Construction of a Negative Ion Mode Linear Time of Flight Mass Spectrometer
|
|
- Edgar Nelson
- 7 years ago
- Views:
Transcription
1 Construction of a Negative Ion Mode Linear Time of Flight Mass Spectrometer Tyler Vanover April 30, 2013 Abstract The aim of this project was to construct a linear time of flight mass spectrometer in negative ion mode. In order to construct a negative ion mode linear TOF it becomes a necessary endeavour to learn about the inner workings of the TOF including the exploration of the source, flight tube, detector and electronics. One of the main issues combating high resolution of the linear TOF is the temporal, spatial and kinetic energy distributions of ions in the source region. The way to remedy this is to employ the technique of delayed pulse extraction, which I explored in detail. Because of a malfunction with an integral part of the delayed pulse extraction technique, I was not able to employ this method to obtain data. However, I did get some response from the detector while operating the TOF in continuous positive ion extraction mode, which proves that ions are being created and that the detector is working. 1 Introduction 1.1 The Linear Time of Flight The Time of Flight mass spectrometer (TOF) is an instrument that is capable of measuring the ratio of mass-to-charge m z of ions, atoms, or molecules (Lu). Its operation is based on the idea that ions with the same kinetic energy 1 2 mv2, but different m z ratios can be separated by their flight time over a given field free path (figure 1) (Fjeldsted). Due to the incredible resolution of the TOF spectrometer that is available today, and by utilizing the various isotopic mass differences of ions, it is possible to determine not only the m z ratio, but also the charge and the mass separately for those ions. Two advantages of using the TOF machine are its tremendously fast data acquisition rates and the ability of detecting trace amounts of ions (Fjeldsted). One attractive feature of the TOF machine is its conceptual simplicity. It represents an excellent application of sophomore physics. Ions which are generated at the source region by a nitrogen laser, for example, are accelerated to an energy q V ( V = P 2 P 1 ) by the electric field E s = V d s, where Es is the electric field of the source, V 2 is the voltage of the field free region, V 1 is the voltage of the repeller plate in the 1
2 Figure 1: Schematic of Linear Time of Flight Mass Spectrometer source, ds is the length of the source region, and L is the length of the flight tube as seen in figure 1. Although the time of flight mass spectrometer is conceptually simple, in practice, it can be quite complicated. The generated ions are spread over a finite volume in space and they do not start with zero kinetic energy. Both of these features reduce the mass resolution. Fortunately, the introduction of what is called the delayed pulse extraction technique corrects for both features and allows all ions to enter the field free region with the same kinetic energy. The resolution m m of the linear TOF is of the order (Guilhaus). Once the ions are in the field free region, they all travel down the tube with the same initial kinetic energy, which is given by The ion velocity, v, can be written as 1 2 mv2 = q V (1) v = 2q V The time of flight, t, of an ion of mass m along the flight tube L is t = L v = m L 2 V m q (2) (3) where t=drift time L=Drift length of field free region m=mass of ion 2
3 Figure 2: Flow Chart of Basic Components q=charge of ion V=Potential difference of plates in the source region 1.2 Basic Components of a Negative Ion Mode Linear Time of Flight Conceptually, the negative ion linear TOF consists of 4 major parts as outlined in figure 2. The high vacuum system is necessary because we do not want the ions to be bumping into other particles such that they can float freely down the flight tube with as little interference as possible. To acquire such a vacuum we have used two turbomolecular pumps in combination with two common laboratory vacuum pumps. Our system has capabilities on the order of 10 8 torr. The ion source plays a major role in the operation of the TOF as this is where the probe tip is placed and ions are created using pulsed laser extraction. The ions are then accelerated through a potential difference and thrust into the mass analyzer. The mass analyzer is simply an elongated tube that has been evacuated where the ions are given time to separate such that the heavier ions reach the detector last and the lighter ions first. There are two main types of analyzers: reflectron and linear. The reflectron offers superior resolution, but is much harder to construct. Our mass analyzer is of the linear type. The detection system is coupled with other electronics responsible for the pulsed laser extraction such that it knows when to detect an incoming ion packet. This is then fed to a data analysis system where a graph of the intensity of the counts is plotted against the mass to charge ratio. Now lets take a look at each component in detail. 3
4 Figure 3: Source Region Figure 4: Schematic of Source Region 2 Source Area The source area is where the ions are generated and given their initial kinetic energies where they are then sent into the flight tube of specified length and lastly received at the detector. Since this is where the ions are generated and given their initial energies, this results in the area where the main source of error in the resolution can be attributed. As ions are created in the source region they will have time, space and kinetic energy distributions that cause error in the mass resolution (Opsal). Figure 3 is our source which is represented as a schematic in figure 4. In figure 3 P 1 represents the repeller voltage plate and P 2 represents the acceleration grid. Note the wires coming from the source flange that carry the current from the power supply. In figure 3 you get a better feel for what the source plates look like as they are stacked on top of one another. The probe tip rests in the middle of the repelling plate and is bombarded by the laser source that is incident at some angle. Since we are interested in constructing a linear TOF in negative ion mode the repelling voltage should be negative while the accelerating grid should have a positive voltage applied to it and the 4
