2 BIOTECH Total Knee Minimal Invasive (and classic) Surgical Technique Page No. Product description 2. Pre-operational planning 3. 1st step Femoral resection 4. P/S Femoral component 7. 2nd step Proximal tibial resection The extramedullary technique 8. Proximal tibial resection The intramedullary technique 10. 3rd step Fixing up the trial femoral component 13. 4th step Trial replacement Checking of the flexion and extension joint gap 16. 5th step Final preparation of the tibia plateau 17. TMB Tibial Component 20. TMB primary tibial surgery 20. TMB Tibial Tray implantation 21. TMB P/S tibial surgery 23. 6th step Mounting of the original implants 24. 7th step Preparation of the patella 27. 1

3 Product Description: Cruciate Retaining and Posterior Stabilized Femoral Components Both, the Cruciate Retaining and the Posterior Stabilized femoral component types are available in 7 different sizes : XXS, XS, S, M, L, XL, XXL (optionally, smaller and bigger sizes are also available) in left and right anatomical designs,cemented and uncemented, made of Cobalt Chrome raw material. Compatibility of sizes Each femoral component is compatible also with the next size of tibial component (one size lower or upper). Tibial base components and inserts Cemented and uncemented tibia plate versions made of Cobalt Chrome are available with Cruciate Retaining or Posterior Stabilized UHMWPE inserts, in different sizes. The wing - designed tibial platform ensures a stable and long-lasting fixation. Patella component The 1 pegged, dome-shaped patellar component is available in cemented UHMWPE version. The correct placement and long-lasting fixation are ensured by the introduction of the peg. 2

4 Pre-operational planning When performing a complete joint replacement, two aims are to be considered primarily: The full shaft adjustment, and the restoration of function. The correct adjustment of the limb is ensured by using the appropriate instrumentation sets. For the sake of restoring full functionality to the joint the surgeon has to properly adjust the arches of the loins. For the right determination of the tibial and femoral mechanical and anatomical axis adjustments, it is recommended to prepare an X-ray of the long leg axis. The mechanical axis extends medially from the femoral head along the centre of the knee, and up to the middle of the ankle. On the basis of the X-ray made, it can be assured that the use of intramedullary devices will not be obstructed by the presence of a bone anomaly or the implanted prosthesis. The tibiofemoral valgus angle can be measured with the help of a goniometer, or with an angle gauge. The angle is usually 5 to 10 degrees. The femoral outer reaction line is at right angles to the mechanical line, and can be found parallel to the joint line. The maximum tolerated angle deviation from the appropriate axis placement should not exceed 3, any higher value could lead to the damage of the prosthesis. For the determination of the approximate size of prosthesis, it is advisable to use an x-ray template. Moreover, the likely use of bone material or bone lining blocks for the reconstruction should be determined pre-operatively. Biotech MTKS Minimally Invasive Total Knee System instruments set, is suitable for implanting the femoral and tibial components (and the patellar component, if needed) by classic and minimal invasive technique. After opening the area by the usual technique, the front surface of the femur is freed up by performing a partial Synovectomy, thereby making it possible to take the correct measurements. Position the retractors in place, and then cut off the meniscuses and the front part of the cruciate ligaments. Remove any osteophytes that may act as obstacles in the way of the aiming device. 3

5 1st step Femoral resection With the help of a drill (Picture 1.) or an awl open up the intramedullary canal. Stay along the axis area of the marrow. Then assemble together the slotted rod ( ) with the T-handle (Picture 2.), and insert the rod carefully into the intramedullary canal (Picture 3.) with rotating movements, until the coned end reaches the isthmus. Now remove the T-handle, and fit the distal femoral cutting guide ( ) on to the rod. (Picture 4.) 4

6 With this, set the desired degree of valgus axis position. In the right position fix the cutting, block through the holes by pins. In the next step remove the rod from intramedullary canal. Then using an oscillating saw perform the distal resection of the femur (Picture 5.) Next, depending on the given intra-operative circumstances, the proximal tibial end resection may be performed, prior to the continuation of the femoral resection. After that, carry on with the measurements, and the final shaping of the femoral component s place. Continuation of the femoral resection: If tissues condition permits, then the next step will be the positioning of the femoral sizer. This will help in determining the right size of the femoral component to be used, adjusting its lateral rotation and preparing the suitable holes for the M.I. cutting block. Make sure that the edges of the block fall at equal distances from the medial and lateral sides of the epicondyles. 5

7 tactile screw for angle stabilisation screw for position stabilisation reference plane 6. a 6.b guide rings Next step, sizing femoral component: put the instrument on the cut-off plane as shows the 6.a picture. Important the correspond contact with cut-off surface!! Now we have to set the tactile on the highest point between the condyles. In this position we can read the size (the appropriate size which is nearest to the reference plane). Finally we set the rotation by the guide rings (Picture 6.c). Angles are between 0ᵒ and 7ᵒ (Left/Right) (Picture 6.d) 6.c 6.d After the lateral rotation is set, through drill into the holes with drilling machine for pins x (Picture 7.b) 7.a 7.b Place the cutting block (classic type: /-0007 Minimal invasive type: /-0017) of the suitable size, onto the pins, in the rotation position that was set 6

8 in the previous step (Mostly 3 ). (Picture 8.). Make sure that the first resection does not reach the femur s anterior cortical. Then perform the 4 resections as indicated on the block. 8.a 8.b 8.c Fixing the cutting block with 2 pin in against rotation. Check the cutting plane with template ( ) 8.d Important: For best results, always use a new 1mm thick saw blade for each operation. During resection, the more force applied to the saw, the more likely it is to get diverted by the sclerotic bones. The use of saw guide blocks reduces the extent of bending of the saw blade. In case metal debris is noticed during sawing, this could be an indication of a wrong technique applied. In such cases the plane of the saw blade should be corrected. Finally, the fixation pins and the cutting block are removed. The planes should be checked and if necessary corrections should be made. P/S Femoral component After completing the outline resections, centre the universal femoral box cutting guide ( ) medially/laterally on the surface of the distal femur, with the anterior flange of the guide laying on the cut surface of the anterior femoral bone (Picture 9.). Secure the guide with 2 to 4 fixation pins. Using an oscillating saw, cut the medial and lateral portions of the bone in the notch. Using the box cutting chisel ( ), remove the remaining bone from the femoral notch (Picture 9. and 10.). 7

9 2nd step A) Proximal resection of the tibia The extramedullary technique The next step is to perform the proximal resection of the tibia. This can be done with the help of an extramedullary aiming device. Assemble the extramedullary tibial cutting block guide, to the telescopic handle ( ) - with the tibial cutting block ( ) on its tip. (Picture 11.) The resection plane should be adjusted in such a way that is most suitable to the anatomical position of the tibia. It can be adjusted at 4.5 or 7 or 10 angle by sliding the scale of the graduated scale found on the ankle clamp of the extramedullary tibia cutting guide rod (Picture 12.). The appropriate angle adjustment can be checked with the help of the resection height gauge that is fitted to the tibia cutting block (Picture 13.). 8

