1 Audiometer and Hearing Aid measurements / mini seminar DELTA Technical-Audiological Laboratory - TAL Edisonsvej Odense C Tlf tal@delta.dk

2 Program 14:00 Welcome and introduction/ Gert Ravn 14:10 SW Development and implementation of speech-gain measurements according IEC / Søren L Jørgensen 14:40 Future EMC measurements on Hearing Aids/ Allan G Kristensen 15:00 New aspects of hearing aid documentation according IEC and updated IEC / Gert Ravn 15:20 Audiometer calibration. Round Robin test results/ Carsten Daugaard 16:00 Closing

3 Audiometer calibration - Agenda Background Audiometers in Denmark Public clinics ENT specialists Private dispensers Issues in practical audiometer calibration Questions reading the standards Deviations in audiometer testing Round robin test Results

4 Background DELTA has as consultants for hearing health care in Denmark traditionally been involved in audiometer calibration. Before 2000 as consultants for technicians in public health care when special questions about calibration arose DELTA has published 2 technical reports in Danish explaining calibration issues For many years DELTA has been represented in WG 10 Audiometric equipment by Gert Ravn (and for a period Carsten Daugaard) Since 2000 DELTA has as a part of the certification of private clinics tested their audiometer s calibration

5 Audiometers in Denmark Two Danish manufactures dominate the market, a small handful other manufactures are represented. ENT specialists, often older audiometers, calibrated by manufactures based on a service contract Approx. 17 public clinics, older clinical audiometers, but also state of the art equipment. calibration is moving from in-house to service contracts Approx 200 Private clinics (some of them also ENT specialists), newer audiometers often portable, yearly calibrated by manufactures based on a service contract and controlled by DELTA as a part of the certification

6 Issues in practical calibration: Speech audiometry Offset value Practical issues in speech calibration Free field calibration Attenuation of background noise RETSPL for new headphones

7 Speech audiometry i Denmark Since Dr. Røjskjær in Odense back in the fifties conducted the first speech tests, this auditory test has been carried out in Denmark Recorded speech material have existed also since the fifthies Late in the eighties the national material Dantale emerged and grew to be the national standard

8 Speech audiometry i Denmark II Based upon the DANTALE material a Discrimination score (DS) and a Speech recognition threshold (SRT) is measured. DS corresponds to the in ISO clause 3.12 defined maximum speech recognition score SRT corresponds to some extend th the in ISO i 3.13 defined speech recognition threshold level, however the threshold for the danish one syllable digits are defined as 2 out of three (66%) where the standard defines 50 % as the threshold Both are as standard presented in headphones with no noise. Thresholds should be obtained in db HL

9 Speech audiometry i Denmark III According to the IEC :1994 standard speech audiometry offset are calibrated to 20 db. In principle every speech material have its own unique threshold, but the 20 db is a universal offset. Speech audiometry are calibrated at 1 khz and presented through TDH-39 Preferably the headphone should have been equalised out Clinicians switching to other headphones report misfits between audiometry and speech audiometry findings

10 ANSI- values for speech audiometry ANSI S offers differentiated offset values for the three transducers Lowering the offset to 12,5 for Insert phones (in 2CC coupler) DELTA investigated if 12,5 db was a meaning full value for insert phones on DANTALE material Or the documented offset of 28,15 db from the reports of the DANTALE material made more sense

11 DELTA test of speech offset Unfortunately no clear conclusion could be drawn from the tests Other than differences between the headphones seems to be limited And that no offered value was spot on So we decided to stick to the 20 db offset, unless really good arguments suggested another value.

12 Attenuation of background noise Requriements of maximum background noise levels are based upon the dampning of TDH-39. In ISO :2010 maximum background levels for TDH-39 are given along with the dampening factor of TDH-39, ER-3A og HDA 200. Attenuation values for ER-3A og Ear 5A is assumed idenical, as the yellow foam plug is the same for both earphones

13 Attenuation of background noise II Attenuation values of new headphones should be published and then requirements for background noise levels can be calculated. In Denmark practise have been that the TDH-39 headphone have been build-in a sound attenuating cup (i.e from a hearing protector) Official attenuation values are not given for these constructions, The supraaural TDH-39 is in this way turned into a circumaural headphone!

14 Dæmpningsværdier for HD280* HDA280 Schalldämmung / Sound attention Frequency d f in db on d f in db on (Hz) B&K 4153 Head and Torso 4128C dB dB dB 6dB dB 10dB dB 12dB dB 20dB dB 25dB dB 27dB * Fra mail korrespondance med Sennheisser

15 RETSPL for different headphones Traditionally all audiometers in Denmark have had TDH39 og ER-3A head/earphones. RETSPL on (6CC) og couplers can be found in ISO (Table 1 og 2) for THD-39 And in ISO389-2 on IEC (711) and IEC (2CC) for ER-3A insert phones HDA200 entered in the High-frequency standard ISO389-5, now it is in u: ISO389-8:2004 table 1-) Ear 5A is commonly accepted as it has the RETSPL according to ER-3A (but are nt really mentioned in the standard) Supraaurale Headphone types, not mentioned above can be measured on the IEC , coupler according to ISO 389-1, where RETSPL for these headphones is given in table 2

16 ISO excerpts : - ISO 389-8: Circumaurale HP (only Sennheisser HDA 200)

17 New headphones emerges New headphones are seen in Denmark: DD45 (TDH39 Clone) and HD280 Rumor has that a new sennheisser is on its way Supraaural headphones are in priciple covered in ISO table 2, where RETSPL for these headphones is given on the IEC , coupler Circumaural headphones are covered in ISO 389-8, but only HDA 200 can be used?

