CM4106 Separation Methods Gas Chromatography: Applications. Hyphenated Techniques
|
|
- Lynn York
- 7 years ago
- Views:
Transcription
1 CM4106 Separation Methods Gas Chromatography: Applications. Hyphenated Techniques Dr. Amalia Muñoz Fundación CEAM. Euphore Laboratories
2 Gas Chromatography Qualitative Analysis The characteristic parameter for the identification of a compound is the RETENTION TIME (RT) or the CORRECTED RETENTION TIME (RT ) RT and RT depend on: Temperature Stationary phase (column) Carrier gas Flow Etc. Time elapsed between the dead point and the peak maximum Time elapsed between the injection point and the peak maximum. Each solute has a characteristic retention time. Small fluctuations could derived in wrong identification It is not correct absolute RT data from literature with those obtained experimentally.
3 Gas Chromatography Qualitative Analysis Relative Retention times It is the ratio between the corrected the retention time of the analyte to identify (RT a ) and a substance used as a reference (RT ref ) RT = r RT RT ' a ' ref To obtain the RT r the reference compound must be added to the sample and to the standards The RT of the reference compound should be closed to the analytes.
4 Gas Chromatography Qualitative Analysis Other method The sample is separated into two aliquots. The first aliquot is injected directly A determined amount of a standard is added into the second aliquot and then injected. If we see a new peak in the second chromatogram, we can ensure that this substance is not in the sample while a signal of some of the peaks appear more intense in the second chromatogram that substance exists in the sample
5 Gas Chromatography Qualitative Analysis It could be simpler Using selective detectors: ECD, NPD, PID, etc. Using hyphenated techniques: GC-MS, GC-FTIR, GC-NMR Using two or more detectors
6 Gas Chromatography Qualitative Analysis Two Detectors A single injection on two GC detectors yields two sets of data in one half of the time Detector A Detector A Injector Injector Detector B Detector B Parallel Dual Detector Configuration Post Column Split Configuration Inyector Detector A Detector B Series Configuration (Detector A must be non-destructive)
7 Gas Chromatography Quantitative Analysis 3 important stages in a GC analysis, 1. The preparation of the sample. 2. The development of the separation and the production of the chromatogram 3. The processing of the data and the presentation of the results. Each stage is equally important and if not carried out correctly the results will be neither precise nor accurate. Sample preparation can be Simple: involving no more that diluting a known weight of sample with mobile phase Much more complex: including an extraction procedure followed by derivatization and then dilution. For some samples the preparation can be the most time consuming and difficult part of the whole analysis.
8 Gas Chromatography Quantitative Analysis Chromatogram The response must be linear Concentration Mass The response factor of each compound is different for each compound Parameters that can be used: Peak Height Peak Area
9 Gas Chromatography Quantitative Analysis Chromatogram Area Normalization PeakAreag % g = 100 areas The sum of the areas of all the peaks corresponds to 100% of the solutes separated. Only true if: All the compounds are eluted Same sensitivity % PeakAreag f g g = 100 ( area f ) i i As the compounds usually do not have the same sensitivity a correction factor should be applied Calibration curve f g = mass area
10 Gas Chromatography Quantitative Analysis Chromatogram Internal Standard An internal standard is a compound, not present in the sample, that is added in a constant amount to samples and calibration standards. The peak of compound must not overlap with the peaks of the analytes. 6 SI Method Area compound/ area SI y = x R 2 = y = 0.497x R 2 = mass compound/ mass SI Advantages: manual injection Disadvantages: To analyse great number of analytes To find a good IS compound A compound B Lineal (compound A) Lineal (compound B)
11 Gas Chromatography Quantitative Analysis Chromatogram External Standard Area compound y = x R 2 = ES Method y = x R 2 = mass compound compound A compound B Lineal (compound A) Lineal (compound B) Advantages: simpler than IS. Disadvantages: Sample injection reproducibility Preferable Automatic injection or sample valve
12 Gas Chromatography DERIVATIZATION Derivatization is the process of chemically modifying a compound to produce a new compound which has properties that are suitable for analysis using a GC WHY? To permit analysis of compounds not directly amenable to analysis due to, for example, inadequate volatility or stability Improve chromatographic behavior or detectability. Derivatization is a useful tool allowing the use of GC and GC/MS to be done on samples that would otherwise not be possible in various areas of chemistry such as medical, forensic, and environmental
13 Gas Chromatography DERIVATIZATION Increases volatility (i.e. sugars): Eliminates the presence of polar OH, NH, & SH groups Derivatization targets O,S, N and P functional groups (with hydrogens available Increases detectability, I.e. steroids/ cholesterol Increases stability Enhances sensitivity for ECD (Electron Capture Detection). The introduction of ECD detectable groups, such as halogenated acyl groups, allows detection of previously undetectable compounds in some cases: derivatization can also be used to decrease volatility to allow analysis of very low molecular weight compounds, to minimize losses in manipulation and to help separate sample peaks from solvent peak.
