1 Abbott PXP Glucose Meters 2013 Resource: Lin Gustafson-MT, Point of Care Coordinator Laboratory

2 Objectives After you complete this Computer-Based Learning (CBL) module, you should be able to: Explain meter operation and maintenance. Describe glucose meter testing protocol. Describe documentation requirements for patient and control results. Explain basic glucose meter troubleshooting.

3 Basic Operation To operate the Abbott PXP glucose meters, you must know: How to use the Scan key. How to use the Clear button. When and how to replace batteries. How to enter the operator ID and the patient ID.

4 Basic Operation Scan Key The Scan key turns on the laser scanner to read bar codes. To operate, press and hold Scan key. Move the green beam over the bar code. A red light will flash. (older model meters have a red beam). The meter beeps when it reads bar code. Hold Scan key until meter verifies bar code ID. Slowly move the meter, if needed, so that the green (or red) beam is directly over the barcode.

5 Basic Operation Scan Key Precautions NEVER look into the bar code scanner. This may cause permanent eye damage. If you hold the Scan key for three seconds or more, the scanner stops. Reposition the scanner and try again. Hold the scanner 3-7 inches from the barcode at a 30 to 35 degree angle. To block glare, umbrella your free hand around the barcode while scanning.

6 Basic Operation Clear Button Use the Clear button to: Clear a bar code scanned by mistake. Back up one space. Clear a numerical code from the monitor after pressing enter. Return to a previous screen. Press Clear two times to clear all numbers shown in the display.

7 Basic Operation Battery Replacement A low battery power message is a CRITICAL alert. Change batteries immediately! Meter requires two AA batteries The battery indicator signal is located in the upper right hand corner of the meter screen.

8 Basic Operation Enter Operator ID To enter your operator ID: Scan your employee badge bar code to enter your operator ID. If your ID is invalid and your certification is current, place the PXP in a docking station to upload the meter again. This reinstalls the list of valid operators. If you have a problem, phone Lin Gustafson, POCC, at Never allow anyone to use your badge bar code. Remember you are legally responsible for all tests performed using your operator ID.

9 Basic Operation Enter Patient ID To enter the patient ID: Scan the patient s billing account number, found on the patient s wrist band. If you can t scan the wrist band, inform the charge nurse immediately to get a new wrist band for the patient. Scanning is for positive patient identification. Scanning prevents manual entry errors and results going to the wrong patient chart. If you must manually enter a number, double check it before you press the Enter button. This should be a rare issue.

10 Patient ID, cont. After scanning the patient s armband: If the patient s demographic information is available, the patient s name, DOB and sex will display under the patient account number on the meter screen. If the information is correct, press 2 to confirm. If the information is incorrect, press 1 to re-enter ID.

11 Patient ID, cont. If the patient s demographic information is unavailable, the message patient data not found displays on the screen under the patient account. Verify that the patient account number is correct. If the patient account number is correct, press 2 to continue. If the patient account number is incorrect, press 1 to re-enter ID.

12 Quality Control (QC) Testing When you open a new set of control solutions: Discard old set of controls. Write the new expiration date on the control bottles using the expiration labels in the meter boxes. For more labels call x22902 (GMC-L) or x26700 (GMC-D) DO NOT write on the barcode. Test low and high levels of controls once every twelve hours. Control solutions expire 90 days after opening or at the expiration date on the bottles, whichever comes first.

13 Performing a QC Test To perform a quality control test: Swirl the control solution several times before applying it to a test strip. Discard first drop to get rid of any air bubbles Apply and hold QC solution to test strip until the meter beeps Keep the meter in a horizontal position whenever performing patient or QC testing.

14 Out-of-Range QC Results If your QC test result is Fail, enter the appropriate comment code. For example: Repeat test (code #1) Procedure error (code #4) Check QC solution expiration date and fluid level. If QC solution is adequate, repeat test using same solution.

15 Out-of-Range QC Results, cont. If QC remains out of range, repeat test using new QC solution. If you are unable to resolve QC problem: Take meter out of service and exchange for a new meter. Notify Lin Gustafson, POCC at

16 Out-of-Range QC Results, cont. Possible causes of out-of-range QC results include: Air bubbles in control solution. Test strips or solutions compromised because of improper storage. QC solution opened for more than 90 days. Controls not mixed prior to testing.

17 Docking the Meter Dock the meter every four hours. Place the PXP glucose meter in the docking station after performing the QC test. A two-way data transfer occurs each time the meter is docked. The meter transmits patient and control results to the Abbott computer system. The computer system transmits new operator and lot information to the meter.

18 Data Upload Procedure Place the PXP in a docking station to start the data upload. Do not remove the meter until the arrows are replaced by the Menu Mode screen. The PXP meter displays: Upload successful, turning off.

19 Patient Blood Glucose Testing When obtaining a capillary finger stick blood glucose sample: 1. Use the patient s middle finger or ring finger. 2. To clean finger, wipe with alcohol. 3. Allow alcohol to air dry. Finger must be dry before puncture. 4. Puncture patient s finger. Avoid sides and tips of finger. Avoid squeezing the finger excessively.

20 Patient Blood Glucose Testing, cont. 5. Wipe away the first drop of blood. 6. Use a second drop for testing. Allow a nice size drop to form before applying to the test strip. Do not allow blood to enter the strip port. To avoid this, perform all testing with the meter in a horizontal position.

21 Patient Blood Glucose Testing Repeat Testing If you have to repeat the test because of an out of range result or error, you can use the same puncture site. 1. Wipe the puncture site with alcohol to get the blood flowing again. 2.Wipe away first drop. 3.Use second drop for repeat testing.