18 DD45 W. Demand (Claus Elberling)

19 Audiometer measurement I TDH-39 Reference A B C D Frekvens db HL (Hz) L R L R L R L R L R ,0 0,0-2,9-3,0-1,7-1,8-1,6-1,8-0,2-0, ,0 0,0 0,6 0,6-0,6-0,7-1,0-1,2-0,3-0, ,0 0,0 1,2 1,3 0,4 0,4-1,0-1,1-0,5-0, ,0 0,0 1,0 1,2 0,1 0,1-1,0-1,1-0,3-0, ,0 0,0 0,7 0,8 0,1 0,2-1,0-1,0-0,2-0, ,0 0,0 0,3 0,7-0,1 0,0-1,2-1,1-1,4-1, ,0 0,0 0,6 0,4 0,1 0,0-0,9-1,0 0,6 0, ,0 0,0 2,6 2,8 0,3 0,4-0,9-1,0-0,3-0, ,0 0,0 2,3 2,5 0,6 0,7-0,7-0,7-0,1-0, ,0 0,0-4,7-7,9 0,9 0,8-1,4-1,5-1,5-1, ,0 0,0 0,7 2,0-0,1 0,3-1,7-1,5-0,4-0,7 Variation sabove 1,5 db Variations above limit

20 Audiometer measurement II EAR-3A Reference A B C Frekvens db HL (Hz) L R L R L R L R ,0 0,0-0,2-0,5-0,1 0,0 0,0 0, ,0 0,0-0,4-0,4-0,3-0,2-0,1-0, ,0 0,0-0,2-0,2 0,2 0,2-0,1-0, ,0 0,0-0,2-0,3 0,2 0,3-0,2-0, ,0 0,0-0,1-0,3 0,7 0,7-0,1-0, ,0 0,0 0,0-0,4 1,7 1,8-0,1-0, ,0 0,0 0,1-0,6 2,8 2,8 0,0 0, ,0 0,0 0,4-0,8 1,8 1,9 0,4 0, ,0 0,0-1,1-1,5-4,4-3,7 1,0 1, ,0 0,0-0,9-0,9 3,3 3,4 0,8 0, ,0 0,0-0,1-0,7-1,7-1,6 0,1 0,2 Variation sabove 1,5 db Variations above limit

21 Audiometer measurement III HDA-200 Reference A B C Frekvens db HL (Hz) L R L R L R L R ,0 0,0-0,1 0,0-0,4-0,5 0,0-0, ,0 0,0-0,2 0,1-0,2-0,2-0,4-0, ,0 0,0-0,3-0,1-0,3-0,3-0,2-0, ,0 0,0-0,4-0,2-0,4-0,6-0,4-0, ,0 0,0-0,5-0,3-0,6-0,6-0,4-0, ,0 0,0-0,7-0,5-1,1-0,7-0,8-0, ,0 0,0-0,7-0,5-1,0-1,0-0,3-1, ,0 0,0-1,8-1,7-1,5-1,8-1,4-2, ,0 0,0-1,0-0,6-0,7-0,9-0,7-1, ,0 0,0-0,2 0,1-0,5 0,3-0,1 0, ,0 0,0-0,1 0,2-0,2-0,4-0,2 0,0 Variation sabove 1,5 db Variations above limit

22 Audiometer measurement IIII BC-71 db HL Frekvens (Hz) Reference A C1 C2 D ,0-0,8-0,5-4,7 0, ,0-0,1-2,9-4,0 0, ,0-0,2-2,4-4,4 0, ,0 0,0-2,4-4,6 0, ,0-0,6-2,1-5,6 0, ,0-0,7-3,0-4,5 0, ,0-0,5-1,2-2,6 0, ,0-3,4 0,8-6,5 0, ,0 2,0-4,4-4,5-0,2 Variation sabove 1,5 db Variations above limit

Instructions for the use of E-A-RTONE 3A. Insert Earphones. Revised 1997 per ANSI S3.6-1996 and ISO 389-2:1994 12/99

Instructions for the use of E-A-RTONE 3A. Insert Earphones. Revised 1997 per ANSI S3.6-1996 and ISO 389-2:1994 12/99 Instructions for the use of E-A-RTONE 3A Insert Earphones Revised 1997 per ANSI S3.6-1996 and ISO 389-2:1994 12/99 INTRODUCTION Congratulations on the purchase of your new E-A-RTONE 3A Insert Earphones.

More information

GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER. Setting The Clinical Standard

GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER. Setting The Clinical Standard GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER Setting The Clinical Standard GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER Tradition of Excellence The GSI AudioStar Pro continues the tradition of

More information

3M E-A-RTONE 3A and 5A Audiometric Insert Earphones

3M E-A-RTONE 3A and 5A Audiometric Insert Earphones 34-8708-1478-6 GZXO$ 3M 3A and 5A Audiometric Insert Earphones www.3m.eu/safety GZXO$ INTRODUCTION Congratulations on the purchase of your new 3A or 5A Insert Earphones. Whether you are performing comprehensive

More information

PURE TONE AUDIOMETRY Andrew P. McGrath, AuD

PURE TONE AUDIOMETRY Andrew P. McGrath, AuD PURE TONE AUDIOMETRY Andrew P. McGrath, AuD Pure tone audiometry is the standard behavioral assessment of an individual s hearing. The results of pure tone audiometry are recorded on a chart or form called

More information

MA33 & MA55. PC Based Audiometers. The ideal choice for health care professionals on the go. www.maico-diagnostics.com. by Maico Diagnostics

MA33 & MA55. PC Based Audiometers. The ideal choice for health care professionals on the go. www.maico-diagnostics.com. by Maico Diagnostics PC Based Audiometers by Maico Diagnostics MA33 & MA55 The ideal choice for health care professionals on the go 6 4 MA33 Pictured www.maico-diagnostics.com Portable - PC Based Audiometers Our PC based Audiometers

More information

Main features * : Internal database > 1000 tests. Suprathreshold tests. High frequency. Musical frequencies. Automatic masking.