14 Gas Chromatography DERIVATIZATION Comments Advantages Disadvantages Silylation Readily volitizes the sample - Wide variety of compounds -Large number of silylating reagents available -Easily prepared -Moisture sensitive -Organic solvents must be aprotic (no protons available) -WAX type columns cannot be used Acylation -Used as the first step to further derivatizations or as a method of protection of certain active hydrogens. -Reduces the polarity of amino, hydroxyl, and thiol groups and adds halogenated functionalities. -Increased detectability by ECD -Derivatives are hydrolytically stable - Increased sensitivity by adding molecular weight -Acylation can be used as a first step to activate carboxylic acids prior to esterfication (alkylation) -Difficult to prepare. - Reaction products often need to be removed before analysis - Moisture sensitive - Reagents are hazardous and odorous Alkylation -Reduces molecular polarity by replacing active hydrogens with an alkyl group. - modify compounds with acidic hydrogens, such as carboxylic acids and phenols. -Reagents containing fluorinated benzoyl groups can be used for ECD -Wide range of alkylation reagents. -Reaction conditions can vary from strongly acidic to strongly basic - Some reactions can be done in aqueous solutions - Alkylation derivatives are generally stable -Limited to amines and acidic hydroxyls - Reaction conditions are frequently severe - Reagents are often toxic Formation of perfluoroderivatives Reagents containing fluorinated benzoyl groups can be used for ECD Wide range of application Easy to prepare Selectivity GC-Quiral Derivatiz.
15 Gas Chromatography DERIVATIZATION Some examples Derivatization Reaction Common Derivatizing Agent Methylation of carboxylic acids Oxime formation of carbonyl functionality Diazomethane, methanol/sulfuric acid PFBHA N-hexyl carbonate, carbamate, and ester formation from hydroxylic, aminic, and carboxylic functionality N-hexyl chloroformate Heptafluorobutyramide formation from aromatic amines Heptafluorobutyramide And much more
16 On-line combination of a chromatographic separation technique with a sensitive and Element-specific detector Chromatographic Techniques Spectroscopic Techniques Separation of analytes Necessity of pure analytes Low security in identification High identification level Hyphenated techniques provide the analyst with structural information on the components present in complex mixtures. This information may be sufficient to identify components
17 Chromatograph Column Interface Spectrometer Column Chromatograph Non-Destructive Detector Interface Spectrometer Chromatograph Detector Column Interface Non-Destructive Spectrometer Chromatograph Column Interface 1 Interface 2 Non-Destructive Spectrometer Spectrometer Chromatograph Column Interface 1 Interface 2 Spectrometer Spectrometer
18 Common hyphenated techniques Gas Chromatography GC-MS GC-FTIR-MS GC-FTIR GC-AAS Mass Spectrometry Infrared Spectroscopy Emission and Absorption Atomic Spectroscopy Nuclear Resonance Spectrometry LC-MS LC-FTIR LC-ICP Liquid Chromatography LC-NMR
19 GC/MS Provide Information about the chemical composition of the analyte Molecular analysis Isotopic analysis Trace analysis Elemental analysis Surface analysis Mass Spectrometer High Detection efficiency and specificity of molecular recognition Molecules are ionized (broken down) into electrically charged particles called ions with a specific mass and charge. Due to that, the speed and direction of the ion may be changed with an electric or magnetic field.
