POLICY AND PROCEDURE

Size: px
Start display at page:

Download "POLICY AND PROCEDURE"

Transcription

1 PURPOSE: To provide guidelines For using the LifePak 20 Defibrillator/Monitor with ADAPITIV Biphasic technology. This device may be used for AED defibrillation, manual defibrillation, synchronized cardioversion, noninvasive pacing and Sp0 2 monitoring. Policy: Only Physician s, PA s, RN s & ARNP s who have been trained and who have demonstrated competency in the operation of the LifePak 20 Defibrillator/Monitor/Pacemaker will be allowed to use the equipment. PROCEDURE: A. Automated External Defibrillator (AED) function 1. Assess patient for unresponsiveness, lack of pulse and respirations. 2. Call for help. 3. Start all necessary BLS/CPR steps following current guidelines. 4. Place LifePak 20 near patient. 5. Turn on LifePak 20, by pressing "Green On" button, keep AED door closed at this time. 6. Remove all clothing from the patient s chest. Remove excessive hair, or medication patch if necessary, ensure that chest is clean and dry. 7. Remove Quick-Combo electrodes from package. 8. Attach Quick-Combo electrodes to patient defibrillation cable on LifePak Place Quick-Combo electrodes on patient s chest, ensuring electrodes are placed correctly. Anterior-Lateral "Quik-Combo" Electrode Placement allows for ECG monitoring, Defibrillation, Synchronized Cardioversion and Noninvasive Pacing. A. Place the heart electrode lateral to patient s left nipple in the midaxillary line, with the center of the electrode in the midaxillary line, if possible. B. Place anterior electrode on the patient s right upper torso, lateral to the sternum and below the clavicle. SUPERCEDES: 4/96, 6/99, 6/01, 4/06, 12/07 Page 1 of 8

2 Note: Follow diagrams on "Quik-Combo" electrode pads. Anterior-Posterior Electrode Placement: Is an alternative position for Noninvasive Pacing, Manual Defibrillation and Cardioversion, but NOT for ECG monitoring or Automated Defibrillation. The ECG signal obtained through electrodes in this position is not a standard lead. Use this configuration if Anterior-Lateral placement is not possible. A. Place the heart electrode over the patient s left precardium. The upper edge of the electrode should be just below the nipple. Avoid placement over the nipple, the diaphragm, or the bony prominence of the sternum if possible. B. Place the posterior electrode behind the heart in the infrascapular area. Note: For patient comfort, place the cable away from the spine. Do not place the electrode over the bony prominences of the spine or scapula. Note: Be sure to place 3 lead ECG electrodes on patient ASAP After "Quik-Combo" pads have been placed. 10. When electrodes have been properly placed, press "Yellow Analyze" button. Ensure no one is touching the patient or bed while AED analyzes the rhythm. 11. If a shock is advised, AED will automatically charge to 200 joules and a voice prompt to clear the patient and push the shock button will be stated. 12. Prior to pushing the "RED Shock" button repeat mnemonic: "Oxygen clear, I'm Clear, You're Clear, Every Body Clear" visually check that no one is touching the patient or the bed and the 0 2 is clear. Call out "Shocking" and press the "RED Shock" button. 13. Immediately following the shock the AED will prompt you to Start CPR. DO NOT push analyze at this time as 2005 ACLS/BLS guidelines require 2 minutes of CPR prior to next defibrillation attempt. Perform 2 minutes of CPR the AED will countdown from 2 minutes and will prompt you to push analyze at the end of 2 minutes of CPR. SUPERCEDES: 4/96, 6/99, 6/01, 4/06, 12/07 Page 2 of 8

3 14. After 2 minutes of chest compressions and ventilations STOP CPR If physician/pa/nurse practitioner not present to interpret ECG rhythm, press the yellow analyze button clear the patient while AED analyzes the rhythm. AED will prompt for defibrillation or state "No Shock Advised". If shock is advised AED will automatically charge to 300 joules. Repeat steps 12 & 13 above. 15. Continue with Automated External Defibrillation as described above until a physician/pa/nurse practitioner arrives and assumes control of the code. When physician/pa/nurse practitioner assumes command then take Lifepak 20 out of AED mode by opening the AED door = push the Grey manual door latch located at the bottom left of the door with your thumb and pull door open, or by pressing the "Energy Select" button if the door had already been opened. Defibrillator may now be used in manual (ACLS) mode and manual defibrillation may be performed on physician/pa/nurse practitioners order. 16. For manual defibrillation in ACLS (manual) mode, press "Energy Select" button to activate ability to select joules for defibrillation. Rotate the speed dial button to highlight desired joules, press speed dial button to select desired joules, press "Yellow Charge" button to charge LifePak 20. Follow step 12 above to defibrillate. 17. Repeat steps 14-16, above if further defibrillation is ordered by physician/pa/nurse practitioner. 18. Press CODE SUMMARY button at the end of code to print summary of events. Note: Immediately after the patient is taken off our LifePak 20 (when Patient transferred out of the area), Be sure to print 2 full code summary strips and staple one strip to the CPR record placed in patient's chart. Staple the second strip to the photocopy of the CPR record that is sent to Sylvester Performance Improvement Office. See code Blue policy & Procedure Code # 166 for details on paperwork distribution. NOTE: Be sure to turn the LifePak 20 OFF immediately after final Code Summary strip has completed printing. DO NOT leave the machine ON as it continues to record the event and will show on the patient s medical record. SPECIAL NOTE: After All Code Blue Tasks Are Complete And Patient Is Transferred, The CHARGE NURSE MUST Call Both Biomedical Engineering at ( or ) And The Education Department At ( or ) For Download Of Code Information From The LifePak 20 Defibrillator. The Patient Information Contained In the Memory may be lost after 24 to 48 hours so must be downloaded ASAP after the code. SUPERCEDES: 4/96, 6/99, 6/01, 4/06, 12/07 Page 3 of 8

