1 Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES A REPORT ON MEDICAL WORKSTATION CALIBRATION By Tom Schulte In years past, the quality of medical images was strictly controlled during the exposure, development, and viewing of hard-copy film. With today s digital imaging systems, the acquisition of digital images is still tightly controlled, following most of the same practices used for film. However, with modern softcopy (computer) display systems, image quality control is often neglected during the display and viewing process. Medical personnel need to know that each soft-copy image they view is an accurate depiction of the tissue and bone that were scanned. The presentation of each image must be repeatable to ensure that every image is shown consistently every time it is viewed. A digital image which has good diagnostic value when viewed on one viewing station should not look different when viewed on another station. This could result in difficulty in reading images, compromised diagnostic value, or even misdiagnosis. Reading difficulty and misdiagnosis due to inconsistent presentation of images can easily be avoided, however, by applying existing technical standards with the medical display calibration tools available today. Types of Medical Images There are two broad categories of patient images used in medical practices; radiology images and medical video images. Radiology images are scanner-sourced images that use some non-visible energy (e.g. X-ray, nuclear, ultrasound, magnetic) to form an interior view of human body parts. Medical video images are camera-sourced images that use visible light to form a surface view of human body parts or samples. Medical video may be motion video or still video images used in endoscopy, surgical, ophthalmology, dermatology, and microscopy applications. Due to differing characteristics between scannersourced radiology images and camera-sourced medical video images, there are different performance requirements for displays used to render the two types of images. In the following sections, we first cover aspects of display requirements for radiology imaging. We then cover medical video. Scanned Radiology Images X-ray scanning systems are designed to produce image brightness levels that are seen to have a linear relationship to the internal densities of the tissue and bone being scanned. However, these systems produce digital image data that has an approximately logarithmic relationship to the internal densities of the scanned tissue or bone. The contrast of darker areas of scanned images presented on electronic monitors must be enhanced to achieve perceptual linearity. If the logarithmic image data produced by these scanning systems were to be displayed on a computer monitor with a logarithmic luminance response*, the monitor would then produce image brightness levels that would be seen to have an approximately linear relationship to the scanned object densities. *A monitor s luminance response is the relationship between its digital drive levels (e.g ) and the resulting screen luminance. SpectraCal, Inc. Page 1

2 There are two important factors to consider, however, when displaying radiology images on computer monitors. 1. Computer monitors do not natively exhibit a logarithmic luminance response. Monitors usually have a luminance response that is approximately a power function, with an exponent (gamma) of 2.2, which is the industry srgb standard for presenting web and other photographic images. Of course, dedicated medical monitors are initially calibrated with the appropriate luminance response for radiology images. 2. The Human Visual System (HVS) exhibits an approximate logarithmic luminance response over much of its range, but not at lower brightness levels. The contrast sensitivity of the HVS depends on factors that include the absolute luminance level, the luminance difference, the ambient lighting environment, among others. Including all factors, human contrast sensitivity is approximately logarithmic above an image luminance of about 20 cd/m 2 (Figure 1). Figure 1: Above approximately 20 cd/m 2 (in yellow), human contrast sensitivity is almost purely logarithmic (linear plot with the log luminance scale). Below 20 cd/m 2 (in blue), contrast sensitivity decreases. At luminance levels below about 20 cd/m 2, however, the contrast sensitivity of the HVS diverges from its near-logarithmic response, with decreasing contrast sensitivity at lower luminance levels (Figure 1). Thus, the contrast of darker areas of scanned images presented on electronic monitors must be enhanced to achieve perceptual linearity between scanned object densities and image brightness. (Medical films on traditional light boxes produce luminance levels that are mostly above 20 cd/m 2 and don t require this contrast enhancement.) Therefore, for accurate presentation of scanned Of course, dedicated medical monitors are initially calibrated with the appropriate luminance response for radiology images, a monitor s luminance response should be approximately logarithmic at higher luminance levels, with additional compensation for reduced HVS contrast sensitivity at lower luminance levels. Inconsistent image presentation from one monitor to another causes inconsistent patient diagnoses and healthcare decisions. Radiology Imaging Standards A wide range of monitor types are used to display scanned radiology images that are used for making patient diagnoses and healthcare decisions. Since a given radiology image may be viewed at different locations within a healthcare organization, on monitors with different native characteristics, images produced by the same image file can present completely different visual appearance, information, and characteristics on those different display devices. This inconsistent image rendering from one electronic display to another in different viewing environments, if uncorrected, causes inconsistent clinical perception. Because the human visual system adapts to overall luminance, different display devices in different locations, with different maximum luminance values, can produce consistent presentation of grayscale images, provided that they have the same luminance ratio and the same luminance response function. This consistent presentation of images across different display devices is the goal of imaging standards for medical radiology displays. SpectraCal, Inc. Page 2

3 A number of technical standards and practice guidelines for medical imaging digital displays have been developed in the United States over the past 30+ years. Although these standards haven t had regulatory authority until recently in a few locales, they have been largely accepted by those medical facilities that have implemented quality control programs for their radiology workstations. DICOM Standard The American College of Radiology (ACR) and the National Electrical Manufacturers Association (NEMA) formed a committee in 1984 that produced the Digital Imaging and Communications in Medicine (DICOM) standard. [1] Part 14 of the DICOM standard specifies a luminance response model for scanned images known as the Grayscale Standard Display Function (GSDF). The intent was to cause radiology images to be displayed on any GSDF-conforming monitor with consistent contrast across the brightness range. The DICOM GSDF is the standard luminance response model for viewing scanned medical images. It is designed to produce approximately the same viewing experience on digital monitors that we would experience when viewing radiology film. Figure 2: The DICOM GSDF function, showing the luminance on a log scale as a function of the JND index, as obtained from Barten s model of human contrast sensitivity. Note that the same perceived contrast (JND range) can result from different absolute luminance differences. The GSDF maps scanned image values to digital driving levels (signal bit levels), such that equal changes in image values produce equal changes in perceived brightness. The brightness changes are measured in terms of Just-Noticeable Differences (JNDs) of brightness. The DICOM GSDF matches a monitor s luminance response to the nonlinear perceptual characteristics of our vision, based on Barten s model of human contrast sensitivity[2,3]. At luminance levels below about 20 cd/m 2, the DICOM GSDF departs from a near-logarithmic model to match the nonlinear perceptual characteristics of the human visual system (Figure 2). Each equal step in image values should produce a perceptually equal brightness change. Figure 3: Measured luminance response of typical uncalibrated monitor, compared to response of DICOM GSDF standard. SpectraCal, Inc. Page 3