5 Figure 5: Kinetic Energy Distribution distance between the plates as noted in figure 1, d s is the accelerating region before the ions are hurled into the flight tube. As I mentioned before, these ions will have temporal, spatial and kinetic energy distribution that interfere with the mass resolution. 2.1 Temporal Distribution As you may have guessed this includes the distributions in the ion time of formation. It also includes limitations of the ion-detection and time-recording devices, but that isnt so much our concern here as we cannot control that aspect. Essentially what this means is that the distributions of ion formation results in ions which enter the drift tube at different times and maintain a constant time difference as they approach the detector (Opsal). Since the mass resolution is given as t 2 t these effects can be minimized by using longer flight tubes. 2.2 Kinetic Energy Distribution From the equations in the introduction you can see that the flight time of the ion is dependent on the kinetic energy of the ion as it leaves the acceleration region which is given by t = m 2KE L where the KE now has an error term of U 0 such that the new KE is given to be KE = q V + U 0 where U 0 refers to the initial kinetic energy from the ions initial components along the flight axis (Opsal). As depicted in figure 5, the ions with energies of q V +U 0 that have the same mass arrive at the detector sooner than ions with an energy of q V.This results in the mass spectrum being skewed to the low-mass side, hence a lower resolution (opsal). 5
6 Figure 6: Delayed Pulse Extraction 2.3 Spatial Distribution The spatial distribution of the ions is really an effect on the initial kinetic energies of the ions. When ions are formed at different locations in the source, it means that the accelerating region will be shorter or longer for the particular ion which results in a smaller or bigger initial kinetic energy as the ion enters the field free region. As these ions enter the field free flight tube there is a region known as the space focus plane where the faster ions that are formed near the rear of the source catch up with the slower ions that are formed near the front of the source, as you can see in figure 5. It turns out that this point is independent of the mass and is located at a distance of 2d s from the grid where the ions exit the accelerating region. There is a way in which we can essentially push this focal plane back to the detector such that the mass resolution is dramatically increased. This method is called delayed pulse extraction. 2.4 Delayed Pulse Extraction As I mentioned above, we want to move the space focus plane back to the detector where the ions are going to be detected to correct for imperfections found in the spatial, KE, and temporal distributions. The technique of delayed pulse extraction will be employed to do so. Figure 6 gives a representation 6
7 of how the delayed pulse extraction works which is not so transparent at first glance. During step 1 one there is no field applied between the two source plates that we saw in figure 6. The laser is fired at the source tip and ions emerge from the sample in what is known as a plume very shortly after being hit with the laser. The ions that you see in step 1 all have the same mass but have different velocities. The result of having these different velocities means that they will have different locations in the source area labeled ds in figure 1 when the electric field is turned on (Matrix Assisted...). The electric field in the accelerating region gets weaker as the ion gets closer to the grid. As a result the ions with the higher velocities will not feel as much of a push toward the grid as the ions with slower velocities. In the end the ions reach the grid with the same velocities hence the same kinetic energy, which plays a key role in the resolution. 3 Flight Tube The flight tube is one of the most simple components of the TOF instrument, at least for the linear TOF. The flight tube is the region of the mass spectrometer that immediately follows the source region. The flight tube is the cavity on the mass spectrometer that is considered a field free region and the point where the initial kinetic energies of all the ions entering the tube should be the same. The ions then traverse the evacuated tube and are received at the detector depending on the mass to charge ratio. Most flight tubes have horizontal and vertical deflection plates that range in voltage and are used to direct the ions onto the detector (Watson). 4 Detection 4.1 The Microchannel Plate (MCP) The components that make up the detector are the mesh grid and the Micro Channel Plate detector (MCP). The mesh grid is simply a steering mechanism where the ions can be directed onto the MCP. The MCP is a very thin ( 2mm) highly resistive material that contains an array of tiny tubes (microchannels) that are densely distributed across the plate(fjeldsted). One of the key characteristics of the tubes is that they are not perpendicular with respect to the surface of the plate, but are rather angled at some angle ( 8 o ) (Fjeldsted). As an ion enters one of the channels, the angle guarantees that the ion will strike the wall of the channel and this impact releases electrons from the walls of the channel that then perpetuate themselves as they propagate down the channel. The purpose of this is to amplify the signal. When the electrons exit the MCP they are then digitized an output to a device such as an oscilloscope (Fjeldsted). Figure 7 is a photo of the detector and figure 8 is a schematic of how it works. 7
8 Figure 7: Detector Figure 8: Microchannel Plate Schematic 8