4 B. Standard Hard Paddles Hands On" Defibrillation 1. Remove the "Quik Combo" Patient defibrillator cable from the front of the LifePak 20 Defibrillator by gently rotating the locking ring on the therapy cable in the direction of the white arrow (counterclockwise) until it stops. Gently pull out the cable. 2. Attach "Hands On" standard hard paddle cable by ensuring white arrow is facing upward and cable cord faces to the right. Gently push connector into receptacle until it snaps into place. 3. If other than 200 joules is desired, press Energy Select button and select the energy to be delivered as described in step 16 above. 4. Remove standard hard paddles from tray on top of LifePak 20 and apply defibrillation gel over entire paddle electrode surface. Never rub the two paddles metal surfaces together as damage to the metal surfaces will occur. 5. Place defibrillator paddles firmly on patient s chest. Standard Paddle Placement A. STERNUM paddle to the patient s right upper torso, lateral to the sternum and below the clavicle. B. APEX paddle lateral to the patient s left nipple in the midaxillary line, with the center of the paddle electrode in the midaxillary line, if possible. 6. Press CHARGE button on APEX paddle. 7. Prior to pushing the "RED Shock" buttons repeat mnemonic: "Oxygen clear, I'm Clear, You're Clear, Every Body Clear" visually check that no one is touching the patient or the bed and the 0 2 is clear. Call out "Shocking" and press both of the "RED Shock" buttons on the paddles simultaneously to defibrillate patient. If shock buttons are not pressed within 60 seconds, stored energy is internally removed. 8. Repeat steps 3 8, if further defibrillation is required as indicated by physician/practitioner. C. External Pacer 1. Press "Green On" button on LifePak Connect 3 ECG electrodes to ECG cable and apply electrodes to patient. 3. Select Lead I, II, or III to monitor. 4. Connect Quick-Combo electrode pads to LifePak 20 and attach to patient s chest. See page 1 #9 for Quik-Combo pad placement. SUPERCEDES: 4/96, 6/99, 6/01, 4/06, 12/07 Page 4 of 8

5 5. Press PACER ON button. Confirm the LED illuminates, indicating that the power is on. 6. Demand pacing message appears in the lower left corner of the monitor screen. 7. Observe the EKG rhythm on monitor screen, confirm that a triangle sense marker appears near the middle of each QRS complex. If sense markers do not appear or are displayed in the wrong location (for example, on the T-wave), select another lead. (It is normal for the sense marker location to vary slightly on each QRS complex.) 8. Select desired pacing rate by pressing the rate button rotate the speed dial to select desired rate. Push speed dial to set rate. Note: The rate button changes the rate in 10 pulse per minute (ppm) increments; the Speed Dial changes the rate in 5 ppm increments. 9. Activate pacing by pressing the current button then slowly rotate speed dial clockwise to increase current until electrical capture occurs. For each delivered pacing stimulus, the pacer indicator flashes off and a positive pace marker displays on the ECG waveform. Note: Increase current slowly to minimize patient discomfort. 10. Observe monitor screen for evidence of electrical pacing capture. Palpate the patients pulse, check blood pressure and compare the Sp02 pulse rate with the set pacing rate to assess for mechanical capture. Consider the use of sedation analgesia if patient is uncomfortable. Note: The current button changes the current in 10 ma increments; the Speed Dial changes the current in 5 ma increments. Use the Speed dial when patient is awake and can feel increment changes, to reduce patient discomfort. 11. To interrupt pacing and view the patient's intrinsic rhythm, press and hold the "Pause" button. This causes the pacer to pace at 25% of the set rate. Release the "Pause" button to resume pacing at the set rate. Avoid holding pause for extended length of time as patient's perfusion will decrease at lower pacing rate. 12. To stop pacing, reduce current to zero or press "Pacer" button. Note: Pacing will automatically stop when the defibrillator is charged or when the energy select button is pushed. SUPERCEDES: 4/96, 6/99, 6/01, 4/06, 12/07 Page 5 of 8

6 D. Synchronized Cardioversion 1. Press "Green On" button. 2. Place ECG and Quik-Combo electrodes on patient as previously described. 3. Select lead with optimum QRS complex amplitude. 4. Press SYNC button. Confirm that the sync LED blinks with each detected QRS complex. 5. Observe monitor screen. Confirm that triangle sense marker appears near the middle of each QRS complex. If the sense markers do not appear or are displayed in the wrong locations (For example on the T-wave), select another lead or reposition the ECG electrodes so SYNC markers occur only on the QRS complex. (It is normal for the sense marker location to vary slightly on each QRS complex.) 6. Press Energy Select button and rotate the speed dial button to highlight desired joules, press speed dial button to select desired joules then Press Yellow Charge button. 7. Shout Oxygen Clear, I m Clear, You're Clear, Everybody Clear when charging is complete and make sure all personnel are clear of patient and bed. 8. Press and HOLD Red Shock button(s) until discharge occurs with next detected QRS complex and then release "Red Shock" button(s). If shock button(s) are not pressed within 60 seconds, stored energy is internally removed. 9. Observe patient and monitor screen. If synchronized cardioversion needs to be reattempted, repeat steps 4 9. Note: When cardioverting you must hold the "Red Shock" button until the shock is delivered, the LifePak 20 will only deliver the shock at the appropriate time on the QRS complex to avoid the R on T phenomenon. E. Sp02 Monitoring Procedure 1. Power to the pulse oximeter is controlled by the defibrillator. When the Defibrillator is turned on; the oximeter turns on and performs a self-test that takes up to 10 seconds. 2. To conserve battery power, the pulse oximeter goes into sleep mode when not in use. Sleep mode is activated within 10 seconds of disconnecting the sensor. The oximeter will return to normal mode after detecting a sensor or a patient SUPERCEDES: 4/96, 6/99, 6/01, 4/06, 12/07 Page 6 of 8

7 signal. The oximeter performs the self test when it returns from sleep mode to active mode. During sleep mode, the screen does not display Sp02 information. 3. The pulse oximeter measures Sp02 levels between 1% and 100%. When Sp02 levels are between 70% and 100%, oximeter measurements are accurate from + 3 digits. Connect the Sp02 cable to the monitor. Attach the sensor to the Sp02 cable and the patient. Press "Green On" button. Observe the pulse bar for fluctuation. Amplitude of the pulse bar indicates relative signal strength. Adjust sensitivity, averaging time, and Sp02 volume as necessary. 4. The Sp02 waveform can be displayed on waveform channel 2 by selecting waveform channel 2 and then selecting Sp02 from the waveform menu. The Sp02 waveform automatically sizes itself to provide optimum waveform viewing. Note: For more information on Sp02 Volume, Sensitivity and Averaging Time please see Medtronic LifePak 20 Operating Instructions Manual, page F. Documentation 1. Be sure to print 2 full code summary strips and attach one to the CPR record placed in patient's chart. Attach the second to the photocopy of the CPR record that is sent to Sylvester Performance Improvement Office. See code Blue policy & Procedure Code # 166 for details on paperwork distribution. 2. Document events of code on CPR record and progress note as needed. G. Daily LifePak 20 Defibrillator/Monitor User Test 1. Defibrillator/monitor user test must be performed at the beginning of each shift by assigned personnel. 2. Steps to perform Defibrillator/monitor test: Check Printer Strip for Successful 03:00am Self Test and unit was plugged into AC power. Unplug Defibrillator Before User Test. Press "Green On" Button. Do Not Push Analyze Open the AED door to convert LifePak 20 to ACLS (manual) mode and stop the analyze prompt. SUPERCEDES: 4/96, 6/99, 6/01, 4/06, 12/07 Page 7 of 8

