Ion Chromatography for PCB and PCA Contamination. Presented By: Foresite Inc.
|
|
- Ashley Walker
- 7 years ago
- Views:
Transcription
1 Ion Chromatography for PCB and PCA Contamination Presented By: Foresite Inc.
2 Objectives Theory of Operation Basic Plumbing Critical Parameters to Monitor Columns, Materials, and Reagents Lab Equipment Making Eluents Making Controls and Standards Equipment Start-Up Making Programs and Schedules In Chromeleon Executing Batch Files in Chromeleon Calibrating the Chromatograph Running Standards and Controls Checking for Linearity Optimizing Graphs Daily Start-up Schedules Daily Logs Sample Extractions (Standard K-Pak TSE) C3 Sample Extraction
3 Introduction to Ion Chromatography Characterizes and Quantifies specific residue species present on sample IPC condoned test method TM Foresite was instrumental in developing this testing methodology Two extraction methods: Standard bath, total board extraction Localized C3 spot extraction
4 What is Ion Chromatography? Qualifies and Quantifies specific process residues by separating ionic and organic compounds suspended in a liquid solution. Separation is achieved through a finely balanced system of a liquid phase element and a charged resin column. Each different species travels through the column (Time of Flight) at different rates, dependant upon the specie s mass and charge. Species exit the column separately and their conductivity is measured by a conductivity meter.
5 Theory of Operation Conductivity {µs} Time
6 Basic Plumbing Flow 1 Eluent (Mobile Phase Carries Sample) 2 Pump (Moves Eluent and Sample) 3 Pulse Dampener (Removes pump oscillation) 4 Sample Valve (Injects sample into Eluent stream) 5 Guard Column (Filter and 20% Separation) 6 Analytical Column (Performs Separation) 7 Suppressor (Removes Counter Species) 8 Conductivity Cell (Measures Designated Species)
7 Basic Plumbing Hardware Eluent inlet From Stock or E.G. Conductivity Cell Measures sample conductivity Pulse Dampener Reduces Pump Noise Sample Valve To Inject Sample Check Valve Prevents Back-flow Suppressor Removes Counter Ions Column Oven Heats Columns Sample Port For Manual Injection Pumps 2-Stage Flow Schematic
8 Column Selection (Why AS-22) AG-22 & AS-22 Carb / Bi-Carb Eluent is easy to make and control Column set is fairly Mature Water Dips do not interfere with eluent peaks SO4 Time of Flight is about 8-9 Min WOA Time of Flight is 15.5 min (Run time) WOA & SO 4 do not Co-Elute! Can Resolve Formate and Acetate without using a separate Tetra Borate Deca-Hydrate Eluent run. Down side - Column is NOT Alcohol Compatible >10%
9 Isocratic vs. Gradient Isocratic Run Steady-State Eluent Baseline is fixed Elution times are fixed Easier to Calibrate Gradient Run Can change Eluent Real time Baseline shifts with eluent Elution Times not as fixed More difficult to Calibrate
10 Typical Anions Many Manufacturers Are Concerned With Today on Electronic Hardware Dionex DX-120 system with a AS4A-SC 4mm Column Cl NO 2 Br NO 3 PO 4 2 WOA
11 Typical Anions manufacturers are concerned with in today s on electronic manufacturing Dionex ICS-2000 system with a AS-22 4mm Column Set
12 Typical Cations manufacturers are concerned with in today s on electronic manufacturing Dionex DX-120 system with a CS12A 4 mm Column Set
13 Critical Perimeters to Monitor Back Pressure PSI Baseline Conductivity µs Elution Times Min Peak Separation Touch Baseline? Peak Profiles Bell Curve Control ppm Values <5% Inaccuracy Column / System Leakage Leak Check Controls and Eluent Age 40F!
14 Lab Equipment The Basics Ion Chromatograph Glass Ware (Misc) Lab Scale (.0001g Resolution) Pipettes (Fixed Volume) Reagents (HPLC Grade) Mixing Vessels KaPak Bags (assorted sizes) Clean Aluminum Foil Verified Clean PVC Gloves Syringes (Verified Clean) Heated Water Bath D.I. Water Source Glass Ware (Volumetric Class A) Calibrated Weights Pipette Tips Weighing Boats Storage / Delivery Vessels Thermal Bag Sealer Lint Free Lab Wipes Chemical Gloves Syringe Filters Line Timer
15 Equipment Large Scale Lab Water System Incoming U.V. Reverse Osmosis Resistance Meter Final U.V. Pre Filter Ion Polish Mid Filter
16 Small Scale Reagent Lab Water System Bench Top System - Low output volume + High Quality Water + Minimal Investment + Small Foot Print + Low Maintenance + Feed in good water get out Very Good Water! Feeding a 18.2 house loop in will generate Ultra-Pure for I.C.
