Physiological Pressure Measurements
|
|
- Hannah June Richardson
- 7 years ago
- Views:
Transcription
1 Physiological Pressure Measurements TBMT09
2 today s lecture Literature: Lecture notes/slides! See also Togawa, Tamura, Öberg: Biomedical Transducers and Instruments Ask: Tryckmätning Blom ch 3
3 Lecture outline Indirect/direct pressure measurement methods Pressure transducers Pressure measurement system
4 The first invasive blood pressure measurement The first invasive attempt to measure blood pressure was made by Stephen Hales in From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970
5 Blood pressure levels Normal blood pressure Hypertension and Hypotension Narrow pressure windows
6 Pressure measurements Indirect methods Arm-occluding cuff Auscultation Palpation Oscillometric method Automatic methods Aplanation tonometry Direct methods Hydraulic systems Catheter tip systems
7 Indirect methods
8 Auscultatory blood pressure measurement Auscultatory blood pressure measurement using a pressure cuff and a stethoscope
9 Effects of height difference on the measured blood pressure From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997
10 The reference level in humans From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997
11 Cuff pressure measurements No sound No sound Fluttering sound Weak noise
12 Korotkoff sounds From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
13 Oscillometric principle of pressure cuff Pressure detection Blood vessel Transducer arm cuff
14 Automatic Oscillometric blood pressure devices Might be unreliable due to geometry of the cuff, pressure detection algorithms and location of the cuff. Can be used for monitoring but diagnostic value of measurements is questionable
15 PPG and BP measurements Zero-balance technique with read light transmission (PPG) and a shaker driven pressure cuff keeping the light signal constant. The transducer signal will follow the time course of the blood pressure From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997
16 Zero-balance technique with light transmission and shaker driven pressure cuff The transmitted red light reflects the blood volume in the finger. The shaker applies a pressure to the finger cuff. By keeping the optical/volume signal constant by varying the shaker signal, the pressure in the cuff will follow instantaneous blood pressure changes
17 Maximal pulsations reflect maximally relieved blood vessels. The DC component of the PPG signal at this point is used as a set value of the feedback system From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997
18 Finapres
19 Micropipettes and BP measurements Zero-balance technique for measuring pressure in blood capillaries using glass micropipettes fill with KCl immersed in the blood. The impedance of the pipette is measured and kept constant by applying a counter pressure from a bellow to the pipette. From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974 The impedance differ very much depending on if the tip of the pipette is filled with high molar KCl or blood/physiological NaCl (se next page). By using a feedback system keeping the impedance constant the pressure applied to the pipette will follow that of the capillaries
20 Impedance of a micropipette Impedance versus counter pressure of a micropipette filled with 3M KCl and immersed in solutions of various concentrations. With a moderate concentration of this solution (as in the case of blood/physiological saline) there is dramatic impedance change around 5 mmhg making the impedance signal suitable as the control signal of the feedback system. From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
21 Tonimetry for intraocular pressuremeasurements Tonimeter with eye microscope
22 When approaching a flat surface to the eye the cornea will be flattened (applanated). In principle the force on the flat surface is the intraocular pressure times the applanated area of the cornea. Knowing the area, the pressure can be measured from the force. Errors in this measurement are the surface adhesion of the tear fluid on one hand and the bending force of the cornea on the other. With the shown Mackay-Marg tonometer only the central (in this case 1.5 mm) of the flat surface is pressure sensitive. From: RC Cobbold. Transducers for biomedical measurements, John Wiley, Pressure increases as surface flattens the cornea. 2. Force corresponding to sum of component from intraocular pressure and bending force. 3. Force reflecting intra ocular pressure only since bending force is outside the pressure sensitive surface
23 Goldman applanation tonimeter With this technique the cornea is applanated to a predetermined area (measured by a split image technique) where in average the forces due to corneal bending and tear fluid adhesion should cancel From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997