8 Press Options Button. Rotate Speed DIAL To Highlight User Test. Push Speed Dial to Select. Rotate Speed DIAL To Highlight YES. Push Speed Dial to Select. Close The DOOR. Allow User Test to Run Wait For Test Strip To Say User Test Successful. Defibrillator Turns Self Off. Plug The Defibrillator Back Into AC Power. H. Biomedical Engineering Bi-Annual preventative maintenance (PM) tests 1. Biomedical Engineering will perform PM test on all LifePak 20 Defibrillator/Monitors every 6 months. Test and documentation to include verification of proper function of standard hands on hard paddles. REFERENCE: - Medtronic LifePak 20 Defibrillator/Monitor with ADAPTIV Biphasic Technology Operating Instructions Manual - AHA Guidelines Sylvester Code Blue Policy # 166 SUPERCEDES: 4/96, 6/99, 6/01, 4/06, 12/07 Page 8 of 8

R Series Quick Reference Guide

R Series Quick Reference Guide R Series Quick Reference Guide 2010 ZOLL Medical Corporation. All rights reserved. CPR Index, OneStep, R Series, Real CPR Help, See-Thru CPR, statpadz, and ZOLL are trademarks or registered trademarks

More information

HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3

HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3 3 How to Use the LIFEPAK 1000 Defibrillator HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3 This section provides an overview of information and instructions for using the LIFEPAK 1000 defibrillator. Modes

More information

Heart of America Medical Center EMR, EMT, EMT-Intermediate & AEMT Protocols

Heart of America Medical Center EMR, EMT, EMT-Intermediate & AEMT Protocols USE OF THE LUCAS DEVICE TO PROVIDE EXTERNAL CHEST COMPRESSIONS IN PATIENTS SUFFERING CARDIAC ARREST Purpose This procedure describes the appropriate methods to apply, operate, and discontinue the LUCAS

More information

LIFEPAK 15 MONITOR/DEFIBRILLATOR OPERATING INSTRUCTIONS

LIFEPAK 15 MONITOR/DEFIBRILLATOR OPERATING INSTRUCTIONS LIFEPAK 15 MONITOR/DEFIBRILLATOR OPERATING INSTRUCTIONS LIFEPAK 15 MONITOR/DEFIBRILLATOR OPERATING INSTRUCTIONS Important Information!USA Rx Only!USA Device Tracking The U.S. Food and Drug Administration

More information

LIFEPAK 15 MONITOR/DEFIBRILLATOR

LIFEPAK 15 MONITOR/DEFIBRILLATOR LIFEPAK 15 MONITOR/DEFIBRILLATOR Setup Options This document describes how to enter Setup mode and change operating settings in the LIFEPAK 15 monitor/defibrillator. Defining Setup Options Setup mode

More information

LIFEPAK 1000 DEFIBRILLATOR OPERATING INSTRUCTIONS

LIFEPAK 1000 DEFIBRILLATOR OPERATING INSTRUCTIONS LIFEPAK 1000 DEFIBRILLATOR OPERATING INSTRUCTIONS LIFEPAK 1000 DEFIBRILLATOR OPERATING INSTRUCTIONS Important This instrument is to be used by authorized personnel only.!usa Rx Only!USA Device Tracking

More information

19.109 AUTOMATED EXTERNAL DEFIBRILLATOR

19.109 AUTOMATED EXTERNAL DEFIBRILLATOR 19.109 AUTOMATED EXTERNAL DEFIBRILLATOR References: American Heart Association Healthcare Provider (CPR/AED) Ohio Revised Code Chapter 2305.235 Immunity of Person Involved with Providing Automated External

More information

Part 2. Copyright 1998 Philips Consumer Communications L.P. All rights reserved. Printed in Mexico. Issue 1AT&T 848229506

Part 2. Copyright 1998 Philips Consumer Communications L.P. All rights reserved. Printed in Mexico. Issue 1AT&T 848229506 2 User's Manual for Two-Line Digital Answering System Telephone with Speakerphone 1872 Fold open this paper for information about this telephone's installation and operation. Please read Part 1 Important

More information

HP Pavilion All-in-One MS200 series PC. Upgrading and Servicing Guide. Printed in

HP Pavilion All-in-One MS200 series PC. Upgrading and Servicing Guide. Printed in HP Pavilion All-in-One MS200 series PC *579907-001* *579907-001* Printed in Upgrading and Servicing Guide Replacing a Wireless Keyboard or Mouse...2 Before You Begin... 2 Replacing the Keyboard or Mouse...

More information

Procedure 17: Cardiopulmonary Resuscitation

Procedure 17: Cardiopulmonary Resuscitation Cardiopulmonary Resuscitation 349 Procedure 17: Cardiopulmonary Resuscitation Introduction Cardiopulmonary arrest (CPA) occurs when a patient s heart and lungs stop functioning. In children, CPA usually

More information

Innovative solutions in resuscitation therapy

Innovative solutions in resuscitation therapy Multifunction Defibrillator Pads Product information Innovative solutions in resuscitation therapy With pads for emergency use, monitoring, as well as special procedures, there are Pads sized and designed

More information

Automated External Defibrillators (AEDs) AP 400.37

Automated External Defibrillators (AEDs) AP 400.37 Administrative Procedure Automated External Defibrillators (AEDs) AP 400.37 Procedure for: Principals, Maintenance and Custodial Staff Adopted: April 12, 2013 Submitted by: Superintendent of Business &

More information

510 / 511 Cardiorespiratory Monitor Preventative Maintenance Test Procedure

510 / 511 Cardiorespiratory Monitor Preventative Maintenance Test Procedure CAS MEDICAL SYSTEMS, INC. 510 / 511 Cardiorespiratory Monitor Preventative Maintenance Test Procedure using the 2301Multi-Function Tester 44 East Industrial Road Branford, CT 06405 USA www.techsrv@casmed.com