Main features * : Internal database > 1000 tests. Suprathreshold tests. High frequency. Musical frequencies. Automatic masking. NEW RNGE OF UDIOMETER PRODUCTS screening, basic diagnosis, clinical diagnosis SIBELSOUND 400 Main features * : dvanced digital technology based on DSP Calculation of hearing loss and diagnosis Internal

More information

PURE TONE AUDIOMETER

PURE TONE AUDIOMETER PURE TONE AUDIOMETER V. Vencovský, F. Rund Department of Radioelectronics, Faculty of Electrical Engineering, Czech Technical University in Prague, Czech Republic Abstract Estimation of pure tone hearing

More information

DAILY BIOLOGIC CHECK OF AUDIOMETERS

DAILY BIOLOGIC CHECK OF AUDIOMETERS DAILY BIOLOGIC CHECK OF AUDIOMETERS An audiometer must be checked before each day of use or if a problem is suspected with the equipment during use. After the audiometer is turned on, the operator should

More information

Guideline for diagnosing occupational noise-induced hearing loss. Part 3: Audiometric standards

Guideline for diagnosing occupational noise-induced hearing loss. Part 3: Audiometric standards Purdy & Williams: Guidelines for audiometry for diagnosis of NIHL Page 1 of 59 Guideline for diagnosing occupational noise-induced hearing loss Part 3: Audiometric standards Suzanne Purdy Head of Speech

More information

SPEECH AUDIOMETRY. @ Biswajeet Sarangi, B.Sc.(Audiology & speech Language pathology)

SPEECH AUDIOMETRY. @ Biswajeet Sarangi, B.Sc.(Audiology & speech Language pathology) 1 SPEECH AUDIOMETRY Pure tone Audiometry provides only a partial picture of the patient s auditory sensitivity. Because it doesn t give any information about it s ability to hear and understand speech.

More information

Speech as input for technical measures

Speech as input for technical measures Speech as input for technical measures Carsten Daugaard DELTA, Edisonvej 24, 5000 Odense C, Denmark, cd@delta.dk Ellen Raben Pedersen Institute of Technology and Innovation, University of Southern Denmark,

More information

Pediatric Whitepaper. Electrophysiological Threshold Estimation and Infant Hearing Instrument Fitting

Pediatric Whitepaper. Electrophysiological Threshold Estimation and Infant Hearing Instrument Fitting Pediatric Whitepaper January 2010 Electrophysiological Threshold Estimation and Infant Hearing Instrument Fitting Merethe Lindgaard Fuglholt, M.A. Oticon A/S, Pediatric Audiology Abstract This paper addresses

More information

4. Have you been exposed to loud noise or listened to music with headphone in the past 24 hours? [ ] Yes [ ] No [ ] Refused [ ] Don t know

4. Have you been exposed to loud noise or listened to music with headphone in the past 24 hours? [ ] Yes [ ] No [ ] Refused [ ] Don t know PhenX Measure: Audiogram Hearing Test (#200100) PhenX Protocol: Audiogram Hearing Test (#200101) Date of Interview/Examination (MM/DD/YYYY): Preliminary Activities 1. Does the examinee have hearing aids

More information

Audiometer Calibration System Manual

Audiometer Calibration System Manual Audiometer Calibration System Manual Larson Davis Audiometer Calibration System Manual IAUDIT.01 Rev J Copyright Copyright 2015 by PCB Piezotronics, Inc. This manual is copyrighted, with all rights reserved.

More information

GUIDELINES ON THE ACOUSTICS OF SOUND FIELD AUDIOMETRY IN CLINICAL AUDIOLOGICAL APPLICATIONS

GUIDELINES ON THE ACOUSTICS OF SOUND FIELD AUDIOMETRY IN CLINICAL AUDIOLOGICAL APPLICATIONS NOTE Although care has been taken in preparing the information supplied by the British Society of Audiology, the BSA does not and cannot guarantee the interpretation and application of it. The BSA cannot

More information

Operating Instructions MA 27

Operating Instructions MA 27 Operating Instructions MA 27 1162-7830 Rev G 03/11 Maico Diagnostics 7625 Golden Triangle Drive Eden Prairie, MN 55344 Toll free (888) 941-4201 Maico Diagnostic GmbH Salzufer 13/14 10587 Berlin, Germany

More information

MA 25 Operating Manual

MA 25 Operating Manual MA 25 Operating Manual MAICO Diagnostics 10393 West 70 th Street Eden Prairie, MN 55344, USA Toll Free 888.941.4201 MA 25 Operating Manual Table of Contents Page 1.... Introduction... 1 2.... Description...

More information

Recommended Procedure. Pure-tone air-conduction and bone-conduction threshold audiometry with and without masking

Recommended Procedure. Pure-tone air-conduction and bone-conduction threshold audiometry with and without masking Recommended Procedure Pure-tone air-conduction and bone-conduction threshold audiometry with and without masking Date: 9 th September 2011 Minor amendments: 6 th February 2012 Date for review: September

More information

Air Conduction Audiometer Series. MA 27 with headphones and built-in carrying case. MA1, MA25 & MA27. www.maico-diagnostics.com

Air Conduction Audiometer Series. MA 27 with headphones and built-in carrying case. MA1, MA25 & MA27. www.maico-diagnostics.com Air Conduction Audiometer Series MA 27 with headphones and built-in carrying case. MA1, MA25 & MA27 www.maico-diagnostics.com MA 1 The super-lightweight MA1 weighs in at only 1 pound. Small enough to hold

More information

GSI AUDIOSTAR PRO. Free Call: 1800 639 263 CLINICAL TWO-CHANNEL AUDIOMETER. Australian Distributer:

GSI AUDIOSTAR PRO. Free Call: 1800 639 263 CLINICAL TWO-CHANNEL AUDIOMETER. Australian Distributer: GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER Australian Distributer: Setting The Clinical Standard Free Call: 1800 639 263 GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER Tradition of Excellence

More information

By Michael Block, PhD

By Michael Block, PhD By Michael Block, PhD IHS offers a diversity of options for obtaining continuing education credit: seminars and classroom training, institutional courses, online studies, and distance learning programs.