20 GC/MS
21 GC/MS Mass Spectrometer Components GC Column/MS Interface Ionizations Source: for example Electron Impact (EI) or Chemical Ionization (CI) Mass Analyzer: for example Magnetic Sector, Quadrupole, Ion Trap, Time of Flight Mass Detector Software/Data display
22 GC/MS Mass Spectrometer Ionization methods Should provide a high ionization efficiency and high stability with minimum kinetic energy distribution and minimum angular dispersion of the ion produced The ion source should produce: Intact molecular ions MW (for MW information) Fragment ions (for structural information) Control over the internal energy transferred to the molecule for control over the degree of fragmentation Should be possible to couple with various types of chromatographs
23 GC/MS Mass Spectrometer Ionization methods Electron Impact, 70 ev Analytes elute here A + B + C + ABC
24 Ionization method Electron Impact (EI) Chemical Ionization (CI) Type of analysis Molecular Molecular Ionization agent Electrons ( 70 ev) Gaseous ions. (Reagent gas: CH 4, NH 3, NF 3, N 2 O) GC/MS Mass Spectrometer Pressure Characteristics Application 10-5 Torr 1 Torr Reproducible spectra, extensive fragmentation Molecular ions and controllable fragmentation VOC; structural elucidation VOC; MW determination Desorption Ionization (DI) Molecular Energetic particles, photons Torr Intact molecular ions from high-mass compounds Condensed phase, high-mass compounds; MW and structure determination Spray Ionization (SI) Molecular Electrical, thermal, and pneumatic energy Torr Intact molecular ions from high-mass compounds Solutions of high-mw compounds; used with LC/MS to determine MW s and structures Glow- Discharge Ionization (GD) Elemental Plasma Torr Very stable, reproducible spectra Elemental analysis of solids Inductively Coupled Plasma (ICP) Elemental Plasma Atmosp. pressure High ionization efficiency Elemental analysis of solutions
25 GC/MS Mass Spectrometer Mass Analyzer Function: to measure the mass-tocharge ratios of ions (m/z), thus providing a mesa to identify them. Depend on the interactions of charged particles with electric or magnetic fields
26 GC/MS Mass Spectrometer Mass Analyzer Quadrupole: Orthogonal DC and RF planes (x/y) are used to separate ion passing through a highly evacuated tube. As an analyte mass fragment pass trough the mass analyzer, the matched x/y field at that instant a very small part of one scan- allows only specific fragment exit into the mass detector. The duration of an entire scan is usually less than 1 second (e.g. 0.6 sec). The m/z range can be adjusted by the analyst MS scanning process occurs very quickly, so each scan includes a measure of all the amounts of each different mass fragment during that scan
27 GC/MS Mass Spectrometer Mass Analyzer Ion trap It is an three-dimensional analog of the quadrupole mass filter. Because forces operate in 3 directions, the electric fields can be used store ions in an electrical bottle. It consist of two end-cap electrodes of hyperbolic cross section that normally are operated at ground potential. A rotationally symmetric electric quadrupole field is generated At a given voltage, ions of a specific mass range are held oscillating in the trap. Initially, the electron beam is used to produce ions and after a given time the beam is turned off. All the ions, (except those selected by the magnitude of the applied rf voltage) are lost to the walls of the trap, and the remainder, continue oscillating within the trap. The potential of the applied rf voltage is then increased, and the ions sequentially assume unstable trajectories and leave the trap via the aperture to the sensor. The ions exit the trap in order of their increasing m/z values.
28 GC/MS Mass Spectrometer Mass Analyzer Time-of-Flight: A beam of ions is accelerated through a known potential V, and the time taken to reach a detector at a distance d, in a linear fligh tube, is measured. If all ions fall through the same potential, V, their velocities must be inversely proportional to the square roots of their masses. The source must be pulsed in order to avoid simultaneous arrivals of ions of different mass-to-charge ratios.