24 Non-contact applanation tonimeter The pneumatic system creates an air jet flow velocity which is controlled by a piston pump. The jet creates a stagnation pressure that displaces/ flattens the corneal surface. The flattened surface is detected by an optical system. Knowing jet velocity and the flattened area the intraocular pressure can be estimated. From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997
25 Direct methods
26 The pressure signal One have to measure not only the values of systolic and diastolic pressure but the whole accurate time course of the signal. Requires: 1. Accuracy 2. Bandwidth
27 Historic background Direct pressure measurement by Fick 1864 From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970
28 Intra-cardiac measurement First intracardiac measurements by Marey & Chauveau 1860 s From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970
29 Example of ancient recording Recordings by Marey &Chauveau 1861 From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970
30 Landois hematogram 1872 From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970
31 Early pressure transducer Pressure transducer with a dome for connecting catheters and a metal membrane linked to strain-gauge sensing elements (Statham 1940`s) From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970
32 Disposable pressure transducer From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997
33 Clinical pressure transducers Modern pressure transducers with a dome (above) and one where the sensing element is included in the Luer tubing (blow)
34 Technical development of pressure transducers From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
35 Illustration of the piezo-resistivity of silicone. Plotting resistance change as a function of material strain p- and n-type have the opposite characteristics. p-type shows a more linear behavior and is therefore preferred From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
36 p-type piezo-resistive elements diffused into silicone material for various geometries Membrane Rod/cantilever From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
37 Catheter and Transducer Changing pressure, p Catheter compliance C radius, r Transducer compliance, Co Displacement of the surface and Volume displacment Mass of the fluid Fluid resistance Damping
38 Hydraulic coupled system From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
39 Resistance, inertance, compliance The flow resistance is given by: 8ηl R = π 4 r The Inertance is given by: L ρl = 2 π r The compliance is given by: C C = C
40 Pressure transfer function The pressure transfer function: which can be written as a second order transfer function: ) ( ) ( sc R sl sc s p s p i + + = ) ( ) ( ω ω ω + + = s D s s p s p i
41 Resonance frequency Identification gives us the resonance frequency: f 0 = 2π 1 LC 0 = 2 1 πρ r lc 0 The resonance frequency is proportional to the radius and inversely proportional to the square root of the catheter length and compliance.
42 Damping The damping is obtained as: D = R C0 4η lc = 3 2 L πρ r 0 The damping depends on the square root of the catheter length and compliance and inversely to the radius to the power of three.
43 Step response of a second order system The response of a second order system as a function of the damping
44 Example under-damped system Correctly measured aortic pressure (upper trace) The same signal measured with un under-damped low band-width hydraulic system (lower trace)
45 Areas of damping and resonance frequency resulting in an accurate recording of the blood pressure signal
46 Hydraulic parallel damping of a pressure measurement system
47 Importance of pressure sensor placements 1 P = P + ρv 2 1 P = P ρv 2 P = P 2 2 Depending on the direction of the catheter tip, the measured pressure can consist of both static and dynamic terms. (according to the Bernoulli equation)
48 A catheter-tip pressure transducer The transducer element is etched in silicon to a pressure sensing membrane
49 Transducer element A detailed view of the transducer element etched in silicon forming a membrane
50 Fiber optic pressure transducer From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
51 Example of optic sensor
52 Principles of a fiber optic pressure transducer From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
53 Fiber optic pressure transducer From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974
54 Fiber optic pressure transducer The principle behind how a fiber optic pressure transducer works. A fiber (1) emits light within a cone in a certain angle. The other fiber (2) receives light within a similar cone. The overlap between these cones is larger the larger the distance is between the fibers and the reflecting metal membrane. Pressure The surface is close pressure on membrane is high Pressure Distance is larger pressure on membrane is smaller
55 Fiber optic miniature pressure transducer If the glass-beam is displaced downwards it approaches a glossy silicone corner surface and more of the light is reflected.