22 Venous / Arterial Blood Testing You must use venous or arterial specimens within 30 minutes of collection. If not used immediately, use only anti-coagulated samples,. Specimen Source Venous Arterial IV start Collect in: Green - or purple - top tube Syringe containing heparin or EDTA Special Instructions Use immediately to avoid specimen clotting. Micro clots interfere with biosensor technology

23 GHS Blood Glucose Ranges Adults: mg/dl Newborn: mg/dl Take action for EVERY result outside these ranges. The meter displays the result and an arrow pointing up (if >400) or down (if <60) and prompts you to input a comment code. Enter comment code #1 (repeat test). Repeat test immediately to verify first result. If the repeat test remains outside of the range, enter an appropriate comment code. Examples: Dr/charge nurse notified (code #11), hypoglycemia protocol (code #9).

24 GHS Blood Glucose Ranges, con t. A STAT lab draw is required for any result remaining >400 mg/dl after a repeat test. A STAT lab draw is recommended for any result: Remaining <60 mg/dl after a repeat test, and Not consistent with physical symptoms, and Technical errors have been ruled out. Click here to review Nursing Policy #6137, Diabetes and Blood Glucose Management. Click here to review Laboratory Policy # , Precision Glucose Systems.

25 Use Proper Test Strip Only use the test strip from the package you scan. Scanning the test strip bar code enters the strip lot number and expiration, control ranges and expiration, and calibration information into the meter. This is important because unique information is associated with each lot number.

26 Patient Results Possible causes for out-of-range patient results Lower than expected: Water or alcohol remaining on puncture site. Venous or arterial blood sample not tested within 30 minutes of collection. Patient is severely dehydrated, severely hypotensive, in shock or in a hyperglycemichyperosmolar state (with or without ketosis). Hematocrit is higher than the acceptable limit for the test strips, as indicated on the test strip package insert

27 Patient Results, con t Possible causes for out-of-range patient results Higher than expected: Serum or plasma used instead of whole blood. Hematocrit is lower than the acceptable limit for the test strips, as indicated on the test strip package insert. Document all patient results in the patient s chart.

28 Patient Test Review 1. Press the On/Off button to turn the monitor on. 2. Press the Menu button. 3. Press 1 to select Data Review 4. Scan operator ID 5. Choose the category of data to review

29 Unregistered Newborns You can find glucose logs for unregistered babies in both the NICU and Well Nursery. You may use an assigned number for the patient ID along with the patient s last name. Place the patient s label on the log as soon as possible. Fax log to laboratory daily:

30 Unregistered ED Patients You can find glucose logs for unregistered ED patients at different locations in the Lawrenceville and Duluth EDs. Please see managers for specific locations. Enter Date and Time (mmdd0000) on log under Number Used as the patient identifier. Example: Enter Jan 8 th at 1400 as: Record results and your operator ID on log. Place the patient s label on log ASAP. Fax log to Laboratory daily:

31 Testing Patients in Isolation Before Testing: Prepare all testing materials prior to entering the patients room. Place meter in a biohazard bag and seal the bag. Use any necessary Personal Protective equipment (PPE) to enter the room. Take ONLY your meter and necessary supplies into the room. DO NOT take entire meter case.

32 Testing Patients in Isolation, cont. Testing: 1. Scan/enter your operator ID. 2. Scan the patients armband. 3. Push the test strip through the bag, making a small hole. 4. Perform test.

33 Testing Patients in Isolation, cont. After Testing: Remove the meter from the room by sliding it out of the biohazard bag to: An isolation cart, -or- Available counter top, -or- Staff member outside of the room. Dispose of the biohazard bag, your PPE, and all testing materials BEFORE exiting the room. Wash your hands and decontaminate the meter AFTER exiting the room.

34 Meter Care To reduce meter down time and extend meter life: Avoid dropping the meter. Do not place meter on beds or cluttered tray tables. If the meter is accidentally dropped, verify correct date and time on meter screen and do QC testing to be sure it is operating properly before doing patient testing. Avoid getting QC solution or patient blood inside meter. Use small drops of control solution and patient blood. Keep meter in a horizontal (flat) position during testing.

35 Meter Care, cont. Avoid getting QC solution or patient blood inside meter, cont. Remove test strips before lifting the meter to read the screen. Excessive specimen may run down the strip, into the strip port opening and onto the circuit board. This results in a strip error where the meter will not recognize the test strip once it has been inserted or removed. Wetting the internal circuits repeatedly damages the meter and causes inaccurate test results.

36 Keeping the Meters Clean Clean the PXP meters with a hospital approved disinfectant wipe after each patient use. Follow directions for use as a disinfectant on back of wipe container. Turn the meter off when cleaning it. Clean the scanner window with an alcohol pad and wipe dry. Soiled or smudged windows do not scan properly DO NOT immerse, flood with any liquid or autoclave the meter.

37 Replacement Meters Replacement meters are available in the Lab. Bring the defective meter to the Lab for exchange along with a description of the problem. A description of the problem is necessary when the meter is sent back to the company.

38 Congratulations! You have completed this Computer-Based Learning (CBL) module. Click on Take Test to continue. Questions? Contact: Lin Gustafson Point of Care Coordinator, Laboratory

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 CentraCare Health (CCH) adopts the following policy/procedure for: St. Cloud Hospital Original: 5/02 Minor Revision: Full Review: 1/15 Replaces:

More information

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM Inform II BLOOD GLUCOSE MONITORING SYSTEM Quick Reference Guide for Healthcare Professionals 2 Table of Contents Important... 4 General Information... 6 ACCU-CHEK Inform II Meter...10 How to Perform Patient

More information

PROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

PROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version. Department Of Pathology POC.514.01- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 1 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.01 LBH 1 OF 7 Printed copies are

More information

A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved

A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved Overview and Objectives This module provides an introduction to

More information

Title: Fingerstick Glucose by Accu-Chek Inform

Title: Fingerstick Glucose by Accu-Chek Inform Title: Fingerstick Glucose by Accu-Chek Inform Target Audience: This module is available to aid in assessing competency for all clinical staff who perform fingerstick glucose testing. Contents Instructions...