More information

Defibtech DDU-100 Semi-Automatic External Defibrillator

Defibtech DDU-100 Semi-Automatic External Defibrillator Defibtech DDU-100 Semi-Automatic External Defibrillator User Manual Software V2.0 U S E O N LY I F PA T I E N T I S : U N C O N S C I O U S U N R E S P O N S I V E N O T B R E A T H I N G 1 PUSH "ON"

More information

MAINTENANCE & TROUBLESHOOTING

MAINTENANCE & TROUBLESHOOTING MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens replace the projection lamp replace the batteries in the remote use the security lock feature troubleshoot the projector Cleaning

More information

Critical Care Therapy and Respiratory Care Section

Critical Care Therapy and Respiratory Care Section Critical Care Therapy and Respiratory Care Section Category: Clinical Section: Clinical Monitoring Title: Overnight Pulse Oximetry Policy #: 08 Revised: 03/00 1.0 DESCRIPTION 1.1 Definition 1.1.1 A pulse

More information

Transcutaneous Electrical Nerve Stimulation Device. Operation Manual Read Before Using

Transcutaneous Electrical Nerve Stimulation Device. Operation Manual Read Before Using Transcutaneous Electrical Nerve Stimulation Device Operation Manual Read Before Using TABLE OF CONTENTS INTRODUCTION TO TENS INDICATIONS AND CONTRAINDICATIONS WARNINGS PRECAUTIONS/ADVERSE REACTIONS ABOUT

More information

5CHAPTER. Setting Up the Polling Place. IMPORTANT: The electronic poll book MUST be set up first and you MUST begin the update by 5:15 am.

5CHAPTER. Setting Up the Polling Place. IMPORTANT: The electronic poll book MUST be set up first and you MUST begin the update by 5:15 am. 5CHAPTER Setting Up the Polling Place Election Day - 5:00 AM TO 6:00 AM 14 IMPORTANT: The electronic poll book MUST be set up first and you MUST begin the update by 5:15 am. n All judges of election and

More information

ACCESSORIES LIFEPAK DEFIBRILLATOR/MONITOR. for use with the. Genuine accessories from Physio-Control

ACCESSORIES LIFEPAK DEFIBRILLATOR/MONITOR. for use with the. Genuine accessories from Physio-Control ACCESSORIES for use with the LIFEPAK 20 DEFIBRILLATOR/MONITOR Genuine accessories from Physio-Control POWER OPTIONS Thank you for choosing Physio-Control as your partner in helping you save lives and improve

More information

ACCESSORIES LIFEPAK 20 DEFIBRILLATOR/MONITOR SERIES. Genuine accessories from Physio-Control

ACCESSORIES LIFEPAK 20 DEFIBRILLATOR/MONITOR SERIES. Genuine accessories from Physio-Control ACCESSORIES LIFEPAK 20 DEFIBRILLATOR/MONITOR SERIES Genuine accessories from Physio-Control POWER OPTIONS Thank you for choosing Physio-Control as your partner in helping you save lives and improve patient

More information

INSTRUCTIONS FOR USE POWERHEART AUTOMATED EXTERNAL DEFIBRILLATOR G3 PLUS 9390A AND 9390E 112-0038-201 A

INSTRUCTIONS FOR USE POWERHEART AUTOMATED EXTERNAL DEFIBRILLATOR G3 PLUS 9390A AND 9390E 112-0038-201 A INSTRUCTIONS FOR USE POWERHEART AUTOMATED EXTERNAL DEFIBRILLATOR G3 PLUS 9390A AND 9390E 112-0038-201 A Notice of Rights 2010 Cardiac Science Corporation. All rights reserved. No part of this documentation

More information

MAINTENANCE & TROUBLESHOOTING

MAINTENANCE & TROUBLESHOOTING MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens clean the fan intake filter replace the projection lamp replace the batteries in the remote control use the Kensington lock feature

More information

Section III. Heart Mate II (2 nd Generation) III.41.1 EMT-BASIC EMT-INTERMEDIATE EMT-PARAMEDIC

Section III. Heart Mate II (2 nd Generation) III.41.1 EMT-BASIC EMT-INTERMEDIATE EMT-PARAMEDIC EMT-BASIC EMT-INTERMEDIATE EMT-PARAMEDIC A Ventricular Assist Device, or VAD, is a mechanical device used to substantially support circulation in a patient with significant cardiac ventricular dysfunction.

More information

Automated External Defibrillator

Automated External Defibrillator Automated External Defibrillator 1. INTRODUCTION An automated external defibrillator (AED) is used to treat victims who experience sudden cardiac arrest. It must only be applied to victims who are unconscious,

More information

1000+ AUDIOMETER OPERATING INSTRUCTIONS

1000+ AUDIOMETER OPERATING INSTRUCTIONS 1000+ AUDIOMETER OPERATING INSTRUCTIONS AMBCO Model 1000+ is a microprocessor controlled pure tone air conduction audiometer with automated screening test feature. INDEX A. Getting Started Patient Instructions

More information

Kiva Technology User s Manual

Kiva Technology User s Manual 2009 Kiva Technology User s Manual For Support Contact Call Center (816) 235-2000: Monday Friday 7:00 am 7:00 pm ILE Support E-Mail: ILESupport@UMKC.EDU URL: http://www.umkc.edu/is/cts/ilesupport/ctssupporttext.asp

More information

>DEFIBRILLATION AEDs (Automatic External Defibrillators)

>DEFIBRILLATION AEDs (Automatic External Defibrillators) > AEDs (Automatic External Defibrillators) > AED 10 124 > AED 20 126 Monitors/Defibrillators > PIC 30 128 > PIC 40 128 > PIC 50 129 > PIC Add-ons/Accessories 129 124 AEDs AED 10 Designed for easy operation

More information

Instructions For Use AED Trainer

Instructions For Use AED Trainer Instructions For Use AED Trainer PART NUMBER Copyright 2007 Cardiac Science Corporation. All rights reserved. The Automated External Defibrillator (AED) Trainer is a device for teaching basic AED rescue

More information

Connecting to the Internet

Connecting to the Internet Connecting to the Internet Connecting the Gateway to the Internet Configuration of TCP/IP Protocol Installing the USB Drivers Connecting Ethernet Network Devices Connecting USB Network Devices Connecting

More information

LIFEPAK 20e DEfibrillator/MONITOR. Works like you work.

LIFEPAK 20e DEfibrillator/MONITOR. Works like you work. LIFEPAK 20e DEfibrillator/MONITOR Works like you work. When cardiac arrest occurs, the clock starts ticking. Respond quickly with confidence. LIFEPAK 20e defibrillator/monitor. Two defibrillators in one.