29 GC/MS Mass Spectrometer Method Quantity measured Mass/Charg e range (Da/charge) Resolution at 1000 (Da/charge) (mass peak witdh) Mass Measurement Accuracy at 1000 Da/charge Dynamic range (number of order of magnitude of concentration over which response varies linearly) Operating Pressure Sector Magnet Momentum/ charge < 5 ppm Time of flight Flight time % Quadrupole ion trap frequency % Quadrupole Filters for m/z % Cyclotron Resonance frequency < 10 ppm
30 GC/MS Mass Spectrometer Mass Detector and data generation One scan includes all the mass detector current at each m/z ratio allowed to exit the mass analyzer. A chromatographic peak ( around10 seconds wide) can be scanned 16 or 17 times. Each scan generates an individual mass spectrum which is saved in the computer s hard drive
31 Coupling GC/MS Direct Introduction Interfaces for GC/MS Jet separator Permselective membrane Molecular effusion Direct coupling Interfaces for LC/MS Total solvent elimination before ionization Partial Solvent elimination before ionization Solvent introduction
32 Coupling GC/MS JET SEPARATOR The GC flow is introduced into an evacuated chamber through a restricted capillary. At the capillary tip a supersonic expanding jet of analyte and carrier molecules is formed and its core area sampled into the mass spectrometer. In an expanding jet, high molecular mass compounds are concentrated in the core flow whereas the lighter and more diffusive carrier molecules are dispersed away, in part through collisions. Thus, sampling of the core flow produces an enrichment of the analyte. J. Abian. JOURNAL OF MASS SPECTROMETRY J. Mass Spectrom. 34, (1999) The jet interface is very versatile, inert and efficient, despite disadvantages of reduced efficiency with more volatile compounds and potential plugging problems at the capillary restrictor.
33 PERMSELECTIVE MEMBRANE Coupling GC/MS It is made of a silicone-rubber membrane that transmits organic non-polar molecules and acts as a barrier for (non-organic) carrier gases. Despite being a very effective enrichment procedure, it also suffers from discrimination effects with more polar analytes and produces significant band broadening of their chromatographic peaks
34 Coupling GC/MS MOLECULAR EFFUSION It is based on the molecular filtering of the gas effluent by means of a porous glass frit. The column effluent passes through a fritted tube situated in a vacuum chamber. Small molecules traverse the microscopic pores in the tube walls and are evacuated whereas high molecular mass molecules are transferred to the ion source. Drawbacks : High dead volume added and high surface area. Shows discrimination effects in the case of smaller molecules
35 Coupling GC/MS DIRECT INTRODUCTION Capillary columns use optimum flow-rates of gas of about 1-2 ml min -1, instead of more than ml min -1 used with packed columns, allowing all the effluent to be directed to the mass spectrometer. This is usually done through a direct coupling where the column exit is introduced into the ion source without a capillary restriction. Extensively used in analytical laboratories. Enrichment interfaces have become unnecessary for GC/MS
36 Coupling LC/MS DIRECT LIQUID INTRODUCTION (DLI) Use of the solvent as the reagent gas in a CI source A major problem in the introduction of liquids through a capillary is that the high vacuum in the ion source produces rapid evaporation of the liquid inside the capillary, eventually leading to flow stoppage through freezing of the solvent. The use of restricted capillaries and the heating of the capillary in part solved this problem Maximum flow-rates accepted by DLI interfaces are in the range μl min -1 and are best suited for micro- and nanobore chromatography (<1 mm i.d.column) Obsolete
37 Coupling LC/MS THERMOSPRAY In the presence of a volatile buffer in the solvent, ions are obtained without any other ionizing source The liquid flowing through the hot capillary is partially evaporated so that an ultrasonic spray of vapour and charged microdroplets is obtained at the probe exit. Ions present in the source are transferred into the mass spectrometer through the ion cone aperture while the main portion of the residual vapours is captured by the high-conductance vacuum line and purged by a rotary pump. Charged microdroplets in the spray are the result of the rapid breakdown of the liquid surface during vaporization and the statistical distribution of electrolyte ions in the droplets. Hence, an equal mixture of positive and negatively charged droplets is formed from a neutral solution. Gas-phase ions are produced in the source from these microdroplets as a result of several processes including ion desolvation and ion evaporation from the charged microdroplets, as well as gas-phase CI processes TO VACUUM