10 12 By sliding the bottom scale (3 or 5 or 7 ) on the ankle clamp of the extramedullary tibia cutting guide rod, varus/valgus adjustments can be performed. (Picture 12.) The height of the tibial horizontal resection is controlled by two possible ways of applying the stylus (Picture 14.): 1. By resting the longer and higher stylus arm on the highest lateral part of the tibia- a 9mm resection is ensured. 2. While by turning the stylus, its smaller and lower arm can be rested on the tibial deepest medial point. The latter position results in the least resection and bone removal volume, ensuring a 2mm resection as compared to the tibial medial plane. (14.Picture) If the damage of the tibia is moderate, then a 2 mm resection from the deepest medial point is performed. If the damage is significant, then make a 9 mm resection laterally, in case you want to implant a cruciate retaining component. It should be taken into consideration, that if the height of the lateral resection exceeds 9mm, then the implantation of a posterior stabilized femoral and tibial component is necessary. 9

11 B) Proximal resection of the tibia The intramedullary technique Prepare the knee by placing it at its maximal flexion position, and determine the tibial plateau midpoint. The point of entry to the intramedullary canal is then identified. It should fall posteriorly to the anterior cruciate ligament. Penetrate the intramedullary canal using a drill. Following the drilling, the resection block is placed in its position, and the Premier II tibial resection guide is fitted on it. Insert the T-handled fluted rod through the tibial resection guide, and push into the intramedullary canal (Picture 15.). This fluted rod will help drive out bone marrow. Proper positioning of the resection guide on to the tibial external cortex should be well checked. Any adjustment needed to the angle of resection of posterior slope (Picture 16.) can be easily achieved by moving the lever to the desired angle (Picture 17.). Available angle options are: 0, 3, 5, 7, and 10 10

12 Place the tibial depth gauge into the resection guide from the lateral plateau side, or the side that is less engaged during this operation. Depth gauge is used to indicate the level of resection to be carried out. The minimal cut is 9mm when the gauge scale 0. (for example if the scale +2 then the cut 11mm). (Picture 18.). 18.Picture For setting the resection level turn the knob at the tip of the device (Picture 19.). To lower the resection guide: turn the knob counter clockwise. To raise the resection guide: turn clockwise

13 For medial and lateral alignment, connect the telescoping alignment rod to the tibial resection guide (Picture 21.). The alignment rod divides the talus into 2 parts at the joint line (Picture 20.). For adjusting the medial-lateral alignment, loosen out the knob, which is present over the posterior slope lever a little bit, and then re-adjust the rod position. When the alignment marks present on both the knob and the guide meet together, the guide will be at 0 of medial lateral rotation. 21 Use drills to fix the cutting block on to the tibia, or alternatively use fixation pins. Remove the resection guide and rod, but not the resection block. (Picture 22.). 12

14 Resect the plateau with an appropriate 1 mm saw blade. Extra rows of holes are available for an additional 2mm resection. 3rd step Fixing up the trial femoral component Place the appropriate size of the chosen trial femur on the already prepared distal femoral part. As the tolerance of the sawing slits is very low, you should place the components with the help of the femoral inserter/extractor forceps ( ). (Picture 23.a, 23.b, 23.c). When you holded the femoral component with forceps, tight the screws as shows 23.b picture. 23.a 23.b c

15 The medial-lateral placing of the femur implant can be then freely defined. Apply strokes on the component using a rider until it is half way fitted (Picture 24). 24 The femoral component impactor/extractor: Catch the implant with the femur forceps. Then tighten the screw and connect the Impactor/Extractor ( ) device to the forceps. Moving quickly the handle till removal the femur implant. (Picture 25.a, 25.b) 25.a 25.b When the correct position is reached, drill through the lateral and medial condyle holes in the femur using the condyle drill bit ( ). (Picture 26). 14

16 Before drilling, check again the position of femoral trial component! 15

17 4th step Trial reposition Checking of the flexion and extension joint gap Place a suitable trial tibial plate ( , -0104) on the dissected tibial plateau using the thinest trial tibial insert and move the limb carefully throughout the entire movement-range of the limb. The flexed and extended joint gap checking has to get through the following basic steps: 1. The soft pressure at extension determines the extent of the femoral distal construction. 2. The extent of the posterior femoral condylar construction is determined by the soft pressure at flexion. The extent of the tibial construction is determined by the pressure at the extension as well as at flexion. A proper leg balance has to be achieved by the construction process. If the surgeon wishes to carry out further fine adjustments to the balance he has to do that in a careful manner with constant checking on the flexion and extension gap. 1. Bending contraction default: In this case the full extension shows that the joint's column is narrower when extended as compared to that at flexion, the extension joint gap has to be sawed to the right size, so that it adapts to the bending joint gap, i.e. the distal femoral resection and the femoral component at a more proximal positioning. Repeat here the distal femoral resection and the femur construction steps. 2. Imperfect narrowness at extension and flexion of joint gap. Detachment due to inadequate soft pressure, or generally inadequate bone cutting. Enlargement of the extension and flexion gap could be achieved by further resection of the proximal tibia. This means that the flexion gap is smaller then the extension gap. This is shown by the enlargement of the trial tibial plateau opening, with increased bending. The flexion gap can be increased with posterior inclination or by decreasing the femoral component size, i.e. with the ventral positioning of the femoral condylars. 16

18 5th Step The tibial plateau final preparation. Chose a tibial plateau size, which properly fits on the rims of the dissected proximal tibia (picture 27). Fix the handle of trial tibial tray with the help of the ridge-headed screw. Insert the straightening rod by vertical bending into the clamp. Place the gauge and straighten out the round shaped rod clamp with the other finger (picture 28). Alternatively: Place the trial component, the femur and the tibia, and move the leg several times through out the entire possible movement range. Let the tibial plateau rotate into the right position. Place a reference point with the help of HF instrument or colour mark (e.g. methyl blue) on the trial plateau's front edge. 17

19 Drill the front slope holes with a 3,2 mm drill bit, and fix the plateau with two fixation pins (picture 29). Place the tibial drill guide ( ) on to the trial tibial tray, and use the 7,8mm drill bit to prepare the initial hole for the tibial stem (picture 30). 18

20 Afterwards take the introducing guide ( ), the handle ( ) and the tibial punch ( ) (picture 31). Place these together on to the appropriate holes on the plateau (picture 32). With mild strokes tap in the chisel until it meets the clamp in the guiding hole. Then remove the chisel, the fixating screw and the plateau. 19

21 TMB Tibial Component Total Miniscal Bearing designs are advantageous in reducing linear polyethylene wear, reduced polyethylene thickness, a larger femoral and tibial contact area which result is minimizing polyethylene damage. Bearing options are compatible with a universal femoral components (Primary/ P/S) Underpart of TMB tibial plateau is the same as the primary tibial plataeau. The preparing of tibial component is same as the primary surgical technique. TMB Primary surgery: The primary tray trial allows TMB primary and TMB P/S. After the preparation of hole, remove the tibial punch. Now the tray trial is on the top of bone with pins (picture 33.) Place the trial TMB rotating tibial trial ( x000)(same size as the tray trial) onto the trial tray (Picture 34.). Check the correct insertion Afterward frame the insert trial into the TMB rotating tibial trial and rotate it easily around the axle (pictures 35. and 36., 37.)