17 Making Carb / Bi-Carb Eluent Baseline conductivity is very important (16µS-18µS) Eluent Baseline conductivity will greatly effect peak times! Each Column Set is Different and will need it s own mix! Make Larger Eluent Volumes to Reduce Mixing Error! Only use Reagent Grade Materials or (HPLC Grade) 16 liters of Eluent Sodium Bi-Carbonate g Sodium Carbonate g Divide by 4 4 liters of Eluent
18 Good Water for I.C.! Good I.C. water has minimal background Ionics Water was polished with a Millipore A10 fed from 18.2 megω House low TOC Sample was shot from a cleaned Boro-silicate vial on a Dionex AS-50 autosampler
19 Calibration of I.C. Calibration and Chromeleon Training will be performed in the Lab
20 I.C. Calibration Basics Amount Table Species Good Linearity is in the 99.9 to 99.7 R 2 Range Linearity
21 Daily Start-Up! First!, Is the Calibration still Valid? First Run of the Day: Cation Control (Dionex) Anion Control (Dionex) Level 3 Cation Standard (4ppm) Level 3 SO 4 / WOA Acid Standard (4ppm) Water Blank / Syringe Blank Bag Blank (1hr in Water Bath) Anion & Cation Controls Every 10 Samples!
22 Documentation (Controls) Standards should be traceable to NIST or NBS
23 Documentation (Calibration) Standards should be traceable to NIST or NBS Controls should have unique documentation codes Pipettes should be verified before use.
24 Event Log (Your System s Diary) Document activity on your system to predict system and column failures
25 Extraction Protocols TSE (Total Standard Extraction) IPC-TM- 650 Standard Uses 75% IPA 25% D.I. (Unique volume for each PCA size) Heated Water Bath Dilution factor is calculated by Dividing Volume into Surface area. For Populated Boards add 10% (Volume of Solution / Total Board Area) Localized Extraction by C3 & Test per IPC-TM- 650 Standard Uses D.I. Water (Steam) Dilution factor for I.C. is based on Cell Aperture (.1sq ) Dilution factor is (2.2mL /.1sq = 22) For Populated Boards add 10% (2.2mL /.11sq = 20)
26 TSE (Total Standard Extraction) IPC Heated Water Bath with Window
27 Extraction Fluid Dispensing Dispenser Dispenser must be repeatable and able to be calibrated! 75% IPA 25% D.I. Water
28 TSE (Total Standard Extraction - FLOAT) Extraction solution is added to a Kapak bag to cover the board Surface. Kapak Bag PCA Platform
29 TSE (Total Standard Extraction) FLOAT Bag Meniscus Bath Meniscus
30 Total Solvent Extraction VS. Localized TSE Good for bare boards Bad if the solder mask is not cured! It will foul the column Not as good on populated boards. They can give anomalous reading due to assembly package contributions. The organics dissolve can foul the Anion columns. Column clean-up is time consuming!!!!! IPA is not compatible with all column sets Is IPC Standard Localized Good for bare boards or populated boards Can check via locations vs. solder mask areas for etch residuals Will not foul the columns
31 Localized Extraction with C3
32 C3 Process Monitoring Tool Purposes of C3: Monitoring tool for production floor Focus on sensitive area of concern (0.1 in 2 ) Performs electrical test and gives immediate clean or dirty reading based on Foresite recommended limits for ionic contamination Localized extraction method for Ion Chromatography Extracts sample from localized testing area using deionized steam Samples can be shipped to a lab for Ion Chromatography analysis
33 Choosing Sensitive Areas to Test with C3 Sensitive areas of circuitry that are prone to failure High impedance devices Low standoff components that could trap residues
34 Capacitor Investigation How Clean is this Spot? Background Chip Capacitors on the battery circuit were causing the 3 volt battery to drain in less than 2 weeks on a new hand held Glucose tester prior to this new product launch. A high humidity screen was developed to separate the good units from the bad until a root cause and corrective action plan could be developed. It was originally thought that the capacitors were cracked, but this turned out not to be true, and standard cleanliness testing showed that this No Clean assembly process was clean and not the problem. Until we tested the failing units in just the area of the capacitor. This showed a high level of sulfate residue on the capacitor. This sulfate was not present in high quantities on the units that passed the screening test.
