Ultrasound. - Dosimetry. Gail ter Haar. Joint Physics Department, Royal Marsden Hospital: Institute of Cancer Research, Sutton, Surrey UK
|
|
- Spencer Terry
- 7 years ago
- Views:
Transcription
1 Ultrasound - Dosimetry Gail ter Haar Joint Physics Department, Royal Marsden Hospital: Institute of Cancer Research, Sutton, Surrey UK
2 Measurable in water Exposure Reduced by: tissue acoustic properties environmental factors tissue geometry Dose Tissue target Difficult to measure
3 Exposure Can be measured in water Can be described by I, p, t, f..
4 DOSE
5 Dose A parameter that : 1. Describes the sound field in tissue 2. Tells us about how well we are ablating tissue 3.Allow treatments to be reproduced either on this machine or on another
6 Dose Actual intensity values in tissue depend on: Depth into tissue, tissue type, frequency, non-linearity
7 What is important Clinically?
8 What is important Clinically? Volume of tissue damaged in a given time
9 What is important in Physics terms?
10 What is important to the Physicist?? Volume of tissue damaged Relative to measurable field parameters
11 Potential Dosimetric parameters: 1. I x t 2. Thermal dose
12 Ixt Intensity x time is a measure of total energy useful for purely thermal lesions
13 Calibration Gail ter Haar
14 Oh No!
15 What is calibration? : Quantitative & Qualitative characterisation of the acoustic field
16 Why calibrate? : Clinical Safety: Select treatment output Limit exposure of sensitive sites Clinical efficacy comparison: between patients (repeatability) between different equipment types
17 What do we need to know? Source: Acoustic: Frequency Diameter Focal length Intensity / Power / P - Exposure: Duration Scan speed
18 Beam profiles (calculated): 0cm Focal depth or 15cm from transducer Intensity W/cm^ Radial Distance (cm) Focus
19 Beam profiles (calculated) : 2.0cm deep Intensity (W/cm^2) cm deep Radial Distance (cm)
20 Pressure We measure : Acoustic power We calculate : Intensity There is no way of measuring intensity directly
21 Intensity : I = p 2 /2ρc pressure density speed of sound
22 Describing the beam profile :
23 Intensity Intensity = energy flux (ie energy crossing a unit area in unit time). Intensity will vary in space and time
24 I o Which intensity? : WATER WATER I free field = I o e -αx x WATER TISSUE I in situ = I o e -αx I o I I free field > in situ
25 Intensity is usually calculated either from: acoustic power & beam dimensions or from pressure measurement Isptp IEC defined quantities (originally for diagnostic fields) are based on hydrophone measurements, but lab practice is often different Ispta Isata IEC 61102,61157, 61689
26 Which intensity? : I in situ I SP I SATA I free field
27 Which intensity? : I SAL Beam profile measured at low pressure, power measured at high pressure Accounts for non-linear propagation
28 Intensity : I SAL = I SP Intensity, spatial average measured under linear conditions I SP = 1.56 W/D 2 = 1.8 I SAL Acoustic power Beam width (linear)
29 Which intensity? : Is I SAL or I sp the correct quantity to use? HEAT
30 Or Or should should we we be be using using pressure? pressure? Which intensity? : Is I SAL or I sp the correct quantity to use? HEAT CAVITATION
31 Measurement problems : Highly focused fields Focal region can be <(10 x 1) mm Rapid spatial variation High Intensity/pressure Non linear propagation (water) High local pressures (can damage measurement devices) Cavitation, thermal damage
32 Transducers Frequencies from 0.5 to 4.0 MHz Spatial peak intensity up to 20kWcm -2 Focal lengths from 3 to 15 cm Phased arrays multiple simultaneous foci
33 What do we measure? : Free Field parameters
34 Free field conditions Acoustic absorber lines walls Courtesy of Dr. Coussios
35 Acoustical absorbers Polyurethane based absorbing rubbers; Two types of material produced: - High quality absorber suitable for use as radiation force balance targets; - Low cost single-ply layer for use as tank wall lining. Courtesy of NPL
36 What do we measure? : Acoustic Pressure Acoustic Beam profile & Acoustic radiation force
37 Pressure & beam profiles
38 Acoustic pressure : Measured using a hydrophone The gold standard GEC Marconi 0.5 mm diameter PVDF Membrane Low intensity 100 Wcm cycles, 1 khz
39 Membrane hydrophone with pre-amplifier. Photos and data courtesy of Precision Acoustics Limited.
40 MEMBRANE HYDROPHONE A sheet of unpoled pvdf film (thickness < 30 µm) is stretched over a ring (diameter ~100 mm) with gold / chromium electrodes vacuum deposited on the surfaces. The sheet is only metallised over a small central area, which forms the active element of the device (diameter typically mm).