More information

Features of Your Cisco Unified IP Phone

Features of Your Cisco Unified IP Phone Cisco Unified IP Phone 8961, page 1 Cisco Unified IP Phone 9951, page 9 Cisco Unified IP Phone 9971, page 18 General Phone Information, page 27 Cisco Unified IP Phone 8961 The following sections describe

More information

PATIENT CARE SERVICES POLICY AND PROCEDURE

PATIENT CARE SERVICES POLICY AND PROCEDURE PATIENT CARE SERVICES POLICY AND PROCEDURE Title: within CHN Facilities Section: Provision of Care, Treatment, and Services Dept. of Origin: Patient Care Services Effective Date: February 02, 1999 Last

More information

Operating Manual Color Changing Fountains

Operating Manual Color Changing Fountains Operating Manual Color Changing Fountains 1-877-80-PONDS www.atlanticwatergardens.com Introduction Thank you for purchasing Atlantic s Color Changing Fountains overflowing features that glow from within.

More information

www.nanoomtech.co.kr ( (

www.nanoomtech.co.kr ( ( ( www.nanoomtech.co.kr ( 4 minutes start brain damage 6 minutes brain death begins 10 minutes biological death Resuscitate within 5 Minutes do we need AEDs? Sudden Cardiac Arrest (SCA) is the leading cause

More information

2015 Interim Resources for BLS

2015 Interim Resources for BLS 2015 Interim Resources for BLS Original Release: November 25, 2015 Starting in 2016, new versions of American Heart Association online courses will be released to reflect the changes published in the 2015

More information

In addition to information on preventing jams and what causes a paper jam, this section includes instructions for clearing the following jams:

In addition to information on preventing jams and what causes a paper jam, this section includes instructions for clearing the following jams: Phaser 4400 Laser Printer Paper Jams In addition to information on preventing jams and what causes a paper jam, this section includes instructions for clearing the following jams: Paper size jam see page

More information

Operating Instructions

Operating Instructions Operating Instructions Vision 5L Digital Holter Recorder Document Number: 010-1646-00 Revision: H Bothell, Washington 98021-8969 (608)-764-1919 Caution: U.S. Federal law restricts this device to sale by

More information

Impact Uni-Vent 754 Portable Ventilator

Impact Uni-Vent 754 Portable Ventilator Impact Uni-Vent 754 Portable Ventilator Description - Indications- Contraindications- Side Effects- Special Considerations- The Uni-Vent 754 Portable Ventilator is a portable electronically controlled,

More information

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS Blood Pressure Project 9 Blood pressure is a measure of the fluid pressure within the circulatory system. This pressure is required to ensure the delivery of oxygen and nutrients to, and the removal of

More information

Department of Emergency and Disaster Medicine Medical University of LODZ

Department of Emergency and Disaster Medicine Medical University of LODZ Electrotherapy in emergency states Department of Emergency and Disaster Medicine Medical University of LODZ defibrillation defibrillation The purpous of defibrillation is to deliver a randomly timed high-energy

More information

Welch Allyn 1500 Patient Monitor Train the Trainer

Welch Allyn 1500 Patient Monitor Train the Trainer Welch Allyn 1500 Patient Monitor Train the Trainer A. Intended Use Depending upon the configuration of your particular device, the Welch Allyn 1500 Patient Monitor can measure the following on Adult, Pediatric,

More information

Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS

Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS Transcutaneous Electrical Nerve Stimulation Device LUMI-TENS Operation Manual Read Before Using LUMI-TENS-INS-LAB-RevA08 TABLE OF CONTENTS INTRODUCTION TO TENS INDICATIONS AND CONTRAINDICATIONS WARNINGS

More information

UPGRADING AND SERVICING GUIDE

UPGRADING AND SERVICING GUIDE UPGRADING AND SERVICING GUIDE HPTouchSmart 610 PC Computer features may vary by model. Removing and Replacing a CD/DVD Drive...2 Removing the CD/DVD Drive... 2 Installing a New CD/DVD Drive... 5 Removing

More information

Success Manual and Cheat Sheet Notes to Pass Your Basic Life Support (BLS) Course

Success Manual and Cheat Sheet Notes to Pass Your Basic Life Support (BLS) Course Success Manual and Cheat Sheet Notes to Pass Your Basic Life Support (BLS) Course Written by: Jay Snaric, MS And Kimberly Hickman, RN CPR St. Louis 44 Meramec Valley Plaza St. Louis MO 63088 www.stlcpr.com

More information

Replacing the Gateway M320 Keyboard

Replacing the Gateway M320 Keyboard Replacing the Gateway M320 Keyboard This package includes a replacement keyboard for your Gateway M320 notebook and these printed instructions. Tools you need You need a small Phillips and a small flat-blade

More information

Replacement Instructions. Warning: During this procedure, keep small parts away from children.

Replacement Instructions. Warning: During this procedure, keep small parts away from children. apple ibook G4 Memory Card Replacement Instructions Follow the instructions in this sheet carefully. Failure to follow these instructions could damage your equipment and void its warranty. Note: Written

More information

Kurzweil KORE 64. ROM Expansion Board for the PC3 and PC3K

Kurzweil KORE 64. ROM Expansion Board for the PC3 and PC3K Kurzweil KORE 64 ROM Expansion Board for the PC3 and PC3K Installation Instructions August 1, 2012 2012 Young Chang Co., Ltd. All rights reserved. Kurzweil is a product line of Young Chang Co., Ltd. Kurzweil,

More information

Configuring the Siemens TC35 modems for use with the MI2292

Configuring the Siemens TC35 modems for use with the MI2292 Configuring the Siemens TC35 modems for use with the MI2292 The following instruction describe how to set up GSM communication between an MI2292 Power Quality Analyser Plus and a computer 1. Equipment

More information

LIFEPAK 12 DEFIBRILLATOR/MONITOR OPERATING INSTRUCTIONS

LIFEPAK 12 DEFIBRILLATOR/MONITOR OPERATING INSTRUCTIONS LIFEPAK 12 DEFIBRILLATOR/MONITOR OPERATING INSTRUCTIONS LIFEPAK 12 DEFIBRILLATOR/MONITOR OPERATING INSTRUCTIONS IMPORTANT INFORMATION!USA Rx Only Device Tracking!USA The U.S. Food and Drug Administration

More information

Removing memory and option cards

Removing memory and option cards These instructions can help you remove memory or option cards. Use the following illustration to locate the connector for the card you want to remove. Hard disk connector 1 Option card connector Firmware

More information

Fi TV Gold Quick Start Guide

Fi TV Gold Quick Start Guide Quick Start Guide Fi TV Gold Quick Start Guide Welcome to Fi TV! We re happy that you chose EPB Fiber Optics to bring the best movies, sports, news and television programming into your business and we

More information

QUICK REFERENCE. Connecting the Cables The power supply automatically switches between 115V and 230V.