More information

Trends In Amplification. Clinical Protocols for Hearing Instrument Fitting in the Desired Sensation Level Method

Trends In Amplification. Clinical Protocols for Hearing Instrument Fitting in the Desired Sensation Level Method Trends In Amplification VOLUME 9, NUMBER 4, 2005 Clinical Protocols for Hearing Instrument Fitting in the Desired Sensation Level Method Marlene Bagatto, AuD,* Sheila Moodie, MClSc,* Susan Scollie, PhD,*

More information

MA 41 Operating Instructions

MA 41 Operating Instructions MA 41 Operating Instructions MAICO Diagnostics 10393 West 70 th Street Eden Prairie, MN 55344, USA Toll Free 888.941.4201 Table of Contents Page 1 Introduction... 1 1.1 Intended Use Statement... 1 1.2

More information

Audio Examination. Place of Exam:

Audio Examination. Place of Exam: Audio Examination Name: Date of Exam: SSN: C-number: Place of Exam: The Handbook of Standard Procedures and Best Practices for Audiology Compensation and Pension Exams is available online. ( This is a

More information

HEARING SCREENING FOR CHILDREN

HEARING SCREENING FOR CHILDREN PURPOSE Audiologic (hearing) screening identifies those persons who are likely to have hearing impairments or disorders that may interfere with body function/structure and/or activity/participation as

More information

TELEPHONE HEADSET SOUND EXPOSURE ASSESSMENT BY THRESHOLD COMPARISON

TELEPHONE HEADSET SOUND EXPOSURE ASSESSMENT BY THRESHOLD COMPARISON Mecánica Computacional Vol XXX, págs. 3201-3216 (artículo completo Oscar Möller, Javier W. Signorelli, Mario A. Storti (Eds. Rosario, Argentina, 1-4 Noviembre 2011 TELEPHONE HEADSET SOUND EXPOSURE ASSESSMENT

More information

HEARING MEASUREMENT 8.1. INTRODUCTION (RATIONALE FOR AUDIOMETRY)

HEARING MEASUREMENT 8.1. INTRODUCTION (RATIONALE FOR AUDIOMETRY) 8 HEARING MEASUREMENT John R. Franks, Ph.D. Chief, Hearing Loss Prevention Section Engineering and Physical Hazards Branch Division of Applied Research and Technology National Institute for Occupational

More information

Operating Instructions MAICO MA 52

Operating Instructions MAICO MA 52 Operating Instructions MAICO MA 52 Table of Contents Page 1 Introduction... 4 2 Description... 5 3 Getting started... 7 3.1 Unpacking your instrument... 7 3.2 Preparing the MA 52 for use... 8 3.3 Getting

More information

PROFESSIONAL BOARD FOR SPEECH, LANGUAGE AND HEARING PROFESSIONS STANDARDS OF PRACTICE IN AUDIOLOGY

PROFESSIONAL BOARD FOR SPEECH, LANGUAGE AND HEARING PROFESSIONS STANDARDS OF PRACTICE IN AUDIOLOGY The Health Professions Council of South Africa PROFESSIONAL BOARD FOR SPEECH, LANGUAGE AND HEARING PROFESSIONS STANDARDS OF PRACTICE IN AUDIOLOGY December 2002 The following are considered to be standard

More information

S-822. Noise and Hearing Conservation RISK MANAGEMENT

S-822. Noise and Hearing Conservation RISK MANAGEMENT RISK MANAGEMENT Noise and Hearing Conservation Resources and information for developing Patient handling programs in health care Employers with workplaces that have excessive levels of noise need to implement

More information

The Correlation of Results between Pure Tone Audiometry and the Virtual Audiometer: A Simulated Air-Conduction Clinic-based Audiometer

The Correlation of Results between Pure Tone Audiometry and the Virtual Audiometer: A Simulated Air-Conduction Clinic-based Audiometer Anne Elizabeth M. Javellana, MD Ray U. Casile, MD Department of Otolaryngology Head and Neck Surgery St. Luke s Medical Center The Correlation of Results between Pure Tone Audiometry and the Virtual Audiometer:

More information

OPEN ACCESS GUIDE TO AUDIOLOGY AND HEARING AIDS FOR OTOLARYNGOLOGISTS

OPEN ACCESS GUIDE TO AUDIOLOGY AND HEARING AIDS FOR OTOLARYNGOLOGISTS HEARING THRESHOLD IN DECIBEL (db HL) OPEN ACCESS GUIDE TO AUDIOLOGY AND HEARING AIDS FOR OTOLARYNGOLOGISTS PURE TONE AUDIOMETRY Pure tone audiometry is generally the 1st quantitative hearing test done

More information

Product Line for Hearing Screening & Diagnostics

Product Line for Hearing Screening & Diagnostics Product Line for Hearing Screening & Diagnostics General Catalogue 2012 Hearing Test Devices www.maico.biz Since 1937 75 Welcome to the World of MAICO Introduction 75 For 75 years MAICO has been an innovator

More information

Practice Standards for Hearing Service Providers

Practice Standards for Hearing Service Providers Practice Standards for Hearing Service Providers The WCB has been unsuccessful in obtaining input from the SASLPA for the development of practice standards and a fee schedule to cover hearing service providers

More information

Step-by-Step RECD Guide

Step-by-Step RECD Guide Precision Audiometric Instruments www.medrx-usa.com Step-by-Step RECD Guide The RECD task involves 4 steps: 1 - Complete Calibration of the Speakers and Probe Tube 2 - Measure an Ear Response 3 - Perform

More information

Audiometer Calibration System

Audiometer Calibration System Audiometer Calibration System User Manual Larson Davis 1681 West 820 North, Provo, UT 84601-1341 Phone: (801) 375-0177, FAX: (801) 375-0182 www.larsondavis.com IAUDIT.01 Rev. C AUDit & AUDIOMETER CALIBRATION

More information

Pure-Tone Assessment and Screening of Children with Middle-Ear Effusion

Pure-Tone Assessment and Screening of Children with Middle-Ear Effusion J Am Acad Audiol 5 : 173-182 (1994) Pure-Tone Assessment and Screening of Children with Middle-Ear Effusion Shlomo Silman* t Carol A. Silvermantt Daniel S. Arick Abstract The purpose of this prospective

More information

The impact of using real ear measures to calculate prescriptive targets on hearing aid follow-up visits

The impact of using real ear measures to calculate prescriptive targets on hearing aid follow-up visits Washington University School of Medicine Digital Commons@Becker Independent Studies and Capstones Program in Audiology and Communication Sciences 2013 The impact of using real ear measures to calculate