4 Each equal step in image values should produce a perceptually equal brightness change, across the monitor s entire luminance range. This equal JND spacing of image brightness steps is called perceptual linearization. The DICOM Grayscale Standard Display Function (GSDF) insures that each image level captured by an X-ray scanner is uniquely perceivable on a GSDFconforming monitor. By standardizing and mapping displays to the contrast sensitivity of the human eye, DICOM calibration insures that a given difference in gray levels appears the same from one display to the next. This is consistent presentation of images. Note that the DICOM Standard defines the GSDF luminance response model for radiology monitors, but doesn t specify any test procedures or acceptance criteria. AAPM and ACR fulfill those needs. The DICOM GSDF insures that each image level captured by an X-ray scanner is uniquely perceivable on a GSDF-conforming monitor. AAPM Guidelines Task Group No. 18 of the AAPM published their On-Line Report No Assessment of Display Performance for Medical Imaging Systems [4]. The intent of the report was To provide standard guidelines to practicing medical physicists, engineers, researchers, and radiologists for the performance evaluation of electronic displays intended for medical use. The scope of the guidelines was limited to display devices that are used to display monochrome medical images. The guidelines provide test procedures and tools for evaluating the performance of radiology imaging displays. These include procedures for qualitative (visual) assessments of display performance as well as procedures for quantitative (measured) assessments of display performance. The guidelines also provide a set of test patterns for the qualitative and quantitative assessments. The TG18 guidelines also recommended quantitative (measured) QC acceptance criteria that were achievable by CRT displays. ACR Standards The American College of Radiology (ACR), in collaboration with the American Association of Physicists in Medicine (AAPM) and the Society for Imaging Informatics in Medicine (SIIM), has published technical standards [5] that, in part, cover radiology diagnostic and clinical displays. The ACR acceptance criteria for radiology diagnostic, mammography, and clinical displays is covered below, in the Luminance Factors section. The ACR Practice Guidelines & Technical Standards recommends QC acceptance criteria that are now achievable by LCD displays. Color in Radiology Images Although grayscale monitors are most commonly used on radiology reading room workstations, color monitors (both medical and commercial grades) are now being heavily used for clinical workstations and for some diagnostic workstations. The use of color has become significant for clinical images obtained from radiography, ultrasound, nuclear, CT, and MRI, and the color information is being used more and more for decision-making. Even when color is not required, commercial grade color monitors, rather than medical grade grayscale monitors, are often deployed in primary diagnostic and clinical review imaging stations for purposes of economy. Using color monitors in these formerly grayscale applications does add the need to test for consistent and accurate presentation of both grayscale and color, to not degrade the diagnostic value of either grayscale or color images presented on these color monitors. Color Standards There is no standard yet for how color radiology images are to be rendered on medical monitors (as opposed to camera-sourced color medical video images). Medical grade color radiology monitors are calibrated using DICOM GSDF. However, DICOM calibration doesn't correct a monitor s color performance. And, DICOM calibration decreases the SpectraCal, Inc. Page 4

5 contrast of camera-sourced color medical video images. As the DICOM GSDF does for grayscale levels, there needs to be a method to create perceptual linearity between color levels in medical images. Practice guidelines and standards for rendering color on medical displays are being established. Just as the JND metric is currently used to create perceptually uniform grayscale steps, the delta E 2000 color difference metric will be used to create perceptually uniform color steps. DICOM, AAPM, and ICC are currently coordinating efforts to develop methods and standards for consistently displaying color images on medical displays and CalMED is ready to implement those standards. Luminance Factors The luminance due to ambient light reflected from a monitor screen reduces the ability to distinguish individual image levels in the darker parts of a scanned image, if the luminance of the darkest image level is close to that of the ambient luminance. The ambient luminance effectively compresses or washes out the darker parts of the image and reduces the contrast in the image (Figure 4). Figure 4: Ambient light, when added to a monitor s native (uncalibrated) luminance response, causes the darkest image levels to be compressed, blending them together. However, a certain amount of ambient room light is required in a reading environment, to avoid eyestrain and to facilitate movement and interaction with other items in the area. It is recommended that about 20 lux of diffuse ambient light be used in diagnostic reading rooms. This is generally sufficient to avoid most screen reflections and still provide sufficient light for the human visual system to adapt to the surrounding environment and the displays. Minimum Luminance, Ambient Offset The lowest luminance elements of an image (Lmin) will not be visible if they are darker than the ambient reflected luminance (Lamb). To insure accurate visual perception of all image luminance levels, the luminance produced on a particular monitor by an image s darkest level needs to be significantly higher than that monitor s ambient luminance level. This means that a monitor s native black level normally needs to be increased, to enable accurate perception of the darkest image elements. Because digital RGB monitors don t usually have a black level control, the image black level needs to be increased before the image pixel values are applied to the monitor as digital drive levels. This is done with a lookup table (LUT) of correction values written to the workstation graphics adapter. According to the latest ACR technical standards for electronic medical imaging, the lowest luminance level from a scanned image (Lmin) needs to be at least four times the luminance level due to reflected ambient light (Lamb). Also, the lowest image luminance level plus the ambient luminance level (L min)*, needs to be at least 1.2 cd/m 2 for mammography displays, 1.0 cd/m 2 for primary diagnostic displays, and at least 0.8 cd/m 2 for all other radiology displays. *The presence of the prime symbol in L min indicates the inclusion of ambient light. Note that higher levels of ambient light can be tolerated when viewing medical films on light boxes, because, as noted above, the lowest luminance levels produced by trans-illuminated films are mostly above 20 cd/m 2 ; much higher than the lowest luminance levels produced by computer monitors displaying scanned medical images. Maximum Luminance SpectraCal, Inc. Page 5