More information

Hemoglobin Testing using the Data Management HemoCue Analyzer

Hemoglobin Testing using the Data Management HemoCue Analyzer The Johns Hopkins Medical Institutions Hemoglobin Testing using the Data Management HemoCue Analyzer A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program Revision: 12/02 copyright

More information

STANDARD OPERATING PROCEDURE FOR NOVA STAT STRIP CONNECTIVITY BLOOD GLUCOSE METERS FOR USE IN LEEDS NHS TRUST

STANDARD OPERATING PROCEDURE FOR NOVA STAT STRIP CONNECTIVITY BLOOD GLUCOSE METERS FOR USE IN LEEDS NHS TRUST STANDARD OPERATING PROCEDURE FOR NOVA STAT STRIP CONNECTIVITY BLOOD GLUCOSE METERS FOR USE IN LEEDS NHS TRUST This Standard Operating Procedure explains the protocol for measuring blood glucose concentration

More information

Key Features in the AgaMatrix Jazz

Key Features in the AgaMatrix Jazz Advanced Blood Glucose Monitoring System OWNER S GUIDE Key Features in the AgaMatrix Jazz No Coding Required - The AgaMatrix Jazz doesn t require you to code your meter. Insert a test strip and you are

More information

VENOUS BLOOD COLLECTION

VENOUS BLOOD COLLECTION Venous Blood Collection p. 1 of 5 Johns Hopkins Medical Laboratories Department of Pathology Core Lab Clinical Services Zd7: Word: JHH Venous Blood Collection Procedure Signature Date Signature Date Written

More information

SCANNING THE USPS SCAN BARCODE

SCANNING THE USPS SCAN BARCODE U.S.P.S. STANDARD OPERATING PROCEDURE SCANNING THE USPS SCAN BARCODE OVERVIEW: Printed on PS Form 8125, Plant-Verified Drop Shipment (PVDS) Verification and Clearance, the USPS Scan barcode will be used

More information

In-Service Plan for Blood Glucose Testing. www.poc.roche.com

In-Service Plan for Blood Glucose Testing. www.poc.roche.com In-Service Plan for Blood Glucose Testing www.poc.roche.com Table of Contents Section 1: Introduction Introduction / How to use this guide 3-4 Section 2: Pre-In-Service Activities

More information

Blood Glucose Monitoring System USER GUIDE. Replaces Owner s Booklet

Blood Glucose Monitoring System USER GUIDE. Replaces Owner s Booklet Blood Glucose Monitoring System USER GUIDE Replaces Owner s Booklet AW 06650201A Rev. date: 07/2009 Contents: Getting to know your system 1 Setting up your meter 4 Testing your blood glucose 8 Attaching

More information

Blood Glucose Monitoring System. Owner's Booklet. Instructions for use.

Blood Glucose Monitoring System. Owner's Booklet. Instructions for use. Blood Glucose Monitoring System Owner's Booklet Instructions for use. Contents covered by one or more of the following U.S. patents: 6,179,979, 6,193,873, 6,284,125, 6,716,577, 6,749,887, 6,797,150, 6,863,801,

More information

INTRODUCTION TO CLINICAL PRACTICE AND CLINICAL SKILLS 2nd Year MEDICAL YEAR 2009/2010

INTRODUCTION TO CLINICAL PRACTICE AND CLINICAL SKILLS 2nd Year MEDICAL YEAR 2009/2010 INTRODUCTION TO CLINICAL PRACTICE AND CLINICAL SKILLS 2nd Year MEDICAL YEAR 2009/2010 POINT OF CARE TESTING This session is designed to teach you the principals of point of care testing. This is common

More information

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of the blood 7 Procedure to test glucose 7 Procedure to

More information

41087.291 Inform II Glucose Meter

41087.291 Inform II Glucose Meter 41087.291 Inform II Glucose Meter Copy of version 1.5 (approved and current) Last Approval or Periodic Review Completed Next Periodic Review Needed On or Before 2/9/2015 2/9/2017 Effective Date 2/9/2015

More information

Nova Max Blood Glucose Monitor Owner s Guide

Nova Max Blood Glucose Monitor Owner s Guide Nova Max Blood Glucose Monitor Owner s Guide Nova Biomedical 200 Prospect Street Waltham, MA 02454-9141 U.S.A. Telephone: 1-800-681-7390 Web site: www.novacares.com Made in the USA by Nova Biomedical Corporation

More information

Caution! Please read complete Owner s Booklet and all product Instructions for Use.

Caution! Please read complete Owner s Booklet and all product Instructions for Use. INTRODUCTION: Liberty Blood Glucose Monitoring System The Liberty System is a simple, accurate way to test whole blood glucose (sugar) level, anytime, anywhere. Our goal is to provide quality healthcare

More information

Suggested BiliChek Usage Protocol

Suggested BiliChek Usage Protocol Suggested BiliChek Usage Protocol I. Subject: BiliChek Non-Invasive Bilirubin Analyzer II. Purpose: To outline and define the use of the BiliChek Non-Invasive Bilirubin Analyzer in the hospital setting

More information

Blood Glucose Monitoring System USER GUIDE

Blood Glucose Monitoring System USER GUIDE Blood Glucose Monitoring System USER GUIDE AW 06629001B Rev. date: 07/2009 Contents: Getting to know your system 1 Setting the time and date, and coding your meter 4 Testing your blood glucose 8 Reviewing