Bioinstrumentation. Kuo-Sheng Cheng, Ph.D. Department of Biomedical Engineering National Cheng Kung University
Bioinstrumentation Medical Instrumentation: Application and Design Third Edition John G. Webster, Editor Kuo-Sheng Cheng, Ph.D. Department of Biomedical Engineering National Cheng Kung University The Importance
More informationHemodynamic Monitoring: Principles to Practice M. L. Cheatham, MD, FACS, FCCM
SUMMARY HEMODYNAMIC MONITORING: FROM PRINCIPLES TO PRACTICE Michael L. Cheatham, MD, FACS, FCCM Director, Surgical Intensive Care Units Orlando Regional Medical Center Orlando, Florida Fluid-filled catheters
More informationUnsteady Pressure Measurements
Quite often the measurements of pressures has to be conducted in unsteady conditions. Typical cases are those of -the measurement of time-varying pressure (with periodic oscillations or step changes) -the
More informationhuidige en toekomstige hemodynamische bewaking
huidige en toekomstige hemodynamische bewaking Stefan De Hert Department of Anesthesiology Ghent University Hospital Ghent University Belgium disclosures? 2 hemodynamic monitoring? ~ adequacy of organ
More informationVarious Technics of Liquids and Solids Level Measurements. (Part 3)
(Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,
More informationPressure Sensor Based Estimation of Pulse Transit Time
International Journal of Information & Computation Technology. ISSN 0974-2239 Volume 4, Number 13 (2014), pp. 1321-1328 International Research Publications House http://www. irphouse.com Pressure Sensor
More information3 rd Russian-Bavarian Conference on Bio-Medical Engineering
3 rd Russian-Bavarian Conference on Bio-Medical Engineering Blood Pressure Estimation based on Pulse Transit Time and Compensation of Vertical Position Dipl.-Inform. Med. Christian Douniama Dipl.-Ing.
More informationUpper Arm Blood Pressure Monitor SC 6800. Instructions for use
Upper Arm Blood Pressure Monitor SC 6800 Instructions for use Contents page Introduction... 3 Features of the SC 6800. 3 Safety instructions... 4 WHO-Interpretation of blood pressure readings... 4 Device
More informationSection 5.0 : Horn Physics. By Martin J. King, 6/29/08 Copyright 2008 by Martin J. King. All Rights Reserved.
Section 5. : Horn Physics Section 5. : Horn Physics By Martin J. King, 6/29/8 Copyright 28 by Martin J. King. All Rights Reserved. Before discussing the design of a horn loaded loudspeaker system, it is
More informationSubminiature Load Cell Model 8417
w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...
More informationAgoraLink Agora for Life Science Technologies Linköpings Universitet Kurs i Fysiologisk mätteknik Biofluidflöden
AgoraLink Agora for Life Science Technologies Linköpings Universitet Kurs i Fysiologisk mätteknik Biofluidflöden Fysiologisk mätteknik Anatomy of the heart The complex myocardium structure right ventricle
More informationAPPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS
APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS Selection and use of Ultrasonic Ceramic Transducers The purpose of this application note is to aid the user in the selection and application of the Ultrasonic
More informationEvaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS
Blood Pressure Project 9 Blood pressure is a measure of the fluid pressure within the circulatory system. This pressure is required to ensure the delivery of oxygen and nutrients to, and the removal of
More informationDirect Arterial Blood Pressure Monitoring Angel M. Rivera CVT, VTS (ECC) Animal Emergency Center Glendale, WI March 2003
Direct Arterial Blood Pressure Monitoring Angel M. Rivera CVT, VTS (ECC) Animal Emergency Center Glendale, WI March 2003 Introduction Direct measurement of arterial blood pressure is obtained via a peripheral
More informationOEM MAXNIBP Frequently Asked Questions
Frequently Asked Questions Why does the monitor sometimes inflate the BP cuff, then shortly thereafter reinflate the cuff? How will I know if the monitor is experiencing motion artifact during a measurement?
More informationSynthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
More information3. Test Methods for Evaluation of ESCR of Plastics
3. Test Methods for Evaluation of ESCR of Plastics A common laboratory request for ESC-prone polymers is to check ESCR performance for quality control, competitive product evaluations, and research and
More informationTHE STRAIN GAGE PRESSURE TRANSDUCER
THE STRAIN GAGE PRESSURE TRANSDUCER Pressure transducers use a variety of sensing devices to provide an electrical output proportional to applied pressure. The sensing device employed in the transducers
More informationHow do you adjust for IOP in a LASIK patient?
