the. College Board AP Physics 1 Course Objectives

Physics 9e/Cutnell. correlated to the. College Board AP Physics 1 Course Objectives Physics 9e/Cutnell correlated to the College Board AP Physics 1 Course Objectives Big Idea 1: Objects and systems have properties such as mass and charge. Systems may have internal structure. Enduring

More information

Germanium Diode AM Radio

Germanium Diode AM Radio Germanium Diode AM Radio LAB 3 3.1 Introduction In this laboratory exercise you will build a germanium diode based AM (Medium Wave) radio. Earliest radios used simple diode detector circuits. The diodes

More information

Samsung Ultrasound H60

Samsung Ultrasound H60 Samsung Medison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer around

More information

So, you want to make a photo-realistic rendering of the Earth from orbit, eh? And you want it to look just like what astronauts see from the shuttle

So, you want to make a photo-realistic rendering of the Earth from orbit, eh? And you want it to look just like what astronauts see from the shuttle So, you want to make a photo-realistic rendering of the Earth from orbit, eh? And you want it to look just like what astronauts see from the shuttle or ISS (International Space Station). No problem. Just

More information

Human Anatomy and Physiology II Laboratory

Human Anatomy and Physiology II Laboratory Human Anatomy and Physiology II Laboratory The Circulation (Two Weeks) 1 This lab involves two weeks work studying the vasculature of the human body. Both weeks involve the exercise in the lab manual entitled

More information

Normal Intracardiac Pressures. Lancashire & South Cumbria Cardiac Network

Normal Intracardiac Pressures. Lancashire & South Cumbria Cardiac Network Normal Intracardiac Pressures Lancashire & South Cumbria Cardiac Network Principle Pressures recorded from catheter tip Electrical transducer - wheatstone bridge mechanical to electrical waveform display

More information

K2 CW Filter Alignment Procedures Using Spectrogram 1 ver. 5 01/17/2002

K2 CW Filter Alignment Procedures Using Spectrogram 1 ver. 5 01/17/2002 K2 CW Filter Alignment Procedures Using Spectrogram 1 ver. 5 01/17/2002 It will be assumed that you have already performed the RX alignment procedures in the K2 manual, that you have already selected the

More information

75cm ODU/SurfBeam 2 Point and Peak Job Aid

75cm ODU/SurfBeam 2 Point and Peak Job Aid Summary This Job Aid covers: 75cm ODU/SurfBeam 2 Point and Peak Job Aid Preparing the Antenna for Pointing and Peaking Configure the SurfBeam 2 Modem and 75cm TRIA Point Elevation Set the Skew Point Azimuth

More information

Upper Extremity Vein Mapping for Placement of a Dialysis Access

Upper Extremity Vein Mapping for Placement of a Dialysis Access VASCULAR TECHNOLOGY PROFESSIONAL PERFORMANCE GUIDELINES Upper Extremity Vein Mapping for Placement of a Dialysis Access This Guideline was prepared by the Professional Guidelines Subcommittee of the Society

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 2011 Interference - 1 INTERFERENCE OF SOUND WAVES The objectives of this experiment are: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves. To observe interference phenomena

More information

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

WAVELENGTH OF LIGHT - DIFFRACTION GRATING PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant

More information

Laboratory Guide. Anatomy and Physiology

Laboratory Guide. Anatomy and Physiology Laboratory Guide Anatomy and Physiology TBME04, Fall 2010 Name: Passed: Last updated 2010-08-13 Department of Biomedical Engineering Linköpings Universitet Introduction This laboratory session is intended

More information

Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle

Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle Objectives Students will record a six-lead ECG from a resting subject and determine the QRS axis of the

More information

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS Blood Pressure Project 9 Blood pressure is a measure of the fluid pressure within the circulatory system. This pressure is required to ensure the delivery of oxygen and nutrients to, and the removal of

More information

Principles of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University

Principles of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University Principles of Medical Ultrasound Pai-Chi Li Department of Electrical Engineering National Taiwan University What is Medical Ultrasound? Prevention: actions taken to avoid diseases. Diagnosis: the process

More information

Sample Questions for the AP Physics 1 Exam

Sample Questions for the AP Physics 1 Exam Sample Questions for the AP Physics 1 Exam Sample Questions for the AP Physics 1 Exam Multiple-choice Questions Note: To simplify calculations, you may use g 5 10 m/s 2 in all problems. Directions: Each

More information

SR2000 FREQUENCY MONITOR

SR2000 FREQUENCY MONITOR SR2000 FREQUENCY MONITOR THE FFT SEARCH FUNCTION IN DETAILS FFT Search is a signal search using FFT (Fast Fourier Transform) technology. The FFT search function first appeared with the SR2000 Frequency

More information

Acceleration levels of dropped objects

Acceleration levels of dropped objects Acceleration levels of dropped objects cmyk Acceleration levels of dropped objects Introduction his paper is intended to provide an overview of drop shock testing, which is defined as the acceleration

More information

Applications of Doppler Ultrasound in Fetal Growth Assessment. David Cole

Applications of Doppler Ultrasound in Fetal Growth Assessment. David Cole Applications of Doppler Ultrasound in Fetal Growth Assessment David Cole Aims The aim of this presentation is to consider the use of Doppler ultrasound to investigate and monitor those pregnancies at risk