9 5 Resolution and Calibration From the equations at the top in the introduction we can see that the mass is a function of the flight time, m(t) (Fjeldsted). With that in mind, the user will always try to keep the voltages applied to the plates, the distance between the plates, and the length of the flight tube constant such that equation 3 can be rewritten as m = At 2 (4) In practice, there is a delay from the time the electronics sends a starting pulse to time that a high voltage is present on the rear ion pulser plate. Unlike delayed pulse extraction, this delay was not intentional. There is also a delay in the time an ion reaches the front surface of the ion detector until the signal is generated that is digitized by the acquisition system (Fjeldsted). Even though these are short delays they are significant and must be accounted for. Because it is not possible to measure the true TOF, we must correct the measured time, t m, by subtracting the sum of the start and stop delay times which will be denoted as t 0 t = t m t 0 (5) Substituting equation into equation gives: m = A(t m t 0 ) 2 (6) As a result, if we want to find the mass of the ion, we have to have values for A and t 0 and to do this a calibration is performed. 5.1 Mass Calibration A solution of compounds whose masses are known with great precision are analyzed and a table is made with the calibrant masses and their respective flight times, t m (Fjeldsted). Here is an example: Calibrant Mass Flight Time Now that we have m and t m for a variety values across the mass range, we can have a computer to use nonlinear regression to find the values of A and t 0 such that the mass can be as close as possible to the real value for all of the mass values in the calibration (Fjeldsted). 9
10 5.2 Data Acquisition The signal that is output by the MCP can be fed to an oscilloscope where intensity (counts) is plotted against the mass-to-charge ratio. For most analyses, the tradeoff is made to widen the range of detected TOFs at the expense of slower data sampling rates. The data can be recorded from the oscilloscope and transferred to a pc. The data is converted to ASCII format and then imported into GRAMS spectral analysis software (Watson). 6 Getting Started The first thing I did for my senior research project was get my hands dirty. I took the entire TOF apart, took an inventory of what we had and attempted to determine the purpose of each component of the TOF. I started by learning how to bring the TOF back to atmospheric pressure (Lab Notebook 47). I then began looking to each component of the TOF starting with the source and detector as I was able to examine those prior to my proposal. I then began to examine the deflection region in the flight tube. 7 The Deflection Region The deflection region is rather important as it is what extracts the ions from the source and collimates them. I took the deflection region off of the TOF and checked for continuity to ensure everything was connected properly. I then documented what each pin on the flange corresponded to (Lab Notebook 56). While working with the deflection region I was also testing the power supplies making sure that each output registered what it was suppose to (Lab Notebook 53). 8 Source Setback After getting a good feel for all of the components on the TOF, I began to reassemble it. Before I put the system back under vacuum, I noticed that we had a problem with the source region. The problem was that when the loading rod was inserted into the source region it was not aligned so that the probe tip would rest in the middle of the repelling plate. The interlock is a series of precisely aligned holes that the loading rod must go through in order to place the probe tip in the center of the repelling plate. If you look back to figure 3 you will notice that the plates P 1 and P 2 are held in place using insulating ceramic rods, which are screwed into the source flange. The ceramic rods showed some signs of damage, which had presumably lead to the plates becoming askew. To remedy this, I realigned the source with a thin washer which pushed the repelling plate in the desired direction just enough such that the probe tip could rest inside the repelling plate. This raised concerns between Dr. Lahamer and I because the 10
11 Figure 9: Original Bench Setup ions must take a very precise path through this deflection region and this could be a source of error in the end. Once this was taken care of I reassembled the system completely and put it back under vacuum. I then began searching for an appropriate laser and setting up the optics for the bench. 9 Nitrogen Laser and Bench Setup The nitrogen laser that I found in the laboratory that was not already part of a system, has a wavelength of 337 nm and a pulse energy of 300µJ. The reason that we are using that particular wavelength is because that is what has been shown to be a good wavelength for producing negative ions, according to Dr. Lahamer. The next order of business was to set up the bench where the laser and optics would remain. Figure 11 gives you an idea of what the original bench setup looked like. The bench setup is extremely important in order for the delayed pulse extraction to perform properly. For the setup you see in figure 10 the laser had its own internal trigger. When the laser fired its beam would reduce aberrations by passing through the iris, it would then pass (mostly) through the reflecting lens, be reflected by the mirror and focused onto the sample tip by the focusing lens. This results in the ionization of the sample material (hopefully). When light passes through the reflecting lens a small portion of the incident light is reflected and that reflected light triggers the photo-detector, which triggers other electronics needed for the delayed pulse extraction. The important thing to take away from this is that the photo-detector determines the period of the laser pulse and this is needed because we want to extract ions from the sample exactly out of phase with the pulse period of the laser. I had a problem getting the photo-detector to trigger the DG535 pulse/delay generator. For this setup it is important for the DG535 to be triggered because it creates a delay in the 11