38 Coupling LC/MS THERMOSPRAY The evaporation of neutral molecules from the initial charged droplet produces a reduction in droplet size and a charge density increment. This excess of electrical energy can produce either fragmentation of the droplets into smaller droplets or desorption of ions into the gas phase by ion evaporation. Alternatively, gas-phase ions can be produced by simple desolvation of liquid-phase ions. These processes produce a plasma of reagent ions mainly derived from the ammonium acetate buffer in the solution. If the proton affinity is relatively high, the analyte will keep its charge or will take it from the reagent plasma. If the analyte proton affinity is low relative to other reagent plasma molecules, it will not be ionized or ionmolecules adducts will be observed. TSP ionization causes simple soft ionization spectra containing mostly molecular information TO VACUUM
39 Coupling LC/MS Atmospheric Pressure ionization Ionization of the column effluent is carried out at atmospheric pressure by any of several procedures including a radioactive source, electrical discharges and high voltage electric fields. The ions produced are continuously sampled through a small aperture and pass into the spectrometer where they are mass analysed. Introduce only a small amount of solvent into the low-pressure region of the MS. Two main commercial source types: Atmospheric Pressure Chemical Ionization (APCI) Electrospray (ESI) source
40 Coupling LC/MS APCI An atmospheric pressure vaporization chamber in which 1-2 μl of a liquid sample is injected through a septum. A hot carrier gas (N 2 ) is introduced into the chamber to help vaporization and to transport the analyte to an area close to a radioactive beta source ( 63 Ni). In this area gas ionization occurred, producing an atmospheric pressure reagent plasma that gave rise to analyte ions through ion- molecule reactions. Ions a then transferred to the mass spectrometer through a small aperture and mass analysed Analysis of medium- and low-polarity compounds or when relatively non-polar solvents have to be used.
41 Coupling LC/MS ESI The liquid sample is introduced into a high potential chamber through the capillary and at the exit an electrically induced spray of charged microdroplets is produced. Ions in these droplets enter the gas phase through evaporative processes (similar to TSP). The ions in the spray are captured through a glass capillary restrictor where they are conducted into the low vacuum area of the mass spectrometer. To promote droplet desolvation, the ionization chamber is continuously supplied with a countercurrent flow of dry nitrogen.
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationElectrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
More informationAMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
More information13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
More informationHow To Use Gc-Ms
The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter
More informationSeparation of Amino Acids by Paper Chromatography
Separation of Amino Acids by Paper Chromatography Chromatography is a common technique for separating chemical substances. The prefix chroma, which suggests color, comes from the fact that some of the
More informationThe photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
More informationGas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
More informationAnalyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
More informationDiffusion and Fluid Flow
Diffusion and Fluid Flow What determines the diffusion coefficient? What determines fluid flow? 1. Diffusion: Diffusion refers to the transport of substance against a concentration gradient. ΔS>0 Mass
More informationOverview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
More informationInfrared Spectroscopy 紅 外 線 光 譜 儀
Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed
More informationChapter 13 Spectroscopy NMR, IR, MS, UV-Vis
Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis Main points of the chapter 1. Hydrogen Nuclear Magnetic Resonance a. Splitting or coupling (what s next to what) b. Chemical shifts (what type is it) c. Integration
More informationThe Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationTHERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles
THERMAL DESORPTION Introduction and Principles Thermal Desorption Thermal desorption is a simple extension of the technique of gas chromatography (GC) It involves the use of heat and a flow of inert gas
More informationOrganic Chemistry Tenth Edition
Organic Chemistry Tenth Edition T. W. Graham Solomons Craig B. Fryhle Welcome to CHM 22 Organic Chemisty II Chapters 2 (IR), 9, 3-20. Chapter 2 and Chapter 9 Spectroscopy (interaction of molecule with
More informationUHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
More informationCHE334 Identification of an Unknown Compound By NMR/IR/MS
CHE334 Identification of an Unknown Compound By NMR/IR/MS Purpose The object of this experiment is to determine the structure of an unknown compound using IR, 1 H-NMR, 13 C-NMR and Mass spectroscopy. Infrared
More informationLecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder
Lecture Chromo-3: Gas Chromatography CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder Outline Introduction Instrument overview Carrier gas Sample
More informationCHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING
CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING Essential Standard: STUDENTS WILL UNDERSTAND THAT THE PROPERTIES OF MATTER AND THEIR INTERACTIONS ARE A CONSEQUENCE OF THE STRUCTURE OF MATTER,