22 When you get the competent thickness, check the motion between the insert trial and the femoral component (picture 38.). The knee is placed in deep flexion! Tibial Tray implantation Cemented and uncemented tibia plate versions made of Cobalt Chrome. Open the TMB Tray introducer ( ) at maximum and place it onto the implant tibial tray (picture 39.). Ensure the holders are below the tray, then tight the screw (picture 40.). At cemented design, cement should be applied to the underside of the tibia implant tray Now insert the iplant into the prepared hole (picture 41.). When fully inserted several hammer blows may be delivered to the top of the TMB Tray Impactor ( ) (pictures 42. and 43.)

23 After all this, check again the position and the rotation with trial components. Place the GAP trial onto the implant tray then put the choosen trial insert onto the GAP (picture 44.). Finally remove the GAP and insert The placement of the original implant: and put the insert implant into the implant tray and rotate it easily (pictures 45 and 46.). Fully rotate the knee from full extension to full flexion (picture 47.)

24 TMB P/S surgery: Difference between the two methods TMB primary and TMB P/S is only in the inserts! Femoral component is P/S.! It is no diference between the tibial implant trays! Place the trial insert onto the TMB rotating tibial trial and determine the necessary insert thickness (pictures 48., 49.) Insertation of tibia tray and position check is the same as TMB primary surgery (page 20.). Finally put the insert implant into the axle of tibial tray implant and check the rotation (pictures 50., 51., 52., 53., 54.)

25 6th Step The placement of the original implant Preparations: Use a piece of the earlier cut out bone as a seal for the intramedullary canal. This can help avoiding post operational bleeding at the site of operation. The cut bone surfaces are inspected carefully once more. These can be finally thoroughly cleaned using a jet washing system and can be then dried out. The original implants (the appropriately chosen sizes) should be handed over sterile by the trained medical staff. Tibial component: The monoblock tibial plateau size is selected according to the trial implant size. The insertion of the tibial tray is performed with the tibial tray inserter ( ) (picture 55). 55. The cement is spread over the lower part of the tibial component and the dissected tibial bone surface, after placing the implant in the proper position (picture 56., 57.)

26 Remove the inserting instrument and then tightly impact the modular component with the help of the tibial impactor. The cement surplus is removed with the help of a scrapper. The components can be implanted simultaneously with the help of the cement mixture, or one after the other. The tibial components are usually implanted first, except in the case of P/S implants, where the femoral component should be implanted first. During the cement solidification phase, the tibial original implant or the tibial trial insert may be placed once more as preferred on the original tibial tray implant. The tibial insert gliding surface condition The original tibial insert gliding surface is fitted on to the metal plate flately from the frontal side, minding the roundly pitted offset on the surface of the tibial metal plate while fitting (picture 58.). 58. The tibial insert is finely tapped with the gradual round tibial insert impactor ( ) to firmly slide into position in the depth of the metal plateau. This kind of slide-in fixation of the tibial inserts from the frontal side helps to avoide any postoperative impingement. A 3,5 mm hex screwdriver ( ) is used to tighten the fixation screw through the tibial insert into the metal plateau (Picture 59.)

27 Femoral component: The low viscosity cement needs to be also spread on the entire internal surface of the femoral component and on all cut out surfaces of the femur. The femoral implant is then held in the inlay instrument and fitted on the cut out femur. (Picture 60.). 60. The femoral component has to be impacted with the femoral inserter/extractor forceps ( ). When the right position is achieved, remove the instrument with loosen the screw (picture 61.). 61. With the femoral impactor ( ) the corrections can be performed. After the positioning of the implant the extra cement is carefully removed, especially at the area of the posterior condyles. The cement solidification process occurs, while the patient's leg is stretched out and a hyperextension is achieved. 26

28 7th Step Preparation of the patella Remove all outer osteophyte, and try to approximately restore the anatomical form. The resection area and the position that arises from that are essential for the patella's proper functionality. The dorsal joint surface is demonstrable by tipping it over. Measure the bone strength with help of patella calliper. A bone layer that corresponds to the used patella implant has to be removed. Fix the patella with the help of the patella resection clamp or pliers and carry out a straight resection. This has to be exactly started under the deep abraded edge of the subcondrial bone, and to be continued in a straight line at the laterally abraded edge till the uppermost point (picture 62.). 62. This way, a plain coating area is created later for the implant. To this surface a patella drill guide is centrally placed with handle, and a fixating screw is introduced in. (picture 63.). With the help of the patella drill (strokes) the central fixing hole is drilled (picture 64.) Out of the available implant sizes we choose the one, which offers the most covering area without protruding out (picture 65.)

29 Measure once again the total strength of the reconstructed bone and joints. It should not be more than the initial strength, in order to avoid the retro-patellal pressure increase. Patella trial position: Check the patella's functioning along with the other prosthesis components. It should be able to freely glide within the trochlea, without slipping to the sides. In case there is a tendency to slip to one side, a gradual retinaculum cleavage has to be performed to restore the gliding path. For the cementing of the original components use the patella clamp, so that it aids the solidification of the cement and keeps it fixed on. The extra cement should be eliminated while solidifying. 28

BIOTECH FUTURE KNEE Minimal Invasive (and classic) Surgical Technique

BIOTECH FUTURE KNEE Minimal Invasive (and classic) Surgical Technique BIOTECH FUTURE KNEE Minimal Invasive (and classic) Surgical Technique Page No. Product description 2. Pre-operational planning 3. 1st step Femoral resection 4. P/S Femoral component 7. 2nd step Proximal

More information

LAMINA SPREADER SURGICAL TECHNIQUE

LAMINA SPREADER SURGICAL TECHNIQUE LAMINA SPREADER SURGICAL TECHNIQUE Balanced and appropriate external rotation of the femoral component is important for tibio-femoral stability in flexion and patello-femoral tracking/function. Depending

More information

Posterior Referencing. Surgical Technique

Posterior Referencing. Surgical Technique Posterior Referencing Surgical Technique Posterior Referencing Surgical Technique INTRO Introduction Instrumentation Successful total knee arthroplasty depends in part on re-establishment of normal lower

More information

Signature* Personalized Patient Care

Signature* Personalized Patient Care Signature* Personalized Patient Care Surgical Technique Addendum Vanguard Complete Knee System MRI-based Positioning Guides CT-based Positioning Guides MRI-based Cut-through Guides One Surgeon. One Patient.