6 The maximum luminance of a scanned radiology image (Lmax) needs to be high enough to yield sufficient image contrast (JND range) to visually detect each image luminance level difference. According to the ACR technical standards, a monitor s maximum luminance needs to be at least 420 cd/m 2 for diagnostic X-ray mammography, 350 cd/m 2 for diagnostic X-ray radiography, and at least 250 cd/m 2 for all other modalities. Multiple monitors on a radiology workstation should have no greater than a 10% difference in maximum luminance between the monitors. Luminance Ratio A monitor s luminance ratio (LR) is the ratio of the monitor s maximum luminance to its minimum luminance, including ambient (L min). The monitor s LR is basically its contrast ratio in the presence of ambient light. Because the human visual system adapts to the average luminance of a still image, and is not able to optimally adapt to very bright and very dark regions at the same time, we need to limit an image s overall luminance ratio to a value for which the HVS has reasonably good adapted contrast response. If we reproduce images with a luminance ratio above about 500, no contrast will be perceived in dark regions of the images. If we reproduce images with a luminance ratio below about 250, however, the image will lack sufficient contrast between individual image levels and will appear washed out. According to the ACR technical standards, a monitor s luminance ratio needs to be about 350 for optimum contrast (on diagnostic and mammography monitors), and at least 250 for acceptable contrast (on clinical review monitors). The luminance ratio should be consistent on all radiology workstations throughout a medical facility to assure consistent presentation of images. Note that a monitor s ability to produce a desired luminance ratio depends primarily on the monitor s maximum luminance capability, along with its minimum luminance level (increased to compensate for ambient light). Medical and high end commercial grade monitors will likely be able to produce a suitable luminance ratio, but many commoditypriced commercial grade monitors may have marginal or insufficient maximum luminance (measured in cd/m 2, inaccurately called brightness) to produce a desired luminance ratio. Different display devices can give consistent presentation of images provided that they have: 1. The same luminance ratio (LR), and 2. The same luminance response model (DICOM GSDF). Bottom Line Because the human visual system adapts to overall brightness, different display devices, with different maximum luminance values, can give consistent presentation of images provided that they have: 3. The same luminance ratio (LR), and 4. The same luminance response model (DICOM GSDF). Grades of Viewing Monitors RGB monitors of various performance grades are used to view scanned images on radiology viewing stations, depending upon each viewing station s purpose, be it for primary diagnostic, clinical review, or technologist modality workstation. Monitors used for medical imaging may be purpose-built medical grade monitors or they may be commercial grade monitors. Both purpose-built medical grade monitors and commercial grade monitors are able to fill the needs for rendering radiology scanned images in various use cases. Commercial Grade Monitors Commercial grade monitors typically have lower luminance output and no luminance stabilization feature, compared to medical grade monitors. They are typically set to produce their maximum luminance output, typically cd/m 2, during initial setup. Without automatic luminance stabilization, commercial monitors are unable to SpectraCal, Inc. Page 6

7 maintain constant luminance during backlight warmup time. When the monitor is first turned on, luminance output tends to be lower until the backlight warms up. They slowly increase to maximum luminance over the first 30 to 60 minutes of run time. Also, a bright commercial monitor that produces 350 cd/m 2 when put into service decays to 250 cd/m 2 in about 18 months. Since cd/m 2 is considered the appropriate luminance range for viewing radiography images, the brightest commercial grade monitors maintain acceptable luminance for only about 18 months. All monitors used to display scanned radiology images require both initial and annual calibration and periodic conformance testing to insure conformance to industry guidelines. Medical Grade Radiology Monitors Medical grade radiology monitors, which are usually used on diagnostic stations in hospital radiology reading rooms, have several features that make them most suitable for use on critical diagnostic reading stations. Medical grade monitors are generally capable of much higher luminance than commercial monitors, typically producing cd/m 2. Medical grade monitors typically have a backlight capable of producing much more luminance than required when the monitor is new. This surplus luminance capability is beneficial because medical grade monitors have a luminance stabilization circuit, with feedback from a backlight sensor, which adjusts the backlight luminance down to the desired output level ( cd/m 2 ). Since a medical grade monitor has extra initial luminance capability, it adjusts its backlight output to compensate for natural luminance decay, keeping the luminance output constant over time. This backlight feedback circuit is capable of maintaining constant luminance not only over the long term life of the monitor (usually about 5 years), but also in the short term, during monitor warm-up. An advantage of brighter displays is that they allow radiologists to work in brighter reading environments, reducing viewing fatigue. The display s minimum luminance should always be regulated, though, to produce the optimum luminance ratio. Per ACR: For brighter monitors, L'min, should be proportionately larger to maintain the same LR. Medical grade radiology monitors may also have a front calibration sensor. The sensor will be integrated with an auto-calibration function within the monitor. Medical Grade Mammography Monitors Medical grade monitors designed for viewing mammography images are generally capable of higher luminance than even general radiology medical monitors, some over 1500 cd/m 2. This is desirable for displaying mammography images, which tend to be much darker than other X-ray images. Like general radiology medical monitors, mammography monitors have a luminance stabilization circuit, with feedback from a backlight sensor, which adjusts the backlight luminance down to the desired output level (420 cd/m 2 for mammography). The front calibration sensor that is optional on general radiology medical monitors is standard on all mammography monitors. An auto-calibration function within the monitor uses the front sensor for DICOM conformance tests and calibration. All monitors used to display scanned radiology images require both initial and annual calibration and periodic conformance testing to insure conformance to industry performance guidelines. Medical grade monitors, with their automatic luminance stabilization circuits, usually require quarterly conformance testing and only annual recalibration, while commercial grade monitors may require quarterly or monthly recalibration to assure continued conformance. SpectraCal, Inc. Page 7

8 Radiology Workstation Test Schedule Both AAPM and ACR recommend a regular schedule of workstation testing. Following is a typical test schedule that ensures continued compliance to industry standards, high reading efficiency, and no contribution to misdiagnoses due to poor image quality. Annual Tests - Performed by medical physicist Test ambient lighting Calibrate display(s) Run integrated calibration software, if available, or Calibrate with CalMED software and certified external meter. Validate display conformance Perform full range of visual tests with test patterns. Conformance test with CalMED software and certified external meter. Document Results Save CalMED conformance report. Quarterly/Monthly Tests - Performed by medical physicist or QC tech Test ambient lighting Analyze display conformance Perform luminance response visual tests (review TG18-QC and TG18-CT patterns). Conformance test with CalMED software and certified external meter. If poor conformance, calibrate display(s) Run integrated calibration software, if available, or Calibrate with CalMED software and certified external meter. Validate display conformance Conformance test with CalMED software and certified external meter. Document Results Save CalMED conformance report. Daily/Weekly Tests - Performed by workstation user Perform luminance response visual tests (review TG18-QC and/or TG18-CT patterns). Radiology Imaging CalMED Calibration The Radiology Imaging Technical QC workflow that is provided with CalMED guides us through the process of calibrating a radiology imaging display by implementing industry standard radiology imaging guidelines from DICOM, AAPM, and ACR. An outline of the workflow (Figure 5) is viewable when you open the left hand CalMED panel. Figure 5: An outline of the Radiology Imaging workflow appears when you open the left hand CalMED panel. Sections 1 and 2 of the workflow, Initial Setup and Workstation Analysis are used to test whether a display conforms to industry standards. Sections 3 and 4, Display Optimization and Workstation Calibration, are used to calibrate a display for conformance. The workflow first guides us through measuring the ambient luminance, setting our minimum luminance target for the display, and then testing the display for conformance to the standard acceptance criteria. If the test shows the display to be in conformance, we can save, view, and/or print a conformance report for the display. If the conformance test shows that the display is not in conformance, we can then proceed to the additional workflow steps to optimize any controls on the display and then calibrate the workstation. This would be followed by a repeat of the SpectraCal, Inc. Page 8