More informationDetermining the Structure of an Organic Compound
Determining the Structure of an Organic Compound The analysis of the outcome of a reaction requires that we know the full structure of the products as well as the reactants In the 19 th and early 20 th
More informationBackground Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
More informationGENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
More informationCoating Technology: Evaporation Vs Sputtering
Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information
More informationMass Spectrometry. Overview
Mass Spectrometry Overview Mass Spectrometry is an analytic technique that utilizes the degree of deflection of charged particles by a magnetic field to find the relative masses of molecular ions and fragments.2
More informationModern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
More informationAccurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis
Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis B. Apicella*, M. Passaro**, X. Wang***, N. Spinelli**** mariadellarcopassaro@gmail.com
More informationChapter 21 Principles of Chromatography and Mass Spectrometry Problems 1,2,3,4,8,9,11,13
Chapter 21 Principles of Chromatography and Mass Spectrometry Problems 1,2,3,4,8,9,11,13 21-1 What is Chromatography Process to separate and purify compounds in a mixture by passing them trough a material
More informationHow To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
More informationChapter 28: High-Performance Liquid Chromatography (HPLC)
Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion
More information0 10 20 30 40 50 60 70 m/z
Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationSpectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
More informationPurification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
More informationFractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
More informationGas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
More information1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
More information# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
More informationName Class Date. In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question.
Assessment Chapter Test A Chapter: States of Matter In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1. The kinetic-molecular
More informationProton Nuclear Magnetic Resonance Spectroscopy
Proton Nuclear Magnetic Resonance Spectroscopy Introduction: The NMR Spectrum serves as a great resource in determining the structure of an organic compound by revealing the hydrogen and carbon skeleton.
More informationVacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
More informationMass spectrometry. What are the principles behind MS? What do all MS instruments have in common?
Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction
More informationIonization Methods in Organic Mass Spectrometry Contents:
Ionization Methods in Organic Mass Spectrometry Contents:. Introduction. Gas-Phase Ionization. Electron Ionization (EI). Chemical Ionization (CI). Desorption Chemical Ionization (DCI). Negative-ion chemical
More informationFor example: (Example is from page 50 of the Thinkbook)
SOLVING COMBINED SPECTROSCOPY PROBLEMS: Lecture Supplement: page 50-53 in Thinkbook CFQ s and PP s: page 216 241 in Thinkbook Introduction: The structure of an unknown molecule can be determined using
More informationNuclear Magnetic Resonance Spectroscopy
Nuclear Magnetic Resonance Spectroscopy Introduction NMR is the most powerful tool available for organic structure determination. It is used to study a wide variety of nuclei: 1 H 13 C 15 N 19 F 31 P 2
More informationEleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
More informationNuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass
Protons, neutrons and electrons Nuclear Structure particle relative charge relative mass proton 1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass Protons and neutrons make up
More informationThe Use of Hydrogen Carrier Gas for GC/MS
Technical Note The Use of Hydrogen Carrier Gas for GC/MS Gas Chromatography/ Mass Spectrometry Highlights Guidelines that can mitigate dangers and leverage benefits of using hydrogen Key safety factors
More informationNMR and other Instrumental Techniques in Chemistry and the proposed National Curriculum.
NMR and other Instrumental Techniques in Chemistry and the proposed National Curriculum. Dr. John Jackowski Chair of Science, Head of Chemistry Scotch College Melbourne john.jackowski@scotch.vic.edu.au
More informationGuide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
More informationTABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ACKNOWLEDGEMENT ABSTRACT LIST OF TABLES LIST OF FIGURES LIST OF SYMBOLS LIST OF APPENDICES
vii TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ii ACKNOWLEDGEMENT iv ABSTRACT v TABLE OF CONTENTS vii LIST OF TABLES xii LIST OF FIGURES xiv LIST OF SYMBOLS xvii LIST OF APPENDICES xviii 1 INTRODUCTION
More informationelectron does not become part of the compound; one electron goes in but two electrons come out.