More information

CoaguChek System. XS Plus. User Manual

CoaguChek System. XS Plus. User Manual CoaguChek System XS Plus User Manual The contents of this manual, including all graphics and photographs, are the property of Roche Diagnostics. No part of this document may be reproduced or transmitted

More information

Mobile App User Guide

Mobile App User Guide Mobile App User Guide Contents 2 Contents Introduction... 4 OneTouch Reveal mobile app...4 Supported Devices and System Requirements...4 Android Legal Notice...4 Google Play Legal Notice...4 Bluetooth

More information

MANITOBA PATIENT SERVICE CENTRE STANDARDS

MANITOBA PATIENT SERVICE CENTRE STANDARDS MANITOBA PATIENT SERVICE CENTRE STANDARDS February 2015 INTRODUCTION These Standards are derived from Z316.7-12 and are approved by the Council of the College of Physicians and Surgeons of Manitoba. These

More information

digital fever alert ear thermometer suitable from birth instruction manual

digital fever alert ear thermometer suitable from birth instruction manual digital fever alert ear thermometer suitable from birth instruction manual digital fever alert ear thermometer instruction manual Please read these instructions carefully before using the thermometer and

More information

STEP-BY-STEP INSTRUCTIONS FOR INVESTIGATIONAL USE. Rapid HCV Antibody Test FOR ORAQUICK RAPID HCV ANTIBODY TEST

STEP-BY-STEP INSTRUCTIONS FOR INVESTIGATIONAL USE. Rapid HCV Antibody Test FOR ORAQUICK RAPID HCV ANTIBODY TEST Before performing testing, all operators MUST read and become familiar with Universal Precautions for Prevention of Transmission of Human Immunodeficiency Virus, Hepatitis B Virus, and other Blood-borne

More information

Preface. Preface. A Healthcare professional should be contacted when Customer Service is not available.

Preface. Preface. A Healthcare professional should be contacted when Customer Service is not available. Preface Preface Thank you for selecting the GE100 Monitoring System. This manual provides all the information you need to operate this product for accurate test results. Please read the entire manual before

More information

FLASH GLUCOSE MONITORING SYSTEM. User s Manual

FLASH GLUCOSE MONITORING SYSTEM. User s Manual FLASH GLUCOSE MONITORING SYSTEM User s Manual Contents Reader Symbols... 1 Important Safety Information... 3 Indications for Use... 3 Contraindications... 4 Getting to Know Your System...........................

More information

Sheep@Purdue. Blood Sampling in Sheep AS-557-W

Sheep@Purdue. Blood Sampling in Sheep AS-557-W Sheep@Purdue AS-557-W Blood Sampling in Sheep Becky Mitchell, Animal Sciences Student Mike Neary, Extension Sheep Specialist Gerald Kelly, Manager of Sheep Teaching and Research Flock Purdue University

More information

Inpatient Cerner Navigation and Documentation For Nursing Students

Inpatient Cerner Navigation and Documentation For Nursing Students Inpatient Cerner Navigation and Documentation For Nursing Students Audience Note: Purpose: Objectives: Cerner PowerChart training is for all students in the following inpatient areas Med/Surg, OSNO, Oncology,

More information

USAccess System- Registrar. Help Guide. Prepared for

USAccess System- Registrar. Help Guide. Prepared for USAccess System- Registrar Help Guide Prepared for United States Department of Agriculture Office of Security 300 7th Street SW, Washington DC 20024 Version 1.0 December 12, 2007 USAccess System Registrar

More information

Sync Guide. Sync Overview. Before the Event. During the Event. After the Event. Greater Giving Event Software

Sync Guide. Sync Overview. Before the Event. During the Event. After the Event. Greater Giving Event Software Sync Guide Sync Overview Timeline Equipment Terms + This guide will instruct you on how to transfer data between Greater Giving Event Software and Auctionpay Master terminal Before, During and After your

More information

INSTRUCTIONS FOR USE HUMIRA 40 MG/0.8 ML, 20 MG/0.4 ML AND 10 MG/0.2 ML SINGLE-USE PREFILLED SYRINGE

INSTRUCTIONS FOR USE HUMIRA 40 MG/0.8 ML, 20 MG/0.4 ML AND 10 MG/0.2 ML SINGLE-USE PREFILLED SYRINGE INSTRUCTIONS FOR USE HUMIRA (Hu-MARE-ah) (adalimumab) 40 MG/0.8 ML, 20 MG/0.4 ML AND 10 MG/0.2 ML SINGLE-USE PREFILLED SYRINGE Do not try to inject HUMIRA yourself until you have been shown the right way

More information

Appendix H IBC Managing Biohazardous Waste SOP

Appendix H IBC Managing Biohazardous Waste SOP Biohazardous waste is managed under the State of Rhode Island s Regulated Medical Waste Regulations (Regulation DEM-OWM-MW-1-2009, amended July, 2010). http://www.dem.ri.gov/pubs/regs/regs/waste/medwaste10.pdf

More information

The Golden Rule of Specimen Collection: The Patient Test Result is Only as Good as the Sample We Get

The Golden Rule of Specimen Collection: The Patient Test Result is Only as Good as the Sample We Get The Golden Rule of Specimen Collection: The Patient Test Result is Only as Good as the Sample We Get Jan Frerichs, MLS (ASCP) The University of Iowa Janice-frerichs@uiowa.edu Importance of Phlebotomist

More information

PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly

PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly PATIENT GUIDE Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Every drainage Weekly Clinician s Signature: ACCESS SYSTEMS Pleural Space Insertion Site Cuff Exit Site Catheter

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Equipment / Procedure: Filename: Assessor s name & date: Reviewer s name & date: Cosmed K4 2012.1201SOP_cosmed_K4 Richard Metcalfe, December 2012 Tom Nightingale, December