9 conformance test, to validate and document the display s conformance to the standard acceptance criteria. For a detailed walkthrough of the CalMED Radiology Imaging workflow, see the paper Medical Display Calibration Walkthrough, available for download at: pers/medicaldisplaycalibrationwalkthrough.pdf Medical Video Standards Images viewed on medical video displays originate from cameras, rather than from scanners. These endoscopic and still image cameras are used in surgery, ER, clinics, dermatology and pathology, among others. Camera-sourced medical images conform to the 1999 srgb standard from the International Electrotechnical Commission (IEC). Consumer digital cameras and internet images also follow the srgb standard, which was designed to be appropriate for brighter viewing environments. The srgb luminance response model starts with a linear section just above black (gamma of 1), then transitions to a power function at higher levels, with a gamma of 2.4. The gamma over the entire luminance range averages to about 2.2. The srgb white point (D65) and color gamut (primary colors) are the same as for studio monitors and HDTV. Medical Video CalMED Calibration The Medical Video workflow that is provided with CalMED guides us through the process of calibrating a medical video display by implementing the srgb industry standard. An outline of the workflow (Figure 6) is viewable when you open the left hand CalMED panel. Figure 6: An outline of the Medical Video workflow appears when you open the left hand CalMED panel. Sections 1 and 2 of the workflow, Initial Setup and Workstation Analysis are used to test whether a display conforms to industry standards. Sections 3 and 4, Display Optimization and Workstation Calibration, are used to calibrate a display for conformance. The workflow first guides us through testing the medical video display for conformance to the srgb standard. If the test shows the display to be in conformance, we can save, view, and/or print a conformance report for the display. If the conformance test shows that the display is not in conformance, we can then proceed to the additional workflow steps to optimize any controls on the display and then calibrate the workstation. This would be followed by a repeat of the conformance test, to validate and document the display s conformance to the srgb standard. For a detailed walkthrough of the CalMED Medical Video workflow, see the paper Medical Display Calibration Walkthrough, available for download at: pers/medicaldisplaycalibrationwalkthrough.pdf SpectraCal, Inc. Page 9

10 Technical Paper Conclusion It is critical to provide consistent presentation of medical images, in order to optimize reading efficiency and diagnostic value and to avoid misdiagnoses due to compromised image quality. The medical industry has developed, and continues to develop, technical standards and practice guidelines that follow the characteristics of the human visual system when rendering images on soft-copy electronic monitors. By maintaining a consistent luminance ratio and luminance response model plus appropriate luminance levels across all displays, you can ensure that medical images will be displayed consistently and accurately on all imaging workstations throughout a medical network. By maintaining conformance to the appropriate industry standards, both radiology imaging and medical video workstations will consistently render their respective images for efficient and accurate reading. The CalMED Radiology Imaging and Medical Video workflows provide easy to use processes for routinely testing workstation conformance to standards and, when necessary, provide fast and accurate calibration processes to bring workstations into conformance. Both CalMED workflows provide conformance reports to fully document a workstation quality assurance program. Display Profiling Solutions Page 10 A Special Report for Imaging Professionals

11 References 1. NEMA: Digital Imaging and Communications in Medicine (DICOM), DICOM PS a - Grayscale Standard Display Function. Available at 2. Barten PGJ. Physical model for contrast sensitivity of the human eye. Proc SPIE Int Soc Opt Eng.1992; 1666: pp Barten PGJ. Contrast Sensitivity of the Human Eye and Its Effects on Image Quality. Bellingham, WA: SPIE Press; Assessment of Display Performance for Medical Imaging Systems, American Association of Physicists in Medicine (AAPM), Task Group 18. Available at 5. ACR AAPM SIIM Technical Standard for Electronic Practice of Medical Imaging, Res Available at 6. ACR AAPM SIIM Practice Parameter for Determinants of Image Quality in Digital Mammography Res , Amended 2014 (Res. 39). Available at SpectraCal, Inc. Page 11

12 About the Author: Tom Schulte is Product Development Manager at SpectraCal. Tom has extensive experience in electronic systems test, calibration and service, as well as electronics test instrument design and usage and has authored numerous technical white papers. Tom was previously an Application Engineer at Sencore for over twenty years, where he was involved in video, audio, and RF test instrument design, plus training and support for electronic test equipment users. About SpectraCal: SpectraCal specializes in the tools and training necessary to achieve images representative of the original subjects for environments from low to high ambient light while achieving the dynamic range, contrast, and colorimetry necessary for the image to have the proper presentation to the viewer. SpectraCal CalMED software was developed to support medical personnel in the step by step process of workstation calibration. The foundation of medical workstation calibration is to understand the elements in a display that require adjustment and how each element inter-relates to the others. From its inception, CalMED has earned rave reviews and has become the preeminent medical workstation calibration software on the market, compatible with virtually all color meters available today. As the medical industry and display technology evolves, CalMED continues to provide the first choice for medical workstation calibration solutions. More Information: For more information on CalMED medical workstation calibration: Visit: info@spectracal.com Phone: SpectraCal, Inc Midvale Avenue N., Suite 100 Shoreline, WA info@spectracal.com SpectraCal, Inc. Page 12

Technical Paper DENTAL MONITOR CALIBRATION

Technical Paper DENTAL MONITOR CALIBRATION Technical Paper DENTAL MONITOR CALIBRATION A REPORT ON DENTAL IMAGE PRESENTATION By Tom Schulte Dental radiographs and oral photographs are often both viewed on the same dental workstation monitor. The

More information

Calibrating Computer Monitors for Accurate Image Rendering

Calibrating Computer Monitors for Accurate Image Rendering Calibrating Computer Monitors for Accurate Image Rendering SpectraCal, Inc. 17544 Midvale Avenue N. Shoreline, WA 98133 (206) 420-7514 info@spectracal.com http://color.spectracal.com Executive Summary

More information

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors Health NYC Quality Assurance Guidelines for Primary Diagnostic Monitors June 2015 Content I. Overview... 1 II. Definitions... 1 III. Acceptance Testing Requirements... 4 IV. Bi-weekly PDM Testing... 4

More information

Medical imaging monitors specification guidelines

Medical imaging monitors specification guidelines Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,

More information

Technical Paper DISPLAY PROFILING SOLUTIONS

Technical Paper DISPLAY PROFILING SOLUTIONS Technical Paper DISPLAY PROFILING SOLUTIONS A REPORT ON 3D LUT CREATION By Joel Barsotti and Tom Schulte A number of display profiling solutions have been developed to correct image rendering errors in

More information

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists

More information

MQSA Quality Control Manual for Monochrome Displays for Mammography

MQSA Quality Control Manual for Monochrome Displays for Mammography MQSA Quality Control Manual for Monochrome Displays for Mammography Ver1.0 Applicable Models LMD-DM50, LMD-DM30 Sony Corporation - 1 - MQSA Quality Control Manual for Monochrome Displays for Mammography