More information

Guidelines for Manual Pure-Tone Threshold Audiometry

Guidelines for Manual Pure-Tone Threshold Audiometry for Manual Pure-Tone Threshold Audiometry Working Group on Manual Pure-Tone Threshold Audiometry Reference this material as: American Speech-Language-Hearing Association. (2005). for Manual Pure-Tone Threshold

More information

1000+ AUDIOMETER OPERATING INSTRUCTIONS

1000+ AUDIOMETER OPERATING INSTRUCTIONS 1000+ AUDIOMETER OPERATING INSTRUCTIONS AMBCO Model 1000+ is a microprocessor controlled pure tone air conduction audiometer with automated screening test feature. INDEX A. Getting Started Patient Instructions

More information

The Accuracy of 0 db HL as an Assumption of Normal Hearing

The Accuracy of 0 db HL as an Assumption of Normal Hearing The Accuracy of 0 db HL as an Assumption of Normal Hearing Introduction An operating assumption of diagnostic audiology is that hearing level of a young adult with no known hearing loss or history of noise

More information

Evaluating sound quality in hearing aids with reference test audiograms

Evaluating sound quality in hearing aids with reference test audiograms 1 Evaluating sound quality in hearing aids with reference test audiograms CARSTEN DAUGAARD, SØREN L. JØRGENSEN, CHRIS V. JØRGENSEN, SØREN V. LEGARTH AND NICK V. ZACHARV DELTA Technical-Audiological laboratory,

More information

CSA Z94.2 Hearing Protection Devices

CSA Z94.2 Hearing Protection Devices THE NEW CANADIAN STANDARD ASSOCIATION STANDARD ON HEARING PROTECTORS: CSA Z94.2 Hearing Protection Devices Alberto Behar, Peng. C.I.H. Ryerson University Fall Symposium October 24, 2013 1 Prepared by the

More information

CHAPTER 11 NOISE AND HEARING CONSERVATION PROGRAM

CHAPTER 11 NOISE AND HEARING CONSERVATION PROGRAM CHAPTER 11 NOISE AND HEARING CONSERVATION PROGRAM INTRODUCTION This program contains information on the effects, evaluation, and control of noise. For assistance in evaluating a noise problem, contact

More information

HyperX Cloud Headset

HyperX Cloud Headset HyperX Cloud Headset Document No. 480KHX-H3CL/WR.A01 HyperX Cloud Headset Page 1 of 12 Introduction Optimized for pro-gaming, HyperX Cloud Headset (KHX-H3CL/WR) is a high-quality communicating device that

More information

UCSC HEARING CONSERVATION PROGRAM Environment, Health & Safety (EH&S) August 2007

UCSC HEARING CONSERVATION PROGRAM Environment, Health & Safety (EH&S) August 2007 UCSC HEARING CONSERVATION PROGRAM Environment, Health & Safety (EH&S) August 2007 Table of Contents Purpose Applicable Regulations and Standards Responsibilities Deans and Directors Department Heads, Managers,

More information

The NAL Percentage Loss of Hearing Scale

The NAL Percentage Loss of Hearing Scale The NAL Percentage Loss of Hearing Scale Anne Greville Audiology Adviser, ACC February, 2010 The NAL Percentage Loss of Hearing (PLH) Scale was developed by John Macrae of the Australian National Acoustic

More information

Western University. Hearing Protection Program. Prepared by: Occupational Health and Safety

Western University. Hearing Protection Program. Prepared by: Occupational Health and Safety Western University Hearing Protection Program Prepared by: Occupational Health and Safety June 1, 2014 1 Table of Content Page 1.0 Purpose 3 2.0 Definitions / Abbreviations 3 3.0 Application 4 3.1 Scope

More information

Leading the Advancement of Hearing Conservation

Leading the Advancement of Hearing Conservation 3M Occupational Health and Environmental Safety Division 3M E-A-Rfit Validation System Leading the Advancement of Hearing Conservation Look How Well It Fits Into Your Hearing Conservation Program Integrating

More information

Manual Pure-Tone Threshold Audiometry

Manual Pure-Tone Threshold Audiometry Guidelines Manual Pure-Tone Threshold Audiometry Guidelines 1977 / II - 221 Manual Pure-Tone Threshold Audiometry ASHA Committee on Audiometric Evaluation Reference this material as: American Speech-Language-

More information

Testing FM Systems on the FP35 Hearing Aid Analyzer -1-

Testing FM Systems on the FP35 Hearing Aid Analyzer -1- Testing FM Systems on the FP35 Hearing Aid Analyzer -1- Testing FM Systems on the FP35 Hearing Aid Analyzer Introduction This workbook describes how to test FM systems with the FP35 Hearing Aid Analyzer

More information

hi HealthInnovations Methods for Prescribing Hearing Device Gain: Reliability and Accuracy

hi HealthInnovations Methods for Prescribing Hearing Device Gain: Reliability and Accuracy hi HealthInnovations Methods for Prescribing Hearing Device Gain: Reliability and Accuracy Dianne J. Van Tasell, Ph.D. Dianne Van Tasell is Adjunct Professor in the Department of Speech, Language, and

More information

Early Detection of Adult Hearing Loss: Next Steps to Clinical Practice

Early Detection of Adult Hearing Loss: Next Steps to Clinical Practice Early Detection of Adult Hearing Loss: Next Steps to Clinical Practice Jonathan Siegel Sumitrajit Dhar Knowles Hearing Center Roxelyn & Richard Pepper Department of Communication Science & Disorders Northwestern

More information

Guidelines on Hearing Checks and Audiometry Under the Safety, Health and Welfare at Work (General Application) Regulations 2007, Control of Noise at

Guidelines on Hearing Checks and Audiometry Under the Safety, Health and Welfare at Work (General Application) Regulations 2007, Control of Noise at Guidelines on Hearing Checks and Audiometry Under the Safety, Health and Welfare at Work (General Application) Regulations 2007, Control of Noise at Work CONTENTS 1. Hearing Checks and Screening Audiometry

More information

Hearing Screening Protocol for School Aged Children

Hearing Screening Protocol for School Aged Children This is an official Northern Trust policy and should not be edited in any way Hearing Screening Protocol for School Aged Children Reference Number: NHSCT/12/577 Target audience: This policy is directed

More information

Once your baby has good head control and can turn towards something interesting, a more advanced behavioural procedure can be used.