35 Capacitors on Assemblies Ion Chromatography C3 all values are ug/in2 Cl Br SO4 WOA Test Time 0805 Cap Good Boards Catalyst Board #1-C3 Site # Pass Catalyst Board #2-C3 Site # Pass Catalyst Board #3-C3 Site # Pass Catalyst Board #4-C3 Site # Pass Catalyst Board #5-C3 Site # Pass Cap Suspect Boards Catalyst Board #1 C3 Site # Failure Catalyst Board #2 C3 Site # Failure Catalyst Board #3 C3 Site # Failure Catalyst Board #4 C3 Site # Failure Catalyst Board #5 C3 Site # Failure 0.24
36 Standard Extraction Results Ion Chromatography C3 all values are ug/in2 Cl Br SO4 WOA Test Time Total Board Good Boards Catalyst Board #1 Total Board N/A N/A Catalyst Board #2 Total Board N/A N/A Catalyst Board #3 Total Board N/A N/A Catalyst Board #4 Total Board N/A N/A Catalyst Board #5 Total Board N/A N/A Total Board Suspect Boards Catalyst Board #1 Total Board N/A N/A Catalyst Board #2 Total Board N/A N/A Catalyst Board #3 Total Board N/A N/A Catalyst Board #4 Total Board N/A N/A Catalyst Board #5 Total Board N/A N/A
37 Capacitor Investigation How Clean is this Spot? Conclusions The Capacitors were failing approximately 3-5% of the screening test. We found that the post plating rinses after the Barrel plating process using MSA (methane sulfonic acid) as the primary plating chemistry was not neutralized or uniformly rinsed The plating residue left on the capacitor surface created an invisible conductive pathway causing the failure of the off circuit due to this moisture absorbing conductive residue when sitting on the shelf in the box waiting for shipment.
Technical Spotlight. The Use of Ion Chromatography in the Printed Circuit Board Industry
The Use of Ion Chromatography in the Printed Circuit Board Industry Keith M Sellers Senior Scientist Trace Laboratories East 5 North Park Dr. Hunt Valley, MD 21030 (410) 584-9099 ksellers@tracelabs.com
More informationfor IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)
for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) IonSwift MAC-100 & MAC-200 Product Manual Page 1 of 16 PRODUCT MANUAL for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) MAC-100, 0.5 x 80mm (P/N 074702)
More informationACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
More informationDetermination of Calcium and Magnesium in Brine
Application Note Determination of Calcium and Magnesium in Brine INTRODUCTION To prevent membrane poisoning, new membrane technology in chlor-alkali cells requires feed brine that is relatively free of
More informationDetermination of Caffeine in Beverages HPLC-1
Determination of in s HPLC-1 Introduction This experiment provides an introduction to the application of High Performance Liquid Chromatography (HPLC) to the solution of complex analytical problems. Cola
More informationCONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
More informationCoating and Extraction of Honeycomb Denuders
QSM Approval: Coating and Extraction of Honeycomb Denuders 1 INTRODUCTION 1.1 The following procedures are used for the coating of honeycomb denuders with citric acid and sodium carbonate solutions, and
More informationResearch for leachables of nuclear grade cation exchange resin
International Journal of Smart Grid and Clean Energy Research for leachables of nuclear grade cation exchange resin Liang-Cheng Chen, Dah-Yu Kao, Tung-Jen Wen, Ping-Ji Huang Institute of Nuclear Energy
More informationDevelop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
More informationWastewater Reuse. Typical treated wastewater is:
Wastewater Reuse Most metal finishing industries have in-house wastewater treatment to economically dispose of the acids, alkali, oils, and dissolved metals in the rinse water and occasional tank solution
More informationEnvironmental Analysis Using the ICS-3000 Ion Chromatography System
Environmental Analysis Using the ICS- Ion Chromatography System Ion chromatography (IC) is well established as a routine method for the analysis of ionic analytes in environmental samples. IC has been
More informationAccurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography
Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography Carl Fisher and Linda Lopez Thermo Fisher Scientific, Sunnyvale,
More informationDissolved and precipitated oxalate
Accepted 2005 Process liquors from bleach plants Dissolved and precipitated oxalate Using Ion Chromatography 0 Introduction In bleach plants of pulp mills with a high degree of system closure, there is
More informationQuantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory
Customer Application Note: 114 Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Sébastien N. Ronkart, Ph.D.; Société wallonne des eaux, rue de la
More informationProtocol: HPLC (amino acids)
Page 1 of 6 1. Chemicals Composition M MW gram volume (ml) product code 0-pthaldialdehyde C 8H 6O 2 134.13 Sigma P0657 Perchloric acid HClO 4 3.5 100.46 Dilute 70% stock solution 1:1 with demi to obtain
More informationProject 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids
Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C
More informationStandard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Grace Analytical Lab 536 South Clark Street 10th Floor Chicago, IL 60605 April 15, 1994 Revision 2 Standard Operating Procedure
More informationHalogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.