12 Figure 10: Oscilloscope Output using HeNe Laser pulse received from the photo-detector. This pulse is then used to trigger the the high voltage supply which is applied to the repelling plate for the extraction of the ions. If you look back to figure 1 the delay in that diagram should occur when the high voltage is applied to the repelling plate. Eventually, I discovered that the problem was not the photo-detector, but rather the laser. I was able to come to this conclusion by using a Helium Neon laser and chopper while observing the output on the oscilloscope. What you see in figure 11 is a good representation of what I should have been seeing using the nitrogen laser. The output from the photo-detector is the result from the incident HeNe laser and the output from the DG535 is exactly out of phase with the pulse from the HeNe laser. Upon seeing this, Dr. Lahamer and I agreed that the photo-detector was not receiving the proper intensity from the nitrogen laser because the signal from the photo-detector was very faint and had no width. Since the light that was reflected from the the reflecting lens was used to trigger the photo-detector, it makes sense that the beam did not have the intensity needed to establish some pulse width. For the image you see in figure 11, I did not use a reflecting lens to obtain that, but rather had the laser directly incident on the photo-detector. The clarity of the image in figure 11 seems to be good evidence that the photodetector was not receiving the appropriate intensity. I could no longer use the setup seen in figure 10 because it relies on the reflecting light from the reflecting lens to trigger the photo-detector, so I had to design a new system. 12
13 Figure 11: Electronics for Delayed Pulse Extraction 10 Improved Bench Setup and Electronics After using the HeNe laser I learned that there was a way that I could trigger the laser with a pulse generator and use the photo-detector to determine the pulse width of the laser. I could then use the DG535 to incorporate a delay that is comprable to the pulse width, such that the high voltage would be applied during the instances the laser is not incident on the sample. Figure 13 gives a schematic of what the electronics for my new system looks like. It should be noted that the bench setup didn t change much. I just removed the photodetector and reflecting lens. It should also be noted that you need to determine the pulse width each time you change the repetition rate. To determine the pulse width of the laser you must use the photo-detector. By using the photodetector you can set the delay to a time similar to that value you found from the photo-detector. So we still need to use the photo-detector for this setup. However, it is only needed when you begin the analysis and it can be set such that the laser is directly incident on the photo-detector. If you decide to change the repetition rate of the laser, you must use the photo-detector again to find the pulse width and set the DG535 delay again. However, it is unlikely that the user will be doing this often. This was one of the biggest accomplishments of my senior project. Once I had overcome this barrier, I was once again hindered; 13
14 The PVX pulse generator was malfunctioning. 11 The PVX Pulse Generator and Remote Pulser Up to this point I had the pulsing electronics working and everything seemed to be properly synced. As you noticed in figure 12, the output from the DG535 goes to the PVX pulse generator. The output from the DG535 is a 5 Volt square wave that triggers the PVX. When the PVX is triggered it outputs a waveform that is exactly like the incoming waveform except the amplitude is much larger. In this case the amplitude would have been on the order of 10 3 Volts. However, the PVX seemed to have an issue with what the manufacturer calls an overcurrent. After making several calls to the manufacturer, they deemed it necessary to send the PVX in for repairs. I was devestated. It seemed that all hope had been lost because we had no way of producing the high voltage waveform necessary to employ the technique of delayed pulse extraction. Dr. Lahamer found a device manufactured by Jordan called a Remote Pulser that worked essentially the same as the PVX. However, after searching online for a product description, I found that the remote pulser is part of a package sold by Jordan and that it was suppose to be used alongside a specific power supply also manufactured by Jordan (Jordan). After speaking with Dan and the manufacturer I discovered that it would be possible to use the remote pulser with our power supplies, but that did not end successfully. In the end, I scrapped the idea of using the remote pulser and decided to try to get a signal using continuous extraction. 12 Continuous Extraction For the continuous extraction the setup becomes much less complex. If you take out the DG535 and the PVX from figure 11 you have the schematic for continuous extraction. However, for continuous extraction, the high voltage power supply is connected directly to the repelling plate and the electronics in figure 11, with the exception of the DG535 and PVX, simply determine the periodic pulse of the laser and simultaneously triggers the oscilloscope. When I employed this method I was basically looking for some kind of signal to let me know that the detector is working, regardless of the eminent poor resolution. Also, while operating the TOF in this mode I ionized some C 60 that Dr. Lahamer had already had in the laboratory in hopes of seeing something on the oscilloscope. I tried both positive and negative ion mode for the continuous extraction, but only got one reading for the positive ion mode. If you look to figure 13 you can see the single peak that I got. This peak was the average of 500 sweeps for 500 points, as you can see by looking at the oscilloscope, so it most certainly is not a fluke. While I know that this may seem relatively unimpressive, it does tell us that ions are being created and that the detector seems to be in working condition, both of which is good news because at the beginning of the semester nobody had any idea what functioned on the linear TOF. Even though we have 14