Once your baby has good head control and can turn towards something interesting, a more advanced behavioural procedure can be used. How do we test the hearing of babies and children? An audiologist will select from a range of different tests to evaluate your child s hearing. The choice of test depends on the information that is needed

More information

National Institutes of Health

National Institutes of Health National Institutes of Health Hearing Conservation Program Division of Occupational Health & Safety May 2013 In compliance with 29CFR1910.95, this document provides information and services for the effective

More information

Current practice in health surveillance for noise

Current practice in health surveillance for noise Health and Safety Executive Current practice in health surveillance for noise Prepared by the Health and Safety Laboratory for the Health and Safety Executive 2013 RR966 Research Report Health and Safety

More information

Stanford University. Hearing Conservation Program

Stanford University. Hearing Conservation Program Stanford University Hearing Conservation Program April 2006 Stanford University HEARING CONSERVATION PROGRAM CONTENTS PAGE 1.0 INTRODUCTION...1 1.1 Program Description...1 1.2 Scope...1 2.0 DEFINITIONS

More information

Behavioural Audiometry for Infants and Young Children Whose hearing loss has been detected in infancy

Behavioural Audiometry for Infants and Young Children Whose hearing loss has been detected in infancy Behavioural Audiometry for Infants and Young Children Whose hearing loss has been detected in infancy Alison King, Principal Audiologist, Paediatric Services, Australian Hearing International Paediatric

More information

Smartphone Based Audiometric Test for Confirming the Level of Hearing; Is It Useable in Underserved Areas?

Smartphone Based Audiometric Test for Confirming the Level of Hearing; Is It Useable in Underserved Areas? J Int Adv Otol 2016; 12(1): 61-6 DOI: 10.5152/iao.2016.1421 Original Article Smartphone Based Audiometric Test for Confirming the Level of Hearing; Is It Useable in Underserved Areas? Levent Renda, Ömer

More information

Hearing Protection Standard OSHA Standard 29 CFR 1910.95

Hearing Protection Standard OSHA Standard 29 CFR 1910.95 Hearing Protection Standard OSHA Standard 29 CFR 1910.95 The purpose of this standard is protect employees over exposure to noise and to prevent hearing loss. This standard also states that employees must

More information

2 Sonomax Hearing Healthcare Inc.

2 Sonomax Hearing Healthcare Inc. February 9, 2007 Version 1.2 Originally prepared for the 32 nd Annual Conference of the National Hearing Conservation Association, held on February 15 17, 2007, in Savannah, Georgia, and published in Spectrum,

More information

SAMPSON COMMUNITY COLLEGE

SAMPSON COMMUNITY COLLEGE 1. PURPOSE SAMPSON COMMUNITY COLLEGE HEARING CONSERVATION PROGRAM 1.1 The purpose of this program is to provide guidelines for employees exposed to occupational noise and to comply with the OSHA Occupational

More information

Hearing Devices Policy and Administration Manual

Hearing Devices Policy and Administration Manual Hearing Devices Policy and Administration Manual Assistive Devices Program, Ministry of Health and Long-Term Care Table of Amendments This page will list all substantive changes to policies and procedures

More information

PRODUCT SHEET OUT1 SPECIFICATIONS

PRODUCT SHEET OUT1 SPECIFICATIONS OUT SERIES Headphones OUT2 BNC Output Adapter OUT1 High Fidelity Headphones OUT1A Ultra-Wide Frequency Response Headphones OUT3 see Stimulators OUT100 Monaural Headphone 40HP Monaural Headphones OUT101

More information

E-A-R/Aearo Technologies E-A-RCAL SM Laboratory 7911 Zionsville Road Indianapolis, IN 46268-1657 eberger@compuserve.com

E-A-R/Aearo Technologies E-A-RCAL SM Laboratory 7911 Zionsville Road Indianapolis, IN 46268-1657 eberger@compuserve.com Original April 24, 2007 Revised June 21, 2007 E-A-R/Aearo Technologies E-A-RCAL SM Laboratory 7911 Zionsville Road Indianapolis, IN 46268-1657 eberger@compuserve.com INTRODUCTION The E-A-RFit TM Validation

More information

ONTARIO INFANT HEARING PROGRAM AUDIOLOGIC ASSESSMENT PROTOCOL. Version 3.1, January 2008

ONTARIO INFANT HEARING PROGRAM AUDIOLOGIC ASSESSMENT PROTOCOL. Version 3.1, January 2008 ONTARIO INFANT HEARING PROGRAM AUDIOLOGIC ASSESSMENT PROTOCOL Version 3.1, January 2008 Otologic Function Unit Mount Sinai Hospital, Toronto, Ontario Contact: Martyn Hyde, PhD Voice 416.586.4510 Fax 416.586.8739

More information

8.Audiological Evaluation

8.Audiological Evaluation 8. A U D I O L O G I C A L E V A L U A T I O N 8.Audiological Evaluation The external ear of the child with Progeria Behavioral testing for assessing hearing thresholds Objective electrophysiologic tests

More information

Your Hearing ILLUMINATED

Your Hearing ILLUMINATED Your Hearing ILLUMINATED INFORMATION FROM YOUR HEARING CARE PROFESSIONAL REDISCOVER your hearing and reconnect 1 with the important things you might have been missing. Your sense of hearing is a vital

More information

PRACTICE STANDARDS AND GUIDELINES FOR HEARING ASSESSMENT OF CHILDREN BY AUDIOLOGISTS

PRACTICE STANDARDS AND GUIDELINES FOR HEARING ASSESSMENT OF CHILDREN BY AUDIOLOGISTS PRACTICE STANDARDS AND GUIDELINES FOR HEARING ASSESSMENT OF CHILDREN BY AUDIOLOGISTS 5060-3080 Yonge Street, Box 71 Toronto, Ontario M4N 3N1 416-975-5347 1-800-993-9459 www.caslpo.com Approved: March 2008