More information

A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)

A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time) chapter 2 Pulsed Ultrasound In diagnostic ultrasound imaging, short bursts, or pulses, of acoustic energy are used to create anatomic images. Continuous wave sound cannot create anatomic images. Analogy

More information

SCREENING COMPRESSION ULTRASOUND FOR LOWER EXTREMITY DVT

SCREENING COMPRESSION ULTRASOUND FOR LOWER EXTREMITY DVT SCREENING COMPRESSION ULTRASOUND FOR LOWER EXTREMITY DVT R. Eugene Zierler, M.D. The D. E. Strandness, Jr. Vascular Laboratory University of Washington Medical Center Division of Vascular Surgery University

More information

ANALYTICAL METHODS FOR ENGINEERS

ANALYTICAL METHODS FOR ENGINEERS UNIT 1: Unit code: QCF Level: 4 Credit value: 15 ANALYTICAL METHODS FOR ENGINEERS A/601/1401 OUTCOME - TRIGONOMETRIC METHODS TUTORIAL 1 SINUSOIDAL FUNCTION Be able to analyse and model engineering situations

More information

Section 5.0 : Horn Physics. By Martin J. King, 6/29/08 Copyright 2008 by Martin J. King. All Rights Reserved.

Section 5.0 : Horn Physics. By Martin J. King, 6/29/08 Copyright 2008 by Martin J. King. All Rights Reserved. Section 5. : Horn Physics Section 5. : Horn Physics By Martin J. King, 6/29/8 Copyright 28 by Martin J. King. All Rights Reserved. Before discussing the design of a horn loaded loudspeaker system, it is

More information

Lecture 1-10: Spectrograms

Lecture 1-10: Spectrograms Lecture 1-10: Spectrograms Overview 1. Spectra of dynamic signals: like many real world signals, speech changes in quality with time. But so far the only spectral analysis we have performed has assumed

More information

Airways Resistance and Airflow through the Tracheobronchial Tree

Airways Resistance and Airflow through the Tracheobronchial Tree Airways Resistance and Airflow through the Tracheobronchial Tree Lecturer: Sally Osborne, Ph.D. Department of Cellular & Physiological Sciences Email: sosborne@interchange.ubc.ca Useful links: www.sallyosborne.com

More information

Engineering with Sound Lesson Plan

Engineering with Sound Lesson Plan Type of lesson: Challenge Teaching Plan: Engineering with Sound Lesson Plan Goal The goal of this lesson is to introduce the students to sound and its properties and have them apply what they learn to

More information

Ultrasound - Vascular

Ultrasound - Vascular Scan for mobile link. Ultrasound - Vascular Vascular ultrasound uses sound waves to evaluate the body s circulatory system and help identify blockages and detect blood clots. A Doppler ultrasound study

More information

Indirect Blood Pressure Measurement

Indirect Blood Pressure Measurement P r o c e d u r e s P r o C A R D I O L O G Y Peer Reviewed Indirect Blood Pressure Measurement Maintenance of appropriate systemic arterial blood pressure is vital for survival. Because many common situations

More information

Computer Networks and Internets, 5e Chapter 6 Information Sources and Signals. Introduction

Computer Networks and Internets, 5e Chapter 6 Information Sources and Signals. Introduction Computer Networks and Internets, 5e Chapter 6 Information Sources and Signals Modified from the lecture slides of Lami Kaya (LKaya@ieee.org) for use CECS 474, Fall 2008. 2009 Pearson Education Inc., Upper

More information

Extracranial Cerebrovascular Duplex Ultrasound Evaluation

Extracranial Cerebrovascular Duplex Ultrasound Evaluation VASCULAR TECHNOLOGY PROFESSIONAL PERFORMANCE GUIDELINES Extracranial Cerebrovascular Duplex Ultrasound Evaluation This Guideline was prepared by the Professional Guidelines Subcommittee of the Society

More information

MD5-26 Stacking Blocks Pages 115 116

MD5-26 Stacking Blocks Pages 115 116 MD5-26 Stacking Blocks Pages 115 116 STANDARDS 5.MD.C.4 Goals Students will find the number of cubes in a rectangular stack and develop the formula length width height for the number of cubes in a stack.

More information

BASIC ELECTRONICS AC CIRCUIT ANALYSIS. December 2011

BASIC ELECTRONICS AC CIRCUIT ANALYSIS. December 2011 AM 5-202 BASIC ELECTRONICS AC CIRCUIT ANALYSIS December 2011 DISTRIBUTION RESTRICTION: Approved for Pubic Release. Distribution is unlimited. DEPARTMENT OF THE ARMY MILITARY AUXILIARY RADIO SYSTEM FORT

More information

Ultrasound in Vascular Surgery. Torbjørn Dahl

Ultrasound in Vascular Surgery. Torbjørn Dahl Ultrasound in Vascular Surgery Torbjørn Dahl 1 The field of vascular surgery Veins dilatation and obstruction (varicose veins and valve dysfunction) Arteries dilatation and narrowing (aneurysms and atherosclerosis)

More information

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Various Technics of Liquids and Solids Level Measurements. (Part 3) (Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,

More information