15 Figure 12: Peak from Positive Ion Continuous Extraction no way of proving it, Dr. Lahamer thinks that the place where the peak is located is generally the area where he has found C 60 in the past. 13 Conclusion As the semester comes to a close, I realize that I have accomplished a lot. I have learned how the linear time-of-flight mass spectrometer works and applied that knowledge to our instrument in hopes of resurrecting it. In doing so, I designed my own method for employing the technique of delayed pulse extraction, which was unfortunately hindered by the malfunction of the PVX pulse generator. I also verified that were were in fact creating ions and that the detector is in working condition. In the future when the PVX has been repaired, it would be great if somebody could continue my work and see if they could get some data from the spectrometer, as the PVX seems to be the only obstace in the way of being able to employ the technique of delayed pulse extraction. 14 Acknowledgments First, I would like to thank Dr. Amer Lahamer for his assistance throughout the semester. Your prior knowledge and guidance has helped immensely over the semester. I would also like to thank Dan Brewer who helped me numerous times with anything and everyhting that I asked of him. I learned a lot from Dan over the course of the semester, mostly in the way of electronics. Many times Dan also helped me locate documents online that helped deepen my sense of 15
16 knowledge and inquiry involving the TOF. I would like to thank Dr. Hodge and Dr. Veillette for the knowledge that they shared through class and conversation. I would also like to thank my friends and colleagues for the support and advice. Lastly, I would like to thank Franceska for all the support and encouragement. 16
17 References [1] Fjeldsted, John. Time-of-Flight Mass Spectrometry: Technical Overview. Review. Agilent Technologies Dec. 2003: Agilent Technologies. Web. technicaloverviews/public/ en11-dec-2003.pdf. [2] Opsal, Richard B., Kevin G. Owens, and James P. Reilly. Resolution in the Linear Time-of-flight Mass Spectrometer. Analytical Chemistry 57.9 (1985): Print. [3] Guilhaus, Michael. Special Feature: Tutorial. Principles and Instrumentation in Time-of-flight Mass Spectrometry. Physical and Instrumental Concepts. Journal of Mass Spectrometry (1995): Print. [4] Watson, R. Craig. Laser-ionization Time-of-flight Mass Spectrometry of High Molecular Mass Inorganic Complexes. Thesis. Virginia Polytechnic Institute and State University, N.p.: n.p., n.d. Print. [5] Matrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry. Swiss Federal Institute of Technology Zurich. N.p., n.d. Web. 20 Jan maldi/tutorial1.pdf. [6] Lu, Cheng. Introduction to Mass Spectrometry. Bioinformatics and Medical Informatics. N.p., n.d. Web. 20 Jan May07_projects/IntroductiontoMassSpectrometry.pdf. [7] Jordan, R. M. JordanTOF Products, Inc. Electron Guns and High Voltage Pulsers from Jordan TOF Products, Inc. Jordan TOF Products, Inc., n.d. Web. 29 Apr
TOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] info@rmjordan.com [e-mail] This
More informationExperiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
More informationNuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass
Protons, neutrons and electrons Nuclear Structure particle relative charge relative mass proton 1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass Protons and neutrons make up
More informationCathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
More informationLab 9: The Acousto-Optic Effect
Lab 9: The Acousto-Optic Effect Incoming Laser Beam Travelling Acoustic Wave (longitudinal wave) O A 1st order diffracted laser beam A 1 Introduction qb d O 2qb rarefractions compressions Refer to Appendix
More informationTime-of-Flight Mass Spectrometry
Time-of-Flight Mass Spectrometry Technical Overview Introduction Time-of-flight mass spectrometry (TOF MS) was developed in the late 1940 s, but until the 1990 s its popularity was limited. Recent improvements
More informationPUMPED Nd:YAG LASER. Last Revision: August 21, 2007
PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow
More informationMass Spectrometry. Overview
Mass Spectrometry Overview Mass Spectrometry is an analytic technique that utilizes the degree of deflection of charged particles by a magnetic field to find the relative masses of molecular ions and fragments.2
More informationEXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS
1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides
More informationModule 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
More informationFTIR Instrumentation
FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation
More information1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002
05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical
More information13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
More informationUltrasonic Wave Propagation Review
Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic
More informationCoating Technology: Evaporation Vs Sputtering
Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information
More informationELECTRON SPIN RESONANCE Last Revised: July 2007
QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron
More informationInterference. Physics 102 Workshop #3. General Instructions
Interference Physics 102 Workshop #3 Name: Lab Partner(s): Instructor: Time of Workshop: General Instructions Workshop exercises are to be carried out in groups of three. One report per group is due by
More informationPosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
More informationOPTICAL FIBERS INTRODUCTION
OPTICAL FIBERS References: J. Hecht: Understanding Fiber Optics, Ch. 1-3, Prentice Hall N.J. 1999 D. R. Goff: Fiber Optic Reference Guide (2 nd ed.) Focal Press 1999 Projects in Fiber Optics (Applications
More informationCHAPTER 11: Flip Flops
CHAPTER 11: Flip Flops In this chapter, you will be building the part of the circuit that controls the command sequencing. The required circuit must operate the counter and the memory chip. When the teach
More informationCHAPTER D1 ION OPTICS OF MAGNETIC/ELECTRIC SECTOR MS
Back to Basics Section D: Ion Optics CHAPTER D1 ION OPTICS OF MAGNETIC/ELECTRIC SECTOR MS TABLE OF CONTENTS QuickGuide...369 Summary...369 Preamble...371 MassAnalysisofIons...371 MagneticSector...371 ElectrostaticAnalyser(ElectricSector)...375
More informationAlignement of a ring cavity laser
Alignement of a ring cavity laser 1 Introduction This manual describes a procedure to align the cavity of our Ti:Sapphire ring laser and its injection with an Argon-Ion pump laser beam. The setup is shown
More informationElectrical Resonance
Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION
More informationE/M Experiment: Electrons in a Magnetic Field.