More information

Key. ➍ Micro USB Port ➎ Operating System Toggle Keys ➏ Foam Screen Protectors. ➊ On/Off switch ➋ Bluetooth Connect Button (flashes when searching)

Key. ➍ Micro USB Port ➎ Operating System Toggle Keys ➏ Foam Screen Protectors. ➊ On/Off switch ➋ Bluetooth Connect Button (flashes when searching) INSTRUCTION MANUAL ➏ ➋ ➊ Product Features ➎ ➍ ➌ Built-in wireless Bluetooth 3.0 keyboard Compatible across ios, Android, and Windows, enabling you to mix and match with devices Ultra-thin, lightweight

More information

Appendix H Managing Biohazardous Waste SOP

Appendix H Managing Biohazardous Waste SOP Biohazardous waste is managed under the State of Rhode Island s Regulated Medical Waste Regulations (Regulation DEM-OWM-MW-1-2009, amended July, 2010). http://www.dem.ri.gov/pubs/regs/regs/waste/medwaste10.pdf

More information

CoaguChek System. User Manual. This is a CLIA waived system.

CoaguChek System. User Manual. This is a CLIA waived system. CoaguChek System XS User Manual This is a CLIA waived system. Table of Contents About this Manual...1 Symbols and Abbreviations...1 User Resources...2 Video...2 Getting Started...2 User Manual...2 Test

More information

BLOOD GLUCOSE METER. Standard Owner s Booklet for Single Patient Use Only

BLOOD GLUCOSE METER. Standard Owner s Booklet for Single Patient Use Only BLOOD GLUCOSE METER Standard Owner s Booklet for Single Patient Use Only Before You Start Testing About the meter and test strips Carefully read and follow the instructions in the Getting Started Guide,

More information

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE TAKING BLOOD FROM INFANTS FOR THE HIV DNA PCR TEST STANDARD OPERATING PROCEDURE ACRONYMS ARV CCMT DBS Antiretrovirals Comprehensive care, management and treatment Dried blood spots This Standard Operating

More information

Wireless Alarm System. Panic Button. User s Manual. Choice ALERT. Control all sensors, accessories and remote transmitters from one location

Wireless Alarm System. Panic Button. User s Manual. Choice ALERT. Control all sensors, accessories and remote transmitters from one location Wireless Alarm System Panic Button User s Manual Choice ALERT Control all sensors, accessories and remote transmitters from one location Table of Contents Important Safeguards 4 Introduction 5 Activating

More information

Table of Contents Function Keys of Your RF Remote Control Quick Setup Guide Advanced Features Setup Troubleshooting

Table of Contents Function Keys of Your RF Remote Control Quick Setup Guide Advanced Features Setup Troubleshooting Congratulations on your purchase of the AT&T U-verse TV Point Anywhere RF Remote Control. This product has been designed to provide many unique and convenient features to enhance your AT&T U-verse experience.

More information

1-800-862-2731 Administration of Meropenem For Child

1-800-862-2731 Administration of Meropenem For Child 1 Drug Name: _Meropenem 1-800-862-2731 Administration of Meropenem For Child Dosage: milligrams every hours over 5 minutes Key Points: FLUSHING: Meropenem Heparin flush 1. Always wash your hands with an

More information

SCM Training For Nursing Staff

SCM Training For Nursing Staff SCM Training For Nursing Staff UK Hosp, Staff Education August 06 1 Resources and Help Support Center: 323-8586 Useful Web Sites Learning Center Web page http://www.mc.uky.edu/learningcenter ICIS Web page

More information

Getting started guide 3G Turbo Stick. 3G Novatel Wireless U760 USB modem

Getting started guide 3G Turbo Stick. 3G Novatel Wireless U760 USB modem Getting started guide 3G Turbo Stick 3G Novatel Wireless U760 USB modem Welcome Thank you for purchasing the 3G Novatel Wireless U760 USB modem from Bell. This two-in-one Turbo Stick with memory storage

More information

Laboratory Biosafty In Molecular Biology and its levels

Laboratory Biosafty In Molecular Biology and its levels Laboratory Biosafty In Molecular Biology and its levels Workshop 16-17 Oct..2012 Guidelines Does not mean optional Laboratory Biosafety The Laboratory Biosafety Manual is an important WHO publication

More information

CONTINUOUS GLUCOSE MONITORING SYSTEM USER S GUIDE

CONTINUOUS GLUCOSE MONITORING SYSTEM USER S GUIDE CONTINUOUS GLUCOSE MONITORING SYSTEM USER S GUIDE CONTINUOUS GLUCOSE MONITORING SYSTEM IMPORTANT CONTACTS AND NUMBERS Dexcom Website: www.dexcom.com Your Transmitter ID: Your Receiver ID: Your Healthcare

More information

Why is Insulin so Important?

Why is Insulin so Important? Insulin Therapy Why is Insulin so Important? If the glucose stays in your blood it doesn t do your cells (body) any good The glucose has to get inside the cells for the body to use it What Does Insulin

More information

Blood Glucose Monitoring System Owner s Booklet Instructions for use

Blood Glucose Monitoring System Owner s Booklet Instructions for use Blood Glucose Monitoring System Owner s Booklet Instructions for use For use with the App on iphone, ipod touch and ipad Blood Glucose Monitoring System Owner s Booklet 1 Owner s Booklet Thanks for choosing

More information

Table of Contents. GETTING STARTED...1 Technical Support...1 Scanner Configuration...2 Login and Select a Mode...4