How do you adjust for IOP in a LASIK patient? Sunita Radhakrishnan, M.D Glaucoma Center of San Francisco Glaucoma Research and Education Group Principle of applanation tonometry Imbert Fick law Applanation
More informationHow to measure absolute pressure using piezoresistive sensing elements
In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article
More informationDigital Systems Ribbon Cables I CMPE 650. Ribbon Cables A ribbon cable is any cable having multiple conductors bound together in a flat, wide strip.
Ribbon Cables A ribbon cable is any cable having multiple conductors bound together in a flat, wide strip. Each dielectric configuration has different high-frequency characteristics. All configurations
More informationLaminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers
Flow and Flow rate. Laminar and Turbulent flow Laminar flow: smooth, orderly and regular Mechanical sensors have inertia, which can integrate out small variations due to turbulence Turbulent flow: chaotic
More informationVascular System The heart can be thought of 2 separate pumps from the right ventricle, blood is pumped at a low pressure to the lungs and then back
Vascular System The heart can be thought of 2 separate pumps from the right ventricle, blood is pumped at a low pressure to the lungs and then back to the left atria from the left ventricle, blood is pumped
More informationHead Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids
Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction Last lab you investigated flow loss in a pipe due to the roughness
More informationGuidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions
Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions Document issued on: March 27, 2006 For questions regarding this document, contact Everette Beers, Ph.D. at 240-276-4200
More informationINSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS
INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS This tutorial provides an overview of signal processing and conditioning for use in instrumentation and automatic control systems.
More informationSurface Profilometry as a tool to Measure Thin Film Stress, A Practical Approach. Gianni Franceschinis, RIT MicroE Graduate Student
1 Surface Profilometry as a tool to Measure Thin Film Stress, A Practical Approach. Gianni Franceschinis, RIT MicroE Graduate Student Abstract-- As the film decreases in thickness the requirements of more
More informationIntroduction to acoustic imaging
Introduction to acoustic imaging Contents 1 Propagation of acoustic waves 3 1.1 Wave types.......................................... 3 1.2 Mathematical formulation.................................. 4 1.3
More informationCharacteristic and use
. Basic principle A PSD basically consists of a uniform resistive layer formed on one or both surfaces of a high-resistivity semiconductor substrate, and a pair of electrodes formed on both ends of the
More informationEndoscope Optics. Chapter 8. 8.1 Introduction
Chapter 8 Endoscope Optics Endoscopes are used to observe otherwise inaccessible areas within the human body either noninvasively or minimally invasively. Endoscopes have unparalleled ability to visualize
More informationRelated concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.
Conductivity of strong and weak electrolytes TEC Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Principle It is possible to
More informationFluid Mechanics: Static s Kinematics Dynamics Fluid
Fluid Mechanics: Fluid mechanics may be defined as that branch of engineering science that deals with the behavior of fluid under the condition of rest and motion Fluid mechanics may be divided into three
More informationPrecision Miniature Load Cell. Models 8431, 8432 with Overload Protection
w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force
More informationAS COMPETITION PAPER 2008
AS COMPETITION PAPER 28 Name School Town & County Total Mark/5 Time Allowed: One hour Attempt as many questions as you can. Write your answers on this question paper. Marks allocated for each question
More informationHow to measure blood pressure using digital monitors
How to measure blood pressure using digital monitors What is high blood pressure? Blood pressure is the pressure exerted by circulating blood upon the walls of blood vessels. Systolic pressure is the blood
More informationSoil Suction. Total Suction
Soil Suction Total Suction Total soil suction is defined in terms of the free energy or the relative vapor pressure (relative humidity) of the soil moisture. Ψ = v RT ln v w 0ω v u v 0 ( u ) u = partial
More informationPressure. Pressure. Atmospheric pressure. Conceptual example 1: Blood pressure. Pressure is force per unit area:
Pressure Pressure is force per unit area: F P = A Pressure Te direction of te force exerted on an object by a fluid is toward te object and perpendicular to its surface. At a microscopic level, te force
More informationProcon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS
Procon Engineering Technical Document PELR 1002 TERMS and DEFINITIONS The following terms are widely used in the weighing industry. Informal comment on terms is in italics and is not part of the formal