More information

Hip Knee Spine Navigation

Hip Knee Spine Navigation GLOBAL MEDACTA KNEE Different needs......your GLOBAL SOLUTION Surgical Technique Hip Knee Spine Navigation GMK REVISION Surgical Technique Hip Knee Spine Navigation INTRODUCTION The GMK Revision can be

More information

Zimmer NexGen LCCK. Surgical Technique for use with LCCK 4-in-1 Instrumentation. Contact your Zimmer representative or visit us at www.zimmer.

Zimmer NexGen LCCK. Surgical Technique for use with LCCK 4-in-1 Instrumentation. Contact your Zimmer representative or visit us at www.zimmer. Zimmer NexGen LCCK Surgical Technique for use with LCCK 4-in-1 Instrumentation DISCLAIMER: This documentation is intended exclusively for physicians and is not intended for laypersons. Information on the

More information

DRAFT. Triathlon TS Knee System. Surgical Protocol. Version 1

DRAFT. Triathlon TS Knee System. Surgical Protocol. Version 1 DRAFT Version 1 Triathlon TS Knee System Table of Contents Acknowledgments..........................................................2 Tibial Canal Preparation....................................................4

More information

PARTIAL KNEE SYSTEM. Frequently Asked Questions. MIRROR PARTIAL KNEE SYSTEM Frequently Asked Questions cayennemedical.

PARTIAL KNEE SYSTEM. Frequently Asked Questions. MIRROR PARTIAL KNEE SYSTEM Frequently Asked Questions cayennemedical. PARTIAL KNEE SYSTEM Frequently Asked Questions MIRROR PARTIAL KNEE SYSTEM Frequently Asked Questions DESIGN RATIONALE What is Ligament Guided Surgery (LGS)? Mirror is a partial knee replacement system

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. Monogram Total Knee Instruments Modular Rotating Hinge Knee System Keel Baseplate Using Monogram IM Revision Instruments

More information

Trabecular Metal Femoral Component. Surgical Technique

Trabecular Metal Femoral Component. Surgical Technique Trabecular Metal Femoral Component Surgical Technique INTRO Trabecular Metal Femoral Component Surgical Technique Introduction The Persona Trabecular Metal Femoral Component is designed to unite stable

More information

Zimmer FuZion Instruments. Surgical Technique (Beta Version)

Zimmer FuZion Instruments. Surgical Technique (Beta Version) Zimmer FuZion Surgical Technique (Beta Version) INTRO Surgical Technique Introduction Surgical goals during total knee arthroplasty (TKA) include establishment of normal leg alignment, secure implant fixation,

More information

NEXGEN COMPLETE KNEE SOLUTION. Epicondylar Instrumentation Surgical Technique For Legacy Posterior Stabilized Knees

NEXGEN COMPLETE KNEE SOLUTION. Epicondylar Instrumentation Surgical Technique For Legacy Posterior Stabilized Knees NEXGEN COMPLETE KNEE SOLUTION Epicondylar Instrumentation Surgical Technique For Legacy Posterior Stabilized Knees INTRODUCTION Successful total knee arthroplasty is directly dependent on reestablishment

More information

Scorpio NRG CR & PS. Single Radius Primary Knee System Surgical Protocol. Anterior Referencing

Scorpio NRG CR & PS. Single Radius Primary Knee System Surgical Protocol. Anterior Referencing Scorpio NRG CR & PS Single Radius Primary Knee System Surgical Protocol AR Anterior Referencing This document is intended to be used by healthcare professionals only. Scorpio NRG AR Single Radius Primary

More information

Triathlon Knee System

Triathlon Knee System Triathlon Knee System The Right Procedure with the Right Implant for Your Patient Table of Contents Acknowledgments..........................................................2 Introduction...............................................................3

More information

Zimmer MOST Options System

Zimmer MOST Options System Zimmer MOST Options System Surgical Technique Modular Options for Severe Bone Loss and Trauma Zimmer MOST Options System Surgical Technique 1 MOST Options System Surgical Technique Table of Contents Introduction

More information

INTUITION INSTRUMENTS SURGICAL TECHNIQUE

INTUITION INSTRUMENTS SURGICAL TECHNIQUE INTUITION INSTRUMENTS SURGICAL TECHNIQUE Introduction This surgical technique provides guidelines for the implantation of the ATTUNE Knee System family of knee implants with the INTUITION Instrumentation.

More information

Zimmer NexGen CR-Flex and LPS-Flex Knees. Surgical Technique with Posterior Referencing Instrumentation

Zimmer NexGen CR-Flex and LPS-Flex Knees. Surgical Technique with Posterior Referencing Instrumentation Zimmer NexGen CR-Flex and LPS-Flex Knees Surgical Technique with Posterior Referencing Instrumentation INTRO Zimmer NexGen CR-Flex and LPS-Flex Knees Surgical Technique with Posterior Referencing Instrumentation

More information

AVANTAGE. Acetabulum System. Surgical Technique

AVANTAGE. Acetabulum System. Surgical Technique AVANTAGE Acetabulum System Surgical Technique 1 Table of Contents Page Pre-operative planning 3 Surgical exposure 3 Acetabular reaming 4 Acetabular gauging 4 Acetabular implant insertion 5 Trial inserts

More information

NexGen Complete Knee Solution Intramedullary Instrumentation Surgical Technique

NexGen Complete Knee Solution Intramedullary Instrumentation Surgical Technique NexGen Complete Knee Solution Intramedullary Instrumentation Surgical Technique For the NexGen Cruciate Retaining & Legacy Posterior Stabilized Knee INTRODUCTION Successful total knee arthroplasty is directly

More information

Unity Knee. Surgical technique Measured resection approach. Think isometry Feel balance

Unity Knee. Surgical technique Measured resection approach. Think isometry Feel balance Unity Knee Surgical technique Measured resection approach Think isometry Feel balance Contents Operative summary 4 Pre-operative assessment 5 Instrumentation rationale 5 Operative technique 6 1. Approach

More information

Zimmer Periarticular Proximal Tibial Locking Plate

Zimmer Periarticular Proximal Tibial Locking Plate Zimmer Periarticular Proximal Tibial Locking Plate Surgical Technique The Science of the Landscape Zimmer Periarticular Proximal Tibial Locking Plate 1 Table of Contents Introduction 2 Locking Screw Technology

More information

Surgical Technique and Reference Guide

Surgical Technique and Reference Guide Surgical Technique and Reference Guide Revision total knee arthroplasty presents many complex problems for the orthopaedic surgeon. Hard tissue defects can be caused by significant bone loss, massive osteolysis,

More information

Vanguard Knee. Surgical Technique

Vanguard Knee. Surgical Technique Vanguard Knee TM Surgical Technique Vanguard Unicompartmental Patellofemoral Arthroplasty Femoral Component Bone and ligament sparing for easier conversion t TKA Minimal bone resection Cobalt chromium

More information

PRIMARY KNEE SYSTEM SIGMA SURGICAL TECHNIQUE. Balanced Surgical Technique INSTRUMENTS