Table of Contents. GETTING STARTED...1 Technical Support...1 Scanner Configuration...2 Login and Select a Mode...4 Barcode Scanner Table of Contents GETTING STARTED...1 Technical Support...1 Scanner Configuration...2 Login and Select a Mode...4 USING THE SCANNER...7 Basic Functionality...7 Shipping or Receiving Mode...8

More information

Policy and Procedure Flushing and / or Blood withdrawal Aspiration Procedure For PICC Line and Midline Catheters

Policy and Procedure Flushing and / or Blood withdrawal Aspiration Procedure For PICC Line and Midline Catheters Policy and Procedure Flushing and / or Blood withdrawal Aspiration Procedure For PICC Line and Midline Catheters Purpose: Blood Withdrawal: To obtain blood samples for laboratory evaluation, eliminating

More information

INSTRUCTIONS FOR USE HUMIRA 40 MG/0.8 ML SINGLE-USE PEN

INSTRUCTIONS FOR USE HUMIRA 40 MG/0.8 ML SINGLE-USE PEN INSTRUCTIONS FOR USE HUMIRA (Hu-MARE-ah) (adalimumab) 40 MG/0.8 ML SINGLE-USE PEN Do not try to inject HUMIRA yourself until you have been shown the right way to give the injections and have read and understand

More information

Instructions for Use PROCRIT (PRO KRIT) (epoetin alfa)

Instructions for Use PROCRIT (PRO KRIT) (epoetin alfa) Instructions for Use PROCRIT (PROKRIT) (epoetin alfa) Use these Instructions for Use if you or your caregiver has been trained to give PROCRIT injections at home. Do not give yourself the injection unless

More information

Virtual Heart User Manual Username Password

Virtual Heart User Manual Username Password Virtual Heart User Manual Username Password These instructions are meant to help you use the Virtual Heart website. Please write down your username and password and store them in a safe place in case you

More information

16.1 REQUIREMENTS AND INSTALLATION

16.1 REQUIREMENTS AND INSTALLATION EXTENDED SERVICE OPTIONS CHAPTER 16 16.0 OVERVIEW The MICR Reader Module is an Extended Service Option that simplifies the input of NSF checks into Debt$Net. When using the MICR Reader Module, you scan

More information

Central Line Blood Draw

Central Line Blood Draw Central Line Blood Draw Welcome to the Central Line dressing blood draw refresher. Please use the navigation below to advance to the next page. The Central Line blood draw module is also available as a

More information

ClikSTAR - Important facts about your new insulin delivery device.

ClikSTAR - Important facts about your new insulin delivery device. ClikSTAR - Important facts about your new insulin delivery device. Instruction for Use ClikSTAR Insulin delivery device Before you start: Read these instructions and follow them completely each time you

More information

PATIENT GUIDE. Understand and care for your peripherally inserted central venous catheter (PICC). MEDICAL

PATIENT GUIDE. Understand and care for your peripherally inserted central venous catheter (PICC). MEDICAL PATIENT GUIDE Understand and care for your peripherally inserted central venous catheter (PICC). MEDICAL Introduction The following information is presented as a guideline for your reference. The best

More information

e 1 0 0 I N S T RUC TION M ANU A L

e 1 0 0 I N S T RUC TION M ANU A L e 1 0 0 I N S T RUC TION M ANU A L The Empire Digital Level Accuracy, Readability, Durability one FEATURES: Easy to read back light Display reads right side up even if the level is upside down Audio beep

More information

Infectious Waste Management Plan

Infectious Waste Management Plan Infectious Waste Management Plan Infectious Waste Management Plan USC Health & Safety Programs Unit 777-5269 POLICY: A. In keeping with the University of South Carolina's policy of providing protection

More information

The OmniPod Insulin Management System

The OmniPod Insulin Management System Caregiver GUIDE The OmniPod Insulin Management System The OmniPod is an easy-to-use, two-part insulin delivery system. If you are a school nurse, daycare provider, or other secondary caregiver for someone

More information

Wireless Laser Barcode Scanner ils 6300BU. User s Manual

Wireless Laser Barcode Scanner ils 6300BU. User s Manual Wireless Laser Barcode Scanner ils 6300BU User s Manual FCC Compliance This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.

More information

The FreeStyle Lite Meter and How It Works

The FreeStyle Lite Meter and How It Works ENGLISH The FreeStyle Lite Meter and How It Works System Check Screen This screen always appears when the meter is turned on so that you can make sure the display is working properly. Do not use the meter

More information

EXERCISE 3: SKIN PUNCTURE ON INFANTS

EXERCISE 3: SKIN PUNCTURE ON INFANTS EXERCISE 3: SKIN PUNCTURE ON INFANTS Textbook: Skills: Chapter 12 Pediatric Procedures Dry Lab Objectives: 1. State the composition of skin puncture blood 2. List five (5) tests that cannot be preformed

More information

Table of Contents. Use. Troubleshooting. Setup. Welcome. 11 How to arm/disarm system/camera(s) 19 Sync Module setup issues. 3 Installing the Blink app

Table of Contents. Use. Troubleshooting. Setup. Welcome. 11 How to arm/disarm system/camera(s) 19 Sync Module setup issues. 3 Installing the Blink app User Guide Table of Contents Welcome Setup Use Troubleshooting 2 What s in the box 2 What you need 3 Installing the Blink app 4 Setting up the Sync Module 5 Connecting the Sync Module to Your Wi-Fi 7 Adding

More information

SD CodeFree BLOOD GLUCOSE MONITORING SYSTEM

SD CodeFree BLOOD GLUCOSE MONITORING SYSTEM Dear SD CodeFree System Owner; Thank you for choosing the SD CodeFree Blood Glucose Monitoring System. Your new SD CodeFree Blood Glucose Monitoring System is an important tool that can help you better

More information

How To Use An Easypad On A Pc Or Macbook Or Ipad (For Pc)