More information

Calibration Best Practices

Calibration Best Practices Calibration Best Practices for Manufacturers SpectraCal, Inc. 17544 Midvale Avenue N., Suite 100 Shoreline, WA 98133 (206) 420-7514 info@spectracal.com http://studio.spectracal.com Calibration Best Practices

More information

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for

More information

Coronis 5MP Mammo. The standard of care for digital mammography

Coronis 5MP Mammo. The standard of care for digital mammography Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment

More information

Nio. Industry-standard diagnostic display systems

Nio. Industry-standard diagnostic display systems Nio Industry-standard diagnostic display systems Diagnostic confidence in grayscale With the Nio diagnostic display system, Barco brings dependable diagnostic imaging to its true potential. Presenting

More information

Understanding Digital Modalities: System Integration and Use

Understanding Digital Modalities: System Integration and Use Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information

More information

Monitor Quality Control Solutions

Monitor Quality Control Solutions Monitor Quality Control Solutions Managing the quality of monitors used every day leads to the improvement of the quality of medical care itself. Detecting changes in a monitor s display quality Controlling

More information

Consistency of Softcopy and Hardcopy: Preliminary Experiences with the new DICOM Extensions for Image Display

Consistency of Softcopy and Hardcopy: Preliminary Experiences with the new DICOM Extensions for Image Display Consistency of Softcopy and Hardcopy: Preliminary Experiences with the new DICOM Extensions for Image Display Marco Eichelberg a, Jörg Riesmeier a, Klaus Kleber b, Jörg Holstein b, Herman Oosterwijk c,

More information

What s New in DICOM. DICOM Workshop @ SPIE MEDICAL IMAGING February 2010. Bas Revet, Philips Healthcare (Chair WG 6)

What s New in DICOM. DICOM Workshop @ SPIE MEDICAL IMAGING February 2010. Bas Revet, Philips Healthcare (Chair WG 6) What s New in DICOM DICOM Workshop @ SPIE MEDICAL IMAGING February 2010 Bas Revet, Philips Healthcare (Chair WG 6) DICOM is the International Standard for Medical Imaging and related information: images,

More information

EIZO Monitor CalMAN Setup Guide

EIZO Monitor CalMAN Setup Guide EIZO Monitor CalMAN Setup Guide CalMAN Display Calibration Software can automatically create optimized calibration 1D or 3D LUTs for EIZO monitors equipped with a self-calibration sensor. CalMAN controls

More information

Monitor Qualit y Control Solutions

Monitor Qualit y Control Solutions Monitor Qualit y Control Solutions Controlling the Quality of Monitors Managing the quality of monitors used every day leads to the improvement of the quality of medical care itself. Q1 A1 Do all monitors

More information

A PACS-Aware DICOM Image Object

A PACS-Aware DICOM Image Object Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry

More information

Barco medical display systems. Images you can trust. Every time. Everywhere.

Barco medical display systems. Images you can trust. Every time. Everywhere. Barco medical display systems Images you can trust. Every time. Everywhere. In healthcare imaging, accuracy is of vital importance. As a medical professional, you count on display technology that delivers

More information

EIZO Sensor Solutions for Medical Monitors. Backlight Sensor / External Sensor / Swing Sensor

EIZO Sensor Solutions for Medical Monitors. Backlight Sensor / External Sensor / Swing Sensor White Paper EIZO Sensor Solutions for Medical Monitors Backlight Sensor / External Sensor / Swing Sensor CONTENTS 1 Introduction... 2 2 What is Monitor Quality Control?... 3 3 Backlight Sensor... 4 3-1

More information

REQUIREMENTS FOR MEDICAL IMAGING MONITORS (part I) TERMINOLOGY NEEDED TO UNDERSTAND MONITOR PERFORMANCE

REQUIREMENTS FOR MEDICAL IMAGING MONITORS (part I) TERMINOLOGY NEEDED TO UNDERSTAND MONITOR PERFORMANCE REQUIREMENTS FOR MEDICAL IMAGING MONITORS (part I) By Ken Compton and Herman Oosterwijk The question of what monitor to use for diagnostic purposes with digital medical images has been a recurring theme

More information

DICOM Digital Imaging and Communications in Medicine

DICOM Digital Imaging and Communications in Medicine DICOM Digital Imaging and Communications in Medicine INF5081 / Vårsemester 2007 Wolfgang Leister Reidar Øksnevad Pedro Cruz G. April 2007 PACS PACS = Picture Archive and Communication System Archive for

More information

A Proposal for OpenEXR Color Management

A Proposal for OpenEXR Color Management A Proposal for OpenEXR Color Management Florian Kainz, Industrial Light & Magic Revision 5, 08/05/2004 Abstract We propose a practical color management scheme for the OpenEXR image file format as used

More information

Technical Paper SPECTRACAL C6 HDR

Technical Paper SPECTRACAL C6 HDR Technical Paper SPECTRACAL C6 HDR HIGH LUMINANCE COLORIMETER By Tom Schulte and Darrell Bird High Dynamic Range (HDR) and Ultra HD TV (UHD) have been big news items in the entertainment industry lately

More information

Contrast ratio what does it really mean? Introduction...1 High contrast vs. low contrast...2 Dynamic contrast ratio...4 Conclusion...

Contrast ratio what does it really mean? Introduction...1 High contrast vs. low contrast...2 Dynamic contrast ratio...4 Conclusion... Contrast ratio what does it really mean? Introduction...1 High contrast vs. low contrast...2 Dynamic contrast ratio...4 Conclusion...5 Introduction Contrast, along with brightness, size, and "resolution"

More information

Medical Imaging Display Colour Space (mrgb) Teleconference 17 July 2014 15:00 (UK) / 10:00 (EST)

Medical Imaging Display Colour Space (mrgb) Teleconference 17 July 2014 15:00 (UK) / 10:00 (EST) Medical Imaging Display Colour Space (mrgb) Teleconference 17 July 2014 15:00 (UK) / 10:00 (EST) The meeting was called to order at 10:00 am (EST) by Craig Revie, acting chair, with the following attendees:

More information

Dolby Vision for the Home

Dolby Vision for the Home Dolby Vision for the Home 1 WHAT IS DOLBY VISION? Dolby Vision transforms the way you experience movies, TV shows, and games with incredible brightness, contrast, and color that bring entertainment to

More information

Dome CXtra User s Guide

Dome CXtra User s Guide Dome CXtra User s Guide 2015 NDS Surgical Imaging, LLC. All rights reserved. This document contains proprietary information of NDS Surgical Imaging. It is exclusive property of NDS Surgical Imaging. It

More information

ACR AAPM SIIM TECHNICAL STANDARD FOR ELECTRONIC PRACTICE OF MEDICAL IMAGING

ACR AAPM SIIM TECHNICAL STANDARD FOR ELECTRONIC PRACTICE OF MEDICAL IMAGING The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

Implementing ISO12646 standards for soft proofing in a standardized printing workflow

Implementing ISO12646 standards for soft proofing in a standardized printing workflow Implementing ISO12646 standards for soft proofing in a standardized printing workflow Aditya Sole Implementing ISO12646 standards for soft proofing in a standardized Gjøvik University College, The Norwegian

More information

Color quality guide. Quality menu. Color quality guide. Page 1 of 6

Color quality guide. Quality menu. Color quality guide. Page 1 of 6 Page 1 of 6 Color quality guide The Color Quality guide helps users understand how operations available on the printer can be used to adjust and customize color output. Quality menu Menu item Print Mode

More information

Spyder 5EXPRESS Hobbyist photographers seeking a simple monitor color calibration solution.