More informationA LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW. 1998 ASME Fluids Engineering Division Summer Meeting
TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW Proceedings of FEDSM 98: June -5, 998, Washington, DC FEDSM98 49 ABSTRACT The pressure
More informationThe V15VxMR. The V15VxMR offers all these outstanding features. Trackability is the stylus' ability to stay in contact with
The V15VxMR (See illustrations on inside back cover.) The Micro-Ridge diamond tip and the beryllium stylus cantilever of the V15VxMR combine to achieve the highest trackability and most detailed and accurate
More informationAP PHYSICS 2012 SCORING GUIDELINES
AP PHYSICS 2012 SCORING GUIDELINES General Notes About 2012 AP Physics Scoring Guidelines 1. The solutions contain the most common method of solving the free-response questions and the allocation of points
More informationRANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA
RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA ABSTRACT Random vibration is becoming increasingly recognized as the most realistic method of simulating the dynamic environment of military
More informationSpring Simple Harmonic Oscillator. Spring constant. Potential Energy stored in a Spring. Understanding oscillations. Understanding oscillations
Spring Simple Harmonic Oscillator Simple Harmonic Oscillations and Resonance We have an object attached to a spring. The object is on a horizontal frictionless surface. We move the object so the spring
More informationTransmission Line and Back Loaded Horn Physics
Introduction By Martin J. King, 3/29/3 Copyright 23 by Martin J. King. All Rights Reserved. In order to differentiate between a transmission line and a back loaded horn, it is really important to understand
More informationLoudspeaker Parameters. D. G. Meyer School of Electrical & Computer Engineering
Loudspeaker Parameters D. G. Meyer School of Electrical & Computer Engineering Outline Review of How Loudspeakers Work Small Signal Loudspeaker Parameters Effect of Loudspeaker Cable Sample Loudspeaker
More informationUltrasonic Reference Blocks
Ultrasonic Reference Blocks Custom and Standard Test Blocks for Nondestructive Testing Introduction 3 IIW Type 1 Calibration Block 4 Miniature Angle Beam (ROMPAS) Calibration Block 4 Distance Calibration
More informationStudents Manual for the Exam. General Engineering and Electrical Civil Engineering Discipline
Students Manual for the Exam General Engineering and Electrical Civil Engineering Discipline -- October March 2014 2013 -- COPYRIGHT NOTICE COPYRIGHTS 2013 NATIONAL CENTER FOR ASSESSMENT IN HIGHER EDUCATION
More informationSensors Collecting Manufacturing Process Data
Sensors & Actuators Sensors Collecting Manufacturing Process Data Data must be collected from the manufacturing process Data (commands and instructions) must be communicated to the process Data are of
More informationNormal Intracardiac Pressures. Lancashire & South Cumbria Cardiac Network
Normal Intracardiac Pressures Lancashire & South Cumbria Cardiac Network Principle Pressures recorded from catheter tip Electrical transducer - wheatstone bridge mechanical to electrical waveform display
More informationOptical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus
Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,
More informationSUSPENSION AND STEERING OVERVIEW
SUSPENSION SUSPENSION AND STEERING OVERVIEW The S40/V50 has a wide track and a long wheelbase for its relative size and weight. This gives the car stable and predictable driving characteristics. It also
More informationSingle-Needle Cross-Over System. Brief description
Single-Needle Cross-Over System Brief description Introduction Brief description of the Single-Needle Cross-Over system from B. Braun Unique pressure-pressure controlled SN process Constant blood flow
More informationComparison of the Response of a Simple Structure to Single Axis and Multiple Axis Random Vibration Inputs
Comparison of the Response of a Simple Structure to Single Axis and Multiple Axis Random Vibration Inputs Dan Gregory Sandia National Laboratories Albuquerque NM 87185 (505) 844-9743 Fernando Bitsie Sandia
More informationAcceleration levels of dropped objects
Acceleration levels of dropped objects cmyk Acceleration levels of dropped objects Introduction his paper is intended to provide an overview of drop shock testing, which is defined as the acceleration
More informationMEASURING AND RECORDING BLOOD PRESSURE
MEASURING AND RECORDING BLOOD PRESSURE INTRODUCTION The blood pressure, along with the body temperature, pulse, and respirations, is one of the vital signs. These measurements are used to quickly, easily,
More informationRLC Resonant Circuits
C esonant Circuits Andrew McHutchon April 20, 203 Capacitors and Inductors There is a lot of inconsistency when it comes to dealing with reactances of complex components. The format followed in this document
More informationINTERFERENCE OF SOUND WAVES
1/2016 Sound 1/8 INTERFERENCE OF SOUND WAVES PURPOSE: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves and to observe interference phenomena with ultrasonic sound waves.