Characterization Techniques for Organic Compounds. When we run a reaction in the laboratory or when we isolate a compound from nature, one of our first tasks is to identify the compound that we have obtained.
More informationINFRARED SPECTROSCOPY (IR)
INFRARED SPECTROSCOPY (IR) Theory and Interpretation of IR spectra ASSIGNED READINGS Introduction to technique 25 (p. 833-834 in lab textbook) Uses of the Infrared Spectrum (p. 847-853) Look over pages
More informationFerda ORHUN Mineral Research and Exploration Institute of Turkey
GAS CHROMATOGRAPHY, EXPERIMENTAL STUDIES ON THE FLOW RATE EFFECTS OF CARRIER GAS AND APPLICATION OF THE METHOD TO HYDROCARBON ANALYSIS OF SOME NATURAL GASES IN TURKEY Ferda ORHUN Mineral Research and Exploration
More informationSimultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
More informationChapter 5 Classification of Organic Compounds by Solubility
Chapter 5 Classification of Organic Compounds by Solubility Deductions based upon interpretation of simple solubility tests can be extremely useful in organic structure determination. Both solubility and
More informationIntelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
More informationCONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
More informationExpectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
More informationFast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
More informationGetting the most from this book...4 About this book...5
Contents Getting the most from this book...4 About this book....5 Content Guidance Topic 1 Atomic structure and the periodic table...8 Topic 2 Bonding and structure...14 Topic 2A Bonding....14 Topic 2B
More informationLC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
More informationAnalysis of Vegetable Oils by High Performance Liquid Chromatography
Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia
More informationQuantitative Evaluation of Perfume Fleuressence Samples using the znose
Quantitative Evaluation of Perfume Fleuressence Samples using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses Conventional electronic noses (enoses) produce a recognizable response
More information23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
More informationChemistry 13: States of Matter
Chemistry 13: States of Matter Name: Period: Date: Chemistry Content Standard: Gases and Their Properties The kinetic molecular theory describes the motion of atoms and molecules and explains the properties
More informationGraphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
More informationGUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005
CAC/GL 56-2005 Page 1 of 6 GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005 CONFIRMATORY TESTS When analyses are
More informationAppendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
More informationIndiana's Academic Standards 2010 ICP Indiana's Academic Standards 2016 ICP. map) that describe the relationship acceleration, velocity and distance.
.1.1 Measure the motion of objects to understand.1.1 Develop graphical, the relationships among distance, velocity and mathematical, and pictorial acceleration. Develop deeper understanding through representations
More informationHow to Quickly Solve Spectrometry Problems
How to Quickly Solve Spectrometry Problems You should be looking for: Mass Spectrometry (MS) Chemical Formula DBE Infrared Spectroscopy (IR) Important Functional Groups o Alcohol O-H o Carboxylic Acid
More informationIntroduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
More informationNuclear Magnetic Resonance notes
Reminder: These notes are meant to supplement, not replace, the laboratory manual. Nuclear Magnetic Resonance notes Nuclear Magnetic Resonance (NMR) is a spectrometric technique which provides information
More informationNew drugs of abuse? A case study on Phenazepam & Methoxetamine
New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied
More informationChapter Test B. Chapter: Measurements and Calculations
Assessment Chapter Test B Chapter: Measurements and Calculations PART I In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1.
More informationUsing Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
More informationElectrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)
Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III) Purpose: Verify the synthesis of the synthesis product, Cr(acac) 3. The presence of any urea containing byproducts will also be determined.