E/M Experiment: Electrons in a Magnetic Field. PRE-LAB You will be doing this experiment before we cover the relevant material in class. But there are only two fundamental concepts that you need to understand.
More informationMagnetic Fields and Their Effects
Name Date Time to Complete h m Partner Course/ Section / Grade Magnetic Fields and Their Effects This experiment is intended to give you some hands-on experience with the effects of, and in some cases
More informationPhysics 41, Winter 1998 Lab 1 - The Current Balance. Theory
Physics 41, Winter 1998 Lab 1 - The Current Balance Theory Consider a point at a perpendicular distance d from a long straight wire carrying a current I as shown in figure 1. If the wire is very long compared
More informationA Guide to Acousto-Optic Modulators
A Guide to Acousto-Optic Modulators D. J. McCarron December 7, 2007 1 Introduction Acousto-optic modulators (AOMs) are useful devices which allow the frequency, intensity and direction of a laser beam
More informationExperiment: Static and Kinetic Friction
PHY 201: General Physics I Lab page 1 of 6 OBJECTIVES Experiment: Static and Kinetic Friction Use a Force Sensor to measure the force of static friction. Determine the relationship between force of static
More information1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.
CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change
More informationFibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques
Fibre Bragg Grating Sensors An ntroduction to Bragg gratings and interrogation techniques Dr Crispin Doyle Senior Applications Engineer, Smart Fibres Ltd. 2003 1) The Fibre Bragg Grating (FBG) There are
More informationPHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.
More informationHow To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
More informationLAB 6: GRAVITATIONAL AND PASSIVE FORCES
55 Name Date Partners LAB 6: GRAVITATIONAL AND PASSIVE FORCES And thus Nature will be very conformable to herself and very simple, performing all the great Motions of the heavenly Bodies by the attraction
More informationThe purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law.
260 17-1 I. THEORY EXPERIMENT 17 QUALITATIVE STUDY OF INDUCED EMF Along the extended central axis of a bar magnet, the magnetic field vector B r, on the side nearer the North pole, points away from this
More informationELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.
More informationExperiment 1: SOUND. The equation used to describe a simple sinusoidal function that propagates in space is given by Y = A o sin(k(x v t))
Experiment 1: SOUND Introduction Sound is classified under the topic of mechanical waves. A mechanical wave is a term which refers to a displacement of elements in a medium from their equilibrium state,
More informationPhysical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect
Objectives: PS-7.1 Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect Illustrate ways that the energy of waves is transferred by interaction with
More informationInterferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)
Interferometers The true worth of an experimenter consists in his pursuing not only what he seeks in his experiment, but also what he did not seek. Claude Bernard (1813-1878) OBJECTIVES To examine the
More informationHelium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment.
Helium-Neon Laser Experiment objectives: assemble and align a 3-mW HeNe laser from readily available optical components, record photographically the transverse mode structure of the laser output beam,
More informationLaboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control
More informationIntroduction to Geiger Counters
Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists
More informationEXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab
EXPERIMENT O-6 Michelson Interferometer Abstract A Michelson interferometer, constructed by the student, is used to measure the wavelength of He-Ne laser light and the index of refraction of a flat transparent
More informationLab 1: The Digital Oscilloscope
PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix
More informationAP1 Waves. (A) frequency (B) wavelength (C) speed (D) intensity. Answer: (A) and (D) frequency and intensity.
1. A fire truck is moving at a fairly high speed, with its siren emitting sound at a specific pitch. As the fire truck recedes from you which of the following characteristics of the sound wave from the
More informationAmplified High Speed Fiber Photodetectors
Amplified High Speed Fiber Photodetectors User Guide (800)697-6782 sales@eotech.com www.eotech.com Page 1 of 7 EOT AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your Amplified
More informationModern Physics Laboratory e/m with Teltron Deflection Tube
Modern Physics Laboratory e/m with Teltron Deflection Tube Josh Diamond & John Cummings Fall 2010 Abstract The deflection of an electron beam by electric and magnetic fields is observed, and the charge
More informationIntroduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
More informationMeasurement of Charge-to-Mass (e/m) Ratio for the Electron
Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic
More informationENGINEERING METROLOGY
ENGINEERING METROLOGY ACADEMIC YEAR 92-93, SEMESTER ONE COORDINATE MEASURING MACHINES OPTICAL MEASUREMENT SYSTEMS; DEPARTMENT OF MECHANICAL ENGINEERING ISFAHAN UNIVERSITY OF TECHNOLOGY Coordinate Measuring
More informationLAB 6 - GRAVITATIONAL AND PASSIVE FORCES
L06-1 Name Date Partners LAB 6 - GRAVITATIONAL AND PASSIVE FORCES OBJECTIVES And thus Nature will be very conformable to herself and very simple, performing all the great Motions of the heavenly Bodies
More informationIf you put the same book on a tilted surface the normal force will be less. The magnitude of the normal force will equal: N = W cos θ
Experiment 4 ormal and Frictional Forces Preparation Prepare for this week's quiz by reviewing last week's experiment Read this week's experiment and the section in your textbook dealing with normal forces
More informationAS COMPETITION PAPER 2008
AS COMPETITION PAPER 28 Name School Town & County Total Mark/5 Time Allowed: One hour Attempt as many questions as you can. Write your answers on this question paper. Marks allocated for each question
More informationMagnetism. d. gives the direction of the force on a charge moving in a magnetic field. b. results in negative charges moving. clockwise.