How To Use An Easypad On A Pc Or Macbook Or Ipad (For Pc) User Guide Congratulations on your new EasyPad 1000. Please follow the instructions in this manual carefully. Statement: If the final version of this product has any differences from the description of

More information

Self-Monitoring of Blood Glucose

Self-Monitoring of Blood Glucose My Doctor Says I Should Monitor My Blood Glucose... What Does This Mean? BD Getting Started Self-Monitoring of Blood Glucose Daily Blood Sugar Monitoring When you have diabetes, managing your blood glucose

More information

myvisiontrack Cleared User s Manual Ophthalmic Medical Device Caution: Federal Law restricts this device to sale by or on order

myvisiontrack Cleared User s Manual Ophthalmic Medical Device Caution: Federal Law restricts this device to sale by or on order myvisiontrack Cleared User s Manual Ophthalmic Medical Device Caution: Federal Law restricts this device to sale by or on order of a licensed healthcare practitioner Copyright Copyright 2014 Vital Art

More information

Sierra Wireless AirCard Watcher Help for Mac OS X

Sierra Wireless AirCard Watcher Help for Mac OS X Sierra Wireless AirCard Watcher Help for Mac OS X Sierra Wireless AirCard Watcher allows you to manage and monitor the connection between your modem and the network. With Watcher, you can: Determine signal

More information

ES&S AutoMARK. Pre-Election Day Checklist

ES&S AutoMARK. Pre-Election Day Checklist ES&S AutoMARK Pre-Election Day Checklist Rear Access Door Touch Screen Monitor Lid Compact Flash Access Door Lid Latches Key-Operated Mode Switch Keypad Ballot Feed Tray ADA Port Headphone Jacks 1 Checking

More information

Thank you for purchasing the Huawei E5332 Mobile WiFi. This quick start guide will help you to get started right away.

Thank you for purchasing the Huawei E5332 Mobile WiFi. This quick start guide will help you to get started right away. Thank you for purchasing the Huawei E5332 Mobile WiFi. This quick start guide will help you to get started right away. You can connect your Huawei E5332 to the internet through any WiFi enabled device

More information

FrontDesk Installation And Configuration

FrontDesk Installation And Configuration Chapter 2 FrontDesk Installation And Configuration FrontDesk v4.1.25 FrontDesk Software Install Online Software Activation Installing State Related Databases Setting up a Workstation Internet Transfer

More information

Sterile Dressing Change with Tegaderm CHG for Central Venous Catheter (CVC)

Sterile Dressing Change with Tegaderm CHG for Central Venous Catheter (CVC) Sterile Dressing Change with Tegaderm CHG for Central Venous Catheter (CVC) The dressing protects your catheter site. It also helps prevent infection at the site. Keep your dressing clean and dry at all

More information

1. Preparing thin and thick blood films with capillary or venous blood

1. Preparing thin and thick blood films with capillary or venous blood Methods in Parasitology 1. Preparing thin and thick blood films with capillary or venous blood Swiss Tropical Institute, Basel April 2005 1.0 Preparing thin and thick blood films Contents 1.1 Preparatory

More information

User manual TAB-10C010-232

User manual TAB-10C010-232 User manual TAB-10C010-232 Congratulations on the purchase of your new tablet! This manual contains important safety and operating information in order to prevent accidents! Please read this manual thoroughly

More information

Universal Remote Instruction Manual

Universal Remote Instruction Manual 24959 Universal Remote Instruction Manual Table of Contents Introduction...3 Button Functions...4 Initial Setup...6 Battery Installation...6 Battery Saver...6 Code Saver...6 Code Setup...6 Direct Code

More information

Digital Persona Fingerprint Reader Installation

Digital Persona Fingerprint Reader Installation Digital Persona Fingerprint Reader Installation The link to download the Fingerprint Reader Software for AXIS-ACH is http://corpcu.com/fingerprint-reader-software This will begin the download for the drivers.

More information

VisTablet Manual. For 12 VT Original and Mini Mouse

VisTablet Manual. For 12 VT Original and Mini Mouse VisTablet Manual For 12 VT Original and Mini Mouse I n d e x 1. Before using your VisTablet ------------------------------------------------------------------------------- Page#3 2. How to install Pen

More information

Blood Collection and Processing SOP

Blood Collection and Processing SOP Brisbane Breast Bank Blood Collection and Processing SOP Breast Pathology Laboratory University of Queensland Centre for Clinical Research Blood Collection We collect 30ml of blood from patients who have

More information

Mobile App User Guide

Mobile App User Guide Mobile App User Guide Contents 2 Contents Introduction... 4 OneTouch Reveal mobile app... 4 Supported Devices and System Requirements...4 Apple Legal Notice...4 Bluetooth Trademark...4 Intended Use...4

More information

User Guide. Quality Check Messages and Codes (continued) (01)00054749001286. Message Cause Action

User Guide. Quality Check Messages and Codes (continued) (01)00054749001286. Message Cause Action Quality Check Messages and Codes (continued) Message Cause Action i-stat Analyzer Error, Use Electronic Simulator Analyzer Error, See Manual No display Cartridge Locked does not disappear after test cycle

More information

Part 2. Copyright 1998 Philips Consumer Communications L.P. All rights reserved. Printed in Mexico. Issue 1AT&T 848229506

Part 2. Copyright 1998 Philips Consumer Communications L.P. All rights reserved. Printed in Mexico. Issue 1AT&T 848229506 2 User's Manual for Two-Line Digital Answering System Telephone with Speakerphone 1872 Fold open this paper for information about this telephone's installation and operation. Please read Part 1 Important