Halogen Free: What, Why and How Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. We will discuss What are Halogens and how we interact with them. Why Halogens, within consumer
More informationDionex IonPac AS19-4µm
Total Thermo Scientific Dionex IonPac AS9-4µm Column Product Manual P/N: September 203 Part of Thermo Fisher Scientific Product Manual for Dionex IonPac AS9-4µm Column Dionex IonPac AG9-4µm Capillary Guard
More informationSUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
More informationGuide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
More informationThe Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection
Application Note 9 The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection INTRODUCTION The accurate measurement of the amount of sugar in final molasses
More informationExtraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
More informationWaters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
More informationfor SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns
for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns Product Manual for SolEx TM HRP Cartridges and SolEx TM HRP RSLC Columns Page 1 of 15 Product Manual for SolEx TM HRP Cartridges SolEx HRP (P/N 074400)
More informationDetermination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*
Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU
More informationEnvironmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
More informationUsing the AS Automated Sampler in the Simultaneous, Sequential, and Concentrate Modes
Technical Note 64 Using the AS Automated Sampler in the Simultaneous, Sequential, and Concentrate Modes INTRODUCTION The AS Autosampler is enhanced with new features to speed throughput and simplify sample
More informationA meaningful, cost-effective solution for polishing reverse osmosis permeate
A meaningful, cost-effective solution for polishing reverse osmosis permeate Electrodeionization or EDI, is a continuous and chemicalfree process of removing ionized and ionizable species from the feed
More informationQUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd
QUANTITATIVE AMINO ACID ANALYSIS Aurélie Lolia Applications Manager, Biochrom Ltd QUANTITATIVE AMINO ACID ANALYSIS Principles of amino acid analysis Ion exchange chromatography The Biochrom 30 physiological
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationcolumns IonPac CS18 Cation-Exchange Column
columns IonPac CS Cation-Exchange Column The IonPac CS is a cation-exchange column of moderate capacity and hydrophobicity, recommended for isocratic and gradient separations of polar amines such as alkanolamines,
More informationApplication Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
More informationHS 1003 Part 2 HS 1003 Heavy Metals Test
HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:
More informationTANNIC ACID. SYNONYMS Tannins (food grade), gallotannic acid, INS No. 181 DEFINITION DESCRIPTION
TANNIC ACID Prepared at the 39th JECFA (1992), published in FNP Add 1 (1992) superseding specifications prepared at the 35th JECFA (1989), published in FNP 49 (1990) and in FNP 52 (1992). Metals and arsenic
More informationHow To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
More informationAutomation of Sample Preparation in Ion Chromatography
Automation of Sample Preparation in Ion Chromatography Praha, nd October 04 Ing. Roman Repáš ACO Sales Manager (IC and SP) Agenda Thermo Scientific Dionex Sample Preparation Products Automation of Sample
More informationIon Exchange Design Hand calculation. Brian Windsor (Purolite International Ltd)
Ion Exchange Design Hand calculation Brian Windsor (Purolite International Ltd) Introduction Before design programmes were introduced, every engineer had to calculate the design by hand using resin manufacturers
More informationANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of carbohydrates in foodstuff (LC/RID method) Exercise guarantor: Assoc. Prof. Karel Cejpek, Ph.D. CONTENT Required knowledge...