Magnetism 1. An electron which moves with a speed of 3.0 10 4 m/s parallel to a uniform magnetic field of 0.40 T experiences a force of what magnitude? (e = 1.6 10 19 C) a. 4.8 10 14 N c. 2.2 10 24 N b.
More informationA More Efficient Way to De-shelve 137 Ba +
A More Efficient Way to De-shelve 137 Ba + Abstract: Andrea Katz Trinity University UW REU 2010 In order to increase the efficiency and reliability of de-shelving barium ions, an infrared laser beam was
More informationThe accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
More informationPhysics 42 Lab 4 Fall 2012 Cathode Ray Tube (CRT)
Physics 42 Lab 4 Fall 202 Cathode Ray Tube (CRT) PRE-LAB Read the background information in the lab below and then derive this formula for the deflection. D = LPV defl 2 SV accel () Redraw the diagram
More informationReflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
More informationLab 4: Magnetic Force on Electrons
Lab 4: Magnetic Force on Electrons Introduction: Forces on particles are not limited to gravity and electricity. Magnetic forces also exist. This magnetic force is known as the Lorentz force and it is
More informationMagnetic Fields and Forces. AP Physics B
Magnetic ields and orces AP Physics acts about Magnetism Magnets have 2 poles (north and south) Like poles repel Unlike poles attract Magnets create a MAGNETIC IELD around them Magnetic ield A bar magnet
More informationThe Sonometer The Resonant String and Timbre Change after plucking
The Sonometer The Resonant String and Timbre Change after plucking EQUIPMENT Pasco sonometers (pick up 5 from teaching lab) and 5 kits to go with them BK Precision function generators and Tenma oscilloscopes
More informationAMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE
AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your Amplified High Speed Fiber Photodetector. This user s guide will help answer any questions you may have regarding the
More informationTheremino System Theremino Spectrometer Technology
Theremino System Theremino Spectrometer Technology theremino System - Theremino Spectrometer Technology - August 15, 2014 - Page 1 Operation principles By placing a digital camera with a diffraction grating
More information1 of 9 2/9/2010 3:38 PM
1 of 9 2/9/2010 3:38 PM Chapter 23 Homework Due: 8:00am on Monday, February 8, 2010 Note: To understand how points are awarded, read your instructor's Grading Policy. [Return to Standard Assignment View]
More informationDiffraction and Young s Single Slit Experiment
Diffraction and Young s Single Slit Experiment Developers AB Overby Objectives Preparation Background The objectives of this experiment are to observe Fraunhofer, or far-field, diffraction through a single
More informationConceptual: 1, 3, 5, 6, 8, 16, 18, 19. Problems: 4, 6, 8, 11, 16, 20, 23, 27, 34, 41, 45, 56, 60, 65. Conceptual Questions
Conceptual: 1, 3, 5, 6, 8, 16, 18, 19 Problems: 4, 6, 8, 11, 16, 20, 23, 27, 34, 41, 45, 56, 60, 65 Conceptual Questions 1. The magnetic field cannot be described as the magnetic force per unit charge
More informationFRICTION, WORK, AND THE INCLINED PLANE
FRICTION, WORK, AND THE INCLINED PLANE Objective: To measure the coefficient of static and inetic friction between a bloc and an inclined plane and to examine the relationship between the plane s angle
More informationHow To Use Gc-Ms
The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter
More informationSynthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
More informationGermanium Diode AM Radio
Germanium Diode AM Radio LAB 3 3.1 Introduction In this laboratory exercise you will build a germanium diode based AM (Medium Wave) radio. Earliest radios used simple diode detector circuits. The diodes
More informationAccurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis
Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis B. Apicella*, M. Passaro**, X. Wang***, N. Spinelli**** mariadellarcopassaro@gmail.com
More informationThe Basics of Scanning Electron Microscopy
The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single
More informationAcousto-optic modulator
1 of 3 Acousto-optic modulator F An acousto-optic modulator (AOM), also called a Bragg cell, uses the acousto-optic effect to diffract and shift the frequency of light using sound waves (usually at radio-frequency).
More informationFunctional Data Analysis of MALDI TOF Protein Spectra
Functional Data Analysis of MALDI TOF Protein Spectra Dean Billheimer dean.billheimer@vanderbilt.edu. Department of Biostatistics Vanderbilt University Vanderbilt Ingram Cancer Center FDA for MALDI TOF
More informationChapter 17: Light and Image Formation
Chapter 17: Light and Image Formation 1. When light enters a medium with a higher index of refraction it is A. absorbed. B. bent away from the normal. C. bent towards from the normal. D. continues in the
More informationPHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.