More informationAPPLICATION NOTE AP050830
APPLICATION NOTE AP050830 Selection and use of Ultrasonic Ceramic Transducers Pro-Wave Electronics Corp. E-mail: sales@pro-wave.com.tw URL: http://www.prowave.com.tw The purpose of this application note
More informationSpecification of 5 Wire Analog Touch Panel
Specification of 5 Wire Analog Touch Panel 5 WIRE ANALOG RESISTIVE TOUCH PANEL A. Application This specification applies to the 5 Wire Analog Resistive Touch Panel. B. Environmental Conditions 1. Operating
More informationBASIC ELECTRONICS AC CIRCUIT ANALYSIS. December 2011
AM 5-202 BASIC ELECTRONICS AC CIRCUIT ANALYSIS December 2011 DISTRIBUTION RESTRICTION: Approved for Pubic Release. Distribution is unlimited. DEPARTMENT OF THE ARMY MILITARY AUXILIARY RADIO SYSTEM FORT
More informationTime Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication
Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication Thomas Reilly Data Physics Corporation 1741 Technology Drive, Suite 260 San Jose, CA 95110 (408) 216-8440 This paper
More informationBlood pressure measurement devices. December 2013
Blood pressure measurement devices December 2013 Contents 1 Executive summary... 3 2 Introduction... 3 3 Blood pressure measurement equipment... 5 3.1 Manual auscultatory technique... 5 3.2 Automated oscillometric
More informationBIOMEDICAL ULTRASOUND
BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic
More informationGMC 2013: Piping Misalignment and Vibration Related Fatigue Failures
GMC 2013: Piping Misalignment and Vibration Related Fatigue Failures www.betamachinery.com Authors/Presenters: Gary Maxwell, General Manager, BETA Machinery Analysis Brian Howes, Chief Engineer, BETA Machinery
More informationCircuits with inductors and alternating currents. Chapter 20 #45, 46, 47, 49
Circuits with inductors and alternating currents Chapter 20 #45, 46, 47, 49 RL circuits Ch. 20 (last section) Symbol for inductor looks like a spring. An inductor is a circuit element that has a large
More informationNormal & Abnormal Intracardiac. Lancashire & South Cumbria Cardiac Network
Normal & Abnormal Intracardiac Pressures Lancashire & South Cumbria Cardiac Network Principle Pressures recorded from catheter tip Electrical transducer - wheatstone bridge mechanical to electrical waveform
More informationRobot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
More informationHarmonic oscillations of spiral springs Springs linked in parallel and in series
.3.26 Related topics Spring constant, Hooke s Law, oscillations, limit of elasticity, parallel springs, serial springs, use of an interface. Principle and task The spring constant D is determined for different
More informationSection Four: Pulmonary Artery Waveform Interpretation
Section Four: Pulmonary Artery Waveform Interpretation All hemodynamic pressures and waveforms are generated by pressure changes in the heart caused by myocardial contraction (systole) and relaxation/filling
More informationApplication Note for SDP600 and SDP1000 Series Measuring Differential Pressure and Air Volume with Sensirion s CMOSens technology
Application Note for SDP600 and SDP1000 Series Measuring Differential Pressure and Air Volume with Sensirion s CMOSens technology Summary The increasing customer requirements for comfort and safety and
More informationPaul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK.