PRIMARY KNEE SYSTEM SIGMA SURGICAL TECHNIQUE. Balanced Surgical Technique INSTRUMENTS SIGMA PRIMARY KNEE SYSTEM Balanced Surgical Technique SURGICAL TECHNIQUE INSTRUMENTS Contemporary total knee arthoplasty demands high performance instrumentation that provides enhanced efficiency, precision

More information

S-ROM NOILES. This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

S-ROM NOILES. This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE S-ROM NOILES This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE CONTENTS INTRODUCTION S-ROM NOILES Rotating Hinge Knee System 2 SURGICAL TECHNIQUE Initial Preparation of

More information

Tension band fixation Christof A. Müller

Tension band fixation Christof A. Müller Tension band fixation Christof A. Müller How to use this handout? The left column is the information as given during the lecture. The column at the right gives you space to make personal notes. Learning

More information

IMAGE TO COME. Unicompartmental High Flex Knee. Zimmer MIS. Lateral Surgical Techniques

IMAGE TO COME. Unicompartmental High Flex Knee. Zimmer MIS. Lateral Surgical Techniques Zimmer MIS Unicompartmental High Flex Knee Lateral Surgical Techniques IMAGE TO COME Spacer Block, Intramedullary, and Extramedullary Options Minimally Invasive Surgical Techniques for the Lateral Compartment

More information

Phoenix Retrograde Femoral Nail System Featuring CoreLock Technology. Surgical Technique

Phoenix Retrograde Femoral Nail System Featuring CoreLock Technology. Surgical Technique Phoenix Retrograde Femoral Nail System Featuring CoreLock Technology Surgical Technique Contents Introduction... Page 1 Indications... Page 2 Design Features... Page 3 Surgical Technique... Page 6 Product

More information

Zimmer Natural Nail System. Tibial Nail Surgical Technique

Zimmer Natural Nail System. Tibial Nail Surgical Technique Zimmer Natural Nail System Tibial Nail Surgical Technique Zimmer Natural Nail System Tibial Nail Surgical Technique 1 Zimmer Natural Nail System Tibial Nail Surgical Technique Table of Contents Product

More information

S-ROM NOILES ROTATING HINGE KNEE SYSTEM SURGICAL TECHNIQUE

S-ROM NOILES ROTATING HINGE KNEE SYSTEM SURGICAL TECHNIQUE S-ROM NOILES ROTATING HINGE KNEE SYSTEM SURGICAL TECHNIQUE TABLE OF CONTENTS SURGICAL TECHNIQUE Key Surgical Steps Summary 4 S-ROM NOILES Rotating Hinge Knee System 6 The S-ROM Hinge System Overview 7

More information

Anterior Lumbar Interbody Fusion (ALIF). Instrument set supports placement of ALIF spacers using anterior or anterolateral approach.

Anterior Lumbar Interbody Fusion (ALIF). Instrument set supports placement of ALIF spacers using anterior or anterolateral approach. Anterior Lumbar Interbody Fusion (ALIF). Instrument set supports placement of ALIF spacers using anterior or anterolateral approach. Technique Guide Instruments and implants approved by the AO Foundation

More information

Zimmer MIS Multi-Reference 4-in-1 Femoral Instrumentation

Zimmer MIS Multi-Reference 4-in-1 Femoral Instrumentation Zimmer MIS Multi-Reference 4-in-1 Femoral Instrumentation Surgical Technique For NexGen Cruciate Retaining & NexGen Legacy Posterior Stabilized Knees Zimmer MIS Multi-Reference 4-in-1 Femoral Instrumentation

More information

7.0 mm Cannulated Screw Technique Guide

7.0 mm Cannulated Screw Technique Guide 7.0 mm Cannulated Screw Technique Guide An Integral Part of the SYNTHES Cannulated Screw System Original Instruments and Implants of the Association for the Study of Internal Fixation AO ASIF 7.0 mm Cannulated

More information

VERSYS HERITAGE CDH HIP PROSTHESIS. Surgical Technique for CDH Hip Arthroplasty

VERSYS HERITAGE CDH HIP PROSTHESIS. Surgical Technique for CDH Hip Arthroplasty VERSYS HERITAGE CDH HIP PROSTHESIS Surgical Technique for CDH Hip Arthroplasty SURGICAL TECHNIQUE FOR VERSYS HERITAGE CDH HIP PROSTHESIS CONTENTS ANATOMICAL CONSIDERATIONS....... 2 PREOPERATIVE PLANNING............

More information

MIS. Surgical Technique M INIMALLY I NVASIVE C UTTING B LOCKS DESIGNED FOR YOU BY YOU. Hip Knee Spine Navigation

MIS. Surgical Technique M INIMALLY I NVASIVE C UTTING B LOCKS DESIGNED FOR YOU BY YOU. Hip Knee Spine Navigation MIS M INIMALLY I NVASIVE C UTTING B LOCKS DESIGNED FOR YOU BY YOU Surgical Technique Hip Knee Spine Navigation MyKnee MIS Surgical Technique Hip Knee Spine Navigation INTRODUCTION This brochure describes

More information

Surgical technique. End Cap for TEN. For axial stabilization and simultaneous protection of soft tissue.

Surgical technique. End Cap for TEN. For axial stabilization and simultaneous protection of soft tissue. Surgical technique End Cap for TEN. For axial stabilization and simultaneous protection of soft tissue. Table of contents Indications and contraindications 3 Implants 4 Instruments 4 Preoperative planning

More information

Patient Fit is Now an Option. Zimmer Natural Nail System

Patient Fit is Now an Option. Zimmer Natural Nail System Patient Fit is Now an Option Zimmer Natural Nail System Taking IM nailing to the next level The Zimmer Natural Nail System is a next generation system of intramedullary nails that is designed to help restore

More information

V-TEK IVP System 2.7 System 4.0

V-TEK IVP System 2.7 System 4.0 V-TEK IVP System Ankle 2.7 Fix System 4.0 Surgical Technique Surgical Technique Titanium osteosynthesis system for tibio-talar and tibio-talo-calcaneal fusion SECTION 1 Ankle Fix System 4.0 Titanium osteosynthesis

More information

Minimally Invasive Knee Surgery (MIS) with Navigation Operative Technique

Minimally Invasive Knee Surgery (MIS) with Navigation Operative Technique Joint Replacements Minimally Invasive Knee Surgery (MIS) with Navigation Operative Technique Mini Mid Vastus Approach Contents - Introduction 01 - Patient Set Up 02 - Surgical Exposure 03 - Navigation

More information

iassist Knee Surgical Technique 2-Pod Version Instruments

iassist Knee Surgical Technique 2-Pod Version Instruments iassist Knee Surgical Technique 2-Pod Version Instruments Table of Contents Page 1. Overview 4 1.1 Indication for Use 4 1.2 Training 4 1.3 Implant Indications 4 1.4 Contraindications 4 1.5 Complications

More information

NCB Distal Femur System. Surgical Technique

NCB Distal Femur System. Surgical Technique NCB Distal Femur System Surgical Technique NCB Distal Femur System Surgical Technique 3 Surgical Technique NCB Distal Femur System Table of Contents Introduction 4 Indications 8 Preoperative Planning

More information

CoNGRUENT RADIAL HEAD PLATE

CoNGRUENT RADIAL HEAD PLATE Mayo Clinic CoNGRUENT RADIAL HEAD PLATE Since 1988 Acumed has been designing solutions to the demanding situations facing orthopedic surgeons, hospitals and their patients. Our strategy has been to know

More information

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee.