More information

Sample Written Program. For HEARING CONSERVATION

Sample Written Program. For HEARING CONSERVATION Sample Written Program For HEARING CONSERVATION 1 HEARING CONSERVATION PROGRAM Co Name Prepared by Date I. PURPOSE This purpose of this hearing conservation program is to prevent occupational hearing loss

More information

COMPENSATION FOR NOISE INDUCED HEARING LOSS

COMPENSATION FOR NOISE INDUCED HEARING LOSS - COMPENSATION FOR NOISE INDUCED HEARING LOSS Of the lesser known health hazards in the mines, noise and resulting noise-induced hearing loss (NIHL) have received little attention in the media, despite

More information

Equinox 2.0 PC-based audiometer. Not just an audiometer - A full counseling and reporting solution

Equinox 2.0 PC-based audiometer. Not just an audiometer - A full counseling and reporting solution Equinox 2.0 PC-based audiometer Not just an audiometer - A full counseling and reporting solution The PC-based Advantage The Equinox 2.0 is a state of the art two channel clinical audiometer. Counseling

More information

OAK GROVE SCHOOL DISTRICT. Hearing Conservation Program

OAK GROVE SCHOOL DISTRICT. Hearing Conservation Program OAK GROVE SCHOOL DISTRICT Hearing Conservation Program CONTENTS PAGE 1.0 INTRODUCTION...1 1.1 Program Description...1 1.2 Scope...1 2.0 DEFINITIONS AND ACRONYMS...2 3.0 RESPONSIBILITIES...3 3.1 Supervisor/Department...3

More information

Effects of Noise Attenuation Devices on Screening Distortion Product Otoacoustic Emissions in Different Levels of Background Noise

Effects of Noise Attenuation Devices on Screening Distortion Product Otoacoustic Emissions in Different Levels of Background Noise Effects of Noise Attenuation Devices on Screening Distortion Product Otoacoustic Emissions in Different Levels of Background Noise Kelsey Nielsen, Au.D. Otolaryngology Associates, P.C. Fairfax, Virginia

More information

Loss Control TIPS Technical Information Paper Series

Loss Control TIPS Technical Information Paper Series Loss Control TIPS Technical Information Paper Series Innovative Safety and Health Solutions SM Introduction Occupational Noise Exposure and Hearing Conservation Noise, or unwanted sound, is one of the

More information

Who are you? Six Components of Hearing Conservation Program. Hearing is Precious!

Who are you? Six Components of Hearing Conservation Program. Hearing is Precious! Six Components of Hearing Conservation Program Helene R. Freed, Ed.M Public Relations Specialist Industrial Hearing Testing Who are you? Are you a student? Do you work in industry in H&S? Do you work in

More information

Environmental Health and Safety* 2809 Daley Drive * Ames, Iowa 50011-3660 * (515) 294-5359 * www.ehs.iastate.edu

Environmental Health and Safety* 2809 Daley Drive * Ames, Iowa 50011-3660 * (515) 294-5359 * www.ehs.iastate.edu Environmental Health and Safety* 2809 Daley Drive * Ames, Iowa 50011-3660 * (515) 294-5359 * www.ehs.iastate.edu Revision 2 October 2008 University Nondiscrimination Statement Iowa State University does

More information

UC Santa Barbara Hearing Conservation Program Manual

UC Santa Barbara Hearing Conservation Program Manual UCSB Hearing Conservation Program Manual Page 1 of 15 UC Santa Barbara Hearing Conservation Program Manual Rev. August 2014 Program Manager: Nick Nieberding Title: Industrial Hygiene Specialist Email:

More information

REGULATIONS FOR THE DEGREE OF MASTER OF SCIENCE IN AUDIOLOGY (MSc[Audiology])

REGULATIONS FOR THE DEGREE OF MASTER OF SCIENCE IN AUDIOLOGY (MSc[Audiology]) 224 REGULATIONS FOR THE DEGREE OF MASTER OF SCIENCE IN AUDIOLOGY (MSc[Audiology]) (See also General Regulations) Any publication based on work approved for a higher degree should contain a reference to

More information

Psychometrically Equivalent Korean Bisyllabic Words Spoken by Male and Female Talkers

Psychometrically Equivalent Korean Bisyllabic Words Spoken by Male and Female Talkers Psychometrically Equivalent Korean Bisyllabic Words Spoken by Male and Female Talkers Richard W. Harris, Eunoak Kim and Dennis L. Eggett (Brigham Young University, Provo, Utah, USA) Richard W. Harris,

More information

C HAPTER T HIRTEEN. Diagnosis and Treatment of Severe High Frequency Hearing Loss. Susan Scollie and Danielle Glista

C HAPTER T HIRTEEN. Diagnosis and Treatment of Severe High Frequency Hearing Loss. Susan Scollie and Danielle Glista C HAPTER T HIRTEEN Diagnosis and Treatment of Severe High Frequency Hearing Loss Susan Scollie and Danielle Glista Providing audible amplified signals for listeners with severe high frequency hearing loss

More information

Generating Element. Frequency Response. Polar Pattern. Impedance. Output Level

Generating Element. Frequency Response. Polar Pattern. Impedance. Output Level Generating Element Frequency Response Polar Pattern Impedance Output Level Weight Pro 7: Dynamic Pro 7 ic: ic Electret Pro 7: 100 Hz - 12 khz Pro 7 ic: 35 Hz - 12 khz Pro 7: Heil Cardioid Pro 7 ic: Cardioid

More information

Operating Instructions Detective Audiometer

Operating Instructions Detective Audiometer Maico Diagnosticsi7625 Golden Triangle DriveiEden Prairie, MN 55344iToll Free 888.941.4201 Maico Diagnostic GmbHiSalzufer 13/14i10587 Berlin, GermanyiTelephone ++030 70 71 46 50 TABLE OF CONTENTS WARRANTY...1

More information

Hearing Test Results from 2009 Audio Engineering Society Convention Testing funded by Shure

Hearing Test Results from 2009 Audio Engineering Society Convention Testing funded by Shure Page 1 of 8 Hearing Test Results from 29 Audio Engineering Society Convention Testing funded by Shure SUMMARY The 127 th Audio Engineering Society (AES) convention was held in New York, New York from October