QUICK REFERENCE. Connecting the Cables The power supply automatically switches between 115V and 230V. QUICK REFERENCE This Quick Reference contains supply loading and general care and maintenance procedures for the Monarch 9906 printer. For more detailed information, refer to the Operator s Handbook available

More information

PC Tab INSTALLATION MANUAL SECURE-IT, INC. EAST LONGMEADOW, MA 01028. 800-451-7592 / 413-525-7039 (fax) 413-525-8807 http://www.secure-it.

PC Tab INSTALLATION MANUAL SECURE-IT, INC. EAST LONGMEADOW, MA 01028. 800-451-7592 / 413-525-7039 (fax) 413-525-8807 http://www.secure-it. PC Tab INSTALLATION MANUAL SECURE-IT, INC. EAST LONGMEADOW, MA 01028 800-451-7592 / 413-525-7039 (fax) 413-525-8807 http://www.secure-it.com IMPORTANT READ ENTIRE MANUAL BEFORE STARTING!!! System Overview

More information

EMS SKILL CARDIAC EMERGENCY: AUTOMATED EXTERNAL DEFIBRILLATION (AED)

EMS SKILL CARDIAC EMERGENCY: AUTOMATED EXTERNAL DEFIBRILLATION (AED) EMS SKILL CARDIAC EMERGENCY: AUTOMATED EXTERNAL DEFIBRILLATION (AED) PERFORMANCE OBJECTIVE Demonstrate competency in assessing signs of cardiopulmonary arrest and performing defibrillation using a semi-automated

More information

There are only four actions that a patient can take to attempt to correct a monitor error:

There are only four actions that a patient can take to attempt to correct a monitor error: MyCareLink Error Codes When an error condition occurs with the MyCareLink Patient Monitor, the display screen provides a four digit code, along with an animated graphic to help patients resolve the error.

More information

W850. Maintenance Guide

W850. Maintenance Guide W850 Maintenance Guide August 2011 www.lexmark.com Contents...3 Cleaning the exterior of the printer...3 Storing supplies...3 Conserving supplies...3 Checking the status of supplies...4 Checking the status

More information

Q3 USER GUIDE. Read Me to Get the Most Out of Your Device...

Q3 USER GUIDE. Read Me to Get the Most Out of Your Device... Q3 USER GUIDE Read Me to Get the Most Out of Your Device... TO GET STARTED 1. Turn The Q3 On 2. Put It On 3. Listen In Model: Q3 Contents Charge Your Earpiece 6 Wearing the Q3 7 Pair Your Earpiece 7 Use

More information

Reocor D External Pacemaker

Reocor D External Pacemaker Cardiac Rhythm Management External Devices Quick Reference Guide Reocor D External Pacemaker Quick Reference Guide Reocor D Note: Before using Reocor D, read its user manual, including all safety instructions.

More information

Project 4.2.1: Heart Rate

Project 4.2.1: Heart Rate Project 4.2.1: Heart Rate Introduction Even before you were born, one of the first things your doctor did when you went for an office visit was listen to your heart. Your heart rate, the number of times

More information

Weather Direct Displays show Lost Forecast (blank boxes in the picture icons)

Weather Direct Displays show Lost Forecast (blank boxes in the picture icons) Weather Direct Displays show Lost Forecast (blank boxes in the picture icons) Many routine events can cause a Lost Forecast situation. Examples include: Power outage Change batteries Internet down in your

More information

ISDN 970 Telephone USER MANUAL. Copyright 1999 Lucent Technologies Inc. All rights reserved. Printed in Mexico. Issue L3 847937794

ISDN 970 Telephone USER MANUAL. Copyright 1999 Lucent Technologies Inc. All rights reserved. Printed in Mexico. Issue L3 847937794 ISDN 970 Telephone USER MANUAL Copyright 1999 Lucent Technologies Inc. All rights reserved. Printed in Mexico. Issue L3 847937794 Table of Contents ABOUT ISDN... 3 ISDN TERMS... 3 A GUIDE TO THIS MANUAL...

More information

9.0 Components Description. 10.0 SENTIO MMG System Quick Setup. 10.1 Plugging in Control Unit. 10.2 Attaching Sensor Harness

9.0 Components Description. 10.0 SENTIO MMG System Quick Setup. 10.1 Plugging in Control Unit. 10.2 Attaching Sensor Harness 9.0 Components Description The main components of the SENTIO MMG System consist of the SENTIO MMG Control Unit with preinstalled SENTIO MMG Software and the SENTIO MMG Sensor Harness. The SENTIO MMG Control

More information

quick reference guide Edge 605/705 GPS-ENABLED BIKE COMPUTER

quick reference guide Edge 605/705 GPS-ENABLED BIKE COMPUTER quick reference guide Edge 605/705 GPS-ENABLED BIKE COMPUTER MB Warning: This product contains a lithium-ion battery. See the Important Safety and Product Information guide in the product box for important

More information

Patient Instruction Manual

Patient Instruction Manual Patient Instruction Manual Preventice Services, LLC and Preventice Technologies, Inc., are independent, wholly owned subsidiaries of Preventice Solutions, Inc. Welcome to BodyGuardian Verité Your physician

More information

Setting Up the MimioCapture Device

Setting Up the MimioCapture Device Setting Up the MimioCapture Device The MimioCapture device enables you to use dry erase markers to draw on the whiteboard and record your writing without the need of a projector. Using MimioStudio Ink

More information

Table of Contents. Introduction. Radio Frequency Control. Initial setup Setting the time. Setting the temperature

Table of Contents. Introduction. Radio Frequency Control. Initial setup Setting the time. Setting the temperature LHZ USER GUIDE LHZ User Guide LHZ would like to thank you for purchasing your LHZ radiator. You have purchased a high quality, German designed and manufactured heating device, which will give you soft

More information

Lenovo IdeaPad Miix 10

Lenovo IdeaPad Miix 10 Lenovo IdeaPad Miix 10 User Guide Read the safety notices and important tips in the included manuals before using your computer. Notes Before using the product, be sure to read Lenovo Safety and General