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee. Technique Guide TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee. Table of Contents Introduction TOMOFIX Osteotomy System 2 AO Principles 4 Indications

More information

High-Flex Solutions for the MIS Era. Zimmer Unicompartmental High Flex Knee System

High-Flex Solutions for the MIS Era. Zimmer Unicompartmental High Flex Knee System High-Flex Solutions for the MIS Era Zimmer Unicompartmental High Flex Knee System Zimmer Unicompartmental High Flex Knee Built On Success In today s health care environment, meeting patient demands means

More information

Technique Guide. VersiTomic. Michael A. Rauh, MD. Anterior Cruciate Ligament Reconstruction

Technique Guide. VersiTomic. Michael A. Rauh, MD. Anterior Cruciate Ligament Reconstruction Technique Guide VersiTomic Anterior Cruciate Ligament Reconstruction Michael A. Rauh, MD The opinions expressed are those of Dr. Rauh and are not necessarily those of Stryker VersiTomic Anterior Cruciate

More information

PERIPROSTHETIC IMPLANTS

PERIPROSTHETIC IMPLANTS PERIPROSTHETIC IMPLANTS PRODUCT OVERVIEW CLINICAL SOLUTIONS Periprosthetic fractures present unique challenges, such as how to gain fixation when the medullary canal is occupied. Special techniques and

More information

SURGICAL TECHNIQUE FIGURE 1

SURGICAL TECHNIQUE FIGURE 1 SURGICAL TECHNIQUE UNI-Elbow Radio Capitellum System Surgical Technique 1 SURGICAL TECHNIQUE 2 Incision The patient is placed under a general or a regional anesthesia. The extremity is prepped and draped

More information

PoLARUS 2 Locking Humeral Rod Surgical Technique

PoLARUS 2 Locking Humeral Rod Surgical Technique PoLARUS 2 Locking Humeral Rod Surgical Technique The Polarus 2 Locking Humeral Rod is designed to provide fixation for 2-, 3-, and some 4-part fractures of the proximal humerus through a minimally invasive

More information

Technique Guide. Screw Removal Set. Instruments for removing Synthes screws.

Technique Guide. Screw Removal Set. Instruments for removing Synthes screws. Technique Guide Screw Removal Set. Instruments for removing Synthes screws. Table of Contents Introduction Screw Removal Set 2 Surgical Technique Preoperative Planning and Preparation 6 Removal of Intact

More information

Structure and Function of the Bones and Noncontractile Elements of the Knee

Structure and Function of the Bones and Noncontractile Elements of the Knee Structure and Function of the Bones and Noncontractile Elements of the Knee BONES OF THE KNEE JOINT Shaft and Distal Femur Linea aspera gives rise to quadriceps. Flatten into supracondylar ridges End in

More information

ACUFEX PINPOINT Anatomic ACL Guide System

ACUFEX PINPOINT Anatomic ACL Guide System *smith&nephew KNEE TECHNIQUE GUIDE ACUFEX PINPOINT Anatomic ACL Guide System Darren L. Johnson, MD Luigi A. Pederzini, MD KNEE HIP SHOULDER EXTREMITIES ACUFEX PINPOINT Anatomic ACL Guide System As described

More information

Staple Fixation System. Implants Instruments Applications

Staple Fixation System. Implants Instruments Applications Staple Fixation System Implants Instruments Applications Contents Introduction 3 Instrument Use - Universal Inserter 4 Instrument Use - Universal Rod 5 Ordering Information - Implants 6 Ordering Information

More information

Know-how brings motion back to life. SM. Solution UC-PLUS. Unicompartmental Knee Prosthesis. Surgical Technique

Know-how brings motion back to life. SM. Solution UC-PLUS. Unicompartmental Knee Prosthesis. Surgical Technique Know-how brings motion back to life. SM TM UC-PLUS Solution Unicompartmental Knee Prosthesis Surgical Technique Table of Contents 1 Introduction 3 2 Characteristics and Advantages of the System 4 3 Indications

More information

Technique Guide. Orthopaedic Cable System. Cerclage solutions for general surgery.

Technique Guide. Orthopaedic Cable System. Cerclage solutions for general surgery. Technique Guide Orthopaedic Cable System. Cerclage solutions for general surgery. Table of Contents Introduction The Orthopaedic Cable System 2 Indications 4 Contraindications 4 Surgical Technique Cerclage

More information

Foot & Ankle. Smart Toe II. Intramedullary Implant. Foot & Ankle

Foot & Ankle. Smart Toe II. Intramedullary Implant. Foot & Ankle Foot & Ankle Smart Toe II Intramedullary Implant Operative Technique Foot & Ankle Smart Toe This publication sets forth detailed recommended procedures for using Stryker Osteosynthesis devices and instruments.

More information

Restoration Modular. Revision Hip System Surgical Protocol

Restoration Modular. Revision Hip System Surgical Protocol Orthopaedics Restoration Modular Revision Hip System Restoration Modular Calcar Body/Conical Distal Stem Femoral Components Using the Restoration Modular Instrument System Revision Hip System Orthopaedics

More information

Zimmer Small Fragment Universal Locking System. Surgical Technique

Zimmer Small Fragment Universal Locking System. Surgical Technique Zimmer Small Fragment Universal Locking System Surgical Technique Zimmer Small Fragment Universal Locking System 1 Zimmer Small Fragment Universal Locking System Surgical Technique Table of Contents Introduction

More information

AM PORTAL SURGICAL TECHNIQUE GUIDE

AM PORTAL SURGICAL TECHNIQUE GUIDE The AperFix System AM PORTAL SURGICAL TECHNIQUE GUIDE The Cayenne Medical AperFix System introduces an innovative, anatomic approach to soft tissue multi-ligament reconstruction of the knee for enhanced

More information

Zimmer PSI Knee System For Use with Persona The Personalized Knee System Surgical Technique

Zimmer PSI Knee System For Use with Persona The Personalized Knee System Surgical Technique Zimmer PSI Knee System For Use with Persona The Personalized Knee System Surgical Technique TOC Zimmer PSI Knee Surgical Technique Zimmer PSI Knee Surgical Technique Table of Contents Intra-Operative Guide

More information

BONE PRESERVATION STEM

BONE PRESERVATION STEM TRI-LOCK BONE PRESERVATION STEM Featuring GRIPTION Technology SURGICAL TECHNIQUE IMPLANT GEOMETRY Extending the TRI-LOCK Stem heritage The original TRI-LOCK Stem was introduced in 1981. This implant was