Spyder 5EXPRESS Hobbyist photographers seeking a simple monitor color calibration solution. Spyder5 User s Guide Spyder 5EXPRESS Hobbyist photographers seeking a simple monitor color calibration solution. Next Generation Spyder5 colorimeter with dual purpose lens cap and counterweight, improved

More information

Digital Mammography Monitors

Digital Mammography Monitors Digital Mammography Monitors Film to Filmless Mammography Diagnosis As the incidence of breast cancer rises, regular mammograms are recommended for early detection. It is vital in the process of early

More information

User s Guide. Version 1.2.00

User s Guide. Version 1.2.00 User s Guide Version 1.2.00 2 MULTIPROFILER - USER S GUIDE Software Updates Occasionally updates and enhancements to the MultiProfiler software will be made available. Visit the MultiProfiler section of

More information

Philips HDR technology White paper

Philips HDR technology White paper Philips HDR technology White paper Philips International B.V. Version 1.0 2015-08-21 1. Introduction Current video formats are still based upon the display capabilities of CRT monitors. They cannot capture

More information

Barco surgical displays. High-accuracy visualization solutions for surgery and endoscopy

Barco surgical displays. High-accuracy visualization solutions for surgery and endoscopy Barco surgical displays High-accuracy visualization solutions for surgery and endoscopy Near-patient surgical displays The complexity of general and minimally invasive surgery places high demands on technology

More information

Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function

Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function PS 3.14-2011 Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function Published by National Electrical Manufacturers Association 1300 N. 17th Street Rosslyn,

More information

Overview. Raster Graphics and Color. Overview. Display Hardware. Liquid Crystal Display (LCD) Cathode Ray Tube (CRT)

Overview. Raster Graphics and Color. Overview. Display Hardware. Liquid Crystal Display (LCD) Cathode Ray Tube (CRT) Raster Graphics and Color Greg Humphreys CS445: Intro Graphics University of Virginia, Fall 2004 Color models Color models Display Hardware Video display devices Cathode Ray Tube (CRT) Liquid Crystal Display

More information

User s Guide. Version 1.0.00

User s Guide. Version 1.0.00 User s Guide Version 1.0.00 Software Updates NEC Display Wall Calibrator - USER S GUIDE Occasionally updates and enhancements to the NEC Display Wall Calibrator software will be made available. Visit the

More information

Color management workflow in Adobe After Effects CS4

Color management workflow in Adobe After Effects CS4 Color management workflow in Adobe After Effects CS4 Technical paper Table of contents 1 Getting started 3 High-definition video workflow 7 Digital cinema workflow 14 Animation/Flash export workflow 19

More information

Coronis 3MP LED Industry-leading diagnostic display system

Coronis 3MP LED Industry-leading diagnostic display system Coronis 3MP LED Industry-leading diagnostic display system Coronis Twenty years of experience and innovation in diagnostic imaging have culminated in the unrivaled Coronis 3MP LED display system. Combining

More information

Scanners and How to Use Them

Scanners and How to Use Them Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

PARCA Certified PACS System Analyst (CPSA2014) Requirements

PARCA Certified PACS System Analyst (CPSA2014) Requirements PARCA Certified PACS System Analyst (CPSA2014) Requirements Copy right notice: Copyright 2014 PACS Administrators in Radiology Certification Association (PARCA). All rights reserved. All rights reserved.

More information

Outline. Quantizing Intensities. Achromatic Light. Optical Illusion. Quantizing Intensities. CS 430/585 Computer Graphics I

Outline. Quantizing Intensities. Achromatic Light. Optical Illusion. Quantizing Intensities. CS 430/585 Computer Graphics I CS 430/585 Computer Graphics I Week 8, Lecture 15 Outline Light Physical Properties of Light and Color Eye Mechanism for Color Systems to Define Light and Color David Breen, William Regli and Maxim Peysakhov

More information

KIP 3000 Series MULTIFUNCTION SIMPLICITY

KIP 3000 Series MULTIFUNCTION SIMPLICITY KIP 3000 Series MULTIFUNCTION SIMPLICITY KIP 3000 SERIES SYSTEMS The KIP 3000 is a highly hl advanced d wide format digital copy system that provides the ability to easily add network printing and scan-to-file

More information

S-Log: A new LUT for digital production mastering and interchange applications

S-Log: A new LUT for digital production mastering and interchange applications S-Log: A new LUT for digital production mastering and interchange applications Hugo Gaggioni, Patel Dhanendra, Jin Yamashita (Sony Broadcast and Production Systems) N. Kawada, K. Endo ( Sony B2B Company)

More information

Reducing eyestrain from video and computer monitors

Reducing eyestrain from video and computer monitors Reducing eyestrain from video and computer monitors Charles A. Poynton www.poynton.com poynton@poynton.com tel +1 416 413 1377 fax +1 416 413 1378 Many of us work for hours each day in front of displays.

More information

MediCal QAWeb. Secured online service for Quality Assurance of medical displays

MediCal QAWeb. Secured online service for Quality Assurance of medical displays MediCal QAWeb Secured online service for Quality Assurance of medical displays The first secured online service for high-grade Quality Assurance MediCal QAWeb is the industry s first online service for

More information

CALIBRATION AND OPERATION OF PANASONIC PLASMA MONITORS JULY 2009

CALIBRATION AND OPERATION OF PANASONIC PLASMA MONITORS JULY 2009 CALIBRATION AND OPERATION OF PANASONIC PLASMA MONITORS JULY 2009 Overview Plasma video monitors provide high resolution images with excellent contrast and dynamic range, low black levels, and saturated

More information

DryView CHROMA Imaging System

DryView CHROMA Imaging System DryView CHROMA Imaging System VERSATILE PRINTING ON FILM OR PAPER One of the most trusted names in medical printing now offers a highly versatile printer that empowers physicians and other healthcare professionals