More information

ZOLL AED Plus Corrective Action CUSTOMER Questions and Answers

ZOLL AED Plus Corrective Action CUSTOMER Questions and Answers ZOLL AED Plus Corrective Action CUSTOMER Questions and Answers Q: How can I tell if my AED Plus is subject to the Corrective Action? A: Look at the unique serial number, containing letters and digits under

More information

Automatic Phone-Out Home Monitoring Systems

Automatic Phone-Out Home Monitoring Systems Automatic Phone-Out Home Monitoring Systems Power Outage and Freeze Alarm Model Number: THP202 Power Outage, Freeze and Flood Alarm Product Description Model Number: THP201 These monitoring systems are

More information

Memory Modules User Guide

Memory Modules User Guide Memory Modules User Guide Copyright 2008 Hewlett-Packard Development Company, L.P. The information contained herein is subject to change without notice. The only warranties for HP products and services

More information

ES&S AutoMARK. Pre-Election Day Checklist

ES&S AutoMARK. Pre-Election Day Checklist ES&S AutoMARK Pre-Election Day Checklist Rear Access Door Touch Screen Monitor Lid Compact Flash Access Door Lid Latches Key-Operated Mode Switch Keypad Ballot Feed Tray ADA Port Headphone Jacks 1 Checking

More information

HEARTSTART XL DEFIBRILLATOR/MONITOR Model: M4735A PRODUCT DESCRIPTION

HEARTSTART XL DEFIBRILLATOR/MONITOR Model: M4735A PRODUCT DESCRIPTION HEARTSTART XL DEFIBRILLATOR/MONITOR Model: M4735A PRODUCT DESCRIPTION Philips Medical Systems M4735A HeartStart XL offers advanced and basic life-support clinicians a compact, lightweight, easy-to-use

More information

Digital Adapter Easy Setup Guide

Digital Adapter Easy Setup Guide Digital dapter Easy Setup Guide HD Digital dapter Easy Setup Guide Installation Operation Troubleshooting Installation Operation Troubleshooting Pace D60Xu What You Need To Know Table Of ontents To bring

More information

Memory Modules. Document Part Number: 407947-001. This guide explains how to replace and upgrade memory in the computer. May 2006

Memory Modules. Document Part Number: 407947-001. This guide explains how to replace and upgrade memory in the computer. May 2006 Memory Modules Document Part Number: 407947-001 May 2006 This guide explains how to replace and upgrade memory in the computer. Contents 1 Adding or replacing memory modules Adding or replacing a memory

More information

PRODUCTIVITY THROUGH INNOVATION 600 CONTROL DIRECT DRIVE TECHNICAL/OPERATION MANUAL

PRODUCTIVITY THROUGH INNOVATION 600 CONTROL DIRECT DRIVE TECHNICAL/OPERATION MANUAL Rev. D PRODUCTIVITY THROUGH INNOVATION 600 CONTROL DIRECT DRIVE TECHNICAL/OPERATION MANUAL 10 BORIGHT AVENUE, KENILWORTH NEW JERSEY 07033 TELEPHONE: 800-524-0273 FAX: 908-686-9317 TABLE OF CONTENTS Page

More information

Router. A. Identify the major parts of the router. B. Complete a written test on safety and operating procedures of the router with 100% accuracy.

Router. A. Identify the major parts of the router. B. Complete a written test on safety and operating procedures of the router with 100% accuracy. Router I. Competencies Given a properly adjusted router, instruction and demonstration of use, each student will be able to: A. Identify the major parts of the router. B. Complete a written test on safety

More information

PeopleGIS Leica RTK GPS Instructions

PeopleGIS Leica RTK GPS Instructions PeopleGIS Leica RTK GPS Instructions Equipment List GG02 Receiver (Antenna) GPS Range Pole Zeno Tablet MiFi MiFi External Battery Carrying Case 1) Turning on the MiFi Power on the MiFi. This little gadget

More information

Michigan Adult Cardiac Protocols CARDIAC ARREST GENERAL. Date: May 31, 2012 Page 1 of 5

Michigan Adult Cardiac Protocols CARDIAC ARREST GENERAL. Date: May 31, 2012 Page 1 of 5 Date: May 31, 2012 Page 1 of 5 Cardiac Arrest General This protocol should be followed for all adult cardiac arrests. Medical cardiac arrest patients undergoing attempted resuscitation should not be transported

More information

Online Monitoring User Guide

Online Monitoring User Guide High Resolution Temperature Sensing Strip Online Monitoring User Guide 888.637.3282 www.nerdata.com Page 1 of 26 Contents The Aurora Online Monitoring System... 3 1. Creating an Account on the Aurora Online

More information

SimPad SkillReporter. User Guide. www.laerdal.com

SimPad SkillReporter. User Guide. www.laerdal.com SimPad SkillReporter User Guide www.laerdal.com 1 Introduction INTRODUCTION SimPad is a wireless touchscreen device for use with Laerdal manikins. This version is installed with SkillReporter software

More information

Lenovo Miix 2 8. User Guide. Read the safety notices and important tips in the included manuals before using your computer.

Lenovo Miix 2 8. User Guide. Read the safety notices and important tips in the included manuals before using your computer. Lenovo Miix 2 8 User Guide Read the safety notices and important tips in the included manuals before using your computer. Notes Before using the product, be sure to read Lenovo Safety and General Information

More information

Ready to save lives. Philips HeartStart XL+ Defibrillator/Monitor

Ready to save lives. Philips HeartStart XL+ Defibrillator/Monitor Ready to save lives Philips HeartStart XL+ Defibrillator/Monitor Ready to respond, revive, Hospital cardiac emergencies are typically stressful and chaotic events. The last thing you want to worry about

More information

Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS

Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS + Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS Inventory Safety and General Information bu001c This unit is intended for indoor use only. Do not operate this unit in direct sunlight,

More information

MINI ELECTRIC SEWING MACHINE OPERATION MANUAL

MINI ELECTRIC SEWING MACHINE OPERATION MANUAL MINI ELECTRIC SEWING MACHINE OPERATION MANUAL 1 Parts & Accessories Takeup Lever(D) Spindle(C) Spool(E) MAIN UNIT Bobbin Holder(V) Bobbin winder Spool(T) Needle Clamp Screw(O) Adjusting Screw(Q) Needle

More information

Copyright 2007 Hewlett-Packard Development Company, L.P. Windows is a U.S. registered trademark of Microsoft Corporation.