41 HYDROPHONE ARRAY (NPL / PRECISION ACOUSTICS) 25 elements, deposited on a pvdf membrane. The minimum element diameter is 0.2 mm with a minimum pitch of 0.3 mm. The overall span of the array is 12 mm.
42 HYDROPHONE ARRAY All 25 channels are available simultaneously. Direct and quick measurement of beam profile (avoids scanning the ultrasonic field with the hydrophone). Now available commercially. However, it is an expensive device.
43 Needle Hydrophones
44 Needle hydrophones Precision Acoustics Courtesy of Dr. Coussios
45 Primary standard for hydrophone calibration Optical interferometer displacement of pellicle (foil) within field measured using a Michelson interferometer; Displacement simply related to the acoustic pressure in the transducer far-field; Hydrophone to be calibrated is placed at the same position in the field and its output voltage determined; Used to calibrate secondary standard hydrophones with uncertainties of ±3% to ±5% (200 khz to 20 MHz); Approaching 20 years old. Courtesy of NPL
46 New primary standard laser interferometer Courtesy of NPL
47 What do we measure? : FWHM (6dB) widths Full width half pressure (6dB) beam width
48 Beam profiles (transverse) : Transducer 20010A3, 06/08/2003 Transverse radial plot, varying x Pressure (mv) FWHM Mean of 3 plots Distance from Max (mm)
49 Beam profiles (axial) : Transducer 20010A2, 17th Sept 2003 Axial plot, varying z pressure (mv) FWHM Mean of 3 plots distance from maximum (mm)
50 Acoustic power
51
52
53 Determination of power Acoustic power can be measured in two main ways: Planar scanning using hydrophones Performed close to output face of scanner head Intensity integrated over beam area Radiation force methods Often useful to perform both, to obtain independent check of results
54 POWER FROM INTENSITY MEASUREMENTS Spatially integrate the temporal average intensity (measured with the hydrophone) over the entire 2-D beam area. Power = I ta da whole beam This is a (time-consuming) alternative to measuring the power using a power balance, but may be the only option for diagnostic ultrasound fields.
55 Radiation force - principles Travelling acoustic wave has associated momentum Net transfer of energy from transducer into medium A target placed in the path of the beam experiences a radiation force, which is proportional to the power contained in the beam Requires target large enough to intercept whole beam
56 Radiation force balances If the target is connected to an appropriate force-measuring device (e.g. a weighing balance), then the power output of the device may be determined very easily Sensitive balance needed Power content of 1 W produces a change in weight of 69 mg Typical scanners will thus produce changes in the range 7-70 mg
57 Absorbing: TARGETS h = 1 Difficult to produce target that absorbs 100% of sound and target heats up, changing its buoyancy and hence weight. Reflecting: θ θ θ h = 2cos 2 θ Reflected ultrasound is absorbed by absorbing materials lining the tank, which can heat up and cause convection currents in the tank.
58 Radiation force : Measured using a force balance Degassed water Reflecting target Counter-balance Pre-focus
59 Radiation force : Measured using a force balance Balanced Forces Pivot point pivots Counter balancing weight Target Sound beam target
60 Radiation force : Measured using a force balance Balanced Forces l F rad = m g l l o Pivot point mg Counter balancing weight l o Target F rad Sound beam
61 Isp (Wcm-2) Isp data from to 01/05/ /11/00 05/12/00 11/12/00 04/01/01 18/01/01 22/01/01 31/01/01 22/02/01 16/03/01 20/03/01 27/03/01 04/04/01 10/04/01 23/04/01 01/05/01 22/05/01 05/06/ db
62 Radiation force : Measured using a force balance Target may be absorbing or reflecting Geometry may be vertical or horizontal
63 POWER MEASUREMENT: POWER BALANCE Microbalance Absorber Target Transducer Control and compensation Produces a signal that is proportional to the force required to keep the target stationary once acoustic power is applied. P = cf h h depends on geometry and size of target. Typical weight-equivalent force from diagnostic beam is just 70µg.