PHYS 222 Spring 2012 Final Exam Closed books, notes, etc. No electronic device except a calculator. NAME: (all questions with equal weight) 1. If the distance between two point charges is tripled, the
More informationForce on Moving Charges in a Magnetic Field
[ Assignment View ] [ Eðlisfræði 2, vor 2007 27. Magnetic Field and Magnetic Forces Assignment is due at 2:00am on Wednesday, February 28, 2007 Credit for problems submitted late will decrease to 0% after
More informationTHE BOHR QUANTUM MODEL
THE BOHR QUANTUM MODEL INTRODUCTION When light from a low-pressure gas is subject to an electric discharge, a discrete line spectrum is emitted. When light from such a low-pressure gas is examined with
More informationErrors Due to Shared Leadwires in Parallel Strain Gage Circuits
Micro-Measurements Strain Gages and Instruments Errors Due to Shared Leadwires in Parallel Strain Gage Circuits TN-516 1. Introduction The usual, and preferred, practice with multiple quarterbridge strain
More information6/2016 E&M forces-1/8 ELECTRIC AND MAGNETIC FORCES. PURPOSE: To study the deflection of a beam of electrons by electric and magnetic fields.
6/016 E&M forces-1/8 ELECTRIC AND MAGNETIC FORCES PURPOSE: To study the deflection of a beam of electrons by electric and magnetic fields. APPARATUS: Electron beam tube, stand with coils, power supply,
More informationO6: The Diffraction Grating Spectrometer
2B30: PRACTICAL ASTROPHYSICS FORMAL REPORT: O6: The Diffraction Grating Spectrometer Adam Hill Lab partner: G. Evans Tutor: Dr. Peter Storey 1 Abstract The calibration of a diffraction grating spectrometer
More informationDevelopment of Optical Wave Microphone Measuring Sound Waves with No Diaphragm
Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and
More informationLab Exercise 1: Acoustic Waves
Lab Exercise 1: Acoustic Waves Contents 1-1 PRE-LAB ASSIGNMENT................. 2 1-3.1 Spreading Factor: Spherical Waves........ 2 1-3.2 Interference In 3-D................. 3 1-4 EQUIPMENT........................
More informationChapter 19 Operational Amplifiers
Chapter 19 Operational Amplifiers The operational amplifier, or op-amp, is a basic building block of modern electronics. Op-amps date back to the early days of vacuum tubes, but they only became common
More informationLab 7: Rotational Motion
Lab 7: Rotational Motion Equipment: DataStudio, rotary motion sensor mounted on 80 cm rod and heavy duty bench clamp (PASCO ME-9472), string with loop at one end and small white bead at the other end (125
More informationHow to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer
How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer Application Note Introduction In a pulsed radar system, one of the key transmitter-side components
More informationView of ΣIGMA TM (Ref. 1)
Overview of the FESEM system 1. Electron optical column 2. Specimen chamber 3. EDS detector [Electron Dispersive Spectroscopy] 4. Monitors 5. BSD (Back scatter detector) 6. Personal Computer 7. ON/STANDBY/OFF
More informationPesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationChapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)
Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know
More informationFraunhofer Diffraction
Physics 334 Spring 1 Purpose Fraunhofer Diffraction The experiment will test the theory of Fraunhofer diffraction at a single slit by comparing a careful measurement of the angular dependence of intensity
More informationGraphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
More informationX Ray Flourescence (XRF)
X Ray Flourescence (XRF) Aspiring Geologist XRF Technique XRF is a rapid, relatively non destructive process that produces chemical analysis of rocks, minerals, sediments, fluids, and soils It s purpose
More informationRF Measurements Using a Modular Digitizer
RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.
More informationNewton s Second Law. ΣF = m a. (1) In this equation, ΣF is the sum of the forces acting on an object, m is the mass of
Newton s Second Law Objective The Newton s Second Law experiment provides the student a hands on demonstration of forces in motion. A formulated analysis of forces acting on a dynamics cart will be developed
More informationPrelab Exercises: Hooke's Law and the Behavior of Springs
59 Prelab Exercises: Hooke's Law and the Behavior of Springs Study the description of the experiment that follows and answer the following questions.. (3 marks) Explain why a mass suspended vertically
More informationChapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
More informationEXPERIMENT III EXPERIMENTS WITH AN ELECTRON BEAM
EXPERIMENT III EXPERIMENTS WITH AN ELECTRON BEAM An electron beam is a collection of free electrons, all traveling in approximately the same direction with the approximately the same velocity. While it
More informationSTATIC AND KINETIC FRICTION
STATIC AND KINETIC FRICTION LAB MECH 3.COMP From Physics with Computers, Vernier Software & Technology, 2000. INTRODUCTION If you try to slide a heavy box resting on the floor, you may find it difficult
More informationLab #4 - Linear Impulse and Momentum
Purpose: Lab #4 - Linear Impulse and Momentum The objective of this lab is to understand the linear and angular impulse/momentum relationship. Upon completion of this lab you will: Understand and know
More information1 One Dimensional Horizontal Motion Position vs. time Velocity vs. time
PHY132 Experiment 1 One Dimensional Horizontal Motion Position vs. time Velocity vs. time One of the most effective methods of describing motion is to plot graphs of distance, velocity, and acceleration
More information