More information

Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function

Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function PS 3.14-2004 Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function Published by National Electrical Manufacturers Association 1300 N. 17th Street Rosslyn,

More information

Color Workflows for Adobe Creative Suite 3. A Self-Help Guide

Color Workflows for Adobe Creative Suite 3. A Self-Help Guide Color Workflows for Adobe Creative Suite 3 A Self-Help Guide Adobe Systems Incorporated 345 Park Avenue, San Jose, CA 95110-2704 USA www.adobe.com Adobe, the Adobe logo, Dreamweaver, Illustrator, Photoshop,

More information

CAR Standards for Teleradiology

CAR Standards for Teleradiology CAR Standards for Teleradiology The standards of the Canadian Association of Radiologists (CAR) are not rules, but are guidelines that attempt to define principles of practice that should generally produce

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Color Management Terms

Color Management Terms Written by Jonathan Sachs Copyright 2001-2003 Digital Light & Color Achromatic Achromatic means having no color. Calibration Calibration is the process of making a particular device such as a monitor,

More information

A Adobe RGB (1998) Color Image Encoding. Version 2005-05 May 2005

A Adobe RGB (1998) Color Image Encoding. Version 2005-05 May 2005 Adobe RGB (1998) Color Image Encoding Version 2005-05 May 2005 ADOBE SYSTEMS INCORPORATED Corporate Headquarters 345 Park Avenue San Jose, CA 95110-2704 (408) 536-6000 http://www.adobe.com Copyright 2005

More information

Digital Image Processing with DragonflyPACS

Digital Image Processing with DragonflyPACS Digital Image Processing with DragonflyPACS The Smarter PACS The smartest image processing system in veterinary medicine. Outstanding image quality that s compliant with DICOM 3.0 standards. Compatible

More information

CARESTREAM PACS Suite (Client, Server, and CD Direct) Version 11.0. DICOM Conformance Statement

CARESTREAM PACS Suite (Client, Server, and CD Direct) Version 11.0. DICOM Conformance Statement CARESTREAM PACS Suite (Client, Server, and CD Direct) Version 11.0 DICOM Conformance Statement Publication # 8G8913 22 April 2009 Table of Contents Table of Contents...2 1 Introduction...5 1.1 Abbreviations

More information

DIVA MEDICAL DISPLAY SOLUTION FEATURE

DIVA MEDICAL DISPLAY SOLUTION FEATURE DIVA MEDICAL DISPLAY SOLUTION FEATURE IPX5 High Image IPX1 Quality Auto Luminance Stabilizer AIS RS-232 TOUCH 1080P CH DIVA Medical s displays have the industry s highest quality image reproduction in

More information

Overview Help Files Viewing Photos

Overview Help Files Viewing Photos User s Guide Overview SpyderGallery is a free mobile App from Datacolor, designed to provide color accurate viewing of images on your mobile device. The Image Viewer can be used with no special preparation;

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Sony B2B. Sony Corporation. S-Log White Paper. S-Log within Digital Intermediate workflow designed for cinema release. Version 1.12.

Sony B2B. Sony Corporation. S-Log White Paper. S-Log within Digital Intermediate workflow designed for cinema release. Version 1.12. Sony Corporation S-Log within Digital Intermediate workflow designed for cinema release Version 1.12.3 1/16 Index 1 S-Log Characteristics... 3 1.1 CHANGING THE SENSITIVITY (PUSH/PULL PROCESS)... 7 1.1.1

More information

WHITE PAPER. Methods for Measuring Flat Panel Display Defects and Mura as Correlated to Human Visual Perception

WHITE PAPER. Methods for Measuring Flat Panel Display Defects and Mura as Correlated to Human Visual Perception Methods for Measuring Flat Panel Display Defects and Mura as Correlated to Human Visual Perception Methods for Measuring Flat Panel Display Defects and Mura as Correlated to Human Visual Perception Abstract

More information

S-Log and ACES workflow in Vegas Pro 12.0

S-Log and ACES workflow in Vegas Pro 12.0 Workflow Paper S-Log and ACES workflow in Vegas Pro 12.0 Revised August 8, 2012 Vegas Pro 12.0 supports the expanded dynamic range of cameras using S-Log encoding, such as the Sony PMW-F3 camera. S-Log

More information

A Adobe RGB Color Space

A Adobe RGB Color Space Adobe RGB Color Space Specification Version DRAFT October 2, 2004 Please send comments to mailto:lars.borg@adobe.com This publication and the information herein are subject to change without notice, and

More information

Otis Photo Lab Inkjet Printing Demo

Otis Photo Lab Inkjet Printing Demo Otis Photo Lab Inkjet Printing Demo Otis Photography Lab Adam Ferriss Lab Manager aferriss@otis.edu 310.665.6971 Soft Proofing and Pre press Before you begin printing, it is a good idea to set the proof

More information

Medical Imaging Display Colour Space (mrgb) Teleconference 16 January 2014 15:00 (UK) / 10:00 (EST)

Medical Imaging Display Colour Space (mrgb) Teleconference 16 January 2014 15:00 (UK) / 10:00 (EST) Medical Imaging Display Colour Space (mrgb) Teleconference 16 January 2014 15:00 (UK) / 10:00 (EST) The meeting was called to order at 10:00 am (EST) by Craig Revie, acting chair, with the following attendees:

More information

anatomage table Interactive anatomy study table

anatomage table Interactive anatomy study table anatomage table Interactive anatomy study table Anatomage offers a unique, life-size interactive anatomy visualization table for the medical community. Anatomage Table offers an unprecedented realistic

More information

WE ARE in a time of explosive growth

WE ARE in a time of explosive growth The Spatial Standard Observer: A new tool for display metrology In the design of displays, beauty is in the eye of the beholder. But until recently, the industry has lacked tools to estimate quality as

More information

PARCA Certified PACS Interface Analyst (CPIA) Requirements

PARCA Certified PACS Interface Analyst (CPIA) Requirements PARCA Certified PACS Interface Analyst (CPIA) Requirements Copyright notice: Copyright 2005 PACS Administrators in Radiology Certification Association (PARCA). All rights reserved. All rights reserved.