64 Shotton Balance Reflecting target
65 Primary standard radiation force balance Sartorius M25-D microbalance; Reflecting and absorbing targets; Power range: 1 mw to 1W; Frequency range: 1 to 20 MHz; Typical uncertainties ±2.5% to ±6%; Therapy-level balance also available, up to 20W. Courtesy of NPL
66 NPL Power Membrane Uses pyro-electric effect Adam Shaw, Institute of Physics Publishing Journal of Physics: Conference Series 1 (2004)
67 Network analyser QA and fault diagnosis first indicator!
68 Network analysis : Electrical properties in water Log Magnitude Real impedance Imaginary impedance SWR Smith Chart
69 Smith Chart
70
71
A Robust Hydrophone for HIFU Metrology
A Robust Hydrophone for HIFU Metrology Claudio I. Zanelli, Samuel M. Howard Onda Corporation, 592 E. Weddell Drive Sunnyvale, CA, 94089 USA Abstract. The high acoustic intensities generated by HIFU systems
More informationBIOMEDICAL ULTRASOUND
BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic
More informationTISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
More informationPIEZOELECTRIC FILMS TECHNICAL INFORMATION
PIEZOELECTRIC FILMS TECHNICAL INFORMATION 1 Table of Contents 1. PIEZOELECTRIC AND PYROELECTRIC EFFECTS 3 2. PIEZOELECTRIC FILMS 3 3. CHARACTERISTICS PROPERTIES OF PIEZOELECTRIC FILMS 3 4. PROPERTIES OF
More informationI. Methodology for Onda's Hydrophone Calibrations
190 ammerwood venue Sunnyvale, 94089 Phone (408) 745-0383 Fax (408) 745-0956 www.ondacorp.com I. Methodology for Onda's ydrophone alibrations alibration is obtained by comparison techniques to reference
More informationAPPLICATION NOTE AP050830
APPLICATION NOTE AP050830 Selection and use of Ultrasonic Ceramic Transducers Pro-Wave Electronics Corp. E-mail: sales@pro-wave.com.tw URL: http://www.prowave.com.tw The purpose of this application note
More informationA pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)
chapter 2 Pulsed Ultrasound In diagnostic ultrasound imaging, short bursts, or pulses, of acoustic energy are used to create anatomic images. Continuous wave sound cannot create anatomic images. Analogy
More informationarxiv:astro-ph/0509450 v1 15 Sep 2005
arxiv:astro-ph/0509450 v1 15 Sep 2005 TESTING THERMO-ACOUSTIC SOUND GENERATION IN WATER WITH PROTON AND LASER BEAMS K. GRAF, G. ANTON, J. HÖSSL, A. KAPPES, T. KARG, U. KATZ, R. LAHMANN, C. NAUMANN, K.
More informationE190Q Lecture 5 Autonomous Robot Navigation
E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator
More informationAPPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS
APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS Selection and use of Ultrasonic Ceramic Transducers The purpose of this application note is to aid the user in the selection and application of the Ultrasonic
More informationSCS Directory Accreditation number: SCS 0115
International standard: ISO/IEC 17025:2005 Swiss standard: SN EN ISO/IEC 17025:2005 METRON SA Calibration Laboratory Via Luserte Sud 7 CH-6572 Quartino Head: Angelo Capone Responsible for MS: Angelo Capone
More informationElectrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge
, Firenze, Italy Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge A. Bulletti, L. Capineri B. Dunn ESTEC Material and Process
More informationDiffraction and Young s Single Slit Experiment
Diffraction and Young s Single Slit Experiment Developers AB Overby Objectives Preparation Background The objectives of this experiment are to observe Fraunhofer, or far-field, diffraction through a single
More informationAntonio Rampoldi Struttura Complessa di Radiologia Interventistica Azienda Ospedale Niguarda Ca Granda, Milano. Terminology
Biophysical principles and clinical applications of MRgFUS Alberto Torresin Struttura Complessa di Fisica Sanitaria Azienda, Milano Università degli Studi di Milano Dip. di Fisica Antonio Rampoldi Struttura
More informationINTERFERENCE OF SOUND WAVES
1/2016 Sound 1/8 INTERFERENCE OF SOUND WAVES PURPOSE: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves and to observe interference phenomena with ultrasonic sound waves.