More informationLead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
More informationDimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
More informationElectrodeionization (EDI)
Electrodeionization (EDI) Fact Sheet Cal Water Industrial Water Purification (800) CAL-WATER EDI is a continuous electro-chemical process of water purification where ion specific membranes, mixed bed resin,
More informationAnalytical Test Report
Analytical Test Report Customer: Address (City, State): Purchase Order: Report Number: Project Number: Date Received: Date of Report: Test Location: Boulder, CO. Assay: Part Number: Amino Acids by HPLC
More informationQ&A. Contract Manufacturing Q&A. Q&A for those involved in Contract Manufacturing using Nelco Electronic Materials
Q&A Q&A for those involved in Contract Manufacturing using Nelco Electronic Materials 1. Do Nelco laminates have any discoloration effects or staining issues after multiple high temperature exposures?
More informationPurification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
More informationDionex ICS-1100 Ion Chromatography System
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Dionex ICS-1100 Ion Chromatography
More informationHigh-Throughput 3-D Chromatography Through Ion Exchange SPE
High-Throughput 3-D Chromatography Through Ion Exchange SPE Application Note 205 Luke Roenneburg and Alan Hamstra (Gilson, Inc.) Introduction 2-dimensional (2-D) separation is the separation of a sample
More informationThe Relationship between ph and Deionized Water
The Relationship between ph and Deionized Water The basics of ph The topic of ph and water has been well documented over the years; however, there is still much confusion about its significance in high
More informationSeparation by Solvent Extraction
Experiment 3 Separation by Solvent Extraction Objectives To separate a mixture consisting of a carboxylic acid and a neutral compound by using solvent extraction techniques. Introduction Frequently, organic
More informationFRACTION COLLECTOR. Microcomputer Controlled CHF122SC
FRACTION COLLECTOR Microcomputer Controlled CHF122SC From open column to HPLC liquid chromatography and general liquid samplings, the Super Fraction Collector, model CHF122SC has a wide range of uses.
More informationPesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationA Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
More informationAutomation of Solid Phase Extraction and Column Chromatographic Cleanup
Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for
More informationReversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
More informationESS Method 130.1: General Auto Analyzer Procedures
ESS Method 130.1: General Auto Analyzer Procedures Environmental Sciences Section Inorganic Chemistry Unit Wisconsin State Lab of Hygiene 465 Henry Mall Madison, WI 53706 Revised October 1992 ESS Method
More information[ Care and Use Manual ]
The Waters GPC Cleanup System uses two Envirogel GPC Cleanup Columns to perform routine cleanup of environmental samples as specified by EPA Method 3640A, GPC Cleanup. The columns, 19 x 150 mm and 19 x
More informationThermo Scientific Dionex AS-AP, AS-DV, and AS-HV Autosamplers. Automation, sample preparation. Autosamplers for Ion Chromatography
Thermo Scientific Dionex AS-AP, AS-DV, and AS-HV Autosamplers Automation, sample preparation Autosamplers for Ion Chromatography Dionex AS-AP Autosampler Advanced Features Perform full-loop, partial-loop,
More informationEUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:
More informationHPLC Basics, Equipment, and Troubleshooting
HPLC Basics, Equipment, and Troubleshooting HPLC Basics, Equipment, and Troubleshooting explains chromatography in practical terms from the ground up HPLC Basics, Equipment, and Troubleshooting HPLC Basics,
More informationTechnical Info Sheet. Ionic contamination testing in a no-clean soldering process. Scope
Scope This document provides information about ionic contamination testing and its relevance in a no-clean. Also so-called C3 testing and ROSE (Resistivity of Solvent Extract) testing are forms of ionic
More informationSIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
CHAPTER 5 SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 Simultaneous determination of Telmisartan
More informationAmino Acid Analyzer L-8900
Amino Acid Analyzer L-8900 VWR - Hitachi Your Partner in Amino Acid Analysis Hitachi has manufactured more than 1800 Amino Acid Analysers for over 40 years. The new Amino Acid Analyser Model L-8900 is
More information1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
More informationLUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)
October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for
More informationSimultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC
Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC Mackenzie Ree and Erik Stoa Department of Chemistry, Concordia College, 901 8 th St S,
More informationKINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)
Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:
More informationStandard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material
PM 2.5 Nylon Filter Cleaning Date: August 25,2009 Page I of8 Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material Environmental and Industrial Sciences Division
More informationProperties of Acids and Bases
Lab 22 Properties of Acids and Bases TN Standard 4.2: The student will investigate the characteristics of acids and bases. Have you ever brushed your teeth and then drank a glass of orange juice? What
More informationTroubleshooting Your RO
Troubleshooting Your RO Summary: There can be many reasons why a RO system suffers a loss in performance, and is unable to produce the proper quantity and/or quality of permeate water. Similar to a doctor
More informationElectrical Conductivity of Aqueous Solutions
Electrical Conductivity of Aqueous Solutions PRE-LAB ASSIGNMENT: Reading: Chapter 4.-4.3 in Brown, LeMay, Bursten & Murphy.. Using Table in this handout, determine which solution has a higher conductivity,.