More information

Digital Color Workflows and the HP DreamColor LP2480zx Professional LCD Display

Digital Color Workflows and the HP DreamColor LP2480zx Professional LCD Display Digital Color Workflows and the HP DreamColor LP2480zx Professional LCD Display Improving accuracy and predictability in color processing at the designer s desk can increase productivity and improve quality

More information

ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY

ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

RGB Color Managed Workflow Example

RGB Color Managed Workflow Example White Paper 23 Level: Basic Date: May 2008 RGB Color Managed Workflow Example Overview Today s abundance of RGB source art such as digital illustrations, photos, and digital art are routinely repurposed

More information

Breast MRI Quality Control

Breast MRI Quality Control Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation

More information

Color Balancing Techniques

Color Balancing Techniques Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction Color balancing refers to the process of removing an overall color bias from an image. For example, if an image appears

More information

Patient Prep Information

Patient Prep Information Stereotactic Breast Biopsy Patient Prep Information Imaging Services Cannon Memorial Hospital Watauga Medical Center Table Weight Limits for each facility Cannon Memorial Hospital Watauga Medical Center

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

The Digital Dog. Exposing for raw (original published in Digital Photo Pro) Exposing for Raw

The Digital Dog. Exposing for raw (original published in Digital Photo Pro) Exposing for Raw Exposing for raw (original published in Digital Photo Pro) The Digital Dog Exposing for Raw You wouldn t think changing image capture from film to digital photography would require a new way to think about

More information

ICC Profiles Guide. English Version 1.0

ICC Profiles Guide. English Version 1.0 English Version.0 Dedicated ICC Profile Types Dedicated ICC profiles are provided for the following three media types: Photo Paper Pro Photo Paper Plus Glossy Matte Photo Paper Compatible Printers A00II

More information

Getting to Know Avreo

Getting to Know Avreo Getting to Know Avreo Avreo develops complete Radiology Workflow Solutions (RWS) that manage every aspect of your workflow from order entry to report distribution within a single application/single database

More information

High Dynamic Range Video The Future of TV Viewing Experience

High Dynamic Range Video The Future of TV Viewing Experience The Future of TV Viewing Experience - White Paper - www.keepixo.com Introduction The video industry has always worked to improve the TV viewing experience. More than a decade ago, the transition from SD

More information

Video Camera Image Quality in Physical Electronic Security Systems

Video Camera Image Quality in Physical Electronic Security Systems Video Camera Image Quality in Physical Electronic Security Systems Video Camera Image Quality in Physical Electronic Security Systems In the second decade of the 21st century, annual revenue for the global

More information

NHS Imaging and Radiodiagnostic activity in England. 2012/13 Release. August 2013

NHS Imaging and Radiodiagnostic activity in England. 2012/13 Release. August 2013 NHS Imaging and Radiodiagnostic activity in England 2012/13 Release August 2013 Commentary This National Statistics release covers Imaging and Radiodiagnostic examinations or tests carried out in the NHS

More information

W O R K F L O W T E S T F O R M S

W O R K F L O W T E S T F O R M S W O R K F L O W T E S T F O R M S E N D - T O - E N D W O R K F L O W I M A G E Q U A L I T Y A S S U R A N C E U S E R G U I D E Isis Imaging Corporation Isis Imaging Corporation Workflow Test Forms Suite

More information

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

First floor, Main Hospital North Services provided 24/7 365 days per year

First floor, Main Hospital North Services provided 24/7 365 days per year First floor, Main Hospital North Services provided 24/7 365 days per year General Radiology (X-ray) Fluoroscopy Ultrasound (Sonography) Nuclear Medicine P.E.T. imaging Computed Tomography (CT scan) Magnetic

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

Digital Mammography Monitor

Digital Mammography Monitor Film Imaging to Digital Imaging for Mammography Diagnosis As the incidence of breast cancer rises, regular mammograms are recommended for early detection. Selecting the Optimum It is vital in the process

More information

Prepared by: Paul Lee ON Semiconductor http://onsemi.com

Prepared by: Paul Lee ON Semiconductor http://onsemi.com Introduction to Analog Video Prepared by: Paul Lee ON Semiconductor APPLICATION NOTE Introduction Eventually all video signals being broadcasted or transmitted will be digital, but until then analog video

More information

Health Care Careers in the Field of Imaging. Shari Workman, MSM,PHR,CIR MultiCare Health System Senior Recruiter/Employment Specialist

Health Care Careers in the Field of Imaging. Shari Workman, MSM,PHR,CIR MultiCare Health System Senior Recruiter/Employment Specialist Health Care Careers in the Field of Imaging Shari Workman, MSM,PHR,CIR MultiCare Health System Senior Recruiter/Employment Specialist What is Health Care Imaging? Technologists working with patients, using

More information

ICC Recommendations for Color Measurement

ICC Recommendations for Color Measurement White Paper #3 Level: Introductory Date: Dec 2004 ICC Recommendations for Color Measurement Introduction In order to prepare a useful device profile based on the International Color Consortium s Specification,

More information

Monochrome Print Features: True 600 dpi print resolution Produces 3,3 A0 size prints or copies per minute 100% efficient No waste toner

Monochrome Print Features: True 600 dpi print resolution Produces 3,3 A0 size prints or copies per minute 100% efficient No waste toner KIP 3100 SERIES KIP 3100 Monochrome Print, Copy & Scan The KIP 3100 system accurately reproduces technical documents at true 600 x 600 dpi resolution. Prints and copies may be delivered to the integrated

More information

Digital Image Basics. Introduction. Pixels and Bitmaps. Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color

Digital Image Basics. Introduction. Pixels and Bitmaps. Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction When using digital equipment to capture, store, modify and view photographic images, they must first be converted to a set

More information

Guidelines On Teleradiology In Malaysia

Guidelines On Teleradiology In Malaysia 1 The following are Teleradiology Standards prepared by our College (MRS) and submitted to the Ministry of Health. The Ministry of Health has not endorsed these standards yet. However, we hope they will

More information

22 (21.5 viewable), 16:9 FHD LED monitor with SuperClear VA technology and HDMI input. VA2265Sh

22 (21.5 viewable), 16:9 FHD LED monitor with SuperClear VA technology and HDMI input. VA2265Sh 22 (21.5 viewable), 16:9 FHD LED monitor with SuperClear VA technology and HDMI input VA2265Sh The ViewSonic VA2265Sh is a 22 (21.5 viewable) widescreen monitor with an LED backlight and glossy, thinbezel

More information

Digital Camera Imaging Evaluation

Digital Camera Imaging Evaluation Digital Camera Imaging Evaluation Presenter/Author J Mazzetta, Electro Optical Industries Coauthors Dennis Caudle, Electro Optical Industries Bob Wageneck, Electro Optical Industries Contact Information

More information

WHICH HD ENDOSCOPIC CAMERA?

WHICH HD ENDOSCOPIC CAMERA? WHICH HD ENDOSCOPIC CAMERA? Colin Dobbyne ORTV 2 Hillcrest Park, Hoyle Road Calverton, Nottingham NG14 6QJ t. +44 (0)115 965 55 77 f: +44 (0)115 965 55 22 e: sales@or-tv.net Welcome Thank you for downloading

More information

A Comprehensive Set of Image Quality Metrics

A Comprehensive Set of Image Quality Metrics The Gold Standard of image quality specification and verification A Comprehensive Set of Image Quality Metrics GoldenThread is the product of years of research and development conducted for the Federal

More information