More informationSound Power Measurement
Sound Power Measurement A sound source will radiate different sound powers in different environments, especially at low frequencies when the wavelength is comparable to the size of the room 1. Fortunately
More informationUltrasonic Wave Propagation Review
Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic
More informationPIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany
ABSTRACT: PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany Today, in-line inspection tools are used routinely
More informationRadiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
More informationDescribing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power
Parameters used to completely characterize a sound wave Describing Sound Waves Chapter 3 Period Frequency Amplitude Power Intensity Speed Wave Length Period Defined as the time it take one wave vibrate
More informationAdvancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications
Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Janet E. Semmens Sonoscan, Inc. 2149 E. Pratt Boulevard Elk Grove Village, IL 60007 USA Phone: (847)
More informationTesting thermo-acoustic sound generation in water with proton and laser beams
International ARENA Workshop DESY, Zeuthen 17th 19th of May 25 Testing thermo-acoustic sound generation in water with proton and laser beams Kay Graf Universität Erlangen-Nürnberg Physikalisches Institut
More informationCopyright 2008 Pearson Education, Inc., publishing as Pearson Addison-Wesley.
Chapter 20. Traveling Waves You may not realize it, but you are surrounded by waves. The waviness of a water wave is readily apparent, from the ripples on a pond to ocean waves large enough to surf. It
More informationSynthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
More informationHunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality
Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL
More informationRobot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
More informationSound absorption and acoustic surface impedance
Sound absorption and acoustic surface impedance CHRISTER HEED SD2165 Stockholm October 2008 Marcus Wallenberg Laboratoriet för Ljud- och Vibrationsforskning Sound absorption and acoustic surface impedance
More informationDevelopment of Optical Wave Microphone Measuring Sound Waves with No Diaphragm
Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and
More informationIntroduction to acoustic imaging
Introduction to acoustic imaging Contents 1 Propagation of acoustic waves 3 1.1 Wave types.......................................... 3 1.2 Mathematical formulation.................................. 4 1.3
More informationWelcome to this presentation on LED System Design, part of OSRAM Opto Semiconductors LED 101 series.
Welcome to this presentation on LED System Design, part of OSRAM Opto Semiconductors LED 101 series. 1 To discuss the design challenges of LED systems we look at the individual system components. A basic
More informationUltrasound Dose Calculations
Ultrasound Dose Calculations One of the advantages of ultrasound therapy remains the reasonably broad range of trails from which effective treatment doses can be established. In principle, there is no
More informationEXPOSURE INDICES IN RF EPIDEMIOLOGY
EXPOSURE INDICES IN RF EPIDEMIOLOGY Paolo Vecchia National Institute of Health, Rome, Italy Advances in Biological effects and Dosimetry of Low Energy Electromagnetic Fields Erice, 28 March - 8 April 1981
More informationOn the beam deflection method applied to ultrasound absorption measurements
On the beam deflection method applied to ultrasound absorption measurements K. Giese To cite this version: K. Giese. On the beam deflection method applied to ultrasound absorption measurements. Journal
More informationInterferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)
Interferometers The true worth of an experimenter consists in his pursuing not only what he seeks in his experiment, but also what he did not seek. Claude Bernard (1813-1878) OBJECTIVES To examine the
More informationFTIR Instrumentation
FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation
More informationIntroduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
More informationSpectrum Level and Band Level
Spectrum Level and Band Level ntensity, ntensity Level, and ntensity Spectrum Level As a review, earlier we talked about the intensity of a sound wave. We related the intensity of a sound wave to the acoustic
More informationINSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS
INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS This tutorial provides an overview of signal processing and conditioning for use in instrumentation and automatic control systems.
More informationDOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM
More informationCO 2 laser for skin resurfacing and benign skin lesions
CO 2 laser for skin resurfacing and benign skin lesions Futuristic design 15 or 20 Watt on tissue Flexible fibre delivery system Low maintenance 610 MedArt A/S is a Danish medical laser company founded
More informationFXA 2008. UNIT G485 Module 4 5.4.3 Ultrasound. Candidates should be able to :
1 Candidates should be able to : ULTRASOUND Describe the properties of ultrasound. ULTRASOUND is any sound wave having a frequency greater than the upper frequency limit of human hearing (20 khz). Describe
More informationINTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
More informationPrinciples of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University
Principles of Medical Ultrasound Pai-Chi Li Department of Electrical Engineering National Taiwan University What is Medical Ultrasound? Prevention: actions taken to avoid diseases. Diagnosis: the process
More informationRock Bolt Condition Monitoring Using Ultrasonic Guided Waves
Rock Bolt Condition Monitoring Using Ultrasonic Guided Waves Bennie Buys Department of Mechanical and Aeronautical Engineering University of Pretoria Introduction Rock Bolts and their associated problems
More informationLab 9: The Acousto-Optic Effect
Lab 9: The Acousto-Optic Effect Incoming Laser Beam Travelling Acoustic Wave (longitudinal wave) O A 1st order diffracted laser beam A 1 Introduction qb d O 2qb rarefractions compressions Refer to Appendix
More informationPipeline External Corrosion Analysis Using a 3D Laser Scanner
Pipeline Technology Conference 2013 Pipeline External Corrosion Analysis Using a 3D Laser Scanner Pierre-Hugues ALLARD, Charles MONY Creaform, www.creaform3d.com 5825 rue St-Georges, Lévis (QC), Canada,
More informationOptical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus
Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,
More informationX-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
More informationLaminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers
Flow and Flow rate. Laminar and Turbulent flow Laminar flow: smooth, orderly and regular Mechanical sensors have inertia, which can integrate out small variations due to turbulence Turbulent flow: chaotic
More information- thus, the total number of atoms per second that absorb a photon is
Stimulated Emission of Radiation - stimulated emission is referring to the emission of radiation (a photon) from one quantum system at its transition frequency induced by the presence of other photons
More information3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage
3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter
More informationThe promise of ultrasonic phased arrays and the role of modeling in specifying systems
1 modeling in specifying systems ABSTRACT Guillaume Neau and Deborah Hopkins This article illustrates the advantages of phased-array systems and the value of modeling through several examples taken from
More information4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:
4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing
More informationTOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] info@rmjordan.com [e-mail] This
More informationDevelopment of MEMS micromirrors for intracavity laser control
Development of MEMS micromirrors for intracavity laser control Walter Lubeigt Centre for Microsystems and Photonics, EEE Department, University of Strathclyde,204 George Street, Glasgow G1 1XW,UK Motivation
More informationTHE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode
THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode Swedish Laser-Medical Society The dose is the most important parameter in laser phototherapy. At a first glance, the dose seem very
More informationScanning Acoustic Microscopy Training
Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed
More informationHuman Exposure Limits
Human Exposure Limits Session 3 0 Version December 2014 Learning objectives In this session we will: Learn about the international exposure limits for workers and the public Learn about methods for assessing
More informationHoles & Selective Laser Sintering
SLS is one of the most accurate 3D printing processes. The process has a layer thickness of 0.1mm. This is the thickness with which a new layer is added to each part. In any direction therefore the maximum
More informationManual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
More informationProper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center
Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster Jesse Garant, JG&A Metrology Center Traditional Metrology and Inspection Tactile Devices (Touch Probe)
More informationRF data receiver super-reactive ASK modulation, low cost and low consumption ideal for Microchip HCS KEELOQ decoder/encoder family. 0.