More informationTheory and Instrumentation of GC Introduction
Theory and Instrumentation of GC Introduction i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference
More informationAn example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry
Limits to Detection/Measurement Ionization is fundamentally a Probabilistic Process Just like radioactive decay So is transmission through the analyzer There is an intrinsic statistical uncertainty Proportional
More informationSolving Spectroscopy Problems
Solving Spectroscopy Problems The following is a detailed summary on how to solve spectroscopy problems, key terms are highlighted in bold and the definitions are from the illustrated glossary on Dr. Hardinger
More informationStrategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
More informationAdvantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites
Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Carmen T. Santasania and David S. Bell SUPELCO, 595 North Harrison Road, Bellefonte, PA 16823 T404045 GUS Abstract
More informationNMR SPECTROSCOPY A N I N T R O D U C T I O N T O... Self-study booklet NUCLEAR MAGNETIC RESONANCE. 4 3 2 1 0 δ PUBLISHING
A N I N T R O D U T I O N T O... NMR SPETROSOPY NULEAR MAGNETI RESONANE 4 3 1 0 δ Self-study booklet PUBLISING NMR Spectroscopy NULEAR MAGNETI RESONANE SPETROSOPY Origin of Spectra Theory All nuclei possess
More informationOrganic Spectroscopy. UV - Ultraviolet-Visible Spectroscopy. !! 200-800 nm. Methods for structure determination of organic compounds:
Organic Spectroscopy Methods for structure determination of organic compounds: X-ray rystallography rystall structures Mass spectroscopy Molecular formula -----------------------------------------------------------------------------
More informationSelective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
More informationContents. Contributors. 1 Quality control and regulation 1 C.J. MOORES
Contents Contributors Preface xiii xiv 1 Quality control and regulation 1 C.J. MOORES 1.1 Introduction 1 1.2 The quality of medicines 2 1.2.1 The meaning of quality 2 1.2.2 Medicines are special 3 1.2.3
More informationORGANIC SAMPLE PREPARATION
ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,
More informationPaper Chromatography: Separation and Identification of Five Metal Cations
Paper Chromatography: Separation and Identification of Five Metal Cations Objectives Known and unknown solutions of the metal ions Ag +, Fe 3+, Co 2+, Cu 2+ and Hg 2+ will be analyzed using paper chromatography.
More information5. Which temperature is equal to +20 K? 1) 253ºC 2) 293ºC 3) 253 C 4) 293 C
1. The average kinetic energy of water molecules increases when 1) H 2 O(s) changes to H 2 O( ) at 0ºC 3) H 2 O( ) at 10ºC changes to H 2 O( ) at 20ºC 2) H 2 O( ) changes to H 2 O(s) at 0ºC 4) H 2 O( )
More informationThe Theory of HPLC. Gradient HPLC
The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims
More informationSymmetric Stretch: allows molecule to move through space
BACKGROUND INFORMATION Infrared Spectroscopy Before introducing the subject of IR spectroscopy, we must first review some aspects of the electromagnetic spectrum. The electromagnetic spectrum is composed
More informationwhere h = 6.62 10-34 J s
Electromagnetic Spectrum: Refer to Figure 12.1 Molecular Spectroscopy: Absorption of electromagnetic radiation: The absorptions and emissions of electromagnetic radiation are related molecular-level phenomena
More informationAromatic plants: from extraction to analysis
Aromatic plants: from extraction to analysis Thierry TALOU Agro-industrial Chemistry Laboratory National Polytechnic Institute of Toulouse ENSIACET 4 allée Emile Monso 31432 Toulouse Olfactory system Olfactory
More informationThe Four Questions to Ask While Interpreting Spectra. 1. How many different environments are there?
1 H NMR Spectroscopy (#1c) The technique of 1 H NMR spectroscopy is central to organic chemistry and other fields involving analysis of organic chemicals, such as forensics and environmental science. It
More informationStates of Matter CHAPTER 10 REVIEW SECTION 1. Name Date Class. Answer the following questions in the space provided.
CHAPTER 10 REVIEW States of Matter SECTION 1 SHORT ANSWER Answer the following questions in the space provided. 1. Identify whether the descriptions below describe an ideal gas or a real gas. ideal gas
More informationAir Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike License. Your use of this material constitutes acceptance of that license and the conditions of use of materials on this
More informationDiscovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
More informationSimultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
More information