Copyright 2007 Hewlett-Packard Development Company, L.P. Windows is a U.S. registered trademark of Microsoft Corporation. Drives User Guide Copyright 2007 Hewlett-Packard Development Company, L.P. Windows is a U.S. registered trademark of Microsoft Corporation. The information contained herein is subject to change without

More information

Orientation Guide www.lucas-cpr.com a product by JOLIFE 900080-00 RevB JOLIFE 2012

Orientation Guide www.lucas-cpr.com a product by JOLIFE 900080-00 RevB JOLIFE 2012 Orientation Guide www.lucas-cpr.com a product by JOLIFE 900080-00 RevB JOLIFE 2012 Refer to the Instructions for Use for complete directions for use, indications, contraindications, warnings, precautions

More information

Start Here USB. (802.11) (Ethernet)

Start Here USB. (802.11) (Ethernet) Start Here 1 USB 802.11 Ethernet USB cable users: Do not connect the USB cable until Section A. USB A USB Wireless (802.11) or wired (Ethernet) network users : you must follow the instructions in this

More information

ACCESSORIES BROCHURE. LIFEPAK 20/20e. DEFIBRILLATOR/MONITOR with CodeManagement Module TM. Genuine Accessories from Physio-Control

ACCESSORIES BROCHURE. LIFEPAK 20/20e. DEFIBRILLATOR/MONITOR with CodeManagement Module TM. Genuine Accessories from Physio-Control ACCESSORIES BROCHURE LIFEPAK 20/20e DEFIBRILLATOR/MONITOR with CodeManagement Module TM Genuine Accessories from Physio-Control LIFEPAK 20 DEFIBRILLATOR/MONITOR 1 LIFEPAK 20 Genuine Accessories from Physio-Control

More information

How to use a SMART Board

How to use a SMART Board 1 How to use a SMART Board Basic instructions on how to set up the SMART Board and how to use the SMART Board with both PC and Mac laptops. Contents Page 2: SMART Board Set Up Instructions Page 3: SMART

More information

FG MOISTURE MONITOR Installation & Operation Manual

FG MOISTURE MONITOR Installation & Operation Manual FG MOISTURE MONITOR Installation & Operation Manual Issue 3.0 7/20/10 1 Contents SERVICE AND TECHNICAL SUPPORT... 2 INSTALLATION:... 3 MOISTURE SENSOR INSTALLATION:... 3 SENSOR CONNECTOR:... 5 MONITOR

More information

How To Upgrade An Aed

How To Upgrade An Aed Software Upgrade Instructions for Powerheart, FirstSave, and CardioVive Automated External Defibrillators (AEDs) Copyright 2010 Cardiac Science Corporation. All rights reserved. Contents About the upgrade

More information

Infant CPR Skills Testing Checklist

Infant CPR Skills Testing Checklist Infant CPR Skills Testing Checklist Student Name Date of Test Scenario: While you are pushing a baby in a stroller at the park, you notice something is wrong with the baby. You do not have a phone nearby.

More information

Electronic Brake Controller Hayes Brake Controller Company ENERGIZE III P/N # 81741B or ENERGIZE XPC P/N #81745 OPERATION MANUAL

Electronic Brake Controller Hayes Brake Controller Company ENERGIZE III P/N # 81741B or ENERGIZE XPC P/N #81745 OPERATION MANUAL Electronic Brake Controller Hayes Brake Controller Company ENERGIZE III P/N # 81741B or ENERGIZE XPC P/N #81745 OPERATION MANUAL ENERGIZE III is for trailers with 2 or 4 electric brakes and vehicles with

More information

Table of Contents. Use. Troubleshooting. Setup. Welcome. 11 How to arm/disarm system/camera(s) 19 Sync Module setup issues. 3 Installing the Blink app

Table of Contents. Use. Troubleshooting. Setup. Welcome. 11 How to arm/disarm system/camera(s) 19 Sync Module setup issues. 3 Installing the Blink app User Guide Table of Contents Welcome Setup Use Troubleshooting 2 What s in the box 2 What you need 3 Installing the Blink app 4 Setting up the Sync Module 5 Connecting the Sync Module to Your Wi-Fi 7 Adding

More information

Four-Line Intercom Speakerphone 944

Four-Line Intercom Speakerphone 944 1 USER S MANUAL Part 2 Four-Line Intercom Speakerphone 944 Please also read Part 1 Important Product Information AT&T and the globe symbol are registered trademarks of AT&T Corp. licensed to Advanced American

More information

Contents Contents Contents Overview Charging your headset Wearing the headset Powering on the headset Connecting your headset for dictation

Contents Contents Contents Overview Charging your headset Wearing the headset Powering on the headset Connecting your headset for dictation 1 Contents Contents Contents 2 Overview 3 Charging your headset 4 Wearing the headset 4 Powering on the headset 4 Connecting your headset for dictation 5 The adapter 5 Adjusting your computer's audio settings

More information

Lexmark T64x Troubleshooting Guide

Lexmark T64x Troubleshooting Guide Lexmark T64x Troubleshooting Guide Online technical support Technical support is available on our Lexmark Web site at www.lexmark.com. Checking an unresponsive printer If the printer is not responding,

More information

OC-3 SERIES OF PRINTERS

OC-3 SERIES OF PRINTERS 110204-003 OC-3 SERIES OF PRINTERS User Guide TABLE OF CONTENTS OC-3 Printer Views... 1 Front and back view... 1 Side views... 1 Setting up your printer... 2 Unpacking your printer... 2 Installing/removing

More information

ACLS Provider Manual Comparison Sheet Based on 2010 AHA Guidelines for CPR and ECC. BLS Changes

ACLS Provider Manual Comparison Sheet Based on 2010 AHA Guidelines for CPR and ECC. BLS Changes ACLS Provider Manual Comparison Sheet Based on 2010 AHA Guidelines for CPR and ECC CPR Chest compressions, Airway, Breathing (C-A-B) BLS Changes New Old Rationale New science indicates the following order:

More information

FiLIP 2 USER GUIDE 2015 Filip Technologies, Inc. All Rights Reserved. APR 2015

FiLIP 2 USER GUIDE 2015 Filip Technologies, Inc. All Rights Reserved. APR 2015 FiLIP 2 USER GUIDE 2015 Filip Technologies, Inc. All Rights Reserved. APR 2015 WELCOME TO THE FiLIP FAMILY FiLIP helps children stay in touch with their parents and helps parents know where their children

More information