More information

Sirus Intramedullary Femur Nail System

Sirus Intramedullary Femur Nail System Sirus Intramedullary Femur Nail System Surgical Technique Anatomical Nail for Reamed and Nonreamed Technique, Choice for Locking and Lateral Entry Point Disclaimer This document is intended exclusively

More information

Aesculap Veterinary Orthopaedics. Targon VET Interlocking Nail

Aesculap Veterinary Orthopaedics. Targon VET Interlocking Nail Aesculap Veterinary Orthopaedics Targon VET Interlocking Nail Flexibility Stability Dynamics The Targon VET is based on new, technology which reverses the known principles of interlocking nails. The rotation-stabilising

More information

Zimmer Segmental System. Distal Femoral Surgical Technique

Zimmer Segmental System. Distal Femoral Surgical Technique Zimmer Segmental System Distal Femoral Surgical Technique INTRO Zimmer Segmental System Distal Femoral Surgical Technique Introduction The Zimmer Segmental System is designed to address significant bone

More information

Original Bledsoe Brace Application Instructions CP Rev M 02/07

Original Bledsoe Brace Application Instructions CP Rev M 02/07 Original Bledsoe Brace Application Instructions CP020101 Rev M 02/07 TOTAL LEG SUPPORT SYSTEM WITH INFINITE R.O.M. & SHORT APPLICATION TIME This device is offered for sale by or on the order of a physician

More information

ROTATING PLATFORM TOTAL KNEE REPLACEMENT BRIGHAM AND WOMEN S HOSPITAL, BOSTON, MA Broadcast May 20, 2004

ROTATING PLATFORM TOTAL KNEE REPLACEMENT BRIGHAM AND WOMEN S HOSPITAL, BOSTON, MA Broadcast May 20, 2004 NARRATOR ROTATING PLATFORM TOTAL KNEE REPLACEMENT BRIGHAM AND WOMEN S HOSPITAL, BOSTON, MA Broadcast May 20, 2004 Thank you for joining us from Brigham and Women s Hospital in Boston, MA, to view the minimally

More information

Unicondylar Sled Prosthesis. with MITUS Instrument Set. Implants & Instruments

Unicondylar Sled Prosthesis. with MITUS Instrument Set. Implants & Instruments Unicondylar Sled Prosthesis with MITUS Instrument Set Implants & Instruments Presented by: Waldemar Link GmbH & Co. KG Barkhausenweg 10 22339 Hamburg, Germany P.O. Box 63 05 52 22315 Hamburg, Germany Tel.:

More information

Quick Reference Chart. Titanium Trochanteric Fixation Nail System. For intramedullary fixation of proximal femur fractures.

Quick Reference Chart. Titanium Trochanteric Fixation Nail System. For intramedullary fixation of proximal femur fractures. Quick Reference Chart Titanium Trochanteric Fixation Nail System. For intramedullary fixation of proximal femur fractures. Titanium Trochanteric Fixation Nail System Implants. Titanium Cannulated Trochanteric

More information

Anatomic ACL Reconstruction Using the CLANCY T Flexible Drill Guide System

Anatomic ACL Reconstruction Using the CLANCY T Flexible Drill Guide System *smith&nephew KNEE TECHNIQUE GUIDE Anatomic ACL Reconstruction Using the CLANCY T Flexible Drill Guide System Asheesh Bedi, MD, Answorth Allen, MD, and David W. Altchek, MD KNEE HIP SHOULDER EXTREMITIES

More information

Zimmer PSI Knee System For Use with the NexGen Complete Knee System Surgical Technique

Zimmer PSI Knee System For Use with the NexGen Complete Knee System Surgical Technique Zimmer PSI Knee System For Use with the NexGen Complete Knee System Surgical Technique TOC Zimmer PSI Knee Surgical Technique Zimmer PSI Knee Surgical Technique Table of Contents Intra-Operative Guide

More information

Osteoarthritis damage to the knee

Osteoarthritis damage to the knee Osteotomy and Unicompartmental Knee Arthroplasty Total joint replacement (arthroplasty) is a common and very successful surgery for people with degenerative arthritis (osteoarthritis) of the knee. Two

More information

Speed-Mat Rectangle Cutter

Speed-Mat Rectangle Cutter Speed-Mat Rectangle Cutter 1 Honeycomb baseboard. 2 Left hold down. 14 3 Bottom hold down. 4 4 Left / right rule. 8 5 8 5 Left / right rule pointer. 1 6 Top / bottom rule. 7 Top / bottom rule pointer.

More information

Zimmer M/L Taper Hip Prosthesis. Surgical Technique

Zimmer M/L Taper Hip Prosthesis. Surgical Technique Zimmer M/L Taper Hip Prosthesis Surgical Technique Zimmer M/L Taper Hip Prosthesis 1 Zimmer M/L Taper Hip Prosthesis Surgical Technique Table of Contents Preoperative Planning 2 Determination of Leg Length

More information

BIOMET OSS. Orthopaedic Salvage System

BIOMET OSS. Orthopaedic Salvage System BIOMET OSS Orthopaedic Salvage System One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and art of medical care is

More information

Stryker Hand Plating System

Stryker Hand Plating System Stryker Hand Plating System Operative Technique VariAx Hand Locking Plate Module Profyle Hand Standard Plating Module Stryker Hand Plating System The Stryker Hand Plating System was developed with one

More information

Table of Contents. System Overview. Surgical Technique. 1.2mm Module mm Module mm Module mm Module... 8

Table of Contents. System Overview. Surgical Technique. 1.2mm Module mm Module mm Module mm Module... 8 hand plating system Table of Contents hand plating system System Overview 1.2mm Module... 2 1.6mm Module... 4 2.0mm Module... 6 2.4mm Module... 8 Cannulated Module... 10 Hand Fusion Module... 12 Instruments...

More information

Anteromedial ACL Reconstruction Surgical Technique

Anteromedial ACL Reconstruction Surgical Technique Anteromedial ACL Reconstruction Surgical Technique for Bone-Tendon-Bone Grafts Advancing ACL Reconstruction with implant absorption and ossification of the implant site 2,3,5 The MILAGRO BR Interference

More information

Expert R/AFN. Retrograde/Antegrade Femoral Nail.

Expert R/AFN. Retrograde/Antegrade Femoral Nail. Expert R/AFN. Retrograde/Antegrade Femoral Nail. Surgical Technique EXPERT Nailing System This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation

More information

Screws Form and function Chris Colton, Judy Orson

Screws Form and function Chris Colton, Judy Orson Screws Form and function Chris Colton, Judy Orson How to use this handout? The left column is the information as given during the lecture. The column at the right gives you space to make personal notes.

More information

Technique Guide. Reamer/Irrigator/Aspirator (RIA). For intramedullary reaming and bone harvesting.

Technique Guide. Reamer/Irrigator/Aspirator (RIA). For intramedullary reaming and bone harvesting. Technique Guide Reamer/Irrigator/Aspirator (RIA). For intramedullary reaming and bone harvesting. Table of Contents Introduction Reamer/ Irrigator/ Aspirator (RIA) 2 Indications 4 Surgical Technique Preparation

More information

Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery.

Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery. Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by

More information

AxSOS Locking Plate System

AxSOS Locking Plate System AxSOS Locking Plate System Operative Technique Proximal Humerus 5 Introduction The AxSOS Locking Plate System is designed to treat periarticular or intra-articular fractures of the Proximal Humerus, Distal

More information

Anteromedial ACL Reconstruction Surgical Technique for Soft Tissue Grafts

Anteromedial ACL Reconstruction Surgical Technique for Soft Tissue Grafts Anteromedial ACL Reconstruction Surgical Technique for Soft Tissue Grafts Advancing ACL Reconstruction with implant absorption and ossification of the Implant site 2,3,5 The MILAGRO BR Interference Screw

More information

The information contained in this document is intended for healthcare professionals only.

The information contained in this document is intended for healthcare professionals only. The information contained in this document is intended for healthcare professionals only. Dall-Miles Cabling System Dall-Miles Recon and Trauma Cable System Trochanteric Reattachment Using the Trochanteric

More information

6.5 / 8.0 mm Cannulated Screw System. Surgical Technique

6.5 / 8.0 mm Cannulated Screw System. Surgical Technique 6.5 / 8.0 mm Cannulated Screw System Surgical Technique 1 6.5/8.0 mm Cannulated Screw System Surgical Technique 6.5/8.0 mm Cannulated Screw System The 6.5/8.0 mm Cannulated Screw System is part of a series

More information

Technique Guide. 3.5 mm LCP Clavicle Hook Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Clavicle Hook Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide 3.5 mm LCP Clavicle Hook Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 3.5 mm LCP Clavicle Hook Plates 2 AO Principles 4 Indications

More information

.org. Tibia (Shinbone) Shaft Fractures. Anatomy. Types of Tibial Shaft Fractures

.org. Tibia (Shinbone) Shaft Fractures. Anatomy. Types of Tibial Shaft Fractures Tibia (Shinbone) Shaft Fractures Page ( 1 ) The tibia, or shinbone, is the most common fractured long bone in your body. The long bones include the femur, humerus, tibia, and fibula. A tibial shaft fracture

More information

Femur. Omega3 System. 95 Supracondylar Plate. Operative Technique. Distal Femur Axially Stable Locking Option

Femur. Omega3 System. 95 Supracondylar Plate. Operative Technique. Distal Femur Axially Stable Locking Option Femur Omega3 System 95 Supracondylar Plate Operative Technique Femur Fractures Distal Femur Axially Stable Locking Option Contents Introduction, Potential Features & Benefits 3 Relative Indications & Contraindications

More information

Structure and Function of the Bones and Noncontractile Elements of the Knee

Structure and Function of the Bones and Noncontractile Elements of the Knee Structure and Function of the Bones and Noncontractile Elements of the Knee Any Questions Going over today: Bones around the knee Menisci LCL/MCL ligaments Biomechanics of knee extension/flexion-is it

More information

Implants and Instruments

Implants and Instruments GEMINI SL Total Knee Replacement cementable / cementless with TiCaP Mobile Bearing Fixed Bearing Fixed Bearing PS MIT-K Navigation Standard Instrument Set Implants and Instruments Presented by: WALDEMAR

More information

Vanguard MPartial Knee. Lateral and Medial Surgical Technique

Vanguard MPartial Knee. Lateral and Medial Surgical Technique Vanguard MPartial Knee Lateral and Medial Surgical Technique One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and

More information

Zimmer Natural Nail System

Zimmer Natural Nail System Zimmer Natural Nail System Cephalomedullary Nail Surgical Technique A solution for everybody Disclaimer This document is intended exclusively for physicians and is not intended for laypersons. Information

More information

Assembly and Surgical Technique. Mini External Fixator

Assembly and Surgical Technique. Mini External Fixator Assembly and Surgical Technique Mini External Fixator Mini External Fixator Table of contents Components and Configurations 2 Indications 3 Single rod assembly 4 Double rod assembly 6 Removal 8 Image

More information

Technique Guide. DHS/DCS Dynamic Hip and Condylar Screw System. Designed to provide stable internal fixation.

Technique Guide. DHS/DCS Dynamic Hip and Condylar Screw System. Designed to provide stable internal fixation. Technique Guide DHS/DCS Dynamic Hip and Condylar Screw System. Designed to provide stable internal fixation. Table of Contents Introduction Dynamic Hip Screw (DHS) 2 Dynamic Condylar Screw (DCS) 3 Indications

More information

Surgical Technique. Antegrade Femoral Nail (AFN)

Surgical Technique. Antegrade Femoral Nail (AFN) Surgical Technique Antegrade Femoral Nail (AFN) Contents Indications/contraindications 4 AFN implants 5 Preparation 7 Surgical technique for the AFN 11 Proximal locking 20 Distal locking 32 Implant removal

More information

Mini External Fixator.

Mini External Fixator. Mini External Fixator. Assembly and Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Image intensifier control This

More information

TransFx External Fixation System Large and Intermediate Surgical Technique

TransFx External Fixation System Large and Intermediate Surgical Technique TransFx External Fixation System Large and Intermediate Surgical Technique TransFx External Fixation System Large and Intermediate Surgical Technique 1 Surgical Technique For TransFx External Fixation

More information

Synergy. Low Back Surgical Technique. Spinal Systems. Developed in conjunction with J. Abbott Byrd III, M.D. Biologic & Structural Innovation

Synergy. Low Back Surgical Technique. Spinal Systems. Developed in conjunction with J. Abbott Byrd III, M.D. Biologic & Structural Innovation Spinal Systems Surgical Technique Synergy IQ Low Back Surgical Technique Developed in conjunction with J. Abbott Byrd III, M.D. Universal system Modular packaging maximizes screw options Low-profile design

More information

SMALL BONE ORTHOPEDICS. Rethinking Possibilities, Reshaping Lives

SMALL BONE ORTHOPEDICS. Rethinking Possibilities, Reshaping Lives SMALL BONE ORTHOPEDICS Rethinking Possibilities, Reshaping Lives Table of Contents System Overview Drill Guides... 4 Guide Pins... 5 Reamers... 4 Trials... 4 Broaches... 5 Impactors... 5 Instruments...

More information

Expert TN. Tibial Nail.

Expert TN. Tibial Nail. Expert TN. Tibial Nail. Surgical Technique Expert Nailing System This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Image intensifier

More information

Solitaire 850L/850/950 INSTALLATION GUIDE. American Steel Mortise Version PK2955

Solitaire 850L/850/950 INSTALLATION GUIDE. American Steel Mortise Version PK2955 INSTALLATION GUIDE Solitaire 850L/850/950 American Steel Mortise Version PK2955 INSTALLATION GUIDE Solitaire 850L/850/950 American Steel Mortise Version 1.0 Introduction...2 2.0 Checking the Door...2 3.0

More information