More informationIon Exchange Determination of Na+ by Displacement and Zn 2+ Using Preconcentration. Reading: Harris pp. 641-647, 699
Ion Exchange Determination of a+ by Displacement and Zn 2+ Using Preconcentration Reading: Harris pp. 641647, 699 Prelab Assignment: 1. The ph of a solution is 4.56. Determine the concentration and use
More informationCORESTA RECOMMENDED METHOD Nº 80
CORESTA RECOMMENDED METHOD Nº 80 USE OF THE PART-FILTER METHOD FOR THE ESTIMATION OF SMOKERS EXPOSURE TO NICOTINE AND NICOTINE-FREE DRY PARTICULATE MATTER (January 2016) 0. INTRODUCTION The CORESTA Smoking
More informationEXPERIMENT # 3 ELECTROLYTES AND NON-ELECTROLYTES
EXPERIMENT # 3 ELECTROLYTES AND NON-ELECTROLYTES Purpose: 1. To investigate the phenomenon of solution conductance. 2. To distinguish between compounds that form conducting solutions and compounds that
More informationAnion chromatography using on-line recycled eluents
Journal of Chromatography A, 1089 (2005) 82 86 Anion chromatography using on-line recycled eluents Takashi Yokoyama, Hitoe Maekubo, Asami Sakai, Michio Zenki Department of Chemistry, Faculty of Science,
More informationWith proper calibration, the upper limit may be extended as needed.
While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the
More informationGas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
More informationColorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
More informationEPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.
EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer HT3 Application Note Introduction The analysis of dissolved gases in ground water is important in
More informationHPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
More informationDetermining the Identity of an Unknown Weak Acid
Purpose The purpose of this experiment is to observe and measure a weak acid neutralization and determine the identity of an unknown acid by titration. Introduction The purpose of this exercise is to identify
More informationCleanliness of Electronic PCB Assemblies Leads to Medical Device Reliability
Cleanliness of Electronic PCB Assemblies Leads to Medical Device Reliability by Mo Ohady and David Estes, Digicom Electronics, Oakland, CA Overcoming the Potential for Failures Component manufacturers
More informationSimultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)
Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction
More informationDetermining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
More informationStandard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2
Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall
More informationHigh sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK
High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen
More informationOptimize Your Process Operations by Improving Process Monitoring
Integral Process Ana Optimize Your Process Operations by Improving Process Monitoring All industrial processes have a common need reliable, accurate, and timely methods for analyzing process chemistry.
More informationCleaning and Cleanliness In IPC Specifications
Cleaning and Cleanliness In IPC Specifications Doug Pauls and Dave Hillman Principal Materials and Process Engineers Rockwell Collins 1 Keynotes: Marketing Vs. Reality What Was Pitched The Grim Reality
More informationDetermination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
More informationHigh Purity Water Resistivity/ Conductivity Measurement
CONDUCTIVITY MEASUREMENT IN HIGH PURITY WATER SAMPLES below 10 µsiemens/cm High Purity Water Resistivity/ Conductivity Measurement Ultra-Pure Water without any chemical impurities will still have a conductivity
More informationCOMMON LABORATORY APPARATUS
COMMON LABORATORY APPARATUS Beakers are useful as a reaction container or to hold liquid or solid samples. They are also used to catch liquids from titrations and filtrates from filtering operations. Bunsen
More informationVitamin C quantification using reversed-phase ion-pairing HPLC
Vitamin C quantification using reversed-phase ion-pairing HPLC Thomas Grindberg and Kristy Williams Department of Chemistry, Concordia College, 901 8 th St S, Moorhead, MN 56562 Abstract Vitamin C, an
More informationThe Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationPPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques
PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques Introduction The amount of dissolved oxygen in process water is continually gaining importance in many industries as a critical parameter
More informationAuditing Contract Manufacturing Processes
Auditing Contract Manufacturing Processes Greg Caswell and Cheryl Tulkoff Introduction DfR has investigated multiple situations where an OEM is experiencing quality issues. In some cases, the problem occurs
More informationThe Theory of HPLC. Gradient HPLC
The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims
More information