More information

The Effects of Ultrasonic Sound Generated by Ultrasonic Cleaning Systems on Human Hearing and Physiology

The Effects of Ultrasonic Sound Generated by Ultrasonic Cleaning Systems on Human Hearing and Physiology The Effects of Ultrasonic Sound Generated by Ultrasonic Cleaning Systems on Human Hearing and Physiology Questions about the effects of ultrasonic energy on hearing and other human physiology arise from

More information

HEARING & HEARING LOSS. Dr I Butler 2015

HEARING & HEARING LOSS. Dr I Butler 2015 HEARING & HEARING LOSS Dr I Butler 2015 DISCLOSURE Sponsorship to attend local and international workshops Cochlear (Southern ENT) Med el TOPICS Anatomy Classification of hearing loss Congenital hearing

More information

NOISE INDUCED HEARING LOSS AMONG INDUSTRIAL WORKERS IN DAR ES SALAAM. B.M. MINJA, N.H. MOSHI and P. RIWA ABSTRACT

NOISE INDUCED HEARING LOSS AMONG INDUSTRIAL WORKERS IN DAR ES SALAAM. B.M. MINJA, N.H. MOSHI and P. RIWA ABSTRACT 238 EAST AFRICAN MEDICAL JOURNAL June 2003 East African Medical Journal Vol. 80 No. 6 June 2003 NOISE INDUCED HEARING LOSS AMONG INDUSTRIAL WORKERS IN DAR ES SALAAM B.M. Minja MBChB (EA), MMed (Surgery)

More information

Active Noise Reduction (ANR) in Hearing Protection: Does it Make Sense for Industrial Applications? E-A-R 01-16/HP. E. H. Berger, M.S.

Active Noise Reduction (ANR) in Hearing Protection: Does it Make Sense for Industrial Applications? E-A-R 01-16/HP. E. H. Berger, M.S. Active Noise Reduction (ANR) in Hearing Protection: Does it Make Sense for Industrial Applications? E-A-R 01-16/HP E. H. Berger, M.S. E A R / Aearo Company E A RCAL SM Laboratory 7911 Zionsville Road Indianapolis,

More information

ACOUSTICAL CONSIDERATIONS FOR EFFECTIVE EMERGENCY ALARM SYSTEMS IN AN INDUSTRIAL SETTING

ACOUSTICAL CONSIDERATIONS FOR EFFECTIVE EMERGENCY ALARM SYSTEMS IN AN INDUSTRIAL SETTING ACOUSTICAL CONSIDERATIONS FOR EFFECTIVE EMERGENCY ALARM SYSTEMS IN AN INDUSTRIAL SETTING Dennis P. Driscoll, P.E. and David C. Byrne, CCC-A Associates in Acoustics, Inc. Evergreen, Colorado Telephone (303)

More information

HEARING SCREENING: PURE TONE AUDIOMETRY

HEARING SCREENING: PURE TONE AUDIOMETRY HEARING SCREENING: PURE TONE AUDIOMETRY QUALIFIED SCREENERS 7-005.01 For the purposes of the school officials verifying that a qualified screener is carrying out the required screening activity, the qualified

More information

Australian Government Hearing Services Program. Hearing Rehabilitation Outcomes for Voucher-Holders

Australian Government Hearing Services Program. Hearing Rehabilitation Outcomes for Voucher-Holders Australian Government Hearing Services Program Hearing Rehabilitation Outcomes for Voucher-Holders 1 July 2015 Introduction The aim of the voucher component of the Australian Government s Hearing Services

More information

Noise: Impact on Hearing; Regulation

Noise: Impact on Hearing; Regulation Noise: Impact on Hearing; Regulation EOH 466A Fall 2008 Mechanism of Hearing Sound waves collected, focused by the outer ear. Humans have little control over muscles in outer ear. Many animals have the

More information

A new method of partial deafness treatment

A new method of partial deafness treatment Signature: Med Sci Monit, 3; 9(4): CS-24 PMID: 129676 WWW.MEDSCIMONIT.COM Case Study Received: 2.12.5 Accepted: 3.2. Published: 3.4.23 A new method of partial deafness treatment Henryk Skarżyński, Artur

More information

Binaural hearing ability with mastoid applied bilateral bone conduction stimulation in normal hearing subjects

Binaural hearing ability with mastoid applied bilateral bone conduction stimulation in normal hearing subjects Binaural hearing ability with mastoid applied bilateral bone conduction stimulation in normal hearing subjects Stefan Stenfelt and Mehrnaz Zeitooni Linköping University Post Print N.B.: When citing this

More information

If you do not use the calculator-generated text, you MUST notify the Rating Job Aids mailbox. Please describe the error in detail.

If you do not use the calculator-generated text, you MUST notify the Rating Job Aids mailbox. Please describe the error in detail. HEARING LOSS CALCULATOR USER GUIDE HL Calculator v4.7 Index Hearing Loss Calculator.....3 How to Navigate the Hearing Loss Calculator User Guide.4 Hearing Loss Calculator - Starting the Program..5 New

More information

Audiometry and Hearing Loss Examples

Audiometry and Hearing Loss Examples Audiometry and Hearing Loss Examples An audiogram shows the quietest sounds you can just hear. The red circles represent the right ear and the blue crosses represent the left ear. Across the top, there

More information

HEARING AID REGULATION 196/2001 [Authority Repealed]

HEARING AID REGULATION 196/2001 [Authority Repealed] PDF Version [Printer-friendly - ideal for printing entire document] HEARING AID REGULATION 196/2001 [Authority Repealed] Published by Quickscribe Services Ltd. Updated To: [deposited August 24, 2001] Important:

More information

Occupational Noise Induced Hearing Loss: Final Program Policy Decision and Supporting Rationale

Occupational Noise Induced Hearing Loss: Final Program Policy Decision and Supporting Rationale Occupational Noise Induced Hearing Loss: Final Program Policy Decision and Supporting Rationale October 2014 1 I Introduction: In September 2012, the WCB Board of Directors added Noise Induced Hearing

More information