The Physics of Flow Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK. Introduction Flow is defined as the quantity of fluid
More informationManufacturing Equipment Modeling
QUESTION 1 For a linear axis actuated by an electric motor complete the following: a. Derive a differential equation for the linear axis velocity assuming viscous friction acts on the DC motor shaft, leadscrew,
More informationExperiment 1: SOUND. The equation used to describe a simple sinusoidal function that propagates in space is given by Y = A o sin(k(x v t))
Experiment 1: SOUND Introduction Sound is classified under the topic of mechanical waves. A mechanical wave is a term which refers to a displacement of elements in a medium from their equilibrium state,
More informationLecture 5 Hemodynamics. Description of fluid flow. The equation of continuity
1 Lecture 5 Hemodynamics Description of fluid flow Hydrodynamics is the part of physics, which studies the motion of fluids. It is based on the laws of mechanics. Hemodynamics studies the motion of blood
More informationForces. Definition Friction Falling Objects Projectiles Newton s Laws of Motion Momentum Universal Forces Fluid Pressure Hydraulics Buoyancy
Forces Definition Friction Falling Objects Projectiles Newton s Laws of Motion Momentum Universal Forces Fluid Pressure Hydraulics Buoyancy Definition of Force Force = a push or pull that causes a change
More informationLecture 24 - Surface tension, viscous flow, thermodynamics
Lecture 24 - Surface tension, viscous flow, thermodynamics Surface tension, surface energy The atoms at the surface of a solid or liquid are not happy. Their bonding is less ideal than the bonding of atoms
More informationInstruction Manual Service Program ULTRA-PROG-IR
Instruction Manual Service Program ULTRA-PROG-IR Parameterizing Software for Ultrasonic Sensors with Infrared Interface Contents 1 Installation of the Software ULTRA-PROG-IR... 4 1.1 System Requirements...
More informationE190Q Lecture 5 Autonomous Robot Navigation
E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator
More informationFigure 2.31. CPT Equipment
Soil tests (1) In-situ test In order to sound the strength of the soils in Las Colinas Mountain, portable cone penetration tests (Japan Geotechnical Society, 1995) were performed at three points C1-C3
More informationFitting Guide. Dallas 800-366-3933 Denver 800-362-4233 Tulsa 800-685-5367
Fitting Guide Dallas 800-366-3933 Denver 800-362-4233 Tulsa 800-685-5367 Sagittal Depth Fitting the Jupiter lens design is based upon sagittal depth. Relative sagittal depth is not just the steepness of
More informationDensity (r) Chapter 10 Fluids. Pressure 1/13/2015
1/13/015 Density (r) Chapter 10 Fluids r = mass/volume Rho ( r) Greek letter for density Units - kg/m 3 Specific Gravity = Density of substance Density of water (4 o C) Unitless ratio Ex: Lead has a sp.
More informationSignal to Noise Instrumental Excel Assignment
Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a
More informationManual Blood Pressure Measurement
Manual Blood Pressure Measurement 1 Peter Mahoney, CETL Clinical Skills Facilitator Measuring Blood Pressure manually can be difficult if not practised regularly. However it is important to be able to
More informationMETHOD STATEMENT HIGH STRIAN DYNAMIC TESTING OF PILE. Prepared by
METHOD STATEMENT HIGH STRIAN DYNAMIC TESTING OF PILE Prepared by Infratech ASTM CO., LTD. Contents Chapter Description Page Contents...... 1 List of Appendix. 1 1. Introduction.. 2 2. Test Method..2 3.