Receiver AC-RX2/CS RF data receiver super-reactive ASK modulation, low cost and low consumption ideal for Microchip HCS KEELOQ decoder/encoder family. Pin-out 38.1 3 Component Side 1 2 3 7 11 13 14 15
More informationRF EXPOSURE LIMITS AND TESTING REQUIREMENTS
RF EXPOSURE LIMITS AND TESTING REQUIREMENTS Jay Moulton Vice President March 12, 2013 1 OVERVIEW Specific Absorption Rate (SAR) and Maximum Permissible Exposure (MPE) Standards and Limits Evaluation Methods
More informationACOUSTIC EMISSION SIGNALS GENERATED BY MONOPOLE (PEN- CIL-LEAD BREAK) VERSUS DIPOLE SOURCES: FINITE ELEMENT MODELING AND EXPERIMENTS
ACOUSTIC EMISSION SIGNALS GENERATED BY MONOPOLE (PEN- CIL-LEAD BREAK) VERSUS DIPOLE SOURCES: FINITE ELEMENT MODELING AND EXPERIMENTS M. A. HAMSTAD Department of Mechanical and Materials Engineering, University
More informationThe Effect of Forced Air Cooling on Heat Sink Thermal Ratings
zpero 1 The Effect of Forced Air Cooling on Heat Sink Thermal Ratings By Paul Bachman, Fellow Engineer & Ronnie Haiduk, Applications Engineer, Crydom, Inc. ABSTRACT A heat sink s ability to dissipate thermal
More informationA METHOD OF PRECISE CALIBRATION FOR PIEZOELECTRICAL ACTUATORS
Uludağ Üniversitesi Mühendislik-Mimarlık Fakültesi Dergisi, Cilt 9, Sayı, 24 A METHOD OF PRECISE CALIBRATION FOR PIEZOELECTRICAL ACTUATORS Timur CANEL * Yüksel BEKTÖRE ** Abstract: Piezoelectrical actuators
More informationUS-SPI New generation of High performances Ultrasonic device
US-SPI New generation of High performances Ultrasonic device Lecoeur Electronique - 19, Rue de Courtenay - 45220 CHUELLES - Tel. : +33 ( 0)2 38 94 28 30 - Fax : +33 (0)2 38 94 29 67 US-SPI Ultrasound device
More informationHOMER Analyzer ISM 2.45 GHz, R-26 Waveguide
HOMER Analyzer ISM 2.45 GHz, R-26 Waveguide Basic Description The HOMER-Series STH 2.45-GHz Analyzer is an automatic impedance and power measurement system based on R-26 (WR-340) waveguide. The system
More informationThe Three Heat Transfer Modes in Reflow Soldering
Section 5: Reflow Oven Heat Transfer The Three Heat Transfer Modes in Reflow Soldering There are three different heating modes involved with most SMT reflow processes: conduction, convection, and infrared
More informationExperiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
More informationLinear Parameter Measurement (LPM)
(LPM) Module of the R&D SYSTEM FEATURES Identifies linear transducer model Measures suspension creep LS-fitting in impedance LS-fitting in displacement (optional) Single-step measurement with laser sensor
More informationA wave lab inside a coaxial cable
INSTITUTE OF PHYSICS PUBLISHING Eur. J. Phys. 25 (2004) 581 591 EUROPEAN JOURNAL OF PHYSICS PII: S0143-0807(04)76273-X A wave lab inside a coaxial cable JoãoMSerra,MiguelCBrito,JMaiaAlves and A M Vallera
More informationDoppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
More informationv = fλ PROGRESSIVE WAVES 1 Candidates should be able to :
PROGRESSIVE WAVES 1 Candidates should be able to : Describe and distinguish between progressive longitudinal and transverse waves. With the exception of electromagnetic waves, which do not need a material
More informationAntenna Properties and their impact on Wireless System Performance. Dr. Steven R. Best. Cushcraft Corporation 48 Perimeter Road Manchester, NH 03013
Antenna Properties and their impact on Wireless System Performance Dr. Steven R. Best Cushcraft Corporation 48 Perimeter Road Manchester, NH 03013 Phone (603) 627-7877 FAX: (603) 627-1764 Email: sbest@cushcraft.com
More informationGrazing incidence wavefront sensing and verification of X-ray optics performance
Grazing incidence wavefront sensing and verification of X-ray optics performance Timo T. Saha, Scott Rohrbach, and William W. Zhang, NASA Goddard Space Flight Center, Greenbelt, Md 20771 Evaluation of
More informationLaser diffuse reflection light scanner with background suppression. Dimensioned drawing
Specifications and description HRTL 3B Laser diffuse reflection light scanner with background suppression Dimensioned drawing We reserve the right to make changes DS_HRTL3B_en.fm en 01-2010/12 50114049
More informationILV. Laser Welding: Line Scanners for Beam Shaping and Guiding. Alfred G. Arlt. Sulzbacher Str. 4 65824 Schwalbach/Taunus DIPL.-ING. ALFRED G.