More information

PA500II Enterprise Mobile Computer

PA500II Enterprise Mobile Computer PA500II Enterprise Mobile Computer - PA500II - Quick Reference Guide 400891G Version 1.0 PA500II Product Introduction & Accessory Kit After opening the box, ensure the following accessories for the PA500II

More information

BLOOD GLUCOSE MONITORING SYSTEM USER GUIDE

BLOOD GLUCOSE MONITORING SYSTEM USER GUIDE BLOOD GLUCOSE MONITORING SYSTEM USER GUIDE Thank you for choosing the CONTOUR blood glucose monitoring system! We are proud to be your partner in helping you manage your diabetes. Our goal is to make this

More information

BODY FAT & HYDRATION % MONITOR SCALE

BODY FAT & HYDRATION % MONITOR SCALE EVALUATION BODY FAT & HYDRATION % MONITOR SCALE INSTRUCTION MANUAL MODEL 12-1190 ver 7/08 BODY FAT & HYDRATION % MONITOR SCALE Introduction and Precaution: MODEL: 12-1190 1. Congratulations for purchasing

More information

Wireless Alarm System. Wireless Keypad. User s Manual. Choice ALERT. Control all sensors, accessories and remote transmitters from one location

Wireless Alarm System. Wireless Keypad. User s Manual. Choice ALERT. Control all sensors, accessories and remote transmitters from one location Wireless Alarm System Wireless Keypad User s Manual 1 2 3 4 Choice ALERT Control all sensors, accessories and remote transmitters from one location Table of Contents Important Safeguards 4 Introduction

More information

How To Prepare and Give a Prefilled Syringe Injection

How To Prepare and Give a Prefilled Syringe Injection Diablo Valley Onc&Hem Med Grp, Inc Phone Number: (925)6775041 How To Prepare and Give a Prefilled Syringe Injection Patient Education Quick Reference Guide The following are instructions for use of prefilled

More information

Troubleshooting: Door Jamming and Door Handing 6. Troubleshooting: Keypad 11. Troubleshooting: Battery 12. Battery FAQ 13

Troubleshooting: Door Jamming and Door Handing 6. Troubleshooting: Keypad 11. Troubleshooting: Battery 12. Battery FAQ 13 Programming and Troubleshooting Guide 1 2 3 4 5 Mastercode 2 Troubleshooting: Door Jamming and Door Handing 6 Troubleshooting: Keypad 11 Troubleshooting: Battery 12 Battery FAQ 13 1 / 13 1 Mastercode 62543

More information

Electronic Documentation/BMV Training For Nursing Students and Instructors. Tammy Galindo MSN/ed, RN Education Coordinator

Electronic Documentation/BMV Training For Nursing Students and Instructors. Tammy Galindo MSN/ed, RN Education Coordinator Electronic Documentation/BMV Training For Nursing Students and Instructors Tammy Galindo MSN/ed, RN Education Coordinator 1 Mission Statement Madera Community Hospital is a not-for-profit community health

More information

9.Pediatric Procedures

9.Pediatric Procedures 9.Pediatric Procedures A. Introduction 1. Pediatric blood collection may be by skin puncture or venipuncture. 2. Skill in pediatric phlebotomy is gained by knowledge of special collection equipment, observation

More information

PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly

PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly PATIENT GUIDE Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Every drainage Weekly Clinician s Signature: ACCESS SYSTEMS Pleural Space Insertion Site Cuff Exit Site Catheter

More information

ZOLL AED Plus Corrective Action CUSTOMER Questions and Answers

ZOLL AED Plus Corrective Action CUSTOMER Questions and Answers ZOLL AED Plus Corrective Action CUSTOMER Questions and Answers Q: How can I tell if my AED Plus is subject to the Corrective Action? A: Look at the unique serial number, containing letters and digits under

More information

HEMOCUE Procedure 1.0. S. Terence Dunn, PhD. 23-Apr-2014 12:33. 23-Apr-2016

HEMOCUE Procedure 1.0. S. Terence Dunn, PhD. 23-Apr-2014 12:33. 23-Apr-2016 County Health Department Laboratory Procedures Title: HEMOCUE Procedure Authors: Version: Authorised By: Date Time Of Last Update: Next Review Due On: Leslie Billetter and S. Terence Dunn, PhD 1.0 S. Terence

More information

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date: PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA

More information

The easy way to accept EFTPOS, Visa and MasterCard payments on the spot. Mobile Users... 2. Charging your PayClip. 2. Downloading the PayClip app.

The easy way to accept EFTPOS, Visa and MasterCard payments on the spot. Mobile Users... 2. Charging your PayClip. 2. Downloading the PayClip app. PayClip User Guide The easy way to accept EFTPOS, Visa and MasterCard payments on the spot. Contents Getting started made easy 2 Information for Merchants....................................................2

More information

Your Guide to Peritoneal Dialysis Module 3: Doing Peritoneal Dialysis at Home

Your Guide to Peritoneal Dialysis Module 3: Doing Peritoneal Dialysis at Home Your Guide to Peritoneal Dialysis Module 3: 6.0959 in Preparing to do PD One of the most important things about PD is to keep the dialysis area and anything that comes in contact with the PD equipment

More information

Ward 29 guide to the safe preparation and administration of intravenous (IV) antibiotics at home

Ward 29 guide to the safe preparation and administration of intravenous (IV) antibiotics at home Ward 29 guide to the safe preparation and administration of intravenous (IV) antibiotics at home Contents Page Important contact numbers 1 General information on preparing and administering IV antibiotics

More information

To maintain a port of entry to venous flow when all available peripheral ports have failed.

To maintain a port of entry to venous flow when all available peripheral ports have failed. I. Purpose: To maintain a port of entry to venous flow when all available peripheral ports have failed. II. General Comments: Since its development, these catheters have been used with increasing frequency

More information