More informationMicro-Optical Sensor Use in Boundary Layer Flows with Polymers and Bubbles
2 nd International Symposium on Seawater Drag Reduction Busan, Korea, 23-26 MAY 2005 Micro-Optical Sensor Use in Boundary Layer Flows with Polymers and Bubbles D. Modarress, P. Svitek (Measurement Science
More informationPre-lab Quiz/PHYS 224 Magnetic Force and Current Balance. Your name Lab section
Pre-lab Quiz/PHYS 224 Magnetic Force and Current Balance Your name Lab section 1. What do you investigate in this lab? 2. Two straight wires are in parallel and carry electric currents in opposite directions
More informationLecture L22-2D Rigid Body Dynamics: Work and Energy
J. Peraire, S. Widnall 6.07 Dynamics Fall 008 Version.0 Lecture L - D Rigid Body Dynamics: Work and Energy In this lecture, we will revisit the principle of work and energy introduced in lecture L-3 for
More informationChapter 11 SERVO VALVES. Fluid Power Circuits and Controls, John S.Cundiff, 2001
Chapter 11 SERVO VALVES Fluid Power Circuits and Controls, John S.Cundiff, 2001 Servo valves were developed to facilitate the adjustment of fluid flow based on the changes in the load motion. 1 Typical
More informationMEASUREMENT OF END FACE GEOMETRY ON FIBER OPTIC TERMINI...2
MEASUREMENT OF END FACE GEOMETRY ON FIBER OPTIC TERMINI...2 IMPORTANCE OF END FACE GEOMETRY...2 FIBER OPTIC CONNECTOR END FACE GEOMETRY MEASUREMENT TECHNIQUES...2 INTERFEROMETRIC MICROSCOPE TYPES...3 MEASUREMENT
More informationSection 2.0 : Construction and Measurement of a Simple Test Transmission Line
Section 2.0 : Construction and Measurement of a Simple Test Transmission Line After deciding to use a Focal 8V 4412 mid-bass driver for my first transmission line design, I started looking for a simple
More informationThe Next Generation of Cable Technology. A technology primer from NORDX/CDT By, Eric d Allmen
A technology primer from NORDX/CDT By, Eric d Allmen Foreword The Telecommunications Industry Association (TIA) and the International Standards Organization (ISO/IEC) are actively engaged in the development
More informationSection Two: Arterial Pressure Monitoring
Section Two: Arterial Pressure Monitoring Indications An arterial line is indicated for blood pressure monitoring for the patient with any medical or surgical condition that compromises cardiac output,
More information4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:
4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing
More informationINTRODUCTION SOME USES OF SPRAY NOZZLES INTRODUCTION TYPES OF SPRAY NOZZLES
SOME USES OF SPRAY NOZZLES It is important that the nozzle you select is appropriate for your particular application Liquid sprays are used in a seemingly endless variety of applications Some of the more
More informationModule 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
More informationCirculation Stations
Circulation Stations This worksheet corresponds with stations around the classrooms. Work in groups of 3 and do the stations in any order. Name: Station 1: Blood smear under a microscope Materials: - Microscope
More informationBuilding Design for Advanced Technology Instruments Sensitive to Acoustical Noise
Building Design for Advanced Technology Instruments Sensitive to Acoustic Noise Michael Gendreau Colin Gordon & Associates Presentation Outline! High technology research and manufacturing instruments respond
More informationGLOBAL COLLEGE OF ENGINEERING &TECHNOLOGY: YSR DIST. Unit VII Fiber Optics Engineering Physics
Introduction Fiber optics deals with the light propagation through thin glass fibers. Fiber optics plays an important role in the field of communication to transmit voice, television and digital data signals
More informationROBOT SYSTEM FOR REMOVING ASBESTOS SPRAYED ON BEAMS
ROBOT SYSTEM FOR REMOVING ASBESTOS SPRAYED ON BEAMS Mitsunori Arai* and Haruo Hoshino Advanced Research Department R&D Institute, Takenaka Corp, Chiba, Japan * Corresponding author (arai.mitsunori@takenaka.co.jp)
More informationWOOD WEAR TESTING USING TRIBOMETER
WOOD WEAR TESTING USING TRIBOMETER Prepared by Duanjie Li, PhD 6 Morgan, Ste156, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard for tomorrow's materials. 2015 NANOVEA INTRO
More information