Laser Welding: Line Scanners for Beam Shaping and Guiding Alfred G. Arlt Ingenieurbüro für Lasertechnik +Verschleißschutz (ILV) Sulzbacher Str. 4 65824 Schwalbach/Taunus aga@ilv-arlt.de lt History of ILV:
More informationPassive Remote Sensing of Clouds from Airborne Platforms
Passive Remote Sensing of Clouds from Airborne Platforms Why airborne measurements? My instrument: the Solar Spectral Flux Radiometer (SSFR) Some spectrometry/radiometry basics How can we infer cloud properties
More informationModule 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
More informationAcousto-optic modulator
1 of 3 Acousto-optic modulator F An acousto-optic modulator (AOM), also called a Bragg cell, uses the acousto-optic effect to diffract and shift the frequency of light using sound waves (usually at radio-frequency).
More informationMECHANICS OF SOLIDS - BEAMS TUTORIAL 2 SHEAR FORCE AND BENDING MOMENTS IN BEAMS
MECHANICS OF SOLIDS - BEAMS TUTORIAL 2 SHEAR FORCE AND BENDING MOMENTS IN BEAMS This is the second tutorial on bending of beams. You should judge your progress by completing the self assessment exercises.
More informationCHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
More informationRobot Sensors. Outline. The Robot Structure. Robots and Sensors. Henrik I Christensen
Robot Sensors Henrik I Christensen Robotics & Intelligent Machines @ GT Georgia Institute of Technology, Atlanta, GA 30332-0760 hic@cc.gatech.edu Henrik I Christensen (RIM@GT) Sensors 1 / 38 Outline 1
More informationPhased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated testing of square and round bars
11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated
More informationStudy the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
More informationEconomic Mass Producible Mirror Panels for Solar Concentrators
Abstract G, G. Burgess, K. Lovegrove and A. Luzzi Centre for Sustainable Energy Systems Australian National University, Canberra 2 Australia E-mail: Glen@faceng.anu.edu.au The Centre for Sustainable Energy
More informationUltrasound Distance Measurement
Final Project Report E3390 Electronic Circuits Design Lab Ultrasound Distance Measurement Yiting Feng Izel Niyage Asif Quyyum Submitted in partial fulfillment of the requirements for the Bachelor of Science
More informationAcoustics: the study of sound waves
Acoustics: the study of sound waves Sound is the phenomenon we experience when our ears are excited by vibrations in the gas that surrounds us. As an object vibrates, it sets the surrounding air in motion,
More information2. THE TEORRETICAL OF GROUND PENETRATING RADAR:
Sixteenth International Water Technology Conference, IWTC 16 2012, Istanbul, Turkey 1 THE USE OF GROUND PENETRATING RADAR WITH A FREQUENCY 1GHZ TO DETECT WATER LEAKS FROM PIPELINES Alaa Ezzat Hasan Ministry
More informationUse of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
More informationUltrasound. Sound waves
Ultrasound Basic Idea Send waves into body which are reflected at the interfaces between tissue Return time of the waves tells us of the depth of the reflecting surface History First practical application,
More informationULTRASONIC THICKNESS GAUGE TT100/110/120/130
ULTRASONIC THICKNESS GAUGE TT100/110/120/130 TT100 TT110 Portable size and easy operation Suitable for any metallic and non-metallic materials ultrasonic can go through Self-compensating of nonlinearity
More informationQUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
More informationEXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
More informationphysics 1/12/2016 Chapter 20 Lecture Chapter 20 Traveling Waves
Chapter 20 Lecture physics FOR SCIENTISTS AND ENGINEERS a strategic approach THIRD EDITION randall d. knight Chapter 20 Traveling Waves Chapter Goal: To learn the basic properties of traveling waves. Slide
More informationLOCATION DEPENDENCY OF POSITIONING ERROR IN A 3-AXES CNC MILLING MACHINE
th International & 26 th All India Manufacturing Technology, Design and Research Conference (AIMTDR 214) December 12 th 14 th, 214, IIT Guwahati, Assam, India LOCATION DEPENDENCY OF POSITIONING ERROR IN
More informationUS-Key New generation of High performances Ultrasonic device
US-Key New generation of High performances Ultrasonic device US-Key connected to a laptop computer US-Key Ultrasound device single channel Features USB2 High Speed connection Ultralow noise preamplifier
More informationA Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer
Presented at IMAPS 2002 Denver, Colorado September 5, 2002 (Best of Session Award) A Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer Jeff Strole*, Scott Corbett*,
More information